Wolf Sanoclav TKL-MCS User manual

DECLARATON OF CONFORMITY „CE“
In accordance to the following directives
97 / 23 / EG for equipment under pressure
used modul: A
Type and model of products;
Autoclave Type : KL – 5 - 3 Autoclave Type : KL – 7 - 3 Autoclave Type : KL – 12 - 2
It is declared that the above mentioned products in conception and type of construction, also in the mode of design traded by us, meet the above directives.
This declaration is going to loose its validity in case of changes not agreed by us.
Used harmonised regulations:
DIN EN 10028 - 7
Used national regulations/technical specifications:
AD – Merkblatt A 2
ADOLF WOLF SANOclav
Bad Überkingen , 07.01.2003
Adolf Wolf .
Directing Manager
Table 8: Survey of the inspection, maintenance and cleaning work
Inspection Maintenance Cleaning
Visual inspection be-
Rubber gas­ket of the vessel lid
Valves
Heater
Autoclave
Sterilization­process
* Please contact us in the event of uncertainty. We will be happy to supply you with
any information required.
for each sterilization process:
Ensure that the gasket is intact before starting!
Visual inspection be­for each sterilization process:
Watch for grime!
Monthly visual in­spection:
Watch for scaling!
The autoclave must be checked every 2-3 years by the manufacturer!
Recalibrate the tem­perature and pres­sure indicator yearly.
Check effective­ness* in conformity with DIN 58946-8 every 6 month.
The rubber gasket is a wearing part.
It must be renewed in the event of dam­age or wear.
For safety reasons
the lid must be returned to the manufacturer when
the valve has become clogged.
Eco-friendly descaling with vinegar at a mod­erate temperature.
After sterilization autoclaves must be
stored in an open condition.
Carefully remove grime with water.
Do not try to clean yourself!
___________
Monthly cleaning of the inside de­pending upon the frequency of use.
Do not use a scouring agent!
Check of Temperature and pressure indicator:
The autoclave is to heat up to 1 bar and 2 bar. Than the follow­ing data have to appear: Pressure: 1bar Temperature: 121°C Pressure: 2bar Temperature: 134°C If there are any deviations the autoclave has to be checked by the manufacturer or, pressure -, respectively temperature­indicator has to be replaced.
_____________________________________________________________________
page 30
4 Optional extras: Media sensor
The media sensor can only be used in conjunction with the MCS control (see fig.2 page 23)
WARNING: Do not open the exhaust steam valve after the sterilization of liquids (media) due to the danger of delayed boiling. Allow the autoclave to cool down natu­rally.
Only open the autoclave, when the chamber temperature has cooled down to at least 70°C the autoclave has no remaining pressure (0 bar)
The media sensor is used to establish the actual temperature of a liquid (medium) that is to be sterilized during the heating-up and cooling-down stages. The sensor is directly immersed in the medium (reference container).
1. Activation of the media sensor (see fig. 2 page 23)
a) To activate the media sensor press the "E" key before
the autoclave is started with the "F" key. The LED indi­cator (6) lights up to confirm that the media sensor has been switched on.
b) Press the "E" key to call the media temperature during
the sterilization process.
c) NOTE: When a media sensor is being used the sterili-
zation time only commences when the medium has reached the setpoint temperature. An additional sterilization time as described under chapter
2.6.2 therefore is not necessary.