WIKA CPH7650 Operating Manual

Operating instructions Betriebsanleitung
Portable pressure calibrator, model CPH7650
Tragbarer Druckkalibrator, Typ CPH7650
EN
DE
Portable pressure calibrator, model CPH7650
EN
Operating instructions model CPH7650 Page 3 - 58
DE
Betriebsanleitung Typ CPH7650 Seite 59 - 113
Further languages can be found at www.wika.com.
© 03/2018 WIKA Alexander Wiegand SE & Co. KG All rights reserved. / Alle Rechte vorbehalten.
®
WIKA
is a registered trademark in various countries.
®
WIKA
ist eine geschützte Marke in verschiedenen Ländern.
Prior to starting any work, read the operating instructions! Keep for later use!
Vor Beginn aller Arbeiten Betriebsanleitung lesen! Zum späteren Gebrauch aufbewahren!
2
WIKA operating instructions, model CPH7650
Contents
Contents
1. General information 5
2. Safety 6
2.1 Intended use . . . . . . . . . . . . . . . . . . . . . . 7
2.2 Personnel qualification . . . . . . . . . . . . . . . . . . . 8
2.3 Special hazards . . . . . . . . . . . . . . . . . . . . . 8
2.4 Use of Lithium-Ion rechargeable batteries . . . . . . . . . . . .10
2.5 Labelling, safety marks . . . . . . . . . . . . . . . . . . .12
3. Specifications 13
4. Design and function 15
4.1 Description . . . . . . . . . . . . . . . . . . . . . . . 15
4.2 Front . . . . . . . . . . . . . . . . . . . . . . . . .16
4.3 Scope of delivery . . . . . . . . . . . . . . . . . . . . .17
4.4 Voltage supply . . . . . . . . . . . . . . . . . . . . . .17
4.5 User interface . . . . . . . . . . . . . . . . . . . . . . 18
4.5.1 Requirements for test assemblies with the CPH7650 . . . . . . . 19
4.5.2 Important instrument settings for calibration using calibration mode . . 19
4.5.3 Zero point setting and offset correction . . . . . . . . . . . .20
4.5.4 Reference pressure sensor CPT6000 . . . . . . . . . . . . 20
5. Transport, packaging and storage 22
5.1 Transport. . . . . . . . . . . . . . . . . . . . . . . . 22
5.2 Packaging . . . . . . . . . . . . . . . . . . . . . . . 22
5.3 Storage . . . . . . . . . . . . . . . . . . . . . . . . 22
6. Commissioning, operation 23
6.1 Menu structure (operating modes) . . . . . . . . . . . . . . . 24
6.2 Explanation of the display . . . . . . . . . . . . . . . . . . 25
6.2.1 Instrument status messages shortly after switching on the CPH7650 . .25
6.2.2 Switching the pressure calibrator off . . . . . . . . . . . . .26
6.2.3 Display contents of the operating modes . . . . . . . . . . . 26
6.2.4 Content of the SETUP menu . . . . . . . . . . . . . . .29
6.3 Operating modes . . . . . . . . . . . . . . . . . . . . .32
6.3.1 MEASURING mode . . . . . . . . . . . . . . . . . . 32
6.3.2 MEASURING mode (with test item) . . . . . . . . . . . . . 34
6.3.3 Mode CALIBRATION . . . . . . . . . . . . . . . . . . 36
6.3.4 CALIBRATION mode (preparing the test points of a calibration). . . . 38
6.3.5 CALIBRATION mode (calibration of a pressure transmitter) . . . . . 41
6.3.6 CALIBRATION mode (calibration of a pressure gauge) . . . . . . 42
6.3.7 SWITCH TEST mode . . . . . . . . . . . . . . . . . . 43
6.4 SETUP additional menu items . . . . . . . . . . . . . . . . 45
EN
3WIKA operating instructions, model CPH7650
Contents
6.4.1 SETUP additional menu items: Functions . . . . . . . . . . . 45
6.4.2 SETUP additional menu items: CPH info . . . . . . . . . . . 46
6.4.3 SETUP additional menu items: Reference sensor . . . . . . . . 47
EN
6.4.4 SETUP additional menu items: Reference sensor list . . . . . . . 48
6.4.5 SETUP additional menu items: CPH configuration . . . . . . . .49
6.4.6 SETUP additional menu items: Interface . . . . . . . . . . . 50
6.4.7 SETUP additional menu items: CLEAR CalProg . . . . . . . . .50
6.5 Connection of the model CPT6000 reference pressure sensor . . . . . 51
6.6 Voltage supply . . . . . . . . . . . . . . . . . . . . . .51
6.7 Charging/discharging the Lithium-Ion rechargeable batteries . . . . . .52
6.8 Pressure measurement . . . . . . . . . . . . . . . . . . .53
6.9 Measuring and sourcing current (4 ... 20 mA) . . . . . . . . . . . 53
7. Maintenance, cleaning and recalibration 54
7.1 Maintenance . . . . . . . . . . . . . . . . . . . . . . 54
7.2 Cleaning . . . . . . . . . . . . . . . . . . . . . . . .54
7.3 Recalibration . . . . . . . . . . . . . . . . . . . . . .54
8. Faults 55
9. Dismounting, return and disposal 56
9.1 Dismounting . . . . . . . . . . . . . . . . . . . . . . 56
9.2 Return. . . . . . . . . . . . . . . . . . . . . . . . . 56
9.3 Disposal . . . . . . . . . . . . . . . . . . . . . . . .56
10. Accessories 57
Declarations of conformity can be found online at www.wika.com.
4 WIKA operating instructions, model CPH7650
1. General information
1. General information
The portable pressure calibrator described in the operating instructions has been manufactured using state-of-the-art technology. All components are subject to stringent quality and environmental criteria during production. Our management systems are certified to ISO 9001 and ISO 14001.
These operating instructions contain important information on handling the instrument. Working safely requires that all safety instructions and work instructions are observed.
Observe the relevant local accident prevention regulations and general safety regulations for the instrument's range of use.
The operating instructions are part of the product and must be kept in the immediate vicinity of the instrument and readily accessible to skilled personnel at any time.
Skilled personnel must have carefully read and understood the operating instructions prior to beginning any work.
The manufacturer's liability is void in the case of any damage caused by using the product contrary to its intended use, non-compliance with these operating instructions, assignment of insufficiently qualified skilled personnel or unauthorised modifications to the instrument.
EN
The general terms and conditions contained in the sales documentation shall apply.
Subject to technical modifications.
Factory calibrations / DKD/DAkkS calibrations are carried out in accordance with international standards.
Further information:
- Internet address: www.wika.de / www.wika.com
- Relevant data sheet: CT 17.02
- Application consultant:
Tel.: +49 9372 132-9986 Fax: +49 9372 132-8767 info@wika.com
5WIKA operating instructions, model CPH7650
1. General information / 2. Safety
Explanation of symbols
WARNING!
EN
... indicates a potentially dangerous situation that can result in serious injury or death, if not avoided.
CAUTION!
... indicates a potentially dangerous situation that can result in light injuries or damage to property or the environment, if not avoided.
Information
... points out useful tips, recommendations and information for efficient and trouble-free operation.
DANGER!
... identifies hazards caused by electrical power. Should the safety instructions not be observed, there is a risk of serious or fatal injury.
2. Safety
WARNING!
Before installation, commissioning and operation, ensure that the appropriate reference pressure sensor has been selected in terms of measuring range, design and specific measuring conditions. Non-observance can result in serious injury and/or damage to the equipment.
Further important safety instructions can be found in the individual chapters of these operating instructions.
6 WIKA operating instructions, model CPH7650
2. Safety
2.1 Intended use
This portable pressure calibrator serves as a calibration instrument for the widest variety of pressure measuring instruments and has been designed for mobile use as well as for stationary workshop and laboratory testing. Through the combination of the integrated electrical pump and the electrical modules, in addition to the traditional measurement of current and voltage signals it also enables you to supply transmitters or sensors with a max. of 30 mA (voltage (idling) = DC 24 V). The pressure calibrator can be used to carry out and document a complete calibration process.
WARNING!
Only use model CPT6000 reference pressure sensors!
Using other sensors could damage both the pressure calibrator and the reference pressure sensor.
To change the sensor, switch off the pressure calibrator and make sure the system is vented. Before switching the instrument on, connect the sensor, otherwise it may not be correctly identified by the instrument.
When the CPH7650 is switched on, the CPT6000 reference pressure sensor must not be under pressure, but rather should be at atmospheric pressure. For overpressure or gauge pressure sensors, there is a pressure compensation vent in the top of the sensor under the plastic fitting. This vent (with integrated diaphragm) must always remain clear!
The instrument has been designed and built solely for the intended use described here, and may only be used accordingly.
EN
The technical specifications contained in these operating instructions must be observed. Improper handling or operation of the instrument outside of its technical specifications requires the instrument to be taken out of service immediately and inspected by an authorised WIKA service engineer.
Handle electronic precision measuring instruments with the required care (protect from humidity, impacts, strong magnetic fields, static electricity and extreme temperatures, do not insert any objects into the instrument or its openings). Plugs and sockets must be protected from contamination.
If the instrument is transported from a cold into a warm environment, the formation of condensation may result in instrument malfunction. Before putting it back into operation, wait for the instrument temperature and the room temperature to equalise.
The manufacturer shall not be liable for claims of any type based on operation contrary to the intended use.
7WIKA operating instructions, model CPH7650
2. Safety
2.2 Personnel qualification
WARNING!
EN
Skilled personnel
Skilled personnel are understood to be personnel who, based on their technical training, knowledge of measurement and control technology and on their experience and knowledge of country-specific regulations, current standards and directives, are capable of carrying out the work described and independently recognising potential hazards.
Special operating conditions require further appropriate knowledge, e.g. of aggressive media.
2.3 Special hazards
Risk of injury should qualification be insufficient!
Improper handling can result in considerable injury and damage to equipment. The activities described in these operating instructions may only be carried out by skilled personnel who have the qualifications described below.
WARNING!
Do not apply a voltage greater than the specified voltage to the instrument (see chapter 3 „Specifications“).
Do not apply any external pressure to the CPH7650.
Make sure that the test probes never contact a voltage source while the test cables are connected to the current terminals.
Do not use the calibrator if it is damaged. Before using the instrument, check that there are no cracks on the case or any missing plastic parts. Pay particular attention to the insulation of the connectors.
Select the proper function and correct measuring range for the measurement.
Observe the operating parameters in accordance with chapter 3 “Specifications”.
Always operate the pressure calibrator within the defined pressure range.
To ensure problem-free operation, only operate the instrument on battery power. Only use the power supply unit for charging the instrument's batteries.
Inspect the test cables for damaged insulation or exposed metal. Check the continuity of the leads. Damaged test leads should be replaced before using the instrument.
When using test probes, keep fingers away from the test probe contacts. Keep your fingers behind the test probes' finger guards.
First connect the common lead, and then the live lead. When disconnecting, remove the live test lead first.
8 WIKA operating instructions, model CPH7650
2. Safety
Do not use the instrument if it is not working properly. The instrument protection might be compromised. If in doubt, have the instrument checked.
Do not operate the instrument in areas with explosive gases, vapours or dust.
When measuring pressure, make sure that the pressure calibrator has been switched to a depressurised state before the reference sensor or the test item is connected or disconnected.
Disconnect test leads before changing to another measurement or source function.
The switching valve (+/- pressure) should only ever be actuated when in a depressurised state.
WARNING!
To avoid false indications, which could lead to possible electric shock or injuries, charge the rechargeable battery as soon as the battery indicator appears.
In order to avoid any possible damage to the instrument or the test equipment, use the correct leads, the correct function and the correct range for the measuring application.
DANGER! Danger to life caused by electric current
Upon contact with live parts, there is a direct danger to life.
Charging using a defective power supply unit (e.g. short circuit from the mains voltage to the output voltage) can result in life-threatening voltages at the instrument!
Only use the power supply unit permitted by WIKA for the model CPH7650 portable pressure calibrator.
Only use a charger that is fully functional or undamaged.
EN
The safety of the operator may be endangered if, for example
there is visible damage to the instrument.
the instrument is no longer working as specified.
the instrument has been stored under unsuitable conditions for an extended period of
time. If there is any doubt, please return the instrument to the manufacturer for repair or maintenance.
9WIKA operating instructions, model CPH7650
2. Safety
2.4 Use of Lithium-Ion rechargeable batteries
WARNING!
EN
Misusing Lithium-Ion batteries can lead to heating, explosion or ignition and result in serious injury. Follow the safety instructions listed below:
Do not solder directly to the Lithium-Ion batteries.
Do not incinerate or heat the Lithium-Ion batteries.
The Lithium-Ion batteries must only ever be connected with the correct polarity.
Never connect the positive terminal and the negative terminal of the Lithium-Ion batteries to each other with any metallic object (such as wire).
Never carry or store the Lithium-Ion batteries together with necklaces, hairpins, or other metal objects.
WARNING!
Lithium-Ion batteries should never be punctured with nails nor hit with a hammer. In addition, Lithium-Ion batteries must never be trodden on or exposed to other strong shocks or vibrations.
Lithium-Ion batteries must never come into contact with water or salt water. Moreover, they must never get wet.
WARNING!
Never take the Lithium-Ion battery apart nor alter it in any way. It contains safety and protection devices which, if damaged, may cause it to generate heat, explode or ignite.
WARNING!
Never place the Lithium-Ion batteries close to fires, ovens or other high-temperature locations. Never leave the Lithium-Ion batteries in direct sunshine or use or store them inside cars in hot weather. Doing so may cause the Lithium-Ion batteries to generate heat, explode or ignite. Using the Lithium-Ion batteries in this manner may also result in a loss of performance and a shortened service life.
Never fit the Lithium-Ion batteries into equipment designed to be hermetically sealed. In some cases hydrogen or oxygen may be discharged from the Lithium-Ion batteries, which may result in rupture, fire or explosion.
10 WIKA operating instructions, model CPH7650
2. Safety
WARNING!
The Lithium-Ion batteries must, without fail, no longer be used if, during operation, charging or storing they give off an unusual smell, feel hot, change colour, change shape, or appear abnormal in any other way. Contact your sales partner if any of these problems are observed.
Never put the Lithium-Ion batteries in microwave ovens, high-pressure containers nor on induction cookers.
Should the Lithium-Ion batteries ever leak and the fluid come into contact with the eyes, do not under any circumstances rub the eyes. Rinse the eyes thoroughly with water and seek immediate medical attention. If the eyes are left untreated, damage to the eyes could occur.
CAUTION!
When the Lithium-Ion batteries wear out, insulate the terminals with adhesive tape or similar materials before disposal.
WARNING!
Follow the instructions listed below for charging the Lithium-Ion batteries. Failure to do so may cause the Lithium-Ion batteries to become hot, explode or ignite and result in serious injury.
To charge the Lithium-Ion batteries, only ever use the specified WIKA battery charger.
Never connect the Lithium-Ion batteries directly to a mains plug or to a car's cigarette lighter.
Never leave the Lithium-Ion batteries in or near fire, nor in direct sunlight. If the Lithium-Ion batteries become hot, the built-in safety device is activated and overcharging prevented. Heating the Lithium-Ion batteries can damage the safety device and can thus lead them to heat up further, to cease to work or to ignite.
EN
WARNING!
Never continue to charge the Lithium-Ion batteries if they do not fully recharge within the specified time. Doing so may cause the Lithium-Ion batteries to become hot, explode or ignite.
11WIKA operating instructions, model CPH7650
2. Safety
2.5 Labelling, safety marks
Product label
EN
6 5 4 3 2
1
Product name
2
Date of manufacture (month/year)
3
Serial number
4
Output signal, power supply
5
Input signal
6
Pressure range
1
Symbols
Before mounting and commissioning the instrument, ensure you read the operating instructions!
This marking on the instruments indicates that they must not be disposed of in domestic waste. The disposal is carried out by return to the manufacturer or by the corresponding municipal authorities.
12 WIKA operating instructions, model CPH7650
3. Specifications
3. Specifications
Measuring ranges
Pressure
Gauge pressure -1 … +20 bar, -1 … +10 bar, 0 … 20 bar, 0 … 10 bar Absolute pressure 0 … 10 bar abs., 0 … 20 bar abs. Overpressure limit Accuracy 0.025 % FS Resolution 5-digit Temperature compensation 15 … 35 °C (59 ... 95 °F) Temperature coefficient 0.002 % of span/°C outside of 15 ... 35 °C (59 ... 95 °F)
Current
Measuring range 0 … 24 mA (max. load 1,000 Ω) Resolution 1 µA Accuracy 0.015 % of reading ±2 µA (simulation and measurement)
Voltage
Measuring range DC 0 … 30 V Resolution 1 mV Accuracy 0.015 % of reading ±2 mV (measurement)
1) Via exchangeable model CPT6000 reference pressure sensors
2) The electrical pump can generate -0.85 … +20 bar (-12 … 290 psi).
