Yokogawa DO202 User Manual

User’s Manual
EXA 202
Fieldbus Communication
IM 12A00A01-61E
IM 12A00A01-61E
2nd Edition
1. INTrOduCTION .................................................................................................. 1-1
2. SAFETy PrECAuTIONS .................................................................................... 2-1
3. FOuNdATION FIELdBuS ..................................................................................3-1
3-1 About Foundation Fieldbus ..............................................................................3-1
3-1-1 Outline ..............................................................................................................................3-1
3-1-2 Internal Structure of EXA ..................................................................................................3-1
3-1-2-1 System/network Management VFD ....................................................................3-1
3-1-2-2 Function Block VFD ............................................................................................3-1
3-1-3 Logical Structure of Each Block .......................................................................................3-1
3-1-4 Wiring System Configuration ............................................................................................3-2
3-2 Getting started .................................................................................................3-2
3-2-1 Connection of Devices ......................................................................................................3-2
3-2-1-1. Fieldbus Preparation ..........................................................................................3-2
3-2-1-2. Cables, terminals and glands ............................................................................3-3
3-2-2 Host Setting ......................................................................................................................3-4
3-2-3 Bus Power ON ..................................................................................................................3-4
3-2-4 Integration of DD ..............................................................................................................3-5
3-2-5 Reading the Parameters ...................................................................................................3-5
3-2-6 Continuous Record of Values ...........................................................................................3-5
3-2-7 Generation of Alarm .........................................................................................................3-5
3-3 Configuration ...................................................................................................3-6
3-3-1 Network Design ................................................................................................................3-6
3-3-2 Network Definition.............................................................................................................3-7
3-3-3 Definition of Combining Function Blocks ..........................................................................3-8
3-3-4 Setting of Tags and Addresses ........................................................................................3-9
3-3-5 Communication Setting...................................................................................................3-10
3-3-5-1 VCR Setting.......................................................................................................3-10
3-3-5-2 Function Block Execution Control .....................................................................3-12
3-3-6 Block Setting ...................................................................................................................3-12
3-3-6-1 Link Object ........................................................................................................3-12
3-3-6-2 Trend Object......................................................................................................3-13
3-3-6-3 View Object .......................................................................................................3-14
3-3-6-4 Function Block Parameters ...............................................................................3-18
3-4 In-process operation ...................................................................................... 3-22
3-4-1 Mode Transition ..............................................................................................................3-22
3-4-2 Generation of Alarm ......................................................................................................3-22
3-4-2-1 Indication of Alarm ............................................................................................3-22
3-4-2-2 Alarms and Events ............................................................................................3-22
3-4-3 Simulation Function ........................................................................................................3-23
3-5 Device status .................................................................................................3-25
3-6 List of parameters for each block of the EXA ................................................. 3-28
3-6-1 Resource Block...............................................................................................................3-28
3-6-2 Analog input Block ..........................................................................................................3-30
3-6-3 Transducer Block ............................................................................................................3-32
3-6-3-1 Transducer Block PH202 ..................................................................................3-32
3-6-3-2 Transducer Block SC202 ..................................................................................3-35
3-6-3-3 Transducer Block ISC202 .................................................................................3-38
3-6-3-4 Transducer Block DO202 ..................................................................................3-41
3-7 Application setting and change of basic parameters .....................................3-43
3-7-1 Applications and selection of basic parameters .............................................................3-43
3-7-2 Setting and change of basic parameters ........................................................................3-44
3-7-3 Setting the AI Function Blocks .......................................................................................3-44
3-7-4 Setting the Transducer Block .........................................................................................3-45
IM 12A00A01-61E 2nd Edition: Sep. 2007(YK) All Rights Reserved, Copyright © 2007, Yokogawa Electric Corporation
IM 12A00A01-61E
3-8 Operation of each parameter in failure mode ................................................3-46
3-8-1 Operation of each parameter in failure mode PH202 ....................................................3-46
3-8-2 Operation of each parameter in failure mode SC202 ....................................................3-48
3-8-3 Operation of each parameter in failure mode ISC202 ...................................................3-50
3-8-4 Operation of each parameter in failure mode DO202 ....................................................3-52
4. PrOFIBUS ........................................................................................................... 4-1
4-1. About Profibus ................................................................................................4-1
4-1-1 Outline ..............................................................................................................................4-1
4-1-2 Internal Structure of EXA ..................................................................................................4-1
4-1-3 Logical Structure of Each BLOCK ....................................................................................4-1
4-1-4 Wiring System Configuration ............................................................................................4-2
4-2. Preparation .....................................................................................................4-4
4-2-1. Cables, terminals and glands ..........................................................................................4-4
4-2-2. Shielding and grounding ..................................................................................................4-4
4-3. GETTING STARTED ......................................................................................4-5
4-3-1 Connection of Devices ......................................................................................................4-5
4-3-2 Host Setting ......................................................................................................................4-6
4-3-3 Bus Power ON ..................................................................................................................4-6
4-3-4 Reading cyclic parameters ...............................................................................................4-6
4-3-5 Reading acyclic parameters .............................................................................................4-7
4-4. Function block parameters and Methods ........................................................4-8
4-4-1. Physical Block Parameters ..............................................................................................4-8
4-4-2. Analog Input Block Parameters .......................................................................................4-8
4-4-3. Transducer block parameters ..........................................................................................4-9
4-4-3-1. Transducer block parameters PH202 ................................................................4-9
4-4-3-2. Function Block Parameters SC202 ..................................................................4-12
4-4-3-3. Function Block Parameters ISC202 .................................................................4-14
4-4-3-4. Function Block Parameters DO202..................................................................4-17
4-4-4 Methods ..........................................................................................................................4-19
APPENdIX 1. LINK MASTEr FUNCTIONS ........................................................... 5-1
A1-1 Link Active Scheduler .................................................................................... 5-1
A1-2 Link Master .................................................................................................... 5-1
A1-3 Transfer of LAS .............................................................................................5-2
A1-4 LM Functions ................................................................................................. 5-3
A1-5 LM Parameters .............................................................................................. 5-4
A1-5-1 LM Parameter List ..........................................................................................................5-4
A1-5-2 Descriptions for LM Parameters.....................................................................................5-6
A1-6 FAQs .............................................................................................................5-8
rEvISION rECOrd ....................................................................................................
1
IM 12A00A01-61E
Introduction 1-1
Manual Name IM No. Instruments mentioned
2-wire pH/ORP Transmitter IM 12B07D02-01E PH202G, PH202S 2-wire Conductivity or Resistivity Transmitter IM 12D08B02-01E SC202G, SC202S 2-wire Inductive Conductivity Transmitter IM 12D06A03-01E ISC202G, ISC202S 2-wire Dissolved Oxygen Transmitter IM 12J05C01-01E DO202G, DO202S
T01.EPS

1. IntroductIon

In the standard user’s manual delivered with the 202 analyzer all necessary information about HART-com­munication is included. This manual describes only those topics that are required for operation of the field­bus communications. For information about instruments related to the EXA202, refer to the following User’s Manuals.
IM 12A00A01-61E
Safety Precautions 2-1
WARNING
CAUTION
IMPORTANT
WARNING
CAUTION

2. SAFETy PrECAUTIONS

• For the protection and safety of the operator and the instrument or the system including the instrument, please be sure to follow the instructions on safety described in this manual when handling this instrument.
In case the instrument is handled in contradiction to these instructions, Yokogawa does not guarantee
safety.
• For the intrinsically safe equipment and explosionproof equipment, in case the instrument is not restored to its original condition after any repair or modification undertaken by the customer, intrinsically safe construction or explosionproof construction is damaged and may cause dangerous condition. Please contact Yokogawa for any repair or modification required to the instrument.
• The following safety symbol marks are used in this Manual:
Indicates a potentially hazardous situation which, if not avoided, could result in death or serious injury.
Indicates a potentially hazardous situation which, if not avoided, may result in minor or moderate injury. It may also be used to alert against unsafe practices.
Indicates that operating the hardware or software in this manner may damage it or lead to system failure.
NOTE Draws attention to information essential for understanding the operation and features.
• Instrument installed in the process is under pressure. Never loosen or tighten the process connector bolts as it may cause dangerous spouting of process fluid.
• During draining condensate or venting gas in transmitter pressure-detector section, take appropriate care to avoid contact with the skin, eyes or body, or inhalation of vapors, if the accumulated process fluid may be toxic or otherwise harmful.
Since draining condensate or bleeding off gas gives the pressure measurement distur-bance, this should
not be done when the loop is in operation.
• If the accumulated process fluid may be toxic or otherwise harmful, take appropriate care to avoid contact with the body, or inhalation of vapors even after dismounting the instrument from process line for maintenance.
This instrument is tested and certified as intrinsically safe type or explosionproof type. Please note that the construction of the instrument, installation, external wiring, maintenance or repair is strictly restricted, and non-observance or negligence of these restriction would result dangerous condition.
IM 12A00A01-61E
Foundation Fieldbus 3-1
System/network management VFD
Function block VFD
PD Tag
Sensor
input
Resource block
Block tag
Parameters
Communication
parameters
VCR
Node address
Function block
execution schedule
AI function
block
AI function
block
Output
AI function
block
Block tag
OUT
Parameters
Transducer
block
Block tag
Parameters
Sensor

3. FOUNdATION FIELdBUS

3-1 About Foundation Fieldbus

3-1-1 Outline

Fieldbus is a bi-directional digital communication protocol for field devices, which offers an advance­ment implementation technologies for process control systems and is widely employed by numerous field devices.
EXA Series Fieldbus communication type employs the specification standardized by The Fieldbus Foundation, and provides interoperability between Yokogawa devices and those produced by other manufacturers. Fieldbus comes with software consisting of three AI function blocks, providing the means to implement flexible instrumentation system.
For information on other features, engineering, design, construction work, startup and maintenance of Fieldbus, refer to http://www.yokogawa.com/fbs/fbs-index.htm.

3-1-2 Internal Structure of EXA

The EXA contains two virtual field devices (VFD) that share the following functions.
3-1-2-1 System/network Management vFd
• Sets node addresses and Physical Device tags (PD Tag) necessary for communication
• Controls the execution of function blocks
• Manages operation parameters and communication resources (Virtual Communication Relationship: VCR)
3-1-2-2 Function Block vFd
(1) resource block
• Manages the status of EXA hardware
• Automatically informs the host of any detected faults or other problems
(2) Transducer block
• Converts sensor output to process values and transfers to AI function block by channels
(3) AI1, AI2, AI3 function block
• Conditions raw data from the Transducer block
• Outputs conditioned process values
• Carries out scaling, damping and square root extraction

3-1-3 Logical Structure of Each Block

Setting of various parameters, node addresses, and PD Tags shown in Figure 3.1 is required before starting operation.
Figure 3.1 Logical Structure of Each Block
IM 12A00A01-61E
3-2 Foundation Fieldbus
Sensor cable gland
Foundation Fieldbus®
cable gland Grounding terminal (connect to safety ground, only if power supply is not grounded)

3-1-4 Wiring System Configuration

The number of devices that can be connected to a single bus and the cable length vary depending on sys­tem design. When constructing systems, both the basic and overall design must be carefully considered to allow device performance to be fully exhibited.

3-2 Getting started

Fieldbus is fully dependent upon digital communication protocol and differs in operation from conventional 4 to 20 mA transmission communication protocol. It is recommended that novice users use field devices in accordance with the procedures described in this section. The procedures assume that field devices will be set up on a bench or an instrument shop.

3-2-1 Connection of Devices

3-2-1-1. Fieldbus Preparation
The Foundation Fieldbus® connections and the sensor connections should be made in accordance with fig­ure 3.2 and 3.3. The terminals are of a plug in style for ease of mounting. The EXA 202 FF is provided with two cable glands. The first is used for the electrode wiring as the other is used for the power/foundation® Fieldbus wiring shown in figure 3.2. To open the EXA 202 for wiring:
1. Loosen the four frontplate screws and remove the cover.
2. The terminal strip is now visible.
3. Connect the power supply to the green connector according figure 3.3. Use the gland on the left for this cable.
4. Connect the sensor input, using the gland on the right (see figure 3.2). Switch on the power. Commission the instrument as required or use the default settings.
5. Replace the cover and secure frontplate with the four screws.
Figure 3.2 Glands to be used for cabling
Figure 3.3 Green connector for power supply
IM 12A00A01-61E
Foundation Fieldbus 3-3
EXA
Power supply
Terminator
HOST
Coupler/terminator
IMPORTANT
3-2-1-2. Cables, terminals and glands
The EXA202 is equipped with terminals suitable for the connection of finished cables in the size range:
0.13 to 2.5 mm (26 to 14 AWG). The glands will form a tight seal on cables with an outside diameter in the
range of 6 to 12 mm (0.24 to 0.47 inches). The following instruments are required for use with Fieldbus devices:
• Power supply
Fieldbus requires a dedicated power supply. It is recommended that current capacity be well over the total value of the maximum current consumed by all devices (including the host). Conventional DC current cannot be used as is.
• Terminator
Fieldbus requires two terminators. Refer to the supplier for details of terminators that are attached to the host.
• Field devices
Connect EXA Fieldbus communication type. Two or more EXA devices or other devices can be connected.
• Host
Used for accessing field devices. A dedicated host (such as DCS) is used for an instrumentation line while dedicated communication tools are used for experimental purposes. For operation of the host, refer to the instruction manual for each host. No details of the host are explained in the rest of this material.
• Cable
Used for connecting devices. Refer to “Fieldbus Technical Information” (TI 38K03A01-01E) for details of instrumentation cabling. Fieldbus uses twisted pair wires. To meet the Electro Magnetic Interference standards a shielded twisted pair is obligated.
Refer to Yokogawa when making arrangements to purchase the recommended equipment. Connect the devices as shown in Figure 3.4. Connect the terminators at both ends of the trunk, with a mini-
mum length of the spur laid for connection. The polarity of signal and power must be maintained.
Figure 3.4 Cabling
NOTE Before using a Fieldbus configuration tool other than the existing host, confirm it does not affect
the loop functionality in which all devices are already installed in operation. Disconnect the relevant control loop from the bus if necessary.
Connecting a Fieldbus configuration tool to a loop with its existing host may cause communication data scrambles resulting in a functional disorder or a system failure.
IM 12A00A01-61E
3-4 Foundation Fieldbus
IMPORTANT
Not used
LM device
Not used
Basic device
Default address
Portable-device address
V (FUN)
V (FUN) + V (NUN)
V (NUN)
0xFF
0xFC
0xFB
0xF8
0xF7
0x00
0x0F
0x10 0x13
0x14
Bridge device
Note 1: LM device: with bus control function (Link Master function) Note 2: BASIC device: without bus control function
EXA

3-2-2 Host Setting

To activate Fieldbus, the following settings are required for the host.
Do not turn off the power immediately after setting. When the parameters are saved to the EEPROM, the redundant processing is executed for an improvement of reliability. If the power is turned off within 60 seconds after setting is made, the modified parameters are not saved and the settings may return to the original values.
Table 3.1 Operation Parameters
Symbol Parameter Description and Settings
V (ST) Slot-Time Set 4 or greater value. V (MID) Minimum-Inter-PDU-Delay Set 4 or greater value. V (MRD) Maximum-Reply-Delay Set so that V (MRD) X V (ST) is 12 or greater V (FUN) First-Unpolled-Node Define the first address that can be used by the host. Set 0x15 or greater. V (NUN) Number-of-consecutive-Unpolled-Node This sets the number of consecutive unpolled nodes. EXA address is
factory-set to 0xEB. Set this address to be within the range of the BASIC device in Figure 3.5.
Figure 3.5 Available Address Range

3-2-3 Bus Power ON

Turn on the power of the host and the bus. First all segments of the display are lit, then the display begins to operate. If the indicator is not lit, check the polarity of the power supply.
Using the host device display function, check that the EXA is in operation on the bus. Unless otherwise specified, the following settings are in effect when shipped from the factory.
PH202 SC202 ISC202 DO202 PD tag PH1001 SC1001 ISC1001 DO1001 Node addr. 232 233 234 235 DEV_TYPE 0x0830 0x0831 0x0832 0x0833
If no EXA is detected, check the available address range and the polarity of the power supply. If the node address and PD tag are not specified when ordering, default value is factory set. If two or more EXA’s are connected at a time with default value, only one EXA will be detected from the host as EXA’s have the same initial address. Separately connect each EXA and set a different address for each.
IM 12A00A01-61E
Foundation Fieldbus 3-5

3-2-4 Integration of dd

If the host supports DD (Device Description), the DD of the EXA needs to be installed. Check if host has the following directory under its default DD directory.
594543\DEV_TYPE (594543 is the manufacturer number of Yokogawa Electric Corporation, and DEV_TYPE is the EXA device number, respectively.) If this directory is not found, DD of EXA has not been included. Create the above directory and copy the DD file (0m0n.ffo,0m0n.sym) (m, n is a numeral) (to be supplied separately) into the directory.
Once the DD is installed in the directory, the name and attribute of all parameters of the EXA are displayed. Off-line configuration is possible by using Capability file (CFF).

3-2-5 reading the Parameters

To read EXA parameters, select the AI1 block of the EXA from the host screen and read the OUT param­eter. The current process value is displayed. Check that MODE_BLOCK of the function block and resource block is set to AUTO.

3-2-6 Continuous record of values

If the host has a function of continuously recording the indications, use this function to list the indications (values). Depending on the host being used, it may be necessary to set the schedule of Publish (the func­tion that transmits the indication on a periodic basis).

3-2-7 Generation of Alarm

If the host is allowed to receive alarms, generation of an alarm can be attempted from EXA. In this case, set the reception of alarms on the host side. EXA’s VCR-7 is factory-set for this purpose. For practical purposes, all alarms are placed in a disabled status; for this reason, it is recommended that you first use one of these alarms on a trial basis. Set the value of link object-3 (index 30002) as “0, 299, 0, 6, 0”. Refer to section 3-3-6-1 Link Object for details.
Since the LO_PRI parameter (index 4029) of the AI1 block is set to “0”, try setting this value to “3”. Select the Write function from the host in operation, specify an index or variable name, and write “3” to it.
The LO_LIM parameter (index 4030) of the AI1 block determines the limit at which the lower bound alarm for the process value is given. In usual cases, a very small value is set to this limit. Set a value higher than the current process value, a lower bound alarm is raised. Check that the alarm can be received at the host. When the alarm is confirmed, transmission of the alarm is suspended.
The above-mentioned items are a description of the simple procedure to be carried out until EXA is con­nected to Fieldbus. In order to take full advantage of the performance and functionality of the device, it is recommended that it be read together with Chapter 5 of relevant User’s Manual, which describes how to use the EXA.
IM 12A00A01-61E
3-6 Foundation Fieldbus

3-3 Configuration

This chapter contains information on how to adapt the function and performance of the EXA to suit specific applications. Because two or more devices are connected to Fieldbus, settings including the requirements of all devices need to be determined. Practically, the following steps must be taken.
(1) Network design
Determines the devices to be connected to Fieldbus and checks the capacity of the power supply.
(2) Network definition
Determines the tag and node addresses for all devices.
(3) definition of combining function blocks
Determines the method for combination between each function block.
(4) Setting tags and addresses
Sets the PD Tag and node addresses one by one for each device.
(5) Communication setting
Sets the link between communication parameters and function blocks.
(6) Block setting
Sets the parameters for function blocks.
The following section describes each step of the procedure in the order given. Using a dedicated configu­ration tool allows the procedure to be signifi-cantly simplified. This section describes the procedure to be assigned for a host which has relatively simple functions.

3-3-1 Network design

Select the devices to be connected to the Fieldbus network. The following instruments are necessary for operation of Fieldbus.
• Power supply
Fieldbus requires a dedicated power supply. It is recommended that current capacity be well over the total value of the maximum current consumed by all devices (including the host). Conventional DC current cannot be used as is. A power conditioner is reguired.
• Terminator
Fieldbus requires two terminators. Refer to the supplier for details of terminators that are attached to the host.
• Field devices
Connect the field devices necessary for instrumentation. EXA has passed the interoperability test conducted by The Fieldbus Foundation. In order to properly start Fieldbus, it is recommended that the devices used satisfy the requirements of the above test.
• Host
Used for accessing field devices. A minimum of one device with bus control function is needed.
• Cable
Used for connecting devices. Refer to “Fieldbus Technical Information” for details of instrumentation cabling. Provide a cable sufficiently long to connect all devices. For field branch cabling, use terminal boards or a connection box as required.
IM 12A00A01-61E
Foundation Fieldbus �-� �-�
Not used
LM device
Not used
Basic device
Default address
Portable-device address
V (FUN)
V (FUN) + V (NUN)
V (NUN)
0xFF
0xFC
0xFB
0xF8
0xF7
0x00
0x0F
0x10 0x13
0x14
Bridge device
Note 1: LM device: with bus control function (Link Master function) Note 2: BASIC device: without bus control function
(EXA 0xEB)
First, check the capacity of the power supply. The power supply capacity must be greater than the sum of the maximum current consumed by all devices to be connected to Fieldbus. The maximum current con­sumed (power supply voltage 9 to �2 V) for EXA is 26.0 mA. The cable must have the spur in a minimum length with terminators installed at both ends of the trunk.

3-3-2 Network Definition

Before connection of devices with Fieldbus, define the Fieldbus network. Allocate PD Tag and node addresses to all devices (excluding such passive devices as terminators).
The PD Tag is the same as the conventional one used for the device. Up to �2 alphanumeric characters may be used for definition. Use a hyphen as a delimiter as required.
The node address is used to specify devices for communication purposes. Because data is too long for a PD Tag, the host uses the node address in place of the PD Tag for communication. A range of 20 to 24� (or hexadecimal 0x14 to 0xF�) can be set.
Addresses of devices with Link Master capabilities are set in a low address range smaller than V(FUN). Addresses of basic devices are set in a higher range bigger than V(FUN) + V(NUN). Specify the adress range used by setting the following two parameters in the LM-device:
Table 3.2 Parameters for Setting Address Range
Symbol Parameters Description
V (FUN) First-Unpolled-Node Indicates the address next to the address range used for the host or other
LM device.
V (NUN) Number-of-consecutive-Unpolled-Nodes Unused address range
The devices within the address range written as “Not used” in Figure �.6 cannot be used on a Fieldbus. For other address ranges, the range is periodically checked to identify when a new device is connected. Care must be taken not to allow the address range to become wider, which can lead to exhaustive consumption of Fieldbus communication performance.
Figure 3.6 Available Range of Node Addresses
To ensure stable operation of Fieldbus, determine the operation parameters and set them to the LM devic­es. While the parameters in Table �.� are to be set, the worst-case values of all the devices to be connect­ed to the same Fieldbus must be used. Refer to the specification of each device for details. Table �.� lists EXA specification values.
IM 12A00A01-61E
3-8 Foundation Fieldbus
LIC100
FIC100
FC100
FI100
EXA
#2
LI100
EXA
#1
Table 3.3 Operation Parameter Values of the EXA to be Set to LM Devices
Symbol Parameters Description and Settings
V (ST) Slot-Time Indicates the time necessary for immediate reply of the device. Unit of time is in octets
(256 µs). Set maximum specification for all devices. For EXA, set a value of 4 or greater.
V (MID) Minimum-Inter-PDU-Delay Minimum value of communication data intervals. Unit of time is in octets (256 µs). Set the
maximum specification for all devices. For EXA, set a value of 4 or greater.
V (MRD) Maximum-Reply-Delay The worst case time elapsed until a reply is recorded. The unit is Slot-time; set the value
so that V (MRD) 3V (ST) is the maximum value of the specification for all devices. For EXA, the setting must be a value of 12 or greater.

3-3-3 Definition of Combining Function Blocks

The input/output parameters for function blocks are combined. For the EXA, three AI blocks output param­eter (OUT) are subject to combination. They are combined with the input of the control block as necessary. Practically, setting is written to the EXA link object with reference to “Block setting” in Section 3-3-6 for details. It is also possible to read values from the host at proper intervals instead of connecting the EXA block output to other blocks.
The combined blocks need to be executed synchronously with other blocks on the communications sched­ule. In this case, change the EXA schedule according to the following table. Enclosed values in the table are factory-settings.
Table 3.4 Execution Schedule of the EXA Function Blocks
Index Parameters Setting (Enclosed is factory-setting)
269(SM) MACROCYCLE_DURATION
276(SM) FB_START_ENTRY.1
277(SM) FB_START_ENTRY.2
278(SM) FB_START_ENTRY.3 279(SM) to
289(SM)
FB_START_ENTRY.4 to 14 Not used.
Cycle (MACROCYCLE) period of control or measurement. Unit is 1/32 ms. (32000 = 1 s)
AI1 block startup time. Elapsed time from the start of MACROCYCLE specified in 1/32 ms. (0 = 0 s)
AI2 block startup time. Elapsed time from the start of MACROCYCLE specified in 1/32 ms. (9600 = 0.3 s)
AI3 block startup time. Elapsed time from the start of MACROCYCLE specified in 1/32 ms. (19200 = 0.6 s)
A maximum of 29 ms is taken for execution of an AI block. Executions of AI blocks should be scheduled sequentially. In no case should two AI function blocks of the EXA be executed at the same time (execution time is overlapped). 29 ms after AI block execution start the out value is available for further processing. Figure 3.8 shows an example of schedule based on the loop shown in Figure 3.7.
Figure 3.7 Example of Loop Connecting Function Block of Two EXA with Other Instruments
IM 12A00A01-61E
Foundation Fieldbus 3-9 3-9
LI100
LIC100
FIC100
FC100
FI100
Commu-
nication
Schedule
OUT
IN
OUT
CAS_IN
BKCAL_OUT
BKCAL_IN
BKCAL_IN
BKCAL_OUT
IN
Unscheduled
Communication
Scheduled Communication
Macrocycle (Control Period)
Figure 3.8 Function Block Schedule and Communication Schedule
When the macrocycle is set to more than 4 seconds, set the following intervals to be more than 1% of the macrocycle.
- Interval between “end of block execution” and “start of sending CD from LAS”
- Interval between “end of block execution” and “start of the next block execution”“end of block execution” and “start of the next block execution”

3-3-4Settingof Tags andAddresses Setting of Tags and Addresses

This section describes the steps in the procedure to set PD Tags and node addresses in the EXA. There are three states of Fieldbus devices as shown in Figure 3.9, and if the state is other than SM_ OPERATIONAL state, no function block is executed. EXA must be transferred back to this state after a tag or address is changed.
UNINITIALIZED
(No tag nor address is set)
Tag clear Tag setting
INITIALIZED
(Only tag is set)
Address clear Address setting
SM_OPERATIONAL
(Tag and address are retained, and
the function block can be executed.)
Figure 3.9 Status Transition by Setting PD Tag and Node Address
EXA has a PD Tag and node address that are set upon shipment from the factory unless otherwise speci­fied. To change only the node address, clear the address once and then set a new node address. To set the PD Tag, first clear the node address and clear the PD Tag, then set the PD Tag and node address again.
Devices whose node address was cleared will await the default address (randomly chosen from a range of 248 to 251, or from hexadecimal F8 to FB). It is necessary to confirm the device ID in order to correctlyorder to correctly specify the device. The device ID of the EXA is 594543083xxxxxxxxx. (The xxxxxxxxx at the end of the above device ID is a total of 9 alphanumeric characters. The characters displyed on the screen when each instrument is powered on, correspond to the xxxxxxxxx and if necessary, should be recorded.)
IM 12A00A01-61E
3-10 Foundation Fieldbus

