3.2 "FUNC-CYCL": Automatic recording with selectable16
Cycle-Time
4.General information about temperature sensors18
4.1 Available temperature sensors18
GB
5.The serial interface18
6.Error and system messages19
7.Calibration services20
8.Specification20
9.Accessories21
Information
This symbol provides you with information, notes an tips.
Warning!
This Symbol warns you against actions that can cause
damage to persons or to the instrument.
2146872 09/2004 GB/D
WIKA Betriebsanleitung Hand-Held Thermometer
WIKA Operating Instructions Hand-Held Thermometer
3
Hand-Held Thermometer
CTH 6200
GB
1. General
In the following chapters detailed information on the hand-held
thermometer CTH 6200 and its proper use can be found.
Should you require further information, or should there be problems which are not dealt with in detail in the operating instructions,
please contact the address listed on the last page.
The factory calibration of the instruments are according to respective international guidelines.
The warranty period for the hand-held thermometer CTH 6200 is
24 months according to the general terms of supply of ZVEI.
All guarantee claims lapse if the appliance is put to improper use
or if the operating instructions are not observed or if an attempt is
made to open the appliance.
We also point out that the contents of these operating instructions
does not form part of an earlier or existing agreement, assurance
or legal relationship or is meant to change these. All obligations of
WIKA Alexander Wiegand GmbH & Co. KG result from the respective sales contract and the general business terms of WIKA
Alexander Wiegand GmbH & Co. KG.
WIKA is a registered trade mark of WIKA Alexander Wiegand
GmbH & Co. KG. Names of companies or products mentioned in
this handbook are registered trade marks of these manufacturers.
We reserve the right to effect reasonable changes on the basis of
technical improvements.
Any reproduction of this manual or parts thereof by any means is
prohibited.
This device has been designed and tested in accordance with
the safety regulations for electronic devices.
However, its trouble-free operation and reliability cannot be
guaranteed unless the standard safety measures and special
safety advises given in this manual will be adhered to when
using the device.
1. Trouble-free operation and reliability of the device can only be
guaranteed if the device is not Subjected to any other climatic
conditions than those stated under "Specification".
2. Device and sensors have to be handled with care (don't throw,
hit, etc.). Protect plugs and sockets from soiling.
3. If the device is transported from a cold to a warm environment
condensation may cause in a failure of the function. In such a
case make sure the device temperature has adjusted to the
ambient temperature before trying a new start-up.
4. If device is to be connected to other devices (e.g. via serial
interface) the circuitry has to be designed most carefully. Internal connection in third party devices may result in not-permissible voltages impairing or destroying the device or another
device connected.
If device is operated with a defective mains power supply (e.g.
short circuit from mains voltage to output voltage) this may
result in hazardous voltages at the device
(e.g. at sensor socket or interface).
GB
5. If there is a risk whatsoever involved in running it, the device
has to be switched off immediately and to be marked accordingly to avoid re-starting.
Operator safety may be a risk if:
■there is visible damage to the device
■the device is not working as specified
■the device has been stored under unsuitable conditions for
a longer period of time.
In case of doubt, please return device to manufacturer for repair
or maintenance.
6. Repairs should only be carried out by the manufacturer. All
other repairs or modifications of the transmitter are unauthorized.
7. Any operation not described in the following instructions or
2146872 09/2004 GB/D
WIKA Betriebsanleitung Hand-Held Thermometer
WIKA Operating Instructions Hand-Held Thermometer
outside the specifications must not be carried out.
5
Hand-Held Thermometer
CTH 6200
GB
1.2 Operation and maintenance advice
■ Operation with battery or accu
If ∆ and ´bAt´ are shown in the secondary display the battery
has been used up and needs to be replaced, or resp. the accu
is empty and has to be recharged. The device will, however,
operate correctly for a certain amount of time. If ´bAt´ is shown
in the upper display the voltage is too low to operate the device;
the battery or accu has been completely used up.
We recommend to take out battery/accu if device is not used for a
longer period of time!
■ Mains operation with power supply
When using a power supply please note that operating voltage
has to be 10.5 to 12 V DC. Do not apply overvoltage! Cheap
12 V-power supplies often have excessive no-load voltage.
We, therefore, recommend using regulated voltage power
supplies. Trouble-free operation is guaranteed by our power
supply unit.
Prior to connecting the power supply to the mains make sure
that the operating voltage stated at the power supply is
identical to the mains voltage.
■ Connecting/Changing Sensors
Switch off device before changing the sensor. Connect sensor
before switching on the device. In such case take the plug not
at the casing but at the buckling protection at the end of the
plug. When connecting the probe the plug will slide in smoothly
if plug is entered correctly. To disconnect sensor plug do not
pull at the cable but at the plug.
