This Warning symbol indicates that danger of injury for persons and the
environment and/or considerable material damage (mortal danger, danger of
injury) will occur if the respective safety precautions are not taken.
This Caution symbol indicates danger for the system and material if the
respective safety precautions are not taken.
This Notice symbol does not indicate safety notices but information for a
better understanding of the facts
2 Operating Instructions - CPC8000
High-End Pressure Controller
CPC8000
Table of Contents
1. General Information .............................................................. 6
All products manufactured by Mensor are warranted to be free of defects in workmanship and
materials for a period of two years from the date of shipment. No other express warranty is given, and
no armation of Seller, by words or actions, shall constitute a warranty. SELLER DISCLAIMS ANY
IMPLIED WARRANTIES OF MERCHANTABILITY OR FITNESS FOR ANY PARTICULAR PURPOSES
WHATSOEVER. If any defect in workmanship or material should develop under conditions of normal
use and service within the warranty period, repairs will be made at no charge to the original purchaser,
upon delivery of the product(s) to the factory, shipping charges prepaid. If inspection by Mensor or
its authorized representative reveals that the product was damaged by accident, alteration, misuse,
abuse, faulty installation or other causes beyond the control of Mensor, this warranty does not apply.
The judgment of Mensor will be nal as to all matters concerning condition of the product, the cause and
nature of a defect, and the necessity or manner of repair. Service, repairs or disassembly of the product
in any manner, performed without specic factory permission, voids this warranty.
MENSOR MAKES NO WARRANTY OF ANY KIND WITH REGARD TO THIS MANUAL, INCLUDING,
BUT NOT LIMITED TO, THE IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR
A PARTICULAR PURPOSE. Mensor shall not be liable for errors contained herein or for incidental or
consequential damages in connection with the furnishing, performance, or use of this material.
1.2 Important Notice
The product specications and other information contained in this manual are subject to
change without notice.
Any reproduction of this manual or parts thereof by any means is prohibited.
1.3 FCC Radio Frequency Emission Notice
This equipment has been tested and found to comply with the limits for a Class A digital device,
pursuant to Part 15 of the FCC Rules. These limits are designed to provide reasonable protection against
harmful interference when the equipment is operated in a commercial environment. This equipment
generates, uses, and can radiate radio frequency energy and, if not installed and used in accordance
with the instruction manual, may cause harmful interference to radio communications. Operation of
this equipment in a residential area is likely to cause harmful interference in which case the user will be
required to correct the interference at his or her own expense.
Use shielded cables to connect external devices to this instrument to minimize RF radiation.
All other brand and product names are trademarks or registered trademarks of their respective companies.
1.5 Software License Agreement
This product contains intellectual property, i.e., software programs, that are licensed for use by the end
user/customer (hereinafter “end user”).
This is not a sale of such intellectual property. The end user shall not copy, disassemble, reverse
engineer or de-compile the software program.
The software programs are provided to the end user “as is” without warranty of any kind,
either express or implied, including, but not limited to, warranties of merchantability and
tness for a particular purpose. The entire risk of the quality and performance of the
software program is with the end user.
Mensor and its suppliers shall not be held to any liability for any damages suered or incurred by the
end user (including, but not limited to, general, special, consequential or incidental damages including
damages for loss of business prots, business interruption, loss of business information and the like),
arising from or in connection with the delivery, use or performance of the software program.
1.6 Mensor Service Plus
Mensor has made a concerted eort to provide complete and current information for the proper use of
the equipment. If there are questions regarding this manual, proper use of the equipment, or if you have
problems using your CPC8000 and you don’t nd the answer in this manual, contact either Mensor or
WIKA for personal assistance. We are ready to help.
Mensor WIKA Alexander Wiegand SE & Co. KG
201 Barnes Drive Alexander-Wiegand-Straße 30
San Marcos, Tx 78666 63911 Klingenberg / Germany
tel: 1-512-396-4200 tel: +49 9372/132-5015
1-800-984-4200 (USA only) fax: 1-512-396-1820 fax: +49 9372/132-8767
email: sales@mensor.com email: CTSales@wika.de
tech.support@mensor.com CTService@wika.de
website: www.mensor.com website: www.wika.de
1.6.1 After the Warranty
Mensor’s concern with the performance of this instrument is not limited to the warranty period. We
provide complete repair, calibration and certication services after the warranty for a nominal fee as
explained in Section 10, Maintenance.
Operating Instructions - CPC8000 7
High-End Pressure Controller
CPC8000
1.6.2 Calibration Services
In addition to servicing our own products, Mensor provides complete pressure calibration services up
to 20,000 psi for many pressure instruments. This service includes a Certicate of Compliance and
Calibration and a record of traceability to the pressure standards of the United States National Institute of
Standards and Technology (NIST).
1.6.3 Accreditations
Mensor is registered to ISO 9001:2008. The calibration program at Mensor is accredited by A2LA, as
complying with both the ISO/IEC 17025:2005 and the ANSI/NCSL Z540-1-1994 standards.
1.7 Packaging for Shipment
If the product must be shipped to a dierent location or returned for any reason through a common carrier
it must be packaged properly to minimize the risk of damage.
The recommended method of packing is to place the instrument in a container, surrounded on all sides
with at least four inches of shock attenuation material such as styrofoam peanuts.
