a. Column Installation
b. Column Equilibration
c. Initial Column Efficiency Determination
II. COLUMN USE
a. Sample Preparation
b. Operating pH Limits
c. Solvents
d. Pressure
e. Temperature
III. SCALING UP/DOWN ISOCRATIC METHODS
IV. TROUBLESHOOTING
V. COLUMN CLEANING, REGENERATING
AND STORAGE
a. Cleaning and Regeneration
b. Storage
VI. CONNECTING THE COLUMN TO THE HPLC
a. Column Connectors and System Tubing Considerations
b. Measuring System Bandspreading Volume
c. Measuring Gradient Delay Volume (or Dwell Volume)
Thank you for choosing a Waters XBridge® Peptide BEH C18, 130Å
or 300Å Column. The XBridge Peptide BEH C18, 130Å and 300Å
packing materials were designed to provide excellent peak shape,
high efficiency and excellent stability. The XBridge Peptide BEH C18,
130Å and 300Å packing materials are manufactured in a cGMP,
ISO 9002 certified planted using ultra pure reagent. Each batch
of XBridge Peptide BEH C18 Column material has been qualified
with a peptide separation and the results are held to narrow
specification ranges to assure excellent, reproducible performance
for peptide separations. Every column is tested and a Performance
Test Chromatogram along with a Certification of Acceptance are
provided with each column.
VII. ADDITIONAL INFORMATION
a. Use of Narrow-Bore (3.0 mm i.d.)
b. Impact of Bandspreading Volume on 2.1 mm i.d.
Column Performance
c. Non-Optimized vs. Optimized LC-MS/MS System:
System Modification Recommendations
Page 2
[ CARE AND USE MANUAL ]
I. GETTING STARTED
Each XBridge Peptide BEH C18, 130Å and 300Å Column
comes with a Certificate of Acceptance and a Performance Test
Chromatogram. The Certificate of Acceptance is specific to each
batch of packing material contained in the Peptide Separation
Technology column and includes the batch number, analysis
of unbonded particles, analysis of bonded particles, and
chromatographic results and conditions. The Performance Test
Chromatogram is specific to each individual column and contains
the information: batch number, column serial number, USP plate
count, USP tailing factor, retention factor, and chromatographic
conditions. These data should be stored for future reference.
a. Column Installation
Note: The flow rates given in the procedure below are for a typical
5 µm packing in a 4.6 mm i.d. column. Scale the flow rate up or
down accordingly based upon the column i.d., length, particle size
and backpressure of the Peptide Separation Technology column
being installed. See Scaling Up/Down Isocratic Separations section
for calculating flow rates when changing column i.d and/or length.
See “Connecting the Column to the HPLC” for a more detailed
discussion on HPLC connections.
b. Column Equilibration
XBridge Peptide BEH C18, 130Å and 300Å Columns are shipped
in 100% acetonitrile. It is important to ensure mobile phase
compatibility before changing to a different mobile phase system.
Equilibrate the column with a minimum of 10 column volumes of
the mobile phase to be used (refer to Table 1 for a listing of empty
column volumes).
To avoid precipitating out mobile phase buffers on your column
or in your system, flush the column with five column volumes of
a water/organic solvent mixture, using the same or lower solvent
content as in the desired buffered mobile phase. (For example,
flush the column and HPLC system with 60% methanol in water
prior to introducing 60% methanol/40% buffer mobile phase).
c. Initial Column Efficiency Determination
1. Perform an efficiency test on the column before using it
in the desired application. Waters recommends using a
suitable solute mixture, as found in the “Performance Test
Chromatogram,” to analyze the column upon receipt.
2. Determine the number of theoretical plates (N) and use this
value for periodic comparisons.
1. Purge the pumping system of any buffer-containing mobile
phases and connect the inlet end of the column to the injector
outlet. An arrow on the column identification label indicates
the correct direction of solvent flow.
2. Flush column with 100% organic mobile phase (methanol or
acetonitrile) by setting the pump flow rate to 0.1 mL/min. and
increase the flow rate to 1 mL/min over 5 minutes.
3. When the mobile phase is flowing freely from the column
outlet, stop the flow and attach the column outlet to the
detector. This prevents entry of air into the detection system
and gives more rapid baseline equilibration.
