WIKA CPP7-H Specifications

Page 1
Pneumatic hand test pump Model CPP7-H
Applications
Simple test pressure generation on site, in the laboratory or in the workshop
Supply of small positive or negative pressures
Specically for the testing, adjustment and calibration
of low-pressure measuring instruments in the areas of: heating, ventilation and air conditioning, clean room
engineering, medical engineering and ltration and dust
removal technology
Calibration technology
WIKA data sheet CT 91.02
Special features
Switchable pressure/vacuum generation
Precise setting through the ne adjustment valve
Simple and ergonomic handling
Compact dimensions
Low weight
Description
Application areas
Hand test pumps serve as pressure generators for the testing,
adjustment and calibration of mechanical and electronic
pressure measuring instruments through comparative measurements. These pressure tests can take place in the laboratory or workshop, or on site at the measuring point.
Easy operation
If one connects the test item and a suciently accurate
reference measuring instrument to the hand test pump, on actuating the pump, the same pressure will act on both measuring instruments. By comparison of the two measured values at any given pressure value, a check of the accuracy
and/or adjustment of the pressure measuring instrument
under test can be carried out. In order to approach the
measuring points exactly, the hand test pump features a ne adjustment valve.
Pneumatic hand test pump model CPP7-H
Easy handling
The CPP7-H is a pneumatic hand test pump for low-pressure ranges up to 7 bar (101.5 psi), with switching to vacuum generation down to -800 mbar (-11.6 psi). The maximum pressure or vacuum achievable is dependent upon the connected test volume. Through the smooth pressure generation and the integrated
ne adjustment valve, very small positive or negative
pressures in the mbar range can be set safely and precisely.
Via a T-connector with a hose connection, the test item and the reference measuring instrument can be easily connected.
12/2016
Data sheets showing similar products and accessories: Hand test pump, pneumatic; model CPP30; see data sheet CT 91.06 Test pumps, hydraulic; CPP series; see data sheet CT 91.05 Hand-held pressure indicator; model CPH6200; see data sheet CT 11.01 ProcessCalibrator; model CPH6000; see data sheet CT 15.01
Page 1 of 4WIKA data sheet CT 91.02
Page 2
Specications Model CPP7-H
Pressure range -0.8 ... +7 bar
(-11.6 ... 101.5 psi)
Pressure transmission medium Air
Pressure connections 1 connection with G ⅛ female thread to the pump body, prepared for connection to a T-connector
with 2 hoses and freely selectable connections to test item
Fine adjustment Fine adjustment valve
Dimensions (L x W x H) 175.3 x 111 x 41 mm (6.90 x 4.37 x 1.61 in) without T-connector
175.3 x 144 x 41 mm (6.90 x 5.67 x 1.61 in) with T-connector
Weight 293 g (0.65 lbs) without T-connector
334 g (0.74 lbs) with T-connector
Dimensions in mm (in)
Isometric view
1
3
2
2
3
1
Connecting hose; length 0.5 m (1.64 ft)
2
Hose connection adapter G ⅛ male to 3 mm quick
connector
3
Pressure connections for reference measuring instrument and test item can be freely selected
1
4
7
5
(optionally G ⅛ female to G ½ female, G ¼, ½ NPT or ¼ NPT)
4
Fine adjustment valve
5
Pump handle
6
Ram
7
T-connector, 2 x G ⅛ female to G ⅛ male
6
Page 2 of 4 WIKA data sheet CT 91.02 ∙ 12/2016
Page 3
Recommended reference pressure measuring instruments
Precision digital pressure gauge model CPG1500
Measuring ranges: up to 1,000 bar (15,000 psi)
Accuracy: 0.05 % of span
For further specications see data sheet CT 10.51
Hand-held pressure indicator model CPH6200
Measuring ranges:
Accuracy:
For further specications see data sheet CT 11.01
Hand-held pressure indicator model CPH6300
Measuring ranges: up to 1,000 bar (14,500 psi)
Accuracy: 0.2 % of span
For further specications see data sheet CT 12.01
Precision hand-held pressure indicator model CPH6400
Measuring ranges: up to 6,000 bar (85,000 psi)
Accuracy: up to 0.025 % of span
For further specications see data sheet CT 14.01
up to 1,000 bar (14,500 psi)
0.2 % of span
ProcessCalibrator model CPH6000
Measuring ranges: up to 8,000 bar (115,000 psi)
Accuracy: up to 0.025 % of span
For further specications see data sheet CT 15.01
Calibration software
WIKA-CAL calibration software for creating calibration
certicates or logger protocols
For specications see data sheet CT 95.10
WIKA data sheet CT 91.02 ∙ 12/2016 Page 3 of 4
Page 4
Calibration case
Consisting of:
Plastic service case with foam insert
Pneumatic hand test pump model CPP7-H
Standard accessories see scope of delivery
Scope of delivery
Hand test pump model CPP7-H
T-connector
4 hose connection adapters
2 connecting hoses; each 0.5 m long (1.64 ft)
Selectable connection adapter (G ½, G ¼, ½ NPT or ¼ NPT)
PTFE sealing tape
Accessories
Adapter
Connection adapter G ⅛ female to G ½ female,
material stainless steel
Connection adapter G ⅛ female to G ¼ female,
Case
Plastic case, incl. foam insert with free spaces for model CPP7-H; dimensions (W x H x D): 350 x 265 x 85 mm (13.78 x 10.53 x 3.35 in)
material stainless steel
Connection adapter G ⅛ female to ½ NPT female,
material stainless steel
Connection adapter, G ⅛ female to ¼ NPT female,
Other
T-connector 2 x G ⅛ female to G ⅛ male, material: brass
Spare hose for model CPP7-H; length 0.5 m (1.64 ft)
material stainless steel
Hose connection adapter G ⅛ male to 4 mm hose
connection
Ordering information
CPP7-H / Standard connection adapter / Additional connection adapter / Case / Further approvals / Additional ordering information
© 12/2016 WIKA Alexander Wiegand SE & Co. KG, all rights reserved.
The specications given in this document represent the state of engineering at the time of publishing. We reserve the right to make modications to the specications and materials.
Page 4 of 4 WIKA data sheet CT 91.02 ∙ 12/2016
WIKA Alexander Wiegand SE & Co. KG
Alexander-Wiegand-Straße 30 63911 Klingenberg/Germany Tel. +49 9372 132-0 Fax +49 9372 132-406 info@wika.de www.wika.de
12/2016 EN
Loading...