If the instrument will be subjected to frequent transport then the optional Transport Case shown in
Section 9, Options, might be an economical solution. This wheeled case is very rugged and provides
complete and long term protection against rough handling.
8 Operating Instructions - CPC8000
High-End Pressure Controller
CPC8000
2. Safety Notices
2.1 User Responsibilities
To ensure safety, the operator must make sure that the following conditions are met:
■ The system is used properly, no dangerous media are used and that all technical specications are
observed.
■ The system is in perfect operating condition.
■ The operation manual is legible and accessible to the user at the system’s location.
■ The system is operated, serviced and repaired only by authorized and qualied personnel.
■ The user receives instruction on industrial safety and environmental protection, and is knowledgeable
of all operating instructions and safety notices relevant to the current procedure.
2.2 General Safety Notices
The system should be operated only by trained personnel who are familiar with this manual
and the operation of the instrument.
A condition for trouble-free and safe operation of this system is proper transport,
proper storage, installation, assembly and proper use as well as careful operation
and maintenance.
Any operation not described in the following instructions should be prohibited. The
system must be handled with care required for an electronic precision instrument
(protect from humidity, impacts, strong magnetic elds, static electricity and
extreme temperatures). Do not insert any objects into the instrument.
The system is powered via the power cable with a voltage that can cause physical
injury. Even after disconnecting the system from the power supply, dangerous
voltages can temporarily occur due to capacitance.
Repairs must be performed only by authorized service personnel.
Additional safety notices are found throughout this manual.
Operating Instructions - CPC8000 9
High-End Pressure Controller
CPC8000
2.3 Warnings and Caution Notices
WARNING: HIGH PRESSURE! High pressure gases are potentially hazardous.
Energy stored in these gases can be released suddenly and with extreme force.
High pressure systems should be assembled and operated only by personnel who
have been trained in proper safety practices.
WARNING: POSSIBLE INJURY! The tubing, valves and other apparatus attached to
the instrument must be adequate for the maximum pressure which will be applied,
otherwise physical injury to the operator or bystanders is possible.
CAUTION: Use the proper pressure medium. Use only clean, dry, noncorrosive
gases. This instrument is not designed for oxygen use.
WARNING: HIGH SOUND LEVELS! Pressures from 1000 psi and up can generate
sound levels above 80 dbA for brief periods when they are exhausted directly to
atmosphere. A muer is provided for connection to the vent port. It is the
operator’s responsibility to measure sound levels at whatever point 1 meter from
the equipment that has the highest sound pressure level. At levels above 80 dbA,
use of protective ear pieces can reduce these higher levels to a safe level.
WARNING: NOT EXPLOSION PROOF! Installation of this instrument in an area
requiring devices rated as intrinsically safe is not recommended.
WARNING: Detachable main power supply cord with inadequate ratings should not
be used. See Section 4.0, Specications, for power supply ratings.
CAUTION: ESD PROTECTION REQUIRED. The proper use of grounded work
surfaces and personal wrist straps are required when coming into contact with
exposed circuits (printed circuit boards) to prevent static discharge damage to
sensitive electronic components.
Additional Warning and Caution notes are included throughout this manual.
10 Operating Instructions - CPC8000
High-End Pressure Controller
CPC8000
3. Product Description
3.1 General Description
The CPC8000 High-End Pressure Controller is our highest performance pressure controller and features
the following:
• 19” rack mount compatibility
• Designed for serviceability and reliability
• Widescreen display with glass capacitive touchscreen
• Front panel door gives access to instrument interior
• Up to three removable transducers 0.01% IS-50 (IntelliScale-50)
• Pressure ranges from 10.36 pressure regulator, up to 10:1 turndown
• Adaptive control algorithm with rate control capability
• Ethernet, RS-232, USB, and IEEE-488 communications
• Remote compatibility with PCS 400, CPC 6000, CPC 3000, old CPC8000 (SCPI)
• Emulation of competitive instrument command sets
The CPC8000 is designed to automate the testing and calibration of all types of pressure devices and
instruments. Up to three removable / interchangeable pressure transducers are available in full scale
(FS) ranges from 0.36 to 6,000 psig, 7.5 to 6015 psia. Each transducer module is congured with its own
calibration parameters on board and have 0.01% IS-50 or optional 0.008% IS-33 accuracy.
The three transducers, in combination with the internal control valve regulator, provide a dynamic output.
The operator can choose to control pressure either using a single, selected transducer or auto-range
control across all three transducers. The ratio between the highest FS range and the lowest FS range
within the CPC8000 cannot exceed 10:1. The three ranges can be chosen to optimise accuracy levels
across the full pressure span of the instrument.
In addition to the capacity for three active ranges, a fourth, barometric transducer is available as an
option. With this option installed a CPC8000, with absolute or gauge pressure transducers, can emulate
pressure of the opposite type.
3.1.1 IntelliScale Transducer Choices
The CPC8000 can be supplied with standard 0.01% IS-50 Transducers or optional 0.008% IS-33 transducers to give a percent of reading accuracy down to 50% or 33% of full scale respectively. Refer to Sec-
tion 4, Specications, for a full description of the accuracy specication and Section 4.7, IntelliScale.
