Yokogawa DO202G User Manual

User’s Manual
Model DO202G [Style: S2], DO202S [Style: S3]
2-wire Dissolved Oxygen Transmitter
IM 12J05C01-01E
IM 12J05C01-01E
4th Edition
TABLE OF CONTENTS

PREFACE

1. INTRODUCTION AND GENERAL DESCRIPTION ......................................... 1-1
1-1. Instrument check .................................................................................... 1-1
1-2. Application ............................................................................................... 1-3
2. GENERAL SPECIFICATIONS ......................................................................... 2-1
2-1. Specifications .......................................................................................... 2-1
2-2. Performance specifications ..................................................................... 2-2
2-3. Model and suffix codes ............................................................................ 2-5
2-4. Control Drawing of DO202S mA HART
®
Specification (IECEx) ............... 2-6
2-5. Control Drawing of DO202S mA HART® Specification (ATEX) ................ 2-7
2-6. Control Drawing of DO202S mA HART® Specification ............................. 2-8
(FM Intrinsically safe design)
2-7. Control Drawing of DO202S mA HART® Specification Specification ...... 2-9
(FM Non-incendive design)
2-8. Reserved for DO202S mA HART® Specification (CSA) ......................... 2-10
2-9. Control Drawing of DO202S FF/PB Specification (IECEx) ......................2-11
2-10. Control Drawing of DO202S FF/PB Specification (ATEX) .................... 2-12
2-11. Control Drawing of DO202S FF/PB Specification ................................ 2-13
(FM Intrinsically safe Entity)
2-12. Control Drawing of DO202S FF/PB Specification ................................ 2-15
(FM Intrinsically safe FISCO)
2-13. Control Drawing of DO202S FF/PB Specification ................................ 2-17
(FM Non-incendive Entity)
2-14. Control Drawing of DO202S FF/PB Specification ................................ 2-18
(FM Non-incendive FNICO)
2-15. Reserved for DO202S FF/PB Specification (CSA) ............................... 2-19
3. INSTALLATION AND WIRING ........................................................................ 3-1
3-1. Installation and dimensions ..................................................................... 3-1
3-1-1. Installation site ......................................................................................................3-1
3-1-2. Mounting methods ................................................................................................3-1
3-2. Wiring of power supply ............................................................................ 3-3
3-2-1. General precautions .............................................................................................3-3
3-2-2. Connection of the power supply ...........................................................................3-3
3-2-3. Switching the instrument on .................................................................................3-3
3-2-4. Preparation ...........................................................................................................3-3
3-2-5. Cables, terminals and glands ...............................................................................3-3
3-3. Wiring of sensors ..................................................................................... 3-3
3-3-1. Wiring the standard galvanic sensor ....................................................................3-3
3-3-2. Wiring other galvanic sensors ..............................................................................3-4
3-3-3. Wiring polarographic sensors ...............................................................................3-4
3-3-4. Wiring sensors with Vp connector ........................................................................3-4
3-3-5.
3-3-6. Additional precautions for installations in hazardous areas .................................3-5
3-3-7. Installation in: Hazardous Area-Non-Incendive ....................................................3-5
Sensor cable connections using junction box (BA10) and extension cable (WF10)
..3-4
4. OPERATION; DISPLAY FUNCTIONS AND SETTING ................................... 4-1
4-1. Operator interface.................................................................................... 4-1
4-2. Explanation of operating keys ................................................................. 4-2
4-3. Passcode protection ................................................................................ 4-2
4-4. Display examples .................................................................................... 4-3
4-5. Display functions ..................................................................................... 4-3
5. PARAMETER SETTING .................................................................................. 5-1
5-1. Maintenance mode .................................................................................. 5-1
5-1-1. Manual activation of hold ......................................................................................5-1
5-1-2. Manual temperature adjustment ...........................................................................5-2
IM 12J05C01-01E 4th Edition: Oct. 2009(YK) All Rights Reserved, Copyright © 2007, Yokogawa Electric Corporation
5-1-3. Manual pressure adjustment ................................................................................5-2
5-2. Commissioning mode .............................................................................. 5-3
5-2-1. Output range .........................................................................................................5-3
5-2-2. Hold.......................................................................................................................5-4
5-2-3. Service ..................................................................................................................5-5
5-3. Notes for guidance in the use of service coded settings ......................... 5-6
5-3-1. Parameter specific functions ................................................................................5-6
5-3-2. Temperature functions ..........................................................................................5-6
5-3-3. Calibration functions .............................................................................................5-6
5-3-4. mA Output functions .............................................................................................5-8
5-3-5. User interface .......................................................................................................5-8
5-3-6. Communication setup .........................................................................................5-10
5-3-7. General ...............................................................................................................5-10
5-3-8. Test and setup mode ..........................................................................................5-10
6. CALIBRATION PROCEDURE......................................................................... 6-1
6-1. General .................................................................................................... 6-1
6-1-1. Calibration methods ..............................................................................................6-1
6-1-2. Diagnostic functions performed during calibration ...............................................6-1
6-2. Calibration procedure using air calibration method ................................. 6-2
6-2-1. Preparation ...........................................................................................................6-2
6-2-2. Procedure for air calibration .................................................................................6-2
6-3. Calibration procedure using water calibration method ............................ 6-3
6-3-1. Preparation ...........................................................................................................6-3
6-3-2. Calibration operation (water calibration method) ..................................................6-3
6-3-3. Procedure for water calibration .............................................................................6-3
6-4. Calibration procedure using manual calibration method ......................... 6-4
6-4-1. Preparation ...........................................................................................................6-4
6-4-2. Procedure for manual slope (sensitivity) calibration .............................................6-4
6-4-3. Procedure for manual zero (offset) calibration .....................................................6-5
7. MAINTENANCE .............................................................................................. 7-1
7-1. Overall dissolved-oxygen metering system ............................................. 7-1
7-1-1. Inspection and maintenance to be implemented periodically ...............................7-1
7-1-2. Inspection and maintenance to be implemented occasion ..................................7-1
7-2. Periodic maintenance for the EXA DO202 transmitter ............................ 7-1
8. TROUBLESHOOTING .................................................................................... 8-1
8-1. Measures in the case of transmitter operation failure.............................. 8-1
8-2. Measures in the case of failure (error) detection ..................................... 8-1
9. SPARE PARTS ................................................................................................ 9-1
APPENDIX 1. USER SETTINGS .........................................................................A1-1
A-1-1. User setting table ................................................................................A1-1
A-1-2. Configuration checklist for DO202 ......................................................A1-2
A-1-3. Hart process variables ........................................................................A1-2
A-1-4. Device description (DD) menu structure .............................................A1-3
APPENDIX 2. QUALITY INSPECTION ...............................................................A2-1
A-2-1. DO202G 2-Wire Dissolved Oxygen Transmitter ................................. A2-1
A-2-2. DO202S 2-Wire Dissolved Oxygen Transmitter .................................. A2-5
A-2-3. DO202G, DO202S Fieldbus Communication .................................... A2-10
A-2-4. DO202G, DO202S Profibus Communication .................................... A2-14
Customer Maintenance Parts List (for DO202G Style: S2) CMPL 12J05C01-02E Customer Maintenance Parts List (for DO202S Style: S3) . CMPL 12J05C01-23E
Revision Record ........................................................................................................i
In this manual a
sign appears if it concerns the DO202G (S)-A, -N, -K.
mA
IM 12J05C01-01E
PREFACE
DANGERDANGER
Electric discharge
The EXA analyzer contains devices that can be damaged by electrostatic discharge. When servicing this equipment, please observe proper procedures to prevent such damage. Replacement components should be shipped in conductive packaging. Repair work should be done at grounded workstations using grounded soldering irons and wrist straps to avoid electrostatic discharge.
Installation and wiring
The EXA analyzer should only be used with equip­ment that meets the relevant international and regional standards. Yokogawa accepts no responsi­bility for the misuse of this unit.
CAUTIONCAUTION
The instrument is packed carefully with shock absorbing materials, nevertheless, the instrument may be damaged or broken if subjected to strong shock, such as if the instrument is dropped. Handle with care.
Although the instrument has a weatherproof construction, the transmitter can be harmed if it becomes submerged in water or becomes exces­sively wet.
Do not use an abrasive material or solvent when cleaning the instrument.
Do not modify the DO202 transmitter.
WARNING
WARNING
Electrostatic charge may cause an explosion haz­ard. Avoid any actions that cause the generation of electrostatic charge, e.g., rubbing with a dry cloth.
Warning label
Notice
• This manual should be passed on to the end user.
• The contents of this manual are subject to change without prior notice.
• The contents of this manual shall not be reproduced or copied, in part or in whole, without permission.
• This manual explains the functions contained in this product, but does not warrant that they are suitable the particular purpose of the user.
• Every effort has been made to ensure accuracy in the preparation of this manual.
However, when you realize mistaken expressions
or omissions, please contact the nearest Yokogawa Electric representative or sales office.
• This manual does not cover the special specifications. This manual may be left unchanged on any change of specification, construction
or parts when the change does not affect the
functions or performance of the product.
• If the product is not used in a manner specified in this manual, the safety of this product may be impaired.
Yokogawa is not responsible for damage to the instrument, poor performance of the instrument or losses resulting from such, if the problems are caused by:
• Improper operation by the user.
• Use of the instrument in improper applications
• Use of the instrument in an improper environment or improper utility program
• Repair or modification of the related instrument by an engineer not authorized by Yokogawa.
Safety and Modification Precautions
• Follow the safety precautions in this manual when using the product to ensure protection and safety of the human body, the product and the system containing the product.
Because the enclosure of the Dissolved Oxygen transmitter Type DO202S-A, -P, -F are made of alu­minium, if it is mounted in an area where the use of category 1 G Zone 0 apparatus is required, it must be installed such, that, even in the event of rare incidents, ignition sources due to impact and friction sparks are excluded.
The following safety symbols are used on the prod-
uct as well as in this manual.
DANGERDANGER
This symbol indicates that an operator must follow
the instructions laid out in this manual in order to avoid the risks, for the human body, of injury, electric shock, or fatalities. The manual describes what special care the operator must take to avoid such risks.
DANGER
Warranty and service
Yokogawa products and parts are guaranteed free from defects in workmanship and material under nor­mal use and service for a period of (typically) 12 months from the date of shipment from the manu­facturer. Individual sales organizations can deviate from the typical warranty period, and the conditions of sale relating to the original purchase order should be consulted. Damage caused by wear and tear, inadequate maintenance, corrosion, or by the effects of chemical processes are excluded from this war­ranty coverage.
WARNINGWARNING
This symbol indicates that the operator must
refer to the instructions in this manual in order to prevent the instrument (hardware) or software from being damaged, or a system failure from occurring.
CAUTIONCAUTION
This symbol gives information essential for
understanding the operations and functions.
This symbol indicates Protective Ground Terminal
This symbol indicates Function Ground Terminal
(Do not use this terminal as the protective ground terminal.)
This symbol indicates Alternating current.
This symbol indicates Direct current.
In the event of warranty claim, the defective goods should be sent (freight paid) to the service depart­ment of the relevant sales organization for repair or replacement (at Yokogawa discretion). The following information must be included in the letter accompa­nying the returned goods:
• Part number, model code and serial number
• Original purchase order and date
• Length of time in service and a description of the process
• Description of the fault, and the circumstances of failure
• Process/environmental conditions that may be related to the installation failure of the device
• A statement whether warranty or non-warranty service is requested
• Complete shipping and billing instructions for return of material, plus the name and phone number of a contact person who can be reached for further information
Returned goods that have been in contact with proc­ess fluids must be decontaminated/disinfected before shipment. Goods should carry a certificate to this effect, for the health and safety of our employees. Material safety data sheets should also be included for all components of the processes to which the equipment has been exposed.
ATEX Documentation
This procedure is only applicable to the countries in European Union.
GB
All instruction manuals for ATEX Ex related prod­ucts are available in English, German and French. Should you require Ex related instructions in your local language, you are to contact your nearest Yokogawa office or representative.
DK
Alle brugervejledninger for produkter relateret til ATEX Ex er tilgængelige på engelsk, tysk og fransk. Skulle De ønske yderligere oplysninger om håndtering af Ex produkter på eget sprog, kan De rette henvendelse herom til den nærmeste Yokogawa afdeling eller forhandler.
I
Tutti i manuali operativi di prodotti ATEX con­trassegnati con Ex sono disponibili in inglese, tedesco e francese. Se si desidera ricevere i man­uali operativi di prodotti Ex in lingua locale, met­tersi in contatto con l’ufficio Yokogawa più vicino o con un rappresentante.
E
Todos los manuales de instrucciones para los pro­ductos antiexplosivos de ATEX están disponibles en inglés, alemán y francés. Si desea solicitar las instrucciones de estos artículos antiexplosivos en su idioma local, deberá ponerse en contacto con la oficina o el representante de Yokogawa más cercano.
NL
Alle handleidingen voor producten die te maken hebben met ATEX explosiebeveiliging (Ex) zijn verkrijgbaar in het Engels, Duits en Frans. Neem, indien u aanwijzingen op het gebied van explosiebeveiliging nodig hebt in uw eigen taal, contact op met de dichtstbijzijnde vestiging van Yokogawa of met een vertegenwoordiger.
SF
Kaikkien ATEX Ex -tyyppisten tuotteiden käyt­töhjeet ovat saatavilla englannin-, saksan- ja ranskankielisinä. Mikäli tarvitsette Ex -tyyppisten tuotteiden ohjeita omalla paikallisella kielellännne, ottakaa yhteyttä lähimpään Yokogawa-toimistoon tai
-edustajaan.
P
Todos os manuais de instruções referentes aos produtos Ex da ATEX estão disponíveis em Inglês, Alemão e Francês. Se necessitar de instruções na sua língua relacionadas com produtos Ex, deverá entrar em contacto com a delegação mais próxima ou com um representante da Yokogawa.
F
Tous les manuels d’instruction des produits ATEX Ex sont disponibles en langue anglaise, allemande et française. Si vous nécessitez des instructions relatives aux produits Ex dans votre langue, veuillez bien contacter votre représentant Yokogawa le plus proche.
D
Alle Betriebsanleitungen für ATEX Ex bezo­gene Produkte stehen in den Sprachen Englisch, Deutsch und Französisch zur Verfügung. Sollten Sie die Betriebsanleitungen für Ex-Produkte in Ihrer Landessprache benötigen, setzen Sie sich bitte mit Ihrem örtlichen Yokogawa-Vertreter in Verbindung.
S
Alla instruktionsböcker för ATEX Ex (explosions­säkra) produkter är tillgängliga på engelska, tyska och franska. Om Ni behöver instruktioner för dessa explosionssäkra produkter på annat språk, skall Ni kontakta närmaste Yokogawakontor eller represent­ant.
GR
SK
CZ
PL
SLO
LT
LV
EST
H
BG
RO
M

1. INTRODUCTION AND GENERAL DESCRIPTION

Introduction 1-1
The Yokogawa EXA 202 is a 2-wire transmitter designed for industrial process monitoring, meas­urement and control applications. This user’s manual contains the information needed to install, set up, operate and maintain the unit cor­rectly. This manual also includes a basic trouble­shooting guide to answer typical user questions.
Yokogawa can not be responsible for the perform­ance of the EXA analyzer if these instructions are not followed.
mA
DISSOLVED OXYGEN TRANSMITTER
MODEL
DO202
SUFFIX
II 1G
SUPPLY OUTPUT
AMB.TEMP.
STYLE No.
Made in Japan Tokyo 180-8750 JAPAN
24V DC 4 20mA DC
-10  55°C
R
LR81741 C
WARNING Substitution of components may impair intrinsic safety
N200
DO202S-A

1-1. Instrument check

Upon delivery, unpack the instrument carefully and inspect it to ensure that it was not damaged dur­ing shipment. If damage is found, retain the original packing materials (including the outer box) and then immediately notify the carrier and the relevant Yokogawa sales office.
Make sure the model number on the nameplate affixed to the side of the instrument agrees with your order. Example of nameplate is shown below.
No. IECEx KEM 06.0055X Zone 0 Ex ia IIC T4 Zone 0 Ex ia IIC T6 for Ta:40 IP65 SEE CONTROL DRAWING
No. KEMA 06ATEX0224 X Ex ia IIC T4 Ex ia IIC T6 for Ta:40 SEE CONTROL DRAWING
IS CL I, DIV 1, GP ABCD AND AEx ia IIC T4 Type 4X
Install per CONTROL DRAWING IKE030-A10 P.5 to P.6
CL I, DIV 1, GP ABCD Ex ia IIC T4 Ex ia IIC T6 for Ta:40 SEE CONTROL DRAWING IP65 Type 3S
AVERTISSEMENT La substitution de composants peut compromeltre la securite intrinseque.
°C
°C
IP65
°C
0344
No. IECEx KEM 06.0055X Ex nA[nL] IIC T4 Ex nA[nL] IIC T6 for Ta:40 IP65 SEE CONTROL DRAWING
No. KEMA 06ATEX0225 EEx nA[nL] IIC T4
II 3 G
EEx nA[nL] IIC T6 for Ta:40 IP65 SEE CONTROL DRAWING
NI CL I, DIV 2, GP ABCD AND CL I, ZN 2, GP IIC T4
Type 4X
Install per CONTROL DRAWING IKE030-A10 P.7 to P.8
Ex nA[nL] IIC NI CL I, DIV 2, GP ABCD T4
R
T6 for Ta:40
LR81741 C
WARNING Substitution of components may impair suitability for class I, Division 2.
IP65 Type 3S
SEE CONTROL DRAWING
DO202S-N
°C
°C
°C
AVERTISSEMENT La substitution de composants peut rendre ce materiel inacceptable pour les emplacements de Classe I, Division 2.
DISSOLVED OXYGEN TRANSMITTER
MODEL
DO202G-F
SUFFIX
SUPPLY OUTPUT
AMB.TEMP.
9 TO 32VDC FF-TYPE113
-10  55°C STYLE No.
Made in Japan Tokyo 180-8750 JAPAN
Figure 1-1. Nameplates
N200
DISSOLVED OXYGEN TRANSMITTER
MODEL
DO202G-P
SUFFIX
SUPPLY OUTPUT
AMB.TEMP.
9 TO 32VDC PROFIBUS-PA
-10  55°C STYLE No.
Made in Japan Tokyo 180-8750 JAPAN
N200
DISSOLVED OXYGEN TRANSMITTER
MODEL SUFFIX
SUPPLY OUTPUT
AMB.TEMP.
STYLE No.
Made in Japan Tokyo 180-8750 JAPAN
DO202S-K
24V DC 4 20mA DC
-10  55°C
Cert No. GYJ081159X Ex ia IIC T4 Ex ia IIC T6 for Ta:40 SEE USER’S MANUAL BEFORE USE
˚C
1-2 Introduction
DISSOLVED OXYGEN TRANSMITTER
MODEL
DO202S-F
SUFFIX
SUPPLY
OUTPUT
AMB.TEMP.
FISCO
17.5VDC
or 24VDC
/380mA/5.32W
/250mA/1.2W
FF-TYPE111 or 511 Li=0 μH, Ci=220pF
-10  55°C STYLE No.
Made in Japan Tokyo 180-8750 JAPAN
0344
DISSOLVED OXYGEN TRANSMITTER
MODEL
DO202S-B
SUFFIX
SUPPLY OUTPUT
AMB.TEMP.
9 TO 32VDC FF-TYPE 113
-10  55°C STYLE No.
Made in Japan Tokyo 180-8750 JAPAN
Figure 1-2. Nameplates
N200
N200
DISSOLVED OXYGEN TRANSMITTER
MODEL
DO202S-P
SUFFIX
SUPPLY
OUTPUT
AMB.TEMP.
FISCO
17.5VDC
or 24VDC
/380mA/5.32W
/250mA/1.2W
PROFIBUS-PA Li=0 μH, Ci=220pF
-10  55°C STYLE No.
Made in Japan Tokyo 180-8750 JAPAN
0344
DISSOLVED OXYGEN TRANSMITTER
MODEL
DO202S-D
SUFFIX
SUPPLY OUTPUT
AMB.TEMP.
9 TO 32VDC PROFIBUS-PA
-10  55°C STYLE No.
Made in Japan Tokyo 180-8750 JAPAN
N200
N200
FISCO field device
IECEx KEM 07.0029X
No. Zone 0 Ex ia IIC T4
IP65 SEE CONTROL DRAWING
No.
KEMA 07ATEX0054 X
Ex ia IIC T4
II 1G
SEE CONTROL DRAWING IP65
IS CL I, DIV 1, GP ABCD AND AEx ia IIC
Type 4X
T4
Install per CONTROL DRAWING IKE031-A10 P.5 to P.8
CL I, DIV 1, GP ABCD
R
Ex ia IIC T4
LR81741 C
WARNING Substitution of components may impair intrinsic safety
SEE CONTROL DRAWING
IP65 Type 3S
DO202S-F/-P
FNICO field device
IECEx KEM 07.0029X
No. Ex nA[nL] IIC T4 Ex nA[nL] IIC T6 for Ta:40 IP65 SEE CONTROL DRAWING
No.
KEMA 07ATEX0055
EEx nA[nL] IIC T4
II 3 G
EEx nA[nL] IIC T6 for Ta:40 IP65 SEE CONTROL DRAWING
NI CL I, DIV 2, GP ABCD AND CL I, ZN 2, GP IIC T4
Type 4X
Install per CONTROL DRAWING
IKE031-A10 P.9 to P.10
Ex nA[nL] IIC NI CL I, DIV 2, GP ABCD T4
R
T6 for Ta:40
IP65 Type 3S
LR81741 C
WARNING Substitution of components may impair suitability for class I, Division 2.
SEE CONTROL DRAWING
AVERTISSEMENT La substitution de composants peut rendre ce materiel inacceptable pour les emplacements de Classe I, Division 2.
DO202S-B/-D
AVERTISSEMENT La substitution de composants peut compromeltre la securite intrinseque.
°C
°C
°C
Note: Check that all the parts are present, including
mounting hardware, as specified in the option codes at the end of the model number. For a description of the model codes, refer to Chapter 2 of this manual under General Specifications.
Basic Parts List: Transmitter DO202
User’s Manual English Optional mounting hardware when specified (See model code)
Introduction 1-3

