Waters XBridge Protein BEH, C4, 300A, 3.5 µm Columns User Manual

[ CARE AND USE MANUAL ]
XBridge Protein BEH C4, 300Å, 3.5, 5, and 10 µm Columns
CONTENTS
I. INTRODUCTION
II. GETTING STARTED
a. Column Installation b. Column Equilibration c. Initial Column Efficiency Determination d. Useful Functional Tests for Benchmarking a New Column
III. COLUMN USE
a. Sample Preparation b. Operating pH Limits c. Solvents d. Pressure c. Temperature
IV. SCALING SEPARATIONS
V. TROUBLESHOOTING
VI. COLUMN CLEANING, REGENERATING, AND STORAGE
a. Cleaning and Regeneration b. Storage
I. INTRODUCTION
Thank you for choosing a Waters reversed-phase Protein Column. The XBridge® Protein BEH C4, 300Å packing material was designed to provide excellent peak shape, high efficiency, and good recovery for biological macromolecules that are too large or too hydrophobic for separation on columns with smaller pores or longer chain bonded phases. The base particle and bonding chemistry are chosen to provide exceptional stability at both high and low pH as well as at high temperature. The XBridge Protein BEH C4, 300Å packing material is manufactured in a cGMP, ISO9002-certified plant using ultra-pure reagents. Each batch of XBridge Protein BEH C4 reversed-phase column material has been qualified with a protein test mixture, and the results are held to narrow specification ranges to ensure reproducible performance. Every column is individually tested for efficiency, and a Performance Test Chromatogram along with a Certification of Acceptance is provided with each column.
VII. CONNECTING THE COLUMN TO THE HPLC
a. Column Connectors and System Tubing Considerations b. Measuring System Bandspreading Volume
VIII. MEASURING GRADIENT SYSTEM VOLUME (OR DWELL VOLUME)
[ CARE AND USE MANUAL ]
II. GETTING STARTED
Each XBridge Protein BEH C4 Column has a Certificate of Acceptance and a Performance Test Chromatogram. The Certificate of Acceptance is specific to each batch of packing material and includes the batch number, analysis of unbonded particles, analysis of bonded particles, and chromatographic results and conditions. The Performance Test Chromatogram is specific to each individual column and contains the information: batch number, column serial number, USP plate count, USP tailing factor, retention factor, and chromatographic conditions. These data should be stored for future reference.
a. Column Installation
Note: The flow rates given in the procedure below are for a typical 3.5 μm packing in a 4.6 mm i.d. column. Scale the flow rate up or down accordingly based upon the column i.d of the column being installed. See “Scaling” section for calculating flow rates when changing column i.d and/or length. See “Connecting the Column to the HPLC” for a more detailed discussion on HPLC connections.
1. Purge the solvent delivery system of any buffer-containing or water-immiscible mobile phases and connect the inlet end of the column to the injector outlet. An arrow on the column identification label indicates the correct direction of solvent flow.
2. Flush column with 100% organic mobile phase (acetonitrile) by setting the pump flow rate to 0.1 mL/min. and increase the flow rate to 1 mL/min over 5 minutes.
3. When the mobile phase is flowing freely from the column outlet, stop the flow and attach the column outlet to the detector. This prevents entry of air into the detection system and gives more rapid baseline equilibration.
4. Gradually increase the flow rate as described in step 2.
5. Once a stable backpressure and baseline have been achieved, proceed to the next section.
Table 1: Empty Column Volumes in mL (multiply by 10 for flush solvent volumes)
Column Internal Diameter
Column Lengt h (mm)
50 0.17 0.83 3.9 14.2 35.3
75 53 100 0.35 1.7 7.9 28.4 70.7 150 0.52 2.5 11. 8 42.5 106 250 0.87 4.2 19.6 70.9 176.7
To avoid precipitating mobile phase buffers on your column or in your system, flush the column with five column volumes of a water/organic solvent mixture, using the same or lower solvent content as in the desired buffered mobile phase. For example, flush the column and HPLC system with 50% acetonitrile in water prior to introducing 50% acetonitrile/50% buffer mobile phase.
