In the following chapters detailed information on the hand-held pressure
indicator CPH6200 and its proper use can be found.
Should you require further information, or should there be problems
which are not dealt within detail in the operating instructions, please
contact the address listed on the last page.
Factory calibration of the instrument is according to relevant international standards.
The warranty period for the CPH6200 hand-held pressure indicator is 24
months according to the general terms of supply of ZVEI.
The guarantee is void if the appliance is put to improper use or if the
operating instructions are not observed or if an attempt is made to open
the appliance.
We also point out that the content of these operating instructions
neither forms part of an earlier or existing agreement, assurance or
legal relationship nor is meant to change these. All obligations of WIKA
Alexander Wiegand GmbH & Co. KG result from the respective sales
contract and the general business terms of WIKA Alexander Wiegand
GmbH & Co. KG.
WIKA is a registered trade mark of WIKA Alexander Wiegand GmbH &
Co. KG. Names of companies or products mentioned in this handbook
are registered trade marks of the manufacturer.
We reserve the right to effect reasonable changes on the basis of technical improvements.
Any reproduction of this manual or parts thereof by any means is
prohibited.
Version key regarding firmware and respective manual.
This device has been designed and tested in accordance with the
relevant safety regulations for electronic devices. However, its
trouble-free operation and reliability cannot be guaranteed unless
the standard safety measures and special safety advise given in this
manual is followed when using the device.
1. Trouble-free operation and reliability of the device can only be
guaranteed so long as the device is not subjected to any climatic
conditions other than those stated under “8. Specification“.
2. The device and sensors must be handled with care (don‘t throw, hit,
etc.). Protect plugs and sockets from contamination.
3. If the device is moved from a cold to a warm environment, condensa
tion may cause the equipment to fail. You should therefore ensure the
device temperature has adjusted to the ambient temperature before
trying to switch it on.
4. If the instrument is to be connected to other devices (e.g. via serial
interface) care must be taken when designing the equipment connections. It is possible that internal wiring within the external device (e.g.
connection of GND to Earth) may cause excessive voltages which
could harm or destroy the instrument or other connected devices.
If the device is operated with a faulty mains power supply (e.g.
short circuit from mains voltage to output voltage) this could result
in dangerous voltages at the device (e.g. at the sensor socket or
serial interface).
5. If there is any risk whatsoever involved in using it, the device must be
switched off immediately and marked accordingly to prevent re-use.
GB
-
Operator safety may be at risk if:
There is visible damage to the device
The device is not working as specified
The device has been stored under unsuitable conditions for an
extended period of time.
If there is any doubt, please return the device to the manufacturer for
repair or maintenance.
6. Customers must not attempt to alter or repair the device themselves.
Please return the device to the manufacturer for any repair or maintenance.
7. Any operation not included in the following instructions or outside the
specifications must not be attempted.
If ´bAt´ is shown in the lower display the battery is low and needs to
be replaced or recharged (using the appropriate battery charger as
shown in the current CPH6200 price list ).
The device will, however, work normally for a limited period. If ´bAt´
is shown in the upper display the voltage is too low to operate the
device; the battery will be completely flat.
A) If the device is not going to be used for some time, we advise that
the battery is removed.
B) The real time clock has to be set again after reconnect to the battery
Mains operation with power supply
When using a power supply please note that the operating voltage
must be 10.5 to 12 V DC. Do not use with higher voltages! Cheap
12 V-power supplies often have excessive no-load voltage, and
we therefore recommend using regulated-voltage power supplies.
Trouble-free operation can be ensured by using our own power
supply unit (see current CPH6200 price list). Before connecting the
power supply to the mains make sure that its operating voltage
matches the local mains voltage.
Connecting/Changing sensors
Only use sensors model CPT6200. Using other sensors can damage
the instrument! Switch off device before changing the sensor and
connect the sensor before turning the device back on, otherwise the
sensor may not be identified correctly.
The CPH6200 and the pressure transmitter are connected using a
discrete connecting cable. When changing the pressure transmitter,
always connect and disconnect at the transmitter end of the cable. To
connect the transmitter, rotate the plug till it locates in the guide slot,
allowing it to plug in. Then tighten the safety collar (by screwing it lightly
in a clockwise direction).
When connecting the cable to the CPH6200, the connector might not
locate properly in the socket. In this instance you should try holding the
plug by the cable strain relief, rather than by the connector housing. If
the plug is correctly located, it will slide in smoothly. To disconnect the
sensor hold it by the plug to release the connector lock. Do not pull the
cable to release it.