1-2. Application

The EXA transmitter is intended to be used for continuous on-line measurement in industrial installations. The unit combines simple operation and microprocessor-based performance with advanced self-diagnostics and enhanced communications capability to meet the most advanced requirements. The measurement can be used as part of an automated process control system. It can also be used to indicate dangerous limits of a process, to monitor product quality, or to function as a simple controller for a BOD/COD system.
Yokogawa designed the EXA analyzer to withstand harsh environments. The transmitter may be installed either indoors or outside because the IP65 (NEMA 4X) housing and cabling glands ensure the unit is ade­quately protected. The flexible polycarbonate window on the front door of the EXA allows pushbutton access to the keypad, thus preserving the water and dust protection of the unit even during routine maintenance operations.
A variety of EXA hardware is optionally available to allow wall, pipe, or panel mounting. Selecting a proper installation site will permit ease of operation. Sensors should normally be mounted close to the transmitter in order to ensure easy calibration and peak performance. If the unit must be mounted remotely from the sensors, WF10 extension cable can be used up to a maximum of box.
The EXA is delivered with a general purpose default setting for programmable items. (Default settings are listed in Chapter 5). While this initial configuration allows easy start-up, the configuration should be adjusted to suit each particular application. An example of an adjustable item is the type of sensor used. The EXA can be adjusted for galvanic and polarographics sensors.
30 metres (100 feet) with a BA10 junction
To record such configuration adjustments, write changes in the space provided in APPENDIX 1 of this manual. Because the EXA is suitable for use as a monitor, a controller or an alarm instrument, program configuration possibilities are numerous.
Details provided in this user’s manual are sufficient to operate the EXA with all Yokogawa sensor systems and a wide range of third-party commercially available probes. For best results, read this manual in conjunction with the corresponding sensor user’s manual.
Yokogawa designed and built the EXA to meet the CE regulatory standards. The unit meets or exceeds stringent requirements of EMC Directive 89/336/EEC, to assure the user of continued accurate performance in even the most demanding industrial installations.
1-4 Introduction

2. GENERAL SPECIFICATIONS

General Specifications 2-1

2-1. Specifications

A. Input specifications
The DO202 accepts output from membrane covered Dissolved Oxygen sensors. These sensors can be Galvanic type, where the sensor generates its own driving voltage or Polarographic type, where the sen­sor uses external driving voltage from the transmitter. The input range is 0-50 A for Galvanic sensors and 0­1 A for Polarographic sensors. For temperature com­pensation the DO202 accepts Pt1000 (DO30 sensor) and 22kNTC elements (OXYFERM and OXYGOLD sensors)
B. Measuring range
- Dissolved Oxygen : 0 - 50 ppm (mg Oxygen
per kg water); 0 - 1999 ppb (mg Oxygen per 1000 kg water) and 0
- 600 % Saturation of Air in Water.
- Temperature : -20 to 150ºC (-4 to 302ºF).
C. Temperature compensation
Automatic using Pt1000 or 22kNTC or manual Range 0 - 100ºC (32 to 212 ºF)
D. Calibration
Semi-automatic one or two point calibration in air or in water using solubility tables of ISO 5814. The influence of pressure and salinity is taken into account when these specifications are entered. Also a two point manual calibration is possible.
E. Transmission Signals
Isolated output of 4-20 mA DC, Burn up (21 mA) or Burn dwon (3.6 mA when HART® comm. non­used, 3.9 mA when HART comm. used) or pulse of 21 mA to signal failure.
Note: The transmitter contains a switched
power supply. The transmitter requires a minimum power voltage in order to work correctly, which is dependant on the load. Please refer to figures 2-1 and 2-2 for the correct power supply
1200.0
1000.0
800.0
775.0
600.0
425.0
400.0
Load Resistance (Ω)
200.0
0.0 12 14 16 18 20 22 24 26 28 30 32 34 36 38 40
Figure 2-1. Supply voltage/ load diagram
17 Volts
14.5 Volts
4 mA
(limit for IS version)
Voltage (V)
22 mA
31.5 V
Terminal voltage (V)
4 mA 7 mA 20 mA
Output Current (mA)
Figure 2-2. Minimum terminal voltage at the DO202
I. Input isolation
1000 VDC
J. Shipping Details
Package size : W x H x D, 290 x 300 x 290
mm (11.5 x 11.8 x 11.5 inch)
Packed weight : approx. 2.5 kg (5lb)
1100.0
Range
Communication
230.0
F. Logbook
Software record of important events and diagnostic data. Available through digital communication.
G. Display
Custom liquid crystal display, with a main display of 3 1/2 digits 12.5 mm high. Message display of 6 alphanumeric characters, 7 mm high.
H. Power supply
Nominal 24 volt DC loop powered system DO202G-A : up to 40 volts DO202S-A, -N: up to 31.5 volts DO202G/S-F, -P: 9-24, 9-17.5 (FISCO) VDC /
26.0 mA
DO202S-B, -D: 9-32 VDC / 26.0 mA
2-2 General Specifications

2-2. Performance specifications

A. Performance in ppm mode
- Linearity : ±0.05 ppm or ±0.8% FS, whichever is greater
- Repeatability : ±0.05 ppm or ±0.8% FS, whichever is greater
- Accuracy : ±0.05 ppm or ±0.8% FS, whichever is greater
B. Performance in ppb mode
- Linearity : ±1 ppb or ±0.8% FS, whichever is greater
- Repeatability : ±1 ppb or ±0.8% FS, whichever is greater
- Accuracy : ±1 ppb or ±0.8% FS, whichever is greater
C. Performance in Temperature
- Linearity : ±0.3ºC
- Repeatability : ±
0.1ºC
- Accuracy : ±0.3ºC
Note on performance specifications:
The specifications are at reference temperature
and with simulated inputs, because the DO202G can be used with many different sensors with their unique characteristics. The following tolerance is added to above performance. mA output tolerance : ± 0.02 mA of "4 - 20 mA"
D. Ambient temperature
Operating temperature -10 to 55ºC
(10 to 130ºF)
Storage temperature -30 to 70ºC
(-20 to 160ºF)
E. Humidity
10 to 90% RH non-condensing
F. Housing
Cast aluminum housing with chemically
resistant coating, cover with flexible polycarbonate window. Case color is off-white with moss green cover. Cable entry is through two polyamide cable glands. Cable glands are provides for wiring up to 2.5 mm.
G. Environmental protection
IP65, NEMA 4X.
H. Mounting
Pipe, wall or panel mounting using optional hardware.
I. Data protection
EEPROM for configuration; Lithium battery for
clock.
J. Watchdog timer: Checks microprocessor.
K. Autoreturn
The transmitter returns into measuring mode automatically when no keystroke is made for 10 minutes.
L. Operation protection
All three levels can be protected by a user programmable 3 digit password
M. EMC Conformity standards
EN 61326-1 Class A, Table 2 (For use in industrial locations) EN 61326-2-3 EN 61326-2-5 (pending)
CAUTION
This instrument is a Class A product, and it is designed for use in the industrial environment. Please use this instrument in the industrial environment only.
N. Explosionproof type
Refer to Control Drawings.
mA
Item
Factory Mutual (FM)
CENELEC ATE X
Item
Factory Mutual (FM)
FM Intrinsically safe Approval
Applicable standard: FM3600, FM3610, FM3810 Intrinsically Safe for Class I, Division 1, Groups ABCD Class I, Zone 0, AEx ia IIC Temp. Class: T4, Amb. Temp.: -10 to 55°C Intrinsically Safe Apparatus Parameters Vmax=31.5 V, Imax=100 mA, Pmax=1.2 W, Ci=22 nF, Li=35 μH
FM Non-incendive safe Approval
Applicable standard: FM3600, FM3611, FM3810 Non-incendive Safe for Class I, Division 2, Groups ABCD, Zone 2 Temp. Class: T4, Amb. Temp.: -10 to 55°C Non-incendive Safe Apparatus Parameters Vmax=31.5 V, Ci=22 nF, Li=35 μH
CENELEC ATEX (KEMA) Intrinsically safe Approval
Applicable standard: EN60079-0, EN50020 EN60079-26 Certificate: KEMA 06ATEX0224 X
Ex ia IIC, Group: II, Category: 1G
Temp. Class: T4, Amb. Temp.: -10 to 55°C T6, Amb. Temp.: -10 to 40°C Ui=31.5 V, Ii=100 mA, Pi=1.2 W, Ci=22 nF, Li=35 μH
CENELEC ATEX (KEMA) Type of protection "n"
Applicable standard: EN60079-0:2006, EN60079-15:2003 Certificate: KEMA 06ATEX0225
EEx nA [nL] IIC, Group: II, Category: 3G
Temp. Class: T4, Amb. Temp.: -10 to 55°C T6, Amb. Temp.: -10 to 40°C Ui=31.5 V, Ci=22 nF, Li=35 μH
FM Intrinsically safe Approval
Applicable standard: FM3600, FM3610, FM3810 Intrinsically Safe for Class I, Division 1, Groups ABCD Class I, Zone 0, AEx ia IIC Temp. Class: T4, Amb. Temp.: -10 to 55°C Intrinsically Safe Apparatus Parameters
FM Non-incendive safe Approval
Applicable standard: FM3600, FM3611, FM3810 Non-incendive Safe for Class I, Division 2, Groups ABCD, Zone 2 Temp. Class: T4, Amb. Temp.: -10 to 55°C Non-incendive Safe Apparatus Parameters
Vmax=24 V, Imax=250 mA,
Entity
Pmax=1.2 W, Ci=220 pF, Li=0 μH Vmax=17.5 V, Imax=380 mA,
FISCO
Pmax=5.32 W, Ci=220pF, Li=0 μH
Vmax=32 V, Pmax=1.2 W,
Entity
Ci=220 pF, Li=0 μH Vmax=32 V, Pmax=5.32 W,
FNICO
Ci=220 pF, Li=0 μH
Description
Description
,
Code
-A
-N
-A
-N
2.EPS
Code
-P
or
-F
-B
or
-D
FM.EPS
General Specifications 2-3
Item
CENELEC ATE X
Entity
CENELEC ATE X
FISCO
CENELEC ATE X
mA
Item
Canadian Standards Association (CSA)
IECEx Scheme
Item
Canadian Standards Association (CSA)
Description
CENELEC ATEX (KEMA) Intrinsically safe Approval
Applicable standard: EN60079-0, EN50020 EN60079-26 Certificate: KEMA 07ATEX0054 X
Ex ia IIC, Group: II, Category: 1G
Temp. Class: T4, Amb. Temp.: -10 to 55°C Ui=24 V, Ii=250 mA, Pi=1.2 W, Ci=220 pF, Li=0 μH
CENELEC ATEX (KEMA) Intrinsically safe Approval
Applicable standard: EN60079-0, EN50020 EN60079-26, EN60079-27 Certificate: KEMA 07ATEX0054 X
Ex ia IIC, Group: II, Category: 1G
Temp. Class: T4, Amb. Temp.: -10 to 55°C Ui=17.5 V, Ii=380 mA, Pi=5.32 W, Ci=220 pF, Li=0 μH
CENELEC ATEX (KEMA) Type of protection "n"
Applicable standard: EN60079-0:2006, EN60079-15:2003 Certificate: KEMA 07ATEX0055
EEx nA [nL] IIC, Group: II, Category: 3G
Temp. Class: T4, Amb. Temp.: -10 to 55°C T6, Amb. Temp.: -10 to 40°C Ui=32 V, Ci=220 pF, Li=0 μH
CSA Intrinsically safe Approval
Applicable standard: C22.2, No. 0-M1991, C22.2, No. 04-M2004, C22.2, No. 157-M1992, C22.2, No. 61010-1
Ex ia Class I, Division 1, Groups ABCD Ex ia IIC
Temp. Class: T4, Amb. Temp.: -10 to 55°C T6, Amb. Temp.: -10 to 40°C Ui(Vmax)=31.5 V, Ii(Imax)=100 mA, Pi(Pmax)=1.2 W, Ci=22 nF, Li=35 μH
CSA Non-incendive safe Approval or type of protection "n"
Applicable standard: C22.2, No.0-M1991, C22.2, No.04-M2004, C22.2, No.157-M1992, C22.2, No.213-M1987, C22.2, No.61010-1 Class I, Division 2, Groups ABCD
Ex nA [nL] IIC
Temp. Class: T4, Amb. Temp.: -10 to 55°C T6, Amb. Temp.: -10 to 40°C Ui(Vmax)=31.5 V, Ci=22 nF, Li=35 μH
IECEx Intrinsically safe
Applicable standard: IEC 60079-0, IEC60079-11, IEC60079-26 Certificate: IECEx KEM 06.0055X
Zone 0 Ex ia IIC
Temp. Class: T4, Amb. Temp.: -10 to 55°C T6, Amb. Temp.: -10 to 40°C Ui=31.5 V, Ii=100 mA, Pi=1.2 W, Ci=22 nF, Li=35 μH
IECEx Type of protection "n"
Applicable standard: IEC 60079-15:2001, IEC 60079-0:2004 Certificate: IECEx KEM 06.0055X
Ex nA [nL] IIC
Temp. Class: T4, Amb. Temp.: -10 to 55°C T6, Amb. Temp.: -10 to 40°C Ui=31.5 V, Ci=22 nF, Li=35 μH
CSA Intrinsically safe Approval
Applicable standard: C22.2, No. 0-M1991, C22.2, No. 04-M2004, C22.2, No. 157-M1992, C22.2, No. 61010-1
Ex ia Class I, Division 1, Groups ABCD Ex ia IIC
Temp. Class: T4, Amb. Temp.: -10 to 55°C
Entity
FISCO
CSA Non-incendive safe Approval or type of protection "n"
Applicable standard: C22.2, No.0-M1991, C22.2, No.04-M2004, C22.2, No.157-M1992, C22.2, No.213-M1987, C22.2, No.61010-1 Class I, Division 2, Groups ABCD
Ex nA [nL] IIC
Temp. Class: T4, Amb. Temp.: -10 to 55°C T6, Amb. Temp.: -10 to 40°C
Entity: FNICO:
Description
Description
Ui(Vmax)=24 V, Ii(Imax)=250 mA, Pi(Pmax)=1.2 W, Ci=220 pF, Li=0 μH Ui(Vmax)=17.5 V, Ii(Imax)=380 mA, Pi(Pmax)=5.32 W, Ci=220 pF, Li=0 μH
Ui(Vmax)=32 V, Ci=220 pF, Li=0 μH Ui(Vmax)=32 V, Ci=220 pF, Li=0 μH
Code
-P
or
-F
-B
or
-D
ATEX.EPS
Code
-A
-N
-A
-N
T12E.EPS
Code
-P
or
-F
-B
or
-D
CSA.EPS
Item
IECEx Scheme
Entity
IECEx Scheme
FISCO
IECEx Scheme
IECEx Intrinsically safe
Applicable standard: IEC 60079-0, IEC60079-11, IEC60079-26 Certificate: IECEx KEM 07.0029X
Zone 0 Ex ia IIC
Temp. Class: T4, Amb. Temp.: -10 to 55°C Ui=24 V, Ii=250 mA, Pi=1.2 W, Ci=220 pF, Li=0 μH
IECEx Intrinsically safe
Applicable standard: IEC 60079-0, IEC60079-11, IEC60079-26, IEC60079-27 Certificate: IECEx KEM 07.0029X
Zone 0 Ex ia IIC
Temp. Class: T4, Amb. Temp.: -10 to 55°C Ui=17.5 V, Ii=380 mA, Pi=5.32 W, Ci=220 pF, Li=0 μH
IECEx Type of protection "n"
Applicable standard: IEC 60079-15:2001, IEC 60079-0:2004 Certificate: IECEx KEM 07.0029X
Ex nA [nL] IIC
Temp. Class: T4, Amb. Temp.: -10 to 55°C T6, Amb. Temp.: -10 to 40°C Ui=32 V, Ci=220 pF, Li=0 μH
Description
mA
NEPSI Certification (DO202S-K)
NEPSI Intrinsically Safe Type
Cert No. GYJ081159X
• Applicable Standard: GB3836.1-2000, GB3836.4-2000
• Type of Protection and Marking Code: Ex ia IIC T4/T6
• Ambient Temperature : T6; –10 to 40°C, T4; –10 to 55°C
Note 1 Entity Parameters
• Intrinsically safe input parameters (terminal + and -):
Maximum Input Voltage (Ui) = 31.5 V Maximum Input Current (Ii) = 100 mA Maximum Input Power (Pi) = 1.2 W Maximum Internal Capacitance (Ci) = 22 nF Maximum Internal Inductance (Li) = 35 H
• Intrinsically safe output parameters and maximum external parameters
(terminal 11 and 17):
Uo=14.4 V, Io=22 mA, Po=187 mW, Co=643 nF, Lo=70 mH
Note 2 Installation
• Electrostatic charges on the display window shall be avoided.
• The external earth connection facility shall be connected reliably.
• The instrument modification or parts replacement by other than authorized representative of Yokogawa Electric Corporation and will void NEPSI Intrinsically safe certification.
• The user shall not change the configuration in order to maintain/ensure the explosion protection performance of the equipment. Any change may impair safety.
• For installation, use and maintenance of the product, the end user shall observe the instruction manual and the following standards:
GB50257-1996 "Code for construction and
acceptance of electric device for explosion atmospheres and fire hazard electrical equipment installation engineering''.
Code
-P
or
-F
-B
or
-D
IEC.EPS
2-4 General Specifications
GB3836.13-1997 "Electrical apparatus for explosive gas atmospheres Part 13: Repair and overhaul for apparatus used in explosive gas atmospheres". GB3836.15-2000 "Electrical apparatus for explosive gas atmospheres- Part 15: Electrical installations in hazardous area (other than mines)" . GB3836.16-2006 "Electrical apparatus for explosive gas atmospheres- Part 16: lnspection and maintenance of electrical installation (other than mines)".
mA
mA-HART® communication A. Input : Two wire system 4-20 mA
B. Power supply : Nominal 24 volt DC loop powered
DO202G : up to 40 volts DO202S : up to 31.5 volts Note: The transmitter contains a switched
power supply, drawing its energy from the 0-4 mA section of the signal. Consequently the 17 volt limit is applied at 4 mA. The characteristic of the unit is such that above about 7 mA on the output, the terminal voltage can drop to 14.5 volts without problem. (see figure 2-2)
C. Transmission: Isolated output of 4 to 20 mA DC. D. Signal : Maximum load 425 at 24 VDC.
(see fi gure 2-1)
Burn to signal failure acc.
NAMUR Recommendation NE43 (18.01.1994)
E. Operating range : 3.9 to 21mA F. Communication
: HART®, 1200 Baud, FSK modulated
on 4 to 20 mA signal
G. Configuration : Local with 6 keys
H. Software : Firmware based on Yokogawa stack.
I. Hardware :
J. Hand Terminal :
JYokogawa HART® Modem F9197UB
Rosemount HHT 275/375
K. Other Control systems
: Yokogawa PRM, Rosemount AMS,
Siemens PDM
L. Output span :
There are three output modes in which the output span can be freely selected: ppm (mg Oxygen per kg of water): 0 - 50 ppm with minimum span of 1 ppm; ppb (mg Oxygen per 1000 kg of water): 0 - 1999 ppb with minimum span of 25 ppb and % Saturation: 0 - 600 % with minimum span of 10 %
M. Cable specification
0.5 mm diameter or 24 AWG over maximum length of 1500 m
N. DD specification
The DO202 Device Description is available
enabling communications with the Handheld Communicator and compatible devices.
PROFIBUS-PA communications A. Input signal: Digital
B. Supply voltage: 9 to 32 V DC C. Operating current: 26.0 mA D. Operating values: According to IEC 1158-2 E. Bus connection
: Fieldbus interface base on
IEC1158-2 according to FISCO­Model
F. Power supply: Power supply is achieved depend-
ant on the application by means of segment coupler
G. Data transfer: According to PROFIBUS- PA
profile class B based on EN 50170 and DIN 19245 part 4
H. GSD file: The actual file can be downloaded
from www.profibus.com Configu­ration: Local with 6 keys
I. Software: Firmware based on Siemens
DPC31 stack.
J. Hardware:
PC- or PCMCIA-interfaces from
Siemens
K. Other control: Siemens PDM systems L Electrical connection:
Terminals acc. to IEC 1158-2
M. Fieldbus-cable-types:
Twisted and shielded two
wire cable according to recommendation based on IEC 1158-2 Cable diameter: 6 to 12 mm (0.24 to 0.47 inch)
FOUNDATION FIELDBUS H1 communications A. Input signal: Digital B. Supply voltage: 9 to 32 V DC C. Operating current: 26.0 mA (base current) D. Operating values: According to IEC 1158-2
E. Bus connection
: Fieldbus interface based on IEC
1158-2 according to FISCO-Model
F. Power supply:
Power supply is achieved
dependant on application by means of segment coupler
G. Data transfer:
FF specification Rev. 1.4 Basic
device
H. Function blocks:
3 x AI, Transducer, Resource
I. Files: Actual file can be downloaded from
our homepage
J. Configuration: locally with 6 keys K. Software: National Instruments:
NI-FBUS configurator
L. Hardware: F-BUS interfaces from National In-
struments (AT-FBUS, PCMIA-FBUS)
M. Other control systems:
YOKOGAWA PRM, DTM

