WIKA TC12-A, TR12-M, TC12-B, TC12-M, TR12-B Operating Instructions Manual

Operating instructions Betriebsanleitung Mode d‘emploi Manual de instrucciones
EN
DE
FR
Models TR12-B, TC12-B Models TR12-M, TC12-MModels TR12-A, TC12-A
Resistance thermometer TR12 and thermocouple TC12
Widerstandsthermometer TR12 und Thermoelement TC12
Sonde à résistance TR12 et thermocouple TC12
Termorresistencia TR12 y termopar TC12
EN
DE
FR
2
14064370.02 08/2018 EN/DE/FR/ES
WIKA operating instructions models TR12, TC12
Operating instructions models TR12 and TC12
Page
3 - 28
Betriebsanleitung Typen TR12 und TC12
Seite
29 - 54
Mode d'emploi types TR12 et TC12
Page
55 - 80
Manual de instrucciones modelos TR12 y TC12
Página
81 - 105
© 08/2013 WIKA Alexander Wiegand SE & Co. KG All rights reserved. / Alle Rechte vorbehalten. WIKA
®
is a registered trademark in various countries.
WIKA
®
ist eine geschützte Marke in verschiedenen Ländern.
Prior to starting any work, read the operating instructions! Keep for later use!
Vor Beginn aller Arbeiten Betriebsanleitung lesen! Zum späteren Gebrauch aufbewahren!
Lire le mode d‘emploi avant de commencer toute opération ! A conserver pour une utilisation ultérieure !
¡Leer el manual de instrucciones antes de comenzar cualquier trabajo! ¡Guardar el manual para una eventual consulta!
EN
14064370.02 08/2018 EN/DE/FR/ES
WIKA operating instructions models TR12, TC12
3
Contents
Contents
1. General information 4
2. Design and function 4
3. Safety 8
4. Transport, packaging and storage 11
5. Commissioning, operation 12
6. Faults 17
7. Maintenance, cleaning and calibration 19
8. Dismounting, return and disposal 20
9. Specifications 21
10. Accessories 27
EN
14064370.02 08/2018 EN/DE/FR/ES
WIKA operating instructions models TR12, TC12
4
1. General information
The thermometers described in the operating instructions have been manufactured using state-of-the-art technology.
These operating instructions contain important information on handling the instrument. Working safely requires that all safety instructions and work instructions are observed.
Observe the relevant local accident prevention regulations and general safety regulations for the instrument’s range of use.
Skilled personnel must have carefully read and understood the operating instructions prior to beginning any work.
Subject to technical modifications.
Further information:
- Internet address: www.wika.de / www.wika.com
- Relevant data sheet: TE 60.16 (TR12-A) TE 60.17 (TR12-B, TR12-M) TE 65.16 (TC12-A) TE 65.17 (TC12-B, TC12-M)
- Application consultant:
Tel.: +49 9372 132-0 Fax: +49 9372 132-406 info@wika.com
Abbreviations RTD “Resistance temperature detector”; resistance thermometer
TC Thermocouple
2. Design and function
2.1 Description
The model TR12-B (resistance thermometer) and model TC12-B (thermocouple) electrical thermometers consist of a module (TR12-M, TC12-M) which is built into a case. The module is made up of a spring-loaded measuring insert (TR12-A, TC12-A) built into a neck tube. The measuring insert (TR12-A, TC12-A) is replaceable.
The active measuring component of the measuring insert is manufactured from a welded tube or from mineral-insulated cable, optionally in combination with ceramic-insulated thermocouple wires. The sensor is embedded in ceramic powder, heat-resistant potting compound, cement compound or thermally conductive paste.
1. General information / 2. Design and function
EN
14064370.02 08/2018 EN/DE/FR/ES
WIKA operating instructions models TR12, TC12
5
If the temperature sensor is designed as a grounded thermocouple, the thermocouple is joined directly to the sheath. Versions with a diameter smaller than 3 mm and with grounded thermocouples should be considered as galvanically connected with earth potential.
The connection side of the measuring insert consists of a transition sleeve with connected bare wires.
This document describes standard versions of instruments. For applications in hazardous areas special instrument designs are required.
For further information for operation in hazardous areas, see the additional information for the corresponding ignition protection type (separate document).
CAUTION! Damage to the instrument
To avoid damage to the instrument, thermometers of this model range must be installed with a thermowell.
Select a suitable thermowell (any type of thermowell can be selected) and take the operational process data (temperature, pressure, density and flow velocity) into account.
The use of special designs without thermowell is possible, but is the responsibility of the operator.
Possible sensor measuring ranges:
Model TR12: -200 ... +600 °C Model TC12: -40 ... +1,200 °C
The following mounting and operating information has been compiled with care. However, it is not possible to consider all potential usage cases.