1)
2)
3 times
EN
Base instrument
Pressure supply -0.85 ... +20 bar (-12 ... +290 psi), via integrated electric pump
Pressure connection for test item
Permissible media For dry, clean and non-aggressive gases
Output
Voltage supply DC 24 V
Power supply
Battery type Lithium-Ion rechargeable battery
Permissible ambient conditions
Operating temperature -10 … +50 °C (14 ... 122 °F) Storage temperature -20 … +60 °C (-4 ... +140 °F) Relative humidity 35 ... 85 % r. h. (non-condensing)
Communication
Interface USB via special interface cable
G ½" female thread
13WIKA operating instructions, model CPH7650
3. Specifications
Base instrument
Case
EN
Case NK-7TM resin Front panel Aluminium Ingress protection IP67 (case closed)
IP40 (case opened) Dimensions 387.4 x 304.8 x 177.8 mm (15.25 x 12 x 7 in) Weight approx. 7 kg (15.5 lbs.)
Certificates
Certificate
Calibration Standard: 3.1 calibration certificate per DIN EN 10204
Option: DKD/DAkkS calibration certificate
Recommended recalibration
1 year (dependent on conditions of use)
interval
Approvals and certificates, see website
Available pressure range and resolution
Pressure range and factors
Gauge pressure -1 ... +20 bar (-14.5 ... +290 psi)
Overpressure limit 40 bar (580 psi) Unit Conversion factor psi 1 300.00 bar 0.06894757 20.684 mbar 68.94757 20,684 kPa 6.894757 2,068.4 MPa 0.00689476 2.0684 kg/cm² 0.07030697 21.092
O (4 °C) 70.3089 21,093
cmH
2
O (20 °C) 70.4336 21,130
cmH
2
O (4 °C) 27.68067 8,304.2
inH
2
O (20 °C) 27.72977 8,318.9
inH
2
O (60 °F) 27.70759 8,312.3
inH
2
mmHg (0 °C) 51.71508 15,515 inHg (0 °C) 2.03602 610.81
3) Electrical pump: -0.85 … +20 bar (-12 … 290 psi)
14 WIKA operating instructions, model CPH7650
3. Specifications / 4. Design and function
104 (4,11)
158 (6,22)
Dimensions in mm (in)
172 (6,76)
390 (15,36)
EN
308 (12,13)
4. Design and function
4.1 Description
The CPH7650 is a compact, portable pressure calibrator, which has been designed for mobile use as well as for stationary workshop and laboratory testing. An integrated electrical pump enables you to generate pressures up to 20 bar (300 psi) and an integrated electrical module also enables you to supply transmitters or sensors with a current of max. 30 mA (voltage (idling) = DC 24 V), in addition to the traditional measurement of current and voltage signals.
The pressure calibrator can be used to carry out and document a complete calibration process. In addition to the features of displaying or measuring and calibration, there is also a switch test. The use of the WIKA-Cal software also allows a complete documentation.
15WIKA operating instructions, model CPH7650
4. Design and function
4.2 Front
EN
1
Test item
2
Overview of the electrical connection
1
2
3 4
5
6
7
8
9
10111213
3
Electrical connection to the model CPT6000 reference sensor
4
Reference sensor CPT6000
5
Connecion of the power supply unit
6
WIKA-Cal connection
7
Fine adjustment / drain valve
8
Switch for pressure / vacuum
9
Pump speed controller
10
Operation
11
Electrical module
12
Display
13
Numeric keypad
16 WIKA operating instructions, model CPH7650
4. Design and function
4.3 Scope of delivery
Portable pressure calibrator model CPH7650
Operating instructions
Test cables
Battery charger
3.1 calibration certificate per DIN EN 10204
Cross-check scope of delivery with delivery note.
4.4 Voltage supply
Charging the battery
In order to avoid false measurements, charge the battery as soon as the battery indicator is displayed. If the battery has run down too far, the instrument will switch itself off automatically.
Only ever use the power supply unit permitted by WIKA for the model CPH7650 portable pressure calibrator.
To avoid measurement uncertainties, only use the CPH7650 without the mains supply connected. The full pump performance can only be guaranteed with the battery fully charged and without the power supply unit connected.
The instrument should be fully charged before being put into operation. The battery capacity status (charge state in %) is displayed shortly after the instrument is switched on. When the power supply unit is connected, the rechargeable battery will be charged, even if the CPH7650 is switched off.
EN
The typical charging time of the rechargeable battery is < 5 hours.
WARNING!
When the power supply unit is no longer being used, the mains plug should be disconnected from the mains socket. Do not leave the rechargeable battery connected to the power supply unit for longer than one day, since overcharging can shorten its service life.
Should the rechargeable battery still not be fully charged after 24 hours, contact the manufacturer. When not being used, a fully charged battery will lose its charge over time.
Extreme temperatures have an adverse effect on battery charging. As a result, the battery may first need to be either cooled or warmed, as appropriate.
When the battery is almost completely discharged, the message “low BAT” appears in the display. To avoid a data loss, the instrument must be charged immediately
17WIKA operating instructions, model CPH7650
4. Design and function
4.5 User interface
EN
6
5
1
SETUP menu
2
Selection and entry confirmation
3
Return to previous level
4
Clear entry
5
Input confirmation
6
Numeric keypad
Switch on via pressing any button Switch off via menu item in main menu
1
2
3
4
18 WIKA operating instructions, model CPH7650
4. Design and function
4.5.1 Requirements for test assemblies with the CPH7650
Before starting any task, the instrument should be switched on briefly to determine that there is sufficient charge in the battery (rechargeable battery capacity in %). The battery capacity is indicated briefly in an instrument status message after switching on (see chapter 6.2.1 “Instrument status messages shortly after switching on the
CPH7650”).
Initially, the test assembly must be physically assembled and, if necessary, connected electrically (see chapter 4.5.4.1 “Connection of the model CPT6000 reference pressure sensor”).
Before switching the CPH7650 on, ensure that the test assembly is not pressurised (system is vented to atmosphere) and that the equipment is correctly assembled and in the correct mounting position.
Only connect test and calibration installations once the system has been depressurised!
Compensation for height differences
If a significant elevation difference exists between the CPT6000 reference pressure sensor and the test item, then the pressure difference, based on a medium column, can be compensated automatically via the menu (see chapter 6.4.5 “SETUP additional menu items: CPH configuration”).
4.5.2 Important instrument settings for calibration using calibration mode
EN
Calibration date
The instrument has an integral real-time clock with date. The current date of a calibration is stated later in the calibration certificate. Before starting a calibration, you must ensure that the internal date of the CPH7650 is correct (see chapter 6.4.5 “SETUP additional menu items: CPH configuration”).
Unit and resolution After selecting one of the main menu items (e.g.: MEASURING, CALIBRATION or SWITCH TEST) from the SETUP menu (press SETUP key), using the menu item “Unit”,
and its associated submenu respectively (move the cursor to “Unit” and press the right or left arrow), you can set the unit and adjust its resolution (see chapter 6.3 “Operating modes”).
19WIKA operating instructions, model CPH7650
4. Design and function
Available units, including their conversion factors in relation to the unit bar
bar 1.00000E+00 mbar 1.00000E-03 hPa 1.00000E-03
EN
psi 6.89475E-02 inHg (0 °C) 3.37690E-02 cmHG (0 °C) 1.33322E-02 MPa 1.00000E+01 kPa 1.00000E-02 Pa 1.00000E-05
O (4 °C) 9.80670E-02
mH
2
O (4 °C) 9.80670E-04
cmH
2
O (4 °C) 9.80670E-05
mmH
2
2
kg/cm
O (60 °C) 2.48800E-03
inH
2
O (0 °C) 1.33322E-03
mmH
2
4.5.3 Zero point setting and offset correction
Zero point setting for overpressure sensors
If the measured value shown on the CPH7650, with an overpressure sensor connected and the test assembly vented to atmosphere, is not equal to zero, then by pressing the CLEAR button twice (within five seconds), the zero point can be corrected (maximum allowable correction value is twice the magnitude of the class accuracy).
9.80665E-01
Offset correction for absolute pressure sensors
For absolute pressure sensors, an offset correction can be made via the menu (see chapter 6.4.3 “SETUP additional menu items: Reference sensor”).
4.5.4 Reference pressure sensor CPT6000
For the model CPH7650 pressure calibrator, there are many reference pressure sensors to choose from, with accuracies of 0.025 % which can be interchanged quickly and without tools. When the pressure calibrator is switched on, the reference pressure sensor attached is recognised automatically, so that no further configuration of the sensor is needed.
20 WIKA operating instructions, model CPH7650
4. Design and function
4.5.4.1 Connection of the model CPT6000 reference pressure sensor
WARNING!
Only use model CPT6000 reference pressure sensors!
Using other sensors could damage both the pressure calibrator and the reference pressure sensor.
To change the sensor, switch off the pressure calibrator and make sure the system is vented. Before switching the instrument on, connect the sensor, otherwise it may not be correctly identified by the instrument.
When the CPH7650 is switched on, the CPT6000 reference pressure sensor must not be under pressure, but rather should be at atmospheric pressure. For overpressure or gauge pressure sensors, there is a pressure compensation vent in the top of the sensor under the plastic fitting. This vent (with integrated diaphragm) must always remain clear!
CAUTION!
Only ever use the original WIKA sensor connection cable in the operation of CPT6000 reference pressure sensors.
WARNING!
The pump performance is independent of the pressure range of the selected reference sensor. The operator must ensure that the CPT6000 reference sensor is not over-pressured.
EN
4.5.4.2 Electrical connection of the CPT6000 reference pressure sensor to the CPH7650
The calibrator and the reference pressure sensor are connected to each other electrically using a separate connection cable. To electrically connect a model CPT6000 reference pressure sensor, the corresponding cable connector must be plugged in at the sensor in accordance with the orientation guide. To disconnect the sensor, do not pull on the cable, but rather only on the connector sleeve. To connect it to the CPH7650, the other end of the cable must also be plugged in in accordance with the orientation guide.
4.5.4.3 Mechanical connection of the CPT6000 reference pressure sensor to the CPH7650
To make the mechanical connection of the CPT6000 reference pressure sensor, it must be placed, connection thread first, in the sensor bracket of the instrument. Then, the sensor must be tightened again. (Tighten = turn clockwise; release = turn anti-clockwise) No further tools are required for this.
21WIKA operating instructions, model CPH7650
5. Transport, packaging, storage
5. Transport, packaging and storage
5.1 Transport
Check the portable pressure calibrator for any damage that may have been caused by
EN
transport. Obvious damage must be reported immediately.
5.2 Packaging
Do not remove packaging until just before mounting. Keep the packaging as it will provide optimum protection during transport (e.g. change in installation site, sending for repair).
5.3 Storage
Permissible conditions at the place of storage:
Storage temperature: -20 ... +60 °C
Relative humidity: 0 ... 85 % r. h. (non-condensing)
Avoid exposure to the following factors:
Direct sunlight or proximity to hot objects
Mechanical vibration, mechanical shock (putting it down hard)
Soot, vapour, dust and corrosive gases
Hazardous environments, flammable atmospheres
Store the portable pressure calibrator in its original packaging in a location that fulfils the conditions listed above. If the original packaging is not available, pack and store the instrument as described below:
1. Wrap the instrument in an antistatic plastic film.
2. Place the instrument, along with shock-absorbent material, in the packaging.
3. If stored for a prolonged period of time (more than 30 days), place a bag containing a desiccant inside the packaging.
WARNING!
Before storing the instrument (following operation), remove any residual media. This is of particular importance if the medium is hazardous to health, e.g. caustic, toxic, carcinogenic, radioactive, etc.
22 WIKA operating instructions, model CPH7650
6. Commissioning, operation
6. Commissioning, operation
By pressing any key, the model CPH7650 portable pressure calibrator will be switched on. The calibrator requires a warm-up for a few minutes (max. 5 minutes) to reach its specified accuracy. Large changes in ambient temperature may make a longer warm-up period necessary. The pressure display of the calibrator should be zeroed before starting the pressure calibration.
Instrument features The instrument features 3 operating modes: MEASURING / CALIBRATION / SWITCH TEST, which offer the user maximum convenience according to its application. For
supplying the test items and to read their measuring signals, there are electrical inputs and outputs available. In the MEASURING (with test item) and CALIBRATION operating modes, the measured values of both the reference pressure sensor and the test item, as well as their deviation, are displayed in both current pressure units and in %. In this way the operator is immediately informed online whether the test item meets the class accuracy or not. The difference between these two modes is that the calibration data in CALIBRATION mode are stored internally and can later be transferred onto printable certificates through software (WIKA-Cal). With respect to the transfer of data to a PC, the model CPH7650 pressure calibrator has a USB interface, selectable via menu.
SETUP menu
Using the SETUP key, one can access the SETUP menu, where the required operating mode (MEASURING / CALIBRATION / SWITCH TEST) can be selected and configured, a stored function can be recalled or a general instrument setting (such as the menu language) can be changed.
EN
23WIKA operating instructions, model CPH7650
6. Commissioning, operation
6.1 Menu structure (operating modes)
Through the SETUP menu, the required operating mode can be easily selected (see drawing below).
EN
Functions CPH-Info ReferenceSensor Ref. Sensor-list CPH-Configuration Interface CLEAR CalData
CPH switch off
SELECTSELECT SELECT
SELECT SELECTSELECT
optional
Operating modes Configuration SETUP menu
It is possible to change the test item display (pressure ↔ electrical signal) via
24 WIKA operating instructions, model CPH7650
6. Commissioning, operation
6.2 Explanation of the display
6.2.1 Instrument status messages shortly after switching on the CPH7650
Directly after the instrument is switched on, the following status messages are displayed briefly:
CPH7650
a b c d e
f
a) The voltage supply of 24 V (available at the upper end of instrument) can, during the
configuration of each operating mode, be switched on or off. If it is not needed for a measurement, then it should be switched off, in order to conserve energy.
b) Current battery capacity (see chapter 6.4.5 “SETUP additional menu items: CPH
configuration”)
c) Height difference in mm
In the “SETUP\CPH configuration” menu, set the height difference between the test item and the CPT6000 reference pressure sensor. This value influences an automatic correction calculation in order to eliminate any pressure difference based on a medium column. This value must be correct for the following measuring procedure and/or be adjusted accordingly in the “SETUP/CPH configuration” menu (see chapter
6.4.5 “SETUP additional menu items: CPH configuration”).
d) Temperature in [°C]
In the “SETUP\CPH configuration” menu, the temperature (ambient temperature) is entered. This value can be adjusted accordingly in the “SETUP\CPH configuration” menu (see chapter 6.4.5 “SETUP additional menu items: CPH configuration”).
e) Current date of the integrated real-time clock
In the “SETUP\CPH configuration” menu, the date of the real-time clock is set, which is later marked on the calibration certificate. This value must be correct for the following measuring procedure in calibration mode and/or be adjusted accordingly in the “SETUP/CPH configuration” menu (see chapter 6.4.5 “SETUP additional menu items: CPH configuration”).
f) Calibration date for the electrical measuring inputs of the CPH7650 (Year/Month/Day)
Following the status messages, the display returns to the screen for the last-selected operating mode (see following chapter 6.2.3 “Display contents of the operating modes”).
EN
25WIKA operating instructions, model CPH7650
6. Commissioning, operation
6.2.2 Switching the pressure calibrator off
The instrument is switched off using the “Switch off CPH” menu item in the first submenu. For this, press the SETUP key, select the “Switch off CPH” menu item and confirm using the SELECT button. Switch off the CPH7650.
EN
Functions CPH-Info ReferenceSensor Ref. Sensor-list CPH-Configuration Interface CLEAR CalData
CPH switch off
Switching off
6.2.3 Display contents of the operating modes
Operating mode: MEASURING
When a CPH7650 with a CPT6000 reference pressure sensor connected to it is first switched on, the instrument (after displaying a brief status message) switches to MEASURING mode (see following figure)
Display: MEASURING with reference pressure sensor only (without test item)
a b
c
a) Measuring range of the CPT6000 reference pressure sensor (which is currently
connected)
b) Current measured value of the reference pressure sensor c) Pressure unit (adjustable via menu)
26 WIKA operating instructions, model CPH7650
6. Commissioning, operation
In MEASURING mode, at the same time as the reference-pressure value, a test item can also be displayed on the screen (see following figure). With respect to the configuration, see chapters 6.3.1 “MEASURING mode” and 6.3.2 “MEASURING mode (with test item)”.
Display: MEASURING with test item
d
a b
c
P I/U
switchable
e
f
Pressure signal (test item) Electrical signal (test item)
a) Measuring range of the test item b) Current measured value of the test item c) Deviation/difference between reference and test item in the current pressure units and
in % of the measuring span (% FS) or % of reading (% rd) d) Pressure unit (of the test item) e) Original output signal of the test item f) Current value of the output signal of the test item
EN
27WIKA operating instructions, model CPH7650
6. Commissioning, operation
Operating mode: CALIBRATION
In the CALIBRATION mode the data shown above the dashed dividing line is the same as in the “MEASURING with test item” mode.
