WIKA CPH62I0-S1, CPH62I0-S2 Operating Instructions Manual

Operating instructions Betriebsanleitung
EN
DE
Intrinsically safe hand-held pressure indicator, model CPH62I0-S1 and CPH62I0-S2
Intrinsically safe hand-held pressure indicator model CPH62I0-S1 with optional model CPT62I0 reference pressure sensor
2
11221801.04 10/2018 EN/DE
WIKA operating instructions, model CPH62I0
EN
DE
Operating instructions model CPH62I0-S1 and CPH62I0-S2 Page 3 - 52
Betriebsanleitung Typ CPH62I0-S1 und CPH62I0-S2 Seite 53 - 102
Further languages can be found at www.wika.com.
© 10/2018 WIKA Alexander Wiegand SE & Co. KG All rights reserved. / Alle Rechte vorbehalten. WIKA
®
is a registered trademark in various countries.
WIKA
®
ist eine geschützte Marke in verschiedenen Ländern.
Prior to starting any work, read the operating instructions! Keep for later use!
Vor Beginn aller Arbeiten Betriebsanleitung lesen! Zum späteren Gebrauch aufbewahren!
3
WIKA operating instructions, model CPH62I0
EN
11221801.04 10/2018 EN/DE
Contents
1. General information 5
2. Short overview 6
2.1 Overview . . . . . . . . . . . . . . . . . . . . . . . . 6
2.2 Description . . . . . . . . . . . . . . . . . . . . . . . 6
2.3 Scope of delivery . . . . . . . . . . . . . . . . . . . . . 7
3. Safety 7
3.1 Explanation of symbols . . . . . . . . . . . . . . . . . . . 7
3.2 Intended use. . . . . . . . . . . . . . . . . . . . . . . 8
3.3 Improper use . . . . . . . . . . . . . . . . . . . . . . 8
3.4 Personnel qualification . . . . . . . . . . . . . . . . . . . 9
3.5 Labelling, safety marks . . . . . . . . . . . . . . . . . . . 9
3.5.1 Product label . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
3.5.2 Explanation of symbols. . . . . . . . . . . . . . . . . . . . . . . . . . . . .10
3.6 Ex marking . . . . . . . . . . . . . . . . . . . . . . . 10
3.6.1 Special conditions for use . . . . . . . . . . . . . . . . . . . . . . . . . . .11
3.6.2 Mains operation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .12
3.6.3 Permitted batteries . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .13
4. Design and function 14
4.1 Display. . . . . . . . . . . . . . . . . . . . . . . . . 14
4.2 Function buttons and operation . . . . . . . . . . . . . . . .15
4.3 Electrical connections . . . . . . . . . . . . . . . . . . . 16
4.4 Voltage supply . . . . . . . . . . . . . . . . . . . . . . 17
4.5 Pressure sensors . . . . . . . . . . . . . . . . . . . . . 18
4.5.1 Available pressure sensors . . . . . . . . . . . . . . . . . . . . . . . . . .18
4.5.2 Connecting/changing pressure sensors . . . . . . . . . . . . . . . . . . .19
4.6 Serial or analogue interface. . . . . . . . . . . . . . . . . . 19
5. Transport, packaging and storage 21
5.1 Transport . . . . . . . . . . . . . . . . . . . . . . . .21
5.2 Packaging and storage . . . . . . . . . . . . . . . . . . . 21
6. Commissioning, operation 22
6.1 Commissioning . . . . . . . . . . . . . . . . . . . . . .22
6.2 Operation . . . . . . . . . . . . . . . . . . . . . . . . 22
6.3 Menu functions . . . . . . . . . . . . . . . . . . . . . .23
6.4 Configuring the instrument . . . . . . . . . . . . . . . . . .26
6.4.1 Pressure units (unit). . . . . . . . . . . . . . . . . . . . . . . . . . . . . .26
6.4.2 Sea-level (SL) and altitude (Alti) correction for absolute pressure sensor 26
6.4.3 Measurement types (rAtE). . . . . . . . . . . . . . . . . . . . . . . . . . .27
6.4.4 Averaging . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .28
Contents
4
WIKA operating instructions, model CPH62I0
EN
11221801.04 10/2018 EN/DE
Declarations of conformity can be found online at www.wika.com.
