Declarations of conformity can be found online at www.wika.com.
4WIKA operating instructions, model CPH7650
14261987.01 03/2018 EN/DE
1. General information
1. General information
■
The portable pressure calibrator described in the operating instructions has been
manufactured using state-of-the-art technology. All components are subject to
stringent quality and environmental criteria during production. Our management
systems are certified to ISO 9001 and ISO 14001.
■
These operating instructions contain important information on handling the instrument.
Working safely requires that all safety instructions and work instructions are observed.
■
Observe the relevant local accident prevention regulations and general safety
regulations for the instrument's range of use.
■
The operating instructions are part of the product and must be kept in the immediate
vicinity of the instrument and readily accessible to skilled personnel at any time.
■
Skilled personnel must have carefully read and understood the operating instructions
prior to beginning any work.
■
The manufacturer's liability is void in the case of any damage caused by using the
product contrary to its intended use, non-compliance with these operating instructions,
assignment of insufficiently qualified skilled personnel or unauthorised modifications to
the instrument.
EN
■
The general terms and conditions contained in the sales documentation shall apply.
■
Subject to technical modifications.
■
Factory calibrations / DKD/DAkkS calibrations are carried out in accordance with
international standards.
... indicates a potentially dangerous situation that can result in serious injury
or death, if not avoided.
CAUTION!
... indicates a potentially dangerous situation that can result in light injuries
or damage to property or the environment, if not avoided.
Information
... points out useful tips, recommendations and information for efficient and
trouble-free operation.
DANGER!
... identifies hazards caused by electrical power. Should the safety
instructions not be observed, there is a risk of serious or fatal injury.
2. Safety
WARNING!
Before installation, commissioning and operation, ensure that the
appropriate reference pressure sensor has been selected in terms of
measuring range, design and specific measuring conditions.
Non-observance can result in serious injury and/or damage to the
equipment.
Further important safety instructions can be found in the individual chapters
of these operating instructions.
6WIKA operating instructions, model CPH7650
14261987.01 03/2018 EN/DE
2. Safety
2.1 Intended use
This portable pressure calibrator serves as a calibration instrument for the widest variety
of pressure measuring instruments and has been designed for mobile use as well as for
stationary workshop and laboratory testing. Through the combination of the integrated
electrical pump and the electrical modules, in addition to the traditional measurement of
current and voltage signals it also enables you to supply transmitters or sensors with a
max. of 30 mA (voltage (idling) = DC 24 V). The pressure calibrator can be used to carry
out and document a complete calibration process.
WARNING!
■
Only use model CPT6000 reference pressure sensors!
■
Using other sensors could damage both the pressure calibrator and the
reference pressure sensor.
■
To change the sensor, switch off the pressure calibrator and make sure
the system is vented. Before switching the instrument on, connect the
sensor, otherwise it may not be correctly identified by the instrument.
■
When the CPH7650 is switched on, the CPT6000 reference pressure
sensor must not be under pressure, but rather should be at atmospheric
pressure. For overpressure or gauge pressure sensors, there is a
pressure compensation vent in the top of the sensor under the plastic
fitting. This vent (with integrated diaphragm) must always remain clear!
The instrument has been designed and built solely for the intended use described here,
and may only be used accordingly.
EN
The technical specifications contained in these operating instructions must be observed.
Improper handling or operation of the instrument outside of its technical specifications
requires the instrument to be taken out of service immediately and inspected by an
authorised WIKA service engineer.
Handle electronic precision measuring instruments with the required care (protect from
humidity, impacts, strong magnetic fields, static electricity and extreme temperatures,
do not insert any objects into the instrument or its openings). Plugs and sockets must be
protected from contamination.
If the instrument is transported from a cold into a warm environment, the formation of
condensation may result in instrument malfunction. Before putting it back into operation,
wait for the instrument temperature and the room temperature to equalise.
The manufacturer shall not be liable for claims of any type based on operation contrary to
the intended use.