Probes may be extremely hot or cold. Precautions must be
taken to prevent personal injury or damage to objects.
6
WIKA Betriebsanleitung Hand-Held Thermometer
WIKA Operating Instructions Hand-Held Thermometer
2146872 09/2004 GB/D
Hand-Held Thermometer
CTH 6200
■ Maintenance advice
The digital instrument and sensors are completely of solid state
principle, not containing any moving parts. Dismantling of ensures invalidates warranty.
If cleaning is necessary, use a cloth damp by a mild soap cleaner,
avoid any solvent or caustic abrasive agent.
As for measurement devices usual, the accuracy of the device
should be tested in regularly time periods (approx. 1 time per year
- see chapter 7).
1.3 Connections
At the top of the device:
are located the socket CH1 for connecting the temperature sensor
of the CPT-series (see chapter 4.1) and the socket for connecting
the interface converter (see chapter 5).
Connection for interface
converter
GB
CH1
At the left side of the device:
is located the mains adapter socket to connect the available
power supply unit (see valid price list of CTH 6200).
2146872 09/2004 GB/D
WIKA Betriebsanleitung Hand-Held Thermometer
WIKA Operating Instructions Hand-Held Thermometer
7
Hand-Held Thermometer
CTH 6200
GB
1.4 Display
1
2
6
3
7
8
4
5
1. Offset: indicates that zero point offset (offset) is activated.
2. Corr: indicates that correction factor is activated.
3. : indicates a low battery and other warnings.
4. Logg: Shown if logger function is selected, flashes if cyclic
logger is running.
5. AL: Flashes if alarm is present.
6. Main display: Currently measured temperature.
7. Min/Max/Hold: shows if a MIN, MAX or Hold value is displayed
in the secondary display.
8. Secondary display: shows MIN, MAX or Hold value acc. to
selected function.
1.5 Basic operation
When switching on the device and the logger function is not off
the time of the integrated clock will shortly be displayed. If a zero
point adjustment was carried out the display shows shortly "nuLL
Corr".
After changing the battery the clock-setting menu is activated
automatically (´CLOC´). Check the clock and adjust, if necessary
(see chapter 2).
8
WIKA Betriebsanleitung Hand-Held Thermometer
WIKA Operating Instructions Hand-Held Thermometer
2146872 09/2004 GB/D
Hand-Held Thermometer
CTH 6200
4
1
2
3
6
5
1. On/Off key
2. Set/Menu: configuration will activated Menu, press for 2 sec
3. Tara:When taking measurements: no function
During configuration: Selection of menu-parameter
4. Max-Function: Pressing ´Max´ shows the maximum of the
measured values. Pressing it again hides it. To clear
the max memory press key ´Max´ for 1 seconds.
5. Min-Function: Pressing ´Min´ shows the minimum of the
measured values. Pressing it again hides it. To clear
the min memory press key ´Min´ for 1 seconds.
GB
6. Hold-Function: By pressing ´Store/Quit´ the last measuring
value will be held in the secondary display. Pressing
it again hides it. (only when main menu logger = ´off´
is chosen).
6. Logger-Functions: Activation with key ´Store/Quit´, if in the
main menu a logger function was chosen before.
(see Chapter 3).
2146872 09/2004 GB/D
WIKA Betriebsanleitung Hand-Held Thermometer
WIKA Operating Instructions Hand-Held Thermometer
9
Hand-Held Thermometer
CTH 6200
GB
Main Menu
Read*
Logg
Set
Conf
Set
AL
Set**
Logg
(2 sec)
Stored data*
Call of Store-Func. data with respective
time (see chapter 3)
Configuration
Unit ** / resolution**
zero point** & scale correction**
Power-off / interface-off
Alarm
Min/Max alarm visual with/without
sound
Logger**
switch from Hold to Logger function:
individual values (Stor) or cyclic (CYCL)
Set
Cloc
(ring - shift)
* appears only if data (Store-Func.) is in the logger.
