*1 In case of unloaded operation of bath only. The maximum temperature varies on circumstances
and operational conditions.
*2 It shows the performance under rated power supply at the room temperature of 23℃±5℃ and with
humidity of 65%±20%.
*3 It does not include projection parts.
(W×D×H)
*3
9.4×11.8×5.9 (inches)
AC230V±10% 50Hz 2.3AAC230V±10% 50Hz 4.5A
●
Bar Alcohol Thermometer
(‑10℃〜110℃ with immersion line) 1
●
Fuse
●
Instruction manual
●
Guarantee
Depth 12 (cm)
24×30×15 (cm)
Φ25×
12.2×14.2×9.1 (inches)
Depth 15 (cm)
31×36×23 (cm)
1
1
1
1
Page 4
2. Safety Information
Safety Symbols
Safety Information
This instruction manual and our products apply various indications for safety. Ignoring these
indications can cause such situations as listed below. Read and understand the following warning and
caution signs in this manual prior to use.
WARNING
CAUTION
(Note 1) Serious injur y : Bodily harm by electric shock, bone fracture or poisoning which may require
hospitalization.
(Note 2) Injury : Bodily harm by electric shock, bone fracture or poisoning which may not require
hospitalization.
(Note 3) Damage : Any damage on equipment, facility, structure, etc.
Indicates the possibility of serious or fatal injury (Note 1).
Indicates the possibility of injury (Note 2) or damage (Note 3) to the
equipment.
Meaning of Graphic Indications
Shows warning or caution.
Specific contents are described aside each sign.
Shows users important information not to do.
Specific contents are described aside each sign.
Shows users important information sure to do.
Specific contents are described aside each sign.
2
Page 5
Safety Information
Safety Precautions
Do not use this unit for any purpose other than its intended use, described in this manual.
Do not use this unit in flammable or explosive gas environments.
This unit is not explosion proof. Never use this unit in flammable or explosive gas environments.
Never fail to ground the unit.
This unit uses a 3-c onductor power cord (including ground wire). Be sur e to ground the unit for
safety.
Do not use this unit if a malfunction occurs.
If smoke or any strange odor should dis burse from this unit, turn the power off im mediately and
pull out the main power cord. Then contact Yamato Scientific. Neglecting this procedure can
result in fire or electric shock. Never try repairing the unit yourself.
Do not bundle the power cord.
Overheat or fire can occur if the power cord is bundled or if an object is on the cord.
Do not damage the power cord.
Forcible bending, pulling wrenching or extending the power cord can cause fire or electric shock.
Do not use any explosive or flammable heat medium.
Never use any explosive or flam mable substances or s uc h c ompounds as a heat m edium which
could cause an explosion or fire.
Avoid water contact.
To avoid electric leak or shock, avoid direct contact with water.
Do not heat with low water.
If you operate this unit with low water, fire can be caused by overheating.
Do not disassemble or remodel the unit
To avoid fire or electric shock, never try disassemble this unit.
3
Page 6
Safety Information
Safety Precautions
If it begins to thunder…
In the event of electrical s torm turn off the main power. Neglecting this procedure can result in
fire, electric shock or other troubles due to thunderbolts.
In case of power failure…
When the power is rest ored after a power f ailure, the unit r esum es operation at the tem perature
previously set as if you turned off the unit.
Indication of temperature and operational range.
This unit is not equipped with a stirring f unction. Therefore, the maximum temperature varies
according to the environments or operational conditions. This unit does not necessarily reach the
maximum operational temperature under such environments as low room temperature.
4
Page 7
Safety Information
Hazardous Materials
Hazardous material are listed below. Never use these materials as samples or
heat media.
Nitroglycol, Nitroglycerin, Nitrocellulose, and other explosive nitric
esters.
Trinitrobenzene, Trinitrotoluene, Picric ac id, and other explosive nitro
compounds.
Peracetic acid, Methyl ethyl ketone peroxide, Benzoyl peroxide, and
other organic peroxides.
Sodium azide, and other metallic azides
Metallic lithium, Metallic potassium, Metallic sodium, Yellow
phosphorus, Phosphorus sulfide, Red phosphorus, Celluloid, Calcium
carbide, Lime phosphate, Magnesium powder, Aluminum powder,
and other combustible metal powders and sodium dithionite
(hydrosulfite).
Potassium chlorate, Sodium chlorate, Ammonium chlorate, and
other chlorates.
Potassium perchlorate, Sodium perchlorate, Ammonia perchlorate,
and other perchlorates.
Potassium peroxide, Sodium peroxide, Barium peroxide, and other
inorganic peroxides.