4. Gradually increase the flow rate as described in step 2.
5. Once a steady backpressure and baseline have been achieved,
proceed to the next section.
Note: If mobile phase additives are present in low concentrations
(e.g., ion-pairing reagents), 100 to 200 column volumes may be
required for complete equilibration. In addition, mobile phases that
contain formate (e.g., ammonium formate, formic acid, etc.) may
also require longer initial column equilibration times.
3. Repeat the test at predetermined intervals to track column
performance over time. Slight variations may be obtained
on two different HPLC systems due to the quality of the
connections, operating environment, system electronics,
reagent quality, column condition and operator technique.
XBridge Peptide BEH C18, 130Å and 300Å Columns
2
Page 3
[ CARE AND USE MANUAL ]
Table 1: Empty Column Volumes in mL (multiply by 10 for flush solvent volumes)
To ensure the continued high performance of XBridge Peptide BEH C18, 130Å and 300Å Columns follow these guidelines:
a. Sample Preparation
1. Sample impurities often contribute to column contamination.
One option to avoid this is to use Waters Oasis® solid-phase
extraction cartridges/columns or Sep-Pak® cartridges of the
appropriate chemistry to clean up the sample before analysis.
2. It is preferable to prepare the sample in the operating mobile
phase or a mobile phase that is weaker (less organic modifier)
than the mobile phase for the best peak shape and sensitivity.
3. If the sample is not dissolved in the mobile phase, ensure that
the sample, solvent and mobile phases are miscible in order to
avoid sample and/or buffer precipitation.
4. Filter sample with 0.2 µm filters to remove particulates. If
the sample is dissolved in a solvent that contains an organic
modifier (e.g., acetonitrile, methanol, etc.) ensure that the
filter material does not dissolve in the solvent. Contact the
filter manufacturer with solvent compatibility questions.
Alternatively, centrifugation for 20 minutes at 8,000 rpm,
followed by the transfer of the supernatant liquid to an
appropriate vial, could be considered.
b. Operating pH Limits
The recommended operating pH range for XBridge Peptide BEH
C18, 130Å and 300Å Columns is 1 to 12. A listing of commonly
used buffers and additives is given in Table 2. Additionally,
the column lifetime will vary depending upon the operating
temperature, the type and concentration of buffer used.
XBridge Peptide BEH C18, 130Å and 300Å Columns
3
Page 4
[ CARE AND USE MANUAL ]
Table 2: Buffer Recommendations for Using XBridge Peptide BEH C18, 130Å and 300Å Columns from pH 1 to 12
Additive/BufferpK
a
Buffer
range
TFA0.3-VolatileYes
Acetic Acid4.76-VolatileYes
Formic Acid3.75-VolatileYes
Acetate
(NH4CH2COOH)
4.763.76 – 5.76VolatileYes
Formate (NH4COOH)3.752.75 – 4.75VolatileYes
Phosphate 12.151.15 – 3.15Non-volatileNo
Phosphate 27.26.20 – 8.20Non-volatileNo
Phosphate 312.311.3 - 13.3Non-volatileNo
Volatility
(±1 pH unit)
Used for
Mass Spec
Comments
Ion pair additive, can suppress MS signal,
used in the 0.02-0.1% range.
Maximum buffering obtained when used with
ammonium acetate salt. Used in 0.1-1.0% range.
Maximum buffering obtained when used with
ammonium formate salt. Used in 0.1-1.0% range.
Used in the 1-10 mM range. Note that sodium
or potassium salts are not volatile.
Used in the 1-10 mM range. Note that sodium
or potassium salts are not volatile.
Traditional low pH buffer, good UV
transparency.
Above pH 7, reduce temperature/concentration
and use a guard column to maximize lifetime.
Above pH 7, reduce temperature/concentration
and use a guard column to maximize lifetime.
4-Methylmorpholine~8.47.4 – 9.4VolatileYesGenerally used at 10 mM or less.
Ammonia (NH4OH)9.28.2 – 10.2VolatileYes
Keep concentration below 10 mM and
temperatures below 30 ˚C.
