Calibration service companies and service industry
■
Measurement and control laboratories
■
Quality assurance
■
Easy on-site calibration
Special features
■
Measuring ranges from -1 … +16 bar to 0 … 1,000 bar
(-14.5 ... 230 psi to 0 ... 14,500 psi)
■
Accuracy: 0.25 % (incl. calibration certicate)
■
Robust case with protective rubber cap
■
Simple operation using four buttons
■
Complete service cases incl. pressure generation
available
Calibration technology
WIKA data sheet CT 09.01
for further approvals see
page 3
Digital pressure gauge model CPG500
Description
General information
The model CPG500 digital pressure gauge enables the
measurement and display of pressure values in a single
instrument. The accuracy of digital measurement technology
and the simplicity of an analogue gauge are combined in this
instrument.
Accuracy
The CPG500 oers an accuracy of 0.25 % of span in eight
pressure measuring ranges. Readings can be displayed in
one of ve standard units.
Sampling rate
With a measuring rate of 100 measurements per second,
the CPG500 features a very high measuring rate. With this,
fast pressure peaks and drops in pressure can be detected.
The bar graph display and drag pointer function integrated
into the display, as well as retrievable MIN/MAX peak values,
enable eective analysis of the measuring point.
∙ 07/2017
Features
In order to prolong the battery life, the CPG500 features an
automatic switch-o function.
With the ZERO function, the display value can be zeroed with
the simple press of a button.
An activatable lter function stabilises pressures which
uctuate strongly, and enables the pressure value to be read
easily.
Complete test and service cases
For maintenance and service applications, various case
systems are available. Service cases with pneumatic or
hydraulic pressure generation are available.
Certied accuracy
For each digital pressure gauge, the accuracy is certied
by a factory calibration certicate which accompanies the
instrument. On request, a DKD/DAkkS calibration certicate
will be provided for this instrument.
Page 1 of 6WIKA data sheet CT 09.01
Data sheets showing similar products:
Precision digital pressure gauge; model CPG1500; see data sheet CT 10.51
Digital pressure gauge for general industrial applications; model DG-10; see data sheet PE 81.66
Hand-held pressure indicator; model CPH6300; see data sheet CT 12.01
Test pumps, hydraulic; CPP series; see data sheet CT 91.05
Test pumps, pneumatic; model CPP30; see data sheet CT 91.06
2 times; > 25 bar ... ≤ 600 bar (> 360 psi ... ≤ 8,700 psi)
1.5 times; > 600 bar(> 8,700 psi)
Pressure connectionG ¼ up to a max. 40 bar (580 psi)
G ½ from > 40 bar(> 580 psi)
Accuracy0.25 % FS ±1 digit
Base instrument
Indicator
Display4 ½ digit, bar graph with drag pointer function, illuminated
Digit height15 mm (0.59 in)
Rotatable case> 270 °
Dimensions50 x 34 mm (1.97 x 1.34 in)
Pressure unitsbar, psi, MPa, kPa, kg/cm²
Functions
Measuring rate10 ms
MemoryMIN/MAX
AutopowerSelectable
Zero functionZero point adjustment
ResetDelete MIN/MAX value
Material
Wetted parts
CaseDie-cast zinc with TPE protective rubber cap
Voltage supply
Power supply2 x 1.5 V AA batteries
Battery lifeapprox. 1,500 hrs.
Battery status displayIcon in display
Permissible ambient conditions
Operating temperature-10 … +50 °C (14 ... 122 °F)
Medium temperature-20 … +80 °C (-4 ... +176 °F)
Storage temperature-20 … +60 °C (-4 ... +140 °F)
Relative humidity< 85 % r. h. (non-condensing)
1) Only for use with safe media per directive 67/548/EEC (article 2, paragraph 2).