6.4.5 Zero point correction sensor 1 (OFS.1) or sensor 2 (OFS.2) . . . . . . . . .28
6.4.6 Scale correction sensor 1 (SCL.1) and sensor 2 (SCL.2) . . . . . . . . . . .28
6.4.7 Power-off function (P.oFF) . . . . . . . . . . . . . . . . . . . . . . . . . . .29
6.4.8 Instrument output (Ovt). . . . . . . . . . . . . . . . . . . . . . . . . . . . .29
6.4.9 Analogue output scaling with dAC.0 and dAC.1 (dAC.) . . . . . . . . . . . .29
6.4.10 Alarm (AL.). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .30
6.4.11 Real-time clock (CLOC) . . . . . . . . . . . . . . . . . . . . . . . . . . . . .30
6.5 Operation of the logger function . . . . . . . . . . . . . . . . 31
6.5.1 Storing individual values (Func-Stor). . . . . . . . . . . . . . . . . . . . .32
6.5.2 Automatic recording with adjustable cycle time “Func-CYCL” . . . . . . . .33
7. Faults 36
8. Maintenance, cleaning and recalibration 39
8.1 Maintenance. . . . . . . . . . . . . . . . . . . . . . . 39
8.2 Battery replacement . . . . . . . . . . . . . . . . . . . .39
8.3 Cleaning . . . . . . . . . . . . . . . . . . . . . . . . 40
8.4 Recalibration. . . . . . . . . . . . . . . . . . . . . . . 40
9. Dismounting, return and disposal 41
9.1 Dismounting . . . . . . . . . . . . . . . . . . . . . . .41
9.2 Return . . . . . . . . . . . . . . . . . . . . . . . . .42
9.3 Disposal . . . . . . . . . . . . . . . . . . . . . . . . 42
10. Specifications 43
10.1 Complete measuring chain . . . . . . . . . . . . . . . . . 43
10.2 Digital indicator model CPH62I0 . . . . . . . . . . . . . . . 44
10.3 Reference pressure sensor model CPT62I0 . . . . . . . . . . .45
10.4 Safety-related characteristic values . . . . . . . . . . . . . . 46
10.5 Ex approvals . . . . . . . . . . . . . . . . . . . . . . 47
10.6 Certificates . . . . . . . . . . . . . . . . . . . . . . . 47
10.7 Dimensions in mm (in) . . . . . . . . . . . . . . . . . . .48
11. Accessories 50 Appendix 1: EU declaration of conformity model CPH62I0 51 Appendix 2: EU declaration of conformity model CPT62I0 52
Contents
5
WIKA operating instructions, model CPH62I0
EN
11221801.04 10/2018 EN/DE
1. General information
1. General information
The models CPH62I0-S1 or CPH62I0-S2 intrinsically safe, hand-held pressure indicators described in the operating instructions have been designed and manufactured using state-of-the-art technology. All components are subject to stringent quality and environmental criteria during production. Our management systems are certified to ISO 9001 and ISO 14001.
These operating instructions contain important information on handling the instrument. Working safely requires that all safety instructions and work instructions are observed.
Observe the relevant local accident prevention regulations and general safety regulations for the instrument's range of use.
The operating instructions are part of the product and must be kept in the immediate vicinity of the instrument and readily accessible to skilled personnel at any time. Pass the operating instructions on to the next operator or owner of the instrument.
Skilled personnel must have carefully read and understood the operating instructions prior to beginning any work.
The general terms and conditions contained in the sales documentation shall apply.
Subject to technical modifications.
Factory calibrations / DKD/DAkkS calibrations are carried out in accordance with international standards.
Further information:
- Internet address: www.wika.de / www.wika.com
- Relevant data sheet:
CT 11.02
- Application consultant:
Tel.: +49 9372 132-0 Fax: +49 9372 132-406
info@wika.de
6
WIKA operating instructions, model CPH62I0
EN
11221801.04 10/2018 EN/DE
2. Short overview
2. Short overview
2.1 Overview
1
Sensor connection
2
Hand-held pressure indicator model CPH62I0-S1 or CPH62I0-S2 with leather case (Ex protective cover)
3
Reference pressure sensor model CPT62I0
2.2 Description
The CPH62I0 hand-held pressure indicator can be used as a calibrator, and for any application where an accurate pressure measurement is required, within hazardous areas. Model CPT62I0 reference pressure sensors with measuring ranges up to 1,000 bar are available for the hand-held pressure indicator. This hand-held pressure indicator automatically recognises the measuring range of the connected pressure sensor and guarantees a high-accuracy pressure measurement.
In addition to pressure sensors for gauge and absolute pressure, differential pressure can also be measured with the CPH62I0-S2 and two connected pressure sensors. Selectable pressure units here are bar, mbar, psi, Pa, kPa, MPa, mmHg or inHg. An integrated data logger and various other functions (such as Min, Max, Hold, Tare, zero
1
2
3
7
WIKA operating instructions, model CPH62I0
EN
11221801.04 10/2018 EN/DE
2. Short overview / 3. Safety
point correction, alarm, power-off, variable measuring rate, sea level etc.) ensure that the hand-held pressure indicator can be used for many different applications.
2.3 Scope of delivery
Hand-held pressure indicator, model CPH62I0-S1 or CPH62I0-S2, incl. 9 V battery
Leather case (Ex protective cover)
One sensor connection cable per channel, approx. 1.1 m (3.3 ft)
Calibration certificate for sensor
Choice of sensors
Cross-check scope of delivery with delivery note.
3. Safety
3.1 Explanation of symbols
WARNING!
... indicates a potentially dangerous situation that can result in serious injury or death, if not avoided.
CAUTION!
... indicates a potentially dangerous situation that can result in light injuries or damage to property or the environment, if not avoided.
DANGER!
... identifies hazards caused by electrical power. Should the safety instructions not be observed, there is a risk of serious or fatal injury.
DANGER!
... indicates a potentially dangerous situation in the hazardous area that can result in serious injury or death, if not avoided.
Information
... points out useful tips, recommendations and information for efficient and trouble-free operation.
8
WIKA operating instructions, model CPH62I0
EN
11221801.04 10/2018 EN/DE
3. Safety
3.2 Intended use
The CPH62I0 hand-held pressure indicator can be used as a calibrator, and for any application where an accurate pressure measurement is required, within hazardous areas.
To use the CPH62I0 in hazardous areas, it is imperative to insert the hand-held pressure indicator into the leather case (Ex protective cover) supplied. The press stud of the leather case (Ex protective cover) must be closed.
The instrument has been designed and built solely for the intended use described here, and may only be used accordingly.
The technical specifications contained in these operating instructions must be observed. Improper handling or operation of the instrument outside of its technical specifications requires the instrument to be taken out of service immediately and inspected by an authorised WIKA service engineer.
Handle electronic precision measuring instruments with the required care (protect from humidity, impacts, strong magnetic fields, static electricity and extreme temperatures, do not insert any objects into the instrument or its openings). Plugs and sockets must be protected from contamination.
The manufacturer shall not be liable for claims of any type based on operation contrary to the intended use.
3.3 Improper use
WARNING! Injuries through improper use
Improper use of the instrument can lead to hazardous situations and injuries.
