3.2 "FUNC-CYCL": Automatic recording with selectable16
Cycle-Time
4.General information about temperature sensors18
4.1 Available temperature sensors18
GB
5.The serial interface18
6.Error and system messages19
7.Calibration services20
8.Specification20
9.Accessories21
Information
This symbol provides you with information, notes an tips.
Warning!
This Symbol warns you against actions that can cause
damage to persons or to the instrument.
2146872 09/2004 GB/D
WIKA Betriebsanleitung Hand-Held Thermometer
WIKA Operating Instructions Hand-Held Thermometer
3
Page 4
Hand-Held Thermometer
CTH 6200
GB
1. General
In the following chapters detailed information on the hand-held
thermometer CTH 6200 and its proper use can be found.
Should you require further information, or should there be problems which are not dealt with in detail in the operating instructions,
please contact the address listed on the last page.
The factory calibration of the instruments are according to respective international guidelines.
The warranty period for the hand-held thermometer CTH 6200 is
24 months according to the general terms of supply of ZVEI.
All guarantee claims lapse if the appliance is put to improper use
or if the operating instructions are not observed or if an attempt is
made to open the appliance.
We also point out that the contents of these operating instructions
does not form part of an earlier or existing agreement, assurance
or legal relationship or is meant to change these. All obligations of
WIKA Alexander Wiegand GmbH & Co. KG result from the respective sales contract and the general business terms of WIKA
Alexander Wiegand GmbH & Co. KG.
WIKA is a registered trade mark of WIKA Alexander Wiegand
GmbH & Co. KG. Names of companies or products mentioned in
this handbook are registered trade marks of these manufacturers.
We reserve the right to effect reasonable changes on the basis of
technical improvements.
Any reproduction of this manual or parts thereof by any means is
prohibited.
This device has been designed and tested in accordance with
the safety regulations for electronic devices.
However, its trouble-free operation and reliability cannot be
guaranteed unless the standard safety measures and special
safety advises given in this manual will be adhered to when
using the device.
1. Trouble-free operation and reliability of the device can only be
guaranteed if the device is not Subjected to any other climatic
conditions than those stated under "Specification".
2. Device and sensors have to be handled with care (don't throw,
hit, etc.). Protect plugs and sockets from soiling.
3. If the device is transported from a cold to a warm environment
condensation may cause in a failure of the function. In such a
case make sure the device temperature has adjusted to the
ambient temperature before trying a new start-up.
4. If device is to be connected to other devices (e.g. via serial
interface) the circuitry has to be designed most carefully. Internal connection in third party devices may result in not-permissible voltages impairing or destroying the device or another
device connected.
If device is operated with a defective mains power supply (e.g.
short circuit from mains voltage to output voltage) this may
result in hazardous voltages at the device
(e.g. at sensor socket or interface).
GB
5. If there is a risk whatsoever involved in running it, the device
has to be switched off immediately and to be marked accordingly to avoid re-starting.
Operator safety may be a risk if:
■there is visible damage to the device
■the device is not working as specified
■the device has been stored under unsuitable conditions for
a longer period of time.
In case of doubt, please return device to manufacturer for repair
or maintenance.
6. Repairs should only be carried out by the manufacturer. All
other repairs or modifications of the transmitter are unauthorized.
7. Any operation not described in the following instructions or
2146872 09/2004 GB/D
WIKA Betriebsanleitung Hand-Held Thermometer
WIKA Operating Instructions Hand-Held Thermometer
outside the specifications must not be carried out.
5
Page 6
Hand-Held Thermometer
CTH 6200
GB
1.2 Operation and maintenance advice
■ Operation with battery or accu
If ∆ and ´bAt´ are shown in the secondary display the battery
has been used up and needs to be replaced, or resp. the accu
is empty and has to be recharged. The device will, however,
operate correctly for a certain amount of time. If ´bAt´ is shown
in the upper display the voltage is too low to operate the device;
the battery or accu has been completely used up.
We recommend to take out battery/accu if device is not used for a
longer period of time!
■ Mains operation with power supply
When using a power supply please note that operating voltage
has to be 10.5 to 12 V DC. Do not apply overvoltage! Cheap
12 V-power supplies often have excessive no-load voltage.
We, therefore, recommend using regulated voltage power
supplies. Trouble-free operation is guaranteed by our power
supply unit.
Prior to connecting the power supply to the mains make sure
that the operating voltage stated at the power supply is
identical to the mains voltage.
■ Connecting/Changing Sensors
Switch off device before changing the sensor. Connect sensor
before switching on the device. In such case take the plug not
at the casing but at the buckling protection at the end of the
plug. When connecting the probe the plug will slide in smoothly
if plug is entered correctly. To disconnect sensor plug do not
pull at the cable but at the plug.
Probes may be extremely hot or cold. Precautions must be
taken to prevent personal injury or damage to objects.
6
WIKA Betriebsanleitung Hand-Held Thermometer
WIKA Operating Instructions Hand-Held Thermometer
2146872 09/2004 GB/D
Page 7
Hand-Held Thermometer
CTH 6200
■ Maintenance advice
The digital instrument and sensors are completely of solid state
principle, not containing any moving parts. Dismantling of ensures invalidates warranty.
If cleaning is necessary, use a cloth damp by a mild soap cleaner,
avoid any solvent or caustic abrasive agent.
As for measurement devices usual, the accuracy of the device
should be tested in regularly time periods (approx. 1 time per year
- see chapter 7).
1.3 Connections
At the top of the device:
are located the socket CH1 for connecting the temperature sensor
of the CPT-series (see chapter 4.1) and the socket for connecting
the interface converter (see chapter 5).
Connection for interface
converter
GB
CH1
At the left side of the device:
is located the mains adapter socket to connect the available
power supply unit (see valid price list of CTH 6200).
2146872 09/2004 GB/D
WIKA Betriebsanleitung Hand-Held Thermometer
WIKA Operating Instructions Hand-Held Thermometer
7
Page 8
Hand-Held Thermometer
CTH 6200
GB
1.4 Display
1
2
6
3
7
8
4
5
1. Offset: indicates that zero point offset (offset) is activated.
2. Corr: indicates that correction factor is activated.
3. : indicates a low battery and other warnings.
4. Logg: Shown if logger function is selected, flashes if cyclic
logger is running.
5. AL: Flashes if alarm is present.
6. Main display: Currently measured temperature.
7. Min/Max/Hold: shows if a MIN, MAX or Hold value is displayed
in the secondary display.
8. Secondary display: shows MIN, MAX or Hold value acc. to
selected function.
1.5 Basic operation
When switching on the device and the logger function is not off
the time of the integrated clock will shortly be displayed. If a zero
point adjustment was carried out the display shows shortly "nuLL
Corr".
After changing the battery the clock-setting menu is activated
automatically (´CLOC´). Check the clock and adjust, if necessary
(see chapter 2).
8
WIKA Betriebsanleitung Hand-Held Thermometer
WIKA Operating Instructions Hand-Held Thermometer
2146872 09/2004 GB/D
Page 9
Hand-Held Thermometer
CTH 6200
4
1
2
3
6
5
1. On/Off key
2. Set/Menu: configuration will activated Menu, press for 2 sec
3. Tara:When taking measurements: no function
During configuration: Selection of menu-parameter
4. Max-Function: Pressing ´Max´ shows the maximum of the
measured values. Pressing it again hides it. To clear
the max memory press key ´Max´ for 1 seconds.
