Waters TQ Detector Quick Start Manual

Page 1
Waters TQ Detector
Quick Start Guide
71500126803/Revision A
Copyright © Waters Corporation 2007.
All rights reserved.
Page 2
Copyright notice
© 2007 WATERS CORPORATION. PRINTED IN THE UNITED STATES OF AMERICA AND IRELAND. ALL RIGHTS RESERVED. THIS DOCUMENT OR PARTS THEREOF MAY NOT BE REPRODUCED IN ANY FORM WITHOUT THE WRITTEN PERMISSION OF THE PUBLISHER.
The information in this document is subject to change without notice and should not be construed as a commitment by Waters Corporation. Waters Corporation assumes no responsibility for any errors that may appear in this document. This document is believed to be complete and accurate at the time of publication. In no event shall Waters Corporation be liable for incidental or consequential damages in connection with, or arising from, its use.
Waters Corporation 34 Maple Street Milford, MA 01757 USA
Trademarks
Connections Insight, ESCi, and Waters are registered trademarks of Waters Corporation. ACQUITY UPLC, IntelliStart, IonSABRE, MassLynx, T-Wave, UPLC, and ZSpray are trademarks of Waters Corporation.
Other trademarks or registered trademarks are the sole property of their respective owners.
Page 3
Customer comments
Waters’ Technical Communications departme nt invites you to tell us of any errors you encounter in this document or to suggest ide as for otherwise improving it. Please help us better understand what you expect from our documentation so that we can continuously improve its accuracy and usability.
We seriously consider every customer comment we receive. You can reach us at tech_comm@waters.com.
Waters Corporation 34 Maple Street Milford, MA 01757 USA
iii
Page 4
Safety considerations
Some reagents and samples used with Waters® instruments ca n po se chemical, biological, and radiological hazard s. Be sure you are aware of the potentially hazardous effects of all substances you work with. Always observe Good Laboratory Practice (GLP) guidelines, published by the U.S. Food and Drug Administration, and consult your organization’s safety representative for guidance.
When you develop methods, follow the “Pro tocol for the Ado ption of Analyt ical Methods in the Clinical Chemistry Laboratory,” Ame rican Journal of Medical Technology, 44, 1, pages 30–37 (1978). This protocol addresses good operating procedures and the techniques necessary to validate system and method performance.
Instrument-specific safety consideratio ns

Solvent leakage hazard

The source exhaust system is designed to be robust and leak-tight. Waters recommends you perform a hazard analysis, assuming a maximum leak into the laboratory atmosphere of 10% HPLC eluate.
Warning:
• To confirm the integrity of the source exhaust system, the source seals must be renewed at intervals not exceeding one year.
• The source seals can withs tand exposure only to specific solvents, see Appendix C in the Wat ers TQ Det ector Oper ator’ s Guide. If you intend to use solvents other than those listed, you must first determine whether those solvents are compatible with the composition of the seals.
iv
Page 5
Flammable solvents hazard
Warning: Where signif icant quantities of flammable solvents are
involved, the source must receive a continuous flow of nitrogen to prevent possible ignition within that enclos ed space.
Never let the nitrogen supply pressure fall below 400 kPa (4 bar, 58 psi) during analyses that require flammable solvents. Connect to the LC output with a gas-f a il co n n e ctor to stop the LC solvent if th e ni t rogen suppl y fa i l s.
High temperature hazard
Warning: The source enclosure can be hot. To avoid burn injuries,
avoid touching the source enclosure when operating or servicing the TQ Detector.
Waters TQ Detector high temperature hazard:
Source enclosure ass em bly
v
Page 6
High voltage hazard
Warning:
• To avoid electric shock, do not remove the TQ Detector’s protective panels. The components they cover are not user-serviceable.
• To avoid non-lethal electric shock, any equipment connected to the ESI and IonSABRE™ APCI probes must be grounded.
• When the TQ Detector is in Operate mode, certain ext ernal surfaces can conduct high voltages. To avoid non-lethal electric shock, make sure the instrument is in Standby mode before touching areas marked with this high voltage warning symbol.
Safety advisories
Consult Appendix A for a comprehensive list of warning and caution advisories.
vi
Page 7
Operating this device
When operating this device, follow standard quality cont rol procedures and the guidelines presented in this section.
Symbols
Symbol Definition
Intended use
The Waters Tandem Quadrupole (TQ) Detector is intended to be used as a research tool to deliver authentica ted mass measurement in both MS and MS/MS modes.
The Waters TQ Detector can be used for general In Vitro Diagnostic applications, only by professionally trained and qualified laboratory personnel.
E C RE P
Authorized Representative of the European Community
The CE symbol serves as confirmation of the conformity of a product with all European Community directives applicable to that product.
For in vitro diagnostic use.
IV D
The Waters TQ Detector is CE-marked according to the European Union In Vitro Diagnostic Device Directive 98/79/EC.
vii
Page 8
Calibration
To calibrate LC systems, follow acceptable calibration method s using at least five standards to generate a standard curve . The concentration range for standards should cover the entire range of quali ty-control samples, typical specimens, and atypical specimens.
To calibrate mass spectrometers, consult the calibration section of the operator’s guide of the instrument you are calibrating.
Quality control
Routinely run three quality-control samples that represent subnormal, normal, and above-normal levels of a compound. Ensure that quality-control sample results fall within an acceptable range, and evaluate precision from day to day and run to run. Data col lected when qualit y control samples are ou t of range might not be valid. Do not r eport the se dat a until you ar e cert ain that the instrument performs satisfactorily.
When analyzing samples from a co m plex matrix such as soil, tissue, serum/plasma, whole blood, etc., note tha t the mat rix components can adversely affect LC/MS results, enhancing or suppres sing ionization. To minimize these matrix effects, Wate rs recommends you adopt the following measures:
viii
Prior to the instrumental analysis, use appropria te sample pretreatment such as protein precipitation, liquid/liquid extraction (LLE), or solid phase extraction (SPE) to remove matrix interferences.
Whenever possible, verify method accuracy and precision using matrix-matched calibrators and QC samples.
Use one or more internal standard compounds, preferably isotopically-labeled analytes.
Page 9
IVD authorized representative information
IVD authorized representative
Waters Corporation (Micromass UK Limited) is registered in the United Kingdom with the Medicines and Healthcare P roducts Regulatory Agency (MHRA) at market Towers, 1 Nine Elms Lane, London, SW8 5NQ. The referenc e number is IVD000167.
Waters Corp o ra tion (Microm a ss UK Ltd.) Floats Road Wythenshawe Manchester M23 9LZ United Kingdom
Telephone: +44-161-946-2400 Fax: +44-161-946-2480 Contact: Quality manager
ix
Page 10
x
Page 11

Table of Contents

Safety considerations .......................................................................................... iv
Instrument-s pe c if i c sa f et y co n si d erations............. ... ............. .. .......................... iv
Flammable solvents hazard ................................................................................ v
High temperature hazard.................................................................................... v
High voltage haza r d......... .. .......................... ... ............. .. .. ............. ... ................... vi
Safety advisories................................................................................................. vi
Operating this device ......................................................................................... vii
Symbols ............................................................................................................. vii
Intended use...................................................................................................... vii
Calibration ....................................................................................................... viii
Quality control ................................................................................................. viii
IVD authorized representative information ................................................. ix
IVD authorized representative .......................................................................... ix
1 Waters TQ Detector Overview ............................................................ 1-1
Overview ............................................................................................................. 1-2
Waters TQ Dete c to r.......................... ... ............. .. ............. .. .............. .. ............. . 1-2
ACQUITY TQD UPLC/MS system .......... .. .. ................................................... 1-4
MassLynx mass spectrometry software and data system ............................. 1-5
ACQUITY UPLC Console......................................................................... ....... 1-6
Ionization techniques and source probes ................................................... 1-7
Electrospray i on i za tion (ESI)..... .. ............. .. .............. .. ............. .. .............. .. ..... 1-7
Combined electrospray ionization and atmospheric p ressure chemical ionization
(ESCi) ......................................................................................................... 1-7
Atmospheric pressure chemical ionization..................................................... 1-8
Atmospheric pressure photoionization ........................................................... 1-8
Table of Contents xi
Page 12
Ion optics ............................................................................................................. 1-9
MS operating modes ....................................................................................... 1-10
MS/MS operating modes ................................................................................ 1-11
Product (daug h t er) ion mode. .............. .. ............. .. ............. ... ............. .. .......... 1-11
Precursor (pare n t) ion mode...... ............. .. ............. ... ............. .. .. .............. .. ... 1-1 2
Multiple reaction monitoring mode .............................................................. 1-12
Constant neutral loss mode........................................................................... 1-13
Sample inlet ...................................................................................................... 1-14
Vacuum system ................................................................................................ 1-14
Rear panel .. ... ............. .. .. .. ... .. .. ............. ... .. .. .. .. .............. .. .. .. .. ... ............. .. .. ... .. .. . 1-15
IntelliStart fluidics system overview ......................................................... 1-16
IntelliStart fluidics system operation ........................................................ 1-17
Operating the IntelliStart fluidics system from the ACQUITY UPLC
Console . .. ............. .. .............. .. ............. .. .............. .. ............. .. .............. .. .. ... 1-1 7
Operating the IntelliStart fluidics system from the Tune window............. 1-18
Programming the MS method to operate the IntelliStart fluidics system. 1-18
2 Preparing the Waters TQ Detector for Operation ......................... 2-1
Starting the TQ Detector ................................................................................. 2-2
Configuring IntelliStart..................... ............................................... .. ............. 2-6
Verifying the instrument’s state of readiness................................................ 2-6
Tuning and calibration information ............................................................... 2-6
Running the TQ Detector at high flow rates.................................................. 2-7
Monitoring the TQ Detector LEDs.................................................................. 2-8
Preparing the IntelliStart fluidics system .................................................. 2-9
Installing the solvent manifold drip tray........................... .. .......................... 2-9
Installing the reservoir bottles..................... .......................................... ....... 2-10
Diverter valve positions................................................................................. 2-11
Purging the infusion syringe......................................................................... 2-13
Rebooting the TQ Detector ........................................................................... 2-14
Rebooting the TQ Detector by pressing the reset button ............................ 2-14
xii Table of Contents
Page 13
Shutting down the TQ Detector .................................................................. 2-15
Putting the TQ Detector in Standby mode for overnight shutdown........... 2-15
Complete TQ Detector shutdown.................................................................. 2-16
Emergency TQ Detector shutdown............................................................... 2-17
3 ESI and ESCi Modes of Operation ..................................................... 3-1
Introduction ....................................................................................................... 3-2
Installing the ESI probe .................................................................................. 3-2
Required materials .......................................................................................... 3-2
Installing the corona pin ................................................................................. 3-5
Required materials .......................................................................................... 3-5
Optimizing the ESI probe for ESCi operation ........................................... 3-7
Required materials .......................................................................................... 3-7
Removing the corona pin ................................................................................ 3-9
Required materials .......................................................................................... 3-9
Removing the ESI probe ............................................................................... 3-11
Required materials ........................................................................................ 3-11
4 Operating the Waters TQ Detector .................................................... 4-1
Setting-up the instrument .............................................................................. 4-2
Required materials .......................................................................................... 4-2
Performing a sample tune .............................................................................. 4-7
Required materials .......................................................................................... 4-7
Developing experiment methods .................................................................. 4-9
Required materials .......................................................................................... 4-9
Verifying the sys t em us in g Sys t em QC ........... ............. .. ... .. .. .. .. .............. .. . 4-11
A Safety Advisories ..................... ..... ..... ................................................... A-1
Warning symbols ............................................................................................... A-2
Task-specific hazard warnings........................................................................ A-2
Warnings that apply to particular instruments, instrument components, and
sample types............................................................................................... A-3
Table of Contents xiii
Page 14
Caution symbol .................................................................................................. A-5
Warnings that apply to all Waters instruments ......................................... A-5
Electrical and handling symbo l s ............. .. ... .. .. .. ............. ... .. .. .. .. ... ............. . A-13
Electrical symbols.......................................................................................... A-13
Handling symbols .......................................................................................... A-14
xiv Table of Contents
Page 15