2. Design and function
EN
14064370.02 08/2018 EN/DE/FR/ES
WIKA operating instructions models TR12, TC12
6
2.2 Technical description of the three variants
Legend:
Neck tube
Thread to the thermowell
Measuring insert
Thread to the connection head
A(U
2
) Insertion length
(tapered thread)
A(L
1
) Insertion length
(parallel thread) NL Nominal length N(M
H
) Neck length
tapered thread
parallel thread
14013854.02
parallel thread
tapered thread
fabricated neck tube
(neck tube welded)
“nipple-union-nipple”
neck tube
tapered thread
parallel thread with counter nut
TR12-B, TC12-B Variant 1
TR12-B, TC12-B Variant 2
TR12-B, TC12-B Variant 3
Thread
TR12-M, TC12-M Module
Thread
Thread Thread Thread
Legend:
Connection head
Neck tube
Connection to thermowell
Measuring insert
Terminal block
Transmitter (option)
Field transmitter
14039769.01
T
max
= +80 °C
T
max
= +150 °C
T
max
= +300 °C
T
max
= undefined
2.3 Neck tube versions
2. Design and function
EN
14064370.02 08/2018 EN/DE/FR/ES
WIKA operating instructions models TR12, TC12
7
2.4 Case- and connection heads
The dimensions of the case- and connection heads are given in the respective data sheet.
2.5 Tightening torques
Connection head, selectable (example)
Thread Tightening torques in Nm
Connection head material Aluminium Stainless steel
1/2 NPT 32 35 3/4 NPT 36 40 M20 x 1.5, with counter nut
1)
23 25
M24 x 1.5, with counter nut
1)
27 30
Tightening torques between connection head and neck tube
Tightening torques for connection to thermowell
Tightening torques for connection to neck tube
1) only for versions with fabricated neck tube
Thread Tightening torques in Nm
R 1/2
1)
50 ... 60
Thread Tightening torques in Nm
1/2 NPT 35 3/4 NPT 40 G 1/2 B 35 G 3/4 B 40 M14 x 1.5 25 ... 30 M18 x 1.5 35 M20 x 1.5 35 ... 40 M27 x 2 40 ... 45
Only ever screw in, or unscrew, the instrument via the spanner flats and to the prescribed torque using an appropriate tool.
The correct torque depends on the dimensions of the connection thread and the sealing used (form/material).
When screwing in or unscrewing the instrument, do not use the connection head as contact surface.
When screwing in the instrument, please observe that the threads are not skewed.
2.6 Scope of delivery
Cross-check scope of delivery with delivery note.
2. Design and function
EN
14064370.02 08/2018 EN/DE/FR/ES
WIKA operating instructions models TR12, TC12
8
3. Safety
3.1 Explanation of symbols
WARNING!
... indicates a potentially dangerous situation that can result in serious injury or death, if not avoided.
CAUTION!
... indicates a potentially dangerous situation that can result in light injuries or damage to property or the environment, if not avoided.
WARNING!
... indicates a potentially dangerous situation that can result in burns, caused by hot surfaces or liquids, if not avoided.
Information
... points out useful tips, recommendations and information for efficient and trouble-free operation.
3.2 Intended use
These resistance thermometers and thermocouples are used for temperature measurement in industrial applications. They can be combined with a variety of thermowell designs, but the operational process data (temperature, pressure, density and flow velocity) must be taken into account. Operation without thermowell is only recommended in certain applications. The replaceable, centrically spring-loaded measuring insert and its extended spring travel enable combination with the widest range of connection head designs.
Neither repairs nor structural modifications are permitted, and any would void the guarantee and the respective certification. The manufacturer shall not be responsible for constructional modifications after delivery of the instruments.
The instrument has been designed and built solely for the intended use described here, and may only be used accordingly. The technical specifications contained in these operating instructions must be observed.
The manufacturer shall not be liable for claims of any type based on operation contrary to the intended use.
3. Safety
EN
14064370.02 08/2018 EN/DE/FR/ES
WIKA operating instructions models TR12, TC12
9
3.3 Responsibility of the operator
The system operator is responsible for selecting the thermometer or thermowell, and for the selection of their materials, so as to guarantee their safe operation within the plant or machine. When preparing a quote, WIKA can only give recommendations which are based on our experience in similar applications.
The safety instructions within these operating instructions, as well as the safety, accident prevention and environmental protection regulations for the application area must be maintained.
The operator is obliged to maintain the product label in a legible condition.
3.4 Personnel qualification
WARNING! Risk of injury should qualification be insufficient
Improper handling can result in considerable injury and damage to equipment.
The activities described in these operating instructions may only be carried out by skilled electrical personnel who have the qualifications described below.
Skilled electrical personnel
Skilled electrical personnel are understood to be personnel who, based on their technical training, know-how and experience as well as their knowledge of country-specific regulations, current standards and directives, are capable of carrying out work on electrical systems and independently recognising and avoiding potential hazards. The skilled electrical personnel have been specifically trained for the work environment they are working in and know the relevant standards and regulations. The skilled electrical personnel must comply with current legal accident prevention regulations.
Operating personnel
The personnel trained by the operator are understood to be personnel who, based on their education, knowledge and experience, are capable of carrying out the work described and independently recognising potential hazards.
