Page 1

[ Care and Use ManUal ]
Pico•Tag amino acid analysis column
I. IntroductIon
Waters exclusive bonding and packing processes, developed for the Pico•Tag®
Column (3.9 mm x 15 cm), P/N 88131, produce a durable, high efficiency
stationary phase ideally suited for the reverse-phase separation of phenyl-
thiocarbamyl amino acids. Our stringent quality control procedures result
in a column which, when used in conjunction with Waters Pico•Tag reagents,
eluents, and HPLC system is guaranteed to provide the speed, resolution, and
sensitivity needed for amino acid analysis.
This information provides the means to ensure that you derive the full benefit
of our research, manufacturing, and quality control. For additional information
on the complete method, refer to Section 4.4 of the Amino Acid Analysis
Manual, P/N 07124, or the Pico•Tag Amino Acid Analysis System Operator’s
Manual, P/N 88140.
WARNING: THE AMPULES OF PHENYLISOTHIOCYANATE WHICH ARE
CONTAINED IN THE PICO•TAG CHEMICAL PACKAGE REACT ON CONTACT WITH
STRONG ACIDS, EMITTING HIGHLY TOXIC CYANIDE FUMES AND/OR OXIDES
OF SULPHUR. CARE MUST BE TAKEN TO ASSURE THAT THESE AMPULES AND
CONTENTS ARE DISPOSED OF PROPERLY AND THAT THEY DO NOT COME INTO
CONTACT WITH ACIDS.
contents
I. IntroductIon
II. PrePArAtIon For oPerAtIon
a. Column Installation
b. Solvent Requirements
c. Equilibration
III. cAre And use
a. Sample and Eluent Preparation and Filtration
b. Precautions
c. Storage Considerations
d. Troubleshooting
e. Service and Applications Information
IV. column eFFIcIency
V. test condItIons
VI. recommended sPAre PArts (steel columns)
Figure 1. Pico•Tag Method Amino Acid Analysis 250 pmol of
Pierce H Standard
Pico•Tag Amino Acid Analysis Column 1
VII. WArrAnty
Page 2

[ Care and Use ManUal ]
II. PrePArAtIon For oPerAtIon
a. Column Installation
Remove the end plugs from your steel column with a 5/16” wrench and save
them for storage after the column has been removed from the system. The
column outlet is indicated by an arrow on the label (show-ing the direction
solvent should flow). Tighten the fittings 1/4-to-1/2 turn beyond finger tight.
DO NOT OVERTIGHTEN - THIS WILL DAMAGE THE FITTING SEAT. A properly
prepared and assembled com-pression fitting in good condition is all that
is required.
Follow the next four steps of this procedure if tubing cutting is required to
connect a new column or to improve the end connections on your existing
fittings.
1. Using a file with a cutting edge (such as the file included in the Startup
Tool Kit, P/N 96146, supplied with each Waters Liquid C hromatography
System) scribe the circumference of the tubing at the desired break.
2. Grasp the tubing on both sides of the scribe mark with cloth-covered
pliers (to prevent marring the tube surface) and gently work the tube
back and forth until it separates.
3. File the ends smooth and assemble as shown.
b. Solvent Requirements
Your Pico•Tag column is shipped containing an aqueous/organic mixture
compatible with the Pico•Tag eluents. It is recommended that only Waters
Pico•Tag eluents be used with this column.
Waters Pico•Tag Eluent A (P/N 88108)
Waters Pico•Tag Eluent B (P/N 88112)
Waters Pico•Tag Diluent (P/N 88119)
c. Equilibration
A necessary step to successful use of your column is the initial solvation
(or wetting) of the packing. Install the column in a column heater and bring
temperature to 38 °C. Purge the column at 1 ml/min with Eluent B, then
equilibrate with Eluent A for fifteen minutes. Equilibration between the mobile
phase and packing is established when a stable baseline can be produced. If
your result is un-satisfactory repeat the equilibration process.
III. cAre And use
Liquid chromatography columns have a finite life influenced by their care and
use, number of injections, sample and solvent cleanliness, frequency of sol-
vent changeover, and handling and storage procedures among other factors. If
a change is observed in the:
4. Slide the compression fitting, followed by the ferrule (large end of the
taper first) over the tube. Be certain to bottom the tube in the fitting
seat for which its use is intended to assure a leak-free connection.
Note: Attach a union in place of the column and flush the lines free of
microparticulates before attaching the column
Figure 5: Ferrule and Compression Assembly
.
Retention of a particular compound resolution between two compounds peak
shape.
Take immediate steps to determine the reason for the changes, and until the
determination is made, the results of any separations using the column must
not be relied upon. Follow generally accepted pro-cedures for quality control
and methods development when using these columns.
Important Note: Before running the first analysis on your new column perform
the Test Sample Separation given in the Test Conditions Section.
Pico•Tag Amino Acid Analysis Column 2
Page 3

