WIKA CTD9100-375 Specifications

Temperature dry-well calibrator Model CTD9100-375
Applications
Testing and calibration of temperature measuring instruments
Also suitable for on-site calibration
Special features
Calibration technology
WIKA data sheet CT 41.32
High accuracy and stability
Possibility to check temperature switches
Low weight and compact design
Simple operation
Description
Versatile in application
Nowadays, fast and simple testing of thermometers is a “must” when it comes to the operational safety of machines and plants. The portable calibrators of the CTD9100 family are particularly suited to local calibration tasks and extremely user-friendly. Due to their compact design and their low weight, the instruments can be taken and used almost anywhere. The CTD9100-375 in particular is characterised by its robust case design and compact dimensions.
Temperature dry-well calibrator model CTD9100-375
Easy to use
The temperature dry-well calibrators of the CTD9100 series work with temperature-controlled metal blocks and inter­changeable inserts.
The calibration temperature, adjusted simply using two keys on the controller, can be very quickly controlled. The set temperature of the heating block is displayed on a large, 4-line, high-contrast LED display. Thus reading errors are virtually eliminated.
The new instrument concept brings together a stable heat source with precision Pt100 temperature measurement. This enables industrial temperature probes to be calibrated even
more eciently. Regular monitoring of temperature probes
helps to recognise failures promptly and shorten downtimes.
Thermometers with dierent diameters can be tted into the
calibrator using inserts, drilled to suit. A new block design, with improved temperature homogeneity at the calibrator's lower range, leads to smaller measurement uncertainties.
Page 1 of 4WIKA data sheet CT 41.32 ∙ 08/2015
Specications Model CTD9100-375
Display
Temperature range t
Accuracy
Stability
1)
2)
… 375 °C (t
amb
± 0.5 ... 0.8 K
± 0.05 K to 100 °C (212 °F)
Resolution 0.1 °C
Temperature control
Heating time 6 min from 20 °C to 300 °C (from 68 °F to 572 °F)
Cooling time 14 min from 300 °C to 60 °C (from 572 °F to 140 °F)
Stabilisation time
3)
5 min
Insert
Immersion depth 100 mm (3.94 in)
Insert dimensions Ø 13 x 100 mm (0.51 x 3.94 in)
Insert material Brass
Voltage supply
Power supply AC 230 V, 50/60 Hz or AC 110 V, 50/60 Hz
Power consumption 250 VA
Power cord for Europe, 230 V
Communication
Interface RS-232
Case
Dimensions (W x H x D) 149 x 74 x 155 mm (5.9 x 2.9 x 6.1 in)
Weight 1.7 kg (3.8 lbs)
1) Is dened as the measuring deviation between the measured value and the reference value.
2) Maximum temperature dierence at a stable temperature over 30 minutes.
3) Time before reaching a stable measuring value.
... 707 °F)
amb
The measurement uncertainty is dened as the total measurement uncertainty (k = 2), which contains the following shares:
accuracy, measurement uncertainty of reference, stability and homogeneity.
CE conformity, approvals, certicates
EC declaration of conformity
EMC directive 2004/108/EC, EN 61326 emission (group 1, class B) and interference immunity (industrial
application)
Low voltage directive 2006/95/EC, EN 61010-1 and EN 61010-2-10, safety requirements for electrical equipment for
measurement, control and laboratory use
Approvals
EAC Import certicate, Eurasian Economic Community
GOST Metrology/measurement technology, Russia
Certicate
Calibration Standard: 3.1 calibration certicate per DIN EN 10204
Option: DKD/DAkkS calibration certicate
Recommended recalibration interval 1 year (dependent on conditions of use)
Approvals and certicates, see website
Page 2 of 4 WIKA data sheet CT 41.32 ∙ 08/2015