WIKA CTH6200 Operating Manual

Operating Instructions Betriebsanleitung
Hand-Held Thermometer
Hand-Held Thermometer
CTH 6200
GB
D
Hand-Held Thermometer CTH 6200
Hand-Held Thermometer CTH 6200
GB
Operating Instructions Hand-Held Thermometer Page 3 - 22
Betriebsanleitung Hand-Held Thermometer Seite 23 - 41
D
2
WIKA Betriebsanleitung Hand-Held Thermometer
WIKA Operating Instructions Hand-Held Thermometer
Hand-Held Thermometer CTH 6200
Contents
Contents
1. General 4
1.1 Safety Instructions 5
1.2 Operation and maintainance advice 6
1.3 Connections 7
1.4 Display 8
1.5 Basic operation 8
2. Configuration 12
2.1 (UNIT) Temperature unit 12
2.2 (rES) Display resolution 12
2.3 (Lin E.751) Sensor curve 13
2.4 (OFFs) Zero point correction / Offset 13
2.5 (Scal) Scale correction 13
2.6 (P.oFF) Power off function 14
2.7 (Out) Serial Interface 14
2.8 (ADR.) Interface-Base address 14
2.9 (AL.) Alarm 14
2.10 (CLOC) Real time clock 14
3. Operation of Logger 15
3.1 "FUNC-STOR": Storing single measurements 15
3.2 "FUNC-CYCL": Automatic recording with selectable 16
Cycle-Time
4. General information about temperature sensors 18
4.1 Available temperature sensors 18
GB
5. The serial interface 18
6. Error and system messages 19
7. Calibration services 20
8. Specification 20
9. Accessories 21
Information This symbol provides you with information, notes an tips.
Warning! This Symbol warns you against actions that can cause damage to persons or to the instrument.
WIKA Betriebsanleitung Hand-Held Thermometer
WIKA Operating Instructions Hand-Held Thermometer
3
Hand-Held Thermometer CTH 6200
GB
1. General
In the following chapters detailed information on the hand-held thermometer CTH 6200 and its proper use can be found.
Should you require further information, or should there be prob­lems which are not dealt with in detail in the operating instructions, please contact the address listed on the last page.
The factory calibration of the instruments are according to respec­tive international guidelines.
The warranty period for the hand-held thermometer CTH 6200 is 24 months according to the general terms of supply of ZVEI. All guarantee claims lapse if the appliance is put to improper use or if the operating instructions are not observed or if an attempt is made to open the appliance. We also point out that the contents of these operating instructions does not form part of an earlier or existing agreement, assurance or legal relationship or is meant to change these. All obligations of WIKA Alexander Wiegand GmbH & Co. KG result from the respec­tive sales contract and the general business terms of WIKA Alexander Wiegand GmbH & Co. KG.
WIKA is a registered trade mark of WIKA Alexander Wiegand GmbH & Co. KG. Names of companies or products mentioned in this handbook are registered trade marks of these manufacturers.
We reserve the right to effect reasonable changes on the basis of technical improvements.
Any reproduction of this manual or parts thereof by any means is prohibited.
© 2004 Copyright WIKA Alexander Wiegand GmbH & Co. KG
4
WIKA Betriebsanleitung Hand-Held Thermometer
WIKA Operating Instructions Hand-Held Thermometer
Hand-Held Thermometer CTH 6200
1.1 Safety Instructions
This device has been designed and tested in accordance with the safety regulations for electronic devices. However, its trouble-free operation and reliability cannot be guaranteed unless the standard safety measures and special safety advises given in this manual will be adhered to when using the device.
1. Trouble-free operation and reliability of the device can only be guaranteed if the device is not Subjected to any other climatic conditions than those stated under "Specification".
2. Device and sensors have to be handled with care (don't throw, hit, etc.). Protect plugs and sockets from soiling.
3. If the device is transported from a cold to a warm environment condensation may cause in a failure of the function. In such a case make sure the device temperature has adjusted to the ambient temperature before trying a new start-up.