Special operating conditions require further appropriate knowledge, e.g. of aggressive media.
3. Safety
EN
14064370.02 08/2018 EN/DE/FR/ES
WIKA operating instructions models TR12, TC12
10
3.5 Labelling, safety marks
Product label (example)
Model
A = measuring insert B = process thermometer M = basic module
Serial number
Approval-related data
Year of manufacture
Information on version (measuring element, measuring range...)
Sensor in accordance with standard (resistance thermometer)
- F =
Thin-film measuring resistor
- W = Wire-wound measuring resistor
Sensor in accordance with standard (thermocouple)
- ungrounded = ungrounded welded
- grounded = welded to the sheath (grounded)
- quasi grounded = The thermometer is, due to its low insulation clearances between resistance sensor and sheath, to be considered as grounded.
Transmitter model (only for design with transmitter)
Before mounting and commissioning the instrument, ensure you read the operating instructions!
TR12-M-IDBZ
1 x Pt100 / A / 3 (F) -50 ... +500 °C
Um = DC 30 V / Pm = 2 W
BVS 07 ATEX E 071 X IECEx BVS 11.0042X II 2 G Ex db IIC T* Gb
EN 60751
0158
WIKA Alexander Wiegand SE & Co.KG, D-63911 Klingenberg Made in Germany 2013
WARNING: DO NOT OPEN WHILE ENERGIZED!
11012345
3. Safety
EN
14064370.02 08/2018 EN/DE/FR/ES
WIKA operating instructions models TR12, TC12
11
4. Transport, packaging and storage
4.1 Transport
Check the instrument for any damage that may have been caused by transport. Obvious damage must be reported immediately.
CAUTION! Damage through improper transport
With improper transport, a high level of damage to property can occur.
When unloading packed goods upon delivery as well as during internal transport, proceed carefully and observe the symbols on the packaging.
With internal transport, observe the instructions in chapter 4.2 “Packaging and storage”.
If the instrument is transported from a cold into a warm environment, the formation of condensation may result in instrument malfunction. Before putting it back into operation, wait for the instrument temperature and the room temperature to equalise.
4.2 Packaging and storage
Do not remove packaging until just before mounting.
Permissible conditions at the place of storage:
Storage temperature: Instruments without built-in transmitter: -40 ... +80 °C Instruments with built-in transmitter: see operating instructions of the transmitter in
question
Humidity: 35 ... 85 % relative humidity (non-condensing)
Avoid exposure to the following factors:
Direct sunlight or proximity to hot objects
Mechanical vibration, mechanical shock (putting it down hard)
Soot, vapour, dust and corrosive gases
Hazardous environments, flammable atmospheres
Store the instrument in its original packaging in a location that fulfils the conditions listed above. If the original packaging is not available, pack and store the instrument as described below:
1. Place the instrument, along with shock-absorbent material, in the packaging.
2. If stored for a prolonged period of time (more than 30 days), place a bag containing a desiccant inside the packaging.
4. Transport, packaging and storage
EN
14064370.02 08/2018 EN/DE/FR/ES
WIKA operating instructions models TR12, TC12
12
5. Commissioning, operation
WARNING! Damage to the measuring instrument by operation outside the upper or lower limits of the operating temperature
Failure to observe the permissible operating temperature, also taking into account convection and radiation, can even cause damage to the thermometer during mounting.
The upper and lower limits of the specified operating temperature range must not be exceeded.
During installation, take care to
Avoid distorting the cable sheath when tightening the pressure screw.
Avoid cutting too deep into the cable sheath.
Use suitable cable.
Be careful of the clamping area of the cable gland.
Thermometers must be earthed if dangerous voltages could be expected at the connection wires (caused, for example, by mechanical damage, electrostatic charge or induction)!
The protection class is not valid with armoured cables (stainless steel sheathed).
Seals should be checked for signs of brittleness and, if necessary, replaced.
5.1 Removal and installation of the measuring insert
The bare wires have a cross-section of approx. 0.22 mm², are 150 mm long and are colour­coded dependent upon the sensor type. The measuring insert is secured against twisting.
Before removing the measuring insert, fully disconnect the electrical connections to the terminal block or transmitter.
After that, the neck tube can be loosened and unscrewed from the head.
5. Commissioning, operation
EN
14064370.02 08/2018 EN/DE/FR/ES
WIKA operating instructions models TR12, TC12
13
Removed measuring insert with neck tube:
To disconnect the measuring insert from the neck tube, loosen the M16 bolt at the top end of the neck tube
and unscrew it.
5. Commissioning, operation
EN
14064370.02 08/2018 EN/DE/FR/ES
WIKA operating instructions models TR12, TC12
14
The installation of the measuring insert is carried out in the reverse order (clean the measuring insert prior to installation). The hexagonal crimped tip of the measuring insert is guided by the screw-in of the hexagonal socket screw. Tightening torque of the screw: 12 ... 14 Nm
5.2 Electrical mounting
Cable glands
Requirements for meeting ingress protection:
Only use cable glands within their indicated clamping area (cable diameter suitable for the cable gland).
