Software and complete service cases (incl. pumps)
available
Model CPH6300-S1 hand-held pressure indicator with
model CPT6200 reference pressure sensor
Calibration technology
WIKA data sheet CT 12.01
Description
Extensive application possibilities
Stainless steel pressure sensors with measuring ranges up to
1,000 bar (14,500 psi) are available for the model CPH6300
hand-held pressure indicator. The robust and waterproof
design makes the CPH6300 ideal for use in adverse
environments. The digital indicator automatically detects
the measuring range of the connected pressure sensor and
guarantees a highly accurate pressure measurement.
Functionality
The CPH6300 can be used for measuring both gauge and
absolute pressure. Dierential pressure measurement is
possible with the 2-channel version CPH6300-S2, and two
connected model CPT6200 reference pressure sensors.
Selectable pressure units here are bar, mbar, psi, Pa, kPa,
MPa, mmHg, inHg, mH
Data sheets showing similar products and accessories:
Hand-held pressure indicator; model CPH6200; see data sheet CT 11.01
Intrinsically safe hand-held pressure indicator (ATEX version); model CPH62I0; see data sheet CT 11.02
Hydraulic test pumps; CPP series; see data sheet CT 91.05
Pneumatic hand test pump; model CPP30; see data sheet CT 91.06
Calibration software; WIKA-Cal; see data sheet CT 95.10
O and one customer-specic unit.
2
∙ 08/2017
An integrated data logger and various other functions such
as Min., Max., Hold, Tare, zero point adjustment, alarm,
power-o, peak value detection (1,000 measurements/s),
average value lter, etc. ensure that the CPH6300 can
be used for many dierent applications. The large, backlit
display and also a long battery life complete the features of
the CPH6300.
Software
In addition to the GSoft data logger evaluation software
for the tabular and graphical representation of the logged
data, WIKA-Cal calibration software for calibration tasks
is also available. WIKA-Cal also oers, over and above
PC-supported calibration, the management of the calibration
and instrument data in an SQL database. A USB interface is
available for the data transfer.
Page 1 of 11WIKA data sheet CT 12.01
Complete test and service cases
For maintenance and service applications, various case
systems are available. These include service cases with or
without pressure generation, rechargeable battery, battery
charger, connection adapter, etc.
Certied accuracy
For each reference pressure sensor, the accuracy for the
complete measuring chain is certied by a factory calibration
certicate which accompanies the instrument. On request,
we can provide a DKD/DAkkS calibration certicate for this
instrument.
Specications
Model CPH6300 hand-held pressure indicator (complete measuring chain)
Resolutiondependant on pressure range (max. 4 1/2-digit)
1)
Accuracy of the measuring chain
Types of pressureGauge pressure, {absolute pressure from 0 ... 25 bar abs. (0 ... 360 psi abs.) and vacuum from
Sensor compatibilityCompatible with model CPT6200 reference pressure sensors
{} Items in curved brackets are optional extras for an additional price.
1) It is dened by the total measurement uncertainty, which is expressed with the coverage factor (k = 2) and includes the following factors: the intrinsic performance of the instrument, the
measurement uncertainty of the reference instrument, long-term stability, inuence of ambient conditions, drift and temperature eects over the compensated range during a periodic
Menu functionsMin./Max. alarm (acoustic/visual), sea level (barometric air pressure), power-o function,
measuring rate, mean value lter
Mean value lter1 ... 120 seconds, adjustable
Data loggerIndividual value logger: Up to 1,000 recordings (with measuring point input, 40 settable
measuring point texts or measuring point nos.) incl. time via button press
Cyclic logger: Automatic recording of up to 10,000 values incl. time
Cycle time: Selectable from 1 ... 3,600 seconds
Real-time clockintegrated clock with date
