The Piccolo® Point-of-Care Chemistry Analyzer provides quantitative in vitro determinations of clinical chemistry analytes in heparinized whole blood, heparinized
plasma, or serum.
1.2Universal precautions
Operator health and safety require that universal precautions be observed at all times
while handling human blood samples or working with the Piccolo
®
Point-of-Care
Chemistry Analyzer in any way. The complete text of the document “OSHA 29 CFR
Part 1910, Occupational Exposure to Bloodborne Pathogens” can be found on the
Internet at www.osha-slc.gov/Preamble/Blood_toc_by_sect.html.
1.3Near-patient testing efficiency
This near-patient portable clinical chemistry system provides the clinician with routine multi-chemistry profiles and electrolytes for panels of blood tests within minutes.
Less than 2 minutes of hands-on time is required to perform the test. The Piccolo
®
Point-of-Care Chemistry Analyzer eliminates the need to transport samples to a central laboratory, and reduces such problems as misplaced samples, inaccurate labeling
and transcription, improper icing and bagging, and sample degradation. Important
criteria for test efficiency include ease of use, along with minimal handling and processing steps. The Piccolo provides automation, easy operation, and reliable results to
operators of almost any skill level.
1.4Overview of the procedure
The Piccolo® system consists of the Piccolo® Pointof-Care Chemistry Analyzer, single-use disposable
reagent discs, and result cards printed by the analyzer or transmitted to a data manager. External and
internal features of the analyzer are described and
illustrated below.
The operator begins the Piccolo
ducing 100 µL of whole blood, plasma, or serum
sample into a self-contained reagent disc and loading the disc into the analyzer. No sample preparation
is required. The operator then enters patient, operator, and physician ID numbers using the keypad and
display screen. The analyzer performs the remainder
of the testing protocol automatically in less than
13 minutes. The used reagent disc is removed and the analyzer is ready for the next
sample.
®
procedure by intro-
H
E
M
Q
C
3+
,
O
G
IP
L
K
L
O
3+
T
B
P
3.0 2.3-5.2
,
IC
K
T
+
6.3
1+
G
L
U
4.5 3.7-5.8
C
R
1
5.4-8.2
E
12* 60-1
C
H
I
O
C
C
L
A
T
+
244 125-27
G
+
B
/D
U
10
N
L
0.3-1.3
L
G
IP
T
/D
B
>180
M
IL
L
M
A
M
0
H
l/L
8.6-11.8
M
Y
E
A
G
M
0
L
M
/D
♦♦♦ 200-1200
7-25
T
G
L
L
A
/D
<10
L
0.1-0.6
M
P
G
A
/D
L
M
L
B
L
G
IP
/D
3.3 2.5-4.4
M
10-1
L
L
S
G
E
/D
M
18
R
20-150
G
IA
O
/D
P
L
E
#
L
/L
D
R
U
:
IS
000
C
G
000000827
U
E
/L
#
N
:
E
L
P
U
O
R
A
/L
A
T
T
G
P
IE
L
/D
A
N
T
C
#
L
T
IE
04/05/96
:
H
N
E
M
T
D
R
IS
T
6054
T
Y
R
#
#
P
000
:
Y
:
E
:
103195-2-1
1
A
2
PICCOLO
M
1:35
A
L
E
A
M
Piccolo Features and Components 1-1
Piccolo Operator’s Manual
The chemical reactions carried out by this analyzer are designed to produce a reaction
that absorbs light at known wavelengths. Analyte concentration is calculated from
light absorbance data. Results can be printed on result cards for purposes of assessment and for inclusion in the patient’s medical chart. Results are stored in the analyzer memory, and can be transmitted to an external computer.
1.5External features
The Piccolo® Point-of-Care Chemistry Analyzer is a
lightweight portable instrument that allows patient
Display
Keypad
testing at the point of care. Its small footprint allows
convenient placement of the analyzer in a near-
Drawer
patient environment. The top surface has a 4-line display screen and keypad, and a recessed carrying handle. A sliding drawer and a slot for inserting result
cards into the internal printer are on the front surface.
The back surface includes connections for AC and DC
power supply, as well as RS232 ports.
