The information in this document is subject to change without notice and should not be construed
as a commitment by Waters Corporation. Waters Corporation assumes no responsibility for any
errors that may appear in this document. This document is believed to be complete and accurate
at the time of publication. In no event shall Waters Corporation be liable for incidental or
consequential damages in connection with, or arising from, its use. For the most recent revision
of this document, consult the Waters Web site (waters.com).
Trademarks
ACQUITY® is a registered trademark of Waters Corporation.
ACQUITY UPLC® is a registered trademark of Waters Corporation.
Alliance® is a registered trademark of Waters Corporation.
Connections INSIGHT® is a registered trademark of Waters Corporation.
DART® is a registered trademark of ionSense Inc.
ESCi® is a registered trademark of Waters Corporation.
EdwardsTM is a trademark of Edwards Limited.
GELoader® is a registered trademark of Eppendorf-Netheler-Hinz GmbH.
iKeyTM is a trademark of Waters Corporation.
ionKeyTM is a trademark of Waters Corporation.
IntelliStartTM is a trademark of Waters Corporation.
LDTDTM is a trademark of Phytronix Technologies Inc.
Leybold® is a registered trademark of Oerlikon Leybold Vacuum GmbH.
LockSprayTM is a trademark of Waters Corporation.
MassLynx® is a registered trademark of Waters Corporation.
June 9, 2016, 715004990 Rev. A
Page ii
Nano LCTM is a trademark of Waters Corporation.
nanoACQUITY® is a registered trademark of Waters Corporation.
NanoFlowTM is a trademark of Waters Corporation.
nanoTile® is a registered trademark of Waters Corporation.
Oerlikon® is a registered trademark of OC Oerlikon Corporation AG.
OpenLynxTM is a trademark of Waters Corporation.
PEEKTM is a trademark of Victrex PLC.
PEEKsilTM is a trademark of SGE Analytical Science Pty Ltd.
RADARTM is a trademark of Waters Corporation.
ScanWaveTM is a trademark of Waters Corporation.
StepWaveTM is a trademark of Waters Corporation.
Swagelok® is a registered trademark of Swagelok Company.
SymbiosisTM is a trademark of Spark Holland Inc.
T-WaveTM is a trademark of Waters Corporation.
THE SCIENCE OF WHAT'S POSSIBLE® is a registered trademark of Waters Corporation.
TRIZAIC® is a registered trademark of Waters Corporation.
TargetLynxTM is a trademark of Waters Corporation.
UNIFI® is a registered trademark of Waters Corporation.
UniSprayTM is a trademark of Waters Corporation.
UPLC® is a registered trademark of Waters Corporation.
UltraPerformance LC® is a registered trademark of Waters Corporation.
Viton® is a registered trademark of DuPont Performance Elastomers LLC.
Waters® is a registered trademark of Waters Corporation.
Xevo® is a registered trademark of Waters Corporation.
ZSprayTM is a trademark of Waters Corporation.
All other trademarks or registered trademarks are the sole property of their respective owners.
Customer comments
Waters’ Technical Communications organization invites you to report any errors that you
encounter in this document or to suggest ideas for otherwise improving it. Help us better
June 9, 2016, 715004990 Rev. A
Page iii
understand what you expect from our documentation so that we can continuously improve its
accuracy and usability.
We seriously consider every customer comment we receive. You can reach us at
tech_comm@waters.com.
Contacting Waters
Contact Waters with enhancement requests or technical questions regarding the use,
transportation, removal, or disposal of any Waters product. You can reach us via the Internet,
telephone, or conventional mail.
Waters contact information
Contacting medium Information
InternetThe Waters Web site includes contact information for Waters locations
Telephone and faxFrom the USA or Canada, phone 800-252-4752, or fax 508-872-1990.
worldwide.
Visit www.waters.com
For other locations worldwide, phone and fax numbers appear in the
Waters Web site.
Conventional mailWaters Corporation
Global Support Services
34 Maple Street
Milford, MA 01757
USA
System manufacturing information
Manufacturer:
Waters Corporation
34 Maple Street
Milford, MA 01757
USA
Manufacturing site:
Waters Technologies Ireland Ltd.
