The information in this document is subject to change without no tice and should not be
construed as a commitment by Waters Corporation. Waters Corporation assumes no
responsibility for any errors that may appear in this document. This document is believed
to be complete and accurate at the time of publication. In no event shall Waters
Corporation be liable for incidental or consequential damages in connection with, or
arising from, its use. For the most recent revision of this document, consult the Waters
Web site (waters.com).
Trademarks
ACQUITY, ACQUITY UltraPerformance LC, ACQUITY UPLC, Alliance, Connections
INSIGHT , ESCi, MassLynx, “THE SCIENCE OF WHAT’S POSSIBLE”, UPLC, Waters,
Waters Quality Parts, and Xevo are registered trademarks of Waters Corporation, and
iKey, IntelliStart, ionKey, ionKey/MS, IonSABRE, NanoFlow, RADAR, T-Wave, and
ZSpray are trademarks of Waters Corporation.
DART is a registered trademark of JEOL USA Inc.
GELoader is a registered trademark of New Brunswick Scientific, Co., Inc.
LDTD is a trademark of Phytronix Technologies Inc.
PEEK is a trademark of Victrex Corporation.
Phillips and Pozidriv are registered trademarks of Phillips Screw Company, Inc.
snoop and Swagelok are registered trademarks of Swagelok Company.
Spark Holland and Symbiosis are trademarks of Spark-Holland BV.
Viton is a registered trademark of DuPont.
Other registered trademarks or trademarks are the sole property of their owners.
January 11, 2016, 715004599 Rev. Biii
Page 4
Customer comments
Waters’ Technical Communications organization invites you to report any errors that you
encounter in this document or to suggest ideas for otherwise improving it. Help us better
understand what you expect from our documentation so that we can continuously improve
its accuracy and usability.
We seriously consider every customer comment we receive. You can reach us at
tech_comm@waters.com.
Contacting Waters
Contact Waters® with enhancement requests or technical questions regarding the use,
transportation, removal, or disposal of any Waters product. You can reach us via the
Internet, telephone, or conventional mail.
Waters contact information:
Contacting mediumInformation
InternetThe Waters Web site includes contact information for
Telephone and faxFrom the USA or Canada, phone 800 252-4752, or
Waters locations worldwide. Visit www.waters.com.
fax 508 872 1990.
For other locations worldwide, phone and fax
numbers appear in the Waters Web site.
Conventional mailWaters Corporation
34 Maple Street
Milford, MA 01757
USA
Safety considerations
Some reagents and samples used with Waters instruments and devices can pose chemical,
biological, or radiological hazards (or any combination thereof). You must know the
potentially hazardous effects of all substances you work with. Always follow Good
Laboratory Practice, and consult your organization’s standard operating procedures.
ivJanuary 11, 2016, 715004599 Rev. B
Page 5
Safety hazard symbol notice
Documentation needs to be consulted in all cases where the symbol is used to find
out the nature of the potential hazard and any actions which have to be taken.
Considerations specific to the Xevo TQ-S micro
Power cord replacement hazard
Warning: To avoid electric shock, use the SVT-type power cord in the United
States and HAR-type (or better) in Europe. The main power cord must only be
replaced with one of adequate rating. For information regarding what cord to use in
other countries, contact your local Waters distributor.
Solvent leakage hazard
The source exhaust system is designed to be robust and leak-tight. Waters recommends
you perform a hazard analysis, assuming a maximum leak into the laboratory atmosphere
of 10% LC eluate.
Warning:
•To confirm the integrity of the source exhaust system, renew the source
O-rings at intervals not exceeding one year.
•To avoid chemical degradation of the source O-rings, which can
withstand exposure only to certain solvents (see page 258), determine
whether any solvents you use that are not listed are chemically
compatible with the composition of the O-rings.
January 11, 2016, 715004599 Rev. Bv
Page 6
Spilled solvent s hazard
Prohibited: To avoid injury or equipment damage caused by spilled solvent, do
not place reservoir bottles on top of the instrument or on its front ledge.
Flammable solvents hazard
Warning: To prevent ignition of flammable solvent vapors in the enclosed space
of a mass spectrometer’s ion source, ensure that nitrogen flows continuously
through the source. The nitrogen supply pressure must not fall below 400 kPa
(4 bar, 58 psi) during an analysis requ iring the use of flammable solvents. Also a
gas-fail device must be installed, to interrupt the flow of LC solvent should the
nitrogen supply fail.
When using flammable solvents, ensu re that a stream of nitrogen continuously flushes the
instrument’s source, and the nitrogen supply pressure remains above 400 kPa (4 bar,
58 psi). You must also install a gas-fail device that interrupts the solvent flowing from the
LC system in the event the supply of nitrogen fails.
Overload hazard
Warning: To prevent personal injury, ensure equipment placed on top of the Xevo
Warning: To avoid injuries from broken glass, falling objects, or
exposure to toxic or biohazardous substances, never place containers
on top of the instrument or on its front covers.
High temperature hazard
Warning: To avoid burn injuries, ensure the source heater is turned off and the ion
block is cool before performing maintenance on these components. The source ion
block, located behind the source enclosure assembly, can become hot.
Xevo TQ-S micro high temperature hazard:
January 11, 2016, 715004599 Rev. Bvii
Page 8
Hazards associated with removing an instrument from service
Warning: To avoid personal contamination with biohazards, toxic
materials, or corrosive materials, wear chemical-resistant gloves
during all phases of instrument decontamination.
Warning: To avoid puncture injuries, handle syringes, fused silica
lines, and borosilicate tips with extreme care.
When you remove the instrument from use to repair or dispose of it, you must
decontaminate all of its vacuum areas. These are the areas in which you can expect to
encounter the highest levels of contamination:
•Source interior
•Waste tubing
•Exhaust system
•Rotary pump oil (where applicable)
The need to decontaminate other vacuum areas of the instrument depends on the kinds of
samples the instrument analyzed and their levels of concentration. Do not dispose of the
instrument or return it to Waters for repair until the authority responsible for approving its
removal from the premises specifies the extent of decontamination required and the level
of residual contamination permissible. That authority must also prescribe the method of
decontamination to be used and the appropriate protection for personnel undertaking the
decontamination process.
You must handle items such as syringes, fused silica lines, and borosilicate tips used to
carry sample into the source area in accordance with laboratory procedures for
contaminated vessels and sharps. To avoid contamination by carcinogens, toxic
substances, or biohazards, you must wear chemical-resistant gloves when handling or
disposing of used oil.
Bottle placement prohibition
Prohibited: To avoid injury from electric shock or fire, and to prevent damage to
the workstation and ancillary equipment, do not place objects filled with
liquid—such as solvent bottles—on these items, or expose them to dripping or
splashing liquids.
viiiJanuary 11, 2016, 715004599 Rev. B
Page 9
FCC radiation emissions notice
Changes or modifications not expressly approved by the party responsible for compliance,
could void the users authority to operate the equipment. This device complies with Part 15
of the FCC Rules. Operation is subject to the following two conditions: (1) this device
may not cause harmful interference, and (2) this device must accept any interference
received, including interference that may cause undesired operation.
Electrical power safety notice
Do not position the instrument so that it is difficult to operate the disconnecting device.
Equipment misuse notice
If the equipment is used in a manner not specified by the manufacturer, the protection
provided by the equipment may be impaired.
Safety advisories
Consult Appendix A for a comprehensive list of warning advisories and notices.
January 11, 2016, 715004599 Rev. Bix
Page 10
Operating this instrument
When operating this instrument, follow standard quality-control (QC) procedures and the
guidelines presented in this section.
Applicable symbols
SymbolDefinition
Manufacturer
Date of manufacture
Part number catalog number
5()
Serial number
Supply ratings
Authorized representative of the European Community
Confirms that a manufactured product complies with all
applicable European Community directives
Australia EMC Compliant
or
Confirms that a manufactured product complies with all
applicable United States and Canadian safety requirements
xJanuary 11, 2016, 715004599 Rev. B
Page 11
SymbolDefinition
Consult instructions for use
Electrical and electronic equipment with this symbol may
contain hazardous substances and should not be disposed
of as general waste.
For compliance with the Waste Electrical and Electronic
Equipment Directive (WEEE) 2012/19/EU, contact W aters
Corporation for the correct disposal and recycling
instructions.
January 11, 2016, 715004599 Rev. Bxi
Page 12
Audience and purpose
This guide is for operators of varying levels of experience. It gives an overview of the
instrument and explains how to prepare it for operation, switch between modes of
operation, and maintain it.
Intended use of the Xevo TQ-S micro
Waters designed the Xev o TQ-S mic r o for use as a research tool to accurately,
reproducibly, and robustly quantify target compounds present at the lowest possible levels
in highly complex sample matrices. The Xevo TQ-S micro is not intended for use in
diagnostic applications.
Calibrating
To calibrate LC systems, follow accepta ble calibration methods using at least five
standards to generate a standard curve. The concentration range for standards should
include the entire range of QC samples, typical specimens, and atypical specimens.
When calibrating mass spectrometers, consult the instrument’s online Help system for
instructions.
Quality control
Routinely run three QC samples that represent subnormal, normal, and above-normal
levels of a compound. If sample trays are the same or very similar, vary the location of the
QC samples in the trays. Ensure that QC sample results fall within an acceptable range,
and evaluate precision from day to day and run to run. Data collected when QC samples
are out of range might not be valid. Do not report these data until you are certain that the
instrument performs satisfactorily.
When analyzing samples from a complex matrix such as soil, tissue, serum/plasma, whole
blood, and other sources, note that the matrix components can adversely affect LC/MS
results, enhancing or suppressing ionization. To minimize these matrix effects, adopt the
following measures:
•Prior to the instrumental analysis, use appropriate sample pretreatment such as
protein precipitation, liquid/liquid extraction (LLE), or solid phase extraction (SPE)
to remove matrix interferences.
•Whenever possible, verify method accuracy and precision using matrix-matched
calibrators and QC samples.
xiiJanuary 11, 2016, 715004599 Rev. B
Page 13
•Use one or more internal standard compounds, preferably isotopically labeled
analytes.
EMC considerations
Canada spectrum management emissions notice
This class A digital product apparatus complies with Canadian ICES-001.
Cet appareil numérique de la classe A est conforme à la norme NMB-001.
ISM Classification: ISM Group 1 Class A
This classification has been assigned in accordance with CISPR 11 Industrial Scientific
and Medical, (ISM) instrument requirements. Group 1 products apply to intentionally
generated and/or used conductively coupled radio-frequency energy that is necessary for
the internal functioning of the equipment. Class A products are suitable for use in all
establishments other than residential locations and those directly connected to a low
voltage power supply network supplying a building for domestic purposes.
