WIKA CPT-2 series Operating Instructions Manual

Operating Instructions
Process pressure transmitter CPT-2x
Process pressure transmitter CPT-2x
Slave for electronic dierential pressure With SIL qualication Ceramic sensor
Contents
WIKA Operating Instructions - Process pressure transmitter CPT-2x
Contents
1 About this document ............................................................................................................... 4
1.1 Function ........................................................................................................................... 4
1.2 Target group ..................................................................................................................... 4
1.3 Symbols used................................................................................................................... 4
2 For your safety ......................................................................................................................... 5
2.1 Authorised personnel ....................................................................................................... 5
2.2 Appropriate use ................................................................................................................ 5
2.3 Warning about incorrect use ............................................................................................. 5
2.4 General safety instructions ............................................................................................... 5
2.5 EU conformity ................................................................................................................... 6
2.6 SIL qualication according to IEC 61508 .......................................................................... 6
2.7 Permissible process pressure .......................................................................................... 6
2.8 Installation and operation in the USA and Canada ........................................................... 6
3 Product description ................................................................................................................. 7
3.1 Conguration .................................................................................................................... 7
3.2 Principle of operation........................................................................................................ 8
3.3 Supplementary cleaning procedures .............................................................................. 13
3.4 Packaging, transport and storage ................................................................................... 14
4 Mounting ................................................................................................................................. 15
4.1 General instructions ....................................................................................................... 15
4.2 Ventilation and pressure compensation .......................................................................... 16
4.3 Combination Master - Slave ........................................................................................... 18
4.4 Level measurement ........................................................................................................ 20
4.5 Dierential pressure measurement ................................................................................ 21
4.6 Interface measurement .................................................................................................. 21
4.7 Density measurement .................................................................................................... 22
4.8 Density-compensated level measurement ..................................................................... 23
4.9 External housing ............................................................................................................ 25
5 Connecting to power supply ................................................................................................. 26
5.1 Preparing the connection ............................................................................................... 26
5.2 Connecting ..................................................................................................................... 26
5.3 Single chamber housing ................................................................................................. 28
5.4 External housing with version IP 68 (25 bar)................................................................... 28
5.5 Connection example ...................................................................................................... 30
6 Functional safety (SIL) .......................................................................................................... 31
6.1 Objective ........................................................................................................................ 31
6.2 SIL qualication .............................................................................................................. 31
6.3 Application area ............................................................................................................. 31
6.4 Safety concept of the parameterization .......................................................................... 32
7 Set up with the display and adjustment module ................................................................ 34
7.1 Parameter adjustment .................................................................................................... 34
8 Diagnosis, asset management and service ........................................................................ 47
8.1 Maintenance .................................................................................................................. 47
8.2 Cleaning - hygienic connection with compression nut .................................................... 47
8.3 Rectify faults ................................................................................................................... 48
Contents
WIKA Operating Instructions - Process pressure transmitter CPT-2x
8.4 Exchange process module on version IP 68 (25 bar) ..................................................... 48
8.5 Instrument repair ............................................................................................................ 49
9 Dismount................................................................................................................................. 51
9.1 Dismounting steps.......................................................................................................... 51
9.2 Disposal ......................................................................................................................... 51
10 Supplement ............................................................................................................................ 52
10.1 Technical data ................................................................................................................ 52
10.2 Calculation of the total deviation ..................................................................................... 62
10.3 Practical example ........................................................................................................... 63
10.4 Dimensions .................................................................................................................... 66
10.5 Trademark ...................................................................................................................... 74
Safety instructions for Ex areas
Take note of the Ex specic safety instructions for Ex applications. These instructions are attached as documents to each instrument with Ex approval and are part of the operating instructions manual.
Editing status: 2017-11-23
1 About this document
WIKA Operating Instructions - Process pressure transmitter CPT-2x
1 About this document
1.1 Function
This operating instructions manual provides all the information you need for mounting, connection and setup as well as important instruc­tions for maintenance, fault rectication, the exchange of parts and the safety of the user. Please read this information before putting the instrument into operation and keep this manual accessible in the im-
mediate vicinity of the device.