Wexford Business Park
June 9, 2016, 715004990 Rev. A
Page iv
Drinagh, Wexford, Ireland
Safety considerations
Some reagents and samples used with Waters instruments and devices can pose chemical,
biological, or radiological hazards (or any combination thereof). You must know the potentially
hazardous effects of all substances you work with. Always follow Good Laboratory Practice
(GLP), and consult your organization’s standard operating procedures as well as your local
requirements for safety.
Considerations specific to the device
Power cord replacement hazard
Warning: To avoid electric shock, use the SVT-type power cord in the United States
and HAR-type (or better) cord in Europe. The main power cord must be replaced only
with one of adequate rating. For information regarding what cord to use in other
countries, contact your local Waters distributor.
Solvent leakage hazard
The source exhaust system is designed to be robust and leak-tight. Waters recommends you
perform a hazard analysis assuming a maximum leak into the laboratory atmosphere of 10% LC
eluate.
Warning: To avoid exposure to toxic substances and biohazards from O-ring leaks in the
source exhaust system, observe these precautions:
• Replace the source O-rings at intervals not exceeding one year.
• Prevent chemical degradation of the source O-rings, which can withstand exposure only to
certain solvents, by determining whether any solvents you use are chemically compatible with
the composition of the O-rings.
Bottle placement prohibition
Warning: To avoid injury from electrical shock or fire, and damage to the equipment, do not
place vessels containing liquid atop the workstation or ancillary equipment or otherwise expose
those units to dripping or splashing liquids.
June 9, 2016, 715004990 Rev. A
Page v
Prohibited: Do not place vessels containing liquid—such as solvent bottles—atop the
workstation or ancillary equipment or otherwise expose those units to dripping or
splashing liquids.
Spilled solvents hazard
Prohibited: To avoid equipment damage caused by spilled solvent, do not place
reservoir bottles directly atop an instrument or device or on its front ledge. Instead,
place the bottles in the bottle tray, which serves as secondary containment in the event
of spills.
Flammable solvents hazard
Warning: To prevent the ignition of flammable solvent vapors in the enclosed space of
a mass spectrometer’s ion source, ensure that these conditions are met:
• Nitrogen flows continuously through the source.
• A gas-fail device is installed, to interrupt the flow of LC solvent should the nitrogen
supply fail.
• The nitrogen supply pressure does not fall below 400 kPa (4 bar, 58 psi) during an
analysis requiring the use of flammable solvents.
When using flammable solvents, ensure that a stream of nitrogen continuously flushes the
instrument’s source, and the nitrogen supply pressure remains above 400 kPa (4 bar, 58 psi).
You must also install a gas-fail device that interrupts the solvent flowing from the LC system in
the event the supply of nitrogen fails.
Glass breakage hazard
Warning: To avoid injuries from broken glass, falling objects, or exposure to toxic substances,
never place containers on top of the instrument or on its front covers.
High temperature hazard
Warning: To avoid burn injuries, avoid touching the source ion block assembly when
operating or servicing the instrument.
June 9, 2016, 715004990 Rev. A
Page vi
Mass spectrometer high temperature hazard
Source ion block assembly
Hazards associated with removing an instrument from service
Warning: To avoid personal contamination with biohazards, toxic materials, and corrosive
materials, wear chemical-resistant gloves when performing this procedure.
Warning: To avoid eye injury from broken fused silica lines, use eye protection when
performing this procedure.
When you remove the instrument from use to repair or dispose of it, you must decontaminate all
of its vacuum areas. These are the areas in which you can expect to encounter the highest levels
of contamination:
• Source interior
• Waste tubing
• Exhaust system
• Rotary pump oil (where applicable)
The need to decontaminate other vacuum areas of the instrument depends on the kinds of
samples the instrument analyzed and their levels of concentration. Do not dispose of the
instrument or return it to Waters for repair until the authority responsible for approving its removal
from the premises specifies the extent of decontamination required and the level of residual
June 9, 2016, 715004990 Rev. A
Page vii
contamination permissible. That authority must also prescribe the method of decontamination to
be used and the appropriate protection for personnel undertaking the decontamination process.
You must handle items such as syringes, fused silica lines, and borosilicate tips used to carry
sample into the source area in accordance with laboratory procedures for contaminated vessels
and sharps. To avoid contamination by carcinogens, toxic substances, or biohazards, you must
wear chemical-resistant gloves when handling or disposing of used oil.