1 Waters TQ Detector Overview

This chapter describes the instrument, including its controls and gas and plumbing connections.
Contents: Topic Page
Overview 1-2 Ionization techniques and source probes 1-7 Ion optics 1-9 MS operating modes 1-10 MS/MS operating modes 1-11 Sample inlet 1-14 Vacuum system 1-14 Rear panel 1-15 IntelliStart fluidics system overview 1-16 IntelliStart fluidics system operation 1-17
1-1
Page 16

Overview

Waters TQ Detector

The Waters® TQ Detector is a tandem quadrupole, atmospheric pressure ionization (API) mass spectrometer. Designed for routine UPLC™/MS/MS analyses in quantitative and qualitative applications, it can operate at fast acquisition speeds compatible with ultra performance LC.
Waters provides these ion sources with the instrument as standard equipment:
ZSpray™ (dual orthogonal sampling) interface.
Multi-mode ESCi chemical ionization (APCI) and electrospray ionization (ESI).
Optional ionization modes are IonSABRE™ APCI and APPI (atmospheric pressure photoionization) .
For TQ Detector specifications, see the Waters TQ Detector Site Preparation
Guide.
Waters TQ Detector:
®
ionization switching for atmospheric pressure
1-2 Waters TQ Detector Overview
TP02592
Page 17
Waters TQ Detector with doors open:
HV
PROBE
NEBULIZER
POWER OPERATE
DESOLVATION
V
H
E
B
O
R
P
IntelliStart technology
TP02627
IntelliStart™ technology monitors LC/MS/MS performance and reports when the TQ Detect or is ready for use.
The IntelliStart software automatically tune s and mass calibrates the TQ Detector and displays performance readbacks . In tegrated with MassLynx™ mass spectrometry software and ACQUITY UPLC™ Console software, IntelliStart enables simplified set-up of the system for use in routine analytical and open access applications.
The IntelliStart fluidics system is built into the TQ Detector. It delivers sample directly to the MS probe from the LC column or from two integral reservoirs. The integral reservoirs can also deliver sample through direct or combined infusion so that you can optimize instrument performance at analytical flow rates.
See also: The TQ Detector online Help for further details of IntelliStart.
Overview 1-3
Page 18

ACQUITY TQD UPLC/MS system

Note: The Waters TQ Detector is designed for compatib ility with the
ACQUITY UPLC system; if you are not using an ACQUITY UPLC system, refer to the documentation relevant to the LC syst em being used.
The ACQUITY TQD UPLC/MS system includes an ACQUITY UPLC system and the Waters TQ Detector.
ACQUITY UPLC system
The ACQUITY UPLC system includes a binary solvent manager, sample manager, column heater, optional sample organizer, op tional detectors, and a specialized ACQUITY UPLC column. Waters MassLynx mass spectrometry software controls the system.
See also:
ACQUITY UPLC System Operator’s Guide
Controlling Contamination in LC/MS System s (part number
715001307). You can find this document on http: //www.waters.com; click Services and Support and then Support Center.
1-4 Waters TQ Detector Overview
Page 19
Waters ACQUITY TQD:
Sample organizer (option al)
manager
Solvent tray
Column heater
TP02597
Sample managerBinary solvent
TQ Detector

MassLynx mass spectrometry software and data system

MassLynx is a high-performance mass spect rometry application that acquires, analyzes, manages, and distributes ultra-violet (UV), evaporative light scattering, analog, and mass spectrometry data.
MassLynx software permits these major operations:
Configuring the instrument.
Creating LC inlet and MS/MS methods that define operating parameters for a run.
Using IntelliStart softwar e to tune and m a ss calibrate the TQ Detector.
Running samples.
Monitoring the run.
Acquiring data.
Processing data.
Overview 1-5
Page 20
•Reviewing data.
•Printing data.
See also: MassLynx 4.1 user documentation and online Help for more
information on installing and using MassLynx software.

ACQUITY UPLC Console

The ACQUITY UPLC Console is a software application with which you configure settings, monitor perfor manc e, run diagnostic tests, and maintain the system and its modules. The ACQUITY UPLC Console functions independently of MassLynx and does not recognize or control the data system.
See also: ACQUITY UPLC System console online Help for details of the TQ
Detector.
TQ Detector ACQUITY UPLC Console page:
1-6 Waters TQ Detector Overview
Page 21

Ionization techniques and source probes

Electrospray ionization (ESI)

In electrospray ionization (ESI), a strong electrical charge is given the eluent as it emerges from a nebulizer. The droplets that compose the resultant aerosol undergo a reduction in size (solvent evaporat ion). As solvent continues to evaporate, the charge density incre ases until the dr oplet su rfac es eject io ns (ion evaporation). The ions can be singly or multiply charged. The multiply charged ions are of particular interest because the TQ Detector separates them according to their mass-to-charge ratios (m/z), permitting the detection of high-molecular-weight compounds.
The instrument can accommodate eluent flow rates of up to 1 mL/min.

Combined electrospray ionization and atmospheric pressure chemical ionization (ESCi)

Combined electrospray ionization and atmospheric pressure chemical ionization (ESCi) is supplied as standard equipment on the TQ Detector. In ESCi, the standard ESI probe is used in conjunction with a corona pi n to allow alternating acquisition of ESI and APCI ionization data, facilitating high throughput and wider compound coverage.
Ionization techniques and source probes 1-7
Page 22
ESCi mode:
Corona pin
TP02695
Sample cone tip
See also: “Electrospray ionization (ESI)” on page 1-7.

Atmospheric pressure chemical ionization

A dedicated high performance atmospheric pressure chemical ionization (APCI) probe is offered as an option.
See also: The Waters TQ Detector Operator’s Guide for full details.

Atmospheric pressure photoionization

Atmospheric pressure photoionization (APPI) is offered as an option. It uses photons generated by a krypton-dischar ge ultra-violet (UV) lamp ( 10.2 eV) to produce sample ions from vaporized LC eluent.
1-8 Waters TQ Detector Overview
Page 23

Ion optics

The TQ Detector’s ion optics operate as follows:
Samples from the LC or Intellistart fluidics system are introduced at atmospheric pressure into the ionization source.
The ions pass through the sample cone into the vacuum system.
The ions pass through the transfer optics to the fir st quadrupole where they are filtered according to their mass-to-charge ratio (m/z).
The mass-separated ions pass into the T-Wave™ collision cell where they either undergo collision-induced dissociation (CID) or pass to the second quadrupole. Any fragment ions are then mass-analyzed by the second quadrupole.
The transmitted ions are detected by the photomultiplier detection system.
The signal is amplified, digitized, and sent to the MassLynx mass spectrometry software.
Ion optics overview:
Sample cone
Sample inlet Transfer optics
T-W ave co llision cell
Conversion dynode
Isolation valve
Z-Spray ion sourc e Quadrupole 1
(MS1)
Quadrupole 2 (MS2)
Detector
Ion optics 1-9
Page 24

