Waters OST C18 User Manual

[ method guidelines ]
XBridge™ ost C18 Method guidelines
*ost: oligonuCleotide separation teChnology
XBridge™ OST C18 columns are based on Waters second generation of hybrid-silica BEH Technology™ particles and can be effectively used for the lab scale purification and analysis of detritylated synthetic oligonucleotides using ion-pair, reversed-phase chromatography. This document provides useful method guidelines for the effective use of this column chemistry for this group of compounds.
Contents
I. PRINCIPLES OF OLIGONUCLEOTIDE SEPARATIONS
II. SAMPLE PREPARATION
III. RECOMMENDED MOBILE PHASES
IV. RECOMMENDED INJECTOR WASH SOLVENT
V. GENERAL CONSIDERATIONS IN DEVELOPING SEPARATIONS
VI. ANALYSIS OF MODIFIED OLIGONUCLEOTIDES
VII. PURIFICATION CONSIDERATIONS
[ method guidelines ]
+
+
+
+
-
-
-
-
-
-
-
-
+
TEA
XBridge™ OST C
18
chain
PO group on Oligo chain
i. p rinCiples o F o l i gonu C l eotide s e parat ions
Separations of detritylated synthetic oligonucleotides on an XBridge™ O ST C18 column are based on ion-pair, reversed-phase chromatographic principles (IP-RP-LC). As shown in Figure 1, the ion-pairing additive in the mobile phase is adsorbed on a hydrophobic sorbent and provides for charge-to-charge interactions with negative charges contained on the oligonucleotide backbone (e.g., phosphate groups).
Figure 1: Proposed Mechanism of IP-RP-LC for Synthetic Oligonucleotide Separations
As a result, an efficient charge-based (length-based) oligonucleotide separation is achieved (Figure 2). Gradient elution using an acetonitrile or methanol eluent displaces both ion-pairing agent and the oligo­nucleotides from the sorbent surface.
Figure 2: Separation of a 15 - 60mer Deoxythymidine Ladder on XBridge™ OST C
18
HPLC system: Waters BioAlliance™ 2796, PDA Detector with micro UV cell Sample Injected: Approximately 100 pmoles of a detritylated 15 – 60mer oligonucleotide ladder diluted in
0.1 M TEAA Column: Waters X Bridge™ OST C18, 2.5 µm (2.1 x 50 mm) Mobile P hases: A: 0.1 M TEAA, B: Acetonitrile / 0.1M TEAA, 20/80, v/v Flow rate: 0.2 mL/min Column Temp.: 60 ˚C Gradient delay: 0.45 mL Gradient: 40 to 62.5% B in 30 minutes (8-12.5% acetonitrile, 0.15% acetonitrile per minute) Detection: 260 nm, 5 scans per second
Separation selectivity and resolution decreases with increasing oligonucleotide length (Figure 2) making the separation of long oligonucleotides challenging. Modified oligonucleotides such as phos­phorothioates and 2-O alkyl modified species are also more difficult to analyze. Special mobile phase may be required (see Section III, Recommended Mobile Phases).
Two commonly used ion-pairing agents for oligonucleotide applications are triethyl ammonium and dimethylbutyl ammonium ions. The final pH of these mobile phases containing either of these ion-pairing reagents is adjusted by the addition of Acetic Acid, or in some cases, Hexafluo­roisopropanol (HFIP). These mobile phases are volatile making them suitable for LC-MS applications.
The ability to adequately resolve synthetic oligonucleotide mixtures by ion-pair, reversed-phase chromatography is significantly affected by the particle size of the material contained in an efficiently packed column (see Figure 3). Consequently, XBridge™ OST C18 columns are efficiently packed with 2.5 micron material to maximize detritylated oligonucleotide component resolution. In order to improve oligonucleotide separation efficiency and speed, elevated separation temperature (e.g. 60 ˚C) is recommended. Elevated temperature will also reduce operating LC System back pressure.