Result
Card
Slot
1.5.1Display and keypad
The keypad and 4-line display screen facilitate interactive communication between
the analyzer and the operator. The display indicates the status of the analyzer, and
presents procedural instructions and error messages. It also reflects information input
via the keyboard so it can be verified or corrected.
The operator inputs data through the 10 numeric
keys, 6 function keys, and 2 arrow keys. Refer to the
figure at the right for the keyboard layout. The specific uses of these keys are described in Sections 2, 3,
and 4.
The
POWER key is used to power the analyzer up or
POWER
CLOSE
OPEN
78 9
5
4
23
1
0
down and to cancel a run in progress. Refer to 2.2.3.
The
OPEN/CLOSE key operates the sliding drawer
where the reagent disc sits. This key is referred to in
the text as either the
containing a reagent disc initiates an analysis. Two of the function keys,
RECALL, provide access to special functions described in detail in Sections 3 and 4.
OPEN or CLOSE key as appropriate in context. Closing a drawer
MENU and
RECALL
MENU
6
EXIT
ENTER
1.5.2Disc drawer
The disc drawer slides in and out to transport the reagent disc into the analyzer and
hold it in place during the analysis. Keep the analyzer drawer closed when not loading or unloading a reagent disc. Always use the
tion of the drawer. Do not push on the drawer to close it. This may damage the instrument.
1-2Piccolo Features and Components
OPEN/CLOSE key to change the posi-
Piccolo Operator’s Manual
1.5.3Result card slot
The result card slot, on the front surface of the analyzer below the disc drawer, provides access into the internal thermal printer. Refer to 1.6, 2.2.2, and Section 4 for
information about printing result cards.
1.5.4Power supply
The analyzer runs on AC or DC power, providing
portability in point-of-care testing environments.
Recessed
handle
AC Power
supply cord
Two power supply components are included. The
DC power supply plugs into the socket on the
lower left back of the analyzer. An AC power cord
connects the DC power adaptor to a grounded
electrical outlet. A “universal” DC-to-AC power
convertor is also available, as well as an accessory
RS232
computer
ports
Power
supply
DC Power supply cord
that can plug into a vehicle’s lighter socket, allowing a 12-volt battery to be used as the power
source.
1.5.5Computer ports
Two RS232 ports support 2-way communication between the analyzer and external
devices. Port 1 [labeled RS232(1)] is used for connecting the analyzer to a computer to
transmit results. Port 2 is used for connecting an optional bar code scanner. Refer to
1.11 for information on linking the analyzer to an external computer.
1.5.6Symbols used in labeling
The following symbols are found on the label on the bottom panel or above the connections on the back panel.
SymbolExplanation
Direct current
!
Caution (refer to accompanying documents)
Serial ports
Piccolo Features and Components 1-3
Piccolo Operator’s Manual
1.5.7Physical and environmental specifications
Height:24.2 cm9.5 inches
Width:15.3 cm6 inches
Depth:29.2 cm11.5 inches
Weight of the analyzer:6.8 kg15 pounds
Weight of the power adapter:0.6 kg1.3 pounds
Mode of operation:Continuous
Protection against ingress of fluids:Ordinary equipment (IPXO)
Ambient operating temperature:15–32°C 59–90°F
Humidity:0–95%, non-condensing
Reaction temperature:37°C 98.6°F
Power requirements:240–100 volts AC 50–60 Hz
or 12–15 volts DC
Thermal protection rating:70°C 178°F
1.6Internal components
Inside the analyzer is a variable speed motor to spin the disc, a spectrophotometer to
measure analyte concentrations, two microprocessors to control testing and analytical
functions, system software on a PCMCA card, and a thermal line printer for imprinting result cards. A heating system maintains the temperature of the disc at 37°C while
reactions are in progress.
1.6.1Optics
The measurement optics consist of a discrete wavelength spectrophotometer with a
xenon lamp as a light source.
1.6.2Microprocessors and memory
The architecture of the instrument consists of two microprocessors: a real-time controller that monitors and controls all the measurements; and an I/O (input/output)
controller for memory management, calculations, and data storage. The two processors cross-check each other’s performance continuously, which allows a very high
level of confidence in the working of the instrument, and consequently of the results
and the integrity of the data.
