WIKA TWG User Manual

INSTRUCTION MANUAL
NI-401E
Rev. 3 07/99
TEMPERATURE SWITCHES
WEATHERPROOF: SERIES TWG EXPLOSIONPROOF: SERIES TAG
One contact
Stem type “B” Stem type “C” Stem type “ F”
A = bulb connection B = cable entry
PESO 2,1 kg dimensions in mm PESO 2,9 kg dimensions in mm
NOTE: dimensions and weights are not binding unless released on certified drawings. CAUTION
Before installing, using or carrying out maintenance on the instrument it is necessary to read and understand the indications given in the attached Instruction Manual.
The instrument must only be installed and maintained by qualified personnel.
Installation must only be carried out following verification of the congruity of the instrument features with the plant
and process requirements.
The functional features of the instrument and its degree of protection are shown on the identification plate fixed to the case.
CONTENTS:
1 - GENERAL NOTES 2 - OPERATING PRINCIPLE 3 - SET POINT REGULATION 4 - SET POINT CALIBRATION 5 - MOUNTING AND CONNECTIONS 6 - INSTRUMENT PLUMBING
X = Bulb sensitive portion Y = Insertion length K = Capillary length
A = bulb connection B = cable entry
7 - PUTTING INTO OPERATION 8 - FUNCTIONAL VERIFICATION 9 - TROUBLESHOOTING 10 - STOPPING AND DISMOUNTING 11 - DEMOLITION
Two contacts
X = Bulb sensitive portion Y = Insertion length K = Capillary length
All data, statements and recommendations supplied with this manual are based on information believed by us to be
reliable. As the conditions of effective use are beyond our control, our products are sold under the condition that the user himself evaluates such conditions before following our recommendations for the purpose or use foreseen by him.
INSTRUCTION MANUAL
1 - GENERAL NOTES
1.1 FOREWORD
The wrong choice of a series or a model, as well as the incorrect installation, lead to malfunction and reduce instrument life. Failure to follow the indications given in this manual can cause damage to the instrument, the environment and persons.
1.2 ALLOWED OVERRANGE
Temperatures exceeding the working range can be occasionally tolerated provided they remain within the limits stated in the instrument features (vacuum or proof temperature). Continuous temperatures exceeding the working range can be applied to the instrument, provided they are clearly stated in the instrument features. The current and voltage values stated in the technical specifications and ratings must not be exceeded. Transitory overranges can have a destructive effect on the switch.
1.3 MECHANICAL VIBRATIONS
Can generally lead to the wearing of some parts of the instrument or cause false actuation. It is therefore recommended that the instrument be installed in a place where there are no vibrations. In cases where this is impossible it is advisable to take measures to lessen the effects (elastic supports, installation with the pin of the microswitch positioned at right angles to the vibration plane, etc.).
1.4 TEMPERATURE
Due to the temperature of both the environment and the process fluid, the temperature of the instrument could exceed the allowed limits (normally from -20° to +70°C). Therefore, in case it does, suitable measures (protection against heat radiation, heated cabinets) must be taken.
2 - OPERATING PRINCIPLE
fig. 1 - Electrical connections and adjustment screws
One contact instruments Two contact instruments
The thermometric bulb is constituted by a rigid constant­volume container. It is connected via capillary tubing to a pressure measuring element (bourdon tube). This system is filled with gas, and its absolute pressure is proportional to absolute temperature (Charles’ law). As a consequence, any change in bulb temperature causes a change in gas pressure acting on sensible element, causing its free tip to undergo an elastic deflection which is used to actuate one or two electrical microswitches adjusted at prefixed set point values.
3 - SET POINT REGULATION
3.1 Each microswitch is independent and can be
regulated by means of a screw (for adjustment) in such a way that it actuates when the temperature reaches (increasing or decreasing) the desired value (set point).
3.2 The instrument is usually supplied with the switches set at 0°C or at the lowest setting range value if this is higher than 0°C (factory calibration).
3.3 The instrument is supplied with an adhesive label showing the set point calibration value. With factory calibration the values are not indicated on the label, as these are temporary and will be modified with the definitive values.
3.4 Prior to installation the instrument must be calibrated and the definitive calibration values written on the adhesive label using a suitable indelible ink pen.
3.5 If the instrument has been expressly ordered with specific calibration, it is a good rule to check the
calibration values stated on the adhesive label, prior to installation.
3.6 The position of the adjustment screws is given in fig. 1.
3.7 The effect that the direction of rotation of the
adjustment screws is described on the adhesive label.
NI-401E
Rev. 3 07/99
1 - Microswitch set point calibration screw 3 - Terminal block 4 - Electrical connection identification plate
Designation of the contacts: C - Common NA - Normally Open NC - Normally Closed
1 – Microswitch 2 set point calibration screw 2 – Microswitch 1 set point calibration screw 3 – Terminal block 4 – Electrical connection identification plate
Microswitch electrical circuit: State of the contacts at initial temperature range (when factory calibrated)
Loading...
+ 4 hidden pages