Wolf Sanoclav TKL-MCS User manual

DECLARATON OF CONFORMITY „CE“
In accordance to the following directives
97 / 23 / EG for equipment under pressure
used modul: A
Type and model of products;
Autoclave Type : KL – 5 - 3 Autoclave Type : KL – 7 - 3 Autoclave Type : KL – 12 - 2
It is declared that the above mentioned products in conception and type of construction, also in the mode of design traded by us, meet the above directives.
This declaration is going to loose its validity in case of changes not agreed by us.
Used harmonised regulations:
DIN EN 10028 - 7
Used national regulations/technical specifications:
AD – Merkblatt A 2
ADOLF WOLF SANOclav
Bad Überkingen , 07.01.2003
Adolf Wolf .
Directing Manager
Table 8: Survey of the inspection, maintenance and cleaning work
Inspection Maintenance Cleaning
Visual inspection be-
Rubber gas­ket of the vessel lid
Valves
Heater
Autoclave
Sterilization­process
* Please contact us in the event of uncertainty. We will be happy to supply you with
any information required.
for each sterilization process:
Ensure that the gasket is intact before starting!
Visual inspection be­for each sterilization process:
Watch for grime!
Monthly visual in­spection:
Watch for scaling!
The autoclave must be checked every 2-3 years by the manufacturer!
Recalibrate the tem­perature and pres­sure indicator yearly.
Check effective­ness* in conformity with DIN 58946-8 every 6 month.
The rubber gasket is a wearing part.
It must be renewed in the event of dam­age or wear.
For safety reasons
the lid must be returned to the manufacturer when
the valve has become clogged.
Eco-friendly descaling with vinegar at a mod­erate temperature.
After sterilization autoclaves must be
stored in an open condition.
Carefully remove grime with water.
Do not try to clean yourself!
___________
Monthly cleaning of the inside de­pending upon the frequency of use.
Do not use a scouring agent!
Check of Temperature and pressure indicator:
The autoclave is to heat up to 1 bar and 2 bar. Than the follow­ing data have to appear: Pressure: 1bar Temperature: 121°C Pressure: 2bar Temperature: 134°C If there are any deviations the autoclave has to be checked by the manufacturer or, pressure -, respectively temperature­indicator has to be replaced.
_____________________________________________________________________
page 30
4 Optional extras: Media sensor
The media sensor can only be used in conjunction with the MCS control (see fig.2 page 23)
WARNING: Do not open the exhaust steam valve after the sterilization of liquids (media) due to the danger of delayed boiling. Allow the autoclave to cool down natu­rally.
Only open the autoclave, when the chamber temperature has cooled down to at least 70°C the autoclave has no remaining pressure (0 bar)
The media sensor is used to establish the actual temperature of a liquid (medium) that is to be sterilized during the heating-up and cooling-down stages. The sensor is directly immersed in the medium (reference container).
1. Activation of the media sensor (see fig. 2 page 23)
a) To activate the media sensor press the "E" key before
the autoclave is started with the "F" key. The LED indi­cator (6) lights up to confirm that the media sensor has been switched on.
b) Press the "E" key to call the media temperature during
the sterilization process.
c) NOTE: When a media sensor is being used the sterili-
zation time only commences when the medium has reached the setpoint temperature. An additional sterilization time as described under chapter
2.6.2 therefore is not necessary.
__________________________________________________
page 28
1 Safety Instructions
„SANOclav" stands for state-of-the-art small-batch autoclaves (small sterilizers). All safety-technical aspects, in so far as they are within the manufacturer's sphere of control, have been taken into account.
The series KL and TKL autoclaves are insufficiently equipped for sterilisation of medical products within the meaning of the European Guideline 93/42/CCE and are therefore not suitable.
For causes outside the manufacturer's sphere of control, no re­sponsibility is accepted for consequential damage, injury to the sterilization product, device, appliance or the surrounding area.
The basic prerequisite for operation ensuring freedom of hazard to a great degree is that:
the devices are only used by persons who can offer the guarantee for proper handling on the basis of their training or their knowledge and practical experience.
These Operating Instructions, as well as any other appro­priate additional documents, must be made available to the operating personnel at all times for their work. Operating personnel are obliged to observe these Operating Instruc­tions and other documents.
Unqualified personnel are not allowed to work at the auto­claves or be in their vicinity.
The autoclaves most not be technically changed in any manner without previously consulting the manufacturer.
