a. Column Installation
b. Column Equilibration
c. Initial Column Efficiency Determination
II. COLUMN USE
a. Guard Columns
b. Sample Preparation
c. Operating pH Limits
d. Solvents
e. Pressure
f. Temperature
III. SCALING UP/DOWN ISOCRATIC METHODS
IV. TROUBLESHOOTING
V. COLUMN CLEANING, REGENERATING AND STORAGE
a. Cleaning and Regeneration
b. Storage
Thank you for choosing a Waters XSelect™ CSH™ HPLC column. XSelect
CSH130 C18, 3.5 µm and 5 µm Columns feature Waters Charged Surface
Hybrid (CSH) Technology which provides excellent peak shape, high
efficiency and loading capacity for basic compounds when using acidic,
low ionic strength mobile phases. This same particle technology is used
in the ACQUITY UPLC CSH130 C18, 1.7 µm and 2.5 µm XP family of
HPLC columns, thus enabling seamless transferability between HPLC and
UPLC® system platforms. The XSelect CSH130 C18 packing materials
were designed and are manufactured in a cGMP, ISO 9001 certified
manufacturing facility using ultra pure reagents. Each batch of XSelect
CSH130 C18 material is tested chromatographically with acidic, basic
and neutral analytes as part of qualification for use in peptide mapping.
The results are held to narrow specification ranges to assure excellent,
reproducible performance. XSelect CSH130 C18 batches are also QC tested
with a gradient separation of a tryptic digest of cytochrome c using 0.1%
formic acid containing eluents. Finally, every shipped column is individually
tested for packed bed efficiency and a Performance Chromatogram and
Certificate of Batch Analysis is available upon request.
VI. CONNECTING THE COLUMN TO THE HPLC
a. Column Connectors and System Tubing Considerations
b. Measuring System Bandspreading Volume
c. Measuring Gradient Delay Volume (or Dwell Volume)
VII. ADDITIONAL INFORMATION
a. Use of Narrow-Bore (2.1 mm i.d.)
b. Impact of Bandspreading Volume on 2.1 mm i.d. Column Performance
c. Non-Optimized vs. Optimized LC/MS/MS System: System
Modification Recommendations
XSelect CSH130 C18 Columns1
[ CARE AND USE MANUAL ]
I. GETTING STARTED
Each XSelect CSH130 C18 Columns comes with a Certificate of Acceptance
and a Performance Test Chromatogram. The Certificate of Acceptance
is specific to each batch of packing material contained in the Peptide
Separation Technology column and includes the batch number, analysis
of unbonded particles, analysis of bonded particles, and chromatographic
results and conditions. The Performance Test Chromatogram is specific to
each individual column and contains the information: batch number, column
serial number, USP plate count, USP tailing factor, retention factor, and
chromatographic conditions. These data should be stored for future reference.
a. Column Installation
Note: The flow rates given in the procedure below are for a typical 5 µm
packing in a 4.6 mm i.d. column. Scale the flow rate up or down accordingly
based upon the column i.d., length, particle size and backpressure of the
Peptide Separation Technology column being installed. See Scaling Up/Down
Isocratic Separations section for calculating flow rates when changing column
i.d and/or length. See Connecting the Column to the HPLC for a more detailed
discussion on HPLC connections.
1. Purge the pumping system of any buffer-containing mobile
phases and connect the inlet end of the column to the injector
outlet. An arrow on the column identification label indicates the
correct direction of solvent flow.
2. Flush column with 100% organic mobile phase (methanol or
acetonitrile) by setting the pump flow rate to 0.1 mL/min. and
increase the flow rate to 1 mL/min over 5 minutes.
b. Column Equilibration
Peptide Separation Technology columns are shipped in 100% acetonitrile.
It is important to ensure mobile phase compatibility before changing to a
different mobile phase system.Equilibrate the column with a minimum of
10 column volumes of the mobile phase to be used (refer to Table 1 for a
listing of empty column volumes).
