Table 8: Survey of the inspection, maintenance and cleaning work
Inspection Maintenance Cleaning
Visual inspection be-
Rubber gasket of the
vessel lid
Valves
Heater
Autoclave
Sterilizationprocess
* Please contact us in the event of uncertainty. We will be happy to supply you with
any information required.
for each sterilization
process:
Ensure that the
gasket is intact
before starting!
Visual inspection befor each sterilization
process:
Watch for grime!
Monthly visual inspection:
Watch for scaling!
The autoclave must
be checked every
2-3 years by the
manufacturer!
Recalibrate the temperature and pressure indicator yearly.
Check effectiveness* in conformity
with DIN 58946-8
every 6 month.
The rubber gasket is
a wearing part.
It must be renewed
in the event of damage or wear.
For safety reasons
the lid must be
returned to the
manufacturer when
the valve has become
clogged.
Eco-friendly descaling
with vinegar at a moderate temperature.
After sterilization
autoclaves must be
stored in an open
condition.
Carefully remove
grime with water.
Do not try to
clean yourself!
___________
Monthly cleaning
of the inside depending upon the
frequency of use.
Do not use a
scouring agent!
Check of Temperature and pressure indicator:
The autoclave is to heat up to 1 bar and 2 bar. Than the following data have to appear:
Pressure: 1bar Temperature: 121°C
Pressure: 2bar Temperature: 134°C
If there are any deviations the autoclave has to be checked by
the manufacturer or, pressure -, respectively temperatureindicator has to be replaced.
_____________________________________________________________________
page 30
4 Optional extras: Media sensor
The media sensor can only be used in conjunction with the
MCS control (see fig.2 page 23)
WARNING: Do not open the exhaust steam valve after
the sterilization of liquids (media) due to the danger of
delayed boiling. Allow the autoclave to cool down naturally.
Only open the autoclave, when
⇒ the chamber temperature has cooled down to at
least 70°C
⇒ the autoclave has no remaining pressure (0 bar)
The media sensor is used to establish the actual temperature of
a liquid (medium) that is to be sterilized during the heating-up
and cooling-down stages. The sensor is directly immersed in
the medium (reference container).
1. Activation of the media sensor (see fig. 2 page 23)
a) To activate the media sensor press the "E" key before
the autoclave is started with the "F" key. The LED indicator (6) lights up to confirm that the media sensor has
been switched on.
b) Press the "E" key to call the media temperature during
the sterilization process.
c) NOTE: When a media sensor is being used the sterili-
zation time only commences when the medium has
reached the setpoint temperature.
An additional sterilization time as described under chapter
2.6.2 therefore is not necessary.
__________________________________________________
page 28