2-3. MODEL AND SUFFIX CODES

Model Suffi x Code Option Code Description
DO202G --------------------- ----------------------- 2-wire Dissolved Oxygen transmitter
Type -A
-P
-F
Language -E
-J
Options
Mounting Hardware
Tag Plate
Conduit Adapter
*1 It can be specifi ed when the suffi x code -A is selected. *2 The housing is coated with epoxy resin.
Hood
-----------------------
-----------------------
-----------------------
-----------------------
-----------------------
/U /PM /H /H2 /SCT /AFTG /ANSI /TB /X1
mA with HART Profi bus FF
English Japanese
Pipe, Wall mounting bracket (Stainless steel) Panel mounting bracket (Stainless steel)
Hood for sun protection (Carbon steel) Hood for sun protection (Stainless steel)
Stainless steel tag plate G 1/2 1/2NPT Screw terminal (*1) Epoxy baked fi nish (*2)
General Specifications 2-5
[Style: S2]
[Style: S3]
Model Suffi x Code Option Code Description
DO202S -------------------- ------------------------- 2-wire Dissolved Oxygen transmitter
Type -A
-K
-P
-F
-B
-N
-D
Language -E
-J
Options
Mounting Hardware
Tag Plate
Conduit Adapter
*1 The housing is coated with epoxy resin. *2 When the instrument with Suffix Code "-B,-N,-D" is used, take measures so that the display window is not exposed to direct
sunlight.
-------------------------
-------------------------
-------------------------
-------------------------
-------------------------
-------------------------
-------------------------
-------------------------
-------------------------
/U /PM
Hood
/H /H2 /SCT /AFTG /ANSI /X1
Intrinsically safe mA with HART (ATEX, CSA, FM) Intrinsically safe mA with HART (NEPSI) Intrinsically safe Profi bus (ATEX, CSA, FM) Intrinsically safe Non-Incendive FF (ATEX, CSA, FM) (*2) Non-Incendive mA with HART (ATEX, CSA, FM) (*2) Non-Incendive Profi bus (ATEX, CSA, FM) (*2)
English Japanese
Pipe, Wall mounting bracket (Stainless steel) Panel mounting bracket (Stainless steel) Hood for sun protection (Carbon steel) Hood for sun protection (Stainless steel) Stainless steel tag plate G 1/2 1/2NPT Epoxy baked fi nish (*1)
FF
(ATEX, CSA, FM)
General Specifications 2-6
_
_
_
_

2-4. Control Drawing of DO202S mA HART® Specification (IECEx)

SENSOR(S) termin als 1 1 -1 8
SENSOR(S) termin als 1 1 -1 8
Intrinsically safe design IECE x sta nd a rd E x ia IIC : T4 for ambie nt te mp . < 5 5° C
Ce rt ific a te n r. IE C Ex KE M 06.0 0 5 5X
(Dissolved Oxygen -transmitter)
DO202S
T6 for ambient temp. < 40°C
+
G
Functional earth
Hazardous area Safe area
Zone 0 or 1
Intrinsically safe design
IECE x sta nd a rd E x ia IIC : T 4 for a m bie nt te mp . < 55°C Certificate nr. IECEx KEM 0 6.0 055 X
Dissolved Oxygen -transmitter)
DO202S
T6 for ambient temp.< 40°C
+
G
Functional earth
Hazardous area
Zone 0 or 1
Functional earth
Ex ia o r ib Ce rt ifie d safety b arrier o r p ower with Rint=300 (HAR T c o m pa tib le)
Uo = 31.5 Volt DC Io = 100 mA
:
24 volts DC N om inal
Sup p ly V oltag e .
Load
Resistance
Ex ia o r ib C ertified R epea t e r Power Supply (HA R T Compatible)
+
Uo = 31.5 Volt DC
Io = 100 mA
Po = 1.2 Watt
Safe area
+
Output
Supply
Electrical data of the DO202S.
- Supply and output circuit (terminals + and -):
Maximum input voltage U
Maximum input power P Effective internal capacitance C Effective internal inductance L
= 31.5 V. Maximum input current Ii = 100 mA.
i
= 1.2 W.
i
= 22 nF.
i
= 35 PH.
i
- Sensor input circuit (terminals 11 through 18): Maximum output voltage U Maximum allowed external capacitance C
= 14.4 V. Maximum output current Io = 22 mA.
o
= 643 nF.(for DO202S-A)
o
Co = 3.5 μF(for DO202S-N)
Maximum allowed external inductance L
= 70 mH. (for DO202S-A)
o
Lo = 160 mH. (for DO202S-N) ・ Barriers and power supply specification must not exceed the maximum values
as shown in the diagram above. These safety descriptions cover most of the commonly used industry standard barriers, isolators and power supplies. The Hand Held Communicator must be of a IECEx certified intrinsically safe type in case it is used on the intrinsically safe circuit in the hazardous area or of a IECEx certified non-incendive type in case it is used in the non-incendive circuit in the hazardous area.

2-5. Control Drawing of DO202S mA HART® Specification (ATEX)

_
_
_
2-7 General Specifications
Intrin s ic ally safe d esign ATE X C la ss I, D iv.1, Group A B CD, T4 for am b ient tem p . < 55 °C T6 for ambient temp. < 40°C Certificate nr. KEMA 06AT EX 0224 X DO 2 0 2S tra ns m itter
Sensor terminals 1 1 -18 Max. cablelength: 60 mtr. Cable dia. : 3…12 mm.
Intrinsically safe design ATE X Class I, Div.1, G ro up A B C D , T 4 for a m bien t tem p. < 55 °C T6 for ambient temp. < 40°C Certificate nr. KEMA 06ATEX 0224 X DO 2 0 2S tr an sm itter
Sensor terminals 1 1 -18 Max. cablelength: 60 mtr. Cable dia.: 3…12 mm.
+
G
For electrical data: see text below.
Classified Location
+
G
For electrical data: see text below.
Classified Location
Functional earth
Functional earth
ATEX Approved safety barrier or power supply with Rint = 300 : (HAR T co m pa tible)
Functional earth
Unclassified Location
Load
Resistance
ATEX Approved Power Supply (HART compatible)
+
Ùnclassified Location
24 volts DC Nominal
Supply Voltage.
+
-
Figure 1
Output
Supply
Figure 2
Electrical data of the DO202S.
- Supply and output circuit (terminals + and -):
Maximum input voltage U
Maximum input power P
= 31.5 V. Maximum input current Ii = 100 mA.
i
= 1.2 W.
i
Effective internal capacitance Ci = 22 nF. Effective internal inductance L
= 35 PH.
i
- Sensor input circuit (terminals 11 through 18): Maximum output voltage U Maximum allowed external capacitance C
= 14.4 V. Maximum output current Io = 22 mA.
o
= 643 nF.(for DO202S-A)
o
Co = 3.5 μF(for DO202S-N) Maximum allowed external inductance Lo = 70 mH. (for DO202S-A) L
= 160 mH. (for DO202S-N)
o
Barriers and power supply specification must not exceed the maximum values
as shown in the diagram above. These safety descriptions cover most of the
commonly used industry standard barriers, isolators and power supplies. The Hand Held Communicator must be of an ATEX certified intrinsically safe type in case it is used on the intrinsically safe circuit in the hazardous area or of an
ATEX certified non-incendive type in case it is used in the non-incendive circuit in the hazardous area.
_
_
_
General Specifications 2-8

2-6. Control Drawing of DO202S mA HART® Specification (FM Intrinsically safe design)

Intrins i c ally safe design FM Class I, Div.1, Group ABCD, T4 for ambient temp. < 55°C T6 for ambient temp. < 40°C DO202S transmitter
Sensor terminals 11-18 Max. cablelength: 60 mtr. Cable dia. : 3…12 mm.
Intrins ic ally safe design FM Class I, Div.1, Group ABCD, T4 for ambient temp. < 55°C T6 for ambient temp. < 40°C DO202S transmitter
+
G
For electrical data: see text below.
Classified Location
+
G
Functional earth
FM Approved safety barrier or power supply with Rint = 300 : (HART compatible)
Functional earth
Unclassified Location
24 volts DC Nominal
Load
Resistance
FM Approved Power Supply (HART compatible)
+
Supp ly Voltage.
+
-
Figure 1
Output
Supply
Sensor terminals 11-18 Max. cablelength: 60 mtr. Cable dia.: 3…12 mm.
For electrical data: see text below.
Classified Location
Functional earth
Ùnclassified Location
Figure 2
Electrical data of the DO202S.
- Supply circuit (terminals + and -): Maximum input voltage Vmax = 31.5 V. Maximum input current Imax = 100 mA. Maximum input power Pi = 1.2 W. Effective internal capacitance Ci = 22 nF. Effective internal inductance Li = 35 PH.
- Sensor input circuit (terminals 11 through 18): Maximum output voltage Vt= 14.4 V. Maximum output current It = 10 mA. Maximum allowed external capacitance Ca = 643 nF. Maximum allowed external inductance L
= 340 mH.
a
If Hand Held Terminal (HHT) is not connected to the power supply lines of the DO202S
(see figure 1):
Any FM Approved barrier or power supply may be used that meets the following requirements. V
or V
oc
d 31.5 V; Isc or Itd 100 mA; Cat 22nF + C
t
; Lat 35P H + L
cable
cable
If HHT is connected to the power supply lines of the DO202S (see figure 2): The Hand Held Terminal must be FM Approved. Refer to the manufacturers control drawing of the HHT and the barrier/power supply to determine the cable parameters. (Voc or Vt ) + V C
t 22nF + C
a
d 31.5 V; (Isc or It ) + I
HHT
cable
+ C
; Lat 35P H + L
HHT
d 100 mA;
HHT
+ L
cable
HHT
When installing this equipment, follow the manufacturer’s installation drawing. Installation should be in accordance with ANSI/ISA RP 12.06.01 “Installation of Intrinsically Safe
Systems for Hazardous (Classified) Locations” and the National Electrical Code (ANSI/NFPA 70). Control equipment connected to the barrier/power supply must not use or generate more than 250 Vrms or Vdc.
Resistance between Intrinsically Safe Ground and earth ground must be less than1.0 Ohm. In case of using cable glands in Outdoor location, they shall be UV rated or made of metal.
WARNING
- Substitution of components may impair Intrinsic Safety
- To prevent ignition of flammable or combustible atmospheres, disconnect power before servicing or read, understand and adhere to the manufacturer’s’live maintenance procedures.
Application Doc. No.: IKE030-A10 P.5 to P.6
_
_
N
N
2-9 General Specifications

2-7. Control Drawing of DO202S mA HART® Specification (FM Non-incendive design)

onincnendive design FM Class I, D iv .2, G roup A B C D , T4 for a m b ien t tem p . < 5 5 °C T6 for ambient temp. < 40°C DO 202 S tran sitte r
+
G
FM A pproved power supply
Voc 31.5 V DC
+
-
Sensor terminals 11-18 Max. cablelength: 60 mtr. Cable dia. : 3… 12 mm .
Sensor terminals 11-18 Max. cablelength: 60 mtr. Cable dia.: 3… 12 mm
For electrical data: see text below .
Classified Location Unclassified Location
onincnendive design FM C lass I, D iv .2, G roup A B C D , T4 for a m b ien t tem p . < 5 5 °C T6 for ambient temp. < 40°C DO202S transmitter
Functional earth
+
G
For electrical data: see text below.
Classified Location
Functional earth
Load
Resistance
FM A pproved
power supply
Voc 31.5 VDC
+
-
Ùnclassified Location
Electrical data of the DO202S.
- Supply circuit (terminals + and -): Maximum input voltage V
= 31.5 V. Maximum input power P
max
= 1.2 W
max
Effective internal capacitance Ci = 22 nF Effective internal inductance Li = 35 μH
Sensor input circuit (terminals 11 through 18): Maximum output voltage V Maximum allowed external capacitance C
Maximum allowed external inductance L
= 14.4 V. Maximum output current It = 10 mA.
t
= 2.29 μF.
a
= 600 mH.
a
The Hand Held Terminal must be FM Approved in case it is used in the classified location.
When installing this equipment, follow the manufacturers installation drawing. Installation shall be in accordance with Article 501.4(B) of the National Electrical Code (ANSI/NFPA 79). Nonincendive field wiring may be installed in accordance with Article 501 of the National Electrical Code
Grounding shall be in accordance with Article 250 of the National Electrical code In case of using cable glands in Outdoor location, they shall be UV rated or made of metal.
WARNING
- Substitution of components may impair suitability for Division 2
- Do not remove or replace while circuit is live unless area is know to be non-hazardous
- Explosion Hazard – Do not disconnect equipment unless area is know to be on-hazardous
- Do not reset circuit breaker unless power has been removed from the equipment or the
area is know to be non-hazardous
Application Doc. No.: IKE030-A10 P.7 to P.8
General Specifications 2-10
_
_
_

2-8. Control Drawing of DO202S mA HART® Specification (CSA)

Sensor terminals 1 1-1 8
Sensor terminals 11-18
Intrin s ic a lly s a fe d e s ign CSA E x ia C lass I, Div.1, G ro up A B C D, T4 for amb ient tem p . < 5 5 °C T6 for am b ient tem p. < 4 0° C
DO202S transmitter
+
G
For electrical data: see text below.
Hazardous area Safe area
Intrinsically safe d esign CSA Ex ia Class I, Div.1, Group ABCD, T4 for ambient temp. < 55°C T6 for am b ient tem p . < 4 0° C
DO202S transmitter
Fuctional earth
Functional earth
+
G
For electrical data: see text below.
Hazardous area
Functional earth
CSA certified safety barrier or power supply with Rint=300 : (HA RT c o m p atible )
Suitable values are:
Vmax = 31.5 VoltDC
Imax = 100 mA
Load
Resistance
CSA certified Power Supply (HA RT co m p a tib le) )
+
Suit ab le value s a r e :
Vma x = 31 .5 V o ltDC Imax = 100 mA Pmax = 1.2 W att
Safe area
24 volts DC N o minal
Supply Voltage.
+
-
Output
Supply
Electrical data of the DO202S.
- Supply and output circuit (terminals + and -) Maximum input voltage V Maximum input power P Effective internal capacitance C
= 31.5 V. Maximum input current I
max
= 1.2 W.
max
= 22nF. Effective internal inductance Li = 35 PH.
i
= 100 mA.
max
- Sensor input circuit (terminals 11 through 18): Maximum output voltage V Maximum allowed external capacitance C Maximum allowed external inductance L
= 14.4 V. Maximum output current Isc = 22 mA.
oc
= 643 nF.
a
= 70 mH.
a
Barriers and power supply should be CSA certified. The specifications must not exceed the
maximum values as shown in the diagram above. Installation should be in accordance with Canadian Electrical Code, Part I.
Maximum safe area voltage should not exceed 250 V
RMS
.
For Class I, Div. 2, Group ABCD the CSA certified barrier is not required, and the
Sensor input circuit (terminals 11 through 18) is non-incendive having the parameters : Maximum output voltage V Maximum allowed external capacitance C Maximum allowed external inductance L
= 14.4 V. Maximum output current Isc = 22 mA.
oc
= 3.5μF.
a
= 160mH.
a
The Hand Held Communicator must be of a CSA certified intrinsically safe type in case it
is used on the intrinsically safe circuit in the hazardous area, or of a CSA certified non-incendive type in case it is used on the non-incendive circuit in the hazardous area.

2-9. Control Drawing of DO202S FF/PB Specification (IECEx)

Ex ia IIC T4 for am bient temp. d 55 qC Ui = 24 V or U i= 1 7,5 V Ii = 250 mA Ii = 380 m A Pi = 1,2 W Pi = 5,32 W
DO202S-F
or DO202S-P
-
+
2-11 General Specifications
Sensor
Connections
Safe area
Apparatus
+
-
Safe area
I.S.
interface
I.S.
certified
Terminator
+
Transm itter
Zone 0 or 1
Hazardous area
-
Transm itter
x Sensor(s) are of a passive type to be regarded as 'simple apparatus'.
x Electrical data of the DO202S-F & DO202S-P:
- Supply and output circuit:
Maximum input voltage Ui = 24 V Maximum input current Ii = 250 mA
Maximum input power Pi = 1.2 W Effective internal capacitance Ci = 220 pF; Effective internal inductance Li = 0 H.
or FISCO field device Maximum input voltage Ui = 17.5 V Maximum input current Ii = 380 mA Maximum input power Pi = 5.32 W Effective internal capacitance Ci = 220 pF; Effective internal inductance Li = 0 H.
I.S.
certified
Terminator
-
+
- Sensor input circuit:
Maximum output voltage Uo = 14.4 V; Maximum output current Io = 22 mA Maximum allowed external capacitance Co = 643 nF Maximum allowed external inductance Lo = 70 mH
x Any I.S. interface may be used that meets the following requirements:
Uo d 24 V Io d 250 mA Po d 1.2 W Co t 220 pF + Ccable; Lo t 0 H + Lcable
or
FISCO power supply
Uo d 17.5 V Io d 380 mA Po d 5.32 W
Co t 220 pF + Ccable; Lo t 0 H + Lcable
x Electrical data of the DO202S-B & DO202S-D (Type of protection “n”)
- Supply and output circuit:
Maximum input voltage Ui = 32 V
Effective internal capacitance Ci = 220pF; Effective internal inductance Li = 0 H.
- Sensor input circuit:
Maximum output voltage Uo = 14.4 V; Maximum output current Io = 22 mA Maximum allowed external capacitance Co = 3.5PF Maximum allowed external inductance Lo = 160 mH
General Specifications 2-12

2-10. Control Drawing of DO202S FF/PB Specification (ATEX)

Ex ia IIC T4 for ambient temp. d 55 qC Ui = 24 V or Ui = 17,5 V Ii = 250 mA Ii = 380 mA Pi = 1,2 W Pi = 5,32 W
DO202S-F
or DO202S-P
-
+
Sensor
Connections
Safe area
Apparatus
+
-
Safe area
I.S.
interface
I.S.
certified
Terminator
+
Transmitter
Zone 0 or 1
Hazardous area
-
+
Transmitter
x Sensor(s) are of a passive type to be regarded as 'simple apparatus'. x Electrical data of the DO202S-F & DO202S-P:
- Supply and output circuit: Maximum input voltage Ui = 24 V Maximum input current Ii = 250 mA Maximum input power Pi = 1.2 W
Effective internal capacitance Ci = 220 pF; Effective internal inductance Li = 0 H. or FISCO field device Maximum input voltage Ui = 17.5 V Maximum input current Ii = 380 mA Maximum input power Pi = 5.32 W Effective internal capacitance Ci = 220 pF; Effective internal inductance Li = 0 H.
I.S.
certified
Terminator
-
- Sensor input circuit: Maximum output voltage Uo =14.4 V; Maximum output current Io = 22 mA Maximum allowed external capacitance Co = 643 nF Maximum allowed external inductance Lo = 70 mH
x Any I.S. interface may be used that meets the following requirements:
Uo d 24 V Io d 250 mA Po d 1.2 W Co t 220 pF + Ccable; Lo t 0 PH + Lcable
or
FISCO power supply Uo d 17.5 V Io d 380 mA Po d 5.32 W Co t 220 pF + Ccable; Lo t 0 PH + Lcable
x Electrical data of the DO202S-B & DO202S-D (Type of protection “n”)
- Supply and output circuit: Maximum input voltage Ui = 32V
Effective internal capacitance Ci = 220pF; Effective internal inductance Li = 0 H.
- Sensor input circuit: Maximum output voltage Uo = 14.4V; Maximum output current Io = 22 mA Maximum allowed external capacitance Co = 3.5 PF Maximum allowed external inductance Lo = 160 mH
2-13 General Specifications

2-11. Control Drawing of DO202S FF/PB Specification (FM Intrinsically safe Entity)

FM Approved barrier
Voc (Vt) d 24 V Ioc (It) d 250 mA Poc (Pt) d 1.2 W Ca t 220pF+ Ccable La t 0 H + Lcable
FM Class I, DIV. 1, Group ABCD T4 for ambient temp. d 55 qC
DO202S-F
or DO202S-P
-
+
+
-
I.S.
certified
Terminator
-
+
Transmitter
Sensor Connections Max. cablelength: 60 mtr. Cable dia. : 3…12 mm.
Sensor
Connections
Terminator
-
+
Transmitter
I.S.
certified
Division 1
Unclassified Location
Classified Location
x Sensor(s) are of a passive type to be regarded as 'simple apparatus', devices which
neither store nor generate voltages over 1.5 V, currents over 0.1 A, power over 25 mW or energy over 20 PJ, or are FM Approvals entity approved and meet connection requirements.
x Electrical data of the DO202S-F & DO202S-P:
- Supply circuit:
Maximum input voltage Vmax = 24 V Maximum input current Imax = 250 mA
Maximum input power Pi = 1.2 W Effective internal capacitance Ci = 220 pF; Effective internal inductance Li = 0 PH.
- Sensor input circuit:
Maximum output voltage Vt = 14.4 V; Maximum output current It = 10 mA
Maximum allowed external capacitance Ca = 643 nF
Maximum allowed external inductance La = 340 mH
x Any FM Approved barrier may be used that meets the following requirements:
Voc or Vt d 24 V Ioc or It d 250 mA Poc or Pt d 1.2 W Ca t 220 pF + Ccable; La t 0 H + Lcable
When installing this equipment, follow the manufacturer’s installation drawing. Installation should be in accordance with ANSI/ISA RP 12.06.01 “Installation of Intrinsically Safe Systems for Hazardous (Classified) Locations” and the National Electrical Code (ANSI/NFPA 70). Associated apparatus connected to the barrier must not use or generate more than 250 Vrms or Vdc.
x Resistance between Intrinsically Safe Ground and earth ground must be less than 1.0
Ohm.
x In case of using cable glands in Outdoor location, they shall be UV rated or made of metal.
WARNING
- Substitution of components may impair Intrinsic Safety
- To prevent ignition of flammable or combustible atmospheres, disconnect power before servicing or read, understand and adhere to the manufacturer’s live maintenance procedures.
General Specifications 2-14
x The cable used to interconnect the devices needs to comply with the following parameters:
Loop resistance R’: 15 … 150 /km; Inductance per unit length L’: 0,4 … 1 mH/km Capacitance per unit length C’: 80 … 200 nF/km (C’ = C’ line/line + 0,5 C’ line/screen if both line are floating) (C’ = C’ line/line + C’ line/screen if the screen is connected to one line) Length of spur cable: max. 30 m Length of trunk cable: max. 1 km Length of splice : max. 1 m
WARNING
- Substitution of components may impair Intrinsic Safety
- To prevent ignition of flammable or combustible atmospheres, disconnect power before servicing or read, understand and adhere to the manufacturer’s live maintenance procedures.
Application Doc. No.: IKE031-A10 P.5 to P.6
2-15 General Specifications