EN
a
b
c
d e
P I/U
switchable
f
g
Pressure signal (test item) Electrical signal (test item)
a) Current measured value of the CPT6000 reference pressure sensor b) Current measured value of the test item c) Deviation between the test item and reference d) Set point of the calibration e) Actual value of the calibration f) P - 01: Test item No. 1
<01>: Test step No. 1
g) IDENT number of the test item h) Current value of the output signal of the test item
h
28 WIKA operating instructions, model CPH7650
6. Commissioning, operation
Operating mode: SWITCH TEST
In the SWITCH TEST mode, along with the reference pressure sensor data (see MEASURING mode), the status and switch points of the pressure switch are also displayed.
a
f
b
c d
e
a) Current measured value of the CPT6000 reference pressure sensor b) Current switching status/status of the pressure switch c) Opening switch point d) Closing switch point e) Hysteresis/separation between opening and closing of the switch f) Pressure unit (adjustable via menu)
EN
6.2.4 Content of the SETUP menu
a b c
d e
f
g h
i
y
k
Functions CPH-Info ReferenceSensor Ref. Sensor-list CPH-Configuration Interface CLEAR CalData
CPH switch off
29WIKA operating instructions, model CPH7650
6. Commissioning, operation
a) Operating mode MEASURING
To measure working or process pressures
For comparative measurements and/or calibrations (without data recording) of
EN
mechanical* and electrical pressure measuring instruments (supply and display of the test item through the CPH7650)
→ For further information, see chapter 6.3.1 “MEASURING mode” and
6.3.2 “MEASURING mode (with test item)”
b) Operating mode CALIBRATION
For on-site calibration of mechanical 1) and electrical pressure measuring instruments (without PC). In this case the data sets (for up to 16 test items, each with up to 32 test steps including date and time) are recorded within the CPH7650.
1) For mechanical dial instruments, the test item's measured value must be entered via the numeric keypad.
→ For more information please refer to chapter 6.3.3 “Mode CALIBRATION”
c) Operating mode SWITCH TEST
For the easy checking of pressure switches, including automatic calculation of the switch hysteresis.
→ For more information please refer to chapter 6.3.7 “SWITCH TEST mode”
d) Operating functions
Tare: Offset correction of the reference pressure value
Min/Max: Minimum/Maximum memory
Alarm: Min/Max alarm (visual and audible)
Filter: Damping/smoothing of the reference sensor signal
→ For further information see chapter 6.4.1 „SETUP additional menu items: Functions“
e) General CPH7650 instrument data
Calibration data for the electrical measuring inputs
Firmware number
Serial number of the instrument
→ For more information please refer to chapter 6.4.2 “SETUP additional menu items:
CPH info”
30 WIKA operating instructions, model CPH7650
6. Commissioning, operation
f) Data for the currently connected reference pressure sensor
Measuring range
Accuracy class
Pressure type of the sensor
Information in the event of reference sensor overpressure
Calibration data for the reference sensor
→ For more information please refer to chapter 6.4.3 “SETUP additional menu items:
Reference sensor”
g) Reference sensor list
List of the stored reference sensors that can be attached and are calibrated.
→ For more information please refer to chapter 6.4.4 “SETUP additional menu items:
Reference sensor list”
h) CPH configuration
Info: On battery capacity
Setting options from: Menu language, system time/system clock, display brightness, Powersave function (automatic energy saving mode; see chapter 6.4.5 “SETUP additional menu items: CPH configuration”)
Input options:
- Ambient temperature during the calibration
- Height difference existing between the reference pressure sensor and test item (see chapter 4.5.1 “Requirements for test assemblies with the CPH7650”).
EN
→ For more information please refer to chapter 6.4.5 “SETUP additional menu items:
CPH configuration”
i) Interface
USB interface, incl. setting of the baud rate
→ For more information please refer to chapter 6.4.6 “SETUP additional menu items:
Interface”
j) CLEAR CalProg
Delete all stored calibration data (clear and reset all memory space)
→ For more information please refer to chapter 6.4.7 “SETUP additional menu items:
CLEAR CalProg”
k) Switching off the model CPH7650 pressure calibrator
→ For more information please refer to chapter 6.2.2 “Switching the pressure calibrator
off”
31WIKA operating instructions, model CPH7650
6. Commissioning, operation
6.3 Operating modes
6.3.1 MEASURING mode
1. Access SETUP menu
EN
Functions CPH-Info ReferenceSensor Ref. Sensor-list CPH-Configuration Interface CLEAR CalData
CPH switch off
Press
(SETUP key)
Selection
(Menu item)
2. Preparing for MEASURING
..Measuring
a
b c
d
e
f
g
3. Mode: MEASURING
Pressure unit
(adjustable via
menu)
SELECT
SELECT
Confirmation
(of the selection)
Selection
(Menu item); configuration, see next page
Confirmation
(of the input)
32 WIKA operating instructions, model CPH7650
6. Commissioning, operation
In order to switch the instrument into MEASURING mode, follow the instructions on the previous page.
The following is a more detailed explanation of point 2 “Preparation for MEASURING”: a) Test item type and test item measuring signal: [----] For measurement without test item b) Start of measuring range and end of measuring range of the test item currently to be
calibrated
c) Unit and resolution (submenu)
Select and confirm (standard units) via
Customer-specific unit; with respect to bar (input via numeric keypad)
Display resolution in operating mode via (back with )
d) Measurement uncertainty of the test item in % FS (i.e. of the span) or % rd (i.e. of
reading) e) Measurement type for the test item (gauge or absolute) f) Test medium (pneumatic → gas or hydraulic → oil) g) Voltage supply for test item (on/off)
If no external supply is required for the test item, “OFF” should be selected to conserve
energy.
EN
XXX
◁▷ ◀▶
0.00 Parameter input via numeric keypad
ENTER
Short info:
Current cursor position; Alter via
Parameter selection from list or menu via
Input confirmation
CLEAR
Clear entry
33WIKA operating instructions, model CPH7650
6. Commissioning, operation
6.3.2 MEASURING mode (with test item)
1. Access SETUP menu
EN
2. Prepare MEASUREMENT
a
b c
d
e f g
Functions CPH-Info ReferenceSensor Ref. Sensor-list CPH-Configuration Interface CLEAR CalData
CPH switch off
SELECT
Press
(SETUP key)
Selection
(Menu item)
Confirmation
(of the selection)
Selection
(Menu item); configuration, see next page
Confirmation
(of the input)
3. Mode: MEASURING
Deviation
SELECT
P I/U
switchable
Pressure signal (test item) Electrical signal (test item)
34 WIKA operating instructions, model CPH7650
6. Commissioning, operation
Should the instrument be switched into “MEASURING” mode (with test item = display of the test signal as an electrical signal or as a pressure), in order to carry out a comparative measurement or calibration without measured-value recording, then follow the instructions on the previous page.
The following is a more detailed explanation of point 2 “Preparation for MEASURING” a) Test item type and test item measuring signal
0 ... 20 mA / 4 ... 20 mA / 0 ... 1 V / 0 ... 2 V / 0 ... 5 V / 0 ... 10 V / or mechanical for dial pressure gauge
If a comparative measurement with a mechanical dial instrument (test item) is being made, then the gauge’s measured value should be entered via the numeric keypad and confirmed with the ENTER key.
b) Start of measuring range and end of measuring range of the test item currently to be
calibrated c) Unit and resolution (submenu)
Select and confirm (standard units) via
Customer-specific unit; with respect to bar (input via numeric keypad)
EN
Display resolution in operating mode via (back with )
d) Measurement uncertainty of the test item in % FS (i.e. of the span) or % rd (i.e. of
reading) e) Measurement type for the test item (gauge or absolute) f) Test medium (pneumatic → gas or hydraulic → oil) g) Voltage supply for test item (on/off)
If no external supply is required for the test item, “OFF” should be selected to conserve
energy.
XXX
◁▷ ◀▶
0.00 Parameter input via numeric keypad
ENTER
Short info:
Current cursor position; Alter via
Parameter selection from list or menu via
Input confirmation
CLEAR
Clear entry
35WIKA operating instructions, model CPH7650
6. Commissioning, operation
Functions CPH-Info ReferenceSensor Ref. Sensor-list CPH-Configuration Interface CLEAR CalData
CPH switch off
6.3.3 Mode CALIBRATION
1. Access SETUP menu
EN
2. Preparation for CALIBRATION
a b c d
e f g h
i y k
l
m
n
SELECT
Press
(SETUP key)
Selection
(Menu item)
Confirmation
(of the selection)
Selection
(Menu item); configuration, see next page
SELECT
Confirmation
(of the input)
3. Mode: Calibration
Deviation
36 WIKA operating instructions, model CPH7650
Pressure signal (test item) Electrical signal (test item)
P I/U
switchable
6. Commissioning, operation
In order to put the instrument into CALIBRATION mode, the procedure on the previous page should be followed.
The following is a more detailed explanation of point 2 “Preparation for CALIBRATION” a) Number of the calibration and therefore the test item (up to 16 calibrations, each with
up to 32 test steps, can be predefined and then recorded) b) Test item type and test item measuring signal
0 ... 20 mA / 4 ... 20 mA / 0 ... 1 V / 0 ... 2 V / 0 ... 5 V / 0 ... 10 V / or mechanical for dial
pressure gauge c) IDENT number of the test item d) Measuring point number of the test item e) Start of measuring range and end of measuring range of the test item currently to be
calibrated f) Unit and resolution (submenu)
Select and confirm (standard units) via
Customer-specific unit; with respect to bar (input via numeric keypad)
Display resolution in operating mode via (back using )
g) Measurement uncertainty of the test item in % FS (i.e. of the span) or % rd (i.e. of
reading) h) Measurement type for the test item (gauge or absolute) i) Test medium (pneumatic → gas or hydraulic → oil) j) Voltage supply for test item (on/off) [If no external supply is required for the test item,
“OFF” should be selected to conserve energy] k) Number of the test point x l) Optional delay time [sec] (see chapter 6.3.4 “CALIBRATION mode (preparing the test
points of a calibration)”) m) Value of the test point x (input via numeric keypad)
(Test point x+1 and x-1 accessible via
)
n) True value of the test item (will be recorded during the calibration)
EN
37WIKA operating instructions, model CPH7650
6. Commissioning, operation
6.3.4 CALIBRATION mode (preparing the test points of a calibration)
1. Test point
(define)
EN
1. Test point
(defined)
ENTER
Calibration/test item No. 1
Select from menu item: “Set point”
No. of the test point Set point of the test point
Entry of the test point (e.g. 0 bar) via numeric keypad and confirmation with
Test point No. 1 = 0 bar
2. Test point (define)
← 2. Call up test point
(with
back to previous test
point)
x. Test point (define)
38 WIKA operating instructions, model CPH7650
6. Commissioning, operation
With this example, the definition of individual test points/pressure steps for a calibration is clarified. It is possible to prepare up to 16 calibrations, each with up to 32 test steps.
Accessing the menu item
Via
and selection of the menu item: CALIBRATION (or see chapter 6.3.3 “Mode CALIBRATION”) Enter the desired test points in the way described on the previous page.
With the calibration of pressure measuring instruments with electrical output signals (pressure transmitters/transmitters) the reference is calibrated to the display (i.e. the pressure is always adjusted so that the reference value matches the set point exactly). As an exact setting of the pressure is not always possible under certain circumstances, the true reference value is also recorded along with the actual value (test item value) and the set point (reference value). With the WIKA-Cal software, this can be listed in the calibration certificate.
If the calibration is to follow DKD/DAkkS guidelines, then the measured value for each subsequent test point should not be recorded until a defined time has passed (for example 30 seconds), consisting of a load change time and settling time (see figure A: Calibration cycle to DKD/DAkkS guideline 6
- 1 for measurement uncertainty > 0.6 % of measuring span).
Z
M1
M2
2 Minutes
30 s 30 s
For Bourdon tube pressure gauges: 5 minutes
Readings
EN
Figure A
39WIKA operating instructions, model CPH7650
6. Commissioning, operation
EN
optional delay time [sec] Input via numeric input keypad and confirmation
ENTER
with
With the entry of such a delay time, the acceptance/recording of the test point is blocked for this duration. (In the above example, after the first test point has been recorded, 30 seconds must pass before the second test point can be recorded.)
Should all test points need to be cleared or reset, since the new calibration consists of fewer test points than the previous test series, the CLEAR button simply needs to be pressed. This will clear and reset the current and all subsequent test points. (This process can take several seconds.)
Should all saved calibration data for all calibrations need to be cleared at a single time, see chapter 6.4.7 “SETUP additional menu items: CLEAR CalProg”.
40 WIKA operating instructions, model CPH7650
6. Commissioning, operation
6.3.5 CALIBRATION mode (calibration of a pressure transmitter)
1. Test point
(e.g. 0 bar)
Deviation
No. of test item
and No. of test
point
2. Test point
(e.g. 1 bar)
ENTER
Generate the specified set point in accordance with the reference display (establish a pressure-free condition/ atmosphere) and with
Record the measured values of the test point
EN
Deviation
No. of test item
and No. of test
point
2. Test point
Deviation
No. of test item
and No. of test
x. Test point
point
ENTER
Generate the specified set point in accordance with the reference display using the pressure generator
Record the measured values of the test point
(with test point)
back to previous
41WIKA operating instructions, model CPH7650
6. Commissioning, operation
6.3.6 CALIBRATION mode (calibration of a pressure gauge)
1. Test point
(e.g. 0 bar)
EN
Deviation
Generate the specified set point in accordance with the test item display using the pressure generator
No. of test item
and No. of test
2. Test point
(e.g. 1 bar)
Deviation
No. of test item
and No. of test
2. Test point
point
point
ENTER
If the set point = 0 bar, the calibration assembly must be brought to a pressure-free condition/vented to atmosphere (test item must indicate 0 bar; if necessary, make a zero point setting) and with
Record the measured values of the test point
Generate the specified set point in accordance with the test item display using the pressure generator
Deviation
No. of test item
and No. of test
x. Test point
ENTER
point
Record the measured values of the test point
(with test point)
back to previous
42 WIKA operating instructions, model CPH7650
6. Commissioning, operation
Functions CPH-Info ReferenceSensor Ref. Sensor-list CPH-Configuration Interface CLEAR CalData
CPH switch off
6.3.7 SWITCH TEST mode
1. Access SETUP MENU
2. Preparing the SWITCH
TEST
a
b
SELECT
Press
(SETUP key)
EN
Selection
(Menu item)
Confirmation
(of the selection)
Selection
(Menu item); configuration, see next page
3. Mode: SWITCH TEST
switching status
Switch points
Current
Hysteresis
SELECT
Confirmation
(of the input)
Before the pressure switch test After the pressure switch test
43WIKA operating instructions, model CPH7650
6. Commissioning, operation
In order to put the instrument into SWITCH TEST mode, the procedure on the previous page should be followed.
EN
The switch test is not suitable for electronic switches (e.g. PNP- or NPN switches), but is only for mechanical, potential-free switches.
The following is a more detailed explanation of point 2 “Preparation for SWITCH TEST” a) Voltage supply for test item (on/off) [If no external supply is required for the test item,
“OFF” should be selected to conserve energy]
b)Unit and resolution (submenu)
Select and confirm (standard unit) via
Customer-specific unit; with respect to bar (input via numeric keypad)
Display resolution in operating mode via (back with )
The calculated measured values of the two switch points and the hysteresis can be reset by pressing the “0” key.
