[ Care and Use ManUal ]
VIRIDIs sFC CoLUMns
Contents
I. IntRoDUCtIon
II. GettInG stARteD
a. Safety Considerations
b. Column Installation
c. Column Equilibration
d. Initial Column Efficiency Determination
III. CoLUMn Use
a. Sample Preparation
b. Solvents
c. Pressure
IV. sCALInG UP/DoWn IsoCRAtIC MetHoDs
V. ConneCtInG tHe CoLUMn to tHe sFC sYsteM
I. IntRoDUCtIon
Thank you for choosing a Viridis® SFC column. The Viridis SFC packing
materials are designed to provide excellent stability, minimal column
bleed, and high mass loading. The Viridis SFC packing materials are
manufactured in an ISO 9000 certified plant using ultra-pure reagents.
Each batch of Viridis SFC material is tested and the results are held to
narrow specification ranges to assure excellent, reproducible performance.
Every column is individually tested and a Performance Chromatogram is
provided with each column along with the Certificate of Batch Analysis.
Table 1. Physical Characteristics
Viridis
SFC
Chemistry
2-Ethylpyridine
Silica Spherical 5 10 0 340 0.90 n/a n/a
Particle
Shape
Spherical 5 10 0 340 0.90 8 No
Particle
Size
(µm)
Pore
Size
(Å)
Surface
Area
(m2/g)
Pore
Volume
(cc /g)
Carbon
Load
(%)
End-
capped
a. Column Connectors and System Tubing Considerations
b. Band Spreading Minimization
VI. tRoUBLesHootInG
VII. CoLUMn CLeAnInG, ReGeneRAtInG AnD stoRAGe
a. Cleaning and Regeneration
b. Storage
1Viridis SFC Columns
[ Care and Use ManUal ]
II. GettInG stARteD
Each Viridis SFC column comes with a Certificate of Analysis and a
Performance Test Chromatogram. The Certificate of Analysis is specific
to each batch of packing material and includes the gel batch number,
analysis of unbonded particle and analysis of bonded particles. The
Performance Test Chromatogram is specific to each individual column
and contains the gel batch number, column serial number, USP plate
count, USP tailing factor, capacity factor, and chromatographic condi-
tions under normal-phase LC conditions. These data should be stored
for future reference.
a. Safety Considerations
An SFC column while being used is, by design, under pressure. The
major safety concern for even a small leak with just CO
is frostbite.
2
Pay attention to any frosting on the column or system connections.
This indicates a leak, usually with temperatures far below 0 °C. Any
small leak with solvent introduced could produce a situation where
the LEL (lower exposure limit) is reached.
b. Column Installation
Note: Scale the flow rate up or down accordingly based upon the i.d.,
length, particle size, and backpressure of the Viridis SFC column being
installed. See “Scaling Up/Down Isocratic Methods” for calculating
flow rates when changing column i.d. and/or length. See “Connecting
the Column to the SFC system” for a more detailed discussion on SFC
connections.
c. Column Equilibration
Viridis SFC columns are shipped dry. Equilibrate the column with a mini-
mum of 10 column volumes of the mobile phase prior to use. (Refer to
Table 2 for a listing of empty column volumes.)
d. Initial Column Efficiency Determination
1. Perform an efficiency test on the column before using it. Waters
recommends using a suitable solute mixture to analyze the column
upon receipt.
2. Determine the number of theoretical plates (N) and use this value
as a benchmark for periodic comparisons.
3. Repeat the test at predetermined intervals to track column
performance over time. Slight variations may be obtained on
two different SFC systems due to the quality of the connections,
operating environment, system electronics, reagent quality and
column condition.
Table 2. Empty Column Volumes in mL (multiply by 10 for flush solvent volumes)
Column
Dimension
50 mm 0.8 4 14 35 98
100 mm 1.7 8 28 70 196
150 mm 2.5 12 42 106 294
250 mm 4 20 70 176 490
4.6 mm 10 mm 19 mm 30 mm 50 mm
Column Internal Diameter (mm)
1. Make sure your co-solvent pump is primed and has an adequate
solvent/modifier supply before performing injections.
2. Connect both the inlet and outlet of the column to the SFC system.
3. If the column is still filled with a solvent, use a low flow rate and
back pressure setting (100 bar) to start pumping CO
and modifier
2
through the column.
4. If you see frosting on the column at the inlet or outlet, tighten the
compression screw on that side. If you see frosting on the column,
turn off the CO
and vent the system. Allow the column to depres-
2
surize fully before disconnecting the inlet or outlet to troubleshoot
the leaking issue.
III. CoLUMn Use
To ensure the continued high performance of Viridis SFC columns,
follow these guidelines:
a. Sample Preparation
1. Sample impurities often contribute to column contamination.
®
Use Oasis
Pak
sample before analysis. Link to www.waters.com/sampleprep.
2. Prepare the sample in the SFC mobile-phase modifier (typically,
methanol) for the best peak shape and sensitivity.
2Viridis SFC Columns
solid-phase extraction cartridges/columns or Sep-
®
cartridges of the appropriate chemistry to clean up the