2
[ method guidelines ]
Figure 3: Effectiveness of Waters BEH Technology™ Hybrid-Silica C18 Particle Size on Deoxythymidine Ladder Separations
HPLC system: Waters BioAlliance™ 2796, PDA Detector with micro UV cell Sample Injected: Approximately 100 pmoles of detritylated 15 – 60mer crude oligonucleotide ladder diluted in 0.1 M TEAA Column: Waters BEH Hybrid-Silica C18 particles (2.1 x 50 mm) Mobile P hases: A: 0.1 M TEAA, B: Acetonitrile / 0.1M TEAA, 20/80, v/v Flow rate: 0.2 mL/min Column Temp.: 60 ˚C Gradient delay: 0.45 mL Gradient: 40 to 62.5% B in 30 minutes (8-12.5% acetonitrile,
0.15% acetonitrile per minute) Detection: 260 nm, 5 scans per second
In addition to ion-pairing, a hydrophobic reversed-phase mechanism also takes place in the oligonucleotide separation. The residual interaction of nucleobases has an impact on overall retention and separation selectivity, especially when using Triethylammonium Acetate (TEAA) ion-pairing mobile phases. Separation of N and N-1mers may be either enhanced or suppressed by the sequence contribution. More potent ion-pairing systems such as Triethylammonium ion with Hexafluo­roisopropanol counter ion provide for more regular “charge-based” separations (Figure 4).
Figure 4: Impact of Ion-pairing System on Separation of a 10-30mer Heterooligonucleotide Ladder
HPLC system: Waters BioAlliance™ 2796, PDA Detector with micro UV cell Sample: 20 mer: TCC C TA GCG T TG AAT TGT CC 25 mer: TC C CTA GCG TTG AAT TGT C CC TTA G 30 mer: TCC CTA GC G TTG AAT TGT CCC TTA GCG GGT Ladder was prepared by hydrolyzing detritylated 20, 25, and 30mer oligonucleotides with a 3’-exonuclease Column: Waters X Bridge™ OST C18, 2.5 µm (4.6 x 50 mm) Mobile p hases: Upper chromatogram: 0.1 M TEAA with acetonitrile gradient; Lower chromatogram: 16.3 mM T EA - 400 mM HFIP with methanol gradient Flow rate: 1.0 mL/min Column Temp.: 60 ˚C Gradient delay: 0.45 mL Detection: 260 nm, 5 scans per second
ii. saMple preparation
1. Dissolve the detritylated synthetic oligonucleotide sample in Mobile Phase A (e.g., 0.1 M TEAA). For example, a 0.05 - 0.2 µmole scale synthesis can be prepared in 0.1 mL of 0.1 M TEAA. Proportionately larger or smaller volumes of 0.1M TEAA are required when dissolving samples from different scale syntheses. Due to the nature of gradient separations, relatively large volumes of sample (in low organic strength eluent) can be injected and concentrated onto the head of the column before beginning the gradient elution program.
2. Samples must be completely in solution and free of particulates before injecting onto the column. Remove all particles from the sample (Controlled Pore Glass Synthesis Support, etc.), which may block the inlet column frit, increase the operating pressure, and shorten the column life time. Sample contamination with high con­centration of salts and/or detergents may also interfere with analysis.
3
[ method guidelines ]
3. To remove particulates the sample may be filtered with a 0.2 μm membrane. Be sure that the selected membrane is compatible and does not dissolve with the selected Mobile Phase diluent. Contact the membrane manufacturer with solvent compatibility questions. An alternative method of particulate removal involves centrifugation for 20 minutes at 8,000 rpm, followed by the transfer of the supernatant liquid to an appropriate vial.
iii . reCo M M ended M o Bile p h ases
The most common ion-pair mobile phase for synthetic oligonucleotide separations is based on Triethylammonium Acetate (TEAA). This mobile phase can be prepared by titrating Glacial Acetic Acid aqueous solution with Triethylamine (TEA).
Note: To maximize column life, it is ESSENTIAL that all prepared OST Mobile Phases be ltered through a solvent compatible, 0.45 µm mem­brane and contained in bottles that are clean and particulate free.
TEAA
1L of 0.1 M TEAA may be prepared as follows:
advisable to use aqueous buffer consisting of 15 mM TEA and 400 mM HFIP (pH 7.9).
TEA-HFIP System 1
1L of 8.6 mM TEA / 100 mM HFIP is prepared as follows:
1) Perform work in a hood
2) Add 10.4 mL of HFIP (16.8 g) into 988.4 g of water and mix well.
3) Slowly add 1.2 mL of TEA.
