3WIKA operating instructions, models CPP700-H and CPP1000-H
1. General information
1. General information
■
The hand test pumps CPP700-H and CPP1000-H described in the operating
instructions have been designed and manufactured using state-of-the-art technology.
EN
All components are subject to stringent quality and environmental criteria during
production. Our management systems are certified to ISO 9001 and ISO 14001.
■
These operating instructions contain important information on handling the instrument.
Working safely requires that all safety instructions and work instructions are observed.
■
Observe the relevant local accident prevention regulations and general safety
regulations for the instrument's range of use.
■
The operating instructions are part of the product and must be kept in the immediate
vicinity of the instrument and readily accessible to skilled personnel at any time. Pass
the operating instructions onto the next operator or owner of the instrument.
■
Skilled personnel must have carefully read and understood the operating instructions
prior to beginning any work.
■
The general terms and conditions contained in the sales documentation shall apply.
4WIKA operating instructions, models CPP700-H and CPP1000-H
14209601.01 06/2017 EN/DE/FR/ES/IT
2. Design and function
2. Design and function
2.1 Overview
Front view
1
7
6
5
1
Test item connection hose
2
CPP700-H: Minimess
3
4
®
system with a G ¼ female pressure gauge bulkhead
481
connector
CPP1000-H: G ¼ female
2
Pressure relief valve
3
Plug screw for fluid reservoir
4
Fine adjustment valve (volume adjustment)
5
Pump handles
6
Toggle control, system and high-pressure generation
7
Fluid reservoir
8
Connection G ½ female, freely rotating for reference instruments
9
Blind plug, do not open!
Rear view
EN
9
7
6
5
14209601.01 06/2017 EN/DE/FR/ES/IT
5WIKA operating instructions, models CPP700-H and CPP1000-H
2. Design and function
2.2 Description
The hand test pumps CPP700-H and CPP1000-H serve as pressure generators for the
testing, adjustment and calibration of mechanical and electronic pressure measuring
instruments through comparative measurements. These pressure tests can take place in
the laboratory or workshop, or on site at the measuring point.
EN
By connecting the pressure measuring instrument to be tested and a sufficiently accurate
reference measuring instrument to the CPP700-H or CPP1000-H, on actuating the pump,
the same pressure will act on both measuring instruments. By comparison of the two
measured values at any given pressure value, a check of the accuracy and/or adjustment
of the pressure measuring instrument under test can be carried out.
The hand test pumps, despite their very compact dimensions, enable simple and
accurate pressure generation. The CPP700-H and CPP1000-H have a fine adjustment
valve for precise adjustment, so as to ensure an accurate comparison test. The reference
instrument is screwed directly to the top of the pump, and the test item is connected using
the connecting hose included in delivery.
2.3 Scope of delivery
CPP700-H
■
Hand test pump, model CPP700-H, oil/water
■
Minimess® test item hose, length 1 m (3.28 ft)
■
Bulkhead connector, G ¼ female
■
Operating instructions
CPP1000-H
■
Hand test pump, model CPP1000-H, oil/water
■
High-pressure test item hose, length 1 m (3.28 ft)
■
Connection, G ¼ female
■
Operating instructions
Cross-check scope of delivery with delivery note.
2.4 Pressure transmission medium
For CPP700-H and CPP1000-H, the following media are suitable as pressure
transmission media:
■
Hydraulic fluids based on mineral oil (water-based hydraulic fluids are not suitable)
■
Simple distilled water
Other pressure transmission media on request.
6WIKA operating instructions, models CPP700-H and CPP1000-H
14209601.01 06/2017 EN/DE/FR/ES/IT
2. Design and function / 3. Safety
Handling of mineral oils
WIKA delivers hydraulic fluid based on mineral oil in containers of 1 litre, labelled “Shell
Tellus S2 M22”, for use in hand test pumps with up to 1,000 bar (14,500 psi). It is no more
hazardous than other common lubricating oils.
Before working with hydraulic fluids, read the material safety data sheet.
The current material safety data sheet can be found at www.wika.de on the
product page of the respective product.
3. Safety
3.1 Explanation of symbols
WARNING!
... indicates a potentially dangerous situation that can result in serious injury
or death, if not avoided.
CAUTION!
... indicates a potentially dangerous situation that can result in light injuries
or damage to property or the environment, if not avoided.
EN
Information
... points out useful tips, recommendations and information for efficient and
trouble-free operation.
3.2 Intended use
Test pumps serve as pressure generators for the testing, adjustment and calibration
of mechanical and electronic pressure measuring instruments through comparative
measurements.