1.2 Target group
This operating instructions manual is directed to trained personnel. The contents of this manual must be made available to the qualied personnel and implemented.
1.3 Symbols used
Information, tip, note
This symbol indicates helpful additional information.
Caution: If this warning is ignored, faults or malfunctions can result. Warning: If this warning is ignored, injury to persons and/or serious
damage to the instrument can result. Danger: If this warning is ignored, serious injury to persons and/or
destruction of the instrument can result.
Ex applications
This symbol indicates special instructions for Ex applications.
List
The dot set in front indicates a list with no implied sequence.
→
Action
This arrow indicates a single action.
1 Sequence of actions
Numbers set in front indicate successive steps in a procedure.
Battery disposal
This symbol indicates special information about the disposal of bat­teries and accumulators.
2 For your safety
WIKA Operating Instructions - Process pressure transmitter CPT-2x
2 For your safety
2.1 Authorised personnel
All operations described in this operating instructions manual must be carried out only by trained specialist personnel authorised by the plant operator.
During work on and with the device the required personal protective equipment must always be worn.
2.2 Appropriate use
CPT-2x is a slave sensor for electronic dierential pressure measure-
ment.
You can nd detailed information about the area of application in chapter "Product description".
Operational reliability is ensured only if the instrument is properly used according to the specications in the operating instructions manual as well as possible supplementary instructions.
2.3 Warning about incorrect use
Inappropriate or incorrect use of this product can give rise to applica­tion-specic hazards, e.g. vessel overll through incorrect mounting or adjustment. Damage to property and persons or environmental contamination can result. Also, the protective characteristics of the instrument can be impaired.
2.4 General safety instructions
This is a state-of-the-art instrument complying with all prevailing regulations and directives. The instrument must only be operated in a technically awless and reliable condition. The operator is responsi­ble for the trouble-free operation of the instrument. When measuring aggressive or corrosive media that can cause a dangerous situation if the instrument malfunctions, the operator has to implement suitable measures to make sure the instrument is functioning properly.
During the entire duration of use, the user is obliged to determine the compliance of the necessary occupational safety measures with the current valid rules and regulations and also take note of new regula­tions.
The safety instructions in this operating instructions manual, the na­tional installation standards as well as the valid safety regulations and accident prevention rules must be observed by the user.
For safety and warranty reasons, any invasive work on the device beyond that described in the operating instructions manual may be carried out only by personnel authorised by the manufacturer. Arbi­trary conversions or modications are explicitly forbidden. For safety reasons, only the accessory specied by the manufacturer must be used.
2 For your safety
WIKA Operating Instructions - Process pressure transmitter CPT-2x
To avoid any danger, the safety approval markings and safety tips on the device must also be observed and their meaning looked up in this operating instructions manual.
2.5 EU conformity
The device fulls the legal requirements of the applicable EU direc­tives. By axing the CE marking, we conrm the conformity of the instrument with these directives.
2.6 SILqualicationaccordingtoIEC61508
The Safety Integrity Level (SIL) of an electronic system is used to as­sess the reliability of integrated safety functions.
For detailed specication of the safety requirements, multiple SIL levels are specied according to safety standard IEC 61508. You can nd detailed information in chapter "Functional safety (SIL)" of the operating instructions.
The instrument meets the specications of IEC 61508: 2010 (Edi­tion 2). It is qualied for single-channel operation up to SIL2. The instrument can be used homogeneously redundant up to SIL3 in multi-channel architecture with HFT 1.
2.7 Permissible process pressure
For safety reasons, the instrument must only be operated within the permissible process conditions. You can nd detailed information on the process conditions in chapter "Technical data" as well as on the type label.
The permissible process pressure range is specied on the type label with "Process pressure", see chapter "Conguration". This applies even if a measuring cell with a measuring range (order-related) higher than the permissible pressure range of the process tting is installed.
A temperature derating, e.g. with anges, can limit the permissible process pressure range.
2.8 Installation and operation in the USA and Canada
This information is only valid for USA and Canada. Hence the follow­ing text is only available in the English language.
