Specifications
AQ6370
Applicable fiber
Measurement wavelength
1)
range
1)
Span
Wavelength accuracy
Wavelength linearity
1), 2), 3)
1), 2), 3)
Wavelength repeatability
Measurement data point
Wavelength resolution
1), 2)
setting
Resolution accuracy
1), 2), 3)
Level sensitivity setting
High dynamic range mode
Level sensitivity
Level accuracy
Level linearity
Level flatness
Maximum input power
Safe max. input pow er
2), 4), 5), 7)
2), 4), 5), 6)
2), 4)
2), 4), 6)
2), 4)
2), 4)
Close-in dynamic range
Polarization dependency
Sweep time
1), 7), 8)
Dimensions
13
(
)
0.51
426
(
)(
16.77
SM (9.5/125 µm),
GI (50/125 µm, 62.5/125 µm)
600 to 1700 nm
0.5 nm to full range and zero span
±0.02 nm (1520 to 1580 nm)
±0.04 nm (1450 to 1520 nm, 1580 to 1620 nm)
±0.1 nm (Full range)
±0.01 nm (1520 to 1580 nm)
±0.02 nm (1450 to 1520 nm, 1580 to 1620 nm)
1), 2)
±0.005 nm (1 min.)
101 to 50001
0.02, 0.05, 0.1, 0.2, 0.5, 1.0 and 2.0 nm
±5 % (1450 to 1620 nm, resolution setting: 0.1
to 2.0 nm,resolution correction: ON,
measurement data point setting: AUTO)
NORM_HOLD, NORM_AUTO, NORMAL,
MID, HIGH1, HIGH2 and HIGH3
Switch (Sensitivity: MID, HIGH1, HIGH2 and
HIGH3)
CHOP (Sensitivity: HIGH1, HIGH2 and
HIGH3)
-90 dBm (1300 to 1620 nm, resolution: 0.05
nm or wider, sensitivity: HIGH3)
-80 dBm (1000 to 1300 nm, resolution: 0.05
nm or wider, sensitivity: HIGH3)
-60 dBm (600 to 1000 nm, resolution: 0.05 nm
or wider, sensitivity: HIGH3)
±0.4 dB (1310/1550 nm, input level: -20 dBm,
sensitivity: MID, HIGH1, HIGH2 and HIGH3)
±0.05 dB (Input level: -50 to +10 dBm,
sensitivity: HIGH1, HIGH2 and HIGH3)
±0.1 dB (1520 to 1580 nm)
±0.2 dB (1450 to 1520 nm, 1580 to 1620 nm)
+20 dBm (Per channel, full span)
+25 dBm (Total safe power)
1,) 2), 9)
45 dB (±0.2 nm from peak at 1523 nm,
resolution: 0.05 nm)
62 dB (±0.4 nm from peak at 1523 nm,
resolution: 0.05 nm)
40 dB (±0.2 nm from peak at 1523 nm,
resolution: 0.1 nm)
57 dB (±0.4 nm from peak at 1523 nm,
resolution: 0.1 nm)
2), 4), 6)
±0.05 dB (1550/1600 nm)
±0.08 dB (1310 nm)
Per sensitivity mode;
NORM_AUTO: 0.5sec
NORMAL: 1 sec
MID: 2 sec
HIGH1: 5 sec
HIGH2: 20 sec
HIGH3: 75 sec
REAR VIEW
(
)
1.26
32
14.8
(
)
0.58
13
)
0.51
)
22120
8.70
(
)
0.79
(
(
18.22
Unit : mm
(approx. inch)
427
)
Function Automatic
Data Internal memory
storage Internal storage
Interface Remote control
Printer
11)
Display
Power requirement
Environmental conditions
Dimensions and mass
Note:
1) Horizontal scale: wavelength display mode
2) At 23±5 °C, with 10/125 µm single mode fiber, after 2 hours of warm-up, after optical alignment with
built-in reference light source
3) After wavelength calibration with built-in reference light source
4) Vertical scale: absolute power displa y mode , resolution setting: 0.05 nm or wider, resolution correction:
OFF
5) With 10/125 µm single mode fiber (B1.1 type defined on IEC60793-2, PC polished, mode field diameter: 9.5 µm, NA: 0.104 to 0.107)
6) Temperature condition changes to 23±3 °C
7) High dynamic range mode: OFF, pulse light measurement mode: OFF, TLS sync sweep: OFF, resolu-
(
)
1.26
32
tion correction: OFF
8) Span: any 100 nm or less, measurement data point: 1001, average number: 1
9) High dynamic range mode: CHOP or SWITCH, resolution correction: OFF
10) Excluding feet and handles
11) Liquid crystal display may include few defective pixels (within 0.002% with respect to the total number
of pixels including RGB).