5. Min-Function: Pressing ´Min´ shows the minimum of the
measured values. Pressing it again hides it. To clear
the min memory press key ´Min´ for 1 seconds.
GB
6. Hold-Function: By pressing ´Store/Quit´ the last measuring
value will be held in the secondary display. Pressing
it again hides it. (only when main menu logger = ´off´
is chosen).
6. Logger-Functions: Activation with key ´Store/Quit´, if in the
main menu a logger function was chosen before.
(see Chapter 3).
2146872 09/2004 GB/D
WIKA Betriebsanleitung Hand-Held Thermometer
WIKA Operating Instructions Hand-Held Thermometer
9
Page 10
Hand-Held Thermometer
CTH 6200
GB
Main Menu
Read*
Logg
Set
Conf
Set
AL
Set**
Logg
(2 sec)
Stored data*
Call of Store-Func. data with respective
time (see chapter 3)
Configuration
Unit ** / resolution**
zero point** & scale correction**
Power-off / interface-off
Alarm
Min/Max alarm visual with/without
sound
Logger**
switch from Hold to Logger function:
individual values (Stor) or cyclic (CYCL)
Set
Cloc
(ring - shift)
* appears only if data (Store-Func.) is in the logger.
** appears only if no data is in the logger (see chapter 3).
10
WIKA Betriebsanleitung Hand-Held Thermometer
WIKA Operating Instructions Hand-Held Thermometer
System-clock
time / day & month / year
set parameter next store
parameter & quit
2146872 09/2004 GB/D
Page 11
Hand-Held Thermometer
CTH 6200
MenuParam. ValuesMeaning
„Menu“ or
SEtSet Configuration: Generic Settings
ConFUnit°C:All temperatures values are in degrees Celsius*
°F:All temperatures values are in degrees Fahrenheit
rES0.1 °:Resolution 0.1 °C*
0.01°
Auto:
Resolution 0.01 °C*
Resolution is selected automatically
LinE.751Measuring / sensor curve according to EN 60751*
USEr
In preparation
OFFS -2.50 °C ... 2.50 °CThe zero point of the measurement will be displaced by this value to*
resp. -4.50 °F ... 4.50 °F compensate for deviations in the probe or in the measuring device.
oFF:Zero displacement inactive (=0.0 °C)
SCAL -2.000 ... 2.000The scale of the measuring will be changed by this factor [in %] to*
compensate deviations of temperature probe or measuring device.
oFF:Scale correction factor inactive ( = 0.000)
P.oFF 1 ... 120Auto Power-Off delay in minutes.
Device will be automatically switched off as soon as this time has
elapsed if no key is pressed / no interface communication takes place.
oFFAuto Power-off function inactive (continuous operation)
OutoFF:Function of the output: No output function, lowest power consumption
SEr:Output is serial interface
dAC:In preparation
Adr.01, 11 .. 91In preparation
SEtSet Alarm: Settings of alarm function
AL.AL.onAlarm Sensor 1
no. SoAlarm Sensor 1 on, without horn-sound
oFFNo alarm function
AL.Lo min range ... AL.HiMin alarm rail (not when AL.oFF)
AL.Hi AL.Lo ... max rangeMax alarm rail (not when AL.oFF)
on, with horn-sound
SEtSet Logger: Configuration of logger function*
LoGGFuncCYCLCyclic: logger function ,cyclic logger`*
StorStore: logger function ,individual value logger`*
oFFNo logger function*
CYCL 0:01 ... 60:00Cycle time of cyclic logger [minutes:seconds]*
SEtSet Clock: Setting of real time clock
CLOCCLOC HH:MMClock: setting ot time hours:minutes
dAtETT.MMDate: setting of time day.month
YEAr YYYYYear: setting of time year
GB
2146872 09/2004 GB/D
WIKA Betriebsanleitung Hand-Held Thermometer
WIKA Operating Instructions Hand-Held Thermometer
11
Page 12
Hand-Held Thermometer
CTH 6200
GB
If keys ´Set´ and ´Store´ are pressed simultaneously for more the 2
seconds, the settings are reset of the settings ex-works.
If the logger memory contains data already, the menus/parameters
marked with (*) can not be invoked! If these should be altered the
logger memory has to be cleared before! (see chapter 3)
2. Configuration
To change device settings, press key: ´Set/Menu´ for 2 seconds.
This will call the configuration menu (main display: "SEt").
Pressing key: ´Set/Menu´ changes between the menus, pressing
key: ´Tara´ jumps to the referring parameters, which can be
selected with key: ´Tara´.
The parameters can be changed with key: ´Min´ or ´Max´. Pressing
key: ´Set/Menu´ again jumps back to the main configuration menu
and saves the settings.
Key: ´Store/Quit´ finishes the configuration and returns to standard measuring operation.
2.1 (Unit) Temperature unit
The measured temperature can be displayed in °C or °F.
2.2 (rES) Display resolution
Standard setting: ´Auto´ ,i.e. the device automatically switches over
to the optimum resolution between 0.1 ° and 0.01 °.
If temperatures to be measured are near the switching threshold, a
fixed resolution may be better, e.g. for easy recording.
In such a case please select the optimum resolution manually.
2.3 (Lin E.751) Sensor curve
This function used the standard conversion of resistance to
temperature following EN60751 (Lin E.751).
12
WIKA Betriebsanleitung Hand-Held Thermometer
WIKA Operating Instructions Hand-Held Thermometer
2146872 09/2004 GB/D
Page 13
Hand-Held Thermometer
CTH 6200
The sensor curve following EN60751 uses the international
temperature scale ITS90 and following formulas:
A zero point correction can be carried out for the measured
temperature:
temperature displayed = temperature measured - offset
Standard setting: ´off´ = 0.0 °, i.e. no zero point correction will be
carried out. The zero point correction is mainly used to compensate for sensor deviations. Unless ´off´ is set, this value will be
displayed shortly after the device is switched on;
during operation it will be identified by means of the offset arrow in
the display.
2.5 (Scal) Scale correction
GB
The scale of the measuring can be influenced by this setting
(factor is in %).
displayed temperature [°C] = measured temperature [°C] * (1 + Scal/100)
or respectively
displayed temperature [°F] = (measured temperature [°C] - 32
°F) * (1 + Scal/100) + 32 °F
Standard setting: ´off´ =0.000, i.e. temperature is nor corrected.
Together with the zero displacement (see above) this factor is
mainly used to compensate for sensor deviations.
Unless the factor is set to ´off´, the Corr arrow in the display shows
an active scale correction.
2146872 09/2004 GB/D
WIKA Betriebsanleitung Hand-Held Thermometer
WIKA Operating Instructions Hand-Held Thermometer
13
Page 14
Hand-Held Thermometer
CTH 6200
GB
2.6 (P.oFF) Power off function
If there won't be pressed any key and no interface communication
takes place for the time of the power off time setting (P.off), the
device will be switched off automatically to save battery power.
If P.oFF = oFF then the automatic switch off is deactivated.
2.7 (Out) Serial interface
The output can be used as serial interface. If none needed, we
suggest to switch the output off, because battery life then is
extended.
2.8 (Adr.) Interface base address
In preparation for use with multiplexer.
2.9 (AL.) Alarm
There are three possible settings: Alarm off (AL. oFF), on with horn
sound (AL. on), on without horn sound (AL. no.So).
Following conditions will display an alarm, when the function is
activated (on or no.So):
■ Value is below lower (AL. Lo) or above upper alarm rail (AL.Hi).
