Digital indicator with easily interchangeable reference
pressure sensors (sensor can also be attached externally)
■
Measuring ranges from 0 ... 250 mbar to 0 ... 8,000 bar
■
Accuracy: 0.025 % (incl. calibration certicate)
■
Calibration and pressure-switch test functions
■
Software and complete service cases (incl. pumps)
available
Calibration technology
WIKA data sheet CT 15.01
ProcessCalibrator model CPH6000
Description
Handling
For a solution that always matches the application, there are
many pressure sensors to choose from, with accuracies of
up to 0.025 % and measuring ranges up to 8,000 bar, which
can be interchanged quickly and without tools. In addition to
being xed to the instrument, the reference pressure sensor
can, optionally, be used externally via an approx. 1.2 m long
connecting cable. If the reference pressure sensor on the
instrument is changed, then the digital indicator will recognise
the new measuring range and so this saves the operator from
having to congure it via the menu.
Functionality
In the setup menu there are 3 operating modes to choose
from:
MEASURING, CALIBRATION and PRESSURE SWITCHTEST. In the CALIBRATION and PRESSURE SWITCHTEST modes, a menu assistant supports the user during
each operation and records, for example, the calibration
data from several calibrations or automatically calculates
the switch hysteresis. To power the test item and to read its
measurement signals, there are electrical inputs and outputs
which are protected from adverse conditions in the eld by
captive protection caps.
Software
For the evaluation and documentation of the calibration data
stored in the CPH6000, WIKA-CAL calibration software
is available. Using this software, the data is automatically
transferred into a printable calibration certicate.
Furthermore, WIKA-CAL also oers, over and above
PC-supported calibration, the management of the calibration
and instrument data in an SQL database. For data transfer,
an RS-232 and a USB interface are available.
Complete test and service cases
For maintenance and service applications, various service
case systems are available. These include service cases
with or without pressure generation, charger, test-cable set,
adapter, etc.
Certied accuracy
For each reference pressure sensor, the accuracy is certied
by a factory calibration certicate which accompanies the
instrument. On request, we are also pleased to provide a
DKD/DAkkS calibration certicate for the instrument from our
own DKD/DAkkS laboratory.
Data sheets showing similar products and accessories:
Test pumps, hydraulic; model CPP series; see data sheet CT 91.05
Test pump, pneumatic; model CPP30; see data sheet CT 91.06
Calibration software; WIKA-CAL; see data sheet CT 95.10
Accuracy of the measuring chain0.1 % FS
Types of pressureRelative pressure, {absolute pressure measuring ranges from 0 ... 25 bar abs. and vacuum measuring ranges from
Sensor compatibilityCompatible with model CPT6000 reference pressure sensors
{} Items in curved brackets are optional extras for an additional price.
1) Up to 10 reference pressure sensors can be supported on each instrument (up to 10 calibration data sets).
2) Calibrated at 23 °C and in vertical mounting position, pressure connection facing downwards.
1 reference pressure sensor (interchangeable without tools) 1); optional: external operation via 1.2 m cable
2)
2)
2)
-1 ... +39 bar}
0.15 % FS
2)
Digital indicator model CPH6000
Display
DisplayLarge TFT colour screen for the display of reference and test signals and additional information
Display resolutionup to 6 digits; adjustable
Pressure unitsmbar, bar, psi, Pa, hPa, kPa, MPa, mmHg, cmHg, inHg, mmH
a customer-specic unit (freely selectable dependant on measuring range)
Functions
Operating modeMEASURING, CALIBRATION and SWITCH-TEST
Measuring rate2 values/s
FunctionsMEASURING, CALIBRATION and SWITCH-TEST
Pressure connection≤ 1,000 bar: G ½ B; {various connection adapters on request}
> 1,000 bar: M16 x 1.5 female, with sealing cone
Material
®
Wetted partsStainless steel (with measuring ranges > 25 bar ... ≤ 1,000 bar ElgiloyInternal transmission uidSynthetic oil (only for measuring ranges up to 25 bar) {Halocarbon oil for oxygen variants}
Sensor specications
Accuracy per year≤ 0.025 % FS (only with CPH6000)
Compensated range0 ... 50 °C
Permissible ambient conditions
Medium temperature
1)
-20 ... +80 °C
4)
Operating temperature-20 ... +80 °C
Storage temperature-40 ... +85 °C
4)
Relative humidity0 ... 95 % r. h. (non-condensing)
Case
MaterialStainless steel
Connection to the CPH6000Option: external operation via 1.2 m connection cable (plug-and-play)
Ingress protectionIP 65 (with cable connected)
Dimensionssee technical drawing
Weightapprox. 230 g
{} Items in curved brackets are optional extras for an additional price.
