Waters Protein-Pak Hi Res IEX Columns User Manual

[ CARE AND USE MANUAL ]
columns
Protein-Pak Hi Res IEX Columns and Standards
CONTENTS
I. INTRODUCTION
II. CONFIGURING AN ACQUITY SYSTEM FOR USE IN IEX PROTEIN SEPARATIONS
III. GETTING STARTED a. eCord Installation
b. Column Connectors c. Column Installation d. Column Equilibration e. Useful Functional Tests for Benchmarking a New Column
IV. COLUMN SPECIFICATIONS AND USE a. Specifications
b. Sample Preparation c. Operating pH d. Solvents e. Pressure f. Flow Rate g. Temperature
I. INTRODUCTION
Thank you for choosing a Protein-Pak™ ion-exchange column.
Waters Protein-Pak Hi Res IEX family of columns were developed
to assist in the UPLC® characterization of recombinant proteins
and monoclonal antibodies found in many of today’s novel
biopharmaceutical therapeutics. T he non-porous, high-ligand
density particles overcome some of the high molecular weight
limitations present using traditional porous, IEX particles. In
addition, use of a multi-layered network of ion-exchange groups
(SP, CM, or Q) on non-porous particles assist the effective diffusion
and binding of charged proteins or other biomolecules onto the
particles to deliver relatively high sample loading capacities and
component resolution while minimizing column fouling.
The available ion-exchange chemistries include a strong-anion
and weak- and strong-cation exchangers (Table 1), all of which are
stable, over a wide pH range (3–10), high salt concentrations and
standard pressures (see Column Specifications and Use section) via
use of polymer-based particles. Only when the Protein-Pak Hi Res
IEX column series is combined with the ACQUITY UPLC® system
will the full performance benefits of this column be realized.
V. TROUBLESHOOTING
VI. COLUMN CLEANING, REGENERATION, AND STORAGE a. Cleaning and Regeneration
b. Storage
VI. INT RODUCING ECORD INTELLIGENT CHIP TECHNOLOGY a. Introduction
b. Installation c. Manufacturing Information d. Column Use Information
VIII. ORDERING INFORMATION
Table 1
Abbreviation Ion-Exchange Ligand
SP Sulfopropyl
CM Carboxymethyl
Q Quaternary ammonium
[ CARE AND USE MANUAL ]
2 mm
II. CONFIGURING AN ACQUITY SYSTEM FOR USE IN IEX PROTEIN SEPARATIONS
Although Waters Protein-Pak Hi Res IEX columns do not require
the high pressure capability of an ACQUITY UPLC system, the full
performance benefits of this column will only be realized with the
low system volume and low detector dispersion of the optimized
ACQUITY UPLC system. However, ion-exchange chromatography
may require modifications to an existing ACQUITY UPLC system.
Please refer to “Size Exclusion and Ion-Exchange Chromatography
of Proteins using the ACQUITY UPLC System”, (715002147A) for
specific recommendations that can be obtained at www.waters.
com/chemcu.
III. GETTING STARTED
An Analysis Report and an Inspection Data Sheet are generated
and available for each shipped Protein-Pak Hi Res IEX column.
The Analysis Report is specific to each batch of packing material
and includes the batch number and chromatographic separation
obtained using defined protein standards. This document is
available upon request http://www.waters.com/waters/form.
htm?id=10048430&ev=10048691&locale=en_US. The
Inspection Data Sheet is included with each shipment and is
specific to each individual column and contains the following
information: batch number, column serial number, plate count,
tailing factor, retention factor, and chromatographic separation
conditions. These data should be stored for future reference.
b1. The ACQUITY UPLC system utilizes tubing and gold plated
compression screws that have been designed to meet
stringent tolerance levels and to minimize extra column
volumes. The columns should be attached to the ACQUITY
UPLC injector with a column stabilizer.
b2. Waters Protein-Pak Hi Res columns require end fittings that
have a 2 mm depth between the ferrule and the end of the
extending stainless steel tubing (Figure A) that must perfectly
seat to the end of the Protein-Pak Hi Res column. Connecting
a different style connector to a Waters Protein-Pak Hi Res
IEX column may leave a gap between the end of the stainless
steel tubing and the column that will result in undesired peak
broadening (Figure B).
To correct this problem, disassemble the finger-tight fitting of
the stabilizer and remake the connection (i.e., 2 mm depth) to the
Waters Protein-Pak Hi Res IEX column.
