WIKA CTD9100-650, CTM9100-150 Operating Manual

Operating instructions Betriebsanleitung Manuale d'uso
Dry-well temperature calibrator, micro calibration bath, multi-function calibrator, models CTD9100, CTB9100, CTM9100
Calibratore di temperatura a secco, microbagno di calibrazione, calibratore multifunzione, modelli CTD9100, CTB9100, CTM9100
EN
DE
IT
Dry-well temperature calibrator,
CTD9100 series
Micro calibration bath,
CTB9100 series
Multi-function calibrator,
model CTM9100
Operating instructions models CTD9100, CTB9100,
EN
CTM9100
Page 3 - 50
Betriebsanleitung Typen CTD9100, CTB9100,
Seite 51 - 98
DE
CTM9100
IT
Manuale d’uso modelli CTD9100, CTB9100, CTM9100 Pagina 99 - 146
Further languages can be found at www.wika.com.
© 07/2012 WIKA Alexander Wiegand SE & Co. KG All rights reserved. / Alle Rechte vorbehalten.
®
WIKA
is a registered trademark in various countries.
®
WIKA
ist eine geschützte Marke in verschiedenen Ländern.
Prior to starting any work, read the operating instructions! Keep for later use!
Vor Beginn aller Arbeiten Betriebsanleitung lesen! Zum späteren Gebrauch aufbewahren!
Prima di iniziare ad utilizzare lo strumento, leggere il manuale d‘uso! Conservare per future consultazioni!
2 WIKA operating instructions, models CTD9100, CTB9100, CTM9100
11263911.05 03/2019 EN/DE/IT
Contents
Contents
1. General information 5
2. Short overview 5
2.1 Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
3. Safety 6
3.1 Explanation of symbols . . . . . . . . . . . . . . . . . . . . . . . . 6
3.2 Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
3.3 Scope of delivery . . . . . . . . . . . . . . . . . . . . . . . . . . 6
3.4 Intended use . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
3.5 Improper use. . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
3.6 Safety instructions for using calibration liquids . . . . . . . . . . . . . . . . 7
3.7 Personnel qualification . . . . . . . . . . . . . . . . . . . . . . . . 8
3.8 Personal protective equipment. . . . . . . . . . . . . . . . . . . . . . 8
3.9 Labelling, safety marks . . . . . . . . . . . . . . . . . . . . . . . . 9
4. Design and function 10
4.1 Overview of the different instrument models . . . . . . . . . . . . . . . . 10
4.2 Isometric views of the CTD9100 series dry-well temperature calibrators. . . . . . . 11
4.3 Isometric views of the CTB9100 series micro calibration baths . . . . . . . . . . 12
4.4 Isometric views of the model CTM9100-150 multi-function calibrator. . . . . . . . 13
4.5 Description of the operating elements . . . . . . . . . . . . . . . . . . 14
4.6 Data interface . . . . . . . . . . . . . . . . . . . . . . . . . . 15
4.7 Interface protocol . . . . . . . . . . . . . . . . . . . . . . . . . 15
4.8 Protective earth monitoring . . . . . . . . . . . . . . . . . . . . . . 15
5. Transport, packaging and storage 16
5.1 Transport . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
5.2 Packaging and storage . . . . . . . . . . . . . . . . . . . . . . . 16
6. Commissioning, operation 16
6.1 Operating position . . . . . . . . . . . . . . . . . . . . . . . . . 16
6.2 Inserts with metal block . . . . . . . . . . . . . . . . . . . . . . . 17
6.3 Preparation of the micro calibration bath . . . . . . . . . . . . . . . . . 17
6.3.1 Properties of the calibration liquid . . . . . . . . . . . . . . . . . 17
6.3.2 Filling the micro calibration bath . . . . . . . . . . . . . . . . . . 18
6.3.3 Operation of the magnetic stirrer . . . . . . . . . . . . . . . . . . 18
6.3.4 Insert for liquids . . . . . . . . . . . . . . . . . . . . . . . 18
6.4 Surface insert (only CTM9100-150) . . . . . . . . . . . . . . . . . . . 19
6.5 Infrared insert (only CTM9100-150) . . . . . . . . . . . . . . . . . . . 19
6.6 Testing of temperature sensors . . . . . . . . . . . . . . . . . . . . 20
6.7 Start-up procedure. . . . . . . . . . . . . . . . . . . . . . . . . 20
6.8 Switching on the calibrator/micro calibration bath . . . . . . . . . . . . . . 20
6.9 Reference and set temperature display . . . . . . . . . . . . . . . . . . 20
6.10 Control of the reference temperature . . . . . . . . . . . . . . . . . . 20
7. Operating elements of the calibrator/micro calibration bath 21
7.1 Selection of the operating modes for the CTM9100-150 . . . . . . . . . . . . 21
7.2 Method of operation in calibration mode within individual operating modes. . . . . . 21
7.3 Calibration (calibration mode) . . . . . . . . . . . . . . . . . . . . . 22
7.4 Setting a temporary set temperature (set-point mode) . . . . . . . . . . . . . 22
11263911.05 03/2019 EN/DE/IT
EN
3WIKA operating instructions, models CTD9100, CTB9100, CTM9100
Contents
7.5 Programming (Main menu) . . . . . . . . . . . . . . . . . . . . . . 23
7.5.1 Menu structure, parameter levels. . . . . . . . . . . . . . . . . . 24
7.5.2 Switching off automatic control . . . . . . . . . . . . . . . . . . 25
7.5.3 Switching on automatic control . . . . . . . . . . . . . . . . . . 25
EN
8. Cooling down the metal blocks or micro calibration baths 32
9. Faults 33
10. Maintenance, cleaning and recalibration 34
11. Dismounting, return and disposal 35
12. Specifications 36
13. Accessories 41
14. Heating and cooling times 42
7.5.4 Switching on manual control . . . . . . . . . . . . . . . . . . . 26
7.5.5 Switching off manual control . . . . . . . . . . . . . . . . . . . 26
7.5.6 Setting and storing fixed set temperatures . . . . . . . . . . . . . . 27
7.5.7 Recalling the stored set temperatures . . . . . . . . . . . . . . . . 28
7.5.8 Setting the ramp control and a temperature profile . . . . . . . . . . . . 28
10.1 Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . 34
10.2 Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
10.2.1 Cleaning calibrators with inserts . . . . . . . . . . . . . . . . . . 34
10.2.2 Cleaning fan guards . . . . . . . . . . . . . . . . . . . . . . 34
10.2.3 Cleaning the micro calibration bath . . . . . . . . . . . . . . . . . 34
10.2.4 External cleaning . . . . . . . . . . . . . . . . . . . . . . . 34
10.3 Recalibration . . . . . . . . . . . . . . . . . . . . . . . . . . 34
11.1 Dismounting . . . . . . . . . . . . . . . . . . . . . . . . . . 35
11.2 Return . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
11.3 Disposal . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
12.1 Dry-well temperature calibrator, CTD9100 series. . . . . . . . . . . . . . 36
12.2 Micro calibration bath, CTB9100 series . . . . . . . . . . . . . . . . . 38
12.3 Multi-function calibrator, model CTM9100-150 . . . . . . . . . . . . . . 40
13.1 CTD9100 series . . . . . . . . . . . . . . . . . . . . . . . . . 41
13.2 CTB9100 series . . . . . . . . . . . . . . . . . . . . . . . . . 41
13.3 Model CTM9100-150 . . . . . . . . . . . . . . . . . . . . . . . 41
14.1 Dry-well temperature calibrator, model CTD9100-COOL . . . . . . . . . . . 42
14.2 Dry-well temperature calibrator, model CTD9100-165 . . . . . . . . . . . . 42
14.3 Dry-well temperature calibrator, model CTD9100-165-X . . . . . . . . . . . 43
14.4 Dry-well temperature calibrator, model CTD9100-450 . . . . . . . . . . . . 43
14.5 Dry-well temperature calibrator, model CTD9100-650 . . . . . . . . . . . . 44
14.6 Micro calibration bath, model CTB9100-165 . . . . . . . . . . . . . . . 45
14.7 Micro calibration bath, model CTB9100-225 . . . . . . . . . . . . . . . 47
14.8 Model CTM9100-150 as a micro calibration bath. . . . . . . . . . . . . . 48
14.9 Model CTM9100-150 as dry-well temperature calibrator . . . . . . . . . . . 49
14.10 Model CTM9100-150 as an infrared black body source . . . . . . . . . . . 49
14.11 Model CTM9100-150 as surface temperature calibrator . . . . . . . . . . . 50
Declarations of conformity can be found online at www.wika.com.