Do not use the lower clamping area with very soft cable types.
Only use round cables (if necessary, slightly oval in cross-section).
Do not twist the cable.
Repeated opening/closing is possible; however only if necessary, as it might have a detrimental effect on the ingress protection
For cables with a pronounced cold-flow behaviour the screw connection must be fully tightened.
5.3 Electrical connection
CAUTION! Danger of short circuit
Damage to cables, wires and connection points can lead to malfunction of the instrument.
Avoid damaging the cables and wires.
Fine-stranded leads with bare ends must be finished with end splices (cable preparation).
Both the internal effective capacitance and inductance must be considered.
The electrical connection is to be made according to the following sensor connections/ pin assignments.
5. Commissioning, operation
EN
14064370.02 08/2018 EN/DE/FR/ES
WIKA operating instructions models TR12, TC12
15
5.3.1 Resistance thermometer
With terminal block
With connection cable
3160629.06
red
1 x Pt100, 2-wire 1 x Pt100, 3-wire 1 x Pt100, 4-wire
white white
red
red
red
red
white
white
white
white
white white
red red
red red
red
2 x Pt100, 2-wire
2 x Pt100, 3-wire
2 x Pt100, 4-wire
red
white
white
red
red
white
black
yellow
yellow
yellow
black
black
white white
red red
red red
black black
black black
yellow
yellow yellow
red white
black
yellow
white
1 x Pt100 2-wire
1 x Pt100 3-wire
1 x Pt100 4-wire
2 x Pt100 2-wire
2 x Pt100 3-wire
red
red red
red red
red
red red
white white
white
white
white
white
black
black
black
yellow
yellow
3160629.06
5. Commissioning, operation
EN
14064370.02 08/2018 EN/DE/FR/ES
WIKA operating instructions models TR12, TC12
16
5.3.2 Thermocouples
With terminal block
With connection cable
Colour coding of cable strands
Sensor type Standard Positive Negative
K IEC 60584 Green White J IEC 60584 Black White E IEC 60584 Violet White N IEC 60584 Pink White
Variant 1
For the electrical specifications (e.g. connection diagrams, tolerance values, etc.), see chapter 6.1 “Electrical connection” or the data sheets TE 60.17 (for TR12) and TE 65.17 (for TC12).
Variant 2
For the electrical specifications (e.g. connection diagrams, tolerance values, etc.) please refer to the relevant operating instructions and/or data sheet for the built-in head-mounted transmitter.
Variant 3
For the electrical specifications (e.g. connection diagrams, tolerance values, etc.) please refer to the relevant operating instructions and/or the relevant data sheet for the built-in field transmitter.
Single thermocouple Dual thermocouple
The colour coding at the positive poles of the instrument decides the correlation of polarity and terminal.
3166822.03
Single thermocouple Dual thermocouple
5. Commissioning, operation
EN
14064370.02 08/2018 EN/DE/FR/ES
WIKA operating instructions models TR12, TC12
17
Variant 1 and 2
Junction between cable gland and connection head M20 x 1.5 thread: Tightening torques 12 Nm ½ NPT thread: Tightening torques 30 Nm
Junction between cable and cable gland Screw the pressure screw tightly into the adapter (use appropriate tools!)
6. Faults
CAUTION! Physical injuries and damage to property and the environment
If faults cannot be eliminated by means of the listed measures, the instrument must be taken out of operation immediately.
Ensure that there is no longer any signal present and protect against being put into operation accidentally.
Contact the manufacturer.
If a return is needed, please follow the instructions given in chapter 8.2 “Return”.
WARNING! Physical injuries and damage to property and the environment caused by hazardous media
Upon contact with hazardous media (e.g. oxygen, acetylene, flammable or toxic substances), harmful media (e.g. corrosive, toxic, carcinogenic, radioactive), and also with refrigeration plants and compressors, there is a danger of physical injuries and damage to property and the environment. Should a failure occur, aggressive media with extremely high temperature and under high pressure or vacuum may be present at the instrument.
For these media, in addition to all standard regulations, the appropriate existing codes or regulations must also be followed.
Wear the required protective equipment (depending on the application; the thermometer itself is basically not dangerous).
For contact details see chapter 1 “General information” or the back page of the operating instructions.
5. Commissioning, operation / 6. Faults
EN
14064370.02 08/2018 EN/DE/FR/ES
WIKA operating instructions models TR12, TC12
18
Faults Causes Measures
No signal/cable break Mechanical load too high or
overtemperature
Replace probe or measuring insert with one of a suitable design
Erroneous measured values
Sensor drift caused by overtemperature
Replace probe or measuring insert with one of a suitable design
Sensor drift caused by chemical attack
Use a suitable thermowell.