Refrain from unauthorised modifications to the instrument.
Do not use the instrument with abrasive or viscous media.
Only use the serial and analogue interfaces outside the hazardous areas.
Only replace the batteries outside hazardous areas.
Within hazardous areas, the accompanying leather case must be used!
Any use beyond or different to the intended use is considered as improper use.
9
WIKA operating instructions, model CPH62I0
EN
11221801.04 10/2018 EN/DE
3. Safety
3.4 Personnel qualification
WARNING! Risk of injury should qualification be insufficient
Improper handling can result in considerable injury and damage to equipment.
The activities described in these operating instructions may only be carried out by skilled personnel who have the qualifications described below.
Skilled personnel
Skilled personnel, authorised by the operator, are understood to be personnel who, based on their technical training, knowledge of measurement and control technology and on their experience and knowledge of country-specific regulations, current standards and directives, are capable of carrying out the work described and independently recognising potential hazards.
Special knowledge for working with instruments for hazardous areas:
The skilled personnel must have knowledge of ignition protection types, regulations and provisions for equipment in hazardous areas.
Special operating conditions require further appropriate knowledge, e.g. of aggressive media.
3.5 Labelling, safety marks
3.5.1 Product label
The operator is obliged to maintain the product label in a legible condition.
Product label for the hand-held pressure indicator
The product label is fixed on the rear of the hand-held.
1
2
3 4
5
1 Product name 2
Approval-related data
3
Article number
4
Date of manufacture
5
Serial number
10
WIKA operating instructions, model CPH62I0
EN
11221801.04 10/2018 EN/DE
3. Safety
Product label for pressure sensor
1
Date of manufacture (month-year)
2
Approval-related data
3
Serial number and article number
4
Pressure measuring range and accuracy
5
Product name
3.5.2 Explanation of symbols
Before mounting and commissioning the hand-held pressure indicator, ensure you read the operating instructions!
Do not dispose of with household waste. Ensure a proper disposal in accordance with national regulations.
3.6 Ex marking
DANGER! Danger to life due to loss of explosion protection
Non-observance of these instructions and their contents may result in the loss of explosion protection.
Observe the safety instructions in this chapter and further explosion instructions in these operating instructions.
Observe the information given in the applicable type examination certificate and the relevant country-specific regulations for installation and use in hazardous areas (e.g. IEC 60079-14, NEC, CEC).
Operation of the serial and analogue interfaces is not permitted in hazardous areas!
Only use model CPT62I0 pressure sensors!
Within hazardous areas, the accompanying leather case must be used!
5
3
4
2
1
11
WIKA operating instructions, model CPH62I0
EN
11221801.04 10/2018 EN/DE
3. Safety
DANGER!
Before reusing the hand-held pressure indicator as an intrinsically safe instrument, prior to insertion into the leather case, make a general check for external damage and functionality of the instrument and the leather case!
Ensure that the instrument is not exposed to environments that enable the ingress of humidity, water, conducting liquids or dust.
Only use the 9 V batteries listed, see chapter 3.6.3 “Permitted batteries”!
Only replace the batteries outside the hazardous area, see chapter
8.2 “Battery replacement”!
The permissible ambient temperature is -10 ... + 50 °C.
The optionally available transport and storage case is not approved for use within hazardous areas! This must always be stored outside the hazardous area.
Check whether the classification is suitable for the application. Observe the relevant national regulations.
For applications that require category 2G instruments (potentially explosive gas atmospheres), the following temperature class classification and ambient temperature ranges apply:
Marking Temperature class Ambient temperature range
Model CPH62I0-S1 or CPH62I0-S2 hand-held pressure indicator
II 2G Ex ib IIC T4 T1 ... T4 Ta = -10 ... +50 °C
Reference pressure sensor model CPT62I0
II 2G Ex ib IIC T4 T1 ... T4 Ta = -20 ... +50 °C II 2G Ex ib IIC T4 Gb
3.6.1 Special conditions for use
Batteries
Only use the listed batteries, see chapter 3.6.3 “Permitted batteries”!
Only replace the batteries outside the hazardous area, see chapter 8.2 “Battery replacement”!
Interface
Operation of the serial interface is not permitted in hazardous areas!
The connection and operation of the interfaces are only permitted outside hazardous areas!
Only use the interface cables specified by WIKA!
12
WIKA operating instructions, model CPH62I0
EN
11221801.04 10/2018 EN/DE
3. Safety
Pressure sensor
Only use model CPT62I0 pressure sensors! The use of other pressure sensors can result in the destruction of the measuring instrument and of the pressure sensor.
When using two model CPT62I0 pressure sensors, make sure that these don't rest on surfaces nor are screwed into items which are at a different potential!
For the electrical connection between the model CPT62I0 pressure sensor and the CPH62I0-S1 or CPH62I0-S2 hand-held pressure indicator, use only the original WIKA sensor connection cable intended for this. This also applies to the extension cable, thus reaching a maximum permissible overall cable length of just under 5 m (16.4 ft).
Analogue output
The use of the analogue output is not permitted in hazardous areas!
Only connect passive voltmeters to the analogue output.
Equipotential bonding
All components (pressure sensor, power supply unit, interface, etc.) connected to the instrument must be at the same potential or connected via equipotential bonding.
For use in potentially explosive atmospheres, only connect pressure sensors! With two connected pressure sensors, take care that these are connected to the same electrical potential, or an equipotential bonding exists.
Non-intrinsically safe use
The hand-held pressure indicator can also be used as a non-intrinsically safe instrument for connection to non-intrinsically safe instruments (e.g. interface cables).
In this operating state also, only use approved accessories!
