Waters Xevo G2 QTof Operator's, Overview And Maintenance Manual

Waters Xevo G2 QTof
Operator’s Overview and Maintenance Guide
Revision B
Copyright © Waters Corporation 20102011
All rights reserved

Copyright notice

© 2010–2011 WATERS CORPORATION. PRINTED IN THE UNITED STATES OF AMERICA AND IN IRELAND. ALL RIGHTS RESERVED. THIS DOCUMENT OR PARTS THEREOF MAY NOT BE REPRODUCED IN ANY FORM WITHOUT THE WRITTEN PERMISSION OF THE PUBLISHER.
The information in this document is subject to change without notice and should not be construed as a commitment by Waters Corporation. Waters Corporation assumes no responsibility for any errors that may appear in this document. This document is believed to be complete and accurate at the time of publication. In no event shall Waters Corporation be liable for incidental or consequential damages in connection with, or arising from, its use.

Trademarks

ACQUITY, ACQUITY UPLC, Connections INSIGHT, ESCi, UPLC, and Waters are registered trademarks of Waters Corporation. IntelliStart, LockSpray, MassLynx, NanoFlow, NanoLockSpray, T-Wave, THE SCIENCE OF WHAT'S POSSIBLE., UNIFI, Xevo, and ZSpray are trademarks of Waters Corporation.
GELoader is a registered trademark of Eppendorf-Netheler-Hinz GmbH. PEEK is a trademark of Victrex plc. POZIDRIV is a registered trademark of Phillips Screw Company, Inc. Swagelok is a registered trademark of Swagelok Company. Super Flangeless and SealTight are trademarks of Upchurch Scientific, Inc. TaperTip is a trademark of New Objective, Inc. Teflon and Viton are registered trademarks of E. I. du Pont de Nemours and
Company. Valco is a trademark of Valco Instruments, Inc. Xylan is a registered trademark of Whitford Corporation. Other trademarks or registered trademarks are the sole property of their
respective owners.
ii

Customer comments

Waters’ Technical Communications department invites you to tell us of any errors you encounter in this document or to suggest ideas for otherwise improving it. Please help us better understand what you expect from our documentation so that we can continuously improve its accuracy and usability.
We seriously consider every customer comment we receive. You can reach us at tech_comm@waters.com.

Contacting Waters

Contact Waters® with enhancement requests or technical questions regarding the use, transportation, removal, or disposal of any Waters product. You can reach us via the Internet, telephone, or conventional mail.
Waters’ contact information:
Contacting medium Information
Internet The Waters Web site includes contact
information for Waters locations worldwide. Visit www.waters.com.
Telephone From the USA or Canada, phone 800
252-HPLC, or fax 508 872-1990. For other locations worldwide, phone and fax
numbers appear on the Waters Web site.
Conventional mail Waters Corporation
34 Maple Street Milford, MA 01757 USA
iii

Safety considerations

Some reagents and samples used with Waters instruments and devices can pose chemical, biological, and radiological hazards. You must know the potentially hazardous effects of all substances you work with. Always follow Good Laboratory Practice, and consult your organization’s safety representative for guidance.

