The Piccolo® Point-of-Care Chemistry Analyzer provides quantitative in vitro determinations of clinical chemistry analytes in heparinized whole blood, heparinized
plasma, or serum.
1.2Universal precautions
Operator health and safety require that universal precautions be observed at all times
while handling human blood samples or working with the Piccolo® Point-of-Care
Chemistry Analyzer in any way. The complete text of the document “OSHA 29 CFR
Part 1910, Occupational Exposure to Bloodborne Pathogens” can be found on the
Internet at www.osha-slc.gov/Preamble/Blood_toc_by_sect.html.
1.3Near-patient testing efficiency
This near-patient portable clinical chemistry system provides the clinician with routine multi-chemistry profiles and electrolytes for panels of blood tests within minutes.
Less than 2 minutes of hands-on time is required to perform the test. The Piccolo®
Point-of-Care Chemistry Analyzer eliminates the need to transport samples to a central laboratory, and reduces such problems as misplaced samples, inaccurate labeling
and transcription, improper icing and bagging, and sample degradation. Test systems
must produce reliable results when used by operators with a wide range of skill levels. An important criterion is ease of use, and automated systems with minimal handling and processing steps, like the Piccolo® Point-of-Care Chemistry Analyzer, are
easy to operate and are widely used in near-patient environments.
1.4Overview of the procedure
The Piccolo® system consists of the Piccolo® Pointof-Care Chemistry Analyzer, single-use disposable
reagent discs, and result cards printed by the analyzer . External and internal features of the analyzer
are described and illustrated below.
The operator begins the Piccolo procedure by introducing whole blood, plasma, or serum sample into
a self-contained reagent disc and loading the d isc
into the analyzer. No premeasuring or sample preparation steps are required. The operator then enters
patient, operator, and physician
numbers using
ID
the keypad and display screen. The analyzer performs the remainder of the testing protocol auto-
HEM
QC
3
+
,
OK
GLOB
LIP
3
TP
+
3
,
ICT
.0
K+
6
1
GLU
.3
+
2
4
.3
CRE
.5
-5
1
5
1
.2
.4
CHOL
2
-8
*
3
I
.2
.7
CT
CA++ LIP
-5
6
2
.8
G/DL
0
BUN
4
-1
4
1
0
G/DL
0
TBIL HEM
.3
>
MM0l/L
1
-
1
1
2
8
AMY
.3
5
0
-
8
MG/DL
2
.6
7
AL
-1
0
MG/DL
♦
7
1
T
-2
♦
.8
ALP
5
♦
<
0
MG/DL
1
.1
0
ALB
-0
MG/DL
2
LIP
.6
0
0
-1
3
MG/DL
1
.3
0
SERIAL#
2
-1
0
MG/DL
0
1
2
8
OPER
0
-
1
2
.5
5
U/L
DISC
:
0
-4
0
GENERAL
0
.4
0
U/L
0
0
#:
0
0
LOT #:
P
U/L
0
A
0
TIENT
G/DL
8
PATIENT
2
7
CHEMISTRY
0
4
/0
DR #:
5
/9
TYPE:
6
6
0
#:
5
0
4
0
0
1
0
12
A
3
1
9
5
-2
P
-
MALE
IC
1
1
:3
C
5
O
AM
L
O
Piccolo Features and Components 1-1
Piccolo Operator’s Manual
matically in about 12 minutes. The used reagent disc is removed and the analyzer is
ready for the next sample.
The chemical reaction s carried o ut by this an alyzer a re designed t o pr oduce a reaction
that absorbs light at known wavelengths. Analyte concentration is calculated from
light absorbance data. Results can be printed on result cards for purposes of assessment and for inclusion in the patient’s medical chart. Results are stored in the analyzer memory, and can be transmitted to an external computer.
