WIKA CPG500 Specifications

Digital pressure gauge Model CPG500
Applications
Calibration service companies and service industry
Measurement and control laboratories
Quality assurance
Easy on-site calibration
Special features
Measuring ranges from -1 … +16 bar to 0 … 1,000 bar (-14.5 ... 230 psi to 0 ... 14,500 psi)
Accuracy: 0.25 % (incl. calibration certicate)
Robust case with protective rubber cap
Simple operation using four buttons
Complete service cases incl. pressure generation available
Calibration technology
WIKA data sheet CT 09.01
for further approvals see page 3
Digital pressure gauge model CPG500
Description
General information
The model CPG500 digital pressure gauge enables the measurement and display of pressure values in a single instrument. The accuracy of digital measurement technology and the simplicity of an analogue gauge are combined in this instrument.
Accuracy
The CPG500 oers an accuracy of 0.25 % of span in eight
pressure measuring ranges. Readings can be displayed in
one of ve standard units.
Sampling rate
With a measuring rate of 100 measurements per second, the CPG500 features a very high measuring rate. With this, fast pressure peaks and drops in pressure can be detected. The bar graph display and drag pointer function integrated into the display, as well as retrievable MIN/MAX peak values,
enable eective analysis of the measuring point.
07/2017
Features
In order to prolong the battery life, the CPG500 features an
automatic switch-o function.
With the ZERO function, the display value can be zeroed with the simple press of a button.
An activatable lter function stabilises pressures which uctuate strongly, and enables the pressure value to be read
easily.
Complete test and service cases
For maintenance and service applications, various case systems are available. Service cases with pneumatic or hydraulic pressure generation are available.
Certied accuracy
For each digital pressure gauge, the accuracy is certied by a factory calibration certicate which accompanies the instrument. On request, a DKD/DAkkS calibration certicate
will be provided for this instrument.
Page 1 of 6WIKA data sheet CT 09.01
Data sheets showing similar products: Precision digital pressure gauge; model CPG1500; see data sheet CT 10.51 Digital pressure gauge for general industrial applications; model DG-10; see data sheet PE 81.66 Hand-held pressure indicator; model CPH6300; see data sheet CT 12.01 Test pumps, hydraulic; CPP series; see data sheet CT 91.05 Test pumps, pneumatic; model CPP30; see data sheet CT 91.06
Specications
Model CPG500
Sensor technology
Measuring range -1 ... +16 bar
Resolution 0.001
Measuring range -1 ... +20 bar -1 ... +40 bar 0 ... 60 bar 0 ... 100 bar
Resolution 0.01
Measuring range 0 ... 350 bar 0 ... 700 bar 0 ... 1,000 bar
Resolution 0.1
Measuring range -14,5 ... 230 psi
Resolution 0.001
Measuring range -14,5 ... 290 psi -14,5 ... 580 psi 0 ... 870 psi 0 ... 1,540 psi
Resolution 0.01
Measuring range 0 ... 5,000 psi 0 ... 10,000 psi 0 ... 14,500 psi
Resolution 0.1
Overload safety 3 times; < 25 bar (< 360 psi)
2 times; > 25 bar ... ≤ 600 bar (> 360 psi ... ≤ 8,700 psi)
1.5 times; > 600 bar(> 8,700 psi)
Pressure connection G ¼ up to a max. 40 bar (580 psi)
G ½ from > 40 bar(> 580 psi)
Accuracy 0.25 % FS ±1 digit
Base instrument
Indicator
Display 4 ½ digit, bar graph with drag pointer function, illuminated
Digit height 15 mm (0.59 in)
Rotatable case > 270 °
Dimensions 50 x 34 mm (1.97 x 1.34 in)
Pressure units bar, psi, MPa, kPa, kg/cm²
Functions
Measuring rate 10 ms
Memory MIN/MAX
Autopower Selectable
Zero function Zero point adjustment
Reset Delete MIN/MAX value
Material
Wetted parts
Case Die-cast zinc with TPE protective rubber cap
Voltage supply
Power supply 2 x 1.5 V AA batteries
Battery life approx. 1,500 hrs.
Battery status display Icon in display
Permissible ambient conditions
Operating temperature -10 … +50 °C (14 ... 122 °F)
Medium temperature -20 … +80 °C (-4 ... +176 °F)
Storage temperature -20 … +60 °C (-4 ... +140 °F)
Relative humidity < 85 % r. h. (non-condensing)
1) Only for use with safe media per directive 67/548/EEC (article 2, paragraph 2).
1)
Stainless steel with sealing NBR
Page 2 of 6 WIKA data sheet CT 09.01 ∙ 07/2017
Loading...
+ 4 hidden pages