3-3-5 Communication Setting

To set the communication function, it is necessary to change the database residing in SM-VFD.
3-3-5-1 VCR Setting
Set VCR (Virtual Communication Relationship), which specifies the called party for communication and resources. EXA has 33 VCRs whose application can be changed, except for the first VCR, which is used for management.
EXA has VCRs of 3 types:
Publisher(-Subscriber) VCR
Publisher-Subscriber VCR’s are designed to link Function Blocks. When a publishing Function Block runs, its output data is stored in the buffer of the Publisher VCR. Then the LAS (LM) sends a CD to this VCR to force it to transfer the data. Subscriber VCRs receive this data and gives this to the subscribing Function Blocks. Typical example is a linkage from an output of an Analog Input (AI) block to the process value input of the PID control block.
Publisher-Subscriber model is one-to-many one-way Communication. Subscribers are able to know whether data is updated since the last publish. This mechanism is important because Data Link Layer transfers data as scheduled regardless the publishing Function Block updates the data in the buffer.
(Client-)Server Model
Client-Server model is universal and used in many communication technologies. An application called “Client” requests another application called “Server” to do a specific action. When the Server finishes the requested action, its result is transferred back to the Client. It is an one-to-one two-way communication. Typical example is a human-machine interface (Client) to read data of a Function Block (Server). The Client sends a Read request to the Server and then the Server sends back the data to the Client. This communi­cation is unscheduled and is handled during the unscheduled interval in the macrocycle. A Client may want to issue many requests at a time. A Client-Server VCR has a queue to store those requests and sends the requests one by one when the node has the token.
Source(-Sink) Model
A Source-Sink VCR is designed to broadcast messages. It is one-to-many one-way communication without any schedule. This model is sometimes called “Report Distribution Model.” A Source VCR transfers a mes­sage in the queue to an assigned global address when the device has the token. Sink VCRs are set to the same global address and receive the same message from a Source. Foundation devices use this model for two specific purposes. One is to report alarms or events detected in the Source and the other is to transmit trends of Source Function Blocks. Alarms are acknowledged through a Client-Server VCR. It is desirable for an alarm logger to receive alarms from all devices with just one VCR. A Sink can receive messages from many Sources if the Sources are configured to send messages to the same global address.
A Source VCR transmits data without established connection. A Sink (QUU) VCR on another device can receive it if the Sink is configured so. A Publisher VCR transmits data when LAS requests so. An explicit connection is established from VCR(s) so that a Subscriber knows the format of published data.
Each VCR has the parameters listed in Table 3.5. Parameters must be changed together for each VCR because modification for each parameter may cause inconsistent operation.
IM 12A00A01-61E
Table 3.5 VCR Static Entry
Sub­index
1 FasArTypeAndRole Indicates the type and role of
2 FasDllLocalAddr Sets the local address to
3 FasDllConfigured
4 FasDllSDAP Specifies the quality of
5 FasDllMaxConfirm
6 FasDllMaxConfirm
7 FasDllMaxDlsduSize Specifies maximum DL Service
8 FasDllResidual
9 FasDllTimeliness
Parameter Description
communication (VCR). The following 3 types are used for EXA. 0x32: Server (Responds to
requests from host.)
0x44: Source (Transmits alarm
or trend.)
0x66: Publisher (Sends AI
block output to other blocks.)
specify VCR in EXA. A range of 0x20 to 0xF7 in hexadecimal.
Sets the node address of the
RemoteAddr
DelayOnConnect
DelayOnData
ActivitySupported
Class
called party for communication and the address (DLSAP or DLCEP) used to specify VCR in that address. For DLSAP or DLCEP, a range of 0x20 to 0xF7 in hexadecimal is used. Addresses in Subindex 2 and 3 need to be set to the same contents of the VCR as the called party (local and remote are reversed).
communication. Usually, one of the following types is set. 0x2B: Server 0x01: Source (Alert) 0x03: Source (Trend) 0x91: Publisher
To establish connection for communication, a maximum wait time for the called party’s response is set in ms. Typical value is 60 seconds (60000).
For request of data, a maximum wait time for the called party’s response is set in ms. Typical value is 60 seconds (60000).
Data unit Size (DLSDU). Set 256 for Server and Trend VCR, and 64 for other VCRs.
Specifies whether connection is monitored. Set TRUE (0xff) for Server. This parameter is not used for other communication.
Not used.
Foundation Fieldbus 3-11
Sub­index
10 FasDllPublisherTime
11 FasDllPublisher
12 FasDllSubsriberTime
13 FasDllSubscriber
14 FmsVfdId Sets VFD for EXA to be used.
15 FmsMaxOutstanding
16 FmsMaxOutstanding
17 FmsFeatures
Parameter Description
Not used.
WindowSize
Not used.
SynchronizaingDlcep
Not used.
WindowSize
Not used.
SynchronizationDlcep
0x1: System/network
management VFD
0x1234: Function block VFD Set 0 to Server. It is not
ServiceCalling
ServiceCalled
Supported
used for other applications. Set 1 to Server. It is not
used for other applications. Indicates the type of
services in the application layer. In the EXA, it is automatically set according specific applications.
3-11
33 VCRs are factory-set as shown in the table 3.6.
IM 12A00A01-61E
3-12 Foundation Fieldbus
Table 3.6 VCR List VCR List
Index (SM) VCR Number Factory Setting
293 1 For system management (Fixed) 294 2 Server (LocalAddr = 0xF3) 295 3 Server (LocalAddr = 0xF4) 296 4 Server (LocalAddr = 0xF7) 297 5 Trend Source (LocalAddr = 0x07, Remote Address=0x111) 298 6 Publisher for AI1 (LocalAddr = 0x20) 299 7 Alert Source (LocalAddr = 0x07, Remote Address=0x110) 300 8 Server (LocalAddr = 0xF9) 301 to 325 to 325 9 to 33 Not set
3-3-5-2 Function Block Execution Control
According to the instructions given in Section 3-3-3, set the execution cycle of the function blocks and schedule of execution.

3-3-6 Block Setting

Set the parameter for function block VFD.
3-3-6-1 Link Object
Link object combines the data voluntarily sent by the function block with VCR. The EXA has 40 link objects. A single link object specifies one combination. Each link object has the parameters listed in Table 3.7. Parameters must be changed together for each VCR because the modifications made to each parameter may cause inconsistent operation.
Table 3.7 Link Object Parameters
Sub­index
1 LocalIndex Sets the index of function block parameters to be combined; set “0” for Trend and Alert. 2 VcrNumber Sets the index of VCR to be combined. If set to “0”, this link object is not used. 3 RemoteIndex Not used in EXA. Set to “0”.
4 ServiceOperation
5 StaleCountLimit
Parameters Description
Set one of the following. Only one link object is used for Alert and/or Trend. 0: Undefined 2: Publisher 6: Alert 7: Trend
Set the maximum number of consecutive stale input values which may be received before the input status is set to BAD. To avoid the unnecessary mode transition caused when the data is not correctly received by subscriber, set this parameter to “2” or more.
Set link objects as shown in Table 3.8.
Table 3.8 Factory-Settings of Link Objects (example)
Index Link Object# Factory Settings
30000 1 AI1.OUT VCR#6 30001 2 Trend VCR#5 30002 3 Alert VCR#7 30003 to 30039 4 to 40 Not used
IM 12A00A01-61E
Foundation Fieldbus 3-13
SMIB
(System Management Information Base)
NMIB
(Network Management Information Base)
AI3
OUT
AI1
OUT
AI2
OUT
FBOD
Alert
Trend
VCR
DLSAP DLCEP
Fieldbus Cable
0xF8 0xF3 0xF4 0xF7
0xF9
0x20 0x220x21
0x07
#5
Device 3
#10
#1 #4
#3
#2
Resource block
Transducer block
Host 1
Host 2
Device 1
Device 2
Link object
#1
#2
#3 #4
#5
#6
#7
#8
#9
3-13
3-3-6-2 Trend Object
It is possible to set the parameter so that the function block automatically transmits Trend. The EXA has ten Trend objects: eight for trends of analog paramenters and two for discrete parameters. A single Trend object specifies the trend of one parameter.
Each Trend object has the parameters listed in Table 3.9. The first four parameters are the items to be set. Before writing to a Trend object, it is necessary to release the WRITE_LOCK parameter.
Table 3.9 Parameters for Trend Objects
Sub-index Parameters Description
1 Block Index Sets the leading index of the function block that takes a trend. 2 Parameter Relative
Index
3 Sample Type Specifies how trends are taken. Choose one of the following 2 types:
4 Sample Interval Specifies sampling intervals in units of 1/32 ms. Set the integer multiple of the function block
5 Last Update The last sampling time. 6 to 21 List of Status 16 samples of status. 21 to 37 List of Samples 16 samples of data.
Sets the index of parameters taking a trend by a value relative to the beginning of the function block. In the EXA AI block, the following three types of trends are possible. 7: PV 8: OUT 19: FIELD_VAL
1:Sampled upon execution of a function block. 2:The average value is sampled.
execution cycle.
Ten trend objects are factory-set as shown Table 3.10.
Table 3.10 Trend Object are Factory-Set
Index Parameters Factory Settings
32000 to 32007 TREND_FLT.1 to TREND_FLT.8 Not used. 32008 to 32009 TREND_DIS.1 to TREND_DIS.2 Not used.
Figure 3.10 Example of Default Configuration
IM 12A00A01-61E
3-14 Foundation Fieldbus
3-3-6-3 ViewObject6-3 View Object
This is the object to form groups of parameters in a block. One advantage of forming groups of parameters is the reduction of load for data transaction. The EXA has four View Objects for each Resource block, Transducer block and AI1, AI2, AI3 function block, and each View Object has the parameters listed in Table
3.12 to 3.14.
Table 3.11 Purpose of Each View Object
Description
VIEW_1 Set of dynamic parameters required by operator for plant
operation. (PV, SV, OUT, Mode etc.)
VIEW_2 Set of static parameters which need to be shown to plant
operator at once. (Range etc.) VIEW_3 Set of all the dynamic parameters VIEW_4 Set of static parameters for configuration or maintenance.
Table 3.12 Indexes of View for Each Block
VIEW_1 VIEW_2 VIEW_3 VIEW_4
Resource Block 40100 40101 40102 40103 Transducer Block 40200 40201 40202 40203 AI1 Function Block 40400 40401 40402 40403 AI2 Function Block 40410 40411 40412 40413 AI3 Function Block 40420 40421 40422 40423
IM 12A00A01-61E
Table 3.13 View Object for Resource Block Table 3.13 View Object for Resource Block
Foundation Fieldbus 3-15
(contineous)
Relative Index
1 ST_REV 2 2 2 2 2 TAG_DESC 3 STRATEGY 2 4 ALERT_KEY 1 5 MODE_BLK 4 4 6 BLOCK_ERR 2 2 7 RS_STATE 1 1 8 TEST_RW 9 DD_RESOURCE 10 MANUFAC_ID 4 11 DEV_TYPE 2 12 DEV_REV 1 13 DD_REV 1 14 GRANT_DENY 2 15 HARD_TYPES 2 16 RESTART 17 FEATURES 2 18 FEATURE_SEL 2 19 CYCLE_TYPE 2 20 CYCLE_SEL 2 21 MIN_CYCLE_T 4 22 MEMORY_SIZE 2 23 NV_CYCLE_T 4 24 FREE_SPACE 4 25 FREE_TIME 4 4 26 SHED_RCAS 4 27 SHED_ROUT 4 28 FAULT_STATE 1 1 29 SET_FSTATE 30 CLR_FSTATE 31 MAX_NOTIFY 1 32 LIM_NOTIFY 1 33 CONFIRM_TIME 4 34 WRITE_LOCK 1 35 UPDATE_EVT 36 BLOCK_ALM 37 ALARM_SUM 8 8 38 ACK_OPTION 2 2 39 WRITE_PRI 40 WRITE_ALM 41 ITK_VER 2 42 SOFT_REV 43 SOFT_DESC 44 SIM_ENABLE_MSG 45 DEVICE_STATUS_1 4 46 DEVICE_STATUS_2 4 47 DEVICE_STATUS_3 4 48 DEVICE_STATUS_4 4
Parameter Mnemonic VIEW1VIEW2VIEW3VIEW
Relative
4
Index
49 DEVICE_STATUS_5 4 50 DEVICE_STATUS_6 4 51 DEVICE_STATUS_7 4 52 DEVICE_STATUS_8 4 53 54 55 56 57 58
Table 3.14 View Object for AI1.AI2.AI3 Function Block
Relative Index
1 ST_REV 2 2 2 2 2 TAG_DESC 3 STRATEGY 2 4 ALERT_KEY 1 5 MODE_BLK 4 4 6 BLOCK_ERR 2 2 7 PV 5 5 8 OUT 5 5 9 SIMULATE 10 XD_SCALE 11 11 OUT_SCALE 11 12 GRANT_DENY 2 13 IO_OPTS 2 14 STATUS_OPTS 2 15 CHANNEL 2 16 L_TYPE 1 17 LOW_CUT 4 18 PV_FTIME 4 19 FIELD_VAL 5 5 20 UPDATE_EVT 21 BLOCK_ALM 22 ALARM_SUM 8 8 23 ACK_OPTION 2 24 ALARM_HYS 4 25 HI_HI_PRI 1 26 HI_HI_LIM 4 27 HI_PRI 1 28 HI_LIM 4 29 LO_PRI 1 30 LO_LIM 4 31 LO_LO_PRI 1 32 LO_LO_LIM 4 33 HI_HI_ALM 34 HI_ALM 35 LO_ALM 36 LO_LO_ALM
Parameter Mnemonic VIEW1VIEW2VIEW3VIEW
SOFTDWN_PROTECT SOFTDWN_FORMAT SOFTDWN_COUNT SOFTDWN_ACT_AREA SOFTDWN_MOD_REV SOFTDWN_ERROR
TOTALS (# BYTES) 22 30 73 35
Parameter Mnemonic VIEW1VIEW2VIEW3VIEW
TOTALS (# BYTES) 31 26 31 46
3-15
4
1 1 2
1
16
2
4
IM 12A00A01-61E
3-16 Foundation Fieldbus
Table 3.15 View Object for Transducer Block PH202 Table 3.15 View Object for Transducer Block PH202
(contineous)
Relative Index
0 BLOCK HEADER 1 ST_REV 2 2 2 2 2 TAG_DESC 3 STRATEGY 2 4 ALERT_KEY 1 5 MODE_BLK 4 4 6 BLOCK_ERR 2 2 7 UPDATE_EVT 8 BLOCK_ALM 9 TRANSDUCER_
10 11 XD_ERROR 1 1 12 COLLECTION_
13 14 PRIMARY_VALUE 5 5 15
16 SENSOR_TYPE_PH 17 SENSOR_MV 18 CAL_POINT_HI 4 19 CAL_POINT_LO 4 20 CAL_MIN_SPAN 4 21 SLOPE 4 22 SLOPE_UNIT 2 23 ZERO 4 24 ZERO_UNIT 2 25 ISOPOTENTIAL_PH 4 26 27 SENSOR_CAL_DATE 8 28 29 SECONDARY_
30 31 SENSOR_TEMP_
32 33 SENSOR_
34 35 TERTIARY_VALUE 5 5 36 TERTIARY_VALUE_
37 TERTIARY_ZERO 4 38 GLASS_IMPEDANCE 4 39 REFERENCE_
40 ALARM_SUM 8 8 41 DEV_ALARM 4 4 42 LOGBOOK1_RESET 43 LOGBOOK1_EVENT 44 LOGBOOK2_RESET 45 LOGBOOK2_EVENT 46 LOGBOOK_CONFIG 23 47 TEST_1 48-59 TEST_2 ... TEST_13 60 STABLE_TIME 61 STABLE_VALUE 62 CALL_MAINT_TIME_
63 CALL_MAINT_TIME_
64 INPUT_1_
65 INPUT_1_
PARAMETER NAME VIEW_1VIEW_2VIEW_3VIEW
_4
DIRECTORY TRANSDUCER_TYPE
DIRECTORY PRIMARY_VALUE_TYPE
PRIMARY_VALUE_ RANGE
SENSOR_CAL_METHOD
SECONDARY_VALUE
VALUE_UNIT SENSOR_TEMP_COMP
MAN_VALUE SENSOR_TYPE_TEMP
CONNECTION_TEMP TERTIARY_VALUE_TYPE
RANGE
IMPEDANCE
COUNTD.
RELOAD
IMPEDANCE_LO_LIM
IMPEDANCE_HI_LIM
2 2 2 2
2
11
1
5 5
2
1 4
2 1
2
11
4
1
1
4
4
Relative Index
66 INPUT_2_
67 INPUT_2_
68 BUFFER1_ID 69 BUFFER1 70 BUFFER2_ID 71 BUFFER2 72 BUFFER3_ID 73 BUFFER3 74 TEMPERATURE_
75 PASSCODE_
76 PASSCODE_
77 PASSCODE_SERVICE 78 SAMPLE_PV 4 79 SAMPLE_PV2 4 80 SAMPLE_TEMP 4 81 ERROR_CONFIG 4 82 CONFIGURATION 4 83 TRANSMITTER_TIME 6 84 SOFTDWN_TEST 1
Table 3.16 View Object for Transducer Block SC202/ISC202
Realtive Index
0 BLK_DATA 1 ST_REV 2 2 2 2 2 TAG_DESC[32] 3 STRATEGY 2 4 ALERT_KEY 1 5 MODE_BLK 4 4 6 BLOCK_ERR 2 2 7 UPDATE_EVT 8 BLOCK_ALM 9 TRANSDUCER_
10 TRANSDUCER_TYPE 2 2 2 2 11 XD_ERROR 1 1 12 COLLECTION_
13 PRIMARY_VALUE_
14 PRIMARY_VALUE 5 5 15 PRIMARY_VALUE_
16 SENSOR_CONST 4 17 CAL_POINT_HI 4 18 CAL_POINT_LO 4 19 CAL_MIN_SPAN 4 20 SENSOR_CAL_
21 SENSOR_CAL_DATE 8 22 SECONDARY_VALUE 5 5 23 SECONDARY_VALUE_
24 SENSOR_TEMP_COMP 1 25
26 SENSOR_TYPE_TEMP 2 27
28 SENSOR_TYPE_COND 2 29 SENSOR_OHMS
PARAMETER NAME VIEW_1VIEW_2VIEW_3VIEW
IMPEDANCE_LO_LIM
IMPEDANCE_HI_LIM
COEFFICIENT
MAINTENANCE
COMMISSIONING
TOTALS (# BYTES) 38 18 65 119
Parameters Mnemonic
DIRECTORY[2]
DIRECTORY[7]
TYPE
RANGE
METHOD
UNIT
SENSOR_TEMP_MAN_ VALUE
SENSOR_CONNECTION_ TEMP
View1View2View3View
2
2
_4
4
4
4
11
1
1
IM 12A00A01-61E
Table 3.16 View Object for Transducer Block SC202/ISC202 (contineous)
Realtive Index
30 XD_MAN_ID[32] 31 TEMPERATURE_
32 CONCENTRATION 5 5 33 TERTIARY_VALUE 5 5 34 REFERENCE_
35 COMP_METHOD 1 1 36 COMP_MATRIX_SEL 1 37 TERTIARY_COMP_
38 TERT_TEMPERATURE_
39 ALARM_SUM 8 8 40 DEV_ALARM 4 4 41 LOGBOOK1_RESET 42 LOGBOOK1_EVENT 43 LOGBOOK2_RESET 44 LOGBOOK2_EVENT 45 LOGBOOK_CONFIG[16] 16 46 TEST_1 58 TEST_13 59 CALIB_SENSOR_
60 MATRIX_TEMP_RANGE 61 SOLUTION_1 62 SOLUTION_2 63 SOLUTION_3 64 SOLUTION_4 65 SOLUTION_5 66 CONCENTRATION_
67 CONCENTRATION_0 68 CONCENTRATION_100 69
70
71
72 E5_LIMIT 4 73 E6_LIMIT 4 74 DISPLAY_RESOLUTION 1 75 PASSCODE_
76 PASSCODE_
77 PASSCODE_SERVICE 2 78 ERROR_CONFIG 4 79 CONFIGURATION 4 80 TRANSMITTER_TIME 6 81 SOFTDWN_TEST 1
Table 3.17 View Object for Transducer Block DO202
Realtive Index
0 BLK_DATA 1 ST_REV 2 2 2 2 2 TAG_DESC[32] 3 STRATEGY 2 4 ALERT_KEY 1 5 MODE_BLK 4 4 6 BLOCK_ERR 2 2 7 UPDATE_EVT 8 BLOCK_ALM
Parameters Mnemonic
COEFF
TEMPERATURE
METHOD
COEFF
CONST
MEASUREMENT
CONCENTRATION_ TABLE_LOW CONCENTRATION_ TABLE_MID CONCENTRATION_ TABLE_HIGH
MAINTENANCE
COMMISSIONING
TOTALS (# BYTES) 43 17 54 96
Parameters Mnemonic
View1View2View3View
4
View1View2View3View
Table 3.17 View Object for Transducer Block DO202
Foundation Fieldbus 3-17
(continues)
Realtive
4
Index
9 TRANSDUCER_
4
10 TRANSDUCER_TYPE 2 2 2 2 11 XD_ERROR 1 1 12 COLLECTION_
4
13 14 PRIMARY_VALUE 5 5 15 PRIMARY_VALUE_
1
16
4
17 18 SAMPLE_CAL 19 ZERO_CURRENT 4 20 SENSITIVITY 4 21 22 AMP_SPAN_
23 AMP_ZERO_
24 SALINITY 4 25 BAR_PRESSURE 4 26 27
28
29 SECONDARY_VALUE 5 5 30 SECONDARY_
31
1
32 SENSOR_TEMP_
33 34 TEMP_SENSOR_CAL 4 35 SENSOR_CURRENT 5 5 36 37 38 ZERO_CAL 39 RESERVED1 40 RESERVED2 41 RESERVED3 42
2
43 PASSCODE_
2
44 PASSCODE_
45 PASSCODE_SERVICE 46 LOGBOOK1_RESET 47 LOGBOOK1_EVENT 48 LOGBOOK2_RESET 49 LOGBOOK2_EVENT 50 51 CALL_MAINT_TIME_
52 CALL_MAINT_TIME_
4
53 ERROR_CONFIG 4 54 CONFIGURATION 4 55 ALARM_SUM 8 8 56 DEV_ALARM 4 4 57 TEST_1 69 TEST_13 70 SOFTDWN_TEST 1
Parameters Mnemonic
DIRECTORY[2]
DIRECTORY[7] PRIMARY_VALUE_TYPE
RANGE PRIMARY_VALUE_UNIT SENSOR_TYPE_OXYGEN
AMP_STABILIZE_TIME
STABILIZE _VALUE
STABILIZE_VALUE
BAR_PRESSURE_UNIT PERCENT_SATURATION_ PRESSURE CHLORINE_ CALIBRATION_RANGES
VALUE_UNIT SENSOR_TEMP_COMP
MAN_VALUE SENSOR_TYPE_TEMP
PERCENT_SATURATION ZERO_CURRENT_LIMIT
TRANSMITTER_TIME[6]
MAINTENANCE
COMMISSIONING
LOGBOOK_CONFIG[25]
RELOAD
COUNTDOWN
TOTALS (# BYTES) 43 12 50 102
View1View2View3View
3-17
4
2
11
2 2
4
4
4
2 4
2
1 4
2
5 5 4
6
25 1
1
IM 12A00A01-61E
3-18 Foundation Fieldbus
3-3-6-4 Function Block Parameters
Function block parameters can be read from the host or can be set. For a list of the parameters of blocks held by the EXA, refer to “3-6 List of parameters for each block of the EXA”. The following is a list of impor­tant parameters with a guide how to set them.
MODE_BLK:
This mode parameter is very important as it gives the state of the block. In O/S (Out_Of_Service) mode the block is out of operation. In this mode it is allowed to update parameters. Manual mode gives the operator the possibility to manually update a selected number of parameters (values, scaling e.g.) in order to test the system. In automatic mode the function block is executed and block parameters are automatically updated. Under normal operating circumstances, set the Auto mode for normal operation. Auto mode is the factory default.
Note: The actual mode is changed by setting the target mode. When the resource block mode is set to
OOS all function blocks in the VFD are set to OOS mode.
CHANNEL:
Transducer blocks convert raw signals into process values. The values are assigned to channels. For the EXA 202 three or four channels are available.
PH202
1: pH, 2: Temperature, 3: ORP/rH
Channel value unit
1 primary_value primary_value_range.unit 2 secondary_value secondary_value_unit 3 tertiary_value tertiary_value_range.unit
SC202
1: Conductivity/Resistivity, 2: Temperature, 3: Second Conductivity/Resistivity, 4: Concentration
Channel value unit
1 primary_value primary_value_range.units 2 secondary_value secondary_value_unit 3 tertiary_value primary_value_range.units 4 concentration always %
ISC202
1: Conductivity, 2: Temperature, 3: Second Conductivity, 4: Concentration
Channel value unit
1 primary_value primary_value_range.units 2 secondary_value secondary_value_unit 3 tertiary_value primary_value_range.units 4 concentration always %
IM 12A00A01-61E
Foundation Fieldbus 3-19
3-19
DO202
1: Dissolved Oxygen, 2: Temperature, 3: Percent Saturation, 4: Sensor Current
Channel Value Unit
1 primary_value primary_value_range.units 2 secondary_value secondary_value_unit 3 percent_saturation % 4 sensor_current nA
XD_SCALE/OUT_SCALE:
Scaling information is used for two purposes. Display devices need to know the range for bar graphs and trending, as well as the units code. Control blocks need to know the range to use internally as percent of span, so that the tuning constants may remain dimensionless. This is converted back to a number with units by using the range of OUT_SCALE. The AI block has the parameter XD_SCALE to define the units expect­ed from the transducer.
Transducer scaling (XD_SCALE) is applied to the value from the channel to produce the FIELD_VAL in per­cent. The XD_SCALE units code must match the channel units code. The EXA transmitter does this automatically when the Service Codes are changed. See table 3.18 to 3.21 for the Service codes and their results. If L_TYPE is set to Indirect or Ind Sqr Root, OUT_SCALE determines the conversion from FIELD_VAL to the output. PV and OUT always have identical scaling. OUT_SCALE provides scaling for PV. The PV is always the value that the block will place in OUT if the mode is Auto.
Table 3.18 Unit Index by XD_SCALE PH202
Channel FF parameters Service code XD_SCALE.UNITS
1 - SC01 (set to 0) pH 2 2029 SC11 (set to 0) ºC 2 2029 SC11 (set to 1) ºF 3 - SC01 (set to 1) mV 3 - SC01 (set to 0), SC02 (set to 1) mV 3 - SC01 (set to 0), SC02 (set to 2) rH
Table 3.19 Unit Index by XD_SCALE SC202
Channel FF parameters Service code XD_SCALE.UNITS
2 2023 SC11 (set to 0) °C (1001) 2 2023 SC11 (set to 1) °F (1002) 1, 3 - SC01 (set to 1) Ω•cm (1295) 1, 3 - SC01 (set to 0) S/cm (1594) 4 - Default % % (1342)
Table 3.20 Unit Index by XD_SCALE ISC202
Channel FF parameters Service code XD_SCALE.UNITS
2 2023 SC11 (set to 0) °C (1001) 2 2023 SC11 (set to 1) °F (1002) 1, 3 - Default S/cm S/cm (1594) 4 - Default % % (1342)
IM 12A00A01-61E
3-20 Foundation Fieldbus
0%, 0ºC
100%, 212ºF
0%, 32ºF
CHANNEL
PV / OUT
OUT_SCALEXD_SCALE
100%, 100ºC
Table 3.21 Unit Index by XD_SCALE DO202
Channel FF parameters Service code XD_SCALE.UNITS
2 FF2030 SC11 (set to 0) °C (1001) 2 FF2030 SC11 (set to 1) °F (1002) 1 FF2016 SC56 (set to 0) ppm (1423) 1 FF2016 SC56 (set to 1) ppb (1424) 1 FF2016 SC56 (set to 2) % (1342) 3 - Default % % (1342) 4 - Default nA nA (1213)
L_TYPE:
Specifies the operation function of the AI block. If set to “Direct”, the input delivered to CHANNEL is directly reflected on OUT. If set to “Indirect”, scaling by XD_SCALE and OUT_SCALE is carried out and is reflected on OUT. If set to “Indirect SQRT”, after scaling by XD_SCALE, the square root is extracted and the value scaled by OUT_SCALE is reflected on OUT.
Example: Channel range is defined as 0 to 100 °C but °F units is required for HOST display. Set the following parameters:
XD_SCALE: EU@0% = 0 °C EU@100% = 100 °C Unit = °C Decimal point = 2 OUT_SCALE: EU@0% = 32 °F EU@100% = 212 °F Unit = °F Decimal point = 2
Figure 3.11 Scaling applied to temperature conversion.
PV_FTIME:
Sets the time constant of the damping function within AI block (primary delay) in seconds.
Alarm Priority:
Indicates the priority of the process alarm. If a value of 3 or greater is set, an alarm is transmitted. The factory default is 0. Four types of alarm can be set: HI_PRI, HI_HI_PRI, LO_PRI, and LO_LO_PRI.
Alarm Threshold:
Sets the threshold at which a process alarm is generated. The factory default setting is a value that does not generate an alarm. Four types of alarm can be set: HI_LIM, HI_HI_LIM, LO_LIM, and LO_LO_LIM.
IM 12A00A01-61E
Equations:
Simulate Convert Cutoff Filter
CHANNEL SIMULATE L_TYPE LOW_CUT PV_FTIME
XD_SCALE Output OUT_SCALE
Mode
Alarms
HI/LO
PV
OUT
FIELD_VAL
Foundation Fieldbus 3-21
3-21
FIELD_VAL = 100 [XD_SCALE]
(channel value - EU@0%)
(EU@100% - EU@0%)
Direct: PV = channel value Indirect: PV = EU@0% + (EU@100% - EU@0%) [OUT_SCALE]
FIELD_VAL
100
Ind Sqr Root: PV = EU@0% + (EU@100% - EU@0%) [OUT_SCALE]
(FIELD_VAL)
100
IM 12A00A01-61E
3-22 Foundation Fieldbus

3-4 In-process operation

This chapter describes the procedure performed when changing the operation of the function block of the EXA in process.