These pressure tests can take place in the laboratory or workshop, or on site at the
measuring point. The model CPP700-H and CPP1000-H hand test pumps have been
specifically developed for test pressure generation on site.
The hand test pumps, despite their very compact dimensions, enable simple
and accurate test pressure generation of up to a max. of 700 bar (10,000 psi) or
1,000 bar (14,500 psi) respectively, with an integrated switching from initial pressure to
high-pressure generation.
14209601.01 06/2017 EN/DE/FR/ES/IT
7WIKA operating instructions, models CPP700-H and CPP1000-H
3. Safety
This instrument is not permitted to be used in hazardous areas!
The instrument has been designed and built solely for the intended use described here,
and may only be used accordingly.
EN
The technical specifications contained in these operating instructions must be observed.
Improper handling or operation of the instrument outside of its technical specifications
requires the instrument to be taken out of service immediately and inspected by an
authorised WIKA service engineer.
Handle the test pump with the required care (protect from humidity, impacts, strong
magnetic fields, static electricity and extreme temperatures, do not insert any objects into
the instrument or its openings). Plugs and sockets must be protected from contamination.
The manufacturer shall not be liable for claims of any type based on operation contrary to
the intended use.
3.3 Improper use
WARNING!
Injuries through improper use
Improper use of the instrument can lead to hazardous situations and injuries.
▶
Refrain from unauthorised modifications to the instrument.
▶
Do not use the instrument within hazardous areas.
▶
Do not use the instrument with abrasive or viscous media.
Any use beyond or different to the intended use is considered as improper use.
3.4 Personnel qualification
WARNING!
Risk of injury should qualification be insufficient
Improper handling can result in considerable injury and damage to
equipment.
▶
The activities described in these operating instructions may only be
carried out by skilled personnel who have the qualifications described
below.
Skilled personnel
Skilled personnel, authorised by the operator, are understood to be personnel who,
based on their technical training, knowledge of measurement and control technology and
on their experience and knowledge of country-specific regulations, current standards and
8WIKA operating instructions, models CPP700-H and CPP1000-H
14209601.01 06/2017 EN/DE/FR/ES/IT
3. Safety
directives, are capable of carrying out the work described and independently recognising
potential hazards.
3.5 Personal protective equipment
The personal protective equipment is designed to protect the skilled personnel from
hazards that could impair their safety or health during work. When carrying out the various
tasks on and with the instrument, the skilled personnel must wear personal protective
equipment.
Follow the instructions displayed in the work area regarding personal protective
equipment!
The requisite personal protective equipment must be provided by the operating company.
Wear safety goggles!
Protect eyes from flying particles and liquid splashes.
3.6 Labelling, safety marks
Product label
The product label is attached to the glass body of the fluid reservoir.
1
Kalibriertechnik / Calibration Technology
Handprüfpumpe / Hand test pump
Druckbereich / Pressure range:
Medium / medium:
Serien-Nr. / serial no.:
Alexander Wiegand SE & Co. KG 63911 Klingenberg / Germany www.wika.com
CPP700-H
0 ... 700 bar
7XXXX
11/2015
1
2
3
4
5
Date of manufacture
2
Model
3
Pressure range
4
Medium
5
Serial no.
EN
Symbol
Before mounting and commissioning the instrument, ensure you read the
operating instructions!
14209601.01 06/2017 EN/DE/FR/ES/IT
9WIKA operating instructions, models CPP700-H and CPP1000-H
4. Transport, packaging and storage
4. Transport, packaging and storage
4.1 Transport
Check the hand test pump for any damage that may have been caused by transport.
EN
Obvious damage must be reported immediately.
CAUTION!
Damage through improper transport
With improper transport, a high level of damage to property can occur.
▶
When unloading packed goods upon delivery as well as during internal
transport, proceed carefully and observe the symbols on the packaging.
▶
With internal transport, observe the instructions in chapter 4.2 “Packaging
and storage”.
If the instrument is transported from a cold into a warm environment, the formation of
condensation may result in instrument malfunction. Before putting it back into operation,
wait for the instrument temperature and the room temperature to equalise.
4.2 Packaging and storage
Do not remove packaging until just before mounting.
Keep the packaging as it will provide optimum protection during transport (e.g. change in
installation site, sending for repair).
Permissible conditions at the place of storage:
■
Storage temperature: 0 ... 55 °C (32 ... 131 °F)
■
Humidity: 35 ... 85 % relative humidity (no condensation)
Avoid exposure to the following factors:
■
Direct sunlight or proximity to hot objects
■
Mechanical vibration, mechanical shock (putting it down hard)
■
Soot, vapour, dust and corrosive gases
■
Hazardous environments, flammable atmospheres
■
Corrosive liquids
Store the hand test pump in its original packaging in a location that fulfils the conditions
listed above. If the original packaging is not available, pack and store the instrument as
described below:
1. Place the instrument, along with the shock-absorbent material, in the packaging.
2. If stored for a prolonged period of time (more than 30 days), place a bag containing a
desiccant inside the packaging.