Electrical power safety notice
Do not position the instrument so that it is difficult to disconnect the power cord.
Safety hazard symbol notice
Documentation needs to be consulted in all cases where the symbol is used to find out the
nature of the potential hazard and any actions which have to be taken.
Equipment misuse notice
If equipment is used in a manner not specified by its manufacturer, protections against personal
injury inherent in the equipment’s design can be rendered ineffective.
Safety advisories
Consult the "Safety advisories" appendix in this publication for a comprehensive list of warning
advisories and notices.
Operating this device
When operating this device, follow standard quality-control (QC) procedures and the guidelines
presented in this section.
Applicable symbols
SymbolDefinition
Manufacturer
June 9, 2016, 715004990 Rev. A
Page viii
SymbolDefinition
REF
Date of manufacture
Authorized representative of the European Community
Confirms that a manufactured product complies with all applicable
European Community directives
Australia EMC compliant
or
Confirms that a manufactured product complies with all applicable United
States and Canadian safety requirements
Consult instructions for use
Alternating current
Electrical and electronic equipment with this symbol may contain
hazardous substances and should not be disposed of as general waste.
For compliance with the Waste Electrical and Electronic Equipment
Directive (WEEE) 2012/19/EU, contact Waters Corporation for the correct
disposal and recycling instructions.
Serial number
Part number catalog number
Audience and purpose
This guide is for operators of varying levels of experience. It gives an overview of the device and
explains how to prepare it for operation, change its modes of operation, and maintain it.
Intended use of the device
Waters designed the Xevo TQ-XS for use as a research tool to accurately, reproducibly, and
robustly quantify target compounds present at the lowest possible levels in highly complex
sample matrices. It is not for use in diagnostic procedures.
June 9, 2016, 715004990 Rev. A
Page ix
When fitted with Waters options (APCI, APGC, APPI, ASAP, ESCi, NanoFlow ESI, TRIZAIC,
UniSpray, UPLC, ionKey), or optional third-party sources (DART, DESI, or LDTD), the Xevo TQXS does not comply with the European Union In Vitro Diagnostic Device Directive 98/79/EC.
Calibrating
To calibrate LC systems, adopt acceptable calibration methods using at least five standards to
generate a standard curve. The concentration range for standards must include the entire range
of QC samples, typical specimens, and atypical specimens.
When calibrating mass spectrometers, consult the calibration section of the operator’s guide for
the instrument you are calibrating. In cases where an overview and maintenance guide, not an
operator’s guide, accompanies the instrument, consult the instrument’s online Help system for
calibration instructions.
Quality control
Routinely run three QC samples that represent subnormal, normal, and above-normal levels of a
compound. If sample trays are the same or very similar, vary the location of the QC samples in
the trays. Ensure that QC sample results fall within an acceptable range, and evaluate precision
from day to day and run to run. Data collected when QC samples are out of range might not be
valid. Do not report these data until you are certain that the instrument performs satisfactorily.
EMC considerations
FCC radiation emissions notice
Changes or modifications not expressly approved by the party responsible for compliance, could
void the user's authority to operate the equipment. This device complies with Part 15 of the FCC
Rules. Operation is subject to the following two conditions: (1) this device may not cause harmful
interference, and (2) this device must accept any interference received, including interference that
may cause undesired operation.
Canada spectrum management emissions notice
This class A digital product apparatus complies with Canadian ICES-001.
Cet appareil numérique de la classe A est conforme à la norme NMB-001.
June 9, 2016, 715004990 Rev. A
Page x
ISM classification: ISM group 1 class A
This classification has been assigned in accordance with IEC CISPR 11 Industrial Scientific and
Medical (ISM) instruments requirements.
Group 1 products apply to intentionally generated and/or used conductively coupled radiofrequency energy that is necessary for the internal functioning of the equipment.
Class A products are suitable for use in all establishments other than residential locations and
those directly connected to a low voltage power supply network supplying a building for domestic
purposes.
There may be potential difficulties in ensuring electromagnetic compatibility in other environments
due to conducted as well as radiated disturbances.
EMC grounding requirements
Notice: To avoid difficulties in ensuring electromagnetic compatibility, if the
instrument's pump control cable is attached to the vacuum hose, ensure that the cable
is grounded to the mass spectrometer.