4 WIKA operating instructions, models CTD9100, CTB9100, CTM9100
11263911.05 03/2019 EN/DE/IT
1. General information / 2. Short overview
1. General information
The calibrator or the micro calibration bath described in the operating instructions has been designed and manufactured using state-of-the-art technology. All components are subject to stringent quality and environmental criteria during production. Our management systems are certified to ISO 9001 and ISO 14001.
These operating instructions contain important information on handling the instrument. Working safely requires that all safety instructions and work instructions are observed.
Observe the relevant local accident prevention regulations and general safety regulations for the instrument’s range of use.
The operating instructions are part of the product and must be kept in the immediate vicinity of the instrument and readily accessible to skilled personnel at any time. Pass the operating instructions onto the next operator or owner of the instrument.
Skilled personnel must have carefully read and understood the operating instructions prior to beginning any work.
The general terms and conditions contained in the sales documentation shall apply.
Subject to technical modifications.
Factory calibrations / DKD/DAkkS calibrations are carried out in accordance with international standards.
Further information:
- Internet address: www.wika.de / www.wika.com
- Relevant data sheet: CT 41.28 CT 41.40 CT 46.30
- Application consultant: Tel.: +49 9372 132-0 Fax: +49 9372 132-406 info@wika.de
EN
2. Short overview
2.1 Overview
3
2
1
1
Temperature block
2
Controller
3
Handle
11263911.05 03/2019 EN/DE/IT
5WIKA operating instructions, models CTD9100, CTB9100, CTM9100
2. Short overview / 3. Safety
2.2 Description
The calibrator or micro calibration bath is a portable unit, for service functions and also for industrial and laboratory tasks. The temperature calibrators and micro calibration baths from WIKA are provided for the calibration of thermometers, temperature switches/thermostats, resistance thermometers
EN
and thermocouples. The operational safety of the delivered instruments is only assured if the equipment is employed for its intended use (verification of temperature sensors). The given limit values should never be exceeded (see chapter 12 “Specifications”). The appropriate instrument should be selected depending on the application; it should be connected correctly, tests carried out and all components serviced. The instrument is manufactured in different versions. The individual version for any single unit can be found on the nameplate on the calibrator/micro calibration bath.
2.3 Scope of delivery
The calibrators/micro calibration baths are delivered in special protective packaging. The packaging must be set aside so that the calibrator or the micro calibration bath can be sent safely back to the manufacturer for recalibration or repair.
Scope of delivery for the model CTD9100 dry-well temperature calibrator
Calibrator
Insert replacement tools
Standard bored insert
Power cord
Calibration certificate
Operating instructions
Scope of delivery for the model CTB9100 micro calibration bath
Micro calibration bath
Transport cover
Sensor basket
Magnetic stirrer
Magnetic lifter
Power cord
Calibration certificate
Operating instructions
Scope of delivery for the model CTM9100 multi-function calibrator
Multi-function calibrator
Insert replacement tools (standard and surface)
Standard bored insert
Transport cover
Sensor basket
Magnetic stirrer
Magnetic lifter
Infrared insert
Surface insert
External reference sensor
Power cord
Calibration certificate
Operating instructions
Cross-check scope of delivery with delivery note.
3. Safety
3.1 Explanation of symbols
DANGER!
... indicates a directly dangerous situation resulting in serious injury or death, if not avoided.
WARNING!
... indicates a potentially dangerous situation that can result in serious injury or death, if not avoided.
CAUTION!
... indicates a potentially dangerous situation that can result in light injuries or damage to equipment or the environment, if not avoided.
6 WIKA operating instructions, models CTD9100, CTB9100, CTM9100
DANGER!
... identifies hazards caused by electric power. Should the safety instructions not be observed, there is a risk of serious or fatal injury.
WARNING!
... indicates a potentially dangerous situation that can result in burns, caused by hot surfaces or liquids, if not avoided.
Information
... points out useful tips, recommendations and information for efficient and trouble-free operation.
11263911.05 03/2019 EN/DE/IT
3. Safety
3.2 Intended use
The calibrator or micro calibration bath is a portable unit, for service functions and also for industrial and laboratory tasks. The temperature calibrators and micro calibration baths from WIKA are provided for the calibration of thermometers, temperature switches/thermostats, resistance thermometers and thermocouples.
This instrument is not permitted to be used in hazardous areas!
The instrument has been designed and built solely for the intended use described here, and may only be used accordingly.
The technical specifications contained in these operating instructions must be observed. Improper handling or operation of the instrument outside of its technical specifications requires the instrument to be taken out of service immediately and inspected by an authorised WIKA service engineer.
Handle electronic precision measuring instruments with the required care (protect from humidity, impacts, strong magnetic fields, static electricity and extreme temperatures, do not insert any objects into the instrument or its openings). Plugs and sockets must be protected from contamination.
The manufacturer shall not be liable for claims of any type based on operation contrary to the intended use.
3.3 Improper use
WARNING! Injuries through improper use
Improper use of the instrument can lead to hazardous situations and injuries.
Refrain from unauthorised modifications to the instrument.
Do not use the instrument within hazardous areas.
Only ever use the supplied mains cable.
Any use beyond or different to the intended use is considered as improper use.
Special hazards
DANGER! Danger of death caused by electric current
Upon contact with live parts, there is a direct danger of death.
The instrument may only be installed and mounted by skilled personnel.
Before replacing the fuse, cleaning, maintenance/servicing and in the event of danger, the calibrator or the micro calibration bath must be disconnected by unplugging the mains cable from the power supply outlet.
The mains socket must be freely accessible at all times!
Thermal fuse
For safety, the calibrator or the micro calibration bath is fitted with an independently-operating temperature fuse, which cuts out the heater power supply if the temperature inside the enclosure is too high. Once the metal block and the liquid bath have cooled down, send in the calibrator or the micro calibration bath to WIKA for examination.
The calibrator and the micro calibration bath have been designed as measuring and control instruments. With any operation of the calibrator/ micro calibration bath not expressly provided for in these operating instructions, additional protective measures must be taken.
Since a malfunction of the calibrator/micro calibration bath can cause personal injury or damage to property, the equipment must be protected by additional electro-mechanical safeguards.
3.4 Safety instructions for using calibration liquids
Calibration liquid, water
Only use distilled water, otherwise the calibrator tank can become strongly scaled and soiled. Collect spilled liquids immediately and dispose of them properly.
EN
11263911.05 03/2019 EN/DE/IT
7WIKA operating instructions, models CTD9100, CTB9100, CTM9100
3. Safety
Calibration liquid, silicone oil
WARNING! Hazardous substances silicone oils
Improper handling can result to poisoning or
EN
personal injury.