10WIKA operating instructions, models CPP700-H and CPP1000-H
14209601.01 06/2017 EN/DE/FR/ES/IT
5. Commissioning, operation
5. Commissioning, operation
Personnel: Skilled personnel
Protective equipment: Safety goggles,
Tools: Syringe and a special needle for filling or cleaning the fluid reservoir
Only use original parts (see chapter 10 “Accessories”).
5.1 Unpacking the hand test pump
As soon as possible after delivery open the packaging of the hand test pump and check
that all the items detailed in the packing list (see chapter 2.3 “Scope of delivery”) are
included.
Unpacking the items, check them for any damage that may have been caused by
transport. If any items are missing, please contact WIKA immediately.
5.2 Ambient conditions
The pressure tests can take place in the laboratory or workshop, or on site at the
measuring point.
5.3 Design
EN
Front view
1
7
6
5
14209601.01 06/2017 EN/DE/FR/ES/IT
2
Rear view
3
4
481
9
7
6
5
11WIKA operating instructions, models CPP700-H and CPP1000-H
5. Commissioning, operation
1
Test item connection hose
CPP700-H: Minimess
connector
CPP1000-H: G ¼ female
EN
2
Pressure relief valve
3
Plug screw for fluid reservoir
4
Fine adjustment valve (volume adjustment)
5
Pump handles
6
Toggle control, system and high-pressure generation
7
Fluid reservoir
8
Connection G ½ female, freely rotating for reference instruments
9
Blind plug, do not open!
5.4 Filling the hand test pump with the pressure transmission medium
1. Open the fluid reservoir
2. Fill the pressure transmission medium in the pump using the syringe or a special
needle. For suitable medium, see chapter 2.4 “Pressure transmission medium”.
CAUTION!
Damage to the hand test pump due to an unsuitable pressure
transmission medium
Unsuitable pressure transmission media attack the sealings and may lead to
leaks and damage of the hand test pump.
▶
Only use the oils supplied by us or authorized spare oils for hydraulic
systems.
▶
Do not use any other liquids such as castor oils, Skydrol, solvents or the
like.
®
system with a G ¼ female pressure gauge bulkhead
7
. To do so, unscrew the plug screws 3.
The maximum permissible temperature of the pressure transmission
medium is 70 °C.
7
3. The fluid reservoir
4. Screw in the plug screw
must not be filled to 100 %.
3
again.
5.5 Replacing the pressure transmission medium
7
1. Open the fluid reservoir
. To do so, unscrew the plug screws 3.
2. Remove the pressure transmission medium from the pump using the syringe or a
special needle.
12WIKA operating instructions, models CPP700-H and CPP1000-H
14209601.01 06/2017 EN/DE/FR/ES/IT
5. Commissioning, operation
3. Rinse the fluid reservoir using water or gentle cleaning agents.
To do so, use a syringe specially designed for cleaning.
⇒
The use of alcohol, spirit, acetone or benzene can cause the container
material to crack or lead to damage of the container.
4. Fill the new pressure transmission medium in the pump using the syringe or a special
needle.
5. The fluid reservoir 7 must not be filled to 100 %.
3
6. Screw in the plug screw
If water is used as test or calibration medium, the test pump must be
emptied after EACH use.
As soon as contamination is visible in the calibration medium, the medium
must be replaced.
5.6 Mechanical mounting
WARNING!
Breakage of the test item hose
Air pockets in the calibration circuit can cause the test item hose to break
during pressure generation.
To prevent this and to maintain a more constant pressure, the following must
be ensured.
▶
Prior to using the hose, fill it with the service liquid.
▶
Fill up with the missing pressure transmission medium, see chapter
5.4 “Filling the hand test pump with the pressure transmission medium”.
again.
EN
CAUTION!
Damage to the hand test pump due to external pressure
To prevent this, the following must be ensured.
▶
Do not connect any external pressure sources to the pump.
▶
Generate pressure only using the hand test pump.
1. Mount the reference pressure measuring instrument on the top side
test pump.
It is sufficient to tighten it by hand.
The reference pressure measuring instrument is sealed by means of the O-ring
⇒
14209601.01 06/2017 EN/DE/FR/ES/IT
integrated in the reference instrument connection.