2-12. Control Drawing of DO202S FF/PB Specification (FM Intrinsically safe FISCO)

FM Approved FISCO barrier
Voc (Vt) d17,5 V Ioc (It) d380 mA Poc (Pt) d5,32 W
FM Class I, DIV. 1, Group ABCD T4 for ambient temp. d 55 qC
DO202S-F
or DO202S-P
-
+
FM Approved Terminator
-
R = 90..100 C = 0..2,2 F
-
+
Transmitter
+
Sensor Connections Max. cablelength: 60 mtr. Cable dia. : 3…12 mm.
Sensor
Connections
-
+
Transmitter
FM Approved Terminator R = 90..100 C = 0..2,2 F
Division 1
Unclassified Location
Classified Location
x Sensor(s) are of a passive type to be regarded as 'simple apparatus', devices which
neither store nor generate voltages over 1.5 V, currents over 0.1 A, power over 25 mW or energy over 20 PJ, or are FM Approvals entity approved and meet connection requirements.
x Electrical data of the DO202S-F & DO202S-P:
- Supply circuit: Vmax = 17.5 V; Imax = 380 mA; Pi = 5.32 W; Ci = 220 pF; Li = 0 H.
- Sensor input circuit: Vt = 14.4 V; It = 10 mA; Ca = 643 nF; La = 340 mH
x Any FM Approved FISCO barrier may be used that meets the following requirements:
Voc or Vt d 17,5 V; Ioc or It d 380 mA; Poc or Pt d 5,32 W When installing this equipment, follow the manufacturer’s installation drawing. Installation should be in accordance with ANSI/ISA RP 12.06.01 “Installation of Intrinsically Safe Systems for Hazardous (Classified) Locations” and the National Electrical Code (ANSI/NFPA 70). Associated apparatus connected to the FISCO barrier must not use or generate more than 250 Vrms or Vdc.
x Resistance between FISCO Intrinsically Safe Ground and earth ground must be less than
1.0 Ohm.
x In case of using cable glands in Outdoor location, they shall be UV rated or made of metal.
x The FISCO concept allows the interconnection of several I.S. apparatus not specifically
examined in such combination. The criterion for such interconnection is that the voltage (Vmax), the current (Imax) and the power (Pi) which I.S. apparatus can receive and remain intrinsically safe, considering faults, must be equal to or greater that the voltage (Voc, Vt), the current (Ioc, It) and the power (Poc, Pt) which can be provided by the FM approved FISCO barrier. In addition, the maximum unprotected residual capacitance (Ci) and inductance (Li) of each apparatus (other than the terminator) connected to the Fieldbus must be less than or equal to 5nF and 10 H respectively.
General Specifications 2-16
x In each I.S. Fieldbus segment only one active source, normally the FM Approved FISCO
barrier, is allowed to provide the necessary power for the Fieldbus system. All other equipment connected to the bus cable has to be passive (not providing energy to the system), except to a leakage current of 50A for each connected device. Separately powered equipment needs a galvanic isolation to insure that the I.S. Fieldbus circuit remains passive.
x The cable used to interconnect the devices needs to comply with the following parameters:
Loop resistance R’: 15 … 150 /km; Inductance per unit length L’: 0,4 … 1 mH/km Capacitance per unit length C’: 80 … 200 nF/km (C’ = C’ line/line + 0,5 C’ line/screen if both line are floating) (C’ = C’ line/line + C’ line/screen if the screen is connected to one line) Length of spur cable: max. 30 m Length of trunk cable: max. 1 km Length of splice : max. 1 m
WARNING
- Substitution of components may impair Intrinsic Safety
- To prevent ignition of flammable or combustible atmospheres, disconnect power before servicing or read, understand and adhere to the manufacturer’s live maintenance procedures.
Application Doc. No.: IKE031-A10 P.7 to P.8
p
2-17 General Specifications

2-13. Control Drawing of DO202S FF/PB Specification (FM Non-incendive Entity)

FM Class I, DIV. 2, Group ABCD
FM Approved
Power Supply
Voc d 32 VDC
T4 for ambient temp. d 55 qC T6 for ambient tem
DO202S-B
or DO202S-D
+
FM Approved Terminator
+
-
R = 90..100 C = 0..2,2 F
-
. d 40 qC
-
+
Transmitter
Sensor Connections Max. cablelength: 60 mtr. Cable dia.: 3…12 mm.
Sensor
Connections
-
+
Transmitter
FM Approved Terminator R = 90..100 C = 0..2,2 F
Division 2
Unclassified Location
x
Sensor(s) are of a passive type to be regarded as 'simple apparatus', devices which
Classified Location
neither store nor generate voltages over 1.5 V, currents over 0.1 A, power over 25 mW or energy over 20 PJ, or are FM Approvals entity approved and meet connection requirements.
x Electrical data of the DO202S-B & DO202S-D:
- Supply circuit: Vmax = 32 V; Pi = 1.2 W; Ci = 220 pF; Li = 0 H
- Sensor input circuit: Vt = 14.4 V; It = 10 mA; Ca = 2.29 PF; La = 600 mH
When installing this equipment, follow the manufacturers installation drawing. Installation shall be in accordance with Article 501.4(B) of the National Electrical Code (ANSI/NFPA 79). Nonincendive field wiring may be installed in accordance with Article
501.4(B)(3)
x Grounding shall be in accordance with Article 250 of the National Electrical code. x In case of using cable glands in Outdoor location, they shall be UV rated or made of metal.
WARNING
- Substitution of components may impair suitability for Division 2.
- Do not remove or replace while circuit is live unless area is know to be non-hazardous
- Explosion Hazard – Do not disconnect equipment unless area is know to be non-hazardous
- Do not reset circuit breaker unless power has been removed from the equipment or the area is know to be non-hazardous
Application Doc. No.: IKE031-A10 P.9
General Specifications 2-18
p

2-14. Control Drawing of DO202S FF/PB Specification (FM Non-incendive FNICO)

FM Class I, DIV. 2, Group ABCD
FM Approved
Power Supply
Voc d 32 VDC
+
T4 for ambient temp. d 55 q C T6 for ambient tem
DO202S-B
or DO202S-D
+
FM Approved Terminator R = 90..100
-
C = 0..2,2 F
-
. d 40 qC
-
+
Transmitter
Sensor Connections Max. cablelength: 60 mtr. Cable dia.: 3…12 mm.
Sensor
Connections
-
+
Transmitter
FM Approved Terminator R = 90..100 C = 0..2,2 F
Division 2
Unclassified Location
x
Sensor(s) are of a passive type to be regarded as 'simple apparatus', devices which
Classified Location
neither store nor generate voltages over 1.5 V, currents over 0.1 A, power over 25 mW or energy over 20 PJ, or are FM Approvals entity approved and meet connection requirements.
x Electrical data of the DO202S-B & DO202S-D:
- Supply circuit: Vmax = 32 V; Pi = 5.32 W; Ci = 220 pF; Li = 0 H
- Sensor input circuit: Vt = 14.4 V; It = 10 mA; Ca = 2.29 PF; La = 600 mH
When installing this equipment, follow the manufacturers installation drawing. Installation shall be in accordance with Article 501.4(B) of the National Electrical Code (ANSI/NFPA 79). Non-incendive field wiring may be installed in accordance with Article 501.4(B)(3)
x Grounding shall be in accordance with Article 250 of the National Electrical code. x In case of using cable glands in Outdoor location, they shall be UV rated or made of metal.
WARNING
- Substitution of components may impair suitability for Division 2.
- Do not remove or replace while circuit is live unless area is know to be non-hazardous
- Explosion Hazard – Do not disconnect equipment unless area is know to be non-hazardous
- Do not reset circuit breaker unless power has been removed from the equipment or the area is know to be non-hazardous
Application Doc. No.: IKE031-A10 P.10

2-15. Control Drawing of DO202S FF/PB Specification (CSA)

CSA Ex ia Class I, DIV. 1, Group ABCD T4 for ambient temp. d 55 qC
Ui = 24 V or Ui = 17,5 V Ii = 250 mA Ii = 380 mA Pi = 1,2 W Pi = 5,32 W
DO202S-F
or DO202S-P
-
+
2-19 General Specifications
Sensor
Connections
Safe area
Apparatus
+
-
Safe area
I.S.
interface
I.S.
certified
Terminator
+
Transmitter
Zone 0 or 1
-
Transmitter
-
+
I.S.
certified
Terminator
Hazardous area
x Sensor(s) are a thermocouple, RTD's, passive resistive switch devices, or is CSA entity
approved and meet connection requirements.
x Electrical data of the DO202S-F & DO202S-P:
- Supply and output circuit:
Maximum input voltage Ui = 24 V Maximum input current Ii = 250 mA
Maximum input power Pi = 1.2 W Effective internal capacitance Ci = 220 pF; Effective internal inductance Li = 0 H. or FISCO field device
Maximum input voltage Ui = 17.5 V Maximum input current Ii = 380 mA Maximum input power Pi = 5.32 W Effective internal capacitance Ci = 220 pF; Effective internal inductance Li = 0 H.
- Sensor input circuit:
Maximum output voltage Uo = 14.4 V; Maximum output current Io = 22 mA Maximum allowed external capacitance Co = 643 nF Maximum allowed external inductance Lo = 70 mH
x Any CSA approved I.S. interface may be used that meets the following requirements:
Uo d 24 V
Io d 250 mA
Po d 1.2 W
Co t 220 pF + Ccable; Lo t 0 H + Lcable
or
FISCO power supply
Uo d 17.5 V
Io d 380 mA
Po d 5.32 W
Co t 220 pF + Ccable; Lo t 0 H + Lcable
Installation should be in accordance with Canadian Electrical Code, Part I or CEC, Part I. Maximum safe area voltage should not exceed 250 Vrms.
x Electrical data of the DO202S-B & DO202S-D (non-incendive):
For Class I, Div.2, Group ABCD the CSA approved I.S. interface is not required, and the sensor input circuit is non-incendive having the parameters:
Maximum output voltage Uo = 14.4 V; Maximum output current Io = 22 mA
Maximum allowed external capacitance Co = 3.5 PF
Maximum allowed external inductance Lo = 160 mH
General Specifications 2-20

3. INSTALLATION AND WIRING

A

3-1. Installation and dimensions

3-1 Installation and Wiring

3-1-1. Installation site

The EXA transmitter is weatherproof and can be installed inside or outside. It should, however, be installed as
The ambient temperature and humidity of the instal­lation environment must be within the limits of the
instrument specifications. (See chapter 2). close as possible to the sensor to avoid long cable runs between sensor and transmitter. In any case, the cable length should not exceed 30 mtr (100 feet). Select an installation site where:
When the instrument with Suffix Code "-B,-N,-D" is
used, take measures so that the display window is
not exposed to direct sunlight.
• Mechanical vibrations and shocks are negligible
• No relay/power switches or RF transmitters are in
the direct environment
• Access is possible to the cable glands (figure 3-1)

3-1-2. Mounting methods

Refer to figures 3-2 and 3-3. Note that the EXA
transmitter
• Panel mounting using optinal bracket, refer to Fig.
• The transmitter is not mounted in direct sunlight or
severe weather conditions
• Panel mounting using two (2) self-tapping screws,
• Maintenance procedures are possible (avoiding
corrosive environments)
• Surface mounting on a plate
• Wall mounting on a bracket
• Pipe mounting using a bracket on a horizontal or
(7.95)
202
(6.4) (6.1)
180
(7)
162
130
(5.1)
50
(2)
155
30
(1.2)
has universal mounting capabilities:
3-2a.
refer to Fig. 3-2b.
(using bolts from the back)
(for example, on a solid wall)
vertical pipe (Nominal pipe diameter 50 A)
Panel thickness
1 to 10
(0.04 to 0.39)
(1.2)
30
60
(2.36)
Unit: mm (inch)
Panel mounting bracket
Hood (Option)
115
Grounding terminal (M4 screw)
9
(0.35)
42
(1.65)
34
(1.3)
56±0.2
(2.20)
Option code : / H
(2.7)
68
Sensor cable inlet cable gland (Pg13.5)
M6 screw (32 )
Transmission signal cable inlet Cable gland : Pg13.5
dapter
G1/2 screw (/AFTG) 1/2 NPT screw (/ANSI)
Figure 3-1. Housing dimensions and layout of glands
77
(3)
(4.5)
(1.26) depth
1.eps
49
(1.93)
2.eps
Approx.
55
(2.2)
156
(6.14)
+1.1 0
185
205
80
3.9
(0.15)
38
(1.5)
(3.15)
PANEL CUTOUT
+1.1
173
0
138
(5.43)
(6.81)
121
(4.76)
Figure 3-2a. Panel mounting diagram
(7.28) (8.07)
5.eps
Installation and Wiring 3-2
Unit: mm (inch)
+1 0
+1
0
18.5
SPACING PANEL CUTOUT
(0.72)
PANEL CUTOUT
Fig. 3-2b. Panel mounting using two (2) self-tapping screws
56
(2.20)
2-Ø6.5
(0.26)
200
(7.87)
4-Ø10
(0.4)
(3)
77
115
70
(2.75)
(4.5)
Figure 3-3. Wall and pipe mounting diagram
3.5 (0.14)
Pipe mounting
(Vertical)
Pipe mounting
(Horizontal)
Nominal 50 A (O.D. Ø60.5 mm)
(2 inch pipe)
4.eps
Figure 3-4. Internal view of EXA wiring compartment
3-3 Installation and Wiring

3-2. Wiring of power supply

3-2-1. General precautions

mA
Do not activate the power supply yet. First make sure that the DC-power supply is according to the specifications given.
WARNINGWARNING
DO NOT USE ALTERNATING CURRENT OR MAINS POWER SUPPLY! !
The cable leading to the distributor (power supply) or safety barrier transports power to and output signal from the transmitter. Use a two conductor shielded cable with a size of at least 1.25 mm diameter of 6 to 12 mm. The cable gland supplied with the instrument accepts these diameters. The maximum length of the cable is 2000 metre, or 1500 metres when using the communications. This ensures the minimum operating voltage for the instrument.
Grounding:
• If the transmitter is mounted on a grounded
surface (e.g. a metal frame fixed in the soil) the shield of the 2-wire cable may NOT be connected to ground at the distributor.
• If the transmitter is mounted on a non-conducting
surface (e.g. a brick wall) it is recommended to ground the shield of the 2-wire cable at the distributor end.
mA

3-2-2. Connection of the power supply

The terminal strip is accessed as was described in section 3-2-1. Use the left-hand gland to insert the supply/output cable to the transmitter. Connect the supply to the terminals marked +, - and G as is indi­cated in figures 3-7.

3-2-3. Switching the instrument on

After all connections are made and checked, the power can be switched on from the distributor. Observe the correct activation of the instrument at the display. If for any reason the display does not indicate a value, consult the trouble shooting section.
mA

3-2-4. Preparation

Refer to figure 3-4. The power/output connections and the sensor connections should be made. Refer to figure 3-7. The terminals are of a plug in style for ease of mounting.
To open the EXA 202 for wiring:
1. Loosen the four frontplate screws and remove
the cover.
2. The terminal strip is now visible.
3. Connect the power supply. Use the gland on the
left for this cable.
4. Connect the sensor input, using the gland on
the right (see fig. 3-5). Switch on the power.
2
and an outside
Commission the instrument as required or use the default settings.
5. Replace the cover and secure frontplate with the four screws.
6. Connect the grounding terminals to protective earth.
7.
The optional hose connection is used to guide the cables coming from an immersion fitting through a protective plastic tubing to the transmitter.

3-2-5. Cables, terminals and glands

The DO202 is equipped with terminals suitable for the connection of finished cables in the size range:
0.13 to 2.5 mm (26 to 14 AWG). The glands will
Sensor cable gland
Power/Output cable gland
Grounding terminal
Figure 3-5. Glands to be used for cabling
form a tight seal on cables with an outside diameter in the range of 6 to 12 mm (0.24 to 0.47 inches).

3-3. Wiring of sensors

3-3-1. Wiring the standard galvanic sensor

The sensor cable has markers on the individual wires. These markings refer to the markers on the terminals.
The temperature sensor has two wires with the markings T1, T2 and/or 11, 12 and must be con­nected to terminal 11 and 12.
The measuring electrode: the cathode is marked with IE and/or 13 and must be connected to terminal 13. The reference electrode: the anode is marked with RE and/or 15 and must be connected to terminal 15. The liquid earth (solution ground) is marked with 16 and must be connected to terminal 16. The overall shield of the cable is marked with 14 and must be connected to terminal 14. Place the jumper in the GALVANIC position.
Installation and Wiring 3-4

3-3-2. Wiring other galvanic sensors

Consult the users manual for the color identification of the sensor cable.
- The temperature sensor has two wires and must be connected to the terminal 11 and 12.
- The measuring electrode: the cathode must be connected to terminal 13.
- The reference electrode: the anode must be connected to terminal 15.
- The overall shield of the cable must be connected to terminal 14 if there is one available.
- Place the jumper in the GALVANIC position.

3-3-3. Wiring polarographic sensors

Consult the owners manual for the color identifica­tion of the sensor cable.
- The temperature sensor has two wires and must be connected to the terminal 11 and 12.
- The measuring electrode: the cathode must be connected to terminal 17.
- The reference electrode: the anode must be connected to terminal 18.
- The overall shield of the cable must be connected to terminal 14 if there is one available.
- Place the jumper in the POLAROGRAPHIC position.

3-3-4. Wiring sensors with Vp connector

Consult the owners manual for the color identifica­tion of the sensor cable
- The temperature sensor has two wires and must be connected to terminal 11 and 12
- Polarographic : the measuring electrode (cathode) must be connected to terminal 17
- Polarographic : the reference electrode (anode) must be connected to terminal 18
- Galvanic : the measuring electrode (cathode) must be connected to terminal 13
- Galvanic : the reference electrode (anode) must be connected to terminal 15
- The overall shield of the cable must be connected to 14
- Place the jumper in the Polarographic or Galvanic position
A-17 pol. 13 galv. B-18 pol. 15 galv. C D E-11 F-12 S-14
temp
Figure 3-7. Terminal indentification labels
3-3-5. Sensor cable connections using junction
box (BA10) and extension cable (WF10)
Where a convenient installation is not possible using the standard cables between sensors and trans­mitter, a junction box and extension cable may be used. The Yokogawa BA10 junction box and the WF10 extension cable should be used. These items are manufactured to a very high standard and are necessary to ensure that the specifications of the system are not compromised. The total cable length
17
1314141615
12
11
11 Re d
12 Blue
15
14 16
14
17
13
12
11
15 Core 16 Screen White Co-axial cable
14 Overall Screen
13 Core 17 Screen Brown Co-axial Cable
WF10 Cable
Figure 3-6. Connections DO
should not exceed 30 metres (including DO sensor cable).
TRANSMITTER / CONVERTER
11
12
17
13
15
16
14
Brown
Screen
Red
C
E
Overall shield
A
D
Blue
B
White
Figure 3-8. Connection of WF10 extension cable and BA10/BP10 junction box
3-5 Installation and Wiring
Extension cable may be purchased in bulk quanti­ties, cut to length. Then it is necessary to terminate the cable as shown below.
Termination procedure for WF10 cable.
1. Slide 3 cm of heat shrink tube (9 x 1.5) over the cable end to be terminated. Strip 9 cm of the outer (black) insulating material,
2. taking care not to cut or damage internal cores.
Figure 3-9a.
3 cm
heat shrink
9 cm
remove insulation
3. Remove loose copper screening, and cut off the cotton packing threads as short as possible.
4. Strip insulation from the last 3 cm of the brown, and the white coaxial cores.
3-3-6. Additional precautions for installations in
hazardous areas
Make sure that the total of capacitance and induct­ances connected to the input terminals of the EXA DO202S do not exceed the limits given in the certifi­cate. This sets a limit to the cable and extensions used. – The intrinsic safe version of the DO202S
instrument can be mounted in Zone 1 (DO202S-B,
-N, -D).
– The sensors can be installed in Zone 0 or Zone 1
if a safety barrier according to the limits given in the system certificate is used.
– Ensure that the total of capacitance and
inductances connected to the terminals of the EXA DO202S do not exceed the limits given in the certificate of the safety barrier or distributor.
– The cable used should preferably have a BLUE
colour or marking on the outside. – Installation for (sensors in Zone 0 or 1): Generally, the distributor with input/output isolation
has no external earth connection. If there is
an earth connection on the distributor and the
external connection of the transmitter is connected
to “protective” earth, the shield of the 2-wire cable
may NOT be connected to “protective” earth at the
distributor too.
3 cm
cotton threads
Figure 3-9b.
5. Extract the coaxial cores from the braid, and trim off the black (low-noise) screening material as short as possible. Insulate the overall screen and drain wire (14) and
6. the 2 coaxial screens with suitable plastic tubing.
7. Strip and terminate all ends with suitable (crimp) terminals and identify with numbers as shown.
Red
11
Blue
12
Black
14
White
15
16
Brown
13
3-3-7. Installation in:
Hazardous Area-Non-Incendive
The EXA DO202S-B, -N, -D may be installed in a Category 3, Zone 2, Div.2 area without the use of safety barriers. Maximum permissible supply voltage
31.5V.
Figure 3-9c.
8. Finally shrink the overall heat shrink tube into position.