1.5External features
The Piccolo® Point-of-Care Chemistry Analyzer is a
lightweight portable instrument that allows patient
testing at the point of care. Its small footprint allows
convenient placement of the analyzer in a near-
Display
Keypad
Drawer
patient environment. The top surface has a 4-line display screen and keypad, and a recessed carrying handle. A sliding drawer and a slot for inserting result
cards into the internal printer are on the front surface.
The back surface has DC power supply and RS232
ports.
Result
Card
Slot
1.5.1Display and keypad
The keypad and 4-line display screen facilitate interactive communication between
the analyzer and the operator. The display indicates the status of the analyzer, and
presents procedural instructions and error m essages. It also reflects info rmation input
via the keyboard so it can be verified or corrected.
The operator inputs data through the 10 numeric
POWER
OPEN
CLOSE
789
5
4
23
1
0
keys, 6 function keys, and 2 arrow keys. Refer to th efigure at the right for the keyboard layout. The specific uses of these keys are described in
Sections 2, 3
and 4.
The
down and to cancel a run in progress. Refer to
The
POWER
OPEN/CLOSE
key is used to power the analyzer up or
2.2.3
key operates the sliding drawer
,
.
where the reagent disc sits. This key is referred to in
the text as either the
OPEN
or
key as appropr i ate in context. Closing a drawer
CLOSE
containing a reagent disc initiates an analysis. Two of the function keys,
RECALL
, provide access to special functions described in detail in
Sections 3
MENU
RECALL
6
and
and 4.
MENU
EXIT
ENTER
1.5.2Disc drawer
The disc drawer slides in and out to transport the reagent disc into the analyzer and
hold it in place during the analysis. Keep the analyzer drawer closed when not loading or unloading a reagent disc. Always use the
OPEN/CLOSE
1-2Piccolo Features and Components
key to change the posi-
Piccolo Operator’s Manual
tion of the drawer.
instrument.
Do not push on the drawer to close it. This may damage the
1.5.3Result card slot
The result card slot, on the front surface of the analyzer below the disc drawer, provides access into the internal thermal printer. Refer to
1.6, 2.2.2
, and
Section 4
for
information about printing result cards.
1.5.4Power supply
The analyzer runs on DC power, providing portability in point-of-care testing environments.
Recessed
handle
AC Power
supply cord
Two power supply components are included. A
DC adapter
plugs into the socket on the lower left
back of the analyzer, allowing the use of a 12-volt
battery as the power source. An
AC power cord
connects the power adaptor to a grounded electrical outlet.
RS232
computer
ports
Power
adaptor
DC Power supply cord
1.5.5Computer ports
Two RS232 ports support 2-way communication between the analyzer and external
devices. Only port 1 [labeled RS232(1)] should be used at this time. Port 2 is reserved
for future use. Refer to
computer.
1.11
for information on linking the analyz er to a n externa l
1.5.6Symbols used in labeling
The following symb ols are found on the label on the bottom panel or above the connections on the back panel.
SymbolExplanation
Direct current
!
=
Caution (refer to accompanying documents)
Serial ports
Piccolo Features and Components 1-3
Piccolo Operator’s Manual
1.5.7Physical and environmental specifications
Height:24.2 cm9.5 inches
Wi dth:15.3 cm6 inches
Depth:29.2 cm11.5 inches
Weight of the analyzer:6.8 kg15 pounds
Weight of the power adapter:0.6 kg1.3 pounds
Mode of operation:Continuous
Protection against ingress of fluids:Ordinary equipment (IPXO)
Ambient operating temperature:15–32°C 59–90°F
Humidity:0–95%, non-co ndensing
Reaction temperature:37°C 98.6°F
Power requirements:250–90 volts AC 50–60 Hz
or 12–15 volts DC
Thermal protection rating:70°C 178°F
1.6Internal components
Inside the analyzer is a variable speed motor to spin the disc, a spectrophotometer to
measure anal yte co ncentrat ions, two mi cr opr oce ssor s to contr o l test ing an d analyt ical
functions, system software on a PCMCA card, and a thermal line printer for imprinting result cards. A heating system maintains the temperature of the disc at 37°C while
reactions are in progress.