■ Sensor error (Sens Erro)
■ Low battery (bAt)
■ Fe 7: System error (always with sound)
In case of an alarm, the alarm arrow flashes, and when polling the
interface the prio-flag is set in the returned interface message.
2.10 (CLOC) Real time clock
The real time clock is used for the logger function: Recorded
values are also containing the point of time, when they were
measured. Please check the settings when necessary.
If the battery was replaced the referring menu ´CLOC´ will automatically be started.
14
WIKA Betriebsanleitung Hand-Held Thermometer
WIKA Operating Instructions Hand-Held Thermometer
2146872 09/2004 GB/D
Page 15
Hand-Held Thermometer
CTH 6200
3. Operation of logger
The device supports two different logger functions:
"Func-Stor": each time when the key: ´Store/Quit´ is pressed a
measurement will be recorded.
"Func-CYCL": measurements will automatically be recorded each
interval, which was set in the logger menu ´CYCL´
until the logger will be stopped or the logger memory
is full. The recording is started by pressing key:
´Store/Quit´ for 2 seconds.
The logger records 1 measurement result each time.
For the evaluation of the data the WIKA Data logger evaluation
software GSOFT (V 1.7 or higher) has to be used. The software
also allows easy configuration and starting of the logger.
When the logger is activated (Func Stor or Func CYCL) the hold
function is no more available, the key: ´Store/Quit´ is solely used
for the operation of the logger functions.
3.1 Func-Stor": Storing single measurements
Each time when key: ´Store/Quit´ is pressed a measurement and
its time stamp will be recorded.
The recorded data can be viewed either in the display (when
calling the configuration an additional menu "REAd LoGG" is
displayed, see below) or by means of the interface and a PC with
GSOFT-software.
GB
Max. number of measurings:99
A measuring max. contains:
■ Current measuring value at the time of recording
■ Min.-Peak, Max.-Peak since last recording
■ Time and date of the recording
After each recording "St. XX" will be displayed for a short time. XX
represents the number of the recording.
2146872 09/2004 GB/D
WIKA Betriebsanleitung Hand-Held Thermometer
WIKA Operating Instructions Hand-Held Thermometer
15
Page 16
Hand-Held Thermometer
CTH 6200
GB
If logger memory contains recordings already:
When key: ´Store/Quit´ is pressed for 2 seconds, the choice for
clearing the logger memory will be displayed:
clearclear nothing
all recordings(cancel menu)
clear the last
recording
The selection can be made by key: ´Min´ and key: ´Max´. Key:
´Store/Quit´ enters the choice.
If the logger memory is full, the display will show:
Viewing Recorded Measurings
Within the "LoGG Stor" function the measurings can be viewed
directly in the display not only by means of a computer (like at
"Func CYCL"): press 2 seconds ´Set/Menu´. The first menu
displayed now is "rEAd LoGG" (read logger data). After pressing
key: ´Tara´ the measurement recorded last will be displayed,
changing between the different values referring to the measurement also is done by pressing key: ´Tara´.
Changing the measurement is done by pressing the key: ´Min´ or
key: ´Max´.
3.2 Func-CYCL": Automatic recording with selectable cycle-time
The Logger-Cycle-Time is setable (p.r.t. Configuration). For
example "CYCL" = 60: A measuring is recorded after each 60
seconds.
Max. number of measurings: 16384
Cycle time:0:01 ... 30:00 (minutes: seconds, min 1 s,
max 1 h), selectable in the configuration
A measuring contains:
■ Current measuring value at time of recording
16
WIKA Betriebsanleitung Hand-Held Thermometer
WIKA Operating Instructions Hand-Held Thermometer
2146872 09/2004 GB/D
Page 17
Hand-Held Thermometer
CTH 6200
Starting a recording:
By pressing ´Store/Quit´ for 2 seconds the recording will be
initiated. After that the display shows “St.XXXX” for a short time
whenever a measuring is recorded. XXXX is the number of the
measuring 1... 16384.
If the logger memory is full, the display will show:
The recording automatically will be stopped.
Stopping the recording manually:
By pressing key: ´Store/Quit´ the recording can be stopped
manually. Then the following choice appears:
Stop theStop no the
recordingrecording
The selection can be made by key: ´Max´ and key: ´Min´.´Store/Quit´, enters the choice.
If you try to switch off the instrument in the cyclic recording
operation. You will be asked once again if the recording is to be
stopped. The device can only be switched off after the recording
has been stopped!
The Auto-Power-Off-function is deactivated during recording!
GB
Clear Recordings:
When key: ´Store/Quit´ is pressed for 2 seconds, the choice for
clearing the logger memory will be displayed:
clear allclear nothing
recordings(cancel menu)
The selection can be made by key: ´Max´ and key: ´Min´ key:
´Store/Quit´ enters the choice.
2146872 09/2004 GB/D
WIKA Betriebsanleitung Hand-Held Thermometer
WIKA Operating Instructions Hand-Held Thermometer
17
Page 18
Hand-Held Thermometer
CTH 6200
GB
4. General information about temperature sensors
■ Heat loss caused by probe construction:
Especially when measuring temperatures which deviate very
much from the ambient temperature, measuring errors often
occur if the heat loss caused by the probe is not considered.
When measuring fluids therefore the probe should be emerged
sufficiently deep and be stirred continuously. When measuring
gases the probe should also emerge as deep as possible in the
gas to be measured (e.g. when measuring in channel/pipes) and
the gas should flow around the probe at sufficient flow.
The tube length should be selected long enough, that temperature
keeps low at the handle.
■
S
elf Heating
The measuring current of the instrument is just 0.3 mA. Because of this comparably low current practically no self heating
effect has to be considered, even at air with low movement the
self heating is <= 0.01 °C.
■ Cooling by Evaporation
When measuring air temperature the probe has to be dry.
Otherwise the cold due to the evaporation causes too low
measurings.
4.1 Available sensors
The device is designed to be connect
ed to all sensors of the
CTH6200 series (see valid price list CTH6200).
5. The serial interface
By means of the serial interface and a suitable electrically isolated
interface adapter the device can be connected to a computer for
data transfer (USB-adapter on request), to visualise the data with
the WIKA Datalogger evaluation software GSOFT (see valid price
t CTH6200). To avoid transmission errors, there are several
lis
security checks implemented e.g. CRC.
18
WIKA Betriebsanleitung Hand-Held Thermometer
WIKA Operating Instructions Hand-Held Thermometer
2146872 09/2004 GB/D
Page 19
Hand-Held Thermometer
CTH 6200
6. Fault and system messages
DisplayMeaningRemedy
Replace battery
Check power supply, replace it when
necessary
Replace battery
Check power supply, replace it when
necessary
Replace battery
Netzgerät überprüfen / austauschen
Disconnect battery or power supply,
wait some time, re-connect
Return to manufacturer for repair
Connect sensor to socket
Return to manufacturer for repair
Is the value exceeding the measuring
range specified?
-> measuring value to high!
Check probe
Return to manufacturer for repair
Check: is the value below the
measuring range specified?
-> measuring value to low!