4. OPERATION; DISPLAY FUNCTIONS AND SETTING

Operation: Display Functions and Settings 4-1

4-1. Operator interface

This section provides a survey of the operation of the EXA operator interface. The basic procedures for obtaining access to the three levels of operation are described briefly. For a step-by-step guide to data entry, refer to the relevant section of this user’s manu­al. Figure 4-1 shows the EXA operator interface.
LEVEL 1: Maintenance
These functions are accessible by pushbutton through a flexible front cover window. The functions make up the normal day-to-day operations that an operator may be required to complete. Adjustment of the display and routine calibration are among the features accessible in this way. (See table 4-1).
Table 4-1. Operations overview
Routine Function Chapter
Maintenance
Commissioning
mA
CALIB
DISP M.TEMP M.PRESS
OUTP HOLD
AIR.CAL H2O.CAL SENS ZERO
LEVEL 2: Commissioning
A second menu is exposed when the EXA front cover is removed and the display board is revealed. Users gain access to this menu by pressing the but­ton marked * in the lower right of the display board. This menu gives access to the service menu. (See table 4-1).
LEVEL 3: Service
For more advanced configuration selections, press the button marked *. Now push the “YES” button. Selecting and entering “Service Code” numbers in the commissioning menu provide access to the more advanced functions. An explanation of the Service Codes is listed in chapter 5 and an over­view table is shown in chapter 11.
Calibration in air (preferred method) Calibration in aerated water Manual sensitivity calibration Manual zero calibration
Read auxiliary data or set message display Adjust manual temperature Adjust manual pressure
Adjust the output range Activate the hold function
6 6 6 6
4 5 5
5 5
Service (access to coded entries from the commissioning level)
SERV Fine tune the specialized functions of the transmitter 5
Note: All three levels may be separately protected by a password. See Service Code 52 in chapter 5
Service Code table for details on setting passwords.
4-2 Operation: Display Functions and Settings
Main display Menu pointer flags
Message display
YES NO
Key prompt flags
Selection keys
YES : Accept setting
NO : Change setting
Adjustment keys
> : Choose digit to
adjust
^ : Adjust digit
ENT : Confirm change
Figure 4-1. DO202 operator interface
NO MODEYES
4-2. Explanation of operating keys
Fail flag
FAI LTEMP.MAN
ENT
MODE
Commissioning
function menu
Commissioning mode
access key
Measure/Maintenance
mode key
ENT
ppm ppb
%sat CAL DISPLAY HOLD
OUTPUT SET HOLD SERVICE
Broken line indicates area that can be
seen through front cover
cover removed or opened. Once this button has been used to initiate the commissioning
MODE key
This key toggles between the measuring and
menu, follow the prompts and use the other
keys as described above. maintenance modes. Press once to obtain access to the maintenance function menu.

4-3. Passcode protection

CALIB DISP Press again to return to the measuring mode.
In Service Code 52, passcode protection can be set for each one of the three operating levels, individu­ally. This procedure should be completed after the
YES/NO keys
These are used to select choices from the menu. YES is used to accept a menu selection.
initial commissioning (setup) of the instrument. The passcodes should then be recorded safely for future reference.
NO is used to reject a selection, or to move
ahead to the next option.
When passcodes have been set, the following addi­tional steps are introduced to the configuration and
DATA ENTRY keys (
)
programming operations:
Is used as a “cursor” key. Each press on this key moves the cursor or flashing digit one place to the right. This is used to select the digit to be changed when entering numerical data.
Maintenance
Press MODE key. The display shows 000 and *PASS* Enter a 3-digit passcode as set in Service Code 52 to obtain access to the Maintenance Mode.
Is used to change the value of a selected digit. Each press on this key increases the value by one unit. The value can not be decreased, so in order to obtain a lower value, increase past nine to zero, then
Commissioning
Press key. The display shows 000 and *PASS* Enter a 3-digit passcode as set in Service Code 52 to obtain access to the Commissioning Mode.
increase to the required number. When the required value has been set using the
and
keys, press to confirm the data entry. Please note that the EXA does not register any change of data until the ENT key is pressed.
Service
From the Commissioning menu, select *Service by pressing YES key. The display shows 000 and *PASS* Enter a 3-digit passcode as set in Service Code 52 to obtain access to the Service Mode.
This is the commissioning mode key. It is
key
used to obtain access to the commissio­ning menu. This can only be done with the
Note: See Service Code 52 for the setting of
passcodes.
Operation: Display Functions and Settings 4-3

4-4. Display examples

The following pages show the sequence of button presses and screens displayed when working in some standard configurations. More or less options will be made available by the configuration of some

4-5. Display function (default)

MODE
FAILTEMP.MAN
ppm ppb %sat
ENT
ENT
CAL DISPLAY HOLD
OUTPUT SET HOLD SERVICE
YES NO
NO MODEYES
service codes, or by choices made in the commis­sioning menu.
The following deviations are possible:
Item marked is omitted when switched off in
*
commissioning mode.
Cell current
YES
NO
Sensitivity
(slope)
YES
NO
Zero current
YES
DISPLAY
NO
DISPLAY
NO
DISPLAY
NO
MODE
YES NO
NO
YES NO
NO
YES
NO
NO
Software release
PPM PPB %SAT CAL DISPLAY HOLD
number
YES
NO
DISPLAY
NO
Process
YES
CAL (SEE CALIBRATION CHAPTER 6)
temperature
YES
YES
NO
NO
NO
DISPLAY
Output
YES
DISPLAY
mA
YES
NO
DISPLAY
NO
% Saturation
YES
SEE MANUAL
NO
TEMPERATURE (CHAPTER 5)
YES
NO
DISPLAY
NO
YES
NO
Unit
NO
YES
SEE MANUAL PRESSURE (CHAPTER 5)
YES
NO
DISPLAY
NO
Process
pressure
YES
DISPLAY
NO

5. PARAMETER SETTING

5-1 Parameter Setting

5-1. Maintenance mode

Standard operation of the EXA instrument involves use of the Maintenance (or operating) mode to set up some of the parameters.
Access to the Maintenance mode is available via the six keys that can be pressed through the flexible window in the instrument cover. Press the MODE key once to enter this dialog mode.
Note: At this stage the user will be prompted for
pass code where this has been previously set up in service code 52 in chapter 5.

5-1-1. Manual activation of hold

mA
Calibration Air See “calibration” section 6.
H
2O See “calibration” section 6.
Sensitivity See “calibration” section 6.
Zero See “calibration” section 6.
mA
Hold Manually switch on/off
“hold” (when enabled in Commissioning menu). See adjustment procedure 5-2-2.
Display See “operation” section 4.
Manual temperature See “manual temperature
adjustment” section 5-1-2.
Manual pressure See “manual pressure
adjustment” section 5-1-3.
YES NO
NO MODEYES
MODE
YES NO
NO
NO
FAILTEMP.MAN
ENT
ENT
CALIBRATE
MODE
ppm ppb %sat CAL DISPLAY HOLD
OUTPUT SET HOLD SERVICE
NO
NO
HOLD
HOLD
NO
YES N O
YES
YES
YES NO YES NO
YES
MEASURE
Parameter Setting 5-2
%SAT
PPB
CAL DISPLAY
PPM
HOLD
MODE
NO 2x
YES
YES NO
TEMP.MAN
ENT
>
Set using >, ,ENT keys.
Manual temperature
5-1-2. Manual temperature adjustment
5-1-3. Manual pressure adjustment
PPM PPB %SAT CAL DISPLAY HOLD
MODE
NO
3x
YES NO
YES
% sat process (process pressure)
Set using >, ,ENT keys.
>
YES
ENT
Calibration pressure Set using >, ,ENT keys.
>
ENT
5-3 Parameter Setting

5-2. Commissioning mode

In order to obtain peak performance from the EXA transmitter, you must set it up for each custom application. Refer to APPENDIX A-1-2 for stand­ard configurations and options.
Output range: mA output is set as default
mA
to 0 - 20ppm. For enhanced resolution in more stable measuring processes, it may be desirable to select for example 5-10 ppm range.

5-2-1 Output range

mA
FAILTEMP.MAN
YES NO
ENT
MODE
ppm ppb %sat CAL DISPLAY HOLD
mA
Hold The EXA DO202 transmitter has the
ability to “HOLD” the output during maintenance periods. This parameter should be set up to hold the last measured value, or a fixed value to suit the process.
Service This selection provides access to
the service menu.
What follows are pictorial descriptions of typical frontplate pushbutton sequences for each param­eter setting function. By following the simple YES/ NO prompts and arrow keys, users can navigate through the process of setting range, setpoints, hold, wash and service functions.
ENT
OUTPUT SET HOLD SERVICE
YES
NO
YES
NO
NO MODEYES
ENT
ENT
Note: When ppb or %sat
is enabled in code 56, the output range is set in a similar way to ppm.
YES
YES
NO
YES
NO
ENT
NO
ENT
ENT
ENT
ENT
YES
NO
NO
ENT
ENT
Parameter Setting 5-4

5-2-2. HOLD

mA
MODE
FAILTEMP.MAN
ppm ppb %sat
YES NO
NO MODEYES
CAL
OUTPUT
DISPLAY
ENT
ENT
HOLD
SET HOLD SERVICE
NO
YES NO
NO
YES NO
NO
YES NO
NO
YES
YES
HOLD
NO
YES
YES NO
YES NO
YES NO
HOLD active last measured value.
NO
HOLD
YES
HOLD
HOLD
HOLD
ENT
ENT
ENT
ENT
Set HOLD "fixed value"
YES NO

5-2-3. SERVICE

5-5 Parameter Setting
MODE
FAILTEMP.MAN
ppm ppb %sat
YES NO
NO MODEYES
CAL
OUTPUT
DISPLAY
ENT
ENT
HOLD
SET HOLD SERVICE
mA
mA
NO
NO
NO
Example: Service Code 01
Select Sensor Type
YES
NO
YES
NO
Polarization voltage
and sensitivity.
Set using >, ,ENT keys.
>
ENT
ENT
ENT
YES
NO
ENT
ENT
for galvanic sensor
for polarographic
With the >, ,ENT keys
ENT
ENT
>
ENT
ENT
YES
ENT
ENT
Parameter Setting 5-6

5-3. Notes for guidance in the use of service coded settings

Don't set or input service code numbers other than the code numbers defined in this manual. Setting an undefined service code may make the transmitter malfunction. When an undefined service code is input by some accident, push the MODE key and escape from the serv­ice level.

5-3-1. Parameter specific functions

Code 1 *S.TYPE Select 0 for DO30G sensor or any other galvanic sensor. Select 1 for Oxyferm,
Oxygold and other polarographic sensors.
*V.POL *SENS Set the sensitivity of the connected sensor.
Code 4 *SAL.TY Enable or disable compensation. This function is particularly useful when
*[Cl-] Set level for salinity compensation.

5-3-2. Temperature functions

Code 10 *T.SENS Select temperature sensor to suit the measuring probe. The default selection is
Code 11 *T.UNIT Select °C or °F for temperature display.
Code 12 *T.ADJ Adjust offset in temperature measurement when the sensor is at a stable known
Code 13 *T.MAN Enable and set manual temperature compensation.
Set the polarization voltage of the connected sensor (only for polarographic sensors).
measuring in estuarine or brackish waters.
the 22k NTC sensor. Refer to the instructions delivered with the DO sensor to check which temperature sensor is used.
value.

5-3-3. Calibration functions

Code 20 *T.SEC Set stability parameters for calibration. A greater time, and/or a smaller step gives *Z.PPM greater stability before a calibration can be accepted. Use caution, however, to avoid *Z.PPB very long calibration times. *Z.SAT *S.PPM
Code 21 *Z.CAL Normally not necessary, zero calibration can be enabled when required. *Z.LIM Zero check limit.
Code 22 *ZERO Calibration values can be entered from independent (laboratory) data. *SENS The sensitivity is the increase of the cell current per ppm and is specified in nA/
ppm at 25˚C (77˚F).
Sensor nominal sensitivity (nA/ppm at 25˚C) DO30(G) 450 OXYFERM 7.5 OXYGOLD 35
5-7 Parameter Setting
Code Display Function Function detail X Y Z Default values
Parameter Specific functions
01 *S.TYPE Sensor type Galvanic 0 1
Polarographic 1
*V.POL (Polarographic) 0.675 V
Polarization Voltage
*SENS Sensor sensitivity Polarographic 7.5 nA/ppm
Galvanic
04 *SAL.TY Salinity comp. Compensation disabled 0 0
Compensation enabled 1
*[Cl-] NaCl conc. Concentration in g/kg water: 0.0
range 1 to 99.9 g/kg
05-09 Not used
Code Display Function Function detail X Y Z Default values
Temperature Measuring Functions
10 *T.SENS Temp. Comp. Pt1000 RTD (DO30G) 0 1
22k NTC (Oxyferm) 1
Pb36 = 2k2 NTC (DO30) 2
11 *T.UNIT Temp units Celsius (°C) 0 0
Fahrenheit (°F) 1
12 *T.ADJ Temp adjust Adjustment +/- 7.5 °C or +/- 13.5 °F
13 *T.MAN Manual TC Automatic temperature comp 0 0
Manual temp compensation 1 25 °C/ 77 °F
14-19 Not used
450 nA/ppm (*S.TYPE=0)
Code Display Function Function detail X Y Z Default values
Calibration Settings
20 *T.SEC Stabilization Stabilization time range 5- 600 s 60 s
Step DO step change range 0-50 ppm
Z.PPM 0-1999 ppb 0.1 ppm (Z.PPM)
Z.PPB 1 ppb (Z.PPB)
Z.SAT 1.0 % SAT (Z.SAT)
S.PPM 0-600 %SAT 0.1 ppm (S.PPM)
(
always PPM
)
21 *Z.CAL Zero calibration Zero calibration disabled 0 0
Zero Calibration enabled 1
*Z.LIM Zero check limit Zero check limit (galv.)
0-19.99 A 19.99 A
Zero check limit (polar.)
0-199.9 nA 199.9 nA
22 *ZERO Zero current Offset range, 0 nA
±Zero Check Limit
*SENS Sensitivity entry Sensitivity range 1-1999 nA/ppm 7.5
450
nA
/ppm
nA
/ppm (*S.TYPE=0)
23-29 Not used
Parameter Setting 5-8
mA

5-3-4. mA output functions

Code 31 *OUTP.F For the DO202 the output may be chosen as linear to input, or configured in a
Code 32 *BURN Diagnostic error messages can signal a problem by sending the output signals
Note: Burnout downscale can only be selected if the HART communication is turned of in service code 60.
Code 35 *TABLE The table function allows the configuration of an output curve by 21 steps

5-3-5. User interface

21 point table to a particular linearization. Enable the table setup in code 31, and configure the table in code 35.
upscale (21 mA) or downscale (3.6 mA when HART comm. non-used, 3.9 mA when HART comm. used). This is called upscale or downscale burnout, from the analogy with thermocouple failure signaling of a burned-out or open circuit sensor. The pulse burnout setting gives a 21 mA signal for the first 30 seconds of an alarm condition. After the “pulse” the signal returns to normal. This allows a latching alarm unit to record the error. In the case of the EXA the diagnostics are extensive and cover the whole range of possible sensor faults.
(intervals of 5%). The following example shows how the table may be configured to linearize the output with a mA curve.
Code 50 *RET When Auto return is enabled, the transmitter reverts to the measuring mode from
anywhere in the configuration menus, when no button is pressed during the set time interval of 10 minutes.
Code 51 *P.MAN In Maintenance mode the pressure can be entered (M. PRESS). *P.% *P.CAL
Code 52 *PASS Passcodes can be set on any or all of the access levels, to restrict access to the
instrument configuration.
Code 53 *Err. xx Error message configuration. Two different types of failure mode can be set.
Hard fail gives a steady FAIL flag in the display.
Soft fail gives a flashing FAIL flag in the display. The call for maintenance is
a good example of where a SOFT fail is useful. A warning that the regular maintenance is due, should not be used to shut down the whole measurement.
Code 55 *CALL.M Call for maintenance is a contact trigger to signal that the system has been in
service for longer than the set time without calibration. The user can set up to 250 days as a routine service interval.
Code 56 *UNIT Select ppm, ppb or % saturation as the main unit, to suit the application.
5-9 Parameter Setting
mA
Code Display Function Function detail X Y Z Default values
Output Functions
30 Not used
31 *OUTP.F mA output functions Linear 0 0 Linear
Table 1
32 *BURN Burn function No burnout 0 0 No Burn.
Burnout downscale 1
Burnout upscale 2
Pulse burnout 3
33, 34 Not used
35 *TABLE Output table for mA
*0% Linearization table for mA in 5% steps.
*5% The measured value is set in the main
*10% display using the >, ^, ENT keys, for
... each of the 5% interval steps.
... Where a value is not known, that value may
*95% be skipped, and a linear interpolation will
*100% take place.
36-39 Not used
Code Display Function Function detail X Y Z Default values
User Interface
50 *RET Auto return No return to meas. from HOLD 0 1
Return to meas after 10 min 1
51 *P.MAN Manual pressure No manual pressure in Maint.mode 0 0
Manual pressure in Maint.mode 1
*P.% Enter Manual process pressure 0-999.9 kPa 101.3 kPa
*P.CAL Enter Cal. pressure in Maint.mode 0-999.9 kPa 101.3 kPa
52 *PASS Password Direct access to MAINT 0 0.0.0
Pass: 111 for #=1, MAINT protected by password #
333 for #=2, Direct access to COMM 0
777 for #=3, COMM protected by password #
888 for #=4, Direct access to SERV 0
123 for #=5, SERV protected by password #
957 for #=6,
331 for #=7,
546 for #=8 and
847 for #=9
53 *Err.01 Error setting Calibration not stable Soft fail 0
Calibration not stable Hard fail 1 1 Hard
*Err.02 Zero out of limits Soft fail 0
Zero out of limits Hard fail 1 1 Hard
*Err.07 Temperature sensor open Soft fail 0
Temperature sensor open Hard fail 1 1 Hard
*Err.08 Temp. sensor shorted Soft fail 0
Temp. sensor shorted Hard fail 1 1 Hard
*Err.09 Cell current abnormal Soft fail 0
Cell current abnormal Hard fail 1 1 Hard
*Err.16 Call for maintenance
Interval time exceeded
Call for maintenance Hard fail 1
Interval time exceeded
55 *CALL.M E16 setting No call for maintenance 0 0
Call for maint. enabled 1
*CALL .M E16 message after 1- 250 days 250
56 *UNIT PV unit Unit of measurement ppm 0 0
Unit of measurement ppb 1
Unit of measurement % sat. 2
57-59 Not used
Soft fail 0 0 Soft
Parameter Setting 5-10

5-3-6. Communication setup

Code 60 *COMM. The settings should be adjusted to suit the communicating device connected
mA
to the output. The communication can be set to HART®.
*ADDR. Select address 00 for point to point communication with 4-20mA transmission.
Addresses 01 to 15 are used in multi-drop configuration (fixed 4mA output).
mA
Code 61 *HOUR The clock/calendar for the logbook is set for current date and time as reference. *MINUT *SECND *YEAR *MONTH *DAY
Code 62 *ERASE Erase logbook function to clear the recorded data for a fresh start. This may be
desirable when re-commissioning an instrument that has been out of service for a while.

5-3-7. General

Code 70 *LOAD The load defaults code allows the instrument to be returned to the default set up
with a single operation. This can be useful when wanting to change from one application to another.

