Welch Allyn 6102, 6154 Directions For Use Manual

Page 1
Welch Allyn In-Floor Scales
Directions for use
Models 6102, 6154
Page 2
© 2017 Welch Allyn. All rights are reserved. To support the intended use of the product described in this publication, the purchaser of the product is permitted to copy this publication, for internal distribution only, from the media provided by Welch Allyn. No other use, reproduction, or distribution of this publication, or any part of it, is permitted without written permission from Welch Allyn.
Welch Allyn assumes no responsibility for any injury to anyone, or for any illegal or improper use of the product, that may result from failure to use this product in accordance with the instructions, cautions, warnings, or statement of intended use published in this manual.
Software in this product is Copyright 2017 Welch Allyn or its vendors. All rights are reserved. The software is protected by United States of America copyright laws and international treaty provisions applicable worldwide. Under such laws, the licensee is entitled to use the copy of the software incorporated with this instrument as intended in the operation of the product in which it is embedded. The software may not be copied, decompiled, reverse-engineered, disassembled, or otherwise reduced to human-perceivable form. This is not a sale of the software or any copy of the software; all right, title, and ownership of the software remain with Welch Allyn or its vendors.
For information about any Welch Allyn product, contact your local Welch Allyn representative: http://
www.welchallyn.com/en/other/contact-us.html.
724518
DIR 80021339 Ver. B Revision date: 2017–01
Welch Allyn, Inc.
4341 State Street Road
Skaneateles Falls, NY 13153 USA
www.welchallyn.com
This manual applies to 901110 NONMASTED SCALES.
Page 3

Contents

Introduction ............................................................................................. 1
Assembly and setup ................................................................................ 7
iii
Intended use ........................................................................................................ 1
Indications for use ................................................................................................ 1
Symbols ............................................................................................................... 1
About warnings and cautions ............................................................................... 2
Controls and indicators ........................................................................................ 4
Scale and handrail installation .............................................................................. 7
Custom setup ...................................................................................................... 7
Scale operation ...................................................................................... 11
Power options .................................................................................................... 11
Audible beeps .................................................................................................... 11
Weighing procedure ........................................................................................... 12
Print a weight measurement ............................................................................. 12
Cleaning and maintenance ................................................................... 13
Cleaning ............................................................................................................. 13
Rechargeable battery replacement .................................................................... 13
Change the printer paper ................................................................................... 14
Maintenance ...................................................................................................... 15
Troubleshooting ................................................................................................. 15
Appendices ............................................................................................ 17
Specifications ..................................................................................................... 17
Product configurations ....................................................................................... 18
Approved accessories ........................................................................................ 18
Warranty ............................................................................................................ 18
Page 4
iv Contents Welch Allyn In-Floor Scales
Page 5

Introduction

Intended use

The Welch Allyn Integrated Scales are intended to be used by clinicians for weighing patients up to 825 lb (375 kg) for models 6102 and 6154.

Indications for use

1
The Welch Allyn In-Floor Scales are used by clinicians to weigh patients.

Symbols

Documentation symbols

WARNING The warning statements in this manual identify conditions or practices that could lead to illness, injury, or death. Warning symbols will appear with a grey background in a black and white document.
Consult Directions for Use

Shipping, storing, and environment symbols

Temperature limit
Caution The caution statements in this manual identify conditions or practices that could result in damage to the equipment or other property, or loss of data.
Keep dry
Separate collection of Electrical and Electronic Equipment. Do not dispose as unsorted municipal waste.
Separate collection of batteries. Do not dispose as unsorted municipal waste.
Page 6
2
Introduction Welch Allyn In-Floor Scales
Humidity limitation
This way up Stacking limit by number
Recyclable
Rechargeable battery Fragile

Miscellaneous symbols

Manufacturer
Atmospheric pressure limitation
Nickel-Cadmium battery
Ni-Cd
Reorder Number
Product Identifier For indoor use only
Serial Number Prescription only or "For Use by or on the
Rated power input, AC Direct Current (DC)
Fuse

About warnings and cautions

Caution statements can appear on the Welch Allyn In-Floor Scales, the packaging, the shipping container, or in this Directions for use.
The Welch Allyn In-Floor Scales are safe for patients and clinicians when used in accordance with the instructions and caution statements presented in this Directions for use.
Before using the device, you must familiarize yourself with all cautions, with the steps to power up the device, and with the sections of this Directions for use that pertain to your use of the device. In addition to reviewing the general cautions presented in the next
order of a licensed medical professional"
Page 7
Directions for use Introduction 3
section, you must also review the more specific cautions that appear throughout the manual in conjunction with setup/startup, operation, and maintenance tasks.
Failure to understand and observe any warning statement in this manual could lead to patient injury or illness.
Failure to understand and observe any caution statement in this manual could lead to damage to the equipment or other property, or loss of patient data.