EC authorized representative
AddressWaters Corporation
Stamford Avenue
Altrincham Road
Wilmslow SK9 4AX UK
Telephone+44-161-946-2400
Fax+44-161-946-2480
ContactQuality manager
June 9, 2016, 715004990 Rev. A
Page xi
Table of contents
General information .......................................................................................................ii
Copyright notice ..................................................................................................................................... ii
Trademarks............................................................................................................................................ ii
1.4.2 System components ............................................................................................................28
1.4.3 System operation................................................................................................................. 28
1.5 Ion optics.......................................................................................................................................28
1.6 MS operating modes ..................................................................................................................... 29
The Xevo TQ-XS is a tandem quadrupole, atmospheric pressure ionization (API) mass
spectrometer. It is designed for routine HPLC/MS/MS and UPLC/MS/MS analyses in quantitative
and qualitative applications, and can operate at fast acquisition speeds compatible with
UltraPerformance LC.
You can use theXevo TQ-XS with the following high-performance ZSpray dual-orthogonal API
sources:
• Standard multi-mode electrospray ionization/atmospheric pressure chemical ionization/
combined electrospray ionization and atmospheric pressure chemical ionization (ESI/APCI/
ESCi)
Requirement: Dedicated APCI operation requires an additional probe.
• Optional atmospheric pressure gas chromatography (APGC)
• Optional TRIZAIC UPLC
• Optional ionKey source
You can also use the Xevo TQ-XS with the following optional third-party sources:
• Direct analysis in real time (DART)
• Desorption electrospary ionization (DESI)
• Laser diode thermal desorption (LDTD)
For additional details, refer to the appropriate manufacturer’s documentation.
Available source options can vary, depending on the software you use to operate the Xevo TQXS. Refer to the MassLynx or UNIFI online Help for more information about supported sources.
For mass spectrometer specifications, see the Waters Xevo TQ-XS Site Preparation Guide
(715005172).
June 9, 2016, 715004990 Rev. A
Page 19
Figure 1–1: Waters Xevo TQ-XS
Source enclosure
1.1IntelliStart technology
IntelliStart technology monitors instrument performance and indicates when the instrument is
ready for use.
The software automatically tunes and mass calibrates the instrument, displays performance
readbacks, and enables simplified setup of the system for use in routine analytical and openaccess applications.
The IntelliStart fluidics system1 is built into the mass spectrometer. It delivers sample directly to
the MS probe from the LC column or from three integral reservoirs. The integral reservoirs can
also deliver sample through direct or combined infusion, enabling you to optimize instrument
performance at analytical flow rates.
See IntelliStart fluidics system and the mass spectrometer’s online Help for further details on
IntelliStart technology.
1
In Waters documents, the term “fluidics” refers to the IntelliStart Fluidics system, which is the instrument’s onboard system that
delivers sample and solvent to the probe of the mass spectrometer. It can also denote plumbing components and fluid pathways
within and between system modules.
June 9, 2016, 715004990 Rev. A
Page 20
1.2ACQUITY UPLC/MS Xevo TQ-XS systems
The Waters Xevo TQ-XS is compatible with the ACQUITY UPLC systems. If you are not using an
ACQUITY UPLC system, refer to the documentation relevant to your LC system.
1.2.1 ACQUITY UPLC system
The ACQUITY UPLC system includes a binary or quaternary solvent manager, sample manager,
column heater or column manager, optional sample organizer, one or more detectors, a
specialized ACQUITY UPLC column, and software to control the system.
For additional information, see the ACQUITY UPLC System Operator's Guide or ControllingContamination in UltraPerformance LC/MS and HPLC/MS Systems (part number 715001307).
You can find these documents on www.waters.com; click Services & Support > Support.
1.2.2
Waters ACQUITY Xevo TQ-XS UPLC/MS system
Figure 1–2: Waters ACQUITY Xevo TQ-XS UPLC/MS System
June 9, 2016, 715004990 Rev. A
Page 21
Probe adaptor
Source enclosure
Source enclosure release
Xevo TQ-XS
Access door to the fluidics
Sample manager
Binary solvent manager or Quaternary solvent manager
Sample organizer (optional on the ACQUITY UPLC system)
Column heater
Solvent tray
Access door to the fluidics valve
Removable panel for ACQUITY arm
Probe high voltage connector
Source interface sliding door
1.2.3 ACQUITY UPLC M-Class system
The ACQUITY UPLC M-Class system is designed for nano-scale and micro-scale separations.