Operating Instructions - CPC8000 11
High-End Pressure Controller
CPC8000
3.2 Front Panel
The front of the CPC8000 has a smooth, uncluttered appearance (see gure 3.2). Its main feature is the
large color display plus an identication label in the lower right corner of the display panel. The display
panel is hinged for easy access to remove or replace the transducer modules inside. The instructions for
accessing the transducers are provided in Section 10.2.2 Transducer removal/replacement. Immediately
to the right of the display panel is a USB port and a power switch.
Hinged Display Panel
Label
USB port
Power switch
Figure 3.2 - Front View
3.2.1 Power Switch
The power switch is a two-state device with an action similar to that of a ball point pen. Push the button
with enough force to latch it in to turn the unit ON. Push it again to release it to turn the system OFF.
If power to the instrument is interrupted while ON it will shut down until the power is
restored, then immediately resume operation.
3.2.2 USB Port
The front panel USB port is the same as the Host USB port on the rear panel. Both are intended for future
expansions or software upgrades
3.2.3 Information Label
The front panel label identies the instrument name, installed transducer ranges in the pressure units
specied by the customer, and the serial number.
3.2.4 Screen
The large (9 inch diagonal) LCD color screen has a glass capacitive touch screen for navigation within
the intuitive operator interface. At power up the main screen is presented. (See gure 3.2.4.1)
3.2.4.1 Main Screen Features
On power up, the main screen will show a numeric data entry key pad on the right 1/3 of the display and
pressure range, control and operating parameters on the left 2/3.
12 Operating Instructions - CPC8000
High-End Pressure Controller
CPC8000
The data entry key pad can be switched between 4 dierent methods for data entry. See Section 6,
Local Operation, for more details. The data entry portion of the main screen is used primarily to input and
adjust control set points and change main screen parameters.
The left 2/3 of the main screen (see gure 3.2.4.1) contains the reading of the internal pressure transducer, current conguration settings, range limits and labels (see Section 6, Local Operation, for more
details). Pressure monitoring calibration and testing functions are performed through Interaction with this
screen; the majority of an operator’s local operation time will be spent working with this screen. A map of
the individual features on the various displays is included in Section 6.1.2.
Main screen
Test / Calibration Parameters
Figure 3.2.4.1 - Main Screen Features
3.2.4.2 Screen Navigation
Individual screens are accessed by a series of touches on tabs, buttons or icons, which open screens
within the hierarchy of the instrument. Nomenclature that designated a screen and how to navigate to
it will be given in this manual in the following structure: “Icon name” [Icon Graphic] / [Tab] / [Button Name].
Data entry
For example the screen used to set the resolution of the instrument transducer would be designated by
the navigational screen hierarchy, starting from the main screen, as follows: Setup [ ] / [Sensor]
/ [Resolution]. This convention is similar to a computer folder system and will be used to describe a
screen and, inherently, how to navigate to it.
All of the screens are set up into two display areas. The larger area (left 2/3 of the screen) shows live
information about the current conguration, the progress of the test being performed and gives an indication of the output of the device. The smaller area (Data Entry) on the right side, has the number key pad,
variable sliders or discrete choice buttons (radio buttons) that all act to input alpha numeric information or
mode changes relating to the test or calibration information being shown on the left.
Operating Instructions - CPC8000 13
High-End Pressure Controller
CPC8000
Buttons on the screen are shown with a gradient background. When inactive the buttons are shown
with the gradient dark on the bottom and light on the top; when selected the button changes color and is
light on the bottom and dark on the top. Buttons are labeled in an intuitive fashion and when selected its
function is obvious.
The setup screens are accessed by pressing the double gear icon [ ] and follows the same format
as the main screen with the right hand side used for data entry and the left hand side used for display.
These setup screens are used for a variety of instrument settings, including, display, transducer, control,
and remote interface settings. This area also contains the application used to calibrate the internal transducers plus various administrative, information and conguration settings. Detailed information about all
setup screens can be found in Section 6, Local Operation.
Touch the left pointing arrow [ ◀ ] in the bottom left corner of any setup screen to return to the main
screen.
Full operational descriptions of all of the screens, symbols, labels and switches are provided in Section
6, Local Operation.
3.3 Main Features
■ Up to three, highly stable removable / interchangeable pressure transducers (see Section 10.2.2,
Transducer Removal/Replacement). Each transducer is a totally self-contained module including its
unique calibration data.
■ IntelliScale calibrations for maximum accuracy specifications.
■ An optional high performance Barometer module can be used for accurate emulation of gauge
pressure with absolute pressure instruments, or used for absolute pressure emulation with gauge
pressure instruments.
■ A large 9- inch diagonal color LCD with glass capacitive touch screen for intuitive operator interface.
■ A very quiet precision pressure regulator with a fast response time.
■ Multiple languages; simply touch one of the national flags on the display to instantly change the
on-screen text language, and the corresponding number and date formats. For more detailed information see Section 7.1.1, Language.
14 Operating Instructions - CPC8000
High-End Pressure Controller
CPC8000
4. Specications
Accuracy specications presented herein are obtained by comparison with primary standards traceable to
the National Institute of Standards and Technology NIST). A calibration data report has been provided with
your new CPC8000 instrument, traceable to NIST. The calibration program at Mensor is accredited to both
ISO/IEC 17025:2005 and Z540-1-1994 by A2LA. Mensor, LP is registered to ISO 9001:2008.
Mensor reserves the right to change these specications without notice.
