b. Preventing Contamination
c. Collecting the Sample
d. Eluting the Derivatives from the Sampler
III. ANALYZING THE DNPH DERIVATIVES
a. Theory of Operation
b. Preventing Contamination
c. Collecting the Sample
d. Calculating Results
IV. APPLICATION EXAMPLES
a. Formaldehyde in Laboratory Air - STEL
b. Glutaraldehyde in Laboratory Air – STEL
I. INTRODUCTION
Waters Sep-Pak® XPoSure™ Aldehyde Samplers are convenient,
reproducible sampling devices for quantifying formaldehyde
concentrations in the workplace and indoor air within a range of
0.001 to 5 parts per million (ppmv).
V. STORAGE AND DISPOSAL OF THE SAMPLERS
a. Storing Unused Samplers
b. Storing Exposed Samplers
c. Disposing of used cartridges
VI. TROUBLESHOOTING
VII. REFERENCES AND BIBLIOGRAPHY
VIII. ORDERING INFORMATION
IX. APPENDIC ES
a. Appendix A: Measuring Acetontrile Purity
b. Appendix B: Synthesizing the DNPH-Derivative Standards
c. Appendix C: Measuring Breakthrough
d. Appendix D: Useful Conversion Factors
d.1. Carbonyl to Hydrazone Conversion Factors
d.2. Equation for converting µg/L to ppmv
d.3. Conversion Factors: µg/L to ppmv
U.S. Environmental Protection Agency; Research Triangle Park,
NC, 1984.
4. Guide to Successful Operation of Your LC System; Waters
Corporation, Milford, MA 1991.
5. ASTM Method E411; Standard Test Method for Trace Quantities
of Carbonyl Compounds with 2,4-Dinitrophenylhydrazine.
If you acetonitrile is unacceptable for your application, contact
your solvent supplier, to purify the acetonitrile. To purify
acetonitrile, distill it from an acidified DNPH solution, using a
procedure analogous to the one described in ASTM Method E411
for the purification of methanol.
5
To measure acetonitrile purity:
1. Clean all glassware by rinsing with acetonitrile and heating
in a 60 °C vacuum oven for at least 30 minutes.
2. Elute a fresh sampler with 3 mL acetonitrile.
3. Within 3 minutes, inject the eluate into the HPLC system to
measure the concentration of DNPH derivatives.
4. Add 1 drop of concentrated HCl to the eluate, and allow to
react at room temperature for 30 minutes.
5. Remeasure the concentration of DNPH derivatives by HPLC.
6. Calculate the difference in the concentration of each DNPH
derivative measured in steps 3 and 5 to yield the contribution
form the acetonitrile.
VIII. ORDERING INFORMATION
Waters Sep-Pak XPoSure Aldehyde Samplers are shipped in boxes
of 20 individually-packaged cartridges. Pouches are supplied for
storage after sampling.
7. Calculate the percent hydrazone contributed by the acetonitrile
relative to the background level. The value for any hydrazone
should not exceed 25% of its value in the blank.
Sep-Pak XPoSure Aldehyde Sampler
7
Page 8
[ CARE AND USE MANUAL ]
[ CARE AND USE MANUAL ]
Example: Measuring Acetonitrile Purity
1. HPLC analysis of a fresh sampler shows the sample contains:
Derivative
FormaldehydeDNPH
AcetaldehydeDNPH
Acetone-DNPH 1.60 ÷ 0.40 x 100 = 400%
Contribution
from
Acetonitrile
0.01 ÷ 0.08 x 100 = 12%
0.02 ÷ 0.12 x 100 = 17%
Divided by
Background
Value
Times
100
Equals
Percent
Relative to
Background
2. Analysis of the concentrations of hydrazones after reacting
with acid yields:
DerivativeConcentration (µg/mL)
Formaldehyde-DNPH0.08
Acetaldehyde-DNPH0.12
Acetone-DNPH0.40
All other hydrazones<0.05
3. The difference between the concentrations of hydrazone from
steps 3 and 5 represents the amount of hydrazone
contributed by the acetonitrile:
DerivativeConcentration (µg/mL)
Formaldehyde-DNPH0.09
Acetaldehyde-DNPH0.14
Acetone-DNPH2.00
All other hydrazones<0.05
4. The percent of the hydrazones contributed by the acetonitrile is:
Derivative
FormaldehydeDNPH
AcetaldehydeDNPH
Acetone-DNPH2.00 µg/mL - 0.40 µg/mL = 1.60 µg/mL
Concentration
after Reaction
with Acid
0.09 µg/mL - 0.08 µg/mL = 0.01 µg/mL
0.14 µg/mL - 0.12 µg/mL = 0.02 µg/mL
Minus
Concentration
in Blank
Equals
Contribution
form Acetonitrile
Since the percent for formaldehyde and acetaldehyde arising
from the acetonitrile is less than 25% of the background level
in the sampler, the acetonitrile is considered clean for these
compounds. If the analysis considers only these compounds, the
acetonitrile is acceptable.