Leak testing/sealing testsPressure rate display, logging via data logger
Voltage supply
Power supply2 x 1.5 V AAA batteries
Battery life> 500 hours of operation (1 sensor with a measuring rate of 4/s)
Battery status indicationIcon in display
Permissible ambient conditions
Operating temperature-25 ... +50 °C (-13 ... +122 °F)
Storage temperature-25 ... +70 °C (-13 ... +158 °F)
Relative humidity0 ... 95 % r. h. (non-condensing)
Communication
InterfaceUSB via interface cable
Analogue outputDC 0 ... 1 V; congurable (selectable via menu alternative to interface)
Weightapprox. 250 g (0.55 lbs) (incl. batteries and protective casing)
O and a customer-specic unit, pre-set to
2
WIKA data sheet CT 12.01 ∙ 08/2017 Page 3 of 11
Reference pressure sensor model CPT6200
Pressure connection
Material
Wetted partsStainless steel or Elgiloy
Internal transmission uidSynthetic oil (only for measuring ranges to 16 bar (250 psi) or ush diaphragm)
Sensor specications
Accuracy
1)
Compensated range0 ... 80 °C (0 ... 176 °F)
Mean temperature coecient≤ 0.2 % of span/10 K (outside of reference conditions)
Permissible ambient conditions
Medium temperature
Operating temperature-20 ... +80 °C (-4 ... +176 °F)
Storage temperature-40 ... +100 °C (-40 ... +212 °F)
Humidity0 ... 95 % r. h. (non-condensing)
Case
MaterialStainless steel
Connection to the CPH6300Standard: via 1 m (3.3 ft) connection cable (plug-and-play)
Ingress protectionIP67
DimensionsSee technical drawing
Weightapprox. 220 g (0.49 lbs)
{} Items in curved brackets are optional extras for an additional price.
1) It is dened by the total measurement uncertainty, which is expressed with the coverage factor (k = 2) and includes the following factors: the intrinsic performance of the instrument, the
measurement uncertainty of the reference instrument, long-term stability, inuence of ambient conditions, drift and temperature eects over the compensated range during a periodic
3) For pressure measuring ranges 0 ... 25 mbar, 0 ... 40 mbar and 0 ... 60 mbar (0 ... 0.4 psi, 0 ... 0.6 psi and 0 ... 0.9 psi) all wetted parts are made of stainless steel, silicium, aluminium,
gold, silicone.
4) As an oxygen version, a ush diaphragm model is not available. In an oxygen version, the model CPT6200 is only available in overpressure ranges ≥ 0.25 bar (≥ 0.4 psi), with media
temperatures between -10 … +50 °C (14 ... 122 °F) and using stainless steel or Elgiloy
5) For pressure measuring ranges 0 ... 25 mbar, 0 ... 40 mbar and 0 ... 60 mbar (0 ... 0.4 psi, 0 ... 0.6 psi and 0 ... 0.9 psi) the medium temperature and storage temperature are limited to
+80 °C (176 °F).
4)
G ½ B; {ush (G 1 for 0.1 ... 1.6 bar (1.5 ... 25 psi)) or various connection adapters on request}
{Halocarbon oil for oxygen applications}; {Listed by FDA for food industry}
≤ 0.2 % of span at reference conditions
4)
-30 ... +100 °C (-22 ... +212 °F)
2)
5)
5)
Optional: up to 5 m (16.4 ft)
®
wetted parts.
Approvals
LogoDescriptionCountry
EU declaration of conformity for CPH6300
■
EMC directive
EN 61326 emission (group 1, class B) and interference immunity (portable equipment)
■
RoHS directive
EU declaration of conformity for CPT6200
■
EMC directive
EN 61326 emission (group 1, class B) and interference immunity (portable equipment)
■
Pressure equipment directive
PS > 200 bar, module A, pressure accessory
■
RoHS directive
EAC
■
EMC directive
■
Pressure equipment directive
GOST
Metrology, measurement technology
KazInMetr
Metrology, measurement technology
-MTSCHS
Permission for commissioning
BelGIM
Metrology, measurement technology
European Union
European Union
Eurasian Economic
Community
Russia
Kazakhstan
Kazakhstan
Belarus
Page 4 of 11WIKA data sheet CT 12.01 ∙ 08/2017
42 (1.65)
86 (3.39)
42 (1.65)
86 (3.39)
LogoDescriptionCountry
UkrSEPRO
Metrology, measurement technology
Uzstandard
Metrology, measurement technology
Ukraine
Uzbekistan
Certicates
Certicate
CalibrationStandard: 3.1 calibration certicate per DIN EN 10204