44 WIKA operating instructions, model CPH7650
6. Commissioning, operation
6.4 SETUP additional menu items
6.4.1 SETUP additional menu items: Functions
a
b
c
d
e
a) Current measured value of the connected CPT6000 reference pressure sensor b) Offset function that influences the current measured value. The value entered is added
to the current measured value. c) (e.g. Ref. 0.000 and Tare: 1.000 → [new] Ref. 1.000) d) Minimum and maximum value memory
The memory is reset by highlighting the value with the cursor (via CLEAR button.
d) Audible and visible alarm function
upper alarm limit: ≥ bar lower alarm limit: ≤ bar
If the current measured value goes outside the set alarm limits, an intermittent alarm tone sounds and the lower status line blinks. Activation via: Move the cursor to the field next to the word Alarm that reads <OFF> and via change it to <ON>. Deactivation via: set back to <OFF>
e) Filter [1-5]:
Damping/smoothing of the reference sensor signal Definition of the numbers: 1 = no additional smoothing ... 5 = strong smoothing
..Functions
0.000
Ref.:
0.000
Tare:
0.000
Min:
0.000
Max:
Alarm:
Filter:
Ref. = rel. Pressure
Offset pressure
ON
10.00
-1.00
0
bar
bar
bar bar
bar bar
ENTER
Select menu item
Input via numeric keypad
Confirmation of input
(Clear deletes the input, or resets the MIN/MAX memory)
) and pressing the
EN
45WIKA operating instructions, model CPH7650
6. Commissioning, operation
6.4.2 SETUP additional menu items: CPH info
EN
Info
CPH7650
CPH6000
a b
c
Cal-Dat.: 2011/05/10
Firmware: 20.05
Serial No: 6000.001
In this menu item general data are listed, such as: a) Calibration date for the calibration of the electrical measuring inputs of the CPH7650
(Year/Month/Day)
b) Firmware version of the CPH7650 c) Serial number of the CPH7650
46 WIKA operating instructions, model CPH7650
6. Commissioning, operation
6.4.3 SETUP additional menu items: Reference sensor
..ReferenceSensor
a
b
c d e
f
Sensor no:
R.-Start: Range-End: Unit: Class:
%FS Pressuretype: Overload:
Read
362A
0.00
10.00 bar
0.025
0 sec
EN
rel
g h
i
Ref-Value: Offset: Cal-Dat.:
Reference sensor
0.001
0.000
2011/05/10
a) Sensor number of the currently connected CPT6000 reference pressure sensor b) Start of measuring range and end of measuring range for the currently connected
CPT6000 reference pressure sensor c) Basic pressure unit of the CPT6000 reference pressure sensor d) Accuracy of the measuring chain of the CPH7650 with connected CPT6000 reference
pressure sensor e) Pressure type of the currently connected CPT6000 reference pressure sensor
(overpressure (gauge pressure) or absolute pressure) f) Length of time for which the CPT6000 reference pressure sensor was unacceptably
overloaded.
If the value here is not equal to zero, then it is highly probable that the instrument no longer meets its specified class accuracy. The only solution for this is an immediate recalibration. (For absolute pressure sensors < 1 bar this function is deactivated, since for this measuring range atmospheric pressure already represents an overload)
g) Current measured value of the connected CPT6000 reference pressure sensor h) This menu option only appears if the CPH7650’s reference pressure sensor is an
absolute pressure sensor.
Through this menu option the measured value of the reference pressure sensor can be
adjusted. This should only be used, however, as close as possible to the absolute zero,
and using a reference that is at least 4 times more accurate. i) Calibration date of the CPT6000 reference pressure sensor (Year/Month/Day)
47WIKA operating instructions, model CPH7650
6. Commissioning, operation
6.4.4 SETUP additional menu items: Reference sensor list
..Ref. Sensor-list
362A Rp52
Kd35 Rp56
----
362A
06 07 08 09 10
10.000
2011/05/10
----
----
----
----
----
0.000
bar
0.025
Selection of a listed sensor with
EN
a
Current
01 02
b
03 04 05
R.-Start: Range-End:
c
Unit: Cal-Dat.: Class:
The CPH7650 pressure calibrator supports up to 10 CPT6000 reference pressure sensors. These are listed in this menu.
a) Currently connected CPT6000 reference pressure sensor b) Sensor list of the supported CPT6000 reference pressure sensors (calibrated with the
instrument)
c) Data of the sensor which has been selected using the cursor
(calibration date: Year/Month/Day)
48 WIKA operating instructions, model CPH7650
6. Commissioning, operation
6.4.5 SETUP additional menu items: CPH configuration
a b
c
d e
ENTER
f
g h
Selection of menu item
Input via numeric keypad
Confirmation of input
(Clear deletes the input)
This menu section lists general instrument settings, such as: a) Input possibility for an ambient temperature b) Input possibility for a height difference between reference pressure sensor and test
item, used in the automatic correction (deduction of a medium column) c) Selection of the menu language (German/English/French/Spanish/Italian) d) Date of the system clock (Year/Month/Day) e) Time of the system clock (Hours/Minutes/Seconds) f) Brightness of the backlighting of the display g) Powersave function (automatic switch-off time for backlighting and internal 24 V test
item power supply).
If the instrument is idle for the set switch-off time (no buttons pressed and no interface
communication), then the backlighting and the 24 V test item power supply will
be switched off, until any button is pressed (except ON/OFF) or the instrument is
addressed over the interface. h) Current battery capacity
At 10 % the low battery charge warning, “low BAT” is shown in the display.
EN
49WIKA operating instructions, model CPH7650
6. Commissioning, operation
6.4.6 SETUP additional menu items: Interface
..Interface
USB
EN
RS232 (8N1)
The USB interface can be switched on and off. In order to extend the battery life, the USB interface should be deactivated when not in use.
ON
9600
SELECT
Select serial interface
Confirm selection
Change between interface type and baud rate
6.4.7 SETUP additional menu items: CLEAR CalProg
If the SETUP menu item CLEAR CalData is selected with the cursor and then the SELECT button pressed 2 x, then all stored calibration data will be deleted or reset.
The status of the deletion process is shown at the left of the lower info screen.
50 WIKA operating instructions, model CPH7650
6. Commissioning, operation
6.5 Connection of the model CPT6000 reference pressure sensor
WARNING!
Only use model CPT6000 reference pressure sensors!
Using other sensors could damage both the pressure calibrator and the reference pressure sensor.
To change the sensor, switch off the pressure calibrator and make sure the system is vented. Before switching the instrument on, connect the sensor, otherwise it may not be correctly identified by the instrument.
When the CPH7650 is switched on, the CPT6000 reference pressure sensor must not be under pressure, but rather should be at atmospheric pressure. For overpressure or gauge pressure sensors, there is a pressure compensation vent in the top of the sensor under the plastic fitting. This vent (with integrated diaphragm) must always remain clear!
CAUTION!
Only ever use the original WIKA sensor connection cable in the operation of CPT6000 reference pressure sensors.
6.6 Voltage supply
The internal Lithium-Ion battery, which can be easily charged with the battery charger supplied with the equipment, serves as the voltage supply for the instrument. To charge the CPH7650 rechargeable batteries, the mains plug of the charger/power supply unit must always be plugged into a mains socket and accessible, so that one can always remove it from the mains socket without difficulty.
EN
CAUTION!
For EMC reasons, never use the CPH7650 with a power supply unit connected.
The instrument should be fully charged before being put into operation. The battery capacity status (charge state in %) is displayed shortly after the instrument is switched on and can be learnt about in chapter 6.4.5 “SETUP additional menu items: CPH configuration”.
When the mains lead is connected to the CPH7650, the battery will be charged, even if the CPH7650 is switched off
51WIKA operating instructions, model CPH7650
6. Commissioning, operation
When the battery charger is no longer being used, the mains plug should be disconnected from the mains socket. Do not leave the battery charger connected to the rechargeable battery for longer than one day, since overloading can shorten its
EN
service life.
Should the rechargeable battery still not be fully charged after 24 hours, contact the manufacturer. When not being used, a fully charged battery will lose its charge over time.
Extreme temperatures have an adverse effect on battery charging. As a result, the battery may first need to be either cooled or warmed, as appropriate.
6.7 Charging/discharging the Lithium-Ion rechargeable batteries
CAUTION!
The temperature range over which the Lithium-Ion battery can be charged is 10 ... 45 °C. Charging the Lithium-Ion battery at temperatures outside of this range may lead to heating or damage. In addition, the performance of the Lithium-Ion battery can be affected and the service life reduced.
WARNING!
In order to charge the Lithium-Ion battery, never use any instrument other than that specified by WIKA. When the Lithium-Ion battery is used in instruments other than the instruments specified by WIKA, the performance and service life of the Lithium-Ion battery may be reduced, and, should the instrument cause an abnormal current to flow, it can cause the Lithium-Ion battery to become hot, explode or ignite and result in serious injury.
CAUTION!
The temperature range over which the Lithium-Ion battery can be discharged is -10 ... +60 °C. Use of the Lithium-Ion battery outside of this temperature range may affect the performance of the battery or may reduce its service life.
52 WIKA operating instructions, model CPH7650
6. Commissioning, operation
6.8 Pressure measurement
For pressure calibration, connect the test item to the test item connection. Choose a reference sensor that is suitable for the pressure range and accuracy.
CAUTION!
Pressure sensors may be damaged and/or injuries may occur to the personnel due to improper application of pressure. For a better understanding with respect to overpressure and burst pressure, follow the specifications laid down in these operating instructions (see chapter 3 “Specifications”). The calibrator display will indicate “OL” when an inappropriate pressure is applied. As soon as “OL” is observed on any pressure display, the pressure should be reduced immediately to prevent damage or possible physical injury. “OL” is displayed if the pressure exceeds the nominal range by 110 %. Use the ZERO button to zero the pressure sensor once it is vented to atmospheric pressure.
Media compatibility
CAUTION!
The calibrator should only be used with clean, dry air! To prevent contamination by the test items, the use of a dirt trap is recommended (see chapter 10 “Accessories”)!
6.9 Measuring and sourcing current (4 ... 20 mA)
To measure current, use the input terminals via the display on the calibrator and select the mA function. Current is measured in mA and percentage of the measuring range. The measuring range of the calibrator is set to 0 % at 4 mA and 100 % at 20 mA.
EN
53WIKA operating instructions, model CPH7650
7. Maintenance, cleaning and recalibration
7. Maintenance, cleaning and recalibration
7.1 Maintenance
This model CPH7650 portable pressure calibrator is maintenance-free.
EN
Repairs must only be carried out by the manufacturer.
CAUTION!
To avoid personal injury or damage to the calibrator, only use accessories supplied by WIKA and ensure that no water finds its way into the case.
7.2 Cleaning
CAUTION!
Before cleaning, vent and switch off the pressure calibrator.
In order not to damage the display or case, do not use any solvent agents or abrasives for cleaning.
Clean the instrument with a moist cloth.
Neither electrical connections nor pressure connections must come into contact with moisture.
For information on returning the instrument see chapter 9.2 “Return”.
7.3 Recalibration
DKD/DAkkS certificate - official certificates:
We recommend that the instrument is regularly recalibrated by the manufacturer, with time intervals of approx. 12 months. In addition, every factory recalibration includes an extensive free-of-charge check of all system parameters with respect to their compliance with the specifications. The basic settings will be corrected if necessary.
54 WIKA operating instructions, model CPH7650
8. Faults
8. Faults
Indication Cause Measures
OL
-OL
No display or undefinable characters, instrument is not responding to a push of a button
Low battery voltage, functioning is only guaranteed for a short period of time
Reading is significantly above or below the measuring range
Battery is empty Charging the battery System error Switch off the instrument, wait for a
Technical fault Send in for repair
Charging the battery
Check: Is the pressure within the permissible measuring range of the sensor?
→ Reduce the applied pressure
short period of time, switch on again
CAUTION!
If faults cannot be eliminated by means of the measures listed above, the hand-held pressure calibrator must be shut down immediately, and it must be ensured that pressure and/or signal are no longer present, and it must be prevented from being inadvertently put back into service. In this case, contact the manufacturer. If a return is needed, please follow the instructions given in chapter 9.2 „Return“.
EN
55WIKA operating instructions, model CPH7650
9. Dismounting, return and disposal
9. Dismounting, return and disposal
WARNING!
EN
9.1 Dismounting
Only disconnect the portable pressure calibrator once the system has been depressurised!
9.2 Return
When returning the instrument, use the original packaging or a suitable transport packaging.
To avoid damage:
1. Wrap the instrument in an antistatic plastic film.
2. Place the instrument, along with shock-absorbent material, in the packaging. Place shock-absorbent material evenly on all sides of the transport packaging.
3. If possible, place a bag containing a desiccant inside the packaging.
4. Label the shipment as carriage of a highly sensitive measuring instrument.
Residual media at the instrument can result in a risk to persons, the environment and equipment. Take sufficient precautionary measures.
WARNING! Strictly observe the following when shipping the instrument:
All instruments delivered to WIKA must be free from any kind of hazardous substances (acids, bases, solutions etc.).
Information on returns can be found under the heading “Service” on our local website.
9.3 Disposal
Incorrect disposal can put the environment at risk. Dispose of instrument components and packaging materials in an environmentally compatible way and in accordance with the country-specific waste disposal regulations.
This marking on the instruments indicates that they must not be disposed of in domestic waste. The disposal is carried out by return to the manufacturer or by the corresponding municipal authorities.
56 WIKA operating instructions, model CPH7650
10. Accessories
10. Accessories
Accessories Order number
Dirt trap 13477103 USB serial adapter 12981878 1 pair of test cables (red/black) 12981975
For further accessories, see current WIKA price list: Calibration technology.
EN
57WIKA operating instructions, model CPH7650
EN
58 WIKA operating instructions, model CPH7650
Inhalt
Inhalt
1. Allgemeines 61
2. Sicherheit 62
2.1 Bestimmungsgemäße Verwendung. . . . . . . . . . . . . . . 63
2.2 Personalqualifikation. . . . . . . . . . . . . . . . . . . . 64
2.3 Besondere Gefahren. . . . . . . . . . . . . . . . . . . . 64
2.4 Verwendung des Lithium-Ionen-Akkus. . . . . . . . . . . . . . 66
2.5 Beschilderung, Sicherheitskennzeichnungen . . . . . . . . . . . 68
3. Technische Daten 69
4. Aufbau und Funktion 71
4.1 Beschreibung . . . . . . . . . . . . . . . . . . . . . . 71
4.2 Front . . . . . . . . . . . . . . . . . . . . . . . . .72
4.3 Lieferumfang . . . . . . . . . . . . . . . . . . . . . . 73
4.4 Spannungsversorgung . . . . . . . . . . . . . . . . . . .73
4.5 Bedienoberfläche . . . . . . . . . . . . . . . . . . . . . 74
4.5.1 Anforderung an Prüfaufbauten mit dem CPH7650 . . . . . . . . . . . . .75
4.5.2 Wichtige Geräteeinstellungen bzgl. einer Kalibrierung via Kalibriermodus 75
4.5.3 Nullpunkteinstellung bzw. Offsetkorrektur . . . . . . . . . . . . . . . . . .76
4.5.4 Referenz-Drucksensor CPT6000. . . . . . . . . . . . . . . . . . . . . . .76
5. Transport, Verpackung und Lagerung 78
5.1 Transport. . . . . . . . . . . . . . . . . . . . . . . . 78
5.2 Verpackung . . . . . . . . . . . . . . . . . . . . . . .78
5.3 Lagerung. . . . . . . . . . . . . . . . . . . . . . . . 78
6. Inbetriebnahme, Betrieb 79
6.1 Menü-Struktur (Arbeitsmodi) . . . . . . . . . . . . . . . . . 80
6.2 Display-Darstellungen . . . . . . . . . . . . . . . . . . . 81
6.2.1 Geräte-Statusmeldung kurz nach dem Einschalten des CPH7650 . . . .81
6.2.2 Druckkalibrator ausschalten . . . . . . . . . . . . . . . . . . . . . . . . .82
6.2.3 Displayinhalte der Arbeitsmodi . . . . . . . . . . . . . . . . . . . . . . . .82
6.2.4 Inhalt des SETUP-Menüs . . . . . . . . . . . . . . . . . . . . . . . . . . .85
6.3 Arbeitsmodi . . . . . . . . . . . . . . . . . . . . . . .88
6.3.1 Modus MESSEN. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .88
6.3.2 Modus MESSEN (mit Prüfling) . . . . . . . . . . . . . . . . . . . . . . . .90
6.3.3 Modus KALIBRIEREN. . . . . . . . . . . . . . . . . . . . . . . . . . . . .92
6.3.4 Modus KALIBRIEREN (die Prüfpunkte einer Kalibrierung vorbereiten) . .94
6.3.5 Modus KALIBRIEREN (die Kalibrierung eines Druckmessumformers) . .97
6.3.6 Modus KALIBRIEREN (die Kalibrierung eines Manometers) . . . . . . .98
6.3.7 Modus SCHALTERTEST . . . . . . . . . . . . . . . . . . . . . . . . . . .99
6.4 SETUP-Zusatzmenüpunkte . . . . . . . . . . . . . . . . 101
DE
59WIKA Betriebsanleitung, Typ CPH7650
Inhalt
6.4.1 SETUP-Zusatzmenüpunkte: Funktionen. . . . . . . . . . . . . . . . . .101
6.4.2 SETUP-Zusatzmenüpunkte: CPH-Info. . . . . . . . . . . . . . . . . . .102
6.4.3 SETUP-Zusatzmenüpunkte: Referenz-Sensor . . . . . . . . . . . . . . 103
6.4.4 SETUP-Zusatzmenüpunkte: Referenz-Sensor-Liste . . . . . . . . . . . 104
6.4.5 SETUP-Zusatzmenüpunkte: CPH-Einstellungen . . . . . . . . . . . . . 105
DE
6.4.6 SETUP-Zusatzmenüpunkte: Schnittstelle . . . . . . . . . . . . . . . . . 106
6.4.7 SETUP-Zusatzmenüpunkte: CLEAR KalProg . . . . . . . . . . . . . . .106
6.5 Anschluss des Referenz-Drucksensors Typ CPT6000 . . . . . . . 107
6.6 Spannungsversorgung . . . . . . . . . . . . . . . . . . 107
6.7 Laden/Entladen des Lithium-Ionen-Akkus . . . . . . . . . . . 108
6.8 Druckmessung. . . . . . . . . . . . . . . . . . . . . 109
6.9 Strom messen und geben (4 ... 20 mA) . . . . . . . . . . . . 109
7. Wartung, Reinigung und Rekalibrierung 110
7.1 Wartung . . . . . . . . . . . . . . . . . . . . . . . 110
7.2 Reinigung . . . . . . . . . . . . . . . . . . . . . . 110
7.3 Rekalibrierung . . . . . . . . . . . . . . . . . . . . . 110
8. Störungen 111
9. Demontage, Rücksendung und Entsorgung 112
9.1 Demontage . . . . . . . . . . . . . . . . . . . . . . 112
9.2 Rücksendung . . . . . . . . . . . . . . . . . . . . . 112
9.3 Entsorgung . . . . . . . . . . . . . . . . . . . . . . 112
10. Zubehör 113
Konformitätserklärungen finden Sie online unter www.wika.de.