The analyzer stores 70 patient results, 70 control results, and system quality control
data. All data stored in memory can be accessed via the Recall function. Refer to
Section 4.
1-4Piccolo Features and Components
Piccolo Operator’s Manual
1.6.3Software
The analyzer software comprises two matched programs. One program controls the
measurement engine itself, i.e., it schedules the flashing of the light source and collects the light intensity data for different cuvettes at different times during the reaction; and it collects all the information generated in the analytical part of the
instrument. The second program processes that information and reports analyte concentration. It also stores data related to each run (time, date, user ID, patient results,
and control data).
Software upgrades are distributed to registered users on a PCMCA card and installed
on site. Refer to 5.3 for instructions on installing a software card.
1.7Piccolo reagent discs
1.7.1Disc structure and function
In the Piccolo® system, all chemistry reactions are performed inside clear plastic
reagent discs, 8 cm in diameter and 2 cm in depth, specially designed to perform all
the steps required to convert a few drops of whole blood, plasma, or serum into a
panel of test results. Each disc contains all the components and reagents needed to
perform one or more tests on a single sample.
A total of 30 cuvettes are located around
the periphery: 4 system cuvettes contain
QC reagent beads for instrument and
chemistry quality control (refer to 1.9); a
minimum and a maximum absorbance
cuvette are employed in calibrating the
spectrophotometer; a specially designed
cuvette detects whether sufficient sample
volume was applied; one cuvette verifies
that a sufficient aliquot of diluted sample
was delivered to the reaction cuvettes; an
empty cuvette captures excess fluids. The
remaining 21 cuvettes contain test-specific lyophilized reagent beads.
The bar code ring attached to the top of the reagent disc contains calibration data specific for the chemistries in the disc. It also contains the disc identification code, lot
number, and expiration date. The analyzer automatically checks the code and rejects
an expired disc. The bar code ring also protects the optical surfaces of the cuvette from
fingerprints and other debris, and helps minimize contamination of the analyzer by
capturing small drops of blood that may be on the disc surface.
The sample port, demarcated by an arrow pointing to a circle molded onto the upper
surface of the disc, provides access to the sample chamber. When sufficient sample
has been loaded into the sample chamber, the sample fill line forms between two
arrows molded on the disc surface.
Bar Code Ring
Cuvettes
Sample Port
Diluent
Container
Sample Fill Line
Piccolo Features and Components 1-5
Piccolo Operator’s Manual
A sample diluent is sealed in a container inside the center of the disc. At the beginning of the reaction cycle, the analyzer opens the container and releases the diluent.
The analyzer separates a heparinized whole blood sample by centrifugation inside the
disc. Plasma and serum samples are unaffected. Precisely measured quantities of sample and diluent are delivered to the mixing chamber. Then centrifugal and capillary
forces deliver the diluted sample to the cuvettes, where it dissolves the reagent beads
and initiates the chemical reactions. Reaction products in the cuvettes are measured
photometrically.
1.7.2Disc storage and handling
• Store all reagent discs as described on their respective labels. When stored as
described as described on their respective labels, all reagents in the disc are stable
until the expiration date printed on the foil pouch and encoded on the bar code
ring. The analyzer will reject an expired disc.
• A disc can be used directly from the refrigerator (stored at 2–8°C) without warming.
• A disc can remain in its sealed pouch at room temperature for a cumulative period
of 48 hours. Longer time at room temperature can cause suppression of chemistries
and disc aborts.
• Do not expose discs, in or out of the foil pouches, to direct sunlight or to temperatures above 32°C (90°F).
• Inspect the unopened foil pouch for tears and punctures. A torn or damaged
pouch may allow moisture to reach the disc and reduce reagent performance.
• Open the disc pouch at the notch on the top right edge of the package. A disc must
be used within 20 minutes of opening the pouch. Once the pouch is opened, do not
place the disc back in the refrigerator for use at a later time.
• Discs are fragile. Handle with care. Do not tap the disc on the table or work bench
to empty the sample port. Do not use a disc that has been dropped.
• Keep discs clean. Handle them only at the edges to avoid smudges on the optical
surfaces. Use a lint-free tissue to remove blood from the disc surface.