Installations in the US shall comply with the relevant requirements of the National Electrical Code (ANSI/NFPA 70).
Installations in Canada shall comply with the relevant requirements of
the Canadian Electrical Code
3 Product description
WIKA Operating Instructions - Process pressure transmitter CPT-2x
3 Product description
3.1 Conguration
The type label contains the most important data for identication and use of the instrument:
4
1
5
10
12
11
13
14
3
2
6
7
8 9
Fig. 1: Layout of the type label (example)
1 Instrument type
2 Field for approvals
3 Power supply and signal output, electronics 4 Protection rating 5 Measuring range 6 Permissible process conditions
7 Material wetted parts 8 Product code
9 Order number 10 Serial number of the instrument 11 Symbol of the device protection class 12 ID numbers, instrument documentation 13 Reminder to observe the instrument documentation 14 SIL identication
This operating instructions manual applies to the following instrument versions:
Hardware from 1.0.0
Software version from 1.0.0
Note:
You can nd the hardware and software version of the instrument as follows:
On the type plate of the electronics module
In the adjustment menu under "Info"
The scope of delivery encompasses:
Instrument CPT-2x - Slave sensor
Ready-made connection cable, unassembled cable gland
Documentation
Quick setup guide CPT-2xSafety Manual (SIL)
Type label
Scope of this operating instructions manual
Scope of delivery
3 Product description
WIKA Operating Instructions - Process pressure transmitter CPT-2x
Documentation instrument parameters (default values)Documentation order-relating instrument parameters (deviation
from default values)
Test certicate for pressure transmittersInstructions for optional instrument featuresEx-specic "Safety instructions" (with Ex versions)If necessary, further certicates
Information:
Optional instrument features are also described in this operating instructions manual. The respective scope of delivery results from the order specication.
3.2 Principle of operation
The CPT-2x slave sensor is combined with a sensor from the instru­ment series for electronic dierential pressure measurement.
1
2
Fig. 2: Example, electronic dierential pressure for level measurement in pres­surized vessel
1 CPT-2x 2 CPT-2x, slave sensor
The sensors are connected via a screened four-wire cable. The measured value from the slave sensor is read in and factored into the calculations. Power supply and parameter adjustment are carried out through the master sensor.
Electronicdierential
pressure
3 Product description
WIKA Operating Instructions - Process pressure transmitter CPT-2x
Information:
The version "Relative pressure climate-compensated" is not suitable.
You can nd further information in chapter "Combination Master - Slave " of this operating instructions.
To reach the Safety Integrity Level (SIL) for the electronic dierential
pressure, both instruments must be SIL-qualied.
The electronic dierential pressure measurement is suitable for the
measurement of the following process variables:
Level
Flow
Dierential pressure
Density
Interface
Level, density-compensated
CPT-2x is suitable for applications in virtually all industries. It is used for the measurement of the following pressure types.
Gauge pressure
Absolute pressure
Vacuum
Measured products are gases, vapours and liquids. Depending on the process tting and measurement setup, measured
products can be also viscous or contain abrasive substances.
Sensor element is the measuring cell with robust ceramic diaphragm. The process pressure deects the ceramic diaphragm and causes a capacitance change in the measuring cell. This capacitance change is converted into an electrical signal and outputted as measured value via the output signal.
16 µm
1
2
3
Fig. 3: Conguration of the ceramic measuring cell
1 Process diaphragm
2 Glass joint
3 Base element
The measuring cell is available in two sizes:
Measured variables
Application area
Measured products
Measuring system pres­sure
10
3 Product description
WIKA Operating Instructions - Process pressure transmitter CPT-2x
ø 28 mm with large process and ange connections as well as with measuring ranges 25 mbar and 100 bar
ø 17.5 mm with small process ttings
A temperature sensor in the ceramic diaphragm of the ø 28 mm or on the ceramic base of the ø 17.5 mm measuring cell detects the actual process temperature. The temperature value is outputted via the master sensor.
Extreme process temperature jumps are also immediately detected by the ø 28 mm measuring cell. The values are compared with that of an additional temperature measurement on the ceramic base body.