14261987.01 03/2018 EN/DE
7WIKA operating instructions, model CPH7650
2. Safety
2.2 Personnel qualification
WARNING!
EN
Skilled personnel
Skilled personnel are understood to be personnel who, based on their technical training,
knowledge of measurement and control technology and on their experience and
knowledge of country-specific regulations, current standards and directives, are capable
of carrying out the work described and independently recognising potential hazards.
Special operating conditions require further appropriate knowledge, e.g. of aggressive
media.
2.3 Special hazards
Risk of injury should qualification be insufficient!
Improper handling can result in considerable injury and damage to
equipment.
The activities described in these operating instructions may only be carried
out by skilled personnel who have the qualifications described below.
WARNING!
■
Do not apply a voltage greater than the specified voltage to the
instrument (see chapter 3 „Specifications“).
■
Do not apply any external pressure to the CPH7650.
■
Make sure that the test probes never contact a voltage source while the
test cables are connected to the current terminals.
■
Do not use the calibrator if it is damaged. Before using the instrument,
check that there are no cracks on the case or any missing plastic parts.
Pay particular attention to the insulation of the connectors.
■
Select the proper function and correct measuring range for the
measurement.
■
Observe the operating parameters in accordance with chapter
3 “Specifications”.
■
Always operate the pressure calibrator within the defined pressure range.
■
To ensure problem-free operation, only operate the instrument on battery
power. Only use the power supply unit for charging the instrument's
batteries.
■
Inspect the test cables for damaged insulation or exposed metal. Check
the continuity of the leads. Damaged test leads should be replaced before
using the instrument.
■
When using test probes, keep fingers away from the test probe contacts.
Keep your fingers behind the test probes' finger guards.
■
First connect the common lead, and then the live lead. When
disconnecting, remove the live test lead first.
14261987.01 03/2018 EN/DE
8WIKA operating instructions, model CPH7650
2. Safety
■
Do not use the instrument if it is not working properly. The instrument
protection might be compromised. If in doubt, have the instrument
checked.
■
Do not operate the instrument in areas with explosive gases, vapours or
dust.
■
When measuring pressure, make sure that the pressure calibrator has
been switched to a depressurised state before the reference sensor or
the test item is connected or disconnected.
■
Disconnect test leads before changing to another measurement or source
function.
■
The switching valve (+/- pressure) should only ever be actuated when in a
depressurised state.
WARNING!
■
To avoid false indications, which could lead to possible electric shock or
injuries, charge the rechargeable battery as soon as the battery indicator
appears.
■
In order to avoid any possible damage to the instrument or the test
equipment, use the correct leads, the correct function and the correct
range for the measuring application.
DANGER!
Danger to life caused by electric current
Upon contact with live parts, there is a direct danger to life.
■
Charging using a defective power supply unit (e.g. short circuit from the
mains voltage to the output voltage) can result in life-threatening voltages
at the instrument!
■
Only use the power supply unit permitted by WIKA for the model
CPH7650 portable pressure calibrator.
■
Only use a charger that is fully functional or undamaged.
EN
The safety of the operator may be endangered if, for example
■
there is visible damage to the instrument.
■
the instrument is no longer working as specified.
■
the instrument has been stored under unsuitable conditions for an extended period of
time.
If there is any doubt, please return the instrument to the manufacturer for repair or
maintenance.
14261987.01 03/2018 EN/DE
9WIKA operating instructions, model CPH7650
2. Safety
2.4 Use of Lithium-Ion rechargeable batteries
WARNING!
EN
Misusing Lithium-Ion batteries can lead to heating, explosion or ignition and
result in serious injury. Follow the safety instructions listed below:
■
Do not solder directly to the Lithium-Ion batteries.
■
Do not incinerate or heat the Lithium-Ion batteries.
■
The Lithium-Ion batteries must only ever be connected with the correct
polarity.