4.1 Measure Specications
Reference pressure transducer model CPR8000/ CPR8800
Accuracy
(1)
0.01% FS
(3)
, 0.01% IS-50
(2)
-or-
(4)
(5)
.
(6)
Pressure ranges
Optional 0.008% IS-33
Optional 0.008% IS-50
0 ... 7.5 psia to 0 ... 6015 psia
0 ... 5 psig to 0 ... 6000 psig
-15 ... 145 psig to -15 ... 6000 psig
See Table 4.6.1 for available Transducer Type, Range, Mode, Accuracy and Calibration interval
Maximum ratio between highest and lowest range is 10:1
Precision
(7)
0.004% FS
Measurement unitspsi, inHg@0C, cmHg@0C, mmHg@0C, inH2O@4C, 20C, bar, mbar, cmH2O@4C and 20C, hPa, kPa,
Pa, MPa, kg/cm sq, and 2 user-dened units. All above in rate (per second, per minute and per hour).
ResolutionOperator selectable from 4 to 7 digits depending on the range, units and selected transducer
Calibration adjustmentsZero adder and Span multiplier or up to 11 point linearization for each transducer
Calibration interval
365 days
(8)
Calibration data storageThe calibration data is stored on the transducer.
Measurement ltersO, low, normal (default), high
Optional barometric reference
Function
The barometric reference can be used to switch pressure types
(9)
, absolute <=> gauge. With gauge
pressure transducers, the measuring range of the transducers must begin with -1 bar (-15 psi) in
order to carry out a complete absolute pressure emulation
Measuring range8 ... 17 psia
Accuracy
Precision
(1) It is dened by the total measurement uncertainty, with the coverage factor (k = 2) and includes the intrinsic performance of the instrument, the measurement uncertainty of the reference instrument, long-term stability, inuence of ambient conditions, drift and temperature eects over the compensated range with recommended zero point adjustment every 30 days
(2) 0.01 % IS-50 accuracy: Between 0 ... 50 % of the full scale, the accuracy is 0.01% of half of the full scale value and between 50 ... 100 % of the full scale, the accuracy is 0.01 % of reading
(3) FS = full span
(4) 0.008 % IS-33 accuracy: Between 0 ... 33 % of the full scale, the accuracy is 0.008% of one third of the full scale value and between 33 ... 100 % of the full scale, the accuracy is 0.008 % of reading
(5) 0.008 % IS-50 accuracy: Between 0 ... 50 % of the full scale, the accuracy is 0.008% of half of the full scale value and between 50 ... 100 % of the full scale, the accuracy is 0.008 % of reading
(6) The minimum calibrated range of absolute transducer(s) is 600 mTorr.
(7) It is dened as the combined eects of linearity, repetability and hysteresis throughout the stated compensated temperature range.
(8) 365 days for 0.01% IS-50, 0.008% IS-33 and 0.008% FS accuracy. For 0.01% FS accuracy, 180 days for pressure ranges below 1 bar (15 psi) gauge or absolute, and -1 ...1 bar (-15 ... 14.5 psi)
bidirectional. 365 days for the remainder of the specied ranges
(9) For a pressure type emulation, we recommend a native absolute pressure transducer, since the zero point drift can be eliminated through a zero point adjustment.
(1)
(7)
.
0.01 % of reading
0.004% FS
.
.
.
.
Operating Instructions - CPC8000 15
High-End Pressure Controller
CPC8000
4.2 Control Specications
External pressure
requirements
Control stability0.002% FS of highest range transducer at the set point
Pressure range ratioMaximum ratio between highest and lowest range is 10:1
Minimum control pressure0.05% FS of highest range transducer or .025 psi over exhaust pressure, whichever is greater
Pressure control rates0.001% of range/sec to 10% of range/sec
Control time60 seconds to stable ag for a 10% pressure change into 150cc volume. Larger volumes can
Supply consumption<2.5 scfh in steady-state control.
Overshoot<0.1% FS of highest range transducer
External volume50 to 300 cc. (Consult factory for volumes exceeding 300 cc). Requires one complete pressure
4.3 General Specications
Source Requirements – 10% over range of highest pressure transducer
Exhaust Requirements – Vacuum source required for any control below atmosphere
lengthen this time. Controlling to pressures less than 0.5 psia will lengthen this time.
cycle after changing the external volume, or after a power cycle. See page 27, “Turning on the
CPC8000”.
SizeSee Section 5.3, Dimensions.
Weight~49 lbs (~22.2 kg) with all internal options.
MountingStandard: desk top case with bezel and handle.
CE markConformity certicate. Complies with EN50081, EN50082, and EN61010-1.
Calibration certicateIncludes calibration certicate(s). The Mensor calibration laboratory is accredited in accordance
with the recognized Internal Standard ISO/IEC 17025:2005 and also meets the requirements
of ANSI/NCSL Z540-1-1994. Accreditation is by the American Association for Laboratory
Accreditation (A2LA).
WarrantyTwo year warranty
Options0.008% IS-33, IS-50 transducer option
Transport/travel case
Barometric reference transducer
Spare Primary, Secondary or Tertiary transducer
19” rackmount kit
WARNING! This is class A equipment for emissions and is intended for use in industrial
environments. In other environments, e.g. residential or commercial installations, it can
interfere with other equipment under certain conditions. In such circumstances the operator
is expected to take the appropriate measures.