MS operating modes

The following table shows the MS operating modes.
MS operating modes :
Operating mode MS1 Collision cell MS2
MS1 Resolving
(scanning)
MS2 Pass all masses Resolving
SIR Resolving (static) Pass all masses
The MS1 mode, in which MS1 is used as the mass filter, is the most common and most sensitive method of per formi ng MS analysis. It i s di rectly analogous to using a single-quadrupole mass spectrometer.
The MS2 mode of operation is used, with collision gas p resent, when switching rapidly between MS and MS/MS operat ion (for example, in survey scan mode). It also provides a useful tool for instrument tuning and calibration before MS/MS analysis and for fault diagnosis.
The selected ion recording (SIR) mode of operation is used as a quantitation mode when no suitable fragment ion can be found to perform a more specific multiple reaction monitoring (MRM) analysis.
Pass all masses
(scanning)
1-10 Waters TQ Detector Overview
Page 25

MS/MS operating modes

The following table shows the MS/MS operating modes.
MS/MS operating modes :
Operating mode MS1 Collision cell MS2
Product (daughter) ion
Static (at precursor mass)
spectrum Precursor
Scanning Static (at product (parent) ion spectrum
MRM Static (at
precursor mass) Constant neutral
loss spectrum
Scanning
(synchronized
with MS2)

Product (daughter) ion mode

Product ion mode is the most commonly used MS/MS operating mode. An ion of interest is selected for fragment ation in the collision cell, thus yielding structural information.
Product ion mode:
Pass all masses Scanning
mass)
Static (at product mass)
Scanning (synchronized with MS1)
MS1
Static (at precursor mass)
Collision cell
Pass all masses
Typical applications
Structural elucidation (for example, peptide sequencing)
MS/MS operating modes 1-11
MS2
Scanning
Page 26
Method development for MRM screening studies: – Identifying product ions for use in MRM transitions. – Optimizing CID tuning conditions to maximize the yield of a specific
product ion to be used in MRM analysis.

Precursor (par ent) ion mode

Precursor ion mode:
MS1
Scanning
Typical application
You typically use the precursor ion mode for structural elucidation–that is, to complement or confirm product scan data–by scanning for all the precursors of a common product ion.

Multiple reaction monitoring mode

MRM mode is a highly selectiv e MS/MS equivalent of SIR. As both MS1 and MS2 are static, greater dwell time on the ions of interest is allowed, and therefore better sensitivity compared to scanning MS/MS. This is the most commonly used acquisition mode for quantitative analysis, allowing the compound of interest to be isolated from the chemical background noise.
Collision cell
Pass all masses
MS2
Static (at product mass)
1-12 Waters TQ Detector Overview
Page 27
Multiple reaction monitoring mode:
MS1
Static (at precursor mass)
Typical application
You typically use MRM to quantify known analytes in complex samples:
Drug metabolite and pharmacokinetic studies.
Environmental, for example, pest icide and herbicide analysis.
Forensic or toxicology, for example, screening for target drugs in sport.
MRM does not produce a spectrum because only one transition is monitore d at a time. As in SI R , a c h romatogram is produce d.

Constant neutral loss mode

Constant neutral loss mode detects the loss of a specific neutral fragment or functional group from an unspecified precur sor or precursors.
The scans of MS1 and MS2 are synchronized. When MS1 transmits a specific precursor ion, MS2 “lo oks” to s ee whether that p recu rsor lo ses a fra gment of a certain mass. If it does, the loss registers at the detector.
In constant neutral loss mode, the spectrum show s the masses of all precursors that actually lost a fragment of a certain mass .
Collision cell
Pass all masses
MS2
Static (at product mass)
MS/MS operating modes 1-13
Page 28
Constant neutral lo ss mode:
Typical application
You typically use constant neutral loss mode to screen mixtures for a specific class of compound that is character ized by a common f ragmentat ion pathway, indicating the presence of compounds containing a common functional group.

Sample inlet

Either of two methods delivers solvent and sample to the installed probe:
An LC system, which delivers the eluent from an LC analysis.
IntelliStart fluidics system, which uses on-board solutions to automate instrument optimization. You can de liver solutions by direct or combined infusion.

Vacuum system

MS1
Scanning
(synchronized with MS2)
Collision cell
Pass all masses
MS2
Scanning
(synchronized with MS1)
An external roughing (rotary vane) pump and an internal split- flow turbomolecular pump combine to create the source vacuum. The turbomolecular pump evacuates the analyzer and ion transfer region.
Vacuum leaks and electrical or vacuum pump failures cause vacuum loss, which protective interlocks guard against. The system monitors turbomolecular pump speed and continuously measures vacuum pressure with a built-in Pirani gauge. The gauge also serves as a switch, discontinuing operation when it senses vacuum loss.
A vacuum isolation valve isolates the source from the mass analyzer, allowing routine source maintenance without venting.
1-14 Waters TQ Detector Overview
Page 29

Rear panel

The following figure shows the rear panel locations of the connectors used to operate the TQ Detector with external devices.
TQ Detector rear panel:
Power cord
Nitrogen inlet
Source vent
Turbo vacuum Collision cell
gas inlet
V~ 200-240 Hz 50-60 VA 1200
APIGAS
6.9BAR MAXIMUM
SOURCEVENT
Waters Corporation
34 Maple Street Milford, MA01757 U.S.A.
VACUUM
COLLISION
GAS
1.0BAR
MAXIMUM
VACUUM
Serial Number
ACN 065444751
+
+
1
1
InjectStart
In
Analog
Out
2
2
-
-
Ground
3
3
Ground
+
+
4
4
In
Event
NotUsed
-
-
5
5
+
6
6
StopFlow
Switch3
Out
7
7
-
Ground
8
8
Ground
+
9
9
Out Out
Switch4
Switch4
10
10
-
IVD
RS232
ETHERNET
Event inputs and outputs
Shielded Ethernet
PUMP
Roughing pump relay switch
Source vacuum
Rear panel 1-15
Page 30

IntelliStart fluidics system overview

The IntelliStart fluidics system is built into the TQ Detector. The system delivers sample directly to the MS probe in one of two ways:
From the LC column
From two integral reservoirs.
Tip: The integral reservoirs can also deliver sample through direct or
combined infusion to enable optimization at analytical flow rates.
The IntelliStart system incorporates a multi-position valve with these attributes:
An input connection from an external LC column.
An input connection from the TQ Detector’s infusion syringe. (The TQ Detector’s infusion syringe is also connected to two reservoirs, A and B. In the software, you specify which reservoir to draw from.)
An output connection to the TQ Detector’s probe.
An output connection to a waste line.
LC flow:
Probe
LC
Column
Syringe
LC Waste
1-16 Waters TQ Detector Overview
Idle
B
Off
A
Off
Reservoir A Reservoir B
Page 31

IntelliStart fluidics system operation

Control of solvent and sample delivery during auto-tuning, auto-calibration, and method development is automatically performed by the sof tware.
IntelliStart configuration requirements can be set in the system console. You can edit the parameters, frequency, and extent of the automation you want IntelliStart to perform.
See also: The TQ Detector online Help for further details of IntelliStart.

Operating the IntelliS tart fluidics system f rom the ACQUITY UPLC Console

To operate the IntelliStart fluidi cs system from the Instrument Console:
1. In the MassLynx window, click MS Console.
2. In the ACQUITY UPLC Console system tree, expand TQ Detector.
3. Click Interactive Display.
4. In the ACQUITY UPLC Console window, click the current flow rate.
5. In the Select Flow Rate dialog box, enter a new flow rate.
IntelliStart fluidics system operation 1-17
Page 32
6. Click OK.
7. Click Reservoir A or B.
8. In the Select Reservoir dialog box, click A or B.
9. Click OK.
Tip: If the selected reservoir is different from the current reservoir, the
system purges as the reservoir changes.
10. Click the current diverter valve position label.
11. In the Select a Flow State dialog box, select the required flow state.
12. Click OK.
13. Click to start.
Tip: A status indicates the amount of fluid in the syringe and the
amount of time remaining before the fluid empties. When the syringe is empty the system becomes idle.
14. Click to refill the syringe or to purge the system.
15. Click to stop the current action.

Operating the IntelliStart fluidics system from the Tune window

To operate the IntelliStart fluidics system from the Tune window:
1. In the MassLynx window, click MS Tune.
2. In the TQ Detector Tune window, click the Fluidics page.
3. Set the Flow Control parameters according to the instructions in the MassLynx online Help.

Programming the MS method to operate the IntelliStart fluidics system

In the MS method, you can program the operation of the system’s multi-position valve to infuse sample during a run. The valve can also divert LC flow to waste as a timed event.
1-18 Waters TQ Detector Overview
Page 33
To program the MS method:
1. In the MassLynx window, click MS Method.
2. In the MS Methods window, click Options > Method Events.
3. In the Method events dialog box, select the appropriate flow state, as shown in the following table.
Flow states :
State LC Syringe
LC TQ probe Waste Combined TQ probe TQ probe Infusion Waste TQ Probe Waste Waste Waste No change No change No change
Tip: At instrument power-up, the LC state is waste. See also: MassLynx online Help topic “Advanced Methods and Events ”.
IntelliStart fluidics system operation 1-19
Page 34
1-20 Waters TQ Detector Overview
Page 35
2 Preparing the Waters TQ
Detector for Operation
This chapter describes how to start and shut-down the TQ Detector.
Contents: Topic Page
Starting the TQ Detector 2-2 Preparing the IntelliStart fluidics system 2-9 Rebooting the TQ Detector 2-14 Shutting down the TQ D etector 2-15
2-1
Page 36