Before reusing the hand-held pressure indicator as an intrinsically safe instrument, prior to insertion into the leather case, make a general check for external damage and functionality of the instrument and the leather case!
Transport and storage cases
The optionally available transport and storage cases are not approved for use within hazardous areas! The cases must always be stored outside the hazardous area.
3.6.2 Mains operation
DANGER! Danger to life due to loss of explosion protection
Non-observance of these instructions and their contents may result in the loss of explosion protection.
Only use model GNG 10/3000 power supply units!
The operation with an external power supply is not permitted in hazardous areas!
13
WIKA operating instructions, model CPH62I0
EN
11221801.04 10/2018 EN/DE
3. Safety
3.6.3 Permitted batteries
DANGER! Danger to life due to loss of explosion protection
Non-observance of these instructions and their contents may result in the loss of explosion protection.
Only use the listed 9 V batteries!
Only replace the batteries outside the hazardous area!
Approved batteries are:
Battery type Manufacturer Battery name
6F22 GB GREENCELL, 9 V (1604G) 6LF22
or
6LR61
GP SUPER alkaline, 9 V (1604A) Duracell DURACELL PLUS, alkaline, 9 V Varta Powerone alkaline, 9 V (No. 4122) Varta INDUSTRIAL, alkaline, 9 V (No. 4022)
14
WIKA operating instructions, model CPH62I0
EN
11221801.04 10/2018 EN/DE
4. Design and function
4. Design and function
4.1 Display
1
Main display: Current measured value for sensor 1
2
Secondary display: Current measured value for sensor 2 or differential value between sensor 1 and sensor 2
3
Logg arrow: Logger is ready Arrow blinking: Automatic recording (Logg CYCL) active
4
Tare arrow: Tare function is activated
5
SL arrow: Altitude correction (sea level) is activated
6
Display arrows for measured value units
7
Indication elements for showing the Min./Max. measured value, difference or hold
6
6
7
1
2
45 3
15
WIKA operating instructions, model CPH62I0
EN
11221801.04 10/2018 EN/DE
4. Design and function
4.2 Function buttons and operation
Pos. Symbol Meaning
7
ON
OFF
On/Off button
Switching the CPH62I0-S1 or CPH62I0-S2 on and off
8
MAX
Display of the respective max. memory value
By pressing the [MAX] button, the maximum value measured will be displayed. Pressing it again hides it. To clear the Max. memory, press the [MAX] button for > 2 seconds.
9
TA R A
Activate the Tare function, zero point correction
Tare function By pressing the [TARA] button, the display will be set to “0”. All measurements from then on are displayed relative to the set tare value. If the tare function is activated, the 'Tare' arrow will be displayed. To deactivate, press and hold the [TARA] button for > 2 seconds.
By activating [TARA] the Min. and Max. memory will be deleted.
Zero point correction (for gauge pressure sensors) When there is no pressure on the pressure ports, the instrument will display a “0”. However, if there is a permanent deviation (when operating in trouble­free ambient conditions), there is a possibility to carry out a permanent zero point correction. In order to carry out a zero point correction, press the [TARA] button for approx. 5 seconds. (Only possible if the display value deviates from the factory calibration by less than 2 %, e.g. 0 ... 25 bar (0 ... 360 psi)
Zero point correction up to 0.5 bar (7.3 psi) possible.
7
12
8
11
9
10
16
WIKA operating instructions, model CPH62I0
EN
11221801.04 10/2018 EN/DE
4. Design and function
Pos. Symbol Meaning
The adjustment can only be carried out if the deviation is less than 500 digits. If a zero point correction has been made, this will be signalled by the message 'nuLL-Corr' being displayed briefly as the instrument is switched on.
Restoring the factory calibration By pressing the [TARA] button for approx. 15 seconds, the factory settings will be restored.
10
QUIT
STORE
Activate hold function or logger functions
(See chapter 6.5 “Operation of the logger function”)
Hold function By pressing the [STORE/QUIT] button, the last measured value will be shown in the lower display. Pressing it again hides the value again (only when logger is deactivated).
Logger function Activated by the [STORE/QUIT] button, only if the logger function has been selected via the main menu (see chapter 6.5 “Operation of the logger function”).
11
MIN
Display the respective Min. memory
By pressing the [MIN] button, the minimum value measured will be displayed. Pressing it again hides it. To clear the Min. memory, press the [MIN] button for > 2 seconds.
12
MENU
SET
Enter configuration
By pressing the [SET/MENU] button for approx. 2 seconds, the settings such as configuration, adjustment, alarm logger and system clock can be accessed.
Differentiation By pressing the [SET/MENU] button, the lower display will show the difference of channel 1 to channel 2 (DIF = CH1 - CH2). Pressing it again undoes this action. (Only with 2-channel version and 2 connected pressure sensors).
Abbreviations, definitions
“XXX” Menu XXX will be selected [XXX] Press button XXX ‚XXX‘ Display of a message 'XXX'
4.3 Electrical connections
On the upper edge of the instrument are located the connection sockets CH1 and CH2 (CH2 only with the 2-channel version) for the connection of model CPT62I0 pressure sensors (see chapter 4.5 “Pressure sensors”), as is the socket for the connection of the WIKA interface cable (see chapter 4.6 “Serial or analogue interface”).
17
WIKA operating instructions, model CPH62I0
EN
11221801.04 10/2018 EN/DE
4. Design and function
The sockets for the connection of the interface can also be used for the function of analogue output. For this, a corresponding analogue connection cable must be used.
The “interface” or “analogue output” operating mode must be congured via menu and aects battery life!
1
Interface connector or optional analogue output
2
Connection channel 1 (only with CPH62I0-S1)
3
Connection channel 2 (only with CPH62I0-S2)
4
Connection channel 1 (only with CPH62I0-S2)
4.4 Voltage supply
DANGER! Danger to life due to loss of explosion protection
Non-observance of these instructions and their contents may result in the loss of explosion protection.