Wear the requisite protective equipment (see chapter 3.6 “Personal protective equipment”).
Observe the information in the material safety data sheet for the corresponding medium.
Before working with silicone oil, read the material data safety sheet. The current material safety data sheet can be found at www.wika.de on the respective product page.
Only use the silicone oil included within the scope of supply or specified in these operating instructions.
When working with silicone oil, sufficient ventilation should be ensured in the room since it can give off pollutants.
Since silicone oil is hygroscopic, after use, always close the calibration bath using the transport cover.
Before transporting with silicone oil, allow the calibrator or micro calibration bath to cool. The transport cover is fitted with a safety valve. If the micro calibration bath is closed when it is still warm, excessive pressure can build up. In order to avoid overpressuring, which could lead to damage to the liquid bath, the safety valve is activated at a pressure of approx.
2.5 bar [36 psi]. This can lead to hot vapour escaping.
WARNING! Risk of burns!
Touching the hot metal block or micro calibration bath, the hot bath liquids or the test item can lead to acute burns.
Before transporting or touching the metal block/liquid bath, make sure that it has cooled down sufficiently.
3.5 Personnel qualification
WARNING! Risk of injury should qualification be insufficient!
Improper handling can result in considerable injury and damage to equipment. The activities described in these operating instructions may only be carried out by skilled personnel who have the qualifications described below.
Skilled personnel
Skilled personnel, authorised by the operator, are understood to be personnel who, based on their technical training, knowledge of measurement and control technology and on their experience and knowledge of country-specific regulations, current standards and directives, are capable of carrying out the work described and independently recognising potential hazards.
Special operating conditions require further appropriate knowledge, e.g. of aggressive media.
3.6 Personal protective equipment
The personal protective equipment is designed to protect the skilled personnel from hazards that could impair their safety or health during work. When carrying out the various tasks on and with the instrument, the skilled personnel must wear personal protective equipment.
Follow the instructions, displayed in the work area, regarding personal protective equipment!
The required personal protective equipment must be provided by the operating company.
Wear safety goggles!
Protect eyes from flying particles and liquid splashes.
Collect spilled liquids immediately and dispose of them properly.
Wear safety goggles!
Do not allow silicone oil to come into contact with the eyes.
Wear protective gloves!
Protect hands from friction, abrasion, cuts or deep injuries and also from contact with hot surfaces and aggressive media.
8 WIKA operating instructions, models CTD9100, CTB9100, CTM9100
Wear protective gloves!
Protect hands from friction, abrasion, cuts or deep injuries and also from contact with hot surfaces and aggressive media.
11263911.05 03/2019 EN/DE/IT
3. Safety
3.7 Labelling, safety marks
Product label (example)
The product label is fixed on the rear of the instrument.
8
7
6
EN
5
4
3
1
Year of manufacture
2
Fuse
3
Notes regarding the safety data sheet
4
Serial no.
5
Power supply
6
Temperature range
7
Model designation
8
Instrument designation
Explanation of symbols
Before mounting and commissioning the instrument, ensure you read the operating instructions!
1
2
Do not dispose of with household waste. Ensure a proper disposal in accordance with national regulations.
11263911.05 03/2019 EN/DE/IT
9WIKA operating instructions, models CTD9100, CTB9100, CTM9100
4. Design and function
4. Design and function
4.1 Overview of the different instrument models
Temperature calibrators
EN
CTD9100-COOL (cooling and heating)
CTD9100-165 (cooling and heating)
CTD9100-450 (heating)
CTD9100-650 (heating)
Micro calibration baths
CTB9100-165 (cooling and heating)
CTB9100-225 (heating)
Multi-function calibrator
CTM9100-150 (cooling and heating)
The calibrator and micro calibration bath consist of a robust, grey-blue-painted steel enclosure, with a carrying handle on top.
The rear part of the enclosure includes a metal block or liquid bath with an opening for the test item, accessible from the top. The metal block/liquid bath incorporate the heating or cooling elements and the temperature sensor for determining the reference temperature. The metal block and the liquid bath are thermally insulated.
The front part contains the complete electronic unit for controlling the reference temperature. Solid state relays (SSR) are used to control the heating and cooling elements. On the front panel is the controller, which is fitted with a 7-segment LED (2 rows of 4 digits) for the reference and set temperature. The micro calibration bath has an additional knob for controlling the stirring speed.
Temperature calibrator, model CTD9100-165 Temperature calibrator, model CTD9100-650
Multi-function calibrator, model CTM9100-150Micro calibration bath, model CTB9100-165
10 WIKA operating instructions, models CTD9100, CTB9100, CTM9100
11263911.05 03/2019 EN/DE/IT
4. Design and function
4.2 Isometric views of the CTD9100 series dry-well temperature calibrators
1
Front and top of the model CTD9100
On the top of the dry-well temperature calibrator, you will find
3
2
the dry well access opening for inserting the inserts.
CTD9100-COOL: Ø 28 x 150 mm [Ø 1.10 x 5.91 in]
CTD9100-165: Ø 28 x 150 mm [Ø 1.10 x 5.91 in]
CTD9100-165-X: Ø 60 x 150 mm [Ø 2.36 x 5.91 in]
CTD9100-450: Ø 60 x 150 mm [Ø 2.36 x 5.91 in]
CTD9100-650: Ø 28 x 150 mm [Ø 1.10 x 5.91 in]
EN
The controller, with display and controls, is located on the front of the calibrator.
Rear of the instrument
On the rear of the equipment is the product label with the key information about the particular model.
CTD9100-COOL: -55 ... +200 °C [-67 ... +392 °F]
CTD9100-165: -35 ... +165 °C [-31 ... +329 °F]
CTD9100-450: 40 ... 450 °C [104 ... 842 °F]
CTD9100-650: 40 ... 650 °C [104 ... 1.202 °F]
1
The correct mains voltage and frequency is also given.
AC 100 ... 240 V, 50 ... 60 Hz
AC 115 V, 50 ... 60 Hz
AC 230 V, 50 ... 60 Hz
8
11263911.05 03/2019 EN/DE/IT
4
Furthermore, the individual serial number (e.g. S/N 550 33 44) as well as the mains voltage and the
5
fuse rating is given.
You will also find the connector for the RS-485 interface here.
Underside of the instrument
On the underside of the instrument are the mains connection socket and the power switch with its fuse-holder. These are located in the centre, at the front. Furthermore, depending on the model, there are one or two air intakes located on the underside of the instrument.
The air intakes must not be obstructed in any way.
1
Temperature block
2
Controller
3
Handle
4
Product label
5
RS-485 interface
6
Fan 1
6
7
7
Fan 2
8
Mains connection socket with power switch
11WIKA operating instructions, models CTD9100, CTB9100, CTM9100
4. Design and function
4.3 Isometric views of the CTB9100 series micro calibration baths
EN
1
3
2
9
On the top of the micro calibration bath are the block opening for filling Ø 60 x 150 mm [Ø 2.36 x 5.91 in].
The controller, with display and controls, is located on the front of the calibrator.
Rear of the instrument
On the rear of the equipment is the product label with the key information about the particular model. Furthermore, the individual serial number (e.g. S/N 550 33 44) as well as the mains voltage and the fuse rating is given.
You will also find the connector for the RS-485 interface here.
Underside of the instrument
On the underside of the instrument are the mains connection socket and the power switch with its fuse-holder.
Front and top of the model CTB9100
1
These are located in the centre, at the front. Furthermore, depending on the model, there are one or two air intakes located on the underside of the instrument.