Starting the TQ Detector

Note: The Waters TQ Detector is designed for compatib ility with the
ACQUITY UPLC system; if you are not using an ACQUITY UPLC system, refer to the documentation relevant to the system being used.
Caution: Using incompatible solvents can cause severe damage to the
instrument.
• Refer to Appendix C of the Waters TQ Detector Operator’s Guide for TQ Detector solvent information.
• Refer to Appendix C of the ACQUITY UPLC System Operator’s Guide for more information on solvent compatibility with ACQUITY.
Starting the TQ Detector entails powering-on the ACQUITY workstation, logging into the workstation, powering-on the TQ Detector and all the other ACQUITY instruments, and starting the MassLynx software.
Requirement: You must power-on and log in to the ACQUITY workstation
first to ensure that it obtains the IP addresses of the system instrum e nts.
To start the TQ Detector:
Warning: During analyses that require flammable solvents, to avoid
ignition of the solvents, never let the nitrogen supply pressure fall below 400 kPa (4 bar, 58 psi).
1. Ensure the n i tr o g e n sup p l y i s connected to th e instrumen t’ s AP I g a s connection.
Note: The nitrogen must be dry and oil-free, with a purity of at least
95%. Regulate the supply at 600 to 690 kPa (6.0 to 6.9 bar, 90 to 100 psi).
See also: Figure titled “TQ Detector rear panel” on page 1-15.
2. Ensure that the collision gas supply is connecte d to the TQ Detector.
Requirement: The collision gas is argon; it must be dry and of high
purity (99.9%). Regulate the supply at 50 kPa (0.5 bar, 7 psi).
3. Power-on the ACQUITY UPLC system workstation, and log in before powering-on the other instruments.
2-2 Preparing the Waters TQ Detector for Operation
Page 37
4. Press the power switch on the top, lef t-hand s ide of the TQ Dete ctor and ACQUITY instruments. Each system instrument “beeps” and runs a series of startup tests.
5. Allow 3 minutes for the embedded PC to initialize. An audible alert sounds when the PC is ready.
The power and status LEDs change as follows:
Each system instrument’s power LED shows green.
During initialization, the binary solvent manager’s and sample
manager’s status LED flashes green.
After the instruments are successfully powered-on, all power LEDs
show steady green. The binary solvent manager’s flow LED, the sample manager’s run LED, and the TQ Detector’s operate LED remain off.
6. Start MassLynx. You can monitor the ACQUITY console for messages and LED indications.
7. In the MassLynx main window’s lower-left corner, click IntelliSt art.
Result: TQ Detector console IntelliStart appears . The TQ Detector is in
Standby mode.
Starting the TQ Detector 2-3
Page 38
8. Click Control > Pump to start the roughing pump. The operate LED remains off.
Tip: There is a 20-second delay, during which th e turbop ump is star ting,
before the roughing pump starts. IntelliStart displays “Instrument in standby”.
2-4 Preparing the Waters TQ Detector for Operation
Page 39
9. Click Resolve or Operate to put the TQ Detector into Operate mode. When the TQ Detector is in good operating conditi on, IntelliStart displays “Ready”.
Tip: Clicking Resolve should prepare the system for operation,
putting the TQ Detector into Operate mode. If clic k i ng Resolve fails to put the instrument into Operate mode, IntelliStart displays corrective actions.
Starting the TQ Detector 2-5
Page 40

Configuring IntelliStart

To configure IntelliStart:
1. In the ACQUITY UPLC Console system tree, click TQ Detector.
2. Click Configure > IntelliStart Configuration.
3. In the IntelliStart Configuration dialog box, in the Checks list, select the checkboxes for the items you want checked during TQ Detector startup. Clear the checkboxes of items you do not want checked.
Tip: To display detailed information for an item, highlight it and click
Properties.
4. Click Apply > OK.

Verifying the instrument’s state of readiness

When the TQ Detector is in good operating condition, the power LED shows constant green and the operate LED is off. You can view any error messages in IntelliStart.
To access IntelliStart:
1. In the ACQUITY UPLC Console system tree, select TQ Detector.
2. In the TQ Detector information window, click IntelliStart.

Tuning and calibration information

The TQ Detector must be tuned and calibrated prior to use, tasks normally performed from IntelliStart.
See also: The online Help topic “Instrument Setup” and Chapter 4,
“Operati n g th e Wa ters TQ Detec tor”.
2-6 Preparing the Waters TQ Detector for Operation
Page 41

Running the TQ Detector at high flow rates

ACQUITY UPLC is run at high flow rates. To optimize desolvation, and thus sensitivity, the ACQUITY TQD system should be run at appropriate gas flows and desolvation temperatures. IntelliStart automatically sets these when you enter a flow rate, according to the following table.
Flow rate versus temperature and gas flow :
Flow rate (mL/min)
0.000 to 0.100 120 250 500
0.101 to 0.300 120 350 600
0.301 to 0.500 150 400 800 >0.500 150 450 900
Note: Under low ambient temperature, high moisture, and high flow rate
conditions, condensation can occur in the instrument’s source.
Source temp (°C)
Desolvation temp (°C)
Desolvation gas flow (L/hr)
Starting the TQ Detector 2-7
Page 42

Monitoring the TQ Detector LEDs

Light-emitting diodes on the TQ Detector indicate its operational status.
Power LED
The power LED, to the top, left-hand side of the TQ Detector’s front panel, indicates when the T Q De te c tor is po we r e d- on o r po we r e d- off.
Operate LED
The operate LED, to the right of the power LED, indicates the operating condition.
Operate LED indications :
LED mode and color Description
Off The instrument is in Standby mode with high
voltages and heaters inoperative.
Constant green The instrument is in Operate mode with heaters
and high voltages operating.
Flashing green Indicates the system is waiting for an instrument
component to reach operational conditio ns. The LED could, for example, flash while the system reaches a programmed temperature set point or vacuum level. The instrument is in work ing order and Operate mode but not yet ready to acquire.
Flashing red Indicates that the sys tem stopped due to error
and is no longer processing any samples or other maintenance requests. Information r egarding the error appears in the data system's contro l panel or the ACQUITY UPLC Console software.
Constant red for 3 minutes
Constant red for more than 3 minutes
2-8 Preparing the Waters TQ Detector for Operation
The instrument is waiting for the instrument to initialize.
Indicates a serious system failure that prevents further operation. Power-off the instrument, and then power-on. If the LED is still steady red for more than 3 minutes, contact your Waters service r epresentative.
Page 43

Preparing the IntelliStart fluidics system

See also: Appendix B of the Waters TQ Detector Operator’s Guide .

Installing the solvent manifold drip tray

Required materials
Chemical-resistant, powder-free gloves
To install the solvent manifold drip tray:
Warning: The solvent ma nifold drip tray can be contaminated
with biohazardous and/or toxic materials. Always wear chemical-resistant, powder-free gloves while performing this procedure.
Install the solvent manifold drip tray as shown below:
TP02685
Preparing the IntelliStart fluidics system 2-9
Page 44

Installing the reservoir bottles

Required materials
Chemical-resistant, powder-free gloves
To install the reservoir bottles:
Warning: The reservoir bottles can be contaminated with
biohazardous and/or toxic materials. Always wear chemical-resistant, powder-free gloves while performing this procedure.
1. Remove the reservoir bottle caps.
2. Screw the reservoir bottles onto the TQ Detector as shown below.
TP02630
2-10 Preparing the Waters TQ Detector for Operation
Page 45

Diverter valve positions

Column and syringe in home position after power-up
After power-up, the flow path between the column and waste is open. The syringe is empty, and the flow path between it and waste is open.
Waste
Column
Probe
Idle
Syringe
B
Off
A
Off
LC Waste
Reservoir A Reservoir B
LC position
In the LC position, the flow path between the LC and probe is open, and the flow path between the syringe and waste is also open.
Probe
B
LC
Column
LC Waste
Idle
Syringe
Off
A
Off
Reservoir A Reservoir B
Preparing the IntelliStart fluidics system 2-11
Page 46
Infusion position in infusion mode
Probe
Infusion
Column
LC Waste
Infusion
Syringe
Reservoir A Reservoir B
B
Off
A
On
Combined position with LC flow and syringe in idle mode
Probe
Combined
Idle
B
Off
A
Off
Column
Syringe
LC Waste
2-12 Preparing the Waters TQ Detector for Operation
Reservoir A Reservoir B
Page 47
Waste position
In the waste position, both the LC flow and the infusion syringe flow are directed to waste. The syringe mode can be only static or dispensing (that is, never drawing).
Probe
Waste
Idle
Column
Syringe
LC Waste

Purging the infusion syringe

Whenever you replace a solution bott le, you sho uld purge t he inf usi on syringe with the solution that you are going to use next.
Tip: Depending on the solutions used, the IntelliStart fluidics system can
require more then one purge cycle to minimize carryover.
To purge the infusion syringe:
1. In the ACQUITY UPLC Console system tree, expand TQ Detector.
2. Click Interactive Display.
B
Off
A
Off
Reservoir A Reservoir B
3. Select the re q u ir e d so l u ti o n reservoir.
4. Click to purge the system.
Tip: System purge takes approximately 2 minutes.
Preparing the IntelliStart fluidics system 2-13
Page 48

Rebooting the TQ Detector

Sometimes you might need to reboot the TQ Detector:
When the Tune window fails to respond.
When MassLynx fails to initialize.
Immediately following software download.

Rebooting the TQ Detector by pressing the reset button

Tip: The reset button shuts down the electronics momentarily and causes the
TQ Detector to reboot.
To reboot the TQ Detector by pressing the reset button:
1. Open the TQ Detector’s front left door.
2. Press the red reset button on the top, left-hand side of the instrument.
TP02687
2-14 Preparing the Waters TQ Detector for Operation
Reset button
Page 49

Shutting down the TQ Detector

Recommendation: You should always leave the TQ Detect or in Operate mode
except when performing routine maintenance. It is not necessary to switch to Standby mode. However, shutting down the TQ Detector is acceptable provided that you consider warm-up time on restarting. If you do shut dow n the TQ Detector, refer to the instructions in this section.
Caution: Buffers left in the system can precipitate and damage
instrument components.
Tip: Set system shutdown parameters in the Shutdown Editor. Consult the
MassLynx online Help for more information.