Only use the listed batteries, see chapter 3.6.3 “Permitted batteries”!
Only replace the batteries outside the hazardous area, see chapter
8.2 “Battery replacement”!
The voltage supply of the instrument is made via a 9 V battery. This is included in the scope of delivery. The battery life is approx. 300 hours of continuous operation with one sensor and a measuring rate of 4/s.
1
2
1
34
Model CPH62I0-S1 Model CPH62I0-S2
18
WIKA operating instructions, model CPH62I0
EN
11221801.04 10/2018 EN/DE
The battery indicator lights up
To avoid false readings, replace the batteries.
If “bAt” is displayed in the lower display, the battery has been run down and must be replaced. However, the instrument function is still ensured for a certain time.
If “bAt” is displayed in the upper display, the battery has been completely run down.
If the instrument is not used for a long time, the battery should be removed. The real-time clock has to be set again once the battery has been reconnected.
The battery must only be used in a proper fashion and must be disposed of properly in line with the current, national regulations. When storing the instrument at over 50 °C (122 °F), the battery must be taken out.
4.5 Pressure sensors
DANGER! Danger to life due to loss of explosion protection
Non-observance of these instructions and their contents may result in the loss of explosion protection. If third-party reference pressure sensors are used, they can damage the hand-held pressure indicator and the reference pressure sensor.
Only use model CPT62I0 reference pressure sensors!
Only ever use the original connection cable from WIKA for the operation of CPT62I0 reference pressure sensor.
4.5.1 Available pressure sensors
The hand-held has been designed so that all model CPT62I0 pressure sensors can be connected without the need for any readjustment. A wide range of interchangeable sensors is therefore available, with ranges of up to 1,000 bar (14,500 psi), see 10 “Specifications”.
4. Design and function
19
WIKA operating instructions, model CPH62I0
EN
11221801.04 10/2018 EN/DE
4.5.2 Connecting/changing pressure sensors
CAUTION! Damage to the instrument
For overpressure or gauge sensors, the pressure compensation vent hole is found at the top of the sensor housing.
This vent (with integrated diaphragm) must always remain clear!
Before switching the instrument on, connect the reference pressure sensor, otherwise it may not be correctly identified by the instrument.
1. To connect or change the reference pressure sensor, switch off the instrument.
2. Connect the hand-held pressure indicator and the pressure sensor to each other electrically using a separate sensor connection cable. Use the 7-pin plug contact on the pressure sensor for this.
3. Connect the 7-pin connector to the reference pressure sensor in accordance with the orientation guide and secure it through the connection sleeve. Turn the connection sleeve clockwise without much force.
4. Connect the 6-pin M-DIN connector to the hand-held on CH1 or CH2 in accordance with the orientation guide.
When connecting the sensor connection cable to the hand-held, the pressure sensor's connector might not locate properly in the socket. In this case you should try holding the connector by the bend protection, rather than by the connector sleeve.
Connect the connector without tilt the threads.
If the connector is positioned correctly, it can be plugged in without any significant effort.
When removing the pressure sensor, do not pull on the sensor connection cable, but only on the connector sleeve.
4.6 Serial or analogue interface
DANGER! Danger to life due to loss of explosion protection
Non-observance of these instructions and their contents may result in the loss of explosion protection. Through working in flammable atmospheres, there is a risk of explosion which can cause death.
Operation of the serial or analogue interface is not permitted in hazardous areas!
4. Design and function
20
WIKA operating instructions, model CPH62I0
EN
11221801.04 10/2018 EN/DE
For data transfer to a computer, a galvanically -isolated interface adapter is available. This is suitable for connection to a USB interface (USB driver needed).
The USB interface cable consists of a USB connector (model A) at one end of the cable and a 3.5 mm stereo jack connector at the other end of the cable. The cable is approx. 2 m (6.6 ft) long.
The CPH62I0-S1 or CPH62I0-S2 intrinsically safe version is in a protective leather case (Ex protective cover). The interface connector is located under the Ex protective cover.
4. Design and function
21
WIKA operating instructions, model CPH62I0
EN
11221801.04 10/2018 EN/DE
5. Transport, packaging and storage
5. Transport, packaging and storage
5.1 Transport
Check the hand-held pressure indicator and the reference pressure sensor for any damage that may have been caused by transport. Obvious damage must be reported immediately.
CAUTION! Damage through improper transport
With improper transport, a high level of damage to property can occur.
When unloading packed goods upon delivery as well as during internal transport, proceed carefully and observe the symbols on the packaging.
With internal transport, observe the instructions in chapter 5.2 “Packaging and storage”.
If the instrument is transported from a cold into a warm environment, the formation of condensation may result in instrument malfunction. Before putting it back into operation, wait for the instrument temperature and the room temperature to equalise.
5.2 Packaging and storage
Do not remove packaging until just before mounting. Keep the packaging as it will provide optimum protection during transport (e.g. change in use, sending for repair).
Permissible conditions at the place of storage:
Storage temperature: -20 ... +70 °C (-4 ... +158 °F)
Humidity: 0 ... 95 % relative humidity (non-condensing)
Avoid exposure to the following factors:
Direct sunlight or proximity to hot objects
Mechanical vibration, mechanical shock (putting it down hard)
Soot, vapour, dust and corrosive gases
Store the instrument in its original packaging in a location that fulfils the conditions listed above. If the original packaging is not available, pack and store the instrument as described below:
1. Wrap the instrument in an antistatic plastic film.
2. Place the instrument, along with the shock-absorbent material, in the packaging.
3. If stored for a prolonged period of time (more than 30 days), place a bag containing a desiccant inside the packaging.