Considerations specific to the Xevo G2 QTof

Solvent-leakage hazard
The source exhaust system is designed to be robust and leak-tight. Waters recommends you perform a hazard analysis, assuming a maximum leak into the laboratory atmosphere of 10% LC eluate.
Warning:
• To confirm the integrity of the source exhaust system, renew the source O-rings at intervals not exceeding one year.
• To avoid chemical degradation of the source O-rings, which can withstand exposure only to certain solvents (see “Solvents used
to prepare mobile phases” on page C-3), determine whether any
solvents you use that are not listed are chemically compatible with the composition of the O-rings.
Flammable-solvents hazard
Warning: To prevent the ignition of accumulated solvent vapors inside
the source, maintain a continuous flow of nitrogen through the source whenever significant amounts of flammable solvents are used during the instrument’s operation.
Never let the nitrogen supply pressure fall below 690 kPa (6.9 bar, 100 psi) during analyses that require flammable solvents. Connect to the LC output with a gas-fail connector to stop the LC solvent if the nitrogen supply fails.
iv
Glass-breakage hazard
Source ion block assembly
Warning: To avoid injuries from broken glass, falling objects, or
exposure to toxic substances, never place containers on top of the instrument or on its front covers.
High-temperature hazard
Warning: To avoid burn injuries, do not touch the source ion block
assembly when operating or servicing the instrument.
Xevo G2 QTof high-temperature hazard:
v
Hazards associated with removing an instrument from service
Warning: To avoid personal contamination with
biohazardous or toxic materials, wear chemical-resistant gloves during all phases of instrument decontamination.
Warning: To avoid puncture injuries, handle syringes, fused silica lines,
and borosilicate tips with care.
When you remove the instrument from use to repair or dispose of it, you must decontaminate all of its vacuum areas. These are the areas in which you can expect to encounter the highest levels of contamination:
Source interior
Waste tubing
Exhaust system
Rotary pump oil (where applicable)
The need to decontaminate other vacuum areas of the instrument depends on the kinds of samples the instrument analyzed and their levels of concentration. Do not dispose of the instrument or return it to Waters for repair until the authority responsible for approving its removal from the premises specifies the extent of decontamination required and the level of residual contamination permissible. Management must also prescribe the method of decontamination to be used and the appropriate protection for personnel undertaking the decontamination process.
You must handle items such as syringes, fused silica lines, and borosilicate tips used to carry sample into the source area in accordance with laboratory procedures for contaminated vessels and sharps. To avoid contamination by carcinogenic, toxic, or biohazardous substances, you must wear chemical-resistant gloves when handling or disposing of used oil.

Safety advisories

Consult Appendix A for a comprehensive list of warning and caution advisories.
vi

Operating this instrument

When operating this instrument, follow standard quality control procedures and the guidelines presented in this section.

Applicable symbols

Symbol Definition
Manufacturer
Authorized representative of the European Community
Confirms that a manufactured product complies with all applicable European Community directives
Australia C-Tick EMC compliant
Confirms that a manufactured product complies with all applicable United States and Canadian safety requirements
Consult instructions for use

Audience and purpose

This guide is for operators of varying levels of experience. It gives an overview of the instrument, and explains how to prepare it, change its modes of operation, and maintain it.

Intended use

Waters designed the orthogonal-acceleration, time-of-flight Xevo™ G2 QTof for use as a research tool to deliver authenticated mass measurement. The Xevo G2 QTof is for research use only and is not intended for use in diagnostic applications.
vii

Calibrating

To calibrate LC systems, follow acceptable calibration methods using at least five standards to generate a standard curve. The concentration range for standards should cover the entire range of QC samples, typical specimens, and atypical specimens.
When calibrating mass spectrometers, consult the calibration section of the operator’s guide for the instrument you are calibrating. In cases where an overview and maintenance guide, not operator’s guide, accompanies the instrument, consult the instrument’s online Help system for calibration instructions.

Quality control

Routinely run three QC samples that represent subnormal, normal, and above-normal levels of a compound. Ensure that QC sample results fall within an acceptable range, and evaluate precision from day to day and run to run. Data collected when QC samples are out of range might not be valid. Do not report these data until you are certain that the instrument performs satisfactorily.

ISM classification

ISM Classification: ISM Group 1 Class A

This classification has been assigned in accordance with CISPR 11 Industrial Scientific and Medical (ISM) instruments requirements. Group 1 products apply to intentionally generated and/or used conductively coupled radio-frequency energy that is necessary for the internal functioning of the equipment. Class A products are suitable for use in commercial, (that is, nonresidential) locations and can be directly connected to a low-voltage, power-supply network.
viii

EC authorized representative

Waters Corporation (Micromass UK Ltd.) Floats Road Wythenshawe Manchester M23 9LZ United Kingdom
Telephone: +44-161-946-2400 Fax: +44-161-946-2480 Contact: Quality manager
ix
x