4. If device is to be connected to other devices (e.g. via serial interface) the circuitry has to be designed most carefully. Inter­nal connection in third party devices may result in not-permis­sible voltages impairing or destroying the device or another device connected.
If device is operated with a defective mains power supply (e.g. short circuit from mains voltage to output voltage) this may result in hazardous voltages at the device (e.g. at sensor socket or interface).
GB
5. If there is a risk whatsoever involved in running it, the device has to be switched off immediately and to be marked accor­dingly to avoid re-starting.
Operator safety may be a risk if:
there is visible damage to the device
the device is not working as specified
the device has been stored under unsuitable conditions for
a longer period of time. In case of doubt, please return device to manufacturer for repair or maintenance.
6. Repairs should only be carried out by the manufacturer. All other repairs or modifications of the transmitter are unautho­rized.
7. Any operation not described in the following instructions or
WIKA Betriebsanleitung Hand-Held Thermometer
WIKA Operating Instructions Hand-Held Thermometer
outside the specifications must not be carried out.
5
Hand-Held Thermometer CTH 6200
GB
1.2 Operation and maintenance advice
Operation with battery or accu If and ´bAt´ are shown in the secondary display the battery has been used up and needs to be replaced, or resp. the accu is empty and has to be recharged. The device will, however, operate correctly for a certain amount of time. If ´bAt´ is shown in the upper display the voltage is too low to operate the device; the battery or accu has been completely used up.
We recommend to take out battery/accu if device is not used for a longer period of time!
Mains operation with power supply
When using a power supply please note that operating voltage has to be 10.5 to 12 V DC. Do not apply overvoltage! Cheap 12 V-power supplies often have excessive no-load voltage. We, therefore, recommend using regulated voltage power supplies. Trouble-free operation is guaranteed by our power supply unit. Prior to connecting the power supply to the mains make sure that the operating voltage stated at the power supply is identical to the mains voltage.
Connecting/Changing Sensors Switch off device before changing the sensor. Connect sensor before switching on the device. In such case take the plug not at the casing but at the buckling protection at the end of the plug. When connecting the probe the plug will slide in smoothly if plug is entered correctly. To disconnect sensor plug do not pull at the cable but at the plug.
Probes may be extremely hot or cold. Precautions must be taken to prevent personal injury or damage to objects.
6
WIKA Betriebsanleitung Hand-Held Thermometer
WIKA Operating Instructions Hand-Held Thermometer
Hand-Held Thermometer CTH 6200
Maintenance advice
The digital instrument and sensors are completely of solid state principle, not containing any moving parts. Dismantling of en­sures invalidates warranty.
If cleaning is necessary, use a cloth damp by a mild soap cleaner, avoid any solvent or caustic abrasive agent.
As for measurement devices usual, the accuracy of the device should be tested in regularly time periods (approx. 1 time per year
- see chapter 7).
1.3 Connections
At the top of the device:
are located the socket CH1 for connecting the temperature sensor of the CPT-series (see chapter 4.1) and the socket for connecting the interface converter (see chapter 5).
Connection for interface converter
GB
CH1
At the left side of the device:
is located the mains adapter socket to connect the available power supply unit (see valid price list of CTH 6200).
WIKA Betriebsanleitung Hand-Held Thermometer
WIKA Operating Instructions Hand-Held Thermometer
7
Hand-Held Thermometer CTH 6200
GB
1.4 Display
1
2
6
3
7
8
4
5
1. Offset: indicates that zero point offset (offset) is activated.
2. Corr: indicates that correction factor is activated.
3. : indicates a low battery and other warnings.
4. Logg: Shown if logger function is selected, flashes if cyclic
logger is running.
5. AL: Flashes if alarm is present.
6. Main display: Currently measured temperature.
7. Min/Max/Hold: shows if a MIN, MAX or Hold value is displayed
in the secondary display.