3-4-1 Mode Transition

When the function block mode is changed to Out_Of_Service, the function block pauses and a block alarm is issued.
When the function block mode is changed to Manual, the function block suspends updating of output val­ues. In this case alone, it is possible to write a value to the OUT parameter of the block for output. Note that no parameter status can be changed.

3-4-2 Generation of Alarm

3-4-2-1 Indication of Alarm
Figure 3.12 Error Identification on Indicator
3-4-2-2 Alarms and Events
Following alarm or event can be reported by EXA as an alert if allowed.
Analog Alerts
(Generated when a process value exceeds threshold)
By AI1 Block Hi-Hi Alarm, Hi Alarm, Low Alarm, Low-Low Alarm By AI2 Block Hi-Hi Alarm, Hi Alarm, Low Alarm, Low-Low Alarm By AI3 Block Hi-Hi Alarm, Hi Alarm, Low Alarm, Low-Low Alarm
Discrets Alerts
(Generated when an abnormal condition is detected)
By Resource Block Block Alarm, Write Alarm By Transducer Block Block Alarm By AI1 Block Block Alarm By AI2 Block Block Alarm By AI3 Block Block Alarm
Update Alerts
(Generated when a important (restorable) parameter is updated)
By Resource Block Update Event By Transducer Block Update Event By AI1 Block Update Event By AI2 Block Update Event By AI3 Block Update Event
IM 12A00A01-61E
An alert has following structure:
Simulation Enable
Not used
1
2
off on
Table 3.22 Alert Object
Subindex
Analog Alert
1 1 1 Block Index Index of block from which alert is generated 2 2 2 Alert Key Alert Key copied from the block 3 3 3 Standard Type Type of the alert 4 4 4 Mfr Type Alert Name identified by manufacturer specific DD 5 5 5 Message Type Reason of alert notification 6 6 6 Priority Priority of the alarm 7 7 7 Time Stamp Time when this alert is first detected 8 8 Subcode Enumerated cause of this alert 9 9 Value Value of referenced data 10 10 Relative Index Relative index of referenced data
11 11 9 Unit Index Unit code of referenced data
Discrete Alert
Update Alert
8 Static Revision Value of static revision (ST_REV) of the block
Parameter Name Explanation
Foundation Fieldbus 3-23
3-23

3-4-3 Simulation Function

The simulation function simulates the input of a function block and lets it operate as if the data was received from the transducer block. It is possible to conduct testing for the downstream function blocks or alarm proc­esses.
A SIMULATE_ENABLE switch is mounted on the FF PCB assembly. This is to prevent the accidental oper­ation of this function. When this is switched on, simulation is enabled. (See Figure 3.13) To initiate the same action from a remote terminal, if REMOTE LOOP TEST SWITCH is written to the SIM_ENABLE_MSG parameter (index 1044) of the resource block, the resulting action is the same as is taken when the above switch is on. Note that this parameter value is lost when the power is turned OFF. In simulation enabled sta­tus, an alarm is generated from the resource block, and other device alarms will be masked; for this reason the simulation must be disabled immediately after using this function.
FF PCB assembly
Figure 3.13 SIMULATE_ENABLE Switch Position
IM 12A00A01-61E
3-24 Foundation Fieldbus
The SIMULATE parameter of AI block consists of the elements listed in Table 3.23 below.
Table 3.23 SIMULATE Parameter
Sub-index Parameters Description
1 Simulate Status Sets the data status to be simulated. 2 Simulate Value Sets the value of the data to be simulated.
3 Transducer Status Displays the data status from the transducer block. It cannot be changed. 4 Transducer Value Displays the data value from the transducer block. It cannot be changed. 5 Simulate En/Disable Controls the simulation function of this block.
1: Simulation disabled (standard) 2: Simulation started
When Simulate En/Disable in Table 3.23 above is set to 2, the applicable function block uses the simula­tion value set in this parameter instead of the data from the transducer block. This setting can be used for propagation of the status to the trailing blocks, generation of a process alarm, and as an operation test for trailing blocks.
IM 12A00A01-61E
Foundation Fieldbus 3-25

3-5 Device status

Device setting status and failures of EXA are indicated by using parameter DEVICE_STATUS_1, DEVICE_STATUS_2 and DEVICE_STATUS_3 (index 1045, 1046 and 1047) in Resource Block.
Table 3.24 Contents of DEVICE_STATUS_1, DEVICE_STATUS_2 and DEVICE_STATUS_3
3-25
DEVICE_STATUS_1
Hexadecimal Display through DD
0x80000000 0x40000000 0x20000000 0x10000000 0x08000000 0x04000000 0x02000000 0x01000000 0x00800000 Sim.enable Jmpr On 0x00400000 RB in O/S mode 0x00200000 0x00100000 0x00080000 Fbus EEPROM error 0x00040000 0x00020000 0x00010000 0x00008000 Link Obj.1 not open 0x00004000 Link Obj.2 not open 0x00002000 Link Obj.3 not open 0x00001000 Link Obj.4 not open 0x00000800 Link Obj.5 not open 0x00000400 Link Obj.6 not open 0x00000200 Link Obj.7 not open 0x00000100 Link Obj.8 not open 0x00000080 Link Obj.9 not open 0x00000040 Link Obj.10 not open 0x00000020 Link Obj.11 not open 0x00000010 Link Obj.12 not open 0x00000008 Link Obj.13 not open 0x00000004 Link Obj.14 not open 0x00000002 Link Obj.15 not open 0x00000001 Link Obj.16 not open
DEVICE_STATUS_3
Hexadecimal Display through DD
0x80000000 0x40000000 0x20000000 0x10000000 0x08000000 Transducer Block is in O/S mode 0x04000000 0x02000000 0x01000000 0x00800000 0x00400000 0x00200000 0x00100000 0x00080000 0x00040000 0x00020000 Simulation is enabled in AI3 Function Block 0x00010000 AI3 Function Block is in Manual mode 0x00008000 AI3 Function Block is inO/S mode 0x00004000 Simulation is enabled in AI2 Function Block 0x00002000 AI2 Function Block is in Manual mode 0x00001000 AI2 Function Block is in O/S mode 0x00000800 AI1 Function Block is not scheduled 0x00000400 Simulation is enabled in AI1 Function Block 0x00000200 AI1 Function Block is in Manual mode 0x00000100 AI1 Function Block is in O/S mode 0x00000080 0x00000040 0x00000020 0x00000010 0x00000008 0x00000004 0x00000002 0x00000001
IM 12A00A01-61E
3-26 Foundation Fieldbus
DEVICE_STATUS_2 PH202
Hexadecimal Display through DD
0x80000000 0x40000000 0x20000000 0x10000000 0x08000000 0x04000000 0x02000000 0x01000000 0x00800000 0x00400000 0x00200000 0x00100000 0x00080000 FF interface checksum error 0x00040000 EXA checksum error (E21) 0x00020000 Internal communication failure 0x00010000 FF interface eeprom failure 0x00008000 EXA eeprom failure (E20) 0x00004000 mismatch between FF- and EXA parameter 0x00002000 0x00001000 0x00000800 0x00000400 0x00000200 0x00000100 calibration timer expired (E16) 0x00000080 reference impedance exceeds low limit (E4.2) 0x00000040 reference impedance exceeds high limit (E5.2) 0x00000020 glass impedance exceeds low limit (E4.1) 0x00000010 glass impedance exceeds high limit (E5.1) 0x00000008 temperature sensor shorted (E8) 0x00000004 temperature sensor open (E7) 0x00000002 tertiary value exceeds limits (E12) 0x00000001 primary value exceeds limits (E9)
DEVICE_STATUS_2 SC202
Hexadecimal Display through DD
0x80000000 0x40000000 0x20000000 0x10000000 0x08000000 0x04000000 0x02000000 0x01000000 0x00800000 0x00400000 0x00200000 0x00100000 0x00080000 FF interface checksum error 0x00040000 EXA checksum error (E21) 0x00020000 Hart communication failure 0x00010000 FF interface eeprom failure 0x00008000 EXA eeprom failure (E20) 0x00004000 mismatch between FF- and EXA parameter 0x00002000 0x00001000 0x00000800 0x00000400 0x00000200 0x00000100 matrix error (E4) 0x00000080 concentration table error (E18) 0x00000040 conductivity exceeds usp limit (E13) 0x00000020 polarization detected (E1) 0x00000010 temperature compensation error (E2) 0x00000008 temperature sensor shorted (E8) 0x00000004 temperature sensor open (E7) 0x00000002 conductivity exceeds low limit (E6) 0x00000001 conductivity exceeds high limit (E5)
IM 12A00A01-61E
Foundation Fieldbus 3-27
3-27
DEVICE_STATUS_2 ISC202
Hexadecimal Display through DD
0x80000000 0x40000000 0x20000000 0x10000000 0x08000000 0x04000000 0x02000000 0x01000000 0x00800000 0x00400000 0x00200000 0x00100000 0x00080000 FF interface checksum error 0x00040000 EXA checksum error (E21) 0x00020000 Hart communication failure 0x00010000 FF interface eeprom failure 0x00008000 EXA eeprom failure (E20) 0x00004000 mismatch between FF- and EXA parameter 0x00002000 0x00001000 0x00000800 0x00000400 0x00000200 0x00000100 matrix error (E4) 0x00000080 concentration table error (E18) 0x00000010 temperature compensation error (E2) 0x00000008 temperature sensor shorted (E8) 0x00000004 temperature sensor open (E7) 0x00000002 conductivity exceeds low limit (E6) 0x00000001 conductivity exceeds high limit (E5)
DEVICE_STATUS_2 DO202
Hexadecimal Display through DD
0x80000000 0x40000000 0x20000000 0x10000000 0x08000000 0x04000000 0x02000000 0x01000000 0x00800000 0x00400000 0x00200000 0x00100000 0x00080000 FF interface checksum error 0x00040000 EXA checksum error (E21) 0x00020000 Internal communication failure 0x00010000 FF interface eeprom failure 0x00008000 EXA eeprom failure (E20) 0x00004000 mismatch between FF- and EXA parameter 0x00002000 0x00001000 0x00000800 0x00000400 0x00000200 0x00000100 0x00000080 0x00000040 Call for maintenance (E16) 0x00000020 Sensor current abnormal (E9) 0x00000010 temperature sensor shorted (E8) 0x00000008 temperature sensor open (E7) 0x00000004 0x00000002 Zero out of limits (E2) 0x00000001
IM 12A00A01-61E
3-28 Foundation Fieldbus

3-6 List of parameters for each block of the EXA

Note: The Write Mode column contains the modes in which each parameter is write enabled. O/S: Write enabled in O/S mode. MAN: Write enabled in Man mode and O/S mode. AUTO: Write enabled in Auto mode, Man mode, and O/S mode.

3-6-1 Resource Block

Relative Index Index
0 1000 Block Header TAG:“RS” Block Tag
1 1001 ST_REV The revision level of the static data associated with the resource
2 1002 TAG_DESC Null AUTO The user description of the intended application of the block. 3 1003 STRATEGY 1 AUTO The strategy field can be used to identify grouping of blocks.
4 1004 ALERT_KEY 1 AUTO The identification number of the plant unit. This information may
5 1005 MODE_BLK AUTO AUTO The actual, target, permitted, and normal modes of the block. 6 1006 BLOCK_ERR This parameter reflects the error status associated with the
7 1007 RS_STATE State of the resource block state machine. 8 1008 TEST_RW AUTO Read/write test parameter-used only for conformance testing and
9 1009 DD_
10 1010 MANUFAC_ID 0x00594543 Manufacturer identification number-used by an interface device
11 1011 DEV_TYPE PH202: 0x0830
12 1012 DEV_REV 3 Manufacturer revision number associated with the resource-used
13 1013 DD_REV 1 Revision of the DD associated with the resource-used by an
14 1014 GRANT_DENY 0 AUTO Options for controlling access of host computer and local control
15 1015 HARD_TYPES Scalar input The types of hardware available as channel numbers.
16 1016 RESTART Allows a manual restart to be initiated. Several degrees of restart
17 1017 FEATURES Soft write lock
18 1018 FEATURE_SEL Soft write lock
19 1019 CYCLE_TYPE Scheduled – 20 1020 CYCLE_SEL Scheduled AUTO Used to select the block execution method for this resource. 21 1021 MIN_CYCLE_T 3200 (100ms) Time duration of the shortest cycle interval of which the resource
22 1022 MEMORY_SIZE 0 Available configuration memory in the empty resource. To be
23 1023 NV_CYCLE_T 0 Interval between writing copies of NV parameters to non-volatile
24 1024 FREE_SPACE 0 Percent of memory available for further configuration. EXA has
Parameter Name Factory Default
Null String identifying the tag of the resource which contains the
RESOURCE
SC202: 0x0831 ISC202: 0x0832 DO202: 0x0833
supported Report supported
supported Report supported
Write Mode Explanation
Information on this block such as Block Tag, DD Revision,
= O/S
Manufacturer’s model number associated with the resource-used
Used to show supported resource block options.
AUTO Used to select resource block options defined in FEATURES.
Execution Time etc.
block. The revision value is incremented each time a static parameter alue in this block is changed.
This data is not checked or processed by the block.
be used in the host for sorting alarms, etc.
hardware or software components associated with a block. It is a bit string, so that multiple errors may be shown.
simulation.
Device Description for this resource.
to locate the DD file for the resource.
by interface devices to locate the DD file for the resource.
by an interface device to locate the DD file for the resource.
interface device to locate the DD file for the resource.
panels to operating, tuning and alarm parameters of the block.
bit0: Scalar input bit1: Scalar outputbit1: Scalar output bit2: Discrete inputbit2: Discrete input bit3: Discrete outputbit3: Discrete output
are possible. They are 1: Run, 2: Restart resource, 3: Restart with initial value specified in FF functional spec. (*1), and 4: Restart processor. *1: FF-891 Foundation TM Specification Function Block*1: FF-891 Foundation TM Specification Function Block
Application Process Part 2.
bit0: Scheduled bit1: Event driven bit2: Manufacturer specified Identifies the block execution methods available for this resource.
is capable.
checked before attempting a download.
memory. Zero means never.
zero which means a preconfigured resource.
IM 12A00A01-61E
Foundation Fieldbus 3-29
3-29
Relative Index Index
25 1025 FREE_TIME 0 Percent of the block processing time that is free to process
26 1026 SHED_RCAS 640000 (2S) AUTO Time duration at which to give up on computer writes to function
27 1027 SHED_ROUT 640000 (2S) AUTO Time duration at which to give up on computer writes to function
28 1028 FAULT_STATE 1 Condition set by loss of communication to an output block, failure
29 1029 SET_FSTATE 1 AUTO 30 1030 CLR_FSTATE 1 AUTO Writing a Clear to this parameter will clear the device fail-safe
31 1031 MAX_NOTIFY 3 Maximum number of unconfirmed notify messages possible. 32 1032 LIM_NOTIFY 3 AUTO Maximum number of unconfirmed alert notify messages allowed. 33 1033 CONFIRM_TIM 640000 (2S) AUTO The minimum time between retries of alert reports. 34 1034 WRITE_LOCK Not locked AUTO If set, no writes from anywhere are allowed, except to clear
35 1035 UPDATE_EVT This alert is generated by any change to the static data. 36 1036 BLOCK_ALM The block alarm is used for all configuration, hardware,
37 1037 ALARM_SUM Enable The current alert status, unacknowledged states, unreported
38 1038 ACK_OPTION 0xFFFF AUTO 39 1039 WRITE_PRI 0 AUTO Priority of the alarm generated by clearing the write lock. 40 1040 WRITE_ALM This alert is generated if the write lock parameter is cleared. 41 1041 ITK_VER 4 Version number of interoperability test by Fieldbus Foundation
42 1042 SOFT_REV EXA software revision number. 43 1043 SOFT_DESC Yokogawa internal use. 44 1044 SIM_ENABLE_
45 1045 DEVICE_
46 1046 DEVICE_
47 1047 DEVICE_
48 1048 DEVICE_
49 1049 DEVICE_
50 1050 DEVICE_
51 1051 DEVICE_
52 1052 DEVICE_
53 1053 SOFTDWN_
54 1054 SOFTDWN_
55 1055 SOFTDWN_
56 1056 SOFTDWN_
57 1057 SOFTDWN_
58 1058 SOFTDWN_
Parameter Name Factory Default
Null AUTO Software switch for simulation function.
MSG
0 Device status (VCR setting etc.)
STATUS_1
0 Device status (failure or setting error etc.)
STATUS_2
0 Device status (function block setting.)
STATUS_3
0 Not used.
STATUS_4
0 Not used.
STATUS_5
0 Not used.
STATUS_6
0 Not used.
STATUS_7
0 Not used.
STATUS_8
1 Not used.
PROTECT
1 Not used.
FORMAT
0 Not used.
COUNT
0 Not used.
ACT_AREA
1, 0, 0, 0, 0, 0,
MOD_REV
ERROR
0, 0, 0 0 Not used.
Write Mode Explanation
additional blocks. EXA does not support this.
block RCas locations. Supported only with PID function.
block ROut locations. Supported only with PID function.
promoted to an output block or a physical contact. When fail-safe condition is set, Then output function blocks will perform their FSAFE actions. Allows the fail-safe condition to be manually initiated by selecting Set.
state if the field condition, if any, has cleared.
WRITE_LOCK. Block inputs will continue to be updated
connection failure or system problems in the block. The cause of the alert is entered in the subcode field. The first alert to become active will set the Active status in the Status attribute. As soon as the Unreported status is cleared by the alert reporting task, another block alert may be reported without clearing the Active status, if the subcode has changed.
states, and disabled states of the alarms associated with the function block.
applied to EXA.
Not used.
IM 12A00A01-61E
3-30 Foundation Fieldbus

3-6-2 Analog input Block

Relative Parameter Factory Write Index Name Default Mode Explanation
0 Block Header TAG: “AI1” or Block Tag Information on this block such as Block Tag, DD “AI2” or “AI3” = O/S Revision, Execution Time etc. 1 ST_REV The revision level of the static data associated with the function block. The
revision value will be incremented each time a static parameter value in the block
is changed. 2 TAG_DESC (blank) AUTO The user description of the intended application of the block. 3 STRATEGY 1 AUTO The strategy field can be used to identify grouping of blocks. This data is not
checked or processed by the block. 4 ALERT_KEY 1 AUTO The identification number of the plant unit. This information may be used in the
host for sorting alarms, etc. 5 MODE_BLK AUTO AUTO The actual, target, permitted, and normal modes of the block. 6 BLOCK_ERR This parameter reflects the error status associated with the hardware or software
components associated with a block. It is a bit string, so that multiple errors may
be shown. 7 PV Either the primary analog value for use in executing the function, or a process
value associated with it. May also be calculated from the READBACK value of an
AO block. 8 OUT Value= The primary analog value calculated as a result of MAN executing the function. 9 SIMULATE Disable AUTO Allows the transducer analog input or output to the block to be manually supplied
when simulate is enabled. When simulation is disabled, the simulate value and
status track the actual value and status. 10 XD_SCALE O/S The high and low scale values, engineering units code, and number of digits to the right of the decimal point used with the value
obtained from the transducer for a specified channel. Refer to Table 3.18 to 3.21
for the unit available. 11 OUT_SCALE O/S The high and low scale values, engineering units code, and number of digits to the right of the decimal point to be used in
displaying the OUT parameter and parameters which have the same scaling as
OUT. 12 GRANT_DENY 0 AUTO Options for controlling access of host computers and local control panels to
operating, tuning and alarm parameters of the block. 13 IO_OPTS 0 O/S Options which the user may select to alter input and output block processing
bit6: Low cutoff 14 STATUS_OPTS Propagate Fault O/S Options which the user may select in the block Forward processing of status 15 CHANNEL AI1: 1 O/S The number of the logical hardware channel that is AI2: 2 connected to this I/O block. This information defines the transducer to be AI3: 3 used going to or from the physical world. 16 L_TYPE Direct (1) MAN Deterines if the values passed by the transducer time of order block to the AI
block may be used directly (Direct) or if the value is in different units and must be
converted linearly (Indirect), or with square root (Ind Sqr Root), using the input
range defined by the transducer and the associated output range. 17 LOW_CUT Linear: 0% AUTO Sets low cut point of output. This low cut value become available by setting “Low cutoff“ to “IO_OPTS“. 18 PV_FTIME 2sec AUTO Time constant of a single exponential filter for the PV, in seconds. 19 FIELD_VAL Raw value of the field device in percent of thePV range, with a status reflecting
the Transducer condition, before signal characterization (L_TYPE) or filtering
(PV_FTIME). 20 UPDATE_EVT This alert is generated by any change to the static data. 21 BLOCK_ALM The block alarm is used for all configuration, hardware, connection failure or
system problems in the block. The cause of the alert is entered in the subcode
field. The first alert to become active will set the Active status in the Status
attribute. As soon as the Unreported status is cleared by the alert reporting task,
another block alert may be reported without clearing the Active status, if the
subcode has changed.
IM 12A00A01-61E
Foundation Fieldbus 3-31
Relative Parameter Factory Write
Index Name Default Mode Explanation
22 ALARM_SUM The current alert status, unacknowledged states, unreported states, and
disabled states of the alarms associated with the function block.
23 ACK_OPTION 0xFFFF AUTO Selection of whether alarms associated with the block will be automatically
acknowledged.
24 ALARM_HYS 0.5% AUTO Amount the PV must return within the alarm limits before the alarm condition
clears. Alarm Hysteresis is expressed as a percent of the PV span. 25 HI_HI_PRI 0 AUTO Priority of the high high alarm. 26 HI_HI_LIM +INF AUTO The setting for high high alarm in engineering units. 27 HI_PRI 0 AUTO Priority of the high alarm. 28 HI_LIM +INF AUTO The setting for high alarm in engineering units. 29 LO_PRI 0 AUTO Priority of the low alarm. 30 LO_LIM -INF AUTO The setting for the low alarm in engineering units. 31 LO_LO_PRI 0 AUTO Priority of the low low alarm. 32 LO_LO_LIM -INF AUTO The setting of the low low alarm in engineering units. 33 HI_HI_ALM The status for high high alarm and its associated time stamp. 34 HI_ALM The status for high alarm and its associated time stamp. 35 LO_ALM The status of the low alarm and its associated time stamp. 36 LO_LO_ALM The status of the low low alarm and its associated time stamp.
3-31
IM 12A00A01-61E
3-32 Foundation Fieldbus