• Write the patient identification number on the
disc surface in the space indicated in the figure
to the right (optional). Do not write anywhere
else on the disc or on the bar code ring.
• Hold reagent discs flat after introducing the
sample or control to avoid spillage.
• The used disc can be replaced in the pouch for
disposal.
• BIOHAZARD: Used reagent discs contain body
fluids. Follow good laboratory working practices. Handle all used discs as if they are contaminated with hepatitis or other
infectious diseases. Check with the appropriate state agency regarding disposal
regulations.
Write Patient ID
Here
1-6Piccolo Features and Components
1.8Result cards
Piccolo Operator’s Manual
Result cards are durable paper strips, about 7.5 inches long by
2.5 inches wide, designed for use with the Piccolo
Chemistry Analyzer, on which test results and other information
can be printed after testing is complete. Result cards are also used
for printing the QC report and the troubleshooting report to assist
in the interpretation of error codes. Additional copies of the result
card can be printed at any time for any analysis held in memory.
Refer to 2.2.2 and Section 4 for information about printing result
cards, QC report cards, and troubleshooting reports. Refer to 1.13
for ordering information.
The result card has an adhesive backing so it can be placed into the
patient’s file. The operator should follow the institution’s procedures for disposition of the result card.
®
Point-of-Care
1.9Intelligent QC (iQC™)
The Piccolo® Point-of-Care Chemistry Analyzer includes design and user interface
features that perform comprehensive system-wide quality control checks during each
run. These features, collectively called “intelligent QC” (iQC™), ensure that operators
at a wide range of skill levels can achieve accurate and reliable results. Two types of
QC reagent beads (instrument and chemistry) are included in each disc. Refer to 1.7.1
for information about disc structure. When these methods confirm that all parameters
are within expected ranges, INSTRU QC:OK CHEM QC:OK is printed on the result
card. Otherwise, no result card is printed and an error message is displayed. Error
messages indicate analyzer or disc malfunctions, and explain why results may not be
available. Whenever an error message is displayed, refer to 5.6 for troubleshooting
procedures.
System and chemistry QC data from each patient sample are stored in the analyzer
memory with the sample results. System and chemistry QC data from each control
run are stored with control results, separately from sample results. Standard information storage and retrieval techniques are employed to ensure the integrity of the data.
All QC data stored in memory can be called up for review at any time. Refer to
Section 4 for complete information and detailed instructions.
iQC™ greatly reduces the requirement for routine control testing. Refer to 2.4 for recommendations and procedure for control testing. Operators requiring assistance setting up control testing procedures should contact Abaxis Technical Support at
800-822-2947.
Piccolo Features and Components 1-7
Piccolo Operator’s Manual
1.9.1Instrument iQC™
The analyzer hardware is subjected to a self-test at power-up. The self-test ensures
that all optics, the flash lamp, and the circuit board components are functioning properly, and also verifies the memory functions. If any component does not meet specification, an error message is displayed. The disc motor, flash lamp, temperature, and
optics are monitored continuously throughout each analysis. The spectrophotometer
is automatically recalibrated at the beginning of each analysis.
Simultaneously with each analysis, advanced optical sensing and electronic systems
monitor reactions involving instrument QC reagent beads to verify the functioning of
the analyzer and the disc
1.9.2Reagent iQC™
Reactions involving chemistry QC reagent beads reveal and quantify any degradation
of the test-specific reagents in the disc due to suboptimal storage conditions. The
value reported is the actual absorbance as a percent of the expected absorbance. The
value must exceed a defined minimum for the reagents to meet performance standards. Otherwise, the run is aborted and an error message is displayed.
A lot-specific cutoff value for chemistry QC absorbance is used to separate the various
tests on the disc into two groups, based on different sensitivity to heat exposure. If the
chemistry QC value falls below this cutoff level but above the defined minimum, the
results from the less heat-sensitive tests are expected to meet performance standards.
The results of these tests are printed. The results of the more heat-sensitive tests may
show some degradation and are suppressed. (This cutoff level is not printed on the
quality control report.)
1.9.3Performance iQC™
Disc checks performed during the analysis include: checking the barcode for current
dating and for the presence of all required calibration factors; confirming that sample
volume is sufficient; confirming the presence of all reagent beads and that all reagent
beads have dissolved in diluted sample; verifying diluent and sample mixing; and
monitoring fluid movement throughout the disc for proper sequence and timing of
the reactions.