For gauge sensors, the pressure compensation vent hole is found
at the top of the sensor housing. This vent (with integral membrane
seal) must be kept clear without fail.
Maintenance advice
Both the CPH6200 and sensors are manufactured using solidstate
technology, and contain no moving parts which could wear. If the
instrument housing has been opened, the warranty becomes invalid.
If cleaning is necessary, use a cloth wetted with mild detergent. Avoid
any solvents, caustic or abrasive substances.
As is normal for all measuring instruments, the accuracy of the device
should be tested at regular intervals (approx. once per year - see
Section 7).
1.3 Electrical connections
On the upper edge of the device:
The connector sockets for attaching the CPH series pressure sensors,
CH1 and CH2 (CH2 only with the 2-channel version), are located on the
top edge of the CPH6200 (see Section 5), as is the socket for the WIKA
interface cable (see Section 4).
Connection for WIKA interface cable
or optional analogue output
GB
CH1 CH2 (only for 2-channel version: CPH6200-S2)
Operation as analogue output: connection via suitable cable.
The output mode has to be configured via menu and influences
battery life!
On the left side of the device:
The mains adapter socket for connecting the power supply unit (see
current CPH6200 price list) is located on the left side of the handheld
unit.
1. Main display: shows the current measured value of CH1.
2. Lower display:
For 2-channel version shows the measured value
of CH2 or DIF (CH1-CH2). For 1 channel version it displays the min,
max or hold function values.
3. Logg:
shown if logger function is selected, flashes if cyclic logger is
running.
4. Tare:
5. SL:
6. Measuring unit:
indicates that tare function is activated.
indicates, that the sea level function is activated.
an arrow points to the chosen measuring unit.
1.5 Basic operation
On turning the device on; if the logger function is selected, the
integrated clock‘s time will be displayed briefly. If a zero point adjustment has been carried out, the display will indicate this by showing
"nuLL- Corr".
After changing the battery the menu for setting the clock is activated
automatically ('CLOC'). Check the clock and adjust if necessary (see
Section 2).
2. Set/Menu: Enter Configuration Menu (press for 2 sec)
3. Tare: Activate tare function, zero point adjustment
4. Store/Quit: Activate hold function or logger function (see Section “3.
5. Min/Max: Display the respective min-/max-memory value in lower
Max-Function: Pressing ´Max´ shows the maximum measured value record-
Min-Function: Pressing ´Min´ shows the minimum measured value record-
Hold-Function:
Logger-Functions: Activated by the
Tare-Functions: By pressing ´Tare´, the display will be set to 0. All measure-
Operation of logger”.)
display
ed. Pressing it again hides it. To clear the max-memory press
the ´Max´ key for > 2 seconds.
ed. Pressing it again hides it. To clear the min-memory press
the ´Min´ key for > 2 seconds.
By pressing ´Store/Quit´ the instantaneous measured value
will be shown in the lower display. Pressing it again hides it.
(Only when the main menu item logger = ‘off‘ is selected).
been selected via the main menu (see Section “3. Operation
of logger”).
ments from then on will be displayed relative to the set tare
value. When the tare function is activated, the arrow ´Tare´
appears in the display. To deactivate the tare function press
´Tare´ for > 2 seconds.
´Store/Quit´
key, only if the logger has
Activating/deactivating tare clears the max- and min-memories.
Zero-Point Adjust: (for gauge pressure sensors only) If there is no pressure
applied to the pressure ports the device will display 0. If
there is a permanent offset (and the device is being used
under steady conditions), a permanent zero point adjustment can be carried out. To carry out the adjustment, press
´Tare´ for approx. 5 seconds. (Please note: A zeropoint
adjustment can only be made if the difference between the
display value and the factory calibration value is less than 2
%! E.g. for a measuring range of 0 ... 25 bar, => zeropoint
adjustment up to 0.5 bar possible). To revert to the factory
calibration, press ´Tare´ for approx. 15 seconds.
GB
A zero-point adjustment can only be carried out if the difference
between the value on display is less than 500 digits!
If a zero-point adjustment has been made, this will be signalled by the
message “nuLL-Corr“ being displayed briefly as the device is turned on.