3-6-3 Transducer Block 3-6-3-1 Transducer Block PH202

Index Parameter name Factory Valid Description Default Range
2000 BLOCK HEADER TAG: “TB” General information about the function block 2001 ST_REV - - The revision level of the static data associated with the function
block. The revision value will be incremented each time a static
parameter value in the block is changed. 2002 TAG_DESC “ “ The user description of the intended application of the block. 2003 STRATEGY 1 The strategy field can be used to identify grouping of blocks. This
data is not checked or processed by the block. 2004 ALERT_KEY 1 The identification number of the plant unit. This information may be
used in the host for sorting alarms, etc. 2005 MODE_BLK AUTO The actual, target, permitted, and normal modes of the block 2006 BLOCK_ERR - This parameter reflects the error status associated with a block. It
is a bit-string, so that multiple errors can be shown. 2007 UPDATE_EVT - The alert is generated by any change to the static data. 2008 BLOCK_ALM - The block alarm is used for all configuration error, hardware
connection failure or system problems in the block. The cause of
the alert is entered in the subcode field. The first alert to become
active will set Active status in Status attribute. 2009 TRANSDUCER_DIRECTORY A directory that specifies the number and starting indices of the
transducers. 2010 TRANSDUCER_TYPE Standard PH, ORP transducer block. pH/ORP 2011 XD_ERROR - The error code in transducer:
- No failure,
- Electronics failure,
- I/O failure,
- Mechanical failure 2012 COLLECTION_DIRECTORY - A directory that specifies the number, starting indices and DD item
IDs of the data collection in each transducer within a transducer
block. 2013 PRIMARY_VALUE_TYPE PH pH, None Type of measurement represented by primary value. 2014 PRIMARY_VALUE - -2 to 16 Primary value of the instrument is pH. 2015 PRIMARY_VALUE_RANGE -2.00 to 16.00 -2 to 16 The pH range of the instrument. (can not change) 2016 SENSOR_TYPE_PH pH/ORP pH/ORP PH is measured with a glass- and a reference electrode. sensor sensor Redox is measured with a metal- and a reference electrode. When
a glass, metal and ref. Electrode are combined one can measure
pH and Redox simultaneously. 2017 SENSOR_MV - - The output of the electrode in mV 2018 CAL_POINT_HI 16 -2 to 16 Highest calibration point 2019 CAL_POINT_LO -2 -2 to 16 Lowest calibration point 2020 CAL_MIN_SPAN 1 1 to 18 Minimum span between two calibration points 2021 SLOPE 100 70 to 110% Sensitivity of the glass electrode 2022 SLOPE_UNIT % % Slope is presented as percentage of theoretical slope (59,16 mV/pH equals 100%) 2023 ZERO 0 -120 to 120 mV, A balanced electrode system gives 0 mV output at pH=7. mV -2 to 16 pH Zero (asymmetry potential) indicates the offset in mV. 2024 ZERO_UNIT mV mV, pH As an alternative to Asymmetry Potential, the Zero point can be
used to define and calibrate the EXA pH transmitter conform to the
DIN standard for instruments No. IEC 746-2. 2025 ISOPOTENTIAL_PH 7 -2 to 16 pH This is the pH value at which the sensor has the same mV output
independent of the process temperature. 2026 SENSOR_CAL_METHOD 0 1 point, not used 2 point 2027 SENSOR_CAL_DATE - till 2104 Date the sensor was last calibrated. 2028 SECONDARY_VALUE - -30 to 140°C , Temperature value
-20 to 280 °F
IM 12A00A01-61E
Foundation Fieldbus 3-33
Index Parameter name Factory Valid Description Default Range
2029 SECONDARY_VALUE_UNIT °C °C, °F Temperature unit 2030 SENSOR_TEMP_COMP automatic Off, manual, Select off when no temperature compensation is required. automatic Select manual when no temperature element is available and the
temperature is stable and select auto when a temperature element
is available pH: man + auto orp: off + auto 2031 SENSOR_TEMP_MAN_VALUE 25 -30 to 140°C, manual temperature value
-20 to 280°F 2032 SENSOR_TYPE_TEMP Pt1000 Pt1000, Pt100, Temperature element used: 5k1, 3kBalco, 8k55, 350, NTC10k, 6k8 2033 SENSOR_CONNECTION 2 2 Only 2-wire connections supported _TEMP 2034 TERTIARY_VALUE_TYPE None None, ORP, rH When a metal electrode is used in combination with a glass and
reference electrode one has the possibility to measure a tertiary
value: 2035 TERTIARY_VALUE - -1500 to 1500 mV, The third value is expresed in mV or rH units 0 to 55 rH 2036 TERTIARY_VALUE_RANGE -1500 -1500 to 1500 mV, The high and low range limits for the tertiary value. to 1500 0 to 55 rH 2037 TERTIARY_ZERO 0 -120 to 120mV Redox offset can be set in mV.
2038 GLASS_IMPEDANCE - 1MΩ to 2 GΩ (HI) Value that indicates the state of the glass membrane. 1kΩ to 1MΩ (LO) Depending on the jumper setting the range is set to High or Low. 2039 REFERENCE_IMPEDANCE - 1MΩ to 2 GΩ (HI) Value that indicates the state of the reference system. 1kΩ to 1MΩ (LO) Depending on the jumper setting the range is set to High or Low.
2040 ALARM_SUM - 2041 DEV_ALARM - Device Alarm is used to give the status of the transmitter. Device_
status in RB displays the device Alarms. 2042 LOGBOOK1_RESET - Reset the pointer to the first (oldest) event in logbook 1. 2043 LOGBOOK1_EVENT - Event whereto the pointer is referenced. When parameter is read,
the pointer is increased by one. 2044 LOGBOOK2_RESET - Reset the pointer to the first (oldest) event in logbook 2. 2045 LOGBOOK2_EVENT - Event whereto the pointer is referenced. When parameter is read,
the pointer is increased by one. 2046 LOGBOOK_CONFIG - Per event one can decide whether it should be logged and in which
logbook (1 or 2) it should be logged. 2047 TEST_1 - Service parameter
.... - 2048 to 2058 are, like 2047 and 2059, service parameters
2059 TEST_13 - Service parameter 2060 STABLE_TIME 5 5.0 to 30.0 Stability criteria used during automatic calibration. 2061 STABLE_VALUE 0.02 0.01 to 1.0 Stability criteria used during automatic calibration. 2062 CALL_MAINT_TIME_ COUNTDOWN 250 1 to 250 The remaining days till maintenance is required. 2063 CALL_MAINT_TIME_RELOAD 250 1 to 250 The interval of the maintenance timer. 2064 2065 2066 2067 2068 BUFFER1_ID 4 0 to 9 The ID of buffer one, used during automatic calibration. 2069 BUFFER1 4.0 -2.0 to 16.0 The first user defined calibration buffer. 2070 BUFFER2_ID 7 0 to 9 The ID of buffer two, used during automatic calibration. 2071 BUFFER2 7.0 -2.0 to 16.0 The second user defined calibration buffer. 2072 BUFFER3_ID 9 0 to 9 The ID of buffer three, used during automatic calibration. 2073 BUFFER3 9.0 -2.0 to 16.0 The third user defined calibration buffer. 2074
-100.0 to 100.0
INPUT_1_IMPEDANCE_LO_LIM INPUT_1_IMPEDANCE_HI_LIM INPUT_2_IMPEDANCE_LO_LIM INPUT_2_IMPEDANCE_HI_LIM
TEMPERATURE_COEFFICIENT
1.00E6 100.0 to 1.0E9 The low limit of the input 1 impedance.
1.00E9 100.0 to 1.0E9 The high limit of the input 1 impedance. 100 100.0 to 1.0E9 The low limit of the input 2 impedance.
200000 100.0 to 1.0E9 The high limit of the input 2 impedance.
0.0 -1.0 to 1.0 The temperature coefficient of the primary value.
3-33
IM 12A00A01-61E
3-34 Foundation Fieldbus
Index Parameter name Factory Valid Description Default Range
2075 PASSCODE_MAINTENANCE 0 0, 111, 333, 777, Passcode used to protect the maintenance menu. 888, 123, 957, 331, 546, 847 2076 PASSCODE_COMMISSIONING 0 0, 111, 333, 777, Passcode used to protect the commissioning menu. 888, 123, 957, 331, 546, 847 2077 PASSCODE_SERVICE 0 0, 111, 333, 777, Passcode used to protect the service menu. 888, 123, 957, 331, 546, 847 2078 SAMPLE_PV Process value of the sample taken. 2079 SAMPLE_PV2 Second process value of the sample taken. 2080 SAMPLE_TEMP Temperature of the sample taken. 2081 ERROR_CONFIG* Soft/Hard fail configuration
(checked = hard; unchecked = soft) 2082 CONFIGURATION* Instrument specific configuration 2083 TRANSMITTER_TIME Transmitter time.
2084 TEST_14 Service parameter.
* Enumerated parameters (4 bytes, 32 bits) where each bit can be set individually.
Error_config
0x00002000 (bit 14), “input 1 impedance exceeds low limit (E4.1)” , 0x00001000 (bit 13), “input 1 impedance exceeds high limit (E5.1)” , 0x00000800 (bit 12), “input 2 impedance exceeds low limit (E4.2)” , 0x00000400 (bit 11) “input 2 impedance exceeds high limit (E5.2)” , 0x00000200 (bit 10), “temperature sensor open (E7)” , 0x00000100 (bit 9), “temperature sensor shorted (E8)” , 0x00800000 (bit 24), “primary value exceeds limits (E9)” ,
0x00010000 (bit 17), “calibration timer expired (E16)”
Configuration
0x08000000 (bit 28), “Process compensation enabled” , 0x00800000 (bit 24), “Aspot check enabled” , 0x00400000 (bit 23), “Slope check enabled” , 0x00200000 (bit 22), “Display resolution 0.01 pH” , 0x00100000 (bit 21), “Auto return enabled” , 0x00080000 (bit 20), “Maintenance timer enabled” , 0x00002000 (bit 14), “Input 1 impedance measurement enabled” , 0x00008000 (bit 16), “Input 1 impedance high” , 0x00004000 (bit 15), “Input 1 impedance compensation enabled” , 0x00000400 (bit 11), “Input 2 impedance measurement enabled” , 0x00001000 (bit 13), “Input 2 impedance high” , 0x00000800 (bit 12), “Input 2 impedance compensation enabled” , 0x00000010 (bit 5), “Take sample”
IM 12A00A01-61E
Foundation Fieldbus 3-35
3-35
3-6-3-2 Transducer Block SC202
Index Parameter name Factory Valid Description Default Range
2000 BLOCK HEADER TAG: “TB” General information about thefunction block 2001 ST_REV - - The revision level of the static data associated with the function
block. The revision value will be incremented each time a static
parameter value in the block is changed 2002 TAG_DESC “ “ The user description of the intended application of the block 2003 STRATEGY 1 The strategy field can be used to identify grouping of blocks.
This data is not checked or processed by the block 2004 ALERT_KEY 1 The identification number of the plant unit. This information may be
used in the host for sorting alarms, etc. 2005 MODE_BLK AUTO The actual, target, permitted, and normal modes of the block 2006 BLOCK_ERR - This parameter reflects the error status associated with a block. It is
a bit-string, so that multiple errors can be shown 2007 UPDATE_EVT - The alert is generated by any change to the static data. 2008 BLOCK_ALM - The block alarm is used for all configuration error, hardware
connection failure or system problems in the block. The cause of
the alert is entered in the subcode field. The first alert to become
active will set Active status in Status attribute. 2009 TRANSDUCER_DIRECTORY A directory that specifies the number and starting indices of the transducers. 2010 TRANSDUCER_TYPE Standard conductivity Conductivity transducer block Conductivity 2011 XD_ERROR - The error code in transducer: No failure, Electronics failure, I/O failure, Mechanical failure 2012 COLLECTION_DIRECTORY -
2013 PRIMARY_VALUE_TYPE Conductivity conductivity / Type of measurement represented by primary value
resistivity 2014 PRIMARY_VALUE - 0 to 2 S/cm Primary value of the instrument is Conductivity 2015 PRIMARY_VALUE_RANGE 0 to 2 S/cm 0 to 2 S/cm The range of the instrument (can not change) 2016 SENSOR_CONST 0.1 0.005 to 50 cm-1 The conductivity cell has a specific cell constant determined by the
2017 CAL_POINT_HI 1.999 0 to 2 S/cm Highest calibration point 2018 CAL_POINT_LO 0 0 to 2 S/cm Lower calibration point 2019 CAL_MIN_SPAN 0,0001 > 0,0001 S/cm Minimum span between two calibration points 2020 SENSOR_CAL_METHOD - 1point, 2point not used 2021 SENSOR_CAL_DATE - till 2104 Date the sensor was last calibrated 2022 SECONDARY_VALUE - -20 to 250 ºC , Temperature value 0 to 500ºF 2023 SECONDARY_VALUE_UNIT ºC ºC, ºF Temperature unit 2024 SENSOR_TEMP_COMP automatic automatic 2025 SENSOR_TEMP_MAN_VALUE - - No manual temp. value can be set. Always Automatic 2026 SENSOR_TYPE_TEMP Pt1000 Pt1000, Pt100, Temperature element used: Ni100, Pb36, 8k55 2027 SENSOR_CONNECTION_ 2 2 Only 2-wire connections supported TEMP 2028 SENSOR_TYPE_COND contact 2-electrode, Either 2-electrode or 4-electrode contacting 2-electrode 4-electrode conductivity cell can be selected 2029 SENSOR_OHMS - Actual cell resistance 2030 XD_MAN_ID “ “ 2031 TEMPERATURE_COEFF 2.1 0 to 3.5%/ºC (%/ºF) Process temperature compensation factor 2032 CONCENTRATION - - Conductivity combined with temperature can be directly related to
2033 TERTIARY_VALUE - 0 to 2 S/cm Second compensated conductivity value
A directory that specifies the number, starting indices and DD item Ids of the data collection in each transducer within a transducer block
dimensions of the cell
the concentration. Concentraion is expressed in percentage
IM 12A00A01-61E
3-36 Foundation Fieldbus
Index Parameter name Factory Valid Description Default Range
2034 REFERENCE_ 25 0 to 100 ºC, Conductivity can be process compensated to a standard TEMPERATURE 32 to 212 ºF reference temperature. Mostly 20ºC or 25ºC is used 2035 COMP_METHOD NaCl NaCl, TC, matrix 2036 COMP_MATRIX_SEL HCl HCl cation (0-80 °C) When matrix compensation is required one can make a
Ammonia, (0-80 °C) selection out of 5 predefined matrices and one user
Ammonia, (0-80 °C) definable matrix
HCl (0-5%, 0-60 °C)
NaOH (0-5%, 0-100 °C),
User defined 2037 TERTIARY_COMP_METHOD NaCl NaCl, TC, matrix Method of process temperature compensation for the
2038 TERT_TEMPERATURE 2.1 0 to 3.5%/ºC Process temperature compensation factor for the _COEFF second conductivity value 2039 ALARM_SUM - 2040 DEV_ALARM - Device Alarm is used to give the status of the
2041 LOGBOOK1_RESET Idle Idle, Reset Reset the pointer to the first (oldest) event in logbook 1 2042 LOGBOOK1_EVENT - Event whereto the pointer is referenced. When parameter is
2043 LOGBOOK2_RESET Idle Idle, Reset Reset the pointer to the first (oldest) event in logbook 2 2044 LOGBOOK2_EVENT - Event whereto the pointer is referenced. When parameter is
2045 LOGBOOK_CONFIG[16] - Per event one can decide whether it should be logged and
2046 TEST_1 -
.... - 2047 to 2057 are, like 2046 and 2058, service parameters
2058 TEST_13 ­ 2059 CALIB_SENSOR_CONST 0.1 0.005~50 The calibrated cell constant [1/cm]. 2060 MATRIX_TEMP_RANGE -20~250, 0~500 The temperature values of the user defined matrix.
2061 SOLUTION_1 0~1.999S/cm, 0~999MΩ•cm The conductivity/resistivity values of solution 1. 2062 SOLUTION_2 0~1.999S/cm, 0~999MΩ•cm The conductivity/resistivity values of solution 2. 2063 SOLUTION_3 0~1.999S/cm, 0~999MΩ•cm The conductivity/resistivity values of solution 3. 2064 SOLUTION_4 0~1.999S/cm, 0~999MΩ•cm The conductivity/resistivity values of solution 4. 2065 SOLUTION_5 0~1.999S/cm, 0~999MΩ•cm The conductivity/resistivity values of solution 5.
2066 CONCENTRATION_ disabled 0=Not initialized, Enabling/Disabling of concentration measurement. MEASUREMENT 1=disabled, 2=enabled 2067 CONCENTRATION_0 0 0~100 The concentration [%] corresponding to the 0 percent table value 2068 CONCENTRATION_100 100 0~100 The concentration [%] corresponding to the 100 percent table value
2069 CONC._TABLE_LOW 0~1.999S/cm, 0~999MΩ•cm The first part of the concentration table (0 till 30 percent). 2070 CONC._TABLE_MID 0~1.999S/cm, 0~999MΩ•cm The second part of the concentration table (35 till 65 percent). 2071 CONC._TABLE_HIGH 0~1.999S/cm, 0~999MΩ•cm The third part of the concentration table (70 till 100 percent). 2072 E5_LIMIT 0.25, 4 0~0.5S, 0~10MΩ The upper range accuracy limit of the analyzer [S or Ohm].
2073 E6_LIMIT 1μ, 1M 0~0.5S, 0~10MΩ The lower range accuracy limit of the analyzer [S or Ohm]. 2074 DISPLAY_RESOLUTION 1 0 to 8 Display resolution, 0, Not initialized (can not change)
2075 PASSCODE_MAINT 0 0, 111, 333, 777, 888, Passcode used to protect the maintenance menu
123, 957, 331, 546, 847 2076 PASSCODE_COMM 0 0, 111, 333, 777, 888, Passcode used to protect the commissioning menu.
123, 957, 331, 546, 847 2077 PASSCODE_SERVICE 0 0, 111, 333, 777, 888, Passcode used to protect the service menu.
123, 957, 331, 546, 847 2078 ERROR_CONFIG* 0=soft, 1=hard Soft/Hard fail configuration (checked = hard unchecked = soft). 2079 CONFIGURATION*
2080 TRANSMITTER_TIME Transmitter time.” 2081 TEST_14 Service parameter.
polarization on, USP off, bitnr. setting autoreturn on
Instrument specific configuration. See below for correct
Method of process temp. compensation for the primary value
second conductivity value
analyser. See separate table for error messages
read, the pointer is increased by one.
read, the pointer is increased by one
in which logbook (1 or 2) it should be logged
1,Auto Ranging, 2, x.xxx μS/cm, MΩ•cm, 3, xx.xx μS/cm, MΩ•cm, 4, xxx.x μS/cm, MΩ•cm,
5, x.xxx mS/cm, kΩ•cm, 6, xx.xx mS/cm, kΩ•cm, 7, xxx.x mS/cm, kΩ•cm, 8, xxxx mS/cm, kΩ•cm
IM 12A00A01-61E
* Enumerated parameters (4 bytes, 32 bits) where each bit can be set individually.
ERROR_CONFIG
0x00008000 (bit 15), “polarization detected (E1)” 0x00000800(bit 11), “conductivity exceeds high limit (E5)” 0x00000400 (bit 10), “conductivity exceeds low limit (E6)” 0x00000200 (bit 9), “temperature sensor open (E7)” 0x00000100 (bit 8), “temperature sensor shorted (E8)” 0x00080000 (bit 19), “conductivity exceeds USP limit (E13)”
CONFIGURATION
0x00004000 (bit 14), “Polarization check enabled” 0x00800000 (bit 23), “Auto return enabled” 0x00000001 (bit 0), “USP enabled”
Foundation Fieldbus 3-37
3-37
IM 12A00A01-61E
3-38 Foundation Fieldbus
3-6-3-3 Transducer Block ISC202
Index Parameter name Factory Valid Description Default Range
2000 BLOCK HEADER TAG: “TB” General information about thefunction block 2001 ST_REV - - The revision level of the static data associated with the function
block. The revision value will be incremented each time a static
parameter value in the block is changed 2002 TAG_DESC “ “ The user description of the intended application of the block 2003 STRATEGY 1 The strategy field can be used to identify grouping of blocks.
This data is not checked or processed by the block 2004 ALERT_KEY 1 The identification number of the plant unit. This information may be
used in the host for sorting alarms, etc. 2005 MODE_BLK AUTO The actual, target, permitted, and normal modes of the block 2006 BLOCK_ERR - This parameter reflects the error status associated with a block. It
is a bit-string, so that multiple errors can be shown 2007 UPDATE_EVT - The alert is generated by any change to the static data. 2008 BLOCK_ALM - The block alarm is used for all configuration error, hardware
connection failure or system problems in the block. The cause of
the alert is entered in the subcode field. The first alert to become
active will set Active status in Status attribute. 2009 TRANSDUCER_DIRECTORY A directory that specifies the number and starting indices of the
transducers. 2010 TRANSDUCER Conductivity conductivity Conductivity transducer block _TYPE Transmitter 2011 XD_ERROR - The error code in transducer: No failure, Electronics failure, I/O failure, Mechanical failure 2012 COLLECTION_DIRECTORY - A directory that specifies the number, starting indices and DD item
Ids of the data collection in each transducer within a transducer
block 2013 PRIMARY_VALUE_TYPE conductivity, conductivity Type of measurement represented by primary value 2014 PRIMARY_VALUE - 0 to 2 S/cm Primary value of the instrument is Conductivity 2015 PRIMARY_VALUE_RANGE 0 to 2 S/cm 0 to 2 S/cm The range of the instrument (can not change) 2016 SENSOR_CONST 1.88 0.2 to 19.99 cm-1 The conductivity cell has a specific cell constant determined by the
dimensions of the cell 2017 CAL_POINT_HI 1.999 0 to 2 S/cm Highest calibration point 2018 CAL_POINT_LO 0 0 to 2 S/cm Lower calibration point 2019 CAL_MIN_SPAN 0,0001 > 0,0001 S/cm Minimum span between two calibration points 2020 SENSOR_CAL_METHOD - 1point, 2point not used 2021 SENSOR_CAL_DATE - till 2104 Date the sensor was last calibrated 2022 SECONDARY_VALUE - -20 to 140 ºC , Temperature value 0 to 280ºF 2023 SECONDARY_VALUE_UNIT ºC ºC, ºF Temperature unit 2024 SENSOR_TEMP automatic automatic auto when a temperature element is available 2025 SENSOR_TEMP_MAN_VALUE - - No manual temp. value possible. Always automatic 2026 SENSOR_TYPE _TEMP NTC30K Pt1000, NTC30k Temperature element used: 2027 SENSOR_CONNECTION_ 2 2 Only 2-wire connections supported TEMP 2028 SENSOR_TYPE_COND toroidal toroidal Non contacting toroidal sensor. 2029 SENSOR_OHMS - Actual cell resistance 2030 XD_MAN_ID “ “ 2031 TEMPERATURE_COEFF 2.1 0 to 3.5%/ºC (%/ºF) Process temperature compensation factor 2032 CONCENTRATION - Conductivity combined with temperature can be directly related to the
concentration. Concentraion isexpressed in percentage 2033 TERTIARY_VALUE - 0 to 2 S/cm Second compensated conductivity value 2034 REFERENCE_TEMPERATURE 25 0 to 100 ºC Conductivity can be process compensated to a standard (32 to 212ºF) reference temperature. Mostly 20ºC or 25ºC is used
IM 12A00A01-61E
Foundation Fieldbus 3-39
Index Parameter name Factory Valid Description Default Range
2035 COMP_METHOD NaCl NaCl, TC, matrix 2036 COMP_MATRIX_SEL H2SO4 H2SO4, 0 -100ºC, 0 - 5%, When matrix compensation is required one can
2037 TERTIARY_COMP_METHOD NaCl NaCl, TC, matrix Method of process temperature compensation for the second conductivity value 2038 TERT_TEMPERATURE 2.1 0 ~ 3.5 %/ºC Process temperature compensation factor for the _COEFF second conductivity value 2039 ALARM_SUM - 2040 DEV_ALARM - Device Alarm is used to give the status of the analyser.
2041 LOGBOOK1_RESET Idle Idle, Reset Reset the pointer to the first (oldest) event in logbook 1 2042 LOGBOOK1_EVENT - Event whereto the pointer is referenced. When parameter
2043 LOGBOOK2_RESET Idle Idle, Reset Reset the pointer to the first (oldest) event in logbook 2 2044 LOGBOOK2_EVENT - Event whereto the pointer is referenced. When parameter
2045 LOGBOOK_CONFIG - Per event one can decide whether it should be logged and
2046 TEST_1 -
.... - 2047 to 2057 are, like 2046 and 2058, service parameters
2058 TEST_13 ­ 2059 CALIB_SENSOR_CONST 1.88 0.2~19.99 cm-1 The calibrated cell constant [1/cm]. 2060 MATRIX_TEMP_RANGE -20~140ºC, 0~280ºF The temperature values of the user defined matrix. 2061 SOLUTION_1 0~1.999 S/cm The conductivity values of solution 1. 2062 SOLUTION_2 0~1.999 S/cm The conductivity values of solution 2. 2063 SOLUTION_3 0~1.999 S/cm The conductivity values of solution 3. 2064 SOLUTION_4 0~1.999 S/cm The conductivity values of solution 4. 2065 SOLUTION_5 0~1.999 S/cm The conductivity values of solution 5. 2066 CONCENTRATION_ disabled MEASUREMENT 2=enabled 2067 CONCENTRATION_0 0 0~100 The concentration [%] corresponding to the 0 percent table value 2068 CONCENTRATION_100 100 0~100 The concentration [%] corresponding to the 100 percent table value 2069 CONC._TABLE_LOW 0~1.999 S/cm The first part of the concentration table (0 till 30 percent). 2070 CONC._TABLE_MID 0~1.999 S/cm The second part of the concentration table (35 till 65 percent). 2071 CONC._TABLE_HIGH 0~1.999 S/cm The third part of the concentration table (70 till 100 percent). 2072 E5_LIMIT 3 (3 S) 0~5 S The upper range accuracy limit of the analyzer [Siemens]. 2073 E6_LIMIT 5e-6 (5 μS) 0~5 S The lower range accuracy limit of the analyzer [Siemens]. 2074 DISPLAY_RESOLUTION 1 0 to 8 Display resolution, 0, Not initialized (can not change)
2075 PASSCODE_MAINT 0 0, 111, 333, 777, Passcode used to protect the maintenance menu
2076 PASSCODE_COMM 0 0, 111, 333, 777, Passcode used to protect the commissioning menu.
2077 PASSCODE_SERVICE 0 0, 111, 333, 777, Passcode used to protect the service menu.
2078 ERROR_CONFIG* 0=soft, 1=hard Soft/Hard fail configuration (checked = hard unchecked = soft). 2079 CONFIGURATION*
2080 TRANSMITTER_TIME Transmitter time.” 2081 TEST_14 Service parameter.
autoreturn on bitnr. setting
H2SO4, 0 -100ºC, 2.5 - 25% HCl, 0 - 60ºC, 0.5 - 5%, one user definable matrix HNO3, 0 - 80ºC, 0.5 - 5% HNO3, 0 -80ºC, 2.5 - 25% NaOH, 0 -100ºC, 0.5 - 5% NaOH, 0 -100ºC, 0.5 - 15% User programmable matrix
0=Not initialized, 1=disabled,
888, 123, 957, 331, 546, 847
888, 123, 957, 331, 546, 847
888, 123, 957, 331, 546, 847
Instrument specific configuration. See below for correct
Method of process temperature compensation for the primary value
, make a selection out of 8 predefined matrices and
See separate table for error messages
is read, the pointer is increased by one.
is read, the pointer is increased by one
in which logbook (1 or 2) it should be logged
Enabling/Disabling of concentration measurement.
1, Auto Ranging, 2, x.xxx μS/cm, 3, xx.xx μS/cm, 4, xxx.x μS/cm, 5, x.xxx mS/cm, 6, xx.xx mS/cm,
7, xxx.x mS/cm, 8, xxxx mS/cm
3-39
IM 12A00A01-61E
3-40 Foundation Fieldbus
* Enumerated parameters (4 bytes, 32 bits) where each bit can be set individually.
ERROR_CONFIG
0x00000800(bit 11), “conductivity exceeds high limit (E5)” 0x00000400 (bit 10), “conductivity exceeds low limit (E6)” 0x00000200 (bit 9), “temperature sensor open (E7)” 0x00000100 (bit 8), “temperature sensor shorted (E8)”
CONFIGURATION
0x00800000 (bit 23), “Auto return enabled”
IM 12A00A01-61E
Foundation Fieldbus 3-41
3-6-3-4 Transducer Block dO202
Index Parameter Factory valid range description Name default
2000 BLK_DATA TAG: “TB” General information about the function block 2001 ST_REV - - The revision level of the static data associated with the
function block. The revision value will be incremented each time a static parameter value in the block is changed.
2002 TAG_DESC “ “ The user description of the intended application of the
block.
2003 STRATEGY 1 The strategy field can be used to identify grouping of
blocks. This data is not checked or processed by the block
2004 ALERT_KEY 1 The identification number of the plant unit. This
information may be used in the host for sorting alarms, etc.
2005 MODE_BLK AUTO The actual, target, permitted, and normal modes of the
block
2006 BLOCK_ERR - This parameter reflects the error status associated with a
block. It is a bit-string, so that multiple errors can be shown. 2007 UPDATE_EVT - The alert is generated by any change to the static data. 2008 BLOCK_ALM - The block alarm is used for all configuration error,
hardware connection failure or system problems in the
block. The cause of the alert is entered in the subcode
field. The first alert to become active will set Active
status in Status attribute. 2009 TRANSDUCER A directory that specifies the number and starting _DIRECTORY indices of the transducers. 2010 TRANSDUCER_TYPE DO DO Yokogawa specific Dissolved Oxygen transducer block 2011 XD_ERROR - The error code in transducer:
- No failure
- Electronics failure
- I/O failure
- Mechanical failure 2012 COLLECTION - A directory that specifies the number, starting indices _DIRECTORY and DD item IDs of the data collection in each
transducer within a transducer block. 2013 PRIMARY_VALUE Dissolved Oxygen Dissolved Oxygen The type of measurement represented by primary _TYPE value. 2014 PRIMARY_VALUE - 0 to 50 ppm Primary measuring value of the instrument including
0 to 1999 ppb status information.
0 to 600 % 2015 PRIMARY_VALUE - 0 to 50 ppm The measuring range of the instrument associated with _RANGE 0 to 1999 ppb the selected unit type.
0 to 600 % 2016 PRIMARY_VALUE_UNIT ppm ppm, ppb, % Select the unit in which the primary value must be
presented 2017 SENSOR_TYPE Polarographic Polarographic, Galvanic Select the type of sensor connected to the instrument, _OXYGEN Galvanic or Polarographic 2018 SAMPLE_CAL - 0 to 50 ppm Enter the correct (current) measured value. This value
0 to 1999 ppb is used to calculate the new sensitivity (slope) 0 to 600 %
2019 ZERO_CURRENT 0 +- ZERO_CURRENT Directly overwrites the current zero current, which is
_LIMIT (param 2037) the leak current of the sensor 2020 SENSITIVITY 7.5 1.000 to 1999 nA/ppm Directly overwrites the sensitivity (slope) of the sensor 2021 AMP_STABILIZE_TIME 60 5 to 600 seconds Stability criteria used during automatic calibration. 2022 AMP_SPAN_STABILIZE 0.1 0 to 50 ppm Stability criteria used during automatic span calibration. _VALUE
IM 12A00A01-61E
3-42 Foundation Fieldbus
Index Parameter Factory valid range description Name default
2023 AMP_ZERO_STABILIZE - 0 to 50 ppm Stability criteria used during automatic zero calibration. _VALUE 0 to 1999 ppb
0 to 600 % 2024 SALINITY 0 0 to 99.9 ppt Salinity value of the process liquid that can be used to
compensate the DO value
2025 BAR_PRESSURE 101.3 0.0 to 999 kPa Manual value of the pressure used to compensate the
DO reading (during calibration) 2026 BAR_PRESSURE_UNIT kPa kPa The engineering unit of barometric pressure value 2027 PERCENT_SATURATION 101.3 0.0 to 999 kPa Manual value of the pressure used to compensate _PRESSURE the DO reading (during measuring) 2028 CHLORINE_CALI- not used BRATION_RANGES 2029 SECONDARY_VALUE - -20 to 150 °C, -4 Temperature value to 302 °F 2030 SECONDARY_VALUE °C °C, °F Temperature unit _UNIT 2031 SENSOR_TEMP_COMP automatic automatic, manual Select manual when no temperature element is available
and the temperature is stable and select auto when a
temperature element is available 2032 SENSOR_TEMP_MAN 25° 0 - 100 °C 32 - 212 °F manual temperature value _VALUE 2033 SENSOR_TYPE_TEMP NTC22K 2034 TEMP_SENSOR_CAL - -20 to 150 °C, -4 Enter the correct (current) measured value. This value
2035 SENSOR_CURRENT - 0 to 1200 nA, The actual cell current of the DO sensor
2036
PERCENT_SATURATION
2037
ZERO_CURRENT_LIMIT
2038 ZERO_CAL - 0 to 50 ppm Enter the correct (current) measured value. This value
2039 RESERVED1 not used 2040 RESERVED2 not used 2041 RESERVED3 not used 2042
TRANSMITTER_TIME[6] 2043 PASSCODE 0 0, 111, 333, 777, 888, Passcode used to protect the maintenance menu. _MAINTENANCE 2044 PASSCODE 0 0, 111, 333, 777, 888, Passcode used to protect the commissioning menu _COMMISSIONING 2045 PASSCODE_SERVICE 0 0, 111, 333, 777, 888, Passcode used to protect the service menu.
2046 LOGBOOK1_RESET - Reset the pointer to the first (oldest) event in logbook 1. 2047 LOGBOOK1_EVENT - Event whereto the pointer is referenced. When
2048 LOGBOOK2_RESET - Reset the pointer to the first (oldest) event in logbook 2. 2049 LOGBOOK2_EVENT - Event whereto the pointer is referenced. When
2050
LOGBOOK_CONFIG[25]
2051 CALL_MAINT_TIME 250 1 to 250 days The interval of the maintenance timer _RELOAD 2052 CALL_MAINT_TIME 250 1 to 250 days The remaining days till maintenance is required. _COUNTDOWN 2053 ERROR_CONFIG* E9,E2,E7,E8 hard fail. Soft/Hard fail configuration (checked = hard;
- 0 to 600 % The DO value expressed as a percentage of the
0 to 199.9 nA, Limit used during zero calibration for the sensor
Transmitter time.
- Per event one can decide whether it should be logged
E16 soft fail unchecked = soft)
NTC22K, Pb36, PT1000
to 302 °F is used to calculate the temperature offset
0 to 50 µA
0 to 19.99 µA current (different for polarographic or galvanic sensor)
0 to 1999 ppb is used to calculate the new offset (zero) 0 to 600 %
123, 957, 331, 546, 847
123, 957, 331, 546, 847
123, 957, 331, 546, 847
Temperature element used
maximum saturation.
parameter is read, the pointer is increased by one.
parameter is read, the pointer is increased by one.
and in which logbook (1 or 2) it should be logged.
IM 12A00A01-61E
Foundation Fieldbus 3-43
Index Parameter Factory valid range description Name default
2054 CONFIGURATION* Instrument specific configuration 2055 ALARM_SUM -
2056 DEV_ALARM - Device Alarm is used to give the status of the transmitter.
Device_status in RB displays the device Alarms. 2057 TEST_1 - Service parameter …. - 2058 to 2068 are, like 2057 and 2070, service
parameters 2070 TEST_14 - Service parameter
* Enumerated parameters (4 bytes, 32 bits) where each bit can be set individually.
Error_config
0x00800000 (bit 23),
“Sensor current abnormal (E9)” 0x00010000 (bit 16), “Call for maintenance (E16)” 0x00004000 (bit 14), “Zero out of limits (E2)” 0x00000200 (bit 9), “Temperature sensor open (E7)” 0x00000100 (bit 8),
“Temperature sensor shorted (E8)”
Configuration
0x00008000 (bit 15), “Zero calibration enabled” 0x00001000 (bit 12), “Auto return on” 0x00000800 (bit 11), “Salinity compensation on” 0x00000400 (bit 10), “Call for maintenance on” 0x00000100 (bit 8), “Manual pressure in mainte­nance menu”