Within only a few measuring cycles the intelligent sensor electronics compensates unavoidable measurement deviations due to tempera­ture shocks. Such shocks cause (depending on the set damping) only slight, brief changes to the output signal.
The measuring cell design depends on the selected pressure type. Relative pressure: the measuring cell is open to the atmosphere.
The ambient pressure is detected in the measuring cell and compen­sated. It thus has no inuence on the measured value.
Absolute pressure: the measuring cell is evacuated and encapsu­lated. The ambient pressure is not compensated and does hence inuence the measured value.
The following presentations show examples for the installation of the ceramic measuring cell into the process tting and the dierent seal concepts.
The recessed installation is particularly suitable for applications with gases, vapours and clear liquids. The measuring cell seal is posi­tioned laterally as well as in addition in front.
Measuring system tem­per
atur
e
Pressure types
Seal concepts
Recessed installation
11
3 Product description
WIKA Operating Instructions - Process pressure transmitter CPT-2x
1
2
5
6
3
4
Fig. 4: Recessed installation of the measuring cell (example: manometer con­nection G½)
1 Measuring cell 2 Seal for the measuring cell 3 Additional, front seal for measuring cell 4 Diaphragm 5 Process tting 6 Seal for the process tting
The front-ush installation is particularly suitable for applications with
viscous and abrasive media and in case of buildup. The measuring cell seal is positioned laterally.
2
3
4
1
5
Fig. 5: Front-ush installation of the measuring cell (example: thread G1½)
1 Seal for the process tting 2 Measuring cell 3 Seal for the measuring cell 4 Process tting 5 Diaphragm
The completely front-ush mounting is particularly suitable for ap-
plications in the paper industry. The diaphragm is in the pulp ow, is hence cleaned and protected against buildup.
Front-ushmountingwith
single seal
Completelyfront-ush
mounting with single seal
12
3 Product description
WIKA Operating Instructions - Process pressure transmitter CPT-2x
1
2
3
4
5
Fig. 6: Front-ush installation of the measuring cell (example: M30 x 1.5)
1 Measuring cell 2 Seal for the measuring cell 3 Seal for the process tting 4 Process tting 5 Diaphragm
The front-ush installation is particularly suitable for applications with viscous media. The additional, front sealing protects the glass joint of the measuring cell against chemical attack and the measuring cell electronics against diusion of aggressive gases from the process.
1
2
3
4
5
Fig. 7: Front-ush installation of the measuring cell with double seal (example: ange connection with extension)
1 Measuring cell 2 Seal for the measuring cell 3 Process tting 4 Additional, front seal for measuring cell 5 Diaphragm
The front-ush, hygienic installation of the measuring cell is particu­larly suitable for food applications. The sealings are installed gap-free. The form seal of the measuring cell protects also the glass joint.
Front-ushmountingwith
double seal
Installation in hygienic
tting
13
3 Product description
WIKA Operating Instructions - Process pressure transmitter CPT-2x
1
2
3
4
5
Fig. 8: Hygienic installation of the measuring cell (example: hygienic tting with compression nut)
1 Measuring cell 2 Form seal for the measuring cell 3 Gap-free seal for process tting 4 Process tting 5 Diaphragm
The front-ush, hygienic installation of the measuring cell acc. to 3A
is particularly suitable for food applications. The sealings are installed gap-free. The additional front sealing for the measuring cell protects also the glass joint. A hole in the process tting is used for leakage
detection.
1
6
2
3
4
5
Fig. 9: Hygienic installation of the measuring cell acc. to 3-A (example: Clamp
connection)
1 Measuring cell 2 Seal for the measuring cell 3 Process tting 4 Additional, front seal for measuring cell 5 Diaphragm 5 Hole for leakage detection
3.3 Supplementary cleaning procedures
The CPT-2x is also available in the version "Oil, grease and silicone- free". These instruments have passed through a special cleaning procedure to remove oil, grease and paint-wetting impairment sub­stances (PWIS).