Option: DKD/DAkkS calibration certicate
Recommended recalibration interval1 year (dependent on conditions of use)
Approvals and certicates, see website
Dimensions in mm (in)
Digital indicator CPH6300-S1Digital indicator CPH6300-S2
163 (6.42)
163 (6.42)
WIKA data sheet CT 12.01 ∙ 08/2017 Page 5 of 11
88.5 (3.48)
0.24)
Dimensions in mm (in)
Reference pressure sensor CPT6200
Ø 27 (Ø 1.06)
27
)
+0.3
+0.01
28.5
(1.12
3 (0.12)
20 (0.79)
3 (0.12)
Ø 17.5
(Ø 0.69)
G 1/2 B
Ø 6 (Ø
Electrical connections
Model CPH6300-S1
12
Model CPH6300-S2
34
1
Connection for interface adapter
2
Connection channel 1 (only with CPH6300-S1)
3
Connection channel 2 (only with CPH6300-S2)
4
Connection channel 1 (only with CPH6300-S2)
1
Operation of the models CPH6300-S1 and CPH6300-S2
1- and 2-channel version with external pressure sensors
1
2
3
4
5
6
7
8
Screen
6
8
7
6
453
1
2
Main display: Current measured value for sensor 1
Secondary display: Current measured value for sensor
2 or dierential value between sensor 1 and sensor 2
Logg arrow: Logger is ready
Arrow blinking: Automatic recording (Logg CYCL) active
Corr arrow: Zero point or slope correction has been
made
User arrow: Measured value is specied in the freelycongurable user unit
Display arrows for measured value units
Indication elements for Min./Max. measured values, as
well as the Tare function and sea-level correction
Battery status indication
Page 6 of 11WIKA data sheet CT 12.01 ∙ 08/2017
Keyboard
2 sec.
Instrument
Illumination
2 sec.
Instrument
Change the secondary
display
CH1 <-> CH2 <-> DIF
(only for 2-channel)
Main menu
Enter conguration
on
on
o
2 sec.
2 sec.
MAX function
on/o
Delete MAX value
MIN function
on/o
Delete MIN value
2 sec.
5 sec.
10 sec.
2 sec.
TARE function
on
TARE function
o
Zero-point
adjustment on
Zero-point
adjustment o
Hold function
on/o
Store measured
value
Clear memory?
if:
A
B
= Press button
2 sec. = Press button for 2 seconds
For more information: See operating instructions
2 sec.
2 sec.
2 sec.
Store cyclically
Stop storage?
Clear memory?
C
A = Logger functions deactivated
B = Logger function “Store measured value” activated via menu
C = Logger function “Store cyclically” activated via menu
WIKA data sheet CT 12.01 ∙ 08/2017 Page 7 of 11
Complete test and service cases
Calibration case with model CPH6300 hand-held
pressure indicator for pressure, consisting of:
■
Plastic service case with foam insert
■
Hand-held pressure indicator model CPH6300
■
Two AAA batteries
■
Sealing set
■
Sensor cable
■
Spaces for several CPT6200 reference pressure sensors
Basic version
Basic version incl. pneumatic pressure generation
Available measuring ranges see specications
Calibration case with model CPH6300 hand-held
pressure indicator and model CPP30 hand test pump
for pressures of -0.95 ... +35 bar (-28 inHg ... 500 psi),
consisting of:
■
Plastic service case with foam insert
■
Hand-held pressure indicator model CPH6300
■
Pneumatic hand test pump model CPP30; -0.95 ... +35 bar
(-28 inHg ... 500 psi)
■
Sealing set
■
Sensor cable
■
Spaces for several CPT6200 reference pressure sensors
Available measuring ranges see specications
Calibration case with model CPH6300 hand-held
pressure indicator and model CPP700-H or CPP1000-H
hand test pump for pressures of 0 ... 700 bar or
0 ... 1,000 bar (0 ... 10,000 psi or 0 ... 14,500 psi),
consisting of:
■
Plastic service case with foam insert
■
Hand-held pressure indicator model CPH6300
■
Hydraulic hand test pump model CPP700-H or
CPP1000-H, 0 ... 700 bar or 0 ... 1,000 bar (0 ... 10,000 psi
or 0 ... 14,500 psi)
■
Sealing set
■
Sensor cable
■
Spaces for several CPT6200 reference pressure sensors
Basic version incl. hydraulic pressure generation
Page 8 of 11WIKA data sheet CT 12.01 ∙ 08/2017
Available measuring ranges see specications
GSoft data logger evaluation software
The GSoft data logger evaluation software is used to display
the logger data of the model CPH6300 hand-held pressure
indicator on a PC in tabular form and as chart.
■
Easy operation with self-explanatory toolbars
■
Data from the pressure and temperature hand-helds
(CTH6200) can be displayed in a single chart (two
separate y-axes)
■
Chart oers a zoom function
■
Operation of the logger function via PC (remote control)
■
Data can be exported (Excel®, etc.)
■
Languages: German, English, French, Spanish and Czech
System requirements
■
IBM compatible PC (Pentium®)
■
At least 20 MB free hard disc space
■
CD-ROM drive
■
At least 32 MB RAM
■
Windows® operating system 95, 98, NT 4.0 (with Service
Pack 3.0 or higher), 2000, XP, Vista or 7
■
Mouse
■
USB port (via interface cable)
Data export e.g. in
an Excel® le
To operate the GSoft software with the CPH6300, GSoft
version 3.0 or later is needed.
Free updates are available for download on www.wika.com.
Windows® is a registered trademark of Microsoft Corporation in the United States and other countries.
Easy operation with self-explanatory icon buttons
Charts toolbarMain toolbar
123456789
1
File functions: Open, save, print
2
Logger functions: Start communication, start logger, stop,
read data
3
Data display: Create chart
4
Interface conguration
5
Program information
6
Settings: Grid and colour settings, manual zooms
7
Zoom: All, left or right y-axis (via mouse), back
8
Rename chart
9
Cursor on/o (Info footer)
10
Legend on/o
11
(Measuring point) Symbols on/o
12
Measurement series (add/delete)
13
Comments on measuring points (add/delete)
10 111213
WIKA data sheet CT 12.01 ∙ 08/2017 Page 9 of 11
WIKA-Cal calibration software
Easy and fast creation of a high-quality calibration certicate
The WIKA-Cal calibration software is used for generating
calibration certicates or logger protocols for pressure
measuring instruments and is available as a demo version for
a cost-free download.
A template helps the user and guides him through the
creation process of a document.
In order to switch from the demo version to a full version of
the respective template, a USB stick with the template has to
be purchased.
The pre-installed demo version automatically changes to
the selected full version when the USB stick is inserted and
remains available as long as the USB stick is connected to
the computer.
■
Creation of calibration certicates for mechanical and
electronic pressure measuring instruments
■
A calibration assistant guides you through the calibration
■
Automatic generation of the calibration steps
■
Generation of 3.1 certicates per DIN EN 10204
■
Creation of logger protocols
■
User-friendly interface
■
Languages: German, English, Italian and more due with
software updates
For further information see data sheet CT 95.10
Calibration certicates can be created with the Cal-Template and logger protocols can be created with the Log-Template.
Cal Demo
Generation of calibration certicates limited to 2 measuring points, with
automatic initiation of pressures via a pressure controller.
Cal Light
Generation of calibration certicates with no limitations on measuring
points, without automatic initiation of pressures via a pressure
controller.
Calibration certificate
Kalibrierzeugnis
General
Allgemein
Calibration certificate
Kalibrierschein
CustomerDemo customer
Kunde
Demo street
DE-12345 Demo city
Customer order No.
Kundenbestellnummer
Order date
Bestelldatum
Date of calibration 01/02/2017
Datum der Kalibrierung
OperatorDemo User
Bearbeiter
Environmental conditions
Umgebungsbedingungen
Temperature (20,0 ± 0,1) °C
Temperatur
Atmospheric pressure (1013,0 ± 0,1) hPa
Atmosphärischer Luftdruck
Rel. humidity (40 ± 1) %
Rel. Luftfeuchte
Measurement conditions
Messbedingungen
Working standard (WS) Pressure controller
Gebrauchsnormal (GN) Druckcontroller
ModelDemo CPC
Typ
Calibration markDemo 12345
Kalibrierzeichen
Scale range(0 ... 10) bar g
Anzeigebereich
Accuracy0.02 % FS
Genauigkeit
Demo laboratory
Tel.: 01234/56789
Demo street 123
Fax: 01234/56789
12345 Demo city
E-Mail: demo@demo.com
Demo country
www.demo.com
Certificate No.00000122
Zeugnis-Nr.