3-7 Application setting and change of basic parameters

3-7-1 Applications and selection of basic parameters

Setting Item (applicable parameters) Summary
Tag No. Sets PD Tag and each block tag.
Up to 32 alphanumeric characters can be set for both tags.
Refer to “Tag and address” in Section 3-3-4 Calibration range setup (XD_SCALE)
Output scale setup (OUT_SCALE)
Output mode setup (L_TYPE)
Output signal low cut mode setup (LOW_CUT) Damping time constant setup (PV_FTIME) Simulation setup (SIMULATE)
Sets the range of input from the transducer block corresponding to the 0% and 100%
points in operation within the AI function block. The calibrated range (0% and 100%) is the
factory default setting. Sets the range unit number of decimals required.
Sets the scale of output corresponding to the 0% and 100% points in operation within
the AI function block. It is possible to set a unit and scale that differs from the calibration
range. Sets the range unit and the number of decimals required.
Selects the operation function of the AI function block. It may be chosen from among
Direct, Indirect, and IndirectSQRT.
Direct:
Indirect:
IndirectSQRT:
If the output falls below the setting of this parameter, the output is set to Zero. It can be
set individually with Direct, Indirect, and IndirectSQRT.
Sets the time constant of the damping (primary delay) function in the AI function block in
seconds.
Performs simulation of the AI function block.
The input value and status for the calibration range can also be set.
It is recommended that this parameter be used for loop checks and other purposes.
Refer to “Simulation Function” in Section 3-4-3.
The output of the transducer block is directly output only via filtering without scaling and square root extraction. Output processed by proportion at the AI function block. Output processed by square root extraction at the AI function block.
IM 12A00A01-61E
3-44 Foundation Fieldbus
IMPORTANT

3-7-2 Setting and change of basic parameters

This section describes the procedure taken to set and change the parameters for each block. Obtaining access to each parameter differs depending on the configuration system used. For details, refer to the instruction manual for each configuration system.
Access the block mode (MODE_BLK) of each block.
Set the Target of block mode (MODE_BLK) to Auto, Man or O/S (*Note 2) according to the Write Mode of the parameter to be set or changed.
When “actual mode” has changed (*Note 1),
data associated with the function block
can be maintenanced.
back to Auto (*Note 2) .
Set the Target (*Note 1) of block mode
Do not turn the power OFF immediately after parameter setting. When the parameters are saved to the EEPROM, the redundant processing is executed for an improvement of reliability. Should the power be turned OFF within 60 seconds after setting of parameters, changed parameters are not saved and may return to their original values.
Note 1: Block mode consists of the following four modes that are controlled by the universal parameter that
displays the running condition of each block. Target: Sets the operating condition of the block. Actual: Indicates the current operating condition. Permit: Indicates the operating condition that the block is allowed to take. Normal: Indicates the operating condition that the block will usually take. Note 2: The following are the operating conditions which the individual blocks will take.
Refer to the “List of parameters for each block of the EXA” for details of the Write Mode for each block.
AI Function Block Transducer Block resource Block
Automatic (Auto) Yes Yes Yes Manual (Man) Yes Out of Service (O/S) Yes Yes Yes

3-7-3 Setting the AI Function Blocks

(1) Setting the output scale
As explained in section 3-3-6-4 the OUT_SCALE can used to convert the channel’s value to a different scale. If the channel’s unit (= XD_SCALE unit) is the same as the output unit DO NOT use scaling or let the OUT_SCALE have the same scaling as XD_SCALE. If L_TYPE is set to Indirect or Ind determines the
conversion from FIELD_VAL to the output. PV and OUT always have identical scaling. OUT_
SCALE provides scaling for PV. The PV is always the value that the block will place in OUT if the mode is Auto.
For AI1 set L_TYPE to Direct IM 12A00A01-61E
Sqr Root, OUT_SCALE
Foundation Fieldbus 3-45
With the EXA, the channel values are displayed on the display indicator, independant of the scaling in the AI blocks.
(2) Setting the output mode
Access the L_TYPE parameter. Set the output mode. 1: Direct (Sensor output value) 2: Indirect (Linear output value) 3: IndirectSQRT (Square root extraction output value)
(3) Setting the damping time constant
Access the PV_FTIME parameter. Set the damping time (in seconds).
(4) Simulation
By optionally setting the input value to the calibration range and status, perform simulation of the AI function block.
Access the Simulate Value parameter.
Set an optional input value.
Access the Simulate Status parameter.
Set the status code.
Access the Simulate En/Disable parameter. Set whether Simulation is enabled or disabled. 2: Enabled 1: Disabled
If simulation is enabled, AI block uses Simulate Status and Simulate Value as the input, and if disabled, the AI block uses Transducer Status and Transducer Value as input. Refer to Section 3.4.3 Simulation Function.

3-7-4 Setting the Transducer Block

To access function specifics of the EXA of the transducer block, the DD (Device Description) for EXA needs to have been installed in the configuration tool used. For integration of DD, refer to “Integration of DD” in Section 3-2-4.
IM 12A00A01-61E
3-46 Foundation Fieldbus

3-8 Operation of each parameter in failure mode 3-8-1 Operation of each parameter in failure mode PH202

• Following table summarizes the value of EXA parameters when LCd display indicates an Alarm.
EXA display
E9
E12
E7
E8
E5.1
E4.1
E5.2
E4.2
E16
E20
E21
error description resource block transducer block transducer block transducer block transducer block transducer block transducer block AI1 (channel = 1) AI2 (channel = 2) AI3 (channel = 3) BLOCK_ERR DEV_ALARM BLOCK_ERR XD_ERROR PV.status SV.status TV.status OUT.status OUT.status OUT.status primary value 0x80000000 INPUT_FAILURE MECHANICAL_FAILURE BAD, SENS BAD, SENS_FAIL exceeds limits _ERR _FAIL tertiary value 0x40000000 INPUT_FAILURE MECHANICAL_FAILURE exceeds limits _ERR temperature 0x20000000 INPUT_FAILURE MECHANICAL_FAILURE BAD, SENS BAD, SENS sensor open _ERR _FAIL _FAIL temperature 0x10000000 INPUT_FAILURE MECHANICAL_FAILURE BAD, SENS BAD, SENS sensor shorted _ERR _FAIL _FAIL glass impedance 0x08000000 INPUT_FAILURE MECHANICAL_FAILURE BAD, SENS exceeds high limit _ERR _FAIL glass impedance 0x04000000 INPUT_FAILURE MECHANICAL_FAILURE BAD, SENS exceeds low limit _ERR _FAIL reference impedance 0x02000000 INPUT_FAILURE MECHANICAL_FAILURE BAD, SENS exceeds high limit _ERR _FAIL reference impedance 0x01000000 INPUT_FAILURE MECHANICAL_FAILURE BAD, SENS exceeds low limit _ERR _FAIL
calibration timer expired NOW_ERR _SPECIFIC _SPECIFIC not used 0x00400000 not used 0x00200000 not used 0x00100000 not used 0x00080000 not used 0x00040000 mismatch between FF 0x00020000 NEEDS_MAINT_ ELECTRONICS_FAILURE BAD, DEV BAD, DEV BAD, DEV_FAIL BAD, DEV_FAIL BAD, DEV_FAIL BAD, DEV_FAIL interface and EXA NOW_ERR _FAIL _FAIL parameter EXA eeprom failure 0x00010000 NEEDS_MAINT_ NOW_ERR _FAIL _FAIL FF interface eeprom LOST_STATIC_ERR, 0x00008000 NEEDS_MAINT_ failure LOST_NV_ERR NOW_ERR _FAIL _FAIL Hart communication 0x00004000 NEEDS_MAINT_ ELECTRONICS_FAILURE BAD, DEV BAD, DEV BAD, DEV_FAIL BAD, DEV_FAIL BAD, DEV_FAIL BAD, DEV_FAIL failure NOW_ERR _FAIL _FAIL EXA checksum error 0x00002000 ELECTRONICS_FAILURE BAD, DEV BAD, DEV BAD, DEV_FAIL BAD, DEV_FAIL BAD, DEV_FAIL BAD, DEV_FAIL FF interface checksum 0x00001000 _FAIL _FAIL error resource block out of OUT_OF_SERVICE_ERR 0x00000800 BAD, NON BAD, NON BAD, NON BAD, NON_SPECIFIC BAD, NON_SPECIFIC BAD, NON_SPECIFIC service _SPECIFIC _SPECIFIC _SPECIFIC transducer block out of 0x00000400 OUT_OF_ BAD, OUT BAD, OUT BAD, OUT_OF BAD, NON_SPECIFIC BAD, NON_SPECIFIC BAD, NON_SPECIFIC service SERVICE_ERR _OF_SERVICE _OF_SERVICE _SERVICE AI1 out of service 0x00000200 AI1 in manual mode 0x00000100 AI1 in simulation mode SIMULATE_ACTIVE_ERR 0x00000080
0x00800000 NEEDS_MAINT_ BAD, NON BAD, NON BAD, NON_SPECIFIC BAD, NON_SPECIFIC
DATA_INTEGRITY_ERROR
DATA_INTEGRITY_ERROR
BAD, DEV BAD, DEV BAD, DEV_FAIL BAD, DEV_FAIL BAD, DEV_FAIL BAD, DEV_FAIL
BAD, DEV BAD, DEV BAD, DEV_FAIL BAD, DEV_FAIL BAD, DEV_FAIL BAD, DEV_FAIL
AI1 not scheduled 0x00000040 AI2 out of service 0x00000020 AI2 in manual mode 0x00000010 AI2 in simulation mode SIMULATE_ACTIVE_ERR 0x00000008 AI3 out of service 0x00000004 AI3 in manual mode 0x00000002 AI3 in simulation mode SIMULATE_ACTIVE_ERR 0x00000001
IM 12A00A01-61E
Foundation Fieldbus 3-47
BAD, SENS_FAIL
BAD, SENS_FAIL
BAD, SENS_FAIL
BAD, SENS_FAIL
BAD, SENS_FAIL
BAD, SENS_FAIL
BAD, SENS_FAIL
BAD, SENS_FAIL
BAD, SENS_FAIL BAD, SENS_FAIL BAD, SENS_FAIL
BAD, SENS_FAIL BAD, SENS_FAIL BAD, SENS_FAIL
BAD, SENS_FAIL BAD, SENS_FAIL
BAD, SENS_FAIL BAD, SENS_FAIL
BAD, SENS_FAIL BAD, SENS_FAIL
BAD, SENS_FAIL BAD, SENS_FAIL
BAD, OUT_OF_SERVICE
BAD, OUT_OF_SERVICE
BAD, OUT_OF_SERVICE
IM 12A00A01-61E
3-48 Foundation Fieldbus

3-8-2 Operation of each parameter in failure mode SC202

• Following table summarizes the value of EXA parameters when LCD display indicates an Alarm.
error description EXA dev_alarm resource block transducer block channel = 1 (AI1) channel = 2 (AI2) channel = 3 (AI3) channel = 4 display
BLOCK_ERR BLOCK_ERR XD_ERROR PV.status SV.status TV.status CONCENTRATION.status OUT.status OUT.status OUT.status OUT.status conductivity exceeds E5 0x80000000 INPUT_FAILURE MECHANICAL BAD, SENS BAD, SENS BAD, SENS_FAIL BAD, SENS BAD, SENS_FAIL BAD, SENS_FAIL high limit _ERR _FAILURE _FAIL _FAIL _FAIL conductivity exceeds E6 0x40000000 INPUT_FAILURE MECHANICAL BAD, SENS BAD, SENS BAD, SENS_FAIL BAD, SENS BAD, SENS_FAIL BAD, SENS_FAIL low limit _ERR _FAILURE _FAIL _FAIL _FAIL temperature sensor E7 0x20000000 INPUT_FAILURE MECHANICAL BAD, SENS BAD, SENS BAD, SENS BAD, SENS_FAIL BAD, SENS BAD, SENS_FAIL BAD, SENS_FAIL BAD, SENS_FAIL open _ERR _FAILURE _FAIL _FAIL _FAIL _FAIL temperature sensor E8 0x10000000 INPUT_FAILURE MECHANICAL BAD, SENS BAD, SENS BAD, SENS BAD, SENS_FAIL BAD, SENS BAD, SENS_FAIL BAD, SENS_FAIL BAD, SENS_FAIL shorted _ERR _FAILURE _FAIL _FAIL _FAIL _FAIL temperature E2 0x08000000 BAD, NON_ BAD, NON BAD, NON_SPECIFIC BAD, NON_ BAD, NON_ BAD, NON_SPECIFIC compensation error SPECIFIC _SPECIFIC SPECIFIC SPECIFIC polarization detected E1 0x04000000 INPUT_FAILURE BAD, NON_ BAD, NON BAD, NON_SPECIFIC BAD, NON_ BAD, NON_ BAD, NON_SPECIFIC _ERR SPECIFIC _SPECIFIC SPECIFIC SPECIFIC conductivity exceeds E13 0x02000000 BAD, NON_ BAD, NON BAD, NON_SPECIFIC BAD, NON_ BAD, NON_ BAD, NON_SPECIFIC usp limit SPECIFIC _SPECIFIC SPECIFIC SPECIFIC concentration table E18 0x01000000 BAD, CONFIG BAD, CONFIG BAD, CONFIG_ERR BAD, CONFIG BAD, CONFIG_ BAD, CONFIG_ERR error _ERR _ERR _ERR ERR matrix error E4 0x00800000 BAD, CONFIG BAD, CONFIG BAD, CONFIG_ERR BAD, CONFIG BAD, CONFIG_ BAD, CONFIG_ERR _ERR not used 0x00400000 _ERR _ERR ERR not used 0x00200000 not used 0x00100000 not used 0x00080000 not used 0x00040000 mismatch between FF 0x00020000 NEEDS_MAINT_ ELECTRONICS BAD, DEV BAD, DEV BAD, DEVL BAD, DEV BAD, DEV_FAIL BAD, DEV_ BAD, DEV_FAIL interface and EXA NOW_ERR _FAIL _FAIL _FAI _FAIL FAIL parameter _FAILURE EXA eeprom failure E20 0x00010000 NEEDS_MAINT_ DATA_INTEGRITY BAD, DEV BAD, DEV BAD, DEV BAD, DEV BAD, DEV_FAIL BAD, DEV_ BAD, DEV_FAIL NOW_ERR _ERROR _FAIL _FAIL _FAIL _FAIL FAIL FF interface eeprom 0x00008000 LOST_STATIC NEEDS_MAINT_ DATA_INTEGRITY BAD, DEV BAD, DEV BAD, DEV BAD, DEV BAD, DEV_FAIL BAD, DEV_ BAD, DEV_FAIL failure _ERR, LOST_ NOW_ERR _ERROR _FAIL _FAIL _FAIL _FAIL FAIL NV_ERR Hart communication 0x00004000 NEEDS_MAINT_ ELECTRONICS BAD, DEV BAD, DEV BAD, DEV BAD, DEV BAD, DEV_FAIL BAD, DEV_ BAD, DEV_FAIL failure NOW_ERR _FAILURE _FAIL _FAIL _FAIL _FAIL FAIL EXA checksum error E21 0x00002000 ELECTRONICS BAD, DEV BAD, DEV BAD, DEV BAD, DEV BAD, DEV_FAIL BAD, DEV_ BAD, DEV_FAIL _FAILURE _FAIL _FAIL _FAIL _FAIL FAIL FF interface checksum 0x00001000 error resource block out of 0x00000800 OUT_OF_ BAD, NON BAD, NON BAD, NON_ BAD, NON BAD, NON_ BAD, NON_ BAD, NON_SPECIFIC service SERVICE_ERR _SPECIFIC _SPECIFIC SPECIFIC _SPECIFIC SPECIFIC SPECIFIC transducer block out of 0x00000400 OUT_OF_ BAD, OUT_OF_ BAD, OUT_ BAD, OUT_ BAD, NON BAD, NON_ BAD, NON_ BAD, NON_SPECIFIC service SERVICE_ERR SERVICE OF_SERVICE OF_SERVICE _SPECIFIC SPECIFIC SPECIFIC AI1 out of service 0x00000200 BAD, OUT_OF _SERVICE AI1 in manual mode 0x00000100 AI1 in simulation mode 0x00000080 SIMULATE_ ACTIVE_ERR AI1 not scheduled 0x00000040 AI2 out of service 0x00000020 BAD, OUT_ OF_SERVICE AI2 in manual mode 0x00000010
AI2 in simulation mode 0x00000008 SIMULATE_ ACTIVE_ERR AI3 out of service 0x00000004 BAD, OUT_ OF_SERVICE AI3 in manual mode 0x00000002 AI3 in simulation mode 0x00000001 SIMULATE_ ACTIVE_ERR
IM 12A00A01-61E
Foundation Fieldbus 3-49
IM 12A00A01-61E
3-50 Foundation Fieldbus

3-8-3 Operation of each parameter in failure mode ISC202

• Following table summarizes the value of EXA parameters when LCD display indicates an Alarm.
error description EXA dev_alarm resource block transducer block channel = 1 (AI1) channel = 2 (AI2) channel = 3 (AI3) channel = 4 display
BLOCK_ERR BLOCK_ERR XD_ERROR PV.status SV.status TV.status CONCENTRATION.status OUT.status OUT.status OUT.status OUT.status conductivity exceeds E5 0x80000000 INPUT_FAILURE MECHANICAL BAD, SENS BAD, SENS BAD, SENS_FAIL BAD, SENS BAD, SENS_FAIL BAD, SENS_FAIL high limit _ERR _FAILURE _FAIL _FAIL _FAIL conductivity exceeds E6 0x40000000 INPUT_FAILURE MECHANICAL BAD, SENS BAD, SENS BAD, SENS_FAIL BAD, SENS BAD, SENS_FAIL BAD, SENS_FAIL low limit _ERR _FAILURE _FAIL _FAIL _FAIL temperature sensor E7 0x20000000 INPUT_FAILURE MECHANICAL BAD, SENS BAD, SENS BAD, SENS BAD, SENS_FAIL BAD, SENS BAD, SENS_FAIL BAD, SENS_FAIL BAD, SENS_FAIL open _ERR _FAILURE _FAIL _FAIL _FAIL _FAIL temperature sensor E8 0x10000000 INPUT_FAILURE MECHANICAL BAD, SENS BAD, SENS BAD, SENS BAD, SENS_FAIL BAD, SENS BAD, SENS_FAIL BAD, SENS_FAIL BAD, SENS_FAIL shorted _ERR _FAILURE _FAIL _FAIL _FAIL _FAIL temperature E2 0x08000000 BAD, NON_ BAD, NON BAD, NON_SPECIFIC BAD, NON_ BAD, NON_ BAD, NON_SPECIFIC compensation error SPECIFIC _SPECIFIC SPECIFIC SPECIFIC concentration table E18 0x01000000 BAD, CONFIG BAD, CONFIG BAD, CONFIG_ERR BAD, CONFIG BAD, CONFIG_ BAD, CONFIG_ERR error _ERR _ERR _ERR ERR matrix error E4 0x00800000 BAD, CONFIG BAD, CONFIG BAD, CONFIG_ERR BAD, CONFIG BAD, CONFIG_ BAD, CONFIG_ERR _ERR not used 0x00400000 _ERR _ERR ERR not used 0x00200000 not used 0x00100000 not used 0x00080000 not used 0x00040000 mismatch between FF 0x00020000 NEEDS_MAINT_ ELECTRONICS BAD, DEV BAD, DEV BAD, DEVL BAD, DEV BAD, DEV_FAIL BAD, DEV_ BAD, DEV_FAIL interface and EXA NOW_ERR _FAIL _FAIL _FAI _FAIL FAIL parameter _FAILURE EXA eeprom failure E20 0x00010000 NEEDS_MAINT_ NOW_ERR _ERROR _FAIL _FAIL _FAIL _FAIL FAIL FF interface eeprom 0x00008000 LOST_STATIC NEEDS_MAINT_ failure _ERR, LOST_ NOW_ERR _ERROR _FAIL _FAIL _FAIL _FAIL FAIL NV_ERR Hart communication 0x00004000 NEEDS_MAINT_ ELECTRONICS BAD, DEV BAD, DEV BAD, DEV BAD, DEV BAD, DEV_FAIL BAD, DEV_ BAD, DEV_FAIL failure NOW_ERR _FAILURE _FAIL _FAIL _FAIL _FAIL FAIL EXA checksum error E21 0x00002000 ELECTRONICS BAD, DEV BAD, DEV BAD, DEV BAD, DEV BAD, DEV_FAIL BAD, DEV_ BAD, DEV_FAIL _FAILURE _FAIL _FAIL _FAIL _FAIL FAIL FF interface checksum 0x00001000 error resource block out of 0x00000800 OUT_OF_ BAD, NON BAD, NON BAD, NON_ BAD, NON BAD, NON_ BAD, NON_ BAD, NON_SPECIFIC service SERVICE_ERR _SPECIFIC _SPECIFIC SPECIFIC _SPECIFIC SPECIFIC SPECIFIC transducer block out of 0x00000400 OUT_OF_ service SERVICE_ERR SERVICE AI1 out of service 0x00000200 BAD, OUT_OF _SERVICE AI1 in manual mode 0x00000100 AI1 in simulation mode 0x00000080 SIMULATE_ ACTIVE_ERR AI1 not scheduled 0x00000040 AI2 out of service 0x00000020 BAD, OUT_ OF_SERVICE AI2 in manual mode 0x00000010 AI2 in simulation mode 0x00000008 SIMULATE_ ACTIVE_ERR AI3 out of service 0x00000004 BAD, OUT_ OF_SERVICE AI3 in manual mode 0x00000002 AI3 in simulation mode 0x00000001 SIMULATE_ ACTIVE_ERR
DATA_INTEGRITY
DATA_INTEGRITY
BAD, DEV BAD, DEV BAD, DEV BAD, DEV BAD, DEV_FAIL BAD, DEV_ BAD, DEV_FAIL
BAD, DEV BAD, DEV BAD, DEV BAD, DEV BAD, DEV_FAIL BAD, DEV_ BAD, DEV_FAIL
BAD, OUT_OF_ BAD, OUT_
OF_SERVICE
BAD, OUT_ BAD, NON BAD, NON_ BAD, NON_ BAD, NON_SPECIFIC
OF_SERVICE _SPECIFIC SPECIFIC SPECIFIC
IM 12A00A01-61E
Foundation Fieldbus 3-51
IM 12A00A01-61E
3-52 Foundation Fieldbus