■
Never connect the positive terminal and the negative terminal of the
Lithium-Ion batteries to each other with any metallic object (such as wire).
■
Never carry or store the Lithium-Ion batteries together with necklaces,
hairpins, or other metal objects.
WARNING!
■
Lithium-Ion batteries should never be punctured with nails nor hit with a
hammer. In addition, Lithium-Ion batteries must never be trodden on or
exposed to other strong shocks or vibrations.
■
Lithium-Ion batteries must never come into contact with water or salt
water. Moreover, they must never get wet.
WARNING!
Never take the Lithium-Ion battery apart nor alter it in any way. It contains
safety and protection devices which, if damaged, may cause it to generate
heat, explode or ignite.
WARNING!
Never place the Lithium-Ion batteries close to fires, ovens or other
high-temperature locations. Never leave the Lithium-Ion batteries in direct
sunshine or use or store them inside cars in hot weather. Doing so may
cause the Lithium-Ion batteries to generate heat, explode or ignite. Using the
Lithium-Ion batteries in this manner may also result in a loss of performance
and a shortened service life.
Never fit the Lithium-Ion batteries into equipment designed to be
hermetically sealed. In some cases hydrogen or oxygen may be discharged
from the Lithium-Ion batteries, which may result in rupture, fire or explosion.
10WIKA operating instructions, model CPH7650
14261987.01 03/2018 EN/DE
2. Safety
WARNING!
The Lithium-Ion batteries must, without fail, no longer be used if, during
operation, charging or storing they give off an unusual smell, feel hot,
change colour, change shape, or appear abnormal in any other way. Contact
your sales partner if any of these problems are observed.
Never put the Lithium-Ion batteries in microwave ovens, high-pressure
containers nor on induction cookers.
Should the Lithium-Ion batteries ever leak and the fluid come into contact
with the eyes, do not under any circumstances rub the eyes. Rinse the eyes
thoroughly with water and seek immediate medical attention. If the eyes are
left untreated, damage to the eyes could occur.
CAUTION!
When the Lithium-Ion batteries wear out, insulate the terminals with
adhesive tape or similar materials before disposal.
WARNING!
Follow the instructions listed below for charging the Lithium-Ion batteries.
Failure to do so may cause the Lithium-Ion batteries to become hot, explode
or ignite and result in serious injury.
■
To charge the Lithium-Ion batteries, only ever use the specified WIKA
battery charger.
■
Never connect the Lithium-Ion batteries directly to a mains plug or to a
car's cigarette lighter.
■
Never leave the Lithium-Ion batteries in or near fire, nor in direct sunlight.
If the Lithium-Ion batteries become hot, the built-in safety device is
activated and overcharging prevented. Heating the Lithium-Ion batteries
can damage the safety device and can thus lead them to heat up further,
to cease to work or to ignite.
EN
WARNING!
Never continue to charge the Lithium-Ion batteries if they do not fully
recharge within the specified time. Doing so may cause the Lithium-Ion
batteries to become hot, explode or ignite.
14261987.01 03/2018 EN/DE
11WIKA operating instructions, model CPH7650
2. Safety
2.5 Labelling, safety marks
Product label
EN
6
5
4
3
2
1
Product name
2
Date of manufacture (month/year)
3
Serial number
4
Output signal, power supply
5
Input signal
6
Pressure range
1
Symbols
Before mounting and commissioning the instrument, ensure you read the
operating instructions!
This marking on the instruments indicates that they must not be disposed of
in domestic waste. The disposal is carried out by return to the manufacturer
or by the corresponding municipal authorities.
12WIKA operating instructions, model CPH7650
14261987.01 03/2018 EN/DE
3. Specifications
3. Specifications
Measuring ranges
Pressure
Gauge pressure-1 … +20 bar, -1 … +10 bar, 0 … 20 bar, 0 … 10 bar
Absolute pressure0 … 10 bar abs., 0 … 20 bar abs.
