WIKA CTB9500 Specifications

Calibration bath Model CTB9400, heating bath Model CTB9500, heating and cooling bath
Applications
Calibration in laboratories and in control and instrumenta­tion workshops
Simultaneous calibration of several sensors
Special features
Calibration technology
WIKA data sheet CT 46.20
Controller with 5.7" colour touchscreen and menu naviga­tion in 11 languages
2 x USB (host, device), Ethernet and RS-232 interface
High-accuracy, reliable temperature control
5-point calibration for control sensors (optional)
Description
Calibration in baths
WIKA calibration baths are an ideal temperature source for the calibration of temperature probes, both in the workshop and the laboratory. For calibration, the test items and the standard thermometer are brought to the same temperature within the bath. As soon as a stable temperature has been reached, the test items are read, or their output signals measured, and compared with the standard thermometer.
In order to achieve a small measurement uncertainty within this comparison, a temperature control unit is needed which guarantees a homogeneous temperature distribution and a constant temperature over time within the calibration range.
These requirements are fullled by these baths.
Along with the bath, the bath uid also plays a signicant
role. To ensure homogeneous temperature distribution, the
selected uid should have a high thermal conductivity and low viscosity. Moreover, the uid should be inert, have a low
vapour pressure, should not decompose chemically, not
Calibration bath model CTB9500
burn, and should maintain its properties over a wide tempe­rature range. In practice, silicone oils have proven themselves here.
Temperature ranges from -45 … +300 °C
The calibration baths are available in two dierent versions:
Model CTB9400 for 28 ... 300 °C
Model CTB9500 for -45 ... +200 °C
The instruments are typically used in calibration laboratories and in control and instrumentation workshops for the calibra­tion of thermometers.
Easy to use
The calibration baths feature a temperature-controlled tank with a usable depth of 200 mm. The maximum immersion depth for the test items of 200 mm reduces heat dissipation errors, resulting in smaller measurement uncertainties.
Page 1 of 7WIKA data sheet CT 46.20 ∙ 06/2013
Specications
Model CTB9400 Model CTB9500
Temperature range 28 … 300 °C -45 … +200 °C
Stability 0.02 K
Display resolution 0.1 °C, (0.01 °C only on exclusive version)
Display units °C/°F - selectable
Voltage supply
Power supply AC 230 V, 50/60 Hz
AC 400 V, 50 Hz, 3~ AC 208 V, 60 Hz, 2~
Power consumption
Current supply max. 14 A
Current supply with 3-phase supply max. 11 A
Fuse 16 A
Fuses with 3-phase 3 x 16 A
Heating power
at 230 V AC 3,000 W
Cooling power
at 20 °C --- 800 W
at 0 °C --- 800 W
at -20 °C --- 500 W
Communication
Interface USB, Ethernet, RS-232
Selectable user languages EN, DE, FR, IT, ES, RU, CN, PT, JP, CZ, PL
Bad dimensions
Minimum lling volume 11.5 litres approx. 22 litres
Bath opening, W x D 270 x 145 mm 280 x 280 mm
Bath depth 200 mm 200 mm
Dimensions, W x D x H 335 x 382 x 433 mm 420 x 565 x 719 mm
Weight (empty) 22 kg 72 kg
AC 230 V, 50/60 Hz AC 208 V 60 Hz, 2~
Accessories Model CTB9400 Model CTB9500
Silicone oil DC 200.05
-40 ... +130 °C, FP = 133 °C
Silicone oil DC 200.10
-35 ... +160 °C, FP = 163 °C
Silicone oil DC 200.20
10 ... 220 °C, FP = 232 °C
Silicone oil DC 200.50
25 ... 250 °C, FP = 280 °C
Mounting xtures X X
USB interface cable X X
Not recommended From -40 … +130 °C usable
Not recommended From -35 … +160 °C usable
From 40 … 220 °C usable From 10 … 200 °C usable
From 40 … 250 °C usable From 25 … 200 °C usable
CE conformity
EMC directive 2004/108/EC, EN 61326 Emission (group 1, class A) and interference immunity (controlled
electromagnetic environment)
Low voltage directive 2006/95/EC, EN 61010-1, EN 61010-2-010
Machinery directive 2006/42/EC, EN 378-2
Approvals and certicates, see website
Page 2 of 7 WIKA data sheet CT 46.20 ∙ 06/2013
Dierences in the rmware
Function Standard Exclusive
Temperature control
Temperature control mode (Internal, Process) X
Programmer X 3 programmes with a maximum of 15 steps
Ramp function X linear
Controller parameter tuning predened
Calibration programme for control sensor
1 point 5 point
(Internal, Process) Monitoring (Level protection, Overtemperature
protection
2)
)
Adjustable alarm limits
VPC (Variable Pressure Control)
3)
Venting program
Compressor automatic control
Set point limit
Display and Operation
Temperature display 5.7" Colour touchscreen 5.7" Colour touchscreen
Display mode graphic, numeric graphic, numeric
Display resolution 0.1 °C 0.1 °C / 0.01 °C
Graphic display for temperature curves Window Window, full screen, scalable
Calendar, date, time
Languages: EN, DE, FR, IT, ES, RU, CN, PT, JP,
CZ, PL
Temperature format, switchable (°C/°F)
Gesture control
Favourites menu X
2nd set point
Connections
Digital interface RS-232
USB interfaces (Host and Device)
Ethernet RJ45 interface
External control signal/ECS STANDBY
4)
Pt100 external sensor connection
Various
Alarm signal optical/acoustic
Autostart (Mains failure automatic)
Plug-and-play
Technical glossary
Remote control via software
Saving/loading of temperature control programmes
X
via USB
1 TAC function available as a 30-day evaluation version. 2 For temperature control devices with integrated overtemperature protection. 3 For temperature control devices with variable-speed pumps or external bypass.
1)
TAC (True Adaptive Control)
X = not available
= present
Page 3 of 7WIKA data sheet CT 46.20 ∙ 06/2013
Loading...
+ 4 hidden pages