Table of Contents

Copyright notice ................................................................................................... ii
Trademarks ............................................................................................................ ii
Customer comments ............................................................................................ iii
Contacting Waters ............................................................................................... iii
Safety considerations .......................................................................................... iv
Considerations specific to the Xevo G2 QTof..................................................... iv
Safety advisories................................................................................................. vi
Operating this instrument ................................................................................ vii
Applicable symbols ........................................................................................... vii
Audience and purpose....................................................................................... vii
Intended use...................................................................................................... vii
Calibrating ....................................................................................................... viii
Quality control ................................................................................................. viii
ISM classification .............................................................................................. viii
ISM Classification: ISM Group 1 Class A ...................................................... viii
EC authorized representative ........................................................................... ix
1 Waters Xevo G2 QTof Overview .......................................................... 1-1
Waters Xevo G2 QTof ........................................................................................ 1-2
IntelliStart technology..................................................................................... 1-2
ACQUITY and nanoACQUITY Xevo G2 QTof UPLC/MS systems............... 1-3
Software and data system ............................................................................... 1-6
Instrument Console ......................................................................................... 1-6
LockSpray source and ionization modes .................................................... 1-6
Electrospray ionization (ESI).......................................................................... 1-7
Atmospheric pressure chemical ionization (APCI) ........................................ 1-8
Combined electrospray and atmospheric pressure chemical
ionization (ESCi)........................................................................................ 1-9
Table of Contents xi
NanoLockSpray source and ionization modes .......................................... 1-9
Combined APPI/APCI source ....................................................................... 1-11
IntelliStart Fluidics system .......................................................................... 1-11
IntelliStart Fluidics system physical layout ................................................ 1-12
System operation ........................................................................................... 1-13
Ion optics ........................................................................................................... 1-14
Leak sensors ..................................................................................................... 1-15
Vacuum system ................................................................................................ 1-15
2 Preparing the Mass Spectrometer for Operation ........................... 2-1
Starting the mass spectrometer .................................................................... 2-2
Verifying the instrument’s state of readiness ................................................ 2-3
Monitoring the mass spectrometer LEDs....................................................... 2-3
Calibration ....................................................................................................... 2-3
Flow rates for the Xevo G2 QTof system ........................................................ 2-4
Preparing the IntelliStart Fluidics system ................................................. 2-4
Installing the reservoir bottles........................................................................ 2-4
Adjusting the solvent delivery tube positions ................................................ 2-7
Purging the pump ............................................................................................ 2-8
Rebooting the mass spectrometer ................................................................. 2-8
Leaving the mass spectrometer ready for operation ............................... 2-8
Emergency shutdown of the mass spectrometer ....................................... 2-9
3 Configuring the LockSpray Source ................................................... 3-1
Configuring the LockSpray source ............................................................... 3-2
Configuring for ESI mode ............................................................................... 3-2
Installing the ESI probe .................................................................................. 3-2
Removing the ESI probe.................................................................................. 3-6
Installing the small-bore capillary option .................................................. 3-7
xii Table of Contents
Configuring for APCI mode .......................................................................... 3-13
Installing the APCI probe ............................................................................. 3-13
Removing the APCI probe ............................................................................. 3-17
Configuring for ESCi mode .......................................................................... 3-18
Optimizing the ESI probe for ESCi operation.............................................. 3-18
4 Configuring the NanoLockSpray Source ......................................... 4-1
Overview of the NanoLockSpray source ..................................................... 4-2
Sample sprayer ................................................................................................ 4-3
Lock-spray sprayer .......................................................................................... 4-3
NanoFlow gas supply....................................................................................... 4-4
Purge gas.......................................................................................................... 4-4
Sprayer platform adjuster assembly............................................................... 4-4
Configuring the NanoLockSpray source ..................................................... 4-4
Source type selection ....................................................................................... 4-5
Advancing and retracting the sprayer platform ....................................... 4-6
Adjusting the sprayer tip position ................................................................ 4-7
Setting up the camera ...................................................................................... 4-8
Optional glass capillary sprayer ................................................................... 4-9
Plumbing the backpressure gas ...................................................................... 4-9
Installing the glass capillary sprayer ........................................................... 4-10
Fitting and loading the glass capillary......................................................... 4-11
5 Maintenance Procedures ..................................................................... 5-1
Maintenance schedule ..................................................................................... 5-3
Spare parts ......................................................................................................... 5-4
Troubleshooting with Connections INSIGHT ............................................ 5-5
Safety and handling ......................................................................................... 5-6
Preparing the instrument for working on the source ............................. 5-7
Table of Contents xiii
Removal and refitting of the source enclosure .......................................... 5-7
Removing the source enclosure from the instrument.................................... 5-7
Fitting the source enclosure to the instrument............................................ 5-10
Installing and removing the corona pin .................................................... 5-11
Installing the corona pin in the source......................................................... 5-11
Removing the corona pin from the source .................................................... 5-13
Operating the source isolation valve ......................................................... 5-14
Removing O-rings and seals ......................................................................... 5-17
Cleaning the instrument case ...................................................................... 5-17