Overpressure limit
Accuracy0.025 % FS
Resolution5-digit
Temperature compensation15 … 35 °C (59 ... 95 °F)
Temperature coefficient0.002 % of span/°C outside of 15 ... 35 °C (59 ... 95 °F)
Current
Measuring range0 … 24 mA (max. load 1,000 Ω)
Resolution1 µA
Accuracy0.015 % of reading ±2 µA (simulation and measurement)
Voltage
Measuring rangeDC 0 … 30 V
Resolution1 mV
Accuracy0.015 % of reading ±2 mV (measurement)
1) Via exchangeable model CPT6000 reference pressure sensors
2) The electrical pump can generate -0.85 … +20 bar (-12 … 290 psi).
1)
2)
3 times
EN
Base instrument
Pressure supply-0.85 ... +20 bar (-12 ... +290 psi), via integrated electric pump
Pressure connection for test
item
Permissible mediaFor dry, clean and non-aggressive gases
Output
Voltage supplyDC 24 V
Power supply
Battery typeLithium-Ion rechargeable battery
Permissible ambient conditions
Operating temperature-10 … +50 °C (14 ... 122 °F)
Storage temperature-20 … +60 °C (-4 ... +140 °F)
Relative humidity35 ... 85 % r. h. (non-condensing)
Communication
InterfaceUSB via special interface cable
14261987.01 03/2018 EN/DE
G ½" female thread
13WIKA operating instructions, model CPH7650
3. Specifications
Base instrument
Case
EN
CaseNK-7TM resin
Front panelAluminium
Ingress protectionIP67 (case closed)
IP40 (case opened)
Dimensions387.4 x 304.8 x 177.8 mm (15.25 x 12 x 7 in)
Weightapprox. 7 kg (15.5 lbs.)
Certificates
Certificate
CalibrationStandard: 3.1 calibration certificate per DIN EN 10204
The CPH7650 is a compact, portable pressure calibrator, which has been designed
for mobile use as well as for stationary workshop and laboratory testing. An integrated
electrical pump enables you to generate pressures up to 20 bar (300 psi) and an
integrated electrical module also enables you to supply transmitters or sensors with
a current of max. 30 mA (voltage (idling) = DC 24 V), in addition to the traditional
measurement of current and voltage signals.
The pressure calibrator can be used to carry out and document a complete calibration
process. In addition to the features of displaying or measuring and calibration, there
is also a switch test. The use of the WIKA-Cal software also allows a complete
documentation.
14261987.01 03/2018 EN/DE
15WIKA operating instructions, model CPH7650
4. Design and function
4.2 Front
EN
1
Test item
2
Overview of the electrical connection
1
2
34
5
6
7
8
9
10111213
3
Electrical connection to the model CPT6000 reference sensor
4
Reference sensor CPT6000
5
Connecion of the power supply unit
6
WIKA-Cal connection
7
Fine adjustment / drain valve
8
Switch for pressure / vacuum
9
Pump speed controller
10
Operation
11
Electrical module
12
Display
13
Numeric keypad
16WIKA operating instructions, model CPH7650
14261987.01 03/2018 EN/DE
4. Design and function
4.3 Scope of delivery
■
Portable pressure calibrator model CPH7650
■
Operating instructions
■
Test cables
■
Battery charger
■
3.1 calibration certificate per DIN EN 10204
Cross-check scope of delivery with delivery note.
4.4 Voltage supply
Charging the battery
In order to avoid false measurements, charge the battery as soon as the battery indicator
is displayed. If the battery has run down too far, the instrument will switch itself off
automatically.
Only ever use the power supply unit permitted by WIKA for the model CPH7650 portable
pressure calibrator.
To avoid measurement uncertainties, only use the CPH7650 without the mains supply
connected. The full pump performance can only be guaranteed with the battery fully
charged and without the power supply unit connected.
The instrument should be fully charged before being put into operation.