4) The pH is approximately 8.3 +/- 0.1.
TEA-HFIP System 2
1 L of 15 mM TEA / 400 mM HFIP is prepared as follows:
1) Perform work in a hood
2) Add 41.56 mL (67.17 g) of HFIP into 956.36 g of water and mix well.
3) Slowly add 2.08 mL (1.52 g) of TEA.
4) The pH of final buffer is approximately 7.9 +/- 0.1.
1) Perform work in a hood.
2) Add 5.6 mL of glacial Acetic Acid into 950 mL of water and mix well.
3) Slowly add 13.86 mL of TEA.
4) The pH should be adjusted to pH 7 +/- 0.5 by careful addition of Acetic Acid.
5) Adjust final volume to 1 L with water.
Alternatively, premixed TEAA can be used [(e.g., Sigma 1 M TEAA (part no. 90357)]. Mix 100 mL with 900 mL of water to prepare 1 L of
0.1 M TEAA mobile phase.
Alternative ion-pairing reagents are recommended for improved sepa­ration of phosphorothioates or when performing LC-MS analyses. An ion-pairing mobile phase based on Triethylamine (TEA) and Hexafluo­roisopropanol (HFIP) as the buffering acid produces an efficient eluent system for improved separations involving these application types.
As indicated below, two ion-pairing systems are useful.
For routine detritylated oligonucleotide applications, aqueous buffer consisting of 8.6 mM TEA and 100 mM HFIP is effective. For applications such as those involving the separation of G-rich oligonucleotides, it is
iV. reCoMMended inJeCtor Wash solVents
Between analyses, the HPLC system injector seals should be washed. A 90% Water / 10% Acetonitrile injector wash solvent is recommended.
V. general Consideration in deVeloping separations
Separation of detritylated synthetic oligonucleotides by ion-pair, reversed-phase chromatography uses very shallow gradients. With both TEAA and TEA-HFIP ion-pairing systems, a rate of strong eluent change between 0.1-0.25 % Acetonitrile (or Methanol) per minute is recommended. However, the formation of shallow gradients can place performance demands on LC pumps and mixers that can compromise the quality of the separation. Consequently, it is strongly advised that Mobile Phase B formulation contain a premix blend of aqueous and organic solvents (e.g., Mobile Phase A= 0.1 M TEAA and Mobile Phase B = Acetonitrile / 0.1M TEAA, 20/80, v/v) to minimize poten­tially inadequate solvent mixing that can compromise component resolution.
4
[ method guidelines ]
As illustrated in Figures 5 through 7, these analyses were performed with the following mobile phases:
Mobile Phase A: 0.1 M TEAA Mobile Phase B: Acetonitrile (ACN) containing 0.1 M TEAA, 20:80 (v:v)
The 0.1% ACN / min gradient change from an initial 5 to 10% Acetonitrile concentration over 50 minutes was programmed as specified in Table 1:
Table 1
Time % A % B Actual Acetonitrile (ACN) Concentration
0 min 75 25 5% 50 min 50 50 10%
Example:
For the initial 5% Acetonitrile concentration:
Initial %B = desired ACN % / Volume Fraction of ACN in Mobile Phase B. So, initial %B = 5% / 0.2 = 25%
For the final 10% Acetonitrile concentration:
The organic solvent concentration at initial sample loading conditions has to be well chosen. If the initial organic solvent strength is too high, then some desired oligonucleotide sequences may be unretained. In the other extreme, when the gradient starts with too low an organic concentration, the analysis is excessively long without the benefit of enhanced component resolution. A suitable gradient separation method can be approximated from the oligonucleotide base (C, G, A, and T) composition. The initial gradient is typically adjusted while keeping the gradient slope constant.