However, the amount of acetone arising for the acetonitrile is
4 times the amount found in the the background level. Therefore,
it is suggested that this lot of acetonitrile may be unacceptable
for use in the analysis of acetone.
b. Appendix B: Synthesizing the DNPH-Derivative Standards
High purity (99%) DNPH derivatives are commercially available
or can be synthesized from DNPH supplied by Aldrich Chemical
Company (70% DNPH and 30% water). To synthesize 98-99%
pure hydrozones:
1. Prepare one liter of 2 M HCl solution: Add 172 mL concentrated
reagent-grade hydrochloric acid (HCl) to a 1 L volumetric
flask. Fill the flask to the mark with distilled deionized water.
2. Saturate the 2 M HCl solution with DNPH: Add 8 g DNPH and
stir for one hour at 20 to 25 °C. Filter through a 0.45 µm
hydrophilic membrane (HVLP) filter (Waters Part number:
WAT200530).
3. Form the hydrazone derivative by adding a two-fold molar
excess of reagent-grade aldehyde or ketone to the filtered
2 M HCl DNPH solution. Stir for 30 minutes to one hour at
20 to 25 °C.
4. Filter the hydrazone slurry. Wash the hydrazone with 50 mL
2 M HCl 3 times. Wash with 50 mL water 3 times. Dry the
filter cake in an oven at 50 to 60 °C overnight.
5. Prepare a standard stock solution of the DNPH-derivatives
by dissolving an accurately weighed amount in acetonitrile.
Prepare a set of calibration standards using the stock
solution. The concentration of the standards should be in the
same range as the expected concentration of the samples.
The solutions are stable for at least one month when stored in
tightly-capped glass vials at 4 °C ± 2 °C.
Sep-Pak XPoSure Aldehyde Sampler
8
Page 9
[ CARE AND USE MANUAL ]
[ CARE AND USE MANUAL ]
c. Appendix C: Measuring Breakthrough
Note: If several aldehydes and ketones are present in significant
concentration, estimate the maximum sample size from the total
concentration of all species. Collection efficiency determinations
are best made during times expected to yield peak formaldehyde
concentrations. This will enable appropriate sampling rates and
intervals to be selected to avoid breakthrough.
Figure 8 shows the predicted total carbonyl concentration versus
the range of sample volumes.
To measure Waters Sep-Pak XPoSure Aldehyde Sampler for
collection efficiency:
1. Connect two unused cartridges together by the Luer fittings
and mark each cartridge for identification.
2. Connect the cartridges to a calibrated pump with a short
length of flexible tubing.
3. Collect the sample.
4. Elute both cartridges and a third blank cartridge.
5. Analyze all three cartridges by HPLC.
6. Subtract the blank value from the values determined from the
other two cartridges.
7. Calculate and sum of the total captured DNPH-derivative from
both cartridges.
8. Divide the amount of DNPH-derivative determined from
the first cartridge by the total amount determined form
cartridges 1 and 2. Multiply by 100. This is the percentage of
DNPH-derivatives captured on the first cartridge. This value
should exceed 95%; otherwise, some of the sample broke
through to the second cartridge.
Figure 8. Total Carbonyl Concentration vs. Range of Sample Volumes.
Collection efficiency for Waters Sep-Pak XPoSure Aldehyde
Sampler is greater than 95% for air sampling rates of up to
1.5 L/min. The sampler may exhibit breakthrough if:
The sampling flow rate is greater than 1.5 L/min
The amount of sample collected is enough to react with
more than 50% of the DNPH (~2.3 µmoles)
Example: Measuring Sample Breakthrough
The expected concentration of formaldehyde is 0.66 ppmv (µL/L).