8
of the hand
13WIKA operating instructions, models CPP700-H and CPP1000-H
5. Commissioning, operation
2. Mount the test item at the end of the test item hose 1.
G 1/4 female thread as a standard
3. Tighten adapters, if used, to a maximum torque of 15 Nm.
EN
A stainless steel set of adapters is available as accessories. The set
contains a set of high-quality nylon sealings or O-rings.
Dirty or chemically contaminated test items should not be fitted as they
contaminate the system unless they are first cleaned.
Alternative mounting of the test item
1
1. Remove the test item hose
from the pump by turning it.
2. Select an adapter suitable for the desired test item (optional set of adapters available
on request).
3. Mount the adapter on the hand test pump (G 1/4 female thread).
4. Mount the test item on the adapter.
This makes it possible to substantially reduce volume and to facilitate the test
⇒
pressure generation.
5.7 Generating pressure
Handles for initial pressure
generation
6
Handles for high-pressure
generation
6
5
5
Pump handles
6
Toggle control, system and high-pressure generation
1. Before generating pressure, press the toggle control
The pump handles 5 are positioned apart (“initial pressure generation” position).
⇒
2. Open the pressure relief valve
14WIKA operating instructions, models CPP700-H and CPP1000-H
2
.
5
6
, if necessary.
14209601.01 06/2017 EN/DE/FR/ES/IT
5. Commissioning, operation
3. Unscrew the fine adjustment valve 4 by turning it counter-clockwise, until it slightly
stops.
2
4. Carefully screw the pressure relief valve
closes securely.
5
5. By means of the pump handle
generated.
The handles can now hardly be pressed together. Depending on the volume of the
⇒
calibration circuit, a pressure of 200 ... 400 bar (2,900 ... 5,800 psi) is generated in
the “initial pressure generation” position.
5
6. Hold the handles
The pump handles 5 are now in the “high-pressure generation” position.
⇒
If too high an initial pressure has been generated, the pump handles
no longer be completely pressed together and the toggle control
longer be actuated
In this case, drain the built-up initial pressure by unscrewing the pressure
relief valve
pressed together and press the toggle control 6.
2
.
, pump the system until an initial pressure is
by turning it clockwise, until the valve
6
can no
5
can
EN
7. Now, by means of the pump handles
reached, but up to a maximum of about 600 bar (8,700 psi).
A test pressure higher than about 600 bar (8,700 psi) is achieved only by turning
⇒
the fine adjustment valve
After increasing the pressure, the value displayed can drop slightly again
for about 30 seconds. The causes can be thermodynamic effects, the hose
connection and the sealings.
Readjust the test pressure using the fine adjustment valve
continues to drop, check the measuring circuit for leaks.
5.8 Reducing pressure
Do not disassemble the reference instrument or the test item until the
pressure relief valve
1. To reduce the pressure, turn the fine adjustment valve counterclockwise.
2. Carefully open the pressure relief valve
Pressure is reduced and there is no pressure on the hand test pump.
⇒
3. Dismount the reference pressure measuring instrument and/or the test item.
4. After the pressure has been relieved, close the pressure relief valve
clockwise.
14209601.01 06/2017 EN/DE/FR/ES/IT
4
clockwise.
2
, pump until the desired test pressure is
5
is open and there is no pressure in the test pump.
2
by turning it clockwise no more than once.
4
. If pressure
2
by turning it
15WIKA operating instructions, models CPP700-H and CPP1000-H
Tools: Syringe and a special needle for filling or cleaning the fluid reservoir
CAUTION!
Physical injuries and damage to property and the environment
If faults cannot be eliminated by means of the listed measures, the hand test
pump must be taken out of operation immediately.
▶
Ensure that pressure or signal is no longer present and protect against
accidental commissioning.
▶
Contact the manufacturer.
▶
If a return is needed, please follow the instructions given in chapter
8.2 “Return”.
For contact details see chapter 1 “General information” or the back page of
the operating instructions.
FaultsCausesMeasures
The hand test pump is
sluggish
Instable output
pressure
System provides
pressure but pressure
decays to zero.
The hand test pump has not been
used for a long time
The drain valve is openClose drain valve
Incorrect sealingInsert the correct sealing
Incorrectly seated sealingPosition the sealing correctly
Sealing is defectiveReplace the sealing with a new
Incorrect adapter or the adapter
has not been mounted properly
If unable to locate a cause.Return the hand test pump for
The drain valve is openClose drain valve
Incorrect operating procedure
being used.