Table 2: Suggested Gradient Conditions for Non-Standard Detritylated Synthetic Oligonucleotide Sequences
Gradient 1
[Stan dard
oligonucleotides (1)]
Initial % ACN 7.0 0 % 5.25% 7.5 0 % Final % ACN 10.75% 9.00% 12. 50 % Gradient Length(4) 15 min 15m in 20min
Gradient 2
[Hig h GC content or s hort
oligonucleotides (2)]
Gradient 3
[Hig h AT content or long
oligonucleotides (3)]
1: Standard oligonucleotides: 10 – 30mers
Final %B = desired ACN % / Volume Fraction of ACN in Mobile Phase B. So, final %B = 10% / 0.2 = 50%
With TEAA mobile phases, the unmodified oligonucleotides elute within a 7-10 % ACN gradient window. However, C and G rich oligo­nucleotide sequences are generally less retained (i.e., elute within a 5-8% ACN gradient window) than A and T rich sequences (i.e., elute within a 8-11% ACN gradient span). When using a shallow gradient, the total length of analysis for an unknown sample sequence may be excessive. Use of a fast scouting gradient with a 1% ACN per minute change is recommended in such cases. Information gathered from this scouting analysis can then be used to create a more appropriate and time efficient set of gradient conditions for the particular sample.
Gradient slope has a direct impact on the achievable oligonucleotide component resolution (along with the type of ion-pairing agent, sequence, and oligonucleotide modification). Steeper gradients (e.g., 1% ACN change per minute on a 4.6 x 50mm column at a 1.0 mL/min flow) are recommended for labeled oligonucleotides or for short, 5-15 mer sequences. Separation of longer sequences are typically performed using more shallow gradient slopes (e.g. 0.15% ACN change per minute on a 4.6 x 50mm column at a 1.0 mL/min flow).
2: Short oligonucleotides: Less than 10mer
3: Long oligonucleotides: 30 – 60mers
4: Assuming use of a 2.1 x 50mm XBridge™ OST C18 column at a flow
of 0.2 mL/min and a separation temperature of 60 ˚C.
The retention of single and dual dye-labeled oligonucleotides is dictated by the nature of label. For example, the retention of 25 mer oligonu­cleotide increases according to the type of label attached as follows: no label<6FAM< <TAMRA<TET<HEX<Cy3.
Vi. analysis oF ModiFied oligonuCleotides
XBridge™ OST C18 columns are suitable for analysis of unmodified as well as modified detritylated oligodeoxyribonucleotides and oligoribo­nucleotides. Phosphorothioate and 2’-O -alkyl modified oligonucleotides can also be analyzed with IP-RP-HPLC method. However, these full length oligonucleotide products are usually more difficult to resolve from their shorter length failure sequences. The recommended ion-pair system for phosphorothioate oligonucleotide analysis is TEA-HFIP (see Recommended mobile phases). An example of a 25mer phosphorothioate oligonucleotide analysis is shown in Figure 5.
5
[ method guidelines ]
Figure 5: Analysis of a Digested 25mer Phosphorothioate Oligonucleotide
HPLC system: Waters BioAlliance™ 2796, PDA Detector with micro UV cell Sample: Detritylated 25mer phosphorot hioate oligonucleotide mix (CTC TCG CAC CCA TC T CTC TCC TTC T ) digested with 3’ exonuclease Column: XBridge™ OST C18, 2.5 µm (2.1 x 50 mm) Mobile p hase: A: 15 mM TEA with 400 mM HFIP B: methanol Flow rate: 0.2 mL/min Column Temp.: 60 ˚C Gradient delay: 0 mL Gradient: 15 to 20% B in 20 minutes (0.25% methanol per minute) Detection: 260 nm, 2 scans per second
Figure 6: Analysis of a 20mer Oligodeoxythymidine Crude Synthesis Mixture
HPLC system: Waters BioAlliance™ 2796, PDA Detector with micro UV cell Sample: ~600 pmol of a detritylated 20mer, ~18 pmol of 19mer, ~4.5 pmol of 17mer was injected on column. Column: XBridge™ OST C18, 2.5 µm (2.1 x 50mm) Mobile p hase: A: 0.1 M TE AA, B: Acetonitrile / 0.1M TEAA, 20/80 (v/v) Flow rate: 0.2 mL/min Column Temp: 60 ˚C Gradient Delay: 0 mL Gradient: 35 to 50% B in 30 minutes (7-10% acetonitrile) Detection: 260 nm, 2 scans per second
Peptide nucleic acids (PNA) can also be analyzed using XBridge™ OST C18 columns. The ion-pairing system recommended for analysis of PNA is similar to those used for peptide analysis (0.1% Trifluoroacetic Acid or Formic Acid).