General warnings and cautions

WARNING Patient injury risk. Make sure the scale is properly cleaned between measurements.
WARNING Patient or operator injury risk. Make sure the scale is positioned on a stable surface and in a way to avoid shock or bumping hazards.
WARNING Improper disposal of batteries may create an explosion or contamination hazard. Never dispose of batteries in refuse containers. Always recycle batteries according to local regulations.
CAUTION Make sure that you routinely perform general maintenance and equipment safety checks on your scale. Remove the scale from service when you notice damaged power cords.
CAUTION Make sure the scale is set in the proper measurement units for your facility. Make sure you record the measurement units as they are displayed.
Page 8
4 Introduction Welch Allyn In-Floor Scales

Controls and indicators

Scale

1 Front panel
2 Hand rail (optional)
3 Weighing platform
Page 9
Directions for use Introduction 5

Front panel buttons

Button Function
ZERO Press this button to reset the scale to zero.
ON Press this button to turn on the scale.
ST
RECALL Press this button to recall the previously stored scale reading.
LB Press this button to display the weight in pounds.
KG Press this button to display the weight in kilograms.
PRINT (DATA) SEND
REWEIGH Press this button to retake the patient's weight.
Press this button in conjunction with ON to enter the setup mode.
Note
Press this button to print the current weight. On models equipped with a RS-232 port, pressing this button when the scale is connected to a
computer sends the current measurement to the computer.
This button is not present if you purchased a scale with the kilogram-only option.
Page 10
6 Introduction Welch Allyn In-Floor Scales
Page 11

Assembly and setup

Use the following instructions to assemble and setup your scale for use.

Scale and handrail installation

Please refer to the instructions that shipped with your scale for installation information.
7

Custom setup

You can customize the scale to best suit your needs. Configurations that can be selected include the automatic shut-off time, weighing units, weight resolution, and more.

Enter custom setup mode

1. Make sure the scale is powered down.
2. Press and hold ST while pressing ON.
3. Once the scale powers up, press and release ST five times.
The scale displays SEt-UP and produces a series of four long beeps.
Page 12
8 Assembly and setup Welch Allyn In-Floor Scales

Set the options

Enter the custom setup as described.
1. Press ST to scroll through the custom setup options.
Note Options indicated with an asterisk (*) have multiple
selections.
2. Press RE CALL once a value is displayed to change the value. When the desired value appears, press ST to set the new value.
3. You can exit the menu at any time by pressing ON to power down the scale. The scale saves all changes.
The options displayed with each press are as follows:
Option displayed Feature
SOFt This option displays the software version of your scale.
dAtE This option displays the release date of the software. The format is MM/DD/YY.
SCALE This option displays the model number of the scale.
AutOFF* This displays the number of seconds before the scale turns off when operating on battery power.
The Cont value prevents the scale from turning off automatically.
AC con*
rES*
UnitS* This option allows you to change the displayed weight unit.
This option causes the scale to remain on when plugged in to AC power. Press RE CALL to turn this option ON or OFF.
This option allows you to change the resolution of weight. Press RE CALL to switch between the following options:
0.1 pounds / 0.05 kilograms
0.1 pounds / 0.1 kilograms
0.2 pounds / 0.1 kilograms
0.5 pounds / 0.2 kilograms
1 pound / 0.5 kilograms
Note Do not change the scale units if you have purchased the kilogram-only option.
Press RE CALL to switch between the following options:
KILOS* This option allows the weight to display in kilograms.
POUNDS* This option allows the weight to display in pounds.
bEEPEr* This option determines the audible signal that occurs when a front panel button is pressed.
Press RE CALL to switch between ON and OFF.
rECALL*
rS-232 Detailed instructions for this feature are available from Welch Allyn. Go to http://
OutPut* This option displays the current data port output option. Detailed instructions for this feature are
Press RE CALL to change the RE CALL button functionality between ON and OFF.
www.welchallyn.com/en/other/contact-us.html to find your local representative.
available from Welch Allyn. Go to http://www.welchallyn.com/en/other/contact-us.html to find your local representative.
Page 13
Directions for use Assembly and setup 9
Set the value to OFF to disable the optional internal printer or the RS-232 port.
PrtUnt*
PrtOPT*
bAud*
pArity*
StPbit*
This option determines whether pounds or kilograms are displayed on a printout. Press RE CALL to switch between the following options:
If the pounds and kilograms indicators are illuminated on the control panel, the printed output is determined by whichever unit is selected on the control panel.
If the kilogram indicator is illuminated, it will only print in kilograms, regardless of the unit is selected on the control panel.
If the pounds indicator is illuminated, it will only print in pounds, regardless of the unit is selected on the control panel.
Press RE CALL to switch between the following options:
PnlPrt: The current weight is printed when PRINT is pressed on the control panel. AutPrt: The current weight is automatically printed when a weight reading occurs.
This option allows you to set the baud rate. Press RE CALL to switch between 1200, 2400, 4800, 9600, and 19200..
Press RE CALL to switch between the following options:
Off: No parity, 8 data bits Odd: Odd parity, 7 data bits EuEn: Even parity, 7 data bits
Press RE CALL to switch between the following options:
1 bit: One stop at the end of a word 2 bits: Two stops at the end of a word
rtS.ctS*
Prtbtn*
rtEPrt*
PC brd
SEt-UP
Press RE CALL to switch between ON for the printer and OFF for RS232 transmissions.
Press RE CALL to turn On or Off the front panel PRINT button.
Press RE CALL to turn On or Off the enabling of printing from a remote location by connecting the receive line to the ground.
This option displays the model number of the printed circuit board. A second press of ST displays the revision of the printed circuit board.
This option is displayed when you have cycled through all the options. Press and hold ON to power down the scale.
Page 14
10 Assembly and setup Welch Allyn In-Floor Scales
Page 15

