Calibration service companies and service industry
■
Measurement and control laboratories
■
Industry (laboratory, workshop and production)
■
Quality assurance
Special features
Calibration technology
WIKA data sheet CT 51.01
for further approvals
see page 3
■
Pt100 probes for -50 ... +250 °C (-58 ... +482 °F)
■
Accuracy: < 0.2 K (complete measuring chain)
■
Integrated data logger
■
GSoft data logger evaluation software available
■
Incl. calibration certicate
Description
Extensive application possibilities
The model CTH6200 hand-held thermometer is suitable for
all precise temperature measurement tasks and impresses
with its simple operation. Its design makes it especially
suitable for the commissioning, maintenance and service/
calibration of temperature instruments and equipment.
Functionality
The digital indicator is the central functional unit of the handheld temperature measuring instrument and provides the
option to connect Pt100 temperature probes of the CTP62x0
series.
Both immersion and penetration probes are available
from stock, along with an extensive range of accessories.
Customer-specic probes are available on request.
An integrated data logger and various other functions (such
as Min, Max, Hold or Alarm) ensure that the hand-held
thermometer can be used for many dierent applications.
Hand-held thermometer model CTH6200
Its simple operation, very handy design and low weight round
o the convenience of this hand-held temperature measuring
instrument.
Complete test and service cases
For the secure transportation and storage of the instrument
and accessories, a rugged service case is available as a
sensible addition. The service case can carry up to two
model CTH6200 digital indicators and several temperature
probes, or the case is available with a model CPH6200 handheld pressure indicator as well.
Certied accuracy
For each temperature probe, the accuracy for the complete
measuring chain is certied by a factory calibration
certicate which accompanies the instrument. On request,
we can provide a DKD/DAkkS calibration certicate for this
instrument.
Data sheets showing similar products:
Hand-held pressure indicator; model CPH6200; see data sheet CT 11.01
Page 1 of 7WIKA data sheet CT 51.01 ∙ 08/2015
Page 2
Specications
Model CTH6200
Hand-held thermometer model CTH6200 (complete measuring chain)
Measuring inputs1
Probe typePt100 (4-wire)
Measuring range-50 ... +250 °C (-58 ... +482 °F)
Resolution0.01 K up to 200 °C (392 °F); then 0.1 K
1)
Accuracy of the measuring chain
1) The accuracy applies for the respective indicator-probe combination following adjustment and calibration, that is the deviation between reference and CTH6200.
Digital indicator model CTH6200
Display
DisplayLarge LC display for display of values and additional information
Indication range-99.99 ... +99.99 °C or -199.9 ... +650.0 °C (-99.99 ... +99.99 °F or -199.9 ... +999.9 °F)
Temperature units°C, °F
Nominal temperature25 °C
Temperature drift≤ 0.002 °C/K
Functions
Functions via key pressMin/Max memory, Hold, Logger (Start/Stop)
Menu functionMin/Max alarm (audible/visual), automatic Power-O function [1 ... 120 min], zero point/slope input,
Data loggers- Individual value logger: Up to 99 recordings incl. time via key press
Voltage supply
Power supply9 V zinc-carbon battery (in scope of delivery), {alternative: 9 V rechargeable battery or mains power
Battery lifeapprox. 20 hours of operation
Current supplySlow measuring cycle: < 1.6 mA; fast: < 7.0 mA; low-power logger funct.: < 0.3 mA
Permissible ambient conditions
Operating temperature-10 ... +50 °C (14 ... 122 °F)
Storage temperature-25 ... +70 °C (-13 ... +158 °F)
Relative humidity0 ... 95 % r. h. (non-condensing)
Communication
Interface (serial)RS-232 or USB via special interface cable
{} Items in curved brackets are optional extras for an additional price.
0.2 K (optional: increased accuracy through specic adjustment)
(dependent on temperature probe)
data logger settings
- Cyclic logger: Automatic recording of up to 16,200 values incl. time
- Cycle time: Selectable from 1 ... 3,600 seconds
supply}
Page 2 of 7WIKA data sheet CT 51.01 ∙ 08/2015
Page 3
Temperature probe model CTP62x0
Probe characteristics
Sensor typeImmersion and penetration probe
Probe typePt100 (4-wire)
Wetted partsImmersion probe and penetration probe: V4A
Connection to the CTH62004-pin mini DIN connector with PVC cable (approx. 1.0 m)
Measuring ranges
Immersion probe CTP6210-50 ... +250 °C (-58 ... +482 °F)
Penetration probe CTP6290-50 ... +250 °C (-58 ... +482 °F)
Sensor input
Current supplySlow measuring cycle: < 1.6 mA; fast: < 7.0 mA; low-power logger funct.: < 0.3 mA
Nominal temperature25 °C (77 °F)
Temperature drift≤ 0.002 °C/K
Approvals
LogoDescriptionCountry
EC declaration of conformity
■
EMC directive 2004/108/EC
EAC
■
Electromagnetic compatibility
■
Low voltage directive
GOST
Metrology, measurement technology
KazInMetr
Metrology, measurement technology
MtschS
Permission for commissioning
Uzstandard
Metrology, measurement technology
European Community
Eurasian Economic Community
Russia
Kazakhstan
Kazakhstan
Uzbekistan
Certicates
Certicate
Calibration
Recommended recalibration interval1 year (dependent on conditions of use)
2) Calibration at 23 °C ±2 °C (73.4 °F ±3.6 °F) ambient temperature.
