Precision hand-held pressure indicator
Model CPH6400
Applications
■
Calibration service companies and service industry
■
Measurement and control laboratories
■
Quality assurance
Special features
■
Digital indicator with easily interchangeable pressure
sensors (plug and play)
■
Measuring ranges from 0 ... 400 mbar up to 0 ... 6,000 bar
(vacuum and absolute pressure ranges also available)
■
Accuracy: 0.025 % (incl. calibration certicate)
■
Simultaneous pressure and temperature measurement
with external Pt100 resistance thermometer
■
Min, max, pressure rate and data logger function
Calibration technology
WIKA data sheet CT 14.01
Description
Features
A wide variety of pressure sensors, oering various
measuring ranges of up to 6,000 bar, are available to
the operator, always ensuring a solution matched to the
application. The individual sensors can be exchanged quickly
and without tools (plug and play) and are automatically
recognised by the digital indicator. Also, as an option, very
accurate temperature measurement with a Pt100 resistance
thermometer is possible. In addition, this instrument
features a large, easily-read display with backlighting (on/o
selectable), a durable housing and simple, clear operation.
Functionality
In addition to mobile and high-precision pressure and
temperature measurement, the CPH6400 also opens up
other helpful applications e.g. recording of pressure and
temperature curves (via the logging function), leak testing
(via the pressure rate function) and peak value determination (via the Min/Max function). These useful functions can
be easily activated or deactivated by pressing the respective
keys.
Precision hand-held pressure indicator model CPH6400
with optional model CPT6400 reference pressure sensor
Software
Using WIKA's PrintCal software, the logged data, stored in
the CPH6400, can be transferred to a PC via a USB connection, where it can be stored in an EXCEL
®
le for documen-
tation and further evaluation. In addition to this, PrintCal
oers the possibility of PC-supported calibration of pressure
measuring instruments using the CPH6400.
Complete test and service set
For maintenance and service applications, various service
case systems are available. These include pressure service
sets with or without pressure pump, charger, adapter, etc.
Certied accuracy
For each reference pressure transmitter and Pt100 resistance thermometer, the accuracy for the complete measuring
chain is certied by a factory calibration certicate which
accompanies the instrument. Alternatively, on request, a
DKD/DAkkS calibration certicate can be supplied.
Data sheets showing similar products and accessories:
ProcessCalibrator; model CPH6000; see data sheet CT 15.01
Model CPH6600 hand-held pressure calibrator with integrated pump; see data sheet CT 16.01
Hand test pumps, hydraulic; CPP series; see data sheet CT 91.05
Hand test pump, pneumatic; Model CPP30; see data sheet CT 91.06
Page 1 of 8WIKA data sheet CT 14.01 ∙ 02/2011
Dimensions in mm
Digital indicator CPH6400
Front view
Side view
Connection for
temperature sensor
(optional)
Dimensions in mm
Reference pressure sensor CPT6400
Electrical connections to the digital
indicator
Top view
Connection for
interface adapter
Sensor connection
Power connection
Page 2 of 8 WIKA data sheet CT 14.01 ∙ 02/2011
Specications
Precision hand-held pressure indicator model CPH6400
(complete measuring chain)
Sensor technology 1 reference pressure sensor (interchangeable without tools)
Measuring range bar 0.4 1.6 6 16 40 100 250 600 1000
Overpressure safety bar 2 10 35 80 80 200 500 1200 1500
Burst pressure bar 2.4 12 42 96 400 800 1200 2400 3000
Accuracy of the measuring chain 0.025 % FS
2)
Measuring range bar 1600 2500 4000 5000 6000
Overpressure safety bar 2300 3500 5000 6000 7000
Burst pressure bar 4000 6000 8000 10000 11000
Accuracy of the measuring chain 0.1 % FS
2)
Type of pressure {In addition to the above-specied pressures; vacuum, bi-directional ranges and absolute
pressures are available}
Pressure units bar, mbar, kPa, psi, mmHg, inHg and kg/cm² (depending on the measuring range, selectable)
Active temperature compensation °C 10 ... 40
Temperature measurement only for 2-channel version
■
Sensor type Pt100, 4-wire
■
Measuring range °C -10 ... +50
■
Resolution °C 0.01
■
Accuracy °C 0.05
Permissible ambient temperature °C 0 ... 50
Calibration
2)
Factory calibration certicate (optional: DKD/DAkkS calibration certicate)
1)
Specications Digital indicator model CPH6400
Display Large graphic display, with backlighting (On/O selectable)
Display resolution up to 6 digits; selectable
Measuring rate (pressure) 5 values/sec
Functions Min, Max memory, pressure rate, tare, oset-correction for gauge pressure ranges, data log-
ger function
Data logger - Cyclic data logger: automatic recording of up to 1000 values (500 pressure values and 500
temperature values)
- Cycle time: adjustable from 1 ... 3600 seconds
Interface USB
Permissible
■
Relative humidity % r. H. 0 ... 85 (non-condensing; at 50 °C)
■
Storage temperature °C -20 ... +60
Power supply
■
Current supply Internal Lithium-Ion rechargeable battery (charging time: < 6 h)
■
Battery operation h approx. 25
Case Polyamide 12, membrane keyboard, transparent panels
Ingress protection IP 65
Weight g approx. 480
CE conformity
■
EMC directive
2004/108/EC, EN 61326 Emission (Group 1, Class B) and Immunity (industrial locations)
Specications Reference pressure sensor model CPT6400
Pressure connection ≤ 1000 bar: G ½ B; {various connection adapters on request}
> 1000 bar: M16 x 1.5 female, with sealing cone
Material Stainless steel for wetted parts (measuring range > 25 bar additional Elgiloy
Internal transmission uid Synthetic oil (only for measuring ranges up to 25 bar)
Permissible temperature ranges
■
Medium °C -20 ... +80
■
Storage °C -40 ... +85
Case Stainless steel
Electrical connection Circular connector, 8-pin
Ingress protection IP 65 (with cable connected)
Weight g approx. 220
CE conformity
■
Pressure equipment directive 97/23/EC, PS > 200 bar; module A, pressure accessory
■
EMC directive
1) Calibration data sets for up to 10 transmitters can be stored in each instrument (up to 10 calibration data sets)
2) Calibrated at 23 °C and in vertical mounting position with pressure connection facing downwards.
{} Items in curved brackets are optional extras for an additional price.
2004/108/EC, EN 61326 Emission (Group 1, Class B) and Immunity (industrial locations)
®
)
Page 3 of 8WIKA data sheet CT 14.01 ∙ 02/2011