1.2 User Features ....................................................................................................................................2
2. Getting Started ........................................................................................................................................7
10.2 Standard Calibrator Solutions.......................................................................................................65
Appendix A - Specifications .....................................................................................................................67
Appendix B - Required Notice..................................................................................................................69
Appendix C - RS232 Connections............................................................................................................71
Appendix D - Temperature Correction Data for Typical Solutions.....................................................73
Appendix E - Conversion Factors............................................................................................................77
Appendix F - Glossary of Terms..............................................................................................................79
ii
SECTION 1 INTRODUCTION
The Model 3200 is a microprocessor based; menu-driven instrument designed to perform
laboratory measurement of conductivity, resistivity, salinity, total dissolved solids and
temperature. The instrument’s menu system makes it simple to use.
The new YSI 3200 series conductivity cells contain internal temperature sensors for temperature
measurement and automatic temperature compensation. The Model 3200 is compatible with YSI
3400 series conductivity cells when used with the YSI Model 3232 cell adapter.
Internal memory for storing up to 100 sets of data and an RS232 port allow you to upload
data from the Model 3200 directly to your computer or serial printer.
1.1 CAPABILITIES
•Adjustable cell constant
0.005 to 100
•Readout units selectable
uS/cm, uS, ppt, ohm-cm, ohm, TDS
• Auto or manual ranges
• Adjustable reference temperature
0 to 100°C
•Adjustable automatic temperature compensation factor
0-10%/°C linear, prestored non-linear curves
• Six cell setup memory, with solution memories
Remembers cell constant, compensation method, multi-point cals and linear or non-linear
solution curves
• 100 data point memory
Stores date, time, temperature and conductivity
• Field software updates
Via RS232 port
• RS232 serial port
Output suitable for import to any spreadsheet
• Serial printer output
• AC line power
• 7-pin mini-DIN probe connector
• Separate connection for standard YSI temperature probes
YSI Incorporated Model 3200 1
Introduction Section 1
1.2 USER FEATURES
• Audio feedback
• Soft key/menu system
• Real time clock
• Auto linear temperature compensation calibration
Two temperature points are tested in a known solution to automatically
determine the solution’s compensation slope. Alternatively, the user may input a
temperature compensation factor directly (0 - 10%/°C).
• Auto non-linear temperature compensation calibration
This is a technique where a known solution is used to calibrate over a wide range of
temperature. The solution is ramped though a temperature range while the instrument
automatically and continuously generates calibration data. The calibration would be done
once and the results stored in non-volatile memory for recall any time that solution is
tested. Data for six solutions can be stored at any one time.
The user may also select from 2 pre-programmed non-linear temperature
compensation curves.
• Single or multi-point cell calibration
The cell constant can be calibrated at several points and stored in permanent
memory. During a reading, the closest calibration point is used as the cell
constant.
• High/low alarm
YSI Incorporated Model 3200 2
Introduction Section 1
1.3 CONTROLS
FRONT PANEL
The front panel of the instrument contains the display and keypad as shown below.
CONDUCTIVITY INSTRUMENT
3200
OPERATION
1.372
12:40
CELLSETUPDIAGRANGE
PM
S
u
24.99°c
Display
Power
ON/OFF
MODE Key
ENTER Key
YSI Incorporated Model 3200 3
Introduction Section 1
p
T
T
Soft k
The 3200 has two main operating modes, Data and Operation. The current mode is displayed in
the dark bar at the top right corner of the display. The following diagram shows the typical
display while the instrument is in the Operation mode.
Conductance: S
Conductivity: S/cm
Resistance: Ω
Resistivity: Ωcm
Tem
The bottom row of the display identifies the function of the four soft keys, which are located on
the keypad below the display. The function of each of these keys changes with each mode.
The [Mode] key and soft keys are used to navigate through the menus. The [Mode] key is used to
cycle between the two main operating modes. Pressing a soft key will bring up a new set of soft
key functions (menu). The [Mode] key is also used to backup through the menus to the top level.
mS/cm
MΩcm
Salinity: ppt
TDS: mg/L
erature: °C
mS
uS
uS/cm
KΩ
MΩ
KΩcm
1.953uS/cm
11:33AM 24.38°C
MESSAGES DISPLAYED HERE
CELL SETUP DIAGNOSTICS RANGE
Operation
K=10.100
25.0°C
1.91%/°C
T CELL
Message line
Mode
Cell constant
emperature
compensation
data
emperature
source
ey
functions
The right side of the display shows the cell constant, temperature compensation data and current
temperature source.