The battery capacity status (charge state in %) is displayed shortly after the instrument is
switched on. When the power supply unit is connected, the rechargeable battery will be
charged, even if the CPH7650 is switched off.
EN
The typical charging time of the rechargeable battery is < 5 hours.
WARNING!
■
When the power supply unit is no longer being used, the mains plug
should be disconnected from the mains socket. Do not leave the
rechargeable battery connected to the power supply unit for longer than
one day, since overcharging can shorten its service life.
■
Should the rechargeable battery still not be fully charged after 24 hours,
contact the manufacturer. When not being used, a fully charged battery
will lose its charge over time.
■
Extreme temperatures have an adverse effect on battery charging. As
a result, the battery may first need to be either cooled or warmed, as
appropriate.
■
When the battery is almost completely discharged, the message “low
BAT” appears in the display. To avoid a data loss, the instrument must be
charged immediately
14261987.01 03/2018 EN/DE
17WIKA operating instructions, model CPH7650
4. Design and function
4.5 User interface
EN
6
5
1
SETUP menu
2
Selection and entry confirmation
3
Return to previous level
4
Clear entry
5
Input confirmation
6
Numeric keypad
Switch on via pressing any button
Switch off via menu item in main menu
1
2
3
4
18WIKA operating instructions, model CPH7650
14261987.01 03/2018 EN/DE
4. Design and function
4.5.1 Requirements for test assemblies with the CPH7650
■
Before starting any task, the instrument should be switched on briefly to determine
that there is sufficient charge in the battery (rechargeable battery capacity in %). The
battery capacity is indicated briefly in an instrument status message after switching
on (see chapter 6.2.1 “Instrument status messages shortly after switching on the
CPH7650”).
■
Initially, the test assembly must be physically assembled and, if necessary, connected
electrically (see chapter 4.5.4.1 “Connection of the model CPT6000 reference
pressure sensor”).
■
Before switching the CPH7650 on, ensure that the test assembly is not pressurised
(system is vented to atmosphere) and that the equipment is correctly assembled and
in the correct mounting position.
■
Only connect test and calibration installations once the system has been
depressurised!
Compensation for height differences
If a significant elevation difference exists between the CPT6000 reference pressure
sensor and the test item, then the pressure difference, based on a medium column, can
be compensated automatically via the menu (see chapter 6.4.5 “SETUP additional menu
items: CPH configuration”).
4.5.2 Important instrument settings for calibration using calibration mode
EN
Calibration date
The instrument has an integral real-time clock with date. The current date of a calibration
is stated later in the calibration certificate. Before starting a calibration, you must ensure
that the internal date of the CPH7650 is correct (see
chapter 6.4.5 “SETUP additional menu items: CPH configuration”).
Unit and resolution
After selecting one of the main menu items (e.g.: MEASURING, CALIBRATION or
SWITCH TEST) from the SETUP menu (press SETUP key), using the menu item “Unit”,
and its associated submenu respectively (move the cursor to “Unit” and press the right
or left arrow), you can set the unit and adjust its resolution (see chapter 6.3 “Operating
modes”).
14261987.01 03/2018 EN/DE
19WIKA operating instructions, model CPH7650
4. Design and function
Available units, including their conversion factors in relation to the unit bar
If the measured value shown on the CPH7650, with an overpressure sensor connected
and the test assembly vented to atmosphere, is not equal to zero, then by pressing the
CLEAR button twice (within five seconds), the zero point can be corrected (maximum
allowable correction value is twice the magnitude of the class accuracy).
9.80665E-01
Offset correction for absolute pressure sensors
For absolute pressure sensors, an offset correction can be made via the menu (see
chapter 6.4.3 “SETUP additional menu items: Reference sensor”).
4.5.4 Reference pressure sensor CPT6000
For the model CPH7650 pressure calibrator, there are many reference pressure sensors
to choose from, with accuracies of 0.025 % which can be interchanged quickly and
without tools. When the pressure calibrator is switched on, the reference pressure sensor
attached is recognised automatically, so that no further configuration of the sensor is
needed.