To avoid precipitating out mobile phase buffers on your column or in your
system, flush the column with five column volumes of a water/organic
solvent mixture, using the same or lower solvent content as in the desired
buffered mobile phase. (For example, flush the column and HPLC system with
60% methanol in water prior to introducing 60% methanol/40% buffer
mobile phase).
c. Initial Column Efficiency Determination
1. Perform an efficiency test on the column before using it in the
desired application. Waters recommends using a suitable solute
mixture, as found in the “Performance Test Chromatogram,” to
analyze the column upon receipt.
2. Determine the number of theoretical plates (N) and use this
value for periodic comparisons.
3. Repeat the test at predetermined intervals to track column
performance over time. Slight variations may be obtained on
two different HPLC systems due to the quality of the connections,
operating environment, system electronics, reagent quality,
column condition and operator technique.
3. When the mobile phase is flowing freely from the column outlet,
stop the flow and attach the column outlet to the detector. This
prevents entry of air into the detection system and gives more
rapid baseline equilibration.
4. Gradually increase the flow rate as described in step 2.
5. Once a steady backpressure and baseline have been achieved,
proceed to the next section.
Note: If mobile phase additives are present in low concentrations (e.g.,
ion-pairing reagents), 100 to 200 column volumes may be required for
complete equilibration. In addition, mobile phases that contain formate
(e.g., ammonium formate, formic acid, etc.) may also require longer initial
column equilibration times.
XSelect CSH130 C18 Columns2
[ CARE AND USE MANUAL ]
Table 1: Empty Column Volumes in mL (multiply by 10 for flush solvent volumes)
Column internal diameter (mm)
Column Length (mm)1.02.14.610193050
500.040.170.833.9143598
1000.080.351.77.82870196
1500.120.522.51242106294
II. COLUMN USE
To ensure the continued high performance of XSelect CSH130 C18 Columns,
follow these guidelines:
a. Guard Columns
Use a Waters guard column of matching chemistry and particle size between
the injector and main column. It is important to use a high performance
matching guard column to protect the main column while not compromising
or changing the analytical resolution.
Guard columns need to be replaced at regular intervals as determined by
sample contamination. When system backpressure steadily increases above
a set pressure limit, it is usually an indication that the guard column
should be replaced. A sudden appearance of split peaks is also indicative
of a need to replace the guard column.
b. Sample Preparation
1. Sample impurities often contribute to column contamination.
One option to avoid this is to use Waters Oasis® solid-phase
extraction cartridges/columns or Sep-Pak® cartridges of the
appropriate chemistry to clean up the sample before analysis.
4. Filter sample with 0.2 µm filters to remove particulates. If the
sample is dissolved in a solvent that contains an organic modifier
(e.g., acetonitrile, methanol, etc.) ensure that the filter material
does not dissolve in the solvent. Contact the filter manufacturer
with solvent compatibility questions. Alternatively, centrifugation
for 20 minutes at 8,000 rpm, followed by the transfer of the
supernatant liquid to an appropriate vial, could be considered.
c. Operating pH Limits
The recommended operating pH range for XSelect CSH130 C18, 3.5 µm
and 5 µm columns are 1 to 11. A listing of commonly used buffers and
additives is given in Table 2. Additionally, the column lifetime will vary
depending upon the operating temperature, the type and concentration of
buffer used.
2. It is preferable to prepare the sample in the operating mobile
phase or a mobile phase that is weaker (less organic modifier)
than the mobile phase for the best peak shape and sensitivity.
3. If the sample is not dissolved in the mobile phase, ensure that
the sample, solvent and mobile phases are miscible in order to
avoid sample and/or buffer precipitation.
XSelect CSH130 C18 Columns3
[ CARE AND USE MANUAL ]
Table 2: Buffer Recommendations for Using XSelect CSH130 C18 Columns from pH 1 to 11
Additive/BufferpKaBuffer range Volatility
TFA0.3VolatileYesIon pair additive, can suppress MS signal, used in the 0.02-0.1% range.