60 WIKA Betriebsanleitung, Typ CPH7650
1. Allgemeines
1. Allgemeines
Der in der Betriebsanleitung beschriebene tragbare Druckkalibrator wird nach dem aktuellen Stand der Technik gefertigt. Alle Komponenten unterliegen während der Fertigung strengen Qualitäts- und Umweltkriterien. Unsere Managementsysteme sind nach ISO 9001 und ISO 14001 zertifiziert.
Diese Betriebsanleitung gibt wichtige Hinweise zum Umgang mit dem Gerät. Voraus­setzung für sicheres Arbeiten ist die Einhaltung aller angegebenen Sicherheitshinwei­se und Handlungsanweisungen.
Die für den Einsatzbereich des Gerätes geltenden örtlichen Unfallverhütungsvorschrif­ten und allgemeinen Sicherheitsbestimmungen einhalten.
Die Betriebsanleitung ist Produktbestandteil und muss in unmittelbarer Nähe des Gerätes für das Fachpersonal jederzeit zugänglich aufbewahrt werden.
Das Fachpersonal muss die Betriebsanleitung vor Beginn aller Arbeiten sorgfältig durchgelesen und verstanden haben.
Die Haftung des Herstellers erlischt bei Schäden durch bestimmungswidrige Verwen­dung, Nichtbeachten dieser Betriebsanleitung, Einsatz ungenügend qualifizierten Fachpersonals sowie eigenmächtiger Veränderung am Gerät.
DE
Es gelten die allgemeinen Geschäftsbedingungen in den Verkaufsunterlagen.
Technische Änderungen vorbehalten.
Werkskalibrierungen / DKD/DAkkS-Kalibrierungen erfolgen nach internationalen Normen.
Weitere Informationen:
- Internet-Adresse: www.wika.de / www.wika.com
- zugehöriges Datenblatt: CT 17.02
- Anwendungsberater:
Tel.: +49 9372 132-9986 Fax: +49 9372 132-8767 info@wika.de
61WIKA Betriebsanleitung, Typ CPH7650
1. Allgemeines / 2. Sicherheit
Symbolerklärung
WARNUNG!
... weist auf eine möglicherweise gefährliche Situation hin, die zum Tod oder zu schweren Verletzungen führen kann, wenn sie nicht gemieden wird.
DE
2. Sicherheit
VORSICHT!
... weist auf eine möglicherweise gefährliche Situation hin, die zu geringfü­gigen oder leichten Verletzungen bzw. Sach- und Umweltschäden führen kann, wenn sie nicht gemieden wird.
Information
… hebt nützliche Tipps und Empfehlungen sowie Informationen für einen effizienten und störungsfreien Betrieb hervor.
GEFAHR!
…kennzeichnet Gefährdungen durch elektrischen Strom. Bei Nichtbeach­tung der Sicherheitshinweise besteht die Gefahr schwerer oder tödlicher Verletzungen.
WARNUNG!
Vor Montage, Inbetriebnahme und Betrieb sicherstellen, dass der richtige Referenz-Drucksensor hinsichtlich Messbereich, Ausführung und spezifi­schen Messbedingungen ausgewählt wurde. Bei Nichtbeachten können schwere Körperverletzungen und/oder Sachschäden auftreten.
Weitere wichtige Sicherheitshinweise befinden sich in den einzelnen Kapiteln dieser Betriebsanleitung.
62 WIKA Betriebsanleitung, Typ CPH7650
2. Sicherheit
2.1 Bestimmungsgemäße Verwendung
Dieser tragbare Druckkalibrator dient als Kalibriergerät für unterschiedlichste Druckmess­geräte und wurde sowohl für den mobilen Einsatz als auch für stationäre Werkstatt- und Laborprüfungen entwickelt. Durch die Kombination der integrierten elektrischen Pumpe und des elektrischen Moduls wird neben dem klassischen Messen des Strom- und Spannungssignals auch eine elektrische Versorgung von Transmittern oder Sensoren mit max. 30 mA (Spannung (Leerlauf) = DC 24 V) ermöglicht. Mit dem Druckkalibrator kann eine vollständige Kalibrierung durchgeführt und dokumentiert werden.
WARNUNG!
Nur Referenz-Drucksensoren Typ CPT6000 verwenden!
Mit anderen Sensoren kann es zur Zerstörung des Druckkalibrators und des Referenz-Drucksensors kommen.
Zum Sensorwechsel den Druckkalibrator ausschalten und vergewissern dass das System belüftet ist. Den Sensor vor dem Einschalten anste­cken, sonst wird er vom Gerät eventuell nicht richtig erkannt.
Zum Zeitpunkt des Einschaltens des CPH7650 darf der Referenz­Drucksensor CPT6000 nicht mit Druck beaufschlagt sein, sondern muss an Atmosphärendruck liegen. Im oberen Teil des Sensorgehäuses unter der Kunststoffverschraubung befindet sich, bei Überdruck- bzw. relativ Drucksensoren, die Öffnung für den Druckausgleich. Diese Öffnung (mit integrierter Membrane) muss unbedingt frei bleiben!
Das Gerät ist ausschließlich für den hier beschriebenen bestimmungsgemäßen Verwen­dungszweck konzipiert und konstruiert und darf nur dementsprechend verwendet werden.
DE
Die technischen Spezifikationen in dieser Betriebsanleitung sind einzuhalten. Eine unsachgemäße Handhabung oder ein Betreiben des Gerätes außerhalb der technischen Spezifikationen macht die sofortige Stilllegung und Überprüfung durch einen autorisierten WIKA-Servicemitarbeiter erforderlich.
Elektronische Präzisionsmessgeräte mit erforderlicher Sorgfalt behandeln (vor Nässe, Stößen, starken Magnetfeldern, statischer Elektrizität und extremen Temperaturen schüt­zen, keine Gegenstände in das Gerät bzw. Öffnungen einführen). Stecker und Buchsen vor Verschmutzung schützen.
Wird das Gerät von einer kalten in eine warme Umgebung transportiert, so kann durch Kondensatbildung eine Störung der Gerätefunktion eintreten. Vor einer erneuten Inbetriebnahme die Angleichung der Gerätetemperatur an die Raumtemperatur abwar­ten.
Ansprüche jeglicher Art aufgrund von nicht bestimmungsgemäßer Verwendung sind ausgeschlossen.
63WIKA Betriebsanleitung, Typ CPH7650
2. Sicherheit
2.2 Personalqualifikation
WARNUNG! Verletzungsgefahr bei unzureichender Qualifikation!
Unsachgemäßer Umgang kann zu erheblichen Personen- und Sachschä­den führen.
DE
Fachpersonal
Das Fachpersonal ist aufgrund seiner fachlichen Ausbildung, seiner Kenntnisse der Mess- und Regelungstechnik und seiner Erfahrungen sowie Kenntnis der landesspezi­fischen Vorschriften, geltenden Normen und Richtlinien in der Lage, die beschriebenen Arbeiten auszuführen und mögliche Gefahren selbstständig zu erkennen.
Spezielle Einsatzbedingungen verlangen weiteres entsprechendes Wissen, z. B. über aggressive Medien.
2.3 Besondere Gefahren
Die in dieser Betriebsanleitung beschriebenen Tätigkeiten nur durch Fachpersonal nachfolgend beschriebener Qualifikation durchführen lassen.
WARNUNG!
Keine größere Spannung als angegeben an das Gerät anlegen (siehe Kapitel 3 „Technische Daten“).
Keinen Druck von außen auf das CPH7650 geben.
Darauf achten, dass die Prüfspitzen nie mit einer Spannungsquelle in Kontakt kommen, wenn die Prüfkabel an die Stromklemmen angeschlos­sen sind.
Den Kalibrator nicht in beschädigtem Zustand verwenden. Vor dem Verwenden des Gerätes prüfen, ob das Gehäuse Risse oder fehlende Kunststoffteile aufweist. Besonders auf die Isolierung der Stecker achten.
Für die Messung die richtige Funktion und den richtigen Messbereich auswählen.
Betriebsparameter gemäß Kapitel 3 „Technische Daten“ beachten.
Druckkalibrator immer innerhalb des definierten Druckbereiches betrei­ben.
Um einen störungsfreien Betrieb zu gewährleisten, das Gerät nur im Akkubetrieb betreiben. Das Netzteil nur zum Laden des Akkus des Gerätes verwenden.
Die Prüfkabel auf eine beschädigte Isolierung oder blankes Metall prüfen. Die Durchgängigkeit der Kabel prüfen. Beschädigte Prüfleitungen austau­schen bevor das Gerät verwendet wird.
Bei Verwendung von Prüfspitzen die Finger von den Prüfspitzenkon­takten fernhalten. Die Finger hinter den Fingerschutz an den Prüfspitzen anlegen.
64 WIKA Betriebsanleitung, Typ CPH7650
2. Sicherheit
Zuerst den Null-Voltleiter anschließen und danach den spannungsführen­den Leiter. Beim Abnehmen zuerst die spannungsführende Prüfleitung abnehmen.
Das Gerät nicht verwenden, wenn es nicht normal funktioniert. Der Geräteschutz kann beeinträchtigt sein. Im Zweifelsfall das Gerät überprü­fen lassen.
Das Gerät nicht im Bereich von explosiven Gasen, Dämpfen oder Staub verwenden.
Bei der Druckmessung sicherstellen, dass der Druckkalibrator drucklos geschaltet ist, bevor der Referenzsensor oder der Prüfling angeschlossen oder abgenommen wird.
Die Prüfleitungen lösen, bevor in eine andere Mess- oder Quellfunktion gewechselt wird.
Das Umschaltventil (+/- Druck) darf nur in drucklosem Zustand betätigt werden.
WARNUNG!
Zur Vermeidung einer falschen Anzeige, die zu einem elektrischen Schlag oder zu Verletzungen führen können, den Akku laden, sobald die Batterieanzeige erscheint.
Um eine mögliche Schädigung des Gerätes oder der Testeinrichtung zu vermeiden, die richtigen Leitungen, die richtige Funktion und den richti­gen Bereich für die Messanwendung verwenden.
DE
GEFAHR! Lebensgefahr durch elektrischen Strom
Bei Berührung mit spannungsführenden Teilen besteht unmittelbare Lebens­gefahr.
Bei Ladevorgang mit einem defekten Netzgerät (z. B. Kurzschluss von Netzspannung zur Ausgangsspannung) können am Gerät lebensgefährli­che Spannungen auftreten!
Nur das von WIKA für den tragbaren Druckkalibrator Typ CPH7650 zugelassene Netzgerät verwenden.
Nur einwandfrei funktionierendes oder unbeschädigtes Ladegerät verwenden.
Die Sicherheit des Benutzers kann durch das Gerät beeinträchtigt werden, wenn es z. B.
sichtbare Schäden aufweist.
nicht mehr wie vorgeschrieben arbeitet.
längere Zeit unter ungeeigneten Bedingungen gelagert wurde.
In Zweifelsfällen das Gerät an den Hersteller zur Reparatur bzw. Wartung einschicken.
65WIKA Betriebsanleitung, Typ CPH7650
2. Sicherheit
2.4 Verwendung des Lithium-Ionen-Akkus
WARNUNG!
Unsachgemäße Verwendung des Lithium-Ionen-Akkus kann zur Erhitzung, Explosion oder Entzündung führen und schwere Verletzungen verursachen. Beachten Sie unbedingt die unten aufgelisteten Sicherheitshinweise:
DE
Löten Sie nicht direkt an den Lithium-Ionen-Akku.
Der Lithium-Ionen-Akku darf weder verbrannt noch erhitzt werden.
Der Lithium-Ionen-Akku darf nur richtig gepolt verbunden werden.
Die positive Klemme und die negative Klemme des Lithium-Ionen-Akkus dürfen nicht über Metallobjekte (z. B. Kabel) miteinander verbunden werden.
Der Lithium-Ionen-Akku darf nicht mit Halsketten, Haarnadeln oder anderen metallischen Objekten getragen oder gelagert werden.
WARNUNG!
Der Lithium-Ionen-Akku darf weder mit Nägeln durchstoßen noch mit einem Hammer geschlagen werden. Außerdem ist es weder erlaubt, auf den Lithium-Ionen-Akku zu treten noch ihn anderen starken Stößen und Erschütterungen auszusetzen.
Der Lithium-Ionen-Akku darf weder mit Wasser noch mit Salzwasser in Berührung kommen. Außerdem darf er nicht nass werden.
WARNUNG!
Der Lithium-Ionen-Akku darf weder auseinander gebaut noch verändert werden. Er ist mit Sicherheits- und Schutzeinrichtungen ausgestattet, die bei Beschädigung dazu führen können, dass dieser Hitze erzeugt, explodiert oder sich entzündet.
WARNUNG!
Der Lithium-Ionen-Akku darf nicht in der Nähe von Feuer, Öfen oder anderen Orten mit hohen Temperaturen liegen. Der Lithium-Ionen-Akku darf nicht in direktem Sonnenlicht liegen und bei heißem Wetter in Autos weder verwendet noch aufbewahrt werden. In diesem Fall könnte der Lithium­Ionen-Akku Hitze erzeugen, explodieren oder sich entzünden. Außerdem könnte dabei auch die Leistung und Lebensdauer des Lithium-Ionen-Akkus sinken.
Der Lithium-Ionen-Akku darf nicht in luftdicht abgeschlossene Geräte einge­setzt werden. In einigen Fällen könnte Wasserstoff oder Sauerstoff von der Zelle freigesetzt werden und so zur Beschädigung des Lithium-Ionen-Akkus, zu Feuer oder Explosion führen.
66 WIKA Betriebsanleitung, Typ CPH7650
2. Sicherheit
WARNUNG!
Der Lithium-Ionen-Akku darf absolut nicht mehr verwendet werden, falls er bei Einsatz, Aufladung oder Lagerung ungewöhnlich riecht, heiß ist, die Farbe oder Form wechselt oder in irgendeiner anderen Weise ungewöhnlich erscheint. Falls eines dieser Probleme auftreten sollte, sofort den jeweiligen Vertriebspartner kontaktieren.
Der Lithium-Ionen-Akku darf nicht in Mikrowellen-Herden, Hochdruck­Containern noch Induktionsherden platziert werden.
Falls der Lithium-Ionen-Akku undicht ist und die Flüssigkeit mit den Augen in Berührung kommen sollte, auf keinen Fall die Augen reiben. Spülen Sie die Augen gut mit Wasser aus und suchen Sie unverzüglich einen Arzt auf. Sollten die Augen nicht behandelt werden, können Verletzungen davon getragen werden.
VORSICHT!
Nach Verschleiß des Lithium-Ionen-Akkus müssen die Klemmen vor der Entsorgung mit Klebeband oder ähnlichem Material isoliert werden.
WARNUNG!
Beachten Sie unbedingt die unten aufgelisteten Hinweise beim Aufladen des Lithium-Ionen-Akkus. Andernfalls könnte sich der Lithium-Ionen-Akku erhitzen, explodieren oder entzünden und schwere Verletzungen verursa­chen.
Verwenden Sie zum Aufladen des Lithium-Ionen-Akkus ausschließlich das von WIKA festgelegte Ladegerät.
Schließen Sie den Lithium-Ionen-Akku weder direkt an eine Steckdose noch an den Zigarettenanzünder eines Autos an.