Differential pressure: Pressing ´Set/Menu´ shows the difference between
the measured values of Channel 1 and Channel 2
(DIF=CH1-CH2) in the lower display. Pressing it again will
mbar, bar, ... Unit: Unit of display
oFF / onSea-level correction: on or off
-200 ... +9999 Altitude: Input of altitude above sea-level [m] (only if on)
Rate: Measuring rate (see chapter 2.3)
SloSlow: Measuring rate (4 Hz filtered, low power consumption)
FAStFast: Measuring rate, filtered (> 100 Hz)
P.dEtPeak detection: fast measuring rate, unfildered (> 100 Hz)
1 ... 120Averaging interval in seconds, used by the averaging function off
oFFAveraging function deactivated
1 ... 120Auto Power-Off delay in minutes. Device will automatically switch
oFFAuto Power-Off function inactive (continuous operation)
oFFFunction of the output: No output function, lowest power
SErOutput is serial interface
dACOutput is analogue output 0 ... 1 V
01, 11 ... 91 Communicationsaddress of interface
CH1, CH2 or
itself off if, during this interval, no key is pressed, or there is no
serial communication.
consumption
Choice of the input to be the source for the analogue output (if
Out = dAC)
Enter desired value at which the analogue output potential should
be 0 V (if Out = dAC)
Enter desired value at which the analogue output potential should
be 1 V (if Out = dAC)
onAlarm function on, with audio alarm
no. SoAlarm function on without audio alarm
oFFNo alarm function
Min-Range ...
... AL.Hi
AL.Lo ...
... Max-Range
CYCLCyclic: logger function ‘cyclic logger’
StorStore: logger function ‘discrete value logger’
oFFNo logger function
1 ... 3600Cycle time for cyclic logger [seconds]
on / oFFLow-power logger with very low power consumption (only if
HH:MMClock: set time hours:minutes
TT.MMDate: set time day.month
YYYYYear: set time year
Min alarm limit
(not when AL.oFF, Min-Range is the lower display range of the
connected sensor)
Max alarm limit
(not when AL.oFF, Max-Range is the upper display range of the
connected sensor)
*
*
*
*
*
*
cyclic logger with slow measuring rate)
*
* If there is existing data in the logger memory, the menus/parameters
cannot be accessed! If these need to be adjusted, the logger memory
must first be cleared! (see Section “3. Operation of logger”.)
** This menu can oly be invoked if a refering sensor is connected to
connection 1. When using a second sensor at connection 2 then
changes are taken over.
2. Configuration
To change device settings, press the ´Set/Menu´ key for 2 seconds. This
will access the configuration menu (main display: “SEt“).
Pressing the ´Set/Menu´ key again scrolls down the menus, pressing
the ´Tare´ key scrolls across to the associated parameters, which can
then be selected with the ´Tare´ key.
The parameters can be altered using the ´Min´ or ´Max´ keys. Pressing
the ´Set/Menu´ key again saves the settings and returns you to the main
configuration menu.
The ´Store/Quit´ key exits the configuration and returns you to standard
measuring mode.
Depending on the pressure range, the pressure value can be displayed
in any one of the following units: mbar, bar, Pa, kPa, MPa, mmHg, inHg
or psi.
2.2 (SL) Sea-level correction for absolute pressure sensors
If an absolute pressure sensor is connected to the instrument, then the
device displays the true absolute pressure measured by the sensor. This
does not necessarily equate to the values given by weather stations,
which give the pressure at sea-level. Usually the sensor is situated
above sea-level and therefore, if the value at sea-level (zero) is to be
measured, the pressure loss resulting from the elevation above sealevel must be considered! To correct the displayed value activate the
“Sea-Level-Function“ (SL). (Setting is only possible, if the abs. Pressure
sensor is connected to sensor socket 1.) Then enter the sensor‘s height
above sea-level in metres (Alti). When activated, the display shows the
SL-arrow and the device displays the absolute pressure value relative to
sea-level.
When two absolute pressure sensors are connected, the sea level
function for both is corresponding to the setting of sensor 1.
2.3 (rAtE) Choice of measuring rates: “rAtE-Slo, -P.dEt, -FASt“
GB
Three options of measuring rate are supported. Two of them work with
high measuring frequencies of over 1000 measurements per second.
Whichever one of them was selected during configuration (see above),
this will be displayed in the lower display: “P.dEt“ or “FASt“.
2.3.1 rAtE-Slo: standard measuring
Measuring rate 4 Hz, averaging and filter functions are active.
Application: Measurement of slowly changing or static pressures, e.g.
leak testing, atmospheric pressure...
Highest accuracy, high noise immunity (EMI and unstable measuring
signals), low power consumption.
Measuring rate > 1000 Hz, the value is displayed unfiltered.
Application with logger function: Measuring of pressure spikes or fluctuating pressures with a resolution of < 1.5 ms. The cyclic logger function
records the arithmetic mean value, the highest and the lowest peak
during the chosen time interval.
Higher power consumption, measuring is sensitive to noise (EMI,..).