4) For oxygen versions, the medium temperature must not exceed 60 °C.
in addition)
4)
CE conformity, approvals, certicates
CE conformity, CPH6000
EMC directive2004/108/EC, EN 61326 emission (group 1, class B) and interference immunity (portable measuring
equipment)
CE conformity CPT6000
Pressure equipment directive97/23/EC; module A
EMC directive
Approvals
GOST-RImport certicate, Russia
GOSTMetrology/measurement technology, Russia
Certicates
CalibrationStandard: 3.1 calibration certicate per DIN EN 10204
Approvals and certicates, see website
2004/108/EC, EN 61326 emission (group 1, class B) and interference immunity (industrial application)
Option: DKD/DAkkS calibration certicate
Page 3 of 10WIKA data sheet CT 15.01 ∙ 03/2014
Dimensions in mm
Digital indicator CPH6000
Manual quickrelease clamp for
easy changeover of
the sensor.
Reference pressure sensor CPT6000
Ø 40
126
42
29.5
17
203
Ø6
Ø 17.5
G ½ B
27 mm ats
M24 x 1.5
Page 4 of 10WIKA data sheet CT 15.01 ∙ 03/2014
Electrical connections to the digital indicator
Power supply 24 V*
(activated via menu)
Measuring input*
Voltage
Electrical connection for: CALIBRATION
2-wire test item
Example:
mA signal test item without its own power
supply (24 V must be activated via menu)
Measuring input*
Current or switch test
Example:
mA signal test item with its own power
supply available
Mains supply/charge socket
(rechargeable battery will be
charged automatically)
Interface USB/RS-232
* Connection: 4 mm test sockets
+
24 V
supply
-
3-wire test item
Example:
mA signal test item without its own power
supply (24 V must be activated via menu)
Electrical connection for: SWITCH-TEST
Potential-free
switches
Example:
V signal test item with its own power
supply available
+
24 V
supply
-
Types of switch:
Normally closed /
normally open
Low-/High-window
Page 5 of 10WIKA data sheet CT 15.01 ∙ 03/2014
User interface
6
1 Setup menu
2 Selection and entry conrmation
3 One step back
The display of the test item measured value is also possible as an original electrical signal (mA/V)
Applications
■
with/without
display of test
item
B) with external reference pressure sensor
Measurement of operating/process pressures
■
Comparative measurements with test items (power supply
and display for the test item through the CPH6000)
■
Maximum and Minimum memory (e.g. for leak testing)
■
Alarm function for safety testing
Operating mode: CALIBRATION
Features
■
mA
Calibration assistant
■
Programmable altitude correction
■
Recordable calibration temperature
A) Record calibration data
Print
USB/RS232
B) Record calibration data on a PC and print the calibration
certicate
Operating mode: SWITCH-TEST
Applications
■
On-site calibration of pressure sensor and pressure
measuring instruments (without PC)
■
A calibration assistant guides you easily through the
calibration (following DKD/DAkkS). With this, the data
sets, including date and time, are recorded within the
CPH6000. Prior to calibration, the calibration steps can
be entered directly on the instrument or downloaded via
WIKA-CAL software.
PC software available
■
WIKA-CAL calibration software for creating calibration
certicates
Features
■
Switch-test assistant
■
Automatic calculation of the hysteresis
With switch-test assistant
Applications
■
Convenient checking of pressure switches
■
A pressure switch-test assistant conveniently guides the
operator through the checking and automatic calculation
of the switch hysteresis.
