WIKA CPH7650 Specifications

Page 1
Portable pressure calibrator Model CPH7650
Applications
Calibration service companies and service industry
Measurement and control laboratories
Quality assurance
Special features
Measurement and electrical pressure generation at the touch of a button with integrated pump, from
-0.85 ... +20 bar (-12 ... 290 psi)
Accuracy: 0.025 % FS (incl. calibration certicate)
Generation/measurement of 4 ... 20 mA, and DC 24 V voltage supply to power transmitters
Robust case design, IP67
Calibration technology
WIKA data sheet CT 17.02
Description
General information
The portable pressure calibrator model CPH7650 is a compact, portable pressure calibrator, which has been designed for mobile use as well as for stationary workshop and laboratory testing. An integrated electrical pump enables you to generate pressures up to 20 bar (300 psi) and an integrated electrical module also enables you to supply transmitters or sensors with a current of max. 30 mA (voltage (idling) = DC 24 V), in addition to the traditional measurement of current and voltage signals.
Accuracy
The exchangeable model CPT6000 reference sensor
oers an accuracy of 0.025 % of span. The temperature
compensation of the built-in sensor safeguards the accuracy
for applications in the eld. Readings can be displayed in one
of 13 standard units.
Portable pressure calibrator, model CPH7650
Features
With the portable pressure calibrator, not only are accurate pressure measurements possible, but also exact current and voltage measurements, as well as a DC 24 V voltage supply for test items are enabled. Pressure values and electrical measured values are simultaneously shown in the display. The calibrator is integrated into a robust, as well as dust and waterproof transport case. Thus the mobile pressure calibrator is particularly suited to use in harsh industrial environments. With the built-in NiMH rechargeable battery, a battery life of approximately 50 hours or 125 pump cycles up to 20 bar (290 psi) is guaranteed. The external universal power supply is designed for an operating voltage of AC 100 ... 240 V.
01/2018
Data sheets showing similar products: ProcessCalibrator; model CPH6000; see data sheet CT 15.01 Precision hand-held pressure indicator; model CPH6400; see data sheet CT 14.01 Hand-held pressure indicator; model CPH6300; see data sheet CT 12.01 Calibration software; WIKA-Cal; see data sheet CT 95.10
Page 1 of 8WIKA data sheet CT 17.02
Page 2
Software
The WIKA-Cal calibration software is available for online calibrations in connection with a PC. Using this software, the data is automatically transferred into a printable calibration
certicate. Furthermore, WIKA-Cal also oers, over and
Certied accuracy
For the CPH7650, the accuracy is certied by a factory calibration certicate which accompanies the instrument. On request, we can provide a DKD/DAkkS calibration certicate
for this instrument. above PC-supported calibration, the management of the calibration and instrument data in an SQL database. A USB interface is available for the data transfer.
Specications
Model CPH7650
Measuring ranges
Pressure
Gauge pressure -1 … +20 bar, -1 … +10 bar, 0 … 20 bar, 0 … 10 bar
Absolute pressure 0 … 10 bar abs., 0 … 20 bar abs.
Overpressure limit
Accuracy 0.025 % FS
Resolution 5-digit
Temperature compensation 15 … 35 °C (59 ... 95 °F)
Temperature coecient 0.002 % of span/°C outside of 15 ... 35 °C (59 ... 95 °F)
Current
Measuring range 0 … 24 mA (max. load 1,000 Ω)
Resolution 1 µA
Accuracy 0.015 % of reading ±2 µA (simulation and measurement)
Voltage
Measuring range DC 0 … 30 V
Resolution 1 mV
Accuracy 0.015 % of reading ±2 mV (measurement)
1) Via exchangeable model CPT6000 reference pressure sensors
2) The electrical pump can generate -0.85 … +20 bar (-12 … 290 psi).
1)
2)
3 times
Base instrument
Pressure supply -0.85 ... +20 bar (-12 ... 290 psi), via integrated electric pump
Pressure connection for test item G ½" female thread
Permissible media For dry, clean and non-aggressive gases
Output
Voltage supply DC 24 V
Power supply
Battery type NiMH rechargeable battery
Permissible ambient conditions
Operating temperature -10 … +50 °C (14 ... 122 °F)
Storage temperature -20 … +60 °C (-4 ... +140 °F)
Relative humidity 35 ... 85 % r. h. (non-condensing)
Page 2 of 8 WIKA data sheet CT 17.02 ∙ 01/2018
Page 3
Base instrument
Communication
Interface USB via special interface cable
Case
Case NK-7TM resin
Front panel Aluminium
Ingress protection IP67 (case closed)
IP40 (case opened)
Dimensions 387.4 x 304.8 x 177.8 mm (15.25 x 12 x 7 in)
Weight approx. 7 kg (15.5 lbs.)