Putting the TQ Detector in Standby mode for overnight shutdown

You might want to shut down the instrument for a relatively brie f pe riod, like overnight, while maintaining the LC flow.
To put the TQ Detector in Standby mode:
1. In the ACQUITY UPLC Console, click to stop the LC flow or, if column flow is required, divert the LC flow to waste as follows:
a. In the ACQUITY UPLC Console system tree, expand TQ Detector. b. Click Interactive Display. c. Click the current diverter valve position label. d. In the Select a Flow State dialog box, select Waste. e. Click OK.
2. Click Standby to shut off heaters and voltages.
Tip: You can create a method to stop the gas flow or lower the
temperature. See the MassLynx online Help for more infor mation on creating me t h ods.
Shutting down the TQ Detector 2-15
Page 50

Complete TQ Detector shutdown

To completely shut down the TQ Detector:
1. Put the TQ Detector in Standby mode.
See also: “Putting the TQ Detector in Standby mode for overnight
shutdown” on page 2-15.
2. In the ACQUITY UPLC Console, click API Gas .
3. Click Col Gas .
4. Click Control > Vent.
Result: The turbomolecular pump is switched off. When the
turbomolecular pump runs down to half its normal operating speed, the vent valve opens and the instrument is automatically vented. The operate LED changes from green to red and then turns off.
5. Exit MassLynx.
6. After the roughing pump shuts off, operate the power button (on the front of the instrument) to power-off the TQ Detec tor.
Warning: The TQ Detecto r’s power switch does not isolate the
instrument from the main power supply. To isolate the instrument, you must disconnect the power cable from the back of the instrument.
7. Disconnect the power cable from the back of the TQ Detector.
8. Power-off all other instruments and the workstation.
Note: The fans inside some instruments run continuously, even after
you power-off the instruments.
2-16 Preparing the Waters TQ Detector for Operation
Page 51

Emergency TQ Detector shutdown

To shut down the TQ Detector in an emergency:
Warning: The TQ Detecto r’s power switch does not isolate the
instrument from the main power supply. To isolate the instrument, you must disconnect the power cable from the back of the instrument.
Caution: Data can be lost during an emergency shutdown.
1. Operate the power button on the front of the TQ Detector .
2. Disconnect the power cable from the back of the TQ Detector.
Shutting down the TQ Detector 2-17
Page 52
2-18 Preparing the Waters TQ Detector for Operation
Page 53
3 ESI and ESCi Modes of
Operation
This chapter describes how to prepare the TQ Detector for the following modes of operation:
ESI (electrospray ionization)
ESCi (combined electrospay and atmospheric pres sur e chemical
ionization)
If your system uses APCI mode, refer to Chapter 6 of the Waters TQ Detector Operator’s Guid e.
Contents: Topic Page
Introduction 3-2 Installing the ESI probe 3-2 Installing the corona pin 3-5 Optimizing the ESI probe for ESCi operation 3-7 Removing the corona pin 3-9 Removing the ESI probe 3-11
3-1
Page 54

Introduction

The ESI and ESCi ionization mode options use th e standar d ESI probe t hat i s fitted to the instrument when it is shipped from the factory. For ESCi operation, the corona pin is used in conjunction with the ESI pr obe. The following sections explain how to install and remove t he ESI probe and cor ona pin.
See also:
“Electro sp ra y i o n iza t i o n (E SI)” on page 1-7.
Table titled “Combined electrospray ionization and atmospheric
pressure chemical ionization (ESCi)” on page 1-7.

Installing the ESI probe

Required materials

Chemical-resistant, powder-free gloves
To install the ESI probe:
Warning: The ACQUITY UPLC system connections, ESI probe,
and source can be contaminated with biohazardous and/or toxic materials. Always wear chemical-resistan t, powder-free gloves while performing this procedure.
Warning: To avoid electric shock, ensure that the instrument is
suitably prepared before commencing this pr ocedure.
1. Prepare the instrument for working on the source.
See also: Chapter 5 of the Waters TQ Detector Operator’s Guide.
Warning: The source can be hot. To avoid burn injuries, take
great care while working with the instrument’s access door open.
2. Open the instrument’s access door.
3-2 ESI and ESCi Modes of Operation
Page 55
Warning: The ESI probe tip is sharp. To avoid puncture wounds,
handle the ESI probe with care.
3. Remove the protective sleeve, if fitted, from the ESI probe tip.
4. Ensure that the contacts o n the ESI probe align with the probe adjuster assembly contacts, and carefully slid e the ESI probe into the hole in the probe adjuster assembly.
ESI probe
Probe adjuster assembly
Probe adjuster assembly contacts
TP02632
5. Secure the ESI probe by tightening the two thumbscrews.
See also: Figure titled “ESI probe mounted on the source enclosure,
showing the connections to the front panel” on page 3-4.
Installing the ESI probe 3-3
Page 56
ESI probe mounted on the source enclosu re, showing th e connections to the front panel:
Vernier prob e adju ste r
Nebulizer gas connection
Desolvation gas connection
ESI probe electrical lead
Probe adjuster assembly electrical lead
Probe adjuster assembly
Thumbscrew
Diverter valve
ESI probe
6. Connect the ESI probe’s PTFE tube to the nebulizer gas connection.
7. Ensure that the probe adjuste r asse mbly’s e lectric al lead is con nected t o the instrument’s probe connection.
8. Connect the ESI probe’s electrical lead to the instrument’s HV connection.
Warning: To avoid electric shock, do not use stainless steel
tubing to connect the diverter valve to the ESI probe; use the PEEK™ tubing supplied with the instrument.
9. Connect the diverter valve to the ESI probe; use tubing of the appropriate internal diameter (ID).
Tip: Two tubes of different ID are supplied with the instrume nt. Requirement: If you are replacing the tubing suppl ied with the
instrument, you must minimize the length of the tube connecting the diverter valve to the ESI probe. This minimizes delays and dispersion.
10. Close the instrument’s access door.
3-4 ESI and ESCi Modes of Operation
Page 57

Installing the corona pin

Required materials

Chemical-resistant, powder-free gloves
Needle-nose pliers
80:20 acetonitrile/water
To install the corona pin:
Warning: The ACQUITY UPLC system connections, ESI probe,
and source can be contaminated with biohazardous and/or toxic materials. Always wear chemical-resistan t, powder-free gloves while performing this procedure.
Warning: To avoid electric shock, ensure that the instrument is in
Standby mode when commencing this procedure.
1. In the ACQUITY UPLC Console, click Stand by , and confirm that the Operate indicator is not illuminated.
Warning: The source can be hot. To avoid burn injuries, take
great care while working with the instrument’s access door open.
2. Open the instrument’s access door.
Warning: The probe tip is sharp. To avoid puncture wounds , take
great care while working with the source enclosure door open if an ESI probe is fitted.
Caution: Do not apply any downward force to the sourc e
enclosure door while the door is open.
3. Unlatch the source enclosure door’s handle and open the door.
Installing the corona pin 3-5
Page 58
4. Use the needle-nose pliers to remove the blanking plug from the corona pin mounting contact.
Tip: Store the blanking plug in a safe location.
Corona pin mounting contact:
Corona pin mounting contact blanking plug
TP02660
Warning: The corona pin tip is sharp. To avoid puncture wounds,
handle the corona pin with care.
Caution: To avoid damaging to the corona pin’s tip and bending
the pin, use the needle nose pliers to grip the corona pin at the end that fits into the mounting contact.
5. Use the needle-nose pliers to fit the corona pin to the mounting contact.
Requirement: Ensure that the corona pin is securely mounted and that
its tip aligns with the sample cone orifice.
3-6 ESI and ESCi Modes of Operation
Page 59
Corona pin:
ESI probe tip
Corona pin
Sample cone tip
TP02695
6. Use the vernier probe adjuster to position the ESI probe tip so that it is pointing approximately mid-way between the tips of the sample cone and corona pin.
See also: Figure titled “ESI probe mounted on the source enclosure,
showing the connections to the front panel” on page 3-4.
7. Close the source enclosure door and fasten the handle.
8. Close the instrument’s access door.