If unable to locate a cause.Return the hand test pump for
The first stroke is somehow
sluggish. This effect has
disappeared further operation.
one
Check the adapter at the test item
side and mount it properly
checking
Ensure that correct operating
procedure is being followed
(see chapter 5.7 “Generating
pressure”)
checking
18WIKA operating instructions, models CPP700-H and CPP1000-H
14209601.01 06/2017 EN/DE/FR/ES/IT
6. Faults / 7. Maintenance and cleaning
FaultsCausesMeasures
System provides
pressure but pressure
decays to lower value
then remains steady.
Insufficient liquid in the hand test
pump.
Air in the systemFill the fluid reservoir with
If unable to locate a cause.Return the hand test pump for
Internal damageReturn the hand test pump for
Incorrect operating procedure
being used.
Fill the fluid reservoir with
the correct fluid (see chapter
5.4 “Filling the hand test pump
with the pressure transmission
medium”)
the correct fluid (see chapter
5.4 “Filling the hand test pump
with the pressure transmission
medium”)
checking
checking
Ensure that correct operating
procedure is being followed
(see chapter 5.7 “Generating
pressure”)
EN
7. Maintenance and cleaning
Personnel: Skilled personnel
Protective equipment: Safety goggles
Tools: Syringe and a special needle for filling or cleaning the fluid reservoir
For contact details see chapter 1 “General information” or the back page of
the operating instructions.
7.1 Maintenance
Repairs must only be carried out by the manufacturer.
Only use original parts (see chapter 10 “Accessories”).
▶
Before adapting the reference instrument and the test item, check the sealings in the
two connections for correct seating and wear.
▶
Replace the sealings, if necessary.
14209601.01 06/2017 EN/DE/FR/ES/IT
19WIKA operating instructions, models CPP700-H and CPP1000-H
7. Maintenance and cleaning
7.2 Cleaning
CAUTION!
Physical injuries and damage to property and the environment
EN
1. Before cleaning, correctly disconnect the instrument from the pressure supply.
2. Use the requisite protective equipment.
3. Clean the instrument with a moist cloth.
Improper cleaning may lead to physical injuries and damage to property and
the environment. Residual media in the dismounted instrument can result in
a risk to persons, the environment and equipment.
▶
Carry out the cleaning process as described below.
CAUTION!
Damage to the instrument
Improper cleaning may lead to damage to the instrument!
▶
Do not use any aggressive cleaning agents.
▶
Do not use any pointed and hard objects for cleaning.
▶
Only clean the fluid reservoir using water or gentle cleaning agents.
The use of alcohol, spirit, acetone or benzol can cause the container
material to crack or lead to damage of the container.
4. Wash or clean the dismounted instrument, in order to protect persons and the
environment from exposure to residual media.
Cleaning the fluid reservoir
Clean the fluid reservoir, see chapter 5.5 “Replacing the pressure transmission medium”.
20WIKA operating instructions, models CPP700-H and CPP1000-H
14209601.01 06/2017 EN/DE/FR/ES/IT
8. Dismounting, return and disposal
8. Dismounting, return and disposal
Personnel: Skilled personnel
Protective equipment: Safety goggles
Tools: Syringe and a special needle for filling or cleaning the fluid reservoir
8.1 Dismounting
WARNING!
Physical injuries and damage to property and the environment
through residual media
When dismounting, there is a danger from aggressive media and high
pressures.
▶
Before storage of the dismounted instrument (following use) wash or
clean it, in order to protect persons and the environment from exposure to
residual media.
▶
Wear the requisite protective equipment (see chapter 3.5 “Personal
protective equipment”).
▶
Observe the information in the material safety data sheet for the
corresponding medium.
▶
Only disconnect the measuring assembly/test and calibration installations
once the system has been depressurised.
EN
Do not disassemble the reference instrument or the test item until the
2
pressure relief valve
1. To reduce the pressure, turn the fine adjustment valve counterclockwise.
2. Carefully open the pressure relief valve
Pressure is reduced and there is no pressure on the hand test pump.
⇒
3. Dismount the reference pressure measuring instrument and/or the test item.
4. Remove the sealings used.
5. Dismount the adapter, if used, from the hand test pump.
6. Remove the test item hose
7. After the pressure has been relieved, close the pressure relief valve
clockwise.
Emptying the fluid reservoir
8. Open the fluid reservoir
9. Remove the pressure transmission medium from the pump using a syringe or a
special needle.
14209601.01 06/2017 EN/DE/FR/ES/IT
7
is open and there is no pressure in the test pump.
2
by turning it clockwise no more than once.
1
from the pump by turning it.
2
. To do so, unscrew the plug screws 3.
by turning it
21WIKA operating instructions, models CPP700-H and CPP1000-H
8. Dismounting, return and disposal
10. Rinse the fluid reservoir using water or gentle cleaning agents.
To do so, use the syringe specially designed for cleaning.
