Yellow Jacket 69336 Operating Manual

Page 1
YELLOW JACKET
ACCUPROBE™ UV and ACCUPROBE™ II
Refrigerant Leak Detector with Heated Sensor
®
Instruction and Operation Manual
Models 69336, 69337, 69338, 69339 and 69354
Design Certified to meet SAE J2791 for R-134a
Made in U.S.A.
Page 2
Introduction
The YELLOW JACKET® ACCUPROBE ACCUPROBE all HCFC refrigerants such as R-22 and R-124. The ultra-sensitive long life sensor with exclusive Heated Electrochemical Sensor Technology (HES) also detects the more current, difficult-to-detect HFC refrigerants such as R-134a, R-404A, R-407C and R-410A (see page 6 for a more complete chart of detectable refrigerants).
The ACCUPROBE UV is equipped with the unique digital SmartAlarm leak size indicator in a hand-held heated sensor leak detector. This feature takes the guesswork out of whether or not to repair a small leak. Unlike the more traditional LED bar graphs that copy or mimic the audio alarm signal, the SmartAlarm digital leak size indicator measures, registers and displays the leak size independently from both the audio alarm and the sensitivity level.
II Hand-held Leak Detectors detect
LED display – the first digital
UV and
The ACCUPROBE UV also features a 3 LED UV light system that works with 395-415nm wavelength leak tracing dyes. The sleek, ergonomic design of these YELLOW JACKET leak detectors makes them easy to use in close areas and extendable into hard-to-reach areas.
When finding leaks, it is important to note that the AccuProbe responds to changes in concentration of refrigerant. For this reason, the detector will stop alarming even though it is held at or near the source of the leak and will not alarm again until the detec­tor senses a change in concentration. To verify the exact location of the source of the leak, always move the probe away from the area of the leak briefly to allow the sensor to reset at a lower con­centration, and then bring it back again until the exact location of the leak source has been verified.
Features
Advanced digital SmartAlarm™ leak size indica-
tor (AccuProbe UV)
Flashing visual alarm indicator at probe end
(AccuProbe UV)
Audio mute button (AccuProbe UV)
Operates on 4 AA alkaline batteries (AccuProbe
UV can also use AC power)
3 LED UV light system that works with 395-
415nm wavelength leak tracing dyes (AccuProbe
UV)
Microcontroller technology
Ultra-high sensitivity to detect 0.06 oz (1.7g)/yr.
of R-134a and 0.03 oz (0.9g)/yr. of R-22
Automatic calibration and reset to ambient
Detects HFC, HCFC and CFC refrigerants (see
page 6)
Long life stable sensor utilizing Heated Electro-
chemical Sensor (HES) technology
3 selectable sensitivity level settings
Sleek ergonomic design
Low battery indicator
Temp Range 24° to 125°F (-4° to + 52°C)
Humidity 0 to 95% RH noncondensing
SAE J2791Certified
CE Marked
True mechanical pump
Specifications
Sensitivity 0.06 oz (1.7g)/yr. of R-134a 0.03 oz (0.9g)/yr. of R-22
Power 4 AA alkaline batteries
Probe Length 17 inches (430mm)
Calibration Automatic
Sensor Life >300 hours
Length(body) 10.5 inches (270mm)
Weight AccuProbe II - 15 ounces (430g); AccuProbe UV - 17 ounces (480g)
Battery Life 4.5 hours continuous
Response Time Instantaneous
2
Page 3
Parts and Controls
UV LEDs
Alarm speaker
Digital SmartAlarm™
UV LED
Power ON/OFF
Audio mute
Heated sensor
Flashing LED
Flexible probe
AccuProbe UV
Tip
Battery Installation
AccuProbe UV
AccuProbe UV
Low battery indicator
Sensitivity level indicators
Sensitivity adjust/select
Power ON/OFF
A/C adapter jack
AccuProbe II
1. Loosen battery door screw located on the bottom of the detector and remove door.
2. Install 4 AA alkaline batteries observing the proper battery polarity as labeled
AccuProbe II
1. Loosen screw located at rear end of unit and pull down hinged battery door to open as shown.
2. Always insert all four batteries with the proper polarity.
inside the battery compartment and shown above.