Check probe
Return to manufacturer for repair
-> Set resolution to 0.1 ° or Auto
-> Set resolution to 0.1 ° or Auto
Return to manufacturer for repair
Do not leave the specified range
No display
or strange
symbols,
device is not
responding
to input
----
Err.1
Err.2
Err.3
Err.4
Err.7
Low battery voltage, device will coninue
to work for a short time
If mains operation:
wrong voltage
Battery empty
Mains operation without battery:
wrong voltage
Battery empty
Mains operation without battery:
wrong voltage
System error
Device defective
Sensorerror: son sensor connected
Sesnor / cable or device defective
Value exceeding measuring range
Wrong probe connected
Sensor / cable or device defective
Value below display range
Wrong probe connected
Sensor / cable or device defective
Value exceeding display range
Value below display range
System error
Out of range
GB
2146872 09/2004 GB/D
WIKA Betriebsanleitung Hand-Held Thermometer
WIKA Operating Instructions Hand-Held Thermometer
19
Page 20
Hand-Held Thermometer
CTH 6200
GB
7. Calibration services
DKD-certificates - other certificates:
If device should be certificated for its accuracy, it is the best
solution to return it with the referring sensors to the manufacturer.
Only the manufacturer is capable to do efficient recalibration if
necessary to get results of highest accuracy!
8. Specifications
Characteristics of the complete measuring chain (digital instrument and probe):
Input1
Sensor typePt 100 (4-wire)
Measuring range ex stock -50 ... +250 °C
Total measurement
uncertainty
Calibration *Calibration report 3.1.B according to DIN EN 10204
Display2 x 4 1/2-digit LCD display for actual value and additional information
Resolution0.01 K up to 200 °C; than 0.1 K
Units°C, °F
CTH 6200 Digital Instrument specific details:
Measuring range-99.99 ... +99.99 °C or -199.9 ... +650.0 °C respectively
Functions via key pressMin-, Max-memory, Hold, Logger (Start/Stopp)
Functions via menuMin-, Max-alarm (acoustic/visual), autom. Power-Off-Function
Data logger- individual value logger: up to 99 recordings incl. time via keypress
Interface (serial)RS 232 interface via interface converter (USB adapter on request)
Power supply9 V-zinc-carbon battery (incl. in scope of delivery),
Power consumptionapprox. 1.25 mA
Nominal temperature25 °C
Temperature drift≤ 0.002 °C/K
Permissible ambient-10 ... +50 °C
temperature
Air humidity0 ... 95 % relative humidity without moisture condensation
Storage temperature-25 ... +70 °C
Housingimpact-resistant ABS, membrane keyboard, transparent panel
Weightapprox. 160 g
CTP 62x0 Temperature sensor specific details (Pt 100):
Probes availableImmersion and penetration probe
Electrical connection4-pole Mini-DIN plug with PVC cable (approx. 1.0 m)
Temperature range
- Immersion probe
CTP 6210-50 ... +250 °C
- Penetration probe
CTP 6290-50 ... +250 °C
Material wetted partsImmersion and penetration probe: stainless steel
DimensionsSensor length: 300 mm; hand grip: 135 mm
1) The total measurement uncertainty is stated for the whole measuring chain after adjustment and
calibration.
*) Calibration at 23 °C ± 2 °C ambient temperature.
{} Items in curved brackets are optional extras for additional price.
1)
1)
0.2 K
(On request a higher accuracy can be reached by an adjustment at a
certain temperature range.)
(dependent on probe)
[1 ... 120 min], Ze ro point / span correction, Logger settings
- cyclic logger: automatic recording up to 16,200 values incl. time,
with cyclic time selectable from 1 ... 3600 seconds
{alternative: 9 V rechargeable battery or mains supply}
Sensor diameter: D = 3 mm
20
WIKA Betriebsanleitung Hand-Held Thermometer
WIKA Operating Instructions Hand-Held Thermometer
2146872 09/2004 GB/D
Page 21
Hand-Held Thermometer
CTH 6200
9. Accessories
Accessories like.
Datalogger-evaluation software GSoft, power supply unit, accu
charger, accus, etx. see valid pricelsit CTH 6200.
GB
2146872 09/2004 GB/D
WIKA Betriebsanleitung Hand-Held Thermometer
WIKA Operating Instructions Hand-Held Thermometer
21
Page 22
Hand-Held Thermometer
CTH 6200
GB
22
WIKA Betriebsanleitung Hand-Held Thermometer
WIKA Operating Instructions Hand-Held Thermometer
2146872 09/2004 GB/D
Page 23
Hand-Held Thermometer
CTH 6200
Inhalt
Inhalt
1. Allgemeines24
1.1 Sicherheitshinweise25
1.2 Betriebs- und Wartungshinweise26
1.3 Anschlüsse27
1.4 Anzeige28
1.5 Bedienung28
2.Konfiguration des Gerätes32
2.1 (UNIT) Anzeigeeinheit32
2.2 (rES) Anzeigeauflösung32
2.3 (Lin E.751) Ausgleichs-Kennlinie33
2.4 (OFFs) Nullpunktkorrektur / Offset33
2.5 (Scal) Steigungskorrektur33
2.6 (P.oFF) Abschaltfunktion34
2.7 (Out) Serielle Schnittstelle34
2.8 (ADR.) Schnittstellen-Adresse34
2.9 (AL.) Alarm34
2.10 (CLOC) Echtzeituhr34
3.Bedienung der Loggerfunktion35
3.1 "FUNC-STOR": Einzelwert speichern35
3.2 "FUNC-CYCL": Autom. Aufzeichnung mit Einstellbarer36
Logger-Zykluszeit
4.Allgemeines zu Temperaturfühlern38
4.1 Verfügbare Temperaturfühler38
D
2143872 09/2004 GB/D
5.Die serielle Schnittstelle38
6.Fehler- und Systemmeldungen39
7.Hinweis zum Kalibrierservice40
8.Technische Daten40
9.Zubehör41
Information
Dieses Zeichen gibt Ihnen Informationen, Hinweise oder Tipps.
Warnung!
Dieses Symbol warnt Sie vor Handlungen, die Schäden an
Personen oder am Gerät verursachen können.
23WIKA Betriebsanleitung Hand-Held Thermometer
Page 24
Hand-Held Thermometer
CTH 6200
1. Allgemeines
D
In den folgenden Kapiteln erhalten Sie nähere Informationen zum
Hand-Held Thermometer CTH 6200 und seinen ordnungsgemäßen
Einsatz. Sollten Sie weitere Informationen wünschen, oder treten
besondere Probleme auf, die in der Betriebsanleitung nicht
ausführlich behandelt werden, erhalten Sie Auskunft unter den auf
der letzten Seite aufgelisteten Adresse.
Bei der Werkskalibrierung der Instrumente wurde sich an entsprechende internationalen Normen orientiert.
Die Gewährleistungszeit für das Hand-Held Thermometer CTH 6200
beträgt 24 Monate nach den Allgemeinen Lieferbedingungen des
ZVEI. Sämtliche Gewährleistungsansprüche verfallen, bei unsachgemäßer Handhabung bzw. bei Nichtbeachtung der Betriebsleitungen oder bei dem Versuch das Gerät zu öffnen.
Außerdem weisen wir darauf hin, dass der Inhalt dieser Betriebsanleitung nicht Teil einer früheren oder bestehenden Vereinbarung,
Zusage oder Rechtsverhältnisses ist oder diese abändern soll.
Sämtliche Verpflichtungen der WIKA Alexander Wiegand GmbH &
Co. KG ergeben sich aus dem jeweiligen Kaufvertrag und den
Allgemeinen Geschäftsbedingungen der WIKA Alexander Wiegand
GmbH & Co. KG.
WIKA ist ein eingetragenes Warenzeichen der WIKA Alexander
Wiegand GmbH & Co. KG.