The air intakes must not be obstructed in any way.
4
5
8
1
Temperature block
2
Controller
3
Handle
4
Product label
5
RS-485 interface
6
Fan 1
7
Fan 2
8
Mains connection socket with power switch
9
Controller for the stirrer
6
7
12 WIKA operating instructions, models CTD9100, CTB9100, CTM9100
11263911.05 03/2019 EN/DE/IT
4. Design and function
4.4 Isometric views of the model CTM9100-150 multi-function calibrator
Front and top of the model CTM9100-150
1
3
11
On the top of the multi-function calibrator, you will find the dry well access opening for inserting the different inserts or filling Ø 60 x 150 mm [Ø 2.36 x 5.91 in].
EN
10
The controller, with display and control elements, is located
2
9
on the front of the calibrator. In addition, the control elements for the external reference are placed on the front.
Rear of the instrument
On the rear of the equipment is the product label with the key information about the particular model. Furthermore, the individual serial number (e.g. S/N 550 33 44) as well as the mains voltage and the fuse rating is given.
You will also find the connector for the RS-485 interface here.
Underside of the instrument
1
On the underside of the instrument are the mains connection socket and the power switch with its fuse-holder. These are located in the centre, at the front. Furthermore, there are two air intakes located on the underside of the instrument.
The air intakes must not be obstructed in any way.
4
5
8
6
7
1
Temperature block
2
Controller
3
Handle
4
Product label
5
RS-485 interface
6
Fan 1
7
Fan 2
8
Mains connection socket with power switch
9
Controller for the stirrer
10
Switch for ext. reference
11
Socket for ext. reference
11263911.05 03/2019 EN/DE/IT
13WIKA operating instructions, models CTD9100, CTB9100, CTM9100
4. Design and function
4.5 Description of the operating elements
Front of the controller
EN
7 8 9
6
5
Overview of the operating elements on the front of the controller
Button [▲]
1
Increasing value being set
Selection of individual menu item
Go back up one menu level
Button [▼]
2
Reduction of value being set
Selection of individual menu item
Go back up one menu level
Button [P]
3
Access to the set temperature adjustment
Access to the menu items and parameters
Input confirmation
LED SET
4
Flashing signals access to the individual menu items and parameters
SV display
5
Set temperature display
Display of particular parameters in the individual
7
8
modes and the menu options
PV display
6
Indication of the current reference temperature
Display of the individual modes, menu items and the
9
parameters
1
2
34
LED OUT 2 a) Heating instrument
Signals the status of the outputs for the fan control
If LED OUT 2 is on, the fan is running at higher speed
If LED OUT 2 is not on, the fan is running at lower speed
b) Heating and cooling instrument
Signals the status of the outputs for the temperature control
If LED OUT 1 is on, the calibrator or the micro calibration bath is cooling
If LED OUT 1 is not on, the calibrator or the micro calibration bath is not cooling
LED OUT 1
Signals the status of the outputs for the temperature control
If LED OUT 1 is on, the calibrator or the micro calibration bath is heating
If LED OUT 1 is not on, the calibrator or the micro calibration bath is not heating
Button [U]
Recall the stored set temperatures
Further definitions
[XXX] Press XXX button
XXX Menu XXX will be displayed
14 WIKA operating instructions, models CTD9100, CTB9100, CTM9100
11263911.05 03/2019 EN/DE/IT
4. Design and function
4.6 Data interface
The instruments are fitted with an RS-485 communications interface. Via this interface one can connect to a PC, level converter or a network. The software protocol used is a MODBUS-RTU protocol, which is used in may monitoring programs available on the market. The transmission speed (baud rate) is factory set to 9,600 baud. Upon request, other data transmission rates are possible. The 5-pin socket has two pins, A and B, that you connect to the corresponding connections on the PC, level converter or network.
Plan view of the 5-pin panel socket
4.7 Interface protocol
The interface protocol is available on request for delivery as a specific additional document.
4.8 Protective earth monitoring
The calibrator is fitted with a protective earth monitor to keep a check on the base isolation of the heater. The monitoring unit works independently of the normal controls and switches off the heating power supply as soon as the calibrator no longer has a connection to the protective earth system of the isolation. Once reconnected to the protective earth system, the monitoring unit automatically switches the power supply back on to the heating circuit.
EN
To connect to a PC, the RS-485 signal must be externally converted to an RS-232 or USB signal. The appropriate converter, including drivers, is available as an option. The computer records all operational data and allows programming of all calibrator configuration parameters.
The minimum requirements for operation with a USB converter are:
IBM compatible PC
Installed operating system, Microsoft® Windows® 98 SE, ME, 2000, XP (Home or Prof.) or 7
A USB interface (USB 1.1 or USB 2.0)
A network connection enables the connection of up to 32 calibrators/micro-baths on the same network. To connect to a network, some factory setting is required. For this, please contact the supplier or WIKA directly.
For access to the programming via the keypad while the serial interface is communicating, the message “buSy” appears on the display, and therefore indicates that it is in an “occupied” state.
11263911.05 03/2019 EN/DE/IT
15WIKA operating instructions, models CTD9100, CTB9100, CTM9100
4. Design and function / 5. Transport, packaging and storage
5. Transport, packaging and storage
5.1 Transport
Check the calibrator or the micro calibration bath for any damage that may have been caused by transport.
EN
Obvious damage must be reported immediately.
CAUTION! Damage through improper transport
With improper transport, a high level of damage to property can occur.
When unloading packed goods upon delivery as well as during internal transport, proceed carefully and observe the symbols on the packaging.
With internal transport, observe the instructions in chapter 5.2 “Packaging and storage”.
If the instrument is transported from a cold into a warm environment, the formation of condensation may result in instrument malfunction. Before putting it back into operation, wait for the instrument temperature and the room temperature to equalise.
5.2 Packaging and storage
Do not remove packaging until just before mounting. Keep the packaging as it will provide optimum protection during transport (e.g. change in installation site, sending for repair).
Permissible conditions at the place of storage:
Storage temperature: -10 ... +60 °C [14 ... 140 °F]
Humidity: 30 ... 95 % relative humidity (no condensation)
Avoid exposure to the following factors:
Direct sunlight or proximity to hot objects
Mechanical vibration, mechanical shock (putting it down hard)
Soot, vapour, dust and corrosive gases
Potentially explosive environments, flammable atmospheres
Store the calibrator or the micro calibration bath in its original packaging in a location that fulfils the conditions listed above.
6. Commissioning, operation
Personnel: Skilled personnel Protective equipment: Protective gloves and safety goggles
Only use original parts (see chapter 13 “Accessories”).
WARNING! Physical injuries and damage to property and the environment caused by hazardous media
Upon contact with hazardous media (e.g. oxygen, acetylene, flammable or toxic substances), harmful media (e.g. corrosive, toxic, carcinogenic, radioactive), and also with refrigeration plants and compressors, there is a danger of physical injuries and damage to property and the environment. Should a failure occur, aggressive media and/ or with high temperature may be present at the instrument.
For these media, in addition to all standard regulations, the appropriate existing codes or regulations must also be followed.
Wear the requisite protective equipment (see chapter 3.6 “Personal protective equipment”).
6.1 Operating position
The operating position of the calibrator or the micro calibration bath is in the vertical orientation, since this guarantees an optimal temperature distribution in the metal block or liquid bath.
Voltage supply
DANGER! Danger to life caused by electric current
Upon contact with live parts, there is a direct danger to life.
Only ever use the supplied mains cable.