3. Reinstall the battery door and tighten the screw.
3
Page 4
Operating Instructions
AccuProbe UV
1. POWER ON: The detector is turned ON and OFF by pressing the POWER button (see figure 1 below).
2. WARM-UP: The detector automatically starts heating the sensor to condition it for use. While in this WARM UP phase – and until ready – the instrument will signal audibly by beeping slowly and visually by flashing the sensitivity LEDs and the zero (0) in the SmartAlarm™ display window. Warm up time is usually about 20 SECONDS or less. For maximum sensitivity, wait an additional one minute after normal warm-up.
3. SEARCH: The detector is ready for leak search­ing when the sensitivity LEDs stop flashing and the beep rate increases. At this time the zero in the display window stops flashing. When a leak is detected, the beeping sound and flashing LED in the probe will increase in frequency, and the SmartAlarm digital LED display will turn on indicating the leak size. If no leak is detected go to HI sensitivity and continue searching.
SmartAlarm™ Display
The SmartAlarm LED Display is a digital leak size indicator that numerically displays the leak size on a scale of 1 to 9 for all HFC and HCFC refrigerants – regardless of the sensitivity setting. This value helps you decide whether or not the leak is large enough to require repair.
For example, when in the HI sensitivity mode, the detector may sound a full audio alarm but the SmartA­larm Display may show a low number – indicating that the leak is very small. In contrast, when in the LO sensitivity mode, a full audio alarm may not sound but the SmartAlarm may show a high number, indicating that the leak is large.
Low battery indicator
UV LED Power ON/OFF
Audio mute
Figure 1
4
Sensitivity level indicators
Sensitivity adjust
Power ON/OFF
SmartAlarm™ DISPLAY
1 TO 3 < 0.1 (2.8g)
4 TO 6 0.1 to 0.5 (2.8-14g)
7 TO 9 > 0.5 (14.1g)
*HFC and HCFC Refrigerants
The maximum value displayed, once the source of the leak is located, indicates the leak size. This value helps you decide whether or not a leak is large enough to require repair. The table above shows the leak rates corresponding to the SmartAlarm numeri­cal display.
LEAK SIZE DISPLAY (OZ/YR.)*
UV LIGHT OPERATION
CAUTION:
EMITS ULTRAVIOLET RADIATION
Before leak checking with the UV light:
(a) Make sure the A/C system is properly charged
(b) Run the A/C system long enough to thoroughly
1. Turn on UV light by pressing the UV light
2. Holding the leak detector approximately 10” to
3. When the UV light shines on the fluorescent
4. The UV LEDs will automatically turn off after
This UV LED during operation
radiates UV light
Avoid direct eye and skin exposure
to UV light
If viewing the UV light is necessary,
please use UV filtered glasses to avoid damage by the UV light
with sufficient dye (see manufacturers specifi­cations for proper dye charge.)
mix and circulate the dye with the refrigerant and lubricating oil.
ON/OFF button (see diagram on left).
14” away, shine the UV light beam slowly over the components, hoses, and metal lines that make up the A/C system.
dye that has escaped from the system, the dye will glow a bright yellow green.
five minutes.
3 UV LED Lights
Page 5
Using the SmartAlarm Display
1. The SmartAlarm will not display a num­ber (1 through 9) until a leak is sensed. Once a leak is sensed, the numbers on the display will gradually increase.
2. Use the SmartAlarm display to "zero in" on the leak source by watching the num­bers climb higher as the leak source is approached.
3. Once the leak source has been located, always wait for the maximum number to be displayed to determine the size of the leak.
4. Lower numbers (approximately 1 to 3) indicate that the leak size is less than
0.1 oz/yr. and may not require repairing at this time – depending on the amount of gas in system. NOTE: Multiple small leaks in a system are cumulative and may require that all system leaks should be repaired.