Firmen- oder Produktnamen, die in diesem Handbuch erwähnt
werden, sind eingetragene Warenzeichen dieser Hersteller.
Zumutbare Änderungen aufgrund technischer Verbesserungen
behalten wir uns vor.
Eine Vervielfältigung dieses Handbuches oder Teilen davon ist
untersagt.
Dieses Gerät ist gemäß den Sicherheitsbestimmungen für
elektronische Messgeräte gebaut und geprüft. Die Funktion
und Betriebssicherheit des Gerätes kann nur dann gewährleistet werden, wenn die nationalen Sicherheits- und Unfallverhütungsvorschriften sowie die Sicherheitshinweise der
Bedienungsanleitung beachtet werden.
1. Die einwandfreie Funktion und Betriebssicherheit des Gerätes
kann nur unter den klimatischen Verhältnissen, die im Kapitel
"Technische Daten" spezifiziert sind, eingehalten werden.
2. Gerät und Sensor müssen pfleglich behandelt werden (nicht
werfen, aufschlagen, etc.). Stecker und Buchsen sind vor Verschmutzung zu schützen.
3. Wird das Gerät von einer kalten in eine warme Umgebung
transportiert, so kann durch Kondensatbildung eine Störung der
Gerätefunktion eintreten. In diesem Fall muss die Angleichung
der Gerätetemperatur an die Raumtemperatur vor einer erneuten Inbetriebnahme abgewartet werden.
4. Konzipieren Sie die Beschaltung beim Anschluss an andere
Geräte besonders sorgfältig. Unter Umständen können interne
Verbindungen in Fremdgeräten (z.B. Verbindung GND mit Erde)
zu nicht erlaubten Spannungspotentialen führen, die das Gerät
selbst oder ein Angeschlossenes Gerät in seiner Funktion beeinträchtigen oder sogar zerstören können.
Bei Betrieb mit einem defekten Netzgerät (z.B. Kurzschluss
von Netzspannung zur Ausgangsspannung) können am Gerät
lebensgefährliche Spannungen auftreten!
D
2146872 09/2004 GB/D
5. Wenn anzunehmen ist, dass das Gerät nicht mehr gefahrlos betrieben werden kann, so ist es außer Betrieb zu setzen und vor
einer Wiederinbetriebnahme durch Kennzeichnung zu sichern.
Die Sicherheit des Benutzers kann durch das Gerät beeinträchtigt sein, wenn es zum Beispiel:
■sichtbare Schäden aufweist.
■nicht mehr wie vorgeschrieben arbeitet.
■längere Zeit unter ungeeigneten Bedingungen gelagert wurde.
In Zweifelsfällen das Gerät grundsätzlich an den Hersteller zur
Reparatur bzw. Wartung einschicken.
6. Reparaturen dürfen nur vom Hersteller durchgeführt werden.
Eingriffe und Änderungen am Gerät sind unzulässig.
7. Ein anderer Betrieb als der in der folgenden Anleitung beschriebene oder außerhalb der Spezifikationen, ist bestimmungswidrig und muss deshalb ausgeschlossen werden.
25WIKA Betriebsanleitung Hand-Held Thermometer
Page 26
Hand-Held Thermometer
CTH 6200
1.2 Betriebs- und Wartungshinweise
D
■ Batteriebetrieb
Wird ∆ und in der unteren Anzeige "bAt" angezeigt, so ist die
Batterie verbraucht und muss erneuert werden bzw. der Akku
leer und muss mit dem dafür vorgesehen Akkuladegerät (siehe
aktuelle Preisliste CTH 6200) geladen werden. Die Gerätefunktion ist jedoch noch für eine gewisse Zeit gewährleistet. Wird in
der oberen Anzeige "bAt" angezeigt, so ist die Batterie ist ganz
verbraucht. Batterie und Akku sind nur sachgemäß zu benutzen und ordnungsgemäß den aktuellen, nationalen Vorschriften
zu entsorgen. Bei Lagerung des Gerätes über 50 °C muss die
Batterie entnommen werden.
Wird das Gerät längere Zeit nicht benutzt, sollte die Batterie/ der
Akku herausgenommen werden.
■ Netzgerätebetrieb
Beim Anschluss eines Netzgerätes muss dessen Spannung
zwischen 10,5 und 12 V DC liegen. Keine Überspannungen
anlegen! Einfache Netzgeräte können eine zu hohe Leerlaufspannung haben, dies kann zu einer Fehlfunktion bzw. Zerstörung des Gerätes führen! Wir empfehlen daher unser Netzgerät
(siehe aktuelle Preisliste CTH 6200) zu verwenden.
Vor dem Verbinden des Netzgerätes mit dem Stromversorgungsnetz ist sicherzustellen, dass die am Netzgerät angegebene Betriebsspannung mit der Netzspannung übereinstimmt.
■ Sensoren anstecken / wechseln
Zum Sensorwechsel Gerät ausschalten. Sensor vor dem Einschalten des Gerätes anstecken. Stecker nicht verkantet anstecken. Bei richtig angesetzten Stecker kann dieser ohne
größeren Kraftaufwand eingesteckt werden. Beim Abstecken
des Sensors nicht am Kabel ziehen, sondern nur an der
Steckerhülse.
Die Temperaturfühler können extrem heiß oder kalt sein.
Entsprechende Vorkehrungen sind zu ergreifen, um eine
Personengefärdung oder materiellen Schaden auszuschließen.
26WIKA Betriebsanleitung Hand-Held Thermometer
2146872 09/2004 GB/D
Page 27
Hand-Held Thermometer
CTH 6200
■ Wartung
Digitalgerät und Sensoren enthalten keinerlei Verschleissteile,
so dass das Öffnen der Geräte-Gehäuse nicht erforderlich ist
und für die Aufrechterhaltung der Gewährleistungsansprüche
unbedingt unterbleiben sollte.
Für die Reinigung der Gehäuse-Oberflächen empfehlen wir ein
weiches / wenig fusselndes, in Seifenwasser angefeuchtetes Tuch.
Wie für Messmittel allgemein üblich, sollten sie in regelmäßigen
Abständen (ca. 1x jährlich) auf Einhaltung der Spezifikationen
überprüft werden (siehe Punkt 7).
1.3 Anschlüsse
Am oberen Ende des Gerätes:
befindet sich die Anschlussbuchse CH1 zum Anschluss der Temperaturfühler der CPT-Serie (Siehe Kapitel 4.1). und die Buchse
zum Anschluss des Schnittstellenconverters (Siehe Kapitel 5).
D
2146872 09/2004 GB/D
Anschluss für Schnittstellenconverter
CH1
An der linken Seite des Gerätes:
befindet sich die Netzbuchse zum Anschluss des erhältlichen
Netzgerätes (Siehe aktuelle Preisliste CTH 6200).
27WIKA Betriebsanleitung Hand-Held Thermometer
Page 28
Hand-Held Thermometer
CTH 6200
D
1. Offset: signalisiert, dass die Nullpunktverschiebung (Offset)
2. Corr: signalisiert, dass die Steigungskorrektur aktiviert ist.
3. : signalisiert schwache Batterie und andere Warnungen.