K= 10.100 indicates that a single fixed cell constant of 10.100 is being used. K= would change to
Km= when using multipoint calibration.
25.0°C indicates that readings are temperature corrected to 25°C using a coefficient of 1.91%/°C,
as indicated just below on the screen. When non-linear temperature compensation is used, N.L. is
displayed here. When a standard non-linear curve is selected, the name of the curve (Pure,
Natural) is displayed here. When no temperature compensation is being used, these two lines of
the 3200 display are blank.
T CELL indicates that the temperature source is a conductivity cell with an internal temperature
sensor (YSI 3200 series). T PROBE is displayed when the temperature source is an external
temperature probe.
YSI Incorporated Model 3200 4
Introduction Section 1
REAR PANEL
The rear panel contains the connections for the power supply, cell, external temperature probe
and RS232 serial port. It also has a mounting post for the 3232 adapter as shown below.
Adapter mounting post
POWER SUPPLY CONNECTION
The power supply connection requires a 12 VDC power supply (included) with at least 300 ma
current. The polarity is marked on the instrument.
CELL CONNECTION
The connector for the cell is a 7-pin mini DIN connector and is marked with an arrow to show
proper alignment. Be sure to align the arrows when plugging in the cell.
YSI 3200 series cells utilize a mini DIN connector that plugs directly into the 3200. If, however,
you have a YSI 3400 series cell, the YSI 3232 cell adapter and a YSI 3220 or YSI 700 series
temperature probe will be required. The YSI 3232 cell adapter mounts on the post on the rear of
the 3200. It has a 7-pin mini-DIN connector and two binding posts. The mini-DIN connector
plugs into the 3200 cell socket and a YSI 3400 series cell (or equivalent) connects to the binding
posts.
RS232 CONNECTION
The RS232 connection is a standard DB9 connector. This connector allows the 3200 to be
connected to the serial port of a computer or serial printer. A standard straight (NOT a null
modem) cable is required for connecting to a computer. To connect to a serial printer, use a
straight 9-pin male to 25 pin female cable (see Appendix - B).
EXTERNAL TEMPERATURE CONNECTION
The external temperature probe connection is used for temperature measurement and
compensation whenever the conductivity cell currently connected does NOT contain an internal
YSI Incorporated Model 3200 5
Introduction Section 1
temperature sensor. All YSI 3200 series cells contain an internal temperature sensor; therefore, an
external temperature probe is NOT necessary when using a 3200 series cell. However, if an
external temperature probe is used with a YSI 3200 series cell, the external temperature probe
will override the temperature sensor in the cell.
YSI 3400 series cells do not contain an internal temperature sensor; therefore, an external
temperature probe (YSI 3220 or YSI 700 series) is needed for temperature measurement or
compensation.
CE COMPLIANCE
Use of an external temperature probe or the YSI Model 3232 cell adapter with this instrument
may cause non-compliance with CE standards. See the Declaration of Conformity.
YSI Incorporated Model 3200 6
SECTION 2 GETTING STARTED
2.1 UNPACKING
When you unpack your new YSI Model 3200 for the first time, check the packing list to make
sure you have received everything you should have. If there is anything missing or damaged, call
the dealer from whom you purchased the Model 3200. If you do not know which of our
authorized dealers sold the system to you, call YSI Customer Service at 800-765-4974 or 937767-7241, and we'll be happy to help you.
2.2 WARRANTY CARD
Please complete the Warranty Card and return it to YSI. This will record your purchase of this
instrument in our computer system. Once your purchase is recorded, you will receive prompt,
efficient service in the event any part of your YSI Model 3200 should ever need repair.
2.3 WHAT YOU NEED
Several things are needed in order to make accurate conductivity measurements using the YSI
3200. The following list shows the basic items required.
• Instrument
• Power Supply
• Conductivity Cell
• Standard Solution(s)
• Beakers
• Rinsing Solution
2.4 SETTING UP
1. Power
Plug the power supply into its mating connector on the back of the instrument.
Depress the (on/off) key to turn the instrument on. The following screen should be displayed.
OPERATION
0.0000uS/cm
11:33AM -10.00°C
CELL SETUP DIAG RANGE
Notice that the units displayed are “uS/cm” indicating that the 3200 is in conductivity mode
(conductivity = conductance x cell constant). If the units are displayed as “uS” (no “/cm”), the
3200 is reading only conductance (as shown on the following page).
