WIKA TWG User Manual

INSTRUCTION MANUAL
NI-401E
Rev. 3 07/99
TEMPERATURE SWITCHES
WEATHERPROOF: SERIES TWG EXPLOSIONPROOF: SERIES TAG
One contact
Stem type “B” Stem type “C” Stem type “ F”
A = bulb connection B = cable entry
PESO 2,1 kg dimensions in mm PESO 2,9 kg dimensions in mm
NOTE: dimensions and weights are not binding unless released on certified drawings. CAUTION
Before installing, using or carrying out maintenance on the instrument it is necessary to read and understand the indications given in the attached Instruction Manual.
The instrument must only be installed and maintained by qualified personnel.
Installation must only be carried out following verification of the congruity of the instrument features with the plant
and process requirements.
The functional features of the instrument and its degree of protection are shown on the identification plate fixed to the case.
CONTENTS:
1 - GENERAL NOTES 2 - OPERATING PRINCIPLE 3 - SET POINT REGULATION 4 - SET POINT CALIBRATION 5 - MOUNTING AND CONNECTIONS 6 - INSTRUMENT PLUMBING
X = Bulb sensitive portion Y = Insertion length K = Capillary length
A = bulb connection B = cable entry
7 - PUTTING INTO OPERATION 8 - FUNCTIONAL VERIFICATION 9 - TROUBLESHOOTING 10 - STOPPING AND DISMOUNTING 11 - DEMOLITION
Two contacts
X = Bulb sensitive portion Y = Insertion length K = Capillary length
All data, statements and recommendations supplied with this manual are based on information believed by us to be
reliable. As the conditions of effective use are beyond our control, our products are sold under the condition that the user himself evaluates such conditions before following our recommendations for the purpose or use foreseen by him.
INSTRUCTION MANUAL
1 - GENERAL NOTES
1.1 FOREWORD
The wrong choice of a series or a model, as well as the incorrect installation, lead to malfunction and reduce instrument life. Failure to follow the indications given in this manual can cause damage to the instrument, the environment and persons.
1.2 ALLOWED OVERRANGE
Temperatures exceeding the working range can be occasionally tolerated provided they remain within the limits stated in the instrument features (vacuum or proof temperature). Continuous temperatures exceeding the working range can be applied to the instrument, provided they are clearly stated in the instrument features. The current and voltage values stated in the technical specifications and ratings must not be exceeded. Transitory overranges can have a destructive effect on the switch.
1.3 MECHANICAL VIBRATIONS
Can generally lead to the wearing of some parts of the instrument or cause false actuation. It is therefore recommended that the instrument be installed in a place where there are no vibrations. In cases where this is impossible it is advisable to take measures to lessen the effects (elastic supports, installation with the pin of the microswitch positioned at right angles to the vibration plane, etc.).
1.4 TEMPERATURE
Due to the temperature of both the environment and the process fluid, the temperature of the instrument could exceed the allowed limits (normally from -20° to +70°C). Therefore, in case it does, suitable measures (protection against heat radiation, heated cabinets) must be taken.
2 - OPERATING PRINCIPLE
fig. 1 - Electrical connections and adjustment screws
One contact instruments Two contact instruments
The thermometric bulb is constituted by a rigid constant­volume container. It is connected via capillary tubing to a pressure measuring element (bourdon tube). This system is filled with gas, and its absolute pressure is proportional to absolute temperature (Charles’ law). As a consequence, any change in bulb temperature causes a change in gas pressure acting on sensible element, causing its free tip to undergo an elastic deflection which is used to actuate one or two electrical microswitches adjusted at prefixed set point values.
3 - SET POINT REGULATION
3.1 Each microswitch is independent and can be
regulated by means of a screw (for adjustment) in such a way that it actuates when the temperature reaches (increasing or decreasing) the desired value (set point).
3.2 The instrument is usually supplied with the switches set at 0°C or at the lowest setting range value if this is higher than 0°C (factory calibration).
3.3 The instrument is supplied with an adhesive label showing the set point calibration value. With factory calibration the values are not indicated on the label, as these are temporary and will be modified with the definitive values.
