WIKA CPT7000 Operating Instructions Manual

Operating instructions Betriebsanleitung
EN
DE
Reference pressure sensor, model CPT7000
Referenz-Drucksensor, Typ CPT7000
Reference pressure sensor, model CPT7000
2
14210181.02 04/2019 EN/DE
WIKA operating instructions reference pressure sensor, model CPT7000
EN
DE
Operating instructions model CPT7000 Page 3 - 22
Betriebsanleitung Typ CPT7000 Seite 23 - 41
© 02/2017 WIKA Alexander Wiegand SE & Co. KG All rights reserved. / Alle Rechte vorbehalten. WIKA
®
is a registered trademark in various countries.
WIKA
®
ist eine geschützte Marke in verschiedenen Ländern.
Prior to starting any work, read the operating instructions! Keep for later use!
Vor Beginn aller Arbeiten Betriebsanleitung lesen! Zum späteren Gebrauch aufbewahren!
3
WIKA operating instructions, reference pressure sensor, model CPT7000
EN
14210181.02 04/2019 EN/DE
Contents
Declarations of conformity can be found online at www.wika.com.
Contents
1. General information 4
2. Design and function 5
2.1 Overview . . . . . . . . . . . . . . . . . . . . . . . . 5
2.2 Description . . . . . . . . . . . . . . . . . . . . . . . 5
2.3 Scope of delivery . . . . . . . . . . . . . . . . . . . . . 5
2.4 Product passport . . . . . . . . . . . . . . . . . . . . . 5
3. Safety 6
3.1 Explanation of symbols . . . . . . . . . . . . . . . . . . . 6
3.2 Intended use . . . . . . . . . . . . . . . . . . . . . . . 6
3.3 Improper use . . . . . . . . . . . . . . . . . . . . . . . 7
3.4 Personnel qualification . . . . . . . . . . . . . . . . . . . 7
3.5 Labelling, safety marks . . . . . . . . . . . . . . . . . . . 8
4. Transport, packaging and storage 9
4.1 Transport . . . . . . . . . . . . . . . . . . . . . . . . 9
4.2 Packaging and storage . . . . . . . . . . . . . . . . . . . 9
5. Commissioning, operation 10
5.1 Mounting the instrument . . . . . . . . . . . . . . . . . . . 10
5.2 Requirements for mounting point . . . . . . . . . . . . . . . . 10
5.3 Mechanical mounting . . . . . . . . . . . . . . . . . . . .11
5.4 Electrical connection . . . . . . . . . . . . . . . . . . . . 11
6. Faults 12
7. Maintenance, cleaning and recalibration 13
7.1 Maintenance . . . . . . . . . . . . . . . . . . . . . . .13
7.2 Cleaning . . . . . . . . . . . . . . . . . . . . . . . . 14
7.3 Recalibration . . . . . . . . . . . . . . . . . . . . . . .14
8. Dismounting, return and disposal 15
8.1 Dismounting . . . . . . . . . . . . . . . . . . . . . . . 15
8.2 Return . . . . . . . . . . . . . . . . . . . . . . . . . 16
8.3 Disposal . . . . . . . . . . . . . . . . . . . . . . . . 16
9. Specifications 17
9.1 Reference pressure sensor . . . . . . . . . . . . . . . . . . 17
9.2 Certificates . . . . . . . . . . . . . . . . . . . . . . . 19
9.3 Dimensions in mm (in). . . . . . . . . . . . . . . . . . . . 20
10. Accessories 21
4
WIKA operating instructions, reference pressure sensor, model CPT7000
EN
14210181.02 04/2019 EN/DE
1. General information
The model CPT7000 reference pressure sensor described in the operating instructions has been designed and manufactured using state-of-the-art technology. All components are subject to stringent quality and environmental criteria during production. Our management systems are certified to ISO 9001 and ISO 14001.
These operating instructions contain important information on handling the instrument. Working safely requires that all safety instructions and work instructions are observed.
These operating instructions for the model CPT7000 reference pressure sensor are valid in combination with the operating instructions for the “model CPH7000 portable process calibrator” (article number 14204467). The CPT7000 can only be used in combination with the CPH7000.
Observe the relevant local accident prevention regulations and general safety regulations for the instrument's range of use.
The operating instructions are part of the product and must be kept in the immediate vicinity of the instrument and readily accessible to skilled personnel at any time. Pass the operating instructions on to the next operator or owner of the instrument.
Skilled personnel must have carefully read and understood the operating instructions prior to beginning any work.
The general terms and conditions contained in the sales documentation shall apply.
Subject to technical modifications.