There may be potential difficulties in ensuring electromagnetic compatibility in other
environments due to conducted as well as radiated disturbances.
Do not use the equipment in close proximity to sources of strong electromagnetic radiation
(for example, unshielded intentional RF sources), as these may interfere with the
equipment’s proper operation.
This equipment complies with the emission and immunity requirements described in the
relevant parts of IEC/EN 61326: Electrical equipment for measurement, control and
Spare parts ............................................................................................................................ 93
Troubleshooting with Connections INSIGHT ................................................................... 94
Safety and handling ............................................................................................................. 95
Preparing the instrument for operations on or inside its source ..................................... 97
Removing and refitting the source enclosure .................................................................... 98
Removing the source enclosure from the instrument....................................................... 98
Fitting the source enclosure to the instrument................................................................ 100
Installing and removing the corona pin ............................................ .... ..... ...................... 101
Installing the corona pin in the source............................................................................ 101
Removing the corona pin from the source...................................................................... 103
Operating the source isolation valve ................................................................................ 104
Removing O-rings and seals .............................................................................................. 107
Cleaning the instrument case ............................................................................................ 108
Emptying the exhaust trap bottle ..................................................................................... 108
Gas ballasting the roughing pump .................................................................................... 110
Checking the roughing pump oil level ............................................................................... 112
Adding oil to the roughing pump ....................................................................................... 113
Cleaning the source components ......................................... ..... .... ..................................... 114
Cleaning the sampling cone assembly ............................................................................... 115
Removing the sampling cone assembly from the source................................................. 115
Disassembling the sampling cone assembly.................................................................... 117
Cleaning the sample cone and cone gas nozzle.............................................................. 120
Assembling the sampling cone assembly....................................................................... 122
Fitting the sampling cone assembly to the source.......................................................... 123
Cleaning the ion block assembly ....................................................................................... 125
Removing the ion block assembly from the source assembly........................................ 125
xviiiJanuary 11, 2016, 715004599 Rev. B
Page 19
Disassembling the source ion block assembly................................................................ 127
Cleaning the ion block components................................................................................ 132
Assembling the source ion block assembly.................................................................... 133
Fitting the ion block assembly to the source assembly................................................... 135
Cleaning the ion guide assembly ....................................................................................... 135
Removing the pumping block assembly and ion guide assembly from the instrument . 136
Removing the ion guide assembly and differential aperture from the pumping block
The Waters® Xevo® TQ-S micro is a triple quadrupole, atmospheric pressure ionization
(API) mass spectrometer. Designed for routine HPLC/MS/MS and UPLC
analyses in quantitative and qualitative applications, it can operate at fast acquisition
speeds compatible with UltraPerformance LC
You can use the TQ-S micro with the following high-performance Waters ZSpray™
sources:
•Standard multimode electrospray ionization/atmospheric pressure chemical
ionization/combined electrospray ionization and atmospheric pressure chemical
ionization (ESI/APCI/ESCi
Requirement: Dedicated APCI operation requires an additional probe
(IonSABRE™ II).
•Optional dual-mode APPI/APCI source
•Optional NanoFlow™ ESI source
•Optional ionKey
®
source
®
) source.
®
applications.
®
/MS/MS
•Optional APGC source
•Optional ASAP probe
For information on installing and removing the optional APGC, and ASAP probe, refer to
the operator’s guide supplements supplied with them.
You can also use the Xevo TQ-S micro with the following optional third-party sources:
•DART
®
•DESI
•LDTD™
For further details, refer to the appropriate manufacturer’s documentatio n.
Note: Available source options can vary depending on the software used to operate the
Xevo TQ-S micro. Refer to the MassLynx
®
online Help for more information about
supported sources.
For mass spectrometer specifications, see the Waters Xevo TQ-S micro Site Preparation
Guide.
24Janua ry 11 , 20 16 , 71 50 0 459 9 R ev . B
Page 25
Xevo TQ-S micro shown with visor down, and visor up:
TP03407
Visor up
Uses and compatibility
IntelliStart technology
IntelliStart™ technology monitors LC/MS/MS performance and reports when the
instrument is ready for use.
The software automatically tunes and mass calibrates the instrument, displays
performance readbacks, and enables simplified setup of the system for use in routine
analytical and open access applications. (See page 29.)
January 11, 2016, 715004599 Rev. B25
Page 26
1 Specifications and Operating Modes
The IntelliStart Fluidics1 system is built into the instrument. It delivers sample directly to
the MS probe from the LC column or from two integral reservoirs. The integral reservoirs
can also deliver sample through direct or combined infusion so that you can optimize
instrument performance at analytical flow rates. See the instrument’s online Help for
further details of IntelliStart.
ACQUITY Xevo TQ-S micro UPLC/MS systems
The Waters Xevo TQ-S micro is compatible with the following ACQUITY UPLC®
systems:
•ACQUITY UPLC
•ACQUITY UPLC H-Class
•ACQUITY UPLC I-Class
•ACQUITY UPLC M-Class (with NanoFlow or ionKey source)
If you are not using one of these systems, refer to the documentation relevant to your LC
system.
The ACQUITY
ACQUITY UPLC H-Class, or ACQUITY UPLC I-Class and the Waters Xevo TQ-S
micro fitted with the ESI/APCI/ESCi source.
The ACQUITY
M-Class system and the W aters Xevo TQ-S micro fitted with either a Nanoflow source, or
an ionKey source.
®
Xevo TQ-S micro UPLC/MS system includes an ACQUITY UPLC,
®
Xevo TQ-S micro UPLC/MS system can include an ACQUITY UPLC
If you are not using your instrument as part of an ACQUITY UPLC system, refer to the
documentation for your LC system.
Note: ACQUITY system options can vary depending on the software used to operate the
Xevo TQ-S micro.
1. In Waters product documentation, the term “fluidics” denotes plumbing connections
and components and the fluid pathways within and among instruments or devices. It also
appears in the product name “IntelliStart™ Fluidics” where it describes a mass
spectrometer’s integral apparatus for delivering sample and solvent directly to the
instrument’s probe. Finally, the term can arise in the context of a component part name, as
in “fluidics drawer”.
26Janua ry 11 , 20 16 , 71 50 0 459 9 R ev . B
Page 27
Uses and compatibility
ACQUITY system core components
Core system components for each ACQUITY system are listed below:
SystemCore components
ACQUITY UPLC•Binary solvent manager
•Sample manager
•Column heater
•UPLC detectors
•Solvent tray
•ACQUITY UPLC column
•Software to control the system
ACQUITY UPLC H-Class•Quaternary solvent manager
•Sample manager with flow-through needle
•Column heater-active
•UPLC detectors
•Solvent tray
•ACQUITY UPLC column
•Software to control the system
ACQUITY UPLC I-Class•Binary solvent manager
•Sample manager
•Column heater-active
•UPLC detectors
•Solvent tray
•ACQUITY UPLC column
•Software to control the system
January 11, 2016, 715004599 Rev. B27
Page 28
1 Specifications and Operating Modes
SystemCore components
ACQUITY UPLC M-Class•µBinary solvent manager
For further instruction, see the ACQUITY UPLC System Operator’s Guide, ACQUITY
UPLC H-Class System Guide, ACQUITY UPLC I-Class System Guide, and Controlling
Contamination in UPLC/MS and HPLC/MS Systems (part number 715001307). You can
find the documents on http://www.waters.com; click Services and Support > Support
Library.
•µSample manager - FL
•Trap Valve Manager (including active column
heater)
•TuV and PDA UPLC detectors
•Solvent tray
•ACQUITY UPLC columns of internal diameters
ranging from 75 µm to 1 mm. The column
hardware and the matched outlet tubing can
withstand pressure of as much as 15,000 psi.
•Software to control the system
28Janua ry 11 , 20 16 , 71 50 0 459 9 R ev . B
Page 29
Xevo TQ-S micro with ACQUITY UPLC system:
Sample organizer (optional)
Solvent tray
Column heater
Xevo TQ-S micro
Sample manager
Binary solvent
manager
Uses and compatibility
Software and data system
MassLynx v4.1 software can control the mass spectrometer. See page 30 for more
information about those applications.
MassLynx software enables these major operations:
•Configuring the system
•Creating LC and MS/MS methods that define operating parameters for a run
•Using IntelliStart software to automatically tune and mass calibrate the mass
•Running samples
•Monitoring the run
•Acquiring data
•Processing data
•Reviewing data
•Printing data
spectrometer
January 11, 2016, 715004599 Rev. B29
Page 30
1 Specifications and Operating Modes
MassLynx v4.1
MassLynx software acquires, analyzes, manages, and distributes mass spectrometry,
ultraviolet (UV), evaporative light scattering, and analog data. OpenLynx
TM
TargetLynx
See the MassLynx v4.1 user documentation and online Help for information about using
MassLynx software.
You configure settings, monitor performance, run diagnostic tests, and maintain the
system and its modules via the MassLynx Instrument Console application.
The Instrument Console software, which functions independently of MassLynx software,
does not recognize or control data systems.
See the online Help for the Instrument Console system for details.
application managers are included as standard software with MassLynx.
TM
and
30Janua ry 11 , 20 16 , 71 50 0 459 9 R ev . B
Page 31
Ionization techniques and source probes
Ionization techniques and source probes
Note: Available source options can vary depending on the software used to operate the
Xevo TQ-S micro. Refer to the MassLynx online Help for more information about
supported sources.
Electrospray ionization (ESI)
In electrospray ionization (ESI), a strong electrical charge is applied to the eluent as it
emerges from a nebulizer. The droplets that compose the resultant aerosol undergo a
reduction in size (solvent evaporation). As solvent continues to evaporate, the charge
density increases until the droplet surfaces eject ions (ion evaporation). The ions can be
singly or multiply charged.
The standard ESI probe accommodates eluent flow rates as high as 2 mL/min, making it
suitable for LC applications in the range 100 µL/min to 2 mL/min.
See page 56 for further details.
Combined ESI and APCI (ESCi)
Combined electrospray ionization and atmospheric pressure chemical ionization (ESCi) is
supplied as standard equipment on the mass spectrometer. In ESCi, the standard ESI probe
is used in conjunction with a corona pin to allow alternating acquisition of ESI and APCI
ionization data, facilitating high throughput and wider compound coverage.
See page 60 for further details.
Atmospheric pressure chemical ionization (APCI)
A dedicated high-performance APCI interface is offered as an option. APCI produces
singly charged protonated or deprotonated molecules for a broad range of nonvolatile
analytes.
The APCI interface consists of the ESI/APCI/ESCi enclosure fitted with a corona pin and
an IonSABRE II probe.
See page 60 for further details.
Dual-mode APPI/APCI source
The optional, combined atmospheric pressure photoionization/atmospheric pressure
chemical ionization (APPI/APCI) source comprises an IonSABRE II probe and the APPI
January 11, 2016, 715004599 Rev. B31
Page 32
1 Specifications and Operating Modes
lamp drive assembly. The APPI lamp drive assembly comprises a UV lamp and a repeller
electrode. In addition, a specially shaped, dual, APPI/APCI corona pin can be used. You
can operate the source in APPI, APCI, or dual mode, which switches rapidly between
APPI and APCI ionization modes.
NanoFlow source
“NanoFlow” is the name given to several techniques that use low flow rate electrospray
ionization. The NanoFlow source allows electrospray ionization in the flow rate range 5 to
1000 nL/min. For a given sample concentration, the ion currents observed approximate
those seen in normal flow rate electrospray. However, for similar experiments,
NanoFlow’s significant reduction in sample consumption accompanies significant
increases in sensitivity.
The following options are available for the spraying capillary:
•Universal nebulizer sprayer (Nano-LC).
This option is for flow injection or for coupling to nano-UPLC. It uses a pump to
regulate the flow rate downward to 100 nL/min. If a syringe pump is used, a
gas-tight syringe is necessary to effect correct flow rates without leakage. A volume
of 250 µL is recommended.
•Borosilicate glass capillaries (nanovials).
Metal-coated, glass capillaries allow the lowest flow rates. Usable for one sample,
they must then be discarded.
•Capillary Electrophoresis (CE) or Capillary Electrochromatography (CEC) sprayer .
This option uses a make-up liquid at the capillary tip that provides a stable
electrospray. The ma ke-up flow rate is less than 1 µL/min.
32Janua ry 11 , 20 16 , 71 50 0 459 9 R ev . B
Page 33
Ionization techniques and source probes
Atmospheric solids analysis probe (ASAP)
The ASAP facilitates rapid analysis of volatile and semivolatile compounds in solids,
liquids, and polymers. It is particularly suited to analyzing low-polarity compounds.The
ASAP directly replaces the electrospray or IonSABRE II probe in the instrument’s source
housing and has no external gas or electrical connections.
See the Atmospheric Solids Analysis Probe Operator’s Guide Supplement for further
details.
Atmospheric pressure gas chromatography (APGC)
The Waters APGC couples an Agilent GC with the Xevo TQ-S micro, making it possible
to perform LC and GC analyses in the same system, without compromising performance.
The APGC provides complementary information to the LCMS instrument, enabling
analysis of compounds of low molecular weight and/or low-to-intermediate polarity.
See the Atmosheric Pressure GC Operator’s Guide Supplement for further details.
ionKey source
The ionKey source integrates UPLC separation into the source of the mass spectrometer.
The source accepts an iKey™ separation device, which contains the fluidic connections,
electronics, ESI interface, heater, e-cord, and chemistry . Inserting the iKey simultaneously
engages the electronic and fluidic connections. This technology eliminates the need to
manually connect electronic cables and tubing, and simplifies the user experience.
See the ACQUITY UPLC M-Class System Guide (part number 715003588) and the
ionKey/MS System Guide (part number 715004028) for further details.
January 11, 2016, 715004599 Rev. B33
Page 34
1 Specifications and Operating Modes
P
W
R
B
A
S
LC
LC
Column
Xevo
TQ-S micro
probe
Pump
Reservoirs
B
Wash
Waste
Selector valve
A
IntelliStart Fluidics system
Functionality
The IntelliStart Fluidics system is a solvent delivery system built into the mass
spectrometer. It delivers sample or solvent directly to the MS probe in one of three ways:
•From the LC column.
•From two integral reservoirs. Use standard reservoir bottles (15-mL) for instrument
setup and calibration. Use 1.5-mL, low-volume vials (sold separately) to infuse
smaller volumes (see page 51).
The reservoirs can also deliver sample through direct or combined infusion to
enable optimization at analytical flow rates.
•From a wash reservoir that contains solvent for automated flushing of the
instrument’s solvent delivery system.
The onboard system incorporates a selector valve, an infusion pump, and two sample
reservoirs mounted on the bottom, right-hand side of the instrument.
Recommendation: Use reservoir A for the calibrant solution and tuning compounds,
and reservoir B for analyte/optimization solution.
IntelliStart Fluidics system:
34Janua ry 11 , 20 16 , 71 50 0 459 9 R ev . B
Page 35
System operation
The software automatically controls solvent and sample delivery during auto-tuning,
auto-calibration, and method development. The selector valve systematically makes
connections between the fluidics components to carry out the operations processed by the
software.
You can set IntelliStart fluidics configuration requirements in the system console. You can
edit the parameters, frequency, and extent of the automation. See the mass spectrometer’s
online Help for further details on IntelliStart software and operation of the solvent delivery
system.
For information on plumbing the IntelliStart Fluidics system, see Appendix D.
IntelliStart Fluidics system
January 11, 2016, 715004599 Rev. B35
Page 36
1 Specifications and Operating Modes
T-W ave collision
cell
mini stepwaveDetector
Conversion dynodeSample inletIon guide 2
QuadrupoleMS2 qua drupole
Ion optics
The mass spectrometer’s ion optics operate as follows:
1.Samples from the LC or Intellistart fluidics system are introduced at atmospheric
pressure into the ionization source.
2.The ions pass through the sample cone into the vacuum system.
3.The ions pass through the transfer optics (the ion guide) to the segmented
quadrupole, where the are filtered according to their mass-to-charge ratios.
4.The mass-separated ions pass into the T-Wave™ collision cell where they either
undergo collision-induced dissociation (CID).
5.The transmitted ions are detected by the photomultiplier detection system.
6.The signal is amplified, digitized, and sent to the mass spectrometry software.
Ion optics overview:
MS operating modes
The following table shows the MS operating modes.
36Janua ry 11 , 20 16 , 71 50 0 459 9 R ev . B
Page 37
MS operating modes
MS operating modes:
Operating modeMS1Collision cellMS2
MSPass all massesResolving
(scanning)
SIRPass all massesResolving
(static)
In MS mode, the instrument can acquire data at scan speeds as high as 20,000 Da/s. Use
this mode for instrument tuning and calibration before MS/MS analysis. See the mass
spectrometer’s online Help for further information.
Use the selected ion recording (SIR) mode for quantitation when you cannot find a
suitable fragment ion to perform a more specific multiple reaction monitoring (MRM)
analysis (see page 38). In SIR and MRM modes, the quadrupole is not scanned, therefore
no spectrum (intensity versus mass) is produced. The data obtained from SIR or MRM
analyses derive from the chromatogram plot (specified mass intensity (SIR), or specified
transition (MRM) versus time).
January 11, 2016, 715004599 Rev. B37
Page 38
1 Specifications and Operating Modes
MS/MS operating modes
The following table shows the MS/MS operating modes, described in more detail in the
following pages.
MS/MS operating modes:
Operating modeMS1Collision cellMS2
Product (daughter)
ion spectrum
Precursor (parent)
ion spectrum
MRMStatic (at precursor
PICS (Product Ion
Confirmation Scan)
RADAR™RADAR mode allows you to create experiments containing both
Constant neutral
loss spectrum
Static (at precursor
mass)
ScanningStatic (at product
mass)
Static (at precursor
mass)
MS and MS/MS functions.
Scanning
(synchronized with
MS2)
Fragment precursor
ions and pass all
masses
Scanning
mass)
Static (at product
mass)
Switching between
static (at product
mass) and scanning
Scanning
(synchronized with
MS1)
38Janua ry 11 , 20 16 , 71 50 0 459 9 R ev . B
Page 39
Product (daughter) ion mode
MS1
Static (at precursor mass)
MS2
Scanning
Collision cell
Fragmenting
precursor ions and
passing all masses
MS1
Scanning
MS2
Static (at product mass)
Collision cell
Fragmenting
precursor ions and
passing all masses
Product ion mode is the most commonly used MS/MS operating mode. You can specify an
ion of interest for fragmentation in the collision cell, thus yielding structural information.
Product ion mode:
Typical applications
Product ion mode is typically used for the following applications:
MS/MS operating modes
•Method development for MRM screening studies:
–Identifying product ions for use in MRM transitions.
–Optimizing CID tuning conditions to maximize the yield of a specific product
ion to be used in MRM analysis.
•Structural elucidation (for example, peptide sequencing)
Precursor (parent) ion mode
Precursor ion mode:
January 11, 2016, 715004599 Rev. B39
Page 40
1 Specifications and Operating Modes
MS1
Static (at precursor mass)
MS2
Static (at product mass)
Collision cell
Fragmenting
precursor ions and
passing all masses
Typical application
You typically use the precursor ion mode for structural elucidation—that is, to
complement or confirm product scan data—by scanning for all the precursors of a
common product ion.
Multiple reaction monitoring mode
Multiple reaction monitoring (MRM) mode is the highly selective MS/MS equivalent of
SIR. Because both MS1 and MS2 are static, greater dwell time on the ions of interest is
possible, so the sensitivity achieved is better, compared with scanning-mode MS/MS. This
mode is the most commonly used acquisition mode for quantitative analysis, allowing the
compound of interest to be isolated from the chemical background noise.
Multiple reaction monitoring mode:
PICS mode
A variation on MRM, PICS allows you to collect a product ion spectrum from the top of
all detected peaks in MRM mode for additional confidence in your peak assignment,
activated by a single check box.
RADAR
In RADAR mode the Xevo TQ-S micro rapidly alternates between MRM and full scan
MS acquisition modes. The instrument tracks target analytes with precision in MRM
mode, while at the same time scanning (in MS mode) the background for all other
components. This enables fast characterization of potential matrix effects, providing a
platform for more robust method development.
40Janua ry 11 , 20 16 , 71 50 0 459 9 R ev . B
Page 41
MS/MS operating modes
Typical application
You typically use RADAR mode during method development prior to performing MRM
or PICS to quantify known analytes in complex samples:
•Drug metabolite and pharmacokinetic studies
•Environmental, for example, pesticide and herbicide analysis
•Forensic or toxicology, for example, screening for target drugs in sports
MRM analysis with no associated RADAR or PICS operation does not produce a
spectrum because only one transition is monitored at a time. As in SIR mode, a
chromatogram is produced.
January 11, 2016, 715004599 Rev. B41
Page 42
1 Specifications and Operating Modes
MS1
Scanning
(synchronized with MS2)
Collision cell
Fragmenting
precursor ions and
passing all masses
MS2
Scanning
(synchronized with MS1)
Constant neutral loss mode
Constant neutral loss mode detects the loss of a specific neutral fragment or functional
group from an unspecified precursor(s).
The scans of MS1 and MS2 are synchronized. When MS1 transmits a specific precursor
ion, MS2 “looks” to see whether that precursor loses a fragment of a certain mass. If it
does, the loss registers at the detector.
In constant neutral loss mode, the spectrum shows the masses of all precursors that
actually lost a fragment of a certain mass.
Constant neutral loss mode:
Typical application
You typically use constant neutral loss mode to screen mixtures for a specific class of
compound that is characterized by a common fragmentation pathway, indicating the
presence of compounds containing a common functional group.
42Janua ry 11 , 20 16 , 71 50 0 459 9 R ev . B
Page 43
Sample inlet
Either of two methods delivers solvent and sample to the installed probe:
•An LC system, which delivers the eluent from an LC analysis.
•IntelliStart Fluidics system, which uses onboard solutions to automate instrument
optimization. You can deliver solutions by direct or combined infusion.
Leak sensors
Leak sensors in the Xevo TQ-S micro and the drip trays of the ACQUITY UPLC system
continuously monitor system components for leaks. A leak sensor stops system flow when
its optical sensor detects about 1.5 mL of accumulated, leaked liquid in its surrounding
reservoir. At the same time, the ACQUITY UPLC Console displays an error message
alerting you that a leak has developed.
See Waters ACQUITY UPLC Leak Sensor Maintenance Instructions for complete details.
Vacuum system
Sample inlet
An external roughing pump and an internal split-flow turbomolecular pump combine to
create the source vacuum. The turbomolecular pump evacuates the analyzer an d ion
transfer region.
Vacuum leaks and electrical or vacuum pump failures cause vacuum loss, the damage
from which is prevented by protective interlocks. The system monitors turbomolecular
pump speed and continuously measures vacuum pressure with a built-in Pirani gauge. The
gauge also serves as a switch, stopping operation when it senses vacuum loss.
A vacuum isolation valve isolates the source from the analyzer region, allowing routine
source maintenance without venting.
January 11, 2016, 715004599 Rev. B43
Page 44
1 Specifications and Operating Modes
Power
Turbo vacuum
Source vent
Nitrogen inlet
Collision cell
gas inlet
(Argon)
Source vacuum
Roughing pump
control
Shielded
Ethernet
Event inputs
and outputs
Waste bottle
electrical connection
nano camera
connection
Rear panel connections
The following figure shows the rear panel locations of the connectors used to operate the
instrument with external devices.
Instrument rear panel:
44Janua ry 11 , 20 16 , 71 50 0 459 9 R ev . B
Page 45
2Prep aring for Operation
This chapter describes how to start and shut down the instrument.
Contents:
TopicPage
Starting the mass spectrometer..................................................................... 46
Preparing the IntelliStart Fluidics system..................................................... 51
Rebooting the instrument.............................................................................. 53
Leaving the mass spectrometer ready for operation..................................... 54
January 11, 2016, 715004599 Rev. B45
Page 46
2 Preparing for Operation
Starting the mass spectrometer
Notice: To avoid causing severe damage to the instrument, use compatible
solvents only. For more details, refer to the following sources:
•See page 257 for solvent information.
•Appendix C of the ACQUITY UPLC System Operator’s Guide for solvent
compatibility with ACQUITY devices.
Starting the mass spectrometer entails powering-on the workstation, logging in,
powering-on the mass spectrometer and all other instruments, and then starting the
MassLynx software.
Requirement: You must power-on and log in to the workstation first to ensure that it
obtains the IP addresses of the system instruments.
See the mass spectrometer’s online help for details on MassLynx and IntelliS tart software.
To start the mass spectrometer:
Warning: To avoid igniting flammable solvents, never let the nitrogen supply
pressure fall below 400 kPa (4.0 bar, 58 psi).
1.On the rear panel, ensure the nitrogen supply is connected to the instrument’s
nitrogen inlet connection (see the figure on page 44).
Requirement: The nitrogen must be dry and oil-free, with a purity of at least 95%
or, for APGC use, at least 99.999%. Regulate the supply at 600 to 690 kPa (6.0 to
6.9 bar, 90 to 100 psi).
For more information on connections, see the figure on page 44.
2.Ensure that the collision gas supply is connected to the instrument’s collison cell
gas inlet.
Requirement: The collision gas is argon; it must be dry and of high purity
(99.997%). Regulate the supply at 50 kPa (0.5 bar, 7 psi).
3.Power-on the workstation, and log in.
4.Press the power switch on the top, right-hand side of the mass spectrometer and the
switches on the top, left-hand sides of the ACQUITY instruments.
Result: Each system instrument runs a series of startup tests.
46Janua ry 11 , 20 16 , 71 50 0 459 9 R ev . B
Page 47
Starting the mass spectrometer
5.Allow 3 minutes for the embedded PC (located inside the mass spectrometer) to
initialize and to sound an alert indicating that the PC is ready.
Tip: The power and status LEDs change as follows:
•Each system instrument’s power LED shows green.
•During initialization, the binary or quaternary solvent manager’s and sample
manager’s status LED flashes green.
•After the instruments successfully power-on, all power LEDs show steady
green. The binary solvent manager’s flow LED, the sample manager’s run
LED, and the mass spectrometer’s Operate LED remain off.
6.Start the MassLynx software, and monito r the Instrument Console software for
messages and LED indications.
January 11, 2016, 715004599 Rev. B47
Page 48
2 Preparing for Operation
7.To evacua te (pump down) the mass spectrometer, follow the procedure below for
MassLynx software.
MassLynx software:
a.Click IntelliStart in the MassLynx main window’s lower left-hand corner.
Result: The mass spectrometer’s console appears. The mass spectrometer is
in Standby mode.
b.Click Control > Pump, to start the roughing pump.
Tip: After a 20-second delay, during which the turbopump is starting, the
roughing pump starts. IntelliStart displays “Instrument in standby”, and the
Operate LED remains off.
c.Wait a minimum of 2 hours for the instrument to be fully pumped down
(evacuated).
Tip: In the Instrument Console, the System Ready indicator shows green when
the instrument is fully pumped down (evacuated).
d.Click Resolve or Operate .
Result: When the mass spectrometer is in good operating condition,
IntelliStart software displays “Ready” in the Instrument Console.
Tip: If clicking Resolve fails to put the instrument into Operate mode,
IntelliStart software displays corrective actions in the Instrument Console.
48Janua ry 11 , 20 16 , 71 50 0 459 9 R ev . B
Page 49
Verifying the instrument’s state of readiness
When the instrument is in good operating condition, the power and Operate LEDs show
constant green. You can view any error messages in IntelliStart software.
Monitoring the instrument LEDs
Light-emitting diodes on the instrument indicate its operational status.
Power LED
The power LED, to the top, right-hand side of the mass spectrometer’s front panel,
indicates when the instrument is powered-on or powered-off.
Operate LED
The Operate LED, on the right-hand side of the power LED, indicates the operating
condition.
See the instrument’s online Help for details of the Operate LED indications.
Starting the mass spectrometer
Tuning and calibration information
You must tune and, if necessary, calibrate the instrument prior to use. You can perform
these tasks using IntelliStart software. For further instruction, see the mass spectrometer’s
online Help.
January 11, 2016, 715004599 Rev. B49
Page 50
2 Preparing for Operation
Running the instrument at different flow rates
The ACQUITY UPLC system runs at high flow rates. To optimize desolvation, and thus
sensitivity , run the ACQUITY Xevo TQ-S micro system at appropriate gas flows and
desolvation temperatures. IntelliStart software automatically sets these parameters when
you enter a flow rate, according to the following table.
Flow rate versus temperature and gas flow:
Flow rate
(mL/min)
0.000 to 0.020150200800
0.021 to 0.100150300800
0.101 to 0.5001505001000
>0.5001506001000
If you are using an APCI interface, IntelliStart software automatically sets the parameters
according to the following table.
Flow rate versus IonSABRE II probe temperature and gas flow:
Flow rate
(mL/min)
0.000 to 0.020400800
0.021 to 0.5005001000
>0.5006001000
Source temp (°C)
IonSABRE II
probe
temperature (°C)
Desolvation
temp (°C)
Desolvation gas
flow (L/h)
Desolvation gas
flow (L/h)
50Janua ry 11 , 20 16 , 71 50 0 459 9 R ev . B
Page 51
Preparing the IntelliStart Fluidics system
Preparing the IntelliStart Fluidics system
For additional information, see page 246 and Appendix D.
Notice: To avoid damaging the mass spectrometer by accidentally spilling solvent
on it, do not store large-volume solvent reservoirs on top of the instrument.
Installing the reservoir bottles
Use standard reservoir bottles (15-mL) for instrument setup and calibration. Use the
Low-volume Adaptor Kit (sold separately) to infuse smaller volumes. The low-volume
vials have a volume of 1.5 mL.
Required material
Chemical-resistant, powder-free gloves
January 11, 2016, 715004599 Rev. B51
Page 52
2 Preparing for Operation
TP03410
Warning: To avoid personal contamination with biohazards or toxic
materials, and to avoid spreading contamination to uncontaminated surfaces,
wear clean, chemical-resistant, powder-free gloves while installing reservoir
bottles.
To install the reservoir bottles:
1.Remove the reservoir bottle caps.
2.Screw the reservoir bottles onto the instrument, as shown below.
3.For each reservoir bottle, ensure that the end of the solvent delivery tube is
positioned so that it is close to, but does not touch, the bottom of the bottle.
To install low-volume vials:
Warning: To avoid personal contamination with biohazards or toxic
materials, and to avoid spreading contamination to uncontaminated surfaces,
wear clean, chemical-resistant, powder-free gloves while installing
low-volume vials.
1.If a standard reservoir bottle is fitted, remove it.
2.Screw the low-volume adaptor into the manifold and tighten it.
3.Screw the low-volume vial into the adaptor.
4.For each low-volume vial, ensure that the end of the solvent delivery tube is
positioned so that it is close to, but does not touch, the bottom of the vial.
52Janua ry 11 , 20 16 , 71 50 0 459 9 R ev . B
Page 53
Purging the infusion pump
TP03415
Reset button aperture
Whenever you replace a solution bottle, purge the infusion pump with the solution that
you are going to use next. See the mass spectrometer’s online Help for details.
Requirement: Ensure that the end of the tubing is fully submerged in the solvent in the
wash reservoir.
Tip: Depending on the solutions used, the instrument’s solvent delivery system can
require more than one purge cycle to minimize carryover.
Rebooting the instrument
The reset button causes the mass spectrometer to reboot.
Reboot the instrument when either of these conditions applies:
•Immediately following a software upgrade.
•The mass spectrometer software fails to initialize.
To reboot the instrument:
Rebooting the instrument
1.Ensure that the mass spectrometer software is closed.
2.Insert a short length (7.5 cm) of PEEK™ tubing, or similar object, into the reset
button aperture at the top, right-hand side of the instrument’s front panel.
3.Remove the PEEK tubing from the reset button aperture.
4.Wait until the reboot sequence ends before starting the mass spectrometer software.
January 11, 2016, 715004599 Rev. B53
Page 54
2 Preparing for Operation
Tip: An audible alert sounds when the reboot sequence is complete.
Leaving the mass spectrometer ready for operation
Leave the mass spectrometer in Operate mode except in the following cases:
•When performing routine maintenance
•When changing the source
•When leaving the mass spectrometer unused for a long period
In these instances, put the mass spectrometer in Standby mode. See the online Help for
details.
Notice: For ionKey/MS operation, to protect the iKey when you leave the mass
spectrometer in Operate mode with no flow, set the capillary voltage to zero.
Emergency shutdown of the mass spectrometer
To shut down the instrument in an emergency:
Warning: To avoid electric shock, isolate the instrument observing the procedure
outlined below. The instrument’s power switch does not isolate it from the main
power supply.
Notice: To avoid losing data, reboot the instrument as described on page 53. Data
can be lost when you perform an emergency shutdown.
1.Operate the power button on the front of the instrument.
2.Disconnect the power cable from the instrument’s rear panel.
54Janua ry 11 , 20 16 , 71 50 0 459 9 R ev . B
Page 55
3Changing the Mode of Operation
This chapter describes how to prepare the mass spectrometer for the following
modes of operation:
•ESI (electrospray ionization)
•ESCi (combined electrospray and atmospheric pressure chemical ionization)
The following sections explain how to install and remove an ESI probe. For further details
on running ESI applications, see page 31.
Installing the ESI probe
Required material
Chemical-resistant, powder-free gloves
Warning: To avoid personal contamination with biohazards or toxic
materials, and to avoid spreading contamination to uncontaminated surfaces,
wear clean, chemical-resistant, powder-free gloves while performing this
procedure. The LC system connections, ESI probe, and source can be
contaminated.
Warning: To avoid electric shock, before beginning this procedure, prepare
the instrument according to the procedure on page 97, “Preparing the
instrument for operations on or inside its source”.
To install the ESI probe:
1.Prepare the instrument for installing the probe according to the procedure on
page 97, “Preparing the instrument for operations on or inside its source”.
Warning: To avoid puncture wounds, handle the ESI probe with care; the
probe tip is sharp.
2.Remove the protective sleeve, if fitted, from the ESI probe tip.
56Janua ry 11 , 20 16 , 71 50 0 459 9 R ev . B
Page 57
TP03129
Location hole of the probe
adjuster assembly
Probe location dowel
Probe label
Probe locking ring
ESI mode
3.With the probe label facing you, carefully slide the ESI probe into the hole in the
probe adjuster assembly, ensuring that the probe location dowel aligns with the
location hole of the probe adjuster assembly.
4.Tighten the probe locking ring to secure the probe in place.
Requirement: Fully tighten the probe locking ring. Doing so ensures a successful
result on the automatic pressure test, which runs when the probe is correctly seated,
and the source enclosure door is closed.
5.Connect the ESI probe’s cable to the high-voltage connector.
January 11, 2016, 715004599 Rev. B57
Page 58
3 Changing the Mode of Operation
Warning: To avoid electric shock, do not use stainless steel tubing or
stainless steel “finger tight” screws to connect the selector valve to the ESI
probe; use the PEEK tubing and natural (beige) colored PEEK “finger tight”
screws supplied with the instrument.
6.Using PEEK tubing equal to 0.004-inch ID, connect port S of the selector valve to
the ESI probe.
Recommendation: To reduce peak broadening, use 0.004-inch ID tubing for
sample flow rates ≤1.2 mL/min; use 0.005-inch ID tubing for sample flow rates
>1.2 mL/min.
Requirements:
•If you are replacing the tubing between the selector valve and the probe,
minimize the length to reduce peak broadening.
•When cutting the tubing to length, cut it squarely (that is, perpendicular to its
horizontal axis).
58Janua ry 11 , 20 16 , 71 50 0 459 9 R ev . B
Page 59
Removing the ESI probe
Required material
Chemical-resistant, powder-free gloves
Warning: To avoid personal contamination with biohazards or toxic
materials, and to avoid spreading contamination to uncontaminated surfaces,
wear clean, chemical-resistant, powder-free gloves while performing this
procedure. The LC system connections, ESI probe, and source can be
contaminated.
Warning: To avoid electric shock, before beginning this procedure, prepare
the instrument according to the procedure on page 97, “Preparing the
instrument for operations on or inside its source”.
To remove the ESI probe:
1.Prepare the instrument for removing the probe according to the procedure on
page 97, “Preparing the instrument for operations on or inside its source”.
ESI mode
2.Disconnect the tubing from the ESI probe.
3.Disconnect the ESI probe cable from the high-voltage connector.
4.Unscrew the probe locking ring.
Warning: To avoid puncture wounds, handle the ESI probe with care; the
probe tip is sharp.
5.Carefully remove the ESI probe from the probe adjuster assembly.
6.If available, fit the protective sleeve to the ESI probe tip.
January 11, 2016, 715004599 Rev. B59
Page 60
3 Changing the Mode of Operation
Sample cone
Corona pin
ESI probe
ESCi mode
T o run ESCi applications, you must fit an ESI probe and corona pin to the ESI/APCI/ESCi
source enclosure.
See “Installing the ESI probe” on page 56, “Installing the corona pin in the source” on
page 101, and “IntelliStart Fluidics System” on page 34.
ESCi mode:
Optimizing the ESI probe for ESCi operation
See the mass spectrometer’s online Help for details on how to optimize the ESI probe for
ESCi operation.
APCI mode
APCI mode, an option for the mass spectrometer, produces singly charged protonated or
deprotonated molecules for a broad range of nonvolatile analytes.
The APCI interface consists of the ESI/APCI/ESCi enclosure fitted with a corona pin and
an IonSABRE II probe. Mobile phase from the LC column enters the probe, where it is
pneumatically converted to an aerosol, rapidly heated, and vaporized or gasified at the
probe tip.
60Janua ry 11 , 20 16 , 71 50 0 459 9 R ev . B
Page 61
APCI mode
IonSABRE II probe
Sample cone
Corona pin
APCI mode:
Hot gas from the IonSABRE II probe passes between the sample cone and the corona pin,
which is typically operated with a discharge current of 5 µA. Mobile phase molecules
rapidly react with ions generated by the corona discharge to produce stable reagent ions.
Analyte molecules introduced into the mobile phase react with the reagent ions at
atmospheric pressure and typically become protonated (in the positive ion mode) or
deprotonated (in the negative ion mode). The sample and reagent ions then pa ss through
the sample cone and into the mass spectrometer.
Installing the IonSABRE II probe
Required materials
•Chemical-resistant, powder-free gloves
•Sharp knife or PEEK tubing cutter
Warning: To avoid personal contamination with biohazards or toxic
materials, and to avoid spreading contamination to uncontaminated surfaces,
wear clean, chemical-resistant, powder-free gloves while performing this
procedure. The LC system connections, IonSABRE II probe, and source can
be contaminated.
Warning: To avoid electric shock, before beginning this procedure, prepare
the instrument according to the procedure on page 97, “Preparing the
instrument for operations on or inside its source”.
January 11, 2016, 715004599 Rev. B61
Page 62
3 Changing the Mode of Operation
TP03129
Location hole of the probe
adjuster assembly
Probe location dowel
Probe label
Probe locking ring
To install the IonSABRE II probe:
1.Prepare the instrument for working on the source (see page 97).
2.With the probe label facing toward you, carefully slide the IonSABRE II probe into
the hole in the probe adjuster assembly, ensuring that the probe location dowel
aligns with the probe adjuster assembly location hole.
3.Tighten the probe locking ring to secure the probe in place.
Requirement: Fully tighten the probe locking ring. Doing so ensures a successful
result on the automatic pressure test, which runs when the probe is correctly seated,
and the source enclosure door is closed.
Warning: To avoid electric shock, do not use stainless steel tubing or
stainless steel “finger tight” screws to connect the selector valve to the
IonSABRE II probe; use the PEEK tubing and natural (beige) colored PEEK
“finger tight” screws supplied with the instrument.
4.Using tubing equal to 0.004-inch ID, connect port S of the selector valve to the
IonSABRE II probe.
Recommendation: To reduce peak broadening, use 0.004-inch ID tubing for
sample flow rates ≤1.2 mL/min; use 0.005-inch ID tubing for sample flow rates
>1.2 mL/min.
Requirements:
62Janua ry 11 , 20 16 , 71 50 0 459 9 R ev . B
Page 63
•If you are replacing the tubing between the selector valve and the probe,
minimize the length to reduce peak broadening.
•When cutting the tubing to length, cut it squarely (that is, perpendicular to its
horizontal axis).
5.Install the corona pin (see page 101).
APCI mode
January 11, 2016, 715004599 Rev. B63
Page 64
3 Changing the Mode of Operation
Removing the IonSABRE II probe
Required material
Chemical-resistant, powder-free gloves
Warning: To avoid personal contamination with biohazards or toxic
materials, and to avoid spreading contamination to uncontaminated surfaces,
wear clean, chemical-resistant, powder-free gloves while performing this
procedure. The LC system connections, IonSABRE II probe, and source can
be contaminated.
Warning: To avoid electric shock, before beginning this procedure, prepare
the instrument according to the procedure on page 97, “Preparing the
instrument for operations on or inside its source”.
To remove the IonSABRE II probe:
1.Prepare the instrument for working on the source (see page 97).
2.Remove the corona pin (see page 101).
3.Disconnect the selector valve tubing from the IonSABRE II probe.
4.Unscrew the probe locking ring.
5.Carefully remove the probe from the probe adjuster assembly.
64Janua ry 11 , 20 16 , 71 50 0 459 9 R ev . B
Page 65
Combined APPI/APCI source
IonSABRE II probe
Sample molecules
APPI lamp drive assembly
Repeller electrodeUV lamp
Sample ions
Sample cone
Photons from
the UV lamp
Operate this optional, replacement source enclosure in APPI, APCI, or dual APPI/APCI
mode. Dual-mode APPI/APCI performs rapid switching between ionization modes.
APPI operation
In atmospheric pressure photoionization (APPI) mode, the source is fitted with an
IonSABRE II probe and the APPI lamp drive assembly is advanced into the source.
APPI mode:
Combined APPI/APCI source
The IonSABRE II probe introduces vaporized sample into the source where photons
generated by an ultraviolet (UV) lamp (mounted in the APPI lamp drive assembly)
produce sample ions. Direct photoionization of a sample molecule occurs when the photon
energy exceeds the ionization potential of the sample molecule.
A repeller electrode (mounted on the APPI lamp drive assembly) deflects and focuses the
sample ions toward the sample cone.
January 11, 2016, 715004599 Rev. B65
Page 66
3 Changing the Mode of Operation
IonSABRE II probe
Retracted APPI lamp drive
assembly
Sample cone
APCI corona pin
APCI operation
The atmospheric pressure chemical ionization (APCI) mode produces singly charged
protonated or deprotonated molecules for a large range of nonvolatile analytes. In APCI
mode, the source is fitted with an APCI corona pin. Unused, the APPI lamp drive
assembly is retracted from the source.
APCI mode:
The IonSABRE II probe introduces vaporized sample into the source. The sample passes
between the sample cone and the corona pin, which typically operates with a discharge
current of 5 µA. The corona discharge generates ions that react with the mobile phase
molecules to produce stable reagent ions. Analyte molecules in the mobile phase react
with the reagent ions at atmospheric pressure and become protonated (in the positive ion
mode) or deprotonated (in the negative ion mode). The sample and reagent ions pass
through the sample cone.
66Janua ry 11 , 20 16 , 71 50 0 459 9 R ev . B
Page 67
Dual-mode operation
Sample cone
IonSABRE II probe
Photons from the UV lamp
Repeller electrode inactive
Corona pin with current applied
Sample cone
IonSABRE II probe
Photons from the UV lamp
Repeller electrode with
voltage applied
Corona pin inactive
Dual-mode operation enables rapid switching between APPI and APCI ionization modes
and allows high-throughput operations (for example, for sample screening).
You replace the standard corona pin with a specially shaped APPI/APCI corona pin, so
that the APPI lamp holder can be advanced into the source for dual operation.
When the source is configured for dual operation in APCI mode, current is applied to the
corona pin, but the repeller electrode is inactive.
Dual operation in APCI mode:
Combined APPI/APCI source
When the source is configured for dual operation in APPI mode, the corona pin is inactive,
and a voltage is applied to the repeller electrode.
Dual operation in APPI mode:
January 11, 2016, 715004599 Rev. B67
Page 68
3 Changing the Mode of Operation
APPI lamp drive assembly
The combined APPI/APCI source components
The combined APPI/APCI source comprises the standard IonSABRE II probe and a
source enclosure with an APPI lamp drive incorporated.
The combined APPI/APCI source enclosure:
Notice: To prevent damage to the corona pin and lamp assembly, ensure that the
lamp assembly does not touch the corona pin when the source enclosure door is
closed.
68Janua ry 11 , 20 16 , 71 50 0 459 9 R ev . B
Page 69
Combined APPI/APCI source
TP03201
APPI lamp drive
assembly
Source enclosure
UV lamp and
repeller electrode
IonSABRE II probe
The UV lamp, which you ignite via a control in the MassLynx Tune window, provides a
constant photon output. You vary the intensity of incident radiation upon the sample
molecules by adjusting the distance between the UV lamp and probe tip.
APPI lamp drive assembly inside the source enclosure:
January 11, 2016, 715004599 Rev. B69
Page 70
3 Changing the Mode of Operation
Installing the combined APPI/APCI source
Required material
Chemical-resistant, powder-free gloves
Warning: To avoid personal contamination with biohazards or toxic
materials, and to avoid spreading contamination to uncontaminated surfaces,
wear clean, chemical-resistant, powder-free gloves while performing this
procedure. The source components can be contaminated.
Warning: To avoid electric shock, before beginning this procedure, prepare
the instrument according to the procedure on page 97, “Preparing the
instrument for operations on or inside its source”.
To install the combined APPI/APCI source:
1.Prepare the instrument for working on the source (see page 97).
Warning: To avoid burn injuries, take great care while working with the
probe and source; these components can be hot.
2.Remove the probe from the currently installed source.
•If you are removing an ESI probe, see page 59.
•If you are removing an IonSABRE II probe, see page 64.
3.Remove the existing source enclosure (see page 98).
4.Install the combined APPI/APCI source enclosure (see page 100).
5.Install the corona pin (see page 101).
6.Connect the APPI drive cable to the instrument’s front panel connector.
70Janua ry 11 , 20 16 , 71 50 0 459 9 R ev . B
Page 71
Combined APPI/APCI source
7.Connect the HT cable to the instrument’s front panel connector.
Notice: To prevent damage to the corona pin and lamp assembly, ensure
that the lamp assembly does not touch the corona pin when the source
enclosure door is closed.
8.Install the IonSABRE II probe to the source, and ensure that it is working correctly
(see page 61).
Tip: An automatic pressure test runs each time you close the source enclosure and
when the instrument starts.
Removing the IonSABRE II probe and APPI/APCI source
enclosure
Required material
Chemical-resistant, powder-free gloves
Warning: To avoid personal contamination with biohazards or toxic
materials, and to avoid spreading contamination to uncontaminated surfaces,
wear clean, chemical-resistant, powder-free gloves while performing this
procedure. The source components can be contaminated.
Warning: To avoid electric shock, before beginning this procedure, prepare
the instrument according to the procedure on page 97, “Preparing the
instrument for operations on or inside its source”.
To remove the combined APPI/APCI source
1.Prepare the instrument for working on the source (see page 97).
Warning: To avoid burn injuries, take great care while working with the
probe and source; these components can be hot.
2.Remove the IonSABRE II probe (see page 64).
3.Disconnect the HT cable from the instrument’s front panel.
4.Disconnect the APPI drive cable from the instrument’s front panel.
5.Remove the source enclosure (see page 98).
6.Remove the corona pin (see page 103).
January 11, 2016, 715004599 Rev. B71
:
Page 72
3 Changing the Mode of Operation
X, Y, Z stage
Microscope Camera
Sprayer enclosure
7.Fit the blanking plug to the pin’s mounting contact.
NanoFlow source
The NanoFlow source enclosure comprises the NanoFlow stage (for x-axis, y-axis, and
z-axis adjustment), the sprayer-enclosure, and a microscope camera.
NanoFlow source, stage and microscope camera:
A sprayer is mounted on an X, Y, Z stage (three-axis manipulator) that slides on a pair of
guide rails that allow its withdrawal from the source enclosure for maintenance and
changes.
A light within the source provides illumination for the spray, which you can observe using
the microscope camera mounted on the corner of the source housing.
72Janua ry 11 , 20 16 , 71 50 0 459 9 R ev . B
Page 73
The low flow rates involved with operating the NanoFlow source prohibit its use with the
instrument’s solvent delivery system.
Installing the NanoFlow source
Required material
Chemical-resistant, powder-free gloves
Warning: To avoid personal contamination with biohazards or toxic
materials, and to avoid spreading contamination to uncontaminated surfaces,
wear clean, chemical-resistant, powder-free gloves while performing this
procedure. The source components can be contaminated.
Warning: To avoid electric shock, ensure that the instrument is prepared for
working on the source before commencing this procedure.
Warning: To avoid burn injuries, ensure that the source heater is turned off
and the block is cool before opening the source. The ion block, which can b e
hot, is exposed when you fit the NanoFlow source.
NanoFlow source
To install the NanoFlow source:
1.Prepare the instrument for working on the source (see page 97).
Warning: To avoid burn injuries, take great care while working with the
probe and source; these components can be hot.
2.Remove the probe from the currently installed source.
•If you are removing an ESI probe, see page 59.
•If you are removing an IonSABRE II probe, see page 64.
3.Remove the existing source enclosure (see page 98).
January 11, 2016, 715004599 Rev. B73
Page 74
3 Changing the Mode of Operation
Stop screw
Retaining screw
Notice: To prevent the sprayer from colliding with the cone and breaking,
always retract the stage before installing the source enclosure or closing the
door.
4.On the NanoFlow source, loosen the stage retaining screw, pull the stop screw, and
slide the stage fully out of the enclosure.
5.Using both hands, fit the NanoFlow source enclosure to the two supporting studs on
the source adaptor housing.
6.Close the source enclosure door.
7.Connect a 1/16-inch PTFE tube between the mass-flow controller output (mounted
beneath the stage on the front of the NanoFlow source) and your sprayer.
Tip: For details regarding how to fit each sprayer, see the corresponding reference:
•Waters Universal NanoFlow Sprayer Installation and Maintenance Guide
(part number 71500110107)
•“Fitting a borosilicate glass capillary (nanovial)” on page 76
8.Connect the probe cable to the high-voltage connector.
74Janua ry 11 , 20 16 , 71 50 0 459 9 R ev . B
Page 75
9.Connect the high-voltage cable to the instrument’s HV connection.
High-voltage cable
NanoFlow source
Tip: The NanoFlow stage contains a high-voltage interlock, so the capillary voltag e
(the voltage applied to the sprayer assembly) and the sampling cone voltage remain
disabled until the sprayer is pushed fully forward in the source.
10. Slide closed the instrument’s source interface door.
January 11, 2016, 715004599 Rev. B75
Page 76
3 Changing the Mode of Operation
Fitting a borosilicate glass capillary (nanovial)
Required materials
•Chemical-resistant, powder-free gloves
•Needle-nose pliers
•Borosilicate glass capillary
•Fused silica syringe needle or GELoader
•Fused silica cutter
Warning: To avoid lacerations, puncture injuries, and possible
contamination with biohazardous and toxic samples, do not touch the
sharp end of the capillary.
Notice: To avoid damaging capillaries, take great care when handling
them. Capillaries are extremely fragile; always hold their blunt end,
never the sharp end, which can easily be damaged.
Warning: To avoid electric shock, ensure that the NanoFlow stage is
fully retracted from the source before beginning this procedure.
®
tip
To fit a borosilicate glass capillary (nanovial):
1.Loosen the stage retaining screw.
2.Pull the stop screw to release the stage.
3.Slide the stage out of the NanoFlow source enclosure and remove the magnetic
cover.
4.Unscrew the retaining screw, and lift the sprayer from the stage.
76Janua ry 11 , 20 16 , 71 50 0 459 9 R ev . B
Page 77
5.Unscrew the union from the end of the sprayer assembly.
Capillary
Union
Foam
Capillary
6.Remove the existing capillary from the sprayer.
NanoFlow source
7.Carefully remove the new borosilicate glass capillary from its case by lifting
vertically while pressing on the foam with two fingers.
8.Load sample into the capillary using a fused silica syringe needle or a GELoader
tip, minimizing any bubbles between the capillary tip and the sample.
January 11, 2016, 715004599 Rev. B77
Page 78
3 Changing the Mode of Operation
5mm
PTFE tubing
Ferrule
Union
Knurled nut
Blue conductive elastomer
Glass capillary
Recommendation: When using a GELoader tip, break the glass capillary in half,
scoring it with a fused silica cutter so that the GELoader can reach the capillary’s
tip.