The voltage supply of the instrument is made via the power cable. This is included in the scope of delivery.
16 WIKA operating instructions, models CTD9100, CTB9100, CTM9100
11263911.05 03/2019 EN/DE/IT
6. Commissioning, operation
6.2 Inserts with metal block
In order to achieve the greatest possible accuracy, the use of exactly matched inserts is necessary. For this, diameter of the specimen must be accurately determined. The bore for the insert is obtained by adding +0.5 mm [+0.02 in].
Inserts
Following use, the inserts should be removed using the replacement tools and then the sleeve and block should be cleaned. This prevents the sleeves becoming jammed in the heating block.
Recommended calibration liquids for the different temperature ranges:
Water as a calibration liquid
Only use distilled or demineralised water, otherwise the calibrator tank can become strongly scaled and soiled.
Silicone oil as a calibration liquid
Only use the silicone oil recommended here.
When working with silicone oil, sufficient ventilation should be ensured in the room since it can give off pollutants.
Since silicone oil is hygroscopic, after use, always close the calibration bath using the transport cover.
Only use clean calibration liquids. The checking of temperature sensors and other temperature measuring devices may lead to the contamination of the calibration liquid. These contaminants, through the rotational movement of the magnetic stirrer, may cause an abrasive effect on the bottom tank.
Wear safety goggles!
Do not allow silicone oil to come into contact with the eyes.
EN
6.3 Preparation of the micro calibration bath
In order to achieve the highest possible accuracy for a micro calibration bath, it should be filled with an appropriate calibration liquid.
6.3.1 Properties of the calibration liquid
Different calibration liquids, due to their specific properties, deliver varying calibration results. A compensation to the calibration liquid used in each case should, if necessary, be
Clean the tank
Before calibration, clean the sensors
Replace the worn magnetic stirrer
Replace the contaminated, cloudy calibration liquid
Wear protective gloves!
Protect hands from friction, abrasion, cuts or deep injuries and also from contact with hot surfaces and aggressive media.
performed by the manufacturer at the factory.
Medium Calibration range Flash point
Distilled water 5 ... 90 °C [51 ... 194 °F] none Dow Corning 200 Fluid with 5 CS -40 ... +123 °C [-40 ... +253 °F] 133 °C [271 °F] Dow Corning 200 Fluid with 10 CS -35 ... +155 °C [-31 ... +311 °F] 163 °C [325 °F] Dow Corning 200 Fluid with 20 CS 7 ... 220 °C [45 ... 428 °F] 232 °C [450 °F] Dow Corning 200 Fluid with 50 CS 25 ... 270 °C [77 ... 518 °F] 280 °C [536 °F]
11263911.05 03/2019 EN/DE/IT
17WIKA operating instructions, models CTD9100, CTB9100, CTM9100
6. Commissioning, operation
6.3.2 Filling the micro calibration bath
1. Firstly, remove the transport cover.
2. Place the test item in the sensor basket.
3. Fill the tank with calibration liquid.
Following maximum filling heights are recommended:
EN
Calibrator model max. filling height
CTB9100-165 / CTM9100-150 130 mm [5.12 in] CTB9100-165 / CTM9100-150
with removable insert CTB9100-225 123 mm [4.84 in] CTB9100-225 with removable insert 105 mm [4.13 in] CTB9100-225-X 115 mm [4.53 in] CTB9100-225-X with removable insert 95 mm [3.74 in]
Following points should be observed at the maximum filling heights:
Measured from the bottom of the sensor basket
Non loaded tank
WIKA standard filling liquid
Pre-filling ex-works with the optimal height
110 mm [4.33 in]
The magnetic stirrer is a consumable.
Liquid bath
6.3.4 Insert for liquids
The insert for liquids consists of:
Insert with leak-proof cover
Sensor basket
Magnetic stirrer and lifter
Replacement tools
The transport cover is fitted with a safety valve. If the micro calibration bath is closed when it is still warm, excessive pressure can build up. In order to avoid overpressuring, which could lead to damage to the liquid bath, the safety valve is activated at a pressure of approx. 2.5 bar [36 psi]. This can lead to hot vapour escaping.
6.3.3 Operation of the magnetic stirrer
The greatest possible homogeneity is achieved by the agitation of the calibration liquid via a magnetic stirrer.
Set the stirring speed to the maximum possible. Turning the wheel upwards will increase the speed, turning it downwards slows down the stirring speed.
Front of the controller with stirrer-speed dial
Insert for liquids
WARNING! Risk of burns!
Hot liquids can lead to acute burns
The insert for liquids must only be removed if the calibrator is used at room temperature.
The fill level must be matched the relevant medium and the temperature.
If the insert for liquids is ordered with a new model CTB9100 micro calibration bath or a new model CTM9100 multi-function calibrator, the instrument will have been matched to the insert for liquids. Should the insert for liquids be ordered afterwards, a re-adjustment of the instrument at the customer’s will be needed.
18 WIKA operating instructions, models CTD9100, CTB9100, CTM9100
11263911.05 03/2019 EN/DE/IT
6. Commissioning, operation
6.4 Surface insert (only CTM9100-150)
The operation of the calibrator with the surface insert is suited to a simple and the most accurate calibration possible of surface probes.
Fit the insert, which is hollow and long from the bottom, into the block using a special replacement tool. The sleeve also has three bores (1 x 3 mm, 1 x 3.1 mm and 1 x 4 mm [1 x 0.12 in, 1 x 0.12 in and 1 x 0.16 in]) directly under the surface, so that the correct surface temperature can be controlled at all times.
Following use, the inserts should be removed using the specific replacement tools and then the sleeve and block should be cleaned. This prevents the sleeves becoming jammed in the heating block.
The calibration of surface thermometers is very difficult and not fully defined. Thermometers mounted on surfaces dissipate heat from the surface and create a cold zone on the surface being measured. In the multi-function calibrators, the calibration temperature is created in a specially designed surface insert and an external reference thermometer measures the temperature directly under the surface. The reference thermometer also determines the temperature of the cold zone through integration of the temperature along the sensitive length of the reference thermometer and thus provides a true temperature calibration of surface temperature sensors.
6.5 Infrared insert (only CTM9100-150)
The operation of the calibrator with the infrared insert is suited to a quick and simple calibration of non-contact thermometers.
Fit the hollow and specially-designed insert into the block using a special replacement tool. The sleeve also has two bores in the edge (1 x 3.5 mm and 1 x 4.5 mm [1 x 0.14 in and 1 x 0.18 in]) for the accurate monitoring of the temperature.
The design and surface finish of the sleeve is important, in order that a defined emissivity of 1 is achieved.
Following use, the inserts should be removed using the specific replacement tools and then the sleeve and block should be cleaned. This prevents the sleeves becoming jammed in the heating block.
With temperatures of < 0 °C [< 32 °F] and higher air humidity, ice or dew can form in the insert. This can falsify the calibration. By capping the measurement opening, the ice or dew formation can be strongly reduced.
Leave the measurement opening closed as long as possible
Open the measurement opening for a short while for the measurement
Remove any ice or dew present by heating
EN
The sleeve is designed so that the supplied external reference provides the best possible result, since the depth of the bore is adjusted to the sensitive length. If a separate external reference is used for comparison calibration, ensure that the sensitive length is known and it lies in the centre of the calibration surface.
Surface insert
11263911.05 03/2019 EN/DE/IT
Insert for infrared measurements
19WIKA operating instructions, models CTD9100, CTB9100, CTM9100
6. Commissioning, operation
6.6 Testing of temperature sensors
To test temperature sensors, connect a separate temperature-measuring instrument to the test item. By comparing the temperature displayed on the external measuring instrument with the reference temperature, there is evidence of the status of the test item. This ensured
EN
that the test item requires a short time until it reaches the temperature of the metal block or liquid bath.