Legen Sie den Lithium-Ionen-Akku weder in die Nähe von Feuer noch in direktes Sonnenlicht. Wenn der Lithium-Ionen-Akku heiß wird, wird die eingebaute Sicherheitseinrichtung aktiviert und verhindert ein Überladen. Das Erhitzen des Lithium-Ionen-Akkus kann die Sicherheitseinrichtung zerstören und kann dazu führen, dass dieser sich weiter erhitzt, kaputt geht oder sich entzündet.
DE
WARNUNG!
Laden Sie den Lithium-Ionen-Akku nicht weiter auf, falls dieser nicht inner­halb der vorgegebenen Zeit vollständig aufgeladen ist. In diesem Fall könnte der Lithium-Ionen-Akku heiß werden, explodieren oder sich entzünden.
67WIKA Betriebsanleitung, Typ CPH7650
2. Sicherheit
2.5 Beschilderung, Sicherheitskennzeichnungen
Typenschild
DE
6 5 4 3 2
1
Produktname
2
Herstellungsdatum (Monat/Jahr)
3
Seriennummer
4
Ausgangssignal, Hilfsenergie
5
Eingangssignal
6
Druckbereich
1
Symbole
Vor Montage und Inbetriebnahme des Gerätes unbedingt die Betriebsanlei­tung lesen!
Bei Geräten mit dieser Kennzeichnung wird darauf hingewiesen, dass diese nicht in den Hausmüll entsorgt werden dürfen. Die Entsorgung erfolgt durch Rücknahme bzw. durch entsprechende kommunale Stellen.
68 WIKA Betriebsanleitung, Typ CPH7650
3. Technische Daten
3. Technische Daten
Messbereiche
1)
Druck
Relativdruck -1 … +20 bar, -1 … +10 bar, 0 … 20 bar, 0 … 10 bar Absolutdruck 0 … 10 bar abs., 0 … 20 bar abs. Überlast-Druckgrenze Genauigkeit 0,025 % FS Auflösung 5-stellig Temperaturkompensation 15 … 35 °C (59 ... 95 °F) Temperaturkoeffizient 0,002 % der Spanne/°C außerhalb von 15 ... 35 °C (59 ... 95 °F)
Strom
Messbereich 0 … 24 mA (max. Last 1.000 Ω) Auflösung 1 µA Genauigkeit 0,015 % vom Messwert ±2 µA (Simulation und Messung)
Spannung
Messbereich DC 0 … 30 V Auflösung 1 mV Genauigkeit 0,015 % vom Messwert ±2 mV (Messung)
1) Über austauschbare Referenzdrucksensoren von Typ CPT6000
2) Die elektrische Pumpe kann -0,85 … +20 bar (-12 … 290 psi) erzeugen.
2)
3-fach
DE
Grundgerät
Druckversorgung -0,85 ... +20 bar (-12 ... +290 psi), über integrierte elektrische
Pumpe
Druckanschluss für Prüfling Innengewinde G ½" Zulässige Medien Trockene, saubere und nicht aggressive Gase
Ausgang
Spannungsversorgung DC 24 V
Hilfsenergie
Akkutyp Lithium-Ionen Akku
Zulässige Umgebungsbedingungen
Betriebstemperatur -10 … +50 °C (14 ... 122 °F) Lagertemperatur -20 … +60 °C (-4 ... +140 °F) Relative Luftfeuchte 35 ... 85 % r. F. (keine Betauung)
Kommunikation
Schnittstelle USB via speziellem Schnittstellenkabel
69WIKA Betriebsanleitung, Typ CPH7650
3. Technische Daten
Grundgerät
Gehäuse
Koffer NK-7TM Harz Frontplatte Aluminium
DE
Schutzart IP67 (Koffer geschlossen)
IP40 (Koffer geöffnet) Abmessungen 387,4 x 304,8 x 177,8 mm (15,25 x 12 x 7 in) Gewicht ca. 7 kg (15,5 lbs.)
Zertifikate/Zeugnisse
Zertifikat
Kalibrierung Standard: Kalibrierzertifikat 3.1 nach DIN EN 10204
Option: DKD/DAkkS-Kalibrierzertifikat
Empfohlenes Rekalibrie-
1 Jahr (abhängig von den Nutzungsbedingungen)
rungsintervall
Zulassungen und Zertifikate siehe Internetseite
Verfügbarer Druckbereich und Auflösung
Druckbereich und Faktoren
Relativdruck -1 ... +20 bar (-14,5 ... +290 psi)
Überlast-Druckgrenze 40 bar (580 psi) Einheit Umrechnungsfaktor psi 1 300,00 bar 0,06894757 20,684 mbar 68,94757 20.684 kPa 6,894757 2.068,4 MPa 0,00689476 2,0684 kg/cm² 0,07030697 21,092 cmH
O (4 °C) 70,3089 21.093
2
cmH
O (20 °C) 70,4336 21.130
2
inH
O (4 °C) 27,68067 8.304,2
2
inH
O (20 °C) 27,72977 8.318,9
2
inH
O (60 °F) 27,70759 8.312,3
2
mmHg (0 °C) 51,71508 15.515 inHg (0 °C) 2,03602 610,81
3) Elektrische Pumpe: -0,85 … +20 bar (-12 … 290 psi)
70 WIKA Betriebsanleitung, Typ CPH7650
3. Technische Daten / 4. Aufbau und Funktion
104 (4,11)
158 (6,22)
Abmessungen in mm (in)
172 (6,76)
390 (15,36)
DE
308 (12,13)
4. Aufbau und Funktion
4.1 Beschreibung
Der CPH7650 ist ein kompakter, tragbarer Druckkalibrator, der sowohl für den mobilen Einsatz als auch für stationäre Werkstatt- und Laborprüfungen entwickelt wurde. Durch eine integrierte elektrische Pumpe können Drücke bis zu 20 bar (300 psi) erzeugt werden, ein integriertes elektrisches Modul ermöglicht neben dem klassischen Messen des Strom- und Spannungssignals auch eine elektrische Versorgung von Transmittern oder Sensoren mit max. 30 mA (Spannung (Leerlauf) = DC 24 V).
Mit dem Druckkalibrator kann eine vollständige Kalibrierung durchgeführt und dokumen­tiert werden. Weitere Features sind neben dem Anzeigen bzw. Messen und Kalibrieren auch ein Schaltertest. In Verbindung mit der Software WIKA-Cal ist auch eine vollständige Dokumentation möglich.
71WIKA Betriebsanleitung, Typ CPH7650
4. Aufbau und Funktion
4.2 Front
1
2
3 4
DE
10111213
1
Prüfling
2
Übersicht elektrischer Anschluss
3
Elektrische Verbindung zum Referenzsensor Typ CPT6000
4
Referenzsensor CPT6000
5
Anschluss Netzgerät
6
Anschluss WIKA-Cal
7
Feinjustierung / Ablassventil
5
6
7
8
9
8
Schalter für Druck / Vakuum
9
Regler Pumpengeschwindigkeit
10
Bedienung
11
Elektrisches Modul
12
Display
13
Nummernblock
72 WIKA Betriebsanleitung, Typ CPH7650
4. Aufbau und Funktion
4.3 Lieferumfang
Tragbarer Druckkalibrator Typ CPH7650
Betriebsanleitung
Prüfkabel
Akku-Ladegerät
Kalibrierzertifikat 3.1 nach DIN EN 10204
Lieferumfang mit dem Lieferschein abgleichen.
4.4 Spannungsversorgung
Akku laden Um falsche Messungen zu vermeiden, den Akku laden, sobald die Batterieanzeige erscheint. Wenn sich der Akku zu sehr entladen hat, schaltet sich das Gerät automatisch aus.
Verwenden Sie ausschließlich das von WIKA für den tragbaren Druckkalibrator Typ CPH7650 zugelassene Netzgerät.
Um Messunsicherheiten auszuschließen das CPH7650 nur ohne angeschlossenes Netzbetrieb verwenden. Die volle Pumpenleistung kann nur bei vollem Akku und ohne angeschlossenes Netzgerät garantiert werden.
Das Gerät sollte vor dem Arbeitseinsatz vollständig geladen sein. Der Akkukapazitätsstatus (Ladezustand in %) wird kurz nach dem Einschalten des Gerätes angezeigt. Ist das Netzteil angeschlossen, so wird der Akku auch im ausgeschal­teten Zustand des CPH7650 geladen.
DE
Die typische Ladezeit des Akkus beträgt < 5 Stunden.
WARNUNG!
Wenn das Netzteil nicht mehr verwendet wird, Netzstecker aus der Steckdose ziehen. Den Akku nicht länger als einen Tag am Netzteil angeschlossen lassen, da eine zu starke Aufladung seine Lebensdauer verkürzen kann.
Sollte der Akku nach 24 Stunden nicht vollständig geladen sein den Hersteller kontaktieren. Bei Nichtgebrauch entlädt sich ein voll aufgelade­ner Akku mit der Zeit.
Extreme Temperaturen haben einen nachteiligen Einfluss auf das Laden des Akkus. Deshalb kann zunächst ein Abkühlen oder Aufwärmen des Akkus erforderlich sein.
Wenn der Akku beinahe vollständig entladen ist, erscheint im Display der Hinweis „low BAT“. Um Datenverluste zu vermeiden muss das Gerät unverzüglich geladen werden
73WIKA Betriebsanleitung, Typ CPH7650
4. Aufbau und Funktion
4.5 Bedienoberfläche
1
DE
6
5
4
1
SETUP-Menü
2
Auswahl und Eingabe-Aktivierung
3
1 Schritt zurück
4
Löschen der Eingabe
5
Eingabe-Quittierung
6
Numerischer Eingabeblock
Einschalten durch Betätigen einer beliebigen Taste Ausschalten durch Menüpunkt im Hauptmenü
2
3
74 WIKA Betriebsanleitung, Typ CPH7650
4. Aufbau und Funktion
4.5.1 Anforderung an Prüfaufbauten mit dem CPH7650
Vor dem Arbeitsbeginn das Gerät kurz einschaltet, um festzustellen, ob der Akku über einen ausreichenden Ladezustand verfügt (Akkukapazität in %). Die Akkukapazität wird durch eine Geräte-Statusmeldung kurz nach dem Einschalten angezeigt (siehe Kapitel 6.2.1 „Geräte-Statusmeldung kurz nach dem Einschalten des CPH7650“).
Als Erstes die Prüfaufbauten mechanisch und ggf. elektrisch verbinden/aufbauen (siehe Kapitel 4.5.4.1 „Anschluss des Referenz-Drucksensors Typ CPT6000“).
Vor dem Einschalten des CPH7650 sicherstellen, dass sich der Prüfaufbau im druck­losen Zustand (System belüftet zu Atmosphäre) befindet und die Geräte die korrekte Einbaulage/Position aufweisen.
Prüf- und Kalibrieraufbauten nur im drucklosen Zustand montieren!
Höhendifferenz-Kompensation
Existiert zwischen Referenz-Drucksensor CPT6000 und Prüfling ein größerer Höhenun­terschied, so kann ein Druckunterschied basierend auf einer Mediumssäule via Menü automatisch kompensiert werden (siehe Kapitel 6.4.5 „SETUP-Zusatzmenüpunkte: CPH-Einstellungen“).
4.5.2 Wichtige Geräteeinstellungen bzgl. einer Kalibrierung via Kalibriermodus
Kalibrierdatum
Das Gerät verfügt über eine integrierte Echtzeituhr mit Datum. Das aktuelle Datum einer Kalibrierung wird später auf dem Kalibrierzeugnis mit angegeben. Vor dem Beginn einer Kalibrierung sicherstellen, dass das interne Datum des CPH7650 korrekt ist (siehe Kapitel 6.4.5 „SETUP-Zusatzmenüpunkte: CPH-Einstellungen“).
DE
Einheit und Auflösung
Nach der Wahl eines der Haupt-Menüpunkte des SETUP-Menüs (SETUP-Taste drücken), wie z. B. MESSEN, KALIBRIEREN oder SCHALTERTEST kann jeweils via dem Menüpunkt „Einheit“ und seinem zugehörigen Untermenü (Cursor auf „Einheit“ und Richtungspfeil rechts oder links drücken) die Einheit und Auflösung eingestellt werden (siehe Kapitel 6.3 „Arbeitsmodi“).
75WIKA Betriebsanleitung, Typ CPH7650
4. Aufbau und Funktion
Verfügbare Einheiten inkl. Umrechnungsfaktor bezogen auf die Einheit bar
bar 1.00000E+00 mbar 1.00000E-03 hPa 1.00000E-03 psi 6.89475E-02 inHg (0 °C) 3.37690E-02
DE
cmHG (0 °C) 1.33322E-02 MPa 1.00000E+01 kPa 1.00000E-02 Pa 1.00000E-05
O (4 °C) 9.80670E-02
mH
2
O (4 °C) 9.80670E-04
cmH
2
O (4 °C) 9.80670E-05
mmH
2
2
kg/cm
O (60 °C) 2.48800E-03
inH
2
O (0 °C) 1.33322E-03
mmH
2
4.5.3 Nullpunkteinstellung bzw. Offsetkorrektur
Nullpunkteinstellung von Überdrucksensoren Ist der Messwert des CPH7650 bei angeschlossenem Überdrucksensor und belüftetem Prüfaufbau ungleich Null, so kann durch ein zweimaliges Drücken der CLEAR-Taste (innerhalb von fünf Sekunden) der Nullpunkt korrigiert werden (maximale zugelassene Korrekturwert ist in Größe der zweifachen Klassengenauigkeit).
9.80665E-01
Offsetkorrektur von Absolutdrucksensoren Bei Absolutdrucksensoren kann eine Offsetkorrektur via Menü vorgenommen werden (siehe Kapitel 6.4.3 „SETUP-Zusatzmenüpunkte: Referenz-Sensor“).
4.5.4 Referenz-Drucksensor CPT6000
Für den Druckkalibrator Typ CPH7650 stehen eine Vielzahl von Referenz-Drucksensoren mit einer Genauigkeit von 0,025 % zur Auswahl, die schnell und ohne Werkzeug am Gerät ausgetauscht werden können. Beim Einschalten des Druckkalibrators wird der angeschlossene Referenz-Drucksensor automatisch erkannt, so dass jegliche Konfigura­tion des Sensors entfällt.
76 WIKA Betriebsanleitung, Typ CPH7650
4. Aufbau und Funktion
4.5.4.1 Anschluss des Referenz-Drucksensors Typ CPT6000
WARNUNG!
Nur Referenz-Drucksensoren Typ CPT6000 verwenden!
Mit anderen Sensoren kann es zur Zerstörung des Druckkalibrators und des Referenz-Drucksensors kommen.
Zum Sensorwechsel den Druckkalibrator ausschalten und vergewissern dass das System belüftet ist. Den Sensor vor dem Einschalten anste­cken, sonst wird er vom Gerät eventuell nicht richtig erkannt.
Zum Zeitpunkt des Einschaltens des CPH7650 darf der Referenz­Drucksensor CPT6000 nicht mit Druck beaufschlagt sein, sondern muss an Atmosphärendruck liegen. Im oberen Teil des Sensorgehäuses unter der Kunststoffverschraubung befindet sich, bei Überdruck- bzw. relativ Drucksensoren, die Öffnung für den Druckausgleich. Diese Öffnung (mit integrierter Membrane) muss unbedingt frei bleiben!
VORSICHT!
Nur das originale WIKA-Sensoranschlusskabel für den Betrieb des Referenz-Drucksensors CPT6000 verwenden.
WARNUNG!
Die Pumpenleistung ist unabhängig vom Druckbereich des ausgewählten Referenzsensors. Der Bediener muss sicherstellen, dass der Referenzsen­sor CPT6000 nicht überdrückt wird.
DE
4.5.4.2 Elektrischer Anschluss des CPT6000 Referenz-Drucksensors am CPH7650
Kalibrator und Referenz-Drucksensor werden mittels eines separaten Anschlusska­bels elektrisch miteinander verbunden. Zum elektrischen Anschluss eines Referenz­Drucksensors Typ CPT6000 die entsprechende Steckverbindung des Kabels am Sensor gemäß der Orientierungsführung zusammenstecken. Beim Abstecken des Sensors nicht am Kabel ziehen, sondern nur an der Steckerhülse. Zum Anschluss an den CPH7650 muss das andere Kabelende ebenfalls gemäß der Orientierungsführung zusammengesteckt werden.
4.5.4.3 Mechanischer Anschluss des CPT6000 Referenz-Drucksensors am CPH7650
Zum mechanischen Anschluss den CPT6000 Referenz-Drucksensor mit dem Anschluss­gewinde voran in die Sensorhalterung am Gerät stecken. Anschließend muss der Sensor noch festgeschraubt werden. (Festziehen = im Uhrzeigersinn drehen; Lösen = gegen den Uhrzeigersinn drehen) Hierfür ist kein weiteres Werkzeug nötig.
77WIKA Betriebsanleitung, Typ CPH7650
5. Transport, Verpackung und Lagerung
5. Transport, Verpackung und Lagerung
5.1 Transport
Tragbaren Druckkalibrator auf eventuell vorhandene Transportschäden untersuchen. Offensichtliche Schäden unverzüglich mitteilen.