Operating Instructions - CPC8000 17
High-End Pressure Controller
CPC8000
4.4 Transducer Removal/Replacement
When replacing transducers the order of position MUST be maintained according to
the pressure limit of each transducer; starting with the highest pressure transducer
on the left (Primary Transducer), and the next lower pressure limit transducer next
(Secondary Transducer), and the low ranged transducer at the far right (Tertiary
Transducer). If the system will be operated with less than the full complement of
three pressure modules, any empty slots must begin at the far right where the
Tertiary Transducer is normally located.
The manifold seal is a metal plate used to seal the pneumatic openings on the manifold at
an unused transducer slot. If there is just a single pressure transducer installed then two
manifold seals are installed. All transducer slots must be sealed with either transducers or
seal plates for the instrument to function. Two manifold seal plates are shipped with each
unit for this purpose.
The D-sub connectors for unused slots can remain unplugged. For more information on transducer removal/replacement see Section 10.2.2 in the Maintenance section.
Up to three transducers can be present within the CPC8000. The full scale pressure of the highest
ranged transducer (the Primary transducer) must fall within the range of the installed regulator module
(see Table 4.5, Regulator Module version and range limits). This primary transducer occupies the leftmost
slot within the CPC8000 (see gure 4.4). The other two slots can be empty or occupied with lower ranged
transducers but their full scale range cannot exceed a ratio of 10:1 of the primary transducer. In other
words, the lowest full scale range in the CPC8000 cannot be less than one tenth of the range of the
primary transducer full scale range (see Section 4.6, Transducer Choices).
Figure 4.4 - Front cut-away view showing three transducers
18 Operating Instructions - CPC8000
High-End Pressure Controller
Regulator Module Range
CPC8000
4.5 Regulator Ranges
The CPC8000 is supplied with one of ve dierent needle valve regulators (NVR) modules. Table 4.5
shows each regulator and its corresponding pressure range limits: LP (low pressure), MP (medium
pressure), SP (standard pressure), HP (high pressure) and EP (extended pressure).
Table 4.5 - Regulator Module Version and Range Limits
Regulator ModulePrimary Range Limits
LP-NVR≤ 90 psi gauge or 105 psi absolute / 6 bar gauge or 7 bar absolute
MP-NVR≤ 1000 psi gauge or 1015 psi absolute / 70 bar gauge or 71 bar absolute
> 90 psi gauge or 105 psi absolute / 6 bar gauge or 7 bar absolute
SP-NVR≤ 2000 psi gauge or 2015 psi absolute / 135 bar gauge or 136 bar absolute
> 1000 psi gauge or 1015 psi absolute / 70 bar gauge or 71 bar absolute
HP-NVR≤ 3000 psi gauge or 3015 psi absolute / 210 bar gauge or 211 bar absolute
> 2000 psi gauge or 2015 psi absolute / 135 bar gauge or 136 bar absolute
EP-NVR≤ 6000 psi gauge or 6015 psi absolute / 400 bar gauge or 401 bar absolute
> 3000 psi gauge or 3015 psi absolute / 210 bar gauge or 211 bar absolute
6000 to 3000 psi
EP-NVR
6000
5000
4000
2000 to 3000 psi
3000
HP-NVR
1000 to 2000 psi
SP-NVR
2000
1000 to 90 psi
1000
MP-NVR
90 to 0 psi
LP-NVR
0
Figure 4.5 - Primary transducer limit for each regulator
The primary transducer (Transducer 1) full scale range must fall within one of the regulator range limits
shown in gure 4.5. The secondary and tertiary transducers may fall outside of these limits but must not
have a full scale value less than 1/10 of the primary transducer full scale range.
Operating Instructions - CPC8000 19
High-End Pressure Controller
CPC8000
4.6 Transducer Choices
Transducer measuring modes can be gauge, absolute, or bi-directional. The CPC8000 will be equipped
with one, two or three transducers of the same mode. Transducer type will be “Standard” (accuracy of
0.01% FS, or 0.01% IS-50) or “Premium” (accuracy of 0.008% IS-33, or 0.008% IS-50). The transducers
must be all “Standard” transducers or all “Premium” transducers and not a mix of both types.
The three transducers, in combination with NVR Module, provide a wide dynamic range (10:1 FS) where
the operator can choose to control output pressure using a single transducer, or “autorange” control
across all three transducers. The three transducer ranges can be chosen to optimize accuracy levels.
Full scale transducer ranges up to 6000 psi are available plus an optional barometric transducer. The
accuracy specications, plus a proprietary needle valve regulator, provides accurate and stable control.
The number of unique readings/sec over the remote interface is dependent on the transducer type. The
transducer update rate of a standard transducer is 30ms, or 33 unique readings/sec. The transducer
update rate of a premium transducer is 100ms, or 10 unique readings/sec. This is a technical limitation
of the premium transducer required to meet performance specications. All remote protocols can query
faster than the transducer update rate.
An optional precision barometer can be used for emulation of gauge pressure with an absolute pressure
instrument, or used for absolute pressure emulation with a gauge pressure instrument. For full range
absolute emulation, the instrument gauge transducers should be ranged down to -15 psig.