Device under test (DUT)
Kalibriergegenstand (KG)
Electrical gauge
Elektrisches Druckmessgerät
ModelD-10
Typ
Serial number demo1234
Seriennummer
Scale range (0 ... 10) bar g
Anzeigebereich
Accuracy0.05 % FS
Genauigkeit
Resolution (scale division) 0.0001 bar
Auflösung (Skalenwert)
Measuring point No. demo tag
Messstellen-Nr
Place of calibration
Ort der Kalibrierung
g = (9.80946 ± 0.01) m/s²
Demo place
Musterort
Calibration certificate
Kalibrierzeugnis
Measurement results
Messergebnisse
Reference value WS Reading DUT
Statement of compliance
Konformitätsaussage
Object keeps the specifications!
Der Kalibriergegenstand hält die Herstellerspezifikationen ein!
Creation of data logger test reports, limited to 5 measured values.
Page 2 / 2
Seite
Log
Creation of data logger test reports without limiting the measured
values.
Page 10 of 11WIKA data sheet CT 12.01 ∙ 08/2017
AccessoriesOrder code
2 AAA rechargeable batteriesCPH-A-63-ZAZZZZZZZZ-Z
2 AAA batteriesCPH-A-63-ZBZZZZZZZZ-Z
Charger for 9 V rechargeable battery and 2 rechargeable AAA batteries (Euro standard)CPH-A-63-Z1ZZZZZZZZ-Z
Charger for 9 V rechargeable battery and 2 rechargeable AAA batteries (UK standard)CPH-A-63-Z2ZZZZZZZZ-Z
Charger for 9 V rechargeable battery and 2 rechargeable AAA batteries (US standard)CPH-A-63-Z3ZZZZZZZZ-Z
Sealing set, consisting of 4 x G ½ USIT seals, 2 x G ¼ USIT seals and plastic boxCPH-A-63-ZDZZZZZZZZ-Z
Sensor connection cable, approx. 1.1 m (3.3 ft)CPH-A-63-ZSZZZZZZZZ-Z
Extension cable for connection of sensors, approx. 3.8 m (12.5 ft) to approx. 5 m (16.4 ft)CPH-A-63-ZVZZZZZZZZ-Z
2 connection cable, 2 m (6.6 ft) with loose ends (sleeves) for connecting the congurable analogue outputCPH-A-63-ZEZZZZZZZZ-Z
USB interface cableCPH-A-63-ZUZZZZZZZZ-Z
Gsoft data logger evaluation softwareCPH-A-63-ZGZZZZZZZZ-Z
Plastic case for 1 x hand-held, 3 x pressure sensors, accessoriesCPH-A-63-ZKZZZZZZZZ-Z
Plastic case for 1 x hand-held, 5 x pressure sensors, 1 x pneumatic hand test pump CPP7-H or CPP30,
accessories
Plastic case for 1 x hand-held, 4 x pressure sensors, 1 x hydraulic hand test pump CPP700-H/CPP1000-H,
accessories
CPH-A-63-ZLZZZZZZZZ-Z
CPH-A-63-ZNZZZZZZZZ-Z
Scope of delivery
■
Hand-held pressure indicator model CPH6300-S1 incl.
two AAA batteries
■
One sensor connection cable per channel
■
3.1 calibration certicate per DIN EN 10204
■
Choice of sensors
Options
■
Hand-held pressure indicator model CPH6300-S2:
2-channel version (dierential pressure measurement
possible via two connected model CPT6200 reference
pressure sensors)
■
DKD/DAkkS calibration certicate
■
Sensors for oxygen applications
Ordering information
CPH6300 / Instrument version / Additional cable for reference pressure sensor / Rechargeable battery and battery charger /
Software / Interface cable / Test pump / Carrying case / Further approvals / Additional order information
CPT6200 / Unit / Pressure range / Accuracy / Process connection / Special design features / Type of certicate / Further
approvals / Additional order information
Model CPH6300-S2 hand-held pressure indicator with
two model CPT6200 reference pressure sensors
The specications given in this document represent the state of engineering at the time of publishing.
We reserve the right to make modications to the specications and materials.