The cleaning is carried out on all wetted parts as well as on surfaces accessible from outside. To keep the purity level, the instruments are immediately packed in plastic foil after the cleaning process. The purity level remains as long as the instrument is kept in the closed original packaging.
Installation in hygienic
ttingacc.to3-A
14
3 Product description
WIKA Operating Instructions - Process pressure transmitter CPT-2x
Caution:
The CPT-2x in this version may not be used in oxygen applications. For this purpose, instruments are available in the special version "Oil and grease-free for oxygen applications".
3.4 Packaging, transport and storage
Your instrument was protected by packaging during transport. Its capacity to handle normal loads during transport is assured by a test based on ISO 4180.
The packaging of standard instruments consists of environment­friendly, recyclable cardboard. For special versions, PE foam or PE foil is also used. Dispose of the packaging material via specialised recycling companies.
Transport must be carried out in due consideration of the notes on the transport packaging. Nonobservance of these instructions can cause damage to the device.
The delivery must be checked for completeness and possible transit damage immediately at receipt. Ascertained transit damage or con­cealed defects must be appropriately dealt with.
Up to the time of installation, the packages must be left closed and stored according to the orientation and storage markings on the outside.
Unless otherwise indicated, the packages must be stored only under the following conditions:
Not in the open
Dry and dust free
Not exposed to corrosive media
Protected against solar radiation
Avoiding mechanical shock and vibration
Storage and transport temperature see chapter "Supplement ­Technical data - Ambient conditions"
Relative humidity 20 … 85 %
With an instrument weight of more than 18 kg (39.68 lbs) suitable and approved equipment must be used for lifting and carrying.
Packaging
Transport
Transport inspection
Storage
Storage and transport temperature
Lifting and carrying
15
4 Mounting
WIKA Operating Instructions - Process pressure transmitter CPT-2x
4 Mounting
4.1 General instructions
Make sure before mounting that all parts of the instrument exposed to the process are suitable for the existing process conditions.
These are mainly:
Active measuring component
Process tting
Process seal
Process conditions in particular are:
Process pressure
Process temperature
Chemical properties of the medium
Abrasion and mechanical inuences
You can nd detailed information on the process conditions in chapter "Technical data" as well as on the type label.
The instrument is suitable for standard and extended ambient condi-
tions acc. to DIN/EN/IEC/ANSI/ISA/UL/CSA 61010-1.
Protect your instrument against moisture ingress through the following measures:
Use a suitable connection cable (see chapter "Connecting to power supply")
Tighten the cable gland
When mounting horizontally, turn the housing so that the cable gland points downward
Loop the connection cable downward in front of the cable gland
This applies mainly to outdoor installations, in areas where high humidity is expected (e.g. through cleaning processes) and on cooled or heated vessels.
To maintain the housing protection, make sure that the housing lid is closed during operation and locked, if necessary.
Make sure that the degree of contamination specied in chapter "Technical data" meets the existing ambient conditions.
On instruments with threaded process tting, the hexagon must be tightened with a suitable wrench. For the proper wrench size see chapter "Dimensions".
Warning:
The housing must not be used to screw the instrument in! Applying tightening force can damage internal parts of the housing.
If there is strong vibration at the mounting location, the instrument version with external housing should be used. See chapter "External housing".
Suitability for the process conditions
Suitability f
or the ambient
conditions
Pr
otection against mois-
ture
Screwing in
Vibrations
16
4 Mounting
WIKA Operating Instructions - Process pressure transmitter CPT-2x
Higher process temperatures often mean also higher ambient temperatures. Make sure that the upper temperature limits stated in chapter "Technical data" for the environment of the electronics hous- ing and connection cable are not exceeded.
1
2
Fig. 10: Temperature ranges
1 Process temperature 2 Ambient temperature
4.2 Ventilation and pressure compensation
Ventilation and pressure compensation are carried out with CPT-2x via a lter element. It is air permeable and moisture-blocking.
Caution:
The lter element causes a time-delayed pressure compensation. When quickly opening/closing the housing cover, the measured value can change for approx. 5 s by up to 15 mbar.
For eective ventilation, the lter element must always be free of buildup.
