WIKA CTH6200 Specifications

Hand-held thermometer Model CTH6200
Applications
Measurement and control laboratories
Industry (laboratory, workshop and production)
Quality assurance
Special features
Calibration technology
WIKA data sheet CT 51.01
for further approvals see page 3
Pt100 probes for -50 ... +250 °C (-58 ... +482 °F)
Accuracy: < 0.2 K (complete measuring chain)
Integrated data logger
GSoft data logger evaluation software available
Incl. calibration certicate
Description
Extensive application possibilities
The model CTH6200 hand-held thermometer is suitable for all precise temperature measurement tasks and impresses with its simple operation. Its design makes it especially suitable for the commissioning, maintenance and service/ calibration of temperature instruments and equipment.
Functionality
The digital indicator is the central functional unit of the hand­held temperature measuring instrument and provides the option to connect Pt100 temperature probes of the CTP62x0 series. Both immersion and penetration probes are available from stock, along with an extensive range of accessories.
Customer-specic probes are available on request.
An integrated data logger and various other functions (such as Min, Max, Hold or Alarm) ensure that the hand-held
thermometer can be used for many dierent applications.
Hand-held thermometer model CTH6200
Its simple operation, very handy design and low weight round
o the convenience of this hand-held temperature measuring
instrument.
Complete test and service cases
For the secure transportation and storage of the instrument and accessories, a rugged service case is available as a sensible addition. The service case can carry up to two model CTH6200 digital indicators and several temperature probes, or the case is available with a model CPH6200 hand­held pressure indicator as well.
Certied accuracy
For each temperature probe, the accuracy for the complete
measuring chain is certied by a factory calibration certicate which accompanies the instrument. On request, we can provide a DKD/DAkkS calibration certicate for this
instrument.
Data sheets showing similar products: Hand-held pressure indicator; model CPH6200; see data sheet CT 11.01
Page 1 of 7WIKA data sheet CT 51.01 ∙ 08/2015
Specications
Model CTH6200
Hand-held thermometer model CTH6200 (complete measuring chain)
Measuring inputs 1
Probe type Pt100 (4-wire)
Measuring range -50 ... +250 °C (-58 ... +482 °F)
Resolution 0.01 K up to 200 °C (392 °F); then 0.1 K
1)
Accuracy of the measuring chain
1) The accuracy applies for the respective indicator-probe combination following adjustment and calibration, that is the deviation between reference and CTH6200.
Digital indicator model CTH6200
Display
Display Large LC display for display of values and additional information
Indication range -99.99 ... +99.99 °C or -199.9 ... +650.0 °C (-99.99 ... +99.99 °F or -199.9 ... +999.9 °F)
Temperature units °C, °F
Nominal temperature 25 °C
Temperature drift ≤ 0.002 °C/K
Functions
Functions via key press Min/Max memory, Hold, Logger (Start/Stop)
Menu function Min/Max alarm (audible/visual), automatic Power-O function [1 ... 120 min], zero point/slope input,
Data loggers - Individual value logger: Up to 99 recordings incl. time via key press
Voltage supply
Power supply 9 V zinc-carbon battery (in scope of delivery), {alternative: 9 V rechargeable battery or mains power
Battery life approx. 20 hours of operation
Current supply Slow measuring cycle: < 1.6 mA; fast: < 7.0 mA; low-power logger funct.: < 0.3 mA
Permissible ambient conditions
Operating temperature -10 ... +50 °C (14 ... 122 °F)
Storage temperature -25 ... +70 °C (-13 ... +158 °F)
Relative humidity 0 ... 95 % r. h. (non-condensing)
Communication
Interface (serial) RS-232 or USB via special interface cable
Case
Material impact-resistant ABS plastic, membrane keyboard, transparent screen
Dimensions see technical drawings
Weight approx. 160 g (0.35 lbs)
{} Items in curved brackets are optional extras for an additional price.
0.2 K (optional: increased accuracy through specic adjustment)
(dependent on temperature probe)
data logger settings
- Cyclic logger: Automatic recording of up to 16,200 values incl. time
- Cycle time: Selectable from 1 ... 3,600 seconds
supply}
Page 2 of 7 WIKA data sheet CT 51.01 ∙ 08/2015
Temperature probe model CTP62x0
Probe characteristics
Sensor type Immersion and penetration probe
Probe type Pt100 (4-wire)
Wetted parts Immersion probe and penetration probe: V4A
Connection to the CTH6200 4-pin mini DIN connector with PVC cable (approx. 1.0 m)
Measuring ranges
Immersion probe CTP6210 -50 ... +250 °C (-58 ... +482 °F)
Penetration probe CTP6290 -50 ... +250 °C (-58 ... +482 °F)
Sensor input
Current supply Slow measuring cycle: < 1.6 mA; fast: < 7.0 mA; low-power logger funct.: < 0.3 mA
Nominal temperature 25 °C (77 °F)
Temperature drift ≤ 0.002 °C/K
Approvals
Logo Description Country
EC declaration of conformity
EMC directive 2004/108/EC
EAC
Electromagnetic compatibility
Low voltage directive
GOST
Metrology, measurement technology
KazInMetr
Metrology, measurement technology
MtschS
Permission for commissioning
Uzstandard
Metrology, measurement technology
European Community
Eurasian Economic Community
Russia
Kazakhstan
Kazakhstan
Uzbekistan
Certicates
Certicate
Calibration
Recommended recalibration interval 1 year (dependent on conditions of use)
2) Calibration at 23 °C ±2 °C (73.4 °F ±3.6 °F) ambient temperature.
Approvals and certicates, see website
2)
Standard: 3.1 calibration certicate per DIN EN 10204 Option: DKD/DAkkS calibration certicate
Page 3 of 7WIKA data sheet CT 51.01 ∙ 08/2015