8. Secondary display: shows MIN, MAX or Hold value acc. to selected function.
1.5 Basic operation
When switching on the device and the logger function is not off
the time of the integrated clock will shortly be displayed. If a zero point adjustment was carried out the display shows shortly "nuLL Corr". After changing the battery the clock-setting menu is activated automatically (´CLOC´). Check the clock and adjust, if necessary (see chapter 2).
8
WIKA Betriebsanleitung Hand-Held Thermometer
WIKA Operating Instructions Hand-Held Thermometer
Hand-Held Thermometer CTH 6200
4
1
2
3
6
5
1. On/Off key
2. Set/Menu: configuration will activated Menu, press for 2 sec
3. Tara: When taking measurements: no function
During configuration: Selection of menu-parameter
4. Max-Function: Pressing ´Max´ shows the maximum of the
measured values. Pressing it again hides it. To clear the max memory press key ´Max´ for 1 seconds.
5. Min-Function: Pressing ´Min´ shows the minimum of the
measured values. Pressing it again hides it. To clear the min memory press key ´Min´ for 1 seconds.
GB
6. Hold-Function: By pressing ´Store/Quit´ the last measuring
value will be held in the secondary display. Pressing it again hides it. (only when main menu logger = ´off´ is chosen).
6. Logger-Functions: Activation with key ´Store/Quit´, if in the
main menu a logger function was chosen before. (see Chapter 3).
WIKA Betriebsanleitung Hand-Held Thermometer
WIKA Operating Instructions Hand-Held Thermometer
9
Hand-Held Thermometer CTH 6200
GB
Main Menu
Read*
Logg
Set
Conf
Set
AL
Set**
Logg
(2 sec)
Stored data*
Call of Store-Func. data with respective time (see chapter 3)
Configuration
Unit ** / resolution** zero point** & scale correction** Power-off / interface-off
Alarm
Min/Max alarm visual with/without sound
Logger**
switch from Hold to Logger function: individual values (Stor) or cyclic (CYCL)
Set
Cloc
(ring - shift)
* appears only if data (Store-Func.) is in the logger. ** appears only if no data is in the logger (see chapter 3).
10
WIKA Betriebsanleitung Hand-Held Thermometer
WIKA Operating Instructions Hand-Held Thermometer
System-clock
time / day & month / year
set parameter next store
parameter & quit
Hand-Held Thermometer CTH 6200
Menu Param. Values Meaning „Menu“ or
SEt Set Configuration: Generic Settings
ConF Unit °C: All temperatures values are in degrees Celsius *
°F: All temperatures values are in degrees Fahrenheit
rES 0.1 °: Resolution 0.1 °C *
0.01° Auto:
Resolution 0.01 °C *
Resolution is selected automatically
Lin E.751 Measuring / sensor curve according to EN 60751 *
USEr
In preparation
OFFS -2.50 °C ... 2.50 °C The zero point of the measurement will be displaced by this value to *
resp. -4.50 °F ... 4.50 °F compensate for deviations in the probe or in the measuring device.
oFF: Zero displacement inactive (=0.0 °C)
SCAL -2.000 ... 2.000 The scale of the measuring will be changed by this factor [in %] to *
compensate deviations of temperature probe or measuring device.
oFF: Scale correction factor inactive ( = 0.000)
P.oFF 1 ... 120 Auto Power-Off delay in minutes.