5-3-8. Test and setup mode

Code 80 *TEST Not used
Note: attempting to change data in service code 80 and above without the proper instructions and
equipment, can result in corruption of the instrument setup, and will impair the performance of the unit.
5-11 Parameter Setting
Code Display Function Function detail X Y Z Default values
Communication
60 *COMM. Communication Set HART® communication Off 0 1.0 On
mA
Set HART® communication On 1
Communication write enable 0 Write
Communication write protect 1 enable
*ADDR. Network address Set address 00 to 15 00
61 *HOUR Clock setup Adjust to current date and time using
mA
*MINUT >, ^ and ENT keys
*SECND
*YEAR
*MONTH
*DAY
62 *ERASE Erase logbook Press YES to clear logbook data
63-69 Not used
Code Display Function Function detail X Y Z Default values
General
70 *LOAD Load defaults Reset configuration to default values
71-79 Not used
Code Display Function Function detail X Y Z Default values
Test and setup mode
80 *TEST Test and setup Not used

6. CALIBRATION PROCEDURE

Calibration Procedure 6-1
Calibration of the dissolved oxygen analyzer is per­formed in the following situations:
• When a new dissolved oxygen sensor is installed.
• When the membrane is replaced and/or the electrolyte solution is replaced
• When the sensor has been disassembled and reassembled for maintenance
• When the measuring error after cleaning exceeds the acceptable deviation from reference method
• When error message E16 indicates the need for maintenance

6-1. General

6-1-1. Calibration methods

There are four methods for calibration of the EXA dissolved oxygen analyzer: air calibration, water calibration, sensitivity and zero (offset) calibration.
Air calibration
This is the most common and easiest way to cali­brate the analyzing system. The calibration is done in AIR.CAL mode in maintenance level.
Water calibration
This is the traditional way to calibrate dissolved oxy­gen analyzers in laboratory environment. . Accurate results can be obtained if the calibration is per­formed carefully. The water calibration may include a zero calibration if so selected in service code 21. Normally only a zero check is performed without calibration, since a good sensor has no zero signal. Water calibration is not possible in salt water unless salinity compensation is activated in service code
04. The calibration is done in H2O.CAL mode in
Note: In normal industrial environments better
results are obtained with air calibration, which is therefore the recommended method.
Manual calibration
This is a calibration method whereby the sensor is not exposed to a calibration fluid, but is calibrated by comparison with a reference method. This refer­ence method is normally a laboratory analysis of the water sample: either by Iodometric titration according ISO 5813 or an electrochemical method according ISO 5814. The calibration is done in SENS/ZERO mode in maintenance level.
6-1-2. Diagnostic functions performed during
calibration
The calibration is a semi-automatic calibration, which means that the sensor output is used for calculation of sensor parameters after the read­ings have stabilized. The criteria for stabilization are set in service code 20. If stable readings are not achieved within one hour error E1 message will appear on the display and the procedure is aborted.
The calculated sensor parameters are compared with preset tolerances and if these conditions are not met, Error E2 (Zero Error) or E3 (Sensitivity Error) will be signalled.
Notes:
1. Error messages that appear initially indicate inadequate settings of the diagnostic variables in service codes 01, 20 and/or 22.
2. Patience improves the accuracy of calibration. When the EXA prompts the user that the preset criteria are met by showing CAL.END message it only means that the minimum criteria are met. When the signal still changes after this initial acceptance, EXA will automatically update the sensor characteristics with the new information.
6-2 Calibration Procedure
6-2. Calibration procedure using air calibration
method

6-2-1. Preparation

Move the sensor to a maintenance site and wash off any dirt on the membrane. Lightly wipe off any remaining water from the membrane with a soft tis­sue. Expose the sensor to the air in an environment where there is no temperature change and no wind

6-2-2. Procedure for air calibration

Press the MODE key. CALIB appears in the display, and the
MODE
FAILTEMP.MAN
ppm ppb %sat CAL
YES NO
ENT
NO MODEYES
ENT
YES
NO
DISPLAY HOLD
OUTPUT SET HOLD SERVICE
YES
from *1
The instruments waits for the reading to stabilise. (The display flashes) When reading Is stable, the CAL END message appears.
(a convenient way to accomplish this environment is to place the sensor in an empty bucket).
Note: If sensors are used without temperature
compensating element, the temperature must be adjusted to ambient temperature in service code 13.
12J6B2-22
YES/NO key flags flash. Press YES.
Note: To start calibration with a zero oxygen (N2 gas.)
NO
YES
NO
YES
ABLE
YES
NO
SC21=1
NO
NO
SC21=0
YES
YES
YES
NO
NO
NO
NO
NO
to *1
YES
NO
NO
YES
YES
NO
Press YES for single point adjustment.
YES
YES
After briefly displaying WAIT,
the display returns
Go to
YES
6-3-3
NO
normal readout.
to the
NO
YES NO
YES
YES
to *1
Go to
YES
6-4-2
NO
Sensor can be returned
to process water.
Go to
YES
6-4-3
NO
Calibration Procedure 6-3
6-3. Calibration procedure using water calibra-
tion method

6-3-1. Preparation

Move the sensor to a maintenance site and wash off any dirt on the membrane. Lightly wipe off any remaining water from the membrane with a soft tissue. Prepare the necessary equipment and reagents to be used for the span and (if required) zero calibration. Equipment for span calibration includes:
• Beaker or bucket
Magnetic stirrer or other means of agitating the water
• Demineralized water or salty process water if salinity compensation is activated
• Air supply pump (aquarium pump)
• Glass diffuser to generate small air bubbles
Note: It takes 15 to 30 minutes of aeration before
it can be assumed that the water is fully saturated with air.
Reagent for zero calibration:
• Sodium sulfite
• demineralized water
The zero oxygen solution contains 20- 30 grams of sodium sulfite per liter.
6-3-2. Calibration operation
(water calibration method)

6-3-3. Procedure for water calibration

MODE
FAILTEMP.MAN
ppm ppb %sat CAL
YES NO
NO MODEYES
YES
NO
OUTPUT
DISPLAY
ENT
SET HOLD
HOLD
SERVICE
ENT
YES
Press the MODE key. CALIB appears in the display, and the
YES
from *1
The instruments waits for the reading to stabilise. (The display flashes). When reading Is stable, the CAL END message appears.
If both zero and span calibration are used, it is bet­ter to start with span calibration in spite of normal calibration sequence with analog instrumentation. Otherwise the sensor must be rinsed off carefully to prevent ingress of oxygen absorbing sulfite in the aerated water. Note: Zero calibration in the sodium sulfite solution
takes a long time. Even after CAL.END prompt it is necessary to wait 40-50 minutes before zero adjustment can be made. Otherwise calibration errors are introduced in the measurement. It is recommended not to do a zero calibration, but a zero check only. A good clean sensor does not have a zero current and does not require zero calibration. If the sensor has a substantial zero current, the sensor has to be disassembled, cleaned and filled with fresh electrolyte solution and a new membrane has to be fitted.
Calibration may take longer than 10 minutes.
Before starting the calibration procedure, check the setting of Auto Return (*RET.) in Service Code 50. It is recommended to set the Auto Return to “0” (No return to meas. from HOLD). After the calibration procedure has been completed, set the Auto Return to the desired setting in Service Code 50.
YES/NO key prompt flags flash. Press YES.
NO
YES
YES
Note: To start calibration with a zero solution (zero-dissolved-oxygen solution.)
NO
SC21=1
YES
NO
NO
NO
SC21=0
YESYES
YES
NO
NO
Go to 6-4-2
Go to 6-4-3
to *1
After briefly displaying WAIT,
the display returns
Sensor can be returned
YES
NO
NO
YES
NO
NO
YES
NO
NO
NO
YES
NO
YES
YES
normal readout.
to process water.
Press YES for single point
NO
adjustment.
to the
NO
to *1
YESYES
YES NO
YESYES
6-4 Calibration Procedure
6-4. Calibration procedure using manual calibra-
tion method

6-4-1. Preparation

A. Cleaning
Confirm that the readings of the analyzer are stable and measured with a clean sensor.
Otherwise move the sensor to a maintenance site and wash off any dirt on the sensor membrane Move the sensor back to the process water and wait for the readings to stabilize.
B. Sampling
The manual calibration procedure involves adjustment of the readings to a reference standard. Normally this reference is a laboratory measurement that analyses a sample that must be identical to the sample measured with the on­line analyzer.

6-4-2. Procedure for manual slope (sensitivity) calibration

The following precautions must be taken into account:
• The sample must be analyzed directly after sampling to avoid change of dissolved oxygen concentration or temperature after sampling
• The reference standard must be analyzed with a laboratory meter that is calibrated prior to sample analysis
• The reference method must be calibrated with the same calibration data as specified in ISO 5814 to avoid indiscrepancies
Most important calibration data are: Dissolved oxygen concentration at saturation conditions as function of temperature, barometric pressure, altitude and salinity.
These data are included in this manual for reference and verification of correct calibration of reference analyzers. DO202 has taken all these variables into account during the semi-automatic calibration methods.
MODE
FAILTEMP.MAN
ppm ppb %sat
YES NO
NO MODEYES
CAL
OUTPUT
DISPLAY
ENT
ENT
HOLD
SET HOLD SERVICE
YES
YES
NO
Select the flashing digit with the key.
Increase its value by pressing the key.
When the correct value is displayed, press
ENT to enter the change.
Press YES to return to the
measuring mode. (Single Point
adjustment only)
YES
YES
NO
After briefly displaying WAIT,
the display returns to
ENT
the normal readout.
YES
ENT
ENT
ENT
NOYES

6-4-3. Procedure for manual zero (offset) calibration

MODE
FAILTEMP.MAN
ppm ppb %sat
YES NO
NO MODEYES
CAL
OUTPUT
DISPLAY
ENT
ENT
YES
HOLD
SET HOLD SERVICE
NO
YES
Calibration Procedure 6-5
YES
NO
YES
ENT
ENT
Select the flashing digit with the key.
Increase its value by pressing the key.
When the correct value is displayed, press ENT to enter the change.
ENT
ENT
NOYES
YES
After briefly displaying WAIT,
the display returns to
the normal readout.
Press YES to return to the
measuring mode. (Single Point
adjustment only)
6-6 Calibration Procedure
Table 6-1. Solubility of oxygen (mg/l) in water as a function of temperature & salinity (ISO 5814)
Temp Solubility of oxygen in water in equilibrium Correction to be substracted for each degree of salinity
with air expressed in grams per kilogram of total salts in water
@101.325kPa[pO2] [∆pO2] °C mg/l (ppm) mg/l (ppm)
0 14.62 0.0875 1 14.22 0.0843 2 13.83 0.0818 3 13.46 0.0789 4 13.11 0.0760 5 12.77 0.0739 6 12.45 0.0714 7 12.14 0.0693 8 11.84 0.0671 9 11.56 0.0650 10 11.29 0.0632 11 11.03 0.0614 12 10.78 0.0593 13 10.54 0.0582 14 10.31 0.0561 15 10.08 0.0545 16 9.87 0.0532 17 9.66 0.0514 18 9.47 0.0500 19 9.28 0.0489 20 9.09 0.0475 21 8.91 0.0464 22 8.74 0.0453 23 8.58 0.0443 24 8.42 0.0432 25 8.26 0.0421 26 8.11 0.0407 27 7.97 0.0400 28 7.83 0.0389 29 7.69 0.0382 30 7.56 0.0371
Table 6-2. Solubility of oxygen (mg/l) at various temperatures and elevations (based on sea level barometric pressure of 760 mm Hg)
Temp Elevation (metres above sea level)
°C 0 300 600 900 1200 1500 1800
0 14.6 14.1 13.6 13.2 12.7 12.3 11.8 2 13.8 13.3 12.9 12.4 12.0 11.6 11.2 4 13.1 12.7 12.2 11.9 11.4 11.0 10.6 6 12.4 12.0 11.6 11.2 10.8 10.4 10.1 8 11.8 11.4 11.0 10.6 10.3 9.9 9.6 10 11.3 10.9 10.5 10.2 9.8 9.5 9.2 12 10.8 10.4 10.1 9.7 9.4 9.1 8.8 14 10.3 9.9 9.6 9.3 9.0 8.7 8.3 16 9.9 9.5 9.2 8.9 8.6 8.3 8.0 18 9.5 9.2 8.9 8.6 8.3 8.0 7.7 20 9.1 8.8 8.5 8.2 7.9 7.7 7.4 22 8.7 8.4 8.1 7.8 7.6 7.3 7.1 24 8.4 8.1 7.8 7.6 7.3 7.1 6.8 26 8.1 7.8 7.6 7.3 7.0 6.8 6.6 28 7.8 7.5 7.3 7.0 6.8 6.6 6.3 30 7.5 7.2 7.0 6.8 6.5 6.3 6.1 32 7.3 7.1 6.8 6.6 6.4 6.1 5.9 34 7.1 6.9 6.6 6.4 6.2 6.0 5.8 36 6.8 6.6 6.3 6.1 5.9 5.7 5.5 38 6.6 6.4 6.2 5.9 5.7 5.6 5.4 40 6.4 6.2 6.0 5.8 5.6 5.4 5.2
Table 6-3. Solubility of oxygen (mg/l) of sea water & fresh water (based on sea level barometric pressure of 760 mm Hg)
Temp Solubility Sea water Fresh water °C mg/l (ppm) mg/l (ppm)
0 11.97 14.62 2 11.36 13.84 4 10.82 13.13 6 10.29 12.48 8 9.84 11.87 10 9.43 11.33 12 9.05 10.83 14 8.69 10.37 16 8.37 9.95 18 8.06 9.54 20 7.77 9.17 22 1.48 8.83 24 7.21 8.53 26 6.93 8.22 28 6.67 7.92 30 6.41 7.63

7. MAINTENANCE

Maintenance 7-1
It is important for maintaining the measurement accuracy of the EXA DO series of wire dissolved­oxygen metering system to perform inspection and maintenance at fixed intervals. It also serves to pre-
7-1-1. Inspection and maintenance to be imple-
mented periodically
Table 7-1. shows the items which are recommended
to be inspected and maintained periodically. vent problems from arising. This chapter describes daily inspection and main­tenance for the purpose of maintaining system per­formance.
7-1-2. Inspection and maintenance to be imple-
mented on occasion
Occasionally implement the maintenance that is not
directly related to measurement functions and main-

7-1. Overall dissolved-oxygen metering system

tenance associated with errors.
Tables 7-1 and 7-2 show the inspection and main­tenance items for equipment composing dissolved­oxygen metering system. The procedure of inspec­tion and maintenance for the dissolved-oxygen transmitter is described in Section 7-2. For inspec­tion and maintenance of the sensors and holders, see each instruction manual respectively.
Table 7-1. Items for periodic inspection and maintenance of dissolved-oxygen metering system
Instrument or apparatus Item Recommended Interval
All equipment a. Calibration a. 1 to 2 months Dissolved oxygen sensor a. Cleaning of membrane a. 1 to 2 months b. Polishing of silver electrode b. 6 to 8 months
surface and replacement of electrolyte c. Replacement of membrane assembly c. 6 to 8 months Dissolved oxygen transmitter a. Inspection of temperature indication a. 6 months
(error check) Float holder a. Sweeping of winch mechanism a. (depends on installation environment) Immersion holder a. Inspection of sensor holding O-ring a. 6 months b. Sweeping of cleaner nozzle b. 1 to 2 months c. Inspection for leakage from the solenoid c. 6 months
valve for cleaning utility
7-2. Periodic maintenance for the EXA DO202
transmitter
The DO202 transmitter requires very little periodic main­tenance. The housing is sealed to IP65
Note: Never used harsh chemicals or solvents. In
the event that the window becomes heavily stained or scratched, refer to the parts list
(Chapter 9) for replacement part numbers. (NEMA 4X) standards, and remains closed in normal operation. Users are required only to make sure the front window is kept clean in order to permit a clear view of the display and allow proper operation of the pushbuttons. If the window becomes soiled, clean it using a soft damp cloth or soft tissue. To deal with more stubborn stains, a neutral detergent may be used.
When you must open the front cover and/or glands, make sure that the seals are clean and correctly fit­ted when the unit is reassembled in order to maintain the housing’s weatherproof integrity against water and water vapor. The DO measurement uses small signal currents and may otherwise be prone to problems caused by exposure of the circuitry to condensation.
Table 7-2. Items for occasional inspection and maintenance of dissolved-oxygen metering
Faulty phenomenon System content of inspection
Error E2 occurrence Check whether the dissolved oxygen sensor membrane is contaminated or not.
Check whether the nozzle is clogged or not.
Error E3 occurrence Polish the silver electrode surface of the dissolved oxygen sensor and replace the
electrolyte.
It is difficult to see the inside of the transmitter Wipe off the dirt on the transparent window using tissue paper or
through the window the like. (Use detergent for hard-to-clean dirt.)

8. TROUBLESHOOTING

8-1 Troubleshooting
This chapter describes the countermeasures for fail­ures, classifying the cases into three categories: dis­solved-oxygen transmitter failure, detection of fail­ure with the self-diagnosis function, and abnormal measured values. The causes for abnormal meas­ured values are not limited to equipment failures. If an abnormal phenomenon occurs, first check the following items:
• Is the property of the measuring solution different
from normal?
8-1. Measures in the case of transmitter opera-
tion failure
If the operation keys do not operate smoothly or the display fails (e.g., a missing character segment), repair of the printed circuit board (digital board) or replacement with a new one is required. After the printed circuit board is replaced, opera­tion checks and parameter settings are necessary. Contact Yokogawa and request board-replacement work.
• Is the dissolved-oxygen sensor properly installed?

8-2. Measures in the case of failure (Error) detection

Table 8-1. Measures in the case of failure (Error) detection
Error Generation Error Content and Causes Measures
No. Mode
E 1 AIR.CAL mode Stability failure Eliminate the causes and press the
H20. CAL mode Even after an hour has elapsed, the (NO) key to perform re calibration.
measured value change does not decrease
within the set value of ppm
• Sensor temperature changes.
• Dissolved-oxygen of the calibration solution changes.
• The value of ppm is not suitable.
E 2 H20. CAL mode Zero error failure Clean the sensor membrane and
MAN.CAL mode The zero error exceeds the set range recalibrate. If the error is detected again,
(ZERO) • Dirt sticks to the sensor membrane. replace the electrolyte and the membrane.
• Membrane abnormality. Electrolyte degradation.
E 3 AIR.CAL mode Sensitivity failure Clean the sensor membrane and
H20. CAL mode The sensitivity exceeds the range of recalibrate. If the error is detected again,
MAN.CAL mode 1 - 1999 nA/ppm. replace the electrolyte and the membrane.
(SENS) • Dirt sticks to the sensor membrane.
• Membrane abnormality. Electrolyte degradation.
E 7 All modes Measured temperature failure (too high) Examine the temperatures of the
It exceeds 150.0 °C (302 °F). measuring solution and sensor and the
• Measuring solution temperature is high CODE 10 setting
• CODE 10 setting is not correct Examine the sensor cable connection
Sensor cable wiring failure status.
Temperature sensor has failed If the temperature sensor fails (abnormal
resistance), replace the sensor.
E 8 All modes Measured temperature failure (too low) Examine the temperatures of the
It falls below -20 °C (or -4 °F). measuring solution and sensor
• Measuring solution temperature is low and the CODE 10 setting.
• CODE 10 setting is not correct Examine the sensor cable connection
• Sensor cable wiring failure status.
• Temperature sensor has failed. If the temperature sensor fails (abnormal
resistance), replace the sensor.
E 9 All modes Input current failure Examine the setting for CODE 01, 022
It does not satisfy the following equation:
input current < 50 A for galvanic sensor
< 1200 nA for polarographic sensor
Troubleshooting 8-2
Error Generation Error content and Causes Measures
No. Mode
E10 All modes EEPROM writing failure Turn off the power and then turn it Electronic circuit failure back on and check whether or not the system returns to normal. If the failure occurs again, contact Yokogawa to request repair. E15 CODE 11 Temperature error correction failure Accurately measure the temperature of The difference between the corrected value the measuring solution (dissolved oxygen and the standard value exceeds the range sensor). corresponding to ±7.5°C (±13 5°F). Examine whether the sensor cable fails
• Temperature entry is not accurate. (partial disconnection or degraded
• Sensor cable failure insulation) and then correct the temperature error again.
E16 All modes Call for maintenance-interval time exceeded Perform maintenance. Reset interval.
mA
E17 *OUTP Output span too small. Re-enter a value within the permissible range.
CODE 35 Min. span is 1ppm, 25ppb or 10%
mA
E18 CODE 35 Table values make no sense. Reprogram. E19 At data entry Configuration failure Re-enter a value within the permissible A value not within the permissible range is entered. range. E20 All modes Instrument initialization failure Contact Yokogawa E21 All modes EPROM comparison failure Contact Yokogawa
Note: It is possible that the error message does
not disappear immediately after corrective measures. Then turn power off and the turn it back on and check whether the system returns to normal.