⇒
EN
11. Screw in the plug screws
8.2 Return
Strictly observe the following when shipping the instrument:
All instruments delivered to WIKA must be free from any kind of hazardous substances
(acids, bases, solutions, etc.) and must therefore be cleaned before being returned.
The use of alcohol, spirit, acetone or benzol can cause the container
material to crack or lead to damage of the container.
3
again.
If water is used as test or calibration medium, the test pump must be
emptied after EACH use.
WARNING!
Physical injuries and damage to property and the environment
through residual media
Residual media at the hand test pump can result in a risk to persons, the
environment and equipment.
▶
With hazardous substances, include the material safety data sheet for the
corresponding medium.
▶
Clean the instrument, see chapter 7.2 “Cleaning”.
When returning the instrument, use the original packaging or a suitable transport
packaging.
To avoid damage:
1. Place the instrument, along with the shock-absorbent material, in the packaging. Place
shock-absorbent material evenly on all sides of the transport packaging.
2. If possible, place a bag, containing a desiccant, inside the packaging.
3. Label the shipment as transport of a highly sensitive measuring instrument.
Information on returns can be found under the heading “Service” on our local
website.
22WIKA operating instructions, models CPP700-H and CPP1000-H
14209601.01 06/2017 EN/DE/FR/ES/IT
8. Dismounting, return and disposal / 9. Specifications
8.3 Disposal
Incorrect disposal can put the environment at risk.
Dispose of instrument components and packaging materials in an environmentally
compatible way and in accordance with the country-specific waste disposal regulations.
Dispose of the hydraulic fluids as described in the material safety data sheet.
The current material safety data sheet can be found at www.wika.de on the
product page of the respective product.
9. Specifications
SpecificationsModel CPP700-HModel CPP1000-H
Pressure range0 ... 700 bar
(0 … 10,000 psi)
Pressure transmission
medium
Fluid reservoir200 cm³
Pressure connections
For reference instrumentG ½ female thread, freely rotating.
For test itemG ¼ female at test connection
Fine adjustmentFine adjustment valve / volume adjustment
Materialanodised aluminium, brass, stainless steel, ABS
SealingsFKM and NBR (standard)
Dimensions (L x W x H)280 x 170 x 120 mm
Weight1.9 kg
1) Other pressure transmission media on request.
Hydraulic fluid based on mineral oil or single distilled water
(water-based hydraulic fluids are not suitable)
The connection can be disconnected using an Allen key. Then a
G ⅜ female thread is available.
hose, length 1 m (3.28 ft),
Minimess
(11.0 x 6.7 x 4.7 in)
(4.2 lbs)
®
1620 system
0 … 1,000 bar
(0 … 14,500 psi)
1)
G ¼ female, freely rotating at
test connection hose, length
1 m (3.28 ft)
EN
Approvals and certificates, see website
For further specifications see WIKA data sheet CT 91.07 and the order documentation.
14209601.01 06/2017 EN/DE/FR/ES/IT
23WIKA operating instructions, models CPP700-H and CPP1000-H
10. Accessories
10. Accessories
Adapter set
■
Adapter and sealing set, “BSP” for test item hose, G ¼ male to G ⅛, G ⅜ and G ½
EN
female
■
Adapter and sealing set, “metric” for test item hose, G ¼ male to M12 x 1.5, M20 x 1.5
and Minimess
■
Adapter and sealing set, “NPT” for test item hose, G ¼ male to ⅛ NPT, ¼ NPT, ⅜ NPT
®
and ½ NPT female
Case
■
Plastic case, incl. foam insert with free spaces for model CPP700-H or CPP1000-H;
Dimensions (W x H x D): 440 x 370 x 140 mm (17.3 x 14.6 x 5.5 in)
Pressure transmission medium
■
Special oil for CPP1000 and CPP1600 test pump series in plastic bottle, contents
1 litre
■
Hydraulic fluid based on “Shell Tellus S2 M22” mineral oil in plastic bottle, content
1 litre
Other
■
Connection adapter, G ¼ male to G ½ female, material: stainless steel
■
Sealing and maintenance set for model CPP700-H or CPP1000-H
■
Spare hose for model CPP700-H or CPP1000-H, length 1 m (3.28 ft)
24WIKA operating instructions, models CPP700-H and CPP1000-H
25WIKA Betriebsanleitung, Typen CPP700-H und CPP1000-H
1. Allgemeines
1. Allgemeines
■
Die in der Betriebsanleitung beschriebenen Handprüfpumpen CPP700-H und
CPP1000-H werden nach dem aktuellen Stand der Technik konstruiert und gefertigt.