Emptying the nitrogen exhaust trap bottle .............................................. 5-18
Inspecting the Varian roughing pump oil level ....................................... 5-20
Adding oil to the Varian roughing pump .................................................. 5-21
Replacing the Varian roughing pump’s oil and oil mist filter ............. 5-23
Emptying the roughing pump’s oil................................................................ 5-23
Replacing the oil mist filter........................................................................... 5-24
To fill the pump with oil ................................................................................ 5-26
Cleaning the source components ................................................................ 5-28
Cleaning the sampling cone assembly ....................................................... 5-29
Removing the sampling cone assembly from the source ............................. 5-29
Disassembling the sampling cone assembly................................................. 5-31
Cleaning the sample cone and cone gas nozzle ............................................ 5-34
Assembling the sampling cone assembly...................................................... 5-36
Fitting the sampling cone assembly to the source ....................................... 5-36
Cleaning the extraction cone ....................................................................... 5-38
Removing the ion block assembly from the source assembly...................... 5-38
Removing the extraction cone from the ion block ........................................ 5-41
Cleaning the extraction cone......................................................................... 5-42
Fitting the extraction cone to the ion block.................................................. 5-44
Fitting the ion block assembly to the source assembly................................ 5-45
xiv Table of Contents
Cleaning the ion block assembly ................................................................. 5-46
Disassembling the source ion block assembly.............................................. 5-46
Cleaning the ion block components .............................................................. 5-53
Assembling the source ion block assembly................................................... 5-55
Cleaning the source hexapole assembly .................................................... 5-57
Removing the ion block assembly, ion block support, and hexapole
from the source assembly ........................................................................ 5-57
Cleaning the hexapole assembly................................................................... 5-59
Fitting the hexapole assembly, PEEK ion block support, and ion
block assembly to the source assembly................................................... 5-61
Replacing the ESI probe tip and gasket .................................................... 5-63
Removing the ESI probe tip and gasket ....................................................... 5-63
Fitting the ESI probe tip and gasket............................................................ 5-65
Replacing the ESI probe sample capillary ............................................... 5-66
Removing the existing capillary.................................................................... 5-66
Installing the new capillary .......................................................................... 5-71
Cleaning the APCI probe tip ........................................................................ 5-75
Replacing the APCI probe sample capillary ............................................ 5-75
Removing the existing capillary.................................................................... 5-75
Installing the new capillary .......................................................................... 5-78
Replacing the reference probe capillary (LockSpray source) ............. 5-82
Removing the existing capillary.................................................................... 5-82
Installing the new capillary .......................................................................... 5-84
Replacing the reference probe capillary (NanoLockSpray source) ... 5-85
Removing the reference probe from the NanoLockSpray source ................ 5-85
Installing the new TaperTip and capillary................................................... 5-88
Cleaning or replacing the corona pin ........................................................ 5-90
Replacing the APCI probe heater ............................................................... 5-91
Removing the APCI probe heater ................................................................. 5-91
Fitting the new APCI probe heater .............................................................. 5-93
Replacing the ion block source heater ...................................................... 5-94
Table of Contents xv
Replacing the LockSpray source’s assembly seals ................................. 5-98
Removing the probe adjuster assembly probe and source
enclosure seals ......................................................................................... 5-98
Fitting the new source enclosure and probe adjuster assembly
probe seals.............................................................................................. 5-100
Replacing the mass spectrometer’s air filters ........................................ 5-102
Replacing the air filter inside the front door.............................................. 5-102
Replacing the air filters on the sides of the instrument............................ 5-104
Replacing the IntelliStart Fluidics tubing ............................................. 5-106
Removing the IntelliStart Fluidics tubing ................................................. 5-107
Plumbing the IntelliStart Fluidics lock-spray system............................... 5-108
Plumbing the IntelliStart Fluidics sample delivery system...................... 5-116
A Safety Advisories .................................................................................. A-1
Warning symbols ............................................................................................... A-2
Task-specific hazard warnings........................................................................ A-2
Specific warnings ............................................................................................. A-3
Caution symbol .................................................................................................. A-5
Warnings that apply to all Waters instruments ......................................... A-6
Electrical and handling symbols ................................................................. A-11
Electrical symbols .......................................................................................... A-11
Handling symbols .......................................................................................... A-12
B External Connections .......................................................................... B-1
Mass spectrometer external wiring and vacuum connections ............. B-2
Connecting the Varian oil-filled roughing pump ..................................... B-3
Making the electrical connections to the Varian oil-filled
roughing pump........................................................................................... B-6
Connecting the Edwards oil-free roughing pump ................................... B-7
Making the electrical connections to the Edwards oil-free
roughing pump......................................................................................... B-10
Connecting to the nitrogen gas supply ..................................................... B-10
xvi Table of Contents
Connecting to the collision cell gas supply ............................................. B-12
Connecting the nitrogen exhaust line ...................................................... B-12
Connecting the liquid waste line ............................................................... B-15
Connecting the EPC ...................................................................................... B-18
Connecting the workstation (systems with no ACQUITY LC) ............ B-19
Connecting Ethernet cables (systems with ACQUITY LC) .................. B-20
Input/output signal connectors .................................................................. B-20
Signal connections ......................................................................................... B-23
Connecting to the electricity source ......................................................... B-26
Connecting the NanoLockSpray source camera .................................... B-26
Installing the camera driver software .......................................................... B-27
C Materials of Construction and Compatible Solvents ................... C-1
Preventing contamination ............................................................................. C-2
Items exposed to solvent ................................................................................ C-2
Solvents used to prepare mobile phases .................................................... C-3
Index ..................................................................................................... Index-1
Table of Contents xvii
xviii Table of Contents