YSI Incorporated Model 3200 7
K=1.000
25.0°C
1.91%/°C
T CELL
Getting Started Section 2
OPERATION
0.0000uS
11:33AM -10.00°C
CELL SETUP DIAG RANGE
It is important to remember that the cell constant is NOT used when displaying conductance. To
change the measurement units, see 3.5 Display Configuration, Measurement Units.
The bottom row of the display shows the soft key menu.
K=1.000
25.0°C
1.91%/°C
T CELL
Soft key
2. Set the Date and Time
Press the [SETUP] soft key to enter the Setup menu.
SETUP
0.0000uS/cm
11:33AM -10.00°C
K=1.000
25.0°C
1.91%/°C
T CELL
DISPLAY TIME/DATE SYSTEM RANGE
Then, press the [TIME/DATE] soft key to enter the Time/Date menu. The following screen will
be displayed.
TIME/DATE
[ 8:34:04] AM
[ 2/15/1997]
[AM/PM] time format
[MM/DD/YY] date format
UP DOWN DIGIT NEXT
The cursor (underline) starts at the first digit of the time field. Use the [UP] or [DOWN] soft key
to change the underlined digit. When you have the correct digit displayed, press the [DIGIT] soft
key to move the cursor to the next digit, then use the [UP] or [DOWN] soft key to change that
digit. After you have entered the correct time, press the [NEXT] soft key to move the cursor to
the date field, then enter the correct date using the [UP], [DOWN] and [DIGIT] soft keys.
After you have entered the correct date and time, press the [MODE] key two times to save and
exit to the Setup menu.
YSI Incorporated Model 3200 8
Getting Started Section 2
3. Connect the Cell
Plug the cell (YSI 3200 series) into the connector on the back of the instrument marked cell.
NOTE: If you are using a YSI 3400 series cell, you will need the following:
• 3232 adapter
• YSI 3220 or 700 series temperature probe
4. Configure the 3200
Next, you must configure the 3200 with the correct cell constant of the cell you are using.
From Operation mode, press the [CELL] soft key to enter the Cell menu.
[0] # Configuration number CELL
[CONFIG 0] Title 0.0000uS offset
[Single] Cell Cal [1.00] K
[Linear] Comp [1.91] %/°C Comp val
[0.68] T.D.S. mult [25.0] °C Comp Temp
ENTER TO SEND CONFIGURATION
CONFIGURE CAL K TEMP PLATIN
Then, press the [CONFIGURE] soft key to enter the Configure menu. Configure is used to select
and configure the current cell and temperature compensation. The 3200 stores six configurations
numbered 0 to 5.
K field (cell constant)
[0] # Configuration number CONFIG
[CONFIG 0] Title 0.0000uS offset
[Single] Cell Cal [1.00] K
[Linear] Comp [1.91] %/°C Comp val
[0.68] T.D.S. mult [25.0] °C Comp Temp
ENTER TO SEND CONFIGURATION
UP DOWN DIGIT NEXT
Press the [UP] or [DOWN] soft key until you have selected configuration 0. The default
configuration is a single fixed cell constant of K = 1, linear temperature compensation corrected
to 25°C using a coefficient of 1.91%/°C.
Press the [NEXT] soft key to move the cursor (underline) to the Title field. Using the [UP],
[DOWN] and [DIGIT] soft keys, you may enter a name for this configuration.
Press the [NEXT] soft key four times to move the cursor to the K field. Using the [UP], [DOWN]
and [DIGIT] soft keys, enter the manufacturer’s stated cell constant for the cell that you are using
(K= 0.1, 1.0, 5.0, 10.0 etc.).
Press the [MODE] key two times to return to the Operation menu.
YSI Incorporated Model 3200 9
Getting Started Section 2
5. Measurement
After setting up the 3200 instrument and cell as described above, the following basic steps should
be used to make measurements.
1. Immerse the cell (YSI 3200 series) in the solution to be measured. If using a YSI 3400
series cell, also place the temperature probe in the solution.
2. Gently tap the cell to remove any air bubbles and dip the cell in the solution 2 or 3 times
to ensure proper wetting. The cell electrodes must be submerged and the electrode
chamber must not contain any trapped air. If using a flow through or fill cell, be certain it
is completely full.
3. Allow time for the temperature to stabilize.
4. Read the display, send the reading to a computer or serial printer (see section 4.3 Send),
or store the reading in memory (see section 4.1 Store).