Table 4.6 - Transducer Choices
Transducer 1
CPR8000 Standard Transducer0.01% FS or 0.01% IS-50
CPR8800 Premium Transducer0.008% IS-33 or 0.008% IS-50
Transducer 2
No second reference transducer
CPR8000 Standard Transducer0.01% FS or 0.01% IS-50
CPR8800 Premium Transducer0.008% IS-33 or 0.008% IS-50
Transducer 3
No third reference transducer
CPR8000 Standard Transducer0.01% FS or 0.01% IS-50
CPR8800 Premium Transducer0.008% IS-33 or 0.008% IS-50
20 Operating Instructions - CPC8000
High-End Pressure Controller
CPC8000
4.6.1 Transducer Type / Range / Mode / Accuracy
Each transducer within the CPC8000 or additional transducers will conform to the limits shown in the following table.
Table 4.6.1 - Transducer Type / Range / Mode / Accuracy
Standard Transducer
(Model CPR8000) Ranges
0 ... 5 to 0 ... <14.5 psigGauge0.01% FS – 180 days
0 ... ≥14.5 to 0 ... 6000 psigGauge0.01% IS-50 – 365 days
0 ... 7.5 to 0 ... <14.5 psiaAbsolute0.01% FS – 180 days
0 ... ≥14.5 to 0 ... 6015 psiaAbsolute0.01% IS-50 – 365 days
Min spanMax span
-5 ... 5 psi-15... +14.5 psiBidirectional0.01% FS – 180 days
0 ... 12 to 0 ... 16.5 psiaAbsolute0.008% IS-33 – 365 days
0 ... 18.4 to 0 ... 33 psiaAbsolute0.008% IS-33 – 365 days
0 ... 36 to 0 ... 50 psiaAbsolute0.008% IS-33 – 365 days
0 ... 80 to 0 ... 110 psiaAbsolute0.008% IS-33 – 365 days
0 ... 160 to 0 ... 220 psiaAbsolute0.008% IS-33 – 365 days
0 ... 240 to 0 ... 500 psiaAbsolute0.008% IS-33 – 365 days
0 ... 700 to 0 ... 1100 psiaAbsolute0.008% IS-50 – 365 days
0 ... 1400 to 0 ... 3300 psiaAbsolute0.008% IS-50 – 365 days
0 ... 4200 to 0 ... 6015 psiaAbsolute0.008% IS-50 – 365 days
ModeAccuracy
ModeAccuracy
Operating Instructions - CPC8000 21
High-End Pressure Controller
CPC8000
4.7 IntelliScale
IntelliScale is a calibration technique that provides another degree of certainty to a pressure
measurement specication. Or said another way, IntelliScale is designed to further reduce the accuracy
specication. It does this by considering the full pressure range as two separate ranges; a specically
dened lower range, and the remaining upper range. It then denes the accuracy of the lower portion
as a percent of the full scale of that lower range, and the accuracy of the upper portion as a percent of
reading at any pressure point within that upper range.
In practice, the standard Mensor IntelliScale specication is 0.01% IS-50. This means that the full scale
lower half (50% of the total range) has an accuracy of 0.01% of that portion of the range, while the upper
half of the total range has an accuracy of 0.01% of the pressure reading. Thus, any pressure within the
lower half of the pressure range has a xed number for the accuracy (.01% of that half scale), while the
accuracy anywhere in the upper half of the full range is a sliding scale number; that is, a percent (0.01%)
of any reading.
For example, a device with a pressure range from 0 to 100 psi with an accuracy specication of IntelliScale 0.01% IS-50 will have an accuracy of 0.005 psi (0.01% x 50 psi FS) on any pressure from 0 psi to 50
psi, and an accuracy of 0.01% of reading (0.01% x R) for any pressure above 50 psi (see gure 4.7.1a).
4.7.1 Optional IntelliScale
An option available for some ranges is IntelliScale 0.008% IS-33, where the lower portion of the full scale
range is 33% of the total range, and the upper portion of the full scale range is 67% of the full range. The
result is an accuracy of 0.00264% (0.008% x 33% FS) of the total full scale pressure for the one third low
end range, and an accuracy of 0.008% of reading (0.008% x R) for any pressure in the upper two thirds
of the transducer range. See gure 4.7.1b.
■ Flat, horizontal location; secure fixed working surface (desk top model) or proper installation in a
sturdy 19” mounting rack or cabinet
■ Sufficient air circulation must be provided at the rear of the instrument to avoid accumulation of heat
Pressure Supply Requirements:
■ Stable supply pressure 10% higher than the full scale of the controller
■ Permissible media: dry, clean air or nitrogen
■ Vacuum: minimum 50 liters per minute (if required)
5.1 Initial Installation
The initial installation of the CPC8000 includes the following steps: Unpack the system, place it in a
suitable work place, connect it, switch it on and congure.
A Quick Start Guide has been placed inside the shipping box containing your instrument. This guide
provides a quick reference to the pneumatic and electrical connections and an introduction to the operator interface. It is intended for experienced operators. All safety precautions within this manual should be
understood and followed.
5.2 Unpacking the system
Your new instrument was subjected to many hours of functional testing before it left the factory. In addition
to testing, the unit was inspected for appearance prior to being packaged for shipment. Upon removal
from its carton please examine the instrument for shipping damage. Report any apparent damage to the
carrier immediately.