Factory calibrations / DKD/DAkkS calibrations are carried out in accordance with international standards.
Further information:
- Internet address: www.wika.de / www.wika.com
- Relevant data sheet: CT 15.51
- Application consultant: Tel.: +49 9372 132-0 Fax: +49 9372 132-406 info@wika.de
1. General information
5
WIKA operating instructions, reference pressure sensor, model CPT7000
EN
14210181.02 04/2019 EN/DE
2. Design and function
2.1 Overview
1
Electrical connection
2
Sealing ring
3
Case; product label
4
Process connection, spanner flats
5
Process connection, thread
2.2 Description
The model CPT7000 reference pressure sensor covers the relative pressure range from
-1 ... 10,000 bar (-14.5 ... 150,000 psi) and the absolute pressure range from 0 ... 40 bar abs. (0 ... 580 psi abs.).
The CPT7000 is operated with the model CPH7000 process calibrator and has also been produced exclusively for this calibrator. Any other usage is not possible.
2.3 Scope of delivery
Reference pressure sensor model CPT7000
Operating instructions
Calibration certificate
Cross-check scope of delivery with delivery note.
2.4 Product passport
The product passport is a web application where various information about an instrument can be retrieved by entering the 11-digit alphanumeric serial number. This includes, for example, instrument configuration, article number, operating instructions, data sheet or calibration certificates.
1
2
3
4
5
2. Design and function
6
WIKA operating instructions, reference pressure sensor, model CPT7000
EN
14210181.02 04/2019 EN/DE
3. Safety
3.1 Explanation of symbols
WARNING!
... indicates a potentially dangerous situation that can result in serious injury or death, if not avoided.
CAUTION!
... indicates a potentially dangerous situation that can result in light injuries or damage to property or the environment, if not avoided.
Information
... points out useful tips, recommendations and information for efficient and trouble-free operation.
3.2 Intended use
The reference pressure sensor serves for the accurate determination of the pressure value. The measured pressure is output as an electrical signal.
Only use the reference pressure sensor in applications that lie within its technical performance limits (e.g. max. ambient temperature, material compatibility, ...). For performance limits see chapter 9 “Specifications”.
The instrument has been designed and built solely for the intended use described here, and may only be used accordingly.
The technical specifications contained in these operating instructions must be observed. Improper handling or operation of the instrument outside of its technical specifications requires the instrument to be taken out of service immediately and inspected by an authorised WIKA service engineer.
The product passport can be retrieved from the product page , directly from the corresponding web application (https://portal.wika.com/serial/) or via the QR code on the product label.
2. Design and function / 3. Safety
7
WIKA operating instructions, reference pressure sensor, model CPT7000
EN
14210181.02 04/2019 EN/DE
Handle electronic precision measuring instruments with the required care (protect from humidity, impacts, strong magnetic fields, static electricity and extreme temperatures, do not insert any objects into the instrument or its openings). Plugs and sockets must be protected from contamination.
The manufacturer shall not be liable for claims of any type based on operation contrary to the intended use.
3.3 Improper use
WARNING! Injuries through improper use
Improper use of the instrument can lead to hazardous situations and injuries.
Refrain from unauthorised modifications to the instrument.
Do not use the instrument within hazardous areas.
Do not use the instrument with abrasive or viscous media.
It is forbidden to open the instrument.
Do not use the CPT7000 if it is damaged. Before use, check for visible defects.
Inspect the test leads for damaged insulation or exposed metal. Damaged test leads should be replaced before using the reference pressure sensor.
Only use the accessories specified and authorized by WIKA.
CAUTION! Damage to the instrument through incorrect pressure media
The pressure medium to be measured can lead to material damage to the pressure sensor.
Check whether the material of the reference pressure sensor is suitable for the pressure medium to be measured.
Any use beyond or different to the intended use is considered as improper use.
3.4 Personnel qualification
WARNING! Risk of injury should qualification be insufficient
Improper handling can result in considerable injury and damage to equipment.
The activities described in these operating instructions may only be carried out by skilled personnel who have the qualifications described below.
3. Safety
8
WIKA operating instructions, reference pressure sensor, model CPT7000
EN
14210181.02 04/2019 EN/DE
Skilled personnel
Skilled personnel, authorised by the operator, are understood to be personnel who, based on their technical training, knowledge of measurement and control technology and on their experience and knowledge of country-specific regulations, current standards and directives, are capable of carrying out the work described and independently recognising potential hazards.
Special operating conditions require further appropriate knowledge, e.g. of aggressive media.