3.4 Prior to installation the instrument must be calibrated and the definitive calibration values written on the adhesive label using a suitable indelible ink pen.
3.5 If the instrument has been expressly ordered with specific calibration, it is a good rule to check the
calibration values stated on the adhesive label, prior to installation.
3.6 The position of the adjustment screws is given in fig. 1.
3.7 The effect that the direction of rotation of the
adjustment screws is described on the adhesive label.
NI-401E
Rev. 3 07/99
1 - Microswitch set point calibration screw 3 - Terminal block 4 - Electrical connection identification plate
Designation of the contacts: C - Common NA - Normally Open NC - Normally Closed
1 – Microswitch 2 set point calibration screw 2 – Microswitch 1 set point calibration screw 3 – Terminal block 4 – Electrical connection identification plate
Microswitch electrical circuit: State of the contacts at initial temperature range (when factory calibrated)
3
INSTRUCTION MANUAL
4 - SET POINT CALIBRATION
In order to proceed with the calibration and the periodical functional verification of the instrument a suitable calibration circuit (fig. 3) and an adequate heat source is required.
4.1 PRELIMINARY OPERATIONS
4.1.1 Weatherproof temperature switches (Series
TWG) (Fig. 2)
Remove the blocking device fixed to the side of the instrument case and the adjustment screw access plate. Remove the cover be rotating it anticlockwise.
fig. 2 - Weatherproof tempeature switch cover blocking
device and plumbing
NI-401E
Rev. 3 07/99
Connection of C and NO terminals
If the circuit is open at the working temperature, the switch closes the circuit as the temperature increases when the desired value is reached.
If the circuit is closed at the working temperature, the switch opens the circuit as the temperature decreases when the desired value is reached.
Connection of C and NC terminals
If the circuit is closed at the working temperature, the switch opens the circuit as the temperature increases when the desired value is reached.
If the circuit is open at the working temperature, the switch closes the circuit as the temperature decreases when the desired value is reached.
4.2.3 The test instrument should have a measurement
range approximately equal to or slightly wider than the temperature switch range and should have an accuracy consistent with the precision required to calibrate the set point.
4.2.4 The temperature switch must be mounted in the
normal installation position, i.e. with the stem or capillary outlet downwards.
fig. 4 - Calibration circuit
a) Plumbing wire b) Plumbing c) Blocking nut d) Blocking bracket e) Adjustment screws access plate
4.1.2 Explosionproof temperature switches (Series TAG) (Fig. 3)
Loosen the locking headless screw situated on the cover using a 1,5 hexagonal key then unscrew the cover. Remove the internal blocking device inserted on the closure plugs and slide out the plugs.
fig. 3 - Explosionproof temperature switch cover blocking
device
4.1.
4.2 CALIBRATION CIRCUIT AND OPERATIONS
4.2.1 Prepare the control circuit as indicated in Fig. 4.
4.2.2 The warning lamps should be connected to contact 1
or 2 in the NO or NC position according to the required contact action.
TS - Temperature switch TT - Test thermometer TB - Thermostatic bath
4.2.5 Avoid forcing the microswitch by hand or with tools.
This could affect the instrument functioning. CAUTION: If the switch is of the kind with adjustable dead
band (letter R in the contact codes) before proceeding with the following operations (4.2.6-4.2.11) it is necessary to proceed with the adjustment of the differential (see attachment NI-705).
4.2.6 Increase the temperature in the circuit up to the
desired set point value for the first microswitch.
4.2.7 Use a wide bladed screwdriver, as indicated on the
adhesive label, until the relative lamp turns on (or turns off).
- If the instrument is equipped with only one contact the
calibration is complete.
- If it is equipped with two contacts continue in the
following manner.
4.2.8 Vary the temperature until the desired set point value
for the second microswitch is reached.
4.2.9 Act on the adjustment screw of the second contact
as in point 4.2.7.
4.2.10 Repeat operations 4.2.6 and 4.2.7 on the first
contact, then operations 4.2.8 and 4.2.9 on the second contact, until the required set point precision is obtained. This is necessary due to the reciprocal influence which the microswitches have with the sensitive element of the instrument.
INSTRUCTION MANUAL
4.2.11 Check the calibration values (varying the temperature in the circuit accordingly) and record them on the adhesive label using a pen with indelible ink.