9.Thread the knurled nut and approximately 5 mm of conductive elastomer over the
blunt end of the capillary.
10. Fit the capillary into the holder (probe).
11. Finger-tighten the nut so that 5 mm of glass capillary protrude from its end.
Tip: Measure the protrusion from the end of the nut to the shoulder of the glass
capillary.
Sprayer assembly:
12. Screw the sprayer back into the assembly.
13. Replace the sprayer cover.
14. On the ES+/- Source tab of the MassLynx MS Tune window, ensure that the
Capillary parameter is set to 0 kV.
78Janua ry 11 , 20 16 , 71 50 0 459 9 R ev . B
Page 79
Notice: To prevent the capillary tip from colliding with the cone or the side
d
2d
3d
Cone aperture diameter
of the source, adjust the sprayer tip position before you push the sprayer
inside the NanoFlow source enclosure.
15. Carefully push the stage back into the NanoFlow source enclosure, using the stop
and handle.
Positioning the borosilicate glass capillary tip
Having obtained a signal, you must adjust the tip position to maximize it. Using the
three-axis manipulator, you can adjust the tip position up and down, left and right, forward
and backward. As a starting point, set the tip so that it is on the center line of the sampling
cone and at a distance between two and three times the diameter of the cone aperture .
Typically this distance is approximately 2 mm.
Capillary tip position:
NanoFlow source
For tuning instructions, see the instrument’s online help.
January 11, 2016, 715004599 Rev. B79
Page 80
3 Changing the Mode of Operation
Microscope
camera
Handle for locking
and unlocking the
iKey separation
device
Front cover
Docking port
for the iKey
separation
device
ionKey source
The ionKey source integrates UPLC separation into the source of the mass spectrometer.
For a complete description, see “ionKey source” on page 33.
The following sections explain how to install or remove an ionKey source.
For further information, see the ACQUITY UPLC M-Class System Guide (part number
715003588) and the ionKey/MS System Guide (part number 715004028).
Installing the ionKey source
The ionKey source enclosure comprises the iKey docking port, the locking handle, the
sprayer-enclosure, and a microscope camera.
ionKey source:
80Janua ry 11 , 20 16 , 71 50 0 459 9 R ev . B
Required materials
•Chemical-resistant, powder-free gloves
•Screwdriver
•¼-inch wrench
Page 81
Warning: To avoid personal contamination with biohazards or toxic
materials, and to avoid spreading contamination to uncontaminated surfaces,
wear clean, chemical-resistant, powder-free gloves while performing this
procedure. The source components can be contaminated.
Warning: To avoid electric shock, prepare the instrument for work performed on
its source before beginning this procedure.
To install the ionKey source:
1.Prepare the instrument for working on its source (see page 97).
Warning: To avoid burn injuries, take great care while working with the
instrument’s source enclosure open; the source can be hot.
2.Remove the probe from the currently installed source:
•If you are removing an ESI probe, see page 59.
•If you are removing an IonSABRE II probe, see page 64.
ionKey source
3.Remove the existing source enclosure (see page 98).
4.Using two hands, fit the ionKey source enclosure to the two supporting studs on the
source adaptor housing.
5.Swing the source enclosure to the closed position, ensuring it locks into place.
Notice: To avoid damaging the ionKey source or µSample manager,
•ensure the µSample manager’s power is off before connecting the data/power
cable;
•ensure that the mass spectrometer is in Standby mode before beginning any
installation or maintenance.
January 11, 2016, 715004599 Rev. B81
Page 82
3 Changing the Mode of Operation
Data/power cable
to PSPI
connector on
µSample
manager
High voltage cable
Options cable
Fluid infusion line
Fluid inlet line
Optional post-column addition (PCA) line
Fluid waste line
ionKey source connections:
82Janua ry 11 , 20 16 , 71 50 0 459 9 R ev . B
6.Connect the data/power cable to the PSPI connector on the rear of the µSample
manager, and use a screwdriver to firmly tighten the connector screws.
Page 83
ionKey source
Data/power
cable to PSPI
connector on
µSample
manager
High voltage
cable
Options
cable
Source connections to mass spectrometer:
7.Connect the high voltage cable (white) to the high voltage supply outlet on the mass
spectrometer.
8.Connect the options cable (blue) to the options port on the mass spectrometer.
January 11, 2016, 715004599 Rev. B83
Page 84
3 Changing the Mode of Operation
Fluid line aperture
Aperture open
(spring-loaded)
Aperture closed
9.Identify each fluid line by the part numbers printed on their shrink-wrap tubing.
Fluid inlet line connected
to injection valve port 6
µSample manager injection valve:
10. Connect the fluid inlet line to port 6 on the injection valve of the µSample manager.
11. Connect the fluid infusion line to port 2 on the fluidics divert valve.
12. Connect the optional post-column addition line to outlet B on the flow control
module of the auxiliary solvent manager.
13. Connect the waste line to a suitable waste container.
Installing ionKey source software
If you are installing an ionKey source on your Xevo TQ-S micro for the first time, you
must install the appropriate MassLynx software SCN and the ACQUITY UPLC M-Class
driver pack. For further details, see the following documents:
•ACQUITY UPLC M-Class Driver Pack Installation and Configuration Guide (part
number 715004448) for detailed installation procedures, and information on using
the ACQUITY Inlet Switch Utility.
•MassLynx software v4.1 and related SCN release notes for detailed information
about installing MassLynx software and SCNs.
January 11, 2016, 715004599 Rev. B85
Page 86
3 Changing the Mode of Operation
Installing the camera in the ionKey source
To install the camera software for the ionKey source:
1.Connect the camera cable from the video output connector on the mass
spectrometer’s rear panel to the video-to-USB converter box.
Notice: To avoid damaging the video converter, make sure the workstation is
powered-off before connecting the converter to the workstation in the next step.
2.Connect the video-to-USB converter box to a USB port on the mass spectrometer’s
workstation.
3.On the Tune page, click Camera Viewer .
4.In Device settings dialog box, specify the parameter settings according to the
following table, and then click OK.
Tip: After you install the camera software, when you select the ionKey camera
viewer for the first time, the device settings dialog box opens. T o subsequently open
the device settings dialog box, in the camera viewer, click View > Camera Options.
Device settings for the camera:
ParameterSetting
Video normPAL_B
Video formatY800 (768 x 576)
Frame rate (FPS)25
Input channel00 Video: Composite
Removing the ionKey source
You can remove the ionKey source, and replace it with a conventional interface.
Alternative: If you are using the ionKey source with an ACQUITY UPLC M-Class
system mounted on an M-Class cart fitted with a universal source holder, you can secure
the source enclosure to the holder. Doing so keeps the enclosure close to the Xevo TQ-S
micro, for when it is next needed; assists with managing the ionKey source’s fluid lines;
and helps prevent contamination of the fluid lines.
86Janua ry 11 , 20 16 , 71 50 0 459 9 R ev . B
Page 87
ionKey source
See the ACQUITY UPLC M-Class documentation for further information about installing
the source holder on the M-Class cart, and securing the source enclosure to the holder.
See M-Class Cart Universal Source Holder (part number 715004884) for further
information about installing and using the universal source holder on the M-Class cart,
and securing the source enclosure to the holder.
See also: The ionKey/MS System Guide.
Required materials
•Chemical-resistant, powder-free gloves
•¼-inch wrench
Warning: To avoid personal contamination with biohazards or toxic
materials, and to avoid spreading contamination to uncontaminated surfaces,
wear clean, chemical-resistant, powder-free gloves while performing this
procedure. The source components can be contaminated.
Warning: To avoid electric shock, prepare the instrument for work performed on
its source before beginning this procedure.
To remove the ionKey source:
1.Prepare the instrument for working on its source (see page 97).
Warning: To avoid burn injuries, take great care while working with the
instrument’s source enclosure open; the source can be hot.
2.Remove the iKey from the docking port (see the ionKey/MS System Guide, part
number 715004028).
3.Shut down the MassLynx software.
4.Power-off the µSample manager.
5.Disconnect the PSPI cable.
6.Using the ¼-inch wrench, loosen and disconnect the fluid waste line (if required)
and fluid inlet lines from the µSample manager.
7.Disconnect the optional post-column addition line from the flow control of the
auxiliary solvent manager.
8.Disconnect the fluid-infusion line from the onboard IntelliStart Fluidics on the mass
spectrometer.
January 11, 2016, 715004599 Rev. B87
Page 88
3 Changing the Mode of Operation
9.Swing open the ionKey source enclosure unit from the source mounting on the mass
spectrometer.
10. Disconnect the high v oltage cable (white) from the high voltage supply outlet on the
mass spectrometer.
11. Disconnect the options cable (blue) from the options port on the mass spectrometer.
12. Carefully lift off the ionKey source module and store safely.
88Janua ry 11 , 20 16 , 71 50 0 459 9 R ev . B
Page 89
4Maintenance Procedures
This chapter provides the maintenance guidelines and procedures necessary to
maintain the instrument’s performance.
Keep to a maintenance schedule, and perform maintenance as required and
described in this chapter.
Spare parts .................................................................................................... 93
Troubleshooting with Connections INSIGHT.............................................. 94
Safety and handling ...................................................................................... 95
Preparing the instrument for operations on or inside its source................... 97
Removing and refitting the source enclosure............................................... 98
Installing and removing the corona pin........................................................ 101
Operating the source isolation valve ............................................................ 104
Removing O-rings and seals......................................................................... 107
Cleaning the instrument case........................................................................ 108
Emptying the exhaust trap bottle.................................................................. 108
Gas ballasting the roughing pump................................................................ 110
Checking the roughing pump oil level ......................................................... 112
Adding oil to the roughing pump ................................................................. 113
Cleaning the source components.................................................................. 114
Cleaning the sampling cone assembly.......................................................... 115
Cleaning the ion guide assembly.................................................................. 135
Cleaning the ion block assembly.................................................................. 125
Cleaning the ion guide assembly.................................................................. 135
Replacing the ESI probe tip and gasket........................................................ 146
Replacing the ESI probe capillary................................................................ 150
January 11, 2016, 715004599 Rev. B89
Page 90
4 Maintenance Procedures
Contents:
TopicPage
Cleaning the IonSABRE II probe tip............................................................ 166
Replacing the IonSABRE II probe sample capillary.................................... 167
Cleaning or replacing the corona pin............................................................ 172
Replacing the IonSABRE II probe heater .................................................... 174
Replacing the ion block source heater.......................................................... 177
Replacing the source assembly seals............................................................ 180
Replacing the air filter.................................................................................. 184
Replacing the roughing pump oil ................................................................. 187
Replacing the roughing pump’s oil demister el ement.................................. 190
APPI/APCI source—changing the UV lamp bulb ....................................... 195
APPI/APCI source—cleaning the lamp window.......................................... 196
APPI/APCI source—replacing the APPI lamp drive seals .......................... 198
Replacing the fluidic lines of the ionKey source.......................................... 206
Cleaning the ionKey source and connectors................................................. 211
Replacing the instrument’s fuses.................................................................. 214
90Janua ry 11 , 20 16 , 71 50 0 459 9 R ev . B
Page 91
Maintenance schedule
The following table lists periodic maintenance schedules that ensure optimum instrument
performance.