3.5 Labelling, safety marks
Product label
1
Serial number
2
Date of manufacture (month-year)
3
Serial number, external sensor
4
Accuracy
5
Pressure measuring range
6
Model
Symbols
Before mounting and commissioning the instrument, ensure you read the operating instructions!
Do not dispose of with household waste. Ensure a proper disposal in accordance with national regulations.
6
5
4
2
1
3
3. Safety
9
WIKA operating instructions, reference pressure sensor, model CPT7000
EN
14210181.02 04/2019 EN/DE
4. Transport, packaging and storage
4.1 Transport
Check the reference pressure sensor for any damage that may have been caused by transport. Obvious damage must be reported immediately.
CAUTION! Damage through improper transport
With improper transport, a high level of damage to property can occur.
When unloading packed goods upon delivery as well as during internal transport, proceed carefully and observe the symbols on the packaging.
With internal transport, observe the instructions in chapter 4.2 “Packaging and storage”.
If the instrument is transported from a cold into a warm environment, the formation of condensation may result in instrument malfunction. Before putting it back into operation, wait for the instrument temperature and the room temperature to equalise.
4.2 Packaging and storage
Do not remove packaging until just before mounting. Keep the packaging as it will provide optimum protection during transport (e.g. change in installation site, recalibration).
Permissible conditions at the place of storage:
Storage temperature: -20 ... +80 °C (-4 ... +176 °F)
Humidity: 0 ... 95 % relative humidity (non-condensing)
Avoid exposure to the following factors:
Direct sunlight or proximity to hot objects
Mechanical vibration, mechanical shock (putting it down hard)
Soot, vapour, dust and corrosive gases
Hazardous environments, flammable atmospheres
Store the instrument in its original packaging in a location that fulfils the conditions listed above. If the original packaging is not available, pack and store the instrument as described below:
1. Wrap the instrument in an antistatic plastic film.
2. Place the instrument, along with the shock-absorbent material, in the packaging.
3. If stored for a prolonged period of time (more than 30 days), place a bag containing a desiccant inside the packaging.
4. Transport, packaging and storage
10
WIKA operating instructions, reference pressure sensor, model CPT7000
EN
14210181.02 04/2019 EN/DE
5. Commissioning, operation
Personnel: Skilled personnel Protective equipment: protective gloves Tools: Spanner or torque spanner
Only use original parts (see chapter 10 “Accessories”).
WARNING! Physical injuries and damage to property and the environment caused by hazardous media
Upon contact with hazardous media (e.g. oxygen, acetylene, flammable or toxic substances), harmful media (e.g. corrosive, toxic, carcinogenic, radioactive), and also with refrigeration plants and compressors, there is a danger of physical injuries and damage to property and the environment. Should a failure occur, aggressive media with extremely high temperature and under high pressure or vacuum may be present at the instrument.
For these media, in addition to all standard regulations, the appropriate existing codes or regulations must also be followed.
5.1 Mounting the instrument
WARNING! Physical injury through aggressive media, high pressures and high temperatures
When mounting, there is a danger from aggressive media, high pressures and high medium temperature.
Observe the information in the material safety data sheet for the corresponding medium.
Connect the pressure sensor once the system has been depressurised.
At medium temperatures above 65 °C (149 °F) wear protective gloves.
Only use the pressure transmitter if it is in perfect condition with respect to safety. Prior to commissioning, the pressure transmitter must be subjected to a visual inspection.
Leaking fluid is indicative of damage.
5.2 Requirements for mounting point
The mounting point must meet the following conditions:
Ambience corresponds to a max. pollution degree of 3.
Humidity ≤ 80 % r. h. (non-condensing).
Protected from weather influences.
Sealing faces are clean and undamaged.
5. Commissioning, operation
11
WIKA operating instructions, reference pressure sensor, model CPT7000
EN
14210181.02 04/2019 EN/DE
Sufficient space for a safe electrical installation.
For information on tapped holes and welding sockets, see Technical information IN 00.14 at www.wika.com.
Permissible ambient and medium temperatures remain within the performance limits. Consider possible restrictions on the ambient temperature range caused by mating connector used.
For performance limits, see chapter 9 “Specifications”
5.3 Mechanical mounting
The max. torque depends on the mounting point (e.g. material and shape). If you have any questions, please contact our application consultant. For contact details see chapter 1 “General information” or the back page of the operating instructions.
1. Seal the sealing faces.
2. At the mounting point, screw the pressure transmitter in hand-tight.
When screwing in, do not cross the threads.
3. Tighten with a torque spanner using the spanner flats. The maximum torque of the CPT7000 is 50 Nm; for flush connections, 40 Nm.