1 Waters Xevo G2 QTof Overview

This chapter describes the instrument, including its controls, sources and IntelliStart™ Fluidics system.
Contents:
Topic Page
Waters Xevo G2 QTof 1-2 LockSpray source and ionization modes 1-6 NanoLockSpray source and ionization modes 1-9 Combined APPI/APCI source 1-11 IntelliStart Fluidics system 1-11 Ion optics 1-14 Leak sensors 1-15 Vacuum system 1-15
1-1

Waters Xevo G2 QTof

The Xevo™ G2 QTof Mass Spectrometry (MS) system is a hybrid, quadrupole, orthogonal acceleration, time-of-flight (oaTOF) mass spectrometer operated by
®
Waters Either of the following high-performance, ZSpray™, dual-orthogonal, API
sources is fitted as standard equipment:
LockSpray™ electrospray ionization/atmospheric pressure chemical
NanoLockSpray™ ESI source, see “NanoLockSpray source and
You can also use the optional combined APPI/APCI source with the Xevo G2 QTof (see the Waters APPI Source Operator’s Guide Supplement).
For the instrument’s specifications, see the Waters Xevo G2 QTof Site Preparation Guide.

IntelliStart technology

informatics software.
ionization/combined electrospray ionization and atmospheric pressure chemical ionization (ESI/APCI/ESCi
and ionization modes” on page 1-6.
ionization modes” on page 1-9.
®
) source, see “LockSpray source
IntelliStart technology monitors instrument performance and reports when it is ready for use.
The console software automatically mass calibrates the mass spectrometer and displays performance readbacks to enable simplified setup of the system for use in routine analytical and open access applications.
1
The IntelliStart Fluidics
system is built into the mass spectrometer. It delivers sample directly to the MS probe from the LC column or from three integral reservoirs. The reservoirs can also deliver sample through direct or combined infusion so that you can optimize instrument performance at analytical flow rates. An additional reservoir contains solvent for the automated flushing of the solvent delivery system.
1. In this document, the term “fluidics” is used to describe plumbing components and fluid pathways within and between instruments and devices.
1-2 Waters Xevo G2 QTof Overview

