Propaq 6.2X-8.4X/Propaq Encore 1.20-1.3X Reference Guide1
Update
2Protocol Systems, Inc.
English
Introduction
Definitions
WARNING
WARNING statements identify conditions or practices that could result
in personal injury.
note...
NOTE statements identify useful tips or points about operation.
Software Version Table
Software Versions
Model and software version
Power Adapter Symbols
Many sections in this update are
accompanied by a Software Version table
indicating which monitor software
versions are affected by the change.
Sections that only affect monitor hardware
do not have this table.
English
The following symbols may appear on the label for the monitor
ac power adapter.
Direct current
NRTL/C
Evaluated to CSA 601-1
and UL2601-1
Propaq 6.2X-8.4X/Propaq Encore 1.20-1.3X Reference Guide3
Update
The Canadian Standards Association has
evaluated this device according to CSA 601-1 and
Underwriters Laboratory Standard UL 2601-1.
This device has met all essential requirements of
the European Medical Device Directive 93/42/EEC
for a Class 1 product.
Alarm Holdoffs
Propaq Audio and Visual Alarm Holdoffs
Software Versions
Propaq Encore 1.20-1.3X
Propaq 100 Series 8.4X
Motion artifact or other factors can cause false vital sign alarms.
To help minimize false alarms, the Propaq briefly delays or
“holds off” triggering alarms for certain vital sign limit
violations.
After the alarm holdoff period begins, if the monitor detects that
the patient’s vital sign has returned to acceptable limits, the
Propaq cancels the alarm holdoff. The next time a vital sign limit
is violated, the Propaq starts a new alarm holdoff period. The
following table summarizes the alarm holdoff periods.
Monitoring Function
HR/PR3 seconds
SpO
2
RR/BR5 seconds
Alarm Holdoff Time Period
10 seconds
When the Propaq is connected to the Acuity Central Monitoring
System, Acuity does not receive any alarm indication from the
monitor until this holdoff period expires.
4Protocol Systems, Inc.
Propaq Audio Alarm Holdoff with Acuity
Software Versions
Propaq Encore 1.20-1.3X
Propaq 100 Series 6.2X-8.4X
When a Propaq is connected to an Acuity System, the audio
alarms at the bedside monitor are delayed for 10 seconds. When
the monitor informs Acuity that an alarm condition has occurred,
Acuity does the following:
•Produces the audio and visual alarm at the Acuity Central
Monitor.
•Sends the bedside monitor a signal that delays the bedside
audio alarm for 10 seconds (the bedside monitor displays the
visual alarm indications without delay).
After the 10-second delay, the bedside audio alarm occurs.
note...
This audio alarm holdoff does
•Neonatal mode (Propaq Encore)
•Apnea (Propaq Encore and Propaq 100 Series)
not
occur for the following:
English
Acuity SpO2 Alarm Holdoff
Software Versions
Propaq 100 Series 6.2X-8.3X
When connected to Propaq 100 Series monitors, the Acuity
System may be configured to hold off the visual and audio alarm
for SpO2 at the Acuity Central Monitor for up to 10 seconds. After
the holdoff period expires, the alarms occur at Acuity.
Propaq 6.2X-8.4X/Propaq Encore 1.20-1.3X Reference Guide5
Update
While this SpO2 alarm holdoff is occurring at Acuity, the
10-second audio alarm holdoff at the bedside Propaq 100 is also
occurring as previously described (see “Propaq Audio Alarm
Holdoff with Acuity”). After the 10-second delay, the bedside
audio alarm occurs. Contact Protocol Systems Technical Support
for more information about this feature.
How to Change Power-On Default Settings
Software Versions
Propaq Encore 1.20-1.3X
Propaq 100 Series 7.XX-8.4X
The power-on default settings for a Propaq Encore or Pr opaq 100
Series monitor are preset at the factory to the “Factory Default
Settings” listed in Appendix B of the
Propaq Encore Reference Guide.
However, you can change any of these power-on default settings.
Change Power-On Settings for Date, Decimal, ECG Filter,
and LCD Contrast
Propaq Reference Guide
and
Whenever you change any of these settings, you also
immediately change the corresponding power-on default setting.
No other steps are needed. (The LCD contrast setting is only
available with the Propaq 100 Series LCD monitors.)
6Protocol Systems, Inc.
Change Other Power-On Settings
To change all other power-on settings such as ECG size, alarm
limits, or printer settings:
1.
Select the desired monitor settings
(for example, select ALARM TONE as HIGH).
2.
From the Main Menu, press:
•Propaq Encore:
SETUP, MORE, MORE, PROGRAM, CURRENT, YES
•Propaq 100 Series:
SYSTEM, (MORE), PROGRAM, CURRENT
The monitor selects all of the current settings as the power-on
default settings.
WARNING
If any alarms are set to OFF and you select CURRENT (to set current
settings as power-on settings), those alarms will be OFF at power-on for
subsequent patients. Consider carefully before setting power-on alarms
to OFF.
note...
The alarm for apnea cannot be turned off at any time.
.
English
.
note...
If you change the patient mode, the alarm limit settings are automatically
changed to the defaults for that mode.
Propaq 6.2X-8.4X/Propaq Encore 1.20-1.3X Reference Guide7
Update
Revert to Factory Power-On Default Settings
To reprogram the monitor to the original factory power-on
default settings (as listed in Appendix B of the
Guide
and
Propaq Encore Reference Guide
1.
From the Main Menu, press:
):
•Propaq Encore:
SETUP, MORE, MORE, PROGRAM, DEFAULT, YES.
•Propaq 100 Series:
SYSTEM, (MORE), PROGRAM, DEFAULT.
note...
This step changes all of the power-on default settings to the factory default
except
settings
settings remain at the current setting.
for Date, Decimal, and ECG Filter (and LCD Contrast). These
Propaq Reference
8Protocol Systems, Inc.
New Right Side Panel Label
The Propaq Encore and Propaq 100 Series monitors have new
right side panel labels. The Acuity connector is now labeled
“Networks ACUITY MODEM.” It allows either dir ect connection
to an Acuity System, or connection to an external modem for
telecommunication to an Acuity System. Refer to the
Propaq Reference Guide
New Propaq Encore Right Side Panel
for more information.
English
Modem-
Defib Sync Connector
Real-Time ECG Output
Connector
Speaker
Battery Charging Light
Connector for Acuity or
Modem-Propaq Option
On/Off Switch
Input Fuse
Power Input
Connector
DEFIB SYNCHROMONITOR
!
!
3A
2AG
12-28V, 3A
EKG x 1000
!
!
New Propaq 100 Series Right Side Panel
On/Off Switch
Input Fuse
Power Input
Connector
MONITOR
3A
2AG
DEFIB SYNCHRO
!
!
Defib Sync Connector
Speaker
Battery Charging Light
Connector for Acuity or
Modem-Propaq Option
12-28V, 3A
Propaq 6.2X-8.4X/Propaq Encore 1.20-1.3X Reference Guide9
Update
Specification Changes
Clarification of Propaq Encore Specification
The temperature probe tolerance has been restated: