Willtek 4032 Datasheet

4 (1)
Willtek 4032 Datasheet

Willtek

4032

STABILOCK Communications

Test System

Built-in spectrum analyzer

AUTORUN

IEEE-488 interface

High-speed measurements

Reliability

Willtek STABILOCK 4032

In excess of 5000 STABILOCK 4032 radio test sets are currently in use worldwide. This success has been due to the versatile design, simple operation and wide variety of system tests which can be made with the “universal” test set.

Applications range from high-speed production tests and quality measurements to servicing and commissioning tests. The wide variety of hardware modules and software options covering frequency ranges up to 2.3 GHz, provide comprehensive capability of analog and digital cellular, cordless and two-way radio systems.

Additional capability includes base station measurements and paging tests.

Systems include TETRA Digital Trunked Radio, Tetrapol as well as CDMA base station testing at 800 MHz and

1900 MHz. The GSM Option of the STABILOCK 4032 now is also capable of testing dual-band mobiles.

Please refer to individual data sheets for further details of system performance.

Typical performance is derived from the standard deviation of many hundreds of production units operating under normal environmental conditions of 20°C.

Built-in spectrum analyzer

Detailed spectrum analysis from 2 MHz to 1 GHz (standard) or up to 2.3 GHz (optional). Additional option for highspeed sweep and a host of useful functions.

AUTORUN

Simple, BASIC-oriented programming language for fully automated tests. Ready-written programs are available for all popular mobile telephone systems.

IEEE-488 interface

Fitted as standard. Ensures remote control via controller or program downloads from LAN servers.

High-speed measurements

Special firmware for final testing during production cuts measuring times by around 30%.

Reliability

All subassemblies are burnt-in tested, with plug-in modules and SMD technology.

Generator (Typical Performance)

Level error

(< –15 dBm, RF socket,

 

f = 1 to 1000 MHz)

Level error (FEX)

(< –20 dBm, RF socket,

 

f = 1000 to 2000 MHz)

 

 

< 0.3 dB

[1.3 dB]

<0.3 dB

[ 1.3 dB]

Analzyer (Typical Performance)

RF power measuring error

(20 to 500 MHz, broadband)

 

4% [

10%]

RF power measuring error

(800 to 920 MHz, broadband)

 

5% [

12%]

 

 

 

 

 

Audio Generator and Analyzer (Typical Performance)

 

 

 

 

 

 

 

AF generator distortion

(f > 3kHz)

 

< 0.6% [

1.0%]

AF volmeter measuring error

(f = 300 Hz to 3 kHz)

 

< 1% [

3.0%]

 

 

 

 

 

Specifications below apply for the basic 4032 up to 999.99 MHz. When the FEX (Frequency Extension) option is fitted please see * section at the end. Please refer to individual data sheets for performance of options.

2

Willtek STABILOCK 4032

AM measurement

 

 

Frequency range

 

2 to 999.9999 MHz

Measuring range

 

0 to 100 %

Input level

 

 

RF socket

 

1 mW to 125 W

RF DIRECT socket

 

0.01 mW to 0.5 W

Resolution

 

0.1%

Measuring accuracy (m

10%)

fmod = 200 Hz to 10 kHz

10% ±2 digits

Demodulation distortion

 

fmod = 300 Hz to 3 kHz

< 1%

Modulation frequency

DC to 10 kHz

Spurious-modulation measurement

Input level

 

 

RF socket

 

1 mW to 125 W

RF DIRECT socket

 

20 mV to 1 V

Measuring range

0 to –40 dB (CCITT-weighted)

 

 

referred to 3 kHz FM dev.,

 

 

3 rad M dev. or 30% AM

Measuring accuracy

 

1 dB

AF Generator

Modulation generator GEN A

 

Frequency range

30 Hz to 30 kHz

Resolution

 

f < 3 kHz

0.1 Hz

f 3 kHz

1 Hz

Frequency error

< 0.01%

Level range (EMF)

0.1 mVrms to 5 Vrms

Resolution

 

