WelchAllyn Xenon 300 User manual

Light Source Menu
Service Manual
Thank you for purchasing the Welch
Allyn
WARNING: All users of this product should be thoroughly trained in the medical procedures appropriate
to the equipment. Furthermore, they should read and understand the instructions for all other equipment used in conjunction with this product.

Table of Contents

Conventions
Warnings and Precautions:
Indications Contraindications. Components
System Description Set-Up and Operation Cleaning Instructions. Lamp Care and Replacement.
...................................................
..................................
.................................................
..;
..............................................
...................................................
........................................
......................................
.......................................
................................
Specifications..................................................
Basic Troubleshooting
...........................................
Warranty......................................................
Ordering Information.
Part Information. Replacement Parts Parts and Service
...........................................
..........................................
.........................................
..........................................
1 1
2
2
3 3
4
4
5
7 8 9
9
9 9
9
WARNING: Extremely high energy light. Some fiber optic bundles will not with­stand the light intensity of this light source. USE ONLY WELCH ALLYN FIBER OPTIC BUNDLE or CUDA PRODUCTS, INC.
.
CLAD ROD (CR) MODEL FIBER OPTIC BUNDLES.
(904-737-7611)

Conventions

Warnings and Precautions:

Consult users manual for important information.
I
I
n
I
A
I
I
I
n
\I/
- ­0
‘I’
\‘/
- _
0
‘I’
POSSIBLE EXPLOSION HAZARD. Equipment not suitable for use in the presence of a flammable thetic
mixture with air or with oxygen or nitrous oxide.
This system should only be used by qualified medical personnel.
WARNING: THERE IS A RISK OF TION. THE RISK OF INJURY IS REDUCED BY ESSARY, BY MINIMIZING EXPOSURE TIME, AND BY AVOIDING CLOSE STATIONARY VIEWING.
WARNING: The high intensity light at the front of the light source and at the tip of a fiber optic bundle may give rise to high temperatures. To minimize the risk of injury, avoid contact with tissues.
WARNING: As with any bright light, do not look into the beam. Permanent
To reduce the chance of direct visual exposure to the light, set the light intensity knob to the minimum level and plug the fiber optic bundle into the light source before turning the power ON.
INJURY
TO TISSUE FROM EXPOSIJRE TO THE INTENSE ILLUMINA-
USING
THE MINIMUM LEVEL OF ILLUMINATION NEC-
eye
injury may result.
anes-
I
I
n
\
f
Lb
When
the unit is on, never disconnect the fiber optic bundle from the light source or the accessory from the
end of the fiber bundle. High intensity light from the end of the fiber optic bundle and light source can
ignite flammable materials (i.e. drapes).
WARNING: TOTAL SYSTEM RISK CURRENT
SHOULD NOT EXCEED SYSTEM RISK CURRENT EXCEEDS
Grounding reliability is achieved only when connected to hospital-use or hospital-grade receptacles. Inspect electrical plug and cord routinely. Do not use if damage is discovered.
THE COVER OF THE LIGHT SOURCE SHOULD NEVER BE REMOVED. ELECTRICAL SHOCK HAZARD DUE TO HIGH INTERNAL VOLTAGE. THERE ARE NO USER SERVICEABLE PARTS INSIDE THIS LIGHT
SOURCE EXCEPT FOR THE LAMP AND FUSE, WHICH ARE ACCESSIBLE THROUGH THE BOTTOM
AND SIDE DOOR PANELS. Refer all service to
100
MICROAMPS. AN ISOLATION TRANSFORMER IS REQUIRED IF TOTAL
100
-
EXCLUDING THE PATIENT APPLIED INSTRIJMENT
MICROAMPS WHEN ACCESSORIES ARE INTERCONNECTED.
Welch Allyn.
-
1-1
WARNING:
PROVIDE VENTILATION TO PREVENT OVERHEATING. Do not cover or drape the light source.
Do not use the turret as a light attenuator result.
The ends of the fiber bundle get hot during use. Allow to cool before removing.
Replace fuses as marked. See Fuse Replacement Section.
or
with the turret misaligned with the light
port
overheating will
Do not operate light source without
CAUTION: IF THIS LIGHT SOLJRCE IS USED DURING INVASIVE SLJRGEKY, THE FAILURE OF A COM­PONENT WITHIN THE SYSTEM, CAUSING LOSS OF ILLUMINATION, MAY DIRECTLY CAUSE RISK TO THE PATIENT. THEREFORE, IT IS RECOMMENDED TO HAVE ALTERNATIVE MEANS OF TION AVAILABLE.
Since the light source uses a custom lamp, always have a spare lamp available for replacement. Welch Allyn Allow the lamp to cool before replacing.
Equipment not protected against the ingress of water. Do not use or store liquids above or on the light
source.