UV detection of eluted oligonucleotide peaks is often performed at 260 nm. Injection of 50 pmol of detritylated oligonucleotide sample on a 2.1 x 50mm XBridge™ OST C18 column yields relatively abundant peaks. Limits of quantitation (LOQ) vary with the type of oligonucleotide, LC system and detector; LOQ generated on Waters 2996 PDA detector equipped with micro UV cell is approximately 1 pmol (2.1 x 50mm XBridge™ OST C18). The Limit of detection (LOD) estimate is shown in Figure 6.
Vii. puriFiCation Considerations
XBridge™ OST C18 columns are designed for laboratory scale oligo­nucleotide purifications and analyses. Sufficient amount of isolated material suitable for molecular biology and other experiments can be prepared in a single injection. For example, a 4.6 X 50 mm XBridge™ OST C18 column can suitably purify approximately 20-200 nmoles of sample in a single injection. It is important to understand that column overloading results in a peak broadening and that some earlier eluting impurities may co-elute with the component of interest. With a proper heart-cutting technique, a good purity of the target oligonucleotide can be obtained without significant yield sacrifice (Figure 7).
Chromatographers frequently develop a separation on the analytical scale before moving to preparative work. The steps required to opti­mize the analytical separation involve:
1) Selecting the appropriate column packing material and mobile phase.
2) Determining the optimal flow rate, gradient during and separation temperature.
3) Determining the amount of material that can be satisfactorily loaded and separated on the analytical scale column.
6
[ method guidelines ]
Once the separation has been optimized, one begins preparing for the preparative separation. The steps to successfully scale a separation from an analytical to a preparative column, containing the same packing material composition, are detailed below
Step A: Calculate the flow rate for use on the Preparative column.
Prep arative Column Flow Rate =
Analytic al Column Flow Rate x (Diameter of Prep Column)
(Diam eter of Anal Column)
2
2
Step B: To get similar chromatography, the gradient elution profile should be created on both columns using the same number of column volumes. When the analytical and preparative columns are of the same length, as is recommended for this application, then the gradient duration should be the same.
Note: This assumes use of the same ow rate linear velocity for both runs as calculated above. For preparative runs, it is also important to note that an initial gradient delay is required to allow the entire sample to load onto the head of the column prior to beginning chromatography.
* XBridge™ OST C18 Custom Column
** Values are only approximate and vary depending on detritylated oligonucleotide length, base composition, and “heart-cutting” fraction collection method used
Figure 7 shows the separation of 90 nmoles of a detritylated 30 mer deoxythymidine crude reaction mixture on a 4.6 x 50mm XBridge OST C18 column. The collection interval is suggested by the lines. Due to partial column overloading, the N -1, N -2... impurities are partially displaced and elute earlier than expected. With the proper hearth-cutting technique, 95-98% purity is typically achieved for 15-35 mer oligo­nucleotides at this purification scale.
Figure 7: Purication of a Detritylated 30mer Deoxythymidine Sample
Step C: The last calculation involves determining how much sample can be loaded on the preparative column. This calculation compares the relative volumes of the two columns assuming that both columns are the same length as recommended for this application.
Preparative Column Sample Load =
Previously Determined Analytical Column Sample Load x (Diameter of Prep Column)2
(Diameter of Anal Column)
2
Table 3: XBridge™ OST C18 Column Selection Guide for Detritylated Oligonucleotide Purication
Column (mm) Approx Mass Load (µmoles)** Flow Rate (mL/min)
2.1 x 50 0.04 0.2
4.6 x 50 0.20 1.0
10.0 x 50 1.0 0 4.5
19.0 x 50* 4.00 16. 0
30.0 x 50* 9.00 40.0
50.0 x 50* 25.00 110 . 0
HPLC system: Waters BioAlliance™ 2796, PDA Detector with micro UV cell Sample: Crude detritylated 30mer oligothymidine, 200 nmole dissolved in 100 µl of mobile p hase A, 45 µl was injected on column Column: XBridge™ OST C18, 2.5 µm (4.6 x 50mm) Mobile p hase: A: 0.1M TEAA with 400 mM HFIP B: Acetonitrile/0.1M TEAA, 20/80 (v/v) Flow rate: 1.0 mL/min Column Temp.: 60 ˚C Gradient delay: 0 mL (compensated) Gradient: 35 to 65% B in 24 minutes (7-13% ACN, 0.25% ACN per minute) Detection: 260 nm, 2 scans per second
Table 4: Ordering Information
Description Particle Size Pore Size Dimension Part No.