1.6.1Optics
The measurement optics consist of a discrete wavelength spectrophotometer with a
xenon lamp as a light source.
1.6.2Microprocessors and memory
The architecture of the instrument consists of two microprocessors: a real-time controller that monitors and controls all the measurements; and an I/O (input/output)
controller for memory management, calculations, and data storage. The two processors cross-check each other’s performance continuously, which allows a very high
level of confidence in the working of the instrumen t, and consequently of the results
and the integrity of the data.
The analyzer stores 150 patient results, 75 control results, and system quality control
data. All data stored in memory can be accessed via the Recall function. Refer to
Section 4
.
1-4Piccolo Features and Components
Piccolo Operator’s Manual
1.6.3Software
The analyzer software comprises two matched programs. One program controls the
measurement engine itself, ie, it schedules the flashing of the light source and collects
the light intensity data for different cuvettes at different times during the rea ction; and
it collects all the information generated in the analytical part of the instrument. The
second program processes that information and reports analyte concentration. It also
stores data related to each run (time, date, user
Software upgrades are distributed to registered users on a PCMCA card and installed
on site. Refer to
5.3
for instructions on installing a softwa re card.
, patient results, and control data).
ID
1.7Piccolo reagent discs
1.7.1Disc structure and function
In the Piccolo system, all chemistry reactions are performed inside clear plastic
reagent discs, 8 cm in diameter and 2 cm in depth, specially designed to perform all
the steps required to convert a few drops of whole blood, plasma, or serum into a
panel of test results. Each disc contains all the components and reagents needed to
perform one or more tests on a single sample.
A total of 30
the periphery: 4 system cuvettes contain
QC reagent beads for instrument and
chemistry quality control (refer to 1.9); a
minimum and a maximum absorbance
cuvette are employed in calibrating the
spectrophotometer; a specially designed
cuvette detects whether sufficient sample
volume was applied; one cuvette verifies
that a sufficient aliquot of diluted sample
was delivered to the reaction cuvettes; an
empty cuvette captures excess fluids. The
remaining 21 cuvettes contain test-specific lyophilized reagent beads.
bar code ring
The
cific for the chemistries in the disc. It also contains the disc identification code, lot
number, and expiration date. The analyzer automat ica lly checks the code and rejects
an expired disc. The bar code ring also protects the optical surfaces of the cuvette fr om
fingerprints and ot her debris, and helps minimize conta mination of the analyzer by
capturing small drops of blood that may be on the disc surface.
sample port,
The
surface of the disc, provides access to the
has been loaded into the sample chamber, the
arrows molded on the disc surface.
cuvettes
are located around
Bar Code Ring
Cuvettes
Sample Port
Diluent
Container
Sample Fill Line
attached to the top of the reagent disc contains calibration data spe-
demarcated by an arrow pointing to a circle molded onto the upper
sample chamber
sample fill line
. When sufficient sample
forms between two
Piccolo Features and Components 1-5
Piccolo Operator’s Manual
A sample
diluent
is sealed in a container inside the center of the disc. At the begin-
ning of the reaction cycle, the analyzer opens the container and releases the diluent.
The analyzer separates a heparinized whole blood sample by centrifugation inside the
disc. Plasma and serum samples are unaffected. Precisely measur ed q uantities of sam ple and diluent are delivered to the
mixing chamber
. Then centrifugal and capillary
forces deliver the diluted sample to the cuvettes, where it dissolves the reagent beads
and initiates the chemical reactions. Reaction products in the cuvettes are measured
photometrically.
1.7.2Disc storage and handling
• Store all reagent discs as described on their respective labels. When stored as
described as described on their respective labels, all reagents in the disc are stable
until the expiration date printed on the foil pouch and encoded on the bar code
ring. The analyzer will reject an expired disc.
• A disc can be used directly from the refrigerator without warming.