Column equilibration may be judged initially by stable pressure and by a stable detector baseline. For a specific application, it is, however, necessary to test the required duration of equilibration. The criteria for adequate equilibration include reproducibility of retention time for major and minor peaks, resolution for critical pairs, and consistent baseline characteristics.
Note: Low concentration mobile phase additives, particularly those with minimal buffering capacity may require extended equilibration and re-equilibration between gradient analyses.
2.1 4.6 10 19 30
c. Initial Column Efficiency Determination
1. Perform an efficiency test on the column before using it in the desired application. Waters recommends using the solute mixture and conditions described in the Performance Test Chromatogram to test the column upon receipt.
b. Column Equilibration
Your XBridge Protein BEH C4, 300Å Column is shipped in 100% acetonitrile. It is important to ensure mobile phase compatibility before changing to a different mobile phase system. Equilibrate the column with a minimum of 10 column volumes of the mobile phase to be used (refer to Table 1 for column volumes).
2. Measure retention of the test compounds and the number of theoretical plates (N).
3.
Repeat the test at predetermined intervals to track column performance over time. Slight variations may be obtained on two diff erent H P L C sy stems due to t he quality of th e connections, operating environment, system electronics, reagent quality, condition of column, and operator technique.
XBridge Protein BEH C4, 300Å, 3.5, 5, 10 �m
2
[ CARE AND USE MANUAL ]
d. Useful Functional Tests for Benchmarking a New Column
The Column Efficiency Test described above is a useful measure of the physical state of the packed bed as well as an indicator of the c hemical integrity of the b onded phas e. It may also be useful to benchmark the column performance with a sample that is more representative of t he intende d ap plic ation. Two tests c an be suggested as starting points for benchmarking a new column and for monitoring a column during its use.
Peptide Mixture Performance Test
Sample: Waters MassPREP™ Peptide Standard Mixture
(P/N 186002337) Reconstitute 1 vial in 100 μL 0.1% TFA:5% Acetronitrile: 94.9% Water Injection Volume: 2.1 mm – 3.3 µL
4.6 mm – 16.0 µL Column: XBridge Protein BEH C
3.5 µm, 2.1 x 50 mm Flow Rate: 2.1 mm – 0.2 mL/min
4.6 mm – 0.96 mL/min Mobile Phase: A: 0.1% TFA in water B: 0.075% in 71.4% acetonitrile/
28.6% water
Gradient:
, 300Å
4
Gradient Time for Different Column Lengths
50 mm 100 mm 150 mm 250 mm %A %B Curve
Initial Initial Initial Initial 100 0 *
30 60 90 150 30 70 6 32 64 96 160 30 70 1 50 100 150
Temperature: 40 °C Detection: 220 nm
250 100 0 1
This chromatogram is typical of the results obtained in Waters laboratories with the method described above, using a XBridge Protein BEH C4, 300Å, 3.5 µm, 2.1 x 50 mm Column. The retention times will double, triple, and be five times greater for the 100 mm, 150 mm and 250 mm columns respectively. The exact results observed in any laboratory will depend on the instrument in use. System volume, gradient generation mechanism, mixing, design of temperature control, detector cell dimensions, detector optical properties, and detector electronic properties all have a direct impact on the observed chromatogram. The pattern should be similar, however, on any well-functioning, modern HPLC. This test is exceptionally valuable for monitoring the life of the column and for troubleshooting separation difficulties that may arise.