2.3.3 rAtE-FASt: fast filtered measuring
Measuring rate > 1000 Hz, but the value is filtered slightly (higher noise
immunity than P.dEt, small peaks will be filtered out), apart from that,
identical behaviour to P.dEt.
2.4 Averaging function
The averaging function acts on the displayed values (LCD and serial
interface). It is completely independent of the averaging within the
logger function, please don‘t mix them up!
The averaging integrates the measured values during a chosen time
interval and then calculates the average display value. It is independent
of the selected measuring rate (slow, fast, peak detect).
Until sufficient values to calculate an average value have been collected
(dependant on selected averaging time), the upper display shows “----“,
the lower display shows a ´countdown´.
During an active low-power-logging procedure the average is always
deactivated.
Operation of min/max-value memory in conjunction with the averaging
function:
If averaging is activated and 'slow' measuring rate is selected (rAtE-
Slo), the min-/max-memory value relates to the average display
value.
If averaging is activated and 'fast' measuring rate is selected (rAtE-
FASt or P.dEt), the min-/max-memory value relates to the internal
measured values (pressure spikes can be detected).
2.5 Zero correction sensor 1 (OFS.1) and sensor 2 (OFS.2)
A zero displacement can be carried out for the measured value:
value displayed = value measured - offset
Standard setting: ‘off‘ = 0.0°, i.e. no zero displacement will be carried
out. Together with the scale correction (see below) this factor is mainly
used to compensate for sensor deviations. Input is in the display unit.
2.6 Span correction sensor 1 (SCL.1) and sensor 2 (SCL.2)
The scale of the measuring can be influenced by this setting (factor is in %):
displayed value = measured value * (1+Scal/100)
Standard setting: ‘off‘ =0.000, i.e. value is not corrected. Together with
the zero displacement (see above) this factor is mainly used to compensate for sensor deviations.
2.7 (P.oFF) Power off function
If no key is pressed, and no serial communication occurs during the
interval set in configuration (P.off), the device will automatically switch
itself off to save battery power.
If P.oFF = oFF then the automatic power off is deactivated.
GB
2.8 (Out) Output
The output can be used as interface (RS-232 or USB) or as analogue
output (0 ... 1 V).
2.8.1 (Adr.) Serial communications address
This parameter has been included in preparation for future operation
with multi-point monitoring systems.
2.8.2 (dAC.) Analogue output – scaling with (dAC.0) and (dAC.1)
Analogue output can not be used during logger recordings
With the dAC.0 and dAC.1 values the output can be rapidly scaled to
your efforts.
Keep in mind not to connect low-resistive loads to the output, otherwise the output value will be wrong and battery life is decreased. Loads
above ca 10 kΩ are uncritical. If the display exceeds the value set by
dAC.1, then the device will apply 1 V to the output. If the display falls
below the value set by dAC.0, then the device will apply 0 V to the
output. In case of an error (Err.1, Err.2, no sensor, etc.) the device will
apply slightly above 1 V to the output.
Plug wiring:
The 3rd contact has to be left floating! Only stereo plugs are
allowed!
2.9 (AL.) Alarm
There are three possible settings: Alarm off (AL.oFF), on with audio
alarm (AL.on), on without audio alarm (AL.no.So).
The following conditions will prompt an alarm to be displayed, if the
alarm function is activated (on or no.So):
Value is below lower alarm limit (AL.Lo) or above upper alarm limit
(AL.Hi).
Sensor error (Sens-Erro)
Low battery (bAt)
Fe 7: System error (always with sound)
When an alarm occurs, the alarm arrow flashes, and, during serial
communication, the 'PRIO'-flag is set within the transmitted message.
If the horn sound of one channel will be switched on/off (on or no.So),
then this horn sound setting will automatically be copied to the other
activated channels.
The real time clock is needed for time-stamping each data point recorded within the logger function. As a result, the clock settings should be
checked regularly. If the battery has been replaced the appropriate
menu 'CLOC' will be accessed automatically.
3. Operation of logger
The device supports two different logger functions:
“Func-Stor“: each time the ´Store/Quit´ key is pressed a data point
will be recorded.
“Func-CYCL“: measurements will automatically be recorded at a
defined interval, set within the logger menu ‘CYCL‘ until
the logger is stopped or the logger memory becomes
full. Logging is started by pressing the ´Store/Quit´ key
for 2 seconds.
The logger records up to 3 measurement values for each interval:
Current or mean value (depending on logger setting, see below), Min-
Peak and Max-Peak. (Sensor 1, Sensor 2, difference)
To evaluate the data, WIKA's Data logger evaluation software GSOFT (V
2.3 or higher) must be used. The software also allows easy configuration
and operation of the logger.