5. Remove the cell (and temperature probe, if used) from the sample solution and rinse with
distilled or deionized water.
YSI Incorporated Model 3200 10
SECTION 3 SYSTEM CONFIGURATION AND OPERATION
3.1 MENU
The following section shows the software flow chart for the 3200. The two operating modes, Data
and Operation, are shown at the top with the soft key functions that are accessed in each mode
below them.
Data
Store
Review
Up
Down
Digit
Store
Recall
Send All
Clear
Clear All
Up
Down
Digit
Send
Up
Down
Digit
Delete
Operation
Cell
Configure
Cal K
Offset
Remove
Measure
Single Point
Multi Point
Send
Temp
L. Temp
Point 1
Point 2
N. L. Temp
Start
End
Send
Platin
Send
Send Setup
Set Com
Set Sen
Setup
Display
Size
Light
Dark
Time/Date
System
Range
Diag
Send
Beep
Range
YSI Incorporated Model 3200 11
System Configuration and Operation Section 3
3.2 CELL CALIBRATION
Before operating the 3200, or whenever you change cells, you must configure the 3200 to match
the cell used. There are two ways to do this. Choose the one that best meets your needs.
1. You may directly enter the manufacturer’s stated (or your manually calculated) cell
constant (K). See 3.2 Cell Calibration, Configure.
or
2. Allow the 3200 to calculate the cell constant and calibrate the system using standard
conductivity solution(s). See 3.2 Cell Calibration, Cal K.
The cell constant of up to six different cells (# 0 - 5) can be stored in non-volatile memory. The
3200 can determine the cell constant for a cell at a single point or up to five points using different
standard solutions.
Remember that Conductance (uS, mS or S) or Resistance (Ω, KΩ or MΩ) readings do NOT
use the cell constant. The 3200 must be set up to read Conductivity (uS/cm, mS/cm or S/cm) or
Resistivity (Ω-cm, KΩ-cm or MΩ-cm) for the cell constant to be used. See 8.2 Conductivity
Principles, Conductivity Fundamentals.
From the Operation mode, press the [CELL] soft key to enter the Cell menu. The bottom row of
the display shows the soft key menu.
[0] # Configuration number CELL
[CONFIG 0] Title 0.0000uS offset
[Single] Cell Cal [1.00] K
[Linear] Comp [1.91] %/°C Comp val
[0.68] T.D.S. mult [25.0] °C Comp Temp
ENTER TO SEND CONFIGURATION
CONFIGURE CAL K TEMP PLATIN
The Cell screen shows the configuration for the currently selected cell. The first two soft keys are
used to select the current cell configuration and calculate the cell constant (K). The cell constant
can be calculated at a single point with a standard solution or up to five points with different
solutions.
The last two soft keys are used to calculate the temperature coefficient of solutions and to
replatinize the electrodes of the cell.
CONFIGURE
Configure allows the user to directly input the cell constant (K), temperature compensation mode
and factor and total dissolved solids (TDS) multiplier. It also allows you to select a standard nonlinear temperature compensation curve for ultrapure or natural water. Alternatively, you may
calibrate the cell constant (K) and measure the temperature compensation factor in menus Cal K
and Temp. See 3.2 Cell Calibration, Cal k and 3.3 Temperature Compensation, Temp.
From the Cell menu, press the [CONFIGURE] soft key to enter the Configure menu. The bottom
row of the display shows the soft key selections.
YSI Incorporated Model 3200 12
System Configuration and Operation Section 3
[0] # Configuration number CONFIG
[CONFIG 0] Title 0.0000uS offset
[Single] Cell Cal [1.00] K
[Linear] Comp [1.91] %/°C Comp val
[0.68] T.D.S. mult [25.0] °C Comp Temp
ENTER TO SEND CONFIGURATION
UP DOWN DIGIT NEXT
Press the [UP] or [DOWN] soft key to select one of the six (0-5) stored configurations. Each
configuration contains the single cell constant, zero offset, multiple cell constants, total dissolved
solids (TDS) multiplier, linear temperature compensation coefficient and non-linear temperature
compensation data.
To change the currently selected configuration, press [NEXT] to move the cursor to the field you
want to change. Then, use the [UP], [DOWN] and [DIGIT] soft keys to change the field, except
Offset. The Offset field can only be changed from the Offset menu. The following table shows
the choices for each field.