Protein Mixture Performance Test
Sample: MassPREP Protein Standard Mixture (P/N 186004900) Dissolved in 0.1% TFA:5% acetonitrile:94.9% water
Protein Sigma P/N Conc. mg/mL
Bovine Ribonuclease A R5500 0.04 Horse Cytochrome c C7752 0.06 Bovine Serum Albumin A8022 0.20 Horse Myoglobin M1882 0.13 Yeast Enolase E6126 0.22 Rabbit Phosphorylase b P6635 0.59
Injection Volume: 2.1 mm – 5.0 µL
4.6mm – 25.0 µL Column:
3.5 µm, 2.1 x 50 mm
Flow Rate: 2.1 mm – 0.2 mL/min
4.6 mm – 0.96 mL/min
XBridge Protein BEH C4, 300Å
,
Figure 1: Typical Peptide Chromatogram Using MassPREP Peptide Standard Mixture
XBridge Protein BEH C4, 300Å, 3.5, 5, 10 �m
3
[ CARE AND USE MANUAL ]
Mobile Phase: A: 0.1% TFA in water B: 0.075% in 71.4% acetonitrile/28.6% water Gradient:
Time (Column Length)
50 mm 100 mm 150 mm 250 m m %A %B Curve
Initial Initial Initial Initial 72 28 *
25 50 75 125 0 100 6 27 54 81 135 0 100 1
45 90 135 225 72 28 1
Temperature: 40 °C Detection: 220 nm
6
1 - RNase A 2 - Cyt. C 3 - BSA 4 - Myoglobin 5 - Enolase 6 - Phosphorylase b
4
5
2
1
Figure 2: Typical Protein Test Mixture Chromatogram using MassPREP Protein Standard Mixture
3
This chromatogram is typical of the results obtained in Waters laboratories with the method described above, using a XBridge Protein BEH C4, 300Å, 3.5 µm, 2.1 x 50 mm Column. The retention times will double, triple, and be five times greater for the 100 mm, 150 mm, and 250 mm columns respectively. The exact results observed in any laboratory will depend on the instrument in use. System volume, gradient generation mechanism, mixing, design of temperature control, detector cell dimensions, detector optical properties, and detector electronic properties all have a direct impact on the observed chromatogram. The pattern should be similar, however, on any well-functioning, modern HPLC. This test is exceptionally valuable for monitoring the life of the column and for troubleshooting separation difficulties that may arise.
III. COLUMN USE
To ensure the continued high performance of XBridge Protein BEH C4, 300Å, 3.5, 5, and 10 µm Columns, follow these guidelines:
a. Sample Preparation
Sample impurities often contribute to column contamination. Samples should be free of particles before injection into the system.
It is preferable to prepare the sample in gradient solvent A or in a mobile phase that is weaker (less organic modifier) than the initial strength mobile phase. This ensures the best peak shape and.
If the sample is not dissolved in the mobile phase, ensure that the sample, solvent and mobile phases are miscible in order to avoid sample and/or buffer precipitation.
Filter sample with 0.2 μm filters to remove particulates. If the sample is dissolved in a solvent that contains an organic modifier (e.g., acetonitrile, methanol, etc.) ensure that the filter material does not dissolve in the solvent. Contact the filter manufacturer with solvent compatibility questions. Alternatively, centrifugation for 20 minutes at 12–50,000 g, followed by the t ran sfer of the supernatant liquid to an appropriate vial, could be considered.
b. Operating pH Limits
The recommended operating pH range for XBridge Protein BEH C4, 300Å, Columns is 1 to 12. A listing of commonly used buffers and additives is given in Table 2. Additionally, the column lifetime will vary depending upon the operating temperature as well as the type and concentration of buffer used.
XBridge Protein BEH C4, 300Å, 3.5, 5, 10 �m
4
[ CARE AND USE MANUAL ]
Table 2: Buffer Recommendations for Using XBridge Protein BEH C4, 300Å Columns from pH 1 to 12
Additive/Buffer pKa Buffer
range (±1
pH unit)
Volatility Used for
Mass
Spec
Comments
TFA 0.3 Volatile Yes Ion pair additive, c an suppress MS signal, used in the 0.02–0.1% range.
Acetic Acid 4.76 Volatile Yes
Formic Acid 3.75 Volatile Yes
Acetate (NH
Formate ( NH
COOH) 4.76 3.76 – 5.76 Volatile Yes
4C2
COOH) 3.75 2.75–4.75 Volatile Yes
4
Maximum buffering obtained when used with ammonium acetate salt. Used in 0.1–1.0% range.
Maximum buffering obtained when used with ammonium formate salt. Used in 0.1–1.0% range
Used in the 1–10 mM range. Note that sodium or potassium salts are not volatile.
Used in the 1–10 mM range. Note that sodium or potassium salts are not volatile
Phosphate 1 2.15 1.15–3.15 Non-volatile No Traditional low pH buffer, good UV transparency.
Phosphate 2 7.2 6.20 –8.20 Non-volatile No
Phosphate 3 12.3 11.3–13.3 Non-volatile No
Above pH 7, reduce temperature/concentration and use a guard column to maximize lifetime
Above pH 7, reduce temperature/concentration and use a guard column to maximize lifetime.
4-Methylmorpholine 8.4 7.4–9.4 Volatile Yes Generally used at 10 mM or less.
Ammonia (NH
Ammonium Bicarbonate
OH) 9.2 8.2–10.2 Volatile Yes Keep concentration below 10 mM and temperatures below 30 ˚C.
4
10.3
(HCO
9.2 (NH
-
)
8.2–11.3 Volatile Yes
3
+
)
4
Used in the 5– 10 mM range (for MS work keep source >150 ˚C). Adjust pH with ammonium hydroxide or acetic acid. Good buffering capacity at pH 10. Note: use ammonium bicarbonate (NH carbonate ([NH
2
4
CO3).
HCO3), not ammonium
4
Ammonium (Acetate) 9.2 8.2–10.2 Volatile Yes Used in the 1–10 mM range.
Ammonium (Formate) 9.2 8.2–10.2 Volatile Yes Used in the 1–10 mM range.
Borate 9.2 8.2–10.2 Non-Volatile No
CAPSO 9.7 8.7–10.7 Non-Volatile No
Reduce temperature/concentration and use a guard column to maximize lifetime.
Zwitterionic buffer, compatible with acetonitrile, used in the 1–10 mM range. Low odor.
Glycine 2.4, 9.8 8.8–10.8 Non-Volatile No Zwitterionic buffer, can give longer lifetimes than borate buffer.
1-Methylpiperidine 10.2 9.3–11.3 Volatile Yes Used in the 1–10 mM range.
CAPS 10.4 9.5–11.5 Non-Volatile No
Triethylamine (as acetate salt)
10.7 9.7–11.7 Volatile Yes
Zwitterionic buffer, compatible with acetonitrile, used in the 1–10 mM range. Low odor.
Used in the 0.1–1.0% range. Volatile only when titrated with acetic acid (not hydrochloric or phosphoric). Used as ion-pair for DNA analysis at pH 7– 9.
Pyrrolidine 11.3 10.3–12.3 Volatile Yes Mild buffer, gives long lifetime.
XBridge Protein BEH C4, 300Å, 3.5, 5, 10 �m
5
[ CARE AND USE MANUAL ]
c. Solvents
To maintain maximum column performance, use high quality chromatography grade solvents. Filter all aqueous buffers prior to use. Pall Gelman Laboratory Acrodisc® filters are recommended. Solvents containing suspended particulate materials can damage the fluidic components of the HPLC system and will generally clog the inlet distribution frit of the column. This will result in higher operating pressure and poor performance.
d. Pressure
XBridge Protein BEH C4, 300Å, 3.5, 5, and 10 µm Columns can tolerate pressures of up to 6,000 psi (400 bar or 40 Mpa) although pressures greater than 4,000-5,000 psi should be avoided in order to maximize column and system lifetimes.
e. Temperature
Temperatures between 20–90 ˚C are recommended for operating XBridge Protein BEH C4, 300Å Columns in order to enhance selectivity, lower solvent viscosity and increase mass transfer rates. However, any temperature above ambient will have a negative effect on lifetime which will vary depending on the pH and buffer conditions used.
IV. SCALING SEPARATIONS
Scale-Up Factors
Scale-up factor =
Consider scaling up from a 4.6 x 150 mm column to a 19 x 150 mm column:
Scale-up factor = = 17.1
Applying the scale-up factor, we can predict that approximately 17-135 mg of sample could be applied to the larger column (packed with the same material as the analytical column). This range is based on an analytical (4.6 mm I.D.) mass load of 1-8 mg.
Flow Rate
Flow rate (prep) = Flow rate (analytical) x
(Diameter prep)
(Diameter analytical)
(Diameter prep)2 x Length prep
(Diameter analytical)2 x Length analytical
(19)2 x 150
2
x 150
(4.6)
2
Particle Size (analytical)
x
2
Particle Size (prep)
The calculated flow rate may be used for the larger column to ensure the same linear velocity of mobile phases as used in the analytical run. However, reasonable flow rates are based on column diameters. Systems will be limited by increasing backpressure with increasing column length and decreasing particle size.
Gradient Duration (GD)
GD (prep) =
GD (anal) x Length (prep)
Length (analytical) Diameter (analytical)
Diameter (prep)2 x Flow Rate (analytical)
x
2
Flow Rate (prep)
Mass Loading
Approximate Mass Loading Capacity (mg) for Preparative OBD Columns (Gradient Mode)
Diameter (mm)
Length (mm) 4.6 10 19 30
50 3 15 45 110 75 165 100 5 25 90 225 150 8 40 135 335 250 13 60 225 560 Reasonable Flow Rate (mL/min) 1.4 6.6 24 60 Reasonable Injection Volume (µL) 20 100 350 880
The calculated prep gradient duration is entered into the pump’s gradient separation over the same number of column volumes as was used in the analytical run.
Reasonable flow rates are based on column diameter. Systems will be limited by increasing backpressure with increasing column length and decreasing particle size.
Reasonable injection volumes are based on column diameter at a length of 50 mm with relatively strong solvents. Increased length is compatible with larger injection, but not proportionately so. Weaker solvents significantly increase injection volume.
Mass loading capacities for peptide purifications depend strongly on the sequence and may be estimated at 5–20% of listed values.
Many factors affect the mass capacity of preparative columns. The listed capacities represent an ‘average’ estimate.
Capacity is:
§
Higher for strongly retained material
§
Higher for simple mixtures
§
Lower where higher resolution is required
XBridge Protein BEH C4, 300Å, 3.5, 5, 10 �m
6
[ CARE AND USE MANUAL ]
AU
AU
AU
AU
§
Very strongly dependent on loading conditions
Limited by loading volume
Limited by diluent solvent strength
Waters Preparative OBD Columns Calculator
§ Convenient scale-up tool provides:
- Mass load scaling
- Gradient scaling with appropriate flow rate scale-up
Minutes
ACQUITY UPLC Protein BEH C4,
XBridge Protein BEH C
XBridge Protein BEH C
XBridge Protein BEH C
300Å, 1.7 µm
, 300Å, 3.5 µm
4
, 300Å, 5 µm
4
, 300Å, 10 µm
4
and predicting volume consumption
- Calculations for split flow ratios for those using mass spectrometer driven chromatography
- Focused gradient UPLC to preparative method transfer
0.40
0.30
0.20
0.10
0.00
0.40
0.30
0.20
0.10
0.00
0.40
0.30
0.20
0.10
0.00
0.40
0.30
0.20
0.10
0.00
2.00 4.00 6.00 8.00 10.00 12.00 14.00 16.00 18.00 20.00
Figure 3: Separation of MassPREP Protein Standard Mixture on ACQUITY UPLC Protein BEH C Protein BEH C methods obtaining using Waters Preparative Chromatography OBD Calculator.
, 300Å, 1.7 μm (Top), XBridge Protein BEH C4, 300Å, 3.5μm, XBridge
4
, 300Å, 5 μm, XBridge Protein BEH C4, 300Å, 10 μm using scaling
4
is to ensure that the elevated pressure resides in the column rather than somewhere else in the system. This is determined by measuring pressure with and without the column attached to the instrument. If the system is occluded, the blockage should be identified and removed. If the pressure increase resides in the column, it is helpful to know whether the problem was associated with a single injection or whether it occurred over a series of injections. If the pressure gradually built up, it is likely that the column can be cleaned as described below (Section V). For future stability, it may be useful to incorporate a stronger regeneration step in the method. If a single sample caused the pressure increase, it likely reflects particulates or insoluble components, such as, lipids. Cleaning is still an option, but using the more aggressive options. The sudden pressure increase suggests that the user should consider some sample preparation, such as filtration or high speed centrifugation.
2. Loss of retention can reflect a change in the column surface chemistry. Before proceeding with diagnostic or corrective measures, check that the mobile phases have been correctly prepared and the correct method has been selected. Then repeat the efficiency test and the functional peptide or protein test. If both the small and large molecule test show loss of retention, it is likely that a significant fraction of the bonded phase has been lost, and the column will require replacement. If the changes are small and reflected only for some proteins, one of the cleaning procedures may be effective.
3.