4.3 FINAL OPERATIONS
4.3.1 Disconnect the instrument from the calibration circuit.
4.3.2 Weatherproof temperature switches (Series TWG)
Take the cover, ensure that the sealing gasket is correctly fitted into its seat, insert the cover onto the case, with the blocking gap positioned in correspondence to the blocking bracket. Turn the cover clockwise closing it tightly. Mount the adjustment screws access plate, then the blocking device as in figure 2.
4.3.3 Explosionproof temperature switches (Series TAG)
Insert the closure plugs of the adjustment screw access holes, block them using the internal device and if necessary seal them with plumbing. Screw on the cover and block it using the headless screw with which it is equipped (Fig. 3).
4.3.4 Replace the supplied protection cap on the cable conduit. IMPORTANT: the protection cap should only be definitively removed during the connection steps (see §5)
5 - MOUNTING AND CONNECTIONS
5.1 MOUNTING Surface mount the instrument by means of the holes
provided, or pipe mount using the appropriate bracket (see figures 6 and 7) in a vertical position (with the stem or capillary outlet downwards). The chosen position must be such that vibrations, accidental shocks or temperature changes are within tolerable limits. The above also applies to direct mounting.
5.2 BULB AND CAPILLARY
With reference to figure 5 unscrew the fitting (3) from the seal press (2) and slide it off from the bulb (5). Mount the fitting (3) on the thermowell and tighten it using the appropriate key. Insert the bulb (5) into the thermowell (4) after covering it with the paste to improve the transmission of heat. Verify that the bulb touches the bottom. Insert the PTFE seal with the relative stainless steel washers into the fitting (3). Screw the seal press (2) onto the fitting (3) taking care not to bend the capillary and relative sheath and tighten until the PTFE seal is tight on the capillary tube. Run the capillary protected by the armor in the established direction, avoiding tight bends, and block using the stainless steel bands. If a large amount of capillary remains this should be rolled up and fixed tightly. The coil must not have a diameter of less than 200 mm.
5.3 ELECTRICAL CONNECTIONS
5.3.1 Set up the cable protection tube according to the applicable standards (especially for explosion-proof
prescriptions). In many applications this is associated to the process piping and is subject to condensation. For this reason it is necessary to provide means to prevent condensation from entering the instrument case. The arrangement shown in figure 6 or 7 is therefore recommended.
5.3.2 Check that there is no power in the lines.
5.3.3 Remove the cover and carry out the cabling and
connections to the terminal block (see fig. 1). Flexible cables with a maximum section of 1.2 mm (16AWG) are recommended using pre-insulated fork thimbles. Do not touch the adjustment screws and do
2
NI-401E
Rev. 3 07/99
not bend the elastic microswitch supports in order to prevent the instrument calibration being altered.
5.3.4 Ensure that no deposits or wire ends remain inside
the case.
5.3.5 Once the connection operations have been
completed, replace the cover and ensure that it is properly sealed and blocked. See Fig. 2 and 3.
6 - INSTRUMENT PLUMBING
6.1 Weatherproof temperatures switches (Series
TWG)
The plumbing, aimed as a guarantee against possible tampering of the calibration and electrical connections, can be carried out using a flexible steel wire (a) inserted into the holes in the locking nut (c) and the bracket (d) provided for this purpose (see figure 2).
6.2 Explosionproof temperature switches (Series
TAG)
Plumbing is not necessary as the cover is blocked with a headless screw, the closure plugs of the adjustment screw access holes are blocked by means of the internal blocking device and the instrument does not have to be opened when installed.
fig. 5 - Mounting the bulb
1) Armored capillary
2) Stuffing nut (SW 12)
3) Rotating fitting (SW 22)
4) Thermowell
7 - PUTTING INTO OPERATION
As the signal transmitted by the instrument is used in a complex system, it is necessary that the means of putting it into operation are established by those in charge of the plant. The instrument starts working as soon as it is connected to an electrical line.