ACQUITY and nanoACQUITY Xevo G2 QTof UPLC/MS systems

The Waters Xevo G2 QTof is compatible with the ACQUITY UPLC® and
®
nanoACQUITY UPLC the documentation relevant to your LC system.
The ACQUITY
®
Xevo G2 QTof UPLC®/MS system includes an ACQUITY UPLC system and the Waters Xevo G2 QTof fitted with the LockSpray ESI/APCI/ESCi source.
The nanoACQUITY Xevo G2 QTof UPLC/MS system includes a nanoACQUITY UPLC system and the Waters Xevo G2 QTof fitted with the NanoLockSpray source.
ACQUITY UPLC system
The ACQUITY UPLC system includes a binary solvent manager, sample manager, column heater, sample organizer, detectors, and a specialized ACQUITY UPLC column. Watersinformatics software controls the system.
For further information, see the ACQUITY UPLC System Operator’s Guide or Controlling Contamination in UPLC/MS and HPLC/MS Systems (part number 715001307). You can find the latter document on http://www.waters.com; click Services and Support > Support.
systems. If you are not using either system, refer to
Waters Xevo G2 QTof 1-3
Waters ACQUITY Xevo G2 QTof UPLC/MS system:
Sample organizer (optional)
Solvent tray
Column heater
Xevo G2 QTof
Sample manager
Binary solvent manager
Access door to the fluidics pump
High voltage connector for the ESI probe
Probe
Source interface sliding door
LockSpray source enclosure
Access door to the fluidics valves
1-4 Waters Xevo G2 QTof Overview
nanoACQUITY UPLC system
Solvent tray
Column heater
Xevo G2 QTof
Sample manager
Access door to the fluidics pump
Access door to the fluidics valve
Source interface sliding door
NanoLockSpray source enclosure
Binary solvent manager
The nanoACQUITY UPLC system includes a binary solvent manager, auxiliary solvent manager, sample manager, column heater, sample organizer, detectors, and a specialized nanoACQUITY UPLC column. Waters informatics software controls the system.
For further information, see the nanoACQUITY UPLC System Operator’s Guide or Controlling Contamination in UPLC/MS and HPLC/MS Systems (part number 715001307). You can find the latter document on http://www.waters.com; click Services and Support > Support.
Waters nanoACQUITY Xevo G2 QTof UPLC/MS system:
Waters Xevo G2 QTof 1-5

Software and data system

Waters informatics software controls the mass spectrometer. The software acquires, analyzes, manages, and distributes data from mass spectrometry, ultraviolet (UV), evaporative light scattering, and other sources.
Waters informatics software enables these major operations:
Configuring the instrument.
Creating LC and MS methods that define operating parameters for a run.
Tuning and mass calibrating the mass spectrometer.
Running samples.
Monitoring sample runs.
Acquiring data.
Processing data.
•Reviewing data.
Printing data.
See the online Help for more information

Instrument Console

The Instrument Console is an area within the Waters informatics software in which you configure settings, monitor performance, run diagnostic tests, and maintain the mass spectrometer. The instrument console functions independently of the data and does not recognize or control the data software. See the online Help for details.

LockSpray source and ionization modes

The LockSpray source uses lock-mass correction to acquire exact mass data. The analyte is introduced into the source through a probe. A reference flow, containing a compound of known mass, flows through a separate ESI probe. An oscillating baffle allows the sprays to be analyzed as two separate data functions. The lock-mass correction calculated from the reference data is then applied to the analyte data set.
You can use the LockSpray source with the ESI, APCI, and ESCi ionization modes. See Chapter 3 “Configuring the LockSpray Source”.
1-6 Waters Xevo G2 QTof Overview
Xevo G2 QTof fitted with LockSpray source:

Electrospray ionization (ESI)

In electrospray ionization (ESI), a strong electrical charge is applied to the eluent as it emerges from a nebulizer. The droplets that compose the resultant aerosol undergo a reduction in size (solvent evaporation). As solvent continues to evaporate, the charge density increases until the droplet surfaces eject ions (ion evaporation). The ions can be singly or multiply charged.
To operate the LockSpray source in ESI mode, you fit the source enclosure with an ESI probe.
LockSpray source and ionization modes 1-7
The standard ESI probe capillary accommodates flow rates of up to 2 mL/min
APCI probe
Sample cone
Corona pin
making it suitable for LC applications in the range 100 µL/min to 2 mL/min. To reduce peak broadening for lower-flow-rate LC applications, such as 1-mm UPLC columns, use the optional, small-bore capillary, which can accommodate a maximum flow rate of 200 µL/min.

Atmospheric pressure chemical ionization (APCI)

Atmospheric pressure chemical ionization (APCI) produces singly-charged protonated or deprotonated molecules for a broad range of nonvolatile analytes.
To operate the LockSpray source in APCI mode, you fit the source enclosure with a corona pin and an APCI probe. Mobile phase from the LC column enters the probe, where it is pneumatically converted to an aerosol, rapidly heated, and vaporized or gasified at the probe tip.
APCI mode:
Hot gas from the APCI probe passes between the sample cone and the corona pin. Mobile phase molecules rapidly react with ions generated by the corona discharge to produce stable reagent ions. Analyte molecules introduced into the mobile phase react with the reagent ions at atmospheric pressure and typically become protonated (in the positive ion mode) or deprotonated (in the negative ion mode). The sample and reagent ions then pass through the sample cone and into the mass spectrometer.
1-8 Waters Xevo G2 QTof Overview

