The YELLOW JACKET® ACCUPROBE
ACCUPROBE
all HCFC refrigerants such as R-22 and R-124. The
ultra-sensitive long life sensor with exclusive
Heated Electrochemical Sensor Technology (HES)
also detects the more current, difficult-to-detect
HFC refrigerants such as R-134a, R-404A, R-407C
and R-410A (see page 6 for a more complete chart
of detectable refrigerants).
The ACCUPROBE UV is equipped with the unique
digital SmartAlarm
leak size indicator in a hand-held heated sensor leak
detector. This feature takes the guesswork out of
whether or not to repair a small leak. Unlike the
more traditional LED bar graphs that copy or mimic
the audio alarm signal, the SmartAlarm digital leak
size indicator measures, registers and displays the
leak size independently from both the audio alarm
and the sensitivity level.
™
II Hand-held Leak Detectors detect
™
LED display – the first digital
™
UV and
The ACCUPROBE UV also features a 3 LED UV
light system that works with 395-415nm wavelength
leak tracing dyes. The sleek, ergonomic design of
these YELLOW JACKET leak detectors makes
them easy to use in close areas and extendable into
hard-to-reach areas.
When finding leaks, it is important to note that the
AccuProbe responds to changes in concentration of
refrigerant. For this reason, the detector will stop
alarming even though it is held at or near the source
of the leak and will not alarm again until the detector senses a change in concentration. To verify the
exact location of the source of the leak, always
move the probe away from the area of the leak
briefly to allow the sensor to reset at a lower concentration, and then bring it back again until the
exact location of the leak source has been verified.
Features
Advanced digital SmartAlarm™ leak size indica-
tor (AccuProbe UV)
Flashing visual alarm indicator at probe end
(AccuProbe UV)
Audio mute button (AccuProbe UV)
Operates on 4 AA alkaline batteries (AccuProbe
UV can also use AC power)
3 LED UV light system that works with 395-
415nm wavelength leak tracing dyes (AccuProbe
UV)
Microcontroller technology
Ultra-high sensitivity to detect 0.06 oz (1.7g)/yr.
of R-134a and 0.03 oz (0.9g)/yr. of R-22
Automatic calibration and reset to ambient
Detects HFC, HCFC and CFC refrigerants (see
page 6)
Long life stable sensor utilizing Heated Electro-
chemical Sensor (HES) technology
3 selectable sensitivity level settings
Sleek ergonomic design
Low battery indicator
Temp Range 24° to 125°F (-4° to + 52°C)
Humidity 0 to 95% RH noncondensing
SAE J2791Certified
CE Marked
True mechanical pump
Specifications
Sensitivity 0.06 oz (1.7g)/yr. of R-134a 0.03 oz (0.9g)/yr. of R-22
1. Loosen battery door screw located on the
bottom of the detector and remove door.
2. Install 4 AA alkaline batteries observing
the proper battery polarity as labeled
AccuProbe II
1. Loosen screw located at rear end of unit
and pull down hinged battery door to
open as shown.
2. Always insert all four batteries with the
proper polarity.
inside the battery compartment and
shown above.
3. Reinstall the battery door and tighten the
screw.
3
Page 4
Operating Instructions
AccuProbe UV
1. POWER ON: The detector is turned ON and
OFF by pressing the POWER button (see figure
1 below).
2. WARM-UP: The detector automatically starts
heating the sensor to condition it for use. While
in this WARM UP phase – and until ready – the
instrument will signal audibly by beeping slowly
and visually by flashing the sensitivity LEDs
and the zero (0) in the SmartAlarm™ display
window. Warm up time is usually about 20
SECONDS or less. For maximum sensitivity,
wait an additional one minute after normal
warm-up.
3. SEARCH: The detector is ready for leak searching when the sensitivity LEDs stop flashing and
the beep rate increases. At this time the zero in
the display window stops flashing. When a leak
is detected, the beeping sound and flashing LED
in the probe will increase in frequency, and the
SmartAlarm digital LED display will turn on
indicating the leak size. If no leak is detected go
to HI sensitivity and continue searching.
SmartAlarm™ Display
The SmartAlarm LED Display is a digital leak size
indicator that numerically displays the leak size on a
scale of 1 to 9 for all HFC and HCFC refrigerants –
regardless of the sensitivity setting. This value helps
you decide whether or not the leak is large enough to
require repair.
For example, when in the HI sensitivity mode, the
detector may sound a full audio alarm but the SmartAlarm Display may show a low number – indicating
that the leak is very small. In contrast, when in the LO
sensitivity mode, a full audio alarm may not sound but
the SmartAlarm may show a high number, indicating
that the leak is large.