3-8-4 Operation of each parameter in failure mode dO202

• Following table summarizes the value of EXA parameters when LCD display indicates an Alarm.
error description dev_alarm EXA resource block transducer block dO value eror BLOCK_ERR BLOCK_ERR XD_ERROR PV.status
calibration not stable 0x80000000 E1 S_QUALITY_BAD zero out of limits 0x40000000 E2 INPUT_FAILURE_ERR MECHANICAL_FAILURE S_SUBS_SENS_FAIL S_LIMIT_NON sensitivity out of limits 0x20000000 E3 S_QUALITY_BAD temp. sensor open 0x10000000 E7 INPUT_FAILURE_ERR MECHANICAL_FAILURE S_SUBS_SENS_FAIL S_LIMIT_NON S_QUALITY_BAD temp. sensor shorted 0x08000000 E8 INPUT_FAILURE_ERR MECHANICAL_FAILURE S_SUBS_SENS_FAIL S_LIMIT_NON S_QUALITY_BAD sensor current abnormal 0x04000000 E9 INPUT_FAILURE_ERR MECHANICAL_FAILURE S_SUBS_SENS_FAIL S_LIMIT_NON calibration timer expired 0x02000000 E16 not used 0x01000000 not used 0x00800000 not used 0x00400000 not used 0x00200000 not used 0x00100000 not used 0x00080000 not used 0x00040000 mismatch between FF S_QUALITY_BAD interface and 0x00020000 EXA parameter S_LIMIT_NON S_QUALITY_BAD EXA eeprom failure 0x00010000 E20 S_LIMIT_NON
failure
S_LIMIT_NON
S_QUALITY_BAD Hart communication
failure
S_LIMIT_NON S_QUALITY_BAD EXA checksum error 0x00002000 E21 S_LIMIT_NON FF interface checksum error resource block out of service transducer block out of service AI1 out of service 0x00000200 AI1 in manual mode 0x00000100 AI1 in simulation mode 0x00000080 SIMULATE_ACTIVE_ERR AI1 not scheduled 0x00000040 AI2 out of service 0x00000020 AI2 in manual mode 0x00000010 AI2 in simulation mode 0x00000008 SIMULATE_ACTIVE_ERR AI3 out of service 0x00000004 AI3 in manual mode 0x00000002 AI3 in simulation mode 0x00000001 SIMULATE_ACTIVE_ERR
0x00008000
0x00004000
0x00001000
0x00000800 OUT_OF_SERVICE_ERR S_BAD_NON_SPECIFIC
0x00000400 OUT_OF_SERVICE_ERR
LOST_NV_ERR
LOST_STATIC_ERR S_QUALITY_BAD
NEEDS_MAINT_NOW_ERR
NEEDS_MAINT_NOW_ERR ELECTRONICS_FAILURE
NEEDS_MAINT_NOW_ERR DATA_INTEGRITY_ERROR
NEEDS_MAINT_NOW_ERR DATA_INTEGRITY_ERROR
NEEDS_MAINT_NOW_ERR ELECTRONICS_FAILURE
S_BAD_NON_SPECIFIC
S_SUBS_DEV_FAIL
S_SUBS_DEV_FAIL
S_SUBS_DEV_FAIL
S_SUBS_DEV_FAIL
ELECTRONICS_FAILURE
S_SUBS_DEV_FAIL
S_BAD_OUT_OF_SERVICE
FF interface eeprom
IM 12A00A01-61E
Foundation Fieldbus 3-53
Temperature value Saturation percentage Cell current AI1 (channel = 1) AI2 (channel = 2) AI3 (channel = 3)
SV.status TV.status QV.status OUT.status OUT.status OUT.status
S_QUALITY_BAD S_QUALITY_BAD S_QUALITY_BAD S_SUBS_SENS_FAIL S_SUBS_SENS_FAIL S_SUBS_SENS_FAIL S_LIMIT_NON S_LIMIT_NON S_LIMIT_NON
S_QUALITY_BAD S_QUALITY_BAD S_QUALITY_BAD S_QUALITY_BAD S_QUALITY_BAD S_SUBS_SENS_FAIL S_SUBS_SENS_FAIL S_SUBS_SENS_FAIL S_SUBS_SENS_FAIL S_SUBS_SENS_FAIL S_LIMIT_NON S_LIMIT_NON S_LIMIT_NON S_LIMIT_NON S_LIMIT_NON S_QUALITY_BAD S_QUALITY_BAD S_QUALITY_BAD S_QUALITY_BAD S_QUALITY_BAD S_SUBS_SENS_FAIL S_SUBS_SENS_FAIL S_SUBS_SENS_FAIL S_SUBS_SENS_FAIL S_SUBS_SENS_FAIL S_LIMIT_NON S_LIMIT_NON S_LIMIT_NON S_LIMIT_NON S_QUALITY_BAD S_QUALITY_BAD S_QUALITY_BAD S_QUALITY_BAD S_SUBS_SENS_FAIL S_SUBS_SENS_FAIL S_SUBS_SENS_FAIL S_SUBS_SENS_FAIL S_LIMIT_NON S_LIMIT_NON S_LIMIT_NON S_BAD_NON_SPECIFIC S_BAD_NON_SPECIFIC S_BAD_NON_SPECIFIC S_BAD_NON_SPECIFIC
S_QUALITY_BAD S_QUALITY_BAD S_QUALITY_BAD S_QUALITY_BAD S_QUALITY_BAD S_SUBS_SENS_FAIL S_SUBS_SENS_FAIL S_SUBS_SENS_FAIL S_SUBS_SENS_FAIL S_SUBS_SENS_FAIL S_LIMIT_NON S_LIMIT_NON S_LIMIT_NON S_LIMIT_NON S_QUALITY_BAD S_QUALITY_BAD S_QUALITY_BAD S_QUALITY_BAD S_QUALITY_BAD S_QUALITY_BAD S_SUBS_SENS_FAIL S_SUBS_SENS_FAIL S_SUBS_SENS_FAIL S_SUBS_SENS_FAIL S_SUBS_SENS_FAIL S_LIMIT_NON S_LIMIT_NON S_LIMIT_NON S_LIMIT_NON S_QUALITY_BAD S_QUALITY_BAD S_QUALITY_BAD S_QUALITY_BAD S_QUALITY_BAD S_QUALITY_BAD S_SUBS_SENS_FAIL S_SUBS_SENS_FAIL S_SUBS_SENS_FAIL S_SUBS_SENS_FAIL S_SUBS_SENS_FAIL S_LIMIT_NON S_LIMIT_NON S_LIMIT_NON S_LIMIT_NON S_QUALITY_BAD S_QUALITY_BAD S_QUALITY_BAD S_QUALITY_BAD S_QUALITY_BAD S_QUALITY_BAD S_SUBS_SENS_FAIL S_SUBS_SENS_FAIL S_SUBS_SENS_FAIL S_SUBS_SENS_FAIL S_SUBS_SENS_FAIL S_LIMIT_NON S_LIMIT_NON S_LIMIT_NON S_LIMIT_NON S_QUALITY_BAD S_QUALITY_BAD S_QUALITY_BAD S_QUALITY_BAD S_QUALITY_BAD S_QUALITY_BAD S_SUBS_SENS_FAIL S_SUBS_SENS_FAIL S_SUBS_SENS_FAIL S_SUBS_SENS_FAIL S_SUBS_SENS_FAIL S_LIMIT_NON S_LIMIT_NON S_LIMIT_NON S_LIMIT_NON S_QUALITY_BAD
S_BAD_NON_SPECIFIC S_BAD_NON_SPECIFIC S_BAD_NON_SPECIFIC S_BAD_NON_SPECIFIC S_BAD_NON_SPECIFIC S_BAD_NON_SPECIFIC
S_BAD_OUT_OF_SERVICE S_BAD_OUT_OF_SERVICE S_BAD_OUT_OF_SERVICE
S_BAD_NON_SPECIFIC S_BAD_NON_SPECIFIC S_BAD_NON_SPECIFIC
S_BAD_OUT_OF_SERVICE
S_BAD_OUT_OF_SERVICE
S_BAD_OUT_OF_SERVICE
IM 12A00A01-61E
Profibus 4-1
Sensor
input
Physical block
Block tag
Parameters
Node address
Output
AI function
block
Block tag
OUT
Parameters
AI function
block
Block tag
OUT
Parameters
AI function
block
Block tag
OUT
Parameters
Block tag
Parameters
Sensor
Transducer
block

4. PrOFIBUS

4-1. About Profibus

4-1-1 Outline

Profibus is a bi-directional digital communication protocol for field devices, which offers an advancement implementation technologies for process control systems and is widely employed by numerous field devices.
EXA Series Profibus communication type employs the specification standardized by the Profibus organisa­tion, and provides interoperability between Yokogawa devices and those produced by other manufacturers. Profibus comes with software consisting of three AI function blocks, providing the means to implement flex­ible instrumentation system.
For information on other features, engineering, design, construction work, startup and maintenance of Profibus, refer to the profibus organisation website: www.profibus.com

4-1-2 Internal Structure of EXA

The EXA contains a structured mapping of function blocks. Each function block serves a specific task.
(1) Physical block
• Manages the status of EXA hardware.
• Automatically informs the host of any detected faults or other problems.
(2) Transducer block
• Converts sensor input to process values which are transfered to AI function block by channels.
(3) AI1, AI2, AI3 function blocks
• Conditions raw data from the Transducer block.
• Outputs conditioned process values
• Carries out scaling, damping and square root extraction.

4-1-3 Logical Structure of Each BLOCK

Figure 4.1 Logical Structure of Each Block
IM 12A00A01-61E
4-2 Profibus
Node adress, block tags and contained parameters within a function block are structured in the EXA device as shown in figure 4.1.

4-1-4 Wiring System Configuration

The number of devices (<32) that can be connected to a single bus and the cable length vary depending on system design. When constructing systems, both the basic and overall design must be carefully considered to allow device performance to be fully exhibited.
Figure 4.2 Internal view of EXA wiring compartment
See Figure 4.4.
IM 12A00A01-61E
Figure 4.3 Glands to be used for cabling
Sensor cable gland
Profibus cable gland
Grounding terminal (connect to safety ground, only if power supply is not grounded)
Profibus 4-3
Figure 4.4 Pinhead connector
Figure 4.5 Gland connection
IM 12A00A01-61E
4-4 Profibus
Power
unit
trunck cable
Spur
Spur
Field
device
Field
device
Junction
box
V
1
V
3
V
2
V
4
V
1 = V2 = V3 = V4
Potential equalisation line (German practice)
Power
unit
trunck cable
Spur
Spur
Field
device
Field
device
Junction
box
High integrity eart, 0.1R or better
Neutral star-point bonding (English practice)
A
B

4-2. Preparation

The EXA 202 Profibus is provided with two cable glands. The first is used for the electrode wiring as the other is used for the power wiring shown in figure 4.3.
To open the EXA 202 for wiring:
1. Loosen the four frontplate screws and remove the cover.
2. The terminal strip is now visible.
3. Connect the power supply according figure 4.4. Use the gland on the left for this cable.
4. Connect the sensor input, using the gland on the right (see figure 4.3). Switch on the power. Commission the instrument as required or use the default settings.
5. Replace the cover and secure frontplate with the four screws.

4-2-1. Cables, terminals and glands

The EXA202 is equipped with terminals suitable for the connection of finished cables in the size range:
0.13 to 2.5 mm (26 to 14 AWG). The glands will form a tight seal on cables with an outside diameter in the
range of 6 to 12 mm (0.24 to 0.47 inches).

4-2-2. Shielding and grounding

Grounding and shielding of the transmitter is necessary for a safe and reliable operation. Please use one of the following schemes (A or B) as these will give proper shielding and grounding. One should pay special attention to instruments that required an external power supply (besides the 9-32V supplied by the bus).
Figure 4.6 Shielding and grounding
IM 12A00A01-61E
Profibus 4-5
EXA
Power supply
Terminator
HOST
Coupler/terminator

4-3. GETTING STArTEd

Profibus is fully dependent upon digital communication protocol (EN 50170 Volume 2 and IEC 61158 for IS areas, Profibus PA) and differs in operation from the conventional 4 to 20 mA transmission communi­cation protocol. It is recommended that novice users use field devices in accordance with the procedures described in this section. The procedures assume that field devices will be set up on a bench or an instru­ment shop.

4-3-1 Connection of devices

The following instruments are required for use with Profibus devices:
• Power supply:
Profibus requires a dedicated power supply. It is recommended that the current capacity be well over the total value of the maximum current consumed by all devices (including the host). Conventional DC current cannot be used as is.
• Terminator:
Profibus requires two terminators. Refer to the suppliers manual for details of terminators.
• Field devices / slaves
Several field devices or slave devices can be connected to the same link. depending on the power consuption, up to 32 devices can be connected. A maximum of 500mA may be used so the theoretical number of devices is 500mA / Power consumption per device.
• Host / Masters
Used for accessing field devices. A dedicated Master Class 1 host (such as PLC) is used for dedicated cyclic communication. A master Class 2 Host is used for device management puposes such as changing the “ device address” or parameter settings. For operation of the host, refer to the instruction manual for each host. No details of the host are explained in the rest of this material.
• Cable:
Used for connecting devices. Refer to IEC 61158 for details of instrumentation cabling. Profibus PA uses twisted pair wires up to 1900 metre. To meet the Electro Magnetic Interference standards a shielded twisted pair is obligated.
For applications in intrinsically safe areas, the transmission method defined in IEC 61158-2 is used with PROFIUS-PA The transmission rate in this case is 31.25 kbit/s. DP transmission via RS 485 to IEC 61158-2 is implemented with the network components (DP/PA coupler or DP/PA link)
Connect the devices as shown in Figure 4.7. Connect the terminators at both ends of the trunk, with a mini­mum length of the spur laid for connection.
The polarity of signal and power must be maintained.
Figure 4.7 Cabling
NOTE Before using a Profibus configuration tool other than the existing host, confirm it does not affect
the loop functionality in which all devices are already installed in operation. Disconnect the relevant control loop from the bus if necessary.
IM 12A00A01-61E
4-6 Profibus
IMPORTANT
IMPORTANT
Connecting a Profibus configuration tool to a loop with its existing host may cause communication data scrambles resulting in a functional disorder or a system failure.

4-3-2 Host Setting

To activate Profibus, the following settings are required for the host.
Do not turn off the power immediately after setting. When the parameters are saved to the EEPROM, the redundant processing is executed for an improvement of reliability. If the power is turned off within 60 sec­onds after settings are made, the modified parameters are not saved and the settings may return to their original values.
For cyclic data communication a GSD file is required. This file contains all necessary information to start this type of communication. Please make sure the gsd file is in the right directory so the information is avail­able to the HOST. Refer to the HOST’s manual for guidance.
For acyclic communication several configuration tools can be used. Each communication tool requires it’s own device driver. We currently support only Siemens Simatic PDM. All parameter lists and methods described in this manual are based on this acyclic communication tool. Please make sure the device driver is in the proper directory. Most HOST systems come with an “install device driver” package which places all required files in the designated folders.

4-3-3 Bus Power ON

Turn on the power of the host and the bus. First all segments of the display are lit, then the display begins to operate. If the indicator is not lit, check the polarity of the power supply.
Using the host device display function, check that the EXA is in operation on the bus. Unless otherwise specified, the following settings are in effect when shipped from the factory.
If no EXA is detected, check the available address. Please set all addresses of the devices in advance or separately connect each EXA and set a different address for each. Make sure to note the address after changing it as from this point on the device can only be accessed through this address.

4-3-4 Reading cyclic parameters

In general, slave devices exchange data cyclically with the master (class 1, e.g. PLC). The EXA (slave) gets the data from the sensor, makes some calculations and makes the outcome available through analog input blocks. The controller device (Master) requests for these process values, makes some calculation (PID, ratio) and sends the result to an actuator. The EXA contains three analog input blocks and therefore three Process values for cyclic data transfer.
The master class 1 device gets the information (of the EXA) by consulting the GSD file. A device specific GSD file should be available for each device and should have an unique identifier. All information necessary for cyclic data transfer is described in the GSD file. The YP01078X.gsd file and a bitmap file YP01078X.dib (device independent bitmap) should be placed in dedicated folders (“directories”). Refer to the manual of the Master Class 1 device for these folders. With these files, the Master Class 1 devices are able to start cyclic data transfer.
IM 12A00A01-61E
Profibus 4-7
Beforehandone must configure which information will be exchanged. According with the Profibus-PA Profile 3.01, there are two kinds of configurations possible. The Identifier byte (or short identifier) and the Extended Identifier Format (or long identifier). The EXA supports both kinds of configurations. The user can choose either “Analog Input (short)” or “Analog Input (long)” and will end up with the same result.
The function blocks of the EXA are in a specific order. The configuration of the cyclic data should be done in the same order.
For the EXA. the process values are mapped as follows:
PH202 SC202 ISC202 DO202
slot (1) “AI1” pH SC1 SC1 Dissolved Oxygen slot (2) “AI2” Temperature Temperature Temperature Temperature slot (3) “AI3” ORP/rH SC2 SC2 Percent Saturation
The GSD file of the EXA 202 specifies 3 modules:
;Empty module
Module = “Empty Module” 0x00 1 EndModule
;Modules for Analog Input
Module = “Analog Input (AI)short” 0x94 2 EndModule
;Modules for Analog Input
Module = “Analog Input (AI)long” 0x42,0x84,0x81,0x81 3 EndModule
;--- Description of the module assignment: --­; SlotDefinition Slot(1) = “AI1” 3 1,2,3 Slot(2) = “AI2” 3 1,2,3 Slot(3) = “AI3” 3 1,2,3
Examples:
- Configuring the output of the AI block 1 (pH) and the output of AI block 2 (temperature):
“Analog Input (short) “ and “Analog Input (short) “ and “Empty Module” or “0x94, 0x94, 0x00”
- Configuring the output of AI block 1 (pH) only:
“Analog Input (long) “ and “Empty module” and “Empty module” or “0x42, 0x84, 0x81, 0x81, 0x00, 0x00”
Note: Most Master Class 1 devices have an advanced Human Machine Interface and will guide you through
these configurations.

4-3-5 Reading acyclic parameters

Communications occur on a peer-to-peer basis. A cyclic communication services for parametrization, opera­tion, monitoring, alarm/error handling and diagnostics of intelligent devices may be handled in parallel to cyclic transfer.
IM 12A00A01-61E
4-8 Profibus

4-4. Function block parameters and Methods

4-4-1. Physical Block Parameters

Parameter Default Alternatives R/W Slot, Index Data Type
(byte,bit) (bytes)
Software revision “R3.01” R 0,24 visible string (16) Hardware revision “R3.01” R 0,25 visible string (16) Device manufacturer ID 37hex (Yokogawa) R 0,26 unsigned16 (2) Device ID PH202 “5945430850” R 0,27 visible string (16) SC202 “5945430851” R 0,27 visible string (16) ISC202 “5945430852” R 0,27 visible string (16) DO202 “5945430853” R 0,27 visible string (16) Device serial number “00000000U9313508” (example) R 0,28 visible string (16) Diagnosis for details see .gsd file R 0,29 bit string (4) Diagnosis extension for details see .gsd file R 0,30 bit string (6) Diagnosis mask R 0,31 bit string (4) Diagnosis mask extension R 0,32 bit string (6) Device certifiacation R 0,33 visible string (32) Write locking Write enable (2457) Write disable (0) R/W 0,34 unsigned16 (2) Factory reset 0 factory reset (1), warm start (2506),
reset bus address to 126 (2712) R/W 0,35 unsigned16 (2)
Descriptor YOKOGAWA PROFIBUS
-PA ANALYZER R/W 0,36 visible string (32)
Device message YOKOGAWA PROFIBUS
-PA ANALYZER R/W 0,37 visible string (32) Device install date “” R 0,38 visible string (16) Ident number selector manufacturer specific
ident no (1) profile specific ident number (0) R/W 0,40 unsigned8 (1) Device configuration “” R 0,52 visible string (32) Init state run (2) R/W 0,53 unsigned8 (1) Device state run (2) R/W 0,54 unsigned8 (1) Global status 0 bit0 - failure, bit1 - maintenance requested,
bit2 - function check, bit3 - limits exceeded, bit4~15 - reserved R 0,55 bit string (2)

4-4-2. Analog Input Block Parameters

Parameter Default Alternatives R/W Slot, Index Data Type
(byte,bit) (bytes)
AI1, AI2, AI3
Static Revision No. R 1,17 unsigned16 (2) Channel PH202: pH(284) Temp (298), ORP/rH(305) SC202: SC1(284) Temp (292), SC2(303), Conc(302) ISC202: SC1(284) Temp (292), SC2(303), Conc(302) DO202: DO(284) Temp (299), %Sat(306), sensor current(305) R/W 1,30 unsigned16 (2) Linearization type Non linearization (0) R/W 1,29 unsigned 8 (1) Unit R 1,28 DS-36 (11) Decimal point 1 R/W 1,28 DS-36 (11) Filter Time Const R/W 1,32 float (4)
Process Value Scale
Lower Value R/W 1,27 float[2] (8) Upper Value R/W 1,27 float[2] (8)
Output Scale
Lower Value R/W 1,28 DS-36 (11) Upper Value R/W 1,28 DS-36 (11)
Output Limits
Upper Limit Alarm Maximum of float R/W 1,37 float (4) Upper Limit Warning Maximum of float R/W 1,39 float (4) Lower Limit Warning Minimum of float R/W 1,41 float (4) Lower Limit Alarm Minimum of float R/W 1,43 float (4) Limit Hysteresis 0.5 % R/W 1,35 float (4)
IM 12A00A01-61E
Profibus 4-9

