WIKA TI.53 Data sheet

Mechanical Temperature Measurement
Bimetal Thermometer, Industrial Grade - All Stainless Steel Construction Type TI.53 - 5” Dial Size, Back Connected
Datasheet TI.53
Applications
vessels, micro-brewing, boilers and water systems/piping
Heating and air-conditioning technology (HVAC) Temperature measurement in harsh and agressive
environments
Product features
Robust industrial design Back connection without external reset NEMA 4X (IP 66) weather protection
Specications
Size
5” (127mm)
Accuracy
+ 1.0% full scale value per ASME B40.3 Grade A
Ranges
From -100 oF (70 oC) to 1000 oF (540 oC) From -50 oC to 550 oC (as single scale) See table on page 2
Working Range
Steady: full scale value Short time: 110% of full scale value
Over/Under Range Protection
< +500 oF (+260 oC): Temporary up to 50% of full scale > +500 oF (+260 oC): Continuous to 800 oF (+427 oC) Intermittant up to 1000 oF (+538 oC)
Connection
Material: 304 stainless steel Center back mount (CBM) - Type TI.53 1/2” NPT Connection
Stem
Material: 304 stainless steel Diameter: 1/4” (6.35 mm)
Length: 2½” to 24” (63.5 mm to 609.6 mm)
Thermometer TI.53
Dial
White aluminum, dished, with black markings
Case
Material: 304 stainless steel Hermetically sealed Weather protection NEMA 4X (IP 66)
Pointer
Black aluminum
Standard Scales
Single: Fahrenheit or Celsius Dual: Fahrenheit (outer) and Celsius (inner)
Window Gasket
Neoprene Silicone for ranges -100 ranges > +550 oF (+260°C)
Window
Flat instrument glass
Weight
16 oz. (454 g), not including the stem Add 1 oz. (28 g) for every 2” (50 mm) of stem length
Dampening
Inert gel to minimize pointer oscillation
o
F (-70 oC) and
Measuring Element
Bi-metal helix
Datasheet TI.53 · 6/2015
Warranty
Limited one year warranty as stated in WIKA’s Terms & Conditions of Sale
Page 1 of 2
R
Optional Extras
ThermowellsSpecial scales and dial markingsAcrylic and safety glass windowsCalibration certication traceable to NIST
Dimensions
Standard versions
STANDARD RANGES
Fahrenheit Single Scale
Dual Scale F & C F Outer, C Inner
Celsius Single Scale
-100/150 F -100/150 F & -70/70 C -50/50 C
-40/120 F 40/120 F & -40/50 C -20/120 C
0/140 F 0/140 F & -20/60 C 0/50 C
0/200 F 0/200 F & -15/90 C 0/100 C
0/250 F 0/250 F & -20/120 C 0/150 C
20/240 F 20/240 F & -5/115 C 0/200 C
25/125 F 25/125 F & -5/50 C
1
0/250 C
50/300 F 50/300 F & 10/150 C 0/300 C
50/400 F 50/400 F & 10/200 C 0/450 C
50/550 F 50/500 F & 10/260 C 100/550 C
150/750 F 150/750 F & 65/400 C
200/1000 F1200/1000 F & 100/540 C
1
Not recommended for continous service over 800°F (425°C)
1
1
1
1
WIKA Type DIAL SIZE A B C S (Stem Length)
53 5” (127 mm) 5-1/4” (8133.4 mm) 15/16” (23.8 mm) 5/16” (7.9 mm) As Specied
Note: Thermowells for temperature instruments are recommended for all process systems where pressure, velocity, or viscous, abrasive and corrosive materials are present individually or in combination. A properly selected thermowell protects the temperature instrument from possible damage resulting from these process variables. Furthermore, a thermowell permits removal of the
temperature instrument for replacement, repair or testing without eecting the process media or the system.
Ordering information
State computer part number (if available) /type number/size/range/connection size and locations/options required. WIKA reserves the right to make changes without prior notice.
Datasheet TI.53 · 6/2015Page 2 of 2
R
WIKA Instrument Corporation
1000 Wiegand Boulevard Lawrenceville, GA 30043 1-888-WIKA-USA /770-513-8200 (in GA) Fax 770-338-5118 info@wika.com www.wika.com
Loading...