Alle Komponenten unterliegen während der Fertigung strengen Qualitäts- und
Umweltkriterien. Unsere Managementsysteme sind nach ISO 9001 und ISO 14001
zertifiziert.
DE
■
Diese Betriebsanleitung gibt wichtige Hinweise zum Umgang mit dem Gerät.
Voraussetzung für sicheres Arbeiten ist die Einhaltung aller angegebenen
Sicherheitshinweise und Handlungsanweisungen.
■
Die für den Einsatzbereich des Gerätes geltenden örtlichen
Unfallverhütungsvorschriften und allgemeinen Sicherheitsbestimmungen einhalten.
■
Die Betriebsanleitung ist Produktbestandteil und muss in unmittelbarer Nähe
des Gerätes für das Fachpersonal jederzeit zugänglich aufbewahrt werden.
Betriebsanleitung an nachfolgende Benutzer oder Besitzer des Gerätes weitergeben.
■
Das Fachpersonal muss die Betriebsanleitung vor Beginn aller Arbeiten sorgfältig
durchgelesen und verstanden haben.
■
Es gelten die allgemeinen Geschäftsbedingungen in den Verkaufsunterlagen.
26WIKA Betriebsanleitung, Typen CPP700-H und CPP1000-H
14209601.01 06/2017 EN/DE/FR/ES/IT
2. Aufbau und Funktion
2. Aufbau und Funktion
2.1 Überblick
Ansicht von vorn
1
7
6
5
1
Prüflingsanschlussschlauch
2
CPP700-H: System Minimess
3
4
®
mit Manometer-Schottverschraubung auf
481
G ¼-Innengewinde
CPP1000-H: G ¼-Innengewinde
2
Druckablassventil
3
Verschlussschraube für Flüssigkeitsreservoir
4
Feinregulierventil (Volumenvariator)
5
Pumpenhandgriffe
6
Umschaltknopf Vor- und Hochdruckerzeugung
7
Flüssigkeitsreservoir
8
Anschluss G ½ innen, freilaufend für Referenzgeräte
9
Blindstopfen, nicht öffnen!
Ansicht von hinten
DE
9
7
6
5
14209601.01 06/2017 EN/DE/FR/ES/IT
27WIKA Betriebsanleitung, Typen CPP700-H und CPP1000-H
2. Aufbau und Funktion
2.2 Beschreibung
Die Handprüfpumpen CPP700-H und CPP1000-H dienen zur Druckerzeugung für
die Überprüfung, Justage und Kalibrierung von mechanischen und elektronischen
Druckmessgeräten durch Vergleichsmessungen. Diese Druckprüfungen können stationär
im Labor, Werkstatt oder vor Ort an der Messstelle stattfinden.
Schließt man das zu prüfende Druckmessgerät und ein hinreichend genaues ReferenzMessgerät an der CPP700-H oder CPP1000-H an, so wirkt bei Betätigung der Pumpe
DE
auf beide Messgeräte der gleiche Druck. Durch Vergleich der beiden Messwerte bei
beliebigen Druckwerten kann eine Überprüfung der Genauigkeit bzw. eine Justage des
zu prüfenden Druckmessgerätes erfolgen.
Die Handprüfpumpen ermöglichen trotz sehr kompakter Abmessungen eine einfache
und exakte Prüfdruckerzeugung. Für präzise Einstellung zur genauen Vergleichsprüfung
besitzen die CPP700-H und CPP1000-H ein Feinregulierventil. Das Referenzgerät wird
direkt oben auf die Pumpe aufgeschraubt und der Prüfling wird über den im Lieferumfang
enthaltenen Anschlussschlauch adaptiert.
2.3 Lieferumfang
CPP700-H
■
Handprüfpumpe Typ CPP700-H Öl/Wasser
■
Minimess® Prüflingsschlauch, Länge 1 m (3,28 ft)
■
Schottverschraubung G ¼ innen
■
Betriebsanleitung
CPP1000-H
■
Handprüfpumpe Typ CPP1000-H Öl/Wasser
■
Hochdruck-Prüflingsschlauch, Länge 1 m (3,28 ft)
■
Anschluss G ¼ innen
■
Betriebsanleitung
Lieferumfang mit dem Lieferschein abgleichen.
2.4 Druckübertragungsmedium
Als Druckübertragungsmedium ist für die CPP700-H und CPP1000-H ausschließlich
geeignet:
■
Hydraulikflüssigkeiten auf Mineralölbasis (Nicht geeignet sind wasserbasierende
Hydraulikflüssigkeiten)
■
Einfach destilliertes Wasser
Andere Druckübertragungsmedien auf Anfrage.
