Hand-held thermometer
Model CTH6200
Applications
■
Calibration service companies and service industry
■
Measurement and control laboratories
■
Industry (laboratory, workshop and production)
■
Quality assurance
Special features
Calibration technology
WIKA data sheet CT 51.01
for further approvals
see page 3
■
Pt100 probes for -50 ... +250 °C (-58 ... +482 °F)
■
Accuracy: < 0.2 K (complete measuring chain)
■
Integrated data logger
■
GSoft data logger evaluation software available
■
Incl. calibration certicate
Description
Extensive application possibilities
The model CTH6200 hand-held thermometer is suitable for
all precise temperature measurement tasks and impresses
with its simple operation. Its design makes it especially
suitable for the commissioning, maintenance and service/
calibration of temperature instruments and equipment.
Functionality
The digital indicator is the central functional unit of the handheld temperature measuring instrument and provides the
option to connect Pt100 temperature probes of the CTP62x0
series.
Both immersion and penetration probes are available
from stock, along with an extensive range of accessories.
Customer-specic probes are available on request.
An integrated data logger and various other functions (such
as Min, Max, Hold or Alarm) ensure that the hand-held
thermometer can be used for many dierent applications.
Hand-held thermometer model CTH6200
Its simple operation, very handy design and low weight round
o the convenience of this hand-held temperature measuring
instrument.
Complete test and service cases
For the secure transportation and storage of the instrument
and accessories, a rugged service case is available as a
sensible addition. The service case can carry up to two
model CTH6200 digital indicators and several temperature
probes, or the case is available with a model CPH6200 handheld pressure indicator as well.
Certied accuracy
For each temperature probe, the accuracy for the complete
measuring chain is certied by a factory calibration
certicate which accompanies the instrument. On request,
we can provide a DKD/DAkkS calibration certicate for this
instrument.
Data sheets showing similar products:
Hand-held pressure indicator; model CPH6200; see data sheet CT 11.01
Page 1 of 7WIKA data sheet CT 51.01 ∙ 08/2015
Specications
Model CTH6200
Hand-held thermometer model CTH6200 (complete measuring chain)
Measuring inputs 1
Probe type Pt100 (4-wire)
Measuring range -50 ... +250 °C (-58 ... +482 °F)
Resolution 0.01 K up to 200 °C (392 °F); then 0.1 K
1)
Accuracy of the measuring chain
1) The accuracy applies for the respective indicator-probe combination following adjustment and calibration, that is the deviation between reference and CTH6200.
Digital indicator model CTH6200
Display
Display Large LC display for display of values and additional information
Indication range -99.99 ... +99.99 °C or -199.9 ... +650.0 °C (-99.99 ... +99.99 °F or -199.9 ... +999.9 °F)
Temperature units °C, °F
Nominal temperature 25 °C
Temperature drift ≤ 0.002 °C/K
Functions
Functions via key press Min/Max memory, Hold, Logger (Start/Stop)
Menu function Min/Max alarm (audible/visual), automatic Power-O function [1 ... 120 min], zero point/slope input,
Data loggers - Individual value logger: Up to 99 recordings incl. time via key press
Voltage supply
Power supply 9 V zinc-carbon battery (in scope of delivery), {alternative: 9 V rechargeable battery or mains power
Battery life approx. 20 hours of operation
Current supply Slow measuring cycle: < 1.6 mA; fast: < 7.0 mA; low-power logger funct.: < 0.3 mA
Permissible ambient conditions
Operating temperature -10 ... +50 °C (14 ... 122 °F)
Storage temperature -25 ... +70 °C (-13 ... +158 °F)
Relative humidity 0 ... 95 % r. h. (non-condensing)
Communication
Interface (serial) RS-232 or USB via special interface cable
Case
Material impact-resistant ABS plastic, membrane keyboard, transparent screen
Dimensions see technical drawings
Weight approx. 160 g (0.35 lbs)
{} Items in curved brackets are optional extras for an additional price.
0.2 K (optional: increased accuracy through specic adjustment)
(dependent on temperature probe)
data logger settings
- Cyclic logger: Automatic recording of up to 16,200 values incl. time
- Cycle time: Selectable from 1 ... 3,600 seconds
supply}
Page 2 of 7 WIKA data sheet CT 51.01 ∙ 08/2015
Temperature probe model CTP62x0
Probe characteristics
Sensor type Immersion and penetration probe
Probe type Pt100 (4-wire)
Wetted parts Immersion probe and penetration probe: V4A
Connection to the CTH6200 4-pin mini DIN connector with PVC cable (approx. 1.0 m)
Measuring ranges
Immersion probe CTP6210 -50 ... +250 °C (-58 ... +482 °F)
Penetration probe CTP6290 -50 ... +250 °C (-58 ... +482 °F)
Sensor input
Current supply Slow measuring cycle: < 1.6 mA; fast: < 7.0 mA; low-power logger funct.: < 0.3 mA
Nominal temperature 25 °C (77 °F)
Temperature drift ≤ 0.002 °C/K
Approvals
Logo Description Country
EC declaration of conformity
■
EMC directive 2004/108/EC
EAC
■
Electromagnetic compatibility
■
Low voltage directive
GOST
Metrology, measurement technology
KazInMetr
Metrology, measurement technology
MtschS
Permission for commissioning
Uzstandard
Metrology, measurement technology
European Community
Eurasian Economic Community
Russia
Kazakhstan
Kazakhstan
Uzbekistan
Certicates
Certicate
Calibration
Recommended recalibration interval 1 year (dependent on conditions of use)
2) Calibration at 23 °C ±2 °C (73.4 °F ±3.6 °F) ambient temperature.
Approvals and certicates, see website
2)
Standard: 3.1 calibration certicate per DIN EN 10204
Option: DKD/DAkkS calibration certicate
Page 3 of 7WIKA data sheet CT 51.01 ∙ 08/2015