[ Care and Use ManUal ]
a. Sample and Eluent Preparation and Filtration
Use HPLC grade solvents, filtered to remove microparticulate matter above
0.45 µm. (Waters Pico•Tag prefiltered Eluents A [P/N 88]108 and B [P/N
88]112 are recommended.) This reduces the problem of plugged filters
and preserves column life. Vacuum filtration or sonification may be used to
remove dissolved gasses which could affect your solvent delivery system.
Always filter prepared samples to prevent excessive pressure buildup due to
particulate matter
• Adequate sample cleanup (using Sep-Pak
®
cartridges designed for this
purpose) prevents alteration of the column chemistry by strongly adsorbing
or precipitating sample components.
• Use of an in-line pre-column filter (P/N 84560), is recommended.
b. Precautions
Normal recommended pressure should not exceed 3500 psi.
• For all silica-based packing materials, stay within a pH range of 2-8
(i.e., avoid using concentrated acids or bases). Maximum column life
(pH): 3.5 - 6.5.
• Filter all aqueous buffers (Waters Pico•Tag eluents are prefiltered).
Avoid using turbid or cloudy buffers. Be sure that any solutions contain-
ing buffers, salts, etc. are compatible with the wetted surfaces of the
column and equipment.
• Protect column from vibration, mechanical shock, and rapid changes in
pressure. Column packings are based on a highly porous and delicate
silica gel alignment. Any thermal, physical or chemical shock (such as
changing solvents rapidly or at high flow rates) can cause the particles
to shift and may result in a loss of efficiency.
• When using water, distill or treat with a Milli-Q
De-ionized water is not ac-ceptable because it contains organic com-
pounds which alter column selectivity.
• Protect the column from rapid changes in solvent composition. DO NOT
change the flow rate faster than 0.5 ml/min increments.
®
or equivalent system.
c. Storage Considerations (more than 72 hours without use)
Leaving the column unused for less than 72 hours does not generally require
storage procedures.
• DO NOT store the column in water alone; this practice will promote
microbial contamination. Store the column in Pico•Tag Eluent B or in
10% organic in water mixture.
• DO NOT allow buffers or other potentially harmful materials to remain
in the system when not being used. Flush and replace with Eluent B or
10% organic in water mixture.
• Return the column to its box with the end plugs firmly in place for
storage. Allowing steel columns to dry out can result in poor chromato-
graphic performance.
d. Troubleshooting
Table 2: Typical Column Problems and Solutions
Problem Cause Solution
Excess pressure
buildup
Loss of resolution,
broad peaks, low
plate counts
• Filters plugged
with Particulates
• Sample precipi-
tates on column
(sample not
soluble in mobile
phase)
• Contaminated
column, insuffi-
cient equilibration
• Column col-
l a p s e a n d vo i d
formation
• Filters partially
plugged
• Clean in an ultrasonic bath
or replace.
• Always alter mobile phases
and samples.
• Slowly purge with a strong
mobile phase that is both
appropriate to dissolve the
contaminate and compat-
ible with the column.
• Slowly purge with very
strong solvent (refer to
Equilibration section)
• Since this may occur the
1st 2 weeks of use, Waters
will replace any column
with this defect. Refer to
warranty for details.
• Replace or clean inlet and
outlet filters.
Pico•Tag Amino Acid Analysis Column 3
e. Service and Applications Information
Waters staff of trained and experienced Service Specialists provides main-
tenance for Waters in-struments on preventative and/or corrective levels.
Contact Waters at 1-800-252-HPLC or your local Waters Representative for
answers to specific chromatography questions in areas such as methods devel-
opment, applications, quality control, and service related matters.
Page 4