• A disc can remain in its sealed pouch at room temperature for a cumulative period
of 48 hours. Longer time at room temperature can cause suppression of chemistries
and disc aborts.
• Do no t expose discs, in or out of the foil pouches, to direct sunlight or to tempera-
tures above 32°C (90°F).
• Inspect the unopened foil pouch for tears and punctures. A torn or damaged
pouch may allow moisture to reach the disc and adversely affect reagent performance.
• Open the disc pouch at the notch on the top right edge of the package. A disc must
be used within 20 minutes of opening the pouch. Once the pouch is opened, do not
place the disc back in the refrigerator for use at a later time.
• Discs are fragile. Handle with care. Do not tap the disc on the table or work bench
to empty the sample port. Do not use a disc that has been dropped.
• Keep discs clean. Handle them only at the edges to avoid smudges on the optical
surfaces. Use a lint-free tissue to remove blood from the disc surface.
• Write the patient identification number on the
disc surface in the space indicated in the figure
Write Patient ID
Here
to the right (optional). Do not write anywhere
else on the disc or on the bar code ring.
• Hold reagent discs flat after introducing the
sample or control to avoid spillage.
• The used disc can be replaced in the pouch for
disposal.
• BIOHAZARD: Used reagent discs contain body
fluids. Follow good laboratory working practices. Handle all used discs as if they
are contaminated with hepatitis or other infectious diseases.
1-6Piccolo Features and Components
Piccolo Operator’s Manual
04/05/96
1:35
AM
PATIENT
TYPE:MALE
PATIENT
#:
103195-2-1
GENERAL CHEMISTRY
12
DISC LOT #:
6054 A
OPER #:
000DR #: 000
SERIAL #: 000000827
ALB 3.32.5-4.4G/DL
ALP L I P
20-150U/L
ALT
<10
10-118
U/L
AMY
♦♦♦ 200-1200U/L
TBIL HEM
0.1-0.6
MG/DL
BUN
>180
7-25
MG/DL
CA++ LIP
8.6-11.8
MG/DL
CHOL
24 4 125-270
MG/DL
CRE I C T
0.3-1.3
MG/DL
GLU
112* 60-110
MG/DL
K+
4. 5 3.7-5.8
MM0l/L
TP
6. 3 5.4-8.2G/DL
GLOB
3. 0 2.3-5.2G/DL
QC OK
HEM
3+ , LIP 3+ , ICT 1+
PICCOLO
1.8Result cards
Result cards are durable paper strips, about 6 inches long by 2.5
inches wide, designed for use with the Piccolo
®
Point-of-Care
Chemistry Analyzer, on which test results and other information
can be printed after testing is complete. Result cards are also used
for printing the QC report and the troubleshooting report to assist
in the interpretation of error codes. Additional copies of the result
card can be printed at any time for any analysis held in memory.
Refer to
2.2.2
cards, QC report cards, and troubleshooting reports. Refer to
and
Section 4
for information about printing result
1.12
for ordering information.
The result card has an adhesive backing so it can be placed into the
patient’s file. The operator should follo w the institution’s procedures for disposition of the result card.
1.9Intelligent QC (iQC™)
The Piccolo® Point-of-Care Chemistry Analyzer includes design an d user interf ace
features that perform compr ehensive sy stem-wide quality control checks during each
run. These features, collectively called “intelligent QC” (iQC™), ensure that operators
at a wide range of skill levels can achieve accu rate and reliable results. Two types of
QC reagent beads (instrument and chemistry) are included in each disc. Refer to
for information about disc structure. When these methods confirm that all parameters
are within expected ranges,
INSTRU QC:OK CHEM QC:OK
is printed on the result
card. Otherwise, no result card is printed and an error message is displayed. Error
messages indicate analyzer or disc malfunctions, and explain why results may not be
available. Whenever an error message is displayed, refer to
5.6
for troubleshooting
procedures.