Apart from any additional components ordered, a shipment consists of:
■ CPC8000 High-End Pressure Controller
■ Power Cord
■ Operating Manual
■ Quick Start Guide
■ Any options ordered
■ Two manifold seal plates (either attached internally or sent externally with the CPC8000)
■ An envelope containing the Calibration Certificate
■ A 3/16 hex screwdriver
■ A microfiber cloth to clean the glass front panel
■ Muffler, if pressure ranges ≥ 1000 psi (70 bar)
Operating Instructions - CPC8000 23
High-End Pressure Controller
19.5
8.6
1.5
16.2
CPC8000
5.3 Dimensions
Desktop:
19.5
8.6
1.5
16.2
Rackmount:
front view
19
16.9
7.0
front viewside view
Adaptor Fitting:
1/4 in. Tube OD x 7/16-20 Male SAE/MS Straight Thread
-or-
6 mm Tube OD x 7/16-20 Male SAE/MS Straight Thread
rear view
side view
161.5
24 Operating Instructions - CPC8000
High-End Pressure Controller
CPC8000
5.4 Turning on the CPC8000
After the pressure connections are secure in accordance with Section 5.5 below, apply power to the power
connector on the rear of the instrument and switch the power switch ON. The instrument will go through an
initialization process and system check. As soon as the system check is completed the system will default
to an operating screen similar to the screen shown in Section 3.2, Front Panel. Allow the instrument at least
30 minutes of warm up time before performing critical pressure measurements.
Earth Ground! Any power adaptors or surge protection devices that negate the
protective earth ground should not be used. The power cord must be accessible
and contain a protective earth ground.
Ventilation! Do not block airow to ventilating fans located on rear of instrument.
The instrument can be set up on a table top or it can be rack mounted. Rack mount adapters are optional.
After the controller has warmed up, cycle the pressure from the minimum to the maximum then back to
the minimum pressure of the primary transducer with the desired volume connected to the Measure/
Control port. This will allow the controller to properly compensate for the dierence in external volume
between what was used at the factory versus what is currently in use. This pressure cycle should occur
after each power cycle, or when the external volume size has been changed.
5.5 Pressure Connections
Warning! The pressure connections must be installed according to the following
instructions, observing the relevant regulations. The installation is to be performed
by trained, authorized personnel, knowledgeable in the safety regulations for
working on pneumatic/hydraulic systems.
Up to 6 pressure connections are on the rear panel. Pressure connections that are not
assigned are plugged.
High-End Pressure Controller - CPC8000
Power supply: 100 ... 120 VAC / 200 ... 240 VAC, 50 ... 60 Hz
Energy consumption: max. 140 VA
Pressure medium: clean dry air or nitrogen
Operating temp. : 15 ... 45°C
Serial number: 910021
Basic instrument: LP-NVR (-14.5 … 90 psi / 0 … 105 psi abs.)
Date of manufacture: 12/2012
Mensor - 201 Barnes Drive - San Marcos, TX 78666
BAROMETRIC
REFERENCE
SUPPLY
EXHAUST
VENT
REF
MEASURE\
CONTROL
Figure 5.5 - Pressure Connections
Operating Instructions - CPC8000 25
High-End Pressure Controller
CPC8000
■ Supply
The pressure supplied to the pressure connection labeled “Supply” should be approximately 10%
higher than the full scale of the highest pressure transducer installed in the controller (see the label on
the lower right of the front panel).
■ Exhaust
The pressure connection labeled “Exhaust” is for the vacuum supply. In a gauge pressure version it
can be left open to atmospheric pressure.
If the instrument’s primary pressure range has the minimum pressure in the vacuum region but
is not negative atmosphere, a vacuum regulator will be included. The vacuum regulator included
(6100500023) will need to be connected in-line between a vacuum pump and the exhaust port of the
instrument in order for the instrument to control properly. This vacuum regulator will need to be set to
10% below the minimum range of the primary transducer of the instrument. The vacuum regulator can
be set using a at-head screwdriver.
■ Vent
The pressure connection labeled “Vent” is the exhaust port where the system pressure is vented to the
atmosphere under certain conditions. Leave this port open or connect the optional muer for noise
reduction.
■ Reference
This is the pressure connection to the reference ports of transducers with ranges <50 psi (4 bar)
gauge.
The Reference port should be left open to the atmosphere and should NEVER be connected to a pressure source.
The higher the supply pressure connected to the Supply port, the higher the pressure which leaves the
system through the exhaust port. If a vacuum pump is connected to the exhaust port, adequate safety
precautions have to be taken to prevent damage to the pump.
If vacuum is connected to the Exhaust port of the controller, negative pressure peaks might occur at the
Measure/Control port for a very short time, when changing from the Measure mode to the Control mode.
Before the system is connected, the user must ensure that suitable protection measures are in place
which exclude overpressuring of the test piece or the system.
■ Measure/Control
The Measure/Control port (when in the Control mode) supplies pressure that is precisely controlled
by the controller. In the Measure mode a pressure applied to the Measure/Control port is measured by
the instrument transducer.
■ Barometric Reference
The Barometric Reference port is connected to the internal barometric transducer and should beleft open to atmospheric pressure.
