ENCLOSED MEDICAL ISOLATION TRANSFORMERS
Enclosed Medical Isolation Transformers
Acme Electric is proud to offer a line of fully
enclosed medical isolation transformers, featuring
Amveco toroidal technology. For medical grade
applications, these units provide additional safety
and protection. When using electronic devices
in a medical environment, Acme’s medical
grade transformers will bring the equipment into
compliance with the UL 60601 medical safety
standard. The transformers operate at 120V
60Hz input with 120V output. They have built
in RFI filtering and in-rush current limiting. The
transformer design utilizes toroidal transformer
technology which offers light weight, high
efficiency, quiet operation, cool overall temperature, and
low stray magnetic field. They are intended for use by either
medical OEM manufacturers looking for stand alone medical
transformer protection or hospitals needing to isolate their
electronic equipment from the power source.
APPLICATIONS
■
Hospitals
■
Medical Facilities
■
OEM Manufacturers
FEATURES
■
Fully enclosed medical grade isolation transformers housed
in white aluminum enclosure
■
Designed for North American 120V 60Hz input operation
■
UL listed to UL 60601-1 and c-UL listed to CSA C22.2
No.60601
■
High efficiency toroidal transformer design yielding overall
compact size and low weight
■
Low leakage design. Less than 100 ÌA leakage current
■
Built in filtering with RFI interference and inrush protection
■
Surge suppression
■
10 ft hospital grade power cord
■
Duplex hospital grade receptacles
■
On/Off circuit breaker
■
Floor standing or wall mount
■
Custom designs available for higher volume requests
SPECIFICATIONS
ENCLOSED MEDICAL ISOLATION TRANSFORMERS
120 VOLT PRIMARY
APPROX. DIMENSIONS
VA CATALOG
NUMBER WIDTH HEIGHT LENGTH REGULATION LBS. (KG.) PLUG OUTLETS
300 AS30327 5.63(14.3) 4.13 (10.5) 10.00 (25.4) 4.5% 10 (4.5) 5-15P (2) 5-15R
600 AS30328 7.13 (18.1) 4.13 (10.5) 12.50 (31.7) 2.9% 17 (7.7) 5-15P (3) 5-15R
900 AS30329 7.13 (18.1) 4.13 (10.5) 12.50 (31.7) 1.5% 26 (11.8) 5-15P (4) 5-15R
1200 AS30330 9.13 (23.2) 4.13 (10.5) 14.00 (35.6) 1.4% 32 (14.5) 5-15P (4) 5-15R
1800 AS30331 9.13 (23.2) 4.13 (10.5) 14.00 (35.6) 1.7% 37 (16.8) 5-20P (4) 5-20R
122
122
—
120 VOLT SECONDARY
INCHES (CM.)
—
60 Hz
ACME ELECTRIC • MILWAUKEE, WI • 800.334.5214 • acmepowerdist.com
LOAD SHIP WEIGHT NEMA DUPLEX
APPROX. HOSP. GRADE