Acetic Acid4.76VolatileYesMaximum buffering obtained when used with ammonium acetate salt. Used in 0.1-1.0% range.
Formic Acid3.75VolatileYesMaximum buffering obtained when used with ammonium formate salt. Used in 0.1-1.0% range.
Acetate (NH4CH2COOH)4.763.76 – 5.76VolatileYesUsed in the 1-10 mM range. Note that sodium or potassium salts are not volatile.
Formate (NH4COOH)3.752.75 – 4.75VolatileYesUsed in the 1-10 mM range. Note that sodium or potassium salts are not volatile.
Phosphate 12.151.15 – 3.15Non-volatileNoTraditional low pH buffer, good UV transparency.
Phosphate 27.26.20 – 8.20Non-volatileNoAbove pH 7, reduce temperature/concentration and use a guard column to maximize lifetime.
Phosphate 312.311.3 - 13.3Non-volatileNoAbove pH 7, reduce temperature/concentration and use a guard column to maximize lifetime.
4-Methylmorpholine~8.47.4 – 9.4VolatileYesGenerally used at 10 mM or less.
Ammonia (NH4OH)9.28.2 – 10.2VolatileYesKeep concentration below 10 mM and temperatures below 30 ˚C.
Ammonium Bicarbonate10.3 (HCO
Ammonium (Acetate)9.28.2 – 10.2VolatileYesUsed in the 1-10 mM range.
Ammonium (Formate)9.28.2 – 10.2VolatileYesUsed in the 1-10 mM range.
Borate9.28.2 – 10.2Non-VolatileNoReduce temperature/concentration and use a guard column to maximize lifetime.
CAPSO9.78.7 – 10.7Non-VolatileNoZwitterionic buffer, compatible with acetonitrile, used in the 1-10 mM range. Low odor.
Glycine2.4, 9.88.8 – 10.8Non-VolatileNoZwitterionic buffer, can give longer lifetimes than borate buffer.
1-Methylpiperidine10.29.3 – 11.3VolatileYesUsed in the 1-10 mM range.
CAPS10.49.5 – 11.5Non-VolatileNoZwitterionic buffer, compatible with acetonitrile, used in the 1-10 mM range. Low odor.
Triethylamine
(as acetate salt)
-
)
8.2 – 11.3VolatileYesUsed in the 5-10 mM range (for MS work keep source >150 ˚C ). Adjust pH with
3
+
9.2 (NH
)
4
10.79.7 – 11.7VolatileYesUsed in the 0.1-1.0% range. Volatile only when titrated with acetic acid (not hydrochloric or phosphoric).
(±1 pH unit)
Used for
Mass Spec
Comments
ammonium hydroxide or acetic acid. Good buffering capacity at pH 10
Note: use ammonium bicarbonate (NH4HCO3), not ammonium carbonate ((NH4)2CO3).
Used as ion-pair for DNA analysis at pH 7-9
d. Solvents
To maintain maximum column performance, use high quality chromatography
grade solvents. Filter all aqueous buffers prior to use. Pall Gelman Laboratory
Acrodisc® filters are recommended. Solvents containing suspended particulate
materials will generally clog the outside surface of the inlet distribution frit of
the column. This will result in higher operating pressure and poor performance.
e. Pressure
XSelect CSH130 C18 columns can tolerate pressures of up to 6,000 psi
(400 bar or 40 Mpa) although pressures greater than 4,000 – 5,000 psi
should be avoided in order to maximize column and system lifetimes.
f. Temperature
Temperatures up to 80 ˚C are recommended for operating XSelect CSH130
C18 columns in order to enhance selectivity, lower solvent viscosity and
increase mass transfer rates. However, any temperature above ambient will
have a negative effect on lifetime which will vary depending on the pH and
buffer conditions used.
XSelect CSH130 C18 Columns4
Loading...
+ 7 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.