4000 VAC Isolation
•
Low Leakage Current
•
Meets Class B, EN55011 Conducted
•
Medical Safety Agency Approvals
•
Both Line and Neutral Fused
•
Fixed Switching Frequency
65 Watts
PBM65 Series
AC/DC Medical Open Frame
Specification
•
Input
Input Voltage • 90-264 VAC
Input Frequency • 47-63 Hz
Input Current • 1.6 A max at 115 VAC
1.0 A max at 230 VAC
Inrush Current • 17 A at 115 VAC
40 A at 230 VAC
Safety Ground • 90 µA max at 115 VAC, 60 Hz
Leakage Current 150 µA max at 230 VAC, 50 Hz
Output
Output Voltage • 5.1 to 28 VDC
Output Voltage • ±5%
Adjustment
Output Power • 65 Watts
Minimum Load • No minimum load required
Line Regulation • ±0.5% max at full load
Load Regulation • ±2% typical
Ripple and Noise • ±1% pk-pk max (See Note 2)
Transient Response • 4% maximum deviation, 500 µs
recovery time to within 1% for a
25% load change
Temperature • ±0.04%/°C
Coefficient
Hold Up Time • 10ms minimum at 110 VAC
Overvoltage • 112-132% with auto recovery
Protection
Overcurrent • 125-200% with auto recovery
Protection
Short Circuit • Protected to short circuit conditions
Protection with auto-recovery
Remote Sense • Compensates for up to 0.5 V drop
General
Efficiency • 75% min for Vo >=12V,
66% min for others
MTBF • 400,000 hrs min MIL-HDBK-217F
Switching Frequency • 3w2 kHz ±5 kHz
Isolation Voltage • 4000 VAC Input to Output
1500 VAC Input to Ground
500 VAC Output to Ground
Size • 5.0” x 3.0” x 1.2”
Weight • 330 g approx
Environmental
Operating • 0 °C to +70 °C
Temperature Derate linearly from 100% load at
+50 °C to 50% load at +70 °C
Cooling • Convection cooled
Relative Humidity • 5-95% non-condensing
Storage • -40 °C to +85 °C
Temperature
EMC & Safety
Safety Approvals • UL2601-1, CSA C22.2 No 601.1,
EN60601-1 CE Mark (LVD)
EMI/EMC • EN55011 (CISPR11) Class B &
FCC Part 18 Level B radiated &
conducted
Intelligent Design Quality Product
XPiQ inc.
PH: 508 429.9883
FAX: 800 226.2100
Email: sales@xpiq.com
Holliston, MA 01746 USA
www.XPiQ.com