Change in peak shape, resolution, or relative retention of
V. TROUBLESHOOTING
peaks. Follow the same steps as for loss of retention (Symptom 2).
4. Carryover and memory effects are defined as the appearance
The first step in systematic troubleshooting is comparison of the column, in its current state, to the column when it was functioning properly. The method suggested in Section I for measuring plate count is an essential first step. This technique detects physical changes to the packed bed and chemical changes in the bonded phase surface. The two functional tests with the Peptide Standard and the Protein Mixture may reveal more subtle changes in surface chemistry that affect the application.
There are several common symptoms of change in the column.
1. An increase in pressure is often associated with lost performance in the application. The first step in diagnosis
of the constituents of one sample in the next gradient analysis. First determine whether the column or the system is the source of carryover. Define a gradient method that includes an “inter
­nal gradient”. That is, the analytical gradient is repeated within a single method. If the protein peaks appear in both gradients, at the same time after start, the protein came from the column in what is often described as a “memory effect”. If the protein peaks only appear when an injection is made, they likely origi
­nate from adsorption to some system component. In that case, follow the instrument manufacturer’s recommendations. Memory effects as a source of carryover may be reduced or eliminated in several ways. First, raising the temperature of the separation
XBridge Protein BEH C4, 300Å, 3.5, 5, 10 �m
7
[ CARE AND USE MANUAL ]
reduces the possibility of incomplete elution of the protein from the column. Second, memory effects may be more pronounced with steep gradients. Keep the gradient slope at 1% per column volume or less. Third, memory effects may be exacerbated by high flow rates. Reduce the flow rate by one half while doubling the gradient time to maintain a constant slope. Fourth, memory effects may be reduced by changing the organic solvent to incorporate propanol, typically 70% pro panol :30% a cetonitril e as strong solvent. Fifth, ca rry over may be reduc ed in routine assay s with a regeneration step including a series of fast gradients from 0–100% acetonitrile. The gradients can be as short as 2 column volumes and 3–5 repetitions may be effective. This “sawtooth” gradient may be appended to each injection. Finally, apparent memory effects may actually reflect the solubility of the protein in the mobile phase. Reducing the amount injected may eliminate the effect.
5. Recovery is often improved by elevating the column temperature.
Note: Useful general information on column troubleshooting problems may be found in HPLC Columns Theory, Technology and Practice, U.D. Neue, (Wiley-VCH, 1997), the Waters HPLC Troubleshooting Guide (Literature code # 720000181EN) or visit the Waters Corporation website for information on seminars (www.waters.com).
VI. COLUMN CLEANING, REGENERATION, AND STORAGE
a. Cleaning and Regeneration
Changes in peak shape, peak splitting, shoulders on the peak, shifts in retention, change in resolution, carryover, ghost peaks, or increasing backpressure may indicate contamination of the column. Choose a cleaning option that may be expected to dissolve the suspected contaminant.
1. All cleaning procedures will be more effective at higher temperatures. The BEH300 C temperatures as high as 90 °C so it is reasonable to conduct cleaning at 70–90 °C.
can be routinely operated at
4
4. Several different cleaning solutions may be injected to strip strongly adsorbed material or particulates from the column. Make the large st injecti on p ossible with the system configuration. With such strong cleaning solutions, it is best to disconnect the detector from the column and to direct the flow to waste.
a. An injection of 1% formic acid.
b. An injection of 10% formic acid.
c. An injection of either 4M urea or 6M guanidine-HCl.
d. If contamination with lipids is suspected, a strong
cleaning option is an injection of tetrahydrofuran.
5. Flow reversal or backflushing is often suggested as part of a cleaning procedure. This should be reserved as a last resort. It may further damage the column or provide a short-lived improvement in performance.
b. Storage
For short-term storage, the column should be stored in the mobile phase with 20–50% acetonitrile. For periods longer than four days at room temp erat ure, store t h e c olumn in 100% acetonitrile. Immediately after use with elevated temperatures and/or at pH extremes, store in 100% acetonitrile for the best column lifetime. Do not store columns in highly aqueous (<20% organic) mobile phases, as this may promote bacterial growth. If the mobile phase contained a buffer salt, flush the column with 10 column volumes of HPLC grade water (see Table 1 for common column volumes) and replace with 100% acetonitrile for storage. Failure to perform this intermediate step could result in precipitation of the buffer salt in the column or system when 100% acetonitrile is introduced. Completely seal column to avoid evaporation and drying out of the bed.