Erroneous measured values (too low)
Entry of moisture into cable or measuring insert
Replace probe or measuring insert with one of a suitable design
Erroneous measured values and response times too long
Wrong mounting geometry, for example mounting depth too deep or heat dissipation too high
The temperature-sensitive area of the sensor must be inside the medium, and surface measurements must be ungrounded
Deposits on the sensor or thermowell
Remove deposits
Erroneous measured values (of thermocouples)
Parasitic voltages (thermal voltages, galvanic voltage) or wrong equalisation line
Use suitable compensating cable
Temporary or sporadic interruptions of the measured value signal
Cable break in connection cable or loose contact caused by mechanical overload
Replace probe or measuring insert with a suitable design, for example equipped with a bend protection spring or a thicker conductor cross­section
Corrosion Composition of the medium
not as expected or modified or wrong thermowell material selected
Analyse medium and then select a more suitable material or replace thermowell regularly
Signal interference Stray currents caused by
electric fields or earth circuits
Use shielded connection leads, and increase the distance to motors and power cables
Earth circuits Eliminate potential differences by
using galvanically isolated barriers or transmitters
6. Faults
EN
14064370.02 08/2018 EN/DE/FR/ES
WIKA operating instructions models TR12, TC12
19
7. Maintenance, cleaning and calibration
For contact details see chapter 1 “General information” or the back page of the operating instructions.
7.1 Maintenance
The thermometers described here are maintenance-free.
Repairs must only be carried out by the manufacturer.
7.2 Cleaning
CAUTION! Physical injuries and damage to property and the environment
Improper cleaning may lead to physical injuries and damage to property and the environment. Residual media in the dismounted instrument can result in a risk to persons, the environment and equipment.
Carry out the cleaning process as described below.
1. Prior to cleaning, disconnect the electrical connections properly.
2. Use the required protective equipment (depending on the application; the thermometer
itself is basically not dangerous).
3. Clean the instrument with a moist cloth.
This applies in particular to thermometers with a case made of plastic to ensure that any risk of electrostatic charge is avoided. Electrical connections must not come into contact with moisture!
CAUTION! Damage to the instrument
Improper cleaning may lead to damage to the instrument!
Do not use any aggressive cleaning agents.
Do not use any hard or pointed objects for cleaning.
4. Wash or clean the dismounted instrument, in order to protect persons and the
environment from exposure to residual media.
7.3 Calibration, recalibration
It is recommended that the measuring insert is recalibrated at regular intervals (resistance thermometers: approx. 24 months, thermocouples: approx. 12 months). This period can reduce, depending on the particular application. The calibration can be carried out by the manufacturer, as well as on site by qualified skilled personnel with calibration instruments.
7. Maintenance, cleaning and calibration
EN
14064370.02 08/2018 EN/DE/FR/ES
WIKA operating instructions models TR12, TC12
20
8. Dismounting, return and disposal
8.1 Dismounting
WARNING! Physical injuries and damage to property and the environment through residual media
Upon contact with hazardous media (e.g. oxygen, acetylene, flammable or toxic substances), harmful media (e.g. corrosive, toxic, carcinogenic, radioactive), and also with refrigeration plants and compressors, there is a danger of physical injuries and damage to property and the environment.
Before storage of the dismounted instrument (following use) wash or clean it, in order to protect persons and the environment from exposure to residual media.
Use the required protective equipment (depending on the application; the thermometer itself is basically not dangerous).
Observe the information in the material safety data sheet for the corresponding medium.
Only disconnect the thermometer once the system has been depressurised.
WARNING! Risk of burns
During dismounting there is a risk of dangerously hot media escaping.
Let the instrument cool down sufficiently before dismounting it!
8.2 Return
Strictly observe the following when shipping the instrument:
All instruments delivered to WIKA must be free from any kind of hazardous substances (acids, bases, solutions, etc.) and must therefore be cleaned before being returned.
WARNING! Physical injuries and damage to property and the environment through residual media
Residual media in the dismounted instrument can result in a risk to persons, the environment and equipment.
With hazardous substances, include the material safety data sheet for the corresponding medium.
Clean the instrument, see chapter 7.2 “Cleaning”.
When returning the instrument, use the original packaging or a suitable transport packaging.
To avoid damage:
1. Place the instrument, along with shock-absorbent material, in the packaging. Place shock-absorbent material evenly on all sides of the transport packaging.
2. If possible, place a bag containing a desiccant inside the packaging.
3. Label the shipment as carriage of a highly sensitive measuring instrument.
8. Dismounting, return and disposal
EN
14064370.02 08/2018 EN/DE/FR/ES
WIKA operating instructions models TR12, TC12
21
Information on returns can be found under the heading “Service” on our local website.
8.3 Disposal
Incorrect disposal can put the environment at risk. Dispose of instrument components and packaging materials in an environmentally compatible way and in accordance with the country-specific waste disposal regulations.
Do not dispose of with household waste. Ensure a proper disposal in accordance with national regulations.