Used in the 5-10 mM range (for MS work keep
source >150 ˚C ). Adjust pH with ammonium
Ammonium
Bicarbonate
10.3 (HCO
9.2 (NH
-
)
3
+
)
4
8.2 – 11.3VolatileYes
hydroxide or acetic acid. Good buffering
capacity at pH 10
Note: use ammonium bicarbonate (NH4HCO3),
not ammonium carbonate ([NH4]2CO3).
Ammonium (Acetate)9.28.2 – 10.2VolatileYesUsed in the 1-10 mM range.
Ammonium (Formate)9.28.2 – 10.2VolatileYesUsed in the 1-10 mM range.
Borate9.28.2 – 10.2Non-VolatileNo
CAPSO9.78.7 – 10.7Non-VolatileNo
Glycine2.4, 9.88.8 – 10.8Non-VolatileNo
Reduce temperature/concentration and use a
guard column to maximize lifetime.
Zwitterionic buffer, compatible with acetonitrile,
used in the 1-10 mM range. Low odor.
Zwitterionic buffer, can give longer lifetimes
than borate buffer.
1-Methylpiperidine10.29.3 – 11.3VolatileYesUsed in the 1-10 mM range.
CAPS10.49.5 – 11.5Non-VolatileNo
Zwitterionic buffer, compatible with acetonitrile,
used in the 1-10 mM range. Low odor.
Used in the 0.1-1.0% range. Volatile only when
Triethylamine
(as acetate salt)
10.79.7 – 11.7VolatileYes
titrated with acetic acid (not hydrochloric or
phosphoric). Used as ion-pair for DNA analysis
at pH 7-9
Pyrrolidine11. 310.3 – 12.3VolatileYesMild buffer, gives long lifetime.
XBridge Peptide BEH C18, 130Å and 300Å Columns
4
Page 5
[ CARE AND USE MANUAL ]
c. Solvents
To maintain maximum column performance, use high quality
chromatography grade solvents. Filter all aqueous buffers prior to
use. Pall Gelman Laboratory Acrodisc® filters are recommended.
Solvents containing suspended particulate materials will
generally clog the outside surface of the inlet distribution frit of
the column. This will result in higher operating pressure and poor
performance.
d. Pressure
XBridge Peptide BEH C18, 130Å and 300Å Columns can tolerate
pressures of up to 6,000 psi (400 bar or 40 Mpa) although
pressures greater than 4,000 – 5,000 psi should be avoided in
order to maximize column and system lifetimes.
e. Temperature
Temperatures between 20 ˚C – 60 ˚C are recommended for
operating XBridge Peptide BEH C18, 130Å and 300Å Columns in
order to enhance selectivity, lower solvent viscosity and increase
mass transfer rates. However, any temperature above ambient will
have a negative effect on lifetime which will vary depending on
the pH and buffer conditions used.
III. SCALING UP/DOWN ISOCRATIC METHODS
The following formulas will allow scale up or scale down, while
maintaining the same linear velocity, and provide new sample
loading values:
If column i.d. and length are altered:
F2 = F1 (r2/r1)
Load2 = Load1 (r2/r1)2 (L2/L1)
Injection volume2 = Injection volume1(r2/r1)2 (L2/L1)
Where: r = Radius of the column
F = Flow rate
L = Length of column
1 = Original, or reference column
2 = New column
2
Information on column troubleshooting problems may be found
in HPLC Columns Theory, Technology and Practice, U.D. Neue,
(Wiley-VCH, 1997), the Waters HPLC Troubleshooting Guide
(Literature code # 720000181EN) or visit the Waters Corporation
website for information on seminars (www.waters.com).
VI. COLUMN CLEANING, REGENERATION
AND STORAGE
a. Cleaning and Regeneration
Changes in peak shape, peak splitting, shoulders on the
peak, shifts in retention, change in resolution or increasing
backpressure may indicate contamination of the column. Flushing
with a neat organic solvent, taking care not to precipitate buffers,
is usually sufficient to remove the contaminant. If the flushing
procedure does not solve the problem, purge the column using the
following cleaning and regeneration procedures.
Use the cleaning routine that matches the properties of the
samples and/or what you believe is contaminating the column
(see Table 3). Flush columns with 20 column volumes each of
HPLC-grade solvents (e.g., 80 mL total for 4.6 x 250 mm column)
listed in Table 3. Increasing mobile phase temperature to 35-55
˚C increases cleaning efficiency. If the column performance is poor
after cleaning and regeneration, call your local Waters office for
additional support.