V. CONNECTING THE COLUMN TO THE HPLC
2. It may be useful to conduct cleaning procedures at one-half the flow rate typic al u sed with that column. In this way the possibility of high pressure events is reduced.
3. The first and simplest cleaning procedure is to run a series of fast gradients from 0–100% acetonitrile. The gradients can be as short as 2 column volumes and 3–5 repetitions may be effective. This “sawtooth” gradient may be appended to each injection to stabilize routine assays.
a. Column Connectors and System Tubing Considerations
Tools needed:
§ 3/8-inch wrench
§ 5/16-inch wrench
Handle the column with care. Do not drop or hit the column on a hard surface as it may disturb the bed and affect its performance.
XBridge Protein BEH C4, 300Å, 3.5, 5, 10 �m
8
[ CARE AND USE MANUAL ]
e
1. Correct connection of 1/16-inch outer diameter stainless steel tu bing leading to and f rom the column is ess ential for high-quality chromatographic results.
2. When using standard stainless steel compression screw fittings, it is important to ensure proper fit of the 1/16-inch outer diameter stainless steel tubing. When tightening or loosening the compression screw, place a 5/16-inch wrench on the compression screw and a 3/8-inch wrench on the hex head of the column endfitting.
Figure 4: Proper Tubing/Column Connection
The presence of a void in the flow stream reduces column performance. This can occur if a Parker ferrule is connected to a Waters style endfitting (Figure 5).
3. If a leak occurs between the stainless steel compression screw fitting and the column endfitting, a new compression screw fitting, tubing and ferrule must be assembled.
4. An arrow on the column identification label indicates correct direction of solvent flow.
Correct connection of 1/16–inch outer diameter stainless steel tubing leading to and from the column is essential for high-quality chromatographic results. To obtain a void-free connection, the tubing must touch the bottom of the column endfitting. It is important to realize that extra column peak broadening due to voids can destroy an otherwise successful separation. The choice of appropriate column connectors and system tubing is discussed in detail below.
0.130 inch
Figure 3: Waters and Parker Ferrule Types
0.090 inches
Due to the absence of an industry standard, various column manufacturers have employed different types of chromatographic column connectors. The chromatographic separation can be negatively affected if the style of the column endfittings does not match the existing tubing ferrule settings. This section explains the differences between Waters style and Parker style ferrules and endfittings (Figure 3). Each endfitting style varies in the required length of the tubing protruding from the ferrule. The XBridge Column is equipped with Waters style endfittings that require a 0.130-inch ferrule depth. If a non-Waters style column is presently being used, it is critical that ferrule depth be reset for optimal performance prior to installing an XBridge Protein BEH C
, 300Å Column.
4
In a proper tubing/column connection (Figure 4), the tubing touches the bottom of the column endfitting, with no void between them.
Note: A void appears if tubing with a Parker ferrule is connected to a Waters style column.
0.090 inches
Figure 5: Parker Ferrule in a Waters Style Endfitting
There is only one way to fix this problem: Cut the end of the tubing with the ferrule, place a new ferrule on the tubing and make a new connection. Before tightening the screw, make sure that the tubing bottoms out in the endfitting of the column.
Conversely, if tubing with a Waters ferrule is connected to a column with Parker style endfitting, the end of the tubing will bottom out before the ferrule reaches its proper sealing position. This will leave a gap and create a leak (Figure 6).
Note: The connection leaks if a Waters ferrule is connected to a column with a Parker style endfitting.
0.130 inches
Figure 6: Waters Ferrule in a Parker Style Endfitting
There are two ways to fix the problem:
1. Tighten the screw a bit more. The ferrule moves forward, and reaches the sealing surface. Do not overtighten since this may end in breaking the screw.