Certicates
Certicate
Calibration Standard: 3.1 calibration certicate per DIN EN 10204
Option: DKD/DAkkS calibration certicate
Recommended recalibration interval 1 year (dependent on conditions of use)
Approvals and certicates, see website
Available pressure range and resolution
Pressure range and factors
Gauge pressure -1 ... +20 bar (-14.5 ... 290 psi)
Overpressure limit 40 bar (580 psi)
Unit Conversion factor
psi 1 300.00
bar 0.06894757 20.684
mbar 68.94757 20,684
kPa 6.894757 2,068.4
MPa 0.00689476 2.0684
kg/cm² 0.07030697 21.092
O (4 °C) 70.3089 21,093
cmH
2
O (20 °C) 70.4336 21,130
cmH
2
O (4 °C) 27.68067 8,304.2
inH
2
O (20 °C) 27.72977 8,318.9
inH
2
O (60 °F) 27.70759 8,312.3
inH
2
mmHg (0 °C) 51.71508 15,515
inHg (0 °C) 2.03602 610.81
3) Electrical pump: -0.85 … +20 bar (-12 … 290 psi)
WIKA data sheet CT 17.02 ∙ 01/2018 Page 3 of 8
Page 4
104 (4,11)
158 (6,22)
390 (15,36)
172 (6,76)
Dimensions in mm (in)
308 (12,13)
Page 4 of 8 WIKA data sheet CT 17.02 ∙ 01/2018
Page 5
Features
Pressure supply
The CPH7650, through an integrated electrical pump, enables the provision of -0.85 ... +20 bar. With this, the operator can supply the test item to be calibrated with pressure on-site and does not need any separate pressure supply.
Current and voltage
In addition to pressure, the portable pressure calibrator also measures and simulates 4 ... 20 mA current loop signals, can measure up to DC 30 V and features an internal DC 24 V voltage supply to power the transmitter being calibrated. The DC 24 V voltage supply is switchable for transmitters with mA as well as V output signal.
Display
It can display two calibration values simultaneously. This means that the external reference sensor and electrical test item values (mA or V) can be displayed simultaneously. The instrument has a large graphic LC display with backlighting.
Pressure ranges
The CPH7650 can be tted with dierent reference pressure
sensors, so that the pressure calibrator is suitable for the
calibration of test items with dierent pressure ranges up to
20 bar.
Each pressure range can be accurately generated with the integrated electric pump at the touch of a button. The ne
adjustment is made with the integrated volume adjuster.
Functions
The CPH7650 has a complete range of useful functions. Switch tests can be carried out and recorded. The deviation of the test item is directly calculated and indicated on the
display. A damping function is available. Up to ve frequently
used instrument settings can be stored and retrieved by pressing a button.
Robustness
The CPH7650 consists of a calibrator, which is built into a rugged service case. With this design, the portable pressure calibrator is suited to extremely harsh operating conditions.
WIKA data sheet CT 17.02 ∙ 01/2018 Page 5 of 8
Page 6
Dirt trap
The CPH7650 should only be used with dry and clean test items. Soiling of the integrated pump, for example, through contaminated test items, can lead to faults on the pump or require the pump to be cleaned.
By using the dirt trap specically developed for the CPH7650,
contamination of the integrated pump can be prevented.
Using the dirt trap
The dirt trap is attached, hand-tight, at the pressure connection of the calibrator. Sealing is achieved via the
O-ring tted on the G ½" male thread of the dirt trap. The
maximum operating pressure is limited to 25 bar.
The test item is mounted to the upper connection of the dirt trap, using a suitable sealing method.
As soon as moisture or dirt accumulates in the transparent chamber, the liquid should be drained via a drain valve and the chamber should be cleaned by subsequent dismantling.
Once the upper dirt trap connection has been unscrewed, the transparent acrylic chamber and the O-rings can be removed and cleaned with a clean cloth.
Input
3
Output to calibrator
4
⅞" or 23 mm spanner
5
Dirt trap
5
4
Page 6 of 8 WIKA data sheet CT 17.02 ∙ 01/2018
3
Page 7
WIKA-Cal calibration software
Easy and fast creation of a high-quality calibration certicate
The WIKA-Cal calibration software is used for generating
calibration certicates or logger protocols for pressure
measuring instruments and is available as a demo version for a cost-free download. A template helps the user and guides him through the creation process of a document. In order to switch from the demo version to a full version of the respective template, a USB stick with the template must be purchased. The pre-installed demo version automatically changes to the selected full version when the USB stick is inserted and remains available as long as the USB stick is connected to the computer.