M-Class system components are optimized for use with sub-2µm particle liquid chromatography
and use reduced fluid volumes. The supported flow rate for a gradient elution ranges from 200
nL/min to 100 µL/min at 15,000 psi.
For further information, see the ACQUITY UPLC M-Class System Guide or ControllingContamination in UltraPerformance LC/MS and HPLC/MS Systems (part number 715001307).
You can find these documents on www.waters.com; click Services & Support > Support.
1.2.4
Non-ACQUITY devices for use with the Xevo TQ-XS
The following non-ACQUITY LC devices are validated for use with the Xevo TQ-XS:
You can use MassLynx software v4.2 to control the mass spectrometer. The software enables
these major operations:
• Configuring the system
• Creating LC and MS/MS methods that define operating parameters for a run
• Using IntelliStart software to automatically tune and mass calibrate the mass spectrometer
• Running samples
• Acquiring data
• Monitoring the run
• Processing data
• Reviewing data
1.2.5.1
1.3
• Printing data
MassLynx software
MassLynx software acquires, analyzes, manages, and distributes mass spectrometry, ultraviolet
(UV), evaporative light scattering (ELS), and analog data. OpenLynx and TargetLynx XS
application managers are included with MassLynx software.
See the MassLynx software user documentation and online Help for information about using
MassLynx software.
You configure settings, monitor performance, run diagnostic tests, and maintain the system and
its modules via the MassLynx Instrument Control application.
The Instrument Control software, which functions independently of MassLynx software, does not
recognize or control data systems.
See the online Help for the Instrument Console system for additional details.
Ionization techniques and source probes
Note: Available source options can vary depending on the software used to operate the Xevo
TQ-XS. Refer to the instrument software's online Help for more information about supported
sources.
June 9, 2016, 715004990 Rev. A
Page 23
1.3.1 Electrospray ionization
In electrospray ionization (ESI), a strong electrical charge is applied to the eluent as it emerges
from a nebulizer. The droplets that compose the resultant aerosol undergo a reduction in size
(solvent evaporation). As solvent continues to evaporate, the charge density increases until the
droplet surfaces eject ions (ion evaporation). The ions can be singly or multiply charged.
To operate the source in ESI mode, you fit the source enclosure with an ESI probe adaptor and
ESI probe assembly.
The standard ESI probe assembly accommodates flow rates of up to 2 mL/min, making it suitable
for LC applications in the range of 100 µL/min to 2 mL/min. To reduce peak broadening for lowerflow-rate LC applications, such as 1-mm UPLC columns, use the optional, small-bore capillary,
which can accommodate a maximum flow rate of 200 µL/min.
See also: ESI, ESCi, and APCI modes for further details.
1.3.2
1.3.3
1.3.4
ESCi
ESCi mode is supplied as standard equipment on the mass spectrometer. In ESCi, the standard
ESI probe adaptor is used in conjunction with a corona pin, to allow alternating acquisition of ESI
and APCI ionization data, which facilitates high throughput and wider compound coverage.
See ESI, ESCi, and APCI modes for further details.
APCI
An optional dedicated high-performance APCI interface is available. APCI produces singly
charged protonated or deprotonated molecules for a broad range of nonvolatile analytes.
The APCI interface consists of the ESI/APCI/ESCi enclosure fitted with a corona pin and an APCI
probe adaptor.
See ESI, ESCi, and APCI modes for further details.
Dual-mode APPI/APCI source
The optional, combined APPI/APCI source comprises an APCI probe adaptor and the APPI lamp
drive assembly. The APPI lamp drive assembly comprises a UV lamp and a repeller electrode. In
addition, a specially shaped, dual, APPI/APCI corona pin can be used. You can operate the
source in APPI, APCI, or dual mode, which switches rapidly between APPI and APCI ionization
modes.
See Combined APPI/APPI source for further details.
June 9, 2016, 715004990 Rev. A
Page 24
1.3.5 UniSpray
The UniSpray source enables the detection of a wide range of compounds in a single analysis. In
contrast to Electrospray ionization, UniSpray uses a grounded capillary, and the resulting spray is
directed at an impactor pin held at a voltage, creating smaller charged droplets, amenable to easy
desolvation.