4. Logg: erscheint falls Loggerfunktion ausgewählt wurde und
5. AL: blinkt, falls Alarm vorliegt
6. Hauptanzeige: zeigt die Temperatur an.
7. Min/Max/Hold: zeigt an, ob sich in der Nebenanzeige um einen
8. Nebenanzeige: zeigt gemäß gewählter Funktionstaste MIN,
1.4 Anzeigen
6
7
8
aktiv ist.
blinkt bei laufendem Logger
Min, Max oder Hold-Wert handelt
MAX oder Holdwerte angezeigt.
1
2
3
4
5
1.5 Bedienung
Beim Einschalten wird, falls eine Loggerfunktion gewählt wurde,
kurz die Uhrzeit angezeigt, falls ein Nullpunktabgleich durchgeführt
worden ist, wird dies mit der Anzeige "nuLL Corr" signalisiert.
Nach einem Batteriewechsel erscheint automatisch das UhrzeitEinstellungs-Menü (´CLOC´). Überprüfen und korrigieren Sie ggf.
die Uhrzeit (siehe Kapitel 2).
28WIKA Betriebsanleitung Hand-Held Thermometer
2146872 09/2004 GB/D
Page 29
Hand-Held Thermometer
CTH 6200
4
1
2
3
6
5
1. Ein-/Ausschalter
2. Set/Menu: Aufruf der Konfiguration (Menü), 2 sek drücken
3. Tara:Während der Messung: ohne Funktion
Während der Konfiguration: Auswahl der Menüparameter
4. Max-Speicher: Die Taste ´Max´ zeigt den maximal gemessenen
Wert. Erneutes Drücken blendet ihn wieder aus. Zum
Löschen des Max-Wertes ´Max´ für 1 sec. drücken.
5. Min-Speicher: Die Taste ´Min´ zeigt den minimal gemessenen
Wert. Erneutes Drücken blendet ihn wieder aus. Zum
Löschen des Min-Wertes ´Min´ für 1 sec. drücken.
6. Hold-Funktion: Durch Drücken der Taste ´Store/Quit´ wird der
letzte Messwert in der unteren Anzeige gehalten.
Erneutes Drücken blendet den Wert wieder aus (nur
falls Logger deaktiviert).
D
2146872 09/2004 GB/D
6. Logger-Funktionen: Aktivierung über ´Store/Quit´-Taste, falls
im Hauptmenü von Hold auf eine Logger Funktion
umgeschaltet wurde (siehe Kapitel 3).
29WIKA Betriebsanleitung Hand-Held Thermometer
Page 30
Hand-Held Thermometer
CTH 6200
D
Menüführung des Hauptmenüs
(2 sec)
Read*
Logg
Set
Conf
Set
AL
Set**
Logg
Set
Cloc
Speicherdaten*
Aufruf der gespeicherten Einzelwertlogger-Daten (Siehe Kapitel 3)
CYCL 0:01 ... 60:00Zykluszeit in [Minuten:Sekunden] bei zyklischem Logger*
SEtSet Clock: Einstellung der Echtzeituhr
CLOCCLOC HH:MMClock: Einstellen der Uhrzeit Stunden:Minuten
dAtETT.MMDate: Einstellen des Datums Tag.Monat
2146872 09/2004 GB/D
YEAr YYYYYear: Einstellen der Jahreszahl
D
31WIKA Betriebsanleitung Hand-Held Thermometer
Page 32
Hand-Held Thermometer
CTH 6200
´Werden die Taste ´Set´ und ´Store´ gemeinsam länger als 2
Sekunden gedrückt, werden die Werkseinstellungen wiederhergestellt.
D
Befinden sich Daten im Loggerspeicher, können Menüpunkte die
mit (*) gekennzeichnet sind nicht aufgerufen werden. Sollen diese
verändert werden, müssen zunächst die Daten gelöscht werden
(siehe Kapitel 3).
Zum Ändern von Einstellungen 2 sec lang ´Set/Menu´-Taste
drücken, dadurch wird das Menü (Hauptanzeige "SEt") aufgerufen.
Mit ´Set/Menu´ wählen Sie das Menü und mit ´Tara´ können Sie zu
den zugehörigen Parametern springen, die Sie dann verändern
können.
Die Einstellung der Parameter erfolgt mit den Tasten ´Min´ oder
´Max´. Erneutes Drücken von ´Set/Menu´ wechselt zurück zum
Hauptmenü und speichert die Einstellungen.
Mit ´Store/Quit´ wird die Konfiguration beendet.
2. Konfigurieren des Gerätes
2.1 (Unit) Anzeigeeinheit
Die gemessene Temperatur kann wahlweise in °C oder °F angezeigt werden.
2.2 (rES) Anzeigenauflösung
Standardeinstellung: ´Auto´, d.h. das Gerät stellt automatisch auf
die günstigste Auflösung zwischen 0,1 ° und 0,01 ° um.
Für Messungen von Temperaturen, die sich nahe an den Umschaltgrenzen befinden, kann es günstiger sein, eine bestimmte
Auflösung beizubehalten, z.B. um das Protokollieren zu erleichtern.
In diesem Falle wählen Sie bitte die entsprechende Auflösung.
2.3 (Lin E.751) Ausgleichs-Kennlinie
Diese Funktion verwendet die Standard Umrechnung Widerstand
zu Temperatur nach EN60751 (Lin E.751).
32WIKA Betriebsanleitung Hand-Held Thermometer
2146872 09/2004 GB/D
Page 33
Hand-Held Thermometer
CTH 6200
Die Kennlinie nach EN60751 benützt die Temperaturskala ITS90
und folgende Berechnungsformel:
Temperaturen <0 °C:
Rneg(T): = 100 [1 + 3,908310
-3
* T - 5,77510-7 * T2 - 4,18310
-12
* (T - 100) * T3]
Temperaturen >=0 °C:
R
(T): = 100 (1 + 3,908310-3 * T - 5,77510-7 * T2)
pos
2.4 (OFFs) Nullpunktkorrektur / Offset
Für die Temperaturmessung kann eine Nullpunktverschiebung
vorgenommen werden:
angezeigte Temperatur = gemessene Temperatur - Offset
Standardeinstellung: ´off´ = 0,0°, d.h. es wird keine Korrektur
vorgenommen. Die Nullpunktkorrektur wird zusammen mit der
Steigungskorrektur (s.u.) vor allem zum Abgleich von Fühlerabweichungen verwendet.
Ist ein anderer Wert als ´off´ eingestellt, wird dies während des
Betriebs durch den Offset-Pfeil im Display gekennzeichnet.
2.5 (Scal) Steigungskorrektur
Die Steigung der Messung kann mit diesem Faktor beeinflusst
werden (Faktor ist in %):
D
2146872 09/2004 GB/D
angezeigte Temperatur [°C] = gemessene Temperatur [°C] * (1
+ Scal/100)
bzw.
angezeigte Temperatur [°F] = (gemessene Temperatur [°F] - 32 °F)
* (1 + Scal/100) + 32 °F
Standardeinstellung: ´off´ =0,000, d.h. es wird keine Korrektur
vorgenommen. Die Steigungskorrektur wird zusammen mit der
Nullpunktkorrektur (s.o.) vor allem zum Abgleich von Fühlerabweichungen verwendet. Ist ein anderer Wert als ´off´ eingestellt,
wird dies während des Betriebs durch den Corr-Pfeil im Display
gekennzeichnet.
33WIKA Betriebsanleitung Hand-Held Thermometer
Page 34
Hand-Held Thermometer
CTH 6200
2.6 (P.oFF) Abschaltfunktion
D
Wird für die Dauer der Abschaltverzögerung keine Taste gedrückt
bzw. keine Schnittstellenkommunikation vorgenommen, so
schalltet sich das Gerät automatisch ab. Die Abschaltverzögerung
ist zwischen 1 und 120 min wählbar.
