1. Correct connection of 1/16 inch outer diameter stainless steel tubing
leading to and from the column is essential for high quality chromato-
graphic results.
2. When using standard stainless steel compression screw fittings, it is
important to ensure proper fit of the 1/16 inch outer diameter stainless
steel tubing. When tightening or loosening the compression screw,
place a 5/16 inch wrench on the compression screw and a 1/2 inch
wrench on the hex head of the column endfitting.
Note: If one of the wrenches is improperly placed on the inner column hex head (or the
cartridge tube flat) during this process, the endfitting will be loosened and leak.
3. If a leak occurs between the stainless steel compression screw fitting
and the column endfitting, a new compression screw fitting, tubing and
ferrule must be assembled.
4. An arrow on the column identification label indicates correct direc-
tion of solvent flow. Correct connection of 1/16 inch outer diameter
stainless steel tubing leading to and from the column is essential for
high-quality chromatographic results. Tubing touc hes the bottom of the
In a proper tubing/column connection (Figure 2), the tubing touches
the bottom of the column endfitting, with no void between them.
Figure 2: Proper Tubing/Column Connection
The presence of a void in the flow stream reduces column perfor-
mance. This can occur if a Parker style ferrule is connected to a
Waters endfitting (Figure 3).
Waters Spherisorb Columns7
[ Care and Use ManUal ]
Void
Figure 3: Parker Ferrule in a Waters Style Endfitting
Figure 5: Single and Double SLIPFREE Connectors
Note: A void appears if tubing with a Parker style ferrule is connected to a Waters
style column.
There is only one way to fix this problem: Cut the end of the tubing
with the ferrule, place a new ferrule on the tubing and make a new
connection. Before tightening the screw, make sure that the tubing
bottoms out in the endfitting of the column. Conversely, if tubing
with a Waters ferrule is connected to a column with Parker style
endfitting, the end of the tubing will bottom out before the ferrule
reaches its proper sealing position. This will leave a gap and create a
leak (Figure 4).
Gap
Figure 4: Waters Ferrule in a Parker Style Endfitting
Note: The connection leaks if a Waters ferrule is connected to a column with a
Parker style endfitting.
There are two ways to fix the problem:
1. Tighten the screw a bit more. The ferrule moves forward, and reaches
the sealing surface. Do not overtighten since this may end in breaking
the screw.
SLIPFREE Connectors Features
• Tubing pushed into endfitting, thereby guaranteeing a void-free
connection
• Connector(s) come(s) installed on tubing
• Various tubing i.d. and lengths available
• Fingertight to 10,000 psi – never needs wrenches
• Readjusts to all column endfittings
• Compatible with all commercially available endfittings
• Unique design separates tube-holding function from sealing function
Table 5. Waters Part Numbers for SLIPFREE Connectors
SLIPFREE Type and
Tubing Length
Single 6 cmPSL 618000PSL 618006PSL 618012
Single10 cmPSL 618002PSL 618008PSL 618014
Single 20 cmPSL 618004PSL 618010PSL 618016
Double 6 cmPSL 618001PSL 618007PSL 618013
Double 10 cmPSL 618003PSL 618009PSL 618015
Double 20 cmPSL 618005PSL 618001PSL 618017
0.005”0.010”0.020”
Tubing Internal Diameter
2. Cut the tubing, replace the ferrule and make a new connection.
Alternatively, replace the conventional compression screw fitting with
®
an all-in-one PEEK
fitting (Waters Part Number PSL613315) that
allows resetting of the ferrule depth. (Note that PEEK fittings are not
recommended for normal-phase applications!) Another approach is to
®
use a Thermo Corporation SLIPFREE
connector to always ensure the
correct fit.
The fingertight SLIPFREE connectors automatically adjust to fit all
compression screw type fittings without the use of tools (Figure 5).
Waters Spherisorb Columns8
[ Care and Use ManUal ]
Diluted/Distorted Sample Band
0.005 inches
0.020 inches
0.040 inches
AU
82
b. Band Spreading Minimization
Figure 6 shows the influence of tubing internal diameter on system
band spreading and peak shape. As can be seen, the larger tubing
diameter causes excessive peak broadening and lower sensitivity.
Figure 6: Effect of Connecting Tubing on System
c. Measuring System Bandspreading Volume and System
Variance
This test should be performed on an HPLC system with a single
wavelength UV detector (not a Photodiode Array [ PDA]).
1. Disconnect column from system and replace with a zero dead
volume union.
2. Set flow rate to 1 mL/min.
3. Dilute a test mix in mobile phase to give a detector sensitivity of 0.5 -
1.0 AUFS (system start up test mix can be used which contains uracil,
ethyl and propyl parabens; Waters P/N WAT034544).
4. Inject 2 to 5 μL of this solution.
5. Measure the peak width at 4.4% of peak height (5-sigma method):
5-sig ma Bandspre ading (μL) = Pea k Width (min) x Flow Rate (mL/min) x (1000 μL /1 mL)
System Varianc e (μL2) = (5-sigma bandspreading)2/ 25
System Volume
Figure 8: Determination of System Bandspreading Volume Using
5-Sigma Method
In a typical HPLC system, the bandspreading volume should be 100 μL
2
± 30 μL (or Variance of 400 μL
+/- 36 μL2). In a microbore (2.1 mm
i.d.) system, the bandspreading volume should be no greater than 20
2
to 40 μL (or variance no greater than 16 μL
to 64 μL2).
d. Measuring System Volume
System volume is important in scaling separations because it
creates an isocratic hold at the start of every run. This hold is often
several column volumes on a small scale, but a fraction of the
volume of a preparative column. Compensation for this volume must
be included in planning a scaling experiment to avoid distorting the
chromatography (Figure 9).