Flow rate is 1.25 L/min for 80 minutes. A sample volume of 100
liters is collected. The theoretical quantity of carbonyl is:
Analyte ppmv Carbonyl Collected
concentration x molecular x air volume = µg Carbonyl
weight
Analyte molar volume at 1 atm/25 °C
This calculates to:
0.66 µL x 30.03 g/mole x 100 L = 81 µg formaldehyde
24.46 L/mole
The actual results are shown in Table 7. To calculate the percent
captured on the first sampler, divide the quantity captured on
sampler 1 by the total quantity captured, then multiply by 100.
Since this value is less than 95%, and the total carbonyl amount
exceeded 2.3 µmoles, breakthrough occurred.
Sep-Pak XPoSure Aldehyde Sampler
9
Page 10
[ CARE AND USE MANUAL ]
[ CARE AND USE MANUAL ]
In the above example, only a single carbonyl source was present.
Under many test conditions more than one carbonyl source may be
present in significant concentrations. These other compounds will
consume DNPH, effectively reducing the capacity of the sampler for
the compound of interest. To assure that the capacity of the sampler
has not been exceeded, compare the DNPH peak areas of the sample
to a similarly eluted blank. T he DNPH peak area in all samples
must be no less than 50% of the DNPH peak area of the blank. This
ensures the sampler capacity has not been exceeded.
Table 7: Breakthrough Example HPLC Results
Sampler
Sampler 175.0675.0091.8
Sampler 26.726.668.2
Blank0.06--
Amount
(µg)
Quantity Captured
Sampler – blank (µg)
Percent Captured
on Sampler
d. Appendix D: Useful Conversion Factors
This appendix contains:
Carbonyl to hydrazone conversion factors
Equation for converting µg/L to ppmv
Conversion factors: µg/L ppmv
Table 9: Conversion Factors,
Carbonyl Weights ↔ Derivative Weights
Carbonyl
Compounds
Formaldehyde6.9980.1429
Adetaldehyde5.0890.19 65
Adetone4.1010.2438
Glutaraldehyde
Carbonyl → Derivative,
(MWD /MWC)
4.5890.2175
Derivative → Carbonyl,
(MWC /MWD)
d.2. Equation for converting µg/L to ppmv
Carbonyl concentrations can be converted from µg/L to ppmv
(µL/L) by using the following expression:
(Result in ppmv) = (Result in µg/L) x 22.41 x T2 x P1
MWC T1 P2
Where values are:
22.41 = Molar volume of an ideal gas at STP (273.15 °K and
1 atm), L/mole
MWC = Molecular weight of carbonyl, g/mole
Obtaining carbonyl concentrations in eluates and air samples
required the use of several constants and conversion factors. The
factors described in this appendix can be used when converting
carbonyl weights to:
Equivalent derivative weights for preparing standard
solutions
Volumes for reporting air samples in ppmv
d.1. Carbonyl to Hydrazone Conversion Factors
Table 8 lists the molecular weights (MW) for some carbonyl
compounds. These values were used to derive the conversion
factors listed in Table 9. Multiply the carbonyl or derivative
weights by the appropriate factor for the desired conversion.
Table 8: Carbonyl and Hydrazone Molecular Weights
Carbonyl
Compounds
Formaldehyde30.03210.15
Adetaldehyde44.0522 4.17
Adetone58.08238.20
Glutaraldehyde
Carbonyl Compounds
Molecular Weight, (MWC)
100.12460.36
Hydrazone Derivative
Molecul ar Weight (MWD)
T1 = Standard temperature, 273.15 °K
T2 = Air sample temperature, ºK
P1 = Standard pressure, 1 atm
P2 = Air sample pressure, atm
d.3. Conversion Factors: µg/L to ppmv
Table 10 lists the factors for converting between µg/L and ppmv at
25 °C and 1 atm. Results are converted between µg/L (or mg/m3)
and ppmv, by multiplying by the appropriate factor.
Table 10: Factors for Converting Between µg/L and ppmv at
25 °C and 1 atm.
Carbonyl
Compounds
Formaldehyde1.230.813
Adetaldehyde1.8 00.555
Adetone2.380.420
Glutaraldehyde
ppmv → µg/Lµg/L p → ppmv
4.090.244
Sep-Pak XPoSure Aldehyde Sampler
10
Page 11
[ CARE AND USE MANUAL ]
Waters, The Science of W hat’s Possible, Symmetry, and Sep-Pak are registered trademarks of Waters Corporation.
XPoSure is a trademark of Waters Corporation. All other trademarks are the property of their respective owners.