Figure A
Figure B
a. eCord Installation
The eCord™ button should be attached to the side of the Waters
ACQUITY UPLC column heater module. The eCord button is
magnetized and does not require specific orientation.
b. Column Connectors
There are two options for connecting the Protein-Pak Hi Res
column to an ACQUITY system. The first option is to use the
reusable fitting on the 150 mm and 100 mm column stabilizers
(sect b1). This fitting/tubing is supplied with ACQUITY UPLC
column heaters. A second alternative is to remove this fitting and
replace it with a non removable fitting (i.e. steel). If the second
option is employed, specific directions must be followed (sec b2).
In a proper tubing/column connection, the tubing touches the
bottom of the column end fitting with no void between them.
(Figure C).
Figure C
Protein-Pak Hi Res IEX Columns and Standards
2
[ CARE AND USE MANUAL ]
c. Column Installation
Note: The flow rates given in the procedure below are for a Protein-
Pak Hi Res CM or SP, 7 µm, 4.6 mm i.d. column.
1. Purge the solvent delivery system of any organic or water-
immiscible mobile phases and connect the inlet end of the column
to the injector outlet. An arrow on the column identification label
indicates the correct direction of solvent flow.
2. Flush column with 100% aqueous buffer (line A1) by setting
the pump flow to 0.2 mL/min.
3. When the buffer is flowing freely from the column outlet attach
the column outlet to the detector. Monitor the system pressure
to ensure the column is within its pressure limitations.
4. Gradually increase the flow rate, by not more than 0.5mL/min
at a time (as described in step 2).
5. Once a stable backpressure has been achieved, proceed to the
next section.
d. Column Equilibration
Protein-Pak Hi Res columns are shipped in 20 mM sodium sulfate
with 0.05% sodium azide. It is important to ensure mobile-phase
compatibility before changing to a different mobile-phase system.
Equilibrate the column with a minimum of 10 column volumes of
the buffer to be used (refer to Table 2 for column volumes).
Table 2. Empty Column Volumes in mL (multiply by 10 for flush solvent volumes).
Column Dimension Approximate Volume
4.6 x 100 mm 1.66 mL
e. Useful Functional Tests for Benchmarking a New Column
It may be useful to benchmark the column performance with
a sample that is representative of the intended application. A
separation of common proteins with an appropriate method is
suggested for your new column. Two standards were specially
designed for this purpose. The IEX Cation Test Standard is
specifically designed with proteins that resolve well on cation-
exchange columns. The IEX Anion Test Standard is specifically
designed with proteins that resolve well on anion-exchange
columns. This test can also be used to monitor the condition of
your column after extended use.
The following sets of conditions and chromatograms show
examples of using the IEX Standards for benchmarking and
monitoring column and system health. T he exact results observed
in any laboratory will depend on the instrument in use. Similar
results can only be expected with a Waters ACQUITY UPLC system.
These tests are valuable for monitoring the life of the column
and for troubleshooting separation difficulties that may arise.
Protein-Pak Hi Res Cation-Exchange Columns and Standards
Sample Diluent: 20 mM Sodium Phosphate Buffer, pH 6–7
The IEX Cation Test Standard is packaged in a Waters TruView
Max Recovery Vial (p/n 186005668cv) and is lyophilized for
prolonged storage and stability. The standard is shipped at room
temperature but it upon arrival, it is recommended to store this
standard at -20 ˚C.
Protein Sigma P/N Conc. mg/mL
Bovine Ribonuclease A R5500 4
Cytochrome C, bovine heart C2037 4
Chicken Lysozyme L6876 4
Column: Protein-Pak Hi Res CM, 4.6 x 100 mm,
or Protein-Pak Hi Res SP, 4.6 x 100 mm
Injection Volume: 10.0 μL
Flow Rate: 1 mL/min
Mobile Phase A: 20 mM sodium phosphate buffer, pH 6.6
Mobile Phase B: 20 mM sodium phosphate buffer,
0.5 M NaCl, pH 6.6
Weak Needle Wash: Mobile phase A
Strong Needle Wash: Mobile phase B
Seal Wash: 90/10 H20/MeOH
Temperature: 30 ˚C
Detection: 280 nm
Gradient:
Time %A %B Curve
1 100 0 6
34 40 60 6
36 40 60 6
38 100 0 6
55 100 0 6
Protein-Pak Hi Res IEX Columns and Standards
3
Loading...
+ 6 hidden pages