See UniSpray source for further details.
1.3.6
1.3.7
Low-flow ESI probe
The optional low-flow ESI probe is fitted with a narrow bore capillary suitable for use with flow
rates from 5 µL/min to 100 µL/min. Its probe tip is optimized for this capillary.
The low-flow ESI probe replaces the standard ESI probe in the instrument’s source housing.
See the Low-flow ESI Probe Operator's Guide for further details.
NanoFlow ESI source
NanoFlow is the name given to several techniques that use low flow rate ESI. The NanoFlow
source allows ESI in the flow rate range of 5 to 1,000 nL/min. For a given sample concentration,
the ion currents observed approximate those seen in normal flow rate electrospray. However, for
similar experiments, NanoFlow’s significant reduction in sample consumption accompanies
significant increases in sensitivity.
The following options are available for the spraying capillary:
• Universal nebulizer sprayer (Nano LC).
This option is for flow injection or for coupling to nano-UPLC. It uses a pump to regulate the
flow rate downward to 100 nL/min. If a syringe pump is used, a gas-tight syringe is necessary
to effect correct flow rates without leakage. A volume of 250 µL is recommended.
1.3.8
• Borosilicate glass capillaries (nanovials).
Metal-coated, glass capillaries allow the lowest flow rates. They are usable for one sample,
and then must be discarded.
• Capillary Electrophoresis (CE) or Capillary Electrochromatography (CEC) sprayer.
This option uses a make-up liquid at the capillary tip that provides a stable electrospray. The
make-up flow rate is less than 1 µL/min.
See NanoFlow ESI source for further details.
Atmospheric solids analysis probe (ASAP)
The ASAP facilitates rapid analysis of volatile and semivolatile compounds in solids, liquids, and
polymers. It is particularly suited to analyzing low-polarity compounds. The ASAP directly
June 9, 2016, 715004990 Rev. A
Page 25
replaces the ESI or APCI probe in the instrument’s source housing and has no external gas or
electrical connections.
See the Atmospheric Solids Analysis Probe Operator's Guide Supplement for further details.
1.3.9 APGC
The Waters APGC couples an Agilent GC with the Xevo TQ-XS. Doing so enables you to perform
LC and GC analyses in the same system, without compromising performance. The APGC
provides complementary information to the LC/MS instrument, enabling analysis of compounds of
low molecular weight and low-to-intermediate polarity.
See the Atmospheric Pressure GC Operator's Guide Supplement for further details.
1.3.10
1.3.11
TRIZAIC UPLC source
The TRIZAIC UPLC source accepts a nanoTile device, which combines the functions of an
analytical column, trapping column, and nanospray emitter. This technology simplifies the
implementation of capillary-scale chromatography and analysis of limited-volume samples.
See the TRIZAIC UPLC System Guide for further details.
ionKey source
The ionKey source integrates UPLC separation into the source of the mass spectrometer. The
source accepts an iKey Separation Device, which contains the fluidic connections, electronics,
ESI interface, heater, e-cord, and chemistry. Inserting the iKey simultaneously engages the
electronic and fluidic connections. This technology eliminates the need to manually connect
electronic cables and tubing, and simplifies the user experience.
See the ACQUITY UPLC M-Class System Guide (part number 715003588) and the ionKey/MSSystem Guide (part number 715004028) for further details.
See also: ionKey source.
1.4
IntelliStart fluidics system
1.4.1 Overview
The IntelliStart fluidics system is a solvent delivery system built into the mass spectrometer. It
delivers sample directly to the MS probe in one of these ways:
June 9, 2016, 715004990 Rev. A
Page 26
• From the LC column.
• From three integral reservoirs. (The reservoirs can also deliver sample, by direct or combined
infusion, to enable optimization at analytical flow rates.)
• From a wash reservoir that contains solvent for automated flushing of the instrument’s solvent
delivery system.
For further information on the IntelliStart fluidics system, see IntelliStart Fluidics Plumbing and the
diagram located on the inside of the fluidics access door (see Waters ACQUITY Xevo TQ-XS
UPLC/MS system).