The iQC system monitors the performance of the reactions. For rate chemistries, the
analyzer confirms that the reactions are linear; that the absorbances from which the
rates are calculated, as well as the rates themselves, are within defined ranges; and
whether the substrate has been depleted. In endpoint chemistries, the analyzer verifies that all measurements are within the dynamic range of the photometer and that
the reaction has reached completion, ie, there are no changes in absorbance as a function of time.
The analyzer measure the levels of hemolysis, lipemia, and icterus. If there was interference with the analyte, information about the interference is printed on the result
card. Refer to 2.3 for additional information.
1-8Piccolo Features and Components
Piccolo Operator’s Manual
1.10Setup and power supply
1.10.1 Shipping verification
Remove the analyzer from the shipping carton and place it on a level surface. Check
the components you received against the following list.
• Piccolo
• Piccolo® Operator's Manual
• DC power adapter and connector cord
•AC power cord
• warranty card
Note: Fill out the warranty card and mail it to Abaxis within 10 days of system installation to
start the warranty period. You will be put on the Abaxis customer list to receive any information pertaining to your analyzer and ancillary products, including product updates.
1.10.2 Powering up
• Use only the power supply components provided with the analyzer. Using other
power supply components will damage the instrument and void the warranty.
• Connect the analyzer only to a grounded electrical outlet.
• To prevent power surges or drain, do not plug the analyzer into the same circuit as
a centrifuge or any other high-current device. Alternatively, use an ancillary surge
protector or uninterruptable power supply (UPS).
®
Point-of-Care Chemistry Analyzer
• Place the analyzer on a level surface free of vibration and sudden jolts, and where
the ambient temperature is 15−32°C (59−90°F). Do not place it near a sunny window or other heat source.
• There should be at least 6 inches clearance behind the analyzer for access to the
power connection and RS232 ports.
• Place the analyzer where airborne matter such as dust, lint, and hair is limited; ie,
do not place it near a fan or where air circulation is high.
Step 1. Connect the power supply.
Refer to the figure at the right. Plug the DC
adapter into the socket on the back of the
analyzer. Connect the AC power cord to the
DC adapter. Plug the AC power cord into a
grounded electrical outlet. The self-test message will appear on the display. The self-test
is described in 1.9.1. If the display is blank,
check all power supply connections.
Performing
self-test
Piccolo Features and Components 1-9
Piccolo Operator’s Manual
Step 2. Allow the analyzer to warm to operating temperature.
Depending on the ambient temperature, warming may take more time than the selftest.
⇒ The analyzer is in standby mode, i.e., ready to run a disc, when the display reads:
Open drawer to
run a disc.
Step 3. Check the date and time.
Check the analyzer date and time to ensure they are correct. Refer to 3.3 to change the
date and time.
Step 4. Link the analyzer to an external computer. (Optional)
Refer to 1.11.
Step 5. Perform control testing.
Perform one or more runs using recommended controls before running patient samples to confirm that the analyzer is functioning to specification. Refer to 2.4 for control
run procedures.
1.10.3 Powering down
The power should remain on unless you are moving the analyzer to a new location.
Always power down using the
The
POWER key does not power down the analyzer when there is a reagent disc in the
drawer.
This message appears when the
shutdown procedure and return the analyzer to standby mode.
Turn analyzer OFF?
Press POWER/OFF to
turn off, or EXIT to
continue operation.
Press
POWER again to initiate system shutdown. The analyzer turns off when the sys-
tem shutdown is complete.
Performing
system shutdown.
POWER key rather than by unplugging the analyzer.
POWER key is pressed. Press EXIT to cancel the system
1-10Piccolo Features and Components
Piccolo Operator’s Manual
1.11Connecting to an external computer
A facility may want to connect the analyzer to an external computer for any or all of
the following reasons: to incorporate testing data into patient records; to provide
records for regulatory compliance; or to connect with automated billing systems. By
default, the analyzer does not transmit data, but it can be configured to transmit data
in simple ASCII text, or in ASTM E 1394-97. Contact Abaxis for detailed information
on connecting the analyzer to a computer.