Scale operation

Power options

The scale can be used with battery power or with AC power.
Battery power
Press ON to turn on the scale.
11
The scale automatically turns off after 45 seconds of non-use when operating on battery power.
The LOW-BATT indicator displays when the battery power is low.
bAttrY displays when the batteries are unable to properly power the scale.
For surface-mounted scales, plug the scale into a power source when either of these messages are displayed.
AC power
When plugged in, the scale continuously runs on AC power automatically.
The scale switches to battery power when AC power is not present.

Audible beeps

The following beeps occur when using the scale.
Sound Meaning
One beep Acknowledgment of a button press.
Two beeps The scale is zeroed and ready to take a measurement.
Three beeps The weight reading is obtained and displayed.
No print, four short beeps
No print, four long beeps
Four long beeps
A print was requested, but valid weight is not displayed on the scale, or PRINT was pressed while the printer is in use.
There is a problem with the printer. Contact your service department.
The battery is too weak to operate the scale. The scale has entered setup mode.
Page 16
12 Scale operation Welch Allyn In-Floor Scales

Weighing procedure

1. Press ON to turn on the scale.
The scale sounds one beep.
2. Wait for the scale to display 0.0.
The scale sounds two beeps.
3. Place the patient on the scale and make sure that his or her weight is evenly distributed.
The scale sounds three beeps and the patient's weight is displayed.
4. If necessary, press REWEIGH to perform another weight measurement.
5. If necessary, press RECALL to display the previously stored scale measurement.
Note A measurement is stored in the scale until a new weight is
taken and zeroes out.
A stationary display of - - indicates a negative weight. Make sure that there is nothing on the scale and press ZERO to set the zero.
If O-LOAd is displayed, the capacity of the scale has been exceeded.

Print a weight measurement

Printers are a purchasable option for some scales. The printer will only print a valid weight. If your printer is set up with the auto-print option, a print is automatically generated after a measurement is taken.
1.
Take a weight measurement as described.
A single beep sounds.
2. Press PRINT (DATA) SEND.
A single beep sounds and the measurement is printed.
3. To print a previously stored scale measurement, press RECALL and then press PRINT (DATA) SEND.
The following errors can occur:
Four short beeps indicate an invalid print request. Make sure that the weight is properly displayed on the scale, or that a measurement is not already in the printing process.
Four long beeps indicate and internal problem with the printer. See the Troubleshooting section.
Page 17

Cleaning and maintenance

Cleaning

1. Clean the scale at least once per day, and the optional handrail after each use.
2. Use one of the following products when cleaning your scale:
70 percent isopropyl alcohol
13
PDI Sani-Cloths
®

Rechargeable battery replacement

Contact your service department.
Page 18
14 Cleaning and maintenance Welch Allyn In-Floor Scales

Change the printer paper

The printer is a purchasable option.
1. Use a screwdriver to loosen the screw on the battery access door and remove the printer access door.
2. Remove the spindle containing the empty printer paper roll from the printer paper well.
3. Replace the printer paper roll in the spindle. Make sure the new printer paper is advancing from the bottom of the roll.
4. Lift the green knob to lift the print head lever.
5. Insert the paper through the paper guide slot, and make sure it goes under and around the black roller bar.
6. Pull the paper under the print head lever and push the green knob down to push the print head lever down.
7. Press PAPER ADVANCE to move the paper through the printer.
Page 19
Directions for use Cleaning and maintenance 15
8. Install the printer access door, and advance the new paper through the slot on the door.
9. Install the screw on the printer access door.