Potassium nitrate, Sodium nitrate, Ammonia nitrate, and other
nitrates.
Explosive
Explosive
Substance
Combustible
Substance
Oxidant
Flammable
Ignitable
Substance
Combustible
Gas
Sodium chlorite and other chlorites.
Calcium hypochlorite and other hypochlorites.
Ethyl ether, Gasoline, Acetaldehyde, Propylene Oxide, Carbon
disulfide, and other flammable substances with a flash point below
minus 30ºC.
Normal hexane, Ethylene oxide, Acetone, Benzene, Methyl ethyl
ketone, and other flamm able substances with a flash point between
minus 30ºC and 0
Methanol, Ethanol, Xylene, Pentyl acetate (amyl acetate), and other
flammable substance with a flash point between 0ºC and 30ºC.
Kerosene, Light oil, Turpentine oil, Isoam yl alcohol, Acetic acid, and
other flammable substances with a flash point between 30ºC and
65ºC
Hydrogen, Acetylene, Ethylene, Methane, Ethane, Propane, Butane
and other flammable gas at 15℃ degree and under 1 atmosphere.
o
C.
5
Page 8
BM110
3. Identification of Parts
BM110
Thermometer
Power Switch
Power Lamp
Temperature Dial
Fuse
Power Cable
Thermometer Holder
Sensor
Heater
6
Page 9
BM210
Identification of Parts
BM210
Thermometer
Power Switch
Drain Plug
Power Lamp
Fuse
Heater
Temperature Dial
Power Cable
Thermometer Holder
Sensor
7
Page 10
Combination with RE210
Identification of Parts
Sample : Combined with RE210
Drawing 2Drawing 1
According to the position or space to set the bath and the body, the arm jack can be set in the front
(drawing 1)or on the right side (drawing 2) as shown above.
8
Page 11
Combination with RE510
Identification of Parts
Sample : Combined with RE510
9
Page 12
How to operate
4. How to Operate
Power Lamp
Temperature Dial
1. Pour adequate water (3000ml for BM110, 5000ml for BM210) into the tank, and place a thermometer
in the holder. Use distilled water. Deionized water should not be used.
2. Connect the unit securely to a power source with AC230V. Always use a grounded outlet. Set the
temperature dial to the desired temperature and turn on the power switch. The power lamp is lit when
the power heater is on.
3. The temperature scale shows an approxim ate figure. When the temperatur e rises close to the set
point, read the thermometer to adjust the temperature. Since there are some differences from the set
point the first time, check if the current temperature accords with the degree to operate.
This unit does not have a stirring function and the tank tem perature varies . We recomm end you to use
any stirrer (our Lab. Stirrer etc.,) together with this unit.
Power Failure
When power supply is restor ed after a power failure, the unit re-starts operation from previously set
temperature.
10
Page 13
Daily maintenance/Weekly maintenance
Do not disjoint the unit
Never modify the unit. There are some highvoltage areas inside the unit, which could cause
electric shock. Request Yamato Scientific for
internal check, adjustment or repair.
Maintenance
5. Maintenance
Daily Maintenance
Turn power switch off and disconnect main power cord before attempting any maintenance.
Once the water bath has cooled to room temperature you may empty the water.
Wipe off any excess dirt on the interior and exterior with a damp cloth. Do not use benzene,
thinner or cleanser.
Weekly Maintenance
The pipe heater should be cleaned at least twice a month. Accumulation of scale build up
results in poor heat transfer and high sheath temperatures.
:
:
When the unit is not in use for a long period of time.
Be sure to disconnect main power and drain the water tank if the unit will not be used for a long
period of time.
Troubleshooting Guide
SituationMake Sure
The power lamp on the panel is not
lit when the power switch is on.
If you have any question, contact Yamato Scientific.
●
Check main power cord is connected firmly.
●
Check main input voltage.
11
Page 14
6. After Sale Service and Warranty
Request for Repair
When you request repair
If any troubles should occur, stop the operation immediately, turn the power off, pull the power cord out
and contact Yamato Scientific’s Technical Service Department.
Necessary information
●
Model Number
●
Serial Number
●
Date of Purchase
●
Distributor Name
●
Information on difficulties
Warranty
●
Keep your warranty card for future references. Check the name of the distributor, date of purchase
and any other contents of warranty.
●
The terms of warranty is two years limited commencing the date of purchase. Repair is made without
charge according to the contents of warranty.
●
Decontamination Statement:
We can not accept any products or parts returned to us for repair or credit that is contaminated with or
has been exposed to potentially infectious agents or radioactive materials.
If you need repair, please call Yamato Scientific for a return authorization number. No product will be
accepted without this number.