Optimizing the ESI probe for ESCi operation

Required materials

80:20 acetonitrile/water
To optimize the ESI probe for ESCi operation:
1. In the MassLynx window, click MS Tune.
2. In the Tune window, click Setup > Inter-scan Setup.
3. In the Inter-scan Setup dialog box, click Reset to Defaults.
Optimizing the ESI probe for ESCi operation 3-7
Page 60
4. Click OK.
5. In the MassLynx Tune window, click Ion Mode > ESCi+.
6. Select box numbers 1 and 2, clear box numbers 3 and 4 (above the peak display).
7. In row 1, set Ion Mode to ES.
8. In row 2, set Ion Mode to APcI.
9. In each row, set Mass to 42 and Span to 5.
10. Start an infusion of 80:20 acetonitrile/water.
11. Use the vernier probe adjuster to ensure that the ESI pr obe tip is pointing approximately mid-way between the tips of the sample cone and corona pin.
12. On the Tune window, observe the 42 Da peak in the ES+ and APcI+ peak displays, and increase the values of Capillary (kV) and Corona [(µA) in the current mode or kV in the voltage mode] to prod uce the mo st intense ESI+ and APcI+ signal.
13. Use the vernier probe adjuster to gradually move the probe bi-directionally to determine the best position for both the ESI+ and APcI+ signals.
Tip: You can find a position between the two modes that yields a
relatively optimized signal.
14. To determine whether you hav e discrete ionization in the ESI or APcI mode, set the Capillary parameter to 0 kV and observe that little or no signal remains in ESI mode. Then set the Corona parameter to 0 µA or 0 kV , and observe that little or no signal remains in APcI mode.
Result: The ESI probe is now optimized for ESCi mode. Tip: If necessary, repeat the above procedure using the analyte of
interest, since ionization pot entials may vary with different samples.
3-8 ESI and ESCi Modes of Operation
Page 61

Removing the corona pin

Required materials

Chemical-resistant, powder-free gloves
Needle-nose pliers
To remove the corona pin:
Warning: The ACQUITY UPLC system connections, corona pin,
ESI probe, and source can be contaminated with biohazardous and/or toxic materials. Always wear chemical-resistant, powder-free gloves while performing this procedure.
Warning: To avoid electric shock, ensure that the instrument is in
Standby mode when commencing this procedure.
1. In the ACQUITY UPLC Console, click Stand by , and confirm that the Operate indicator is not illuminated.
Warning: The source can be hot. To avoid burn injuries, take
great care while working with the instrument’s access door open.
2.
Open the instrument’s access door.
Warning: The probe tip is sharp. To avoid puncture wounds , take
great care while working with the source enclosure door open if an ESI probe is fitted.
Caution: Do not apply any downward force to the sourc e
enclosure door while the door is open.
3. Unlatch the source enclosure door’s handle and open the door.
Removing the corona pin 3-9
Page 62
Warning: The corona pin tip is sharp. To avoid puncture wounds,
handle the corona pin with care.
Caution: To avoid damaging to the corona pin’s tip and bending
the pin, use the needle nose pliers to grip the corona pin at the end that fits into the mounting contact.
4. Use the needle-nose pliers to remove the corona pin from its mounting contact.
Tip: Store the corona pin in a safe location. See also: Figure titled “Corona pin:” on page 3-7.
5. Use the needle-nose pliers to fit the blanking plug to the corona pin mounting contact.
See also: Figure titled “Corona pin mounting contact:” on page 3-6.
6. Close the source enclosure door and fasten the handle.
7. Close the instrument’s access door.
3-10 ESI and ESCi Modes of Operation
Page 63

Removing the ESI probe

Required materials

Chemical-resistant, powder-free gloves
To remove the ESI probe:
Warning: The ACQUITY UPLC system connections, ESI probe,
and source can be contaminated with biohazardous and/or toxic materials. Always wear chemical-resistan t, powder-free gloves while performing this procedure.
Warning: To avoid electric shock, ensure that the instrument is
suitably prepared before commencing this pr ocedure.
1. Prepare the instrument for working on the source.
See also: Chapter 5 of the Waters TQ Detector Operator’s Guide.
Warning: The ESI probe and source can be hot. To avoid burn
injuries, ta ke g re a t care while wo rking with th e i n st rument’s access door open.
2. Open the instrument’s access door.
3. Disconnect the diverter valve tubing from the ESI probe.
4. Disconnect the ESI probe’s electrical lead from the high voltage connection.
5. Disconnect the ESI probe’s PTFE tube from the nebulizer gas connection.
6. Undo the two thumbsc rews s ecur ing the ESI pro be to the probe adjust er assembly.
Warning: The ESI probe tip is sharp. To avoid puncture wounds,
handle the probe with care.
7. Carefully remove the ESI probe from the probe adjuster assembly.
Removing the ESI probe 3-11
Page 64
8. If available, fit the protective sl eeve to the ESI probe tip.
9. Close the instrument’s access door.
3-12 ESI and ESCi Modes of Operation
Page 65
4 Operating the Waters TQ
Detector
This chapter is an introduction to operating your TQ Detector; it explains these tasks:
Setting-up your TQ Detector.
Performing a sample tune.
Developing instrument methods.
Verifying the system.
Contents: Topic Page
Setting-up the instrument 4-2 Performing a sample tune 4-7 Developing experiment me thods 4-9 Verifying the system using System QC 4-11
4-1
Page 66

Setting-up the instrument

The IntelliStart instrument setup calibrates the instrument and then, by default, performs a sample tune. If calibration is unneces sary, you can perform only a sample tune.
See also: “Performing a sample tune” on page 4-7. Tip: Instrument setup need only be performed every three to six months,
depending on your usage requirements. In the following example, sodium cesium iodide is used as the calibrant
solution and sulfadimethoxine the tune sample.
Tip: You can substitute solutions suitable for your requirements. See also: The TQ Detector online Help for further details of IntelliStart.

Required materials

Sodium cesium iodide solution (2 ng/µL)
Sulfadimethoxine solution (100 pg/µL)
To prepare the IntelliStart fluidics system: See also: “IntelliStart fluidics syst em operation” on page 1-17.
1. Ensure that IntelliStart fluidics system’s reservoir A is filled with sodium cesium iodide solution.
2. Ensure that reservoir B is filled with sulfadimethoxine solution.
Requirement: Ensure that there is enough soluti on in each r eservoir f or
approximately 5 minutes of operation during the set-up procedure.
Recommendation: In general, place calibrant solution in reservoir A
and tune sample in reservoir B.
3. In the MassLynx window, click MS Console.
4. In the ACQUITY UPLC Console system tree, expand TQ Detector.
5. Click Interactive Display.
6. In the ACQUITY UPLC Console window, click the current flow rate.
7. In the Select Flow Rate dialog box, enter a 50 µL/min flow rate.
4-2 Operating the Waters TQ Detec tor
Page 67
8. Click OK.
9. Click to purge the system.
Tip: System purge takes approximately 2 minutes.
10. Repeat step 9.
Rationale: Purging the IntelliStart fluidics system twice ensures that
any traces of previously used solutions are removed from the system.
11. Click on the currently-selected reservoir display.
12. In the Select Reservoir dialog box, select the reservoir not currently selected.
13. Click OK.
Result: The other reservoir is selected, and the system is purged.
14. When the purge finishes, click to purge the sys tem again.
To specify the instrument set-up parameters:
1. In the ACQUITY UPLC Console system tree, click TQ Detector.
2. Ensure that Ion Mode is ES+.
3. In the ACQUITY UPLC Console system tree, click IntelliStart.
4. Ensure that Instrument Setup is selected.
5. If you require system pre-checking, click Pre-checks.
Rationale: If Pre-checks is selected, when IntelliStart starts the
instrument set-up, it dete rmines whethe r t he existi ng cali brati on is stil l valid. If so, it does not perform a full calibration but proc eeds to sample tuning the instrument.
See also: TQ Detector online Help topic “IntelliStart flow diagram”.
6. Click Start to open the IntelliStart Setup Paramete rs dialog box.
7. In the Instrument Setup tab’s Reference drop-d own li st , click Naics.
Rationale: Naics is the calibration reference file for sodium ces i um
iodide when working in ES+ ion m ode.
Setting-up the instrument 4-3
Page 68
8. Click Fill from reference file.
Result: The default mass values appear in the In strument Tune Mas ses
text boxes.
Rule: You must click Fill from reference file every time you select a new
calibration reference file.
Instrument Setup Parameters dialog box Instrument Setup tab with default parame ters:
Tip: You can use alternative reference solutions to calibrate at higher
masses.
9. The tune and calibration results are written to the files in the Instrument Tune and Calibration text boxes; you can use the default files, enter new names, or Browse for files.
4-4 Operating the Waters TQ Detec tor
Page 69
To specify the sample tune parameters:
1. In the IntelliStart Setup Parameters dialog box, click the Sample Tune tab.
2. Click Copy Instrument Setup masses.
Rationale: The Tune Masses from the Instrument Setup tab are copied
into the Sample Tune Masses boxes.
Instrument Setup Parameters dialog box Sample Tune tab:
3. In the Tune text box, enter sulfadimethoxine.ipr.
Rationale: The sample tune results are written to this file.
Setting-up the instrument 4-5
Page 70
Instrument Setup Parameters dialog box Sample Tune tab with example sample tune parameters:
To start instrument set-up:
1. Click Start .
Result: A message appears reminding you to ensure that the calibrant
solutions, calibration parameters and LC flow are set correctly.
2. Click OK.
Result: Instrument set-up starts. An autotune on the calibrant is
followed by automatic calibration. The ACQUITY UPLC Console displays the progress of the setup.
4-6 Operating the Waters TQ Detec tor
Page 71
Example ACQUITY UPLC Console display during calibration:
IntelliStart creates tune and calibration files , which are saved as specified on the Instrument Setup Parameters dialog box’s Instrument Setup tab. Once calibration is complete, the sample tune starts on the four masses defined in the Intell iStart Setup Parameters dialog box. When the sample tune is complete, the results are written to the Sulfadimethoxine.ipr file.

Performing a sample tune

In the following example, sulfadimethoxine is used as the tuning sample.

