WIKA CPP4000 Specifications

Hydraulic comparison test pump Model CPP4000-X
Applications
Simple hydraulic test pressure generation on site, in the laboratory or in the workshop
For the testing, adjustment and calibration of all types of pressure measuring instruments
Hydraulic pressure generation up to 4,000 bar
Special features
Calibration technology
WIKA data sheet CT 91.09
Precisely adjustable dual-area spindle pump for lling, pressure generation and ne adjustment of pressure
Proven technology of the model CPB3800HP dead-weight
tester
Compact dimensions
Low weight
Description
Application areas
Test pumps serve as pressure generators for the testing, adjustment and calibration of mechanical and electronic pressure measuring instruments through comparative measurements. These units can be used in the laboratory or workshop, or on site at the measuring point.
Easy operation
By connecting the test item and a suciently accurate
reference pressure measuring instrument to the test pump, the same pressure will act on both measuring instruments. By comparison of the two measured values at any given pressure
value, a check of the accuracy and/or adjustment of the
pressure measuring instrument under test can be carried out.
Easy handling
The model CPP4000-X comparison test pump is a hydraulic comparison test pump for generating pressure up to 4,000 bar. It is technically identical to the instrument base of
the model CPB3800HP dead-weight tester. The integrated
dual-area spindle pump with switching valves enables rapid
lling of the test system and smooth pressure generation.
At the same time, the precisely adjustable spindle pump also
serves for ne pressure adjustment. A control schematic for
pressure generation on the instrument base facilitates quick and easy operation.
The two test connections are equipped with G ½ female loose union connections with double ended metal cone. Two
threaded adapters mounted on M16 x 1.5 male thread are
included in the scope of delivery. Further thread adapters
are available for the calibration of instruments with dierent
connection threads.
Compact instrument design
The CPP4000-X is also particularly notable for its compact dimensions, which are not altered during operation, since the spindle runs within the pump body. With those dimensions, the exceptionally robust ABS plastic case and the low weight associated with these, the CPP4000-X is also well suited for on-site applications.
Data sheets showing similar products: Pneumatic hand test pump; model CPP7; see data sheet CT 91.04 Pneumatic hand test pump; model CPP30; see data sheet CT 91.06 Hydraulic hand test pump; models CPP700-H and CPP1000-H; see data sheet CT 91.07
Hydraulic test pumps; models CPP1000-X to CPP7000-X; see data sheet CT 91.05 Hydraulic comparison test pump; model CPP1200-X; see data sheet CT 91.08
Page 1 of 4WIKA data sheet CT 91.09 ∙ 02/2015
Specications Model CPP4000-X
Pressure range 0 ... 4,000 bar
Pressure transmission medium Hydraulic uid based on VG22 mineral oil or Sebacate oil (optional)
Reservoir 170 cm³
Pressure connections
Test connection 2 x G ½ B female thread union nut with metal cone;
equipped with two thread adapters M16 x 1.5 as standard
Distance between test connections 290 mm
Piston of spindle pump
Piston diameter, low-pressure piston 31.2 mm
Piston diameter, high-pressure piston 6.35 mm
Displacement approx. 5 cm³
Material
Wetted parts Austenitic stainless steel, high-tensile brass, nitrile rubber
Permissible ambient conditions
Operating temperature 0 ... 40 °C
Instrument base
Dimensions 401 x 397 x 155 mm (W x D x H), for details, see technical drawings
Weight 13.5 kg
1) Other pressure transmission media on request.
Storage case (option)
Material Wood
Dimensions 664 x 614 x 380 mm (W x D x H)
Weight 8.5 kg
CE conformity
Pressure equipment directive 97/23/EC (PS > 1,000 bar; module A, pressure accessory)
Approvals and certicates, see website
Recommended reference pressure measuring instruments:
Precision hand-held pressure indicator model CPH6400
Measuring ranges: up to 6,000 bar Accuracy: up to 0.025 % of span
For further specications see data sheet CT 14.01
ProcessCalibrator model CPH6000
Measuring ranges: up to 8,000 bar Accuracy: up to 0.025 % of span
For further specications see data sheet CT 15.01
Page 2 of 4 WIKA data sheet CT 91.09 ∙ 02/2015