EMF

5 V

10 mV

EMF

1 V

1 mV

EMF

0.1 V

0.1 mV

EMF

10 mV

10 V

Level error

 

f = 100 Hz to 10 kHz

< 3%

f = 30 Hz to 30 kHz

< 10%

Distortion

 

f = 30 Hz to 3 kHz

< 0.5%

f > 3 kHz

< 1%

Output impedance (balanced)

 

f = 300 Hz to 3 kHz

< 10

f = 30 Hz to 30 kHz

< 40

Output impedance

 

(unbalanced)

600 ±5%

Permissible load impedance

> 200

AF Analyzer

AF voltmeter

Frequency range

30 Hz to 30 kHz or to CCITT P 53A

Measuring range

0.1 mV to 20 V

Resolution

 

Level < 0.1 V

0.1 mV

Level < 1 V

1 mV

Level < 10 V

10 mV

Level < 20 V

100 mV

Measuring accuracy

 

f = 300 Hz to 3 kHz

3%

f = 50 Hz to 15 kHz

6%

Source impedance

> 100 k or 600 ±3%

Input capacitance

20 pF

AF counter

 

Frequency range

30 Hz to 30 kHz

Input level

5 mV to 20 V

Resolution

 

f < 300 Hz

0.1 Hz

f < 10 kHz

1 Hz

f 10 kHz

10 Hz

Measuring accuracy

0.01% ±1 digit

Distortion meter

 

Input level

0.1 to 20 V

Test frequency

1 kHz ±5 Hz

Measuring range

0 to 99%

Resolution

0.1%

Measuring accuracy

 

d = 1 to 90%

5% of meas. value ±3 digits

SINAD meter

 

Input level

0.1 to 20 V

Measuring range

1 to 46 dB

Resolution

 

SINAD < 30 dB

0.1 dB

SINAD 30 dB

0.5 dB

Measuring accuracy for SINAD < 30 dB

0.8 dB ±1 digit

Scope & Analyzer

Spectrum analyzer

 

 

Frequency range

2 to 999.9999 MHz

Frequency accuracy

 

 

better than 2% of sweep width

Input level range for measuring accuracy 3 dB

in the frequency range 0.5 × fc

f 2 × fc

RF socket

 

–70 to +47 dBm

RF DIRECT socket

 

–90 to +13 dBm

Sweep width

200 kHz, 2 MHz, 10 MHz

Sweep time

 

 

Sweep width 2 MHz and 10 MHz

approx. 500 ms

Sweep width 200 kHz

 

approx. 2 s

Evaluation bandwidth

 

 

Sweep width 2 MHz and 10 MHz

30 kHz

Sweep width 200 kHz

 

6 kHz

Inherent noise on RF DIRECT socket

Sweep width 2 MHz and 10 MHz

–95 dBm

Sweep width 200 kHz

 

–105 dBm

Oscilloscope

 

Inputs external

Zi = 1 M /40 pF (AC/DC)

Inputs internal RX mod, TX demod, duplex demod, AF voltmeter, residual distortion

Frequency range

DC (3 Hz) to 20 kHz

Level error

< 10% + 0.2 div

Grating

6 x 10 div

Horizontal deflection

100 s/div to 500

s/div

Vertical deflection

2 mV/div to 10 V/div or

 

160 Hz/div to 8 kHz/div (FM)

 

0.16 rad/div to 8 rad/div (

M)

 

0.8 %/div to 40 %/div (AM)

Trigger

± slope selectable trigger level

Operating modes

auto, norm, one-shot, freeze,

time measurement (max. resolution 2.5 s)

Selective-call encoder and decoder

Standard tone sequences

 

ZVEI 1

CCIR

VDEW

ZVEI 2

EEA

NATEL

EIA

EURO

CCITT

User-defined tone sequences

Sequence of up to 30 tones can be stored by user. Also double tones and underlying continuous tone (with GEN B option).

Encoder

Operating modes

Single-tone sequence (max. 30 tones). Double-tone sequence (with GEN B option) (singletone and double-tone sequences can be transmitted continuously).

Acknowledgement call (max. 15 double tones) from response time of < 100 ms acknowledgement call only possible with optional duplex FM/ M stage

Frequency error

1 × 10Hz

Decoder

Decoding of each tone of tone sequences

(max. 30 tones). Continuous decoding can be set.

4

Loading...
+ 5 hidden pages