Ist P.oFF = oFF, so ist die Abschaltverzögerung deaktiviert.
2.7 (Out) Serielle Schnittstelle
Der Ausgang kann als serielle Schnittstelle verwendet werden.
Wird er nicht benötigt empfehlen wir, den Ausgang abzuschalten,
da dadurch der Stromverbrauch des Gerätes verringert wird.
2.8 (Adr.) Schnittstellen-Adresse
In Vorbereitung für Multiplexerbetrieb.
2.9 (AL.) Alarm
Es sind 3 Einstellungen möglich: aus (AL. oFF), an mit Hupe (AL.
on), an ohne Hupe (AL. no.So).
Bei folgenden Bedingungen wird bei aktiver Alarmfunktion (on oder
no.So) ein Alarm ausgegeben:
■ untere (AL. Lo) bzw. obere Alarmgrenze (AL.Hi) unter- bzw.
überschritten.
■ Sensorfehler (Sens Error)
■ schwache Batterie (bAt)
■ Fe 7: Systemfehler (wird immer mit Hupe gemeldet)
Im Alarmfall blinkt die Anzeige, bei Schnittstellenzugriffen wird das
´PRIO´-Flag gesetzt.
2.10 (CLOC) Echtzeituhr
Die Echtzeituhr wird für die zeitliche Zuordnung der Loggerdaten
benötigt. Kontrollieren Sie deshalb bei Bedarf die Einstellungen.
Nach einem Batteriewechsel wird nach dem Einschalten des
Gerätes automatisch das Uhrzeit-Einstellungs-Menü gestartet.
34WIKA Betriebsanleitung Hand-Held Thermometer
2146872 09/2004 GB/D
Page 35
Hand-Held Thermometer
CTH 6200
3. Bedienung der Loggerfunktion
Grundsätzlich besitzt das Gerät zwei verschiedene Loggerfunktionen, die man über das Hauptmenü aktiviert:
"Func-Stor": jeweils ein Messergebnis wird aufgezeichnet, wenn
die ´Store/ Quit´-Taste gedrückt wird.
"Func-CYCL": Messergebnisse werden automatisch im Abstand
der eingestellten Zykluszeit aufgezeichnet, die Aufzeichnung wird mit 2 sek. lang ´Store/ Quit´ drücken
gestartet.
Der Logger zeichnet jeweils ein Messergebnis pro Datensatz auf.
Zur Auswertung der "Func-CYCL"-Daten benötigen sie die Auswerte-Software GSOFT, mit der auch der Logger sehr einfach
gestartet und eingestellt werden kann.
Bei aktivierter Loggerfunktion: "Func Stor" oder "Func CYCL"
(siehe Menüführung des Hauptmenüs) steht die Hold Funktion
nicht zur Verfügung.
3.1 Func-Stor": Einzelwerte speichern
Jeweils ein Messergebnis wird aufgezeichnet, wenn die ´Store/
Quit´-Taste gedrückt wird. Die gespeicherten Daten können in der
Anzeige selbst betrachtet werden (bei Aufrufen der Konfiguration
erscheint ein zusätzliches Menü: "REAd LoGG") oder mit Hilfe der
Schnittstelle in einen PC eingelesen werden.
D
2146872 09/2004 GB/D
Speicherbare Datensätze: 99
Ein Datensatz besteht max. aus:
■ Messwert zum Zeitpunkt des Speicherns
■ Min.-Peak, Max-Peak seit dem letzten Speichern
■ Uhrzeit und Datum zum Zeitpunkt des Speicherns
Bei jedem Speichern wird kurz "St. XX" angezeigt. XX ist dabei die
Nummer des Messergebnisses.
35WIKA Betriebsanleitung Hand-Held Thermometer
Page 36
Hand-Held Thermometer
CTH 6200
Wenn bereits Daten gespeichert wurden:
Wird Taste ´Store´ 2 sek. lang gedrückt, wird die Auswahl zum
Löschen des Loggerspeichers angezeigt:
D
Die Auswahl erfolgt mit Tasten ´Min´ bzw. ´Max´. Mit ´Store/Quit´
wird die Auswahl quittiert.
Falls der Loggerspeicher voll ist, erscheint:
Einzelwerte betrachten
Im Gegensatz zur zyklischen Loggerfunktion können Einzelwerte
auch direkt in der Anzeige betrachtet werden: 2 sek. lang ´Set/Menu´ drücken: als erstes Menü wird jetzt "rEAd LoGG" (Logger
auslesen) angeboten. Nach Drücken der Taste ´Tara´ wird das
letzte Messergebnis gezeigt, das Wechseln zwischen den Werten
eines Messergebnisses erfolgt ebenfalls mit ´Tara´.
Das Wechseln zu anderen Messergebnissen erfolgt mit den Tasten
´Max´ oder ´Min´.
Alle Datensätzenichts löschen
löschen(Vorgang abbre-
chen)
den zuletzt aufgezeichneten Datensatz löschen
3.2 Func-CYCL": Autom. Aufzeichnung mit einstellbarer Zykluszeit
Die Logger-Zykluszeit ist einstellbar (siehe Konfiguration).
Beispielsweise "CYCL" = 1:60: alle 60 Sekunden wird ein
Messergebnis abgespeichert.
Speicherbare Messergebnisse: 16384
Zykluszeit: 0:01...30:00 (Minuten:Sekunden, min 1s,
max 1h), einstellbar in der Konfiguration
Ein Messergebnis besteht aus:
■ Messwert zum Zeitpunkt des Speicherns
36WIKA Betriebsanleitung Hand-Held Thermometer
2146872 09/2004 GB/D
Page 37
Hand-Held Thermometer
CTH 6200
Loggeraufzeichnung starten:
Durch 2 Sekunden Drücken der Taste ´Store/Quit´ wird die
Aufzeichnung gestartet. Danach wird bei jeder Aufzeichnung kurz
die Anzeige “St.XXXX“ angezeigt. XXXX steht hierbei für die
Nummer des Datensatzes 1 ... 16384.
Falls der Loggerspeicher voll ist, erscheint:
Loggeraufzeichnung Stoppen:
Durch kurzes Drücken von ´Store/Quit´ kann die Aufzeichnung
gestoppt werden. Es erscheint dann eine Sicherheitsabfrage:
Die Auswahl erfolgt mit Tasten ´Min´ bzw. ´Max´. Mit ´Store/Quit´
wird die Auswahl quittiert.
D
Die Aufzeichnung wird automatisch angehalten.
AufzeichnungDie Aufzeichnung
stoppennicht stoppen
2146872 09/2004 GB/D
Wird versucht ein mit zyklischer Aufzeichnung laufendes Messgerät auszuschalten, so wird automatisch nachgefragt, ob die
Aufzeichnung gestoppt werden soll. Nur bei gestoppter Aufzeichnung kann das Gerät abgeschaltet werden. Die Auto-Power-Off
Funktion ist bei laufender Aufzeichnung deaktiviert.
Loggeraufzeichnung löschen:
Wird die Taste ´Store/Quit´ für 2 Sekunden gedrückt, so wird, falls
Loggerdaten vorhanden, die Auswahl zum Löschen des Loggerspeichers angezeigt:
Alle Datensätzenicht löschen
löschen(Vorgang abbre-
chen)
Die Auswahl erfolgt mit Tasten ´Min´ bzw. ´Max´. Mit ´Store/Quit´
wird die Auswahl quittiert.