Low
battery
indicator
UV LED
Power
ON/OFF
Audio
mute
Figure 1
4
Sensitivity
level
indicators
Sensitivity
adjust
Power
ON/OFF
SmartAlarm™
DISPLAY
1 TO 3 < 0.1 (2.8g)
4 TO 6 0.1 to 0.5 (2.8-14g)
7 TO 9 > 0.5 (14.1g)
*HFC and HCFC Refrigerants
The maximum value displayed, once the source of
the leak is located, indicates the leak size. This value
helps you decide whether or not a leak is large
enough to require repair. The table above shows the
leak rates corresponding to the SmartAlarm numerical display.
LEAK SIZE
DISPLAY (OZ/YR.)*
UV LIGHT OPERATION
CAUTION:
EMITS ULTRAVIOLET RADIATION
Before leak checking with the UV light:
(a) Make sure the A/C system is properly charged
(b) Run the A/C system long enough to thoroughly
1. Turn on UV light by pressing the UV light
2. Holding the leak detector approximately 10” to
3. When the UV light shines on the fluorescent
4. The UV LEDs will automatically turn off after
This UV LED during operation
radiates UV light
Avoid direct eye and skin exposure
to UV light
If viewing the UV light is necessary,
please use UV filtered glasses to
avoid damage by the UV light
with sufficient dye (see manufacturer’s specifications for proper dye charge.)
mix and circulate the dye with the refrigerant
and lubricating oil.
ON/OFF button (see diagram on left).
14” away, shine the UV light beam slowly
over the components, hoses, and metal lines
that make up the A/C system.
dye that has escaped from the system, the dye
will glow a bright yellow green.
five minutes.
3 UV LED
Lights
Page 5
Using the SmartAlarm Display
1. The SmartAlarm will not display a number (1 through 9) until a leak is sensed.
Once a leak is sensed, the numbers on the
display will gradually increase.
2. Use the SmartAlarm display to "zero in"
on the leak source by watching the numbers climb higher as the leak source is
approached.
3. Once the leak source has been located,
always wait for the maximum number
to be displayed to determine the size of
the leak.
4. Lower numbers (approximately 1 to 3)
indicate that the leak size is less than
0.1 oz/yr. and may not require repairing
at this time – depending on the amount
of gas in system. NOTE: Multiple small
leaks in a system are cumulative and
may require that all system leaks should
be repaired.
AccuProbe II
1. TURN ON: Press the ON/OFF button once to
turn on and again to turn off.
2. WARM UP: The detector automatically starts
heating the sensor. During the heating cycle,
the detector will sound a slow “beep.” Warm
up is generally less than 20 seconds. For maximum sensitivity, wait an additional one minute
after normal warm-up.
3. READY: The detector is ready to begin
searching for leaks when the green sensitivity
LED turns on. The audio “beep” increases in
frequency.
Sensitivity
level
indicators
Power
ON/OFF
Sensitivity
adjust
Adjusting the sensitivity levels
To choose another sensitivity level, press the
Sensitivity button. The LED below each level
will change indicating the new setting.
The leak detector will default to the NORM
sensitivity level automatically once the unit
comes out of the warm-up cycle and the green
LED turns on.
Using the Leak Standard
Use the leak standard to determine that the
leak detector performs to specifications.
1. Lift off the plastic sealing cap on the top
of the LEAK STANDARD.
2. POWER ON the unit. After WARM UP
and when READY expose the sensor
directly to the small hole in the top of the
bottle cap. The beep rate should increase
to an alarm. For the AccuProbe, the
SmartAlarm should display a number
greater than 2. If the SmartAlarm fails to
display 2 or above, leave the detector on
for approximately 15 to 30 seconds
longer and retest. This indicates that
the sensor and electronic circuit are
functioning properly.
3. Replace plastic cap seal after leak test.
Note: Replace the leak standard when the
green color is no longer visible.
Removable sealing cap
Non-removable cap with
leak opening
Non-ozone depleting
chemical has green color
Leak Standard Bottle
AccuProbe II Keypad
5
Page 6
Low battery indication
Replace the batteries when the red low battery
LED comes on (see page 3). Follow instructions under section titled "BATTERY INSTALLATION" on page 3.
Audio alarm mute (AccuProbe UV)
To silence the audio alarm,
press the MUTE button. Press
the MUTE button again to
restore the audio alarm.
Sensor failure mode
If the sensor is not working correctly, the
AccuProbe Leak Detector will not come out of
the warm-up mode. (Some competitive units
without this function will not alert you that the
sensor is malfunctioning or has failed.)
If the AccuProbe detector does not come out
of warm-up, first be sure the sensor is plugged
in all the way. If that does not correct the situation, replace the sensor.
Push straight on (do not twist) to install sensor
Unscrew tip to
replace filter
Filter
Maintenance
Batteries: Replace the batteries when the
red low battery LED turns on. See “BATTERY
INSTALLATION” on page 3.