Figure 1–3: IntelliStart fluidics system:
Reservoir C
Reservoir B
Reservoir A
Pump
Wash bottle, located in solvent tray
To waste system
LC
Column
Diverter valve
Probe
7-port selector valve
June 9, 2016, 715004990 Rev. A
Page 27
1.4.2 System components
The onboard system incorporates a 7-port selector valve, a multi-position diverter valve, a pump,
and three sample reservoirs.
The sample reservoirs are mounted on the instrument’s front panel. When you select a solvent
from the instrument console, an LED illuminates the appropriate reservoir. You can
simultaneously illuminate all three reservoirs or extinguish the LEDs for light-sensitive samples.
Recommendation: Use reservoir A for the calibrant solution, reservoir B for tuning
compounds, and reservoir C for analyte/optimization solution.
1.4.3
System operation
The software automatically controls solvent and sample delivery during auto-tuning, autocalibration, and method development, via the instrument console.
See the mass spectrometer’s online Help for additional details on IntelliStart software and
operation of the instrument’s solvent delivery system.
1.5Ion optics
The mass spectrometer’s ion optics operate as follows:
1.Samples from the LC or instrument’s solvent delivery system are introduced at atmospheric
pressure into the ionization source, where they are ionized.
2.The ions pass through the sample cone into the vacuum system.
3.The resulting ion beam passes through the source sampling orifice, undergoing a certain
amount of expansion.
4.The ion beam then passes into the entrance of the StepWave transfer optics. The entrance
is large enough to efficiently capture ions in the expanded beam. The StepWave transfer
optics comprise two stages. The first stage (conjoined ion guide) focuses the ion beam and
directs it to the second stage (T-Wave ion guide). The off-axis design ensures that any
neutral materials entering the source sampling orifice are actively extracted from the
system.
5.The ions then pass to the first quadrupole, where they can be filtered according to their
mass-to-charge ratio (m/z).
6.The mass-separated ions pass into the T-Wave/ScanWave collision cell, where they
undergo collision-induced dissociation (CID) or pass to the second quadrupole. Any
fragment ions can then be mass-analyzed by the second quadrupole.
7.The transmitted ions are detected by the photomultiplier detection system.
8.The signal is amplified, digitized, and sent to the mass spectrometry software:
June 9, 2016, 715004990 Rev. A
Page 28
Quadrupole 1 (MS1)
T-Wave/ScanWave collision cell
Quadrupole 2 (MS2)
Conversion dynode
Detector assembly
Photomultiplier tube
Source sampling orifice
Isolation valve
Z-Spray ion source
Sample inlet
Sample cone
Conjoined ion guide
StepWave
T-Wave ion guide
1.6MS operating modes
The following table shows the MS operating modes.
June 9, 2016, 715004990 Rev. A
Page 29
Table 1–1: MS operating modes:
Operating modeMS1Collision cellMS2
MSPass all massesResolving (scanning)
SIRPass all massesResolving (static)
MS1Resolving (scanning)Pass all masses
In MS mode, the instrument can acquire data at scan speeds as high as 20,000 Da/s. Use this
mode for instrument tuning and calibration before MS/MS analysis. See the mass spectrometer’s
online Help for further information.
Use the selected ion recording (SIR) mode for quantitation when you cannot find a suitable
fragment ion to perform a more specific multiple reaction monitoring (MRM) analysis (see MS/MS
operating modes for further details) . In SIR and MRM modes, neither quadrupole is scanned,
therefore no spectrum (intensity versus mass) is produced. The data obtained from SIR or MRM
analyses derive from the chromatogram plot (specified mass intensity versus time).
1.7MS/MS operating modes
The following table shows the MS/MS operating modes.
Table 1–2: MS/MS operating modes:
Operating modeMS1Collision cellMS2
Product (daughter)
ion spectrum
Precursor (parent)
ion spectrum
MRMStatic (at precursor
Constant neutral
loss spectrum
Constant neutral
gain spectrum
ScanWave
daughter scan
Static (at precursor
mass)
ScanningStatic (at product
mass)
Scanning
(synchronized with
MS2)
Scanning
(synchronized with
MS2)
Static (at precursor
mass)
Fragment precursor ions
and pass all masses
Scanning
Scanning
mass)
Static (at product
mass)
Scanning
(synchronized with
MS1)
Scanning
(synchronized with
MS1)
(synchronized with
collision cell)
June 9, 2016, 715004990 Rev. A
Page 30
Loading...
+ 183 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.