37WIKA Betriebsanleitung Hand-Held Thermometer
Page 38
Hand-Held Thermometer
CTH 6200
4. Allgemeines zu Temperaturfühlern
D
■ Wärmeableitung durch Fühlerkonstruktion:
Insbesondere bei Messung von Temperaturen die extrem von
der Umgebungstemperatur abweichen, treten messunsicherheiten auf, wenn die Wärmeableitung durch den Fühler nicht
berücksichtigt wird. Bei Messungen in Flüssigkeiten sollte deswegen ausreichend tief eingetaucht und anschließend gerührt
werden. Bei Messungen von Gasen sollte das Fühlerrohr möglichst weit in das zu messende Gas hineinragen (bspw. bei
Kanalmessungen) und das Gas sollte den Fühler möglichst
kräftig umspülen.
Bei Messung von hohen Temperaturen sollte die Fühlerrohrlänge
ausreichend lang gewählt werden, damit die Temperatur am
Handgriff niedrig bleibt.
■ Eigenerwärmung
Der verwendete Sensorstrom beträgt lediglich 0,3 mA. Dadurch
ist in der Praxis die Sensorerwärmung selbst von sehr kleinen
Sensorelementen an ruhender Luft (Worst Case) <= 0,01 °C.
■ Verdunstungskälte
Bei Messungen der Lufttemperatur sollte der Fühler trocken
sein, ansonsten wird eine zu niedrige Temperatur gemessen.
(Abkühlung durch Verdunstung).
4.1 Verfügbare Temperaturfühler
Das Messgerät ist so konzipiert, dass alle Sensoren der CTH-Serie
angesteckt werden können (siehe aktuelle Preisliste CTH 6200).
5. Die serielle Schnittstelle
Mit einem galv. getrennten Schnittstellenconverter (siehe aktuelle
Preisliste CTH 6200) kann das Gerät an eine RS232-Schnittstelle
eines PC angeschlossen werden (auf Anfrage USB-Adapter
möglich), um die Daten mit der WIKA Datenlogger-Auswertesoftware GSOFT (siehe aktuelle Preisliste CTH 6200) zu visualisieren.
Die Übertragung ist durch aufwendige Sicherheitsmechanismen
gegen Übertragungsfehler geschützt (CRC).
38WIKA Betriebsanleitung Hand-Held Thermometer
2146872 09/2004 GB/D
Page 39
Hand-Held Thermometer
CTH 6200
6. Fehler- und Systemmeldungen
AnzeigeBedeutungAbhilfe
Keine
Anzeige
bzw. wirre
Zeichen,
Gerät
reagiert
nicht auf
Tastendruck
----
Err.1
Err.2
Err.3
Err.4
Err.7
Batteriespannung schwach, Funktion ist
nur noch kurze Zeit ge-währleistet
Bei Netzgerätebetrieb:
falsche Spannung
Batterie ist leer
Bei Netzgerätebetrieb:
falsche Spannung
Batterie ist leer
Bei Netzgerätebetrieb:
falsche Spannung / Polung
Systemfehler
Gerät defekt
Es ist kein Sensor angesteckt
Angesteckter Sensor oder
Gerät ist defekt
Messbereich ist überschritten
Falscher Fühler angeschlossen
Sensor oder Gerät defekt
Messbereich ist unterschritten
Falscher Fühler angeschlossen
Sensor oder Gerät defekt
Anzeigebereich überschritten
Anzeigebereich unterschritten
Systemfehler
Gerät ausserhalb zul. Arbeitstemperatur
Neue Batterie einsetzten
Netzgerät überprüfen / austauschen
Neue Batterie einsetzen
Netzgerät überprüfen / austauschen
Neue Batterie einsetzen
Netzgerät überprüfen / austauschen
Batterie und Netzgerät abklemmen,
kurz warten, wieder anstecken
Zur Reparatur einschicken
Sensor an Fühlerbuchse anschließen?
Zur Reparatur einschicken
liegt Temperatur über zul. Bereich?
-> Messwert ist zu hoch!
Fühler überprüfen
Zur Reparatur einschicken
liegt Temperatur unter zul. Bereich?
-> Messwert ist zu tief!
Fühler überprüfen
Zur Reparatur einschicken
-> Auflösung auf 0,1 ° oder Auto stellen
-> Auflösung auf 0,1 ° oder Auto stellen
Zur Reparatur einschicken
Zulässige Arbeitstemperatur einhalten
D
2146872 09/2004 GB/D
39WIKA Betriebsanleitung Hand-Held Thermometer
Page 40
Hand-Held Thermometer
CTH 6200
7. Hinweis zum Kalibrierservice
D
DKD-Schein - amtliche Bescheinigungen:
Soll das Messgerät einen Kalibrierschein erhalten, ist dieses mit
dem zugehörigen Fühler zum Hersteller einzuschicken.
Nur der Hersteller kann die Grundeinstellungen überprüfen und
wenn notwendig korrigieren.
8. Technische Daten
Kenndaten der kompletten Messkette (Digitalgerät und Fühler):
Messeingänge1
FühlertypPt 100 (4-Leiter)
Messbereiche ab Lager-50 ... +250 °C
Gesamt-Messunsicherheit 1)0,2 K
Kalibrierung *Abnahmeprüfzeugnis 3.1.B entsprechend DIN EN 10204
Anzeigegroßes LCD-Display zur Anzeige von 2 x 4 1/2-stelligen Werten und
Auflösung0,01 K bis 200 °C; dann 0,1 K
Temperatureinheiten°C, °F
CTH 6200 Digitalgerät spezifische Angaben:
Messbereiche-99,99 ... +99,99 °C bzw. -199,9 ... +650,0 °C (abhängig vom Fühler)
Funktionen via Tastendruck Min-, Max-Speicher, Hold, Logger (Start/Stopp)
Funktionen via MenüMin-, Max-Alarm (akustisch/visuell),
Datenlogger- Einzelwertlogger: bis zu 99 Aufzeichnungen inkl. Uhrzeit
Schnittstelle (seriell)RS 232-Schnittstelle via Schnittstellenkonverter
Stromversorgung9V-Zink-Kohle-Batterie (Lieferumfang), {alternativ: 9V-Akku oder
Stromaufnahmeca. 1,25 mA
Nenntemperatur25 °C
Temperaturdrift≤ 0,002 °C/K
zul. Umgebungstemperatur -10 ... +50 °C
Relative Feuchte0 ... 95 % r.F. (nicht betauend)
Lagertemperatur-25 ... +70 °C
Gehäuseschlagfestes ABS-Kunststoff, Folientastatur, Klarsichtscheibe
Masseca. 160 g
CTP 62x0 Fühlerspezifische Daten (Pt 100):
SensorartEintauch- und Einstechfühler
Sensoranschluss4-poliger Mini-DIN-Stecker mit PVC-Kabel (ca. 1,0 m)
Messbereich
- Eintauchfühler CTP 6210 -50 ... +250 °C
- Einstechfühler CTP 6290 -50 ... +250 °C
Messstoffberührte TeileEintauchfühler und Einstechfühler: V4A
MaßeSensorlänge: 300 mm; Handgriff: 135 mm
(Ist eine erhöhte Genauigkeit gefordert, so lässt sich diese für einen
engeren Temperaturbereich durch gezielte Justage erreichen.)