4-4-3. Transducer block parameters

4-4-3-1. Transducer block parameters PH202
Parameter default Alternatives Unit r/W Slot, Index data Type
(byte,bit) (bytes)
Primary value Type 111(pH) R 4, 27 unsigned16 (2) Measured Value pH R 4, 28 DS-33 (5) Status R 4, 28 DS-33 (5) Sensor type pH pH/ORP sensor(160) pH/ORP sensor(160) R 4, 30 unsigned16 (2) Sensor mV mV R 4, 31 float (4)
Secondary value Parameters
Value ºC R 4, 42 DS-33 (5) Status R 4, 42 DS-33 (5)
Secondary value unit ˚C ˚F R/W 4, 43 unsigned16 (2)
Sensor temp comp Automatic(2) Off(0), Manual(1) R/W 4, 44 unsigned8 (1) Sensor temp man value ºC R/W 4, 45 float (4) Temperature sensor Pt1000(148) Pt100(128), 5k1(150), 3kBalco(145), R/W 4, 46 unsigned16 (2) 8k55(151), 350(153), PTC10k(158),
6k8(157) Temp. connection type 1 R 4, 47 unsigned8 (1) Process compensation Disabled(0) Enabled(1) R/W 4, 167(1.0) unsigned32 (4) Temperature Coefficient mV/
10ºC R/W 4, 159 float (4)
Tertiary value Parameters
Tertiary value type None(1) ORP(2), rH(3) R 4, 48 unsigned16 (2) Value mV R 4, 49 DS-33 (5) Status R 4, 49 DS-33 (5)
Calibration Parameters
Calibration High pH R/W 4, 32 float (4) Calibration Low pH R/W 4, 33 float (4) Cal min span pH R 4, 34 float (4) Slope check Enabled(1) Disabled(0) R/W 4, 167(1.1) unsigned32 (4) Slope % R/W 4, 35 float (4) Slope unit %(1342) %(1342) R 4, 36 unsigned16 (2) Zero check Enabled(1) Disabled(0) R/W 4, 167(1.0) unsigned32 (4) Zero mV R/W 4, 37 float (4) Zero unit mV(1243) pH(1422) R/W 4, 38 unsigned16 (2) Isopotential pH pH R/W 4, 39 float (4) Tertiary zero mV R/W 4, 51 float (4) Calibration method not calibrated(0) one point(107), two point(108), other(255) R/W 4, 40 unsigned8 (1) Date of Calibration R 4, 41 DS-21 Stable Time 5.0 sec S R/W 4, 97 float (4) Stable Value 0.02pH pH R/W 4, 98 float(4)
diagnostic Settings
Glass impedance ohm R 4, 52 float (4) Glass impedance Lo Lim ohm R/W 4, 101 float (4) Glass impedance Hi Lim ohm R/W 4, 102 float (4) Glass impedance check Enabled(1) Disabled(0) R/W 4, 167(2.2) unsigned32 (4) Glass impedance circuit High(1) Low(0) R/W 4, 167(2.0) unsigned32 (4) Glass impedance comp. Enabled(1) Disabled(0) R/W 4, 167(2.1) unsigned32 (4) Glass impedance exceeds low limit (E4.1) Hard fail(1) Soft fail(0) R/W 4, 166(2.2) unsigned32 (4) Glass impedance exceeds high limit (E5.1) Hard fail(1) Soft fail(0) R/W 4, 166(2.3) unsigned32 (4) Reference impedance ohm R 4, 53 float (4) Reference impedance low limit ohm R/W 4, 103 float (4) Reference impedance high limit ohm R/W 4, 104 float (4) Reference impedance check
Enabled(1) Disabled(0) R/W 4, 167(2.5) unsigned32 (4)
IM 12A00A01-61E
4-10 Profibus
4-4-3-1. Transducer block parameters PH202 (continued)
Parameter default Alternatives Unit r/W Slot, Index data Type
(byte,bit) (bytes)
Reference impedance circuit Reference impedance
compensation Disabled(0) Enabled(1) R/W 4, 167 (2.4) unsigned32 (4) Ref. impedance exceeds low limit (E4.2) Hard fail(1) Soft fail(0) R/W 4, 166 (2.4) unsigned32 (4) Ref. impedance exceeds high limit (E5.2) Hard fail(1) Soft fail(0) R/W 4, 166 (2.5) unsigned32 (4) Temp. sensor open (E7) Hard fail(1) Soft fail(0) R/W 4, 166 (2.6) unsigned32 (4) Temp. sensor shorted (E8) Hard fail(1) Soft fail(0) R/W 4, 166 (2.7) unsigned32 (4) Primary value exceeds limits (E9) Hard fail(1) Soft fail(0) R/W 4, 166 (1.0) unsigned32 (4) Calibration timer expired (E16) Soft fail(0) Hard fail(1) R/W 4, 166 (1.7) unsigned32 (4)
Passcode Configuration
Passcode Maintenance - 000, 111, 333, 777, 888, 123, 957,
Passcode Commissioning - 000, 111, 333, 777, 888, 123, 957,
Passcode Service - 000, 111, 333, 777, 888, 123, 957,
display Settings
Display precision 0.1 pH(0) 0.01 pH(1) R/W 4, 167 (1.2) unsigned32 (4) Auto return Enabled(1) Disabled(0) R/W 4, 167 (1.3) unsigned32 (4)
Logbook Configuration
power up Logbook 2(3), Not Logged(1), Logbook 1(2) R/W 4, 60 unsigned8 (1) power down Logbook 2(3), Not Logged(1), Logbook 1(2) R/W 4, 61 unsigned8 (1) system error Logbook 1(2), - R 4, 62 unsigned8 (1) defaults loaded Not Logged(1), Logbook 1(2), Logbook 2(3) R/W 4, 63 unsigned8 (1) logbook erased Not Logged(1), Logbook 1(2), Logbook 2(3) R/W 4, 64 unsigned8 (1) init performed Logbook 1(2), - R 4, 65 unsigned8 (1) error on Not Logged(1), Logbook 1(2), Logbook 2(3) R/W 4, 66 unsigned8 (1) error off Not Logged(1), Logbook 1(2), Logbook 2(3) R/W 4, 67 unsigned8 (1) temperature adjust Logbook 1(2), Not Logged(1), Logbook 2(3) R/W 4, 68 unsigned8 (1) manual temperature Not Logged(1), Logbook 1(2), Logbook 2(3) R/W 4, 69 unsigned8 (1) aspot changed Logbook 1(2), Not Logged(1), Logbook 2(3) R/W 4, 70 unsigned8 (1) slope changed Logbook 1(2), Not Logged(1), Logbook 2(3) R/W 4, 71 unsigned8 (1) aspot 2 changed Logbook 1(2), Not Logged(1), Logbook 2(3) R/W 4, 72 unsigned8 (1) temperature coefficient Not Logged(1), Logbook 1(2), Logbook 2(3) R/W 4, 73 unsigned8 (1) input 1 imp after cal Logbook 1(2), Not Logged(1), Logbook 2(3) R/W 4, 74 unsigned8 (1) input 2 imp after cal Logbook 1(2), Not Logged(1), Logbook 2(3) R/W 4, 75 unsigned8 (1) delta ph changed Not Logged(1), Logbook 1(2), Logbook 2(3) R/W 4, 76 unsigned8 (1) delta t changed Not Logged(1), Logbook 1(2), Logbook 2(3) R/W 4, 77 unsigned8 (1) isothermal point changed Not Logged(1), Logbook 1(2), Logbook 2(3) R/W 4, 78 unsigned8 (1) zeropoint changed Logbook 1(2), Not Logged(1), Logbook 2(3) R/W 4, 79 unsigned8 (1) new sensor installed Logbook 1(2), Not Logged(1), Logbook 2(3) R/W 4, 80 unsigned8 (1) one point calibration Not Logged(1), Logbook 1(2), Logbook 2(3) R/W 4, 81 unsigned8 (1) two point calibration Not Logged(1), Logbook 1(2), Logbook 2(3) R/W 4, 82 unsigned8 (1)
Buffer Solution 1
Buffer1 ID 4.00 R/W 4, 105 unsigned8 (1) Buffer1 at 0°C 4.00 R/W 4, 106 float (4) Buffer1 at 5°C 4.00 R/W 4, 107 float (4) Buffer1 at 10°C 4.00 R/W 4, 108 float (4) Buffer1 at 15°C 4.00 R/W 4, 109 float (4)
Low(0) High(1) R/W 4, 167 (2.3) unsigned32 (4)
331, 546, 847 R/W 4, 160 unsigned16 (4)
331, 546, 847 R/W 4, 161 unsigned16 (4)
331, 546, 847 R/W 4, 162 unsigned16 (4)
IM 12A00A01-61E
Profibus 4-11
4-4-3-1. Transducer block parameters PH202 (continued)
Parameter default Alternatives Unit r/W Slot, Index data Type
(byte,bit) (bytes)
Buffer1 at 20°C 4.00 R/W 4, 110 float (4) Buffer1 at 25°C 4.01 R/W 4, 111 float (4) Buffer1 at 30°C 4.02 R/W 4, 112 float (4) Buffer1 at 35°C 4.02 R/W 4, 113 float (4) Buffer1 at 40°C 4.04 R/W 4, 114 float (4) Buffer1 at 45°C 4.05 R/W 4, 115 float (4) Buffer1 at 50°C 4.06 R/W 4, 116 float (4) Buffer1 at 55°C 4.08 R/W 4, 117 float (4) Buffer1 at 60°C 4.09 R/W 4, 118 float (4) Buffer1 at 65°C 4.11 R/W 4, 119 float (4) Buffer1 at 70°C 4.13 R/W 4, 120 float (4) Buffer1 at 75°C 4.15 R/W 4, 121 float (4) Buffer1 at 80°C 4.16 R/W 4, 122 float (4)
Buffer Solution 2
Buffer2 ID 7.00 R/W 4, 123 unsigned8 (1) Buffer2 at 0°C 6.98 R/W 4, 124 float (4) Buffer2 at 5°C 6.95 R/W 4, 125 float (4) Buffer2 at 10°C 6.92 R/W 4, 126 float (4) Buffer2 at 15°C 6.90 R/W 4, 127 float (4) Buffer2 at 20°C 6.88 R/W 4, 128 float (4) Buffer2 at 25°C 6.87 R/W 4, 129 float (4) Buffer2 at 30°C 6.85 R/W 4, 130 float (4) Buffer2 at 35°C 6.84 R/W 4, 131 float (4) Buffer2 at 40°C 6.84 R/W 4, 132 float (4) Buffer2 at 45°C 6.83 R/W 4, 133 float (4) Buffer2 at 50°C 6.83 R/W 4, 134 float (4) Buffer2 at 55°C 6.83 R/W 4, 135 float (4) Buffer2 at 60°C 6.84 R/W 4, 136 float (4) Buffer2 at 65°C 6.84 R/W 4, 137 float (4) Buffer2 at 70°C 6.85 R/W 4, 138 float (4) Buffer2 at 75°C 6.85 R/W 4, 139 float (4) Buffer2 at 80°C 6.86 R/W 4, 140 float (4)
Buffer Solution 3
Buffer3 ID 9 R/W 4, 141 unsigned8 (1) Buffer3 at 0°C 9.46 R/W 4, 142 float (4) Buffer3 at 5°C 9.40 R/W 4, 143 float (4) Buffer3 at 10°C 9.33 R/W 4, 144 float (4) Buffer3 at 15°C 9.28 R/W 4, 145 float (4) Buffer3 at 20°C 9.23 R/W 4, 146 float (4) Buffer3 at 25°C 9.18 R/W 4, 147 float (4) Buffer3 at 30°C 9.14 R/W 4, 148 float (4) Buffer3 at 35°C 9.10 R/W 4, 149 float (4) Buffer3 at 40°C 9.07 R/W 4, 150 float (4) Buffer3 at 40°C 9.04 R/W 4, 151 float (4) Buffer3 at 50°C 9.01 R/W 4, 152 float (4) Buffer3 at 55°C 8.99 R/W 4, 153 float (4) Buffer3 at 60°C 8.96 R/W 4, 154 float (4) Buffer3 at 65°C 8.94 R/W 4, 155 float (4) Buffer3 at 70°C 8.92 R/W 4, 156 float (4) Buffer3 at 75°C 8.90 R/W 4, 157 float (4) Buffer3 at 80°C 8.89 R/W 4, 158 float (4)
IM 12A00A01-61E
4-12 Profibus
4-4-3-2. Function Block Parameters SC202
Parameter default Alternatives Unit r/W Slot, Index data Type
(byte,bit) (bytes)
Primary value Parameters
Primary value Type Conductivity(113) conductivity(113), resistivity(116) R 4, 27 unsigned16 (2) Measured Value S/cm 4, 28 Status 4, 28 Conductivity sensor type 2 electrode(146) contact 2-electrode(146),
contact 4-electrode(163) R/W 4, 42 unsigned16 (2)
Secondary value Parameters
Value -20~250, 0~500 °C R 4, 36 DS-33 (5) Status 4, 36 Secondary value unit °C(1001) °C(1001), °F(1002) R/W 4, 37 unsigned16 (2) Sensor temp comp Automatic(3) auto(3) R 4, 38 unsigned8 (1) Sensor temp man value 25 0 °C R 4, 39 float (4) Temperature sensor Pt1000(148) Pt1000(148), Pt100(128), Ni100(149), 8k55(151), Pb36/JIS6k(152) R/W 4, 40 unsigned16 (2) Temp. connection type 2 2 R 4, 41 unsigned8 (1)
Tertiary value Parameters
Value S/cm R 4, 47 DS-33 (5) Status 4, 47
Compensation Parameters
Reference temperature 25 0~100, 32~212 °C R/W 4, 48 float (4) Compensation method NaCl(1) NaCl(1), TC(2), matrix(3) R/W 4, 49 unsigned8 (1) Temperature coefficient 2.1 0~3.5 %/°C R/W 4, 45 float (4) Matrix selection HCl cation(1), Ammonia pure water(2), Morpholine pure water(3), HCl 0-5%(4), Natrium hydroxide 0-5%, user defined(9) R/W 4, 50 unsigned8 (1) Tertiary comp. method NaCl(1) NaCl(1), TC(2), matrix(3) R/W 4, 51 unsigned8 (1) Tertiary temp. coefficient 2.1 0~3.5 R/W 4, 52 float (4)
Concentration Parameters
Concentration value 0 R 4, 46 DS-33 (5) Concentration status 4, 46 Concentration Measurement Concentration 0% 0 0~100 % R/W 4, 121 float (4) Concentration 100% 100 0~100 % R/W 4, 122 float (4) Concentration Table 0% 0 0~1.999, 0~999MΩ•cm S/cm R/W 4, 123 float (4) Concentration Table 5% 0.00005 S/cm R/W 4, 124 float (4) Concentration Table 10% 0.0001 S/cm R/W 4, 125 float (4) Concentration Table 15% 0.00015 S/cm R/W 4, 126 float (4) Concentration Table 20% 0.0002 S/cm R/W 4, 127 float (4) Concentration Table 25% 0.00025 S/cm R/W 4, 128 float (4) Concentration Table 30% 0.0003 S/cm R/W 4, 129 float (4) Concentration Table 35% 0.00035 0~1.999, 0~999MΩ•cm S/cm R/W 4, 130 float (4) Concentration Table 40% 0.0004 S/cm R/W 4, 131 float (4) Concentration Table 45% 0.00045 S/cm R/W 4, 132 float (4) Concentration Table 50% 0.0005 S/cm R/W 4, 133 float (4) Concentration Table 55% 0.00055 S/cm R/W 4, 134 float (4) Concentration Table 60% 0.0006 S/cm R/W 4, 135 float (4) Concentration Table 65% 0.00065 S/cm R/W 4, 136 float (4) Concentration Table 70% 0.0007 0~1.999, 0~999MΩ•cm S/cm R/W 4, 137 float (4) Concentration Table 75% 0.00075 S/cm R/W 4, 138 float (4) Concentration Table 80% 0.0008 S/cm R/W 4, 139 float (4) Concentration Table 85% 0.00085 S/cm R/W 4, 140 float (4) Concentration Table 90% 0.0009 S/cm R/W 4, 141 float (4) Concentration Table 95% 0.00095 S/cm R/W 4, 142 float (4) Concentration Table 100%
Disabled(1) Enabled(2) 4, 120 unsigned8(1)
0.001 S/cm R/W 4, 143 float (4)
IM 12A00A01-61E
Profibus 4-13
4-4-3-2. Function Block Parameters SC202 (continued)
Parameter default Alternatives Unit r/W Slot, Index data Type
(byte,bit) (bytes)
Calibration Parameters
Nominal cell constant 0.1 0.005~50 1/cm R/W 4, 30 float (4) Calibrated cell constant 0.1 0.005~50 1/cm R 4, 89 float (4) Calibration method Not calibrated(0) 1 point(107), 2 point(108) R/W 4, 34 unsigned8 (1) Diagnostic Settings Polarisation check Enabled(1) Disable(0) 4, 151(2.1) Polarization detected (E1) Hard fail(1) Soft fail(0) 4, 150(2.0)
E5 Limit 0.25 0~0.5, 0~10MΩ S R/W 4, 144 float (4) E6 Limit 0.000001 0~0.5, 0~10MΩ S R/W 4, 145 float (4)
Conductivity exceeds high limit (E5) Hard fail(1) Soft fail(0) 4, 150(2.4) Conductivity exceeds low limit (E6) Hard fail(1) Soft fail(0) 4, 150(2.5) Temp. sensor open (E7) Hard fail(1) Soft fail(0) 4, 150(2.6) Temp. sensor shorted (E8) USP Disabled Conductivity exceeds USP limit (E13) Soft fail(0) Hard fail(1) R/W 4, 150(1.4) unsigned32 (4)
Passcode Configuration
Passcode Maintenance 0 0, 111, 333, 777, 888, 123, 957, 331, 546, 847 R/W 4, 147 unsigned16 (2) Passcode Commissioning 0 0, 111, 333, 777, 888, 123, 957, 331, 546, 847 R/W 4, 148 unsigned16 (2) Passcode Service 0 0, 111, 333, 777, 888, 123, 957, 331, 546, 847 R/W 4, 149 unsigned16 (2)
display Settings
Display Resolution Auto Ranging(1) auto(1), x.xxx µS/cm(2), xx.xx µS/cm(3)
Auto return Enabled(1) Disable(0) R 4, 151(1.0)
Logbook Configuration
Power up Logbook 2(3) off(1), logbook1(2), logbook2(3) R/W 4, 59 unsigned8 (1) Power down Logbook 2(3) R/W 4, 60 unsigned8 (1) System error Logbook 1(2) R/W 4, 61 unsigned8 (1) Defaults loaded Not logged(1) R/W 4, 62 unsigned8 (1) Logbook erased Not logged(1) R/W 4, 63 unsigned8 (1) Init performed Logbook 1(2) R/W 4, 64 unsigned8 (1) Error on Not logged(1) R/W 4, 65 unsigned8 (1) Error off Not logged(1) R/W 4, 66 unsigned8 (1) Temperature adjust Logbook 1(2) R/W 4, 67 unsigned8 (1) Cell constant Logbook 1(2) R/W 4, 68 unsigned8 (1) Air calibration Logbook 1(2) R/W 4, 69 unsigned8 (1) Calibration Logbook 1(2) R/W 4, 70 unsigned8 (1) Reference temperature Not logged(1) R/W 4, 71 unsigned8 (1) Temperature coefficient 1 Not logged(1) R/W 4, 72 unsigned8 (1) Matrix selection Not logged(1) R/W 4, 73 unsigned8 (1) Temperature coefficient 2 Not logged(1) R/W 4, 74 unsigned8 (1)
User defined matrix
Temperature 1 0 -20~250, 0~500 °C R/W 4, 90 float (4) Temperature 2 25 °C R/W 4, 91 float (4) Temperature 3 50 °C R/W 4, 92 float (4) Temperature 4 75 °C R/W 4, 93 float (4) Temperature 5 100 °C R/W 4, 94 float (4) Solution 1 at Temp. 1 0.031 0~1.999, 0~999MΩ•cm S/cm R/W 4, 95 float (4) Solution 1 at Temp. 2 0.053 S/cm R/W 4, 96 float (4) Solution 1 at Temp. 3 0.076 S/cm R/W 4, 97 float (4) Solution 1 at Temp. 4 0.0975 S/cm R/W 4, 98 float (4) Solution 1 at Temp. 5 0.119 S/cm R/W 4, 99 float (4) Solution 2 at Temp. 1 0.061 0~1.999, 0~999MΩ•cm S/cm R/W 4, 100 float (4)
Hard fail(1) Soft fail(0) 4, 150(2.7)
xxx.x µS/cm(4), x.xxx mS/cm(5), xx.xx mS/
cm(6) xxx.x mS/cm(7), xxxx mS/cm(8) R 4, 146 unsigned8 (1)
IM 12A00A01-61E
4-14 Profibus
4-4-3-2. Function Block Parameters SC202 (continued)
Parameter default Alternatives Unit r/W Slot, Index data Type
(byte,bit) (bytes)
Solution 2 at Temp. 2 0.101 S/cm R/W 4, 101 float (4) Solution 2 at Temp. 3 0.141 S/cm R/W 4, 102 float (4) Solution 2 at Temp. 4 0.182 S/cm R/W 4, 103 float (4) Solution 2 at Temp. 5 0.223 S/cm R/W 4, 104 float (4) Solution 3 at Temp. 1 0.086 0~1.999, 0~999MΩ•cm S/cm R/W 4, 105 float (4) Solution 3 at Temp. 2 0.145 S/cm R/W 4, 106 float (4) Solution 3 at Temp. 3 0.207 S/cm R/W 4, 107 float (4) Solution 3 at Temp. 4 0.264 S/cm R/W 4, 108 float (4) Solution 3 at Temp. 5 0.318 S/cm R/W 4, 109 float (4) Solution 4 at Temp. 1 0.105 0~1.999, 0~999MΩ•cm S/cm R/W 4, 110 float (4) Solution 4 at Temp. 2 0.185 S/cm R/W 4, 111 float (4) Solution 4 at Temp. 3 0.286 S/cm R/W 4, 112 float (4) Solution 4 at Temp. 4 0.339 S/cm R/W 4, 113 float (4) Solution 4 at Temp. 5 0.41 S/cm R/W 4, 114 float (4) Solution 5 at Temp. 1 0.127 0~1.999, 0~999MΩ•cm S/cm R/W 4, 115 float (4) Solution 5 at Temp. 2 0.223 S/cm R/W 4, 116 float (4) Solution 5 at Temp. 3 0.319 S/cm R/W 4, 117 float (4) Solution 5 at Temp. 4 0.408 S/cm R/W 4, 118 float (4) Solution 5 at Temp. 5 0.495 S/cm R/W 4, 119 float (4)
4-4-3-3. Function Block Parameters ISC202
Parameter default Alternatives Unit r/W Slot, Index data Type
(byte,bit) (bytes)
Primary value Parameters
Primary value Type Conductivity(113) conductivity(113), resistivity(116) R 4, 27 unsigned16 (2) Measured Value S/cm 4, 28 Status 4, 28 Conductivity sensor type toroidal(147) R/W 4, 42 unsigned16 (2)
Secondary value Parameters
Value -20~140, 0~280 °C R 4, 36 DS-33 (5) Status 4, 36 Secondary value unit °C(1001) °C(1001), °F(1002) R/W 4, 37 unsigned16 (2) Sensor temp comp Automatic(3) auto(3) R 4, 38 unsigned8 (1) Sensor temp man value 25 0 °C R 4, 39 float (4) Temperature sensor NTC30k(154) Pt1000(148) R/W 4, 40 unsigned16 (2) Temp. connection type 2 2 R 4, 41 unsigned8 (1)
Tertiary value Parameters
Value S/cm R 4, 47 DS-33 (5) Status 4, 47 Compensation Parameters Reference temperature 25 0~100, 32~212 °C R/W 4, 48 float (4) Compensation method NaCl(1) NaCl(1), TC(2), matrix(3) R/W 4, 49 unsigned8 (1) Temperature coefficient 2.1 0~3.5 R/W 4, 45 float (4) Matrix selection H2SO4 0.5-5.0%, 0-100ºC(1), 2.5-25%,
0-100ºC(2) HCl 0-5%, 0-60ºC(3), 1-20%,
0-60ºC(4) HNO3 0.5-5.0%, 0-80ºC(5),
2.5-25%, 0-80ºC(6) NaOH 0.5-5.0%,
0-100ºC(7), 0.5-15%, 0-100ºC(8)
user defined(9) R/W 4, 50 unsigned8 (1) Tertiary comp. method NaCl(1) NaCl(1), TC(2), matrix(3) R/W 4, 51 unsigned8 (1) Tertiary temp. coefficient 2.1 0~3.5 R/W 4, 52 float (4)
Concentration Parameters
Concentration value 0 %/°C R 4, 46 DS-33 (5) Concentration status 4, 46 Concentration Measuremen
t Disabled(1) Enabled(2) % 4, 120 unsigned 8 (1)
IM 12A00A01-61E
Profibus 4-15
4-4-3-3. Function Block Parameters ISC202 (continued)
Parameter default Alternatives Unit r/W Slot, Index data Type
(byte,bit) (bytes)
Concentration 0% 0 0~100 % R/W 4, 121 float (4) Concentration 100% 100 0~100 % R/W 4, 122 float (4) Concentration Table 0% 0 0~1.999 S/cm R/W 4, 123 float (4) Concentration Table 5% 0.05 S/cm R/W 4, 124 float (4) Concentration Table 10% 0.1 S/cm R/W 4, 125 float (4) Concentration Table 15% 0.15 S/cm R/W 4, 126 float (4) Concentration Table 20% 0.2 S/cm R/W 4, 127 float (4) Concentration Table 25% 0.25 S/cm R/W 4, 128 float (4) Concentration Table 30% 0.3 S/cm R/W 4, 129 float (4) Concentration Table 35% 0.35 0~1.999 S/cm R/W 4, 130 float (4) Concentration Table 40% 0.4 S/cm R/W 4, 131 float (4) Concentration Table 45% 0.45 S/cm R/W 4, 132 float (4) Concentration Table 50% 0.5 S/cm R/W 4, 133 float (4) Concentration Table 55% 0.55 S/cm R/W 4, 134 float (4) Concentration Table 60% 0.6 S/cm R/W 4, 135 float (4) Concentration Table 65% 0.65 S/cm R/W 4, 136 float (4) Concentration Table 70% 0.7 0~1.999 S/cm R/W 4, 137 float (4) Concentration Table 75% 0.75 S/cm R/W 4, 138 float (4) Concentration Table 80% 0.8 S/cm R/W 4, 139 float (4) Concentration Table 85% 0.85 S/cm R/W 4, 140 float (4) Concentration Table 90% 0.9 S/cm R/W 4, 141 float (4) Concentration Table 95% 0.95 S/cm R/W 4, 142 float (4) Concentration Table 100%
Calibration Parameters
Nominal cell constant 1.88 0.005~50 1/cm R/W 4, 30 float (4) Calibrated cell constant 0.1 0.005~50 1/cm R 4, 89 float (4) Calibration method Not calibrated(0) 1 point(107), 2 point(108) R/W 4, 34 unsigned8 (1) Diagnostic Settings
E5 Limit 3 0~0.5Ω S R/W 4, 144 float (4) E6 Limit 0.000005 0~0.5Ω S R/W 4, 145 float (4)
Conductivity exceeds high limit (E5) Hard fail(1) Soft fail(0) 4, 150(2.4) Conductivity exceeds low limit (E6) Hard fail(1) Soft fail(0) 4, 150(2.5) Temp. sensor open (E7) Hard fail(1) Soft fail(0) 4, 150(2.6) Temp. sensor shorted (E8) Passcode Configuration Passcode Maintenance 0 0, 111, 333, 777, 888, 123, 957, 331, 546, 847 R/W 4, 147 unsigned16 (2) Passcode Commissioning 0 0, 111, 333, 777, 888, 123, 957, 331, 546, 847 R/W 4, 148 unsigned16 (2) Passcode Service 0 0, 111, 333, 777, 888, 123, 957, 331, 546, 847 R/W 4, 149 unsigned16 (2)
display Settings
Display Resolution Auto Ranging(1)
Auto return Enabled(1) Disable(0) R 4, 151(1.0)
Logbook Configuration
Power up Logbook 2(3) off(1), logbook1(2), logbook2(3) R/W 4, 59 unsigned8 (1) Power down Logbook 2(3) R/W 4, 60 unsigned8 (1) System error Logbook 1(2) R/W 4, 61 unsigned8 (1) Defaults loaded Not logged(1) R/W 4, 62 unsigned8 (1) Logbook erased Not logged(1) R/W 4, 63 unsigned8 (1) Init performed Logbook 1(2) R/W 4, 64 unsigned8 (1) Error on Not logged(1) R/W 4, 65 unsigned8 (1) Error off Not logged(1) R/W 4, 66 unsigned8 (1) Temperature adjust Logbook 1(2) R/W 4, 67 unsigned8 (1) Cell constant Logbook 1(2) R/W 4, 68 unsigned8 (1) Air calibration Logbook 1(2) R/W 4, 69 unsigned8 (1)
1 S/cm R/W 4, 143 float (4)
Hard fail(1) Soft fail(0) 4, 150(2.7)
auto(1), xxx.x µS/cm(4) x.xxx mS/cm(5), xx.xx
mS/cm(6) xxx.x mS/cm(7), xxxx mS/cm(8)
R 4, 146 unsigned8 (1)
IM 12A00A01-61E
4-16 Profibus
4-4-3-3. Function Block Parameters ISC202 (continued)
Parameter default Alternatives Unit r/W Slot, Index data Type
(byte,bit) (bytes)
Calibration Logbook 1(2) R/W 4, 70 unsigned8 (1) Reference temperature Not logged(1) R/W 4, 71 unsigned8 (1) Temperature coefficient 1 Not logged(1) R/W 4, 72 unsigned8 (1) Matrix selection Not logged(1) R/W 4, 73 unsigned8 (1) Temperature coefficient 2 Not logged(1) R/W 4, 74 unsigned8 (1)
User defined matrix
Temperature 1 0 -20~140, 0~280 °C R/W 4, 90 float (4) Temperature 2 25 °C R/W 4, 91 float (4) Temperature 3 50 °C R/W 4, 92 float (4) Temperature 4 75 °C R/W 4, 93 float (4) Temperature 5 100 °C R/W 4, 94 float (4) Solution 1 at Temp. 1 0.0338 0~1.999 S/cm R/W 4, 95 float (4) Solution 1 at Temp. 2 0.0470 S/cm R/W 4, 96 float (4) Solution 1 at Temp. 3 0.0575 S/cm R/W 4, 97 float (4) Solution 1 at Temp. 4 0.0637 S/cm R/W 4, 98 float (4) Solution 1 at Temp. 5 0.0680 S/cm R/W 4, 99 float (4) Solution 2 at Temp. 1 0.0635 0~1.999 S/cm R/W 4, 100 float (4) Solution 2 at Temp. 2 0.0923 S/cm R/W 4, 101 float (4) Solution 2 at Temp. 3 0.1120 S/cm R/W 4, 102 float (4) Solution 2 at Temp. 4 0.1260 S/cm R/W 4, 103 float (4) Solution 2 at Temp. 5 0.1380 S/cm R/W 4, 104 float (4) Solution 3 at Temp. 1 0.0950 0~1.999 S/cm R/W 4, 105 float (4) Solution 3 at Temp. 2 0.1350 S/cm R/W 4, 106 float (4) Solution 3 at Temp.3 0.1660 S/cm R/W 4, 107 float (4) Solution 3 at Temp.4 0.1890 S/cm R/W 4, 108 float (4) Solution 3 at Temp. 5 0.2060 S/cm R/W 4, 109 float (4) Solution 4 at Temp. 1 0.1240 0~1.999 S/cm R/W 4, 110 float (4) Solution 4 at Temp. 2 0.1780 S/cm R/W 4, 111 float (4) Solution 4 at Temp. 3 0.2200 S/cm R/W 4, 112 float (4) Solution 4 at Temp. 4 0.2490 S/cm R/W 4, 113 float (4) Solution 4 at Temp. 5 0.2730 S/cm R/W 4, 114 float (4) Solution 5 at Temp. 1 0.1540 0~1.999 S/cm R/W 4, 115 float (4) Solution 5 at Temp. 2 0.2180 S/cm R/W 4, 116 float (4) Solution 5 at Temp. 3 0.2700 S/cm R/W 4, 117 float (4) Solution 5 at Temp. 4 0.3070 S/cm R/W 4, 118 float (4) Solution 5 at Temp. 5 0.3360 S/cm R/W 4, 119 float (4)
IM 12A00A01-61E
Profibus 4-17
4-4-3-4. Function Block Parameters dO202
Parameter default Alternatives Unit r/W Slot, Index data Type
(byte,bit) (bytes)
Primary value parameters
Primary value Type Measured Value ppm R 4,28 DS-33 (5) Status R 4,28 DS-33 (5) Sensor type DO polarographic (2) galvanic (1) R/W 4,31 unsigned16 (2) Primary value unit ppm (1423) % saturation (1342), ppb (1424) R 4,30 unsigned16 (2)
Secondary value Parameters
Value °C R 4,43 DS-33 (5) Status R 4,43 DS-33 (5) Secondary value unit °C (1001) °F (1002) R/W 4,44 unsigned16 (2) Sensor temp comp auto (2) manual (1) R/W 4,45 unsigned8 (1) Sensor temp man value 25 °C R/W 4,46 float (4) Temperature sensor NTC22k (160) Pt1000 (148), Pb36 (152) R/W 4,47 unsigned16 (2)
Other value Parameters
Sensor current nA R 4,49 float (4) %Saturation % R 4,50 float (4)
Calibration Parameters
Zero calibration disabled (0) enabled (1) R/W 4,97 (1.0) unsigned32 (4) Zero current 0.0 -199.9 ~ 199.9 nA R/W 4,33 float (4) Zero current limit 199.9 nA R/W 4,51 float (4) Sensitivity 7.5 1.0 ~ 1999.9 nA/
Stable time 60 5 ~ 600 s R/W 4,35 float (4) Stable value (sensitivity) 0.1 0 ~ 50.0 ppm R/W 4,36 float (4) Stable value (zero) 0.1 0 ~ 50.0 ppm R/W 4,37 float (4) Calibration pressure 101.3 (kPa) 0 ~ 999 kPa R/W 4,39 float (4)
Compensation Parameters
Salinity compensation disabled (0) enabled (1) R/W 4,97 (1.4) unsigned32 (4) Salinity 0.0 0 ~ 99.9 ppt R/W 4,38 float (4) %Saturation pressure 101.3 (kPa) 0 ~ 999 kPa R/W 4,41 float (4)
diagnostic Settings
Calibration not stable (E1) Hard fail (1) Soft fail (0) R/W 4,96 (1.0) unsigned32 (4) Zero out of limits (E2) Hard fail (1) Soft fail (0) R/W 4,96 (1.1) unsigned32 (4) Temp. sensor open (E7) Hard fail (1) Soft fail (0) R/W 4,96 (1.6) unsigned32 (4) Temp. sensor shorted (E8) Primary value exceeds limits (E9) Hard fail (1) Soft fail (0) R/W 4,96 (2.0) unsigned32 (4) Calibr. timer expired (E16) Soft fail (0) Hard fail (1) R/W 4,96 (2.7) unsigned32 (4)
Passcode Configuration
Passcode Maintenance 000 111,333,777,888,123,957,331,546,847 R/W 4,62 float (4) Passcode Commisioning 000 111,333,777,888,123,957,331,546,848 R/W 4,63 float (4) Passcode Service 000 111,333,777,888,123,957,331,546,849 R/W 4,64 float (4)
display Settings
Manual pressure in maintenance menu disabled (0) enabled (1) R/W 4,97 (1.7) unsigned32 (4) Auto return enabled (1) disabled (0) R/W 4,97 (1.3) unsigned32 (4)
Logbook Configuration
Power up logbook2 (3) not logged (1), logbook1 (2) R/W 4,69 unsigned8 (1) Power down logbook2 (3) not logged (1), logbook1 (2) R/W 4,70 unsigned8 (1) System error logbook1 (2) - R 4,71 unsigned8 (1) Defaults loaded not logged (1) logbook1 (2), logbook2 (3) R/W 4,72 unsigned8 (1) Logbook erased not logged (1) logbook1 (2), logbook2 (3) R/W 4,73 unsigned8 (1) Init performed logbook1 (2) - R 4,74 unsigned8 (1) Error on not logged (1) logbook1 (2), logbook2 (3) R/W 4,75 unsigned8 (1) Error off not logged (1) logbook1 (2), logbook2 (3) R/W 4,76 unsigned8 (1) Temperature adjust logbook1 (2) not logged (1), logbook2 (3) R/W 4,77 unsigned8 (1)
dissolved oxyg. (65520)
Hard fail (1) Soft fail (0) R/W 4,96 (1.7) unsigned32 (4)
R 4,27
ppm R/W 4,34 float (4)
unsigned16
(2)
IM 12A00A01-61E
4-18 Profibus
4-4-3-4. Function Block Parameters dO202 (continued)
Parameter default Alternatives Unit r/W Slot, Index data Type
(byte,bit) (bytes)
Manual temperature logbook1 (2) not logged (1), logbook2 (3) R/W 4,78 unsigned8 (1) Manual calibration (0%) logbook1 (2) not logged (1), logbook2 (3) R/W 4,83 unsigned8 (1) Manual calibration (100%) logbook1 (2) not logged (1), logbook2 (3) R/W 4,79 unsigned8 (1) Air calibration (0%) logbook1 (2) not logged (1), logbook2 (3) R/W 4,92 unsigned8 (1) Air calibration (100%) logbook1 (2) not logged (1), logbook2 (3) R/W 4,80 unsigned8 (1) H2O calibration (0%) logbook1 (2) not logged (1), logbook2 (3) R/W 4,82 unsigned8 (1) H2O calibration (100%) logbook1 (2) not logged (1), logbook2 (3) R/W 4,81 unsigned8 (1) Zero setting changed logbook1 (2) not logged (1), logbook2 (3) R/W 4,84 unsigned8 (1) Sensitivity changed logbook1 (2) not logged (1), logbook2 (3) R/W 4,90 unsigned8 (1) Delta t for calibation changed not logged (1) logbook1 (2), logbook2 (3) R/W 4,86 unsigned8 (1) Delta PV for calibration changed (0%) not logged (1) logbook1 (2), logbook2 (3) R/W 4,85 unsigned8 (1) Delta PV for calibration changed (100%) not logged (1) logbook1 (2), logbook2 (3) R/W 4,91 unsigned8 (1) Salinity compensation changed not logged (1) logbook1 (2), logbook2 (3) R/W 4,87 unsigned8 (1) %Saturation pressure changed not logged (1) logbook1 (2), logbook2 (3) R/W 4,88 unsigned8 (1) Calibration pressure changed not logged (1) logbook1 (2), logbook2 (3) R/W 4,89 unsigned8 (1) New sensor installed logbook1 (2) not logged (1), logbook2 (3) R/W 4,93 unsigned8 (1)
IM 12A00A01-61E
Profibus 4-19

4-4-4 Methods

The EXA supports methods. A method is a tool to provide the user with a “step by step” user interface for changing settings and providing information.
The following screendumps are derived from a Siemens PDM package.
display and y-t diagram methods
The EXA provides three process values simultaneously. These can be viewed with the display method for bargraphs or with Y-t diagram.
Clock method
This method can be used to set the time of the EXA.
IM 12A00A01-61E
4-20 Profibus
Status method
This method can be used to obtain general information of the EXA like serial number and software revision. This method also provides the user with the current status of the EXA. This can be usefull when the EXA shows errors.
Logbook method
One of the powerfull features of the EXA is the logbook functionality. All events can be stored in one of the two logbooks. To read (upload) the logbook information, simply use this method and the user is provided with all stored events. Each logbook can obtain up to 50 events.
IM 12A00A01-61E
APPENDIX 1. LINK MASTER FUNCTIONS 5-1
LM
Node address:
0x15
SlotTime = 5
LM
Node address:
0x16
SlotTime = 5
Node address: 0x14 SlotTime = 5
Basic device
Node address:
0xF1
Basic device
Node address:
0xF2
Basic device
Node address:
0xF3
Basic device
Node address:
0xF4
LAS
There are 3 LMs on this segment.
LM

APPENdIX 1. LINK MASTEr FUNCTIONS

A1-1 Link Active Scheduler

A link active scheduler (LAS) is a deterministic, centralized bus scheduler that can control communications on an H1 fieldbus segment. There is only one LAS on an H1 fieldbus segment. A EXA202 supports the following LAS functions.
• PN transmission: Identifies a fieldbus device newly connected to the same fieldbus segment. PN is short for Probe Node.
• PT transmission: Passes a token governing the right to transmit, to a fieldbus device on the same segment. PT is short for Pass Token.
• CD transmission: Carry out a scheduled transmission to a fieldbus device on the same segment. CD is short for Compel Data.
• Time synchronization: Periodically transmits the time data to all fieldbus devices on the segment and returns the time data in response to a request from a device.
• Live list equalization: Sends the live list data to link masters on the same segment.
• LAS transfer: Transfers the right to be the LAS on the segment to another link master.