28WIKA Betriebsanleitung, Typen CPP700-H und CPP1000-H
14209601.01 06/2017 EN/DE/FR/ES/IT
2. Aufbau und Funktion / 3. Sicherheit
Umgang mit Mineralölen
WIKA liefert Hydraulikflüssigkeit auf Mineralölbasis in Behältern zu 1 Liter, beschriftet
mit „Shell Tellus S2 M22“ zur Verwendung in Handprüfpumpen mit bis zu 1.000 bar
(14.500 psi). Es ist nicht gefährlicher als andere Schmieröle.
Bevor mit Hydraulikflüssigkeiten gearbeitet wird, das Sicherheitsdatenblatt
durchlesen.
Das aktuelle Sicherheitsdatenblatt befindet sich unter www.wika.de auf der
Produktseite des jeweiligen Produkts.
3. Sicherheit
3.1 Symbolerklärung
WARNUNG!
... weist auf eine möglicherweise gefährliche Situation hin, die zum Tod oder
zu schweren Verletzungen führen kann, wenn sie nicht gemieden wird.
VORSICHT!
... weist auf eine möglicherweise gefährliche Situation hin, die zu
geringfügigen oder leichten Verletzungen bzw. Sach- und Umweltschäden
führen kann, wenn sie nicht gemieden wird.
DE
Information
... hebt nützliche Tipps und Empfehlungen sowie Informationen für einen
effizienten und störungsfreien Betrieb hervor.
3.2 Bestimmungsgemäße Verwendung
Prüfpumpen dienen zur Druckerzeugung für die Überprüfung, Justage und Kalibrierung
von mechanischen und elektronischen Druckmessgeräten durch Vergleichsmessungen.
Diese Druckprüfungen können stationär in Labor, Werkstatt oder vor Ort an der
Messstelle stattfinden. Die Handprüfpumpen Typ CPP700-H und CPP1000-H wurden
speziell für die Prüfdruckerzeugung vor Ort entwickelt.
Die Handprüfpumpen ermöglichen trotz sehr kompakter Abmessungen eine einfache und
exakte Prüfdruckerzeugung bis max. 700 bar (10.000 psi) bzw. 1.000 bar (14.500 psi) mit
integrierter Umschaltung von Vordruck- auf Hochdruckerzeugung.
14209601.01 06/2017 EN/DE/FR/ES/IT
29WIKA Betriebsanleitung, Typen CPP700-H und CPP1000-H
3. Sicherheit
Dieses Gerät ist nicht für den Einsatz in explosionsgefährdeten Bereichen zugelassen!
Das Gerät ist ausschließlich für den hier beschriebenen bestimmungsgemäßen
Verwendungszweck konzipiert und konstruiert und darf nur dementsprechend verwendet
werden.
Die technischen Spezifikationen in dieser Betriebsanleitung sind einzuhalten. Eine
unsachgemäße Handhabung oder ein Betreiben des Gerätes außerhalb der technischen
DE
Spezifikationen macht die sofortige Stilllegung und Überprüfung durch einen autorisierten
WIKA-Servicemitarbeiter erforderlich.
Prüfpumpe mit erforderlicher Sorgfalt behandeln (vor Nässe, Stößen, starken
Magnetfeldern, statischer Elektrizität und extremen Temperaturen schützen, keine
Gegenstände in das Gerät bzw. Öffnungen einführen). Stecker und Buchsen vor
Verschmutzung schützen.
Ansprüche jeglicher Art aufgrund von nicht bestimmungsgemäßer Verwendung sind
ausgeschlossen.
3.3 Fehlgebrauch
WARNUNG!
Verletzungen durch Fehlgebrauch
Fehlgebrauch des Gerätes kann zu gefährlichen Situationen und
Verletzungen führen.
▶
Eigenmächtige Umbauten am Gerät unterlassen.
▶
Gerät nicht in explosionsgefährdeten Bereichen einsetzen.
▶
Gerät nicht für abrasive und viskose Messstoffe verwenden.
Jede über die bestimmungsgemäße Verwendung hinausgehende oder andersartige
Benutzung gilt als Fehlgebrauch.
3.4 Personalqualifikation
WARNUNG!
Verletzungsgefahr bei unzureichender Qualifikation
Unsachgemäßer Umgang kann zu erheblichen Personen- und
Sachschäden führen.
▶
Die in dieser Betriebsanleitung beschriebenen Tätigkeiten nur durch
Fachpersonal nachfolgend beschriebener Qualifikation durchführen
lassen.
30WIKA Betriebsanleitung, Typen CPP700-H und CPP1000-H
14209601.01 06/2017 EN/DE/FR/ES/IT
Loading...
+ 86 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.