Table 3: Cleaning and Regeneration Sequence or Options
Polar
Samples
1. Water
2. Methanol
3. Isopropanol
Proteinaceous Samples
Option 1: Inject repeated 100 µL aliquots of
dimethylsulfoxide (DMSO) using a reduced
flow rate delivering 50% Eluent A and 50%
Eluent B
Option 2: gradient of 10% to 90% B where:
A = 0.1% trifluoroacetic acid (TFA) in water
B = 0.1% trifluoroacetic acid (TFA) in
acetonitrile (CH3CN)
Option 3: Flush column with 7 M guanidine
hydrochloride, or 7 M urea
IV. TROUBLESHOOTING
Changes in retention time, resolution, or backpressure are
often due to column contamination. See the Column Cleaning,
Regeneration and Storage section of this Care and Use Manual.
Note: To avoid potentially damaging precipitation within your
column (e.g., if your separation eluent contains phosphate buffer),
be certain to flush column with 5 to 10 column volumes of water
BEFORE using suggested organic eluent column wash procedures.
XBridge Peptide BEH C18, 130Å and 300Å Columns
5
Page 6
[ CARE AND USE MANUAL ]
b. Storage
For periods longer than four days at room temperature, store the
column in 100% acetonitrile. Immediately after use with elevated
temperatures and/or at pH extremes, store in 100% acetonitrile
for the best column lifetime. Do not store columns in highly
aqueous (<20% organic) mobile phases, as this may promote
bacterial growth. If the mobile phase contained a buffer salt, flush
the column with 10 column volumes of HPLC grade water (see
Table 1 for common column volumes) and replace with 100%
acetonitrile for storage. Failure to perform this intermediate step
could result in precipitation of the buffer salt in the column or
system when 100% acetonitrile is introduced. Completely seal
column to avoid evaporation and drying out of the bed.
Note: If a column has been run with a mobile phase that contains
formate (e.g., ammonium formate, formic acid, etc.) and is then
flushed with 100% acetonitrile, slightly longer equilibration times
may be necessary when the column is re-installed and run again
with a formate-containing mobile phase.
V. CONNECTING THE COLUMN TO THE HPLC
a. Column Connectors and System Tubing Considerations
Tools needed:
3/8 inch wrench
5/16 inch wrench
Handle the column with care. Do not drop or hit the column on a
hard surface as it may disturb the bed and affect its performance.
1. Correct connection of 1/16 inch outer diameter stainless steel
tubing leading to and from the column is essential for highquality chromatographic results.
2. When using standard stainless steel compression screw
fittings, it is important to ensure proper fit of the 1/16 inch
outer diameter stainless steel tubing. When tightening or
loosening the compression screw, place a 5/16 inch wrench on
the compression screw and a 3/8 inch wrench on the hex head
of the column endfitting.
Note: If one of the wrenches is placed on the column tube flat
during this process, the endfitting will be loosened and leak.
3. If a leak occurs between the stainless steel compression screw
fitting and the column endfitting, a new compression screw
fitting, tubing and ferrule must be assembled.
4. An arrow on the column identification label indicates correct
direction of solvent flow.
Correct connection of 1/16 inch outer diameter stainless steel
tubing leading to and from the column is essential for high-quality
chromatographic results. To obtain a void-free connection, the
tubing must touch the bottom of the column endfitting. It is
important to realize that extra column peak broadening due to
voids can destroy an otherwise successful separation. The choice
of appropriate column connectors and system tubing is discussed
in detail below.
Figure 1. Waters and Parker Ferrule Types.
Waters Ferrule Setting
.130”
Parker Ferrule Setting
.090”
Due to the absence of an industry standard, various column
manufacturers have employed different types of chromatographic
column connectors. The chromatographic separation can be
negatively affected if the style of the column endfittings does not
match the existing tubing ferrule settings. This section explains
the differences between Waters style and Parker style ferrules and
endfittings (Figure 1). Each endfitting style varies in the required
length of the tubing protruding from the ferrule. The XBridge
Column is equipped with Waters style endfittings that require a
0.130 inch ferrule depth. If a non-Waters style column is presently
being used, it is critical that ferrule depth be reset for optimal
performance prior to installing a XBridge column.