20WIKA operating instructions, model CPH7650
14261987.01 03/2018 EN/DE
4. Design and function
4.5.4.1 Connection of the model CPT6000 reference pressure sensor
WARNING!
■
Only use model CPT6000 reference pressure sensors!
■
Using other sensors could damage both the pressure calibrator and the
reference pressure sensor.
■
To change the sensor, switch off the pressure calibrator and make sure
the system is vented. Before switching the instrument on, connect the
sensor, otherwise it may not be correctly identified by the instrument.
■
When the CPH7650 is switched on, the CPT6000 reference pressure
sensor must not be under pressure, but rather should be at atmospheric
pressure. For overpressure or gauge pressure sensors, there is a
pressure compensation vent in the top of the sensor under the plastic
fitting. This vent (with integrated diaphragm) must always remain clear!
CAUTION!
Only ever use the original WIKA sensor connection cable in the operation of
CPT6000 reference pressure sensors.
WARNING!
The pump performance is independent of the pressure range of the selected
reference sensor. The operator must ensure that the CPT6000 reference
sensor is not over-pressured.
EN
4.5.4.2 Electrical connection of the CPT6000 reference pressure sensor to the
CPH7650
The calibrator and the reference pressure sensor are connected to each other electrically
using a separate connection cable. To electrically connect a model CPT6000 reference
pressure sensor, the corresponding cable connector must be plugged in at the sensor in
accordance with the orientation guide. To disconnect the sensor, do not pull on the cable,
but rather only on the connector sleeve.
To connect it to the CPH7650, the other end of the cable must also be plugged in in
accordance with the orientation guide.
4.5.4.3 Mechanical connection of the CPT6000 reference pressure sensor to the
CPH7650
To make the mechanical connection of the CPT6000 reference pressure sensor, it must
be placed, connection thread first, in the sensor bracket of the instrument. Then, the
sensor must be tightened again.
(Tighten = turn clockwise; release = turn anti-clockwise)
No further tools are required for this.
14261987.01 03/2018 EN/DE
21WIKA operating instructions, model CPH7650
5. Transport, packaging, storage
5. Transport, packaging and storage
5.1 Transport
Check the portable pressure calibrator for any damage that may have been caused by
EN
transport. Obvious damage must be reported immediately.
5.2 Packaging
Do not remove packaging until just before mounting.
Keep the packaging as it will provide optimum protection during transport (e.g. change in
installation site, sending for repair).
5.3 Storage
Permissible conditions at the place of storage:
■
Storage temperature: -20 ... +60 °C
■
Relative humidity: 0 ... 85 % r. h. (non-condensing)
Avoid exposure to the following factors:
■
Direct sunlight or proximity to hot objects
■
Mechanical vibration, mechanical shock (putting it down hard)
■
Soot, vapour, dust and corrosive gases
■
Hazardous environments, flammable atmospheres
Store the portable pressure calibrator in its original packaging in a location that fulfils
the conditions listed above. If the original packaging is not available, pack and store the
instrument as described below:
1. Wrap the instrument in an antistatic plastic film.
2. Place the instrument, along with shock-absorbent material, in the packaging.
3. If stored for a prolonged period of time (more than 30 days), place a bag containing a
desiccant inside the packaging.
WARNING!
Before storing the instrument (following operation), remove any residual
media. This is of particular importance if the medium is hazardous to health,
e.g. caustic, toxic, carcinogenic, radioactive, etc.
22WIKA operating instructions, model CPH7650
14261987.01 03/2018 EN/DE
6. Commissioning, operation
6. Commissioning, operation
By pressing any key, the model CPH7650 portable pressure calibrator will be switched
on. The calibrator requires a warm-up for a few minutes (max. 5 minutes) to reach its
specified accuracy. Large changes in ambient temperature may make a longer warm-up
period necessary. The pressure display of the calibrator should be zeroed before starting
the pressure calibration.