5.4 Electrical connection
To electrically connect a model CPT7000 to the CPH7000 process calibrator, the corresponding plug connection of the cable must be plugged in at the sensor. To connect it to the CPH7000, the other end of the cable must also be plugged in.
1. To connect or change the reference pressure sensor, must not switch off the CPH7000 process calibrator.
2. Connect the process calibrator and the reference pressure sensor to each other electrically using a separate sensor connection cable. Use the 7-pin plug contact on the pressure sensor for this.
3. Connect the 7-pin connector to the reference pressure sensor in accordance with the orientation guide and secure it through the connection sleeve. Turn the connection sleeve clockwise without much force.
4. Connect the 6-pin M-DIN connector to the process calibrator on channel 1 or channel 2 in accordance with the orientation guide.
5. Commissioning, operation
12
WIKA operating instructions, reference pressure sensor, model CPT7000
EN
14210181.02 04/2019 EN/DE
When connecting the sensor connection cable to the hand-held, the pressure sensor’s connector might not locate properly in the socket. In this event you should try holding the connector by the bend protection, rather than by the connector sleeve.
Connect the connector without tilt the threads.
If the connector is positioned correctly, it can be plugged in without any significant effort.
To disconnect the pressure sensor, do not pull on the sensor connection cable, but rather only on the connector sleeve.
The voltage supply is provided via the model CPH7000 process calibrator.
6. Faults
Personnel: Skilled personnel Protective equipment: protective gloves Tools: Spanner or torque spanner
WARNING! Physical injuries and damage to property and the environment caused by hazardous media
Upon contact with hazardous media (e.g. oxygen, acetylene, flammable or toxic substances), harmful media (e.g. corrosive, toxic, carcinogenic, radioactive), and also with refrigeration plants and compressors, there is a danger of physical injuries and damage to property and the environment. Should a failure occur, aggressive media with extremely high temperature and under high pressure or vacuum may be present at the instrument.
For these media, in addition to all standard regulations, the appropriate existing codes or regulations must also be followed.
Use the requisite protective equipment.
5. Commissioning, operation / 6. Faults
13
WIKA operating instructions, reference pressure sensor, model CPT7000
EN
14210181.02 04/2019 EN/DE
CAUTION! Physical injuries and damage to property and the environment
If faults cannot be eliminated by means of the listed measures, the instrument must be taken out of operation immediately.
Ensure that pressure or signal is no longer present and protect against accidental commissioning.
Contact the manufacturer.
If a return is needed, please follow the instructions given in chapter
8.2 “Return”.
For contact details see chapter 1 “General information” or the back page of the operating instructions.
Faults Causes Measures
No output signal Cable break Visually check the continuity of
the cable. If there is any damage to the cable, replace it with a new one.
Leakage at the process connection
Hairline crack in the pressure port
Consult with manufacturer, if required, replace instrument.
7. Maintenance, cleaning and recalibration
Personnel: Skilled personnel Protective equipment: protective gloves Tools: Spanner or torque spanner
For contact details see chapter 1 “General information” or the back page of the operating instructions.
7.1 Maintenance
The CPT7000 reference pressure sensor is maintenance-free. Repairs must only be carried out by the manufacturer. Only use original parts (see chapter 10 “Accessories”).
6. Faults / 7. Maintenance, cleaning and recalibration
14
WIKA operating instructions, reference pressure sensor, model CPT7000
EN
14210181.02 04/2019 EN/DE
7.2 Cleaning
CAUTION! Physical injuries and damage to property and the environment
Improper cleaning may lead to physical injuries and damage to property and the environment. Residual media in the dismounted instrument can result in a risk to persons, the environment and equipment.
Use the requisite protective equipment.
Carry out the cleaning process as described below.
1. Before cleaning, correctly disconnect the instrument from the pressure supply, switch it off and disconnect it from the voltage supply (CPH7000).
2. Use the requisite protective equipment.
3. Clean the instrument with a moist cloth. Electrical connections must not come into contact with moisture!
CAUTION! Damage to the instrument
Improper cleaning may lead to damage to the instrument!
Do not use any aggressive cleaning agents.
Do not use any pointed and hard objects for cleaning.
Do not use solvents or abrasives for cleaning.
4. Wash or clean the dismounted instrument, in order to protect persons and the environment from exposure to residual media.
7.3 Recalibration
DKD/DAkkS certificate - official certificates:
We recommend that the instrument is regularly recalibrated by the manufacturer, with time intervals of approx. 12 months.
7. Maintenance, cleaning and recalibration