9. Specifications
9.1 Models TR12-A, TR12-M
Output signal Pt100
Temperature range Measuring range -200 ... +600 °C Measuring element
(measuring current: 0.1 ... 1.0 mA)
1)
Pt100 measuring resistor
Connection method 1 x 2-wire, 1 x 3-wire, 1 x 4-wire, 2 x 2-wire, 2 x 3-wire,
2 x 4-wire
2)
Tolerance value of the measuring element per EN 60751 Class B
Class A Class AA
Wire-wound Thin film
-200 ... +600 °C
-100 ... +450 °C
-50 ... +250 °C
-50 ... +500 °C
-30 ... +300 °C 0 ... 150 °C
Measuring insert (replaceable)
Material Stainless steel 1.4571, 316/316L Diameter Standard: 3 mm
3)
, 6 mm, 8 mm (with sleeve)
Option (on request): 1/8" 3) (3.17 mm), 1/4" (6.35 mm), 3/8" (9.53 mm)
Spring travel approx. 20 mm Response time (in
water, per EN 60751)
t
50
< 10 s t90 < 20 s (measuring insert diameter 6 mm: The thermowell required for operation increases the response time dependent upon the actual parameters for the thermowell and the process.)
Use resistance thermometers with shielded cable, and, if the lines are longer than 30 m or leave the building, ground the shield on at least one end of the lead. For a correct determination of the overall measuring deviation, both sensor and transmitter measuring deviations have to be considered.
1) For detailed specifications for Pt100 sensors, see Technical information IN 00.17 at www.wika.com.
2) Not for 3 mm diameter
3) Not for 2 x 4-wire connection method
8. Dismounting, return and disposal / 9. Specifications
EN
14064370.02 08/2018 EN/DE/FR/ES
WIKA operating instructions models TR12, TC12
22
Neck tube (only model TR12-M)
Material Stainless steel 1.4571, 316/316L Connection thread to the
thermowell
G 1/2 B, G 3/4 B, 1/2 NPT, 3/4 NPT, M14 x 1.5, M18 x 1.5, M20 x 1.5, M27 x 2
Connection thread to the head
M20 x 1.5 with counter nut
M24 x 1.5 with counter nut
1/2 NPT
3/4 NPT
Neck length
min. 150 mm, standard neck length
200 mm
250 mm
other neck lengths on request
Ambient conditions
Ambient and storage temperature
-60
4)
/ -40 ... +80 °C
Ingress protection IP00 per IEC/EN 60529 Vibration resistance
6 g peak-to-peak, wire-wound measuring resistor or thin film (standard)
20 g peak-to-peak, thin-film measuring resistor (option)
50 g peak-to-peak, thin-film measuring resistor (option)
5)
4) Special version on request (only available with selected approvals), other ambient and storage temperature on request
For further specifications see WIKA data sheets TE 60.16, TE 60.17 and the order documentation.
9.2 Model TR12-B
Output signal Pt100
Temperature range Measuring range -200 ... +600 °C Measuring element
(measuring current: 0.1 ... 1.0 mA)
6)
Pt100 measuring resistor
Connection method 1 x 2-wire, 1 x 3-wire, 1 x 4-wire, 2 x 2-wire, 2 x 3-wire,
2 x 4-wire
Tolerance value of the measuring element per EN 60751 Class B
Class A Class AA
Wire-wound Thin film
-200 ... +600 °C
-100 ... +450 °C
-50 ... +250 °C
-50 ... +500 °C
-30 ... +300 °C 0 ... 150 °C
Use resistance thermometers with shielded cable, and, if the lines are longer than 30 m or leave the building, ground the shield on at least one end of the lead. For a correct determination of the overall measuring deviation, both sensor
and transmitter measuring deviations have to be considered.
5) For measuring insert diameter < 8 mm
6) For detailed specifications for Pt100 sensors, see Technical information IN 00.17 at www.wika.com.
9. Specifications
EN
14064370.02 08/2018 EN/DE/FR/ES
WIKA operating instructions models TR12, TC12
23
Output signal 4 ... 20 mA, HART® protocol, FOUNDATION™ Fieldbus and PROFIBUS
®
PA
7)
Transmitter model (selectable versions) T15 T32 T53 TIF50, TIF52 Data sheet TE 15.01 TE 32.04 TE 53.01 TE 62.01 Output
4 ... 20 mA x x x
HART® protocol x x
FOUNDATION™ Fieldbus and PROFIBUS® PA
x
Connection method
1 x 2-wire, 3-wire or 4-wire x x x x
Measuring current < 0.2 mA < 0.3 mA < 0.2 mA < 0.3 mA
Measuring insert (replaceable)
Material Stainless steel 1.4571, 316/316L Diameter Standard: 3 mm
8)
, 6 mm, 8 mm (with sleeve)
Option (on request): 1/8" 8) (3.17 mm), 1/4" (6.35 mm), 3/8" (9.53 mm)
Spring travel approx. 20 mm Response time (in
water, per EN 60751)
t
50
< 10 s t90 < 20 s (measuring insert diameter 6 mm: The thermowell required for operation increases the response time dependent upon the actual parameters for the thermowell and the process.)