Instrument features
The instrument features 3 operating modes: MEASURING / CALIBRATION / SWITCH
TEST, which offer the user maximum convenience according to its application. For
supplying the test items and to read their measuring signals, there are electrical inputs
and outputs available.
In the MEASURING (with test item) and CALIBRATION operating modes, the measured
values of both the reference pressure sensor and the test item, as well as their deviation,
are displayed in both current pressure units and in %. In this way the operator is
immediately informed online whether the test item meets the class accuracy or not.
The difference between these two modes is that the calibration data in CALIBRATION
mode are stored internally and can later be transferred onto printable certificates through
software (WIKA-Cal).
With respect to the transfer of data to a PC, the model CPH7650 pressure calibrator has a
USB interface, selectable via menu.
SETUP menu
Using the SETUP key, one can access the SETUP menu, where the required operating
mode (MEASURING / CALIBRATION / SWITCH TEST) can be selected and configured,
a stored function can be recalled or a general instrument setting (such as the menu
language) can be changed.
EN
14261987.01 03/2018 EN/DE
23WIKA operating instructions, model CPH7650
6. Commissioning, operation
6.1 Menu structure (operating modes)
Through the SETUP menu, the required operating mode can be easily selected (see
drawing below).
It is possible to change the test item display (pressure ↔ electrical signal) via
24WIKA operating instructions, model CPH7650
14261987.01 03/2018 EN/DE
6. Commissioning, operation
6.2 Explanation of the display
6.2.1 Instrument status messages shortly after switching on the CPH7650
Directly after the instrument is switched on, the following status messages are displayed
briefly:
CPH7650
a
b
c
d
e
f
a) The voltage supply of 24 V (available at the upper end of instrument) can, during the
configuration of each operating mode, be switched on or off. If it is not needed for a
measurement, then it should be switched off, in order to conserve energy.
b) Current battery capacity (see chapter 6.4.5 “SETUP additional menu items: CPH
configuration”)
c) Height difference in mm
In the “SETUP\CPH configuration” menu, set the height difference between the test
item and the CPT6000 reference pressure sensor. This value influences an automatic
correction calculation in order to eliminate any pressure difference based on a
medium column. This value must be correct for the following measuring procedure
and/or be adjusted accordingly in the “SETUP/CPH configuration” menu (see chapter
6.4.5 “SETUP additional menu items: CPH configuration”).
d) Temperature in [°C]
In the “SETUP\CPH configuration” menu, the temperature (ambient temperature) is
entered. This value can be adjusted accordingly in the “SETUP\CPH configuration”
menu (see chapter 6.4.5 “SETUP additional menu items: CPH configuration”).
e) Current date of the integrated real-time clock
In the “SETUP\CPH configuration” menu, the date of the real-time clock is set, which is
later marked on the calibration certificate. This value must be correct for the following
measuring procedure in calibration mode and/or be adjusted accordingly in the
“SETUP/CPH configuration” menu (see chapter 6.4.5 “SETUP additional menu items:
CPH configuration”).
f) Calibration date for the electrical measuring inputs of the CPH7650 (Year/Month/Day)
Following the status messages, the display returns to the screen for the last-selected
operating mode (see following chapter 6.2.3 “Display contents of the operating
modes”).
EN
14261987.01 03/2018 EN/DE
25WIKA operating instructions, model CPH7650
6. Commissioning, operation
6.2.2 Switching the pressure calibrator off
The instrument is switched off using the “Switch off CPH” menu item in the first submenu.
For this, press the SETUP key, select the “Switch off CPH” menu item and confirm using
the SELECT button. Switch off the CPH7650.