22
WIKA operating instructions, model CPH62I0
EN
11221801.04 10/2018 EN/DE
6. Commissioning, operation
6. Commissioning, operation
Personnel: Skilled personnel
DANGER! Danger to life from explosion!
Through working in flammable atmospheres, there is a risk of explosion which can cause death.
Only carry out set-up work in non-hazardous environments!
WARNING! Physical injuries and damage to property and the environment caused by hazardous media
Upon contact with hazardous media (e.g. oxygen, acetylene, flammable or toxic substances), harmful media (e.g. corrosive, toxic, carcinogenic, radioactive), and also with refrigeration plants and compressors, there is a danger of physical injuries and damage to property and the environment. Should a failure occur, aggressive media with extremely high temperature and under high pressure or vacuum may be present at the instrument.
For these media, in addition to all standard regulations, the appropriate existing codes or regulations must also be followed.
6.1 Commissioning
Before switching the instrument on, connect the reference pressure sensor, otherwise it may not be correctly identified by the instrument, see chapter
4.5.2 “Connecting/changing pressure sensors”.
Before switching on connect the reference pressure sensor(s) to the intended female connector of the hand-held and make sure that a fully charged 9 V battery is inserted. The connection sockets are marked on the instrument case with 1 or 2 correspondingly (only with CPH62I0-S2). Next to these are located the serial or analogue interface.
6.2 Operation On turning the instrument on, if the logger function is selected, the integrated clock‘s
time will be displayed briefly. If a zero point correction has been carried out, the display will indicate this by showing “nuLL-Corr”.
23
WIKA operating instructions, model CPH62I0
EN
11221801.04 10/2018 EN/DE
6. Commissioning, operation
After changing the battery the menu for setting the clock is displayed automatically
(‘CLOC’). Check the clock and adjust if necessary (see chapter 6.4.11 “Real-time clock (CLOC)”).
6.3 Menu functions
Menu Parame-
ter
Values Meaning
[MENU]
or
SEt ConF
Set Configuration: General settings
Unit
mbar, bar, ...
Display of unit
1) 2)
SL
oFF / on
Sea level: Switch sea-level correction on or off
1) 2)
ALti
-200 ... +9999
Altitude: Input of altitude above sea level [metre] (only if SL is activated)
1) 2)
rAtE
Rate: Measuring rate (see chapter 6.4.3 “Measurement types (rAtE)”)
1)
SLo
Slow: Slow measurement (4 Hz filtered, low power consumption)
1)
FASt
Fast: Fast measurement, filtered (> 1,000 Hz)
1)
P.dEt
Peak detection: fast measuring rate, unfiltered (> 1,000 Hz)
1)
t.AU6
1 ... 120
Time in seconds, calculated via the averaging function
3)
oFF
Averaging deactivated
3)
P.oFF
1 ... 120
Auto power-off delay in minutes. If no button is pressed and there is no data transfer via the interface, the instrument will switch itself off after this interval.
oFF
Auto power-off function deactivated (continuous operation)
Ovt
oFF
No output function, lowest power consumption
SEr
Instrument output is serial interface
dAC
Instrument output is analogue output 0 ... 1 V
Adr.
01, 11 ... 91
Communications address of the interface (only with Out =
SEr)
dAC.
CH1, CH2 or CH DIF
Measuring input which should be used for the analogue output (only with Out = dAC)
dAC.0
e.g. -5.00 ...+5.00 mbar
Zero point setting for Out = dAC: Input of the measured value at which the analogue output should output 0 V (only for Out = dAC)
dAC.1
e.g. -5.00 ...+5.00 mbar
Scale setting for Out = dAC: Input of the measured value at which the analogue output should output 1 V (only for Out = dAC)
1) If there is data in the logger memory, this menu item cannot be accessed. If these should be changed, the data must first be deleted (see chapter 6.5 “Operation of the logger function”).
2) This menu can only be selected if an appropriate sensor is connected to connection 1. When using a second corresponding sensor on connection 2 then the settings are adopted.
3) If the automatic recording is activated, this menu item cannot be accessed. If these should be changed, the cyclic data logger must first be stopped (see chapter 6.5 “Operation of the logger function”).
24
WIKA operating instructions, model CPH62I0
EN
11221801.04 10/2018 EN/DE
6. Commissioning, operation
Menu Parame-
ter
Values Meaning
[MENU]
or
SEt CAL
Set Calibration: Adjustment of sensors
3)
OFS.1
Sensordep., e.g. -5.00 ... +5.00 mbar
The zero point of sensor 1 will be displaced by this value to compensate deviations of the probe or the measuring instrument.
oFF
Zero point offset is deactivated (= 0.000)
SCL.1
e.g. -5.00 ...+5.00 mbar
The measuring scale of sensor 1 will be changed by this factor [%] to compensate deviations of the probe or the measuring instrument.
oFF
Scale correction factor deactivated (= 0.000)
OFS.2
Sensordep., e.g. -5.00 ... +5.00 mbar
The zero point of sensor 2 will be displaced by this value to compensate for deviations of the probe or the measuring instrument.
oFF
Zero point offset is deactivated (= 0.000)
SCL.2
e.g. -5.00 ...+5.00 mbar
The measuring scale of sensor 2 will be changed by this factor [%] to compensate deviations of the probe or the measuring instrument.
oFF
Scale correction factor deactivated (= 0.000)
SEt AL
.
Set Alarm: Alarm function configuration
4)
AL.