1)
Stainless steel with sealing NBR
Page 2 of 6WIKA data sheet CT 09.01 ∙ 07/2017
Page 3
Base instrument
Case
Dimensions79 x 79 x 33 mm (3.11 x 3.11 x 1.30 in)
Ingress protectionIP67
Weightapprox. 400 g (0.882 lbs)
Approvals
LogoDescriptionCountry
EU declaration of conformity
■
EMC directive
EN 61326 emission (group 1, class B) and interference immunity (industrial application)
■
Pressure equipment directive
■
RoHS conformity
EAC
■
EMC directive
■
Pressure equipment directive
-MTSCHS
Permission for commissioning
KazInMetr
Metrology, measurement technology
BelGIM
Metrology, measurement technology
UkrSEPRO
Metrology, measurement technology
Uzstandard
Metrology, measurement technology
European Union
Eurasian Economic
Community
Kazakhstan
Kazakhstan
Belarus
Ukraine
Uzbekistan
Certicates
Certicate
CalibrationStandard: 3.1 calibration certicate per DIN EN 10204
Option: DKD/DAkkS calibration certicate
Recommended recalibration interval1 year (dependent on conditions of use)
Approvals and certicates, see website
WIKA data sheet CT 09.01 ∙ 07/2017 Page 3 of 6
Page 4
44.5 (1.75)
89.5 (3.52)
Dimensions in mm (in)
Front view (with protective rubber cap)
SW 27
1)
Side view (with protective rubber cap)
1)
ED sealing
Ø 19 (0.75)
1) Drawing with G ¼ pressure connection
Front foil
9
8
7
12 (0.47)
21.25 (0.84)
Bar graph indicates the current pressure graphically
1
Currently set unit
2
Display of the measuring range or the MIN/MAX value
3
1
2
Clear the MIN/MAX values
4
Conrmation function in the menu
Hold down to access the menu
5
Zero point adjustment carried out by simply pressing
3
4
Change display value to MIN/MAX or FS (Full Scale)
6
With a single press, switch the digital pressure gauge on/
7
o
Press for a while in order to switch the backlighting on
Battery status
8
Pressure indication
9
6
Page 4 of 6WIKA data sheet CT 09.01 ∙ 07/2017
5
Page 5
Complete test and service cases
Basic version incl. pneumatic pressure generation
Calibration case with model CPG500 digital pressure
gauge and model CPP40 hand test pump, for pressures
-0.95 ... +40 bar (13.8 ... 580 psi), consisting of:
■
Plastic service case with foam insert
■
Digital pressure gauge model CPG500
■
Pneumatic hand test pump model CPP40; -0.95 ... +40 bar
(13.8 ... 580 psi)
Available measuring ranges see specications
Calibration case with model CPG500 digital pressure
gauge and model CPP700-H or CPP1000-H hand test
pumps, for pressures 0 ... 700 bar (0 ... 10,000 psi) or
0 ... 1,000 bar (0 ... 14,500 psi), consisting of:
Basic version incl. hydraulic pressure generation
Recommended pressure generation
Pneumatic hand test pump model CPP40
Pressure range:-0.95 ... +40 bar (13.8 ... 580 psi)
Model CPP700-H or model CPP1000-H hydraulic hand test
pumps
■
Plastic service case with foam insert
■
Digital pressure gauge model CPG500
■
Hydraulic hand test pump model CPP700-H or
CPP1000-H, 0 … 700 bar (0 ... 10,000 psi) or
0 ... 1,000 bar (0 .... 14,500 psi)
Available measuring ranges see specications
Pressure range:
0 ... 700 bar or 0 ... 1,000 bar
(0 ... 10,000 psi or 0 .... 14,500 psi)
Further specications see data sheet CT 91.07
WIKA data sheet CT 09.01 ∙ 07/2017 Page 5 of 6
Page 6
Scope of delivery
■
Digital pressure gauge model CPG500
■
Operating instructions
■
3.1 calibration certicate per DIN EN 10204
■
2 x AA batteries
■
Protective rubber cap for case
Option
■
DKD/DAkkS certied accuracy
Accessories
Connection adapters
■
Various pressure adapters
Pressure generation
■
Pneumatic test pumps
■
Hydraulic test pumps
Test case
■
Various calibration cases incl. pressure generation
Ordering information
Model / Unit / Measuring range / Process connection / Test pump / Transport case / Type of certicate / Further approvals /
The specications given in this document represent the state of engineering at the time of publishing.
We reserve the right to make modications to the specications and materials.