When a CPH7650 with a CPT6000 reference pressure sensor connected to it is first
switched on, the instrument (after displaying a brief status message) switches to
MEASURING mode (see following figure)
Display: MEASURING with reference pressure sensor only (without test item)
a
b
c
a) Measuring range of the CPT6000 reference pressure sensor (which is currently
connected)
b) Current measured value of the reference pressure sensor
c) Pressure unit (adjustable via menu)
14261987.01 03/2018 EN/DE
26WIKA operating instructions, model CPH7650
6. Commissioning, operation
In MEASURING mode, at the same time as the reference-pressure value, a test item can
also be displayed on the screen (see following figure). With respect to the configuration,
see chapters 6.3.1 “MEASURING mode” and 6.3.2 “MEASURING mode (with test item)”.
Display: MEASURING with test item
d
a
b
c
PI/U
switchable
e
f
Pressure signal (test item)Electrical signal (test item)
a) Measuring range of the test item
b) Current measured value of the test item
c) Deviation/difference between reference and test item in the current pressure units and
in % of the measuring span (% FS) or % of reading (% rd)
d) Pressure unit (of the test item)
e) Original output signal of the test item
f) Current value of the output signal of the test item
EN
14261987.01 03/2018 EN/DE
27WIKA operating instructions, model CPH7650
6. Commissioning, operation
Operating mode: CALIBRATION
In the CALIBRATION mode the data shown above the dashed dividing line is the same
as in the “MEASURING with test item” mode.
EN
a
b
c
d
e
PI/U
switchable
f
g
Pressure signal (test item)Electrical signal (test item)
a) Current measured value of the CPT6000 reference pressure sensor
b) Current measured value of the test item
c) Deviation between the test item and reference
d) Set point of the calibration
e) Actual value of the calibration
f) P - 01: Test item No. 1
<01>: Test step No. 1
g) IDENT number of the test item
h) Current value of the output signal of the test item
h
28WIKA operating instructions, model CPH7650
14261987.01 03/2018 EN/DE
6. Commissioning, operation
Operating mode: SWITCH TEST
In the SWITCH TEST mode, along with the reference pressure sensor data (see
MEASURING mode), the status and switch points of the pressure switch are also
displayed.
a
f
b
c
d
e
a) Current measured value of the CPT6000 reference pressure sensor
b) Current switching status/status of the pressure switch
c) Opening switch point
d) Closing switch point
e) Hysteresis/separation between opening and closing of the switch
f) Pressure unit (adjustable via menu)
For comparative measurements and/or calibrations (without data recording) of
EN
mechanical* and electrical pressure measuring instruments (supply and display of
the test item through the CPH7650)
→ For further information, see chapter 6.3.1 “MEASURING mode” and
6.3.2 “MEASURING mode (with test item)”
b) Operating mode CALIBRATION
■
For on-site calibration of mechanical 1) and electrical pressure measuring
instruments (without PC). In this case the data sets (for up to 16 test items, each
with up to 32 test steps including date and time) are recorded within the CPH7650.
1) For mechanical dial instruments, the test item's measured value must be entered via the numeric keypad.
→ For more information please refer to chapter 6.3.3 “Mode CALIBRATION”
c) Operating mode SWITCH TEST
■
For the easy checking of pressure switches, including automatic calculation of the
switch hysteresis.
→ For more information please refer to chapter 6.3.7 “SWITCH TEST mode”
d) Operating functions
■
Tare: Offset correction of the reference pressure value
■
Min/Max: Minimum/Maximum memory
■
Alarm: Min/Max alarm (visual and audible)
■
Filter: Damping/smoothing of the reference sensor signal
→ For further information see chapter 6.4.1 „SETUP additional menu items: Functions“
e) General CPH7650 instrument data
■
Calibration data for the electrical measuring inputs
■
Firmware number
■
Serial number of the instrument
→ For more information please refer to chapter 6.4.2 “SETUP additional menu items:
CPH info”
30WIKA operating instructions, model CPH7650
14261987.01 03/2018 EN/DE
Loading...
+ 86 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.