1.12Bar Code Scanner Option/Requirements
For ease of entry of patient identification, the Piccolo Point-of-Care Blood Analyzer
can be used with any Bar Code Scanner that meets the following requirements:
1.12.1 Communication Interface
The communication interface must be a serial connection using a female 9-pin D-SUB
connector. The TX (transmit) should be on pin 2, and the RX (receive) should be on
pin 3.
1.12.2 Communication Parameters
Data is received at 9600 Baud, No Parity, 8 Bits, and 1 Stop Bit (9600,N,8,1).
1.12.3 Barcode Formats
The scanner should auto discriminate at a minimum Code 39 and Code 128 symbologies. There can be no preamble in the output data. The postamble in the output data
shall have CR (Carriage Return [0x0D]), LF (Line Feed [0x0A]) and NULL [0x00]).
1.13Consumables and ancillaries
Contact Abaxis or your authorized distributor to order reagent discs, result cards,
controls, and sample collection equipment and supplies.
1.14Customer service and technical support
Call Abaxis Customer Service and Technical Support at 800-822-2947 with questions
regarding the operation the Piccolo
®
Point-of-Care Chemistry Analyzer.
Piccolo Features and Components 1-11
Piccolo Operator’s Manual
Section 2:Testing Procedure and
Interpretation of Results
2.1Sample requirements
• The Piccolo® Point-of-Care Chemistry Analyzer accepts heparinized whole blood,
heparinized plasma, or serum samples.
• Lithium heparin is the only anticoagulant recommended for use with the Piccolo®
Point-of-Care Chemistry Analyzer.
• For use of the CLIA ‘88 Waived Lipid Panel Disc, the only sample that may be used
is lithium heparinized whole blood.
• A sample size of 90–120 µL is required.
• Whole blood must be analyzed within 60 minutes of collection, or separated into
plasma or serum.
• To prevent hemolysis, do not refrigerate or shake whole blood.
• If not analyzed immediately, plasma or serum (separated) may be stored at room
temperature for no longer than 5 hours after centrifugation. If storage for more
than 5 hours is required, the sample should be refrigerated in the stoppered tube at
2–8°C (36–46°F) for no longer than 48 hours; or stored at −10°C for up to 5 weeks in
a freezer without a self-defrost cycle. Under these conditions, there will be no clinically important changes in most analyte concentrations.
• For accurate interpretation of results, the sample should be collected from a patient
who has fasted for at least 12 hours (to avoid lipemic samples).
• Operator health and safety require that Universal Precautions be observed at all
times while handling human blood samples or working with the Piccolo® Pointof-Care Chemistry Analyzer in any way. The complete text of the document
“OSHA 29 CFR Part 1910, Occupational Exposure to Bloodborne Pathogens” can
be found on the Internet at www.osha-slc.gov/Preamble/Blood_toc_by_sect.html.
2.2Testing procedure
• Wear powder-free gloves while handling reagent discs or operating the analyzer.
Powder may disrupt the optical components.
• If necessary, power up the analyzer before beginning the procedure. The power-up
procedure is described in 1.10.2.
• Ensure that the ambient temperature is 15–32°C (59–90°F).
• The analyzer is in standby mode, i.e., ready to run a disc, when the display reads:
Open drawer to
run a disc.
Testing Procedure and Interpretation of Results 2-1
Piccolo Operator’s Manual
2.2.1Preparing the reagent disc
• Refer to 1.7 for complete information about Piccolo® reagent discs, including handling instructions. Please become thoroughly familiar with this information before
beginning the procedure.
• Refer to 2.1 for sample handling and storage requirements.
• The analysis must begin immediately (no more than 10 minutes) after dispensing
the sample into the reagent disc.
Step 1. Dispense the sample.
Use a micropipette (one is included with the Piccolo) or other transfer device to
dispense approximately 100 µL of sample into the disc via the sample port.
Write Patient ID
Here
a.Fill the sample port.
Expel air bubbles from the tip of the micropipette. Place the micropipette in the sample port
and tilt it until it is perpendicular to the disc
90 µL
surface. Push down on the plunger with a slow,
continuous motion.