2.7.1 Trouble shooting for the MCS control
Table 7: Reason for error message and possible cause
Display Description Cause
Err 1
Err 2
Err 3
Err 4
Err 5
- - -
F
F
- - -
Table 8: Action after an error message
Display Acti on
Err 1
Err 2
Err 3
Err 4
Err 5
- - -
F
F
- - -
Operation was aborted during the heating-up stage
Operation was aborted during the sterilization period
Operation was interrupted Defective (broken) chamber sensor Operation was interrupted Defective (broken) media sensor
Operation was interrupted
Arises when the device is switched on or after Err 3 and Err 4
Arises when the device is switched on or after Err 3 and Err 4
Repeat sterilization. Consult manufacturer if message reappears.
Repeat sterilization. Consult manufacturer if message reappears.
Autoclave must be repaired by the manufacturer
Autoclave must be repaired by the manufacturer
Check rubber gasket for proper seat in the lid and then repeat steriliza­tion. Consult manufacturer if message reappears.
Autoclave must be repaired by the manufacturer
Autoclave must be repaired by the manufacturer
for example: Power failure
for example: Power failure
The chamber or media temperature dropped by more than 2°C below the setpoint temperature during a sterili­zation period of more than 2 minutes
Broken chamber sensor
Short-circuited chamber sensor
These safety instructions do not claim to be complete. In the event of queries and problems please consult our specialists who will be happy to help you and supply corresponding information.
__________________________________________________
page 1
All error messages must be confirmed with the "F" key. This resets the autoclave to its initial state. The messages "Err 3", "Err 4" and "F---" cannot be followed by another operation. The
messages "F
---
" or "F
" appear when the "F" key (device
---
ON/OFF) is depressed.
__________________________________________________
page 26
WARNING: Transport and installation:
1. Ensure that the room is well illuminated (at least 300 Lug). Incorrectly read measured values can result in damage to the sterilization product or inadequate re­sults.
2. To avoid damage due to radiated heat, autoclaves of the KL and TKL series must be spaced by at least
0.5 m from heat sensitive walls. The base must be heat-proof.
3. Unsuitable (sensitive to heat) wall panelling can be damaged in the event of uncontrolled emission of hot steam. We recommend the use of our condense collecting container (see Chapter 3 "Accessories").
4. There is a danger of autoclaves toppling over if they are placed on a sloping or soft base.
5. The risk of short circuits is greatly increased if mois­ture-sensitive devices (PC, meters, etc.) are set up in the same room as the autoclaves.
6. The vapours of aggressive media (e.g. acids, lees, various solvents, etc.) can corrode the autoclaves.
1. Fill the vessel with at least 1½ litre of water, but never more than ¾ of the specified maximum vessel volume. The auto­clave must never run dry.
2. Charge and close the autoclave in the manner described in Chapters 2.6.1-1.
3. Open the exhaust steam valve right up to its stop point to ensure proper venting during the heating-up stage. The autoclave can now be switched on with its mains switch.
4. Setting the sterilization temperature ("C" key) and time ("D" key): Press the given setpoint key (see fig. 1) and adjust the required value with the edit keys. The actual temperature is displayed after the settings have been completed.
5. Start the sterilization process by pressing the "F" key. WARNING: Renewed depression of the "F" key terminates the sterilization process (e.g. in the event of an incorrect en­try).
B: During operation
WARNING: Please observe the "Safety Instructions"
and "Acceptable Risks" (Chapters 1. and 1.2), but particularly the section "During Operation".
__________________________________________________
page 3
6. The autoclave is adequately vented when a strong jet
steamis emitted by the pressure control valve (approx. 4 minutes at 99-100°C thermometer display). Close the exhaust steam valve.
__________________________________________________
page 24
8. In case of sterilization of adhesive fluids an Operating Test of the Safety Valve has do be made before and after the sterilization (see chapter 6.2)
9. In case of sterilization of fluids (in open containers) we recom­mend the use of a media sensor (see chapter 4)
2.7 Operation of small-batch autoclaves, type „TKL-MCS“
WARNING: During operation:
1. An autoclave without MCS control must be con­stantly supervised during the sterilization process. It may be necessary to readjust the steam pressure control valve to avoid a short-fall or transgression of the operating pressure.
2. Inadequate venting of the autoclave prior to steriliza­tion will result in inadequate sterilization results. Please refer to the corresponding details in these Operating Instructions.