[ Care and Use ManUal ]
IV. column eFFIcIency
Waters measures column efficiency by using the 5-sigma method. Plate count,
as an expression of efficiency, is determined by measuring the peak width at
4.4% of the peak height. Unlike half-peak height and tangent methods used by
other manufacturers, this stringent method takes into con-sideration naturally
occurring peak asymmetry.
Figure 3. Column Efficiency
V. test condItIons
Columns are thoroughly tested in our quality control laboratories for adher-
ence to our specifica-tions. Since slight variations in your results will occur
depending on the equipment used, test sam-ple makeup and equipment
settings and condition, perform the test sample run given here for your new
column and record the results (retention time and the settings used) before
attempting the first analysis. Use these results for comparison throughout the
life of your column.
NOTE: BE SURE TO RECORD RESULTS AND INSTRUMENT SETTINGS (AND
CONFIGURATIONS) TO ALLOW EXACT REPRODUCTION AND COMPARISON IN
THE FUTURE.
Test Sample: Acenaphthene (0.05% in acetonitrile)
Solvent: Pico•Tag Eluent B
Flow Rate: 1.0 ml/min
Chart Speed: 10 cm/min
Injection Volume: 15 µl
Detection: 254 nm @ 0.5 AUFS
VI. recommended sPAre PArts (steel columns)
Item P/N Quantity
Pico•Tag Reagent Kit 8 8123 1
Phenylisothiocyanate (PITC) 8 8120 1
Triethylamine (TEA) 8 8121 1
Protein Hydrolyzate Standards 8 8122 1
3.9 mm Filter Replacement Kit 84095 1
Filter Insert Assembly (only)* 84056 2
Retainers (only)* 84057 2
* Included as components of 3.9 mm Filter Replacement Kit
VII. WArrAnty
Waters warrants its quality manufactured columns in accordance with the
following terms and conditions. Waters will repack or replace the column (at
our discretion) at no cost if it fails to perform satisfactorily and Waters is noti-
fied within 90 days of receipt. Columns being returned must have prior Return
Authorization granted by Waters Customer Service Department. Approval is
subject to the following exclusions:
• Physical damage to the column due to misuse or abuse.
• Chemical damage to the packing material due to operating with incom-
patible solvents, buffers, or pH.
• Physical damage to the packing material due to operation at incorrect
temperatures or pressures.
• High internal pressure in the column due to improper solvent or sample
filtra-tion practices causing particulate buildup or precipitation in the
column or end fittings.
Pico•Tag Amino Acid Analysis Column 4
Page 5

[ Care and Use ManUal ]
Sales Offices:
Austria and European Export
(Central South Eastern Europe,
CIS and Middle East) 431 877 18 07
Australia 2 9933 1777
Belgium 32 2 726 1000
Brazil 55 11 5094 3788
Canada 800 252 4752
China 8621 6495 6999
CIS/Russia +7 495 3367000
Czech Republic 42 02 617 11384
Denmark 45 46 59 8080
Finland +358 9 5659 6288
France (33) 1 30 48 72 00
Germany 49 6196 400600
Hong Kong 852 29 64 1800
Hungary 36 1 350 5086
India and India Subcontinent
91 80 2 837 1900
Ireland 353 1 448 1500
Italy 39 02 274 211
Japan (81) 3 3471 7191
Korea (82) 2 820 2700
Mexico 5255 5200 1860
The Netherlands +31 (0)76-50 87 200
Norway 47 63 84 60 50
Poland (48) 22 833 4400
Puerto Rico 787 747 8445
Singapore 65 6273 1221
Spain 34 93 600 93 00
Sweden 46 8 555 11500
Switzerland 41 62 889 2030
Taiwan 886 2 2543 1898
United Kingdom 44 208 238 6100
©2007 Waters Corporation. Waters, The Science of W hat’s
Possible, Pico. Tag, and Sep-Pak are trademarks of Waters Corporation. Milli-Q is a trademark of Millipore Corporation.
November 2007 WAT088126 Rev 2 VW-PDF
Pico•Tag Amino Acid Analysis Column 5
Waters Corporation
34 Maple Street
Milford, MA 01757 U.S.A.
T: 1 508 478 2000
F: 1 508 872 1990
www.waters.com