GB
If the logger is selected (Func-STOR or Func-CYCL) the hold function is
deactivated; the ´Store/Quit´ key is then used solely for the operation of
the logger functions.
Min- and Max-Peak are, respectively, the minimum and the maximum
measured values during the last measurement interval. Therefore both
the instantaneous pressure values and the measured fluctuations can be
better analysed.
3.1 “Func-Stor“: Storing discrete measurements
In this mode, each time the ´Store/Quit´ key is pressed a measurement and its time stamp will be recorded. The data recorded can be
viewed either on the display (an additional menu item “REAd-LoGG“ is
displayed when scrolling through the Configuration menu) or through the
Sensor 2* Min-Peak, Max-Peak since last data point
Instantaneous difference (Sensor 1 - Sensor 2)* at that data point
Min-Peak, Max-Peak of the difference (Sensor 1 - Sensor 2)*, Min-
Peak, Max-Peak since last data point
Time and date stamp of that data point
* Only for 2-channel version CPH6200-S2
After each recording “St. XX“ will be displayed for a short time. XX
represents the number of the data point.
If the logger memory already contains data:
When the ´Store/Quit´ key is pressed for 2 seconds, the option to clear
the logger memory will be displayed:
clear clear nothing
all data points (cancel menu)
clear the last
data point
Use the ´Min´ or the ´Max´ key to display the required option. The
´Store/Quit´ key selects that choice.
If the logger memory is full, the display will show:
Viewing Recorded Measurements
Within the “LoGG-Stor“ function the measurements can be viewed
directly on the instrument display as well as through a computer (“FuncCYCL“ only allows for the data to be viewed on computer). Press the
´Set/Menu´ key for 2 seconds: The first menu displayed is now “rEAdLoGG“ (read logger data). By then pressing the ´Tare´ key, the last
measurement recorded will be displayed; pressing the ´Tare´ key again
scrolls between the different values related to this data point.
Pressing the ´Min´ key or ´Max´ key selects different data points.
3.2 Func-CYCL“: Automatic datalogging with adjustable
cycle-time
The Logger cycle-time is adjustable (see Configuration). For example,
setting “CYCL“ = 60: A measurement is made every 60 seconds.
When the slow measurement “rAtE-Slo“ is chosen, additionally a low
power function is available: “Lo.Po“.
If “Lo.Po“ is on, the device only will take a measurement at the point of
time of the recording. In between the recordings the measuring shut‘s
down. This decreases the power consumption enormously and therefore is recommended e.g. for long time recordings where no mains
adapter is available.
By pressing ´Store/Quit´ for 2 seconds the logger operation will be
called. The display will show:
By pressing ´Store/Quit´ again the recording will be initiated.
After that the display shows ‘St.XXXX‘ for a short time whenever a
measuring is recorded. XXXX is the number of the measuring 1 ... 4000
or 10000.
If the logger memory is full, the display will show:
The recording automatically will be stopped.
If the Low-Power-Logger-function is selected („Lo.Po = on“) the device
switches itself off as soon as the memory becomes full.
Stopping the logging manually:
By pressing the ´Store/Quit´ key, recording can be manually interrupted.
The following choice then appears:
Stop Do not stop
recording recording
The selection can be made by key: ´Max´ and key: ´Min´.
Key ´Store/Quit´ enters the choice.
If you try to switch off the instrument during a logging cycle, you
will also be asked if you wish to stop recording. The device can only
be switched off after the recording has been stopped!
The Auto-Power-Off-function is deactivated during logging!
When the ´Store/Quit´ key is pressed for 2 seconds, the logger operation will be called.
The display will show:
By pressing the key ´Max´ or key ´Min´ the display will change to:
When ´Store/Quit´ is pressed, the choice for clearing the logger memory
will be displayed:
clear clear nothing
all recordings (cancel menu)
clear the last
recording sequence
The selection can be made by key ´Max´ or key ´Min´.
´Store/Quit´ enters the choice.
GB
4. The serial interface
For data transfer to a computer there are two electrically-isolated interface adapters available for the connection of the CPH6200 to a RS-232
or USB interface (USB-driver necessary).
WIKA‘s Data logger evaluation software GSOFT (see current CPH6200
price list) is used to display the data. To avoid transmission errors, there
are several security features implemented (e.g. CRC).
5. Available pressure sensors
The instrument has been designed so that all sensors model CPT6200
can be connected without the need for any configuration or recalibration. There is therefore a wide range of interchangeable sensors
DKD-certificates - other certificates:
If the instrument requires calibration, we recommend returning it, along with its associated sensors, to the manufacturer. Only the manufacturer is able to verify the instrument‘s
basic settings and configuration, and if necessary correct them.