There is no possibility for calibrating earthed thermocouples, since the heating block is earthed and this would therefore lead to false measurement results.
6.7 Start-up procedure
If the calibrator is not used for a long time, because of the materials used (magnesium oxide), moisture can penetrate the heating elements. Following transport or storage of the calibrator in moist environments, the heating elements must be taken to higher temperatures slowly.
During the drying process,it should be assumed that the calibrator has not yet reached the required insulation voltage for protection class I. The start-up set point is T [248 °F] with a holding time of T
= 15 min.
n
= 120 °C
anf
Once the set temperature has been reached, through short switch impulses, the radiated heating energy from the metal block or liquid bath is delivered, so that the temperature within remains constant.
Reference and set temperature display
6.10 Control of the reference temperature The red LED OUT 1 indicates that the heater is turned on.
6.8 Switching on the calibrator/micro calibration bath
1. Connect to the mains using the mains connector supplied.
2. Switch on the mains switch.
The controller will be initialised. The upper PV display will show tEst. In the lower SV display, the version number (e.g. rL 2.2) will be shown.
After approx. 5 secs, the initialisation will be complete and the calibration mode will automatically be displayed. The built-in heating or cooling elements will temper the metal block automatically from room temperature to the controller’s set temperature.
6.9 Reference and set temperature display
Upper PV display:
The red 4-digit, 7-segment display shows the current temperature of the metal block or liquid bath.
Lower SV display:
The green, 4-digit, 7-segment display shows the current set temperature of the metal block or the liquid bath.
Display with LED OUT 1
During the heating phase a steady light indicates that heating energy is being supplied. A flashing LED indicates that the reference temperature (set temperature) has almost been reached and therefore the heating energy is only being supplied in short bursts.
Control achieved via PID algorithm
To ensure good temperature stability, the cycle time of the controller is set low and the control output is usually raised.
20 WIKA operating instructions, models CTD9100, CTB9100, CTM9100
11263911.05 03/2019 EN/DE/IT
7. Operating elements of the calibrator/micro calibration bath
7. Operating elements of the calibrator/micro calibration bath
There are three modes available for operation
Calibration mode
In this standard operating state the calibration of test items can be made.
Set-point mode
In this mode, the set temperature is entered.
Main menu
In this mode, all settings, such as specifying the desired temperature or setting the control parameters are made.
7.1 Selection of the operating modes for the CTM9100-150
To operate the multi-function calibrator in the desired mode of operation, first specify the correct linearisation for the controller. In the main menu (via the [P] button) select the desired function using the arrow keys.
LI Operates as a micro calibration bath DB Operates as a dry block Ir Operates as an infrared black body source SU Operates as a surface temperature calibrator
Select the liquid used depending upon the desired calibration temperature.
Dry-block operating mode
1. Cleaning the tank (if needed)
2. Inserting the insert (aluminium)
3. Set the speed of the magnetic stirrer to “0”.
4. In order to ensure the correct linearisation, set DB in the controller. For this, press the [P] button for approx. 5 secs. and confirm the appropriate setting in the main menu with [P].
5. Set the switch to the left of the controller to external reference. Connect the delivered external reference in to the socket provided and slide into the appropriate bore in the insert.
The insert has several bores into which the thermometer probes being calibrated and the external reference, for comparative calibration, can be inserted. The block is either heated or cooled to the desired calibration temperature. Once a stable temperature has been reached, the temperature probes to be calibrated can be compared with the reference thermometer.
EN
The desired function is accepted using the [P] button. On the main screen, the desired linearisation (LI, DB, Ir or
SU) is displayed in the PV line every 5 secs.
7.2 Method of operation in calibration mode within individual operating modes
Micro calibration bath operating mode (possible with or without insert for liquids)
1. Insert the magnetic stirrer and sensor basket.
2. Fill the micro calibration bath (see chapter 6.3.2 “Filling
the micro calibration bath”).
3. Set the speed of the magnetic stirrer, in order to achieve
the best homogeneity possible.
4. In order to ensure the correct linearisation, set LI in the
controller. For this, press the [P] button for approx. 5 secs. and confirm the appropriate setting in the main menu with [P].
5. Set the switch to the left of the controller to internal
reference.
Angled thermometers, large-diameter thermometers or thermometers with special designs cannot be calibrated with a dry-well calibrator. For this reason the multi-function calibrator also has the possibility to function as a stirred liquid bath. The liquid is circulated using a magnetic stirrer and thus provides very good temperature distribution within the bath.
11263911.05 03/2019 EN/DE/IT
Infrared operating mode
1. Cleaning the tank (if needed)
2. Insert the insert (hollow, ceramic coated)
3. Set the speed of the magnetic stirrer to “0”.
4. In order to ensure the correct linearisation, set Ir in the controller. For this, press the [P] button for approx. 5 secs. and confirm the appropriate setting in the main menu with [P].
5. Set the switch to the left of the controller to external reference. Connect the delivered external reference into the socket provided and slide it into into the appropriate bore on the outer edge of the insert.
The measuring spot of the pyrometer being calibrated must be smaller than the diameter of the infrared insert.
Surface operating mode
1. Cleaning the tank (if needed)
2. Insert the insert (hollow, fitted with a collar at the top)
3. Set the speed of the magnetic stirrer to “0”.
4. In order to ensure the correct linearisation, set SU in the controller. For this, press the [P] button for approx. 5 secs. and confirm the appropriate setting in the main menu with [P].
5. Set the switch to the left of the controller to external reference. Connect the delivered external reference in to the socket provided and slide into the appropriate bore directly under the surface of the insert.
21WIKA operating instructions, models CTD9100, CTB9100, CTM9100
7. Operating elements of the calibrator/micro calibration bath
The calibration of surface thermometers is very difficult and not fully defined. Thermometers mounted on surfaces dissipate heat from the surface and create a cold zone on the surface being measured. In the multi-function calibrator, the calibration temperature is created in a specially designed surface insert and an external reference thermometer
EN
measures the temperature directly under the surface. The reference thermometer also determines the temperature of the cold zone through integration of the temperature along the sensitive length of the reference thermometer and thus provides a true temperature calibration of surface temperature sensors. The sleeve is designed so that the supplied external reference provides the best possible result, since the depth of the bore is adjusted to the sensitive length. If a separate external reference is used for comparison calibration, ensure that the sensitive length is known and it lies in the centre of the calibration surface.
7.3 Calibration (calibration mode)
As soon as the calibrator or micro calibration bath is switched on, following the initialisation, it defaults to calibration mode. On the upper PV display, the current reference temperature is indicated. On the lower SV display is the set temperature.
The LED OUT 1 signals the status of the outputs for the heating control:
If the LED OUT 1 is on, the temperature will increase.
If the LED OUT 1 is not on, the heating is switched off.
a) Heating instrument
The LED OUT 2 signals the status of the outputs for the fan control:
If LED OUT 2 is on, the fan is running at higher speed.
If LED OUT 2 is not on, the fan is running at lower speed.
b) Heating and cooling instrument
The LED OUT 2 signals the status of the outputs for the cooling control:
If the LED OUT 2 is on, the temperature will decrease.
If the LED OUT 2 is not on, the cooling is switched off.
Two possibilities for setting the set temperature
Either a temporary set temperature is set (see chapter
7.3 “Calibration (calibration mode)”) or a fixed set temperature (see chapter 7.4 “Setting a temporary set temperature (set-point mode)”) is stored in the main menu.