In a proper tubing/column connection (Figure 2), the tubing
touches the bottom of the column endfitting, with no void
between them.
Figure 2. Proper Tubing/Column Connection.
The presence of a void in the
flow stream reduces column
performance. This can occur if
a Parker ferrule is connected
to a Waters style endfitting
(Figure 3).
Note: A void appears if tubing with a Parker ferrule is connected to
a Waters style column.
XBridge Peptide BEH C18, 130Å and 300Å Columns
6
Page 7
Diluted/Distorted Sample Band
[ CARE AND USE MANUAL ]
Figure 3. Parker Ferrule in a Waters Style Endfitting.
Void
There is only one way to fix this problem: Cut the end of the tubing
with the ferrule, place a new ferrule on the tubing and make a new
connection. Before tightening the screw, make sure that the tubing
bottoms out in the endfitting of the column.
Conversely, if tubing with a Waters ferrule is connected to a
column with Parker style endfitting, the end of the tubing will
bottom out before the ferrule reaches its proper sealing position.
This will leave a gap and create a leak (Figure 4).
Note: The connection leaks if a Waters ferrule is connected to a
column with a Parker style endfitting.
Figure 4. Waters Ferrule in a Parker Style Endfitting.
SLIPFREE connector features:
Tubing pushed into endfitting, thereby guaranteeing a
void-free connection
Connector(s) come(s) installed on tubing
Various tubing i.d’s and lengths available
Fingertight to 10,000 psi – Never needs wrenches
Readjusts to all column endfittings
Compatible with all commercially available endfittings
Unique design separates tube-holding function from
sealing function
Table 5: Waters Part Numbers for SLIPFREE Connectors
SLIP FREE TypeTubing Internal Diameter
Tubing Length0.005”0.010”0.020”
Single 6 cmPSL 618000PSL 618006P SL 618012
Single 10 cmPSL 618002PSL 618008 P SL 618014
Single 20 cmPSL 61800 4PSL 618010P SL 618016
Double 6 cmP SL 618001PS L 618007PSL 618013
Double 10 cmPSL 618003PSL 61800 9PSL 618 015
Double 20 cmPSL 618005P SL 618001PSL 618017
Gap
There are two ways to fix the problem:
1. Tighten the screw a bit more. The ferrule moves forward, and
reaches the sealing surface. Do not overtighten since this may
end in breaking the screw.
2. Cut the tubing, replace the ferrule and make a new connection.
Alternatively, replace the conventional compression screw fitting
with an all-in-one PEEK™ fitting (Waters Part Number PSL613315)
that allows resetting of the ferrule depth. Another approach is to
use a SLIPFREE® connector to always ensure the correct fit. The
fingertight SLIPFREE connectors automatically adjust to fit all
compression screw type fittings without the use of tools (Figure 5).
Figure 5. Single and Double SLIPFREE Connectors.
Band Spreading Minimization
Figure 6 shows the influence of tubing internal diameter on
system band spreading and peak shape. As can be seen, the
larger tubing diameter causes excessive peak broadening and
lower sensitivity.
Figure 6. Effect of Connecting Tubing on System.
0.005 inches
0.020 inches
0.040 inches
XBridge Peptide BEH C18, 130Å and 300Å Columns
7
Page 8
[ CARE AND USE MANUAL ]
b. Measuring System Bandspreading Volume and
System Variance
This test should be performed on an HPLC system with a single
wavelength UV detector (not a Photodiode Array [PDA]).
1. Disconnect column from system and replace with a zero dead
volume union.
2. Set flow rate to 1 mL/min.
3. Dilute a test mix in mobile phase to give a detector sensitivity
of 0.5–1.0 AUFS (system start up test mix can be used which
contains uracil, ethyl and propyl parabens; Waters Part Number
WAT034544).
4. Inject 2 to 5 µL of this solution.
5. Measure the peak width at 4.4% of peak height
(5-sigma method):
5-sigma Bandspreading (µL) =
Peak Width (min) x Flow Rate (mL/min) x (1000 µL/1 mL)
System Variance (µL2) = (5-sigma bandspreading)2/25
1. Replace the column with a zero dead volume union.
2. Prepare mobile phase A (pure solvent, such as methanol) and
mobile phase B (mobile phase A with a UV absorbing sample,
such as (v/v) 0.1% acetone in methanol).