Required materials

Sulfadimethoxine solution (100 pg/µL)
To perform a sample tune:
1. Prepare the IntelliStart fluidics system with sulfadimethoxine solution in reservoir B.
See also: “To prepare the IntelliStart fluidics system:” on page 4-2.
Performi ng a sample tune 4-7
Page 72
2. In the ACQUITY UPLC Console system tree, click TQ Detector.
3. Ensure that the Ion Mode is ES+.
4. In the ACQUITY UPLC Console system tree, click IntelliStart.
5. Clear the Instrument Setup check box.
6. If you require system pre-checking, click Pre-checks.
Rationale: If Pre-checks is selected, when IntelliStart starts the sampl e
tune, it determines whether the existing tune is still valid. If so, no sample tune is performed.
See also: TQ Detector online Help topic “IntelliStart flow diagram”.
7. Click Start .
8. In the IntelliStart Setup Parameters dialog box’s Sample Tune tab, specify the sample tune parameters as described in “To specify the
sample tune parameters:” on page 4-5.
9. Click Start .
Result: A message appears reminding you to ensure that the tune
solutions, tune parameters, and LC flow are set correctly.
10. Click OK.
Result: Sample tune starts. The ACQUITY UPLC Console displays the
progress of the tune. When the sample tune is complet e , the results are written to the Sulfadimethoxine.ipr file.
4-8 Operating the Waters TQ Detec tor
Page 73

Developing experiment methods

IntelliStart allows you to automatically develop quantitative MRM or SIR methods for compounds of interest. Up to four compounds can be handled in a single process. Up to three transitions can be selected for each compound.In this example, methods for sulfadimethoxine are create d.

Required materials

Sulfadimethoxine solution (100 pg/µL)
To create a method:
1. Prepare the IntelliStart fluidics system with sulfadimethoxine solution in reservoir B.
See also: “To prepare the IntelliStart fluidics system:” on page 4-2.
2. In the ACQUITY UPLC Console system tree, click TQ Detector.
3. Ensure that the Ion Mode is ES+.
4. In the ACQUITY UPLC Console system tree, click IntelliStart.
5. Clear the Instrument Setup check box.
6. Click Develop Method.
7. If you require system pre-checking, click Pre-checks.
See also: TQ Detector online Help topic “IntelliStart flow diagram”.
8. Click Start .
9. In the IntelliStart Setup Parameters dialog box, click the Method Develope r ta b .
Note: The masses, tune file and IntelliStart fluidics system parameters
already set in IntelliStart are used.
10. Select the parameters shown in the following figure.
Developing experiment methods 4-9
Page 74
Instrument Setup Parameters dialog box Method Developer tab:
In this case, Sulfadimethoxine.exp is the method file created. The Validation pane selections save optimization data for validation
purposes and creates an autotune report file (Sulfadimethoxine.xml).
11. Click Start .
Result: A message appears reminding you to ensure that the sample
solutions, sample tune parameters and LC flow are set correctly.
12. Click OK.
Result: Method development starts. The ACQUITY UPLC Console
displays the progress of the method development. When the method development is complete, the method is written to the Sulfadimethoxine.exp file.
The Validation pane selection saves optimization data for validation purposes, and creates an autotune report file (Sulfadimethoxine.xml).
A green check mark indicates a successful run; a red cross indicates a failure.
4-10 Operating the Waters TQ Detector
Page 75

Verifying the system using System QC

The System QC function uses pre-defined LC/MS methods to test the system by injecting a diagnostic sample on-column. The sample helps identify any problems with the configured system by providing these data:
Sensitivity (signal-to-noise rat io)
Response (peak area, peak height)
Chromatographic performance (peak width, retention time)
System QC can be activated in several ways:
Manually, from the ACQUITY UPLC Console.
Manually, as part of an ACQUITY UPLC Console Autotune sequence.
Automatically, based on a scheduled date and time.
The following example shows you how to perform a manual verification.
See also: The MassLynx online Help for details of how to perform verifica tion
manually, as part of an Autotune sequence or automatically based on a scheduled dat e and ti me.
Multiple injections are processed using the OpenLynx Application- Ma nager. Measurements are reported for each chosen parameter across the repeat analyses. If results are within user-defined tolerances, the LC/MS system is ready to use.
The System Ready results are logged via the ACQUITY UPLC Console but can also appear in a printed report. The raw data and OpenLynx browser report are also stored for re cords.
Requirement: For System QC to work, OpenLynx must be installed with
MassLynx. OpenLynx is normally installed by default. If it has not been installed, run the installation DVD, select Modify, and select OpenLynx when prompted.
Before running System QC you need a suitable tune file, method f ile and inle t file. You can use IntelliSt art to creat e a tune file and devel op a me thod fo r t he sample you want to use for System QC.
See also: “Performing a sample tune” on page 4-7 and “Developing
experiment methods” on page 4-9.
Verifying the system using System QC 4-11
Page 76
An inlet file must be created independently.
See also: The MassLynx online Help. Requirement: IntelliStart reads the tune, inlet, and method files from the
System QC project (SystemQC.pro). You must save the tune, inlet, and method files to the Aqudb folder of the System QC project before performing the verifica tion.
To perform a manual verification using System QC:
1. In the ACQUITY UPLC Console, click System QC.
2. Click Start .
Instrument Setup Parameters dialog box System QC tab with defau lt parameters:
4-12 Operating the Waters TQ Detector
Page 77
3. In the IntelliStart Setup Parameters dialog box’s System QC tab, enter the number of pre-run and run injections, the vial (that is , the position of your sample in the Autosampler) and the injection volume.
Tip: At least two injections are required for a run, because some of the
validation parameters use standard deviation. A pre-run s ample is used to clean and stabilize the system. You need not perform a pre-run prior to a run.
Recommendation: Perform at least three injections.
4. Enter a tune, LC and inlet file, or use the browse button to select a file.
5. Select whether you are using a Scanning or SIR/MRM method. If using a scanning method, enter a Mass on which to perform peak detection.
6. Select Print Report to print a report of the results.
7. Select the Acceptance parameters.
Instrument Setup Parameters dialog box System QC tab Acceptance parameters :
Parameter Description
Retention Time Enter the expected retention time in the Set
Point box and enter the tolerance. If, for any sample, the peak top is outside the range, the test reflects a fa i l u re .
Peak Area The variation in peak area is calculated by
percentage Relative Standard Deviation (%RSD).
%RSD = (Standard Deviation/Average Area) × 100%
Enter a tolerance in %RSD, above which the test reflects a fa i l u re .
Peak Height This is similar to Peak Area but using Peak
Height instead.
Peak Width Enter the expected peak wi dth in the Set Point
box and enter the tolerance. If, for any sam ple, the peak width falls outside the range, the test reflects a failure.
Verifying the system using System QC 4-13
Page 78
Instrument Setup Parameters dialog box System QC tab Acceptance parameters (Continue d):
Parameter Description
Signal/Noise In Signal/Noise enter the signal-to-noise ratio
below which the test reflects a failure. In Noise Range, enter the range in retention
time window over which the noise should be calculated.
Select how the noise is calculated, either RMS or Peak-to- P eak.
Tip: The signal is the peak height.
Tip: The calculations do not include injections where no peak was found.
For example, if only eight out of ten injections evidence detected peaks, the acceptance test results would be based on eight values rather than ten. However, the System QC process would always be marked as “failed” due to the missing peaks.
8. Click Start .
Result: A message appears reminding you to ensure that the sample
vial, QC parameters and LC flow are set correctly.
9. Click OK.
Result: System QC starts. The ACQUITY UPLC Console displays its
progress.
4-14 Operating the Waters TQ Detector
Page 79

A Safety Advisories

Waters instruments display hazard symbols designed to alert you to the hidden dangers of operating and maintaining the instruments. Their corresponding user guides also include the hazard symbols, with accompanying text statements describing the hazards and telling you how to avoid them. This appendix presents all the safety symbols and statements that apply to the entire line of Waters products.
Contents: Topic Page
Warning symbols A-2 Caution symbol A-5 Warnings that apply to all Waters instruments A-5 Electrical and handling symbols A-13
A-1
Page 80

Warning symbols

Warning symbols alert you to the risk of death, injury, or seriously adverse physiological reactions associated with an instrument’s use or misuse. Heed all warnings when you install, repair, and operate Waters instruments. Waters assumes no liability for the failure of those who install, repair, or operate its instruments to comply with any safety precaution.

Task-specific hazard warnings

The following warning symbols alert you to risks that can arise when you operate or maintain an instrument or instrument component. Such risks include burn injuries, electric shocks, ultraviolet radiation exposures, and others.
When the following symbols appear in a manual’s narratives or procedures , their accompanying text identifies the specif ic risk and explains how to avoid it.
Warning: (Ge neral risk of danger. When this symbol appears on an
instrument, consult the instrument’s user documentation for important safety-related information before you use the instrument.)
Warning: (Risk of burn injury from contacting hot surfaces.)
Warning: (R isk of electric shock.)
Warning: (Risk of fire)
Warning: (R isk of needle puncture.)
Warning: (Risk of injury caused by moving machinery.)
Warning: (Risk of exposure to ult raviolet r a di a ti o n . )
Warning: (Risk of contacting corrosive substances.)
Warning: (Risk of exposure to a toxic substance.)
A-2 Safety Advisories
Page 81
Warning: (Risk of personal exposure to laser radiation.) Warning: (Risk of exposure to biological agents that can pose a serious
health threat.)