Creation of calibration certicates for mechanical and
electronic pressure measuring instruments
A calibration assistant guides you through the calibration
Automatic generation of the calibration steps
Generation of 3.1 certicates per DIN EN 10204
Creation of logger protocols
User-friendly interface
Languages: German, English, Italian and more due with
software updates
For further information see data sheet CT 95.10
Calibration certicates can be created with the Cal-Template and logger protocols can be created with the Log-Template.
Cal Demo
Generation of calibration certicates limited to 2 measuring points, with
automatic initiation of pressures via a pressure controller.
Cal Light
Generation of calibration certicates with no limitations on measuring
points, without automatic initiation of pressures via a pressure controller.
Calibration certificate
Kalibrierzeugnis
General
Allgemein
Calibration certificate
Kalibrierschein
Customer Demo customer
Kunde
Demo street DE-12345 Demo city
Customer order No.
Kundenbestellnummer
Order date
Bestelldatum
Date of calibration 01/02/2017
Datum der Kalibrierung
Operator Demo User
Bearbeiter
Environmental conditions
Umgebungsbedingungen
Temperature (20,0 ± 0,1) °C
Temperatur
Atmospheric pressure (1013,0 ± 0,1) hPa
Atmosphärischer Luftdruck
Rel. humidity (40 ± 1) %
Rel. Luftfeuchte
Measurement conditions
Messbedingungen
Working standard (WS) Pressure controller
Gebrauchsnormal (GN) Druckcontroller
Model Demo CPC
Typ
Calibration mark Demo 12345
Kalibrierzeichen
Scale range (0 ... 10) bar g
Anzeigebereich
Accuracy 0.02 % FS
Genauigkeit
Demo laboratory
Tel.: 01234/56789
Demo street 123
Fax: 01234/56789
12345 Demo city
E-Mail: demo@demo.com
Demo country
www.demo.com
Certificate No. 00000122
Zeugnis-Nr.
Device under test (DUT)
Kalibriergegenstand (KG)
Electrical gauge
Elektrisches Druckmessgerät
Model D-10
Typ
Serial number demo1234
Seriennummer
Scale range (0 ... 10) bar g
Anzeigebereich
Accuracy 0.05 % FS
Genauigkeit
Resolution (scale division) 0.0001 bar
Auflösung (Skalenwert)
Measuring point No. demo tag
Messstellen-Nr
Place of calibration
Ort der Kalibrierung
g = (9.80946 ± 0.01) m/s²
Demo place
Musterort
Calibration certificate
Kalibrierzeugnis
Measurement results
Messergebnisse
Reference value WS Reading DUT
Statement of compliance
Konformitätsaussage
Object keeps the specifications!
Der Kalibriergegenstand hält die Herstellerspezifikationen ein!
Demo laboratory
Page 1 / 2
Demo street 123
Seite
12345 Demo city Demo country
Referenzwert GN Ablesung KG
bar bar
↑ ↓
0.00000 0.0000 0.0010
10.00000 10.0010 10.0020
Tel.: 01234/56789 Fax: 01234/56789 E-Mail: demo@demo.com www.demo.com
Certificate No. 00000122
Zeugnis-Nr.
Log Demo
Creation of data logger test reports, limited to 5 measured values.
Log
Creation of data logger test reports without limiting the measured values.
Page 2 / 2
Seite
WIKA data sheet CT 17.02 ∙ 01/2018 Page 7 of 8
Page 8
Scope of delivery
Portable pressure calibrator model CPH7650
Operating instructions
Test cables
Battery charger
3.1 calibration certicate per DIN EN 10204
Option
DKD/DAkkS certied accuracy
Accessories
Connection adapters
Various pressure adapters
Software
WIKA-Cal calibration software
Ordering information
CPH7650 / Reference sensor / Unit / Pressure type / Measuring range / Accuracy / Type of certicate / Electrical module /
Connection adapter / Further approvals / Additional ordering information
© 01/2018 WIKA Alexander Wiegand SE & Co. KG, all rights reserved.
The specications given in this document represent the state of engineering at the time of publishing. We reserve the right to make modications to the specications and materials.
Page 8 of 8 WIKA data sheet CT 17.02 ∙ 01/2018