3. Equilibrate the system with mobile phase A until a stable
baseline is achieved.
4. Set the detector wavelength to the absorbance maximum of the
probe (265 nm for acetone).
5. Program a 0-100% B linear gradient in 10 min at 2 mL/min (the
exact conditions are not critical; just make sure the gradient
volume is at least 20 mL) with a hold at 100% B.
6. Determine the dwell time by first locating the time at the
midpoint of the formed gradient (t
) (half the vertical
1/2
distance between the initial and final isocratic segments as
shown in Figure 8).
Figure 8. Determination of Gradient Delay Volume.
1.0
Figure 7. Determination of System Bandspreading Volume Using
5-Sigma Method.
System Volume
5
4.4 %h
In a typical HPLC system, the Bandspreading Volume should be no
greater than 100 µL ± 30 µL (or Variance of 400 µL2 ± 36 µL2).
In a microbore (2.1 mm i.d.) system, the Bandspreading Volume
should be no greater than 20 to 40 µL (or Variance no greater
than 16 µL2 to 64 µL2).
c. Measuring Gradient Delay Volume (or Dwell Volume)
For successful gradient-method transfers the gradient delay
volumes should be measured using the same method on both HPLC
systems. The procedure below describes a method for determining
the gradient delay volumes.
0.8
0.6
Au
0.4
0.2
0.0
1/2 Vertical
Distance
t
1/2
Time
7. Subtract half the gradient time (1/2 tg) (10 min/2 = 5 min in
this example) from the gradient midpoint (t
) to obtain the
1/2
dwell time (tD).
8. Convert the dwell time (tD) to the dwell volume (VD) by
multiplying by the flow rate (F).
Dwell Volume VD = (t
—1/2 tg) x F
1/2
For fast gradient methods, the gradient delay volume (or dwell
volume) should be less than 1 mL. If the gradient delay volume
is greater than 1 mL, see System Modification Recommendations
section on how to reduce system volume.
XBridge Peptide BEH C18, 130Å and 300Å Columns
8
Page 9
[ CARE AND USE MANUAL ]
VII. ADDITIONAL INFORMATION
a. Use of Narrow-Bore Columns
This section describes how to minimize extra column effects
and provides guidelines on maximizing the performance of a
narrow-bore column in an HPLC system. A 2.1 mm i.d. column
requires modifications to the HPLC system in order to eliminate
excessive system bandspreading volume. Without proper system
modifications, excessive system bandspreading volume causes
peak broadening and has a large impact on peak width as peak
volume decreases.
b. Impact of Bandspreading Volume on 2.1 mm i.d.
Column Performance
System with 70 µL bandspreading: 10,000 plates
System with 130 µL bandspreading: 8,000 plates (same column)
Note: Flow splitters after the column will introduce additional
band-spreading.
System optimization, especially in a system that contains
a flow splitter, can have dramatic effects on sensitivity and
resolution. Optimization includes using correct ferrule depths
and minimizing tubing inner diameters and lengths. An example
is given in Figure 9 where system optimization resulted in a
doubling of sensitivity and resolution of the metabolite in an
LC-MS/MS system.
Figure 9. Non-Optimized vs. Optimized LC-MS/MS System.
7.007.50
Non-optimized LC/MS/MS SystemOptimized System
8.00
7.007.508.00
c. Non-Optimized vs. Optimized LC-MS/MS System:
System Modification Recommendations
1. Use a microbore detector flow cell with 2.1 mm i.d. columns.
Note: Detector sensitivity is reduced with the shorter flow cell path
length in order to achieve lower bandspreading volume.
2. Minimize injector sample loop volume.
3. Use 0.009 inch (0.25 mm) tubing for rest of connections
in standard systems and 0.005 inch (0.12 mm) tubing for
narrowbore (2.1 mm i.d.) systems.
4. Use perfect (pre-cut) connections (with a variable depth inlet
if using columns from different suppliers).
5. Detector time constants should be shortened to less than