8 - FUNCTIONAL VERIFICATION
This will be carried out according to the Client’s control procedures. Series TWG instruments can be verified. Series TAG instruments can only be verified on the plant if test equipment suitable for the environment is available and if there is no current in the electric line. If this is not the case it is necessary to stop operation, dismount by means of the three piece joints and carry out the verification in a test room. Verification consists in checking the calibration value and possibly regulatory the adjustment bush (see §4).
5) Bulb
6) Stainless steel washer
7) PTFE washer
INSTRUCTION MANUAL
9 - TROUBLESHOOTING IMPORTANT NOTE: operations involving replacement of essential components must be carried out at our
workshop, especially for instruments with explosionproof certificate; this is to guarantee the user the total and correct restoration of the product’s original characteristics.
MALFUNCTION PROBABLE CAUSE REMEDY
Wear of contact surfaces between
microswitch pin and sensing element tip.
Wear of contact surfaces between
microswitch cradle and adjustment screw.
Possible encrustations or corrosions on
Set point shift
Poor Repeatability
Slow Response
Failed or Undue Actuation
Undue Actuation
10 - STOPPING AND DISMOUNTING
Before proceeding with these operations ensure that the plant or machines have been put into the conditions foreseen to allow these operations.
10.1 Remove the power supply (signal) from the electrical line.
With reference to figure 5.
10.2 Loosen and remove the stuffing nat (2) being careful
not to bend the capillary and protective sheath (1).
10.3 Loosen and remove the fitting (3) then extract the bulb (5) from the thermowell (4) holding it by the capillary tube (1), without bending it.
Do not waste in to environment the process fluid if polluting or harmful to people.
With reference to figures 6 and 7.
above listed surfaces.
Permanent deformation of the sensitive
element due to fatigue or non-tolerated over-ranges.
Loss of filling fluid. Assembling screws loosened. Encrustations on bulb or thermowell Check and clean encrustated surfaces
Microswitch contacts damaged Electrical joints loosened Electrical line interrupted, or short circuit Accidental impacts or excessive
mechanical vibrations
10.4 Unscrew the three piece joint (8) (electrical cable
tubing).
10.5 Remove the instrument cover and disconnect the
electrical cables from the terminal block and earth screws.
10.6 Remove the screws fixing the case to the panel (or
pipe) and remove the instrument, taking care to slide the electrical conductors out from the case.
Replace the instrument cover. Insulate and protect the conductors remaining on the plant. Temporarily plug the thermowell.
11 - DEMOLITION
The instruments are mainly made of stainless steel and aluminium and therefore, once the electrical parts have been dismounted and the parts coming into contact with fluids which could be harmful to people or the environment have been properly dealt with, can be scrapped.
Recalibrate Recalibrate Clean surfaces and recalibrate. Check
housing tightness.
Recalibrate or replace the sensitive
element with another made of a suitable material. If necessary apply a fluid separator.
Replace sensing element with another
one of same type
Check terminal screws, microswitch,
electrical subassembly, fastening screws
Replace microswitch Check all electrical joints Check state of electrical line Modify installation arrangement
NI-401E
Rev. 3 07/99
INSTRUCTION MANUAL
NI-401E
Rev. 3 07/99
WEATHERPROOF EXPLOSIONPROOF
fig. 6 - Example of connections - fig. 7 - Example of connections -
1) Weatherproof temperature switch series TWG
2) Capillary spacer
3) Thermowell
4) Bulb
5) Process piping
6) Derivation
7) Flexible armour
8) Cable gland
9) Bracket for 2” pipe
10) 2” pipe
fig. 8 - Thermometric wells: example of installation
Minimum dimensions 3"; for lesser dimensions provide for an increase in diameter up to 3".
Minimum dimensions 6"; for lesser dimensions provide for an increase in diameter up to 6".
1) Explosionproof temperature switch series TAG
2) Capillary spacer
3) Thermowell
4) Bulb
5) Process piping
6) Derivation
7) Blocking joint
8) Cable gland
9) Bracket for 2” pipe
10) 2” pipe
ETTORE CELLA SPA Viale de Gasperi, 48 - Casella Postale (P.O. Box) 96 - I 20010 Bareggio (MILANO) ITALY
Telefoni +39 029036.1146/1237/1241 – FAX +39 029036.1331
e-mail: cella@ecellaspa.com
Loading...