There may be few pixels on the liquid crystal display that do not emit all the time or remains ON all the
time. Note that these are not malfunctions.
Standard Accessories
Name Q'ty
Power cable 1
User's manual (1set) 1
measurement
Setting of
measuring conditions
Display
Traces
Marker/Search
Analysis
Other
External
File type
Category
Optical connector
10)
Macro program function (64 programs, 200 steps)
Center wavelength setting, Span setting, Measurement
data point setting, Wav elength resolution setting, Sensitivity
setting, High dynamic range mode setting, Averaging
number setting (1 to 999 times), Automatic measuring
conditions setting, Sweep between line markers, zero span
sweep, A utomatic measurement data point setting, Pulse
light measurement, External trigger measurement, Sweep
trigger, Sweep status output, Analog output, TLS
synchronized sweep, Air/vacuum wavelength
measurement, Pass/Fail judgment with template
Level scale setting (0.1 to 10 dB/div. and linear), Vertical
sub scale setting (0.1 to 10 dB/div. and linear), Ref erence
level and position setting, V ertical division number setting
(8, 10 or 12), Frequency horizontal scale display,
Horizontal scale zoom in/out display, Measurement
condition display, Noise mask, Data table, Label, Split
display, % display, dB/nm (power spectral density)
display, dB/km display, Template display,
7 independent traces, Write/Fix setting, Display/Blank
setting, Max./Min. hold, calculation between traces , Roll
(Sweep) averaging (2 to 100 times), Normalized, Curve
fit/Peak curve fit/Marker curve fit, Trace copy function,
Trace clear function
Delta marker (Max. 1024), V ertical/Horizontal line marker,
Peak search, Next peak search, Bottom search, Next
bottom search, Auto search, Search between horizontal
line markers, Search in the zooming area
Spectral width (threshold, envelope, RMS, Peak RMS,
notch), WDM (OSNR) analysis, EDF A-NF analysis , Filter
peak/bottom analysis, WDM filter peak/bottom analysis,
DFB-LD analysis, FP-LD analysis, LED analysis, SMSR
analysis, Power analysis, PMD analysis, Pass/Fail
judgment with template, Auto analysis, Analysis between
horizontal line markers, Analysis in the zooming area
Self optical alignment function with built-in light source,
Self wavelength calibration function
64 Traces, 64 programs , 3 template lines
Max. 128MByte
USB storage (memory/HDD) Capability, FAT32 format
CSV(text)/Binary, BMP/TIFF
GPIB, RS-232C and Ethernet (TCP/IP)
AQ6317 series compliant commands (IEEE488.1) and
IEEE488.2 full support
GPIB x2 (standard/controller), RS-232C, Ethernet,
USB1.1 x2, PS/2 (keyboard), SVGA output, Analog
output port, Trigger input port, Trigger output port
Free space optical input: Requires A Q9447 (*) connector
adapter
PC contact built-in light source output: Requires A Q9441
(*) Universal adapter
Built-in high-speed thermal printer (Factory option)
10.4-inch color LCD (Resolution: 800 x 600)
100 to 240 VAC, 50/60Hz, approx. 150VA
Operating temperature: +5 to +35 °C
Storage temperature: -10 to +50 °C
Humidity: 80 %RH or less (no condensation)
Approx. 426 (W) x 221 (H) x 459 (D) mm, Approx. 27kg
(without printer option)