DE
5.2 Verpackung
Verpackung erst unmittelbar vor der Montage entfernen. Die Verpackung aufbewahren, denn diese bietet bei einem Transport einen optimalen Schutz (z. B. wechselnder Einbauort, Reparatursendung).
5.3 Lagerung
Zulässige Bedingungen am Lagerort:
Lagertemperatur: -20 ... +60 °C
Relative Feuchtigkeit: 0 ... 85 % r. F. (nicht kondensierend)
Folgende Einflüsse vermeiden:
Direktes Sonnenlicht oder Nähe zu heißen Gegenständen
Mechanische Vibration, mechanischer Schock (hartes Aufstellen)
Ruß, Dampf, Staub und korrosive Gase
Explosionsgefährdete Umgebung, entzündliche Atmosphären
Den tragbaren Druckkalibrator in der Originalverpackung an einem Ort lagern, der die oben gelisteten Bedingungen erfüllt. Wenn die Originalverpackung nicht vorhanden ist, dann das Gerät wie folgt verpacken und lagern:
1. Das Gerät in eine antistatische Plastikfolie einhüllen.
2. Das Gerät mit dem Dämmmaterial in der Verpackung platzieren.
3. Bei längerer Einlagerung (mehr als 30 Tage) einen Beutel mit Trocknungsmittel der Verpackung beilegen.
WARNUNG!
Vor der Einlagerung des Gerätes (nach Betrieb) alle anhaftenden Messstoff­reste entfernen. Dies ist besonders wichtig, wenn der Messstoff gesund­heitsgefährdend ist, wie z. B. ätzend, giftig, krebserregend, radioaktiv, usw.
78 WIKA Betriebsanleitung, Typ CPH7650
6. Inbetriebnahme, Betrieb
6. Inbetriebnahme, Betrieb
Durch das Betätigen einer beliebigen Taste wird der tragbare Druckkalibrator Typ CPH7650 eingeschaltet. Der Kalibrator benötigt einen Warmlauf von einigen Minuten (max. 5 Minuten), um die angegebene Genauigkeit zu erreichen. Große Schwankungen der Umgebungstemperatur können eine längere Warmlaufphase erfordern. Vor dem Start der Druckkalibrierung sollte die Druckanzeige des Kalibrators genullt werden.
Gerätemerkmale
Das Gerät verfügt über 3 Arbeitsmodi: MESSEN / KALIBRIEREN / SCHALTERTEST, die dem Anwender jeweils größtmöglichen Komfort entsprechend seiner Anwendung bieten. Zur Versorgung von Prüflingen und zum Einlesen deren Messsignale stehen elektrische Ein- und Ausgänge zur Verfügung. Im Arbeitsmodus MESSEN (mit Prüfling) und KALIBRIEREN werden die Messwerte des Referenz-Drucksensors und des Prüflings, sowie deren Abweichung in der aktuellen Druckeinheit und in % dargestellt. Hierdurch wird der Anwender direkt online informiert, ob der Prüfling die Klassengenauigkeit einhält oder nicht. Der Unterschied zwischen diesen beiden Modi ist, dass die Kalibrierdaten im KALIBRIEREN-Modus intern gespei­chert werden und später mittels Software (WIKA-Cal) in ein druckfähiges Zeugnis überführbar sind. Bezüglich des Datentransfer zu einem PC verfügt der Druckkalibrator Typ CPH7650 über eine USB-Schnittstelle, die via Menü auswählbar sind.
SETUP-Menü
Via SETUP-Taste gelangt man in das SETUP-Menü, indem der gewünschte Arbeits­modus (MESSEN / KALIBRIEREN / SCHALTERTEST) gewählt und konfiguriert, eine hinterlegte Funktion aufgerufen oder allgemeine Geräteeinstellungen wie z. B. die Menüsprache verändert werden kann.
DE
79WIKA Betriebsanleitung, Typ CPH7650
6. Inbetriebnahme, Betrieb
Prüfling 24V/50mA
..SCHALTER-TEST
..KALIBRIEREN
Prüflingsart
6.1 Menü-Struktur (Arbeitsmodi)
Mittels SETUP-Menü lässt sich der gewünschte Arbeitsmodus leicht einstellen (siehe nachfolgende Zeichnung).
DE
SELECTSELECT SELECT
..MESSEN
Einheit:
Klasse:
Messart: Medium: Versorgung:
4-20 mAPrüfling:
0.000MB-Anfang:
6.000MB-Ende: bar
0.50 %FS
rel
Gas
24V
Kalibrierung:
ID Nr: MS Nr:
Einheit: Klasse:
Messart: Medium:
Haltezeit: Soll: Ist:
Kalibrierprogramm
123456 456789
SELECT SELECTSELECT
4-20 mAPrüfling:
0.000MB-Anfang:
6.000MB-Ende:
%FS
0.00000
0.00000
0.50
Gas 24VVersorgung:
1
bar
rel
1Prüfpunkt: 0
optional
Arbeitsmodi Konfiguration SETUP-Menü
Wechsel der Prüflingsanzeige möglich (Druck ↔ elektrisches Signal) via
Einheit:
AusVersorgung:
bar
80 WIKA Betriebsanleitung, Typ CPH7650
6. Inbetriebnahme, Betrieb
6.2 Display-Darstellungen
6.2.1 Geräte-Statusmeldung kurz nach dem Einschalten des CPH7650
Direkt nach dem Einschalten des Gerätes erscheint für einen kurzen Moment folgende Statusmeldungen:
a b c
d
e
f
a) Die Spannungsversorgung von 24 V (verfügbar am oberen Ende des Gerätes) kann
während der Konfiguration eines jeden Arbeitsmodus ein- bzw. ausgeschaltet werden. Wird sie für eine Messung nicht benötigt, so sollte sie ausgeschaltet werden, um Energie zu sparen.
b) Aktuelle Akkukapazität (siehe Kapitel 6.4.5 „SETUP-Zusatzmenüpunkte: CPH-Einstel-
lungen“)
c) Höhendifferenz in mm
Im Menü „SETUP-\CPH-Einstellungen“ eingestellte Höhendifferenz zwischen Prüfling und CPT6000 Referenz-Drucksensor. Dieser Wert bewirkt eine automatische Korrek­turberechnung, um eine Druckdifferenz, basierend auf einer Mediumssäule, zu eliminieren. Dieser Wert muss für den nachfolgenden Messvorgang stimmen bzw. im Menü „SETUP-\CPH-Einstellungen“ entsprechend angepasst werden (siehe Kapitel
6.4.5 „SETUP-Zusatzmenüpunkte: CPH-Einstellungen“).
d) Temperatur in [°C]
Im Menü „SETUP-\CPH-Einstellungen“ eingestellte Temperatur (Umgebungstem­peratur). Dieser Wert kann im Menü „SETUP-\CPH-Einstellungen“ entsprechend angepasst werden (siehe Kapitel 6.4.5 „SETUP-Zusatzmenüpunkte: CPH-Einstellun­gen“).
e) Aktuelles Datum der integrierten Echtzeituhr
Im Menü „SETUP-\CPH-Einstellungen“ eingestelltes Datum der Echtzeituhr, das später mit auf dem Kalibrierzeugnis vermerkt wird. Dieser Wert muss für den nachfolgenden Messvorgang im Kalibriermodus stimmen bzw. im Menü „SETUP-\ CPH-Einstellungen“ entsprechend angepasst werden (siehe Kapitel 6.4.5 „SETUP­Zusatzmenüpunkte: CPH-Einstellungen“).
f) Kalibrierdatum der elektrischen Messeingänge des CPH7650 (Jahr/Monat/Tag)
Nach der Statusmeldung erscheint die Oberfläche des zuletzt eingestellten Arbeits­modus (siehe nachfolgendes Kapitel 6.2.3 „Displayinhalte der Arbeitsmodi“).
Info
CPH7650
CPH6000
Spannung 24V: EIN
Akkukapazität: 100%
Höhendifferenz: 0 mm
Temperatur: 23.0 °C
Akt-Dat.: 2011/05/10
Kal-Dat.: 2011/05/10
DE
81WIKA Betriebsanleitung, Typ CPH7650
6. Inbetriebnahme, Betrieb
6.2.2 Druckkalibrator ausschalten
Das Ausschalten erfolgt über den Menüpunkt „CPH ausschalten“ im ersten Untermenü. Hierzu einmalig die SETUP-Taste drücken, den Menüpunkt „CPH ausschalten“ auswäh­len und mit der SELECT-Taste bestätigen. CPH7650 ausschalten.
DE
Ausschalten
6.2.3 Displayinhalte der Arbeitsmodi Arbeitsmodus: MESSEN
Nach dem erstmaligen Einschalten des CPH7650 mit angeschlossenem CPT6000 Referenz-Drucksensor befindet sich das Gerät (nach einer kurzen Statusmeldung) im Modus: MESSEN (siehe nachfolgendes Bild)
Displayanzeige: MESSEN nur mit Referenz-Drucksensor (ohne Prüfling)
a b
a) Messbereich des Referenz-Drucksensors CPT6000 (der aktuell angeschlossen ist) b) Aktueller Messwert des Referenz-Drucksensors c) Druckeinheit (via Menü einstellbar)
82 WIKA Betriebsanleitung, Typ CPH7650
c
6. Inbetriebnahme, Betrieb
Im Modus MESSEN kann parallel zum Referenz-Druckmesswert auch ein Prüfling zur Anzeige gebracht werden (siehe nachfolgendes Bild). Bezüglich der Konfiguration siehe Kapitel 6.3.1 „Modus MESSEN“ und 6.3.2 „Modus MESSEN (mit Prüfling)“.
Displayanzeige: MESSEN mit Prüfling
MESSEN
Referenz
0.000 10.000
0.000
d
Prüfling
a b
4-20 mA
4.0160
13:28
bar
e
f
DE
c
P I/U
umschaltbar
Abw.:
0.006
0.10
bar
%FS
Drucksignal (Prüfling) Elektrisches Signal
(Prüfling)
a) Messbereich des Prüflings b) Aktueller Messwert des Prüflings c) Abweichung/Differenz zwischen Referenz und Prüfling in der aktuellen Druckeinheit
und in % von der Messspanne (% FS) bzw. % vom Messwert (% rd) d) Druckeinheit (des Prüflings) e) Original Ausgangsignal des Prüflings f) Aktueller Wert des Ausgangssignals des Prüflings
83WIKA Betriebsanleitung, Typ CPH7650
6. Inbetriebnahme, Betrieb
Arbeitsmodus: KALIBRIEREN
Im Modus KALIBRIEREN sind die Daten über der gezogenen Trennlinie dieselben wie im Modus „MESSEN mit Prüfling“.
a
DE
b
c
d e
P I/U
umschaltbar
f
g
Drucksignal (Prüfling) Elektrisches Signal
a) Aktueller Messwert des CPT6000 Referenz-Drucksensors b) Aktueller Messwert des Prüflings c) Abweichung zwischen Prüfling und Referenz d) Sollwert der Kalibrierung e) Istwert der Kalibrierung f) P - 01: Prüfling Nr. 1
< 01 >: Prüfschritt Nr. 1
g) IDENT-Nummer des Prüflings h) Aktueller Wert des Ausgangssignals des Prüflings
h
(Prüfling)
84 WIKA Betriebsanleitung, Typ CPH7650
6. Inbetriebnahme, Betrieb
Arbeitsmodus: SCHALTERTEST
Im Modus SCHALTERTEST werden neben dem Referenz-Drucksensor (siehe Modus MESSEN) Angaben zum Status und zu den Schaltpunkten des Druckschalters angezeigt.
a
f
b
c d
e
a) Aktueller Messwert des CPT6000 Referenz-Drucksensors b) Aktueller Schaltzustand/Status des Druckschalters c) Schaltpunkt des Öffnens d) Schaltpunkt des Schließens e) Hysterese/Abstand zwischen Öffnen und Schließen des Schalters f) Druckeinheit (via Menü einstellbar)
6.2.4 Inhalt des SETUP-Menüs
DE
a b c
d e
f
g h
i j
k
85WIKA Betriebsanleitung, Typ CPH7650
6. Inbetriebnahme, Betrieb
a) Arbeitsmodus MESSEN
Zum Messen von Arbeits- bzw. Prozessdrücken
Für Vergleichsmessungen bzw. Kalibrierung (ohne Speicherung der Daten) von mechanischen* und elektrischen Druckmessgeräten (Versorgung und Anzeige des Prüflings durch das CPH7650)
DE
→ Weitere Informationen siehe Kapitel 6.3.1 „Modus MESSEN“ und 6.3.2 „Modus
MESSEN (mit Prüfling)“
b) Arbeitsmodus KALIBRIEREN
Zur Kalibrierung von mechanischen 1) und elektrischen Druckmessgeräten vor Ort (ohne PC). Hierbei werden die Datensätze von bis zu 16 Prüflingen mit jeweils bis zu 32 Prüfschritten inkl. Datum und Zeit im CPH7650 gespeichert.
1) Bei mechanischen Zeigerinstrumenten muss der Messwert des Prüflings via numerischem Eingabeblock eingegeben werden.
→ Weitere Informationen siehe Kapitel 6.3.3 „Modus KALIBRIEREN“
c) Arbeitsmodus SCHALTERTEST
Für die komfortable Überprüfung von Druckschaltern inkl. automatischer Berech­nung der Schalthysterese
→ Weitere Informationen siehe Kapitel 6.3.7 „Modus SCHALTERTEST“
d) Arbeitsfunktionen
Tara: Offsetkorrektur des Referenzdruckmesswertes
Min/Max: Minimum/Maximum Speicher
Alarm: Min/Max-Alarm (visuell und akustisch)
Filter: Beruhigung/Glättung des Referenzsensorsignals
→ Weitere Informationen siehe Kapitel 6.4.1 „SETUP-Zusatzmenüpunkte: Funktionen“
e) Allgemeine CPH7650 Gerätedaten
Kalibrierdaten der elektrischen Messeingänge
Firmware-Nummer
Seriennummer des Gerätes
→ Weitere Informationen siehe Kapitel 6.4.2 „SETUP-Zusatzmenüpunkte: CPH-Info“
86 WIKA Betriebsanleitung, Typ CPH7650
6. Inbetriebnahme, Betrieb
f) Daten des aktuell angeschlossenen Referenz-Drucksensors
Messbereich
Genauigkeitsklasse
Druckart des Sensors
Info falls Referenzsensor überlastet wurde
Kalibrierdaten des Referenzsensors
→ Weitere Informationen siehe Kapitel 6.4.3 „SETUP-Zusatzmenüpunkte: Referenz-
Sensor“
g) Referenz-Sensor-Liste
Liste der gespeicherten Referenz-Sensoren, die angeschlossen werden können und kalibriert sind.
→ Weitere Informationen siehe Kapitel 6.4.4 „SETUP-Zusatzmenüpunkte: Referenz-
Sensor-Liste“
h) CPH-Einstellungen
Info: zur Akkukapazität
Einstellmöglichkeiten von: Menüsprache, Systemzeit/Systemuhr, Display­Helligkeit, Powersave-Funktion (automatischer Energiesparmodus siehe Kapitel
6.4.5 „SETUP-Zusatzmenüpunkte: CPH-Einstellungen“)
Eingabemöglichkeiten:
- Umgebungstemperatur während der Kalibrierung
- Vorhandene Höhendifferenz zwischen Referenz-Drucksensor und Prüfling (siehe Kapitel 4.5.1 „Anforderung an Prüfaufbauten mit dem CPH7650“).
DE
→ Weitere Informationen siehe Kapitel 6.4.5 „SETUP-Zusatzmenüpunkte: CPH-Einstel-
lungen“
i) Schnittstelle
USB-Schnittstelle inkl. Einstellung der Baudrate
→ Weitere Informationen siehe Kapitel 6.4.6 „SETUP-Zusatzmenüpunkte: Schnittstelle“
j) CLEAR KalProg
Löschen aller gespeicherten Kalibrierdaten (Freigabe und Zurücksetzen aller Speicherplätze)
→ Weitere Informationen siehe Kapitel 6.4.7 „SETUP-Zusatzmenüpunkte: CLEAR
KalProg“
k) Ausschalten des Druckkalibrators Typ CPH7650
→ Weitere Informationen siehe Kapitel 6.2.2 „Druckkalibrator ausschalten“
87WIKA Betriebsanleitung, Typ CPH7650
6. Inbetriebnahme, Betrieb
6.3 Arbeitsmodi
6.3.1 Modus MESSEN
1. SETUP-Menü aufrufen
DE
Drücke (SETUP-Taste)
Auswahl (Menüpunkt)
2. MESSEN vorbereiten
a
b c
d
e
f
g
3. Modus: MESSEN
Druckeinheit
(via Menü
einstellbar)
..MESSEN
Prüfling:
MB-Anfang: MB-Ende: Einheit:
Klasse:
Messart: Medium: Versorgung:
Prüflingsart
- - - - l
0.000
0.000 bar
0.00
%FS
Gas
AUS
SELECT
Bestätigung (der Auswahl)
Auswahl (Menüpunkt); Konfiguration,
rel
siehe nächste Seite
SELECT
Bestätigung (der Eingabe)
88 WIKA Betriebsanleitung, Typ CPH7650
6. Inbetriebnahme, Betrieb
Um das Gerät in den Modus MESSEN zu überführen, ist wie auf der vorhergehenden Seite zu verfahren.