7.4 Setting a temporary set temperature (set-point mode)
In this operating state, a stored set temperature is temporarily changed.
1. Press button [P] briefly.
On the upper PV display, the currently active set value memory is displayed, e.g. SP2 (set point 2). On the lower SV display is the corresponding set temperature.
2. By pressing button [▲] the set temperature is increased.
By pressing button [▼] the set temperature is reduced.
3. By pressing button [P] again, the new set point is
accepted.
Display in HEATING calibration mode
Temporary set temperature setting
The LED OUT 2 signals the status of the outputs for the fan/ cooling control.
By pressing button [▲] or [▼] the value is increased or decreased by 0.01 °C [0.01 °F]. If the keys are held down for at least 1 sec., however, the value increases or decreases more rapidly, and after 2 secs. even faster, so that the desired value can be reached very quickly. If, in set-point mode, no button has been pressed for approx. 15 secs., an automatic return to calibration mode occurs.
Display in FAN or COOLING calibration mode
22 WIKA operating instructions, models CTD9100, CTB9100, CTM9100
11263911.05 03/2019 EN/DE/IT
7. Operating elements of the calibrator/micro calibration bath
7.5 Programming (Main menu)
In this menu element, all settings can be made.
1. Press button [P] for approx. 5 secs. This opens the main menu.
2. Select the required main menu using buttons [▲] and [▼] (see overview).
3. Press button [P] to accept the selected menu point.
EN
approx. 5 secs.
Calibration mode
Main menu
Set-point mode
Menu structure (main menu)
11263911.05 03/2019 EN/DE/IT
23WIKA operating instructions, models CTD9100, CTB9100, CTM9100
7. Operating elements of the calibrator/micro calibration bath
7.5.1 Menu structure, parameter levels As shown in the menu structure, via OPEr, the groups and parameter levels are accessible, in which the settings can
then be made.
EN
Calibration mode
approx. 5 secs.
Set-point mode
Main menu
Return to another level
If, in the main menu, no button has been pressed for approx. 15 secs., an automatic return to one level above in calibration mode occurs. A return can be made by pressing button [▲] or [▼].
Group level Parameter level
Parameter input
Set point 1
Operating modes for CTM9100-150
Set point 2
Control off
Manual control
Micro calibration bath
Dry well calibrator
Infrared Dwell time
Surface Falling ramp
Parameter level
Parameter input
Set point 3
Set point 4
Rising ramp
Menu structure
24 WIKA operating instructions, models CTD9100, CTB9100, CTM9100
11263911.05 03/2019 EN/DE/IT
7. Operating elements of the calibrator/micro calibration bath
7.5.2 Switching off automatic control
For certain tasks, it is advantageous to switch off control (e.g. to make settings on the calibrator or micro calibration bath). In calibration mode, press button [P] for approx. 5 secs., this opens the main menu. The upper PV display will show OPEr. In the lower SV display, the LED SET will blink.
Display in main menu
Press button [▲] or [▼] until OFF is displayed.
7.5.3 Switching on automatic control
The control is switched off when the following display is shown: On the upper PV display, the current reference temperature and OFF are indicated alternately. On the lower SV display, the currently-selected set temperature is displayed.
Display with setting Control OFF
The control is switched back on by pressing button [P] for approx. 5 secs., this opens the main menu. The upper PV display will show ’rEG. In the lower SV display, the LED SET will blink.
EN
Control OFF menu
Confirm with button [P]. On the upper PV display, the current reference temperature and OFF are indicated alternately. On the lower SV display, the currently-selected set temperature is displayed.
Display with setting Control OFF
The control is now switched off and the reference temperature will fall steadily, unless it is corrected.
Display ‘rEG
Confirm switching the control on by pressing button [P].
Control is now activated. The calibrator or micro calibration bath will default to calibration mode and the set temperature will be reached.
11263911.05 03/2019 EN/DE/IT
25WIKA operating instructions, models CTD9100, CTB9100, CTM9100
7. Operating elements of the calibrator/micro calibration bath
7.5.4 Switching on manual control
Switch off the automatic control of the calibrator or micro calibration bath, and reach the required temperature with manual control. Press button [P] for approx. 5 secs., this opens the main menu.
EN
The upper PV display will show OPEr. In the lower SV display, the LED SET will blink.
Display in main menu
Press button [▲] or [▼] until OPLO is displayed. The upper PV display will show OPLO. In the lower SV display, the LED SET will blink.
By pressing button [▲], the output power is increased. By pressing button [▼], the output power is decreased.
By pressing button [▲] or [▼] the value is increased or decreased by 0.1 %. If the keys are held down for at least 1 sec., however, the value increases or decreases more rapidly, and after 2 secs. even faster, so that the desired value can be reached very quickly.
7.5.5 Switching off manual control
The manual control is switched on if the following display is shown: On the upper PV display, the current reference temperature is indicated. On the lower SV display, an H and the current set output power in % is displayed.
Manual control OPLO menu
Confirm by pressing the button [P]. On the upper PV display, the current reference temperature is indicated. On the lower SV display, an H and the current set output power in % is displayed.
Display with setting manual control OPLO
Display with setting manual control OPLO
The manual control is switched back off by pressing button [P] for approx. 5 secs., this opens the main menu. The upper PV display will show ’rEG. In the lower SV display, the LED SET will blink.
Display in main menu
Confirm switching the automatic control on by pressing button [P].
26 WIKA operating instructions, models CTD9100, CTB9100, CTM9100
11263911.05 03/2019 EN/DE/IT
7. Operating elements of the calibrator/micro calibration bath
7.5.6 Setting and storing fixed set temperatures
In order to store the set temperature in the calibrator or micro calibration bath, the appropriate set-point memory must be opened.
In calibration mode, press button [P] for approx. 5 secs., this opens the main menu. The upper PV display will show OPEr. In the lower SV display, the LED SET will blink.
Operator menu OPEr
If button [P] is pressed again, this opens the groups level. The upper PV display will show OPEr. In the lower SV display, ’SP is shown and additionally LED SET will blink.
Select one of the four set-point memories SP1, SP2, SP3 or SP4 using button [▲] or [▼]. By pressing button [P] the corresponding set-point memory will then be opened. In the upper PV display, the selected set-point memory e.g. SP3 will blink. On the lower SV display, the corresponding currently­selected set temperature is displayed.
Input in set-point memory SP3
By pressing button [▲], the set temperature is increased. By pressing button [▼], the set temperature is decreased.
By pressing button [▲] or [▼] the value is increased or decreased by 0.01 °C [0.01 °F]. If the keys are held down for at least 1 sec., however, the value increases or decreases more rapidly, and after 2 secs. even faster, so that the desired value can be reached very quickly.
EN
’SP Group
If button [P] is pressed again, this opens the parameter level.
The upper PV display will show ’SP. In the lower SV display, the set-point memory SP1 and additionally LED SET will blink.
Parameter for set-point memory SP1
By pressing button [P], the newly-set temperature is accepted. The set-point memory is left and the screen returns to the parameter level. To return to calibration mode, press button [▲] or [▼] for a long time.
If no button has been pressed for approx. 15 secs., an automatic return to one level above in calibration mode occurs.
11263911.05 03/2019 EN/DE/IT
27WIKA operating instructions, models CTD9100, CTB9100, CTM9100
7. Operating elements of the calibrator/micro calibration bath
7.5.7 Recalling the stored set temperatures
The set temperature can be called up from calibration mode. Press button [U] for approx. 2 secs. The current set-point memory will be opened. On the upper PV display, the current reference temperature is indicated.