1.7.1
System and chemistry QC data from each patient sample are stored in the analyzer
memory with the sample results. System and chemistry QC data from each control
run are stored with control results, separately from sample results. Standard information storage and retrieval techniques are employed to ensure the integrity of the data.
All QC data stored in memory can be called up for review at any time. Refer to
Section 4
for complete information and detailed inst ructions.
iQC™ greatly reduces the requirement for routine control testing. Refer to
ommendations and procedure for control testing. Operators requiring assistance setting up control testing procedures should contact Abaxis Technical Support.
Piccolo Features and Components 1-7
2.4
for rec-
Piccolo Operator’s Manual
1.9.1Instrument iQC™
The analyzer hardware is subjected to a self-test at power-up. The self-test ensures
that all optics, the flash lamp, and the circuit board components are functioning properly, and also verifies the memory functions. If any component does not meet specification, an error message is displayed. The disc motor, flash lamp, temperature, and
optics are monitored continuously throughout each analysis. The spectrophotometer
is automatically recalibrated at the beginning of each a na lysis.
Simultaneously with each analysis, advanced optical sensing and electronic systems
monitor reactions involving instrument QC reagent beads to verify the functioning of
the analyzer and the disc
1.9.2Reagent iQC™
Reactions involving chemistry QC reagent beads reveal and quantify any degradation
of the test-specific reagents in the disc due to suboptimal storage conditions. The
value reported is the actual absorbance as a percent of the expected absorbance. The
value must exceed a defined minimum for the reagents to meet performance standards. Otherwise, the run is aborted and an error message is displayed.
A lot-specific cutoff value for chemistry QC absorbance is used to separate the various
tests on the disc into two groups, based on differen t sensitivity to heat exposure. If the
chemistry QC value falls below this cutoff level but above the defined minimum, the
results from the less heat-sensitive tests are expected to meet performance standards.
The results of these tests are printed. The results of the more heat-sensitive tests may
show some degradation and are suppressed. (This cutoff level is not printed on th e
quality control report.)
1.9.3Performance iQC™
Disc checks performed during the analysis include: checking the barcode for current
dating and for the presence of all required calibration factors; confirming tha t sample
volume is sufficient; confirming the presence of all reagent beads and that all reagent
beads have dissolved in diluted sample; verifying diluent and sample mixing; and
monitoring fluid movement throughout the disc for proper sequence and timing of
the reactions.
The iQC system monitors the performance of the reactions. For rate chemistries, the
analyzer confirms that the reactions are linear; that the absorbances from which the
rates are calculated, as well as the rates themselves, are within defined ranges; and
whether the substrate has been depleted. In endpoint chemistries, the analyzer verifies that all measurements are within the dynamic range of the photometer and that
the reaction has reached completion, ie, there are no changes in absorbance as a function of time.
The analyzer measure the levels of hemolysis, lipemia, and icterus. If there was interference with the analyte, information about the interference is printed on the result
card. Refer to
2.3
for additional information.
1-8Piccolo Features and Components
Piccolo Operator’s Manual
1.10Setup and power supply
1.10.1 Shipping verification
Remove the analyzer from the shipping carton and place it on a level surface.
Check the components you received against the following list.
• Piccolo
• Piccolo
• DC power adapter and connector cord
• AC power cord
• warranty card
Note: Fill out the warranty card and mail it to Abaxis within 10 days of system installation to
start the warranty period. You will be put on the Abaxis custom er list to receive any information pertaining to your analyzer and ancillary products, including product updates.
1.10.2 Powering up
• Use only the power supply components provided with the analyzer. Using other
power supply components will damage the instru ment and void the warranty.
®
Point-of-Care Chemistry Analyzer
®
Operator's Manual
• Connect the analyzer only to a
• To pr event power surges or drain, do not plug the analyzer into the same circuit as
a centrifuge or any other high-current device. Alternatively, use an ancillary surge
protector or uninterruptible power supply (UPS).
• Place the analyzer on a level surface free of vibration and sudden jolts, and where
the ambient temperature is 15−32°C (59−90°F). Do not place it near a sunny window or other heat source.