Example: To select the standard non-linear temperature compensation curve for ultrapure water,
press [NEXT] three times, to move the cursor to the Comp field, then, press the [UP]
or [DOWN] soft key to cycle through the selections until Pure is displayed. Press
[MODE] two times to exit to the Operation menu.
Field Range / options Description
Title A - Z or 0 - 9 User defined name for this configuration.
Cell Cal Single or Multi Selects between a single fixed cell constant
and multiple cell constants.
Comp Linear, N. Linear, Pure, Natural
or Off
Selects between a linear temperature
compensation coefficient, non-linear
temperature compensation curve, standard
non-linear temperature compensation curve
(pure or natural) or no temperature
compensation.
TDS mult 0.55 - 0.90 Factor used to calculate the total dissolved
solids from the conductivity.
uS offset 0.0000 - 0.0150 Zero offset. Used only when measuring
conductance below 10uS to offset cell lead
insulation resistance. This value can only be
changed from the Offset menu.
K 0.001 - 100.999 Single fixed cell constant value (K). This
value is only used when the Cell Cal field is
set to Single.
YSI Incorporated Model 3200 13
System Configuration and Operation Section 3
Field Range / options Description
%/°C comp val 0.0 - 10 Linear temperature compensation value. This
value is only used when the Comp field is set
to Linear.
°C comp temp 0.0 - 100 Reference temperature that readings are
compensated to when temperature
compensation is used.
After you have finished changing the current configuration, press [MODE] two times to return to
Operation mode.
From the Config menu, the [ENTER] key may be used to send the current configuration data to a
computer or serial printer via the RS232 port.
CAL K
From the Cell menu, press the [CAL K] soft key to enter the Calibrate K menu. This menu is used
to measure the cell constant (K) of the current cell and store it in the currently selected
configuration. The cell constant can be measured at a single point using a single standard solution
or at multiple points using different value standard solutions. Alternatively, you may directly
input the cell constant in Cell Configuration. See 3.2 Cell Calibration, Configure.
[0] # Configuration number CALIBRATE K
[CONFIG 5] Title 0.0000uS offset
[Single] Cell Cal [1.00] K
[Linear] Comp [1.91] %/°C Comp val
[0.68] T.D.S. mult [25.0] °C Comp Temp
OFFSET SINGLE PT MULTI PT SEND
Single Point
To calculate and store the cell constant (K) at a single point:
1. From the Calibrate K menu, press the [SINGLE PT] soft key.
2. Clean the container and cell (and temperature probe, if used) with deionized water.
3. Rinse the cell (and temperature probe, if used) with some of the standard solution.
4. Place the cell (and temperature probe, if used) in the standard solution and wait for the
temperature reading and measured K value to stabilize.
5. Using the [UP], [DOWN] and [DIGIT] soft keys, enter the value of the standard solution
at the current temperature. Temperature compensation is OFF during cell calibration.
Example: YSI 3161 Conductivity Calibrator is 1000uS at 25°C, but at 23°C you would
enter 961uS.
6. Press [ENTER] to store the current cell constant (K). Press [MODE] three times to exit.
7. Remove the cell (and temperature probe, if used) from the solution and rinse with
deionized water.
YSI Incorporated Model 3200 14
System Configuration and Operation Section 3
r
Entered standard solution
Measured K
[961.00] uS/cm Conductivity
1.024 measured K 23.0°C
[1.00] K Cell
ENTER TO STORE
UP DOWN DIGIT
CAL CELL K
Current temperature
Multi Point
To calculate and store the cell constant (K) at multiple points using different value standard
solutions:
1. From the Calibrate K menu, press the [MULTI PT] soft key.
2. Clean the container and cell (and temperature probe, if used) with deionized water.
3. Rinse the cell (and temperature probe, if used) with some of the standard solution.
4. Place the cell (and temperature probe, if used) in the solution and wait for the temperature
reading and measured K value to stabilize.
5. Using the [UP], [DOWN] and [DIGIT] soft keys, enter the value of the standard solution
at the current temperature. Temperature compensation is OFF during cell calibration.
Example: YSI 3161 Conductivity Calibrator is 1000uS at 25°C, but at 23°C you would
enter 961uS.
6. Press [ENTER] to store the current cell constant (K). Stored values are displayed at the
bottom of the screen.
7. Repeat steps 3 - 6 for the remaining solutions.
8. If you have entered less than 5 solutions, enter 0.0uS for each remaining (unused) point
and press [ENTER] to store each one.