** appears only if no data is in the logger (see chapter 3).
10
WIKA Betriebsanleitung Hand-Held Thermometer
WIKA Operating Instructions Hand-Held Thermometer
System-clock
time / day & month / year
set parameter next store
parameter & quit
2146872 09/2004 GB/D
Hand-Held Thermometer
CTH 6200
MenuParam. ValuesMeaning
„Menu“ or
SEtSet Configuration: Generic Settings
ConFUnit°C:All temperatures values are in degrees Celsius*
°F:All temperatures values are in degrees Fahrenheit
rES0.1 °:Resolution 0.1 °C*
0.01°
Auto:
Resolution 0.01 °C*
Resolution is selected automatically
LinE.751Measuring / sensor curve according to EN 60751*
USEr
In preparation
OFFS -2.50 °C ... 2.50 °CThe zero point of the measurement will be displaced by this value to*
resp. -4.50 °F ... 4.50 °F compensate for deviations in the probe or in the measuring device.
oFF:Zero displacement inactive (=0.0 °C)
SCAL -2.000 ... 2.000The scale of the measuring will be changed by this factor [in %] to*
compensate deviations of temperature probe or measuring device.
oFF:Scale correction factor inactive ( = 0.000)
P.oFF 1 ... 120Auto Power-Off delay in minutes.
Device will be automatically switched off as soon as this time has
elapsed if no key is pressed / no interface communication takes place.
oFFAuto Power-off function inactive (continuous operation)
OutoFF:Function of the output: No output function, lowest power consumption
SEr:Output is serial interface
dAC:In preparation
Adr.01, 11 .. 91In preparation
SEtSet Alarm: Settings of alarm function
AL.AL.onAlarm Sensor 1
no. SoAlarm Sensor 1 on, without horn-sound
oFFNo alarm function
AL.Lo min range ... AL.HiMin alarm rail (not when AL.oFF)
AL.Hi AL.Lo ... max rangeMax alarm rail (not when AL.oFF)
on, with horn-sound
SEtSet Logger: Configuration of logger function*
LoGGFuncCYCLCyclic: logger function ,cyclic logger`*
StorStore: logger function ,individual value logger`*
oFFNo logger function*
CYCL 0:01 ... 60:00Cycle time of cyclic logger [minutes:seconds]*
SEtSet Clock: Setting of real time clock
CLOCCLOC HH:MMClock: setting ot time hours:minutes
dAtETT.MMDate: setting of time day.month
YEAr YYYYYear: setting of time year
GB
2146872 09/2004 GB/D
WIKA Betriebsanleitung Hand-Held Thermometer
WIKA Operating Instructions Hand-Held Thermometer
11
Hand-Held Thermometer
CTH 6200
GB
If keys ´Set´ and ´Store´ are pressed simultaneously for more the 2
seconds, the settings are reset of the settings ex-works.
If the logger memory contains data already, the menus/parameters
marked with (*) can not be invoked! If these should be altered the
logger memory has to be cleared before! (see chapter 3)
2. Configuration
To change device settings, press key: ´Set/Menu´ for 2 seconds.
This will call the configuration menu (main display: "SEt").
Pressing key: ´Set/Menu´ changes between the menus, pressing
key: ´Tara´ jumps to the referring parameters, which can be
selected with key: ´Tara´.
The parameters can be changed with key: ´Min´ or ´Max´. Pressing
key: ´Set/Menu´ again jumps back to the main configuration menu
and saves the settings.
Key: ´Store/Quit´ finishes the configuration and returns to standard measuring operation.
2.1 (Unit) Temperature unit
The measured temperature can be displayed in °C or °F.
2.2 (rES) Display resolution
Standard setting: ´Auto´ ,i.e. the device automatically switches over
to the optimum resolution between 0.1 ° and 0.01 °.
If temperatures to be measured are near the switching threshold, a
fixed resolution may be better, e.g. for easy recording.
In such a case please select the optimum resolution manually.
2.3 (Lin E.751) Sensor curve
This function used the standard conversion of resistance to
temperature following EN60751 (Lin E.751).
12
WIKA Betriebsanleitung Hand-Held Thermometer
WIKA Operating Instructions Hand-Held Thermometer
2146872 09/2004 GB/D
Hand-Held Thermometer
CTH 6200
The sensor curve following EN60751 uses the international
temperature scale ITS90 and following formulas:
A zero point correction can be carried out for the measured
temperature:
temperature displayed = temperature measured - offset
Standard setting: ´off´ = 0.0 °, i.e. no zero point correction will be
carried out. The zero point correction is mainly used to compensate for sensor deviations. Unless ´off´ is set, this value will be
displayed shortly after the device is switched on;
during operation it will be identified by means of the offset arrow in
the display.
2.5 (Scal) Scale correction
GB
The scale of the measuring can be influenced by this setting
(factor is in %).
displayed temperature [°C] = measured temperature [°C] * (1 + Scal/100)
or respectively
displayed temperature [°F] = (measured temperature [°C] - 32
°F) * (1 + Scal/100) + 32 °F
Standard setting: ´off´ =0.000, i.e. temperature is nor corrected.
Together with the zero displacement (see above) this factor is
mainly used to compensate for sensor deviations.
Unless the factor is set to ´off´, the Corr arrow in the display shows