XBridge™ OST C18 2.5 μm 135 Å 2.1 x 50 mm 186003952 XBridge™ OST C18 2.5 μm 135 Å 4.6 x 50 mm 186003953 XBridge™ OST C18 2.5 μm 135 Å 10.0 x 50 mm 186003954 Custom
XBridge™ OST C
18
-- -- -- 186003955
7
[ method guidelines ]
sales oFFiC es
Australia
Waters Australia Pty. Limited Tel: 2 9933 1777 Fax: 2 9898 1455
Austria and European Export (Central South Eastern Europe, CIS and Middle East)
Tel: 431 877 18 07 Fax: 431 877 18 08
Belgium
Waters S.A.-N.V. Tel: 32 2 726 1000 Fax: 32 2 726 1100
Brazil
Waters Technology Tel: 55 11 5094 3788 Fax: 55 11 5093 6413
Canada
Waters Limited Tel: 800 252 4752 Fax: 905 678 9237
China
Waters Technologies (Shanghai) Limited Tel: 8621 6495 6999 Fax: 8621 6495 1999
CIS/Russia
Tel: +7 495 727 4490 Fax: +7 495 336 7000
Czech Republic
Tel: 42 02 617 11384 Fax: 42 02 617 11386
Denmark
Waters A/S Tel: 45 46 59 8080 Fax: 45 46 59 8585
Finland
Waters Finland Tel: (09) 5659 6288 Fax: (09) 5659 6282
France
Waters S.A Tel: (33) 1 30 48 72 00 Fax: (33) 1 30 48 72 01
Germany
Waters GmbH Tel: 49 6196 400600 Fax: 49 6196 4006010
Hungary
Tel: 36 1 350 5086 Fax: 36 1 350 5087
Hong Kong
Waters China Ltd. Tel: 852 29 64 1800 Fax: 852 25 49 6802
India
Waters India Pvt. Ltd. Tel: 91 80 2 837 1900 Fax: 91 80 2 839 2157
Ireland
Waters Chromatography Ireland Ltd. Tel: 353 1 448 1500 Fax: 353 1 448 1510
Italy
Waters S.P.A. Tel: 39 02 274 211 Fax: 39 02 250 1827
Japan
Nihon Waters K.K. Tel: (81) 3 3471 7191 Fax: (81) 3 3471 7118
Korea
Waters Korea Limited, Tel: (82) 2 820 2700 Fax: (82) 2 820 2730
Mexico
Waters S.A. DE C.V. Tel: 5255 5200 1860 Fax: 5255 5524 9376
The Netherlands
Waters Chromatography B.V. Tel: +31 (0)76-50 87 200 Fax: +31 (0)76-50 87 280
Norway
Waters AS Tel: 47 63 84 60 50 Fax: 47 63 84 60 51
Poland
Tel: (48) 22 833 4400 Fax: (48) 22 833 0987
Puerto Rico
Waters Technologies Corporation Tel: 787 747 8445 Fax: 787 747 8448
Singapore
Waters Asia Ltd. Tel: 65 6273 1221 Fax: 65 6273 6116
Spain
Waters Cromatografía, S.A. Tel: 34 93 600 93 00 Fax: 34 93 600 93 60
Sweden
Waters Sverige AB Tel: 46 8 555 11500 Fax: 46 8 555 11550
Switzerland
Waters AG Tel: 41 62 889 2030 Fax: 41 62 889 2059
Taiwan
Tel: 886 2 2543 1898 Fax: 886 2 2543 1918
United Kingdom
Waters U.K. Ltd. Tel: 44 208 238 6100 Fax: 44 208 207 7070
All Other Countries
Waters Corporation
34 Maple Street Milford, MA 01757 USA Tel: 508 478 2000 800 252 4752 Fax: 508 872 1990
http://www.waters.com
Waters, XBridge, BEH Technology and BioAlliance are trademarks of Waters Corporation.
©2008 Waters Corporation. April 2008 715001476 Rev B SC
Waters Corporation
34 Maple Street Milford, MA 01757 U.S.A. T: 1 508 478 2000 F: 1 508 872 1990 www.waters.com
Loading...