9. SPARE PARTS

See Customer Maintenance Parts List.
9-1 Spare Parts
Figure 9-1. Exploded view

APPENDIX 1. USER SETTINGS

A-1-1. User setting table

FUNCTION SETTING DEFAULTS USER SETTINGS
Parameter specific functions
01 *S.TYPE 1 Polarographic
*V.POL 0.675 V
*SENS 7.5 nA/ppm
04 *SAL.TY 0 Off
*[Cl-] 0 g/kg
Temperature measuring functions
10 *T.SENS 1 22k NTC
11 *T.UNIT 0 °C
12 *T.ADJ None
13 *T.MAN 0 Off
Calibration parameter functions
20 *T.SEC 60 sec.
*Z.PPM 0.1 ppm
*Z.PPB 1 ppb
*Z.SAT 1.0 %
*S.PPM 0.1 ppm
21 *Z.CAL 0 Off
*Z.LIM 199.9 nA (polar)
22 *ZERO 0 A (nA)
*SENS 7,5 nA/ppm
mA
mA outputs
31 *OUTP.F 0 Linear ppm
32 *BURN 0 No Burn
35 *TABLE 0 to 20 ppm
User Interface
50 *RET 1 on
51 *P.MAN 0 off
*P.% 101.3 kPa
*P.CAL 101.3 kPa
52 *PASS 0.0.0 all off
53 *Err.01 1 hard fail
*Err.02 1 hard fail
*Err.07 1 hard fail
*Err.08 1 hard fail
*Err.09 1 hard fail
*Err.16 0 soft fail
55 *CALL.M 0 Off
*CALL.M 250 days
56 *UNIT 0 ppm
Communication
mA
60 *COMM. 1.0 on/write enable
*ADDR. 00
mA
61 *HOUR
62 *ERASE
General
70 *LOAD
Test and setup mode
80 *TEST
Appendix 1-1
1-2 Appendix

A-1-2 Configuration Checklist for DO202

Standard configuration Options Reference for change
Measured variable(s)
Primary inputs DO and Temp
DO units ppm ppb. % saturation code 56
Temperature unit Celsius Fahrenheit code 11
mA
mA Outputs
Analog output 4-20mA for ppm ppb or %sat code 56, 31
Output linearization disabled ppm, ppb or %sat code 35
Burn out disabled burn low (3.6) / high (21) on mA output code 32
mA
Communication
Digital interface HART disabled code 60
Communication software external HHC contact factory
Interface
Variables on display ppm and temp %sat, °C, SE, ZR, REL “display”
Password protection disabled for mains/ comm./ serv level code 52
Autoreturn return to measure in 10 minutes enable or disable code 50
Add. functions in MAINT disabled manual pressure code 51
Diagnostics
Check on Slope active 1-1999 na/ppm code 22
Check on zero disabled enable or disable code 02, 21, 22
Check on stability 0.1 ppm per 60 s 0-50 ppm in 5-600 s code 20
Compatibility
DO sensor SM31 DO30G code 01, 10
Temperature sensor 22k NTC Pt 1000 code 10
Sensor principle polarographic galvanic code 01, “wiring”
Manual temp. comp. disabled disable or enable code 13
Special Features
Salinity compensation disabled comp. for 1-99.9 ppt NaCI code 04
Temperature calibration none adjustment ±7.5 °C or ±13.5°F code 12
Zero calibration disabled disable or enable code 21
Atm. press. compensation manual
Soft fail alarm disabled possible for E1..E2, E7..E9, E16 code 53
Logbook disabled 2 volumes of 50 events code 61,62

A-1-3 HART process variables

mA
Through HART the user has access to the process variables: PV: User selected process value: ppm, ppb or % saturation SV: Temperature TV: % Saturation QV: sensor current
with two 0- 20 or 4- 20 mA outputs and three contact outputs with 24 VDC power supply. Full communication with the DO202 can be done through various brands of HART Handheld termi­nals (HHT) or through Pactware. This program and the Yokogawa DTM library can be downloaded from YOKOGAWA WEB sites free of charge by
Yokogawa customers. Each of these process variables can be decoded into current output or contact output through the HIM: Hart
Yokogawa recommends the use of MH-02 modem
for HART-RS232 conversion. Interface Module. Yokogawa offers standard a HIM
Appendix 1-3

A-1-4 Device description (DD) menu structure

mA
The Device Description (DD) is available from Yokogawa or the HART® foundation. An example is shown below of the ON LINE menu structure. This manual makes no attempt to explain the operation of the Hand Held Communicator (HHC). For detailed operating instructions, refer to the HHC instruction manual and the on-line help structure.
Online menu Level 1 menu Level 2 menu Level 3 menu Level 4 menu Level 5 menu
Device setup Process variables PV %sat Temp PV % rnge More process var. Cell cur Sens. Zero cur Diag/Service/Cal Status Hold Hold func. Hold type Hold fix Hold status
Manual temperature Temp.meas. Man.temp.
Logbook Logbook config. Event1...Event27 Logbook1 Logbook2
Calibrate Zero cal. Sample cal. Zero cal. Temp. cal. Loop test
Basic setup Tag Device information Date Descriptor Message Write protect Manufacturer Dev id
Detailed setup DO measurement PV meas. Sensor type Sensor Pol. volt
Pressure specific %satpres Cal.pres
Salinity comp. Sal.comp. Salinity Temp.specific Temp.sens Temp.unit Temp.adj
Cal. specific Sens. Zero cur. Zero lim. Stab.time Stab.Zero Stab.Span
Output function mA funct. Burn funct. mA-table
User Interface Error prog. Error1...Error16
Review Model Maintenance Maint.Timer Manufacturer M.time Primary value Distributor M.due in Analog output Tag Lower range value Descriptor Display Auto.Ret Upper range value Message Man.press.en. Date Passcodes Maintenance Dev id Commissioning Write protect Service Universal rev Fld dev rev Software rev Hardware rev Poll addr Num req preams
(Note):HART protocol DD files can be downloaded by following URL.
http://www.yokogawa.com/an/download/an-dl-fieldbus-001en.htm
APPENDIX 2. QUALITY INSPECTION

A-2-1. DO202G 2-Wire Dissolved Oxygen Transmitter

Appendix 2-1
Quality Inspection
DO202G Dissolved Oxygen Transmitter
Standards
1. Scope
This inspection standard applies to the DO202G Dissolved Oxygen Transmitter.
2. Inspection Items
2.1 Insulation resistance test
2.2 Current output test
2.3 Temperature indication check * 2.4 Galvanic indication check * 2.5 Polarograph indication check
Note: Items marked with an asterisk (*) may only be confirmed by a test certificate.
3. Inspection Methods, Standards and Conditions
z Connect the testing circuit as shown in Figure 1. Allow the instrument to warm up for at least
5 minutes before conducting the tests. For the connections for the insulation resistance test, follow the instructions in Section 3.1.
z Performance tests should be done in the inspection mode where the tests from Section 3.2
through Section 3.5 take place in sequence and cannot be retraced. If the reconfirmation of a test is needed, turn off the power to the transmitter, turn on the power again, and enter the inspection mode to restart the tests.
3.1 Insulation Resistance Test
(1) Apply 500 V DC between the power supply terminals shorted together (+ and –) and the
earth terminal (G). The insulation resistance must be 100 M or greater.
(2) Apply 500 V DC between the input terminals shorted together (11 to 18) and the earth
terminal (G). The insulation resistance must be 100 M or greater.
3.2 Current Output Test
Connect the testing circuit as shown Figure 1 and set the equipment as follows:
Decade resistance box: 1019.5 [] DC generator 1: 0 A DC generator 2: 0 A DC source: 24 V DC
User the following procedure to enter the inspection mode. (1) Entering Service Code 87.
a. Press the [] key. The message display will show “OUTP.” b. Press the [NO] key twice. The message display will show “SERV.” c. Press the [YES] key. The data display will show “00” with the first digit of 0 flashing. d. Press the [
flashing.
e. Press the [>] key once. The data display will show “80” with the second digit of 0
flashing.
f. Press the [
flashing.
g. Press the [ENT] key. The message display will show “PASS.”
] key eight times. The data display will show “80” with the first digit of 8
] key seven times. The data display will show “87” with the second digit of 7
QIS 12J05C01-01E
1st Edition: Feb. 2007
2-2 Appendix
(2) Setting Password 070.
(3) Confirming the date and time.
This is the inspection mode. When the message display shows “mA,” check the current outputs. Each time the [ENT] key is pressed, the value on the data display increase in steps of 4.0 like “4.0,” and “8.0.” Check the current outputs. Corresponding to the data display, the current output must be within the range shown in Table 1.
2/3
a. Press the [>] key once. The data display will show “000” with the second digit of 0
flashing.
b. Press the [
7 flashing. c. Press the [ENT] key. The message display will show “HIF.” d. Press the [YES] key.
a. Press the [ENT] key. The data display will show the date in day, month, year format. b. Press the [ENT] key. The data display will show the time in hour, minute, second format. c. Press the [ENT] key. The message display will show “mA.”
Table 1
] key seven times. The data display will show “070” with the second digit of
Data Display Current Output (mA DC)
4.0 4 ±0.02
12.0 12 ±0.02
20.0 20 ±0.02
3.3 Temperature Indication Check
Following Section 3.2, press the [ENT] key until the message display shows “22kNTC.” in this state, change the resistance of the decade resistance box and check the data display. The value on the data display must be within the range shown in Table 2.
Table 2
Temperature (°C)
-9 99.6k -9 ±0.3
25 22k 25 ±0.3
140 653.8 140 ±0.3
3.4 Galvanic Indication Check
Following Section 3.3, press the [ENT] key until the message display shows “GALVA.” Then turn SW1 on. Change the current of the DC generator 1 and check the data display. The value on the data display must be within the range shown in Table 3.
Table 3
Reference (A) Data Display (A)
10 10 ±0.06 30 30 ±0.1 50 50 ±0.2
3.5 Polarograph Indication Check
Press the [ENT] key until the message display shows “POLAR.1.” Then turn SW1 off. Change the current of the DC generator 2 and check the data display. The value on the data display must be within the range shown in Table 4.
Decade Box Resistance (:)
Data Display (°C)
QIS 12J05C01-01E
Appendix 2-3
3/3
Table 4
Reference (nA) Data Display (nA)
2 2 ±0.02 6 6 ±0.04
12 12 ±0.07
Press the [ENT] key. The message display will show “POLAR.2.” Change the current of the DC generator 2 and check the data display. The value on the data display must be within the range shown in Table 5.
Table 5
Reference (nA) Data Display (nA)
20 20 ±0.1 60 60 ±0.3
100 100 ±0.5
Press the [ENT] key. The message display will show “POLAR.3.” Change the current of the DC generator 2and check the data display. The value on the data display must be within the range shown in Table 6.
Table 6
Reference (nA) Data Display (nA)
200 200 ±1 600 600 ±3
1000 1000 ±5
Press the [ENT] key until the message display shows “READY.” Press the [ENT] key again to finish the tests.
DO202G
SUPPLY TEMP
+
+
DC Source
24 V DC
G 11 12 16 15 13 18 14 17
-
+
DC
Ammeter
-
300 :
­Decade
Resistance
Box
Figure 1 Testing Circuit and Test Equipment
+
-
+
DC
Generator 1
+SH
-
GALV SENS POL SENS
SW1
-
+
DC
Generator 2
Shorting bar
QIS 12J05C01-01E
2-4 Appendix

A-2-2. DO202S 2-Wire Dissolved Oxygen Transmitter

Appendix 2-5
Quality Inspection
DO202S Dissolved Oxygen Transmitter
Standards
1. Scope
This inspection standard applies to the DO202S Dissolved Oxygen Transmitter.
2. Inspection Items
2.1 Insulation resistance test * 2.2 Dielectric strength test
2.3 Current output test
2.4 Temperature indication check * 2.5 Galvanic indication check * 2.6 Polarograph indication check
Note: Items marked with an asterisk (*) may only be confirmed by a test certificate.
3. Inspection Methods, Standards and Conditions
z Connect the testing circuit as shown in Figure 1. Allow the instrument to warm up for at least
5 minutes before conducting the tests. For the connections for the insulation resistance test, follow the instructions in Section 3.1, and for the connections for the dielectric strength test, follow the instructions in Section 3.2.
z Performance tests should be done in the inspection mode where the tests from Section 3.3
through Section 3.6 take place in sequence and cannot be retraced. If the reconfirmation of a test is needed, turn off the power to the transmitter, turn on the power again, and enter the inspection mode to restart the tests.
3.1 Insulation Resistance Test
(1) Apply 500 V DC between the power supply terminals shorted together (+ and –) and the
earth terminal (G). The insulation resistance must be 100 M or greater.
(2) Apply 500 V DC between the input terminals shorted together (11 to 18) and the earth
terminal (G). The insulation resistance must be 100 M or greater.
3.2 Dielectric Strength Test
Apply 600 V AC, an AC voltage of substantially sinusoidal waveform with a frequency of 50 Hz or 60 Hz, between the terminals shown below, for at least 2 seconds. The insulation must withstand this voltage. (The sensed current should be 10 mA.)
(1) Between the power supply terminals shorted together (+ and –) and the earth terminal (G)
(2) Between the input terminals shorted together (11 to 18) and the earth terminal (G)
(3) Between the input terminals shorted together (11 to 18) and the power supply terminals
shorted together (+ and –)
3.3 Current Output Test
Connect the testing circuit as shown Figure 1 and set the equipment as follows:
Decade resistance box: 1019.5 DC generator 1: 0 A DC generator 2: 0 A DC source: 24 V DC
User the following procedure to enter the inspection mode.
QIS 12J05C01-21E
1st Edition: Feb. 2007
2-6 Appendix
(1) Entering Service Code 87.
(2) Setting Password 070.
a. Press the [] key. The message display will show “OUTP.” b. Press the [NO] key twice. The message display will show “SERV.” c. Press the [YES] key. The data display will show “00” with the first digit of 0 flashing. d. Press the [
flashing.
e. Press the [>] key once. The data display will show “80” with the second digit of 0
flashing.
f. Press the [
flashing.
g. Press the [ENT] key. The message display will show “PASS.”
a. Press the [>] key once. The data display will show “000” with the second digit of 0
flashing.
b. Press the [
7 flashing. c. Press the [ENT] key. The message display will show “HIF.” d. Press the [YES] key.
] key eight times. The data display will show “80” with the first digit of 8
] key seven times. The data display will show “87” with the second digit of 7
] key seven times. The data display will show “070” with the second digit of
2/4
(3) Confirming the date and time.
a. Press the [ENT] key. The data display will show the date in day, month, year format. b. Press the [ENT] key. The data display will show the time in hour, minute, second format. c. Press the [ENT] key. The message display will show “mA.”
This is the inspection mode. When the message display shows “mA,” check the current outputs. Each time the [ENT] key is pressed, the value on the data display increase in steps of 4.0 like “4.0,” and “8.0.” Check the current outputs. Corresponding to the data display, the current output must be within the range shown in Table 1.
Table 1
Data Display Current Output (mA DC)
4.0 4 ±0.02
12.0 12 ±0.02
20.0 20 ±0.02
3.4 Temperature Indication Check
Following Section 3.3, press the [ENT] key until the message display shows “22kNTC.” In this state, change the resistance of the decade resistance box and check the data display. The value on the data display must be within the range shown in Table 2.
Table 2
Temperature (°C)
–9 99.6k –9 ±0.3 25 22k 25 ±0.3
140 653.8 140 ±0.3
Decade Box Resistance (:)
Data Display (°C)
QIS 12J05C01-21E
3.5 Galvanic Indication Check
Following Section 3.4, press the [ENT] key until the message display shows “GALVA.” Then turn SW1 off. Change the current of the DC generator 1 and check the data display. The value on the data display must be within the range shown in Table 3.
Table 3
Reference (A) Data Display (A)
10 10 ±0.06 30 30 ±0.1 50 50 ±0.2
3.6 Polarograph Indication Check
Press the [ENT] key until the message display shows “POLAR.1.” Then turn SW1 on. Change the current of the DC generator 2 and check the data display. The value on the data display must be within the range shown in Table 4.
Table 4
Reference (nA) Data Display (nA)
2 2 ±0.02 6 6 ±0.04
12 12 ±0.07
Appendix 2-7
3/4
Press the [ENT] key. The message display will show “POLAR.2.” Change the current of the DC generator 2 and check the data display. The value on the data display must be within the range shown in Table 5.
Table 5
Reference (nA) Data Display (nA)
20 20 ±0.1 60 60 ±0.3
100 100 ±0.5
Press the [ENT] key. The message display will show “POLAR.3.” Change the current of the DC generator 2 and check the data display. The value on the data display must be within the range shown in Table 6.
Table 6
Reference (nA) Data Display (nA)
200 200 ±1 600 600 ±3
1000 1000 ±5
Press the [ENT] key until the message display shows “READY.” Press the [ENT] key again to finish the tests.
QIS 12J05C01-21E
2-8 Appendix
4/4
DO202S
SUPPLY TEMP
+
+
DC Source
24 V DC
G 11 12 16 15 13 18 14 17
-
+
DC
Ammeter
-
300 :
­Decade
Resistance
Box
Figure 1 Testing Circuit and Test Equipment
+
-
+
DC
Generator 1
+SH
-
GALV SENS POL SENS
SW1
-
+
DC
Generator 2
Shorting bar
QIS 12J05C01-21E
Appendix 2-9
2-10 Appendix

A-2-3. DO202G, DO202S 2-Wire Dissolved Oxygen Transmitter (Fieldbus Communication)

Quality Inspection Standards
1. Scope
This inspection standard applies to the DO202G and DO202S Dissolved Oxygen Transmitters (Fieldbus specification).
2. Inspection Items
2.1 Insulation resistance test * 2.2 Dielectric strength test
2.3 Temperature indication check * 2.4 Galvanic indication check * 2.5 Polarographic indication check * 2.6 Fieldbus communication functional check
Note: Items marked with an asterisk (*) may only be confirmed by a test certificate.
3. Inspection Methods, Standards and Conditions
z Connect the testing circuit as shown in Figure 1. Allow the instrument to warm up for at least
5 minutes before conducting the tests. For the connections for the insulation resistance test, follow the instructions in Sections 3.1 and for the connections for the dielectric strength test, follow the instructions in Section 3.2.
z Use test equipment shown in Figure 1, or equivalent, for the tests.
DO202G, DO202S Dissolved Oxygen Transmitter (Fieldbus Communication)
3.1 Insulation Resistance Test
(1) Apply 500 V DC between the power supply terminals shorted together (+ and –) and the
earth terminal (G). The insulation resistance must be 100 M or greater.
(2) Apply 500 V DC between the input terminals shorted together (11 to 18) and the earth
terminal (G). The insulation resistance must be 100 M or greater.
3.2 Dielectric Strength Test (Required Only for DO202S)
Apply 600 V AC, an AC voltage of substantially sinusoidal waveform with a frequency of 50 Hz or 60 Hz, between the terminals shown below, for at least 2 seconds. The insulation must withstand this voltage. (The sensed current should be 10 mA.)
(1) Between the power supply terminals shorted together (+ and –) and the earth terminal (G)
(2) Between the input terminals shorted together (11 to 18) and the earth terminal (G)
(3) Between the input terminals shorted together (11 to 18) and the power supply terminals
shorted together (+ and –)
3.3 Temperature Indication Check
Connect the instruments as shown in Figure 1, and set them as follows.
Connect a shorting bar between terminals POL and SENS SW1: OFF Decade resistance box: 22 k DC generator 1: 0 A DC generator 2: 0 nA DC source: 24 V DC
QIS 12J05C01-61E
1st Edition: Apr. 2007
In this state, change the resistance value of the decade resistance box as shown in Table 1. The corresponding temperature indication must be within the range.
Table 1
Reference
Temperature (°C)
Resistance of
Resistance Box ()
Data Display (°C)
–9 99.6k –9 ±0.3 25 22k 25 ±0.3
140 653.8 140 ±0.3
3.4 Galvanic Indication Check
Turn on SW1. Set the sensor type to galvanic. Setting Service Code 01 to 0.
a. Press the [] key. The message display will show “SERV.” b. Press the [YES] key. The data display will show “00” with the first digit of 0 flashing. c. Press the [>] key once. The second digit of 0 flashes. d. Press the [
] key once. The data display will show “01” with 1 flashing.
e. Press the [ENT] key. The message display will show “ STYPE.” f. Press the [
] key once. The data display will change from “1” to “0.”
g. Press the [ENT] key until the message display shows “SERV.” h. Press the [] key. The instrument returns to the measurement mode.
Set the display to A current indication by following the steps below.
a. Press the [MODE] key. The message display will show “CALIB.” b. Press the [NO] key once. The message display will show “DISP.” c. Press the [YES] key once. The message display will show temperature indication. d. Press the [NO] key five times. The message display will show A current indication.
Appendix 2-11
2/3
Change the current of DC generator 1 as shown in Table 2. The corresponding indication must be within the range.
Table 2
Reference (A) Data Display (A)
10 10 ±0.06 30 30 ±0.10 50 50 ±0.20
Press the [MODE] key to return to normal mode.
3.5 Polarographic Indication Check
Turn off SW1. Set the sensor type to polarograph. Setting Service Code 01 to 1.
a. Press the [] key. The message display will show “SERV.” b. Press the [NO] key twice. c. Press the [YES] key. The data display will show “00” with the first digit of 0 flashing. d. Press the [>] key once. The second digit of 0 flashes. e. Press the [
] key once. The data display will show “01” with 1 flashing.
f. Press the [ENT] key. The message display will show “ STYPE.” g. Press the [
] key once. The data display will change from “0” to “1.”
h. Press the [ENT] key until the message display shows “ SERV.”
QIS 12J05C01-61E
2-12 Appendix
Set the display to nA current indication by following the steps below.
Change the current of DC generator 2 as shown in Table 3. The corresponding indication must be within the range.
Press the [MODE] key to return to normal mode.
3/3
i. Press the [] key. The instrument returns to the measurement mode.
a. Press the [MODE] key. The message display will show “CALIB.” b. Press the [NO] key once. The message display will show “DISP.” c. Press the [YES] key once. The message display will show temperature indication. d. Press the [NO] key five times. The message display will show nA current indication.
Table 3
Reference (nA) Data Display (nA)
200 200 ±1 600 600 ±3
1000 1000 ±5
3.6 Fieldbus Communication Functional Check
Check for normal function using Fieldbus equipment specified by Yokogawa.
DO202
CR
CR
SUPPLY
+
L
L
R
+
DC Source 24VDC
TEMP
G111216 15+13
Decade
Resistance
Box
DC
Generator
1
RRC
+
18
+-+-
C
SH
GALV SENS POL SENS
14 17
SW1
DC
Generator
2
R=50Ω±1Ω C=2μF±0.4μF L=5mH±5% R
L=50Ω±1%
Shorting Bar
Fieldbus Equipment Specified by Yokogawa
Figure 1 Testing Circuit and Test Equipment
QIS 12J05C01-61E
Appendix 2-13
2-14 Appendix

A-2-4. DO202G, DO202S 2-Wire Dissolved Oxygen Transmitter (Profibus Communication)

