WIKA CPG500 Specifications

Page 1
Digital pressure gauge Model CPG500
Applications
Calibration service companies and service industry
Measurement and control laboratories
Quality assurance
Easy on-site calibration
Special features
Measuring ranges from -1 … +16 bar to 0 … 1,000 bar (-14.5 ... 230 psi to 0 ... 14,500 psi)
Accuracy: 0.25 % (incl. calibration certicate)
Robust case with protective rubber cap
Simple operation using four buttons
Complete service cases incl. pressure generation available
Calibration technology
WIKA data sheet CT 09.01
for further approvals see page 3
Digital pressure gauge model CPG500
Description
General information
The model CPG500 digital pressure gauge enables the measurement and display of pressure values in a single instrument. The accuracy of digital measurement technology and the simplicity of an analogue gauge are combined in this instrument.
Accuracy
The CPG500 oers an accuracy of 0.25 % of span in eight
pressure measuring ranges. Readings can be displayed in
one of ve standard units.
Sampling rate
With a measuring rate of 100 measurements per second, the CPG500 features a very high measuring rate. With this, fast pressure peaks and drops in pressure can be detected. The bar graph display and drag pointer function integrated into the display, as well as retrievable MIN/MAX peak values,
enable eective analysis of the measuring point.
07/2017
Features
In order to prolong the battery life, the CPG500 features an
automatic switch-o function.
With the ZERO function, the display value can be zeroed with the simple press of a button.
An activatable lter function stabilises pressures which uctuate strongly, and enables the pressure value to be read
easily.
Complete test and service cases
For maintenance and service applications, various case systems are available. Service cases with pneumatic or hydraulic pressure generation are available.
Certied accuracy
For each digital pressure gauge, the accuracy is certied by a factory calibration certicate which accompanies the instrument. On request, a DKD/DAkkS calibration certicate
will be provided for this instrument.
Page 1 of 6WIKA data sheet CT 09.01
Data sheets showing similar products: Precision digital pressure gauge; model CPG1500; see data sheet CT 10.51 Digital pressure gauge for general industrial applications; model DG-10; see data sheet PE 81.66 Hand-held pressure indicator; model CPH6300; see data sheet CT 12.01 Test pumps, hydraulic; CPP series; see data sheet CT 91.05 Test pumps, pneumatic; model CPP30; see data sheet CT 91.06
Page 2
Specications
Model CPG500
Sensor technology
Measuring range -1 ... +16 bar
Resolution 0.001
Measuring range -1 ... +20 bar -1 ... +40 bar 0 ... 60 bar 0 ... 100 bar
Resolution 0.01
Measuring range 0 ... 350 bar 0 ... 700 bar 0 ... 1,000 bar
Resolution 0.1
Measuring range -14,5 ... 230 psi
Resolution 0.001
Measuring range -14,5 ... 290 psi -14,5 ... 580 psi 0 ... 870 psi 0 ... 1,540 psi
Resolution 0.01
Measuring range 0 ... 5,000 psi 0 ... 10,000 psi 0 ... 14,500 psi
Resolution 0.1
Overload safety 3 times; < 25 bar (< 360 psi)
2 times; > 25 bar ... ≤ 600 bar (> 360 psi ... ≤ 8,700 psi)
1.5 times; > 600 bar(> 8,700 psi)
Pressure connection G ¼ up to a max. 40 bar (580 psi)
G ½ from > 40 bar(> 580 psi)
Accuracy 0.25 % FS ±1 digit
Base instrument
Indicator
Display 4 ½ digit, bar graph with drag pointer function, illuminated
Digit height 15 mm (0.59 in)
Rotatable case > 270 °
Dimensions 50 x 34 mm (1.97 x 1.34 in)
Pressure units bar, psi, MPa, kPa, kg/cm²
Functions
Measuring rate 10 ms
Memory MIN/MAX
Autopower Selectable
Zero function Zero point adjustment
Reset Delete MIN/MAX value
Material
Wetted parts
Case Die-cast zinc with TPE protective rubber cap
Voltage supply
Power supply 2 x 1.5 V AA batteries
Battery life approx. 1,500 hrs.
Battery status display Icon in display
Permissible ambient conditions
Operating temperature -10 … +50 °C (14 ... 122 °F)
Medium temperature -20 … +80 °C (-4 ... +176 °F)
Storage temperature -20 … +60 °C (-4 ... +140 °F)
Relative humidity < 85 % r. h. (non-condensing)
1) Only for use with safe media per directive 67/548/EEC (article 2, paragraph 2).