AccuProbe II
1. TURN ON: Press the ON/OFF button once to turn on and again to turn off.
2. WARM UP: The detector automatically starts heating the sensor. During the heating cycle, the detector will sound a slow “beep.” Warm up is generally less than 20 seconds. For maxi­mum sensitivity, wait an additional one minute after normal warm-up.
3. READY: The detector is ready to begin searching for leaks when the green sensitivity LED turns on. The audio “beep” increases in frequency.
Sensitivity level indicators
Power ON/OFF
Sensitivity adjust
Adjusting the sensitivity levels
To choose another sensitivity level, press the Sensitivity button. The LED below each level will change indicating the new setting.
The leak detector will default to the NORM sensitivity level automatically once the unit comes out of the warm-up cycle and the green LED turns on.
Using the Leak Standard
Use the leak standard to determine that the leak detector performs to specifications.
1. Lift off the plastic sealing cap on the top of the LEAK STANDARD.
2. POWER ON the unit. After WARM UP and when READY expose the sensor directly to the small hole in the top of the bottle cap. The beep rate should increase to an alarm. For the AccuProbe, the SmartAlarm should display a number greater than 2. If the SmartAlarm fails to display 2 or above, leave the detector on for approximately 15 to 30 seconds longer and retest. This indicates that the sensor and electronic circuit are functioning properly.
3. Replace plastic cap seal after leak test. Note: Replace the leak standard when the green color is no longer visible.
Removable sealing cap
Non-removable cap with leak opening
Non-ozone depleting chemical has green color
Leak Standard Bottle
AccuProbe II Keypad
5
Page 6
Low battery indication
Replace the batteries when the red low battery LED comes on (see page 3). Follow instruc­tions under section titled "BATTERY IN­STALLATION" on page 3.
Audio alarm mute (AccuProbe UV)
To silence the audio alarm, press the MUTE button. Press the MUTE button again to restore the audio alarm.
Sensor failure mode
If the sensor is not working correctly, the AccuProbe Leak Detector will not come out of the warm-up mode. (Some competitive units without this function will not alert you that the sensor is malfunctioning or has failed.)
If the AccuProbe detector does not come out of warm-up, first be sure the sensor is plugged in all the way. If that does not correct the situa­tion, replace the sensor.
Push straight on (do not twist) to install sensor
Unscrew tip to
replace filter
Filter
Maintenance
Batteries: Replace the batteries when the
red low battery LED turns on. See “BATTERY INSTALLATION” on page 3.
Sensor filter replacement: Unscrew the
sensor tip as shown to replace the filter. For optimum performance, replace filter whenever it becomes visibly dirty with grease or oil or every 2-3 months (depending on use).
Note: Never clean dirty filters with a solvent or soap and water. Always replace with a new filter supplied with the leak detector or they can be re-ordered from your supplier or distributor.
Sensor replacement: Remove sensor by
pulling out of socket. Install the new sensor by aligning the notch in the sensor cover with the raised keyway on the sensor socket holder (see Figure 2).
Note: Do not force sensor into socket. Misalignment can damage the sensor pins.