Indications

For
use
with the Welch
lamp
module in place.
replacement lamp number 04980. USE CAUTION BECAUSE LAMP MAY BE HOT.
Allyn
Headlight System
ILLUMINA-
Use
only

Contraindications

Neonate Ophthalmic Laryngoscopes Endoscopes
1-2
transillumination
procedures

Set-Up and Operation

LJnpack
1.
2.
Plug power cord into light source power
3.
Select type of fiber optic bundles port (ACMI, Wolf, Olympus,
4.
Insure that the ends of the fiber optic bundle are clean before each
unit and visually inspect it for damage that may have occurred during
module
and the remaining end into a hospital grade receptacle.
Storz)
and rotate turret to align with light port. use.
Any residue on the ends of the fiber
shipment.
optic will absorb the high intensity light, and will overheat and damage the fiber optic bundle. Insert fiber optic bundle into light port and connect the headlight luminare.
5.
level.
This prevents inadvertent shining into
6.
When the unit is on, NEVER disconnect the fiber optic bundle from the light source or the headlight from the
the
patient’s or user’s eyes when turned on.
Set light intensity to minimum
fiber optic bundle. High intensity light from the end of the fiber optic bundle and light source can ignite flam­mable materials (i.e. drapes).
Turn the unit on.
7. a.
Set light intensity to minimum level necessary for adequate visualization.
9.
Turn unit off when not in use.
10. Allow to cool before removing fiber optic bundle.
The ends of the fiber optic bundle get hot during use.
With the output of high intensity (Xenon) light sources, it is possible to convey a large amount of light energy and to concentrate this energy in a very small area. This may result in thermal injury to the tissue.
The risk of injury is increased under the following
1.
Prolonged close contact or close stationary viewing of the mucosa or tissue.
2.
Using a high intensity light source at high output levels.
3.
Using multiple instruments to illuminate the same area.
conditions:
The following recommendations will reduce the risk of thermal injury:
1.
Use
the minimum level of illumination necessary for adequate visualization.
2.
When possible, avoid close stationary viewing.
3. Minimize exposure time.

Cleaning Instructions

NOTE: Welch Allyn Xenon 300 Light Source should not be sterilized or immersed.
1.
Prior to cleaning, turn power off and disconnect the power cord from both the wall outlet and the Welch Allyn Xenon 300 Light Source.
2. The external surfaces may be cleaned by wiping with a cloth dampened with mild soap and water or by using the following cleaning/disinfecting solutions. DO NOT IMMERSE.
l 70% isopropyl alcohol l 10% mild bleach solution
3.
The power cord may be wiped clean using the same a clean cloth. Care must be taken not to get the plug prongs wet. DO NOT IMMERSE.
4. Do not plug the unit into a wall outlet until it is thoroughly dry.
CAUTION: The ends of the fiber optic bundle must be clean before each use. High intensity light will be
absorbed by any residue (including cleaning and disinfecting solutions) on the surface of the fiber optic. Absorbed light will overheat the residue causing damage to the fiber optic bundle.
cleaning/disinfecting
solutions listed above. Wipe dry with
l-4

Lamp Care and Replacement

1-5
The
lamp in the Welch
symptoms indicating the need for replacement are:
1.
Low light output
2.
After the unit is turned on, there are four or more ignition attempts without lamp going into opera-
tion.
3.
The lamp life indicator reads more than 500 hours.
Allyn
Xenon 300 Light Source will typically provide more than 500 hours of service.
The
I__
The light source is designed for simple and quick changing of the lamp module, which houses the lamp.
To remove the old lamp module:
1.
Turn unit off and unplug power cord from its power source.
2.
Wait 5 minutes to allow the lamp to cool.
3.
Unlock
4.
Grasp the two handles and pull the entire lamp module out of the light source.
To insert new lamp module:
1. Orient the new lamp module so that the window of the lamp is facing the front of the light source.
2.
Carefully align the connection pins with the brass sockets on the lamp module and push firmly to insure proper connection.
and open access door located on the side of the
\
SERIAL NUMBER
light
source.
I
I
n
3.
Close and lock the access door.
CAUTION: DO
not incinerate used lamps to avoid risk of explosion
1-5