Caution:
Do not use a high-pressure cleaner. The lter element could be dam­aged, which would allow moisture into the housing.
The following paragraphs describe how the lter element is arranged in the dierent instrument versions.
The lter element is mounted into the electronics housing. It has the following functions:
Ventilation of the electronics housing
Atmospheric pressure compensation (with relative pressure meas­uring ranges)
Turn the housing so that the lter element points downward after the instrument is installed. This provides better protection against buildup.
Temperature limits
Filter elements
Instruments in non-Ex and Ex-ia version
17
4 Mounting
WIKA Operating Instructions - Process pressure transmitter CPT-2x
1 2 3
4 4 4
Fig. 11: Position of the lter element - non-Ex, Ex-ia version
1 Plastic, stainless steel housing (precision casting) 2 Aluminium housing 3 Stainless steel housing (electropolished) 4 Filter element
With the following instruments a blind plug is installed instead of the lter element:
Instruments in protection IP 66/IP 68 (1 bar) - ventilation via capil­laries in non-detachable cable
Instruments with absolute pressure
The lter element is integrated in the process assembly. It is located in
a rotatable metal ring and has the following function:
Atmospheric pressure compensation (with relative pressure meas­uring ranges)
Turn the metal ring in such a way that the lter element points downward after installation of the instrument. This provides better protection against buildup.
1
2
Fig. 12: Position of the lter element - Ex-d version
1 Rotatable metal ring 2 Filter element
Instruments with absolute pressure have a blind plug mounted instead of the lter element.
The lter element is mounted into the electronics housing. It has the
following functions:
Ventilation of the electronics housing
Turn the housing so that the lter element points downward after the instrument is installed. This provides better protection against buildup.
Instruments in Ex-d ver­sion
Instruments with Second Line of Def
ense
18
4 Mounting
WIKA Operating Instructions - Process pressure transmitter CPT-2x
The process assembly of instruments with Second Line of Defense (gastight leadthrough) is completely encapsulated. An absolute pres­sure measuring cell is used so that no ventilation is required.
1
2
Fig. 13: Position of the lter element - gastight leadthrough
1 Filter element
The lter element is mounted into the electronics housing. It has the following functions:
Ventilation of the electronics housing
Atmospheric pressure compensation (with relative pressure meas­uring ranges)
Turn the housing so that the lter element points downward after the instrument is installed. This provides better protection against buildup.
1
Fig. 14: Position of the lter element - IP 69K version
1 Filter element
Instruments with absolute pressure have a blind plug mounted instead of the lter element.
4.3 Combination Master - Slave
In principle, any sensor combination within the instrument series is allowed. The following requirements must be fullled:
Conguration, Master sensor suitable for electronic dierential pressure
Pressure type is identical for both sensors, i.e. relative pressure/ relative pressure or absolute pressure/absolute pressure
Master sensor measures the higher pressure
Measurement setup as shown in the following chapters
Instruments in IP 69K version
19
4 Mounting
WIKA Operating Instructions - Process pressure transmitter CPT-2x
The measuring range of each sensor is selected such that it ts the measuring loop. For this, the max. recommended turn down must be noted. See chapter "Technical data". It is absolutely necessary the the measuring ranges of Master and Slave correspond.