Keyway alignment
6
Pull straight out (do not twist) to remove sensor
Figure 2
Page 7
Replacement Parts
AccuProbe UV
and
AccuProbe II
AccuProbe UV
AccuProbe II
Parts kit (sensor, filters and leak standard bottle) 69383
Sensor and filter 69384
Sensor filters (package of 5) 69385
Leak standard bottle 69386
A/C Adapter Input: 115V 60Hz UL listed 69380
A/C AdapterEU/UK Plug Input: 230V 50Hz CE & TÜV Approved
A/C AdapterAU/NZ Plug Input: 230V 50Hz CE & TÜV Approved
Battery cover and screw 69388
Carrying case - blow molded with inserts 69387
Carrying pouch 69361
Troubleshooting Guide
PROBLEM CHECK REPAIR OR REPLACE
No power
Stays in "warm up" mode
No detection
Slow recovery after detection
No beeping
Check for weak or reversed
batteries
Sensor not plugged into socket
correctly
Sensor open/defective
Check sensor with leak standard
bottle
Check if the filter is dirty or sensor
opening is plugged
Check if filter is dirty or sensor
opening is plugged
Nothing Press mute button (if equipped)
Replace batteries
Make sure sensor is pushed all the
way down into socket
Replace sensor
Replace sensor
Replace filter or clean out opening
Replace filter or clean out opening
to turn speaker back on
69381
69378
Partial list of detectable SNAP* listed refrigerants
R-12 ALTERNATIVES R-22 ALTERNATIVES
R-134a, R-401A (MP-39), R-401B (MP-66), R-401C (MP-52), R
-406A (GHG)
R-414A (GHG-X4), R-414B (Hot Shot), R-416A (Frig C, FR-12)
R-409A (FX-56), Freeze 12, Free Zone, GHG-X5, GHG-HP, IKON 12
R-502, R-500 ALTERNATIVES
R-402A&B, R-404A, R-407A, R-408A, R-411A&B, R-507 R-290, R-600A, R-170/R-290, R-600A/R-290
* SNAP (Significant New Alternatives Program) an EPA program for ozone depleting refrigerants for mobile and stationary A/C systems
R-407C, R-410A, R-410B, R-507
R-113, R-13B & R-503 ALTERNATIVES
R-403B, R-508A, R-508B
HC REFRIGERANTS (not SNAP approved)
7
Page 8
24 MONTH LIMITED WARRANTY
Ritchie Engineering guarantees YELLOW JACKET AccuProbe Leak Detectors to be free of defective material and workmanship that would affect the life of the product under normal use for the purpose for which it was designed. This warranty does not cover items that have been altered, abused, misused, improperly maintained or returned solely in need of field service maintenance. This warranty excludes the sensor, which is warranted for one year.
If found defective, we will upon compliance with the following instructions, credit, replace or repair at our option, the defective leak detector provided it is returned within 24 months of the
Cross Sensitivity to Automotive Chemicals
Some automotive solvents and chemicals have similar hydrocarbon properties as R-134a and may elicit a positive response (<30 seconds) from the AccuProbe. Before leak checking, clean up any chemicals in the list below that elicit a positive response.
Brand/Chemical Name Response
Dextron Transmission Fluid heated to 160° F N
Quaker State Motor Oil heated to 160° F N
Rain-X Windshield Wash Fluid N
Ford Silicone Lubricant N
Ford Rust Inhibitor (when wet) Y
Ford Gasket Adhesive (when wet) Y
Loctite Natural Blue Degreaser (undiluted) Y
Ford Brake Parts Cleaner (when wet) Y Yes
Ford Silicon Rubber (when uncured) Y Yes
Motorcraft Antifreeze heated to 160° F Y Yes
Gunk Liquid Wrench (when wet) Y Yes
Ford Spot Remover (when wet) Y Yes
Ford Pumice Lotion (with mineral solvent) Y Yes
Ford Motorcraft Brake Fluid Y Yes
Ford Carburetor Cleaner (when wet) Y Yes
date of sale. ACCUPROBE leak detectors have a date of manufacture serial number located on the label on the bottom of the unit.
Correction in the manner provided above shall constitute a fulfillment of all liabilities with respect to the quality, material and workmanship of the product.
THE FOREGOING WARRANTY IS EXCLUSIVE AND IN LIEU OF ALL OTHER WARRANTIES OF QUALITY, WHETHER WRITTEN, ORAL OR IMPLIED.
Clears <30 seconds
N/A
N/A
N/A
N/A
Yes
Yes
Yes
YELLOW JACKET Products Division
For tips on searching
for leaks, visit
www.yellowjacket.com
Copyright © 2006-2012 Printed in U.S.A. P/N 400399_Rev.C
Telephone: 800-769-8370 or 952-943-1333
Ritchie Engineering Co., Inc.
10950 Hampshire Avenue South
Bloomington, MN 55438-2623
Fax: 800-322-8684 or 952-943-1605
E-mail: custserv@yellowjacket.com
Web: www.yellowjacket.com
Loading...