26 Operating Instructions - CPC8000
High-End Pressure Controller
CPC8000
5.6 Pneumatic Connections and Pressure Fittings
The CPC8000 is supplied with ve 1/4 in. tube OD x 7/16-20 male SAE/MS straight thread adaptors or
ve 6 mm tube OD x 7/16-20 male SAE/MS straight thread adaptors (per order). The 7/16-20 male SAE/
MS straight thread end of the adaptors should be connected to the instrument’s 7/16-20 female SAE/MS
straight thread connections. The adaptors need only be “wrench tight” for proper sealing. Connections
should be made according to your specic requirements ( for example, a vacuum pump is only required
when controlling to a subatmospheric pressure).
For functional testing and to become familiar with the operation of the CPC8000, a volume
close to the volume expected during normal operation can be connected to the Measure/
Control port.
WARNING! The user must ensure that released pressure medium does not endanger personnel, environment or the instrument.
Test items with large volumes or long piping runs with small diameter tubing can have a
negative eect on the control performance.
Open
Clean dry air or Nitrogen
regulated pneumatic
supply pressure set at
110% of the highest full
scale internal range
High-End Pressure Controller - CPC8000
Power supply: 100 ... 120 VAC / 200 ... 240 VAC, 50 ... 60 Hz
Energy consumption: max. 140 VA
Pressure medium: clean dry air or nitrogen
Operating temp. : 15 ... 45°C
Serial number: 910021
Basic instrument: LP-NVR (-14.5 … 90 psi / 0 … 105 psi abs.)
Date of manufacture: 12/2012
Mensor - 201 Barnes Drive - San Marcos, TX 78666
Vacuum
pump, if required for
sub-atmospheric
pressure, otherwise
open
Open or
muer
Figure 5.6 - Pneumatic Setup
Open
Device under test
Operating Instructions - CPC8000 27
High-End Pressure Controller
CPC8000
5.7 Electrical & Communication Connections
WARNING! The electrical installation must be done according to the following
instructions while observing the relevant regulations. It is to be performed by
individuals familiar with the safety regulations for working on electrical connections.
5.7.1 Notices Regarding the Electrical Connections
5.7.1.1 Power Supply
WARNING! Before connecting the power supply, make sure that the power voltage
agrees with the voltage specication of the CPC8000. Switch o the system before
connecting the power.
The 3-pin power cable supplied is tted with a ground lead. The system should only be operated from a
3-pin socket with the ground lead properly connected.
5.7.1.2 Interfaces
WARNING! The interface cables must not be longer than 3 meters and must be
separate from cables with voltages greater than 60 volts.
5.7.1.3 Ethernet Interface
The Ethernet jack is a standard Ethernet interface used for remote communication.
5.7.1.4 USB Host Interface
The USB Host interface is a service port used to upgrade software and download information from the
device. It has the same function as the USB Host port on the front of the CPC8000.
5.7.1.5 USB Device Interface
The USB Device jack is a standard USB Type B receptacle interface used for remote communication. The
USB Driver can be downloaded at http://www.mensor.com/download_software_drivers_en_um.WIKA
5.7.1.6 RS-232 Interface
The RS-232 interface is a 9-pin D-Sub socket used for remote communication.
5.7.1.7 IEEE-488 Interface
The IEEE-488 interface is an IEEE-488 socket used for remote communication.
5.7.1.8 Input/Output Port
CAUTION! The Input/Output port has eight digital inputs that will accept either 3.3 VDC or 5 VDC signaling. Each input consumes about 8-10mA. If a voltage greater than 6VDC is
applied to an input it will be permanently damaged.
The Input/Output port also has eight relay outputs. Both normally open and normally closed contacts are
available for each output on the port. The contact ratings for the relays are as follows:
The instrument can be set to respond to the digital inputs and to change the outputs based on certain
conditions. See section 7.5.2.5.8.
Operating Instructions - CPC8000 29
High-End Pressure Controller
CPC8000
6. Local Operation
6.1 General
This section describes the procedures for operating the CPC8000 from the front panel. Instructions for
operating the instrument from an external computer are covered in Section 8, Remote Operation. By
following the procedures provided in these two Sections and in Section 7, Setup, you can expect your
CPC8000 to deliver maximum accuracy and dependability for many years of exceptional service.
When the instrument is powered on it will take about 30 seconds to boot up. At the end of the boot
process the display screen similar to gure 6.1 will appear. This is the main operating screen. At this point
the CPC8000 is ready for operation. However, for critical tests and measurements allow the unit to warm
up for at least 15 minutes with power on to ensure rated accuracy.
Notice that the portion of the screen with a blue eld covers about two-thirds of the total screen area. This
area contains the pressure output reading, the setpoint, the active transducer range, units of measure,
limits and a choice between the control modes: Measure, Control or Vent.
The sidebar to the right (black eld) is used for entering data, adjusting environmental settings, dening
parameters for remote operations, and making choices in the conguration. Several dierent sidebars
can be accessed from the main screen to achieve dierent objectives.
This section of the manual will explain features of this main operating screen and the associated
sidebars.
Initialized Display - Ready for Local Operation
Main Screen (Blue)
Test / Calibration Parameters
Figure 6.1 - Initialized Display
Sidebar (Black)
Data entry
Beyond the main screen are a series of setup screens used to adjust parameters and conditions, such as
performing calibrations, running programmed routines, operating from a remote computer, etc. All such
setup screens are on black backgrounds, and have their own sidebars (also black) for choice menus and
data entry. Explanations and instructions for using these setup screens are covered in Section 7, Setup.
30 Operating Instructions - CPC8000
Loading...
+ 84 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.