2. Cut the tubing, replace the ferrule and make a new connection.
Alternatively, replace the conventional compression screw fitting with an all-in-one PEEK™ fitting (Waters Part Number PSL613315) that allows resetting of the ferrule depth. Another approach is to
XBridge Protein BEH C4, 300Å, 3.5, 5, 10 �m
9
[ CARE AND USE MANUAL ]
4.4% h
use a Keystone, Inc. SLIPFREE® connector to always ensure the correct fit. The finger tight SLIPFREE connectors automatically adjust to fit all compression screw type fittings without the use of tools (Figure 7).
Figure 7: Single and Double SLIPFREE Connectors
SLIPFREE Connector Features:
§ Tubing pushed into endfitting, thereby guaranteeing a void-free connection
§ Connector(s) come(s) installed on tubing
§ Various tubing i.d.s and lengths available
§ Fingertight to 10,000 psi – never needs wrenches
§ Readjusts to all column endfittings
§ Compatible with all commercially available endfittings
§ Unique design separates tube-holding function from sealing function
Table 3: Waters Part Numbers for SLIPFREE Connectors
SLIPFR EE Type Tubing Internal Diameter
Tubing Length
Single 6 cm PSL 618000 P SL 618006 P SL 618012 Single 10 cm PSL 618002 PSL 618008 PSL 618014 Single 20 cm P SL 618004 P SL 618010 PSL 618016 Double 6 cm PSL 618001 PSL 618007 P SL 618013 Double 10 cm PSL 618003 PSL 618009 P SL 618015 Double 20 cm PSL 618005 PSL 618001 P SL 618017
0.005” 0.010” 0.020”
Minimizing Band Spreading
Figure 8 shows the influence of tubing internal diameter on system band spreading and peak shape. As can be seen, the larger tubing diameter causes excessive peak broadening and lower sensitivity.
0.005 inches
0.020 inches
0.040 inches
Diluted/Distorted Sample Band
Figure 8: Effect of Connecting Tubing on System
b. Measuring System Variance and System Volume
This test should be performed on an HPLC system with a UV detector.
1. Disconnect column from system and replace with a zero dead volume union.
2. Set flow rate to 1 mL/min.
3. Dilute a test mix in mobile phase to give a detector sensitivity of 0.5–1.0 AUFS (system start up test mix can be used which contains uracil, ethyl, and propyl parabens; Waters Part Number WAT034544).
4. Inject 2 to 5 μL of this solution.
5. Measure the peak width at 4.4% of peak height (5-sigma method):
5-sigma Bandspreading (μL) = Peak Width (min) x Flow
Rate (mL/min) x (1000 μL/1 mL)
2
System Variance (μL
System Volume
) = (5-sigma bandspreading)2 / 25
Figure 9: Determination of System Bandspreading Volume Using 5-Sigma Method
XBridge Protein BEH C4, 300Å, 3.5, 5, 10 �m
10
[ CARE AND USE MANUAL ]
In a typical HPLC system, the Bandspreading Volume should be no
2
greater than 100 μL ± 30 μL (or Variance of 400μL
± 36μL2).
In a microbore (2.1 mm i.d.) system, the bandspreading volume should be no greater than 20 to 40 μL (or variance no greater
2
than 16μL
to 64μL2).
VIII. MEASURING SYSTEM VOLUME (OR DWELL VOLUME)
1. Remove column. 2 . Use acetonitrile as A, and acetonitrile with 0.05 mg/mL uracil
as B (eliminates non-additive mixing and viscosity problems).
3. Monitor 254 nm.
4. Use the flow rate in the original method and the intended flow rate on the target instrument.
5. Collect 100% A baseline for 5 min.
6. At 5.00 min, program a step to 100% B, and collect data for an additional 5 min.
7. Measure absorbance difference between 100% A and 100% B.
8. Measure time at 50% of that absorbance difference.
9. Calculate time difference between start of step and 50% point.
10. Multiply time difference by flow rate.
Figure 10: Determination of System Volume
©2014 Waters Corporation. Waters, T he Science of What’s Possible and XBridge are registered trademarks of Waters Corporation. MassPREP is a trademark of Waters Corporation.All other trademarks are property of their respective owners.
February 2014 WAT715001869 Rev D VW-PDF
Waters Corporation 34 Maple Street Milford, MA 01757 U.S.A. T: 1 508 478 2000 F: 1 508 872 1990 www.waters.com
Loading...