Take care not to overfill the sample chamber. A
> 90 µL
< 120 µL
100 µL sample will fill the sample chamber and
form a line between the two arrows molded on
the disc. More than 120 µL of sample will over-
<90 µL
fill the chamber, and less than 90 µL will underfill. Discard the pipette tip into a biohazard
> 120 µL
container.
b. Fill the sample chamber.
Tilt the disc to 45° with the sample port above the fill line, so that the entire sample flows into the sample chamber. Clean the reagent disc. Use a lint-free tissue to
remove any sample spilled on the outside of the disc, taking care that the tissue
does not withdraw any sample from the sample port. Dispose of the tissue in a
biohazard container.
c.Carry the prepared disc to the analyzer.
Hold the disc by its edges in a flat position.
2-2Testing Procedure and Interpretation of Results
Piccolo Operator’s Manual
2.2.2Running a patient sample
• The analyzer must be in standby mode to begin the procedure.
• Use the numeric and arrow keys to enter patient, operator, and physician IDs
while the analysis is proceeding. The → key inserts a dash; the ← key deletes the
character to the left.
• The reference range set and patient, operator, and physician IDs must be entered
before results can be calculated. If you wait until the run is complete before entering these numbers, the analyzer will require an additional one minute to calculate
the results.
• If an error message is displayed at any time during the run, refer to 5.6 for troubleshooting procedures associated with the specific error code.
Step 2. Open the drawer and insert the disc.
Press
OPEN. The following messages are displayed
sequentially:
Opening drawer...
Close drawer to
start analysis.
Place the disc in the drawer. Press
Closing drawer...
CLOSE. The display reads:
The analysis begins when the drawer closes.
Step 3. Select the reference range set.
By default, the reference range set for the previous sample is displayed. The reference range set for the current sample must be selected for correct interpretation of
results. Use the numeric keys to select the correct reference range set. Press
when the correct number appears to the right of Select.
Select:1 MALE
1- Male 3- Special
2- Female 4- Control
ENTER to accept.
ENTER
Testing Procedure and Interpretation of Results 2-3
Piccolo Operator’s Manual
Step 4. Enter the patient ID.
A patient ID (up to 14 characters) must be entered to continue. Press
Input Patient #:
_
_____________
ENTER when finished.
Step 5. Enter the operator ID.
Enter the operator ID for the current sample, up to 14 characters. Press
Input Operator #:
_____________
_
ENTER when finished.
Step 6. Enter the physician ID.
Enter the physician ID for the current sample, up to 14 characters. Press
Input DR #:
_____________
_
ENTER.
ENTER.
ENTER.
ENTER when finished.
Step 7. Automatic sample processing.
The analyzer processes the sample and calculates results in less than 13 minutes
with no further operator input. During the run, the analyzer displays the following message, including time remaining to complete the analysis (XX:XX) and
patient ID.
Results ready in
XX:XX
for patient #:
XXXXXXXXXXXXXX
Sample processing is complete when the analyzer beeps and displays these messages sequentially:
Analysis complete. Open
drawer or insert card
to print results.
Insert card to
print results.
2-4Testing Procedure and Interpretation of Results
Piccolo Operator’s Manual
Step 8. Print results.
To end the procedure without printing results, press
If a Run Cancelled error message is displayed, refer to 5.6 for an explanation
and further instructions.
Place the result card in the result card slot. Remove the card when directed to do
so on the display. These messages appear sequentially:
Printing results...
Remove card.
Opening drawer...
EXIT and continue to step 9.
Remove disc
and close drawer.
If the results card shows one or more rows of diamonds (♦♦♦) in place of analyte
concentration, this indicates suppressed results. Refer to 2.3.2 B. In this situation, insert a second results card when prompted.
Step 9. Return the analyzer to standby mode.
Remove the reagent disc from the drawer and dispose of it, following your lab’s
biohazard procedures. Press
CLOSE to return the analyzer to standby mode.
2.2.3Canceling an analysis in progress
Press POWER to cancel an analysis in progress. The display requests verification that
the analysis should be canceled:
Cancel analysis?
POWER/OFF to cancel
run, or EXIT to
continue analysis.
Testing Procedure and Interpretation of Results 2-5
Loading...
+ 51 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.