EN
On the lower SV display, the set-point memory (SP1, SP2, SP3 or SP4) is shown for 2 secs., and then the
corresponding currently-selected set temperature is displayed.
It initially displays the set-point memory (SP1, SP2, SP3 or SP4).
Then it displays the stored set temperature.
A simple temperature profile can be generated. After switching on the calibrator or micro calibration bath, the temperature profile will run automatically.
Ramp control and temperature profile
Heating gradient, SLor
The heating gradient, SLor, is active if the reference temperature is lower than the set temperature. Each calibrator model has a maximum heating output, and thus only settings < than this heating power make sense and actually increase the time taken to achieve a set temperature.
Display on calling the set temperature
In order to retrieve a different stored set point, press button [U] again. The selected temperature value will be immediately adopted and approached.
7.5.8 Setting the ramp control and a temperature profile
Using the ramp control, the time can be defined over which the target temperature is achieved. This time may be shorter or longer than that which the calibrator or the micro calibration bath would normally require. By changing the set temperature or turning on the calibrator or micro calibration bath, it is automatically determined which of the gradients to use (heating gradient, SLor, or cooling gradient, SLoF). In the calibrator or in the micro calibration bath, the dwell time, dur.t, can be programmed so that the set temperature reached is automatically switched from the set-point memory SP1 to the set temperature in set-point memory SP2 after a predetermined time.
Calibrator model
Setting for SLor
(heating/cooling)
CTD9100-165 < 7 °C/min [< 13 °F/min] CTB9100-165 / CTM9100-150
with silicone oil 10 CS CTB9100-165 / CTM9100-150
with distilled water CTM9100-150 as calibrator < 3 °C/min [< 5 °F/min] CTM9100-150 as infrared calibrator < 3 °C/min [< 5 °F/min] CTM9100-150 as surface calibrator < 3 °C/min [< 5 °F/min]
Calibrator model
< 3 °C/min [< 5 °F/min]
< 5 °C/min [< 9 °F/min]
Setting for SLor
(heating)
CTD9100-450 / CTD9100-650 < 35 °C/min [< 63 °F/min] CTB9100-225 with silicone oil 20 CS < 22 °C/min [< 40 °F/min] CTB9100-225 with distilled water < 12 °C/min [< 22 °F/min]
11263911.05 03/2019 EN/DE/IT
28 WIKA operating instructions, models CTD9100, CTB9100, CTM9100
7. Operating elements of the calibrator/micro calibration bath
Cooling gradient SLoF
The cooling gradient, SLoF, is active if the reference temperature is higher than the set temperature. Only settings that are under the heating power of the calibrator have an effect on the cooling gradient.
Calibrator model
Setting for SLoF
(heating/cooling)
CTD9100-165 < 5 °C/min [< 9 °F/min] CTB9100-165 / CTM9100-150
with silicone oil 10 CS CTB9100-165 / CTM9100-150
with distilled water CTM9100-150 as calibrator < 4 °C/min [< 7 °F/min] CTM9100-150 as infrared calibrator < 4 °C/min [< 7 °F/min] CTM9100-150 as surface calibrator < 4 °C/min [< 7 °F/min]
Calibrator model
< 6 °C/min [< 11 °F/min]
< 4 °C/min [< 7 °F/min]
Setting for SLoF
(heating)
CTD9100-450 / CTD9100-650 up to 300 °C [572 °F]
300 °C to 100 °C [572 °F to 212 °F] CTB9100-225 with silicone oil 20 CS 200 °C to 50 °C [392 °F to 122 °F]
50 °C to 30 °C [122 °F to 86 °F] CTB9100-225 with distilled water 90 °C to 50 °C [194 °F to 122 °F]
50 °C to 30 °C [122 °F to 86 °F]
< 10 °C/min [< 18 °F/min] < 5 °C/min [< 9 °F/min]
< 4 °C/min [< 7 °F/min] < 0,5 °C/min [< 1 °F/min]
< 2 °C/min [< 4 °F/min] < 0,5 °C/min [< 1 °F/min]
The heating and cooling gradients and the dwell time can be set via the parameter level, ’rEG. This is achieved by pressing button P for approx. 5 secs., this opens the main menu. The upper PV display will show OPEr. In the lower SV display, the LED SET will blink.
Operator menu OPEr
If button [P] is pressed again, this opens the groups level. The upper PV display will show OPEr. In the lower SV display, ’SP is shown and additionally LED SET will blink.
EN
The dwell time, dur.t, is active when the set temperature, SP1, has been reached. Subsequently, the calibrator or micro
calibration bath automatically switches to set temperature SP2.
If settings have been made for these three parameters, the calibrator or the micro calibration bath will not use the new values until you either change the set temperature or until the calibrator or the micro calibration bath has been turned off and then on again. Another approach would be to switch the automatic control off before the parameter change (see chapter 7.5.2 “Switching off automatic control”), and then turning it back on (see chapter 7.5.3 “Switching on automatic control”).
’SP group
Select the group, 'rEG, by pressing button [▼]. The upper PV display will show OPEr. In the lower SV display, ’rEG is shown and additionally LED
SET will blink.
’rEG group
11263911.05 03/2019 EN/DE/IT
29WIKA operating instructions, models CTD9100, CTB9100, CTM9100
7. Operating elements of the calibrator/micro calibration bath
If button [P] is pressed again, this opens the parameter level.
The upper PV display will show ‘rEG. In the lower SV display, SLor will blink.
EN
Parameter for heating gradient, SLor
7.5.8.1 Setting the heating gradient
The heating gradient, SLor, is active if the reference temperature is lower than the set temperature. The setting range stretches from 99.99 °C/min to 0.00 °C/min [99.99 °F/min to 0.00 °F/min].
The function is deactivated if SLor = InF (in no Function) is set.
You are at the parameter level (as described in chapter
7.5.1 “Menu structure, parameter levels”).
The upper PV display will show ‘rEG. In the lower SV display, SLor will blink.
By pressing button [▲], the heating gradient, SLor, will be increased. By pressing button [▼], the heating gradient, SLor, will be decreased.
By pressing button [▲] or [▼] the value is increased or decreased by 0.1. If the keys are held down for at least 1 sec., however, the value increases or decreases more rapidly, and after 2 secs. even faster, so that the desired value can be reached very quickly.
By pressing button [P], the newly-set heating gradient, SLor, is accepted. The display returns to the parameter level and the other parameters can be set.
If no button has been pressed for approx. 15 secs., an automatic return to one level above in calibration mode occurs.
After the setting has been made, the calibrator or the micro calibration bath will not use the new values until you either change the set temperature or the calibrator or the micro calibration bath has been turned off and then on again.
7.5.8.2 Setting the cooling gradient
The cooling gradient, SLoF, is active if the reference temperature is higher than the set temperature. The setting range stretches from 99.99 °C/min to 0.00 °C/min [99.99 °F/min to 0.00 °F/min].
The function is deactivated if SLoF = InF (in no Function) is set.
Parameter for heating gradient, SLor
Press button [P]. On the upper PV display, SLor will blink. On the lower SV display, the corresponding currently-
selected heating gradient is displayed.
Input of the heating gradient
30 WIKA operating instructions, models CTD9100, CTB9100, CTM9100
You are at the parameter level (as described in chapter
7.5.1 “Menu structure, parameter levels”). The upper PV display will show ‘rEG. In the lower SV display, SLor will blink.
Parameter for heating gradient, SLor
11263911.05 03/2019 EN/DE/IT
Loading...
+ 118 hidden pages