8. Specifications
Input1: for CPH6200-S1;2: for CPH6200-S2
Pressure range in bar0.10.16 0.25 0.40.61.01.62.54.06.0
Overpressure safety in bar 11.5224510101735
Burst pressure in bar222.42.44.86121220.542
Resolution in mbar0.11
Pressure range in bar10162540601001602504006001000
Overpressure safety in bar 358050801202003205008001200 1500
Burst pressure in bar42962504005508001000 1200 1700 2400 3000
Resolution in bar0.010.11
Type of pressuregauge {absolute 0.25 up to 16 bar & vacuum on request}
Accuracy of the measuring chain
Calibration*factory calibration certificate (optional: DKD calibration certificate)
Displaylarge LCD display for display of 2 values and additional information
Range of displaymax. -19999 up to 99999, depending on sensors used
Unitscan be selected from the following: mbar, bar, Pa, kPa, MPa, mmHg, inHg, psi
CPH6200 Digital Instrument specific details:
Functions via key pressMin-, Max-memory, Hold, Tare, Offset-correction, Logger (Start/Stop)
Functions via MenuMin-, Max-alarm (acoustic**/visual), Sea-level (barom. air pressure), Power-Off-
Data logger
Interface 1) (serial)RS-232 or USB interface via special interface cables
Power supply9V-zinc-carbon battery, alternative**: 9V rechargeable battery or mains supply
Power consumption
Working temperature0 ... 50 °C
Air humidity0 ... 95 % relative humidity without moisture condensation
Storage temperature-20 ... +70 °C
HousingImpact-resistant ABS, membrane keyboard, transparent panels (CPH62I0 with
WeightApprox. 160 g
EMV- / CE-conformityThe CPH6200 device corresponds to the essential protection requirements estab-
1) For the intrinsically safe version, the use of the interface is not permitted within an explosive atmosphere.
*) Calibrated in vertical mounting position with lower pressure connection.
**) Not available for the intrinsically safe version CPH62I0.
11221780.1.3 01/2009 GB/D/F/E
{} Items in curved brackets are optional extras at an additional price.
0.2 % FS +/- 1 digit at reference temperature of 20 °C (optional 0.1 %)
(dependent on range)
Function, measuring rate: 4/sec (“slow“); > 1000/sec (“fast“); > 1000/sec unfiltered (“peak-detect“) [via “peak-detect“ and min-/max-memory: pressure peaks
of 1.5 msec can be detected]
- discrete value-logger: up to 99 data points (incl. time-stamp) via key press
- cyclic-logger: automatic logging of up to 10000 values incl. time.
cycle-time: selectable between 1 ... 3600 seconds
lished in the regulations of the council for the approximation of legislation of the
member countries regarding electromagnetic compatibility (89/336/EWG).
Pressure connectionG½ B; {flush diaphragm (G1 for 0.1 up to 1.6 bar) or various connection
Material wetted partsWetted parts stainless steel, case stainless steel;
Internal transmission fluidSynthetic oil, (only for pressure ranges up to 16 bar or flush
One year stability0.2 % of span at reference conditions
Permissible
- medium temperature
- ambient temperature0 ... 50 °C
- storage temperature-40 ... +80 °C
Compensated range0 ... 70 °C
Temperature coefficients
- mean TC of zero signal0.2 %/10 K
- mean TC of span0.2 %/10 K
Connection to CPH6200via 1 m cable (plug and play); on request: up to 5 m
Weightapprox. 220 g
2) O-ring made of Viton or EPDM for flush diaphragm with integrated cooling element.
3) The oxygen version must not be used with medium temperatures higher than 60 °C.
Cannot be manufactured for absolute pressure ranges < 1 bar abs.
{} Items in curved brackets are optional extras at an additional price.
For accessories such as pressure pumps (pneumatic/hydraulic), flexible
pressure hose, pressure adapters, data logger-evaluation software Gsoft
or calibration software, power supply unit, battery charger, rechargeable
batteries, etc. see current “WIKA Testing and Calibration Technology
pricelist”.
