Waters XBridge Peptide BEH C18 130A and 300A Columns User Manual

[ CARE AND USE MANUAL ]
XBridge Peptide BEH C18, 130Å and 300Å Columns
CONTENTS I. GETTING STARTED
a. Column Installation b. Column Equilibration c. Initial Column Efficiency Determination
II. COLUMN USE
a. Sample Preparation b. Operating pH Limits c. Solvents d. Pressure e. Temperature
III. SCALING UP/DOWN ISOCRATIC METHODS
IV. TROUBLESHOOTING
V. COLUMN CLEANING, REGENERATING AND STORAGE
a. Cleaning and Regeneration b. Storage
VI. CONNECTING THE COLUMN TO THE HPLC
a. Column Connectors and System Tubing Considerations b. Measuring System Bandspreading Volume c. Measuring Gradient Delay Volume (or Dwell Volume)
Thank you for choosing a Waters XBridge® Peptide BEH C18, 130Å or 300Å Column. The XBridge Peptide BEH C18, 130Å and 300Å packing materials were designed to provide excellent peak shape, high efficiency and excellent stability. The XBridge Peptide BEH C18, 130Å and 300Å packing materials are manufactured in a cGMP, ISO 9002 certified planted using ultra pure reagent. Each batch of XBridge Peptide BEH C18 Column material has been qualified with a peptide separation and the results are held to narrow specification ranges to assure excellent, reproducible performance for peptide separations. Every column is tested and a Performance Test Chromatogram along with a Certification of Acceptance are provided with each column.
VII. ADDITIONAL INFORMATION
a. Use of Narrow-Bore (3.0 mm i.d.) b. Impact of Bandspreading Volume on 2.1 mm i.d. Column Performance c. Non-Optimized vs. Optimized LC-MS/MS System: System Modification Recommendations
[ CARE AND USE MANUAL ]
I. GETTING STARTED
Each XBridge Peptide BEH C18, 130Å and 300Å Column comes with a Certificate of Acceptance and a Performance Test Chromatogram. The Certificate of Acceptance is specific to each batch of packing material contained in the Peptide Separation Technology column and includes the batch number, analysis of unbonded particles, analysis of bonded particles, and chromatographic results and conditions. The Performance Test Chromatogram is specific to each individual column and contains the information: batch number, column serial number, USP plate count, USP tailing factor, retention factor, and chromatographic conditions. These data should be stored for future reference.
a. Column Installation
Note: The flow rates given in the procedure below are for a typical
5 µm packing in a 4.6 mm i.d. column. Scale the flow rate up or
down accordingly based upon the column i.d., length, particle size
and backpressure of the Peptide Separation Technology column
being installed. See Scaling Up/Down Isocratic Separations section
for calculating flow rates when changing column i.d and/or length.
See “Connecting the Column to the HPLC” for a more detailed
discussion on HPLC connections.
b. Column Equilibration
XBridge Peptide BEH C18, 130Å and 300Å Columns are shipped in 100% acetonitrile. It is important to ensure mobile phase compatibility before changing to a different mobile phase system. Equilibrate the column with a minimum of 10 column volumes of the mobile phase to be used (refer to Table 1 for a listing of empty column volumes).
To avoid precipitating out mobile phase buffers on your column or in your system, flush the column with five column volumes of a water/organic solvent mixture, using the same or lower solvent content as in the desired buffered mobile phase. (For example, flush the column and HPLC system with 60% methanol in water prior to introducing 60% methanol/40% buffer mobile phase).
c. Initial Column Efficiency Determination
1. Perform an efficiency test on the column before using it in the desired application. Waters recommends using a suitable solute mixture, as found in the “Performance Test Chromatogram,” to analyze the column upon receipt.
2. Determine the number of theoretical plates (N) and use this value for periodic comparisons.
1. Purge the pumping system of any buffer-containing mobile phases and connect the inlet end of the column to the injector outlet. An arrow on the column identification label indicates the correct direction of solvent flow.
2. Flush column with 100% organic mobile phase (methanol or acetonitrile) by setting the pump flow rate to 0.1 mL/min. and increase the flow rate to 1 mL/min over 5 minutes.
3. When the mobile phase is flowing freely from the column outlet, stop the flow and attach the column outlet to the detector. This prevents entry of air into the detection system and gives more rapid baseline equilibration.
4. Gradually increase the flow rate as described in step 2.
5. Once a steady backpressure and baseline have been achieved, proceed to the next section.
Note: If mobile phase additives are present in low concentrations
(e.g., ion-pairing reagents), 100 to 200 column volumes may be
required for complete equilibration. In addition, mobile phases that
contain formate (e.g., ammonium formate, formic acid, etc.) may
also require longer initial column equilibration times.
3. Repeat the test at predetermined intervals to track column performance over time. Slight variations may be obtained on two different HPLC systems due to the quality of the connections, operating environment, system electronics, reagent quality, column condition and operator technique.
XBridge Peptide BEH C18, 130Å and 300Å Columns
2
[ CARE AND USE MANUAL ]
Table 1: Empty Column Volumes in mL (multiply by 10 for flush solvent volumes)
Column internal diameter (mm)
Column Length (mm) 1.0 2.1 4.6 10 19 30
50 0.04 0.17 0.83 3.9 14 35 100 0.08 0.35 1.7 7.8 28 70 150 0.12 0.52 2.5 12 42 106 250 - 0.87 4.2 20 70 176
II. COLUMN USE
To ensure the continued high performance of XBridge Peptide BEH C18, 130Å and 300Å Columns follow these guidelines:
a. Sample Preparation
1. Sample impurities often contribute to column contamination. One option to avoid this is to use Waters Oasis® solid-phase extraction cartridges/columns or Sep-Pak® cartridges of the appropriate chemistry to clean up the sample before analysis.
2. It is preferable to prepare the sample in the operating mobile phase or a mobile phase that is weaker (less organic modifier) than the mobile phase for the best peak shape and sensitivity.
3. If the sample is not dissolved in the mobile phase, ensure that the sample, solvent and mobile phases are miscible in order to avoid sample and/or buffer precipitation.
4. Filter sample with 0.2 µm filters to remove particulates. If the sample is dissolved in a solvent that contains an organic modifier (e.g., acetonitrile, methanol, etc.) ensure that the filter material does not dissolve in the solvent. Contact the filter manufacturer with solvent compatibility questions. Alternatively, centrifugation for 20 minutes at 8,000 rpm, followed by the transfer of the supernatant liquid to an appropriate vial, could be considered.
b. Operating pH Limits
The recommended operating pH range for XBridge Peptide BEH C18, 130Å and 300Å Columns is 1 to 12. A listing of commonly used buffers and additives is given in Table 2. Additionally, the column lifetime will vary depending upon the operating temperature, the type and concentration of buffer used.
XBridge Peptide BEH C18, 130Å and 300Å Columns
3
Loading...
+ 6 hidden pages