[1,2,DIF]
on
Alarm sensor is on; is indicated acoustically
no.So
Alarm sensor is on; is indicated acoustically
oFF
No alarm function
AL.Lo
[1,2,DIF]
Min-Range ... ... AL.Hi
Min. alarm limit (not when AL.oFF, Sensor min. is the lower display range limit of the connected sensor)
AL.Hi
[1,2,DIF]
AL.Lo ... ... Max-Range
Max. alarm limit (not when AL.oFF, sensor max. is the upper display range limit of the connected sensor)
SEt Lo66
Set Logger: Logger function configuration
1)
Func
CYCL
Cyclic: Logger function “cyclic logger”
Stor
Store: Logger function “individual value logger”
oFF
No logger function
CYCL
1 ... 3600
Cycle time for cyclic logger [seconds]
Lo.Po
on / oFF
Low-power logger with very low current supply (only with cyclic logger and slow measurement)
SEt CLOC
Set Clock: Setting of real-time clock
CLOC
HH:MM
Setting the time hours:minutes
dAtE
TT.MM
Setting the time, day.month
YEAr
YYYY
Setting the year
1) If there is data in the logger memory, this menu item cannot be accessed. If these should be changed, the data must first be deleted (see chapter 6.5 “Operation of the logger function”).
2) This menu can only be selected if an appropriate sensor is connected to connection 1. When using a second corresponding sensor on connection 2 then the settings are adopted.
3) If the automatic recording is activated, this menu item cannot be accessed. If these should be changed, the cyclic data logger must first be stopped (see chapter 6.5 “Operation of the logger function”).
4) If an alarm function limit is crossed (over or under), this is signalled by a “hooting” and a beeping.
25
WIKA operating instructions, model CPH62I0
EN
11221801.04 10/2018 EN/DE
6. Commissioning, operation
MENU
SET
MENU
SET
MENU
SET
MENU
SET
MENU
SET
MENU
SET
MENU
SET
QUIT
STORE
QUIT
STORE
QUIT
STORE
QUIT
STORE
QUIT
STORE
QUIT
STORE
TAR A
TAR A
TAR A
TAR A
TAR A
TAR A
(Press for 2 seconds.)
Read
Lo66
Set
Conf
Set
CAL
Set
AL
Set
Lo66
Set
CLOC
(ring - shift)
Stored data
1)
Recall of the stored individual value logger data, see chapter 6.5 “Operation of the logger function”
Configuration
Unit 2) / sea level 2) / measuring rate 2) / power-off / interface address
Adjustment
3)
Of zero point and span
Alarm
3)
Min./Max. alarm visual
Logger
2)
Change over from hold on the logger function: Individual value (SToR) or cyclic (CYCL)
System-clock
Time / day & month / year
1) Appears only if data has been stored in the individual value logger memory
2) Appears only if no data has been stored in the logger, see chapter 6.5 “Operation of the logger function”
3) If the automatic recording is activated, this menu item cannot be accessed. If these should be changed, the cyclic data logger
must first be stopped (see chapter 6.5 “Operation of the logger function”).
MAX
MIN
TA RA
QUIT
STORE
Set parameter Next parameter Store or quit
Menu tree
26
WIKA operating instructions, model CPH62I0
EN
11221801.04 10/2018 EN/DE
6. Commissioning, operation
6.4 Configuring the instrument
1. Press the [SET/MENU] button for 2 seconds.
The main menu SEt” is accessed.
2. Keep pressing the [SET/MENU] button until the desired function is displayed.
3. With the [TARA] button, select the parameters.
4. With the [MIN] or [MAX] buttons, set the parameter.
5. With the [STORE/QUIT] button, confirm the entry.
6. Press the [SET/MENU] button.
Return to main menu.
6.4.1 Pressure units (unit)
Depending on the measuring range of the current pressure sensor, the pressure value can be displayed in any one of the following units: mbar, bar, Pa, kPa, MPa, mmHg, inHg or psi.
6.4.2 Sea-level (SL) and altitude (Alti) correction for absolute pressure sensor
With a connected absolute pressure sensor, the instrument measures absolute pressure. This is not to be confused with the “air pressure at sea level” given by the weather station. With this pressure indication, the altitude-dependent atmospheric pressure drop is calculated. The instrument is capable of making this air pressure altitude correction.
Setting of SL and Alti
1. Press the [SET/MENU] button for 2 seconds.
The main menu SEt” is accessed.
2. With the [TARA] button, select the SL parameters.
3. With the [MIN] or [MAX] buttons, activate the SL parameter.
4. With the [STORE/QUIT] button, confirm the entry.
5. With the [TARA] button, select the Alti parameters.
6. With the [MIN] or [MAX] buttons, enter the altitude.
7. With the [STORE/QUIT] button, confirm the entry.
8. Press the [SET/MENU] button.
Return to main menu.
The setting is only possible, if an absolute pressure sensor is connected to sensor connection 1.
When the “Sea Level” function is activated, the arrow for 'SL' is indicated below in the display. Once the altitude of the fixed location above sea level has been entered, the instrument displays the absolute pressure at sea level.
27
WIKA operating instructions, model CPH62I0
EN
11221801.04 10/2018 EN/DE
6. Commissioning, operation
When two absolute pressure sensors are connected, the “Sea Level” function for both pressure sensors corresponds to the setting of pressure sensor 1.
6.4.3 Measurement types (rAtE)
The instrument supports three different measurement types for various purposes. Two of these operate with an increased measuring rate of > 1,000 measurements/s.
6.4.3.1 Standard measurement (rAtE-SLo)
The measuring rate is 4 Hz. The averaging and filter functions are active.
Application range
Measurement of slowly changing or static pressures, e.g. calibration, leak testing, atmospheric pressure measurement,
Highest measurement accuracy, insensitive to disturbances, low power consumption.
6.4.3.2 Peak value detection (rAtE-P.dEt)
The measuring rate is > 1,000 Hz and the measuring signal is displayed unfiltered.