3. A short-fall of the operating temperature will result in inadequate sterilization, while an excessive operating temperature can damage the product being steril­ized. Consequently, it must be ensured that auto­claves that do not feature an MCS control must never be operated unattended.
4. The autoclave lid and walls become very hot during operation. Contact with the autoclave in this condi­tion can result in burns (except for the plastic han­dles).
5. To achieve the required sterilizing effect, the sterili­zation period for liquids must be extended by the ad­ditional time listed in table 6 page 19.
6.
__________________________________________________
page 5
Steam release valve
Safety valve
Pressure gauge
Manometer main switch microcomputer
Fig. 1 TKL – Autoclave with MCS-controller
__________________________________________________
page 22
B: After sterilizing liquids:
1. Let the autoclave cool down to at least 70°C chamber temperature.
Liquid sterilization products will boil over on account of delayed boiling and cause serious burns if the steam pressure is vented instead of allowing the autoclave to cool down slowly.
Only open the autoclave, when
the chamber temperature has cooled down to at least 70°C the autoclave has no remaining pressure (0 bar).
2. In case of sterilizing adhesive fluids an Operating Test of
the Safety Valve is necessary before and after the ster­ilization.(see chapter 6.2)
After a result of delayed boiling adhesive fluid can clog up the valves. In such an event the autoclave lid must be sent to the manufacturer so that the valves can be checked.
Continued operation with clogged valves can result in the explosion of the pressure vessel.
Risk to Life !
C: After sterilization
WARNING: Please observe the "Safety Instructions"
and "Acceptable Risks" (Chapters 1. and 1.2), but particularly the section "After Sterilization".
1. Do not vent the steam when sterilization has been completed but rather allow the autoclave to cool down slowly to at least 70°C. Liquid sterilization products will boil over on account of delayed boiling if the steam pressure is vented. In unfortu­nate circumstances such effect can gravely scald attending operators.
2. Consequently, as a result of the delayed boiling effect, the autoclave must only be opened after
it has cooled down to below 70°C. there is no remaining pressure in it (0 bar)
3. In case of sterilizing adhesive fluids an Operating Test of the Safety Valve is necessary before and after the sterilization. (see chapter 6.2)
After a result of delayed boiling adhesive fluid can clog up the valves. In such an event the autoclave lid must be sent to the manufacturer so that the valves can be checked.
Continued operation with clogged valves can result in the explosion of the pressure vessel. Risk to Life!
____________________________________________________ page 7
__________________________________________________
page 20
WARNING: Electromagnetic compatibility: 0
1. To guarantee proper sterilization after a mains failure, the sterilization must be repeated for the set time when the power has returned.
2. Do not complete any changes (e.g. a different power supply cable) without previously consulting the manufacture as this can adversely influence the autoclave's immunity to malfunctions.
2 Details concerning use
2.1 Fields of application
The autoclaves of type „KL + TKL..“ are for:
Sterilization in high-pressure steam
C: After sterilization
WARNING: Please observe the „Safety Instructions“
and „Acceptable Risks“ (chapter 1 and 1.2), but par­ticularly the section „After Sterilization“.
Turn off the heater when the sterilization time has ended
The steam pressure control valve can now be carefully
opened and the pressure slowly vented until there is no re­maining pressure in the chamber (manometer = 0 bar)
Let the autoclave cool down to at least 80°C
Before opening the lid check again if there is no pressure
remaining in the chamber
Now you can open the autoclave and take off the steriliza­tion products.
2.6.2 Sterilization of liquids
A: Preparation for operation
Do not sterilize in those autoclaves:
packed solids liquids in closed containers
further the following dangerous liquids:
explosive
high inflammatory
easy inflammatory
inflammatory (if the maximum allowed temperature
is above the flamepoint)
very toxic
toxic
effective for burning
__________________________________________________
page 9
WARNING: Please observe the "Safety Instructions" and "Ac­ceptable Risks" (Chapters 1. and 1.2), but particularly the sec­tion "Preparations for Operation".
The same preparations are required as described in Section
2.6.1 A, except that the following additional and modified pre-
paratory work is necessary:
1. Do place the container holding the liquid to be sterilized deep
but not directly on the autoclave base.
CAUTION: Do not use closed containers. Closed containers may lead to a retardation of ebullition (delayed boiling).
__________________________________________________
page 18
2.2 Transport and installation
WARNING: Please observe "Safety Instructions" and
"Acceptable Risks" (Chapters 1. and 1.2), but par­ticularly the section "Transport and Installation".