1)
Stainless steel with sealing NBR
Page 2 of 6 WIKA data sheet CT 09.01 ∙ 07/2017
Page 3
Base instrument
Case
Dimensions 79 x 79 x 33 mm (3.11 x 3.11 x 1.30 in)
Ingress protection IP67
Weight approx. 400 g (0.882 lbs)
Approvals
Logo Description Country
EU declaration of conformity
EMC directive EN 61326 emission (group 1, class B) and interference immunity (industrial application)
Pressure equipment directive
RoHS conformity
EAC
EMC directive
Pressure equipment directive
- MTSCHS
Permission for commissioning
KazInMetr
Metrology, measurement technology
BelGIM
Metrology, measurement technology
UkrSEPRO
Metrology, measurement technology
Uzstandard
Metrology, measurement technology
European Union
Eurasian Economic Community
Kazakhstan
Kazakhstan
Belarus
Ukraine
Uzbekistan
Certicates
Certicate
Calibration Standard: 3.1 calibration certicate per DIN EN 10204
Option: DKD/DAkkS calibration certicate
Recommended recalibration interval 1 year (dependent on conditions of use)
Approvals and certicates, see website
WIKA data sheet CT 09.01 ∙ 07/2017 Page 3 of 6
Page 4
44.5 (1.75)
89.5 (3.52)
Dimensions in mm (in)
Front view (with protective rubber cap)
SW 27
1)
Side view (with protective rubber cap)
1)
ED sealing
Ø 19 (0.75)
1) Drawing with G ¼ pressure connection
Front foil
9
8
7
12 (0.47)
21.25 (0.84)
Bar graph indicates the current pressure graphically
1
Currently set unit
2
Display of the measuring range or the MIN/MAX value
3
1
2
Clear the MIN/MAX values
4
Conrmation function in the menu
Hold down to access the menu
5
Zero point adjustment carried out by simply pressing
3
4
Change display value to MIN/MAX or FS (Full Scale)
6
With a single press, switch the digital pressure gauge on/
7
o
Press for a while in order to switch the backlighting on Battery status
8
Pressure indication
9
6
Page 4 of 6 WIKA data sheet CT 09.01 ∙ 07/2017
5
Page 5
Complete test and service cases
Basic version incl. pneumatic pressure generation
Calibration case with model CPG500 digital pressure gauge and model CPP40 hand test pump, for pressures
-0.95 ... +40 bar (13.8 ... 580 psi), consisting of:
Plastic service case with foam insert
Digital pressure gauge model CPG500
Pneumatic hand test pump model CPP40; -0.95 ... +40 bar (13.8 ... 580 psi)
Available measuring ranges see specications
Calibration case with model CPG500 digital pressure gauge and model CPP700-H or CPP1000-H hand test pumps, for pressures 0 ... 700 bar (0 ... 10,000 psi) or 0 ... 1,000 bar (0 ... 14,500 psi), consisting of:
Basic version incl. hydraulic pressure generation
Recommended pressure generation
Pneumatic hand test pump model CPP40
Pressure range: -0.95 ... +40 bar (13.8 ... 580 psi)
Model CPP700-H or model CPP1000-H hydraulic hand test pumps
Plastic service case with foam insert
Digital pressure gauge model CPG500
Hydraulic hand test pump model CPP700-H or CPP1000-H, 0 … 700 bar (0 ... 10,000 psi) or
0 ... 1,000 bar (0 .... 14,500 psi)
Available measuring ranges see specications
Pressure range:
0 ... 700 bar or 0 ... 1,000 bar
(0 ... 10,000 psi or 0 .... 14,500 psi)
Further specications see data sheet CT 91.07
WIKA data sheet CT 09.01 ∙ 07/2017 Page 5 of 6
Page 6
Scope of delivery
Digital pressure gauge model CPG500
Operating instructions
3.1 calibration certicate per DIN EN 10204
2 x AA batteries
Protective rubber cap for case
Option
DKD/DAkkS certied accuracy
Accessories
Connection adapters
Various pressure adapters
Pressure generation
Pneumatic test pumps
Hydraulic test pumps
Test case
Various calibration cases incl. pressure generation
Ordering information
Model / Unit / Measuring range / Process connection / Test pump / Transport case / Type of certicate / Further approvals /
Additional ordering information
© 11/2011 WIKA Alexander Wiegand SE & Co. KG, all rights reserved.
The specications given in this document represent the state of engineering at the time of publishing. We reserve the right to make modications to the specications and materials.
Page 6 of 6 WIKA data sheet CT 09.01 ∙ 07/2017
WIKA Alexander Wiegand SE & Co. KG
Alexander-Wiegand-Straße 30 63911 Klingenberg/Germany Tel. +49 9372 132-0 Fax +49 9372 132-406 info@wika.de www.wika.de
07/2017 EN
Loading...