9. When you have measured all the solutions and stored all 5 points, press [MODE] two
times to exit to the Cell menu.
10. Press the [CONFIGURE] soft key to enter the Configure menu, then press [NEXT] two
times to move the cursor to the Cell Cal field. Press the [UP] or [DOWN] soft key to
select Multi cell calibration. Press [MODE] two times to exit to Operation mode.
11. Remove the cell (and temperature probe, if used) and rinse with deionized water.
Entered standard solution
Current temperature
[961.00] uS/cm conductivityCAL MULT K
10.11 measured K 23.0°C
[1] K point index
10.110 K value
95.48uS conductance
ENTER TO STORE
UP DOWN DIGIT NEXT
Point numbe
Stored value
After calculating and storing multiple cell constants, the Operation screen shows the current cell
constant used during measurements. Notice that the K= changes to Km= to show that multiple
cell constants are in use.
YSI Incorporated Model 3200 15
System Configuration and Operation Section 3
t
OPERATION
3.150uS/cm
11:33AM 24.38°C
CELL SETUP DIAG RANGE
Km=1.124
25.0°C
1.50%/°C
T CELL
Offset
When measuring conductance below 10uS (resistance above 100KΩ), the insulation resistance of
the cell leads may be a source of error. This error may be greatly reduced by allowing the 3200 to
calculate and store a zero offset value.
From the Calibrate K menu, press the [OFFSET] soft key to enter the Offset menu.
OFFSET
Offset value is only used below 10uS
Current offset
0.0000uS offset
To measure, connect a clean, dry cell
and press MEASURE
REMOVE MEASURE
Current cell constan
To measure and store the offset for the current cell in the currently selected configuration,
connect a clean, dry cell to the 3200, then, press the [MEASURE] soft key. The current offset is
displayed on the screen.
OFFSET
Offset value is used below 10uS
Current offset
To remove a stored offset (set the offset to zero), press the [REMOVE] soft key.
0.0015uS offset
To measure, connect a clean, dry cell
and press MEASURE
REMOVE MEASURE
Send
From the Calibrate K menu, the [SEND] soft key may be used to send the current cell constant
data to a computer or serial printer via the RS232 port.
3.3 TEMPERATURE COMPENSATION
There are four options for temperature compensation on the YSI 3200.
1. Directly enter a linear temperature compensation factor (0 - 10%/°C)
2. Select a standard non-linear compensation curve (ultrapure, natural).
YSI Incorporated Model 3200 16
System Configuration and Operation Section 3
3. Measure a linear temperature compensation factor.
4. Measure a non-linear temperature compensation curve.
The previous section, 3.2 Cell Calibration, describes the first two options. The following section,
Temp, describes how options 3 and 4 are used.
TEMP
From the Cell menu, press the [TEMP] soft key to enter the Cell Temp menu. This menu is used
to calculate the linear or non-linear temperature compensation coefficient of a particular solution
and store it in the currently selected configuration. Although the cell used is recorded, the
temperature compensation information can be used with any cell. Non-linear temperature
compensation [N.L. Temp] records multiple points over a temperature range and stores the
overall curve. Linear temperature compensation [L. Temp] measures a solution at only two points
and calculates a linear temperature coefficient in %/°C. Alternatively, you may input a linear
temperature compensation factor (0 - 10%/°C). See 3.2 Cell Calibration, Configure.
[5] # Configuration number CELL TEMP
[CONFIG 5] Title 0.0000uS offset
[Single] Cell Cal [1.00] K
[Linear] Comp [1.50] %/°C Comp val
[0.68] T.D.S. mult [25.0] °C Comp Temp
L. TEMP N. L. TEMP SEND
Linear Temperature Compensation
From the Cell Temp menu, press the [L. TEMP] soft key to enter the Lin Temp menu. This menu
enables the user to calculate the linear temperature compensation coefficient of a solution by
measuring it at two temperature points.
NOTE : If you already know the linear temperature compensation coefficient of the solution, you
may enter it directly in the Configure menu. See 3.2 Cell Calibration, Configure.
[25.00] °C p1 LIN TEMP
[0.00] uS p1 22.61 °C
[25.00] °C p2 895.1 uS
[0.00] uS p2
[0.00] %/°C calculated
ENTER TO STORE
POINT 1 POINT 2
To calculate and store the linear temperature compensation coefficient:
1. Verify that the current configuration contains the correct cell constant for the current cell.
2. Clean the container and YSI 3200 Series cell (or YSI 3400 Series and temperature probe)
with deionized water.