Combined electrospray and atmospheric pressure chemical ionization (ESCi)

In combined electrospray and atmospheric pressure chemical ionization
©
(ESCi pin, to allow alternating acquisition of ESI and APCI ionization data, facilitating high-throughput processing and wider compound coverage.
) mode, the standard ESI probe is used in conjunction with a corona

NanoLockSpray source and ionization modes

The NanoLockSpray source allows electrospray ionization performed in the flow rate range of 5 to 1000 nL/min. For a given sample concentration, the ion currents for similar experiments approximate those in normal flow rate electrospray. However, because sample consumption is greatly reduced, the sensitivity gains are significant when similar scan parameters are used.
Lock-mass correction with the NanoLockSpray source works as the LockSpray source does in electrospray ionization mode.
The NanoLockSpray source enclosure consists of a sprayer—universal, borosilicate glass capillary, or CE (see below)—mounted on a ZSpray, three-axis manipulator.
A light within the source provides illumination for the spray, which you can observe using the video camera mounted on the corner of the source housing.
See Chapter 4 “Configuring the NanoLockSpray Source”.
NanoLockSpray source and ionization modes 1-9
Xevo G2 QTof fitted with NanoLockSpray source:
The following options are available for the spraying capillary:
Universal NanoFlow™ nebulizer sprayer. This option, for flow injection or coupling to nanoACQUITY, uses a
pump to regulate the flow rate as low as 100 nL/min.
Borosilicate glass capillary NanoFlow (nanovials). This option uses metal-coated glass capillaries, which allow the lowest
flow rates. Usable for one sample only, they must then be discarded.
NanoFlow capillary electrophoresis (CE) sprayer. This option uses a make-up liquid at the CE capillary tip, which allows a
stable electrospray to occur. The make-up flow rate is less than 1µL/min.
1-10 Waters Xevo G2 QTof Overview

Combined APPI/APCI source

Atmospheric pressure photoionization (APPI) uses photons generated by a discharge UV lamp (~10.2 eV) to produce sample ions from vaporized LC eluent. Direct photoionization of the sample molecule occurs when the photon energy exceeds the ionization potential of the sample molecule.
The optional dual-mode (APPI/APCI) ionization source comprises an APPI source enclosure, which is used in conjunction with a standard APCI probe. You can operate the source in APPI or dual-mode, which switches rapidly between ionization modes, facilitating high-throughput analyses.
For further details, see the Waters APPI Source Operator’s Guide Supplement.

IntelliStart Fluidics system

The IntelliStart Fluidics system is built into the instrument; it controls how sample is delivered to the source.
For standard flow applications, the system delivers sample directly to the mass spectrometer source in one of three ways:
From the LC column.
From three integral reservoir bottles. Use standard reservoir bottles (30 mL) for instrument setup and calibration. Use low-volume vials (1.5 mL) to infuse smaller volumes.
Tip: The reservoir bottles can also deliver sample through direct or
combined infusion to enable optimization at analytical flow rates.
From a wash reservoir, which contains solvent for automated flushing of the instrument’s solvent delivery system.
For nanoACQUITY, the valves and pumps that make up the IntelliStart Fluidics system introduce dead volume, which can cause unacceptable peak broadening. For this reason, the nanoACQUITY is plumbed directly to the NanoFlow sprayer using a suitable short piece of silica tubing.
For reference flows for both the LockSpray and NanoLockSpray source, the IntelliStart Fluidics system delivers lock mass solution from reservoir bottle B or, for extended operating hours, from a separate, external bottle of lock mass solution.
Combined APPI/APCI source 1-11

IntelliStart Fluidics system physical layout

A
B
C
AA
BB
C
AA
BB
C
Water sWater s
A
A
B
B
C
C
Lock-spray selector valve
Sample selector valve
Divert valve
Sample pump
Lock-spray pump
Sample reservoir bottles (A, B and C)
Tube guides Flow sensor
Grounded union
Access doors
The IntelliStart Fluidics system comprises the components shown in the following figure.
IntelliStart Fluidics system components:
1-12 Waters Xevo G2 QTof Overview
Loading...
+ 218 hidden pages