Approvals and certicates, see website
2)
Standard: 3.1 calibration certicate per DIN EN 10204
Option: DKD/DAkkS calibration certicate
Page 3 of 7WIKA data sheet CT 51.01 ∙ 08/2015
Page 4
36 (1,42)
71 (2,8)
Dimensions in mm (in)
Temperature probe model CTP62x0Digital indicator model CTH6200
829555-1
135 (5.31)L
142 (5,59)
26 (1,02)
L = 300 (11.81)
Electrical connections
Top viewSide view (left)
Ø D
CTP6210
Immersion
probe
Ø = 3 (0.12)
CTP62 90
Penetration
probe
Ø = 3 (0.12)
Interface con-
Connection for
temperature
probe
Page 4 of 7WIKA data sheet CT 51.01 ∙ 08/2015
nector
Connection for
mains power
supply
Page 5
Operating functions model CTH6200
Display
1
2
3
4
5
6
7
Oset: Zero oset is activated
1
Corr: Slope correction is activated
2
Main display: Current temperature display
3
Symbol: Indicates a low battery and other warnings
4
Secondary display: Display according to the selected
5
function keys Min, Max or Hold value
Logg: Arrow appears,logger functionwas selected via
6
menu
Arrow blinking: Automatic recording (Logg CYCL) active
AL: Indicates an alarm state
7
Keyboard
2 sec.
Instrument
on/o
Main menu
Enter conguration
2 sec.
2 sec.
MAX function
on/o
Delete MAX value
MIN function
on/o
Delete MIN value
Key only active within
the menu
Hold function
on/o
Store measured
value
2 sec.
Clear memory?
if:
A
B
= Press key
2 sec. = Press key for 2 seconds
For more information: see operating instructions
2 sec.
2 sec.
A = Logger functions deactivated
B = Logger function “Store measured value” activated via menu
C = Logger function “Store cyclically” activated via menu
Store cyclically
Stop storage?
Clear memory?
C
Page 5 of 7WIKA data sheet CT 51.01 ∙ 08/2015
Page 6
GSoft data logger evaluation software
The GSoft data logger evaluation software is used to
display the logger data from the model CTH6200 hand-held
thermometer on a PC in tabular form and as a chart.
■
Easy operation with self-explanatory toolbars
■
Data from the pressure and temperature hand-helds can
be displayed in a single chart (two separate y-axes)
■
Chart oers a zoom function
■
Operation of the logger function via PC (remote control)
■
Data can be exported (Excel®, etc.)
■
Languages: German, English, French, Spanish and Czech
System requirements
■
IBM compatible PC (Pentium®)
■
At least 20 MB free hard disc space
■
CD-ROM drive
■
At least 32 MB RAM
■
Windows® operating system 95, 98, NT 4.0 (with Service
Pack 3.0 or higher), 2000, XP, Vista or 7
■
Mouse
■
One free serial interface or USB connection (via interface
cable)
Data export e.g. in
an Excel® le
Windows® is a registered trademark of Microsoft Corporation in the United States and other countries.
Easy operation with self-explanatory icon keys
Charts toolbarMain toolbar
1122334455678
1. File functions: open, save, print
2. Logger functions: Start communication, start logger, stop,
read data
3. Data display: Create chart
4. Interface conguration
5. Program information
1. Settings: Grid and colour settings, manual zooms
2. Zoom: All, left or right y-axis (via mouse), back
3. Rename chart
4. Cursor on/o (info footer)
5. Legend on/o
6. (Measuring point) Symbols on/o
7. Measurement series (add/delete)
8. Comments on measuring points (add/delete)
Page 6 of 7WIKA data sheet CT 51.01 ∙ 08/2015
Page 7
Scope of delivery
■
Hand-held thermometer model CTH6200 incl. 9 V battery
■
3.1 calibration certicate per DIN EN 10204
■
Choice of temperature probes
Options
■
Immersion probe
■
Penetration probe
■
Other probes on request
■
DKD/DAkkS calibration certicate
Accessories
Test case
■
Robust transport case
■
Various test and service cases
Software
■
GSoft data logger evaluation software
■
Appropriate interface cable
Voltage supply
■
Power supply unit
■
Rechargeable battery and battery charger
Example of a service case (temperature and pressure)
08/2015 EN
Ordering information
Model / Reference temperature probe with handle / Accuracy / Power supply unit / Rechargeable battery and charger /
Software / Interface cable / Transport case / Calibration (with factory calibration certicate for 1 probe) / Additional ordering
The specications given in this document represent the state of engineering at the time of publishing.
We reserve the right to make modications to the specications and materials.