Sensor filter replacement: Unscrew the
sensor tip as shown to replace the filter. For
optimum performance, replace filter whenever
it becomes visibly dirty with grease or oil or
every 2-3 months (depending on use).
Note: Never clean dirty filters with a solvent
or soap and water. Always replace with a
new filter supplied with the leak detector or
they can be re-ordered from your supplier
or distributor.
Sensor replacement: Remove sensor by
pulling out of socket. Install the new sensor by
aligning the notch in the sensor cover with the
raised keyway on the sensor socket holder (see
Figure 2).
Note: Do not force sensor into socket.
Misalignment can damage the sensor pins.
Keyway alignment
6
Pull straight out (do not twist) to remove sensor
Figure 2
Page 7
Replacement Parts
AccuProbe UV
and
AccuProbe II
AccuProbe UV
AccuProbe II
Parts kit (sensor, filters and leak standard bottle) 69383
Sensor and filter 69384
Sensor filters (package of 5) 69385
Leak standard bottle 69386
A/C Adapter Input: 115V 60Hz UL listed 69380
A/C Adapter—EU/UK Plug Input: 230V 50Hz
CE & TÜV Approved
A/C Adapter—AU/NZ Plug Input: 230V 50Hz
CE & TÜV Approved
Battery cover and screw 69388
Carrying case - blow molded with inserts 69387
Carrying pouch 69361
Troubleshooting Guide
PROBLEMCHECKREPAIR OR REPLACE
No power
Stays in "warm up"
mode
No detection
Slow recovery after
detection
No beeping
Check for weak or reversed
batteries
Sensor not plugged into socket
correctly
Sensor open/defective
Check sensor with leak standard
bottle
Check if the filter is dirty or sensor
opening is plugged
Check if filter is dirty or sensor
opening is plugged
Nothing Press mute button (if equipped)
Replace batteries
Make sure sensor is pushed all the
way down into socket
Replace sensor
Replace sensor
Replace filter or clean out opening
Replace filter or clean out opening
to turn speaker back on
69381
69378
Partial list of detectable SNAP* listed refrigerants
R-12 ALTERNATIVES R-22 ALTERNATIVES
R-134a, R-401A (MP-39), R-401B (MP-66), R-401C (MP-52), R
* SNAP (Significant New Alternatives Program) an EPA program for ozone depleting refrigerants for mobile and stationary A/C systems
R-407C, R-410A, R-410B, R-507
R-113, R-13B & R-503 ALTERNATIVES
R-403B, R-508A, R-508B
HC REFRIGERANTS (not SNAP approved)
7
Page 8
24 MONTH LIMITED WARRANTY
Ritchie Engineering guarantees YELLOW
JACKET AccuProbe Leak Detectors to be free of
defective material and workmanship that would
affect the life of the product under
normal use for the purpose for which it was
designed. This warranty does not cover items that
have been altered, abused, misused, improperly
maintained or returned solely in need of field
service maintenance. This warranty excludes the
sensor, which is warranted for one year.
If found defective, we will upon compliance with
the following instructions, credit, replace or
repair at our option, the defective leak detector
provided it is returned within 24 months of the
Cross Sensitivity to Automotive Chemicals
Some automotive solvents and chemicals have similar hydrocarbon properties as R-134a and
may elicit a positive response (<30 seconds) from the AccuProbe. Before leak checking, clean
up any chemicals in the list below that elicit a positive response.
Brand/Chemical Name Response
Dextron Transmission Fluid heated to 160° F N
Quaker State Motor Oil heated to 160° F N
Rain-X Windshield Wash Fluid N
Ford Silicone Lubricant N
Ford Rust Inhibitor (when wet) Y
Ford Gasket Adhesive (when wet) Y
Loctite Natural Blue Degreaser (undiluted) Y
Ford Brake Parts Cleaner (when wet) Y Yes
Ford Silicon Rubber (when uncured) Y Yes
Motorcraft Antifreeze heated to 160° F Y Yes
Gunk Liquid Wrench (when wet) Y Yes
Ford Spot Remover (when wet) Y Yes
Ford Pumice Lotion (with mineral solvent) Y Yes
Ford Motorcraft Brake Fluid Y Yes
Ford Carburetor Cleaner (when wet) Y Yes
date of sale. ACCUPROBE leak detectors have a
date of manufacture serial number located on the
label on the bottom of the unit.
Correction in the manner provided above shall
constitute a fulfillment of all liabilities with
respect to the quality, material and workmanship
of the product.
THE FOREGOING WARRANTY IS
EXCLUSIVE AND IN LIEU OF ALL OTHER
WARRANTIES OF QUALITY, WHETHER
WRITTEN, ORAL OR IMPLIED.