3. Rinse the cell (and temperature probe, if used) with some of the solution to be analyzed.
Stored points
Current
Calculated temperature
coefficient
YSI Incorporated Model 3200 17
System Configuration and Operation Section 3
4. Place the cell (and temperature probe, if used) in the solution at the first desired
temperature point and wait for the reading to stabilize. Ensure that the cell and
temperature probe are submerged and that no air bubbles are trapped in the cell.
5. Press the [POINT 1] soft key to store point 1.
6. Place the cell (and temperature probe, if used) in the same solution at the second desired
temperature point and wait for the reading to stabilize.
7. Press the [POINT 2] soft key to store point 2.
8. Remove the cell (and temperature probe, if used) and rinse with deionized water.
The linear temperature compensation coefficient has now been calculated and stored under the
current cell configuration.
NOTE: Any non-linear temperature compensation curve stored under the current configuration
will be retained.
Non-linear Temperature Compensation
From the Cell Temp menu, press the [N. L. TEMP] soft key to enter the NL Temp menu. This
menu enables the user to calculate a non-linear temperature compensation curve for a solution by
measuring it while the temperature is changed.
NL TEMP
[20.00] °C High 17.34 °C
[0.00] °C Low 1.34 S
CURRENT CURVE WILL BE ERASED
START END
Current
Temperature end
To calculate and store the non-linear temperature compensation curve:
1. Verify that the current configuration contains the correct cell constant for the current cell.
2. Clean the container and YSI 3200 Series cell (or YSI 3400 Series and temperature probe)
with deionized water.
3. Rinse the cell (and temperature probe, if used) with some of the solution to be analyzed.
4. Place the cell (and temperature probe, if used) in the solution at one end of the
temperature range that you want to compensate over and wait for the reading to stabilize.
5. Press the [START] soft key.
6. Change the temperature of the solution while the cell (and temperature probe, if used) is
measuring it.
7. When the temperature reaches the end of the range you want to compensate over, press
the [END] soft key.
8. Remove the cell (and temperature probe, if used) and rinse with deionized water.
The non-linear temperature compensation curve has now been calculated and stored under the
current cell configuration.
YSI Incorporated Model 3200 18
System Configuration and Operation Section 3
Send
From the Cell Temp menu, the [SEND] soft key may be used to send the current temperature
compensation data to a computer or serial printer via the RS232 port.
3.4 PLATINIZATION
The 3200 can be used to replatinize the electrodes of the cell. See section 6.2 Platinization for
details.
3.5 DISPLAY CONFIGURATION
The 3200 display is extremely flexible. The user may select from two screen sizes and may select
what data is displayed in each field.
SETUP
3.150uS/cm
11:33AM 24.38°C
K=10.100
25.0°C
1.50%/°C
T CELL
DISPLAY TIME/DATE SYSTEM RANGE
From Operation mode, press the [SETUP] soft key, then press [DISPLAY] to enter the Setup
Disp menu. The default display is the large display with three user selectable fields. Conductivity
is in the large field and temperature and time are in the smaller fields.
SETUP DISP
13.15uS/cm
11:33AM 24.38°C
ENTER TO CHANGE UNITS
SIZE NEXT LIGHT DARK
To change the display size, press the [SIZE] soft key. The small display contains six user
selectable fields. The first three fields in the small display correspond to the fields of the large
display. The following is an example of the small display.
K=10.100
25.0°C
1.50%/°C
T CELL
1006 uS/cm 11:33AM
24.38°C 994.7Ωcm
YSI Incorporated Model 3200 19
SETUP DISP
K=0.098
25.0°C
System Configuration and Operation Section 3
684
ENTER TO CHANGE UNITS
SIZE NEXT LIGHT DARK
mg/L
TDS
0.4
ppt
SAL
1.91%/°C
T CELL
MEASUREMENT UNITS
Each of the six fields on the small display (three fields on the large display) may be setup by the
user. Selection of fields is important, even if only the large display is used, because, when a
reading is stored in the 3200 memory, the date, time, temperature and conductivity are stored,
along with the first two additional fields (fields other than date, time, temperature and
conductivity) currently on the display. These same selected fields are also used when logging data
to a computer or serial printer.