Programmed time = 5 min
0.70
0.65
0.60
0.55
0.50
0.45
0.40
0.35
0.30
0.25
0.20
0.15
0.10
0.05
0.00
50%
02468101214161
Figure 9: Determination of Gradient Delay Volume
Flow Rate = 1.5 mL/min
5.69 min
-5.00 min
0.69 min
Time = 5.69 minutes
System Volume:
0.69 min x 1.5 mL/min = 1.04 mL
0 Min.
Waters Spherisorb Columns9
[ Care and Use ManUal ]
1. Remove column.
2. Use acetonitrile as mobile phase A, and acetonitrile with 0.05 mg/mL
uracil as mobile phase B (eliminates non-additive mixing and viscosity
problems).
3. Set UV detector at 254 nm.
4. Use the flow rate in the original method and the intended flow rate on
the target instrument.
5. Collect 100% A baseline for 5 minutes.
6. Program a step change at 5 minutes to 100% B, and collect data for an
additional 5 minutes.
7. Measure absorbance difference between 100% A and 100% B.
8. Measure time at 50% of that absorbance difference.
9. Calculate time difference between start of step and 50% point.
10. Multiply time difference by flow rate.
vii. additional inFormation
a. Use of Narrow-bore (<3.0 mm i.d.) Columns
This section describes how to minimize extra column effects and
provides guidelines on maximizing the performance of a narrowbore
column in an HPLC system. A 3.0 mm i.d. narrow-bore column
usually requires no system modifications. A 2.1 mm i.d. column,
however, requires modifications to the HPLC system in order to
eliminate excessive system bandspreading volume. Without proper
system modifications, excessive system bandspreading volume
causes peak broadening and has a large impact on peak width as
peak volume decreases.
System with 130 L
bandspreading: 8,000 plates
78 min7.578 min7.5
Non-optimized LC/MS/MS SystemOptimized System
Figure 10: Non-Optimized vs. Optimized LC/MS/MS System
System with 70 L bandspreading:
10,000 plates (same column)
c. Non-Optimized vs. Optimized LC/MS/MS System:
System Modification Recommendations
1. Use a microbore detector flow cell with ≤2.1 mm i.d. columns.
Note: Detector sensitivity is reduced with the shorter flow cell path length in
order to achieve lower bandspreading volume.
2. Minimize injector sample loop volume.
3. Use 0.009 inch (0.25 mm) tubing between pump and injector.
4. Use 0.009 inch (0.25 mm) tubing for rest of connections in standard
systems and 0.005 inch (0.12 mm) tubing for narrowbore (2.1 mm i.d.)
systems.
5. Use perfect (pre-cut) connections (with a variable depth inlet if using
columns from different suppliers).
6. Detector time constants should be shortened to less than 0.2 seconds.
b. Impact of Bandspreading Volume on 2.1 mm i.d.
Column Performance
Note: Flow splitters after the column will introduce additional bandspreading.
System optimization, especially in a system that contains a flow
splitter, can have dramatic effects on sensitivity and resolution.
Optimization includes using correct-depth ferrules and minimizing
tubing diameter and lengths. An example is given in Figure 10
where system optimization resulted in a doubling of sensitivity and
resolution of the metabolite in an LC/MS/MS system.
Waters Spherisorb Columns10
[ Care and Use ManUal ]
d. Guard Cartridges and Columns Assembly
These endfittings and accessories can be used with 4.6 mm, 4.0 mm, or
3.0 mm i.d. cartridges.
DescriptionQty.Part No.
Removable Column Endfitting2/pkPSS614100
Frit Assembly (2 µm)5/pkPSS614103
Frit Assembly (0.5 µm)5/pkPSS614104
Column Coupler2/pkPSS614102
Long Tail Endfitting2/pkPSS614101
Extended Endfitting for use with 10 mm Integral Guard1/pkPSS614108
Nylon Column Plugs for Storage of Complete Column1/pkWAT015674
Nylon Column Caps for Storage of
Replacement Cartridge Column
Inline 10 mm Guard Cartridge Holder Kit
for use with above items
2
1
10/pkPSS614113
PSS830008
3
4
Inline Guard Cartridge Holder
10 mm Guard Cartridge (4.6 mm i.d.)
Inline Guard Cartridge Holder
PSS 830008
Extended Endfitting for use with 10 mm Guard Cartridges
10 mm Guard Cartridge (4.6 mm i.d.)
Cartridge
®
Extended Endfitting
for use with 10 mm Guard
PSS 614108
Waters Spherisorb
30 mm Guard Cartridge Coupled to Spherisorb Cartridge
1. 30 mm Stand Alone Guard/Column (endfittings not included)
2. Extended endfitting for use with 10 mm Integral Guard: PSS614108
3. 10 mm Integral Guard Column
4. Column Coupler: PSS614102
Austria and European Export (Central South Eastern Europe, CIS and Middle East) 43 1 877 18 07, Australia 61 2 9933 1777, Belgium 32 2 726 1000, Brazil 55 11 4134 3788,