Fuse Replacement

1.
If light
source
fails to
operate
after power is
turned on,
turn unit off prior to checking
2. Unplug the light source.
3.
Remove fuse covers by using flat blade screw driver and check for blown fuse.
4. Replace the blown fuses with
5. Replace
fuse covers.
s.
the same
size and rating.
the
fuses.
1-6

Specifications

Electrical
Lamp Operating Power: 300 watts DC, nominal Lamp Current: Input Voltage: Lamp: Lamp Life: 500 hours
Dimensions
Size: Weight: 12.5 lbs (5.7kg)
Fiber Optic Bundle Interface
Turret Fiber Bundle Port accepts ACMI, Wolf, Olympus. and Storz
Environment
Operating: Transport/Storage:
Fuses
5mm
fuse, T
Approvals
UL 2601-l CAN/CSA C22.2 IEC 601-l IEC 601-l-2 CE
The CE mark on this product indicates it has been tested to and conforms with the provisions noted within the
42/EEC
Declarations of Conformity are held by Welch Allyn’s European Representative.
European Regulatory Manager Welch Allyn LTD. Kells Road, County Tel +353 46 79060 FAX +353 46 27128
Equipment Classification
Class 1 Equipment Continuous Operation
8.0L
8 amp 250V.
No601.1-M90
Medical Device Directive.
Navan,
Meath,
Republic of Ireland
21 Amps DC, nominal
100-120/200-240
CermaxB
12.75”
10°C to
-2OoC
Xenon Lamp
(324mm)W
4OoC,
to
49oC,
VAC,
x 4.75”
95% R.H.,
95% R.H. Max,
50/60
(121mm)
500hPa -
Hz, 5.5A
H x 12.5”
1060
hPa,
5OOhPa - 1060hPa,
(318mm)
Altitude
Altitude
D
93/
Patents
Patents Pending
NOTE: This light source complies with current required standards for electromagnetic interference and should not present problems to other equipment or be affected by, other devices. As a precaution, avoid using this device in close proximity to other equipment.
Equipment Disposal: Do not incinerate this equipment at the end of useful life for risk of lamp explosion. Equip­ment contains no hazardous materials and can be discarded without environmental risks.
1-7

Basic Troubleshooting

1-8
Possible Problem
Power
Power does not come on.
plug is not plugged into the
Power switch is off.
receptacle.
Plug
in power cord.
Turn on
power
switch.
No light output.
Lamp flickers or dims.
Fuse is blown. Lamp door is not closed tightly. Lamp module is not installed incorrectly.
Power plug is not plugged into receptacle. Power switch is off. Fuse is blown. Lamp door is not closed tightly.
Lamp module is installed incorrectly.
Turret misaligned with light port.
Lamp burned out.
Fiber bundle is not connected. Light attenuator is set to minimum.
Lamp is getting old.
Replace fuse. Close lamp door completely. Reinstall lamp module correctly.
Plug in power cord. Turn on power switch. Replace fuse. Close lamp door completely. Reinstall lamp module. Set turret to detent position (light port
aligned). Replace lamp. Connect fiber bundle correctly. Set the light attenuator to minimum
level needed for adequate visualization. Replace lamp.
Light source turns off
after a few minutes of operation.
Low light output
Several ignition attempts needed to start lamp.
I-8
Obstructed air intake: overheating causes thermal switch to trip.
Fan not running: overheating causes thermal switch to trip.
Accessory damaged. Lamp is old. Light transmission rod is damaged or con-
taminated.
Three or fewer ignition attempts are normal.
Allow unit to cool (5-10 minutes). Remove obstruction(s).
Contact Welch
Replace accessory. Replace lamp. Clean with isopropyl alcohol and allow
to dry. Use care not to chip edges. Contact Replace
attempts are required.
lamp if four or more ignition
Allyn.
Welch Allyn
for repair.
The Welch
1-9
Allyn other than misuse, neglect, damage in shipment, or normal wear. This warranty is in effect for one nal date of
Allyn
Xenon 300 Light Source is
will repair or replace, free of charge,
purchase.
guaranteed
any
parts of its own manufacture that prove to be defective for reasons
by Welch Allyn against all manufacturing defects. Welch

Ordering Information

Part Information

year from
origi-
49801

Replacement Parts

04980
76500

Parts and Service

Welch Allyn, Inc.
ATTN: Technical Services
4341 State Street Road
Skaneateles Falls, NY 13153 l-800-535-6663
Welch Allyn Xenon 300 Light Source 100-120/200-240 VAC, 50-60 Hz, 5.5A 50-60 Hz USA Plug type
Replacement Lamp Module USA
Power Cord
1-9