• There should be at least 6 inches clearance behind the analyzer for access to the
power connection and RS232 ports.
• Place the ana lyzer where airborne matter such as dust, lint, and hair is limited; ie,
do not place it near a fan or where air circulation is high.
Step 1. Connect the power supply.
Refer to the figure at the right. Plug the DC
adapter into the socket on the back of the
analyzer . Connect the AC power cord to the
DC adapter. Plug the AC power cord into a
grounded
sage will appear on the display. The self-test
is described in
check all power supply connections.
electrical outlet. The self-test mes-
1.9.1
. If the display is blank,
grounded
electrical outlet.
Performing self-test
Piccolo Features and Components 1-9
Piccolo Operator’s Manual
Step 2. Allow the analyzer to warm to operating temperature.
Depending on the ambient temperature, warming may take more time than the selftest.
⇒⇒⇒⇒ The analyzer is in standby mode, ie, ready to run a disc, when the display reads
Open drawer to run a
disc
Step 3. Check the date and time.
Check the analyzer date and time to ensure they ar e corr ect. Ref er to
date and time.
Step 4. Link the analyzer to an external computer. (Optional)
Refer to
1.11
.
Step 5. Perform control testing.
Per form one or more runs using recommended controls before running patient samples to confirm that the analyzer is functioning to specification. Refer to
run procedures.
3.2
to change the
2.4
for control
:
1.10.3 Powering down
The power should remain on unless you are moving the analyzer to a new location.
Always power down using the
The
key does not power down the analyzer when there is a reagent disc in the
POWER
drawer.
This message appears when the
shutdown procedure and return the analyzer to standby mode.
Turn analyzer off?
Press POWER to turn
off, or EXIT to
continue operation.
Press
again to initiate system shutdown. The analyzer turns off when the sys-
POWER
tem shutdown is complete.
Performing system
shutdown.
key rather than by unplugging the analyzer.
POWER
key is pressed. Press
POWER
to cancel the system
EXIT
#
1-10Piccolo Features and Components
Piccolo Operator’s Manual
1.11Connecting to an external computer
An institution may wish to connect the analyzer to an external computer for any or all
of the following reasons: to incorporate testing data into patient records; to provide
records for regulatory compliance; or to connect with automated billing sys tems. The
instrument provides two RS232 ports for connection to external devices. Data from
the analyzer are exported to the external computer as text files (*.txt). Contact Abaxis
for technical support.
1.12Consumables and ancillaries
Contact Abaxis or your authorized distributor to order reagent discs, result cards,
controls, and sample collection equipment and supplies.
1.13Customer service and technical support
Call Abaxis Customer Service and Technical Support at 800-822-2947 with questions
®
regarding operation of and problems with the Piccolo
Analyzer.
Point-of-Care Chemistry
Piccolo Features and Components 1-11
Piccolo Operator’s Manual
Section 2: Testing Procedure and
Interpretation of Results
2.1Sample requirements
• The Piccolo® Point-of-Care Chemistry Analyzer accepts heparinized whole blood,
heparinized pla s ma, or serum samples.
• Lithium heparin is the only anticoagulant recommended for use with the Piccolo
Point-of-Care Chemistry Analyzer.
• A sample size of 90–120µL is required.
• Whole blo od must be analyzed within 60 minutes of collection, or separated into
plasma or serum.
• To prevent hemolysis, do not refrigerate or shake whole blood.
• If not analyzed immediately, plasma or serum may be stored at room temperature
for no longer than 5 hours after centrifugation. If storage for more than 5 hours is
required, the sample should be refrigerated in the stoppered tube at 2– 8°C
(36– 46°F) for no longer than 48 hours; or stored at −10°C for up to 5 weeks in a
freezer without a self-defrost cycle. Under these conditions, there will be no clinically important changes in most analy te concentrations.
• For accurate interpretation of glucose results, the sample should be collected f rom
a patient who has fasted for at least 12 hours.