2.5Nullpunktkorrektur Sensor 1 (OFS.1) bzw. Sensor 2 (OFS.2)39
2.6Steigungskorrektur Sensor 1 (SCL.1) bzw. Sensor 2 (SCL.2)39
2.7(P.oFF) Abschaltverzögerung39
2.8(Out) Geräteausgang39
2.8.1(Adr.) Schnittstellen-Adresse39
2.8.2(dAC.) Analogausgang-Skalierung mit dAC.0 und dAC.140
2.9(AL.) Alarm40
2.10(CLOC) Echtzeituhr41
3.Bedienung der Loggerfunktion41
3.1„Func-Stor“: Einzelwerte speichern41
3.2„Func-CYCL“: Autom. Aufzeichnung mit einstellbarer Zykluszeit
4.Die serielle Schnittstelle45
5.Verfügbare Drucksensoren45
6.Fehler- und Systemmeldungen46
7.Hinweis zum Kalibrierservice47
8.Technische Daten47
9.Zubehör49
D
43
11221780.1.3 01/2009 GB/D/F/E
WIKA Betriebsanleitung Hand-Held Druckmessgerät27
D
Hand-Held Druckmessgerät
CPH6200-S1 / CPH6200-S2
11221780.1.3 01/2009 GB/D/F/E
WIKA Betriebsanleitung Hand-Held Druckmessgerät28
1. Allgemeines
In den folgenden Kapiteln erhalten Sie nähere Informationen zum
Hand-Held Messgerät CPH6200 und seinen ordnungsgemäßen Einsatz.
Sollten Sie weitere Informationen wünschen, oder treten besondere
Probleme auf, die in der Betriebsanleitung nicht ausführlich behandelt
werden, erhalten Sie Auskunft unter den auf der letzten Seite aufgelisteten Adressen.
Bei der Werkskalibrierung der Instrumente wurde sich an entsprechende
internationalen Normen orientiert.
Die Gewährleistungszeit für das Hand-Held Messgerät CPH6200
beträgt 24 Monate nach den Allgemeinen Lieferbedingungen des ZVEI.
Sämtliche Gewährleistungsansprüche verfallen, bei unsachgemäßer
Handhabung bzw. bei Nichtbeachtung der Betriebsleitungen oder bei
dem Versuch das Gerät zu öffnen.
Außerdem weisen wir darauf hin, dass der Inhalt dieser Betriebsanleitung nicht Teil einer früheren oder bestehenden Vereinbarung, Zusage
oder Rechtsverhältnisses ist oder diese abändern soll. Sämtliche
Verpflichtungen der WIKA Alexander Wiegand GmbH & Co. KG ergeben
sich aus dem jeweiligen Kaufvertrag und den Allgemeinen Geschäftsbedingungen der WIKA Alexander Wiegand GmbH & Co. KG.
WIKA ist ein eingetragenes Warenzeichen der WIKA Alexander Wiegand
GmbH & Co. KG.
Firmen- oder Produktnamen, die in diesem Handbuch erwähnt werden,
sind eingetragene Warenzeichen dieser Hersteller.
Zumutbare Änderungen aufgrund technischer Verbesserungen behalten
wir uns vor.
Eine Vervielfältigung dieses Handbuches oder Teilen davon ist untersagt.
Firmware - Betriebsanleitung-Versionsschlüssel
ManualFirmware
V 1.1V 4.0 - V 4.9
V 1.2V 5.0 - V 6.0
V 1.3> V 6.0
Dieses Gerät ist gemäß den Sicherheitsbestimmungen für elektronische Messgeräte gebaut und geprüft. Die Funktion und Betriebssicherheit des Gerätes kann nur dann gewährleistet werden, wenn
die nationalen Sicherheits- und Unfallverhütungsvorschriften sowie
die Sicherheitshinweise der Betriebsanleitung beachtet werden.
1. Die einwandfreie Funktion und Betriebssicherheit des Gerätes
kann nur unter den klimatischen Verhältnissen, die im Kapitel „8.
Technische Daten“ spezifiziert sind, eingehalten werden.
2. Gerät und Sensor müssen pfleglich behandelt werden (nicht werfen,
aufschlagen, etc.). Stecker und Buchsen sind vor Verschmutzung zu
schützen.
3. Wird das Gerät von einer kalten in eine warme Umgebung transpor
tiert, so kann durch Kondensatbildung eine Störung der Gerätefunktion eintreten. In diesem Fall muss die Angleichung der Gerätetemperatur an die Raumtemperatur vor einer erneuten Inbetriebnahme
abgewartet werden.
4. Konzipieren Sie die Beschaltung beim Anschluss an andere Geräte
besonders sorgfältig. Unter Umständen können interne Verbindungen in Fremdgeräten (z. B. Verbindung GND mit Erde) zu nicht
erlaubten Spannungspotentialen führen, die das Gerät selbst oder ein
angeschlossenes Gerät in seiner Funktion beeinträchtigen oder sogar
zerstören können.