The following ambient conditions must be fulfilled to ensure optimal autoclave operation:
Ensure good lighting conditions.
The base must be level, firm and resistant to heat.
Ensure effective ventilation.
TKL-MCS autoclaves must be connected to a socket with
earthed conductor (see Table 3 "Technical Data").
Maintain the following ambient parameters for TKL-MCS autoclaves:
Table 2: Ambient conditions
Ambient conditions
Altitude for operation
Ambient temperature
Voltage fluctuations
Pollution
only indoors
up to maximum 2000 m mean sea level
from 5°C to 40°C
maximum ±10% of rated value
maximum Pollution Severity Degree 2 acc. to IEC 664
2.6.1 Sterilization of solids
A: Preparations for operation
WARNING: Please observe the "Safety Instructions"
and "Acceptable Risks" (Chapters 1. and 1.2), but particularly the section "Preparations for Operation".
1. The autoclave's lid must only be opened when the device is
in a pressureless state.
If necessary, open the steam pressure valve. The closing ever can be raised to a vertical position as soon as no more steam is being emitted by the valve. This releases the lid as it is lowered into the vessel.
Turn the lid by 90° with its closing lever and then tilt towards the valve side. The lid can then be removed.
2. Fill the vessel with at least 1,5 litre of water, but never more
than 3/4 of the specified maximum vessel volume.
The autoclave must never run dry.
The inserts are selected according to the products that are to be sterilized: Sieve tray, sieve basket and test tube rack (see Chapter 3 "Accessories").
3. The pressure vessel is closed by returning the lid in the
same manner.
Turn the lid by 90° so that the cross bracing is inserted precisely in the recessions in the edge of the vessel.
Adjust the closing lever to its horizontal closing position. An accurately inserted lid must be uniformly seated on the vessel opening.
__________________________________________________
page 11
__________________________________________________
page 16
Do not complete any changes (e.g. a different power supply cable) without previously consulting the manufacture as this can ad­versely influence the autoclave's immunity to malfunctions.
2.5 General details concerning sterilization
with high-pressure steam
Table 4a
Field of application Requirements
concerning Emitted interference Interference immunity
Living area* EN 55011 EN 50082-1
Industrial area EN 55011 EN 50082-2
* Living area: Office rooms, workshops and laboratories, small enterprises
Table 4b: Details concerning severity with regard to interference immunity
Details on electromagnetic compatibility Inspection values
Interference immunity towards static electricity discharges
Tested acc. to IEC 801-2, EN 61000-4-2 Air discharge:
Interference immunity towards transientinterference voltage (burst)
Coupling to the power supply line of the device: Tested acc. to IEC 801-4, prEN 61000-4-4
Interference immunity towards surge voltage
Coupling to the power supply line of the device:
Tested acc. to IEC 801-5, draft VDE 0843-5
Interference immunity towards voltage dip, brief interruptions and voltage fluctuations
Voltage dip with 60% reduction for more than 100 ms; angle in 45° increments
Voltage dip with 30% reduction for more than 10 ms; angle in 45° increments
±
Voltage reduction Tested acc. to IEC 801-11, EN 61000-4-11:
10% for more than 30 s
Contact dis­charge:
L - N: L - PE: N - PE:
No interference
No interference No interference
8 kV 4 kV
2 kV
2 kV 4 kV 4 kV
WARNING: Always check whether sterilization with high-pressure steam is suitable and which temperature and sterilizing time is required to determinate the given product.
Prior to each sterilization process it is necessary to check whether the selected sterilization method is suitable to kill off the germs. It is also necessary to establish whether the product that is to be sterilized will withstand the sterilization conditions. For instance, instruments with rubber or plastic parts can be significantly damaged by a temperature of e.g. 143°C (3 bar).
The subsequent Table 5 can function as a guideline for the op­eration of an autoclave. The specified values are purely for ori­entation and merely relate the sterilization time to temperature and pressure.
Table 5: Ratio between sterilization time and steam pressure and steam temperature
Sterilization time* in minutes 20 15 12 10 6 3 1
Stream pressure: bar 0,5 1,0 1,5 2,0 2,5 3,0 (4)
Stream temperature: °C 110 120 127 133 138 143 (151)
The steam temperature depends upon the steam pressure. The temperature display is significant for sterilization, and pressure display for the operating conditions inside the autoclave.
Interference immunity towards high-frequency radiation
Acc. to IEC 801-3, prEN 55024-3
3 V/m
_____________________________________________
page 13
__________________________________________________
page 14