®
• Operator health and safety require that Universal Precautions be observed at all
times while handling hum an blood samples or working with the Piccolo® Pointof-Care Chemistry Analyzer in any way. The complete text of the document
“OSHA 29 CFR Part 1910, Occupational Exposure to Bloodborne Pathogens” can
be found on the Internet at www.osha-slc.gov/Preamble/Blood_toc_by_ sect.html.
2.2Testing procedure
• Wear powder-free gloves while handling reagent discs or operating the analyzer.
Powder may disrupt the optical components.
• If necessary , power up the analyzer befor e beginning the procedure. Th e power-up
procedure is described in
• Ensure that the ambient temperature is 15–32°C (59–90°F).
The analyzer is in standby mode, ie, ready to run a disc, when the display reads
•
Open drawer
to run a disc
1.10.2
.
:
Testing Procedure and Interpretation of Results 2-1
Piccolo Operator’s Manual
Write Patient ID
Here
90 µL
> 90 µL
< 120 µL
<90 µL
> 120 µL
2.2.1Preparing the reagent disc
• Refer to
dling instructions. Please become thoroughly familiar with this information before
beginning the procedure.
1.7
for complete information about Piccolo ® reagent discs, in clud ing han-
• Refer to
2.1
for sample handling and storage requirements.
• The analysis must begin immediately (no more than 10 minutes) after dispensing
the sample into the reagent disc.
Step 1. Dispense the sample.
Use a micropipette or other transfer device to dispense approximately 100 µL of sample into the disc via the sample port.
a.Fill the sample port.
Expel air bubbles from the tip of the micropipette. Place the micropipette in the sample port
and tilt it until it is perpendicular to the disc
surface. Push down on the plunger with a slow,
continuous motion.
Take care not to overfill the sample chamber. A
90 µL sample will fill the sample chamber and
form a line between the two arrows molded on
the disc. More than 120 µL of sample will overfill the chamber. Discard the pipette tip into a
biohazard container.
b. Fill the sample chamber.
Tilt the disc to 45° with the sample port above the fill line, so that the entire sample flows into the sample chamber. Clean the reagent disc. Use a lint-fr ee tis sue to
remove any sample spilled on the outside of the disc, taking care that the tissue
does not withdraw any sample from the sample port. Dispose of the tissue in a
biohazard container.
c.(Optional) Label the disc.
Write the patient
on the disc surface in the space indicated in the figure above.
ID
Do not write anywhere else on the disc or on the bar code ring.
d. Carry the prepared disc to the analyzer.
Hold the disc by its edges in a flat position.
2-2Testing Procedure and Interpretation of Results
Piccolo Operator’s Manual
2.2.2Running a patient sample
• The analyzer must be in standby mode to begin the procedure.
• Use the numeric and arrow keys to enter patient, operator , and physician
ID
the analysis is proceeding. The → key inserts a dash; the ← key deletes the character to the left.
• The reference range set and patient, operator, and physician
s must be entered
ID
before results can be calculated. If you wait until the run is complete before entering these numbers, the analyzer will require an additional one minute to calculate
the results.
• If an error message is displa yed at any time during the run, refer to
5.6
for trouble-
shooting procedures associated with the specific error code.
Step 2. Open the drawer and insert the disc .
Press
. The following messages are displayed sequen-
OPEN
tially:
Opening drawer...
Close drawer to start
analysis.
s while
Place the disc in the drawer. Press
Closing drawer...
. The display reads:
CLOSE
The analysis begins when the drawer closes.
Step 3. Select the reference range set.
By default, the reference range set for the previous sample is displayed. The reference
range set for the current sample must be selected for correct interpretation of results.
Use the numeric keys to select the correct reference range set. Press enter when the
correct number appears to the right of
Select Set:_ X
1- Male 2- Female
3- Special
ENTER to accept.
Select Set
.
Testing Procedure and Interpretation of Results 2-3
Loading...
+ 46 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.