Bei Betrieb mit einem defekten Netzgerät (z. B. Kurzschluss von
Netzspannung zur Ausgangsspannung) können am Gerät lebensgefährliche Spannungen auftreten!
5. Wenn anzunehmen ist, dass das Gerät nicht mehr gefahrlos betrie
ben werden kann, so ist es außer Betrieb zu setzen und vor einer
Wiederinbetriebnahme durch Kennzeichnung zu sichern.
-
-
D
Die Sicherheit des Benutzers kann durch das Gerät beeinträchtigt sein,
wenn es zum Beispiel:
Sichtbare Schäden aufweist.
Nicht mehr wie vorgeschrieben arbeitet.
Längere Zeit unter ungeeigneten Bedingungen gelagert wurde.
In Zweifelsfällen das Gerät grundsätzlich an den Hersteller zur Reparatur
bzw. Wartung einschicken.
6. Es dürfen am Gerät keine Veränderungen oder Reparaturen vom
Kunden vorgenommen werden. Zur Wartung oder Reparatur muss
das Gerät zum Hersteller eingesandt werden.
7. Ein anderer Betrieb als der in der folgenden Anleitung beschriebene
oder außerhalb der Spezifikationen, ist bestimmungswidrig und muss
11221780.1.3 01/2009 GB/D/F/E
WIKA Betriebsanleitung Hand-Held Druckmessgerät29
deshalb ausgeschlossen werden.
D
Hand-Held Druckmessgerät
CPH6200-S1 / CPH6200-S2
11221780.1.3 01/2009 GB/D/F/E
WIKA Betriebsanleitung Hand-Held Druckmessgerät30
1.2 Betriebs- und Wartungshinweise
Batteriebetrieb
Wird in der unteren Anzeige "bAt" angezeigt, so ist die Batterie
verbraucht und muss erneuert werden bzw. der Akku ist leer und
muss mit dem dafür vorgesehen Akkuladegerät (siehe aktuelle Preisliste CPH6200) geladen werden. Jedoch ist noch für eine gewisse Zeit
die Gerätefunktion gewährleistet. Wird in der oberen Anzeige "bAt"
angezeigt, so ist die Batterie ganz verbraucht. Batterie und Akku
sind nur sachgemäß zu benutzen und ordnungsgemäß den aktuellen,
nationalen Vorschriften zu entsorgen. Bei Lagerung des Gerätes über
50 °C muss die Batterie entnommen werden.
Wird das Gerät längere Zeit nicht benutzt, sollte die Batterie / der Akku
herausgenommen werden.
Die Uhrzeit muss nachher jedoch wieder neu eingestellt werden.
Netzgerätebetrieb
Beim Anschluss eines Netzgerätes muss dessen Spannung
zwischen 10,5 und 12 V DC liegen. Keine Überspannungen anlegen!
Einfache Netzgeräte können eine zu hohe Leerlaufspannung haben,
dies kann zu einer Fehlfunktion bzw. Zerstörung des Gerätes
führen! Wir empfehlen daher unser Netzgerät (siehe aktuelle Preisliste CPH6200) zu verwenden.
Vor dem Verbinden des Netzgerätes mit dem Stromversor-gungsnetz ist sicherzustellen, dass die am Netzgerät angegebene
Betriebsspannung mit der Netzspannung übereinstimmt.
Sensoren anstecken/wechseln
Nur Sensoren des Types CPT6200 verwenden! Mit anderen Sensoren
kann es zur Zerstörung des Messgerätes und des Sensors kommen.
Zum Sensorwechsel Gerät ausschalten. Sensor vor dem Einschalten
des Gerätes anstecken, sonst wird er vom Gerät evtl. nicht richtig
erkannt.
Digitalgerät und Sensor werden mittels eines separaten Verbindungskabels elektrisch miteinander verbunden. Für den Sensorwechsel sollte
bevorzugt der 7-polige Steckkontakt am Sensor benutzt werden.
Zum Anschluss eines Sensors stecken Sie die 7-polige Steckverbindung
am Sensor gemäß der Orientierungsführung zusammen und sichern Sie
diese durch die Überwurfhülse (Überwurfhülse ohne größeren Kraftaufwand in Uhrzeigersinn drehen).
Beim Anstecken des Kabels am Digitalgerät kann es vorkommen, dass
der Stecker nicht richtig in die Buchse einrastet. In diesem Fall den
Stecker beim Anstecken nicht an Steckerhülse sondern am Knickschutz
halten. Stecker nicht verkantet anstecken. Bei richtig angesetztem
Loading...
+ 70 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.