Device will be automatically switched off as soon as this time has
elapsed if no key is pressed / no interface communication takes place.
oFF Auto Power-off function inactive (continuous operation)
Out oFF: Function of the output: No output function, lowest power consumption
SEr: Output is serial interface dAC: In preparation
Adr. 01, 11 .. 91 In preparation
SEt Set Alarm: Settings of alarm function
AL. AL. on Alarm Sensor 1
no. So Alarm Sensor 1 on, without horn-sound oFF No alarm function
AL.Lo min range ... AL.Hi Min alarm rail (not when AL.oFF) AL.Hi AL.Lo ... max range Max alarm rail (not when AL.oFF)
on, with horn-sound
SEt Set Logger: Configuration of logger function *
LoGG Func CYCL Cyclic: logger function ,cyclic logger` *
Stor Store: logger function ,individual value logger` * oFF No logger function *
CYCL 0:01 ... 60:00 Cycle time of cyclic logger [minutes:seconds] *
SEt Set Clock: Setting of real time clock
CLOC CLOC HH:MM Clock: setting ot time hours:minutes
dAtE TT.MM Date: setting of time day.month YEAr YYYY Year: setting of time year
GB
WIKA Betriebsanleitung Hand-Held Thermometer
WIKA Operating Instructions Hand-Held Thermometer
11
Hand-Held Thermometer CTH 6200
GB
If keys ´Set´ and ´Store´ are pressed simultaneously for more the 2 seconds, the settings are reset of the settings ex-works.
If the logger memory contains data already, the menus/parameters marked with (*) can not be invoked! If these should be altered the logger memory has to be cleared before! (see chapter 3)
2. Configuration
To change device settings, press key: ´Set/Menu´ for 2 seconds. This will call the configuration menu (main display: "SEt"). Pressing key: ´Set/Menu´ changes between the menus, pressing key: ´Tara´ jumps to the referring parameters, which can be selected with key: ´Tara´. The parameters can be changed with key: ´Min´ or ´Max´. Pressing key: ´Set/Menu´ again jumps back to the main configuration menu and saves the settings. Key: ´Store/Quit´ finishes the configuration and returns to stand­ard measuring operation.
2.1 (Unit) Temperature unit
The measured temperature can be displayed in °C or °F.
2.2 (rES) Display resolution
Standard setting: ´Auto´ ,i.e. the device automatically switches over to the optimum resolution between 0.1 ° and 0.01 °. If temperatures to be measured are near the switching threshold, a fixed resolution may be better, e.g. for easy recording. In such a case please select the optimum resolution manually.
2.3 (Lin E.751) Sensor curve
This function used the standard conversion of resistance to temperature following EN60751 (Lin E.751).
12
WIKA Betriebsanleitung Hand-Held Thermometer
WIKA Operating Instructions Hand-Held Thermometer
Hand-Held Thermometer CTH 6200
The sensor curve following EN60751 uses the international temperature scale ITS90 and following formulas:
Temperatures <0 °C: R
(T): = 100 [1 + 3,908310-3 * T - 5,77510-7 * T2 - 4,18310
neg
-12
* (T - 100) * T3]
Temperatures >=0 °C: R
(T): = 100 (1 + 3,908310-3 * T - 5,77510-7 * T2)
pos
2.4 (OFFs) Zero point correction / Offset
A zero point correction can be carried out for the measured temperature:
temperature displayed = temperature measured - offset
Standard setting: ´off´ = 0.0 °, i.e. no zero point correction will be carried out. The zero point correction is mainly used to compen­sate for sensor deviations. Unless ´off´ is set, this value will be displayed shortly after the device is switched on; during operation it will be identified by means of the offset arrow in the display.
2.5 (Scal) Scale correction
GB
The scale of the measuring can be influenced by this setting (factor is in %).
displayed temperature [°C] = measured temperature [°C] * (1 + Scal/100)
or respectively
displayed temperature [°F] = (measured temperature [°C] - 32 °F) * (1 + Scal/100) + 32 °F
Standard setting: ´off´ =0.000, i.e. temperature is nor corrected. Together with the zero displacement (see above) this factor is mainly used to compensate for sensor deviations. Unless the factor is set to ´off´, the Corr arrow in the display shows an active scale correction.
WIKA Betriebsanleitung Hand-Held Thermometer
WIKA Operating Instructions Hand-Held Thermometer
13