Quality Inspection Standards
1. Scope
This inspection standard applies to the DO202G and DO202S Dissolved Oxygen Transmitters (Profibus specification).
2. Inspection Items
2.1 Insulation resistance test * 2.2 Dielectric strength test
2.3 Temperature indication check * 2.4 Galvanic indication check * 2.5 Polarographic indication check * 2.6 Profibus communication functional check
Note: Items marked with an asterisk (*) may only be confirmed by a test certificate.
3. Inspection Methods, Standards and Conditions
z Connect the testing circuit as shown in Figure 1. Allow the instrument to warm up for at least
5 minutes before conducting the tests. For the connections for the insulation resistance test, follow the instructions in Sections 3.1 and for the connections for the dielectric strength test, follow the instructions in Section 3.2.
z Use test equipment shown in Figure 1, or equivalent, for the tests.
DO202G, DO202S Dissolved Oxygen Transmitter (Profibus Communication)
3.1 Insulation Resistance Test
(1) Apply 500 V DC between the power supply terminals shorted together (+ and –) and the
earth terminal (G). The insulation resistance must be 100 M or greater.
(2) Apply 500 V DC between the input terminals shorted together (11 to 18) and the earth
terminal (G). The insulation resistance must be 100 M or greater.
3.2 Dielectric Strength Test (Required Only for DO202S)
Apply 600 V AC, an AC voltage of substantially sinusoidal waveform with a frequency of 50 Hz or 60 Hz, between the terminals shown below, for at least 2 seconds. The insulation must withstand this voltage. (The sensed current should be 10 mA.)
(1) Between the power supply terminals shorted together (+ and –) and the earth terminal (G)
(2) Between the input terminals shorted together (11 to 18) and the earth terminal (G)
(3) Between the input terminals shorted together (11 to 18) and the power supply terminals
shorted together (+ and –)
3.3 Temperature Indication Check
Connect the instruments as shown in Figure 1, and set them as follows.
Connect a shorting bar between terminals POL and SENS SW1: OFF Decade resistance box: 22 k DC generator 1: 0 A DC generator 2: 0 nA DC source: 24 V DC
QIS 12J05C01-71E
1st Edition: Sep. 2007
In this state, change the resistance value of the decade resistance box as shown in Table 1. The corresponding temperature indication must be within the range.
Table 1
Reference
Temperature (°C)
Resistance of
Resistance Box ()
Data Display (°C)
–9 99.6k –9 ±0.3 25 22k 25 ±0.3
140 653.8 140 ±0.3
3.4 Galvanic Indication Check
Turn on SW1. Set the sensor type to galvanic. Setting Service Code 01 to 0.
a. Press the [] key. The message display will show “SERV.” b. Press the [YES] key. The data display will show “00” with the first digit of 0 flashing. c. Press the [>] key once. The second digit of 0 flashes. d. Press the [
] key once. The data display will show “01” with 1 flashing.
e. Press the [ENT] key. The message display will show “ STYPE.” f. Press the [
] key once. The data display will change from “1” to “0.”
g. Press the [ENT] key until the message display shows “SERV.” h. Press the [] key. The instrument returns to the measurement mode.
Set the display to A current indication by following the steps below.
a. Press the [MODE] key. The message display will show “CALIB.” b. Press the [NO] key once. The message display will show “DISP.” c. Press the [YES] key once. The message display will show temperature indication. d. Press the [NO] key five times. The message display will show A current indication.
Appendix 2-15
2/3
Change the current of DC generator 1 as shown in Table 2. The corresponding indication must be within the range.
Table 2
Reference (A) Data Display (A)
10 10 ±0.06 30 30 ±0.10 50 50 ±0.20
Press the [MODE] key to return to normal mode.
3.5 Polarographic Indication Check
Turn off SW1. Set the sensor type to polarograph. Setting Service Code 01 to 1.
a. Press the [] key. The message display will show “SERV.” b. Press the [NO] key twice. c. Press the [YES] key. The data display will show “00” with the first digit of 0 flashing. d. Press the [>] key once. The second digit of 0 flashes. e. Press the [
] key once. The data display will show “01” with 1 flashing.
f. Press the [ENT] key. The message display will show “ STYPE.” g. Press the [
] key once. The data display will change from “0” to “1.”
h. Press the [ENT] key until the message display shows “ SERV.”
QIS 12J05C01-71E
2-16 Appendix
Set the display to nA current indication by following the steps below.
Change the current of DC generator 2 as shown in Table 3. The corresponding indication must be within the range.
Press the [MODE] key to return to normal mode.
3/3
i. Press the [] key. The instrument returns to the measurement mode.
a. Press the [MODE] key. The message display will show “CALIB.” b. Press the [NO] key once. The message display will show “DISP.” c. Press the [YES] key once. The message display will show temperature indication. d. Press the [NO] key five times. The message display will show nA current indication.
Table 3
Reference (nA) Data Display (nA)
200 200 ±1 600 600 ±3
1000 1000 ±5
3.6 Profibus Communication Functional Check
Check for normal function using Profibus equipment specified by Yokogawa. In the tests of Item 3.3 to 3.5, check the communication function using Profibus communication.
DO202
+
CR
CR
R=50 Ω± Ω C=2μF±0.4μF
SUPPLY
TEMP
G111216 15+13
Decade
Resistance
Box
DA/PA Coupler
DC
Generator
1
+
+
18
+-+-
DC
Source
24VDC
SH
GALV SENS ● ● POL SENS
14 17
SW1
DC
Generator
2
Shorting Bar
Profidbus Equipment Specified by Yokogawa
Figure 1 Testing Circuit and Test Equipment
QIS 12J05C01-71E
Appendix 2-17
Customer
1
5
2
8
9
10
14
Item Part No. Qty Description 1 Cover Assembly K9315CA 1 Polyurethane Coating K9315CN 1 Epoxy-polyester Coating 2 Internal Works Assembly with amplifier assembly K9662AD 1 For mA + HART K9662AE 1 For FF K9662AF 1 For Profibus
3 Housing Assembly K9661HA 1 Polyurethane Coating K9661HB 1 Epoxy-polyester Coating 4 L9811FV 2 Cable Gland Assembly 5 K9660AQ 1 Flat Cable 6 A1726JD 1 Pin Terminal Unit 3 terminals type 7 K9184AA 1 Screw Terminal Unit when /TB specified 8 K9661HR 2 Stud *9 Analog Board Assembly K9661XA 1 For mA + HART K9661XC 1 For FF/Profibus
*10 1 Digital/Display Board K9661GU 1 For mA + HART K9661GW 1 For FF/Profibus 11 K9660YQ 1 Screw Assembly to fix amplifier 12 K9660YP 1 Stainless steel screw assembly to fix cover
13 Adapter Assembly K9414DH 1 For G1/2 screw when /AFTG specified (2 units). K9414DJ 1 For 1/2NPT screw when /ANSI specified (2 units). *14 FF Board Assembly K9661ML 1 For FF K9661NL 1 For Profibus
* Do not exchange these parts. Call service personnel.
12
11
3
4
6,7
13
Maintenance Parts List
Model DO202G [Style: S2] Disolved Oxygen Transmitter
©Copyright 2007, 1st Edition: Mar, 2007 (YK)
Subject to change without notice.
CMPL 12J05C01-02E
2nd Edition : Aug.2007 (YK)
2
2
(Option Code : /SCT)
Option Code : /H
/H2
(Option Code : /PM)
3
Item Parts No. Qty Description 1 K9171SS 1 Universal Mount Set (/U)
K9311BT 1 Tag Plate (/SCT) 2 K9311KA 1 3 K9311KG 1
4
K9660JA 1
4
Fitting Assembly (/PM)
Hood Assembly (/H) Hood Assembly (/H2)
1
Pipe/Wall Mounting Hardware
(Option Code : /U)
Panel Mounting Hardware
1
Hood to sun protection
CMPL 12J05C01-02E
2nd Edition : Aug.2007 (YK)
Customer
1
5
2
8
9
10
14
Item Part No. Qty Description 1 Cover Assembly K9315CA 1 Polyurethane Coating K9315CN 1 Epoxy-polyester Coating 2 Internal Works Assembly with amplifier assembly
1 For mA + HART 1 For FF 1 For Profibus 1 For mA + HART (Non-incendive)
3 Housing Assembly
1 Polyurethane Coating
1 Epoxy-polyester Coating 4 L9811FV 2 Cable Gland Assembly 5 1 Flat Cable 6 1 Pin Terminal Unit 3 terminals type
8 2 Stud 9 Analog Board Assembly
1 For mA + HART
1 For FF/Profibus
1 For mA + HART (Non-incendive)
10 1 Digital/Display Board
1 For mA + HART
1 For FF/Profibus 11 1 Screw Assembly to fix amplifier 12 K9660YP 1 Stainless steel screw asssembly to fix cover
13 Adapter Assembly K9414DH 1 For G1/2 screw when /AFTG specified (2 units). K9414DJ 1 For 1/2NPT screw when /ANSI specified (2 units). 14 FF Board Assembly
1 For FF
1 For Profibus
12
11
3
4
6
13
Maintenance Parts List
Model DO202S [Style: S3] Disolved Oxygen Transmitter
©Copyright 2008, 1st Edition: Feb, 2008 (YK)
Subject to change without notice.
CMPL 12J05C01-23E
2nd Edition : Nov.2008 (YK)
2
2
(Option Code : /SCT)
Option Code : /H
/H2
(Option Code : /PM)
3
Item Parts No. Qty Description 1 K9171SS 1 Universal Mount Set (/U)
K9311BT 1 Tag Plate (/SCT) 2 K9311KA 1 3 K9311KG 1
4
K9660JA 1
4
Fitting Assembly (/PM)
Hood Assembly (/H) Hood Assembly (/H2)
1
Pipe/Wall Mounting Hardware
(Option Code : /U)
Panel Mounting Hardware
1
Hood to sun protection
CMPL 12J05C01-23E
2nd Edition : Nov.2008 (YK)

Revision Record

Manual Title : Model DO202G [Style: S2], DO202S [Style: S3] 2-wire Dissolved Oxygen Transmitter Manual Number : IM 12J05C01-01E
Edition Date Remark (s)
1st Mar. 2007 Newly published
2nd Oct. 2007 PREFACE-1, Some of warning description modified; P. 1-1, Some of nameplate
in Figure 1-1 changed; P. 1-2, Some of nameplate in Figure 1-2 changed; P. 2-2,
EN 61000-3-3 deleted from "M. Regulatory compliance"; P. 2-3, Certificate no. of
CENELEC ATEX and IECEx Scheme explosionproof added, CSA explosionproof
description added; P. 2-4, P. layout changed; P. 2-5, Note added to Model and
suffix codes; P. 2-8, Control Drawing for mA HART Specification (FM Intrinsically
safe design) corrected; P. 2-9, Control Drawing for mA HART Specification
(FM Non-incendive design) corrected; P. 2-10, Control Drawing for mA HART
Specification (CSA) corrected; P. 2-11, Control Drawing for FF/PB Specification
(IECEx) corrected; P. 2-12, Control Drawing for FF/PB Specification (ATEX)
corrected; P. 2-13, Control Drawing for FF/PB Specification (FM Intrinsically safe
Entity) corrected; P. 2-15, Control Drawing for FF/PB Specification (FM Intrinsically
safe FISCO) corrected; P. 2-17, Control Drawing for FF/PB Specification (FM Non-
incendive Entity) corrected; P. 2-18, Control Drawing for FF/PB Specification (FM
Non-incendive FNICO) corrected; P. 2-19, Control Drawing for FF/PB Specification
(CSA) corrected; P. 5.1, Subsection 5-1-1, "Manual activation of hold:" Made some
revisions; P. 5-3, Subsection 5-2-1, "Output range:" Made some revisions; P. 5-
6, Section 5-3, "Notes for guidance in the use of service code settings:" Added
some cautions; P. 5-7, Code 01 "*SENS" and Code 22 "*SENS:" Added default
value of galvanic sensor; P. 6-2, Subsection 6-2-2, "Procedure for air calibration:";
Made some revisions; P. 6-3, Subsection 6-3-3, "Procedure for water calibration:"
Made some revisions; P. 6-5, Subsection 6-4-3, "Procedure for manual zero
(offset) calibration:" Made some revisions; APPENDIX 2 QUALITY INSPECTION
added; CMPL 12J05C01-02E, -22E revised to 2nd edition because some part no.
changed.
3rd Apr. 2008 Style of DO202S changed to S3 and related description changed as follows.
P.2-2, "FM Approved is pending" deleted, because FM approved; P.2-5, Style of
DO202S changed to S3 for FM approval; P.3-1, Some of dimensions in Figure 3-1
corrected; P.1-3 (Appendix), Note of HART protrocol DD files URL added; CMPL
12J05C01-23E 1st edition added for DO202S style S3.
4th Oct. 2009 PREFACE, "Zone 0" added to Warning label explanation; P.1-1, Name plate of
DO202S-K (NEPSI) added to Figure 1-1; P.2-2, Some revision of M. Regulatory
compliance (description for EMC revised); P.2-3 to 2-4, Some revision of IECEx
Intrinsically safe description ("Zone 0" added), and NEPSI Certification added to
page 2-3 and 2-4; P.2-5, NEPSI suffix code of "-K" added to the DO202S MS-
code; CMPL 12J05C01-23E of DO202S(S3) revised to 2nd edition (some parts
no. deleted).
User’s Manual
Model DO202G [Style: S2], DO202S [Style: S3] 2-Wire Dissolved Oxygen Transmitter

Supplement

Thank you for selecting our Model DO202G [Style: S2] and/or DO202S [Style: S3] 2-Wire Dissolved Oxygen Transmitter. User's Manual, IM 12J05C01-01E, 4th Edition, supplied with the product, some revisions/additions have been made. Please replace the corresponding pages in your copy with the attached, revised pages.
Revisions:
- PREFACE, "How to dispose the batteries" added.
- Page 2-2, Description of Profibus added to EMC conformity standard.
All Rights Reserved, Copyright © 2007, 2nd Edition: May. 2010 (YK)
Subject to change without notice.
IM 12J05C01-01E
4th Edition
IM 12J05C01-01E
PREFACE
DANGERDANGER
CAUTIONCAUTION
WARNINGWARNING
Electric discharge
The EXA analyzer contains devices that can be damaged by electrostatic discharge. When servicing this equipment, please observe proper procedures to prevent such damage. Replacement components should be shipped in conductive packaging. Repair work should be done at grounded workstations using grounded soldering irons and wrist straps to avoid electrostatic discharge.
Installation and wiring
The EXA analyzer should only be used with equip­ment that meets the relevant international and regional standards. Yokogawa accepts no responsi­bility for the misuse of this unit.
The instrument is packed carefully with shock absorbing materials, nevertheless, the instrument may be damaged or broken if subjected to strong shock, such as if the instrument is dropped. Handle with care.
Although the instrument has a weatherproof construction, the transmitter can be harmed if it becomes submerged in water or becomes exces­sively wet.
Do not use an abrasive material or solvent when cleaning the instrument.
Do not modify the DO202 transmitter.
Electrostatic charge may cause an explosion haz­ard. Avoid any actions that cause the generation of electrostatic charge, e.g., rubbing with a dry cloth.
Warning label
Because the enclosure of the Dissolved Oxygen transmitter Type DO202S-A, -P, -F are made of alu­minium, if it is mounted in an area where the use of category 1 G Zone 0 apparatus is required, it must be installed such, that, even in the event of rare incidents, ignition sources due to impact and friction sparks are excluded.
Notice
• This manual should be passed on to the end user.
• The contents of this manual are subject to change without prior notice.
• The contents of this manual shall not be reproduced or copied, in part or in whole, without permission.
• This manual explains the functions contained in this product, but does not warrant that they are suitable the particular purpose of the user.
• Every effort has been made to ensure accuracy in the preparation of this manual.
However, when you realize mistaken expressions
or omissions, please contact the nearest Yokogawa Electric representative or sales office.
• This manual does not cover the special specifications. This manual may be left unchanged on any change of specification, construction
or parts when the change does not affect the
functions or performance of the product.
• If the product is not used in a manner specified in this manual, the safety of this product may be impaired.
Yokogawa is not responsible for damage to the
instrument, poor performance of the instrument or losses resulting from such, if the problems are caused by:
• Improper operation by the user.
• Use of the instrument in improper applications
• Use of the instrument in an improper environment or improper utility program
• Repair or modification of the related instrument by an engineer not authorized by Yokogawa.
Safety and Modification Precautions
• Follow the safety precautions in this manual when using the product to ensure protection and safety of the human body, the product and the system containing the product.
How to dispose the batteries:
This is an explanation about the new EU Battery Directive (DIRECTIVE 2006/66/EC). This directive is only valid in the EU. Batteries are included in this product. Batteries incorporated into this product can­not be removed by yourself. Dispose them together with this product. When you dispose this product in the EU, contact your local Yokogawa Europe B.V.office. Do not dispose them as domestic house­hold waste. Battery type: silver oxide battery
Notice: The symbol (see above) means they shall be
sorted out and collected as ordained in ANNEX II in DIRECTIVE 2006/66/EC.
2-2 General Specifications
CAUTION
Item
Description
Code
Factory Mutual (FM)
-N
-A
-N
-A
CENELEC ATEX
2.EPS
FM Non-incendive safe Approval
Applicable standard: FM3600, FM3611, FM3810 Non-incendive Safe for Class I, Division 2, Groups ABCD, Zone 2 Temp. Class: T4, Amb. Temp.: -10 to 55°C Non-incendive Safe Apparatus Parameters
Vmax=31.5 V, Ci=22 nF, Li=35 μH
FM Intrinsically safe Approval
Applicable standard: FM3600, FM3610, FM3810 Intrinsically Safe for Class I, Division 1, Groups ABCD Class I, Zone 0, AEx ia IIC Temp. Class: T4, Amb. Temp.: -10 to 55°C Intrinsically Safe Apparatus Parameters Vmax=31.5 V, Imax=100 mA,
Pmax=1.2 W, Ci=22 nF, Li=35 μH
CENELEC ATEX (KEMA) Type of protection "n"
Applicable standard: EN60079-0:2006, EN60079-15:2003 Certificate: KEMA 06ATEX0225
EEx nA [nL] IIC, Group: II, Category: 3G
Temp. Class: T4, Amb. Temp.: -10 to 55°C T6, Amb. Temp.: -10 to 40°C
Ui=31.5 V, Ci=22 nF, Li=35 μH
CENELEC ATEX (KEMA) Intrinsically safe Approval
Applicable standard: EN60079-0, EN50020 EN60079-26 Certificate: KEMA 06ATEX0224 X
Ex ia IIC, Group: II, Category: 1G
Temp. Class: T4, Amb. Temp.: -10 to 55°C T6, Amb. Temp.: -10 to 40°C
Ui=31.5 V, Ii=100 mA, Pi=1.2 W, Ci=22 nF, Li=35 μH
Item
Description
Code
Factory Mutual (FM)
FM Non-incendive safe Approval
Applicable standard: FM3600, FM3611, FM3810 Non-incendive Safe for Class I, Division 2, Groups ABCD, Zone 2 Temp. Class: T4, Amb. Temp.: -10 to 55°C Non-incendive Safe Apparatus Parameters
FM Intrinsically safe Approval
Applicable standard: FM3600, FM3610, FM3810 Intrinsically Safe for Class I, Division 1, Groups ABCD Class I, Zone 0, AEx ia IIC Temp. Class: T4, Amb. Temp.: -10 to 55°C Intrinsically Safe Apparatus Parameters
FM.EPS
Vmax=32 V, Pmax=1.2 W,
Ci=220 pF, Li=0 μH
Vmax=32 V, Pmax=5.32 W,
Ci=220 pF, Li=0 μH
Entity
FNICO
-F
-P
or
-D
-B
or
Vmax=24 V, Imax=250 mA,
Pmax=1.2 W, Ci=220 pF, Li=0 μH
Vmax=17.5 V, Imax=380 mA,
Pmax=5.32 W, Ci=220pF, Li=0 μH
Entity
FISCO
2-2. Performance specifications
A. Performance in ppm mode
- Linearity : ±0.05 ppm or ±0.8% FS, whichever is greater
- Repeatability : ±0.05 ppm or ±0.8% FS, whichever is greater
- Accuracy : ±0.05 ppm or ±0.8% FS, whichever is greater
B. Performance in ppb mode
- Linearity : ±1 ppb or ±0.8% FS, whichever is greater
- Repeatability : ±1 ppb or ±0.8% FS, whichever is greater
- Accuracy : ±1 ppb or ±0.8% FS, whichever is greater
C. Performance in Temperature
- Linearity : ±0.3ºC
- Repeatability : ±0.1ºC
- Accuracy : ±0.3ºC
Note on performance specifications:
The specifications are at reference temperature
and with simulated inputs, because the DO202G can be used with many different sensors with their unique characteristics. The following tolerance is added to above performance. mA output tolerance : ± 0.02 mA of "4 - 20 mA"
L. Operation protection
All three levels can be protected by a user programmable 3 digit password
M. EMC Conformity standards ,
EN 61326-1 Class A, Table 2 (For use in industrial locations) EN 61326-2-3 EN 61326-2-5 (Profibus communication may be
influenced by strong electromagnetic field.)
This instrument is a Class A product, and it is designed for use in the industrial environment. Please use this instrument in the industrial environment only.
N. Explosionproof type
mA
Refer to Control Drawings.
D. Ambient temperature
Operating temperature -10 to 55ºC
(10 to 130ºF)
Storage temperature -30 to 70ºC
(-20 to 160ºF)
E. Humidity
10 to 90% RH non-condensing
F. Housing
Cast aluminum housing with chemically
resistant coating, cover with flexible polycarbonate window. Case color is off-white with moss green cover. Cable entry is through two polyamide cable glands. Cable glands are provides for wiring up to 2.5 mm.
G. Environmental protection
IP65, NEMA 4X.
H. Mounting
I. Data protection
J. Watchdog timer: Checks microprocessor.
K. Autoreturn
Pipe, wall or panel mounting using optional hardware.
EEPROM for configuration; Lithium battery for
clock.
The transmitter returns into measuring mode automatically when no keystroke is made for 10 minutes.
IM 12J05C01-01E
Loading...