Field of application in combination with logger function
Measurement of pressure spikes or fluctuating pressures with a resolution of < 1 ms.
The cyclic logger function records the arithmetic mean value, the highest and the lowest pressure during the chosen time interval.
In this setting, there is an increased power consumption and the measurement is susceptible to interference (also electromagnetic interference).
6.4.3.3 Fast measurement = Fast (rAtE-FASt)
The measuring rate is > 1,000 Hz and the measuring signal is displayed filtered. As a result, it is less sensitive to interference and the short-term spikes are filtered out. Other
than that, this function is identical to rAtE-P.dEt.
28
WIKA operating instructions, model CPH62I0
EN
11221801.04 10/2018 EN/DE
6. Commissioning, operation
6.4.4 Averaging
The averaging function acts on the displayed values (display and interface). It is completely independent of the averaging within the logger function (do not confuse them!). The averaging integrates the measured values during a chosen time interval and then calculates the average display value. The function is independent of the selected measuring rate (fast/slow measurement).
So long as a sufficiently long set time (in seconds) has not yet been measured in order to calculate the mean value, “----” is shown in the display - in the lower display, a 'countdown' is displayed.
During the low-power logger operation, the averaging function is always deactivated.
Function of Min./Max. value memory in conjunction with the averaging function:
If averaging is activated and slow measurement, rAtE-SLo, is selected, the Min./ Max. memory value relates to the average display values.
If averaging is activated and fast measuring function is selected (rAtE-FASt or rAtE-P.dEt), the Min./Max. memory value relates to the internal measured values (> 1,000 Hz measuring rate).
6.4.5 Zero point correction sensor 1 (OFS.1) or sensor 2 (OFS.2)
A zero point offset can be carried out for the respective measurement:
Displayed value = measured value – offset
Default setting
'off' = 0.0, i.e. no correction is made. The zero point correction, together with the slope correction, is mainly used for the adjustment of sensor deviations. The entry is made in the display unit.
6.4.6 Scale correction sensor 1 (SCL.1) and sensor 2 (SCL.2)
The scale of the corresponding measurement can be influenced by this factor (factor is in %):
Displayed value = (measured value – offset) * (1+Scal/100)
Default setting
'off' = 0.000, i.e. no correction is made. The scale correction, together with the zero point correction, is mainly used for the adjustment of sensor deviations.
29
WIKA operating instructions, model CPH62I0
EN
11221801.04 10/2018 EN/DE
6. Commissioning, operation
6.4.7 Power-off function (P.oFF)
If no button is pressed and no serial communication occurs during the switch-off delay, the instrument will automatically switch itself off. The switch-off delay can be set between 1 and 120 min. If P.oFF= oFF then the power-off function is deactivated.
6.4.8 Instrument output (Ovt)
DANGER! Danger to life due to loss of explosion protection
Non-observance of these instructions and their contents may result in the loss of explosion protection. Through working in flammable atmospheres, there is a risk of explosion which can cause death.
Operation of the serial or analogue interface is not permitted in hazardous areas!
The output can be used as a USB or RS-232 interface or as an analogue output (0 ... 1 V).
6.4.9 Analogue output scaling with dAC.0 and dAC.1 (dAC.)
DANGER Danger to life due to loss of explosion protection
Non-observance of these instructions and their contents may result in the loss of explosion protection. Through working in flammable atmospheres, there is a risk of explosion which can cause death.
The use of the analogue output is not permitted in hazardous areas!
WARNING! Damage to property through incorrect measuring instruments
By using incorrect measuring instruments, this damage could occur to the hand-held.
Only connect passive voltmeters to the analogue output.
Analogue output cannot be used during logger recordings.
With dAC.0 and dAC.1 the analogue output can be easily scaled.
It must be ensured that the analogue output is not loaded too heavily, otherwise the output value can be corrupted and the power consumption of the instrument increases correspondingly.
30
WIKA operating instructions, model CPH62I0
EN
11221801.04 10/2018 EN/DE
6. Commissioning, operation
Loads up to approx. 10 kΩ are harmless.
If the display exceeds the value set with dAC.1, then 1 V will be output.
If the display drops below the value set with dAC.0, then 0 V will be output.
In the event of an error (Err.1, Err.2, ----, etc.) an analogue signal slightly over 1 V will be output.
Jack connector wiring
WARNING! Damage to property through incorrect jack connector
Using an incorrect jack connector or incorrect wiring can cause damage to the hand-held.
The 3rd connection must not be used.
Only stereo jack connectors are permitted.
Only use the original connection cable from WIKA.
6.4.10 Alarm (AL.) There are 3 possible settings: off = “AL.oFF”, on = “AL.on”, or on = “AL.no.So”. Under the following conditions, an alarm is given when the alarm function “AL.on” or “AL. no.So” is active:
Value is below lower alarm limit AL.Lo or above upper alarm limit AL.Hi.
Sensor error (Sens-Erro)
Low battery, 'bAt'
Err.7: System error
In the event of an alarm, the display flashes, with interface access the 'PRIO' flag is set.
6.4.11 Real-time clock (CLOC)
The real-time clock is needed for the time allocation of the logger data. If required, check the settings.
Checking the real-time clock
1. Press the [SET/MENU] button for 2 seconds.
The main menu SEt” is accessed.
2. Keep pressing the [SET/MENU] button until SEt-CLOC is displayed.
3. With the [TARA] button, select the CLOCparameters.
4. With the [MIN] or [MAX] buttons, enter the time.
5. With the [TARA] button, select the dAtEparameters.
6. With the [MIN] or [MAX] buttons, enter the day and month.
7. With the [TARA] button, select the YEArparameters.
Loading...
+ 74 hidden pages