Measurement result = Measured value of Master (total pres­sure) - measured value of Slave (static pressure)
Depending on the application, individual combinations can result, see following examples:
Data
Application: Level measurement Medium: Water Vessel height: 12 m, hydrostatic pressure = 12 m x 1000 kg/m3 x
9.81 m/s2 = 117.7 kPa = 1.18 bar
Superimposed pressure: 1 bar Total pressure: 1.18 bar + 1 bar = 2.18 bar
Instrument selection
Nominal measuring range Master: 2.5 bar Nominal measuring range Slave: 1 bar Turn Down: 2.5 bar/1.18 bar = 2.1 : 1
Data
Application: Level measurement Medium: Water Vessel height: 500 mm, hydrostatic pressure = 0.50 m x 1000 kg/m3 x
9.81 m/s2 = 4.9 kPa = 0.049 bar
Superimposed pressure: 350 mbar = 0.35 bar Total pressure: 0.049 bar + 0.35 bar = 0.399 bar
Instrument selection
Nominal measuring range Master: 0.4 bar Nominal measuring range Slave: 0.4 bar Turn Down: 0.4 bar /0.049 bar = 8.2 : 1
Data
Application: Dierential pressure measurement Medium: Gas Static pressure: 0.8 bar Dierential pressure on orice: 50 mbar = 0.050 bar Total pressure: 0.8 bar + 0.05 bar = 0.85 bar
Instrument selection
Nominal measuring range Master: 1 bar Nominal measuring range Slave: 1 bar Turn Down: 1 bar/0.050 bar = 20 : 1
Example - large vessel
Example - small vessel
Example-oricein
pipeline
20
4 Mounting
WIKA Operating Instructions - Process pressure transmitter CPT-2x
The measuring result (level, pressure dierence) as well as measured value Slave (static or superimposed pressure) are outputted by the sensor. Depending on the instrument version, output as 4 … 20 mA signal or digitally via HART, Probus PA or Foundation Fieldbus.
To reach the Safety Integrity Level (SIL) for the electronic dierential pressure, both instruments must be SIL-qualied.
4.4 Level measurement
The master/slave combination is suitable for level measurement in a pressurized vessel
Keep the following in mind when setting up the measuring system:
Mount the master sensor below the min. level
Do not mount the master sensor close to the lling stream or emptying area
Mount the master sensor so that it is protected against pressure shocks from the stirrer
Mount the slave sensor above the max. level
2
Fig. 15: Measurement setup, level measurement in pressurized vessel
1 CPT-2x 2 CPT-2x, slave sensor
Output measured values
Measurement setup
21
4 Mounting
WIKA Operating Instructions - Process pressure transmitter CPT-2x
4.5 Dierentialpressuremeasurement
The master/slave combination is suitable for dierential pressure
measurement Take note of the following instructions for the measurement setup, for
example in gases:
Mount the instruments above the measuring point
Possible condensation can then drain o into the process line.
1
2
Fig. 16: Measurement setup for dierential pressure measurement of gases in
pipelines
1 CPT-2x 2 CPT-2x, slave sensor
4.6 Interface measurement
The master/slave combination is suitable for interface measurement
Requirements for a functioning measurement are:
Vessel with changing level
Products with steady density
Interface always between the measurement points
Total level always above the upper measurement point
The mounting distance h of the two sensors should be at least 10 %, better 20 %, of the nal value of the sensor measuring range. A bigger distance increases the accuracy of the interface measurement.
Measurement setup
Measurement setup
22
4 Mounting
WIKA Operating Instructions - Process pressure transmitter CPT-2x
h
1,0
0,8
1
2
Fig. 17: Measurement setup with interface measurement, h = distance between the two measuring points
1 CPT-2x 2 CPT-2x, slave sensor
The interface measurement is possible in open as well as in closed vessels.
4.7 Density measurement
The master/slave combination is suitable for density measurement. Requirements for a functioning measurement are:
Vessel with changing level
Distance between the measurement points as large as possible
Level always above the upper measuring point
Measurement setup
23
4 Mounting
WIKA Operating Instructions - Process pressure transmitter CPT-2x
h
1
2
Fig. 18: Measurement setup for density measurement, h = distance between the two measuring points
1 CPT-2x 2 CPT-2x, slave sensor
The mounting distance h of the two sensors should be at least 10 %, better 20 %, of the nal value of the sensor measuring range. A bigger distance increases the accuracy of the density measurement.
Slight density changes cause only slight changes of the measured dierential pressure. The measuring range must hence be selected accordingly.
The density measurement is possible in open as well as in closed vessels.
4.8 Density-compensated level measurement
The master/slave combination is suitable for density-compensated
level measurement Keep the following in mind when setting up the measuring system:
Mount the master sensor below the min. level
Mount the slave sensor above the master sensor
Mount both sensors away from the lling stream and emptying and protected against pressure shocks from the stirrer
Measurement setup
Loading...
+ 53 hidden pages