1006 uS/cm 9999uS
65.68°F 97.8 Ω
684
ENTER TO CHANGE UNITS
SIZE NEXT LIGHT DARK
mg/L
TDS
0.4
ppt
SAL
SETUP DISP
K=0.098
25.0°C
1.91%/°C
T CELL
From the Setup Disp screen, press the [NEXT] soft key to select the field you want to change,
then press [ENTER] repeatedly to change the field until the information you want is displayed.
Each field may contain one of the following:
• Conductance, S, mS, uS
• Conductivity, S/cm, mS/cm, uS/cm
• Resistance, Ω (ohms), KΩ, MΩ
• Resistivity, Ω-cm, KΩ-cm, MΩ-cm
• Total Dissolved Solids, mg/L
• Salinity, ppt
• Time
• Date
• Temperature, °C
• Temperature, °F
• Blank
YSI Incorporated Model 3200 20
System Configuration and Operation Section 3
CONTRAST
The display contrast is adjusted from the Setup Display menu.
SETUP DISP
13.15 uS/cm
11:33AM 24.38°C
ENTER TO CHANGE UNITS
SIZE NEXT LIGHT DARK
Press the [LIGHT] or [DARK] soft keys to change the display contrast as necessary.
K=0.100
25.0°C
1.50%/°C
T CELL
TIME/DATE
From Operation mode, press the [SETUP] soft key, then press [TIME/DATE] to enter the
Time/Date screen. This screen is used to set the time and date values and format.
TIME/DATE
[ 8:34:04] AM
[ 2/15/1997]
[AM/PM] time format
[MM/DD/YY] date format
UP DOWN DIGIT NEXT
The first two fields are used to enter the current time and date. Use the [UP], [DOWN] and
[DIGIT] soft keys to enter the correct time. Press [NEXT] to change to the date field, then enter
the correct date. The default time and date formats are AM/PM and Month/Day/Year. To change
these formats, use the [NEXT] soft key to select the field, then use the [UP] or [DOWN] soft key
to change the format to 24Hr or Day/Month/Year. After you have entered the correct values,
press [MODE] to save and exit to the Setup menu. Press [MODE] again to return to the Operation
menu.
ALARMS
The 3200 has both high and low user programmable alarms. When an alarm is triggered, the 3200
screen will display “
send the information to a computer or serial printer via the RS232 port.
To activate the alarm(s):
YSI Incorporated Model 3200 21
ALARM! READINGS OUT OF LIMITS” and can also be set to beep and/or
System Configuration and Operation Section 3
1. From Operation mode, press the [SETUP] soft key, then, press [SYSTEM] to enter the Setup
Sys menu. The following menu will be displayed.
[Yes] key beep SETUP SYS
[Off] alarm status
[0.00] low [Off] beep on alarm
[0.00] high [On] send on alarm
[Conductance] uS alarm source
UP DOWN DIGIT NEXT
2. Use the [NEXT] soft key to select the Alarm status field, then press [UP] or [DOWN] to
change the Alarm status to “On”.
3. Press [NEXT] to select the Low field, then use the [UP], [DOWN] and [DIGIT] soft keys to
enter the set point for the low alarm.
4. Press [NEXT] to select the High field and use the [UP], [DOWN] and [DIGIT] soft keys to
enter the set point for the high alarm.
5. Press [NEXT] to select the alarm source, then press [UP] or [DOWN] to change the source.
The source may be conductance, conductivity or resistance.
6. Press [NEXT] to select “beep on alarm”, then press [UP] or [DOWN] to turn the alarm beeper
on or off.
7. Press [NEXT] to select “send on alarm”. When this parameter is enabled (On), readings will
be sent to a computer or serial printer whenever the alarm is triggered.
Press [MODE] once, to return to the Setup menu, or two times to return to Operation mode.
NOTE: To setup the communications parameters or select the report format and which data fields
will be sent over the RS232 serial port, see 4.3 Send, Send Setup.
BEEPER
The 3200 contains a beeper to indicate when a key has been pressed. The beeper may be turned
on or off from the Setup Sys menu. Press the [SETUP] soft key, while Operation mode, then
press [SYSTEM] to enter the Setup Sys menu .
[No] key beep SETUP SYS
[Off] alarm status
[0.00] alarm low [Off] beep on alarm
[0.00] alarm high [On] send on alarm
[Conductance] uS alarm source
UP DOWN DIGIT NEXT
Beeper status
YSI Incorporated Model 3200 22
Loading...
+ 60 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.