Neck tube
Material Stainless steel 1.4571, 316/316L Connection thread to
the thermowell
G 1/2 B, G 3/4 B, 1/2 NPT, 3/4 NPT, M14 x 1.5, M18 x 1.5, M20 x 1.5, M27 x 2
Connection thread to the head
M20 x 1.5 with counter nut
M24 x 1.5 with counter nut
1/2 NPT
3/4 NPT
Neck length
min. 150 mm, standard neck length
200 mm
250 mm
other neck lengths on request
7) Protect the temperature transmitter from temperatures > 85 °C.
8) Not for 2 x 4-wire connection method
9. Specifications
EN
14064370.02 08/2018 EN/DE/FR/ES
WIKA operating instructions models TR12, TC12
24
Ambient conditions
Ambient and storage temperature
-60
9)
/ -40 ... +80 °C
Ingress protection IP66 per IEC/EN 60529
The specified ingress protection only applies with corresponding thermowell, connection head, cable gland and appropriate cable dimensions
Vibration resistance
6 g peak-to-peak, wire-wound measuring resistor or thin film (standard)
20 g peak-to-peak, thin-film measuring resistor (option)
50 g peak-to-peak, thin-film measuring resistor (option)
10)
9) Special version on request (only available with selected approvals), other ambient and storage temperature on request
10) For measuring insert diameter < 8 mm
For further specifications see WIKA data sheet TE 60.17 and the order documentation.
9.3 Models TC12-A, TC12-M
Output signal thermocouple
Recommended max. operating temperature
Type K
Type J
Type E
Type N
1,200 °C 800 °C 800 °C 1,200 °C
Thermocouple per DIN EN 60584-1
11)
Types K, J, E, N
Measuring point
Ungrounded welded (ungrounded)
Grounded
Tolerance value of the measuring element
per EN 60584-1
per ASTM E230 (only for types K and J)
Class 1 and 2 Standard and special
Measuring insert (replaceable)
Material Inconel 600, others on request Diameter Standard: 3 mm, 4.5 mm, 6 mm, 8 mm
Option (on request): 1/8" (3.17 mm), 1/4" (6.35 mm), 3/8" (9.53 mm)
Spring travel approx. 20 mm Response time (in
water, per EN 60751)
t
50
< 5 s t90 < 10 s (measuring insert diameter 6 mm: The thermowell required for operation increases the response time dependent upon the actual parameters for the thermowell and the process.)
11) For detailed specifications for thermocouples, see Technical information IN 00.23 at www.wika.com.
9. Specifications
EN
14064370.02 08/2018 EN/DE/FR/ES
WIKA operating instructions models TR12, TC12
25
Neck tube (only model TC12-M)
Material Stainless steel 1.4571, 316/316L Connection thread to
the thermowell
G 1/2 B, G 3/4 B, 1/2 NPT, 3/4 NPT, M14 x 1.5, M18 x 1.5, M20 x 1.5, M27 x 2
Connection thread to the head
M20 x 1.5 with counter nut
M24 x 1.5 with counter nut
1/2 NPT
3/4 NPT
Neck length
min. 150 mm, standard neck length
200 mm
250 mm
other neck lengths on request
Ambient conditions
Ambient and storage temperature -60
12)
/ -40 ... +80 °C
Ingress protection IP00 per IEC/EN 60529 Vibration resistance 50 g, peak-to-peak
12) Special version on request (only available with selected approvals), other ambient and storage temperature on request
For further specifications see WIKA data sheets TE 65.16, TE 65.17 and the order documentation.
9.4 Model TC12-B
Output signal thermocouple
Recommended max. operating temperature
Type K
Type J
Type E
Type N
1,200 °C 800 °C 800 °C 1,200 °C
Thermocouple per DIN EN 60584-1
13)
Types K, J, E, N
Measuring point
Ungrounded welded (ungrounded)
Grounded
Tolerance value of the measuring element
per EN 60584-1
per ASTM E230 (only for types K and J)
Class 1 and 2 Standard and special
13) For detailed specifications for thermocouples, see Technical information IN 00.23 at www.wika.com.
9. Specifications
EN
14064370.02 08/2018 EN/DE/FR/ES
WIKA operating instructions models TR12, TC12
26
Output signal 4 ... 20 mA, HART® protocol, FOUNDATION™ Fieldbus and PROFIBUS
®
PA
14)
Transmitter model (selectable versions) T16 T32 T53 TIF50, TIF52 Data sheet TE 16.01 TE 32.04 TE 53.01 TE 62.01 Output
4 ... 20 mA x x x
HART® protocol x x
FOUNDATION™ Fieldbus and PROFIBUS® PA
x
Galvanic isolation x x x x
Measuring insert (replaceable)
Material Ni alloy 2.4816 (Inconel 600), others on request Diameter Standard: 3 mm, 4.5 mm, 6 mm, 8 mm
Option (on request): 1/8" (3.17 mm), 1/4" (6.35 mm), 3/8" (9.53 mm)
Spring travel approx. 20 mm Response time (in
water, per EN 60751)
t
50
< 5 s t90 < 10 s (measuring insert diameter 6 mm: The thermowell required for operation increases the response time dependent upon the actual parameters for the thermowell and the process.)