Nachfolgend nähere Erklärung zu Punkt 2 “MESSEN vorbereiten“: a) Prüflingsart bzw. Prüflingsmesssignal: [----] für Messen ohne Prüfling b) Messbereichsanfangswert und -endwert des aktuell zu kalibrierenden Prüflings c) Einheit und Auflösung (Untermenü)
Auswahl und Bestätigung (Standard-Einheit) via
Kundenspezifische Einheit; bez. auf bar (Eingabe via numerischem Eingabeblock)
Auflösung der Anzeige im Arbeitsmodus via (zurück mit )
d) Messunsicherheit des Prüflings in % FS (bzw. auf die Spanne) oder % rd (bzw. auf den
Messwert) e) Messart des Prüflings (relativ oder absolut) f) Prüfmedium (Pneumatik → Gas oder Hydraulik → Öl) g) Spannungsversorgung für Prüfling (an/aus)
Falls keine Fremdversorgung für den Prüfling benötigt wird, sollte „AUS“ gewählt
werden, um Energie zu sparen.
DE
XXX
◁▷ ◀▶
0.00 Parameter-Eingabe mittels numerischem Eingabeblock
Kurzinfo:
ENTER
Aktuelle Cursorposition; Wechsel via
Parameter-Auswahl aus Liste bzw. Menü via
Eingabe-Quittierung
CLEAR
Löschen der Eingabe
89WIKA Betriebsanleitung, Typ CPH7650
6. Inbetriebnahme, Betrieb
..MESSEN
Prüfling:
MB-Anfang: MB-Ende:
Klasse:
Messart: Medium: Versorgung:
Einheit:
Prüflingsart
0.000
6.000
0.50
rel Gas 24V
bar
%FS
4-20 mA
6.3.2 Modus MESSEN (mit Prüfling)
1. SETUP-Menü aufrufen
DE
2. MESSEN vorbereiten
Drücke (SETUP-Taste)
Auswahl (Menüpunkt)
SELECT
Bestätigung (der Auswahl)
a
b c
d
e f g
Auswahl (Menüpunkt); Konfiguration, siehe nächste Seite
3. Modus: MESSEN
90 WIKA Betriebsanleitung, Typ CPH7650
Abweichung
SELECT
Bestätigung (der Eingabe)
MESSEN
Referenz
0.000 10.000
0.000
Prüfling
4-20 mA
13:28
bar
4.0160
P I/U umschaltbar
Drucksignal (Prüfling) Elektrisches Signal (Prüfling)
Abw.:
0.006
0.10
bar
%FS
6. Inbetriebnahme, Betrieb
Soll das Gerät in den Modus MESSEN (mit Prüfling = Anzeige des Prüflingssignals als elektrisches Signal oder als Druck) überführt werden, um eine Vergleichsmessung oder Kalibrierung ohne Messwertspeicherung durchzuführen, so ist wie auf der vorhergehen­den Seite zu verfahren.
Nachfolgend nähere Erklärung zu Punkt 2 „MESSEN vorbereiten“ a) Prüflingsart bzw. Prüflingsmesssignal
0 ... 20 mA / 4 ... 20 mA / 0 ... 1 V / 0 ... 2 V / 0 ... 5 V / 0 ... 10 V / bzw. mechanisch für Zeigermanometer
Wird eine Vergleichsmessung mit einem mechanischem Zeigerinstrument (Prüfling) durchgeführt, so ist der aktuelle Messwert des Zeigerinstrumentes via numerischem Eingabeblock einzugeben und mit der ENTER-Taste zu bestätigen.
b) Messbereichsanfangswert und -endwert des aktuell zu kalibrierenden Prüflings c) Einheit und Auflösung (Untermenü)
Auswahl und Bestätigung (Standard-Einheit) via
Kundenspezifische Einheit; bez. auf bar (Eingabe via numerischem Eingabeblock)
DE
Auflösung der Anzeige im Arbeitsmodus via (zurück mit )
d) Messunsicherheit des Prüflings in % FS (bzw. auf die Spanne) oder % rd (bzw. auf den
Messwert) e) Messart des Prüflings (relativ oder absolut) f) Prüfmedium (Pneumatik → Gas oder Hydraulik → Öl) g) Spannungsversorgung für Prüfling (an/aus)
Falls keine Fremdversorgung für den Prüfling benötigt wird, sollte „AUS“ gewählt
werden, um Energie zu sparen.
XXX
◁▷ ◀▶
0.00 Parameter-Eingabe mittels numerischem Eingabeblock
Kurzinfo:
ENTER
Aktuelle Cursorposition; Wechsel via
Parameter-Auswahl aus Liste bzw. Menü via
Eingabe-Quittierung
CLEAR
Löschen der Eingabe
91WIKA Betriebsanleitung, Typ CPH7650
6. Inbetriebnahme, Betrieb
..KALIBRIEREN
Kalibrierung: Prüfling: ID Nr: MS Nr: MB-Anfang: MB-Ende:
Klasse:
Messart: Medium: Versorgung: Prüfpunkt: Haltezeit: Soll: Ist:
Einheit:
Kalibrierprogramm
4-20 mA
0.000
6.000
0.50
rel Gas 24V
1 0
0.00000
0.00000
bar
%FS
1
123456 456789
6.3.3 Modus KALIBRIEREN
1. SETUP-Menü aufrufen
DE
2. KALIBRIEREN vorbereiten
Drücke (SETUP-Taste)
Auswahl (Menüpunkt)
SELECT
Bestätigung (der Auswahl)
a b c d
e f g h
i j k
l
m
n
Auswahl (Menüpunkt); Konfiguration, siehe nächste Seite
3. Modus: Kalibrieren
92 WIKA Betriebsanleitung, Typ CPH7650
Abweichung
SELECT
Bestätigung (der Eingabe)
KALIBRIEREN
Referenz
0.000 10.000
0.000
Prüfling
4-20 mA
13:41
bar
4.011
Abw.:
P I/U umschaltbar
0.004
0.07
0.000 bar
Soll:
0.000 bar
Ist : P-01:
123456
ID Nr:
01
Drucksignal (Prüfling) Elektrisches Signal (Prüfling)
bar
%FS
6. Inbetriebnahme, Betrieb
Um das Gerät in den Modus KALIBIEREN zu überführen, ist wie auf der vorhergehenden Seite zu verfahren.
Nachfolgend nähere Erklärung zu Punkt 2 „KALIBRIEREN vorbereiten“ a) Nummer der Kalibrierung bzw. des Prüflings (bis zu 16 Kalibrierungen mit jeweils bis
zu 32 Prüfschritten können vordefiniert bzw. gespeichert werden)
b) Prüflingsart bzw. Prüflingsmesssignal
0 ... 20 mA / 4 ... 20 mA / 0 ... 1 V / 0 ... 2 V / 0 ... 5 V / 0 ... 10 V / bzw. mechanisch für
Zeigermanometer c) IDENT-Nummer des Prüflings d) Messstellennummer des Prüflings e) Messbereichsanfangswert und -endwert des aktuell zu kalibrierenden Prüflings f) Einheit und Auflösung (Untermenü)
Auswahl und Bestätigung (Standard-Einheit) via
Kundenspezifische Einheit; bez. auf bar (Eingabe via numerischem Eingabeblock)
Auflösung der Anzeige im Arbeitsmodus via (zurück mit )
DE
g) Messunsicherheit des Prüflings in % FS (bzw. auf die Spanne) oder % rd (bzw. auf den
Messwert) h) Messart des Prüflings (relativ oder absolut) i) Prüfmedium (Pneumatik → Gas oder Hydraulik → Öl) j) Spannungsversorgung für Prüfling (an/aus) [Falls keine Fremdversorgung für den
Prüfling benötigt wird, sollte „AUS“ gewählt werden, um Energie zu sparen] k) Nummer des Prüfpunktes x l) Optionale Wartezeit [sek] (siehe Kapitel 6.3.4 „Modus KALIBRIEREN (die Prüfpunkte
einer Kalibrierung vorbereiten)“) m) Wert des Prüfpunktes x (Eingabe via numerischem Eingabeblock)
(Prüfpunkt x+1 bzw. x-1 via
aufrufbar)
n) Wahrer Wert des Prüflings (wird bei der Kalibrierung gespeichert)
93WIKA Betriebsanleitung, Typ CPH7650
6. Inbetriebnahme, Betrieb
6.3.4 Modus KALIBRIEREN (die Prüfpunkte einer Kalibrierung vorbereiten)
1. Prüfpunkt
(definieren)
Kalibrierung/Prüfling Nr. 1
DE
Auswahl vom Menüpunkt: „Sollwert“
Nr. des Prüfpunktes Sollwert des Prüfpunktes
Eingabe des Prüfpunktes (z. B.
1. Prüfpunkt
(definiert)
ENTER
0 bar) via numerischem Einga­beblock und Bestätigung mit
Prüfpunkt Nr. 1 = 0 bar
2. Prüfpunkt (definieren)
← 2. Prüfpunkt aufrufen
(mit
zurück zum vorherge-
henden Prüfpunkt)
x. Prüfpunkt (definieren)
94 WIKA Betriebsanleitung, Typ CPH7650
6. Inbetriebnahme, Betrieb
Mit diesem Beispiel wird erklärt wie die einzelnen Prüfpunkte/Druckstufen vor einer Kalib­rierung definiert werden. Es können bis zu 16 Kalibrierungen mit bis zu 32 Prüfschritten vorbereitet werden.
Aufruf des Menüpunktes Via
6.3.3 „Modus KALIBRIEREN“) Die gewünschten Prüfpunkte eingeben, wie auf der vorhergehenden Seite beschrieben.
Bei der Kalibrierung von Druckmessgeräten mit elektrischem Ausgangssignal (Druck­messumformer/Transmitter) wird nach Anzeige der Referenz kalibriert, d. h. der Druck ist immer so einzustellen, dass der Referenzwert exakt dem Sollwert entspricht. Da unter Umständen ein exaktes Einstellen des Druckes nicht immer möglich ist, wird neben dem Istwert (Prüflingswert) und dem Sollwert (Referenzwert) auch der wahre Referenzwert gespeichert. Mittels der Software WIKA-Cal kann dies im Kalibrierzeugnis aufgelistet werden.
und Auswahl des Menüpunktes: KALBRIEREN (bzw. siehe Kapitel
Wird nach DKD/DAkkS-Richtlinie kalibriert, so dürfen die Messwerte des jeweils nächsten Prüfpunktes erst nach einer bestimmten Zeit (bestehend aus Laständerungszeit und Beharrungszeit) von z. B. 30 Sekunden übernommen/gespeichert werden (siehe Abbildung A: Kalibrierzyklus nach DKD/DAkkS-Richtlinie 6 - 1 für Messunsicherheit > 0,6 % der Messspanne).
DE
M1
M2
Abbildung A
2 Minuten
30 s 30 s
Bei Federmano­metern: 5 Minuten
Ablesungen
Z
95WIKA Betriebsanleitung, Typ CPH7650
6. Inbetriebnahme, Betrieb
DE
optionale Wartezeit [sek] Eingabe via numerischem Eingabeblock und Bestätigung mit
Durch die Eingabe einer Wartezeit wird die Übernahme/Speicherung des Prüfpunktes für diese Dauer gesperrt. (Im obigen Beispiel müssen nach der Speicherung des ersten Prüfpunktes 30 Sekunden vergehen bis der zweite Prüfpunkt gespeichert werden kann.)
Sollten alte Prüfpunkte gelöscht bzw. zurückgesetzt werden, da die neue Kalibrierung weniger Prüfpunkte als die vorherige Prüfreihe besitzt, ist einfach die CLEAR-Taste zu drücken. Hierdurch wird der aktuelle und alle nachfolgenden Prüfpunkte gelöscht bzw. zurückgesetzt. (Dieser Vorgang kann einige Sekunden dauern.)
Sollen alle gespeicherten Kalibrierdaten sämtlicher Kalibrierungen auf einmal gelöscht werden siehe Kapitel 6.4.7 „SETUP-Zusatzmenüpunkte: CLEAR KalProg“.
ENTER
96 WIKA Betriebsanleitung, Typ CPH7650
6. Inbetriebnahme, Betrieb
6.3.5 Modus KALIBRIEREN (die Kalibrierung eines Druckmessumformers)
1. Prüfpunkt (z. B. 0 bar)
Abweichung
Vorgegebener Sollwert nach
Nr. Prüfling und
Nr. Prüfpunkt
2. Prüfpunkt (z. B. 1 bar)
ENTER
Anzeige der Referenz generie­ren (drucklosen Zustand herstel­len/Atmosphäre) und mit
Messwerte des Prüfpunktes speichern
DE
Abweichung
Nr. Prüfling und
Nr. Prüfpunkt
2. Prüfpunkt
Abweichung
Nr. Prüfling und
Nr. Prüfpunkt
x. Prüfpunkt
ENTER
Vorgegebener Sollwert mittels Druckerzeugung nach Anzeige der Referenz generieren
Messwerte des Prüfpunktes speichern
(mit gehenden Prüfpunkt)
zurück zum vorher-
97WIKA Betriebsanleitung, Typ CPH7650
6. Inbetriebnahme, Betrieb
6.3.6 Modus KALIBRIEREN (die Kalibrierung eines Manometers)
1. Prüfpunkt (z. B. 0 bar)
DE
Abweichung
Nr. Prüfling und
Nr. Prüfpunkt
2. Prüfpunkt (z. B. 1 bar)
ENTER
Vorgegebener Sollwert mittels Druckerzeugung nach Anzeige des Prüflings generieren
Bei Sollwert = 0 ist der Kalibrier­aufbau in drucklosen Zustand zu bringen/Atmosphäre (Prüfling muss 0 bar anzeigen; ggf. Null­punkteinstellung) und mit
Messwerte des Prüfpunktes speichern
Abweichung
Vorgegebener Sollwert mittels
Nr. Prüfling und
Nr. Prüfpunkt
Druckerzeugung nach Anzeige des Prüflings generieren
2. Prüfpunkt
Abweichung
Nr. Prüfling und
Nr. Prüfpunkt
ENTER
x. Prüfpunkt
98 WIKA Betriebsanleitung, Typ CPH7650
Messwerte des Prüfpunktes speichern
(mit gehenden Prüfpunkt)
zurück zum vorher-
6. Inbetriebnahme, Betrieb
..SCHALTER-TEST
Prüfling 24V/50mA
6.3.7 Modus SCHALTERTEST
1. SETUP-MENÜ aufrufen
2. SCHALTERTEST
vorbereiten
a
b
Versorgung:
Einheit:
Aus
bar
SELECT
Drücke (SETUP-Taste)
DE
Auswahl (Menüpunkt)
Bestätigung (der Auswahl)
Auswahl (Menüpunkt); Konfiguration, siehe nächste Seite
3. Modus: SCHALTERTEST
Aktueller
Schaltzustand
Schaltpunkte
Hysterese
SELECT
Bestätigung (der Eingabe)
Vor dem Druckschaltertest Nach dem Druckschaltertest
99WIKA Betriebsanleitung, Typ CPH7650
6. Inbetriebnahme, Betrieb
Um das Gerät in den Modus SCHALTERTEST zu überführen, ist wie auf der vorherge- henden Seite zu verfahren.
Der Schaltertest ist nicht für elektronische Schalter, wie z. B.: PNP- oder NPN-Schalter geeignet, sondern nur für mechanische, potentialfreie Schal­ter.
DE
Nachfolgend nähere Erklärung zu Punkt 2 „SCHALTERTEST vorbereiten“ a) Spannungsversorgung für Prüfling (an/aus) [Falls keine Fremdversorgung für den
Prüfling benötigt wird, sollte „AUS“ gewählt werden, um Energie zu sparen]
b) Einheit und Auflösung (Untermenü)
Auswahl und Bestätigung (Standard-Einheit) via
Kundenspezifische Einheit; bez. auf bar (Eingabe via numerischem Eingabeblock)
Auflösung der Anzeige im Arbeitsmodus via (zurück mit )
Die ermittelten Messewerte der beiden Schaltpunkte und der Hysterese können durch das Drücken der Ziffer: „0“ zurückgesetzt werden.
100 WIKA Betriebsanleitung, Typ CPH7650
Loading...