Maintenance

Routinely perform the following preventive maintenance to keep your scale in working order.
1. Check the calibration annually or as required.
2. Inspect the scale for cracks or loose mounting hardware. Replace or repair as necessary.
3. Visually inspect the scale enclosure for damage or loose or missing hardware. Replace or repair as necessary.
4. If equipped, inspect the AC line cord for abrasions or other signs of wear.
5. Do not expose the scale to excessive water or moisture.
6. Do not store the scale where heavy objects can be placed on it.
7. Replace the batteries annually or as required.
8. Do not service or perform maintenance while the scale is in use with a patient.

Troubleshooting

This section presents tables of information to help you troubleshoot issues with the scale.
To use these tables, locate the specific problem with your scale in the left column of the table. The remainder of the row explains possible causes and suggests actions that can resolve the issue.
Issue
The weight measurement is not accurate
The weight measurement takes too long to display
The scale displays O-LOAd
The measurement does not change when weight is applied to the scale
Possible cause Suggested action
There may be a mechanical obstruction on the platform
The platform may be in motion Wait for the platform to settle.
The object on the scale may be larger than the maximum value allowed
The load cell transducer, connector, or cable may be damaged
Make sure the platform is not touching a foreign object or has anything restricting its downward motion.
If the weight value is within the specified range, contact your service department.
Contact your service department.
The scale buttons are not responding
The error message E-FAIL is displayed
The buttons may be damaged Check for visible signs of
damage.
Contact your service department.
The scale's memory has failed Contact your service department.
Page 20
16 Cleaning and maintenance Welch Allyn In-Floor Scales
Issue Possible cause Suggested action
The error message r-FAIL is displayed
The error message A-FAIL is displayed
The error message C-FAIL is displayed
The scale's memory has failed. Contact your service department.
The scale's memory has failed Contact your service department.
The scale's memory has failed. Contact your service department.
Page 21

Appendices

Specifications

Item Specification
Scales 6102, 6154
17
Resolution (factory default)
Weighing capacity
Readout
Power source
Platform dimensions
0.1 lb / 100 g
6102, 6154: 825 lb (375 kg)
Large, bright light-emitting diode display.
Selectable weight displayed in pounds or kilograms.
Indicator light to show selected units.
Battery
Ni-Cd
7.2V DC, 10.8wh, 1500mAH minimum
AC power
115V ± 10% VAC, 6 watts
Fuse
1/8A 250 V S.B.
6102: 32 in. x 36 in.
6154: 54 in. x 36 in.
Page 22
18
Appendices Welch Allyn In-Floor Scales

Product configurations

The Welch Allyn In-Foor Scales are available in the following configurations:
Model 6102, Flush-Mounted In-Floor Scale
Model Description
6102-KP Flush-Mounted In-Floor Scale, kilogram only, printer
6102-KX Flush-Mounted In-Floor Scale, kilogram only, data port
6102-XP Flush-Mounted In-Floor Scale, standard weight, printer
6102-XP Flush-Mounted In-Floor Scale, standard weight, data port
Model 6154, Flush-Mounted In-Floor Scale
Model Description
6154-KP Flush-Mounted In-Floor Scale, kilogram only, printer
6154-KX Flush-Mounted In-Floor Scale, kilogram only, data port
6154-XP Flush-Mounted In-Floor Scale, standard weight, printer
6154-XP Flush-Mounted In-Floor Scale, standard weight, data port

Approved accessories

Models 6102 and 6154
Part number
61022 Handrail
845010W Wall-mounted height gauge
846112 Printer
23709 Thermal printer paper (box, 15 rolls)
360007 Battery pack 7.2V
Description

Warranty

Welch Allyn will warranty the weight scale to be free of defects in material and workmanship and to perform in accordance with manufacturer specifications for the period of one year from the date of retail purchase.
The warranty period shall start on the date of purchase. The date of purchase is: 1) the invoiced ship date if the device was purchased directly from Welch Allyn, 2) the date
Page 23
Directions for use Appendices 19
specified during product registration, 3) the date of purchase of the product from a Welch Allyn authorized distributor as documented from a receipt from said distributor.
This warranty does NOT cover damages caused by misuse or abuse, including but not limited to:
Failure caused by unauthorized repairs or modifications
Damage caused by shock or dropping during transportation
Damage caused by improper use of the power supply
Failure caused by improper operation not consistent with the instructions stated in this Directions for use