Neck tube
Material Stainless steel 1.4571, 316/316L Connection thread to
the thermowell
G 1/2 B, G 3/4 B, 1/2 NPT, 3/4 NPT, M14 x 1.5, M18 x 1.5, M20 x 1.5, M27 x 2
Connection thread to the head
M20 x 1.5 with counter nut
M24 x 1.5 with counter nut
1/2 NPT
3/4 NPT
Neck length
min. 150 mm, standard neck length
200 mm
250 mm
other neck lengths on request
14) Protect the temperature transmitter from temperatures > 85 °C.
9. Specifications
EN
14064370.02 08/2018 EN/DE/FR/ES
WIKA operating instructions models TR12, TC12
27
Ambient conditions
Ambient and storage temperature
-60
15)
/ -40 ... +80 °C
Ingress protection IP66 per IEC/EN 60529
The specified ingress protection only applies with corresponding thermowell, connection head, cable gland and appropriate cable dimensions
Vibration resistance 50 g, peak-to-peak
15) Special version on request (only available with selected approvals), other ambient and storage temperature on request
For further specifications see WIKA data sheet TE 65.17 and the order documentation.
10. Accessories
The seals can be ordered from WIKA, indicating the WIKA order number and/ or the designation (see table).
WIKA order number
Designation Suitable for
threads
11349981 per DIN 7603 form C 14 x 18 x 2 -CuFA G ¼, M14 x 1.5 11349990 per DIN 7603 form C 18 x 22 x 2 -CuFA G ⅜, M18 x 1.5 11350008 per DIN 7603 form C 21 x 26 x 2 -CuFA G ½, M20 x 1.5 11350016 per DIN 7603 form C 27 x 32 x 2.5 -CuFA G ¾, M27 x 2
Legend: CuFA = Copper, max. 45HB
a
; filled with asbestos-free sealing material
9. Specifications / 10. Accessories
EN
14064370.02 08/2018 EN/DE/FR/ES
WIKA operating instructions models TR12, TC12
28
DE
WIKA Betriebsanleitung Typen TR12, TC12
29
14064370.02 08/2018 EN/DE/FR/ES
Inhalt
Inhalt
1. Allgemeines 30
2. Aufbau und Funktion 30
3. Sicherheit 34
4. Transport, Verpackung und Lagerung 37
5. Inbetriebnahme, Betrieb 38
6. Störungen 43
7. Wartung, Reinigung und Kalibrierung 45
8. Demontage, Rücksendung und Entsorgung 46
9. Technische Daten 47
10. Zubehör 53
DE
WIKA Betriebsanleitung Typen TR12, TC12
30
14064370.02 08/2018 EN/DE/FR/ES
1. Allgemeines
Die in der Betriebsanleitung beschriebenen Thermometer werden nach dem aktuellen Stand der Technik gefertigt.
Diese Betriebsanleitung gibt wichtige Hinweise zum Umgang mit dem Gerät. Voraus­setzung für sicheres Arbeiten ist die Einhaltung aller angegebenen Sicherheitshinweise und Handlungsanweisungen.
Die für den Einsatzbereich des Gerätes geltenden örtlichen Unfallverhütungsvorschrif­ten und allgemeinen Sicherheitsbestimmungen einhalten.
Das Fachpersonal muss die Betriebsanleitung vor Beginn aller Arbeiten sorgfältig durchgelesen und verstanden haben.
Technische Änderungen vorbehalten.
Weitere Informationen:
- Internet-Adresse: www.wika.de / www.wika.com
- zugehöriges Datenblatt: TE 60.16 (TR12-A) TE 60.17 (TR12-B, TR12-M) TE 65.16 (TC12-A) TE 65.17 (TC12-B, TC12-M)
- Anwendungsberater:
Tel.: +49 9372 132-0 Fax: +49 9372 132-406 info@wika.de
Abkürzungen RTD englisch: „Resistance temperature detector“; Widerstandsthermometer
TC englisch: „Thermocouple“; Thermoelement
2. Aufbau und Funktion
2.1 Beschreibung
Die elektrischen Thermometer Typ TR12-B (Widerstandsthermometer) bzw. Typ TC12-B (Thermoelement) bestehen aus einem Modul (TR12-M, TC12-M) welches an ein Gehäuse angebaut ist. Das Modul besteht aus einem federnd gelagertem Messeinsatz (TR12-A, TC12-A) eingebaut in ein Halsrohr. Der Messeinsatz (TR12-A, TC12-A) ist auswechselbar.
Der messaktive Teil des Messeinsatzes ist hergestellt aus einem angeschweißten Röhrchen oder aus mineralisolierter Leitung, optional in Kombination mit keramikisolierten Thermodrähten. Der Sensor ist eingebettet in Keramikpulver, hitzebeständiger Verguss­masse, Zementkitt oder Wärmeleitpaste.
1. Allgemeines / 2. Aufbau und Funktion
Loading...
+ 78 hidden pages