A1-2 Link Master

A link master (LM) is any device containing a link active scheduler. There must be at least one LM on a segment. When the LAS on a segment has failed, another LM on the same segment starts working as the LAS.
Figure A1-1. Example of Fieldbus configuration-3 LMs on Same Segment
IM 12A00A01-61E
5-2 APPENDIX 1. LINK MASTER FUNCTIONS
LM
Node address:
0x15
SlotTime = 5
LM
Node address:
0x16
SlotTime = 5
Node address: 0x14 SlotTime = 5
Basic device
Node address:
0xF1
Basic device
Node address:
0xF2
Basic device
Node address:
0xF3
Basic device
Node address:
0xF4
LAS
LAS
In the event that the current LAS in this segment (node address 0x14) fails, the LM with the address of 0x15 takes its place to become the LAS.
LM
Not used
LM device
Not used
Basic device
Default address
Portable-device address
V (FUN)
V (FUN) + V (NUN)
V (NUN)
0xFF
0xFC
0xFB
0xF8
0xF7
0x00
0x0F
0x10 0x13
0x14
Bridge device
Sub-
index
Element
Device1Device2Device
3
Description
1
3
6
SlotTime MaxResponse
Delay MinInterPdu
Delay
4
3
4
8
6
8
10
3
12
20
5
10
Capability value for V(ST)
Capability value for V(MRD)
Capability value for V(MID)
DlmeBasicInfo (EXA202 Index 361 (SM))
EXA202
Subindex
Element
ConfiguredLinkSettingsRecord (EXA202 Index 369 (SM))
Description
Setting
(Default)
V (ST) V (MRD) V (MID)
20
6
12
(4095) (5) (12)
SlotTime MaxResponseDelay MinInterPduDelay
1 3 6

A1-3 Transfer of LAS

There are two procedures for an LM to become the LAS:
• If the LM whose value of [V(ST)3V(TN)] is the smallest on a segment, with the exception of the current LAS, judges that there is no LAS on the segment, in such a case as when the segment has started up or when the current LAS has failed, the LM declares itself as the LAS, then becomes the LAS. (With this procedure, an LM backs up the LAS as shown in the following figure.)
• The LM whose value of [V(ST)3V(TN)] is the smallest on a segment, with the exception of the current LAS, requests the LAS on the same segment to transfer the right of being the LAS, then becomes the LAS.
Figure A1-2. Backup of LAS
To set up a EXA202 as a device that is capable of backing up the LAS, follow the procedure below.
NOTE: When changing the settings in a EXA202,
add the EXA202 to the segment in which an LAS is running. After making changes to the settings, do not turn off the power to the EXA202 for at least 60 seconds.
(1) Set the node address of the EXA202.
In general, use an address from 0x14 to [V(FUN) – 1].
(2) In the LAS settings of the EXA202, set the
values of V(ST), V(MRD), and V(MID) to the same as the respective lowest capability values in all the devices within the segment. An example is shown below.
In this case, set SlotTime, MaxResponseTime, and MinInterPduDelay as follows:
Figure A1-3. Node Address ranges
IM 12A00A01-61E
(3) In the LAS settings of the EXA202, set the
Subindex
Element
ConfiguredLinkSettingsRecord (EXA202 Index 369 (SM))
Description
Default Value
V (FUN) V (NUN)
0x25 0xBA
FirstUnpolledNodeId NumConsecUnpolledNodeId
4 7
No. Function Description
LM initialization
Startup of other nodes (PN and Node Activation SPDU transmissions)
PT transmission (including final bit monitoring)
CD transmission
Time synchronization
Domain download server
Live list equalization
LAS transfer
Reading/writing of NMIB for LM
Round Trip Delay Reply (RR) Reply to DLPDU
Long address
1
2
3
4
5
6
7
8
9
10
11
When a fieldbus segment starts, the LM with the smallest [V(ST) × V(TN)] value within the segment becomes the LAS. At all times, each LM is checking whether or not a carrier is on the segment.
Transmits a PN (Probe Node) message, and Node Activation SPDU message to devices which return a new PR (Probe Response) message.
Passes a PT (Pass Token) message to devices included in the live list sequentially, and monitors the RT (Return Token) and final bit returned in reply to the PT.
Transmits a CD (Compel Data) message at the scheduled times.
Supports periodic TD (Time Distribution) transmissions and transmissions of a reply to a CT (Compel Time).
Sets the schedule data. The schedule data can be equalized only when the Domain Download command is carried out from outside the LM in question. (The version of the schedule is usually monitored, but no action takes place, even when it changes.)
Transmits SPDU messages to LMs to equalize live lists.
Transfers the right of being the LAS to another LM.
See Section A1-5.
Not yet supported in the current version.
Not yet supported in the current version.
values of V(FUN) and V(NUN) so that they include the node addresses of all nodes within the same segment. (See also Figure A1-3.)

A1-4 LM Functions

APPENDIX 1. LINK MASTER FUNCTIONS 5-3
IM 12A00A01-61E
5-4 APPENDIX 1. LINK MASTER FUNCTIONS
362 363
364 365 366
367 368
369
0 1 MaxSchedulingOverhead 2 DefMinTokenDelegTime 3 DefTokenHoldTime 4 TargetTokenRotTime 5 LinkMaintTokHoldTime 6 TimeDistributionPeriod 7 MaximumInactivityToClaimLasDelay 8
LasDatabaseStatusSpduDistributionPeriod
0 1 Element1 2 Element2 3 Element3 4 Element4 5 Element5 6 Element6 7 Element7 8 Element8
0 1 SlotTime 2 PerDlpduPhlOverhead 3 MaxResponseDelay 4 FirstUnpolledNodeId 5 ThisLink 6 MinInterPduDelay 7 NumConseeUnpolledNodeId 8 PreambleExtension 9 PostTransGapExtension 10 MaxInterChanSignalSkew 11 TimeSyncClass 0 1 SlotTime 2 PerDlpduPhlOverhead 3 MaxResponseDelay 4 FirstUnpolledNodeId 5 ThisLink 6 MinInterPduDelay 7 NumConseeUnpolledNodeId 8 PreambleExtension 9 PostTransGapExtension 10 MaxInterChanSignalSkew 11 TimeSyncClass
0x04
0 100 300 4096 400 5000 8 6000 – – 0x0000×16, 0x012c×16 0x012c×5, 0x0000×27 0x0000×32 0x0000×32 0x0000×32 0x0000×32 0x0000×31, 0x012c 0x012c×32 0x02 0x01
4095 4 5 37 0 12 186 2 1 0 4
LAS: True = 0xFF; non-LAS: False = 0x00
0x01 (basic device); 0x02 (LM) Settings for LAS
RW RW
RW
R
RW
RW
R
RW
Meanings of Access column entries: RW = read/write possible; R = read only
DLME_LINK_MASTER_CAPABILITIES_VARIABLE DLME_LINK_MASTER_
INFO_RECORD
PRIMARY_LINK_MASTER_FLAG_VARIABLE LIVE_LIST_STATUS_ARRAY_VARIABLE MAX_TOKEN_HOLD_
TIME_ARRAY
BOOT_OPERAT_FUNCTIONAL_CLASS CURRENT_LINK_
SETTING_RECORD
CONFIGURED_LINK_ SETTING_RECORD
Index
(SM)
Sub-parameter Name
Access
(Sub Index)
Parameter Name
Default Factory
Setting
Remarks

A1-5 LM Parameters

A1-5-1 LM Parameter List

The tables below show LM parameters of a EXA202.
IM 12A00A01-61E
APPENDIX 1. LINK MASTER FUNCTIONS 5-5
370
371
372
373 374
375
376
377 378
0 1 ChannelStatisticsSupported 2 MediumAndDataRatesSupported 3 IecVersion 4 NumOfChannels 5 PowerMode 0 1 channel-1 2 channel-2 3 channel-3 4 channel-4 5 channel-5 6 channel-6 7 channel-7 8 channel-8 0 1 InterfaceMode 2 LoopBackMode 3 XmitEnabled 4 RcvEnabled 5 PreferredReceiveChannel 6 MediaTypeSelected 7 ReceiveSelect
0 1 NumOfSchedules 2 NumOfSubSchedulesPerSchedule 3 ActiveScheduleVersion 4 ActiveSheduleOdIndex 5 ActiveScheduleStartingTime 0 1 Version 2 MacrocycleDuration 3 TimeResolution 0 1 Version 2 MacrocycleDuration 3 TimeResolution
0x00 0x4900000000000000 1 (0x1) 1 (0x1) 0 (0x0)
0 (0x0) 128 (0x80) 128 (0x80) 128 (0x80) 128 (0x80) 128 (0x80) 128 (0x80) 128 (0x80)
0 (0x0) 0 (0x0) 1 (0x1) 1 (0x1) 1 (0x1) 73 (0x49) 1 (0x1)
0 1 0 0 0
0 0 0
0 0 0
Read/write impossible. Get-OD possible. Read/write impossible. Get-OD possible.
R
R
R
RW
R
R
R
Index
(SM)
Sub-parameter Name
Access
(Sub Index)
Parameter Name
Default Factory
Setting
Remarks
PLME_BASIC_ CHARACTERISTICS
CHANNEL_STATES
PLME_BASIC_INFO
LINK_SCHEDULE_ACTIVATION_VARIABLE LINK_SCHEDULE_LIST_
CHARACTERISTICS_ RECORD
DLME_SCHEDULE_ DESCRIPTOR.1
DLME_SCHEDULE_ DESCRIPTOR.2
DOMAIN.1 DOMAIN.2
IM 12A00A01-61E
5-6 APPENDIX 1. LINK MASTER FUNCTIONS
Meaning
Bit
Position
Value
Description
1
0
0
LAS Schedule in Non-volatile Memory
Last Values Record Supported
Link Master Statistics Record Supported
Whether the LAS schedule can (= 1) or cannot (= 0) be saved to the non-volatile memory
Whether to support (= 1) or not to support (= 0) LastValuesRecord.
Whether to support (= 1) or not to support (= 0) DlmeLinkMasterStatisticsRecord.
B3: 0x04
B2: 0x02
B1: 0x01
Element
Sub-
index
Descrip-
tion
Size
[bytes]
1 2 2 2 2 4 2 2
V(MSO) V(DMDT) V(DTHT) V(TTRT) V(LTHT) V(TDP) V(MICD) V(LDDP)
MaxSchedulingOverhead DefMinTokenDelegTime DefTokenHoldTime TargetTokenRotTime LinkMaintTokHoldTime TimeDistributionPeriod MaximumInactivityToClaimLasDelay LasDatabaseStatusSpduDistributionPeriod
1 2 3 4 5 6 7 8
0x00 00 84 00 00 00 00 00 00 00 00 00 00 00
00 00 00 00 00 00 00 00 00 00 00 00 00 00 00 00 00 00
Bit correspondences: 0 0 0 0 0 0 0 0 0 0 0
0×00
0 0 0 0 0 1 0 0 0 0 1 0 0...
0×10 0×15
Element
Sub-
index
Descrip-
tion
Size
[bytes]
2 1 1 1 2 1 1 1 1 1 1
V(ST) V(PhLO) V(MRD) V(FUN) V(TL) V(MID) V(NUN) V(PhPE) V(PhGE) V(PhIS) V(TSC)
SlotTime PerDlpduPhlOverhead MaxResponseDelay FirstUnpolledNodeId ThisLink MinInterPduDelay NumConsecUnpolledNodeId PreambleExtension PostTransGapExtension MaxInterChanSignalSkew TimeSyncClass
1 2 3 4 5 6 7 8 9
10
11

A1-5-2 descriptions for LM Parameters

The following describes LM parameters of a EXA202 transmitter.
NOTE: Do not turn off the power to the EXA202
for 60 seconds after making a change to its parameter settings.
(1) dlmeLinkMasterCapabilitiesvariable
(2) dlmeLinkMasterInforecord
(5) MaxTokenHoldTimeArray
An 8- by 64-byte array variable, in which each set of 2 bytes represents the delegation time (set as an octet time) assigned to a device. The delega­tion time denotes a time period that is given to a device by means of a PT message sent from the LAS within each token circulation cycle.
The leading 2 bytes correspond to the device address 0x00, and the final 2 bytes to the device address 0xFF. Specify the subindex to access this parameter.
(6) BootOperatFunctionalClass
Writing 1 to this parameter in a device and restart­ing the device causes the device to start as a basic device. On the contrary, writing 2 to this parameter and restarting the device causes the device to start as an LM.
(3) PrimaryLinkMasterFlagvariable
Explicitly declares the LAS. Writing “true” (0xFF) to this parameter in a device causes that device to attempt to become the LAS. However, a request of writing “true” to this parameter in a device is rejected if the value of the same parameter in any other device that has a smaller node address within the same segment is true.
(4) LiveListStatusArrayvariable
A 32-byte variable, in which each bit represents the status of whether a device on the same seg­ment is live or not. The leading bit corresponds to the device address 0x00, and final bit to 0xFF. The value of LiveListStatusArrayVariable in the case where devices having the addresses 0x10 and 0x15 in the fieldbus segment is shown below.
IM 12A00A01-61E
(7) CurrentLinkSettingrecord and Config-
uredLinkSettingsrecord
CurrentLinkSettingRecord indicates the bus parameter settings currently used. ConfiguredL inkSettingsRecord indicates the bus parameter settings to be used when the device becomes the LAS. Thus, when a device is the LAS, its CurrentLinkSettingRecord and ConfiguredLinkSetti ngsRecord have the same values.
APPENDIX 1. LINK MASTER FUNCTIONS 5-7
Element Description
Sub-
index
Size
[bytes]
2
1 1
1 2 1
1
1 1 1
Indicates the capability value for V(ST) of the device.
V(PhLO) Indicates the capability
value for V(MRD) of the device.
V(TN), node address V(TL), link-id Indicates the capability
value for V(MID) of the device.
Indicates the capability value for V(TSC) of the device.
V(PhPE) V(PhGE) V(PhIS)
SlotTime
PerDlpduPhlOverhead MaxResponseDelay
ThisNode ThisLink MinInterPduDelay
TimeSyncClass
PreambleExtension PostTransGapExtension MaxInterChanSignalSkew
1
2 3
4 5 6
7
8 9
10
Element Value Description
Sub-
index
Size
[bytes]
1
8
2
1
1
Statistics data are not supported.
Wire medium, voltage mode, and 31.25 kbps are supported.
IEC 4.3 is supported.
0: Bus-powered; 1: Self-powered
Channel Statistics Supported
Medium AndData Rates Supported
IceVersion
NumOf Channels
Power Mode
0
0x4900000000000000
0x0403
1
0
1
2
3
4
5
Element
Value
Description
Sub-
index
Size
[bytes]
1
1 1 1 1 1 1 1
In Use, No Bad since last read, No Silent since last read, No Jabber since last read, Tx Good, Rx Good
Unused Unused Unused Unused Unused Unused Unused
Channel 1
Channel 2 Channel 3 Channel 4 Channel 5 Channel 6 Channel 7 Channel 8
0x00
0x80 0x80 0x80 0x80 0x80 0x80 0x80
1
2 3 4 5 6 7 8
Element
Value
Description
Sub-
index
Size
[bytes]
1
1
1 1 1
1
1
0: Half duplex; 1: Full duplex
0: Disabled; 1: MAU; 2: MDS
Channel 1 is enabled. Channel 1 is enabled. Channel 1 is used for
reception. Wire medium, voltage
mode, and 31.25 kbps are selected.
Channel 1 is used for reception.
InterfaceMode
LoopBackMode
XmitEnabled RcvEnebled PreferredReceive
Channel MediaType
Selected
ReceiveSelect
0
0
0x01 0x01 0x01
0x49
0x01
1
2
3 4 5
6
7
Element Description
Sub-
index
Size
[bytes]
1
1
2
2
6
Indicates the total number of LAS schedules that have been downloaded to the domain.
Indicates the maximum number of sub-schedules an LAS schedule can contain. (This is fixed to 1 in the Yokogawa communication stacks.)
Indicates the version number of the schedule currently executed.
Indicates the index number of the domain that stores the schedule currently executed.
Indicates the time when the current schedule began being executed.
NumOf Schedules
NumOfSub SchedulesPer Schedule
ActiveSchedule Version
ActiveSchedule OdIndex
ActiveSchedule StaringTime
1
2
3
4
5
(8) dlmeBasicInfo
(9) PlmeBasicCharacteristics
(11) PlmeBasicInfo
(12) LinkScheduleActivationvariable
Writing the version number of an LAS sched­ule, which has already been downloaded to the domain, to this parameter causes the correspond­ing schedule to be executed. On the other hand, writing 0 to this parameter stops execution of the active schedule.
(10) ChannelStates
(13) LinkScheduleListCharacteristicsrecord
IM 12A00A01-61E
5-8 APPENDIX 1. LINK MASTER FUNCTIONS
Element Description
Sub-
index
Size
[bytes]
2
4
2
Indicates the version number of the LAS schedule downloaded to the corresponding domain.
Indicates the macro cycle of the LAS schedule downloaded to the corresponding domain.
Indicates the time resolution that is required to execute the LAS schedule downloaded to the corresponding domain.
Version
Macrocycle Duration
TimeResolution
1
2
3
(14) dlmeScheduledescriptor
This parameter exists for the same number as the total number of domains, and each describes the LAS schedule downloaded to the correspond­ing domain. For the domain to which a schedule has not yet been downloaded, the values in this parameter are all zeros.
(15) domain
Read/write: impossible; get-OD: possible Carrying out the GenericDomainDownload com­mand from a host writes an LAS schedule to the domain.
A1-6 FAQs Q1. When the LAS stops, a EXA202 does not
back it up by becoming the LAS. Why?
A1-1. Is that EXA202 running as an LM? Check
that the value of BootOperatFunctionalClass (index 367) is 2 (indicating that it is an LM).
Q3. On a segment where a EXA202 works as
the LAS, another device cannot be con­nected. Why?
A3-1. Check the following bus parameters that indi
­cate the bus parameter as being the LAS for the EXA202 and the capabilities of being the LAS for the device that cannot be connected:
• V(ST), V(MID), and V(MRD) of EXA202:
ConfiguredLinkSettingsRecord (index 369)
• V(ST), V(MID), and V(MRD) of problematic
device: DlmeBasicInfo
Then, confirm that the following conditions are
met:
EXA202 Problematic Device V(ST) > V(ST)
V(MID) > V(MID) V(MRD) > V(MRD)
A3-2. Check that the node address of the problem
­atic device does not lie within either 0x00 to 0x10 or the range of unused (unpolled) node addresses determined by the EXA202’s LM parameter settings, which is 0x00 to 0x10 or V(FUN) to V(FUN) + V(NUM). (Refer to Sec­tion 3-3-2, “Network Definition.”)
A1-2. Check the values of V(ST) and V(TN) in all
LMs on the segment and confirm that the fol­lowing condition is met:
EXA202 Other LMs V(ST) 3 V(TN) < V(ST) 3 V(TN)
Q2. How can I make a EXA202 become the
LAS?
A2-1. Check that the version numbers of the active
schedules in the current LAS and the EXA202 are the same by reading:
LinkScheduleListCharacteristicsRecord
(index 374 for a EXA202)
- ActiveScheduleVersion (subindex 3) A2-2. Make the EXA202 declare itself as and be
come the LAS by writing:
• 0x00 (false) to PrimaryLinkMasterFlagVari able in the current LAS; and
• 0xFF (true) to PrimaryLinkMasterFlagVari able (index 364) in the EXA202.
IM 12A00A01-61E
-
-
-
Revision Re�o��Re�o��
Manual Title : EXA202 Fieldbus Communication Manual Number : IM 12A00A01-61E
E�ition Date Rema�k (s)
1st Apr. 2007 Newly published 2nd Sep.2007 Revised as follows
p1-1 IM No. to be refered revised; p3-2 Figure No. to be refered corrected; p3-4 Some error correction of Table 3.1; p3-7, p3-12 Section No. to be refered corrected; p3-19 Error of reference Table No. for XD_SCALE corrected; p3-20 Some error correction; p3-30 Error of reference Table No. for XD_ SCALE corrected; p3-34, p3-37, p3-40 Some error correction; p4-8 Some error correction of Sec 4-4-1; p5-3 Some error correction.
IM 12A00A01-61E
User’s
EXA 202
Manual
Fieldbus Communication

Supplement

Thank you for selecting our EXA202 Fieldbus Communication. User's Manual, IM12A00A01-61E, 2nd Edition, supplied with the product, some revisions/additions have been made. Please replace the corresponding pages in your copy with the attached, revised pages.
Revisions:
- Page 3-19, Some revision of Table 3.19 and 3.20, because Unit Index Code has been corrected.
- Page 3-54, Operational Precaution with FieldMate added.
All Rights Reserved, Copyright © 2009, 2nd Edition: Oct. 2009 (YK)
Subject to change without notice.
IM12A00A01-61E
2nd Edition
Foundation Fieldbus 3-19 3-19
DO202
1: Dissolved Oxygen, 2: Temperature, 3: Percent Saturation, 4: Sensor Current
Channel Value Unit
1 primary_value primary_value_range.units
2 secondary_value secondary_value_unit
3 percent_saturation %
4 sensor_current nA
XD_SCALE/OUT_SCALE:
Scaling information is used for two purposes. Display devices need to know the range for bar graphs and trending, as well as the units code. Control blocks need to know the range to use internally as percent of span, so that the tuning constants may remain dimensionless. This is converted back to a number with units by using the range of OUT_SCALE. The AI block has the parameter XD_SCALE to define the units expect­ed from the transducer.
Transducer scaling (XD_SCALE) is applied to the value from the channel to produce the FIELD_VAL in per­cent. The XD_SCALE units code must match the channel units code. The EXA transmitter does this automatically when the Service Codes are changed. See table 3.18 to 3.21 for the Service codes and their results. If L_TYPE is set to Indirect or Ind Sqr Root, OUT_SCALE determines the conversion from FIELD_VAL to the output. PV and OUT always have identical scaling. OUT_SCALE provides scaling for PV. The PV is always the value that the block will place in OUT if the mode is Auto.
Table 3.18 Unit Index by XD_SCALE PH202
Channel FF parameters Service code XD_SCALE.UNITS
1 - SC01 (set to 0) pH
2 2029 SC11 (set to 0) ºC
2 2029 SC11 (set to 1) ºF
3 - SC01 (set to 1) mV
3 - SC01 (set to 0), SC02 (set to 1) mV
3 - SC01 (set to 0), SC02 (set to 2) rH
Table 3.19 Unit Index by XD_SCALE SC202
Channel FF parameters Service code XD_SCALE.UNITS
2 2023 SC11 (set to 0) °C (1001)
2 2023 SC11 (set to 1) °F (1002) 1, 3 - SC01 (set to 1) Ω•cm (1295)
1, 3 - SC01 (set to 0) S/cm (1680)
4 - Default % % (1342)
Table 3.20 Unit Index by XD_SCALE ISC202
Channel FF parameters Service code XD_SCALE.UNITS
2 2023 SC11 (set to 0) °C (1001)
2 2023 SC11 (set to 1) °F (1002)
1, 3 - Default S/cm S/cm (1680)
4 - Default % % (1342)
IM 12A00A01-61E
3-54 Foundation Fieldbus
u Operational Precaution
This document supplements information regarding Operational Precaution. Operate the product carefully based on the following note.
Display on the FieldMate (*)
With using Yokogawa’s FieldMate on the Fieldbus communication of the EXA202 Series instrument, even when unit setting is changed on the instrument, units on AI function blocks on the FieldMate are not changed. Process values on the FieldMate are changed to their new process values for the new units. After changing settings on the instrument, upload the new settings to the FieldMate.
Uploading function on the FieldMate (*)
With using Yokogawa’s FieldMate on the Fieldbus communication of the EXA202 Series instrument, uploading the instrument’s setting to the FieldMate starts with showing an indicator window. On this window, progression of uploading is indicated by a bar and a percentage. Occasionally this percentage on the window doesn’t reach 100%, although uploading is finished. In this case, just close the indicator window.
(*) FieldMate is a communication tool for HART and Fieldbus.
IM 12A00A01-61E
Loading...