Warnings that apply to particular instruments, instrument components, and sample types

The following warnings can appear in the user manuals of particular instrument s a nd on labels affixed to them or th e i r component parts.
Burst warning
This warning applies to Waters instruments fitted with nonmetallic tubing.
Warning: Pressurized nonmetallic, or po lymer, tubing can burst.
Observe these precautions when working around such tubing:
• Wear eye protection.
• Extinguish all nearby flames.
• Do not use tubing that is, or has been, stressed or kinked.
• Do not expose nonmetallic tubing to incompatible compounds like tetrahydrofuran (THF) and nitric or sulfuric acids.
• Be aware that some compounds, like methylene chloride and dimethyl sulfoxide, can cause nonmetallic tubing to swell, which significantly reduces the pressure at which the tubing can rupt ure.
Mass spectrometer flammable solvents warning
This warning applies to instrum ents operated with flammable solvents.
Warning: Where significant quantities of flammable solvents are
involved, a continuous flow of nitrogen into the ion source is required to prevent possible ignition in that enclosed sp ace.
Ensure that the nitrogen supply pressure never falls below 400 kPa (4 bar, 58 psi) during an analysis in which flammable solvents are used. Also ensure a gas-fail connection is connected to the HPLC system so that the LC solvent flow stops if the nitrogen supply fails.
Warning symbols A-3
Page 82
Mass spectrometer shock hazard
This warning applie s t o al l Wa te r s m ass spectrometer s.
Warning: To avoid electric shock, do not remove the mass
spectrometer’s protective panels. The components they cover are not user-serviceable.
This warning applies to certain instruments when they are in Operate mode.
Warning: High volt ages can be present at certain external surfaces of
the mass spectrometer when the instrument is in Operate mode. To avoid non-lethal electric shock, make sure the instrument is in Standby mode before touching areas marked with this high voltage warning symbol.
Biohazard warning
This warning applies to Waters instruments that can be used to process material that might contain biohazards: substances that contain biological agents capable of producing harmful effects in humans.
Warning: Waters's instruments and software can be used to analyze or
process potentially infectious human-sourced products, inactivated microorganisms, and other biological materials. To avoid infecti on with these agents, assume that all biological fluids are infectious, observe good laboratory practices and, consult your organization’s biohazard safety representative regarding thei r proper use and handling. Sp ecific precautions appear in the latest edition of the US National Institutes of Health (NIH) publication, Biosafety in Microbiological and Biomedical Laboratories (BMBL).
A-4 Safety Advisories
Page 83
Chemical hazard warning
This warning applies to Waters instruments that can process corrosive, toxic, flammable, or other types of hazardous material .

Caution symbol

The caution symbol signifies that an instrument’s use or misuse can damage the instrument or compromise a sample’s integrity. The following symbol and its associated statement are typical of the kind that alert you to the risk of damagin g t he instrument or sample.
Warning: Waters instruments can be used to analyze or
process potentially hazardous substances. To avoid injury with any of these materials, familiarize yourself with the materials and their hazards, observe Good Laborator y Practices (GLP), and consult your organization’s safety representative regarding proper use and handling. Guidelines are provided in the latest edi tion of the National Research Council's publication, Prudent Practices in the
Laboratory: Handling and Disposal of Chemica ls.
Caution: To avoid damage, do not use abrasives or sol vents to clean the
instrument’s case.

Warnings that apply to all Waters instruments

When operating this device, follow standard quality cont rol procedures and the the equipment guidelines in this section.
Caution symbol A-5
Page 84
Attention: Changes or modifications to this unit not expressly approved by the party responsible for c ompliance could void the user’s aut hority to operat e the equipment.
Important: Toute modification sur cette unité n’ayant pas été expressément approuvée par l’autorité respons able de la conformité à la réglementation peut annuler le droit de l’utilisateur à exploiter l’équipement.
Achtung: Jedwede Änderungen oder Modifikationen an dem Gerät ohne die ausdrückliche Genehmigung der für die ordnungsgemäße Funktionstüchtigkeit verantwortlichen Personen kann zum Entzug der Bedienungsbefugnis des Systems führen.
Avvertenza: eventuali modific he o al terazioni apportate a questa unità e non espressamente approvate da un ente responsabile per la conformità annulleranno l’autorità dell’utente ad operare l’apparecchiatura.
Atencion: cualquier cambio o modificación efectuado en esta unidad que no haya sido expresamente aprobado por la parte respons able del cumplimiento puede anular la autorización del usuario para utilizar el equipo.
A-6 Safety Advisories
Page 85
Warning: Use caution when working with any polymer tubing under pressure:
• Always wear eye protection when near pressurized polymer tubing.
• Extinguish all nearby flames.
• Do not use tubing that has been severely stressed or kinked.
• Do not use nonmetallic tubing with tetrahydrofura n (THF) or concentrated nitric or sulfuric acids.
• Be aware that methylene chloride and dimethyl sulfoxide cause nonmetallic tubing to swell, whic h greatly reduces the rupture pressure of the tubing.
Attention: Manipulez les tubes en polymère sous pression avec precaution:
• Portez systématiquement des lunettes de protection lorsque vous vous trouvez à proximité de tubes en polymère pressurisés.
• Eteignez toute flamme se trouvant à proximité de l’instrument.
• Evitez d'utiliser des tubes sévèrement déformés ou endommagés.
• Evitez d'utiliser des tubes non métalliques avec du tétrahydrofurane (THF) ou de l'acide sulfurique ou nitrique concentré.
• Sachez que le chlorure de méthylène et le diméthylesulfoxyde entr aînent le gonflement des tuyaux non métalliques, ce qui réduit considérablement leur pression de rupture.
Vorsicht: Bei der Arbeit mit Polymerschläuchen unter Druck ist besondere Vorsicht angebracht:
• In der Nähe von unter D ruck stehenden Polymerschläuchen stets Schutzbrille tragen.
• Alle offenen Flammen in der Nähe löschen.
• Keine Schläuche verwenden, die stark geknickt oder überbeansprucht sind.
• Nichtmetallische Schläuche nicht für Tetrahydrofuran (THF) oder konzentrierte Salpeter- oder Schwefelsäure verwenden.
• Durch Methylenchlorid und Dimethylsulfoxid können nichtmetallische Schläuche quellen; dadurch wird der Berstdruck des Schlauches erheblich reduziert.
Warnings that apply to all Waters instruments A-7
Page 86
Attenzione: prestare attenzione durante l’utilizzo dei tubi di polime ro pressurizzati:
• Indossare sempre o cchiali da lavoro protet tivi nei pres si di tubi di polimero pressurizzati.
• Estinguere ogni fonte di ig nizione circostante.
• Non utilizzare tubi soggetti che hanno subito sollecitazioni eccessive o son stati incurvati.
• Non utilizzare tubi non metallici con tetraidrofurano (THF) o acido solforico o nitrico concentrato.
• Tenere presente che il cloruro di metilene e il dimetilsolfossido provocano rigonfiamento nei tubi non met allici, rid ucendo notevol mente la r esistenza alla rottura dei tubi stessi.
Advertencia: se recomienda precaución cuando se trabaje con tubos de polímero sometidos a presión :
• El usuario deberá protegerse siempr e los ojos cuando trabaje cerca de tubos de polímero sometidos a presión.
• Si hubiera alguna llama las proximidades.
• No se debe trabajar con tubos que se hayan doblado o sometido a altas presiones.
• Es necesario utilizar tubos de metal cuando se trabaje con tetrahidrofurano (THF) o ácidos nítrico o sulfúrico concentrados.
• Hay que tener en cuenta que el cloruro de metileno y el sulfóxido de dimetilo dilatan los tubos no metál icos, lo que reduce la presión de r uptura de los tubos.
A-8 Safety Advisories
Page 87
Warnings that apply to all Waters instruments A-9
Page 88
Warning: The user shall be made aware that if the equipment is used in a manner not specified by the manufacturer, the protection provided by the equipment may be impaired.
Attention: L’utilisateur doit être informé que si le matériel est utilisé d’une façon non spécifiée par le fabricant, la protection assurée par le matériel risque d’être défectueuses.
Vorsicht: Der Benutzer wird darauf aufmerksam gemacht, dass bei unsachgemäßer Verwenddung des Gerätes unter Umständen nicht ordnungsgemäß funktionieren.
Attenzione: l’utente deve essere al corrente del fat to che, se l’apparecchiatura viene usta in un modo specificato dal produttore, la protezione fornita dall’apparecchiatura potrà essere invalidata.
Advertencia: el usuario deberá saber que si el equipo se utiliza de forma distinta a la especificada por el fabricante, las medidas de protección del equipo podrían ser insuficientes.
A-10 Safety Advisories
Page 89
Warning: To protect against fir e hazard, replace fuses with thos e of the same type and rating.
Attention: Remplacez toujours les fusibles par d’ autres du même type et de la même puissance afin d’éviter tout risque d’incendie.
Vorsicht: Zum Schutz gegen Feuergefahr die Sicherungen nur mit Sicherungen des gleichen Typs und Nennwertes ersetzen.
Attenzione: per una buona protezione contro i rischi di incendio, sostituire i fusibili con altri dello stesso tipo e amperaggio.
Advertencia: sustituya los fusibles por otros del mismo tipo y características para evitar el riesgo de incendio.
Warnings that apply to all Waters instruments A-11
Page 90
Warning: To avoid pos sible elect rical shock, di sconnect the p ower cord before servicing the instrument.
Attention: Afin d’éviter toute poss ibilité de commotion électrique, débranchez le cordon d’alimentation de la prise avant d’effectuer la maintenance de l’instrument.
Vorsicht: Zur Vermeidung von Stromschlägen sollte das Gerät vor der Wartung vom Netz getrennt werden.
Attenzione: per evitare il rischio di scossa elettrica, scollegare il cavo di alimentazione prima di svolgere la manutenzione dello strumento.
Precaución: para evitar descargas eléctricas, desenchufe el cable de alimentación del instrumento antes de realizar cualquier reparación.
A-12 Safety Advisories
Page 91

Electrical and handling symbols

Electrical symbols

These can appear in inst rument user manuals and on the instrument’s front or rear panels.
Electrical power on
Electrical power off
Standby
Direct current
Alternating current
Protective conductor terminal
Frame, or chassis, terminal
Fuse
Recycle symbol: Do not dispose in municipal waste.
Electrical and handling symbols A-13
Page 92

Handling symbols

These handling symbols and their associ ated t ext can app ear on lab els affixed to the outer packaging of Waters instrument and component shipments.
Keep upr ight!
Keep dry!
Fragile!
Use no hooks!
A-14 Safety Advisories
Loading...