The maintenance frequencies shown apply to instruments that normally receive moderate
use.
Maintenance schedule:
ProcedureFrequencyFor information...
Clean the instrument case.As required.See page 108.
Empty the exhaust trap bottle in
the instrument exhaust line.
Replace the oil-free (scroll)
pump’s seals.
Gas ballast the roughing pump.ESI – weekly.See page 110.
Maintenance schedule
Check daily, empty as
required.
Annually.See Edwards
See page 108.
document XDS35i
Instruction Manual
A730-01-880.
Inspect and adjust the roughing
pump oil level.
Clean the source components.When they are visibly
Clean or replace the ESI probe tip. When sensitivity
Replace the ESI probe capillary.When sensitivity
Weekly.See page 112.
fouled, the background or
high-peak contaminants
are unacceptably high, or
sensitivity decreases to
unacceptable levels.
decreases to unacceptable
levels.
decreases to unacceptable
levels or sample flow is
inconsistent.
January 11, 2016, 715004599 Rev. B91
See page 114.
See page 184.
See page 150.
Page 92
4 Maintenance Procedures
Maintenance schedule:
ProcedureFrequencyFor information...
Clean the IonSABRE II probe tip.
(Options using the IonSABRE II
probe only.)
When sensitivity
decreases to unacceptable
levels or when significant
See page 166.
chemical interference is
present.
Replace the IonSABRE II probe
capillary. (Options using the
IonSABRE II probe only.)
When sensitivity
decreases to unacceptable
levels or sample flow is
See page 167.
inconsistent.
Clean or replace the corona pin
(APCI and ESCi modes).
When the corona pin is
corroded or black, the
See page 172.
corona current cannot be
stabilized, or the
sensitivity decreases to
unacceptable levels.
Replace the IonSABRE II probe
heater. (Options using the
IonSABRE II probe only.)
If the heater fails to heat
when the instrument is
pumped down
See page 174.
(evacuated).
Replace the ion block heater
cartridge.
If the heater fails to heat
when the instrument is
See page 177.
pumped down
(evacuated).
Replace the source assembly seals. Annually.See page 180.
Replace the instrument’s air filters. Annually.See page 184.
Change the roughing pump oil.Annually.See page 187.
Replace ionKey source fluid lines.As required or during
Clean the ionKey source surface,
fluid connectors, or electronic
connectors.
92Janua ry 11 , 20 16 , 71 50 0 459 9 R ev . B
periodic maintenance.
As required or during
periodic maintenance.
See page 206
See page 211.
Page 93
Spare parts
Maintenance schedule:
ProcedureFrequencyFor information...
Replace the roughing pump’s
demister element.
Clean the APPI/APCI source UV
lamp window.
Change the APPI/APCI source UV
lamp bulb.
Replace the APPI lamp drive
assembly O-rings.
Spare parts
Replace only spare parts, which are the parts mentioned in this document. For details
about spare parts, use the W aters Quality Parts
& Support page.
Annually.
See page 190.
Tip: Applications that
contaminate the roughing
pump oil reduce this
period, which must be
determined from
experience.
When the window
See page 196.
becomes visibly dirty or
when the sensitivity
decreases to unacceptable
levels.
When the bulb fails.See page 195.
Annually.See page 198.
®
Locator on the W aters Web site’s Services
January 11, 2016, 715004599 Rev. B93
Page 94
4 Maintenance Procedures
Troubleshooting with Connections INSIGHT
Connections INSIGHT® is an “intelligent” device management (IDM) Web service that
enables W aters to provide proactive service and support for the ACQUITY UPLC system.
To use Connections INSIGHT, you must install its service agent software on your
workstation. In a client/server system, the service agent must also be installed on the
computer from which you control the system. The service agent software automatically
and securely captures and sends information about the support needs of your system
directly to Waters.
If you encounter a performance issue when using the Instrument Console, you can
manually submit a Connections INSIGHT request to Waters customer support.
Alternatively, you can use Remote Desktop, a real-time collaboration option that controls
the two-way connection with the ACQUITY UPLC system by enabling the Connections
INSIGHT iAssist service level.
Consult these sources for more information about Connections INSIGHT and Connections
INSIGHT iAssist:
•http://www.waters.com
•Connections INSIGHT Installation Guide (part number 715001399)
•Connections INSIGHT User's Guide (part number 715001400)
•Your sales representative
•Your local Waters subsidiary
•Waters Customer Support
To submit a Connections INSIGHT request:
1.Open the Connections INSIGHT software tray application, and select the iHelp tab.
2.In the Connections INSIGHT Request dialog box, type your name, telephone
number, e-mail address, and a description of the problem.
3.Click Submit, and allow approximately 5 minutes to save the service profile.
Result: A .zip file containing your Connections INSIGHT profile is forwarded to
Waters customer support for review.
Note: Saving a service profile or plot file from the Instrument Console can require
as much as 150 MB of file space.
94Janua ry 11 , 20 16 , 71 50 0 459 9 R ev . B
Page 95
Safety and handling
Bear in mind the following safety considerations when performing maintenance
procedures:
Warning: To avoid personal contamination, always wear
chemical-resistant, powder-free gloves while handling instrument
components. The components can be contaminated with biohazards or toxic
materials.
Warning: To prevent injury, always observe Good Laboratory Practice
when handling solvents, changing tubing, or operating the instru ment. Know
the physical and chemical properties of the solvents used (see the Material
Safety Data Sheets for the solvents in use).
Warning: To avoid electric shock,
•do not remove the instrument’s panels. There are no user-serviceable
parts inside the instrument.
•ensure that the instrument is in Standby mode before commencing any
maintenance.
Safety and handling
Warning: To avoid burn injuries, take great care while working with the
probe and source; they can be hot.
Warning: To avoid injury, ensure that these criteria are met when
performing maintenance inside the source enclosure:
•The instrument is in Standby mode.
•LC flow is diverted to waste or set to Off.
•Desolvation gas flow is stopped.
January 11, 2016, 715004599 Rev. B95
Page 96
4 Maintenance Procedures
Notice: To avoid damaging the iKey:
•Handle it with care. The component parts are fragile.
•For recommendations regarding the maximum pressure to subject the device
to, see the iKey Separation Device Care and Use Manual (part number
720004897EN).
•Do not apply electrospray potential to the emitter without flow.
•Do not drop it.
•Do not immerse it in liquid.
•Do not freeze or overheat it. Keep the iKey within the allowed temperature
ranges during operation and in storage.
•Use the iKey sheath to protect the device when it is not in use.
•Do not bend or pull the capillary connection tubing at the ionKey source
module coupling.
•Avoid excess voltage, which can erode the emitter over time.
•Do not touch the electrospray emitter, for it can bend.
•Decompress the iKey before you remove it from the source.
See Appendix A for safety advisory information.
96Janua ry 11 , 20 16 , 71 50 0 459 9 R ev . B
Page 97
Preparing the instrument for operations on or inside its source
Preparing the instrument for operations on or inside its
source
For safety reasons, you must follow this procedure before working on the source (for
example, when changing the probe, operating the source isolation valve, and maintaining
the source).
To use MassLynx software to prepare the instrument for operations on or
inside its source:
1.In the Instrument Console, click Stop Flow to stop the LC flow or, if column
flow is required, divert the LC flow to waste as follows:
a.In the Instrument Console system tree, expand Xevo TQ-S micro, Interactive
Fluidics.
b.Click Control .
c.Select Waste as the flow state.
2.In the Instrument Console, click Standby , and confirm that the Operate
indicator is not illuminated.
3.Wait 3 minutes to allow the desolvation gas flow to cool the probe and source.
4.In the Instrument Console, click API to stop the desolvation gas flow.
5.Lift the visor on the front of the instrument so that it is clear of all the source
components and probe.
January 11, 2016, 715004599 Rev. B97
Page 98
4 Maintenance Procedures
Removing and refitting the source enclosure
The optional combined APPI/APCI, NanoFlow and ionKey sources are supplied as a
complete source enclosure. To fit them, you must first remove the standard source
enclosure.
Removing the source enclosure from the instrument
Required material
Chemical-resistant, powder-free gloves
Warning: To avoid personal contamination with biohazardous and/or toxic
materials, always wear chemical-resistant, powder-free gloves while
performing this procedure. The source components can be contaminated.
To remove the source enclosure:
1.Prepare the instrument according to the procedure on page 97.
Warning: To avoid burn injuries, take great care while working with the
probe and source; they can be hot.
2.Remove the probe from the source.
•If you are removing an ESI probe, see page 59.
•If you are removing an IonSABRE II probe, see page 64.
3.Disconnect the probe adjuster and options cables from the instrument’s connectors.
Warning: To avoid puncture wounds, take great care while working with
the source enclosure open if a corona pin is fitted. The corona pin tip is
sharp.
4.Pull the source enclosure release (located at the bottom, right-hand side) outward,
and swing open the enclosure.
98Janua ry 11 , 20 16 , 71 50 0 459 9 R ev . B
Page 99
Removing and refitting the source enclosure
TP03164
Supporting stud
Source enclosure
Cable storage positions
5.Using both hands, grasp the source enclosure and lift it vertically off the two
supporting studs on the source adaptor housing.
6.Store the cables neatly by plugging them into the cable-storage positions on the rear
of the source enclosure.
January 11, 2016, 715004599 Rev. B99
Page 100
4 Maintenance Procedures
Fitting the source enclosure to the instrument
Required material
Chemical-resistant, powder-free gloves
Warning: To avoid personal contamination with biologically hazardous, or
toxic materials, and to avoid spreading contamination to uncontaminated
surfaces, wear clean, chemical-resistant, powder-free gloves when working
with the source components.
Warning: To avoid puncture wounds, take great care while fitting the
source enclosure to the source if a corona pin is fitted (the pin tip is sharp).
To fit the source enclosure:
1.Using both hands, fit the source enclosure to the two supporting studs on the source
adaptor housing.
Notice: To prevent the sprayer from colliding with the cone and,
consequently, breaking when you use a NanoFlow source, always retract the
stage before closing the source enclosure door.
2.Close the source enclosure.
3.Connect the probe adjuster and options cables to the instrument’s connectors.
100January 11, 2016, 71500459 9 Rev . B
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.