Page 7 of 10WIKA data sheet CT 15.01 ∙ 03/2014
Complete test and service cases
Calibration case with model CPH6000
ProcessCalibrator for pressures consisting of:
■
Plastic service case with foam insert
■
ProcessCalibrator model CPH6000
■
Sensor cable for external operation of sensor
■
Test-cable set with connection terminals
■
Charger
■
Interface cable
■
Sealing set
■
Spaces for several CPT6000 reference pressure sensors
Basic version
Basic version incl. pneumatic pressure generation
Available measuring ranges see specications.
Calibration case with model CPH6000
ProcessCalibrator and model CPP30 hand test pump
for pressures -0.95 ... +35 bar consisting of:
■
Transport case with model CPH6000 ProcessCalibrator
■
Model CPP30 pneumatic hand test pump;
-0.95 ... +35 bar
■
Sensor cable for external operation of sensor
■
Test-cable set with connection terminals
■
Charger
■
Interface cable
■
Sealing set
■
Spaces for several CPT6000 reference pressure sensors
Available measuring ranges see specications.
Calibration case with model CPH6000
ProcessCalibrator and model CPP1000-L hand spindle
pump for pressures up to 1,000 bar consisting of:
■
Transport case with model CPH6000 ProcessCalibrator
■
Model CPP1000-L hydraulic hand spindle pump up to
1,000 bar
■
Sensor cable for external operation of sensor
■
Test-cable set with connection terminals
■
Charger
■
Interface cable
■
Sealing set
■
Spaces for several CPT6000 reference pressure sensors
Basic version incl. hydraulic pressure generation
Page 8 of 10WIKA data sheet CT 15.01 ∙ 03/2014
Available measuring ranges see specications.
WIKA-CAL calibration software
Easy and fast creation of a high-quality calibration
certicate
The WIKA-CAL calibration software is used for generat-
ing calibration certicates or logger protocols for pressure
measuring instruments and is available as a demo version for
a cost-free download.
A template helps the user and guides him through the
creation process of a document.
In order to switch from the demo version to a full version of
the respective template, a USB key with the template has to
be purchased.
The pre-installed demo version automatically changes to
the selected full version when the USB key is inserted and is
available as long as the USB key is connected to the computer.
■
Creation of calibration certicates for mechanical and
electronic pressure measuring instruments
■
A calibration assistant guides you through the calibration
■
Automatic generation of the calibration steps
■
Generation of 3.1 certicates in accordance with
DIN EN 10204
■
Creation of logger protocols
■
User-friendly interface
■
Languages: German, English, Italian and more due with
software updates
For further information see data sheet CT 95.10
Calibration certicates can be created with the Cal-Template and logger protocols can be created with the Log-Template.
Cal Demo
Generation of calibration certicates limited to 2 measuring
points, with automatic initiation of pressures via a pressure
controller.
Cal Light
Generation of calibration certicates with no limitations on
measuring points, without automatic initiation of pressures
via a pressure controller.
Log Demo
Creation of data logger test reports, limited to 5 measured
values.
Log
Creation of data logger test reports without limiting the
measured values.
Page 9 of 10WIKA data sheet CT 15.01 ∙ 03/2014
Scope of delivery
■
ProcessCalibrator model CPH6000
■
Battery charger
■
Test-cable set with various connection terminals
■
3.1 calibration certicate per EN 10204
■
Choice of sensors
Accessories
Connection adapters
■
Various pressure adapters
■
MINIMESS® quick-connect process connection system
Power supply
■
Battery charger
Options
■
DKD/DAkkS certied accuracy
■
Sensors for oxygen applications
Connection cables
■
Test-cable set with various connection terminals
■
USB or RS-232 interface cable
Pressure generation
■
Pneumatic test pumps
■
Hydraulic test pumps
Test cases
■
Various calibration cases incl. pressure generation
Software
■
WIKA-CAL calibration software
Ordering information
CPH6000 / Ingress protection / Additional cable for reference pressure sensor / Software / Interface cable / Test pump /
Transport case / Additional order information
CPT6000 / Operating pressure range / Unit / Measuring range / Accuracy / Process connection / Special design features / Type
The specications given in this document represent the state of engineering at the time of publishing.
We reserve the right to make modications to the specications and materials.